Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq,
OxiBand, Dura-Y, Max-Fast, and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
T o obtain information about a warranty, if any , contact Nellcor’ s Technical Services Department, or your local
representative.
Purchase of this instrument confers no express or implied license under any Nellcor Puritan Bennett patent to
use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or mor e of the foll owing U.S. Patents and for eign equivalents: 4,653,498; 4,802,486; 4,869,254;
4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,865,736; 6,463,310;
6,708,049; Re.35, 122; and foreign equivalents.
Figure 49: Main PCB Battery Charger Schematic Diagram ............................117
Figure 50: Main PCB Isolation for MP-506 Schematic Diagram .....................119
Figure 51: Main PCB Isolation for External Interface Schematic Diagram ......121
Figure 52: Main PCB DC Battery Input and DC/DC Converter Schematic Diagram
123
Figure 53: Main PCB Interface and Speaker Control Schematic Diagram .....125
Figure 54: Main PCB MCU and Storage Control Schematic Diagram ............127
Figure 55: Main PCB Parts Locator Diagram .............................. .................... 129
Figure 56: Front Panel PCB 7-Segment Display Schematic Diagram ............131
Figure 57: Front Panel PCB LED Display and Buttons Schematic Diagram ...133
Figure 58: Front Panel PCB Interface and LED Drivers Schematic Diagram .135
Figure 59: Front Panel Parts Locator Diagram ...............................................137
Figure 60: Power Supply Schematic Diagram ................................................139
Figure 61: Power Supply Parts Locator Diagram ........................... ................. 141
Warnings are identified by the WARNING symbol shown above and a line
above and below the warning text. The word WARNING and all warning text
are boldfaced.
W arnings alert the user to potential serious outcomes (death, injury, or adverse
events) to the patient or user.
Cautions
Notes
Cautions are identified by the CAUTION symbol shown above. The word
CAUTION and the caution text are boldfaced.
Cautions alert the user to exercise care necessary for the safe and effective use
of the N-550B.
Notes are identified by the Note symbol shown above.
Notes provide additional useful information.
N-550B1
Introduction
Manual Overview
This manual contains information for servicing the Nellcor® model N-550B
pulse oximeter. Only qualified service personnel should service this product.
Before servicing the N-550B, read the operator's manual carefully for a
thorough understanding of operation.
WARNING: Explosion hazard. Do not use the N-550B pulse oximeter in
the presence of flammable anesthetics.
Description Of N-550B Pulse Oximeter
The N-550B Pulse Oximeter is indicated for the continuous noninvasive
monitoring of functional oxygen saturation of arterial hemoglobin (SpO
pulse rate. The N-550B is intended for use with neonatal, pediatric, and adult
patients during both no-motion and motion conditions and for patients who
are well or poorly perfused in hospitals, hospital-type facilities, intra-hospital
transport, and home environments. For prescription use only.
2) and
Note:Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical care
areas, within the hospital plus hospital-type facilities. Hospital-type
facilities include physician office based facilities, sleep labs, skilled
nursing facilities, surgicenters, and subacute centers.
Intra-hospital transport includes transport of a patient within the
hospital or hospital-type facility.
Use with any particular patient requires the selection of an appropriate oxygen
transducer as described in the N-550B Operator's Manual.
Motion performance claims are applicable to models MAX-A, MAX-AL,
MAX-P, MAX-N, and MAX-I Nellcor O
XIMAX
TM
oximetry sensors.
Digital displays are provided for oxygen saturation and pulse rate, and a
10-segment light-emitting diode (LED) bar indicates pulse amplitude. The
controls and indicators for the N-550B are illustrated and defined in Figure 1
and Figure 2 on page 3.
To perform test and troubleshooting procedures and to understand the
principles of operation and circuit analysis sections of this manual, you must
know how to operate the N-550B. Refer to the N-550B operator's manual. To
understand the various Nellcor sensors that work with the N-550B, refer to the
individual sensor's directions for use.
