LEICA DMI6000 B User Manual

Leica DMI6000 B
Operating Manual
1
Published October 2004 by:
Leica Microsystems Wetzlar GmbH Ernst-Leitz-Straße D-35578 Wetzlar (Germany)
Responsible for contents: Bernard Kleine (Marketing CM, Compound Microscopy, Product Manage­ment) Holger Grasse (Safety Officer according to MPG §30) In case of questions, please contact the hotline:
Phone +49(0)6441-292286 Fax +49(0)6441-292255 E-mail: MQM-Hotline@leica-microsystems.com
2
Leica DMI6000 B
Operating Manual
3
Copyrights
Copyrights
All rights to this documentation are held by Leica Microsystems Wetzlar GmbH. Reproduc­tion of text or illustrations (in whole or in part) by print, photocopy, microfilm or other method (in­cluding electronic systems) is not allowed with­out express written permission from Leica Mi­crosystems Wetzlar GmbH.
The term "Windows" may appear in the following text without further identification. It is, however, a registered trademark of Microsoft Corpora­tion. The names of companies and products used herein may be trademarks of their respec­tive owners.
The instructions contained in the following doc­umentation reflect state-of-the-art technology and knowledge standards. We have compiled the texts and illustrations as accurately as pos­sible. Nevertheless, no liability of any kind may be assumed for the accuracy of this manual’s contents. Still, we are always grateful for com­ments and suggestions regarding potential mis­takes within this documentation.
The information in this manual is subject to modifi­cation at any time and without notification.
4
Contents
Contents
1. Important Notes about this Manual ...... 7
2. Intended Purpose of the Microscope ... 8
3. Safety Notes ............................................... 9
3.1 General Safety Notes ............................... 9
3.2 Electrical Safety ........................................ 10
4. Overview of the Leica DMI
4.1 Specifications ............................................ 12
4.2 Glossary....................................................... 16
5. Unpacking the Microscope .................... 22
6. Assembling the Microscope .................. 25
6.1 Assembly Tools .......................................... 25
6.2 Installation of the Transmitted Light
Illumination Carrier (DL) ........................... 26
6.3 Installation of the DIC Module
and DIC Objective Prisms ........................ 27
6.4 Installation of Specimen Stages............. 28
6.5 Installation of Condensers ....................... 34
6.6 Installation of Eyepieces .......................... 39
6.7 Installation of Objectives ......................... 39
6.8 Installation of Filters
in the Illumination Arm ............................. 40
6.9 Installation of the
Transmitted Light Lamp Housing ............ 40
6.10 Installation and Replacement of the
Transmitted Light Lamps: .........................
Lamp Housing 107 or 107/2 ...................... 41
6000 B......... 12
6.11 Installation of Lamp Housing Mount
and Mirror Housing ................................... 42
6.12 Installation and Replacement
of Incident Light Lamps ............................ 44
6.13 Equipping the Incident Light
Tu rret Disk ................................................... 48
6.14 Installation of the Polarizer and
Analyzer....................................................... 50
6.15 Optional Accessories ............................... 52
6.16 Connection to the
Electronics Box CTR6000 ......................... 53
6.17 Connection to the Computer ................... 54
6.18 Connection to the Power Supply ............ 54
7. Start-up........................................................ 55
7.1 Functional Principle .................................. 55
7.2 Switching on the Microscope ................. 59
7.3 The LeicaScreen........................................ 60
7.4 The Function Buttons on the Stand ....... 61
7.5 The SmartMove Remote Control
Module......................................................... 64
7.6 Illumination ................................................. 65
7.6.1 Transmitted Light ............................ 65
7.6.2 Incident Light - Fluorescence ...... 68
7.7 Checking Phase Contrast Rings ............. 69
7.8 Setting the Motorized Polarizer ............. 70
7.9 Adjusting the Light Sources .................... 71
5
Contents
8. Operation .................................................... 74
8.1 Switching On .............................................. 74
8.2. Contrast Methods ...................................... 76
8.2.1 Brightfield (TL)................................. 76
8.2.2 Phase Contrast (TL) ........................ 77
8.2.3 Darkfield (TL) ................................... 77
8.2.4 Polarization (TL) .............................. 78
8.2.5 Differential
Interference Contrast (TL) ............ 79
8.3 Fluorescence .............................................. 80
8.4 Combination Methods .............................. 81
8.5 Focusing ...................................................... 82
8.6 Tubes............................................................ 84
8.7 Eyepieces .................................................... 85
8.8 Objectives ................................................... 85
8.9 Stages and Object Displacement ........... 88
8.10 Magnification Changer............................. 89
8.11 Light Sources ............................................. 89
8.12 Aperture and
Field Diaphragm ......................................... 90
9. Troubleshooting......................................... 91
10. Care of the Microscope ........................... 95
10.1 Dust Cover .................................................. 95
10.2 Cleaning....................................................... 95
10.3 Handling Acids and Bases ...................... 96
11. Major Consumable and Replacement
Parts ............................................................ 97
12. Dimensions ................................................. 98
13. Abbreviations and Pictograms ............... 99
14. Index ............................................................ 101
15. EU Declaration of Conformity ................. 104
6
1. Important Notes about this Manual
1. Important Notes about this Manual
Caution!
This operating manual is an essential com­ponent of the microscope, and must be read carefully before the microscope is assem­bled, put into operation or used.
Text symbols, pictograms and their meanings:
(1.2)
p.20
This operating manual contains important in­structions and information for the operational safety and maintenance of the microscope and accessories. It must therefore be kept safely for future reference. A separate manual is available on CD-ROM cov­ering the operation of the Leica Application Suite (LAS).
Numbers in parentheses, such as "(1.2)", corre­spond to illustrations (in the example, Figure 1, Item 2).
Numbers with pointer arrows (for example p.20), point to a certain page of this manual.
Caution! Special safety instructions within this manu­al are indicated with the triangle symbol shown here, and have a gray background.
Caution! The microscope and accessories can
!
*
be damaged when operated incorrectly.
Explanatory note.
Item not contained in all configurations.
7
2. Intended Purpose of the Microscope
2. Intended Purpose of the Microscope
The Leica DMI 6000 B microscope described in these instructions is intended for biological rou­tine and research applications. This includes the examination of samples taken from the human body to provide information on physiological or pathological states or congenital abnormalities, or determing the safety and compatibility with po­tential recipients, or monitoring therapeutic measures.
The Leica DMI of Leica’s proven inverted research micro­scopes. It is designed for cellular and tissue ex­amination, micromanipulation and microinjec­tion techniques, microdissection, and confocal microscopy. The Leica DMI universal deployment, all contrast methods such as darkfield, brightfield, phase contrast, DIC, flu­orescence, and modulation contrast are integral to the microscope and can be adapted or changed quickly and easily. Variable illumination and imaging beam paths, as well as HCS optics, modular accessories and a comprehensive range of peripherals, complement the Leica DMI 6000 B inverted research stand.
6000 B is a further development
6000 B is suitable for
The above-named microscope complies with the Council Directive 98/79/EEC concerning in vitro diagnostics. It also conforms to the Council Directives 73/23/EEC concerning electrical ap­paratus and 89/336/EEC concerning electromag­netic compatibility for use in an industrial envi­ronment.
Caution!
The manufacturer assumes no liability for damage caused by, or any risks arising from using the microscope for other purposes than those for which it is intended or not us­ing it within the specifications of Leica Mi­crosystems Wetzlar GmbH. In such cases the declaration of conformity shall cease to be valid.
Caution!
This (IVD) device is not intended for use in the patient environment defined by DIN VDE 0100-710. Neither is it intended for combin­ing with medical instruments according to EN 60601-1. If a microscope is electrically connected to a medical instrument accord­ing to EN 60601-1, the requirements de­fined in EN 60601-1-1 shall apply.
8
3. Safety Notes
3. Safety Notes
3.1 General Safety Notes
This safety class 1 device is constructed and tested in accordance with EN 61010-2-101:2002, EN 61010-1:2001, IEC 1010-1:2001, Safety regulations for electrical measuring, con­trol, and laboratory devices.
In order to maintain this condition and to en­sure safe operation, the user must follow the instructions and warnings contained in this operating manual.
Caution!
Caution!
The devices and accessories described in this operating manual have been tested for safety and potential hazards. The responsible Leica affiliate or the main plant in Wetzlar must be consulted whenev­er the device is altered, modified, or used in conjunction with non-Leica components that are outside of the scope of this manual.
Unauthorized alterations to the device or noncompliant use shall void all rights to any warranty claims!
9
3. Safety Notes
3.2 Electrical Safety General specifications
Caution!
Leica CTR6000 Electronics Box
For indoor use only. Supply voltage: Frequency: Power input: Fuses:
Ambient temperature: Relative humidity: Overvoltage category: Pollution degree:
Microscope
For indoor use only. Supply voltage: Frequency: Power input: Fuses: Ambient temperature: Relative humidity: Overvoltage category: Pollution degree:
ebq 100 Supply Unit*
90-250V~ 50-60 Hz max. 290VA T6.3 A (IEC 60127-2/3) 15-35°C max. 80% to 30°C II 2
90-250V~ 50-60 Hz See CTR6000 See CTR6000 15-35°C max. 80% to 30°C II 2
Power plugs may only be plugged into an outlet equipped with a grounding contact.
Do not interfere with the grounding function by using an extension cord without a ground wire. Any interruption of the ground wire in­side or outside of the device, or release of the ground wire connection, can cause the device to become hazardous. Intentional ground interruption is not permitted!
Caution!
Peripheral devices with their own or sepa­rate power supplies that are connected to the microscope can have the same protec­tive conductor potential by connecting them to the ground screw on the back of the Leica CTR6000 electronics box. For connections without a ground connector, Leica Service must be consulted.
For indoor use only. Supply voltage: Frequency: Power input: Fuses: Ambient temperature: Relative humidity: Overvoltage category: Pollution degree: (See enclosed manual)
10
90-250V~ 50-60 Hz max. 155VA 2xT2A (IEC 127) 10-36°C max. 80% to 30°C II 2
Caution!
Never use any fuses as replacements other than those of the types and the current rat­ings listed here. Using patched fuses or bridging the fuse holder is not permitted. The use of incorrect fuses may result in a fire hazard.
Caution!
The microscope’s electrical accessory com­ponents are not protected against water. Water can cause electric shock.
Caution!
Protect the microscope from excessive tem­perature fluctuations. Such fluctuations can lead to the accumulation of condensation, which can damage the electrical and optical components. Ambient temperature: 15-35°C.
Caution!
3. Safety Notes
Before exchanging the fuses or lamps, be absolutely certain to switch off the main power switch and remove the power cable.
11
4. Overview of the Instrument
4. Overview of the Leica DMI 6000 B
4.1 Specifications
Contrast Methods
Transmitted Light Axis
Incident Light Axis
•Transmitted light (DL): BF, DF, PH, DIC, Pol
• Incident light (IL): Fluo
• Combination (DL/IL): Fluo/DIC, Fluo/PH
•Intermediate pupil: IMC (integrated modulation contrast) IPH (integrated pos./neg. phase contrast)
• Automatic Illumination Manager
(aperture, field diaphragm, intensity, process switching)
• Automatic, color-neutral intensity control
• Manual or motorized shutter
• Lamp housing mount for interchangeable lamp housings.
