Nellcor N-40 Service manual

SERVICE MANUAL
NPB-40 Handheld Pulse Oximeter
Caution: Federal law (U.S.) restricts this device to sale by or on the order of a physician. To contact Nellcor Puritan Bennett’s representative: In the United States, call 1-800-635-5267 or
925-463-4000; outside of the United States, call your local Nellcor Puritan Bennett representative.
© 2001 Nellcor Puritan Bennett, Inc. All rights reserved. 036048D-0601
0123
Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588 USA Telephone Toll Free 1.800.NELLCOR
Tyco Healthcare UK LTD Fareham Road Gosport PO13 0AS U.K. Tel: +44.1329.224000
Puritan Bennett Technical Services Department, or your local Nellcor Puritan
Bennett representative. Notice: Purchase of this instrument confers no express or implied license under any
Nellcor Puritan Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Oxisensor II is a trademark of Nellcor Puritan Bennett, Inc. Covered by one or more of the following U.S. Patents and foreign equivalents:
4,621,643; 4,653,498; 4,700,708; 4,770,179; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; and Re. 35,122.
TABLE OF CONTENTS
List of Figures List of Tables
Section 1: Introduction................................................................................. 1-1
1.1 Manual Overview............................................................................ 1-1
1.2 Equipment Description ................................ ................................ ... 1-1
Section 2: Routine Maintenance ................................................................ 2-1
2.1 Maintenance Overview ................................................................... 2-1
2.2 Battery Maintenance ...................................................................... 2-1
2.3 Cleaning ................................ ................................ ............................ 2-2
Section 3: Performance Verification .......................................................... 3-1
3.1 Performance Verification ................................................................... 3-1
Section 4: Troubleshooting ................................................................ ......... 4-1
4.1 Introduction..................................................................................... 4-1
4.2 Who Should Perform Repairs......................................................... 4-1
4.3 Repair Level Supported.................................................................. 4-1
4.4 How to Use This Section ................................ ................................ 4-1
4.5 Obtaining Replacement Parts ........................................................ 4-1
4.6 Troubleshooting Guide ................................................................... 4-2
Section 5: Disassembly Guide .................................................................... 5-1
5.1 Introduction..................................................................................... 5-1
5.2 System Disassembly and Reassembly .......................................... 5-1
Section 6: Spare Parts.................................................................................. 6-1
6.1 Introduction..................................................................................... 6-1
Section 7: Packing for Shipment................................................................. 7-1
7.1 General Instructions........................................................................ 7-1
7.2 Packing NPB-40 in Original Carton ................................ ................ 7-1
7.3 Packing in a Different Carton ......................................................... 7-1
Section 8: Specifications ................................................................ ............. 8-1
8.1 Performance ................................................................................... 8-1
8.2 Electrical ......................................................................................... 8-2
8.3 Environmental................................................................................. 8-2
8.4 Physical ......................................................................................... 8-3
Technical Supplement.................................................................................. S-1
S1 Introduction..................................................................................... S-1
S2 Pulse Oximetry Principles of Operation .......................................... S-1
S3 Circuit Analysis ............................................................................... S-2
S4 Schematic Diagrams ................................ ................................ ...... S-17
2.2.1 Battery Replacement ......................................................... 2-1
2.2.2 Battery Disposal ................................................................ 2-2
S2.1 Overview ................................................................ ........... S-1
S2.2 Functional Versus Fractional Saturation............................ S-1
S2.3 Measured versus Calculated Saturation............................ S-2
S3.1 Overall Unit Block Diagram Analysis ................................ . S-3
S3.2 CPU PCB Theory of Operation.......................................... S-5
S3.3 LCD PCB Theory of Operation .......................................... S-11
iii
Table of Contents
LIST OF FIGURES