The latest versions of the operator’s manual and the service manual are posted
on the Internet at:
Spare parts and accessories are posted on the Internet at:
4
Routine Maintenance
Cleaning
CAUTION: Do not spray, pour, or spill any liquid on the N-550B, its
accessories, connectors, switches, or openings in the chassis.
For surface-cleaning and disinfecting follow your institution's procedures or:
•The N-550B may be surface-cleaned by using a soft cloth dampened
with either a commercial, nonabrasive cleaner or a solution of 70%
alcohol in water, and lightly wiping the surfaces of the N-550B.
•The N-550B may be disinfected using a soft cloth saturated with a
10% solution of chlorine bleach in tap water .
Before attempting to clean an SpO
enclosed with the sensor. Each sensor model has cleaning instructions specific
to that sensor.
Periodic Safety Checks
The N-550B requires no routine service or calibration other than changing the
battery at least every two years. See Battery Removal on page 66. The
following checks should be performed at least every two years by a qualified
service technician.
1. Inspect the equipment for mechanical and functional damage.
2. Inspect safety labels for legibility. If the labels are damaged, contact
Nellcor’s Technical Services Department, 1.800.635.5267, or your local
Nellcor representative.
2 sensor, read the directions for use
Functional Checks
1. If the N-550B has been visibly damaged or subjected to mechanical shock
(for example, if dropped), perform the performance tests. See
Performance Tests on page 8.
N-550B5
Routine Maintenance
Battery
2. Perform the electrical safety tests detailed in Safety Tests on page 26. If
the N-550B fails these electrical safety tests, refer to Troubleshooting on
page 43.
3. Inspect the fuses for proper value and rating (F1 and F2 = 2 amp,
250 volts).
Nellcor recommends replacing the N-550B's battery every two years. When
the N-550B is going to be stored for two months or more, remove the battery
prior to storage. To replace or remove the battery, refer to Disassembly Guide
on page 53.
If the N-550B has been stored for more than 30 days, charge the battery as
described in Battery Charge on page 8. A fully discharged battery requires
6 hours to receive a full charge. The battery is being charged whenever the
N-550B is plugged into an AC power source.
6
Performance Verification
Introduction
This section discusses the tests used to verify performance following repairs
or during routine maintenance. All tests can be performed without removing
the N-550B cover. All tests except the battery char ge and battery performance
tests must be performed as the last operation before the N-550B is returned to
the user.
If the N-550B fails to perform as specified in any test, repairs must be made to
correct the problem before the N-550B is returned to the user.
Equipment Needed
Table 1: Equipment Needed
EquipmentDescription
Digital multimeter (DMM)Fluke Model 87 or equivalent
Durasensor
(sensor)
O
XIMAX
Pulse oximeter testerSRC-MAX (must meet current EN
Safety analyzerMust meet current AAMI ESI/1993 &
Pulse oximetry cableDOC-10
Data interface cableEIA-232 cable (optional)
StopwatchManual or electronic
®
oxygen transducer
oxygen transducer
DS-100A
MAX-A
61326-1, FCC Part 15, and UL 1010-1,
or IEC 61010-1, or CSA 22.2 No.
1010-1 specifications)
IEC 60601-1/1998 specifications
N-550B7
Performance Verification
Performance Tests
Battery Charge
Note:This section is written using Nellcor factory-set defaults. If your
institution has preconfigured custom defaults, those values will be
displayed. Factory defaults can be restored. See Menu Item 10 (Restore Factory Default Settings as Power-On Default) on page 38.
Perform the following procedure to fully charge the battery.
1. Connect the N-550B to an AC power source.
2. Verify that the N-550B is off and that the
3. Charge the battery for at least 6 hours in standby.
Power-Up Performance
The power-up performance tests verify the following N-550B functions:
•Power-On Self-Test (POST) on page 9
•Power-On Defaults and Alarm Limit Ranges on page 11
CAUTION: If any indicator or display element does not light, or the
speaker does not sound, do not use the N-550B. Instead, contact qualified
service personnel, your local Nellcor representative, or Nellcor's
Technical Services Department, 1.800.635.5267.
