Nellcor N-550 User manual

Page 1
Page 2
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada. Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor,
OxiCliq, OxiBand, Dura-Y, MAX-FAST, SatSeconds, PediCheck, Oxismart and O
XIMAX are
trademarks of Nellcor Puritan Bennett Inc. To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department,
or your local representative. Purchase of this instrument confers no express or implied license under any Nellcor Puritan
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign equivalents: 4,621,643; 4,653,498; 4,700,708; 4,770,179; Re. 35,122; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; 5,485,847; 5,533,507; 5,662,106; 5,853,364, and 6,083,172.
Page 3

Contents

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warnings .................................................................................1
Cautions ..................................................................................2
Notes .......................................................................................3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use for the N-550 .....................................................5
Description of Controls, Indicators, and Symbols . . . . . . . . . . . . . . . . . . 7
Identification of Front Panel Buttons and Symbols ..................7
Identification of Rear Panel Components ................................8
N-550 Symbols ........................................................................8
Description of Controls ............................................................9
Description of Displays and Indicators ..................................11
Description of Audible Indicators ...........................................13
Setting up the N-550 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
List of Components ...............................................................16
Connecting the N-550 to AC Power ......................................17
Connecting a Sensor to the N-550 ........................................17
Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Operating the N-550 on Battery Power .................................19
Low Battery Indicator .............................................................20
Using the N-550 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Turning on the N-550 ............................................................21
Discussion ....................................................................21
Procedure ....................................................................21
Sensor Attached ..........................................................24
No Sensor Attached .....................................................26
Sensor Message ...................................................................26
N-550 i
Page 4
Contents
Setting the Pulse Beep Volume ............................................ 27
Setting the Alarm Volume ..................................................... 28
Setting Alarm Silence Duration .............................................28
Discussion ................................................................... 28
Procedure .................................................................... 29
Disabling Audible Alarms ......................................................30
Discussion ................................................................... 30
Procedure .................................................................... 30
Verify Patient Settings ...........................................................31
Alarm Limits Changed Indicator ............................................33
Setting Alarm Limits ..............................................................34
Discussion ................................................................... 34
Procedure .................................................................... 34
Setting SatSeconds Duration ................................................36
Discussion ................................................................... 36
Procedure .................................................................... 36
Setting the Data Port Baud Rate ...........................................37
Discussion ................................................................... 37
Procedure .................................................................... 37
Setting the Data Port Protocol ..............................................39
Clearing Trend Information ...................................................40
N-550 Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Trend Data Operation ........................................................... 43
Trend Data ............................................................................43
#1: Trend Print .............................................................44
#2: Trend Clear ............................................................44
#3: Not Used ................................................................44
#4: Baud Rate ..............................................................44
#5: Data Port Printout .................................................. 45
Option 1 ..............................................................45
Option 2 ..............................................................45
Using the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Overview ............................................................................... 47
Connecting to the Data Port ..................................................47
Data Port Pinouts ..................................................................48
ii
Page 5
Contents
Data Port Setup .....................................................................49
Discussion ....................................................................49
Procedure ....................................................................50
Nurse Call Interface ...............................................................53
Setting Nurse Call RS-232 Polarity ..............................54
Setting Nurse Call Relays Normally Open/Closed .......54
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Printing N-550 Real-Time Data .............................................57
Trend Data Printout ...............................................................59
Column Headings ........................................................60
Data Source .................................................................61
Device/Software Revision Level ..................................61
Alarm Limits .................................................................61
N-550 Mode .................................................................62
Data Column Headings ................................................62
Time .............................................................................62
Patient Data .................................................................63
Operating Status ..........................................................63
Sensors and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Selecting a Sensor ................................................................65
Biocompatibility Testing .........................................................68
Optional Accessories .............................................................68
Foot Switch ..................................................................69
Visual Alarm Indicator ..................................................69
Pole Mount Bracket ......................................................70
Performance Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Performance Verification .......................................................71
N-550 Performance Considerations ......................................71
Dysfunctional Hemoglobins .........................................72
Anemia .........................................................................72
Saturation .....................................................................72
Pulse rates ...................................................................72
Sensor Performance Considerations ....................................72
N-550 iii
Page 6
Contents
Operator’s Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Error Codes ...........................................................................77
Corrective Action ...................................................................78
EMI (Electro-magnetic Interference) ..................................... 81
Obtaining Technical Assistance ............................................82
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Returning the N-550 ..............................................................85
Service .................................................................................. 85
Periodic Safety Checks .........................................................86
Cleaning ................................................................................ 86
Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Description of Alarms ............................................................87
Factory Defaults ....................................................................88
SatSeconds Display ..............................................................89
Describing SatSeconds .........................................................90
SatSeconds “Safety Net” ......................................................92
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
Oximetry Overview ................................................................93
Automatic Calibration ............................................................94
Functional versus Fractional Saturation ................................95
Measured versus Calculated Saturation ............................... 95
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
Performance .........................................................................97
Electrical .............................................................................101
Environmental Conditions ...................................................102
Physical Characteristics ......................................................104
Compliance ......................................................................... 105
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
iv
Page 7

Figures

Contents
Figure 1: Front Panel Buttons and Symbols .........................7
Figure 2: Rear Panel Symbols ..............................................8
Figure 3: Data Port Pin Layout ...........................................49
Figure 4: Real-Time Data Printout ......................................59
Figure 5: Trend Data Printout .............................................60
Figure 6: Alarm Response with SatSeconds ......................91
Figure 7: Oxyhemoglobin Dissociation Curve ....................96
N-550 v
Page 8
Contents

Tables

Table 1: Data Port Pinouts .................................................. 48
Table 2: Nellcor Oximetry Sensor Models
and Patient Weights .........................................66
Table 3: Operator’s Menu ...................................................75
Table 4: Error Codes ...........................................................78
Table 5: Factory Default Settings ........................................88
Table 6: Tone Definition ......................................................98
vi
Page 9

Safety Information

Warnings

Warnings are identified by the WARNING symbol shown above.
W arnings alert the user to potential serious outcomes (death, injury, or adverse events) to the patient or user.
WARNING: Explosion hazard. Do not use the N-550 pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Pulse oximetry readings and pulse signals can be affected by certain ambient environmental conditions, sensor application errors, and certain patient conditions. See the appropriate sections of the manual for specific safety information.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.
N-550 1
Page 10
Safety Information

Cautions

Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and effective use of the N-550.
CAUTION: When connecting the N-550 to any instrument, verify proper operation before clinical use. Both the N-550 and the instrument connected to it must be connected to a grounded outlet. Accessory equipment connected to the N-550's data interface must be certified according to IEC Standard 60950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port (N-550 data port connector) configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC Standard 60601-1-2. The N-550 accuracy may degrade if it is connected to secondary I/O devices when the instrument is not connected to earth reference.
CAUTION: Do not lift the N-550 by the sensor cable or power cord because the cable or cord could disconnect from the N-550, causing damage to the N-550 or injuring the patient.
CAUTION: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
2
Page 11

Notes

Safety Information
Notes are identified by the Note symbol shown above.
Notes provide additional useful information.
N-550 3
Page 12
Blank Page
Page 13

Introduction

WARNING: The N-550 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. Do not make any clinical judgments based on the oximeter's measurements only.
Intended Use for the N-550
The N-550 Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO with neonatal, pediatric, and adult patients during both no-motion and motion conditions and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only.
2) and pulse rate. The N-550 is intended for use
Note: Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility .
Use with any particular patient requires the selection of an appropriate oxygen transducer (sensor) as described in this Operator's Manual.
Motion performance claims are applicable to models MAX-A, MAX-AL, MAX-P, MAX-N, and MAX-I Nellcor O
XIMAX
TM
oximetry sensors.
N-550 5
Page 14
Blank Page
Page 15

Description of Controls, Indicators, and Symbols

Identification of Front Panel Buttons and Symbols

Figure 1: Front Panel Buttons and Symbols
1 — SpO
2 Sensor Port 12 — SatSeconds Alarm Limit
Button 2 — Power On/Off Button 13 — SpO 3 — %SpO 4 — Pulse Amplitude
2 Display 14 — Motion Indicator
15 — Sensor Off Indicator
2 Alarm Limit Button
Indicator 5 — Pulse Rate Display 16 — Sensor Message Indicator 6 — Alarm Silence Button 17 — Pulse Search Indicator 7 — Alarm Silence Indicator 18 — Data In Sensor Indicator 8 — Adjust Up Button 19 — Low Battery Indicator 9 — Adjust Down Button 20 — AC Power Indicator 10 — Pulse Rate Alarm Limit
21 — Speaker
Button
11 — SatSeconds
N-550 7
TM
Display
Page 16
Description of Controls, Indicators, and Symbols

