Leica DM750 User Manual

Page 1
Leica DM750 Manual
Page 2
Manufacturer information
Published in November 2010 by:
Leica Microsystems (Schweiz) AG Industry Division Max Schmidheiny Strasse 201 CH-9435 Heerbrugg (Switzerland)
In case of questions, please contact the hotline: mqm-hotline@leica-microsystems.com
Leica DM750 Manual 2
Page 3
Chapter overview
Safety regulations 5
The Leica DM750 17
Ready! 20
Set! 30
Go! 43
Care of the Microscope 45
Troubleshooting 48
Dimensions 51
Leica DM750 Manual 3
Page 4
Contents
Safety regulations 5
Safety concept 6 Symbols used in this instruction manual 7 Important Notes 8 Intended Purpose of the Microscope 9 Instructions on use 10 Health risks and dangers of use 12 Information for the person responsible for the instrument 13 Care instructions 14 Accessories, maintenance and repair 15 Electrical data and ambient conditions 16
The Leica DM750 17
Introduction 18 Unpacking 19
Ready! 20
Substage Illumination 21 Attaching the viewing Tubes 22 Leica EZ Viewing Tube – integrated eyepieces 23 Leica Standard Viewing Tube – separated eyepieces 25 Eyecups 26 Installing Objectives 27 Installing the Substage Condenser 28
Set! 30
Turning on the microscope 31 Complete Condenser Centering 32 Using the Condenser 33 Prepare to View a Specimen Slide 34 Focusing 35 Viewing Tube Adjustment 36 Koehler Configuration 38 Oil Immersion Technique 40 Time Delay Shutoff 42
Go! 43
Ready! Set! Go! 44
Care of the Microscope 45
General maintenance 46
Troubleshooting 48
Troubleshooting 49
Dimensions 51
Leica DM750 Manual 4
Page 5

Safety regulations

Leica DM750 Manual 5
Page 6

Safety concept

The individual modules of the Leica DM microscopy series include an interactive CD-ROM with all relevant user manuals in several languages. Keep it in a safe place, and readily accessible to the user. User manuals and updates are also available for you to download and print from our web site at www.leica-microsystems.com.
This user manual describes the special func­tions of the individual modules of the Leica DM microscopy series and contains important instructions for their operational safety, main­tenance, and accessories.
The "Safety concept" booklet contains additi­onal safety information regarding the service work, requirements and the handling of microscope, accessories and electrical accesso­ries as well as general safety instructions. You can combine individual system articles with articles from external suppliers. Please read the user manual and the safety requirements of the supplier.
Before installing, operating or using the instru­ments, read the user manuals listed above. In particular, please observe all safety instructions.
To maintain the unit in its original condition and to ensure safe operation, the user must follow the instructions and warnings contained in these user manuals.
Leica DM750 Manual 6
Page 7

Symbols used in this instruction manual

Warning of a danger
This symbol indicates especially impor-
tant information that is mandatory to
read and observe.
Failure to comply can cause the following:
Ϙ Hazards to personnel
Ϙ Functional disturbances or damaged inst-
ruments
Warning of hazardous electrical voltage
This symbol indicates especially impor-
tant information that is mandatory to
read and observe.
Failure to comply can cause the following:
Ϙ Hazards to personnel
Ϙ Functional disturbances or damaged
instruments
Danger due to hot surface.
This symbol warns against touching
accessible hot surfaces, e.g. those of
light bulbs.
Important information
This symbol indicates additional infor-
mation or explanations that intend to
provide clarity.
Explanatory notes
ϐ This symbol within the text points to additi-
onal information and explanations.
Instructions for disposing of the inst­rument, accessory components, and
consumables.
Leica DM750 Manual 7
Page 8

