All rights to this documentation and the
software it describes are owned by Leica
Microsystems Wetzlar GmbH. Copying of text
and illustrations – in full or in part – by printing,
photostat, microfilm or other techniques, including electronic systems, is only permitted
subject to the express written consent of Leica
Microsystems Wetzlar GmbH.
The information contained in the following
documentation represents the latest stage of
technology and knowledge. We have composed
the texts and illustrations with great care.
However, as it is impossible to eliminate the risk
of error completely, we cannot accept any kind
of liability for the correctness of the contents of
this manual. Nevertheless, we are always
grateful to be notified of any errors.
The information in this manual may be altered
without prior notice.
4
Contents
Important notes on this manual .......................7
General safety information ...............................8
General technical data...................................... 125
Main wearing and spare parts, tools ............. 127
EU Conformity declaration ............................... 128
Important notes on this manual
This manual is an integral part of the Leica
DM IRB microscope and must be read carefully
before you start using the microscope.
This manual contains important instructions and
information on the operating safety and
maintenance of the system. It must therefore be
kept in a safe place.
Text symbols and their meaning:
(1.2)
→ p. 20
The manual is multi-lingual. Due to the spiral
binding you can turn the language version you
want to the front.
The DM IR is available both as a life sciences
microscope and as a metallographic/industrial
microscope. In cases where the function and
operation are identical, the same text and
illustrations are used in both the separate
instruction manuals.
Numbers in brackets, e. g. (1.2) refer to
illustrations, in this example Fig. 1, item 2.
Numbers with an arrow, e.g. → p. 20 refer to a
particular page in this manual.
Special safety information is indicated by the
triangular symbol on the left and is given a
grey background.
Caution! Operation errors can damage the
!
*
microscope and/or its accessories.
Warning of hot surface.
Explanatory note.
Not part of all configurations.
7
General safety information
This instrument of safety class 1 has been built
and tested according to EN
VDE 0411-1, safety standards for electrical
measurement, control and laboratory equipment.
To keep the microscope in this safe condition,
it is essential to note the advice and warnings given in this manual.
The mains plug must only be inserted into a
grounded outlet.
If an extension cord is used, it must be grounded
as well. Any interruption of the ground
connector inside or outside the microscope or
disconnecting the ground connector can make
the microscope potentially dangerous. Intentional interruption is forbidden!
n. b.:
61 010-1/IEC
1010-1/
Make sure that only fuses of the specified type
and rating are used as replacements. It is
forbidden to use mended fuses or to shortcircuit the fuse holder.
The instruments and accessories described
in this manual have been safety-tested and
checked for possible hazards.
Before modifying the instrument in any way
or combining it with non-Leica products not
dealt with in this manual, it is essential to
consult the Leica agency for your area or the
main factory in Wetzlar!
Any unauthorized alteration to the microscope or use for which it was not intended
will automatically terminate any warranty
claim.
n. b.:
Using the ground connection, any accessories connected to the microscope
which have their own and/or a different
power supply can be given the same ground
conductor potential. Please consult our
servicing personnel if you intend to connect
units without a ground conductor.
8
n. b.:
n. b.:
n. b.:
The electric accessories of the microscope
are not waterproof. If water gets inside them,
it may cause electrical shock.
Do not put the microscope and its accessories too near a water supply or anywhere
else where water may get inside them.
Before changing fuses or lamps, always turn
the mains switch off and disconnect the
mains cable.
n. b.:
Protect the microscope from major temperature fluctuations. These may lead to
condensation which can damage the electric
and optical components.
Avoid skin contact when using immersion
oil! Ask the supplier for a safety information
sheet!
n. b.:
9
Intended application
The new DM IRB is the logical further
development of the successful inverted
research microscope from Leica. It is used
for examinations of cells and tissue, for
micromanipulation and microinjection techniques all the way through to microdissection or
confocal microscopy. The DM IRB has universal
application potential, incorporating all the
contrasting techniques of brightfield, darkfield,
phase contrast, DIC, fluorescence and
Hoffmann modulation contrast, which are all
easy to use and switch between. Variable
illumination and imaging light paths, HCS optics,
modular accessories and a wide range of
peripherals make the large DM IRB research
microscope from Leica a versatile and powerful
product.