1-1 NPB-40 Handheld Pulse Oximeter ................................................ 1-1
1-2 NPB-40 Front Panel....................................................................... 1-2
2-1 NPB-40 Battery Installation ........................................................... 2-2
3-1 NPB-40 Self-Test Front Panel Display ................................ .......... 3-1
3-2 Printer Interface Setup................................................................... 3-3
3-3 Typical Average Data Print Out ..................................................... 3-4
3-4 Typical Data Print Out ................................................................... 3-5
4-1 Typical Error Code Display ............................................................ 4-2
5-1 Opening the NPB-40 Case ............................................................ 5-2
5-2 Top Case Removal ........................................................................ 5-2
5-3 Unlocking CPU PCB Connector J3 ............................................... 5-3
5-4 Separating LCD PCB from CPU PCB............................................ 5-3
6-1 NPB-40 Expanded View ................................................................ 6-2
7-1 Repacking the NPB-40 .................................................................. 7-2
S2-1 Oxyhemoglobin Dissociation Curve .............................................. S-2
S3-1 NPB-40 Block Diagram ................................................................. S-4
S3-2 CPU PCB Block Diagram .............................................................. S-6
S3-3 LCD PCB Block Diagram............................................................... S-12
S4-1 CPU PCB Parts Locator Diagram.................................................. FO-1
S4-2 LCD PCB Parts Locator Diagram .................................................. FO-2
S4-3 CPU PCB Schematic Diagram ...................................................... FO-3
S4-4 LCD PCB Schematic Diagram....................................................... FO-4
S4-5 Front Panel Keypad Schematic Diagram ...................................... FO-5
LIST OF TABLES
4-1 Troubleshooting Guide .................................................................. 4-2
4-2 NPB-40 Error Codes ..................................................................... 4-5
iv
SECTION 1: INTRODUCTION
1.1 Manual Overview
1.2 Equipment Description
1.1 MANUAL OVERVIEW
This manual contains information for servicing the NPB-40 handheld pulse oximeter. Only qualified service personnel should service this product. Before servicing the NPB-40, read the operator’s manual carefully for a thorough understanding of how to operate the NPB-40.
1.2 EQUIPMENT DESCRIPTION
The Nellcor Puritan Bennett NPB-40 handheld pulse oximeter is used for noninvasive spot-check measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by the SpO2 sensor). The NPB-40 is for attended monitoring only, and must be used in the continuous presence of a qualified healthcare provider. The NPB-40 can be used on adult, pediatric, and neonatal patients. It can be used in mobile environments when protected from excessive moisture such as direct rainfall. The NPB-40 is powered by four “AA” cell alkaline batteries. An external Hewlett-Packard HP82240B printer can be used with the NPB-40 to printout readings stored in the NPB-40 memory. This printer is available from Nellcor Puritan Bennett. The NPB-40 handheld pulse oximeter is illustrated in Figure 1-1.
Figure 1-1: NPB-40 Handheld Pulse Oximeter
The NPB-40 is operated using a four-key keypad and an LCD display on the front panel as shown in Figure 1-2. Refer to the NPB-40 operator’s manual for complete operating instructions.
1-1
Section 1: Introduction
13
12 11
10
1
Pulse Amplitude indicator
2
1
2
3 4
5
9
8
6
7
Pulse Rate display
3
Pulse Search display
4
Low Battery indicator
5
Sensor disconnect indicator
6
Power On/Off key
7
Display Light/Beep On/Off key
8
Store Data/Print key
9
Shift key
10
Store Data icon
11
Shift icon
12
Print icon
13
SpO2% display
Figure 1-2: NPB-40 Front Panel
1-2
SECTION 2: ROUTINE MAINTENANCE
2.1 Maintenance Overview
2.2 Battery Maintenance
2.3 Cleaning
2.1 MAINTENANCE OVERVIEW
The NPB-40 requires no routine service or calibration. The performance verification tests in Section 3, Performance Verification, may be used following repairs or during routine maintenance (if required by your local institution).
2.2 BATTERY MAINTENANCE
Warning: The NPB-40 uses four “AA” cell alkaline batteries for operation. If batteries are not replaced when recommended or are not disposed of properly, serious personal injury or damage to the NPB-40 could result.
Caution: When inserting the positive end of each battery, exercise caution to not damage the small coiled spring contact.
Nellcor Puritan Bennett recommends that you comply with the following guidelines for battery replacement and battery disposal as a minimum.
2.2.1 Battery Replacement
Caution: The NPB-40 could be damaged by batteries that are left unused in the unit and begin to leak. Never store the NPB-40 with the batteries installed for a prolonged period of time.
NPB-40 batteries should be replaced whenever a low battery indication is observed on the unit. Remove the batteries if you will be storing the NPB-40 for longer than one month. Refer Figure 2-1 for replacing the batteries.
Note: For easier battery installation, insert the negative end of the battery
first when installing each battery.