Identification of Rear Panel Components

1 — Data Port Connector 3 — AC Power Connector
N-550 Symbols
2 — Foot Switch and Visual
Alarm Connector
Figure 2: Rear Panel Symbols
The symbols that are located on the rear panel of the N-550 are as follows:
1 — Data Interface 3 — Foot Switch 2 — Caution - Do not connect
while power is on
4 — Equipotential Connector
(ground)
4 — Equipotential Terminal
(ground)
8
Page 17
Description of Controls, Indicators, and Symbols
The symbols that are located on the front panel of the N-550 are as follows:
1 — T ype BF Applied Part -
Not Defibrillator Proof 2 — Pulse Rate 7 — Data In Sensor 3 — Motion 8 — Low Battery 4 — Sensor Off 9 — AC Power/Battery Charge 5 — Sensor Message

Description of Controls

Note: Each button press, except the Power On/Off button, should
result in either a valid or an invalid button tone. If the button pressed fails to emit a tone, contact qualified service personnel.
6 — Pulse Search
The Power On/Off button is used to turn the N-550 on or off.
N-550 9
Page 18
Description of Controls, Indicators, and Symbols
The Alarm Silence button is used to silence current alarms for the alarm silence duration period. When an alarm has been silenced, pressing the button again reactivates, or “unsilences” the alarm. It is also used to view and adjust alarm silence duration and alarm volume.
The Adjust Up button is used to increase alarm limit values, alarm silence duration, pulse beep volume, alarm volume, and data port baud rate. The Adjust Up button is used to select the communication protocol and time settings. Press the Adjust Up button one time to increase the display by one digit. Holding the Adjust Up button down for more than one second cause the display digits to scroll.
The Adjust Down button is used to decrease alarm limit values, alarm silence duration, pulse beep volume, alarm volume, and data port baud rate. The Adjust Down button is used to select the communication protocol and time settings. Press the Adjust Down button one time to decrease the display by one digit. Holding the Adjust Down button down for more than one second cause the display digits to scroll.
The SpO When the SpO
2 Alarm Limit button is used to view the SpO2 alarm limit.
2 Alarm Limit button is pressed at the same time as the
Pulse Rate Alarm Limit button for approximately three seconds, the menu options are enabled.
The Pulse Rate Alarm Limit button is used to view the pulse rate alarm limit. When the Pulse Rate Alarm Limit button is pressed at the same time as the SpO
2 Alarm Limit button for approximately three
seconds, the menu options are enabled.
The SatSeconds Alarm Limit button is used to view the SatSeconds alarm limit. When pressed, the menu options are enabled. The Adjust Up and Adjust Down buttons are used to change the SatSeconds limit settings.
10
Page 19
Description of Controls, Indicators, and Symbols

Description of Displays and Indicators

The Pulse Amplitude Indicator (blip bar). A 10-segment LED that indicates pulse beat and shows the relative pulse amplitude. As the detected pulse becomes stronger, more bars light with each pulse.
The %SpO
2 Display. Shows the saturation level of oxygenated
hemoglobin. The display value flashes zero during loss-of-pulse alarms and flashes the SpO
2 value when the SpO2 is outside the alarm
limits. During pulse search, the N-550 continues to update the display. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the SpO
2 value (100.).
The Pulse Rate Display . Shows the pulse rate in beats per minute. It flashes zeros during loss-of-pulse alarms and flashes the beats per minute value in red when the pulse rate is outside of the alarm limit. During pulse search, the N-550 continues to update the display. Pulse rates outside of the pulse rate range (20 to 250 bpm) are displayed as the closest value within the range. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the pulse rate value (112.).
The AC Power Indicator. Lights continuously when the N-550 is connected to AC power. It also indicates that the battery is charging. It is off when the N-550 is being powered by its internal battery.
The Low Battery Indicator. Lights cont inuously to indicate that 15 or fewer minutes of battery capacity remain. The Low Battery indicator flashes when the battery is critically low.
N-550 11
Page 20
Description of Controls, Indicators, and Symbols
The Alarm Silence Indicator. Lights continuously when an audible alarm has been silenced. It flashes when the alarm silence duration has been set to OFF.
The Motion Indicator. Lights continuously whenever the O algorithm detects the presence of artifact independent of its severity or the impact on the SpO indicator and the pulse search indicator are simultaneously lit, it is an indication that the artifact is significant and/or has been persistent.
The Pulse Search Indicator. Lights continuously prior to initial acquisition of a pulse signal and during prolonged and challenging monitoring conditions. It flashes during a loss-of-pulse signal.
The Sensor Off Indicator is lit when either the sensor is invalid, or no longer on the patient’ s finger. It usually indicates that the sensor is not on the patient.
The Sensor Message Indicator. Lights when the N-550 cannot determine an SpO
2 level or a pulse rate. The Sensor Message
recommendations for improving the signal are:
®
XIMAX
2 or pulse rate values. Whenever the motion
12
Reposition sensor
Check or change adhesive wrap
Choose alternate site
Warm site
Cover sensor
Use forehead, nasal, or ear sensor (adult patients only)
•Use O
XIMAX adhesive sensor
Secure cable
Secure with headband (M
AX-FAST)
Remove nail polish
Loosen sensor (too tight)
Isolate external interference (electrosurgical device, cell phone)
Clean site (MAX-R)
Page 21
Description of Controls, Indicators, and Symbols
The Data In Sensor Indicator. The indicator blinks for approximately one minute when initially connected to the N-550 to indicate that the attached O
XIMAX sensor contains a patient sensor event record. The
indicator lights continuously to indicate that the attached sensor memory is full. The indicator does not light when there is no data in the sensor, ev en though a valid sensor is connected to the N-550.
TM
The SatSeconds
Indicator. Fills in clockwise as the SatSeconds alarm management approaches the SatSeconds alarm limit threshold. All segments of the SatSeconds indicator flash during a SatSeconds alarm. When a SatSeconds setting other than OFF is selected, the green LED at the top of the SatSeconds indicator will light. The green LED at the 12-o'clock position indicates that SatSeconds alarm management is engaged.

Description of Audible Indicators

Following are descriptions of N-550 audible indicators.
Power-On Self-Test Pass
A 1-second tone indicating that the N-550 has been turned on and has successfully completed the power-on self-test
Valid Button Press A short, medium-pitched tone indicating
that an appropriate button has been pressed
Invalid Button Press A short, low-pitched tone indicating that a
button has been pressed that is not appropriate for the current state of the N-550
High Priority Alarm A high-pitched, fast-pulsating tone
indicating loss of pulse with no patient motion
Medium Priority Alarm
A medium-pitched, normal-pulsating tone indicating an SpO
2 or pulse rate limit
violation
N-550 13
Page 22
Description of Controls, Indicators, and Symbols
Low Priority Alarm A low-pitched, slow-pulsating tone
indicating a sensor disconnect, low battery, or N-550 failure
Alarm Silence Reminder
Three beeps that sound approximately every 3 minutes when alarms are silenced with the alarm silence duration set to OFF
Pulse Beep A single beep sounds for each detected
pulse. The pitch changes as monitored
2 values increase or decrease.
SpO
Volume Setting Tone A continuous tone that is used to adjust
the alarm volume
Confirmation T o ne Three beeps sound to indicate that default
settings have been saved or reset to factory defaults or trend data has been deleted
14
Page 23
Setting up the N-550
WARNING: Explosion hazard. Do not use the N-550 pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Pulse oximetry readings and pulse signals can be affected by certain ambient environmental conditions, sensor application errors, and certain patient conditions. See the appropriate sections of the manual for specific safety information.
WARNING: To ensure patient safety, do not place the N-550 in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING: Disconnect the N-550 and Nellcor sensor from the patient during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns. The N-550 may affect the MRI image; the MRI unit may affect the accuracy of oximeter measurements.
WARNING: To ensure accurate performance and prevent device failure, do not subject the N-550 to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure.
N-550 15
Page 24
Setting up the N-550
WARNING: Do not use an N-550, sensor, cables, or connectors that appear to be damaged.
WARNING: The N-550 is not defibrillator-proof. However, it may remain attached to the patient during defibrillation or while an electrosurgical unit is in use, but the readings may be inaccurate during use and shortly thereafter.
WARNING: In the USA, do not connect the N-550 to an electrical outlet controlled by a wall switch because the N-550 may be accidentally turned off.
WARNING: Use only the DOC-10 pulse oximetry cable with the N-550. Use of another sensor cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the sensor port. Do not connect any device other than a Nellcor-approved sensor to the sensor connector.

List of Components

1 — N-550 Pulse Oximeter
1 — Nellcor Sensor or Assortment Pack
1 — DOC-10 Pulse Oximeter Cable
1 — N-550 Operator's Manual
1 — Hospital-Grade Power Cord or power cord appropriate for
1 — Sensor Accuracy Grid
1 — Quick Guide
16
country of sale
Page 25
Connecting the N-550 to AC Power
1 — Power Connector
1. Plug the female connector of the power cord into the N-550 AC
power connector (1).
2. Plug the male connector of the power cord into a properly
grounded AC outlet.
Setting up the N-550
3. Verify that the AC POWER INDICATOR is lit.