10
The microscope and its components
Key subassemblies
The following views of the whole microscope
show and name important subassemblies of the
microscope and its accessories.
Fig. 1 – 2
1 Binocular phototube, 2 Eyepiece adapter tube, 3 Eyepieces, 4 Tube mount (tube change interface), 5 Tube port for photo/TV
connection, 6 Beamsplitter switch, 7 Mains switch, 8 Brightness adjustment, 9 Lateral TV port, 10 Coaxial coarse and fine
drive, 11 Fluorescence module, 12 ICT prism adjustment, 13 Sextuple objective nosepiece, 14 Centring buttons for incident light
field diaphragm, 15 Field diaphragm adjustment, 16 Filters, 17 Aperture diaphragm adjustment, 18 Lamphousing mount (or
mirrorhousing for two lamphousings), 19 Lamphousing, 20 Stage plate, 21 Analyser, 22 Tube lens module (Bertrand lens and
magnification changer), 23 Switch rod for lateral TV port, 24 Transmitted light illumination column, 25 Condenser, 26 Transmitted
light lamphousing, 27 Transmitted light field diaphragm, 28 SLR port, 29 Second lamphousing
Fig. 1 View from the rightFig. 2 View from the left
16
3
5
2
1
20
4
28
24
25
13
19
22
23
26
27
17
18
29
14
15
9
10
11
6
12
21
8
7
11
The stand
There are 5 basic versions of the DM IRB stand,
which allow over 50 microscope variants to be
configured.
These 5 basic versions are:
– Manual or electronic stand
– With or without integrated fluorescence axis
– With or without SLR front port or bottom port
– Lateral photo port 100 % or 80 %
– With or without integrated magnification
changer
The variants and their components, differences
and uses will be explained individually in this
manual. The function and operation of all
microscopy techniques and the necessary
accessories for the Leica DM IRB will be
described and explained in detail in the Opera-
tion section of this manual.
First of all, here is a general overview:
The stand with bottom port cannot be equipped
with an SLR front port. This variant is only
produced at a customer’s specific request.
The electronic stand additionally offers a
motorised objective nosepiece, electronic
focusing, IC objective prism coding, LCD display
of microscope functions and (optional)
motorised filter cube changer (RF4-mot module)
with electric dark flap, control panel, etc. All the
above-mentioned stands are also available as
fluorescence stands with integrated fluorescence axis. All the fluorescence stands (including the manual versions) can be fitted with
the RF4-mot module.
Tube mount
The interface between the stand and the tube is
called tube mount or tube change interface.
The tube mount is compatible with DM IR tubes
and the IR/R tube adapter which allows the use
of DM R tubes.
Stands
The basic stand has a photo port on the left for
the adaption of: TV camera, SLR camera or
photomicro system. The variants offered send
either 100 % or 80 % of the light to this photo
port.
Besides the lateral photo port, the SLR stand
also has another port facing the front (front
port*) which can be equipped with either an
SLR camera or a TV camera with c-mount
connection.
12
Tube
The tube, or its tube lens, produces the primary
image together with the objective.
DM IR tubes consist of a basic part, the
binocular part and the tube change ring. The
trinocular tube also has a photo/TV port. A
switchable mirror either directs the light 100 %
to the eyepieces or 100 % to the photo port, or
splits it 50 % /50 %.
Brightness adjustment
Aperture diaphragm
A 12 V 100 W transformer is built into the stand
for stepless regulation of brightness with the
brightness control.
Coarse and fine control
The coarse and fine focus control allows fast
and precise focusing of the microscope image.
Focusing is done by a vertical movement of the
objective nosepiece. The vertical movement
range is 9 mm.
Mains switch
The mains switch is used for switching the
microscope power supply on and off.
Incident light fluorescence device
The variant with incident light fluorescence
device contains the integrated fluorescence
axis and the lamp mount or a mirrorhousing
for adaption of a second lamphousing. The
fluorescence stand also comprises the fluorescence module which takes 4 filter cubes.
This module is also available as a motorised variant under the name RF4-mot module (see
Technical Description).