2-1
Section 2: Routine Maintenance
2.2.2 Battery Disposal
2.3 CLEANING
Figure 2-1: NPB-40 Battery Installation
When NPB-40 batteries have been replaced, dispose of the old batteries. Always follow local ordinances when disposing of the NPB-40 batteries.
Warning: Never dispose of NPB-40 batteries by burning. The batteries could explode in fire and cause serious personal injury.
Caution: Do not immerse the NPB-40 or accessories in liquid or use caustic or abrasive cleaners.
To clean the NPB-40, dampen a soft cloth with a commercial nonabrasive cleaner and wipe the unit with the cloth.
2-2
SECTION 3: PERFORMANCE VERIFICATION
3.1 Performance Verification
3.1 PERFORMANCE VERIFICATION
The performance of the NPB-40 can be verified using the following procedure. Before performing this procedure, the NPB-40 must have fresh batteries installed. If any of the required observations cannot be obtained, do not return the NPB-40 to service before referring to Section 4, Troubleshooting.
You will need an SRC-2 pulse oximeter tester to perform this procedure. If the performance of the printing function will be verified, you will need an HP82240B printer.
The NPB-40 is to be turned off and any sensors disconnected before performing this procedure.
1. Press the On/Off key on the front panel keypad.
2. The NPB-40 will perform a self-test. Verify that all display segments are lit during the self-test as shown in Figure 3-1. Also verify that a low pitched audio tone is heard at the end of the self test. Also verify that you then hear a high, a low, and a high pitch error tone indicating that a sensor is not connected. Verify that dashes are shown on the display and the SENSOR DISCONNECT indicator is flashing.
Figure 3-1: NPB-40 Self-Test Front Panel Display
3. Wait for 20 to 40 seconds. Verify that two triple beeps are heard. Verify that the NPB-40 then shuts itself off.
4. Connect an SRC-2 pulse oximeter tester to the NPB-40 sensor connector.
Note: If the SRC-2 is connected directly to the NPB-40 sensor
connector, its controls will be facing the opposite direction of the NPB-40 controls. Connecting the SRC-2 through a sensor extension cable, such as the Nellcor Puritan Bennett model EC­4 or EC-8, will make it easier to perform this procedure.
3-1
Section 3: Performance Verification
Set the SRC-2 controls as follows:
RATE 112 LIGHT HIGH 1 MODULATION HIGH LOCAL/REMOTE/RCAL LOCAL
5. Press the On/Off key on the NPB-40 front panel keypad.
6. When the self-test is complete, verify that the PULSE SEARCH indicator illuminates momentarily and verify that the NPB-40 is displaying a pulse rate of 112 (±3 bpm), that the pulse amplitude indicator is displaying properly, that the SpO2% indicator is displaying 81 (±2 %), that a pulse beep can be heard, and that the PULSE SEARCH indicator is off.
7. Set the following SRC-2 controls as indicated:
RATE 38 LIGHT HIGH 2 MODULATION LOW
Wait 30 seconds and verify that the NPB-40 is displaying a pulse rate of 38 (±3 bpm), that the pulse amplitude indicator is displaying properly, that the SpO2% indicator is displaying 81 (±2 %), that a pulse beep can be heard, and that the PULSE SEARCH indicator is off.
8. Set the following SRC-2 controls as indicated:
RATE 201 LIGHT LOW
Wait 30 seconds and verify that the NPB-40 is displaying a pulse rate of 201 (±3 bpm), that the pulse amplitude indicator is displaying properly, that the SpO2% indicator is displaying 81 (±2 %), that a pulse beep can be heard, and that the PULSE SEARCH indicator is off.
9. Set the following SRC-2 control as indicated: MODULATION HIGH Wait 30 seconds and verify that the NPB-40 is displaying a pulse rate of
201 (±3 bpm), that the pulse amplitude indicator is displaying properly, that the SpO2% indicator is displaying 81 (±2 %), that a pulse beep can be heard, and that the PULSE SEARCH indicator is off.
10. Press the Shift key followed by the Display Light/Beep On/Off key. Verify that the SHIFT indicator illuminates when the Shift key is pressed then verify that the pulse beep stops and the SHIFT indicator extinguishes when the Display Light/Beep On/Off key is pressed.
11. Press the Shift key followed by the Display Light/Beep On/Off key and verify that the pulse beep can be heard again at a low level.
3-2
Section 3: Performance Verification
12. Press the Shift key followed by the Display Light/Beep On/Off key and verify that the pulse beep can again be heard again at a higher level.