• Automatic, electronic condenser identification
• Aautomatic Illumination Manager
(aperture, field diaphragm, intensity, process switching)
• Automatic, color-neutral intensity control
• Motorized shutter (switching speed < 50ms)
• Lamp housing mount for up to 3 interchangeable light sources
• Motorized 6-place filter turret
• Fluorescence Intensity Manager (FIM) (reduction of incident illumination intensity)
• Mechanical booster lens for central boosting of fluorescence or uniform distribution
• Motorized Excitation Manager to monitor fluorescence emission when using double and triple filter cubes
• Ultrafast filter wheel for 3 excitation wavelengths (switching speed < 50 ms)
12
Tube
• Ergonomic with or without camera port at left
•2 switching positions: 100%VIS and 50%VIS / 50%CAM
• Optional Bertrand lens
• Eye spacing adjustment
• Height and angle adjustment (10° - 45°)
4. Overview of the Instrument
Magnification Changer
Objective Turret
Stages
• Motorized
•3 switching positions (choice of magnifications: 1x, 1.5x, 1.6x or 2.0x)
• Effective for all camera ports
or
• Manual
•2 switching positions (choice of magnifications: 1x, 1.5x, 1.6x or 2.0x)
• Effective on tube port
• Motorized and coded
• 6x for objectives with M25 thread and 45mm parfocal distance
• For DIC: motorized/coded Wollaston prism carousel
• Anti-vibration locking
Fixed regular stages
• Ceramic-coated stage plate (248mm x 204mm)
•Heated stage plate (from 3°C above room temperature to 60°C) (248 x 212mm)
•Temperature-controlled stage plate (0°C to 60°C) (248mm x 212mm)
Fixed micromanipulation stages
• Ceramic-coated stage plate (248mm x 204/122mm)
•Heated stage plate (from 3°C above room temperature to 60°C) (248mm x 204/122mm)
•Temperature-controlled stage plate (0°C to 60°C) (248mm x 204/122mm)
Regular manual 3-plate cross-stage
• Positioning range: 83mm x 127mm
• 20 optional inserts (standard, heating, cooling) for a variety of applications, size of inserts:160mm x 110mm (compatible with scanning stages)
Manual micromanipulation 3-plate cross-stage
• Positioning range: 40mm x 40mm
•3 optional inserts for a variety of applications
Motorized micromanipulation 3-plate cross-stage
• Positioning range: 40mm x 40mm
Scanning stage IM 120 x 100 (motors on top)
• 1mm, 2mm, 4mm spindle pitch (higher resolution v. higher speed)
• 20 optional inserts (standard, heating, cooling) for a variety of applications, size of inserts:160mm x 110mm
13
4. Overview of the Instrument
Scanning stage IM 120 x 100 (motors on bottom)
• 1mm, 2mm, 4mm spindle pitch (higher resolution vs. higher speed)
• 20 optional inserts (standard, heating, cooling) for a variety of applications, size of inserts:160mm x 110mm
Condensers
Z focus
Observation Ports
• Motorized and coded
• Motorized or manual aperture diaphragm
• Contrast methods: BF, DF, PH, DIC, Pol, IMC
• Automatic method switching
• Condenser turret with 7 positions for contrast methods
•2 condenser housings (S1-S28 and S70)
• Condenser heads: S1/1.4 oil, S1/0.9 dry, S23/0.53, S28/0.55
• Condenser heads can be swung out
• Condenser S70 with additional lens for low magnifications
• All condensers suitable for magnifications from 1.25x to 100x
• With or without motorized or manual polarizer
•With or without motorized or coded Wollaston prism disk
• Motorized and coded
• 9mm travel (1mm below, 8mm above the stage)
• Maximum travel speed: 5mm/s
•5 focus steps: 0.05 µm, 0.1 µm, 0.7 µm, 1.5 µm, 5.0 µm
• Electronic focus repositioning
• Automatic lowering prior to objective change
• Electronic parfocality
• Motorized and coded
• Left side ports (100%, 80%, or 50% transmission)
• Left side port dichroic splitting at 680 nm
• Right side ports (100%, 80%, or 50% transmission)
• Bottom port
Optional
• Top port with 2 switching positions
• 100% to eyepieces
• 50% to eyepieces/ 50% to port
14
4. Overview of the Instrument
Controls
Electronics Box Leica CTR6000
Interfaces
•7 fixed control buttons for illumination and apertures
•7 variable function buttons behind the focus controls
•3 fixed control buttons for focus steps
•2 focus hand wheels
•7 buttons for fluorescence cubes and shutters
•4 buttons for magnification changers and ports
• SmartMove: ergonomic controller for x, y, z and 4 additional vari­able function buttons
• Separate control unit for all motorized and electronic elements of the microscope such as:
• Objective turret
• Ffocus
• Ports
• Magnification changer
• Fluorescence
• Condenser
• Motorized stages With
• Power supply for 100W halogen lamp
• Power supply for SmartMove
•2 x RS232C
•2 x USB
•4 x external/internal peripherals
Software Tools
• Leica Application Suite (LAS) for Windows For:
• Microscope and camera configuration
• Microscope and camera control
• Image acquisition
TM
2000, XP with plug-ins
15
4. Overview of the Instrument
4.2 Glossary The Stand
Four basic versions of the Leica DMI stand (DMI 3000 B, DMI 4000 B, DMI 5000 B, DMI 6000 B) are available which can be combined into a wide range of microscope variants.
The basic building blocks of the Leica DMI6000 B stand are:
• DMI6000 B electronic stand
• Integrated fluorescence axis with motorized filter cube changer (6 positions)
• With or without bottom port
• Lateral camera ports, 100%, 80%, or 50%
• Optional top port at the left side of the tube
• Integrated motorized tube lens changer
• Optional Bertrand lens
The individual variants and their components, differences, and applications are described in this manual. The function and performance of all microscopy techniques and required accesso­ries of the Leica DMI6000 B will be described in detail in the section of this manual that covers the operation of the microscope.
Tube
The tube and tube lens create the primary image together with the objective. The tubes are an integral part of the stand and consist of a basic body and a binocular section. The trinocular tube also features a photo/video port. A switchable mirror diverts either 100% of the light to the eyepieces or camera port, or splits it, with 50% each going to the eyepieces and camera port. A Bertrand lens is also available as an option.
Eyepieces
The eyepieces create an enlarged, virtual image of the actual intermediate image created by the objective. The eyepiece serves as a magnifier in this respect.
Intensity Controller
The stand contains a 12V 100 W transformer for continuous regulation of the intensity via the in­tensity controller. The intensity can be adjusted using the controls (1.6).
Focus Wheel
The focus wheel allows quick, precise focusing of the microscopic image. Focusing is realized by the vertical travel of the objective turret, with a total range of 9mm.
Incident Light-Fluorescence Unit
The stand features an integrated fluorescence axis and a motorized filter cube changer with 6 positions.
Aperture Ddiaphragm
The aperture diaphragm determines the resolu­tion, depth of field, and contrast of the micro­scopic image. The best resolution is obtained when the apertures of the objective and the condenser are roughly the same.
The aperture diaphragm in the illumination light path is not for setting the image bright­ness. This should be done only with the inten­sity controls or neutral filters.
Caution:
16
4. Overview of the Instrument
Condenser
The condenser is a lens system that gathers light and projects it onto the specimen from above. The condenser is designed for the utiliza­tion of the numerical aperture in the objective.
Condenser Height Adjustment
The markings of the transmitted-light column indi­cate the height to be set for the used condenser.
Stages and Accessories
The stage is designed to accommodate the specimens to be observed. Mechanical and mo­torized 3-plate cross-stages are available for the Leica DMI 6000 B.
Motorized Objective Turret and Objectives
The motorized objective turret is designed to ac­commodate the objectives. The L-objectives with their long working distances especially take into consideration the correction of varying container bottom thicknesses. All microscope objectives are usable, from
1.25:1 to 100 : 1 magnification. All objectives in
the Leica product range with a 25mm thread and coverslip correction are compatible. For perfor­mance data on Leica objectives, please refer to the most current valid objective lists available from your Leica representative.
Transmitted Light Illumination Unit
The transmitted light illumination unit consists of an illumination carrier and the transmitted light illumination column. The transmitted light illumi­nation carrier does not contain a lamp housing, but a filter module for two swing-in filters and, depending on the condenser used, a field dia­phragm.
Filter
The filters are generally used to enhance the contrast of the specimen and are installed in the illumination carrier. A selection of various filters can be interchanged as required.
Field Diaphragm
The field diaphragm is used to realize Koehler il­lumination.
Lamp Housing for Transmitted Light
The lamp housings 107/2 and 107 (both for 12V 100W halogen) are available for the Leica DMI 6000 B. For their descriptions and applications, please refer to the section of this manual de­scribing the use of the microscope. The letter L indicates a lamp housing designed for left-handed operation.
Lamp Housing for Incident Light
The 106 z L lamp housing (for halogen or xenon) is available for the Leica DMI 6000 B. For their descriptions and applications, please refer to the section of this manual describing the use of the microscope. The letter L indicates a lamp housing designed for left-handed operation.
Leica CTR6000 Electronics Box
The Leica CTR6000 electronics box contains the power supply for the lamp and the circuit boards required to control the motorized functions of the stand.
17
4. Overview of the Instrument
18
17
1
14
Fig. 1 Leica DMI 6000 B left view 1 Eyepiece 2 Eyepiece tube 3 Top port 4 Intermediate pupil interface 5 LeicaScreen 6 Light intensity 7 Field diaphragm 8 TL/IL switching 9 Aperture diaphragm 10 Focus wheel
16
2
15
3
4
5
678910111213
11 Variable function buttons 12 Right side port 13 Booster lens 14 Lamp mount 15 Condenser head 16 Condenser base 17 Field diaphragm 18 Transmitted light lamp housing 19 DIC objective prism disk
18
4. Overview of the Instrument
11
4
5
Fig. 2 Leica DMI 6000 B right view 1 E-focus control buttons 2 Focus wheel 3 Variable function buttons 4 Opener for drawer 5 Drawer 6 Right side port
12 3
67812910
7 Analyzer slot 8 Centering window 9 Field diaphragm centering 10 Incident light lamp housing 11 Objective turret 12 Stage with attachable mechanical stage
19
4. Overview of the Instrument
4
3
5
6
2
Fig. 3 Leica DMI 6000 B front view 1 LeicaScreen 2 Front control panel 3 Port switching 4 Top port 5 Manual transmitted light filters 6 Field diaphragm centering
20
1
Fig. 3b SmartMove remote control module 1 Travel in x 2 Travel in y 3 Focus
Fig. 3a Front control panel
Fig. 4 Overall view of Leica DMI 6000 B with SmartMove remote control module
4 Variable function buttons
(preassigned at factory)
4. Overview of the Instrument
3
1
2
4
21
5. Unpacking the Microscope
5. Unpacking the Microscope
The microscope is delivered in several packages.
The stand package contains the following com­ponents:
• Stand with integrated incident light axis, objective turret and tube
• Illumination arm
• Specimen stage
• CD with Leica Application Suite (LAS) software package
• Instructions and list of microscope presets (identification sheet)
The system package contains the microscope's accessories:
• Eyepieces
• Objectives
• Condenser
• Lamp housings with accessories
• Assembly tools
• Additional accessories such as filter cubes, etc. depending on feature set
The Leica CTR6000 electronics box, the Smart­Move remote control module and the ebq 100 supply unit are supplied in separate packages.
22
Please carefully compare the contents of the delivery to the packing slip, delivery note, or in­voice. We urgently recommend storing a copy of these documents with the manual to ensure that you have information on the time and scope of delivery handy for subsequent orders or service work. Please make sure that no small parts remain in the packing material. Parts of our packing material are marked by symbols to simplify recycling.
First, carefully remove all components from the transportation and packaging materials.
Do not put the instrument into operation in the event of visible damage to the compo­nents or packing material.
Caution!
5. Unpacking the Microscope
Caution!
Do not connect the microscope or periph­erals to an AC power source at this time under any circumstances!
Installation Location
Work with the microscope should be performed in a dust-free room, which is free of oil vapors and other chemical vapors, as well as extreme humidity. At the workplace, large temperature fluctuations, direct sunlight, and vibrations should be avoided. These may adversely affect measurements and long-term observations.
Allowable ambient conditions Temperature 15-35°C Relative humidity maximum 80% up to 30°C
Note:
If at all possible, avoid touching the lens surfac­es of the objectives. If fingerprints do appear on the glass surfaces, remove them with a soft leather or linen cloth. Even small traces of finger perspiration can damage the surfaces in a short time. See the chapter "Care of the Microscope"
p. 95, for additional instructions.
Microscopes in warm and warm-damp climatic zones require special care in order to prevent the build up of fungus. See the chapter "Care of the Microscope" p. 95, for additional instructions.
Electrical components must be placed at least 10 cm away from the wall and away from flammable substances.
Caution!
23
5. Unpacking the Microscope
Transport
For shipping or transporting the microscope and its accessory components, the original packag­ing should be used.
As a precaution to prevent damage from vibra­tions, the following components should be dis­assembled and packaged separately:
• Unscrew the objectives.
• Remove the eyepieces.
• Remove the condenser.
• Remove the specimen stage.
• Remove the transmitted light arm.
• Remove the lamp housings.
• Remove the lamp housing mount.
• Disassemble the burner of 106 z lamp housing.
• Remove the filter cube.
• Remove all moving or loose parts.
24
6. Assembling the Microscope
6. Assembly
The microscope components are logically as­sembled in this order:
•Transmitted light illumination carrier
• DIC module and DIC objective prisms*
• Specimen stage
• Condenser with condenser head
• Eyepieces
• Objectives
•Transmitted light lamps
• Lamp housing mount (mirror housings)
• Incident light lamps
• Assembly of incident light turret disk*
• Polarizer and analyzer*
When using intermediate systems and optical accessories, the sequence may vary. In this case, read Chapter "6.15 Optional Accessories" p. 52.
6.1 Assembly Tools
If possible, the microscope should be assem­bled and set up with the assistance of Leica sales or service personnel. A small number of universal screwdrivers which are included in the scope of delivery are re­quired for assembly (Fig. 7).
Fig. 7 Assembly tools 1 Phillips screwdriver*
2
3mm Allen key
3 1.5mm centering key* 4 2mm centering key* 5 3mm hex key* 6 2.5mm hex key* (short type) 7 2.5mm hex key*
* depending on scope of delivery
2
5
1
3
4
6
2
7
25
6. Assembly
6.2 Installation of the Transmitted Light Illumination Carrier (DL)
Wipe the installation surface (8.3) with a dry cloth. Tip the illumination carrier (8.1) back slightly and install it so that the pin (8.2) engages the groove in the support surface (8.4).
Set the DL illumination carrier upright and fasten it with the 4 screws.
When fastening the transmitted light illumina­tion carrier, do not hold it so as to ensure its op­timal alignment with the optical axis. The tilt angle of the illumination carrier can be varied with the knurled screw (9.1) or fixed verti­cally. Connect the electronics cable to one of the sockets, EXT1 - EXT4.
The transmitted light lamp housing for 12V 100W halogen lamps is a separate component. For in­structions on replacing the halogen lamp Ch. 6.10, p. 41.
26
6. Assembly
6.3 Installation of the DIC Module
and DIC Objective Prisms
If your microscope is not equipped with DIC, please continue with Chapter 6.4.