Note:The N-550B should complete the POST function within 12 seconds.
5 (in pulse rate left window)low priority alarm
5 (in pulse rate center window)medium priority alarm
0 (in pulse rate right window)high priority alarm
n (in SpO
n (in SpO
N-550B9
2 left window)
2 center window)
Performance Verification
3. The N-550B automatically starts the Power-On Self-Test (POST), which
tests N-550B circuitry and functions.
CAUTION: During POST (immediately after power-up), confirm that all
display segments and indicators light, and the speaker sounds a
one-second pass tone.
4. While performing POST, the self-test display appears for approximately
five seconds. During this time:
•All indicators illuminate
•All segments of all numeric digits light and change from red to green
•All segments of the Pulse Amplitude Display light
•All segments of the
SatSeconds indicator light
5. Once the display test portion of POST is complete, the N-550B version is
displayed for approximately two seconds.
10
Performance Verification
Note:The version above is only a sample. Check your N-550B for the
current version installed.
The version number is needed when calling Nellcor's Technical
Services Department, 1.800.635.5267, or your local Nellcor
representative for technical assistance. Write down the number and
have it available prior to requesting technical assistance.
6. If the N-550B detects an internal problem during the POST, an alarm tone
sounds and the N-550B displays an Error Code and corresponding
number. See Troubleshooting on page 43.
7. Upon successful completion of the POST, the N-550B sounds a
one-second tone indicating that the N-550B has passed the test.
WARNING: If you do not hear the POST pass tone, do not use the
N-550B.
WARNING: Ensure that the speaker is clear of any obstruction. Failure
to do so could result in an inaudible alarm tone.
Note:In addition to serving as the POST pass verification, the POST pass
tone also functions as an audible confirmation that the speaker is
performing properly. If the speaker does not function, the alarm
sounds cannot be heard.
Power-On Defaults and Alarm Limit Ranges
See Figure 3 on page 9 for the location of the N-550B controls.
Note:When observing or changing default limits, a three-second time-out is
in effect. That is, if no action is taken within three seconds, the
N-550B automatically returns to the normal mode.
N-550B11
Performance Verification
1. Turn the N-550B on by pressing the Power On/Off button.
2. Wait for POST to be completed.
3. Press and release the
4. Verify that the N-550B emits a single beep and the %SpO
SpO2 Alarm Limit button.
2 display
indicates an upper alarm limit of “100” for about three seconds.
5. Verify that “HI” is displayed in the
Pulse Rate display.
Note:“HI” in the display window indicates an upper alarm limit that is
being adjusted, and a “LO” in the window indicates that a low alarm
limit is being adjusted.
Figure 4: Adjusting High %SpO2 Alarm Limit
12
6. Press the SpO2 Alarm Limit button. Within 3 seconds, press and hold
the Adjust Down button. Verify that the %SpO2 display reduces to a
minimum of one point above the %SpO
2 low alarm limits setting.
Note:A decimal point to the right of the value in either window display
indicates that the alarm limits are not power-on default values.
7. Press the
seconds). Verify that the N-550B emits two beeps and that the %SpO
SpO2 Alarm Limit button two times rapidly (twice within three
2
display indicates “85” for three seconds. Verify that “LO” is displayed in
the pulse rate window.
Performance Verification
Figure 5: Adjusting Low %SpO2 Alarm Limit
8. Press the SpO2 Alarm Limit button two times rapidly. Within three
seconds, press and hold the
%SpO2 display reduces to a minimum of “20.” Press and hold the Adjust
Up
button and verify that the %SpO2 display cannot be raised past the
upper alarm limit setting minus one point.
Adjust Down button and verify that the
9. Turn the N-550B off by pressing the
10. Turn the N-550B on by pressing the
Power On/Off button.
Power On/Off button.
11. Wait for POST to be completed. The %SpO
will be reset to the default values.