The aperture diaphragm determines the
resolution, field depth and contrast of the
microscope image. The best resolution is
achieved when the apertures of the objectives
and the condenser are roughly the same.
The aperture diaphragm in the illuminationlight path is not intended for adjusting the
brightness of the image. This should only be
done by turning the brightness adjustment
knob or using the neutral light damping filter.
Condenser
The condenser is a lens system through which
the light is collected and focused on the
specimen underneath. The condenser is designed to utilise the numerical aperture of the
objective.
Condenser height adjustment
The markings on the transmitted light illumination column indicate the height settings of the
different condensers.
n. b.:
13
Specimen stages and accessories
Transmitted light illumination unit
The specimen stage supports the specimens
that are to be examined through the microscope. Several options are available to
accommodate the wide variety of specimens
examined, such as object guides, extension
plates, specimen clips, scanning stage, heating
stage, etc.
Objective nosepiece and objectives
The objective nosepiece is used to hold the
objectives. L objectives with long working
distances, for example, are specially corrected
to respect different thicknesses of vessel bases.
All microscope objectives from magnification
2.5x to 100x can be used. All objectives from the
DM L and DM R range with 25 mm thread are
compatible. The performance data of Leica
objectives can be found in the chapter
“Technical Data; Performance Data” or on the
relevant objective lists available from your Leica
agency.
IR/R tube adapter
The tube adapter is used to adapt tubes from the
DM R range.
The transmitted light illumination unit consists of
the lamphousing and the transmitted light
illumination column. The transmitted light
lamphousing comprises a precentred 12 V 100 W
halogen lamp and a filter module for three
swing-in filters.
Field diaphragm
The field diaphragm is used to produce Koehler
illumination.
Lamphousings
A variety of lamphousings are offered for the
DM IRB (for halogen, mercury or xenon lamps).
The description and area of application can be
found in the operation section of this manual.
Filters
Filters are generally used to enhance the
contrast of the specimen and are assembled in
the illumination column. A variety of different
filters are easily changed.
Eyepieces
A magnified, virtual image of the actual
intermediate image produced by the objective is
produced with the eyepieces. They act as a
magnifier.
14
Installation site
The microscope should be used in a dust-free
room which is free of oil and chemical fumes
and extreme humidity. Also, the workplace
should not be exposed to major temperature
fluctuations, direct sunlight or vibrations. These
may impair measurements or photographs of the
microscope image.
Ambient conditions:
Temperature10 –36 °C
Relative humidity0– 80 % to 30 °C
Microscopes in warm and humid climates need
special care to prevent build-up of fungus.
Further details are given in the chapters
“Maintenance” and “Storage”.
n. b.:
Lamphousings* and power units* must be
placed at least 10 cm away from the wall and
from flammable objects.
15
Unpacking
Installation
Please compare the delivery carefully with the
packing note, delivery note or invoice. We
strongly recommend that you keep a copy of
these documents with the manual, so that you
have information on the time and scope of the
delivery later when ordering more equipment or
when the microscope is serviced. Make sure
that no small parts are left in the packing
material. Some of our packing material has
symbols indicating environment-friendly recycling.
n. b.:
Keep the packing material for storage and
transport of the microscope and its accessories.
Try to avoid touching the lens surfaces of the
optics. Any fingermarks on the glass surfaces
should be removed with a soft leather or
linen cloth. Even small amounts of finger
perspiration can attack the surfaces of
optical instruments within a short time. Further information is given in the Maintenance
and Cleaning chapters.
n. b.:
● First take all the components out of the
transport and packing material.
● Put the basic DM IRB stand on a desk which
has enough room for it.
On no account should the microscope be
connected to the power socket yet!
n. b.:
16
Assembly
Assembly tools
Installation and assembly of the microscope
should preferably be carried out together with a
member of Leica sales or service staff.
Only a few ordinary screwdrivers are required
for assembly, and these are supplied with the
microscope.
Assembly of the transmitted light illumination
column
Wipe off the interface surface (4.3) with a dry
cloth. Tilt the illumination column (4.1) slightly to
the back and insert so that the pin (4.2) engages
in the groove of the interface surface (4.4).