In the Leica DMI 6000 B microscope, the DIC prisms are already installed in the DIC disk be­low the objective turret (Fig. 10b).
Proceed as follows when making changes to the IC prism disk:
• Remove the front cover (Fig. 11) below the
objective revolver after releasing the socket screws (Fig. 10a).
Fig. 10a Removing the front cover
Fig. 11 Front cover, DIC prism disk
• Insert the DIC prism disk (Fig. 10b) in its receptacle and tighten the two socket screws. Note: insert the prism disk with the electronics board facing down.
Replacing Individual IC Prisms:
• Release the two socket screws and remove the prism disk.
• Place the prism against the stop pin (10b.3), place the washer between the screw and the prism, and tighten gently to prevent undue tension. Insert the prism so that its identifying letter, e.g. ID, is facing upward and is legible.
• After installing the prisms, replace the prism disk into its receptacle.
Fig. 10b DIC objective prism turret (coded and motorized) 1 IC objective prism in frame 2 Identification letter (ID) 3 Orientation pin
Fig. 12 IC objective prism 1 Objective prism in frame 2 Screw and washer
1
2
321
27
6. Assembly
6.4 Installation of Specimen Stages
A wide range of specimen stages are available. The most important are the following:
• Fixed stage (248mm x 204mm): normal, heating and temperature-controlled
• Fixed micromanipulation stage (248mm x 204/ 112mm): normal, heating, temperature­controlled
• Regular manual 3-plate cross-stage, position­ing range: 83mm x 127mm
• Manual micromanipulation 3-plate cross-stage positioning range: 40mm x 40mm
• Motorized micromanipulation 3-plate cross­stage positioning range: 40mm x 40mm
• Manual rotating stage
• Scanning stage IM 120 x 100 (motors on top)
• Scanning stage IM 120 x 100 (motors on bottom)
Fig. 14 Mechanical 3-plate stage
Fig. 15 Micromanipulation stage with attachable mechan-
ical stage
Fig. 13 Fixed stage (normal)
28
Fig. 16 3-plate micromanipulation stage
6. Assembly
The assembly of these stages is identical. The stages are solidly attached to the microscope by three screws. In the case of fixed stages, an at­tachable mechanical stage may be installed (Fig. 18). These are supplied in a separate pack­age.
Multiple-plate stages are supplied separately. Like the fixed stages, these stages are mounted as follows:
• If the screws for the stage are already in the stand, remove them first. In most cases, the screws will be found in the packing material of the stand.
Caution!
!
The screw lengths may vary. When using screws of different lengths, use the shorter of the three screws in the front hole and the equally long ones in the rear holes.
• Use a clean cloth to remove dust and packing material residue from the stand’s contact surface with the stage.
• Align the stage so that the pair of holes faces back toward the illumination axis and the single hole faces forward toward the tube.
• Align the mounting holes in the stage with the holes in the support surface. If the holes are covered, in the case of 3-plate cross-stages or scanning stages, please shift the upper stage plate until the opening becomes visible.
• First, tighten the single front screw with the included 3mm hex screwdriver. Be sure to use
shortest of the three screws in the front
the hole, as an excessively long screw can interfere with the focus travel. (If you have a rotating stage, please continue reading under "Rotating Stage and Insert Frame for Cover­slips").
Fig. 17 Fixed micromanipulation stage Fig. 18 Attachable mechanical stage for fixed microman-
ipulation stage
29
6. Assembly
• Next, firmly tighten the two rear screws.
• Finally, give the front screw a final firm tightening.
Fixed Stage
Attachable mechanical stages that are designed to accept a variety of culture dishes are also available for fixed stages. These mechanical stages may be attached to either side of the fixed stage (Fig. 17).
Two screws are located at the underside (right or left) of the attachable mechanical stage. Tighten these screws in the threaded holes on the underside of the fixed stage with the 3mm hex screwdriver. Retighten these screws from time to time after frequent use.
The attachable mechanical stage has been pre­adjusted in the factory. In the event that the at­tachable mechanical stage runs out of focus when moving from right to left, this can be cor­rected by Leica’s technical service.
Next, remove one or more of the ordered insert frames (Fig. 20) from their packaging and place the insert frame into the precise retention sys­tem. The stage, the attachable mechanical stage, and the insert frame are now ready for use.
Some (not all) inserts are provided with self-ad­hesive scales to allow the coordinates to be read.
Apply these scales to the recesses of the at­tachable mechanical stage.
Fig. 20 a, b, c
Inserts for attachable mechanical stage (micromanipulation stage)
a
Fig. 19 a, b
Inserts for attachable mechanical stage (fixed stage)
30
a
b
b
c
Manual Fixed Micromanipulation Stage
To install the attachable mechanical stage for the manual fixed micromanipulation stage (Fig. 24), proceed as you would for the normal attachable mechanical stage.
The insert frames (Fig. 20a to c) differ at this point. These are held by two screws on the at­tachable mechanical stage and changed by re­leasing the screws.
Fig. 21
Inserts for fixed stages
6. Assembly
Fig. 24 Installation of attachable mechanical stage
Fig. 25 Installation of attachable mechanical stage
Fig. 22
Glass insert for 3-plate cross-stage and scanning stage
Fig. 23
Heating insert P
31
6. Assembly
Motorized 3-plate or Scanning Stages
3-plate stage: After installing the stage, connect the included stage cable to the socket on the stage, then with the CRT6000 box. The correct place on the box is called "XY Stage".
Scanning stage: After installing the stage, con­nect the included stage cable first to the X and Y sockets of the stage and the DM STC control unit for the scanning stage: XY Stage
Next, connect the DM STC control unit to the CRT6000 box. Connect the Y-cable as follows: In­sert the small round plug into the COAX CTRL socket and the second flat plug in the XYZ-Con- trol socket of the CRT6000 box. Connect the plug with the two cables to the SmartMove.
A variety of inserts (including heating ones) are available for the normal 3-plate and scanning stag­es. Install these inserts diagonally from above into the corner with the spring clips. The insert will click into place when installed properly.
Caution:
!
Press the spring clip into place only from the side.
Do not press the insert onto the spring clips di­agonally from above, as the insert will not be aligned parallel to the stage and may be bent in the process.
32
6. Assembly
Rotating Stage and Insert Frames for Coverslips
The rotating stage (Fig. 30) is also mounted with 3 screws (30.2). Rotate the stage to make all of the threaded holes accessible. Insert the screws (30.2).
Caution:
!
Use additional washers (30.3) for the rear holes. Tighten the screws only lightly, as the rotating stage must be centered first: Insert the adjusting aid into the rotating stage for this purpose. Acti­vate the Bertrand lens and focus, or use a fo­cusing telescope (Fig. 32). Move the stage until the bright circle is in the middle of the field of view. Next, tighten the stage, swing the Ber­trand lens out and remove the adjusting aid.
To insert glass slides in insert frames (31.1), press on the center of the leaf spring (31.2) and insert the coverslip in the direction of the arrow. Clamp the insert frame in the attachable me­chanical stage (30.1).
6.5 Installation of Condensers
Fig. 30 Rotating stage 1 Attachable mechanical stage 2 Screws for stage mounting 3 Washers
2
1
Fig. 31 1 Insert frame for coverslips 2 Leaf springs
3
2
1
2
Fig. 29 a, b Mounting screws for 3-plate cross-stage
ab
Fig. 32 Focusing telescope
33
6. Assembly
All the Leica DMI 6000 B microscope condens­ers feature 7-position turrets and can be equipped individually with suitable ring dia­phragms for phase contrast (PH), darkfield (DF) or IC prisms for DL interference contrast (DIC). Light rings and condenser prisms are generally already installed in the turret at the factory, making the following assembly steps unneces­sary. Please continue on
page 37, Installation
of Condensers.
To change the components, proceed as follows:
Installing the Light Rings
Change the light rings only with the instrument’s power turned off. To change the light rings, start by removing the complete condenser from the transmitted light illumination axis. Simply re­lease the socket-head screw at the right side of the condenser mount. The condenser can now be removed very easily with Leica’s own con­denser quick release. Open the condenser cov-
er at the top right side. You will now have access to the various numbered openings for the light rings.
Install light rings for phase contrast (designated by the ID numbers 0, 1, 2, 3 and the focal inter­cept S of the corresponding condenser head) and the DF diaphragm (designated by D for dark­field and the focal intercept S of the correspond­ing condenser head) in the positions of the turret disk as follows:
Fig. 33 Condenser base S1-S28
34
Fig. 34
Condenser head S1
Fig. 35
Condenser head S28
6. Assembly
• Select a position and make sure that the two mounting screws have been released to the point that they no longer extend into the position. To adjust the screws, turn the desired light ring position into the beam path. You can now turn the screws using the two adjusting keys.
• Next, take the special condenser tool (Fig. 38).
• If possible, install the light rings 0 to 3 in ascending order. The numbering of the openings is located at the edge of the crown gear (4 large openings: 1-4; 3 small openings: 5-7).
• Grasp the light ring to be installed with the condenser tool (the lettering must face upward and be legible) so that the groove of the tool just reaches over the edge of the light ring and the upper edge of the light ring is lying flat in the holder of the tool. Press the cheeks of the tool to grasp the light ring.
• Place the light ring in the desired opening and make a note of the opening and light ring designation for the subsequent configuration of the Leica Application Suite (LAS).
• Insert the light ring at a slight angle from above so that the frame slides under the sp­ring clip of the receptacle. Do not press the spring clip down under any circumstances. This can destroy the clip or result in an unstable position of the light ring.
• Once the light ring is properly positioned, release the tool and install the remaining light rings in the same way.
• Use the front openings to coarsely center the light rings; the screws must not extend beyond the outer edge of the disk under any circumstances.
• Perform the fine adjustment with the Bertrand lens or telescope after switching the unit on.
Fig. 36 Phase contrast rings
Fig. 37 Condenser prisms
35
6. Assembly
Please continue reading if you also have to in­stall IC prisms. Otherwise, skip to the next sec­tion.
Installation of IC prisms
The IC prisms are installed at the factory. To change the components, proceed as follows:
Change the IC prisms only with the instrument’s power turned off. To change the IC prisms, start by removing the complete condenser from the transmitted light illumination axis. Simply re­lease the socket-head screw at the right side of the condenser mount. The condenser can now be removed very easily with Leica’s own con­denser quick release. Open the condenser cov­er at the top right side. You will now have access to the various numbered openings for the IC prisms.
• IC prisms can only be installed in the large positions of the condenser turret with guide grooves.
• Select a position and make sure that the two mounting screws have been released to the point that they no longer extend into the position. It is advisable to remove one of the screws when installing a prism, as a single screw is sufficient for centering. The opposing pressure resulting from the use of two screws may destroy the prism.
• Next, take the special condenser tool (Fig. 38).
• If possible, install the prisms 0 to 3 in ascending order. The numbering of the openings is located at the edge of the crown gear.
• Grasp the prism to be installed with the condenser tool (the lettering must face upward and be legible) so that the groove of the tool just reaches over the edge of the prism and the upper edge of the prism is lying flat in the holder of the tool. The numbers K2 to K16 should be positioned toward the end of the tool. Press the cheeks of the tool to grasp the light prism.
Fig. 38 Condenser tool Fig. 39
Inserting the condenser prisms and phase rings
36
6. Assembly
• Place the prism in the desired opening and make a note of the opening and prism designation for the subsequent configuration of the Leica Application Suite (LAS).
• Insert the prism at a slight angle from above so that the frame slides under the spring clip of the receptacle. Do not press the spring clip down under any circumstances. This can destroy the clip or result in an unstable position of the prism. Two guide hooks are located on the underside of the prisms. These must fit into the two grooves of the opening. Only one of the two possible alignments is correct.
• The ID, e.g. K10, must be visible when the prism is installed and must be oriented toward the center of the condenser. The prism will also fit with the ID facing outward, but DIC imaging will not be possible in this position.
• Use the adjusting key to tighten the centering screws to the point that they no longer extend over the edge of the disk. Use only the left centering screw to position the prism (see ICT operation). The right centering screw must not restrict the range of adjustment under any circumstances.
• Use the front openings to coarsely center the prisms; the screws must not extend beyond the outer edge of the disk under any circumstances.
• Perform the fine adjustment with the Bertrand lens or telescope after switching the unit on.
Installation of Condensers
The installation procedure is identical for all condensers S1 to S70 (motorized or manual/cod­ed).
Release the socket-head screw at the right side of the condenser holder. Place the condenser on the retaining pins of the illumination arm and move the condenser to the correct height. Use the markings on the column and condenser to determine the correct position.
Once you have reached the correct position, tighten the socket-head screw.
Fig. 40
Installation of condenser on transmitted light illumination arm
• If necessary, carefully remove dust or fingerprints from the prisms.
• Once the prism is properly positioned, release the tool and install the remaining prisms in the same way.
37
6. Assembly
Condenser Heads
Four different condenser heads are available:
1) S1/1.40 oil
2) S1/0.90 dry
3) S23/0.53
4) S28/0.55
Condensers 3 and 4 are screwed directly into the condenser body. A spacer ring (42.2) must be screwed into the thread at the bottom of the condenser body prior to installing condensers 1 and 2. The S1 condenser heads will fit into this ring.
The S70 condenser is delivered complete with a condenser head, which makes additional as­sembly unnecessary.