Connecting a Sensor to the N-550

WARNING: Use only the DOC-10 pulse oximetry cable with the N-550. Use of another sensor cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the sensor port. Do not connect any device other than a Nellcor-approved sensor to the sensor connector.
N-550 17
Page 26
Setting up the N-550
.
1 — SpO2 Sensor Port
1. Connect a DOC-10 pulse oximetry cable to the SpO2 Sensor Port
(1) on the front of the N-550.
2. Connect an SpO
2 Sensor to the other end of the DOC-10 pulse
oximetry cable. Plug the sensor connector firmly into the DOC-10 pulse oximetry cable.
18
Page 27

Battery Operation

WARNING: Dispose of an old battery by following local guidelines for disposal of lead acid batteries.
Operating the N-550 on Battery Power
The N-550 has an internal battery that may be used to power the N-550 during transport or when AC power is not available. A new, fully charged battery will provide at least 1.5 to 2 hours of monitoring time under the following conditions: no audible alarms sound and no serial output devices are attached.
Note: Whenever the N-550 is connected to AC power, the battery is
being charged. Therefore, it is recommended that the N-550 remain connected to AC power when not in use. This will make a fully charged battery availabl e for use at any t ime .
The N-550 cannot operate with a dead battery (even when plugged in). Before attempting to turn on an N-550 with a depleted battery, first plug the N-550 into an AC outlet to allow the battery to charge for a few minutes. The N-550 may then be powered on.
To charge a low or dead battery, connect the N-550 to AC power. A full charge of a dead battery takes 11 hours while the N-550 is turned off or 12 hours while the N-550 is on.
When all of the following conditions are present for 15 minutes, the N-550 will automatically shut down:
N-550 is running on battery power
No buttons have been pressed
No pulse has been detected (for example, when no patient is connected to the sensor or the sensor is disconnected)
N-550 19
Page 28
Battery Operation
No alarms are present (other than low battery or a

Low Battery Indicator

The Low Battery Indicator lights and a low priority alarm begins to sound when 15 minutes but not more than 20 minutes of monitoring time remain on the existing battery charge. This alarm cannot be silenced while running on battery power. Connecting the N-550 to AC power will silence the alarm. If the N-550 is not connected to AC power within approximately 15 minutes, the N-550 will shut off.
Note: As the battery is used and recharged over a period of time, the
It is recommended that qualified service personnel replace the internal battery every 24 months.
non-correctable error)
amount of time between the onset of the low battery alarm and the N-550 shut-off may become shorter.
20
CAUTION: If the N-550 is to be stored for 2 months or longer, notify service personnel to remove the battery from the N-550 prior to storage. Recharge the battery when it has not been charged for 2 or more months.
The Low Battery Indicator flashes and a high priority alarm begins to sound when the battery reaches the lowest battery voltage at which an N-550 can support normal operation. This alarm cannot be silenced while running on battery power. If the N-550 is not connected to AC power, the N-550 will shut off after 10 seconds.
Page 29
Using the N-550
Turning on the N-550

Discussion

Before using the N-550 in a clinical setting, you must verify that the N-550 is working properly and is safe to use. Proper working condition can be verified by successful completion of the Power-On Self-Test (POST), described in the following section.
WARNING: The N-550 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the N-550 should be observed to verify normal operation in the configuration in which it is used.
CAUTION: If any indicator or display element does not light, or the speaker does not sound, do not use the N-550. Instead, contact qualified service personnel, your local Nellcor representative, or Nellcor's Technical Services Department, 1.800.635.5267.
Note: The N-550 should complete the POST function within 12
seconds.

Procedure

1. Turn on the N-550 by pressing and holding the Power On/Off
button for more than one second.
N-550 21
Page 30
Using the N-550
2. The N-550 displays/sounds:
Display Sound
5 (in pulse rate left window) one beep tone 5 (in pulse rate center window) one beep tone 0 (in pulse rate right window) one beep tone n (in SpO n (in SpO
2 left window) none 2 center window) none
3. The N-550 automatically starts the Power-On Self-Test (POST),
which tests N-550 circuitry and functions.
CAUTION: During POST (immediately after power-up), confirm that all display segments and indicators light, and the speaker sounds a 1-second pass tone.
22
4. While performing POST, the self-test display appears for
approximately 2 to 4 seconds. During this time:
All indicators illuminate
All segments of all numeric digits light and change from red to green
Page 31
Using the N-550
All segments of the Pulse Amplitude Display light
All segments of the SatSeconds indicator light
5. Once the display test portion of POST is complete, the N-550 software version is displayed for approximately 2 seconds.
Note: The device version illustrated is only a sample. The device
version identifies the hardware configuration and the software revision.
Device version numbers are often needed when calling Nellcor's Technical Services Department, 1.800.635.5267, or your local Nellcor representative for technical assistance. Write down the numbers and have them available prior to requesting technical assistance.
6. If the N-550 detects an internal problem during the self-test, an alarm tone sounds and the N-550 displays an Error Code and the corresponding number. See Troubleshooting on page 77.
N-550 23
Page 32
Using the N-550
7. Upon successful completion of the POST, the N-550 sounds a one-second tone indicating that the N-550 has passed the test.
WARNING: If you do not hear the POST pass tone, do not use the N-550.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.

Sensor Attached

24
Note: In addition to serving as the POST pass verification, the POST
pass tone also functions as an audible confirmation that the speaker is performing properly. If the speaker does not function, the alarm sounds cannot be heard.
The Pulse Search indicator and the Sensor Off indicator light and the N-550 displays dashes in the %SpO
2 and Pulse Rate displays while
the N-550 is searching for a valid pulse.
Page 33
Using the N-550
1 —Sensor Off 2 — Pulse Search Indicator
When a valid pulse is detected, the N-550 enters the Monitoring Mode and displays patient parameters. Look for movement of the blip bar and, if the pulse beep volume is not 0, listen for pulse beeps to verify that displayed measurement values are current and accurate.
When a sensor is connected to the N-550 and is not connected to a patient, the display reads dashes and the N-550 remains in the Pulse Search mode.
N-550 25
Page 34
Using the N-550

No Sensor Attached

Sensor Message

When the sensor is not attached the N-550 displays dashes (---) and the Pulse Search indicator is not lit, indicating that the N-550 failed to detect a sensor.
The Sensor Message feature is an indication that the sensor position or site needs to be considered. The Sensor Message indicator lights when the N-550 cannot determine an SpO
2 level or a pulse rate. The
Sensor Message recommendations for improving the signal are:
26
Reposition sensor
Check or change adhesive wrap
Choose alternate site
Warm site
Cover sensor
Use forehead, nasal, or ear sensor (adult patients only)
•Use O
XIMAX adhesive sensor
Secure cable
Secure with headband (M
AX-FAST)
Remove nail polish
Loosen sensor (too tight)
Page 35
Isolate external interference (electrosurgical device, cell phone)
Clean site (MAX-R)

Setting the Pulse Beep Volume

Using the N-550
1 — SpO2 Sensor Port
WARNING: Use only Nellcor- approved sensors and sensor cables.
1. Connect a DOC-10 pulse oximetry cable to the SpO
2 Sensor Port
(1) on the front of the N-550.
2. Connect an SpO
2 Sensor to the other end of the DOC-10 pulse
oximetry cable.
3. Place the SpO
2 Sensor on the patient or yourself.
4. Turn on the N-550.
5. Press and hold the Adjust Up or Adjust Down button to increase or decrease the pulse beep tone volume.
N-550 27
Page 36
Using the N-550

Setting the Alarm Volume

With the N-550 in the normal monitoring mode:
1. Press and hold the Alarm Silence button until the alarm volume display appears.
2. While continuing to press the Alarm Silence button, press the Adjust Up button or Adjust Down button to increase or decrease the alarm volume.

Setting Alarm Silence Duration

Discussion

WARNING: Do not silence an audible alarm or decrease its volume if patient safety could be compromised.
Alarms can be silenced for a preset period called alarm silence duration. To view the current setting, press and hold the Alarm Silence button until 30 SEC, 60 SEC, 120 SEC, or OFF is displayed. To adjust the settings, press and hold the Alarm Silence button until 30 SEC, 60 SEC, 120 SEC, or OFF is displayed. Continue holding the
28
Page 37
Using the N-550
Alarm Silence button, and use the Adjust Up button or Adjust Down button to increase or decrease the value. Possible values are 30, 60, 90, or 120 seconds, or OFF. The OFF selection is discussed under Disabling Audible Alarms on page 30.
If the Alarm Silence button is pressed during the alarm silence duration, the alarm silence duration is ended and the audible alarms are re-enabled.
Visual indications of an alarm condition cannot be turned off. For example, if the %SpO be silenced for the alarm silence duration, but the %SpO
2 lower alarm limit is exceeded, the alarm can
2 value will
continue to flash.
If the alarm condition is still present when the alarm silence duration has elapsed, the alarm will sound.
The power-on default setting for audible alarm silence duration is set at the factory to 60 seconds. The default setting can be adjusted by service personnel as described in the N-550 service manual.