13. Press and release the Display Light/Beep On/Off key on the front panel keypad and verify that the LCD backlight comes on, remains on and the LCD is adequately illuminated.
14. Press the Display Light/Beep On/Off key on the front panel keypad and verify that the LCD backlight is off.
15. Set the following SRC-2 control as indicated: LOCAL/REMOTE/RCAL REMOTE
16. On the NPB-40, verify that the PULSE SEARCH indicator illuminates after a few seconds and that the number “0” is displayed in the SpO2% and pulse rate displays.
17. Set the following SRC-2 control as indicated: LOCAL/REMOTE/RCAL LOCAL
18. On the NPB-40, verify that the PULSE SEARCH indicator illuminates momentarily and verify that the NPB-40 is displaying a pulse rate of 201 (±3 bpm), that the pulse amplitude indicator is displaying properly, that the SpO2% indicator is displaying 79 to 83, that a pulse beep can be heard, and that the PULSE SEARCH indicator is off.
Note: If an HP82240B printer is not available, skip steps 19 through
27 and proceed to step 28.
19. Refer to the printer operator’s manual as needed. Verify that paper is properly installed in the printer and turn the printer on.
20. Align the NPB-40 with the printer as shown in Figure 3-2.
0°–5°
0°–5°
1"–2" (2.5–5.1 cm.)
Figure 3-2: Printer Interface Setup
3-3
Section 3: Performance Verification
21. On the NPB-40, press the Shift key followed by the Store Data/Print key on the front panel keypad.
Note: In the following steps, printed SpO2% values will be 79 to 83
22. Verify that the PRINT icon is illuminated on the NPB-40 display, that the printer begins printing and that the printer prints out a summary report similar to that shown in Figure 3-3.
and pulse rate values will be 201, ±3 bpm. Figures 3-3 and 3-4 are examples of typical print outs. The NPB-40 will not display pulse or SpO2% values while printing.
Figure 3-3: Typical Average Data Print Out
23. With the SRC-2 connected to the NPB-40 and still set up as in step 18, press only the Store Data/Print key on the front panel keypad and verify that the LCD display indicates “1-Id” and the STORE DATA icon illuminates.
24. Press only the Store Data/Print key on the front panel keypad and verify that the LCD display indicates “2-Id” and the STORE DATA icon illuminates.
25. Press only the Store Data/Print key on the front panel keypad and verify that the LCD display indicates “3-Id” and the STORE DATA icon illuminates.
26. Align the NPB-40 with the printer as shown in Figure 3-2 and press the Shift key on the NPB-40 front panel keypad, followed by the Store Data/Print key.
27. Verify that the PRINT icon is illuminated on the NPB-40 display, that the printer begins printing and that the printer prints out data similar to that shown in Figure 3-4. As data stored in steps 24, 25, and 26 is printed out, the NPB-40 display will indicate “1-Id”, “2-Id”, and “3-Id”, respectively.
3-4
Section 3: Performance Verification
Figure 3-4: Typical Data Print Out
28. Press the On/Off key on the front panel keypad.
29. Verify that the NPB-40 shuts off.
3-5
(This page intentionally left blank)
SECTION 4: TROUBLESHOOTING
4.1 Introduction
4.2 Who Should Perform Repairs
4.3 Repair Level Supported
4.4 How to Use This Section
4.5 Obtaining Replacement Parts
4.6 Troubleshooting Guide
4.1 INTRODUCTION
This section provides information for troubleshooting the NPB-40 and helping you to isolate a failure to the front panel keypad, the CPU PCB, or the LCD PCB. A troubleshooting guide is provided in Paragraph 4.6, “Troubleshooting Guide” that lists possible difficulties, along with probable causes, and recommended actions to correct each difficulty. The Technical Supplement at the end of this manual provides information on how the components of the NPB-40 function.
4.2 WHO SHOULD PERFORM REPAIRS
Only qualified service personnel should remove and replace components of the NPB-40. Repairs to the NPB-40 are limited to the repair level identified in Paragraph 4.3, “Repair Level Supported.” If your facility does not have qualified service personnel, contact the Nellcor Puritan Bennett Technical Services Department or your local Nellcor Puritan Bennett representative.
4.3 REPAIR LEVEL SUPPORTED
Besides the batteries, the NPB-40 has five replaceable components, the case top with the front panel keypad, the case bottom, the CPU PCB, the LCD PCB, and the battery compartment door.