Fig. 41 Condenser on transmitted light illumination arm
Fig. 42 Installation of condenser heads S1 1 Condenser base 2 Spacer ring 3 Condenser head
38
Fig. 43 Installation of condenser head S28
1
2
3
6. Assembly
6.6 Installation of Eyepieces
The eyepieces are inserted into the eyepiece tubes.
Note:
We recommend running a teach-in via the Leica Application Suite (LAS) software when using eyepieces not included in the scope of delivery. This will ensure that the total magnification shown in the LeicaScreen is correct.
Fig. 44 Eyepieces
6.7 Installation of Objectives
The positions in the objective turret disk are num­bered (Fig. 45). Depending on your equipment, the individual objectives have already been assigned to specific positions at the factory. For details on the exact positions of the objec­tives, please refer to the enclosed identifica­tion sheet.
Caution:
!
Close vacant threads in the nosepiece with dust protection caps!
Please note that the front lenses of the objec­tives point upward and are therefore more vul­nerable to contamination than those of upright microscopes. Check the front lenses for cleanliness frequently.
Note:
We recommend running a parfocality compen­sation via the Leica Application Suite (LAS) soft­ware.
Fig. 45a Objective turret Fig. 45b Objective turret, loaded
39
6. Assembly
6.8 Installation of Filters in the Illumination Arm
The Leica DMI 6000 B is generally equipped with a holder for two 40mm diameter filters. The fil­ters are installed at the factory. To change filters yourself, proceed as follows:
Release the screw (46.1) and remove the cover. Place the filter in the holder. Place the cover on the transmitted light illumination carrier and fasten with the locking screw.
• Mark the lever with the provided adhesive labels.
Fig. 46 Unscrewing the filter holder cover and inserting
filters in the transmitted light illumination arm
1 Screw
1
6.9 Installing the Transmitted Light Lamp Housing
• Place the lamp housing in the transmitted light
lamp housing mount (fig. 47) and fasten it with the clamping screw on the side.
• Thread the cable through the transmitted light
illumination arm (Fig. 48).
• Connect the lamp housing cable to the power
supply for transmitted light on the Leica CTR6000 electronics box (Fig. 49.1).
For instructions on changing the lamp, please see Chapter 6.10.
These instructions also apply to installing an Hg lamp on the transmitted light axis. For descriptions of the lamp housings and replacement of the burner, please see Chapter 6.12,
Fig. 47 Mounting the lamp housing on the
transmitted light illumination arm
p. 44ff.
Fig. 48 Lamp housing cabling (cable duct)
40
Fig. 49 Connecting the lamp housing to the
Leica CTR6000 electronics box
1
6. Assembly
6.10 Installation and Replacement of the Trans­mitted-Light Lamps: 107 or 107/2 Lamp Housing
This lamp housing is used with a 12V 100W halo­gen lamp, which is already mounted. In case the lamp has to be removed:
Changing the 12V 100W Halogen Lamp
Caution!
Make sure that the lamp housing has been disconnected from the power supply. Unplug the power plug and the power supply during assembly.
Caution!
Light sources pose a potential irradiation risk (glare, UV-radiation, IR-radiation). Therefore, lamps have to be operated in closed housings.
• Lift the housing off (Fig. 50b).
• Remove the lamp.
Caution!
Do not remove the new lamp’s dust cover until you have installed the lamp. Avoid fin­gerprints on the lamp.
• Insert the new 12V 100W lamp (Fig. 51) with the dust cover straight into the socket until it stops. Be sure that the lamp is inserted straight.
• Remove the lamp’s dust cover.
• Replace the housing and fasten it in place using the fastening screw.
Fig. 50b
Removing housing
• Remove the fastener screw on the housing (Fig. 50a).
Fig. 50a
Lamp housing 107/2 Releasing the fastening screw
Fig. 50c
Lamp housing 107/2 opened 1 Mount with
halogen lamp
2 Collector
1
2
41
6. Assembly
Fig. 51
Inserting lamp with cover
a Right b Wrong
6.11 Installation of Lamp Housing Mount and Mirror Housing
Place lamp housing mount (Fig. 53) or mirror housing on rear wall. Mount from front with socket-head screws.
a
b
Fig. 53 Lamp housing mount
Fig. 52 Leica DMI 6000 B rear panel 1 Installation point for lamp housing mount
or mirror housing
2 Holes for lamp housing mount or mirror housing screws
2 2
1
42
Fig. 54 Lamp housing 106z 1 Collector adjustment 2 Vertical lamp adjustment 3 Horizontal lamp adjustment 4 Adapter ring
4
2
3 1
If a booster lens is included in the scope of de­livery, insert it into the rear stand opening at the left or right, depending on the stand model.
The booster slide has several positions:
1. Slide pulled out: no effect
2. Filter position: wavelength of the installed filter acti­vated
3. Depending on orientation of slide:
a) symbol visible:
• center orientation The intensity of the fluorescence is increased by 50% in the center of the field of view (approx. 30% of the field).
b) symbol
visible: The overall intensity is reduced by 25%. The entire field of view is evenly illuminated, however.
6. Assembly
Fig. 56 Booster lens in stand 1 Booster lens
1
Fig. 55 Booster lens
Fig. 57 Hg-mercury burner
43
6. Assembly
6.12 Installation and Replacement of Incident Light Lamps
Caution!
Light sources pose a potential irradiation risk (glare, UV-radiation, IR-radiation). Therefore, lamps have to be operated in closed housings.
Make sure that the lamp housing has been disconnected from the power supply. Unplug the power plug and the power supply during assembly.
During assembly work on xenon burners, al­ways wear the supplied protective gloves and face protection (Fig. 58) (risk of explosion).
Never touch the glass parts of the burner with bare hands. Never look directly into the beam path (blinding hazard).
Lamp Housing 106 z
This lamp housing is suitable for use with a 12V 100W halogen lamp or a variety of gas discharge lamps.
Caution!
Make sure to follow the instructions and safety notes of the lamp supplier. Before changing lamps allow at least 30 min­utes for cooling down!
Fig. 58
Protective gloves and mask
44
Fig. 59 Lamp housing 106 z L with Hg 100W lamp 1 Collector focusing 2 Vertical lamp adjustment 3 Horizontal lamp adjustment 4 Hg lamp mount 5 Reflector adjustment (not visible)
2
5
3 1
4
6. Assembly
Inserting Gas Discharge Lamps (Hg and Xe) in the 106z Lamp Housing
Hg and Xe lamps are powered by separate sup­ply units. Please also read the separate instruction manu­al provided with these supply units.
The following gas discharge lamps may be used and require different supply units and lamp mounts (Fig. 60, 61):
Type Typical Bulb Life*
100W high-pressure mercury burner (direct current) 200 hours 100W high-pressure mercury burner (direct current, type 103 W/2) 300 hours 75W high-pressure xenon burner (direct current) 400 hours
* Please observe the data sheets of the lamp manufacturer.
Fig. 60 Lamp mounts for Hg 100 gas discharge lamp 1 Upper clamping system 2 Lower clamping system 3 Cooling element
Hg 100
1
2
Fig. 61 Lamp mounts for gas discharge lamp Xe 75 1 Upper clamping system 2 Lower clamping system 3 Cooling element 4 Protective cover of Xe 75 burner
a
3
Xe 75
b
3
1
4
2
45
6. Assembly
Caution!
Make sure to follow the safety notes on page 44.
•To open the 106 z lamp housing, unscrew the
fastening screws (63.8) on the cover. Loosen the contact plug somewhat and pull it out of the socket (63.9). Flip the cover up (63.1).
• Loosen the mounting screws (63.8) on the
lamp socket and pull the socket out.
• Remove the transport anchorage (red plastic
rod in place of the burner) in the lamp mount. To do so, remove the lower clamp (60.1, 61.1). Pull up the cooling element (61.3, 60.3) and turn it to the side. Detach the lower clamp system (61.2, 60.2) and remove the transport anchorage.
Caution!
Do not remove the burner’s dust cover until you have installed the lamp. Avoid fingerprints on the lamp. Sweat from your fingers on the glass will shorten the life of the lamp significantly.
• Install the burner in reverse order.
Caution!
Xe 75 burner: Remove the burner’s dust cover (61.4) after you have installed the burner.
Fig. 63 106 z lamp housing (on the side, open) 1 Cover raised 2 Collector 3 12V 100W lamp or
gas discharge lamp in mount
4 Reflector (mirror) 5, 6, 7 Adjusting screw for x-y reflector 8 Locking screws for lamp mount 9 Socket for contact plug
Fig. 62 Rear panel of ebq 100 supply unit 1 Lamp connection
1
46
1
2
4
5
3
6
7
898
• Insert the lamp mount, with the burner
installed, into the lamp housing and tighten it with the screws (63.8).
•Test the adjustment of the collector (63.2):
Do not touch the power supply while performing these actions. When closing the lamp housing, make sure that the pins of the contact plug engage in their sockets (63.9). Tighten the screws of the cover and press the contact plug home.
• Place the lamp housing in the incident light
lamp housing mount (fig. 53) and fasten it with the clamping screw on the side.
• Connect the lamp housing to the external
power supply (62.1).
6. Assembly
Caution!
The burner must be adjusted immediately after lighting.
47
6. Assembly
6.13 Equipping the Incident Light Turret Disk
The fluorescence drawer is located on the right side of the stand. Before opening this drawer, remove the cap below the drawer covering the analyzer slot. Remove the analyzer if it is already in the slot.
The replacement of individual cubes is more convenient with the microscope switched on. The position to be changed then automatically turns to the outside and you can be sure that the cube is positioned in the correct holder. You can therefore postpone installing the filter cubes un­til after the microscope has been switched on.
You can also insert the filter cubes while the in­strument is switched off. Press the white button next to the drawer. The drawer will glide out into its initial position. This is the only position in which the inner disk for the cubes can be turned to a free or desired position.
Fig. 66 Opening the fluorescence drawer
Fig. 67 Open fluorescence drawer
The positions in the turret disk are numbered. Depending on your equipment, the individual fil­ter and reflector cubes have already been as­signed to specific positions at the factory. For details, check the identification sheet included with your order.
Fig. 64 Filter cube,
front side
Fig. 65 Filter cube,
back side
48
Fig. 68 Inserting or removing a filter cube
6. Assembly
Now open the drawer several mm further until it clicks into its end position. The disk will no long­er turn in this position.
You can now insert a filter block. Proceed as fol­lows:
•With the holder facing you squarely, insert the
filter or reflector cubes into the holder in accordance with the included identification sheet.
• The fluorescence cubes are suitable for both
upright and inverted microscopes. When using them with inverted microscopes, insert them so that the writing is upside down along the lower edge.
To do so, place the filter or reflector cube on the left side and press it to the right into the mounting (Fig. 68).
• Make sure that the cube is correctly seated. A
loose cube can be destroyed or block the disk.
Replacing cubes with the instrument switched on:
• Remove the analyzer or the cap of the analyzer slot.
• Press and hold the SET button on the right side of the stand and at the same time, press the button on the front panel for the cube you would like to insert or replace.
• The filter changer will then rotate to the correct position to insert or replace the cube when you open the drawer by pressing the white button on the right side of the stand. The following message will appear in the top line of the LeicaScreen.
Load.
To insert the cubes, proceed exactly as de­scribed above.
• For the next cube, close the drawer to the point that the disk is once again free to turn. Once you have reached the next position, open the drawer fully once again. Continue in this way for all of the cubes.
• Once all filter and reflector cubes have been inserted, close the drawer and replace the analyzer or cap.
49
6. Assembly
6.14 Installation of the Polarizer and Analyzer
Installed at the factory. To change the components, proceed as follows:
Manual condenser: Attach the single or triple position holder to the top of the manual con­denser. The holder has a guide that must be in­serted in the opening next to the screw threads. The holder must be positioned so that the polar­izer or filter to be used covers the opening of the condenser. Insert the polarizer or filter with the correct side facing up into the holder (λ: lambda and polariz­er; POL: polarizer only). A click mechanism will indicate proper seating. The polarizer must turn easily between the two stops (approx. 30°).
Fig. 70 Mechanical polarizer holder 1 Mechanical polarizer 2 Mechanical analyzer
1
50
Fig. 71 Condenser with motorized polarizer
2
Analyzer for Incident Light and Transmitted Light.
• Remove the cap (Fig. 72) on the right side of the stand (under the fluorescence drawer).
• Insert the analyzer into the receptacle until it latches in place (Fig. 73.1).
Fig. 72 Analyzer slot cap
6. Assembly
Fig. 73 Inserting the analyzer 1 Slot 2 Analyzer
Fig. 74 Inserting the analyzer
1
2
51
6. Assembly
Fig. 75 C-Mount 0.63x
6.15 Optional Accessories Camera
Connecting a camera A camera can be installed using a C-mount or Vario mount.
• Place the C-mount or Vario mount onto one of the camera ports and secure it with the locking screw at the side.
• Screw on the camera.
Note:
When using a C-mount or Vario mount, run a teach-in via the Leica Application Suite (LAS) software.
Connecting multiple cameras Two or more cameras – for example a digital and an analog camera – can be adapted as required.
Fig. 76 C-mount 0.5x
52
• When using a DC type camera, connect the camera to the PCI card of your PC.
• When using a DFC type camera, connect the camera to the FireWire card of your PC.
Note:
Please read the separate operating manual of your digital camera.
