z Indicates a potential hazard or unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION
z Indicates a potential hazard or unsafe practice that, if not avoided, could result in
death or serious injury.
NOTE
z Provides maintenance tips or other useful information.
1-1
1.1.1 Warnings
WARNING
z All installation operations, expansions, changes, modifications and repairs of this
product should be conducted by authorized personnel only.
z Always disconnect the equipment with the charger stand and remove the batteries
before disassembling the equipment.
z Dispose of the packaging material according to local waste control regulations and
your hospital’s waste disposal protocols. Keep the packaging material out of
children’s reach.
1.1.2 Cautions
CAUTION
z Make sure that no electromagnetic radiation interferes with the performance of the
equipment when preparing to carry out performance tests. Mobile phone, X-ray
equipment and MRI devices are possible sources of interference as they may emit
higher levels of electromagnetic radiation.
z Before connecting the charger stand to the AC mai n s, ch eck that the voltage and
frequency ratings of the AC mains meet the specifications indicated on the
equipment’s label or in this manual.
z Protect the equipment from damage caused by drop, impact, strong vibration or
other mechanical force during servicing.
1.1.3 Notes
NOTE
z Refer to Operation Manual for detailed operation and other information.
1-2
1.2 Equipment Symbols
Direct Current (DC)
Attention: Consult this manual before maintenance.
Auxiliary output connector
Audio pause
Battery door locked/unlocked
Power supply connector
Left/Right button
Power button
Up button
Down button
Date of manufacture
Serial number
Safety Class II equipment
Type BF applied part, defibrillation protected
The following definition of the WEEE label applies to EU member states
only.
This symbol indicates that this product should not be treated as household
waste. By ensuring that this product is disposed of correctly, you will help
prevent bringing potential negative consequences to the environment and
human health. For more detailed information with regard to returning and
recycling this product, please consult the distributor from whom you
purchased it.
* For system products, this label may be attached to the main unit only.
1-3
FOR YOUR NOTES
1-4
2 Theory of Operation
2.1 Introduction
This pulse oximeter is designed to monitor or measure the oxygen saturation and pulse rate of
single adult, pediatric and neonatal patient.
The pulse oximeter also:
Presents audible and visual alarms in case of patient or equipment problems.
Enables the real-time displaying, reviewing, storing and exporting of SpO
values.
Supports Pitch Tone, which means the pitch of pulse tone rises as the oxygen saturation
level increases and falls as the oxygen saturation level decreases.
Operates on either alkaline batteries or a lithium-ion battery.
Offers wired or wireless communication with a personal computer.
and PR
2
2.2 System Connections
2.2.1 Mounting the Pulse Oximeter
The pulse oximeter can be mounted on a wall bracket or on a trolley support. The wall
bracket or trolley support can be ordered optionally. Each type of mounting bracket is
delivered with a complete set of mounting hardware and instructions for use. To install the
pulse oximeter, refer to the instrutions for installation.
CAUTION
z Use mounting brackets we supply or approve. If other compatible mounting
bracket is used, be sure it can be safely applied to the pulse oximeter.
z The mounting bracket should be installed by our qualified service personnel, or
mechanical engineers who have adequate knowledge on it.
z If other mounting solution is used, the installation personnel and the customer
should verify if it can be safely applied to the pulse oximeter, and the customer
assume the responsibility for any risk resulting from that.
2-1
2.2.2 Connectors for Peripheral Devices
The connectors for peripheral devices are located at the top, right side and bottom of the
pulse oximeter as shown in the figure.
Multifunctional connector
It is a DB 9 connector which is used to connector a SpO
sensor (including reusable
2
sensor, disposable sensor and veterinary sensor) to measure the oxygen saturation or
connect a personal computer through a PC communication cable to export trend data.
Infrared port
It is a port through which a personal computer is communicated using a infrared adapter
to export data in real time without affecting patient monitoring.
Power supply connector
It is a female power socket which is used to connect the charger stand’s male power
plug . To avoid being mixed with other power supply adapters, it has nonstandard size.
2-2
2.3 Main Unit
The pulse oximeter consists of main board, SpO
charger stand. The main board is composed of a main control unit and a power module.
The following diagram shows the structure of the main unit.
board, display, speaker, batteries and
2
2.3.1 Main Control Unit
The main control unit provides the system with resources and support. It controls the LCD
screen, alarm LED, speaker, button operation and data storage. It also implements
communication with the SpO
The functional block diagram of the main control unit is shown below:
board, power module and external interfaces.
2
2-3
2.3.1.1 Power Supply
The power module provides the main control unit with 3.1 V, 2.5 V and 0.9 V power supple.
Label Voltage Applicable parts
DVDD 3.1 V Alarm LEDs, Flash memory, infrared light-emitting IC,
PC communication cable, LCD boost up circuit,
prestage audio optional amplifier.
VDDE 2.5 V DSP peripheral voltage, SDRAM, LCD logic, infrared
logic, buttons and resetting.
VDDINT 0.9 V DSP kernel voltage
2-4
2.3.1.2 Core Control Unit
The Core control unit consists of CPU, SDRAM and Flash memory.
The CPU is ADI’s DSP BF531. Its kernel running frequency is up to 400 MHz and external
frequency up to 133MHz. The kernel voltage is 0.8 V and the current consumed at 50 MHz is
as low as 26mA. The oscillating frequency of the CPU clock is 11.0592 MHz and its
expected frequency is 55MHz, which can be implemented through internal PLL frequency
multiplication. The SDRAM provides space for program running and the Flash memory
provides space for storing program, data, lingual library and configuration information. The
BF531 starts directly from the Flash memory.
2.3.1.3 Man-machine Interfaces
The functions of the buttons are listed below:
Button In measurement mode In menu mode
Audio Pause
button
Up button Increases the beat
Down button Decreases the beat
Left button Enters the main menu Enters a submenu or confirm
Right button Locks/Unlocks buttons. Returns to the previous menu
LCD screen
It is a 2.4” standard QVGA (320×240) TFT LCD with a 36-pin connector. The LCD is
connected with the BF531 through a bus. 4 LEDs in series are used to backlight the
LCD. The maximum current is 15mA and driving voltage 13.2V. The main control
board provides DC power supply for the LCD and the backlight board.
Pauses audible alarms. Pauses audible alarms.
Moves the cursor upwards or
volume.
volume.
increases the value of selected
menu item by one.
Moves the cursor downwards
or decreases the value of selected
menu item by one.
the selection.
or exits the current menu.
Alarm lamp
The alarm lamp gives visual alarm signals which meet applicable requirements. It
consists of 4 LEDs in parallel. The alarm LEDs receive electrical signal sent by the main
board and convert it into optical signal which is then sent to the panel through a light
conducting bar. The alarm LEDs light up in red and yellow.