Erect the TL illumination column and secure
with the 4 screws.
When screwing on the TL illumination column,
do not hold onto it so that optimal alignment to
the optical axis is guaranteed. The angle of tilt of
the illumination column can be varied or
clamped securely in the vertical position with
the knurled screw (5.1).
1
3
2
4
6
5
17
The lamphousing for transmitted light
illumination for 12 V 100 W halogen lamps with
single-lens aspherical collector and heat
protection filter is an integral part of the
transmitted light illumination column. The halogen lamp is preassembled. The chapter on
Troubleshooting includes a description of how
to assemble and change halogen lamps.
The cable on the illumination column can then
be connected to the 12 V 100 W socket on the
back of the microscope stand.
Assembly of condensers
The technical description of the condensers can
be found in the chapter “Technical description”.
All condensers of the Leica DM IRB are
equipped with a 6-position rotating disc (6.2 and
8.3) and can be individually fitted with the
corresponding annular diaphragms for phase
contrast (PH), darkfield (DF) or IC prisms for TL
interference contrast (ICT) (10).
Usually the annular diaphragms are already
inserted in the condenser disc in the factory, so
you will not normally have to fit them yourself.
The condenser disc (11.5) is removed from the
condenser by slackening the screw (11.4) on the
underneath of the condenser.
Fig. 4 Assembly of transmitted light illumination column
1 Transmitted light illumination column, 2 Pin of TL illu-
mination column, 3 Support surface, 4 Groove of support
surface, 5 Drill holes for fixing screws
3
5
2
5
4
1
18
Fig. 5 Transmitted light illumination column, back view
1 Knurled screw for clamping the transmitted light illumi-
nation column
1
Insert the light rings for Phaco (identified by the
code numbers 0, 1, 2, 3 and the intercept
distance S of the corresponding condenser top,
e. g. 2 S1) and the DF diaphragm (identified by D
for darkfield and the intercept distance S of the
corresponding condenser top, e. g. D S1, see
table “Technical Description”) in the slots of the
disc as follows:
– Slightly unscrew the two centring screws
(10.11) using the supplied centring key (12.1).
– Insert the diaphragms so that the mount fits
under the spring (10.3) of the slot.
– When the light rings are assembled, their
identification code must be visible i.e. pointing
upwards (12.3, 12.4 and 12.5).
– Insert the light rings in the order 0, 1, 2, 3. The
DF diaphragm can only be inserted in a large
hole.
– Using the centring keys, screw the centring
screws back in until they no longer protrude
over the outer edge of the disc.
– Fit IC condenser prisms if appropriate (see
assembly of ICT objective prisms).
– Insert the plastic labels (10.7) in the disc
(12.2), allocating them to the corresponding
diaphragms.
– Mark any empty holes with white labels.
Insert the disc into the condenser with notches
(10.6) facing upwards – towards the aperture
diaphragm (6.3 and 8.4) – and screw down (11.4).
Fig. 6 Condenser 0.30 S70
1
Condenser top 0.30 S70 (not for use with condenser base
(8.1)), 2 Condenser disc, 3 Aperture diaphragm, 4 Filter holder,5 Condenser clamp screw
Fig. 9 Condenser tops for condenser base (8.1)
1 Condenser top 0.53 S23, 2 Condenser top 0.90 S1,
3 Condenser top P 1.40 OIL S1, 4 Spacer ring for assembling
9.2 and 9.3
4
2
3
1
2
3
4
Fig. 10 6-position condenser disc, empty
1 Condenser disc with slots for light rings and IC condenser
prisms, 2 Guide groove for IC condenser prisms (2
concealed, 3 Spring, 4 Holes for centring keys, 5 Spaces for
label plates, 6 Notches, 7 Label plates, 8 Light ring for
darkfield, 9 IC condenser prism with 2 guide cams, 10 Light
ring for phase contrast, 11 Centring screws
nd
groove is
20
6
10
9
4
8
11
4
5
7
1
2
3
Assembly of IC condenser prisms
The IC condenser prisms are assembled at the
factory. The following steps are only necessary
in case of a retrofit:
Remove the condenser disc (11.5) by slackening
the screw (11.4) on the underneath of the
condenser.