6.16 Connection to the Electronics Box CTR6000
Note:
The CTR6000 electronics box must not be used with other stands. The serial number of the associ­ated stand has been recorded on the back of the electronics box.
• Connect the Microscope (77.6) socket to the back of the stand (78.5) using the 25-pin microscope cable.
• Connect the SmartMove remote control module to the XYZ-Control socket (77.5).
6. Assembly
• Connect the motorized stage, if present, to the XY-Stage socket (77.2).
• Connect the lamp power cable (78.7) to the 12V,
max 100W socket (77.7).
Caution!
Make sure that the plugs are correctly in­serted and secured to prevent overheating of the sockets.
Fig. 77 Rear view of Leica CTR6000 1 AC power socket 2 XY Stage socket for motorized stage 3 Direct interface socket optional 4Z Control for separate focus control 5 XYZ Control for SmartMove 6 Microscope socket for microscope 7 12V, max 100W for the lamp power cable of stand 8 DL: reset button
8
5
4 3
6
2
1
7
Fig. 78 Rear view of stand 1 RS232 ports 2 2 x USB 3 4 x EXT. 4 XYZ control for SmartMove 5 Connection for Leica CTR6000 electronics box 6 Condenser cable 7 Lamp power cable
7 6
1 2
5 4
3
53
6. Assembly
6.17 Connection to the Computer
Note:
To start the Leica Application Suite (LAS), make sure that the COM1 serial port is not in use by another program or driver. This is frequently the case when using Palms or other PDAs or when using external modems or other devices. The devices in question must therefore always be disabled before using the Leica Application Suite (LAS) software.
• Please use the included serial cable. Connect the COM1 port of your PC with the RS232C port (78.1) on the back of the stand.
6.18 Connection to the Power Supply
• Once all installation work is complete, connect the Leica CTR6000 electronics box to an AC power outlet with the included power cable (socket 77.1).
• If you are using the external ebq 100 supply unit, connect it to an AC power outlet at this time (socket 79.1).
54
Fig. 79 Rear panel of ebq 100 supply unit 1 AC power socket
1
7. Start-up
7. Start-up
7.1 Functional Principle
Because of its intelligent automation, the Leica DMI6000 B can be controlled using a variety of con­trol elements.
1. Intelligent Automation
• Switching between contrast methods at the touch of a button. Light rings, DIC prisms, etc. are automatically positioned in the beam path.
• The microscope recognizes the selected objective and associated contrast method. The intensity (INT), aperture diaphragm (AP), and field diaphragm (FD) are always set to suitable values.
• The INT, AP, and FD values are always based on the currently activated illumination axis (transmitted light or incident light).
• The INT, AP, and FD values can be adjusted individually. Manual adjustments overwrite the previous settings. The current setting is stored and is retained from one session to the next when power is switched off.
2. Controls
• SmartMove knobs for stage and focus control.
• Fixed function buttons on stand for INT, AP, and FD, as well as for switching between transmitted light and incident light axis.
•Variable function buttons on stand and SmartMove These function buttons have functions suitable to the configuration of your microscope as­signed to them at the factory. The functions can be reprogrammed and/or adapted to your specific requirements, however.
• Complete control of microscope and camera via software (Leica Application Suite (LAS)
55
7. Start-up
Note: (Reset Function)
The microscope can be reset to its factory de­fault programming:
• With the stand switched off, press the top three variable function buttons on the left side of the stand.
• Switch on the power for the stand.
• Hold the buttons until the initialization is complete.
• The standard information display will now appear in the LeicaScreen (Fig. 81 and 82, p. 59).
• Switch the instrument off and back on. The settings are now saved.
The table on the following page provides an overview of the microscope functions and their controls.
56
7. Start-up
Function Fixed Variable SmartMove Software
Function Function Function Rotary Buttons Buttons Buttons Knobs Stand Stand
Select contrast method -++-+
Change transmitted light/incident light axis +---+
Change to objective -++-+
Teach-in parfocality ----+ Change operating mode (dry/imm) -++-+
Illumination Manager +++-+
Magnification Changer +---+
Focusing +--+1)+
Set stops +---+ Go to stop +---+ Change step increment (coarse/fine) -++-+
XY stage positioning ---++
Change speed ----+ Stage positions (store/go to) ----+
Change to filter/reflector cube + (+) + - +
Side and bottom ports + (+) + - +
DIC fine adjustment +---+
+ always possible (+) optional
- not possible
1)
Focusing alternatively via wheels
57
7. Start-up
Possible Assignments for Variable Function Buttons on Stand and SmartMove For Leica DMI6000 B: Function Button Function BF Brightfield transmitted light
PH Phase contrast transmitted light ICT Interference contrast transmitted light DF Darkfield transmitted light POL Polarization transmitted light CHANGE TL Cycle through all contrast methods
INT Increase intensity (transmitted light) INT Reduce intensity (transmitted light) AP Open aperture diaphragm (transmitted light) AP Close aperture diaphragm (transmitted light) FD Open field diaphragm (transmitted light) FD Close field diaphragm (transmitted light) SHUTTER TL Open/close TL shutter
FLUO Fluorescence (last filter cube) CUBE 1-6 Select filter cube in position 1-6 CHANGE CUBE CW Change cube clockwise (1 4) CHANGE CUBE CCW Change cube counterclockwise (4 1)
INT FLUO Increase intensity (fluorescence) INT FLUO Reduce intensity (fluorescence) FD FLUO Open field diaphragm (fluorescence) FD FLUO Close field diaphragm (fluorescence)
CHG FW Toggle filter functions IFW Activate external filter wheel ExMan Activate Excitation Manager
COMBI Combination method (PH fluorescence or ICT fluorescence) CHANGE COMBI Cycle through all combination methods
CHANGE OBJ CW Cycle through objectives clockwise CHANGE OBJ CCW Cycle through objectives counterclockwise Z FINE Activate fine focus Z COARSE Activate coarse focus XY PRECISE Activate precise stage XY FAST Activate fast stage BTP on/off Bottom port on/off DRY/IMM Switch dry/immersion CHANGE FLT Switch TL filter CHANGE CS Switch to confocal application
58
7. Start-up
7.2 Switching on the Microscope
• Switch on the power of the CTR6000 electronics box at the On/Off switch (80.1). The signal lamp (80.2) is lit green when the unit is ready. All motorized microscope components will then run through an initialization phase.
Note:
If a PC is connected, switch on the electronics box first, and then the computer.
After the initialization (Fig. 81) is complete, the LeicaScreen will display the microscope’s cur­rent settings (Fig. 82).
If a component has not been installed correctly, the LeicaScreen will display an error message. See Troubleshooting chapter, p. 91.
Components such as diaphragms, condensers, light and phase contrast rings have been pre­centered at the factory. It may be necessary to correct the centering after the microscope has been transported and assembled. Before performing the required steps, please fa­miliarize yourself with the LeicaScreen and the controls.
Caution!
After turning on the gas discharge lamp, the burner must be immediately adjusted. There­fore, do not turn on the power supply unit yet. First, work in transmitted light in order to familiarize yourself with the microscope’s controls.
Fig. 80
Front side Leica CTR6000
1 On/Off switch 2 Signal lamp
Fig. 81
LeicaScreen Initialization
Fig. 82
LeicaScreen after initialization
2 1
59
7. Start-up
7.3 The LeicaScreen
The screen displays the microscope’s current settings. The content of the display depends on the features of the individual microscope. For information on the abbreviations used, please turn to the table of abbreviations
p. 99.
The screen has a number of areas and lines.
Line 1: Contrast method Line 2: Objective/magnification Line 3: Illumination/diaphragms Line 4: Active ports Line 5: Focus/stops
The content of the display changes according to the active function.
Fig. 83 Arrangement of the function buttons — overview 1 Four variable function buttons 2 Illumination Manager 3 Front control panel 4 Focus control buttons 5 Three variable function buttons 6 SmartMove knobs 7 SmartMove function buttons
Pictograms
Contrast Method
Objective/ Magnification
Illumination Diaphragm
Ports/Eyepiece
Focus/Stops
60
6
6
7
543321
7
7.4 The Function Buttons on the Stand
7. Start-up
A number of function buttons are located on both sides of the stand. These can be broken down into fixed and variable buttons. The vari­able function buttons have different functions depending on the features of the individual mi­croscope.
Fixed Function Buttons on the Left Side
The TL/IL button (84.1) toggles between the inci­dent light and transmitted light axis. The contrast method last used with a given axis is restored when switching. The INT buttons (84.3) adjust the light intensity. The adjustment can be made in coarse or fine steps. Pressing both INT buttons at the same time toggles between coarse and fine adjust­ment. "Intensity fine" will appear in the display when fine adjustment is selected. The AP buttons (84.2) for the aperture dia- phragm and FD (84.4) for the field diaphragm open and close their respective diaphragms.
Note:
Changes to the light intensity as well as aperture and field diaphragm settings are stored for the individual objectives and contrast methods.
Variable Ffunction Buttons on the Stand
84 Fixed function buttons (left side of stand) 1 Toggle transmitted light/incident light 2 Aperture diaphragm 3 light intensity 4 field diaphragm
3
2
4
1
61
7. Start-up
The variable function buttons are assigned func­tions at the factory that are appropriate to the features of your microscope. They are labeled accordingly. For details on button assignments, please refer to the included identification sheet. For information on the abbreviations used, please refer to the list p. 58.
Note:
The Leica Application Suite (LAS) software is required for changing the button assignments.
Function Buttons on the Ffront Panel (Fig. 87)
Possible Functions*:
BF PH ICT DF POL CHANGE TL INT INT AP AP FD FD SHUTTER TL FLUO CUBE 1 CUBE 2 CUBE 3 CUBE 4 CUBE 5 CUBE 6
CHANGE CUBE CW CHANGE CUBE CCW INT FLUO INT FLUO FD FLUO FD FLUO CHG FW IFW ExMan COMBI CHANGE COMBI
CHANGE OBJ CW CHANGE OBJ CCW Z FINE Z COARSE XY PRECISE XY FAST BTP ON/OFF DRY/IMM CHANGE FLT CHANGE CS
Fig. 85 Function buttons (left side of stand) 1 Variable function buttons 2 Open/close aperture diaphragm 3 TL/IL switching 4 Open/close field diaphragm 5 Increase/decrease light intensity
54321
Fig. 86 Function buttons (left side of stand) 1 Variable function buttons
* See page 58 for abbreviations
62
1
7. Start-up
100% of the light goes to the eye­piece (87.1).
Toggle function for the side ports
(87.2). This function depends on the
individual microscope configuration. Note: Switching to the bottom port: via the variable function buttons; switching to the top port: manual.
SHUTTER Opens and closes the shutter (87.3).
Switches between the possible mag-
nifications of the magnification
changer (87.4).
The magnification changer is set to the
1x
magnification 1x (87.5).
CUBE The CUBE 1 to CUBE 6 (87.6) buttons
permit the direct selection of individ­ual filter cubes, provided the select­ed cube is valid for the selected method. Press the CUBE 3 and CUBE 4 but­tons at the same time to display the assignments of the variable function buttons. To reset the display, press the buttons again or wait 3 seconds.
Focus Buttons (Fig. 88) Z
Z
SET + Z
SET + Z
Moves the Z drive in the indicated di­rection.
Sets the upper focus stop.
Sets the lower stop.
Fig. 87 Front control panel 1 100% light to eyepiece 2 Toggle ports 3 Shutter 4 Switch magnifications 5 Magnification 1x 6 Selecting filter cubes
4
2 5
13
66
Fig. 88 1 Focus control buttons 2 Open filter drawer
2
1
63
7. Start-up
SET + Cube Button 1-6
The selected cube is moved to the loading position for replacement. "Load" appears on the screen. Press the opening button (88.2) to open the drawer and position the appropriate cube. The filter cube will remain in this position after the drawer is closed.
7.5 The SmartMove Remote Control Module
SmartMove Knobs
Use the knobs 89.1 and 89.2 to move the stage in X and Y directions. Knob 89.3 focuses the image.
The height of the knobs can be adjusted to a comfortable working position by turning 89.4.
Variable Function Buttons on SmartMove
The variable function buttons are assigned func­tions at the factory that are appropriate to the features of your microscope. They are labeled accordingly. For details on button assignments, please refer to the included identification sheet. For information on the abbreviations used, please refer to the list p. 58.
Fig. 89 SmartMove remote control module 1 Travel in x 2 Ttravel in y 3 Focus 4 Individual adjustment of button height 5 Variable function buttons (factory preset)
Note:
The Leica Application Suite (LAS) software is required for changing the button assignments.
7.6 Illumination
64
1
3
4
2
5
7. Start-up
7.6.1 Transmitted Light
If your microscope has not yet been set up for Koehler illumination, please continue with the "Koehler Illumination" section.
• Select an objective with moderate magnification (10x-20x).
• Activate the transmitted light axis with the TL/ IL button (84.1).
• Press the BF button to activate the brightfield contrast method (one of the variable function buttons on the stand).
• Place a specimen on the stage.
• Focus the specimen using the SmartMove or the focus wheels.
• Adjust the light intensity with the INT buttons (84.3).
• Close the field diaphragm with the FD button (84.4) or manually until the edge of the diaphragm appears in the field of view.
• Using the condenser height adjuster (90.2), adjust the condenser until the edge of the field diaphragm appears in sharp relief.
• Open the field diaphragm just enough for it to disappear from the field of view (91d).