Procedure

With the N-550 in the normal monitoring mode:
1. Press the Alarm Silence button until XX SEC is displayed.
Durations are: OFF, 30, 60, 90, and 120 seconds.
Note: For steps 2 and 3, the Adjust Up and Adjust Down buttons
must be pressed while pressing the Alarm Silence button.
N-550 29
Page 38
Using the N-550
2. While pressing the Alarm Silence button, press and hold the Adjust Up button to increase alarm silence duration to Off, 30, 60, 90, or 120 seconds.
3. While pressing the Alarm Silence button, press and hold the Adjust Down button to decrease alarm silence duration to OFF, 30, 60, 90, or 120 seconds.
Note: Releasing the Adjust Up or Adjust Down button sets the alarm
silence duration.

Disabling Audible Alarms

Discussion

Setting the alarm silence duration to OFF means that the N-550 will produce no audible alarms.

Procedure

30
V isual indications of an alarm condition are not affected by disabling the audible alarms.
The ability to set the alarm silence duration to OFF can be enabled or disabled by qualified service personnel as described in the service manual.
WARNING: Do not silence the audible alarm or decrease its volume if patient safety could be compromised.
With the N-550 in the normal monitoring mode:
Page 39
Using the N-550
1. Press the Alarm Silence button until XX SEC is displayed.
2. While pressing the Alarm Silence button, press and hold the
Adjust Up button until OFF is displayed. Release buttons.
3. The Alarm Silence indicator will flash, indicating that the alarm
sounds are disabled. The N-550 will sound three beeps approximately every three minutes to warn the user that the alarm sound has been silenced.

Verify Patient Settings

With the N-550 in the normal monitoring mode:
1. Press the SpO
upper alarm limit.
N-550 31
2 Alarm Limit button to view the current %SpO2
Page 40
Using the N-550
2. Press the SpO2 Alarm Limit button twice to view the current %SpO
2 lower alarm limit.
32
3. Press the Pulse Rate Alarm Limit button to view the current Pulse Rate upper alarm limit.
4. Press the Pulse Rate Alarm Limit button twice to view the current Pulse Rate lower alarm limit.
Page 41
Using the N-550
5. Press the SatSeconds Alarm Limit button to view the current
SatSeconds setting.
Note: The SatSeconds (12 O’clock) indicator lights, indicating that
SatSeconds units are being reviewed. The SatSeconds (12 O’clock) indicator lights for all SatSeconds except Off.

Alarm Limits Changed Indicator

If an alarm limit is changed from the N-550’s power-on default setting, a decimal point appears after the applicable displayed value, during patient monitoring and when alarm limits are viewed. The decimal point remains on the display until the N-550 is turned off or the limit is returned to the default value.
N-550 33
Page 42
Using the N-550
Alarm limits that have been changed from the default setting are identified by a decimal point (.) after the displayed reading (%SpO or Pulse Rate).

Setting Alarm Limits

Discussion

Alarm limits determine the upper and lower points of patient data at which the N-550 will sound an alarm.

Procedure

With the N-550 in the normal monitoring mode:
2
1. Press SpO
2 Alarm Limit button to view the current %SpO2 upper
alarm limit.
2. Press the Adjust Up button or Adjust Down button to increase or decrease the alarm limit setting.
Note: When an alarm limit is changed from power-on default, the
N-550 displays a decimal point (.) after the changed parameter.
34
Page 43
Using the N-550
3. Press the SpO2 Alarm Limit button twice to view the current
%SpO
2 lower alarm limit.
4. Press the Adjust Up button or Adjust Down button to increase or
decrease the alarm limit setting.
5. Press the Pulse Rate Alarm Limit button to view the current upper
alarm limit for Pulse Rate.
6. Press the Adjust Up button or Adjust Down button to increase or
decrease the alarm limit setting.
N-550 35
Page 44
Using the N-550
7. Press the Pulse Rate Alarm Limit button twice to view the current lower alarm limit for Pulse Rate.
8. Press Adjust Up button or Adjust Down button to increase or decrease the alarm limit setting.

Setting SatSeconds Duration

Discussion

Procedure

36
Refer to Describing SatSeconds on page 90 for a description of the SatSeconds function.
With the N-550 in the normal monitoring mode:
1. Press the SatSeconds Alarm Limit button. The current SatSeconds setting is displayed.
Page 45
Note: The possible settings for SatSeconds are Off, 10, 25, 50, and
100 seconds.
2. Press the Adjust Up button or Adjust Down button to select the
desired SatSeconds setting.

Setting the Data Port Baud Rate

Using the N-550

Discussion

Procedure

The baud rate determines the speed at which the N-550 sends data to the attached equipment (printer or portable computer). The baud rate is determined by the capabilities of the attached equipment.
With the N-550 in the normal monitoring mode:
1. Simultaneously, press the SpO
Rate Alarm Limit button for 3 seconds. Menu option 1 will be displayed.
2 Alarm Limit button and Pulse
N-550 37
Page 46
Using the N-550
2. Press the Adjust Up button until menu item option 4 is displayed.
3. Press the SatSeconds Alarm Limit button. Current baud rate is displayed.
Press the Adjust Up button or Adjust Down button to select the desired baud rate. Possible settings are:
24 (2,400 bps)
96 (9,600 bps)
192 default (19,200 bps).
38
Page 47

Setting the Data Port Protocol

With the N-550 in the normal monitoring mode:
Using the N-550
1. Simultaneously, press the SpO
2 Alarm Limit button and Pulse
Rate Alarm Limit button for 3 seconds. Menu option 1 will be displayed.
2. Press the Adjust Up button until menu item option 5 is displayed.
3. Press the SatSeconds Alarm Limit button. Current protocol is
displayed.
N-550 39
Page 48
Using the N-550
4. Press the Adjust Up button or Adjust Down button to select the desired protocol. Possible settings are:
•1 - ASCII
2 - External equipment communications. Refer to the external
equipment manuals for the interfacing instructions.

Clearing Trend Information

40
With the N-550 in the normal monitoring mode:
1. Simultaneously press and hold the SpO
2 Alarm Limit button and
Pulse Rate Alarm Limit button for at least 3 seconds. Option 1 will be displayed.
2. Press the Adjust Up button to select Option 2.
Page 49
Using the N-550
3. Press the SatSeconds Alarm Limit button to clear the trend data.
4. The N-550 emits 3 beeps, indicating that data is cleared.
N-550 41
Page 50
Blank Page
Page 51
N-550 Trend

Trend Data Operation

From the initial measurement of a patient, trend data (a data point) is stored in memory every 4 seconds. Up to 50 alarm limit changes can also be stored in trend data. The N-550 can store up to 24 hours of trend data.
The N-550 trend data will be lost if the coin cell battery fails or is removed. The coin cell battery is located on the main circuit board.
CAUTION: Changing alarm limit settings uses trend memory space. Change alarm limits only as needed.
Note: Trend memory always contains the most recent 24 hours of
data, with newly collected data overwriting the oldest data on a rolling basis. The N-550 continues to record data points as long as the N-550 is powered on and an initial patient measurement has been made, with “blank” data points collected if no sensor is connected to the N-550 or patient. “Blank” data will overwrite older patient data if the memory becomes full. Therefore, if you want to save old patient data, it is important that you turn your N-550 off when you are not monitoring a patient, and that you download the trend memory before it fills up and overwrites the old data with new data (or “blank” data).

Trend Data

Trend data information may be retrieved or cleared through the N-550 data port using options available in a display menu.
N-550 43
Page 52
N-550 Trend

#1: Trend Print

To access the menu options, simultaneously press the SpO2 Alarm Limit and Pulse Rate Alarm Limit buttons until Option 1 appears on the display .
Then, using the Adjust Up button and Adjust Down button, you may scroll through the available menu options as follows:
Allows printing of the existing trend data. The output may be viewed on a PC using the Hyper Terminal program. You must press the
SatSeconds Alarm Limit button to initiate printing. Refer to Trend Data Printout on page 59.

#2: Trend Clear

#3: Not Used

#4: Baud Rate

44
Clears the existing trend data. To activate, press the SatSeconds Alarm Limit button.
Allows the interconnection to various printers. Refer to Setting the Data Port Baud Rate on page 37. The baud rate selections are:
24 (2,400 baud rate)
Page 53

#5: Data Port Printout

Selections are as follows:
Option 1
Printout in ASCII characters.
Option 2
External equipment communications. Refer to the external equipment manual for the interfacing instructions.
N-550 Trend
96 (9,600 baud rate)
192 default (19,200 baud rate)
N-550 45
Page 54
Blank Page
Page 55

Using the Data Port

Overview

Patient data can be obtained through the data port on the back of the N-550 by connecting it to an attached PC or serial printer.
When connecting the N-550 to a printer or PC, verify proper operation before clinical use. Both the N-550 and the printer or PC must be connected to a grounded AC outlet. The N-550 protocol setting must be ASCII.
Any printer or PC connected to the N-550's data port must be certified according to IEC Standard 60950. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects a printer or PC to the data output port configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC Standard 60601-1-2.