Important Notes

Description
The Leica DM750 microscope meets today's state of the art of technology. Nevertheless, hazards may still arise during operation. The potential risks are described below.
Before installing, operating or using the
instrument, it is mandatory to read this user manual. In particular, please observe all safety instructions.
Contact address
Leica Microsystems (Schweiz) AG Industry Division Max Schmidheiny Strasse 201 CH-9435 Heerbrugg (Switzerland)
User Manual
This user manual includes important instruc­tions related to operating safety, maintenance and accessories,
Your Leica DM750 microscope comes
with an interactive CD-ROM with all relevant user manuals. Keep it in a safe place, and readily accessible to the user. User manu­als and updates are also available for you to download and print from our website at www.leica-microsystems.com.
Accessories from third-party suppliers
You can combine individual system articles with articles from external suppliers. Please read the user manual and the safety requirements of the supplier.
Original condition
To maintain the unit in its original condition and to ensure safe operation, the user must follow the instructions and warnings contained in these user manuals.
Legal requirements
Adhere to general and local regulations rela­ting to accident prevention and environmental protection.
EC Declaration of Conformity
Electrically operated accessories are construc­ted based on the state of the art of technology and are provided with an EC Declaration of Conformity.
Leica DM750 Manual 8
Page 9

Intended Purpose of the Microscope

The Leica DM750 microscope, for which
this instruction manual has been writ­ten, is intended for biological routine and research applications. This includes exami­ning specimens taken from the human body for the purpose of gaining information about physiological or pathological conditions or inborn anomalies, or testing for safety and compatibility for potential recipients, or for monitoring therapeutic measures.
The above-named microscope complies with the Council Directive 98/79/EEC concerning in vitro diagnostics.
Caution! The Leica DM750 microscope
is not intended for use with biologically
hazardous materials.
Caution! The manufacturer assumes
no liability for any noncompliant use or use outside of the specifications from Leica Microsystems CMS GmbH as well as any risks that may arise from such use. In such cases, the Declaration of Conformity shall be invalid.
Caution! This (IVD) instrument is not
intended for use in a patient envi­ronment as defined by DIN VDE 0100-710. It is also not intended to be combined with medical devices as defined by EN 60601-1. If a microscope is electrically connected to a medi­cal instrument in accordance with EN 60601-1, the requirements defined in EN 60601-1-1 shall apply. Not suitable for examining potentially infectious specimens.
Leica DM750 Manual 9
Page 10

Instructions on use

The Leica DM750 microscope may be
used only in closed rooms and must be placed on a solid substrate.
The Leica DM750 microscope can be used in clean rooms without any problems.
Always position the Leica DM750
microscope so that you can disconnect it from the power supply at any time. The power cable must remain accessible at all times, because it is intended as a power disconnect device.
Place of use
Only use the instruments in closed, dust free rooms and between +10°C and +40°C. Protect the devices from oil, chemicals and extreme humidity. If using the devices outdoors, protect them from dust and moisture. Never use electri­cal devices outdoors. Install electrical devices at least 10 cm from the wall and away from flam­mable substances.
Avoid large temperature fluctuations, direct sunlight and vibrations. These conditions can distort measurements and micrographic images.
In warm and warm-damp climatic zones,
the individual components require special care in order to prevent the build-up of fungus.
Non-intended use
Never install any other plug (NEMA
5-15P) or unscrew any mechanical components unless expressly instructed to do so in the instructions.
The devices and accessories described in
this instruction manual have been tested
for safety and potential hazards.
Leica DM750 Manual 10
Page 11
Instructions on use (continued)
The responsible Leica affiliate must be
consulted whenever the instrument is altered, modified or used in conjunction with non-Leica components that are outside of the scope of this manual!
Unauthorized alterations to the instru-
ment or noncompliant use shall void all rights to any warranty claims.
Transport
If at all possible, use the original packaging for shipping or transporting individual modules.
In order to prevent damage from vibrations, disassemble all moving parts that (according to the user manual) can be assembled and disassembled by the customer and pack them separately.
Disposal
Once the product has reached the end of its service life, please contact Leica Service or Sales about disposal. Please observe and ensure compliance with the national laws and regulations that implement, for example, the EC Directive WEEE.
Like all electronic devices, this inst-
rument, its accessory components and consumables must never be disposed of with general household waste. Disposal must comply with locally applicable laws and regu­lations.
Integration in third-party products
When installing Leica products into third-party products, the manufacturer of the complete system or its dealer is responsible for following all applicable safety instructions, laws and guidelines.
Leica DM750 Manual 11
Page 12