The condenser height setting is dependent on the thickness of the specimen and may require adjustment for each new specimen.
Koehler Illumination
Suitable aperture and field diaphragm values have been preset for each objective. The con­denser has also been centered at the factory. However, it may be necessary to readjust the condenser in some cases. Therefore, check the condenser centering.
The following procedure is provided for the transmitted light-brightfield illumination. All required functions can be executed at the touch of a button with the Leica DMI 6000 B electronic microscope. (See Chapter 8, Opera­tion).
Preparation:
• Configure the microscope as follows: Set up the illumination, condenser, objectives and eyepieces correctly. (Please make sure that the objectives are properly screwed in and check the eyepiece settings.)
• Switch the microscope on and wait for the initialization phase to complete (automatic functions only).
•You will need either an empty Petri dish (preferably with a glass bottom) with a marking in the middle or a stained specimen on a slide with a coverslip.
• Switch to the 10x objective (if not present, the 20x objective).
Note:
65
7. Start-up
• Make sure that the condenser is at the correct height. The condenser height adjustment lets you set the condenser head to the height of the nominal free working distance. (For an S23 condenser, for example, the distance between the surface of the stage and the front lens of the condenser is approx. 23mm).
• Hold a piece of white paper (approx. 3-10cm) under the light source (field diaphragm). A light ring should appear on the paper – if not, check the power cable, the light source and the fuse of the supply unit (CTR box) and make sure that all of the parts are correctly connected to one another.
• Open the field diaphragm as far as possible until the light ring reaches its maximum diameter.
• Next, hold the paper under the condenser, directly on the stage. Open the aperture diaphragm as far as possible, until the light ring has reached its maximum brightness. In order to achieve maximum brightness, make sure that no port is activated. The full light should be directed to the VIS port.
• Check the magnification changer to make sure that the 1x tube lens is selected.
• Adjust the lenses of the eyepieces so that circle is visible in the eyepieces (not two!). If you wear eye glasses, remove the antiglare hoods from the eyepiece tubes (or fold them back).
• Make sure that the focus on the eyepieces is set to ±0 (turn the upper part of the eyepiece tubes until the silver ring is just covered).
•You should see light when looking through the eyepieces at this point. If the light is too bright, reduce it as required.
Remove all unneeded components from the light path.
one
• Swing all filters (in the filter magazine of the lamp housing or the filter holder of the condenser) out of the beam path.
• Set the condenser disk to the brightfield position.
• If your microscope is equipped for DIC:
• Remove the polarizer.
• Remove the analyzer.
• Remove the objective prism (move the
magazine to the "empty" or "brightfield" position).
• If your microscope is equipped for fluorescence:
• Select an empty filter position (or a filter
with low transmission in the visible range, e.g. filter A).
Now to begin with the actual Koehler illumina­tion:
• Place your specimen on the stage and focus so that you can see its details as clearly as possible. You probably will not get a perfect image at this point, as the illumination will not be optimal (90a).
• Next, attempt to get a sharp image (or at least a part of the image at the edge) by carefully
the condenser up and down (90.2). Try
moving this with a variety of field diaphragm settings until you get a clear, sharp image (91.b). This may take a while!
•To center the sharp image, insert the centering keys in the openings provided at either side of the top part of the condenser (90.1). Move the image into the center of the field of view (91.c). Next, open the field diaphragm until the image fills nearly the entire field of view. The black edges of the image should have the same distance to the outer edge of the field of view on all sides. If
66
7. Start-up
not, recenter the image with the centering screws. Adjust the height of the condenser until the edges are sharp. Now open the field diaphragm until the image fills the entire field of view and the black edges have disappeared completely (91.d).
• The last step is the adaptation of the contrast settings. To improve the contrast, close the aperture diaphragm – if you close it too far, however, the resolution of the image details will deteriorate. To see the aperture diaphragm, remove an eyepiece tube and look directly into the tube. Your eye should be around 10 to 20 cm from the tube. Change the size of the aperture diaphragm until its image is clearly visible in the pupil of the objective.
• Set the aperture diaphragm to cover 2/3 to 4/5 of the pupil diameter. You will now have the optimal balance between resolution and contrast.
Fig. 90 Condenser centering 1 Centering openings 2 Height adjuster 3 Prism and phase ring centering
2 1
3
Fig. 91 Koehler Illumination a Field diaphragm not focused, not centered b Field diaphragm focused, but not centered c Field diaphragm focused and centered
d Illumination field diameter = visible field diameter
21
3
Diameter is too small, however
(Koehler illumination)
a
cd
b
67
7. Start-up
7.6.2 Incident Light - Fluorescence
Suitable aperture and field diaphragm values have been preset for each objective. The inci­dent light module has also been centered at the factory.
However, it may be necessary to readjust the in­cident light module in some cases after trans­porting and setting up the stand. Therefore, check the field diaphragm centering. The following procedure is provided for the inci­dent light-brightfield illumination.
• Select an objective with moderate magnification (10x-20x).
• Activate the incident light axis with the TL/IL button (84.1).
• Press the IL-BF / Fluo button to activate the brightfield contrast method (one of the variab­le function buttons on the stand).
• Place a specimen on the stage.
Adjusting the Field Diaphragm
• Close the field diaphragm with the FD button (84.4) or manually until the edge of the diaphragm (round or rectangular) appears in the field of view.
• If the limits of the field diaphragm are not in the center of the field of view, move the position of the field diaphragm to the center with the two centering screws (92.1) on the right side of the stand.
• Use the function buttons FD (84.4) to open the field diaphragm to the point that they just disappear from the field of view.
•We recommend the use of a rectangular field diaphragm when using a digital camera. Match the size of the diaphragm to the chip size of the camera.
• Focus the specimen using the SmartMove or the focus wheels.
• Adjust the light intensity with the INT buttons (84.3).
68
Fig. 92 Adjusting the field diaphragm (incident light-
fluorescence)
1 Adjusting screws for moving the field diaphragm
1
7. Start-up
7.7 Checking Phase Contrast Rings
If your microscope is equipped for phase con­trast, light rings to match your objectives will be installed in the condenser. The light rings are already centered in the facto­ry. However, the centering should be rechecked.
Note:
Each objective has its own light ring assigned to it in the condenser. The test must therefore be performed for each objective. When selecting an objective suitable for phase contrast, the ap­propriate light ring will automatically be select­ed as well.
• Press the BF (brightfield) button (one of the variable function buttons on the stand).
• Instead of an eyepiece, place a focusing telescope (Fig. 93) in the observation tube or activate the Bertrand lens (pull rod (94.1) on tube).
• Select the phase contrast objective with the lowest magnification (one of the variable function buttons on the SmartMove).
• Focus on the specimen using the SmartMove or the focus wheels.
Fig. 94 1 Activating the Bertrand lens 2 Focusing the Bertrand lens
2
1
Fig. 93 Focusing telescope 1 Adjustable eyelens 2 Clamping ring for fixing the focus position
1
2
Fig. 95 Phase contrast centering procedure PH=phase contrast ring, LR=light ring
a Condenser in brightfield (BF) position b Condenser in phase contrast (PH) position
Light ring (LR) not centered
c Light ring and phase ring centered
ab c
PH LR
69
7. Start-up
• Focus the ring structure (95a) by loosening the clamping ring (93.2) somewhat and moving the eyelens (93.1), or focus the Bertrand lens (94.2).
• Retighten the clamping ring.
• Press the PH (phase contrast) button (one of the variable function buttons behind the focus wheels). The light ring will be selected in the condenser.
• If the light ring and the phase ring are not shown as arranged in Fig. 95c, the light ring must be centered.
• Insert the centering keys into the openings provided on both sides of the condenser (90.3).
•Turn the centering keys until the dark ring (phase ring in the objective) is congruent with the slightly narrower bright ring (light ring in condenser) (95 c).
7.8 Setting the Motorized Polarizer
• Select the POL method (one of the variable function buttons on the stand).
• Insert the centering key in the opening provided on the condenser (Fig. 96).
• Set up optimal darkening. (The analyzer must be in place.)
• Remove the centering keys.
Fig. 96 Condenser with motorized polarizer 1 Centering key for polarizer
1
Caution!
The centering keys must be removed from the centering openings before changing ob­jectives. They may block the condenser.
• Repeat the process for all additional phase contrast objectives.
• Remove the centering keys after centering.
70
Fig. 97 Condenser centering 1 Centering openings 2 Height adjuster 3 Prism and phase ring centering
2 1
3
21
3
7. Start-up
7.9 Adjusting the Light Sources
Transmitted Light Axis (TL) with Lamp Housing 107/2
The lamp housing 107/2 with a 12V 100W halo­gen lamp is fixed. Centering the lamp is not re­quired.
Lamp Housing 107 L for 12V 100W Halogen Lamp
The lamp can be adjusted using the screws (98.2) and the button (98.3).
• Place a sheet of white paper under the field diaphragm.
• Adjust the lamp to create an evenly bright spot on the paper.
Incident Light Axis (IL) with Lamp Housing 106 z
• When a supply unit is used, it is turned on first.
• Activate the incident light axis with the TL/IL function button. FLUO will appear on the LeicaScreen.
• Insert the lamp adjustment reflector (Fig. 99) in the filter turret in place of a filter cube. Make a note of the designation of the replaced filter cube.
•Turn the reflector into the beam path. The reflector is correctly positioned when the LeicaScreen shows the designation of the replaced filter cube.
Fig. 98 Lamp housing 107 L 1 Locking screws for housing 2 Screw for vertical adjustment 3 Button for horizontal adjustment 4 Collector focusing
4
12
Fig. 99 Reflector cube for lamp adjustment
3
71
7. Start-up
Caution!
Never look directly into the beam path! Beware of the glare hazard when switching to reflector BF or Smith!
Centering the Hg 100W and Xe 75 W Mercury Lamps
• The adjustment window shows the direct image of the arc and its mirror image. These are generally not in alignment with one another.
Caution!
Light sources pose a potential irradiation risk (glare, UV-radiation, IR-radiation).
In the lamp housing 106 z, the direct image of the filament (in halogen lamps) or the arc (in gas dis­charge lamps) and its reflection are focused sep­arately and adjusted in relation to one another.
An adjustment window (2.8, p. 19) in which the light source is visible is located on the right side of the microscope.
Adjust the lamp as follows while observing the light source in the adjustment window.
• Focus the direct image with the collector (100.6).
• Use the adjusting buttons to pivot the arc’s mirror image on the rear side of the lamp housing (100.2,100.4) to the side or completely out of the beam path. The arc’s focused image remains visible (Fig. 101).
• Use the adjusting buttons (100.1 and 100.5) to place the direct arc image in the middle of the centering plane, whereby the bright tip of the arc, the focal spot, should lie slightly outside the center (Fig. 102).
Fig. 100 Lamp housing 106 z L 1 Lamp adjustment, vertical 2 Vertical reflector adjustment 3 Focusing the reflector image 4 Horizontal reflector adjustment 5 Lamp adjustment, horizontal 6 Collector focusing 7 Screw
1
6
7
5
72
2 3 4
7. Start-up
• Then pivot the arc’s mirror image with the adjusting knobs (100.2) and (100.4) and focus it using the reflector (100.3).
• Use the adjusting knobs (100.2) and (100.4) to orient the mirror image symmetrically to the direct image (Fig. 103). The V-shaped irradiation of the direct image and mirror image arcs can be superimposed.
Caution!
The bright tips of the arcs, the focal spots, must never be projected onto each other, as this results in a danger of explosion by over­heating.
Caution!
The structure of the arc can no longer be made out clearly in lamps that have been in service for a long time. The image and mirror image can no longer be superimposed ex­actly. In this case, align both images.
Fig. 101 Direct arc image focused but not centered
(in reality, the image is less focused)
Fig. 102 Direct arc image in target position
(in reality, the image is less focused)
• Using the collector, defocus the image with the knob (100.6) until the arc image and mirror image are no longer recognizable and the image is homogeneously illuminated.
• Replace the lamp adjustment reflector with the original filter cube.
Fig. 103 Direct arc image and mirror image in target
position (in reality, the image is less focused)
73
8. Operation
8. Operation
8.1 Switching On
When using a gas discharge lamp, the ebq 100 external supply unit must be turned on separate­ly (104.1).
Switch on the power of the CTR6000 electronics box at the On/Off switch (105.1). The signal lamp (105.2) is lit green when the unit is ready.
Note:
If a PC is connected, switch on the electronics box first, and then the computer.
All motorized microscope components will then run through an initialization phase.
Note:
Fig. 104 Front panel of ebq 100 supply unit 1 Power switch 2 Lamp status
In the case of faulty initialization ("Init Error" message on LeicaScreen), see Troubleshooting chapter, p. 91.
All of the user's previous settings are restored during the initialization.
Caution:
!
The focal position and the lower stop are also retained from one session to the next when power is switched off.
Fig. 105
Front side Leica CTR6000
1 On/Off switch 2 Signal lamp
74
2 1
1 2
After the initialization is complete, the LeicaS­creen will display the status screen with mi­croscope's current settings. Fig. 107 is an ex­ample.
Note: (Reset Function)
The microscope can be reset to its factory de­fault programming:
• With the stand switched off, press the top three variable function buttons on the left side of the stand.
• Switch on the power for the stand.
• Hold the buttons until the initialization is complete.
• The standard information display will now appear in the LeicaScreen (Fig. 106 and 107).