– Using the centring keys (12.1), slightly un-
screw the two centring screws (10.11).
– IC condenser prisms can only be inserted into
the large holes of the condenser disc which
have guide grooves (10.2).
– Insert the IC condenser prisms in ascending
order, e. g. K1, K2 and so that the mount fits
under the spring (10.3) in the slot and the
2 guide cams engage in the grooves of the
condenser disc (10.2).
– When the prisms are inserted, their
identification code, e. g. K10, must be visible
and pointing towards the
centre of the disc
(12.6 and 12.7).
– Using the centring keys, screw the centring
screws back in until they no longer protrude
over the outer edge of the disc. The prism is
adjusted with the left centring screw only (see
operation of ICT). The right centring screw
must never restrict the adjustment range.
– Assemble the light rings and DF diaphragm if
appropriate (see previous section).
– Insert the label plates (10.7) corresponding to
the relevant IC condenser prism.
– Mark any empty holes with white labels.
– Remove any finger marks or dust on the
prisms carefully.
– Put the condenser disc back in the condenser
Fig. 12 6-position condenser disc, fully equipped
1 Centring keys for centring screws (in working position),
2 Label plates, 3, 4 Light rings for phase contrast, 5 Light ringfor darkfield, 6, 7 IC condenser prisms, 8 Hole for brightfield
7
6
8
5
3
4
11
2
21
Condenser top
The base and top of condenser 0.30 S70 form a
self-contained unit (Fig. 6).
The condenser top 0.30 S70 (13.4) cannot be
used with the condenser base (8.1).
The condenser top 0.53 S23 (8.2 and 9.1) is
screwed straight on to the condenser base (8.1).
A spacer ring (9.4 and 11.3) must be used for
assembling the condenser tops 0.90 S1 and
P 1.40 OIL S1 (9.2 and 9.3).
Assembly of the condensers to the illumination
column
Condenser 0.30 S70
Tilt the TL illumination column to the back (13.1).
Insert the condenser 0.30 S70 (13.4) from below
into the dovetail guide of the illumination column
(13.2), with the condenser top pointing towards
the microscope stage. Adjust the height of the
condenser until its upper edge is flush with the
condenser height marking S70 on the illumination column. Secure the condenser with the
supplied hexagonal screwdriver. Erect the TL
illumination column.
Fig. 14 Assembly of condenser 0.30 S70
1 Condenser 0.30 S70 in working position (upper edge of
condenser is flush with condenser height marking S70)
1
Condensers 0.53 S23 and 0.90 S1
5
3
1
4
2
6
7
4
3
1
2
5
With the illumination column tilted to the back,
insert the condenser holder (15.4) into the
dovetail guide of the illumination column from
below (15.2). The condenser height adjustment
should point to the left. Adjust the height of the
condenser holder until its upper edge coincides
with the condenser height marking S23 or S1 on
the illumination column (16.1), depending on the
condenser top used. Secure the condenser holder with the hexagonal screwdriver and clamp
screw (15.5). Mount the base part of the
condenser with the dovetail guide (8.6) to the
slide change mechanism (7.1) of the condenser
holder (17). The condenser top should point
downwards and the aperture diaphragm control
towards the front (17.3). Slacken the clamp
screw (17.5) and push the condenser back as far
as the stop. Retighten the clamp screw (17.5)
slightly.
Fig. 15 Assembly of condenser holder
1 Transmitted light illumination column, 2 Dovetail guide,
3 Condenser height markings S1, S23 and S70, 4 Condenserholder, 5 Clamp screw for securing the condenser holder,
6 Clamp screw for field diaphragm module, 7 Transmitted light
lamphousing
Fig. 16 Assembly of condenser holder
1 Condenser holder in working position for condenser
0.53 S23 (upper edge of condenser holder coincides with
condenser height marking S23)
1
Fig. 17 Assembly of 0.53 S23 condenser
1 Dovetail guide of the condenser, 2 Sliding condenser
changer, 3 Aperture diaphragm adjustment, 4 Condenser top
0.53 S23, 5 Condenser clamp screw
23
Assembly of field diaphragm
Assembly of filters and filter holder
To enable Koehler illumination when using
condensers 0.53 S23 and 0.90 S1, a field
diaphragm has to be assembled. Insert the field
diaphragm module (18.1) into the mount (Fig. 18)
from below. The diaphragm adjustment (18.2)
should point in the direction of the tube. Secure
with clamp screw (18.3).