Connecting to the Data Port

The N-550 data port may be connected to the printer or PC by using a cable terminated with an AMP connector (AMP part number 747538-1), ferrule (AMP part number 1-747579-2), and compatible pins (AMP part number 66570-2). The cable should be no more than 25 feet (7.6 meters) in length. The external ITE (Information Technology Equipment) device must be certified to UL-1950 or IEC-60950.
N-550 47
Page 56
Using the Data Port
The cable used must have a braided shield providing 100% coverage, such as a Belden cable (Belden part number 9609) or equivalent. The shield must have a 360-degree connection to the metal shell on the N-550's DB-15 connector and to the connector on the PC or serial printer. Do not create sharp bends in the cable, as this may tear or break the shielding.
No hardware flow control is used. However, in the ASCII mode, XON/XOFF flow control is supported.

Data Port Pinouts

TXD represents the Transmit Data line, and RXD is the Receive Data line.
The pinouts for the data port are listed in Table 1.
Table 1: Data Port Pinouts
Pin Signal Name
1 RXD+ (RS-422 positive input) 2 RXD (RS-232 input) 3 TXD (RS-232 output) 4 TXD+ (RS-422 positive output)) 5 Signal Ground (isolated from Earth Ground) 6 NC (No connection) 7 Normally Open (relay closure nurse call, normally open) 8 Normally Closed (relay closure nurse call, normally
closed)
9 RXD- (RS-422 negative output) 10 Signal Ground (isolated from earth ground) 11 Nurse Call (RS-232 level output)
48
Page 57
Using the Data Port
Table 1: Data Port Pinouts
Pin Signal Name
12 TXD- (RS-422 negative output) 13 NC 14 NC 15 Nurse Call Common
The pin layouts (as viewed from the rear panel of the N-550) are illustrated in Figure 3. The conductive shell is connected to earth ground when connected to a PC or printer.
Figure 3: Data Port Pin Layout
Pins 2, 3, 5, and 11 provide data in RS-232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format. TXD+ and TXD­are the differential transmit data pair. RXD+ and RXD- are the differential receive pair.

Data Port Setup

Discussion

Use the Data Port Setup procedure to configure the N-550 data port baud rate and protocol.
N-550 49
Page 58
Using the Data Port

Procedure

With the N-550 in the normal operating mode:
1. Simultaneously press and hold the SpO
2 Alarm Limit button and
Pulse Rate Alarm Limit button for at least 3 seconds. Option 1 will be displayed.
2. Press the Adjust Up button until Option 4 is displayed.
50
3. Press the SatSeconds Alarm Limit Button to select Option 4. The selected baud rate will be displayed.
Page 59
Using the Data Port
Note: The available selections are:
24 = 2400 baud rate
96 = 9600 baud rate
192 = 19200 baud rate
4. Press the Adjust Up button or the Adjust Down button to select
the desired baud rate.
5. Press the SatSeconds Alarm Limit button to save the desired baud
rate.
6. Simultaneously press and hold the SpO
2 Alarm Limit button and
Pulse Rate Alarm Limit button for at least 3 seconds. Option 1 will be displayed.
N-550 51
Page 60
Using the Data Port
7. Press the Adjust Up button until Option 5 is displayed.
8. Press the SatSeconds Alarm Limit Button to select Option 5. The selected protocol will be displayed.
Note: The available selections are:
1 = Real time ASCII
2 = External equipment communications. Refer to the
external equipment manuals for the interfacing instructions.
52
Page 61
9. Press the Adjust Up button or the Adjust Down button to display
10. Press the SatSeconds Alarm Limit button to save the desired

Nurse Call Interface

Using the Data Port
the desired protocol.
protocol.
WARNING: The nurse call feature should not be used as the primary source of alarm notification. The audible and visual alarms of the pulse oximeter, used in conjunction with clinical signs and symptoms, are the primary sources for notifying medical personnel that an alarm condition exists.
WARNING: The nurse call feature is not functional whenever the pulse oximeter alarms are silenced.
Caution: The nurse call function needs to be tested after it has been set up in your facility. The nurse call feature should be tested whenever setting up the N-550 pulse oximeter in a location that uses nurse call. If an attached O
XIMAX sensor is disconnected
from the patient, the N-550 remains in the Pulse Search Mode for awhile, then the N-550 displays “--- “ (3 dashes) in the %SpO
2
and pulse rate display. One way to test the nurse call function is to create an alarm condition (for example, sensor disconnect) and verify that your facility's nurse call system is activated.
N-550 53
Page 62
Using the Data Port
The nurse call feature of the N-550 is operational when the N-550 is powered by AC power or battery power. The nurse call feature of the N-550 works in conjunction with the nurse call system of your institution when the N-550 sounds an audible alarm.
The N-550 provides two different types of nurse call interfaces: an RS-232 format and relay closure. Both interfaces function when the N-550 is operating either on AC power or battery power.
The remote location is signaled anytime there is an audible alarm. If the audible alarm has been turned off or silenced, the nurse call function is also turned off.
Pin 11 on the data port is the RS-232 lev el nurse call signal and pin 5 or 10 is ground (see Table 1 on page 48). When there is no alarm condition, the voltage between pins 10 and 11 is -5 to -12 VDC. Whenever the N-550 is in an alarm condition, the output between pins 10 and 11 is +5 to +12 VDC. This is the default condition (normally low). There is a service menu to change the default condition. Refer to the N-550 service manual for the procedure.
Pins 7 and 15 provide a relay that closes when an alarm is sounding on the N-550. Pins 8 and 15 provide a relay that opens when an alarm is sounding. Pin 15 is a common lead for both relays.
Note: When the relay is closed there is approximately 27 ohms of
resistance.
Setting Nurse Call RS-232 Polarity
The nurse call polarity can be set to a high signal on an N-550 alarm condition or a low signal on an N-550 alarm condition. Refer to the N-550 service manual for setting the Nurse Call RS-232 polarity.

Setting Nurse Call Relays Normally Open/Closed

Data port pins 7 and 15 provide a relay that closes (nominally 27 ohms) when an alarm is sounding on the N-550. Pins 8 and 15
54
Page 63
Using the Data Port
provide a relay that opens when an alarm is sounding. Pin 15 is a common lead for both relays. The relay operates whether the N-550 is operating on AC power or battery.
N-550 55
Page 64
Blank Page
Page 65

Printing

Printing N-550 Real-Time Data
Real-time data is continuously sent to the data port on the back of the N-550. Patient data can be obtained through the data port by connecting to a computer or serial printer. When a real-time printout or display is being transmitted to a printer or computer, a new line is printed/displayed every 2 seconds. Column headings are printed/ displayed every 25 lines, or if one of the values in the column heading is changed.
Note: If the data output stops transmitting, the N-550 must be turned
off and then turned back on.
When connecting the N-550 to a printer or computer, verify proper operation before clinical use. Both the N-550 and the printer or PC must be connected to a grounded AC outlet.
Any printer or computer connected to the N-550's data port must be certified according to IEC Standard 60950. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects a printer or PC to the data output port configures a medical system and is therefore responsible that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC Standard 60601-1-2.
N-550 57
Page 66
Printing
1 — Data Port Connector
1. Connect the serial printer to the N-550's Data Port connector (1).
2. Turn on the printer.
3. Connect the N-550 to an AC outlet.
58
4. Turn on the N-550.
The printer will start printing real time trend data or the PC will start displaying real time data at a rate of one line every 2 seconds.
Page 67
Figure 4: Real-Time Data Printout
Printing

Trend Data Printout

The format of data displayed when a trend printout is requested is the same as the real-time data. The only differences are that “TREND” is displayed in the top row instead of the “CRC: XXXX” software verification number and there is no “Status” column.
Readings are displayed in 4-second intervals. The values on each row are an average for the 4-second period.
At the end of the printout, an “Output Complete” line indicates that the transmission was successful. If the “Output Complete” line is not present, a corruption of the data may have been detected and the data should be ignored.
N-550 59
Page 68
Printing

Column Headings

Figure 5: Trend Data Printout
Once a trend printout has begun, the N-550 must be turned off and back on again before a new trend printout can begin.
Column headings are displayed or printed after every 25 lines, or if one of the values in the column heading changes.
60
Page 69

Data Source

Data in the highlighted box above represents the model number of the N-550, in this case the N-550.