8. Operation
• Switch the instrument off and back on. The settings are now saved.
Fig. 106 LeicaScreen initialization Fig. 107 LeicaScreen following initialization
75
8. Operation
8.2. Contrast Methods
All contrast processes can be selected and con­trolled using the variable function buttons and the Leica Application Suite (LAS) software. The only exceptions to this are processes involving com­ponents that require manual operation (e.g. sys­tems with manual analyzers). The following sec­tion describes the use of the function buttons on the stand. For instructions on the use of the soft­ware, please refer to the separate manual.
Fig. 108 Function buttons (left side of stand) 1 Variable function buttons 2 Open/close aperture diaphragm 3 TL/IL switching 4 Open/close field diaphragm 5 Increase/decrease light intensity
54321
8.2.1 Brightfield (TL)
Note:
If all positions of the filter turret are occupied, filter cube "A" can be swapped for filter cube "A-TL" using the Leica Application Suite (LAS). TL contrast methods are possible with that filter cube.
• Use the TL/IL function button to switch to transmitted light (TL).
• Select the BF (brightfield) contrast method by pressing the variable button BF. Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see the identification sheet.) BF will appear on the LeicaScreen. Motorized condensers will now move to the brightfield position. Coded condensers must be switched manually. The fluorescence filter turret will automatically go to an empty position or to the "A-TL" filter cube.
• Insert a transmitted light specimen.
Fig. 109 Function buttons (left side of stand) 1 Variable function buttons
1
76
• Rotate an appropriate objective into place.
• Focus the image with the knob on the SmartMove or the focusing wheel and adjust the intensity with the INT function buttons.
8. Operation
8.2.2 Phase Contrast (TL)
• Use the TL/IL function button to switch to transmitted light (TL).
• Select the BF (phase contrast) contrast method by pressing the variable button PH. Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see the identification sheet.) PH will appear on the LeicaScreen. Motorized condensers will now switch to the correct light ring. Coded condensers must be switched manually.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place. Objectives that are suitable for phase contrast are engraved with PH.
• Focus the image with the knob on the SmartMove or the focusing wheel and adjust the intensity with the INT function buttons.
8.2.3 Darkfield (TL)
• Use the TL/IL function button to switch to transmitted light (TL).
• Select the DF (darkfield) contrast method by pressing the variable button BF. Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see the identification sheet.) DF will appear on the LeicaScreen. Motorized condensers will now switch to the darkfield ring. Coded condensers must be switched manually.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
• Focus the image with the knob on the SmartMove or the focusing wheel and adjust the intensity with the INT function buttons.
Note:
When selecting the phase contrast method, the aperture diaphragm is opened fully and cannot be adjusted.
Note:
The maximum usable objective aperture for dark field is 0.70 for the condenser S1 and 0.40 for the condenser S23/S28.
When selecting the darkfield method, the aper­ture diaphragm is opened fully and cannot be adjusted.
77
8. Operation
8.2.4 Polarization (TL)
• Use the TL/IL function button to switch to transmitted light (TL).
• Select the POL (polarization) contrast method by pressing the variable button POL. Alternatively: press the variable button CHANGE TL .
(For details on button assignments, please see the identification sheet.)
POL will appear on the LeicaScreen.
Manual Method:
• Move the polarizer on the condenser into the beam path.
• Insert the analyzer into the right side of the stand until it clicks into position (Fig. 110).
• Bring the polarizer and analyzer into cross position until they reach maximum darkness.
• Place a specimen on the stage and select a suitable objective.
Motorized Method:
• If the microscope is equipped with the rel­evant components, the polarizer will be acti­vated automatically in the condenser when the POL contrast method is selected. The ana­lyzer cube is also automatically positioned in the beam path.
Combined Methods:
• The Leica DMI6000 B microscope permits purely mechanical and motorized components — such as a mechanical analyzer and motor­ized polarizer — to be combined.
78
Fig. 110 Inserting the analyzer
8. Operation
8.2.5 Differential Interference Contrast (TL)
• Use the TL/IL function button to switch to transmitted light (TL).
• Select the DIC contrast method by pressing the variable button DIC. Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see the identification sheet.) DIC will appear on the LeicaScreen.
• The polarizer in the condenser and the suit­able condenser prism are automatically posi­tioned in the beam path. The corresponding objective prism and the analyzer cube are also positioned automatically.
• Place a DIC specimen on the stage.
• Rotate an appropriate objective into place.
• Focus the image with the knob on the SmartMove or the focusing wheel and adjust the intensity with the INT function buttons.
Manual alternative:
• Move the polarizer on the condenser into the beam path manually.
• Insert the analyzer manually into the right side of the stand until it clicks into position (Fig.
110). Adjust the objective and condenser prisms manually until a valid combination appears on the display.
• Use the knurled wheel below the objective turret for fine adjustment (Fig. 111).
• Use the knurled wheel below the objective turret for fine adjustment (Fig. 111).
Fig. 111 DIC disk with knurled wheel for fine adjustment
79
8. Operation
8.3 Fluorescence
• Use the TL/IL function button to switch to fluorescence FLUO.
• Place a specimen on the stage and select a suitable objective.
• The current fluorescence filter cube will be displayed on the LeicaScreen.
•You may protect your specimen from fading by closing the incident light shutter. To do so, press the SHUTTER button (87.3) on the front panel. The following pictogram will appear on the LeicaScreen:
Changing the Fluorescence Filter Cube
Fixed function buttons on the front panel:
CUBE 1 to CUBE 6 or Cube CCW
• Focus the image with the knob on the SmartMove or the focusing wheel and adjust the intensity with the INT function buttons.
Options
• The intensity of the fluorescence can be increased by using the booster lens (Fig. 112) on the left rear side of the stand (Fig. 113). If bright fluorescence is required in the center of the field of view, slide the booster lens into the receptacle with the marking
1.4x facing the user. If a homogeneous distribution over the entire field of view is required, turn the booster lens 180° so that the marking
0.7x is facing forward.
• For multiple fluorescence, we recommend using the Excitation Manager and/or the ultra­fast internal filter wheel. Emissions and excitation wavelengths can thus be changed in milliseconds. They are controlled by the function buttons.
Variable function buttons on the front
panel and SmartMove: CUBE CW or CUBE CCW
Leica Application Suite (LAS) software
Fig. 112 Booster lens
80
Fig. 113 1 Booster lens in stand
1
8.4 Combination Methods
8. Operation
Up to two combination methods are possible de­pending on the features of the individual micro­scope:
FLUO/PH and FLUO/DIC
• Select the combination method by pressing the variable button COMBI . Alternatively: press the variable button CHANGE COMBI . (For details on button assignments, please see the identification sheet.) The content of the display changes accordingly.
• Place a specimen on the stage and select a suitable objective.
• Select the desired filter cube using the fixed function buttons on the front panel.
• The illumination settings for the fluorescence and transmitted light axes can be adjusted separately.
Note:
manual analyzer (Fig. 110) must be used for
The the FLUO/DIC method as described in Chapter
8.2.5, p. 79.
•Toggle the illumination axes with the TL/IL function button. The content of the LeicaScreen changes accordingly.
FLUO > DIC
The transmitted illumination is activated.
FLUO < DIC
The fluorescence illumination is activated.
81
8. Operation
8.5 Focusing
Note:
The parfocality teach-in has already been per­formed at the factory. However, it may be neces­sary to perform another teach-in after installing the objectives when setting the microscope up. We recommend checking parfocality ting the stops and performing a teach-in with the Leica Application Suite (LAS) if necessary.
Focusing the Image
The focusing is controlled by the knobs (116.3, p. 88) on the SmartMove remote control module.
Alternatively, use the focus wheels on either side of the stand.
The current Z position is shown on the Leica­Screen. In the case of motorized stages, the Z drive will travel to its lowest position prior to the stage initialization when switching the micro­scope on.
The focus buttons Z the stand (Fig. 114) permit fast focusing or low­ering of the objectives.
and Z on the right side of
before set-
Set the lower focus stop by pressing and holding the SET button and pressing the Z well. The display will show . Pressing the button combination again will de­lete the stop. The display will show .
The lower focus stop can also be set using the Leica Application Suite (LAS). The lower stop is the same for cannot be traversed.
In addition, a focus position that cannot be tra­versed can also be set. To do so, press and hold the SET button and press the Z The display will show . Pressing the button combination again will de­lete the stop. The display will show .
The focus position can also be set using the Lei­ca Application Suite (LAS).
Set the focus position for the dry objective at the highest magnification. The focus positions will then be set automatically for all other objec­tives, taking parfocality and working distances into account.
button as well.
 
 
button as
all objectives and
Setting Stops
Fig. 114 1 Focus control buttons
1
82
Set the stops via
Fixed function buttons on stand
Leica Application Suite (LAS).
Summary of pictograms
8. Operation
Lower focus stop not set
Lower focus stop set
Focus position not set
Focus position set
 
Going to the Stops
Go to the lower stop by pressing and holding the
button.
Z
Go to the focus position by pressing and holding
button.
the Z These functions can be assigned to variable function buttons on the stand or SmartMove, or they can be controlled via software.
Go to Stops via
Fixed function buttons on stand
Variable function buttons on stand and
SmartMove
It is possible to toggle between Fine and Coarse step increments. The Fine value varies to suit the tive. Suitable values have been predefined. The assignments can be changed with the Leica Ap­plication Suite (LAS). When selecting Coarse, the positioning speed is the same for to the maximum speed.
The assignment of a specific step increment to an objective not only applies to the Z drive, but also to the step increments assigned to the stage when Precise ( p. 88) is selected.
all objectives. Coarse corresponds
Note:
current objec-
Switch between Fine and Coarse via
Variable function buttons on stand and
SmartMove
Leica Application Suite (LAS).
8.6 Tubes
Leica Application Suite (LAS).
Note:
When going to the stops with the Z tons, hold the button until the stop has been reached.
Setting the Step Increments
and Z but-
83
8. Operation
Note:
Close any unused tube openings, as otherwise stray light can interfere with observation.
The
button on the front control panel switches 100% of the light to the eyepieces. Use the
Adjusting the Viewing Distance
• Adjust the viewing distance of the eyepieces so that a congruent total image is seen (Fig. 115).
Adjusting the Vviewing Angle
• Ergotubes feature a tilting binocular section for a 10°-45° viewing angle adjustment range.
Beam Splitting in Photo Tubes
The beam splitting is set manually by pulling out a control bar.
Button Observation Photo VIS 100 % 0 % 50/50 50 % 50 % BL activation of Bertrand lens*
Light distribution via
Manual control bar.
button, also on the front control panel, to select the side ports. Depending on the configuration, the screen will now display
- The active port (right or left) and
- The percentage of light going to the port (100%, 80%, 50%).
Optional: The bottom port selection function can be as­signed to one of the variable function buttons on the stand or the SmartMove. The top port can only be selected manually.
Select ports via
Fixed function buttons on stand (side
ports)
Variable function buttons on stand and
SmartMove (bottom port)
Manual action (top port).
8.7 Eyepieces
Fig. 115 Tube setting
Selecting Ports
84
Note:
The eyepiece’s aperture protector must be re­moved or folded back, during microscopy while wearing eyeglasses. We recommend removing bifocals and spectacles with progressive-addition lenses when using the microscope.
8. Operation
We recommend running a teach-in via the Leica Application Suite (LAS) software when using eyepieces not included in the scope of delivery. This will ensure that the total magnification shown in the LeicaScreen is correct.
8.8 Objectives
Changing Objectives
• For the adjustable tubes with documentation output, choose the 100% VIS position.
Eyepieces with Inlaid Reticle
• Focus the reticle by adjusting the eyelens.
• Focus on the object through this eyepiece.
• Then, close that eye and focus on the object
by adjusting only the second ocular.
Correction for Vision Problems
• With your right eye, look through the right
eyepiece and bring the specimen into sharp focus.
• Then, with your left eye, view the same
specimen and rotate the left eyepiece tube until the object is brought into sharp focus. Do not change the Z position in the process!
The objectives can be selected with the function buttons on the stand or the SmartMove, or by manually turning the objective turret. When changing objectives manually, please make sure that the turret clicks into position.
The positions of the objectives in the objective turret have been specified at the factory and must be observed when installing the objec­tives. (also see Objectives, p. 39).
When selecting an objective, the microscope automatically selects:
• The optimal setting for the field diaphragm
• The optimal setting for the aperture diaphragm
• The light intensity for the current contrast method
The objective magnification and total magnifica­tion are displayed on the LeicaScreen.
Note:
• For immersion objectives use the appropriate immersion medium.
85
8. Operation
OIL: only use optical immersion oil
according to DIN/ISO standards.
Cleaning p. 96. W: Water immersion. IMM: Universal objective for water, glycerol,
oil immersion.
Caution!
Follow safety instructions for immersion oil!
Select objectives via
Variable function buttons on stand and
SmartMove
Leica Application Suite (LAS) Software
Manual selection
Changing the Operating Modes "dry" (DRY) and "immersion" (IMM)
Each objective is assigned to a specific objec­tive category:
1) Dry objectives (DRY)
2) Immersion objectives (IMM)
• First, select the operating mode (Imm or Dry) using the function buttons. The operating mode may also be selected in the Leica Application Suite (LAS).