Fig. 18␣ ␣ Assembly of field diaphragm
1 Field diaphragm module, 2 Field diaphragm adjustment,
3 Clamp screw for securing the field diaphragm module
3
1
2
The Leica DM IRB is equipped with a holder with
spaces for 3 filters with 40 mm diameter.
The filters are already fitted into the holder at
the factory. If you are retrofitting filters yourself,
assemble as follows:
● Slacken the clamp screws (Fig. 19.1) and
remove the filter holder.
● Put the filters into the holder (20).
● Mount the filter holder onto the transmitted
light illumination column and secure in position with the clamp screws.
Fig. 19 Assembly of filters
1 Clamp screw for securing the filter holder
1
24
Fig. 20 Assembly of filter holder for 3 filters
Assembling the ICT objective prisms
Assembling the IC module and IC objective
prisms
The IC prism disc with the IC prisms ordered by
the customer are already assembled in the
microscope at the factory. In case you want to
retrofit the IC prism disc, please proceed as
follows:
Fig. 21 IC objective prism disc without fixing knurl
1 IC objective prism in mount, 2 Code letter (e. g. A), 3 Washerand screw, 4 Stop pin
1
2
3
4
Remove the front cover (22.2) under the
objective nosepiece (23.1) after slackening the
Allen screws (22.4).
Insert the IC prism disc (22.1) in the mount and
tighten with the two Allen screws. n. b.: insert
the prism disc with the prism mount pointing
downwards.
Retrofitting individual IC prisms:
Please align prisms against the stop pin (21.4)
and only screw down lightly to avoid strain.
Insert the prisms so that the code letter, e. g. A
points upwards and is readable.
Label the position of the prism on the front of the
ICR prism disc with a label plate (22.5).
Examples of prisms:
Prism A – for objectives N PLAN 5x, 10x.
Prisms D and D1 – both for objectives N PLAN
20x, 50x, 100x and HC PL FLUOTAR 5x – 100x.
Fig. 22 Assembly of IC objective prism module
1 IC objective prism module, 2 Cover, 3 Fixing screw, 4 Holefor fixing screws, 5 Label plates, 6 Knurled fixing knob
3
1
4
2
5 6
Abb. 23 Assembly of IC objective prism module
1 Objective nosepiece, 2 Mount for IC objective prism module,
3 Stop pins
1
2
3
3
25
Differences between prism D and D1
Inserting the analyser
Prism D is the standard prism with greater
shearing and therefore higher detection sensitivity for minute topological and refractive index
variations in the specimen. Prism D1 has smaller
shearing than prism D and a lower detection
sensitivity for topological and refractive index
variations.
However, prism D1 is better at resolving details
of fine specimen structures.
Remove the blind slide and insert the analyser
st
(24.2) from the left as far as the 1
clickstop.
Inserting the polariser
The polariser is inserted into the filter holder of
the condenser. In addition a whole-wave
compensator is applied to the back of the
polariser. It is activated by turning the polariser
over, in order to enable colour contrasting in
polarisation or interference contrast (the com-
The fluorescence module (Fig. 26) is part of the
fluorescence stand, but is also available as a
retrofit kit. To retrofit the fluorescence module,
remove the blind cover from the stand. The
fluorescence module can be fitted with up to
four different filter cubes (26.3). They are
inserted into the dovetail mount (26.2) of the
fluorescence module with their engraving facing
downwards (towards the turret plate). The
fluorescence module is inserted on a dovetail
guide into its slot on the stand by pushing it as
far as the stop. One part of the fluorescence
module is the anti-glare protection (27.1), which
can be inserted between the tube and the stage.
Proceed in the same way if you are inserting a
motorised filter cube changer instead of the
manual filter module. Also read the manual for
the electronic version.