Device/Software Revision Level

The next data field tells the user the software level (Version 1.00.00) and a software verification number (CRC: XXXX). Neither of these numbers should change during normal operation. The numbers may change if the N-550 is serviced and receives a software upgrade.
Printing

Alarm Limits

The last data field in the top line indicates the upper and lower alarm limits for %SpO alarm setting (OFF). The SatSeconds setting may be OFF , 10, 25, 50, or 100 depending on the SatSeconds alarm setting. In the example above, the low alarm limit for SpO is 100%. Pulse Rate alarm limits are 40 to 170 bpm. SatSeconds alarm limit is set to off.
N-550 61
2 and for the pulse rate (PR), and the SatSeconds
2 is 85% and the upper alarm limit
Page 70
Printing
N-550 Mode
The N-550 mode is ADULT.

Data Column Headings

Actual column headings are in the third row of the column heading line. Patient data that is presented in the chart, from left to right, are the time that the patient data on the row was obtained, the current %SpO minute (bpm), the current Pulse Amplitude (PA), and the operating status of the N-550.
2 value being measured, the current Pulse Rate in beats per

Time

62
The Time column represents the N-550 real-time clock. Refer to the N-550 service manual to set the N-550 real-time clock.
Page 71

Patient Data

Printing
Patient data is highlighted in the display above. Parameter values are displayed directly beneath the heading for each parameter. In this example, the %SpO
2 is 100, and the pulse rate is 190 beats per
minute. An asterisk indicates that the parameter is outside the set limits. If no data for a parameter is available, three dashes (- - -) are displayed.
PA represents pulse amplitude. The number can range from 0 to 254. There are no alarm parameters for this value. It can be used for trending information as an indication of a change in pulse volume, relative pulse strength, or circulation.

Operating Status

The Status column indicates alarm conditions and operating status of the N-550. In this example, Pulse High (PH) means that the pulse rate upper alarm limit has been exceeded. The status codes are listed below. As many as four codes can be displayed at one time in the Status column.
Code Meaning
AO Alarm Off
N-550 63
Page 72
Printing
Code Meaning
AS Alarm Silence LB Low Battery
LM Loss of Pulse w/Motion
LP Loss of Pulse
MO Patient MOtion
PH Pulse Rate Upper Limit Alarm PL Pulse Rate Lower Limit Alarm PS Pulse Search SH Saturation Upper Limit Alarm SL Saturation Lower Limit Alarm SD Sensor Disconnect SO Sensor Off
Note: A sensor disconnect and sensor off will also cause three dashes
(- - - ) to be displayed in the patient data section of the printout.
64
Page 73

Sensors and Accessories

Selecting a Sensor

WARNING: Before use, carefully read the sensor directions for use, including all warnings, cautions, and instructions.
WARNING: Do not use a damaged sensor or pulse oximetry cable. Do not use a sensor with exposed optical components.
WARNING: Use only Nellcor sensor s and pulse oximetry cables with the N-550. Other sensors or pulse oximetry cables may cause improper N-550 performance.
Sensors and Accessories
WARNING: Use only one pulse oximetry cable to increase the length of the sensor. Use of more than one pulse oximetry cable may have an adverse effect on performance. Do not attach any cable that is intended for computer use to the sensor port.
WARNING: Pulse oximetry readings and pulse signal can be affected by certain ambient environmental conditions, sensor application errors, and certain patient conditions.
WARNING: Tissue damage can be caused by incorrect application or duration of use of an SpO sensor site as directed in the sensor directions for use.
N-550 65
2 sensor. Inspect the
Page 74
Sensors and Accessories
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
Table 2: Nellcor Oximetry Sensor Models and Patient Weights
WARNING: Do not immerse or wet the sensor.
When selecting a sensor, consider the patient's weight and activity level, the adequacy of perfusion and the available sensor sites, the need for sterility, and the anticipated duration of monitoring. For more information, refer to Table 2 or contact your local Nellcor representative.
For a complete and up-to-date listing of all sensors applicable to the N-550, refer to the Sensor Accuracy Grid posted on the Internet at:
Follow the sensor sterilization procedures in the applicable sensor directions for use (DFU).
XIMAX Sensor
O
O
XIMAX MAX-FAST adhesive forehead
sensor, single-patient-use
O
XIMAX Softcare nonadhesive sensor,
single-patient-use, preteen infant
O
XIMAX Softcare nonadhesive sensor,
single-patient-use, adult
O
XIMAX Softcare nonadhesive sensor,
single-patient-use, preteen infant
O
XIMAX adhesive sensor,
-pateint-use, adult
single
XIMAX adhesive sensor,
O
-pateint-use, adult, longer cable
single 36 inches (91.44 cm)
XIMAX adhesive sensor,
O
-pateint-use, neonatal/adult
single
Patient Size
Model
MAX-FAST
SC-PR <1.5 kg (3.3 lbs)
SC-NEO 1.5 to 5 kg (3.3 to
SC-A >40 kg (8 8 lbs)
MAX-A >30 kg (66 lbs)
MAX-AL >30 kg (66 lbs)
MAX-N <3 kg or >40 kg
>=greater than <=less than
>10 kg (22 lbs)
11 lbs)
(<6.6 lbs or >88 lbs)
66
Page 75
Sensors and Accessories
Table 2: Nellcor Oximetry Sensor Models and Patient Weights
Patient Size
XIMAX Sensor
O
Model
>=greater than <=less than
O
XIMAX adhesive sensor,
-pateint-use, pediatric
single
O
XIMAX adhesive sensor,
-pateint-use, infant
single
O
XIMAX adhesive sensor,
-pateint-use, adult nasal
single
O
XIMAX OxiCliq
-patient-use, adult, reusable cable
single
O
XIMAX OxiCliq nonadhesive sensor,
-patient-use, neonatal/adult,
single
®
nonadhesive sensor,
reusable cable
XIMAX OxiCliq nonadhesive sensor,
O
-patient-use, pediatric, reusable
single cable
O
XIMAX OxiCliq nonadhesive sensor,
-patient-use, infant, reusable
single cable
O
XIMAX Durasensor
®
finger-clip sensor,
reusable, adult
O
XIMAX Oxiband
®
sensor, reusable,
neonatal/adult
MAX-P 10 to 50 kg (22 to
110 lbs)
MAX-I 3 to 20 kg (6.6 to
44.1 lbs)
MAX-R >50 kg (110 lbs)
OxiCliq A >30 kg (66 lbs)
OxiCliq N <3 kg or >40 kg
(<6.6 lbs or >88 lbs)
OxiCliq P 10 to 50 kg (22 to
110 lbs)
OxiCliq I 3 to 20 kg (6.6 to
44.1 lbs)
DS
OXI
-100A
-A/N
>40 kg (88 lbs)
<3 kg or >40 kg (<6.6 lbs or >88 lbs)
O
XIMAX Oxiband sensor, reusable,
pediatric/infant
OXI
-P/I
3 kg to 40 kg (6.6 lbs to 88 lbs)
N-550 67
Page 76
Sensors and Accessories
Table 2: Nellcor Oximetry Sensor Models and Patient Weights
XIMAX Sensor
O
Model
Patient Size >=greater than <=less than
OXIMAX Dura-Y reusable
For use with the Dura Ear clip (Reusable, nonsterile) Pedi
-Check
clip (Reusable, nonsterile)
®
TM
pediatric spot-check

Biocompatibility Testing

Biocompatibility testing has been conducted on Nellcor sensors in compliance with ISO 10993-1, Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing. The sensors have passed the recommended biocompatibility testing and are therefore in compliance with ISO 10993-1.

Optional Accessories

multisite sensor,
-Y sensor:
D
-YS
D
-YSE
-YSPD
D
>1 kg (>2.2 lbs)
>30 kg (66 lbs)
3 kg to 40 kg (6.6 lbs to 88 lbs)
68
Several optional accessories are offered with the N-550. Contact Nellcor's Technical Services Department, 1.800.635.5267, or your local Nellcor representative for information about these accessories.
Foot switch
Visual alarm indicator
Pole mount bracket
Page 77

Foot Switch

Sensors and Accessories
Accessories for the N-550 are listed on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/Apartweb/main/
PartAcceMenu.html
An optional foot switch is available for the N-550. The foot switch may be used to silence N-550 alarms. The foot switch is conn ected to the rear of the N-550.
CAUTION: The N-550 must be turned off when connecting or disconnecting the foot switch.
1 — Foot switch connector

Visual Alarm Indicator

An optional visual alarm indicator is available for the N-550. The visual alarm indicator mounted on the top of the N-550 with double­back tape. The visual alarm indicator is connected to the rear of the N-550.
CAUTION: The N-550 must be turned off when connecting or disconnecting the visual alarm indicator.
N-550 69
Page 78
Sensors and Accessories

Pole Mount Bracket

1 — Visual alarm indicator
70
An optional pole mount bracket is available for the N-550. The pole mount bracket may be used to attach the N-550 to an IV pole. The pole mount bracket is attached to the rear of the N-550.
Page 79

Performance Considerations

WARNING: Pulse oximetry readings and pulse signals can be affected by certain ambient environmental conditions, sensor application errors, and certain patient conditions. See the appropriate sections of the manual for specific safety information.