12. Press the
a single beep, the
and that the %SpO
Pulse Rate Alarm Limit button. V erify that the N-550B emits
Pulse Rate display indicates an alarm limit of “170”
2 display window shows “HI” for about 3 seconds.
2 and pulse rate alarm limits
Figure 6: Adjusting High Pulse Rate Alarm Limit
13. Press the Pulse Rate Alarm Limit Button. Within 3 seconds press and
hold the
reduces to a minimum of one point above the pulse rate low limit alarm
setting.
N-550B13
Adjust Down button. Verify that the Pulse Rate display
Performance Verification
14. Press the Pulse Rate Alarm Limit button two times (twice within 3
seconds). Verify that the N-550B emits two beeps and that the
Rate
display indicates an alarm limit of “40” for 3 seconds. Verify that
“LO” is displayed in the %SpO
2 display.
Pulse
Figure 7: Adjusting Low Pulse Rate Alarm Limit
15. Press the Pulse Rate Alarm Limit button two times rapidly. Within 3
seconds press and hold the
Rate
display cannot be raised past the upper alarm limit setting minus
one point.
Adjust Down button. Verify that the Pulse
16. Turn the N-550B off by pressing the
17. Turn the N-550B on by pressing the
18. Wait for POST to be completed. The %SpO
Power On/Off button.
Power On/Off button.
2 and pulse rate alarm limits
will be reset to the default values.
19. Press the
three seconds press and hold the
Rate
Pulse Rate Alarm Limit button two times rapidly. Within
Adjust Up button. Verify that the Pulse
display cannot be adjusted above the upper alarm limit setting
minus one point.
20. Press the SatSeconds Alarm Limit button. Verify that the N-550B emits a
single beep and that the
“OFF” for three seconds. Verify that “SAS” is displayed in the %SpO
Pulse Rate display indicates an alarm limit of
2
display.
14
Performance Verification
Figure 8: Adjusting SatSeconds Alarm Limit
21. Press the SatSeconds Alarm Limit button. Within three seconds press the
Adjust Up button or the Adjust Down button. Verify that the Pulse
Rate
display changes between 10, 25, 50, 100, and OFF.
22. Press the
23. Press the
24. Press and release the
display indicates an alarm limit of “100.”
25. Press the
%SpO2 display indicates an alarm limit of “85.”
26. Press and release the
Power On/Off button to turn the N-550B off.
Power On/Off button to turn the N-550B on.
%SpO2 Alar m Limit button. Verify that the %SpO2
%SpO2 Alarm Limit button two times rapidly. Verify that the
Pulse Rate Alarm limit button. Verify that the
Pulse Rate display indicates an alarm limit of “170.”
27. Press the
the Pulse Rate display indicates an alarm limit of “40.”
28. Press the SatSeconds Alarm Limit button. Verify that the
display indicates an alarm limit of “OFF.”
Pulse Rate Alarm Limit button two times rapidly. Verify that
Pulse Rate
29. Press the
N-550B15
Power On/Off button to turn the N-550B off.
Performance Verification
General Operation
LED Excitation Test
The following tests are check overall system performance:
•LED Excitation Test below.
•Operation with a Live Subject on page 17.
This procedure uses normal system components to test circuit operation. A
Nellcor O
XIMAX oxygen transducer (sensor), model MAX-A, is used to
examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Sensor Port
1. Connect the N-550B to an AC power source.
2. Connect a DOC-10 pulse oximetry cable to the N-550B sensor port (1).
3. Connect a MAX-A sensor to the sensor-input cable.
4. Press the
Power On/Off button to turn the N-550B on.
5. Leave the sensor open with the LEDs and photo detector visible.
6. After the N-550B completes its normal power-up sequence, verify that
the sensor LED is brightly lit.
16
Performance Verification
7. Slowly move the sensor LED in proximity to the photo detector element
of the sensor (close the sensor slowly). Verify, as the LED approaches the
optical sensor, that the LED intensity decreases.
8. Open the sensor and notice that the LED intensity increases.
9. Repeat step 7 and the intensity will again decrease. This variation is an
indication that the microprocessor is in proper control of LED intensity.