4.4 HOW TO USE THIS SECTION
Failures of the case bottom, the battery compartment door, and the case top, not including the front panel keypad, are determined by visually inspecting these components for cracks or deformations and for mechanical failures such as the screw holes stripping out in the case top. The case top and the front panel keypad are replaced together. Use the troubleshooting guide provided in Paragraph 4.6, “Troubleshooting Guide” to isolate failure to the front panel keypad, the CPU PCB, or the LCD PCB. Once a failure has been isolated, refer to Section 5, Disassembly Guide for instructions for removing and replacing a component of the NPB-40.
4.5 OBTAINING REPLACEMENT PARTS
Nellcor Puritan Bennett Technical Services Department provides technical assistance information and replacement parts. To obtain replacement parts, contact the Nellcor Puritan Bennett Technical Services Department or your local Nellcor Puritan Bennett representative. Refer to parts by the part names and part numbers listed in Section 6, Spare Parts.
4-1
Section 4: Troubleshooting
4.6 TROUBLESHOOTING GUIDE
If you encounter a problem that cannot be resolved through a visual inspection, refer to Table 4-1 which provides a list of symptoms, probable causes, and recommended actions to take to correct the problem. It is recommended that corrective actions be performed in the order presented. For a symptom that is not listed in Table 4-1, contact the Nellcor Puritan Bennett Technical Services Department or your local Nellcor Puritan Bennett representative.
If an error code is shown on the front panel LCD display, as shown in Figure 4-1, turn the NPB-40 off and back on again. If the error code still persists, refer to Table 4-2 for the indicated failure and corrective action.
Figure 4-1: Typical Error Code Display
Once you have performed the recommended action, reassemble the NPB-40, refer to Section 3, Performance Verification, and conduct a performance verification before returning the NPB-40 to service. If the symptom persists, continue troubleshooting.
Table 4-1: Troubleshooting Guide
Symptom Probable Cause Corrective Action
The unit does not turn on when the On/Off key is
The batteries are missing.
Open the battery compartment and if batteries are missing, install new batteries.
pressed.
The batteries are installed incorrectly or are backwards.
Open the battery compartment and if batteries are not installed correctly, remove and reinstall the batteries.
The batteries are at
Install new batteries.
or near a voltage too (Continued on next page)
low for the NPB-40 to
operate.
4-2
Section 4: Troubleshooting
Table 4-1: Troubleshooting Guide (Continued)
Symptom Probable Cause Corrective Action
The unit does not turn on when the
The sensor is
defective.
Replace the sensor.
On/Off key is pressed. (continued)
The front panel
keypad is defective.
Open the NPB-40, disconnect the case top from J3 on the CPU PCB and connect an ohmmeter between the flex circuit conductor for J3, pin 5 and the conductor for J3, pin 3. Observe a short when the On/Off key is pressed and an open when not pressed. If incorrect, replace case top. If all keys function correctly, replace CPU PCB.
Caution: Unlock J3 before attempting to remove flex circuit conductor.
Flex circuit between
front panel and CPU
PCB is disconnected.
A CPU PCB
component has
failed.
Inspect the flex circuit between the front panel and the CPU PCB and reconnect if loose.
Inspect the CPU PCB components and circuit board for cracking, burning, or damage, and replace the CPU PCB if any are found. If any failed components are observed, replace CPU PCB.
Replace the CPU PCB with a known good PCB.
One or more keys on the front panel keypad does not work.
The front panel
keypad is defective.
Open the NPB-40, disconnect the case top from J3 on the CPU PCB and connect an ohmmeter lead to the flex circuit conductor for J3, pin 3. Refer to the front panel schematic diagram and individually connect the other ohmmeter lead to each conductor for the other keys. Observe a short when the key is pressed and an open when not pressed. If incorrect, replace case top. If correct, replace the CPU PCB.
Caution: Unlock J3 before attempting to remove flex circuit conductor.
Flex circuit between
front panel and CPU
PCB is disconnected.
A CPU PCB
component has
failed. (Continued on next page)
Inspect the flex circuit between the front panel and the CPU PCB and reconnect if loose.
Inspect the CPU PCB components and circuit board for cracking, burning, or damage and replace the CPU PCB if any are found.
4-3
Loading...
+ 38 hidden pages