Performance Verification

The performance of the N-550 can be verified by following the procedures outlined in the Performance Verification section of the N-550 service manual. Qualified service personnel should perform these procedures before using the N-550 for the first time in a clinical setting.

N-550 Performance Considerations

Certain patient conditions can affect the measurements of the N-550 and cause the loss of the pulse signal.
Inaccurate measurements can be caused by:
prolonged patient movement
venous pulsations
intravascular dyes, such as indocyanine green or methylene blue
defibrillation
N-550 71
Page 80
Performance Considerations

Dysfunctional Hemoglobins

Dysfunctional hemoglobins such as carboxyhemoglobin, methemoglobin, and sulphemoglobin are unab le to carry oxygen. SpO
2 readings may appear normal; however, a patient may be
hypoxic because less hemoglobin is available to carry oxygen. Further assessment beyond pulse oximetry is recommended.

Anemia

Anemia causes decreased arterial oxygen content. Although SpO2 readings may appear normal, an anemic patient may be hypoxic. Correcting anemia can improve arterial oxygen content. The pulse oximeter may fail to provide an SpO 5 gm/dl.
2 if hemoglobin levels fall below

Saturation

The N-550 will only measure saturation levels between 1 and 100%.

Pulse rates

The N-550 will only measure pulse rates between 20 and 250 beats per minute. Detected pulse rates outside the range of 20 to 250 beats per minute are displayed as the closest value within the range.

Sensor Performance Considerations

WARNING: Pulse oximetry readings and pulse signal can be affected by certain ambient conditions, sensor application errors, and certain patient conditions.
72
Page 81
Performance Considerations
Inaccurate measurements can be caused by:
incorrect application of the sensor
placement of the sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line
ambient light
prolonged patient movement
Loss-of-pulse signal can occur for the following reasons:
the sensor is applied too tightly
a blood pressure cuff is inflated on the same extremity as the one with the sensor attached
there is arterial occlusion proximal to the sensor
Use only Nellcor sensors and sensor cables.
WARNING: The use of accessories, sensor s, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the N-550 pulse oximeter.
Select an appropriate sensor, apply it as directed, and observe all warnings and cautions presented in the directions for use accompanying the sensor. Clean and remove any substances such as nail polish from the application site. Periodically check to ensure that the sensor remains properly positioned on the patient.
WARNING: Tissue damage can be caused by incorrect application or inappropriate duration of use of an SpO
2 sensor.
Inspect the sensor site as directed in the sensor directions for use.
N-550 73
Page 82
Performance Considerations
High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance of an SpO light, ensure that the sensor is properly applied and cover the sensor site with opaque material.
Caution:Failure to cover the sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
If patient movement presents a problem, try one or more of the following remedies to correct the problem.
2 sensor. To prevent interference from ambient
verify that the sensor is properly and securely applied
move the sensor to a less active site
use an adhesive sensor that tolerates some patient motion
use a new sensor with fresh adhesive backing
If poor perfusion affects performance, consider using the O
XIMAX
MAX-R sensor; it obtains measurements from the nasal septal anterior ethmoid artery, an artery supplied by the internal carotid artery. This sensor may obtain measurements when peripheral perfusion is relatively poor.
74
Page 83

Operator’s Menu

This operator’s menu provides a quick reference to the functions on the N-550. The functions that are Bold are the default settings.
Ta ble 3: Operator’s Menu
Menu
1 - Trend print (tabular
2 - Trend clear Clearing Trend
3 - Language English ­4 -Baud rate: Setting the Data Port
-242400 baud -
- 96 9600 baud -
- 192 19200 baud -
5 - Data Port Protocol: Setting the Data Port
- 1ASCII -
- 2 External equipment
Sub-
Menu
Function
N-550 trend only)
communications.
Reference
Paragraph
Trend Data Printout on page 59
Information on page 40
Baud Rate on page 37
Protocol on page 39
Refer to the external equipment manuals for the interfacing instructions.
N-550 75
Page 84
Blank Page
Page 85

Troubleshooting

WARNING: If you are uncertain about the accuracy of any measurement, check the patient's vital signs by alternate means; then make sure the N-550 is functioning correctly.
WARNING: The cover should be removed only by qualified service personnel. There are no user-serviceable parts inside.
CAUTION: Do not spray, pour, or spill any liquid on the N-550, its accessories, connectors, switches, or openings in the chassis.

Error Codes

When the N-550 detects an error condition, it may display the letters "EEE" followed by an error code.
When an error code is displayed, turn the N-550 off, wait 10 seconds, and turn the N-550 on. If the error code is listed in Table 4, follow the action(s) listed. If the action does not correct the error condition notify service personnel. If the error code is not listed in Table 4, notify service personnel.
When the N-550 detects a defective sensor connected to the N-550 the N-550 displays an error code of “Sen Err.” The sensor should be replaced and the N-550 power should be recycled.
N-550 77
Page 86
Troubleshooting
Table 4: Error Codes

Corrective Action

If you experience a problem while using the N-550 and are unable to correct it, contact Nellcor’s Technical Services Department or your local Nellcor representative. The N-550 service manual, which is for use by qualified service personnel, provides additional troubleshooting information.
Error Code
513 1 — Charge battery
2 — Notify service personnel.
514 1 — Restart the N-550.
2 — Notify service personnel.
525 1 — Restart the N-550.
2 — Notify service personnel.
526 1 — Restart the N-550.
2 — Notify service personnel.
528 1 — Restart the N-550.
2 — Notify service personnel.
Action
78
Following is a list of possible problems and suggestions for correcting them.
Page 87
Troubleshooting
1. There is no response to the Power On/Off button.
If the N-550 is operating on AC power, the fuse may be blown. Notify service personnel to check and, if necessary, replace the fuse.
If the N-550 is on battery power , the battery fuse may require replacement, the battery may be missing, or the battery may be discharged. Charge the battery or notify service personnel to replace the battery or the battery fuse, as required.
2. One or more display segments or indicators do not light during the power-on self-test.
Do not use the N-550; contact qualified service personnel or
your local Nellcor representative.
3. The N-550 does not sound a tone indicating successful completion of the Power-On Self-Test (POST).
The N-550 has failed the power-on self-test. Do not use the
N-550. This tone not only indicates the successful completion of POST, but it confirms that the audible alarm is functional. Contact qualified service personnel or your local Nellcor representative.
4. The Pulse Search indicator is lit for more than 10 seconds while the sensor is connected to the patient.
Check the sensor directions for use to determine if an
appropriate sensor is being used and if it is applied properly. Check sensor and sensor cable connections. Test the sensor on someone else. Try another sensor or sensor cable.
N-550 79
Page 88
Troubleshooting
Perfusion may be too low for the N-550 to track the pulse. Check the patient. Test the N-550 on someone else. Change the sensor site. Try another type of sensor.
Excessive patient motion may be preventing the N-550 from tracking the pulse. Keep the patient still, if possible. Verify that the sensor is securely applied, and replace it if necessary. Change the sensor site. Use a type of sensor that tolerates more patient movement (for example, an adhesive senso r).
The sensor may be too tight, there may be excessive ambient light, or the sensor may be on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. Reposition the sensor, as necessary.
Excessive environmental motion or electromagnetic interference may be preventing the N-550 from tracking the pulse. Remove the source of interference or try to stabilize the environment, or do both.
5. The Pulse Search indicator lights after successful measurements have been ma de.
80
Check the patient.
Perfusion may be too low for the N-550 to track the pulse.
Test the N-550 on someone else. Change the sensor site. Try another type of sensor. Refer to Sensor Performance Considerations on page 72.
Prolonged patient motion may be preventing the N-550 from
tracking the pulse. Verify that the sensor is securely applied and replace it if necessary. Change the sensor site. Use a type of sensor that tolerates more patient movement (for example, an adhesive sensor).
The sensor may be too tight, there may be excessive ambient
light, or the sensor may be on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. Reposition the sensor, as necessary.
Excessive environmental motion or electromagnetic
interference may be preventing the N-550 from tracking the
Page 89
Troubleshooting
pulse. Remove the source of interference or try to stabilize the environment, or do both.
6. The letters EEE, followed by a number, appear on the display.
This is an error code. To confirm, press the Power On/Off button to turn the N-550 off, then press the button again to turn it back on. If the display shows the error code once again, record the number and provide that information to qualified service personnel, or your local Nellcor representative.
Error Code “EEE 513" is displayed when the battery is discharged to a critically low level. Turn th e N-550 off and let it charge for about 10 minutes and then turn the N-550 back on. If the error code is still present, turn the N-550 off and let it continue to charge. If the N-550 has been charged for 30 minutes and the error code is still present, notify service personnel.
Refer to the N-550 service manual for a complete listing of error codes.