10. Press the
Operation with a Live Subject
Patient monitoring involves connecting the sensor to a live subject for a
qualitative test.
1. Ensure that the N-550B is connected to an AC power source.
Power On/Off button to turn the N-550B off.
1. Sensor Port
2. Connect a DOC-10 pulse oximetry cable to the N-550B sensor port (1).
3. Connect a Nellcor Durasensor oxygen transducer (sensor), model
DS-100A, to the sensor cable.
4. Clip the DS-100A to the subject as recommended in the sensor's
directions for use.
5. Press the
Power On/Off button to turn the N-550B on and verify that the
N-550B is operating.
N-550B17
Performance Verification
Alarm Volume Control
6. The N-550B should stabilize on the subject's physiological signal in about
15 to 30 seconds. Verify that the oxygen saturation and pulse rate values
are reasonable for the subject.
1. Press and hold the Alarm Silence button for more than three seconds.
Verify the following:
•“Alarm Silence Duration Time” is displayed for approximately three
seconds.
•After three seconds:
•a steady tone is heard at the default volume setting
•the %SpO2 dis play indicates “VOL”
•the
2. Press the
button, until an alarm volume setting of “1” is displayed. Verify that the
alarm volume has decreased but is still audible.
3. Press the
increase the alarm volume setting to a maximum value of “10.” Verify
that the volume increases.
4. Press the
button, until a comfortable audio level is obtained.
Pulse Rate display indicates the default setting of “4”
Adjust Down button, while holding the Alarm Silence
Adjust Up button, while holding the Alarm Silence button, to
Adjust Down button, while holding the Alarm Silence
18
5. Release the
Alarm Silence button. The tone stops.
Pulse Tone Volume Control
1. When a valid pulse is acquired, press the Adjust Up button and verify
Performance Verification
that the sound level of the beeping pulse tone increases.
2. Press the
Adjust Down button and verify that the beeping pulse tone
decreases until it is no longer audible.
3. Press the
Adjust Up button to return the beep volume to a comfortable
level.
Pulse Oximetry Functional Tests
These tests utilize the pulse oximetry functional tester (SRC-MAX) to verify
the performance of the N-550B pulse oximeter.
All of these tests should be done in sequence.
1 — DOC-10 Cable Connector6 — % Modulation Select Button
2 — Red LED Drive Indicator7 — % SpO
3 — Not Used For N-550B8 — Light Level Selection Button
4 — Not Used For N-550B9 — Pulse Rate Selection Button
N-550B19
2 Select Button
Performance Verification
Introduction
5 — Battery Low Indicator10 — Infrared LED Drive Indicator
The SRC-MAX functional tester allows qualified technicians to functionally
test Nellcor O
XIMAX technology-based pulse oximeters and OEM OXIMAX
technology-based monitors. The technician must perform the test setup
procedure before performing tests 1 through 4. The following is a brief
description of each test:
•Test Setup — This procedure establishes the baseline for all the other
tests. The T est Setup procedure must be performed before performing
any or all of the SRC-MAX tests.
Initial Setup
•Test #1: BPM — This procedure simulates an O
XIMAX sensor
attached to a patient indicating 60 bpm and 200 bpm. The test setup
procedure sets up Test 1 for 60 bpm.
•Test #2: SpO
2 — This procedure simulates an OXIMAX sensor
attached to a patient, indicating 75 percent blood oxygen saturation
and 90 percent blood oxygen saturation. The test setup procedure sets
up Test 2 for 75 percent blood oxygen saturation.
•Test #3: Modulation — This procedure simulates an O
XIMAX sensor
attached to a patient indicating low and high pulse strength. The test
setup procedure sets up Test 3 for low pulse strength.
•Test #4: Light — This procedure simulates an O
XIMAX sensor
attached to a patient indicating low and high light level passing
through the patient at the sensor site. The test setup procedure sets up
Test 4 for low light level.
20
1. Sensor Port
Loading...
+ 115 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.