Assembly of the lamp mount, mirror housing,
lamphousing, illumination telescope
1. Insert the lamp mount or mirror housing in
the back panel and screw down with Allen
screws. Engage the guide pin of the lamp
mount (29.1) in the back panel of the
microscope stand (28.2).
2. Hold the lamphousings 107/2, 107, 106 z
against the lamphousing mount and secure
with the fixing screw (Fig. 31).
3. We recommend using the illumination
telescope for gas discharge lamps. This is
inserted between the lamp mount and the
lamphousing 106 z (30.4) and magnifies the
image of the focal point of the lamp by the
factor 2x in the entrance pupil of the
objective. This results in a significantly
higher illumination intensity for fluorescence.
Fig. 26 Fluorescence module
1 Rotatable turret, 2, 4 Dovetail mounts for filter cubes (the
numbers 1␣ –␣ 4 are markings of the assembly positions),
3 Filter cubes, 4 Display of the position in the light path,
5 Switch rod with BG 38 and light stop
3
1
4
2
5
Fig. 27 Anti-glare protection
27
4. Connect the lamp plug to the connecting
socket in the stand (28.3).
5. Insert light filters, 50 mm Ø into the 2 spaces
in the lamphousing mounts (29.4).
!
n. b.:
Connect the appliance cable to the mains
socket on the microscope stand (28.4)!
Fig. 28 Back view of microscope
1 Space for assembling a lamphousing mount or mirror
housing, 2 Hole for guide pin, 3 Socket for lamp plug, 4 Mains
socket, 5 Potential equalisation
2
3
Fig. 30 Mirror housing and illumination telescope
1 Lever for mirror switching, 2 Lateral lamphousing mount
with fixing screw, 3 Back lamphousing mount with Allen
screw, 4 Illumination telescope for gas discharge lamps
32
1
5
4
Fig. 29 Lamp mount
1 Guide pin, 2 Lateral lamphousing mount, 3 Dovetail ring for
mounting to stand, 4 2 spaces for light filters, 5 Allen screws
for fixing
5
4
2
1
3
Fig. 31 Lamphousing 106 z L
1 Collector adjustment, 2 Vertical lamp adjustment, 3 Horizon-tal lamp adjustment, 4 Mount ring
2
28
4
1
4
3
1
Assembling and exchanging incident light
lamps
Exchanging the 12 V 100 W halogen lamp:
n. b.:
Disconnect the lamp and lamphousing from
the power supply. Pull out the mains plug.
Fig. 32a –c Lamphousing 107 L
1 Mount with halogen lamp, 2 Collector
Lamphousing 107 L
Slacken the fixing screw on the cover and lift off
the cover (Fig. 32a and 32b). Move the collector
to the front and pull the defect 12 V 100 W lamp
out of the base towards the front. Without
removing its protective cover, put a new lamp
into the base, without tilting, as far as it will go.
n. b.:
Leave the protective cover on the lamp until it
is in position.
Avoid making finger marks or wipe off
immediately. Close the lamphousing.
2
1
29
Lamphousing 106 z L
Slacken the fixing screw on the lid (33.10). Pull
the cut-out plug slightly out of the socket and flip
up lid (33.11; 33.1).
Move the collector to the front and lift the defect
lamp out of the base (33.1; 33.2; 33.3). For
convenience, the lamp holder can be removed
from the lamphousing as well. To do this,
slacken the fixing screws on the lamp holder
(33.10) and pull out lamp holder (Fig. 34).
Without removing its protective cover, put a new
lamp into the base, without tilting, as far as it will
go.
n. b.:
It is important to leave the protective cover
on the lamp until it is in position.
Avoid making finger marks or wipe off
immediately. Close the lamphousing.
Fig. 33 Lamphousing 106 z L
1 Lid, flipped up, 2 Collector, 3 12 V 100 W halogen lamp or gasdischarge lamp in holder, 4, 9 Cover fixing screws, 5 Reflector,
6, 7, 8 x-y adjustment screw for reflector, 10 Fixing screws forlamp mount, 11 Socket for contact plug
1
5
6
7
8
9
1011
10
2
3
4
30
Fig. 34 12 V 100W lamp holder with halogen lamp
Loading...
+ 100 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.