• The objective turret is lowered to its bottom stop. This is to permit the application of the immersion liquid when changing from a dry to an immersion objective. It also permits the removal of the liquid when changing to dry mode. The current objective remains in the beam path.
• Next, press the button for the objective you intend to use.
Note:
If the Imm or Dry operating mode buttons are pressed accidentally, the original mode can be restored by pressing the appropriate button.
Change operating mode via
Variable function buttons on stand and
SmartMove
Note:
It is possible to use objectives for both operating modes. The mode can be assigned in the Leica Applica­tion Suite (LAS).
Changing the Operating Mode
86
Leica Application Suite (LAS) Software
8. Operation
When replacing objectives, you must perform a teach-in for the new objectives in the Leica Ap­plication Suite (LAS). A parfocality teach-in should also be performed.
Color Coding of Objectives
The magnification of each objective is indicated by a color ring in accordance with DIN/ISO stan­dards:
100x 63x 40x 25x 16x 10x 6.3x 4x 2.5x 1.6x 125x 50x 32x 20x 5x 150x 160x
white dark- light- dark- light- yellow orange red brown gray
Note:
Note:
blue blue green green
For lockable immersion objectives lock these by pushing the front part upwards until it stops (ap­prox. 2mm). Then, after a gentle turning motion to the right, the objective is locked. For objectives with corrective mounts turn the knurl to adjust the objective to the thickness of the cover glass.
8.9 Stages and Object Displacement
Object Displacement Using SmartMove
Immersion objectives are marked by an addi­tional, lower color ring.
black oil or Imm (universal objective for
oil, water or glycerin)
white water orange glycerin
The various engraved markings of the objectives provide information on their applications:
black or brightfield objectives, dark blue strain-free
reen phase contrast objectives,
g
strain-free
87
8. Operation
The positioning of the stage is controlled by the knobs (116.1, 116.2) on the SmartMove remote control module.
Setting the Step Increments
The positioning speed of the stage can be varied by switching between the Fast and Precise step increments. When selecting Fast, the positioning speed is the same for The Precise speed varies to suit the
all objectives.
current ob-
jective.
Switch between Precise and Fast
via
Variable function buttons on stand and
SmartMove
Leica Application Suite (LAS)
Storing and Restoring Stage Positions
A variety of stage positions can be stored tempo­rarily in the Leica Application Suite (LAS). The XY position is stored, not the Z position. In addition to a loading position (Load), 5 stage positions can be set temporarily. When switch­ing the microscope on, the stage will travel to a previously-defined starting position.
Temporarily Store and
Restore Stage Positions via
Leica Application Suite (LAS)
8.10 Magnification Changer
Fig. 116 SmartMove remote control module 1 Travel in x 2 Travel in y 3 Focus 4 Individual adjustment of button height 5 Variable function buttons (factory preset)
88
1
3
4
2
5
8. Operation
A motorized magnification changer can be used optionally. The following magnification factors can be selected:
1x; 1.5x; 1,6x 2x
The selected factor is displayed on the LeicaS­creen and in the relevant field of the Leica Ap­plication Suite (LAS), and is taken into account when calculating the total magnification.
Pressing the left button (117.1) switches be­tween the possible magnification factors; press­ing the right button selects the factor 1x.
8.11 Light Sources
Change magnification via
fixed function buttons on stand
Leica Application Suite (LAS) Software
• Adjust the intensity with the function buttons (118.4). The INT function buttons are always assigned to the currently active transmitted light (TL) or incident light (IL) axis.
• For TL and IL: The setting can be made in coarse and fine steps. Pressing both INT (118.2) buttons as the same time toggles between coarse and fine adjustment. The light intensity displayed on the LeicaScreen changes accordingly.
Coarse adjustment: 0-20 Fine adjustment: 0-255
• The intensity is individually adjusted and stored for each objective and contrast method.
• FLUO: The intensity can be adjusted in 5 fixed levels.
100% / 55% / 30% / 17% / 10%
(FIM=Fluorescence Intensity Manager)
Adjust intensity via
Fig. 117 Front control panel 1 Function buttons for magnification changer
1
Fixed function buttons on stand
Variable function buttons on stand and
SmartMove
Leica Application Suite (LAS) Software
8.12 Aperture and Field Diaphragm
1
89
8. Operation
Both diaphragms have been set to suitable val­ues for the current objective and contrast meth­od at the factory.
• The diaphragms can be adjusted at any time with the AP (aperture diaphragm) (118.1) and FD (field diaphragm) (118.3) function buttons. The values displayed on the LeicaScreen change accordingly. The function buttons are assigned to the currently active transmitted light (TL) or incident light (IL) axis.
Caution:
The old values will be overwritten by the current ones!
Caution:
When using PH or DF, the aperture diaphragm is fully open and
cannot be closed.
Adjust diaphragms via
Fixed function buttons on stand
Variable function buttons on stand and
SmartMove
Leica Application Suite (LAS) Software
90
Fig. 118 Fixed function buttons, left side of stand 1 Aperture diaphragm 2 Transmitted light/incident light 3 Field diaphragm 4 Light intensity
1234
9. Troubleshooting
9. Troubleshooting
Problem
Stand
The microscope does not respond.
Illumination
The image is completely dark.
The image is unevenly or not uniformly illuminated.
Cause/Remedy
Make sure that the AC outlet has power.Make sure that the CTR6000 electronics box is
connected to an AC outlet.
Check the cable connections.Inform Service and have the supply unit fuse
checked.
Open the shutter (Check the connections of the lamp housings
p. 63).
on the microscope (transmitted light/ fluorescence).
Make sure that the lamps are connected to
the power supply and are not defective.
Inform Service and have the ebq 100 supply
unit fuse checked.
Remove all unneeded filters from the light
path.
Center the lamp (Replace the old lamp (
p. 71ff).
p. 41f, 44ff).
The illumination flickers.
The lamp does not illuminate immediately upon being switched on.
Be sure that there is no loose connection at
the power supply.
Replace the old lamp (
The ebq 100 must be switched-on repeatedly.Hot Hg lamps should cool down before
p. 41f, 44ff).
switching on again.
91
9. Troubleshooting
Problem
Brightfield
The specimen can not be brought into focus.
Darkfield
No definite DF contrast is possible.
Cause/Remedy
Use the correct immersion medium.Place the specimen on the stage with the
coverslip facing down.
Make sure that the cover glass thickness is
correct and that it suits the indication on the objective.
Make sure that you are using an objective
with coverslip correction.
Adjust the correction ring on the objective if
present.
Be sure that a DF objective is being used.The objective aperture is too high:
Maximum 0.7 for condenser S1 Maximum 0.4 for condenser S2328 If necessary, reduce the objective aperture using the iris diaphragm on the objective.
Check the condenser centering.
The image is unevenly or not uniformly illuminated.
Undesirable stray light.
92
The magnification is too weak.
Use a higher magnification.
Remove the condenser head or condenser
lenses.
Clean the specimen and neighboring lenses
p 95).
(
9. Troubleshooting
Problem
Phase contrast
No phase contrast is possible.
Polarization
No polarization contrast is possible.
Transmitted light interference contrast
No transmitted light interference contrast is possible.
Cause/Remedy
The specimen is too thick.The cover glass is not placed evenly.Check the centering of the light rings
(p. 69).
Bring the polarizer and analyzer into cross
position until they reach maximum darkness (without specimen). (p. 78).
The specimen is too thick or too thin.Embedding medium or specimen are of
birefringent material. Rotate the specimen.
The difference in the refractive indices of the
specimen and the embedding medium is too small.
The cover glass is too thick.Check the Koehler illumination (p. 65).Bring the polarizer and analyzer into cross
position until they reach maximum darkness (without specimen). (p. 78).
Check whether the suitable condenser prism
and corresponding objective prism are selected (manual alternative p. 79).
Make sure that the IC prisms are correctly
seated (p. 36).
93
9. Troubleshooting
Problem
Fluorescence
The image is completely dark (no fluorescence).
The fluorescence is too weak.
LeicaScreen
Init Error!
Cause/Remedy
Open the shutter (Select the incident light axis (IL) (Check the antigen-antibody combination.Insert a new lamp (
Insert the booster (Center the lamp (Insert a new lamp (
Check the cable connections.Check whether the cover of the filter disk has
p. 63).
p. 44ff).
p. 80).
p. 71ff).
p. 44ff).
p. 61).
clicked into position.
Check the installed objectives, filter cubes,
etc.
Switch the microscope off and back on.
94
10. Care of the Microscope
10.2 Cleaning
Caution!
10. Care of the Microscope
Unplug the power supply before performing cleaning and maintenance work! Protect electrical components from mois­ture!
Microscopes in warm and warm-damp climatic zones require special care in order to prevent fungus contamination. The microscope should be cleaned after each use, and the microscope optics should be kept strictly clean.
10.1 Dust Cover
Note:
To protect against dust, cover the microscope and accessories with the dust cover after each use.
Caution!
Let lamps cool down before covering the stand with a dust cover. The dust cover is not heat-resistant. In addition condensation water may occur.
Caution:
!
Residual fiber and dust can create unwanted background fluorescence.
Cleaning Coated Parts
Dust and loose dirt particles can be removed with a soft brush or lint-free cotton cloth.
Stubborn dirt can be removed with all commonly available aqueous solutions, naphtha or alcohol. For cleaning coated parts, use a linen or leather cloth that is moistened with one of these sub­stances.
Caution:
!
Acetone, xylene or nitro-containing thinner can harm the microscope and thus must not be used.
Test cleaning solutions of unknown composition first on a less visible area of the unit. Be sure that coated or plastic surfaces do not become matted or etched.
Cleaning the Stage
Rub the stage with paraffin oil or acid-free Vase­line to remove light spots on the stage.
95
10. Care of the Microscope
Cleaning Glass Surfaces
Remove dust on glass surfaces with a fine, dry, and fat-free hair brush, by blowing with a blow bag or vacuum suction.
Remove stubborn dirt on glass surfaces with a clean cloth dampened with distilled water. If the dirt still can not be removed, use pure alcohol, chloroform or benzine.
Cleaning Objectives
Caution!
The objective may not be unscrewed during cleaning. If damage appears on inner sur­faces, the objectives must be sent to your Leica subsidiary for repair. We also advise against cleaning the inside surfaces of the eyepieces.
The front lenses of objectives are cleaned as described under "Cleaning Glass Surfaces". The upper lens is cleaned by being blown off with a pneumatic pump.
Removing Immersion Oil
Caution!
Follow safety instructions for immersion oil!
First, wipe off the immersion oil with a clean cotton cloth, and then re-wipe the surface sev­eral times with ethyl alcohol.
10.3 Handling Acids and Bases
For examinations using acids or other aggres­sive chemicals, particular caution must be tak­en.
Caution:
!
Be absolutely certain to prevent the optics and mechanical parts from coming into con­tact with these chemicals.
96
11. Major Consumable and Replacement Parts
11. Major Consumable
and Replacement Parts
Order No. Material No. Name Used for
Replacement Lamp 11 500 974 Halogen lamp 12V 100W 107/2 lamp housing 11 500 137 High-pressure
11 500 138 High-pressure
11 500 321 High-pressure
11 500 139 High-pressure
Screw cap for unused objective receptacles 020-422-570-000 Screw cap M 25 Objective turret
Cover for unused objective DIC disk opening 11 090-144-020-088 Cover for DIC microscope stand
Dust and light protection cover for analyzer slot 11 020-437-101-013 Analyzer slot cover microscope stand
mercury burner 50 W 106 z lamp housing
mercury burner 100W 106 z lamp housing
mercury burner 100W 106 z lamp housing
(103 W/2)
xenon burner 75 W 106 z lamp housing
Dust and light protection cover for camera port openings 11 020-387-556-009 Analyzer slot cover microscope stand
Replacement eyecup (diaphragm protection) for HC PLAN eyepiece 021-500-017-005 HC PLAN eyecup 10x/25 eyepiece 021-264-520-018 HC PLAN eyecup 10x/22 eyepiece 021-264-520-018 HC PLAN eyecup 10x/20 eyepiece
Immersion oil conforming to DIN/ISO standards, fluorescence-free 11 513 787 110 ml OIL and IMM objectives 11 513 522 100 ml and oil condenser heads 11 513 788 500 ml
97
12. Dimensions
12. Dimensions
Space Requirements
Height Compensation Plate*
A height compensation plate was developed to raise the viewing height by 20mm or to raise the side camera ports for oversize cameras or spin­ning disks, or to use the microscope with an in­active bottom port on workbenches without openings. The height compensation plate is available in two versions (12mm and 25mm).
98
13. Abbreviations and Pictograms
13. Abbreviations and Pictograms
Contrast method
Magnification
Illumination
Ports/Eyepiece
Focus
6 6
6
5 6
5
Lower focus stop not set
Lower focus stop set
Focus position not set
Focus position set
Shutter open
Shutter closed
Transmitted light filter
Field diaphragm, rectangular
Field diaphragm, round
Aperture diaphragm
Light distribution
99
13. Abbreviations and Pictograms
AP Aperture diaphragm BF Brightfield COMBI Combination method CUBE Fluo cube DF Darkfield incident/transmitted light DIC Differential Interference Contrast FD Field diaphragm FLUO Fluorescence axis (incident light) ICR Interference contrast, incident light ICT Interference contrast, transmitted light IL Incident light INT Intensity PH Phase contrast POL Polarization, incident/transmitted light TL Transmitted light
100
Loading...