EMI (Electro-magnetic Interference)

CAUTION: This device has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2, EN60601- 1-2, and the Medical Device Directive 93/42/EEC. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care environments (for example, electrosurgical units, cellular phones, mobile two-way radios, electrical appliances, and high-definition television), it is possible that high levels of such interference due to close proximity or strength of a source may result in disruption of performance of this device.
N-550 81
Page 90
Troubleshooting
The N-550 is designed for use in environments in which the pulse can be obscured by electromagnetic interference. During such interference, measurements may seem inappropriate or the N-550 may not seem to operate correctly.
Disruption may be evidenced by erratic readings, cessation of operation, or other incorrect functioning. If this occurs, the site of use should be surveyed to determine the source of this disruption, and the following actions taken to eliminate the source:
Turn equipment in the vicinity off and on to isolate the offending equipment.
Reorient or relocate the interfering equipment.
Increase the separation between the interfering equipment and this equipment.
The N-550 generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference with other devices in the vicinity.
If assistance is required, contact Nellcor’s Technical Services Department, 1.800.635.5267, or your local Nellcor representative.

Obtaining Technical Assistance

For technical information and assistance, or to order parts or a service manual, contact Nellcor’s Technical Services Department,
1.800.635.5267, or your local Nellcor representative. The service manual includes block diagrams and a parts list required by qualified personnel when servicing the N-550.
When calling Nellcor’s Technical Services Department,
1.800.635.5267, or your local Nellcor representative, you may be asked to tell the representative the software version number of your N-550.
The device version appears in the N-550 display each time the N-550 successfully completes the power-on self-test. Write the number down and have it available whenever requesting technical assistance.
82
Page 91
Troubleshooting
The most recent revision of this manual and the service manual are available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/
ProductManuals.html
N-550 83
Page 92
Blank Page
Page 93

Maintenance

CAUTION: Follow local governing ordinances and recycling instructions regarding the disposal or recycling of device components and accessories.

Returning the N-550

Contact Nellcor’s T echn ical Services Department, 1.800.635.5267, or your local Nellcor representative for shipping instructions including a Returned Goods Authorization (RGA) number. Unless otherwise instructed by Nellcor’s Technical Services Department, it is not necessary to return the sensor or other accessory items with the N-550. Pack the N-550 in its original shipping carton. If the original carton is not available, use a suitable carton with appropriate packing material to protect it during shipping.
Return the N-550 by any shipping method that provides proof of delivery.

Service

WARNING: The cover should be removed only by qualified service personnel. There are no user-serviceable parts inside.
The N-550 requires no routine service or calibration other than changing the battery at least every 24 months. Refer to the N-550 service manual for the battery changing procedure.
If service is necessary, contact qualified service personnel or your local Nellcor representative.
N-550 85
Page 94
Maintenance

Periodic Safety Checks

It is recommended that the following checks be performed every 24 months.
Inspect the equipment for mechanical and functional damage.
Inspect the safety-relevant labels for legibility.

Cleaning

CAUTION: Do not spray, pour, or spill any liquid on the N-550, its accessories, connectors, switches, or openings in the chassis.
For surface-cleaning and disinfecting the N-550, follow your institution's procedures or:
•The N-550 may be surface-cleaned by using a soft cloth dampened with either a commercial, nonabrasive cleaner or a solution of 70% alcohol in water, and lightly wiping the surfaces of the N-550.
86
•The N-550 may be disinfected using a soft cloth saturated with a solution of 10% chlorine bleach in tap water.
Before attempting to clean an SpO enclosed with the sensor. Each sensor model has cleaning instructions specific to that sensor.
2 sensor, read the directions for use
Page 95

Technical Information

Description of Alarms

The N-550 has three levels of audible alarms.
1. High-priority alarm: A high-pitched, fast-pulsating tone indicating loss of pulse with no patient motion.
2. Medium-priority alarm: A medium-pitched, normal-pulsating tone indicating an SpO
During a medium-priority alarm, the green display turns red and flashes with the patient parameter that violated the limit (%SpO Pulse Rate). If the alarm is a SatSeconds alarm, the SatSeconds indicator (clock) will flash.
2 or pulse rate limit violation.
2 or
3. Low-priority alarm: A low-pitched, slow-pulsating tone indicating a sensor disconnect, low battery, or N-550 failure.
During a low-battery alarm, the Low Battery indicator illuminates and the alarm tone sounds immediately, even if the alarms are silenced or set to OFF.
N-550 87
Page 96
Technical Information

Factory Defaults

The N-550 is shipped with factory default settings. Authorized technical personnel using the procedures described in the N-550 service manual can change default settings.
Table 5: Factory Default Settings
Factory
Parameter Range
Default Setting
2 Upper Alarm
%SpO Limit
2 Lower Alarm
%SpO Limit
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Alarm Silence
Lower Alarm Limit plus 1 to 100%
20% to Upper Alarm Limit minus 1
Lower Alarm Limit plus 1 to 250 bpm
30 bpm to Upper Alarm Limit minus 1
On or Off On
Reminder Alarm Silence
Duration Alarm Silence
Off, 30, 60, 90, 120 seconds
Audible reminder
Restrictions
No audible reminder
Do not allow alarms off
Alarm Sound Selector 1, 2, 3 1
100%
85%
170 bpm
40 bpm
60 seconds
Audible reminder
88
Alarm Volume 1 to 10 4 Data Port Baud Rate 2400, 9600, 19200 19200 Data Port Protocol 1, 2 1 (ASCII)
Page 97
Technical Information
Table 5: Factory Default Settings
Factory
Parameter Range
Default Setting
Display 0, 1 0 (On) In-Sensor Trend Mode 0, 1, 2 0 (Event SpO Language English English Pulse Beep Volume 0 to 10 4 RS-232 Nurse Call
Priority SatSeconds Off, 10, 25, 50, 100 Off Silence Alarms 0, 1 1 (Off)

SatSeconds Display

When the N-550 SatSeconds technology detects an SpO2 value outside the alarm limit, the SatSeconds indicator LEDs begin to light (fill) clockwise. When the SpO SatSeconds indicator LEDs will extinguish counterclockwise.
When all of the SatSecond indicator LEDs are lit, indicating that the
SatSeconds setting has been reached, an audible alarm sounds and the SatSeconds indicator LEDs flash. As with traditional alarm
management, the audible alarm may be silenced by pressing the Alarm Silence button.
Normally high,
Normally low
normally low
2 value is within the set limits, the
2)
Note: When a SatSeconds setting other than OFF is selected, the
green LED at the top of the SatSeconds indicator will light. The green LED indicates that SatSeconds have been engaged.
N-550 89
Page 98
Technical Information

Describing SatSeconds

With traditional alarm management, upper and lower alarm limits are set for monitoring oxygen saturation. During monitoring, as soon as an alarm limit is violated by as little as one percentage point, an audible alarm immediately sounds. When the %SpO near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarms can be distracting.
The N-550 utilizes Nellcor SatSeconds alarm management. With the SatSeconds technique, upper and lower alarm limits are set in the same way as traditional alarm management. However, the clinician also sets a SatSeconds limit that allows monitoring of %SpO the selected lower alarm limit and above the selected upper alarm limit for a period of time before an audible alarm sounds.
2 level fluctuates
2 below
The SatSeconds limit controls the time that the %SpO
2 level may fall
below the alarm limit before an audible alarm sounds.
The method of calculation is as follows:
The number of percentage points that the %SpO alarm limit is multiplied by the number of seconds that the %SpO
2 falls outside the
2
level remains outside that limit. This can be stated as an equation:
Points x Seconds = SatSeconds
Where:
Points = SpO
Seconds = number of seconds that SpO
2 percentage points outside of the limit
2 remains at that
point outside of the limit
The alarm response time, assuming a SatSeconds limit set at 50 and a lower alarm limit set at 90, is described and illustrated below.
90
Page 99
Technical Information
In this example, the %SpO2 level drops to 88 (2 points) and remains there for a period of 2 seconds (2 points x 2 seconds = 4). The %SpO then drops to 86 for 3 seconds and then to 84 for 6 seconds. The resulting SatSeconds are:
2
%SpO
2 Seconds SatSeconds
2 x 2 = 4 4 x 3 = 12 6 x 6 = 36
Total SatSeconds =
52
After approximately 10.9 seconds, the SatSeconds alarm will sound, because 50 SatSeconds will have been exceeded. See arrow (
) in
Figure 6.
Figure 6: Alarm Response with SatSeconds
N-550 91
Page 100
Technical Information
Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Often, %SpO
2 levels may fluctuate above
and below the alarm limit, reentering the non-alarm range several times.
During such fluctuation, the N-550 integrates the number of %SpO points, both positive and negative, until either the SatSeconds limit (SatSeconds setting) is reached, or the %SpO normal range and remains there.

SatSeconds “Safety Net”

The SatSeconds “Safety Net” is for patients whose saturation levels frequently go below the limit but do not stay below the limit long enough for the SatSeconds setting to be reached. When 3 or more limit violations occur in 60 seconds, an alarm will sound even if the SatSeconds setting has not been reached.
2
2 level returns within a
92
Loading...