Health risks and dangers of use

Health risks
Workplaces with microscopes facilitate
and improve the viewing task, but they also impose high demands on the eyes and holding muscles of the user. Depending on the duration of uninterrupted work, asthenopia and musculoskeletal problems may occur. For this reason, appropriate measures for reduction of the workload must be taken:
Ϙ Optimum workplace layout
Ϙ Frequent changes of activity
Ϙ Thorough training of the personnel, giving
consideration to ergonomic and organiza­tional aspects
The ergonomic design and construction of the Leica microscopy series are intended to reduce the exertion of the user to a minimum.
Danger of infection
Direct contact with eyepieces is a
potential transmission method for
bacterial and viral infections of the eye.
The risk can be kept to a minimum by
using personal eyepieces for each indivi-
dual or detachable eyecups.
Dangers during use
Ϙ The Leica DM750 microscope may only be
connected to a grounded socket.
Ϙ The Leica DM750 microscope may not be
operated unless it is in proper functioning condition.
LED Klasse 1 /LED 1 Laser Product
The illumination meets laser class 1 require­ments in accordance with IEC 60825-1 "Safety of Laser devices".
Caution: Hazardous radiation may occur if
controls, adjustment jigs, or procedures other than those outlined in this User Manual are used!
Class 1
LED Product
LED class 1
Leica DM750 Manual 12
Page 13

Information for the person responsible for the instrument

Information for the person responsible for the instrument
Ϙ Ensure that the Leica DM750 microscope is
used only by qualified personnel.
Ϙ Ensure that this user manual is always
available at the Leica DM750 microscope.
Ϙ Carry out regular inspections to make
certain that the authorized users are adhering to safety requirements.
Ϙ When instructing new users, do so
thoroughly and explain the meanings of the warning signs and messages.
Ϙ Assign individual responsibilities for star-
ting, operating and servicing the instru­ment and monitor the observance of these responsibilities.
Ϙ Do not use the Leica DM750 microscope
unless it is in perfect condition.
Ϙ Inform your Leica representative or Leica
Microsystems (Schweiz) AG, Industry Division, 9435 Heerbrugg, Switzerland, immediately of any product defect that could potentially cause injury or harm.
Ϙ If you use accessories made by third-party
manufacturers with the Leica DM750 microscope, be sure that each such manu­facturer confirms the safety-engineering, harmless usability of the product and observe the product's user manual.
Ϙ Modifications and maintenance of the Leica
DM750 microscope may only be performed by professionals expressly authorized by Leica.
Ϙ Only original Leica replacement parts may
be used in servicing the product.
Ϙ After service work or technical modifica-
tions, the unit must be reconfigured with observance to our technical requirements.
Ϙ If the unit is modified or serviced by
unauthorized persons, is improperly maintained (as long as maintenance was not carried out by us), or is handled impro­perly, Leica will not accept any liability.
Ϙ The electric installation in the building
must conform to the national standard, e.g. current-operated ground leakage protec­tion (fault-current protection) is suggested.
Leica DM750 Manual 13
Page 14

Care instructions

General instructions
Ϙ Protect the Leica DM750 microscope
against damp, vapors, acids, alkalis, and corrosive substances. Do not store chemi­cals in the vicinity.
Ϙ Protect the Leica DM750 microscope from
oil and grease. Never grease or oil mechani­cal parts or sliding surfaces.
Ϙ Follow the instructions of the disinfectant
manufacturer.
Ϙ It is advisable to enter a service agreement
with Leica Service.
Cleaning coated parts and plastic parts
Ϙ Dust and dirt particles should be removed
with a soft brush or lint-free cotton cloth.
Ϙ Remove coarse debris with a moistened
disposable cloth.
Ϙ Clinging dirt can be removed using petro-
leum ether or alcohol
Ϙ Acetone, xylene or nitro-containing thin-
ners must NOT be used.
Ϙ Never use chemicals to clean colored
surfaces or accessories with rubberized parts. This could damage the surfaces, and specimens could be contaminated by abra­ded particles.
Cleaning glass surfaces
Ϙ Remove dust using a dry and grease-free
brush made from hair, by blowing with bellows, or by using a vacuum.
Ϙ Remove dirt using a clean cloth that has
been moistened with distilled water.
Ϙ Remove tough dirt using alcohol or petro-
leum ether.
Leica DM750 Manual 14
Page 15

Accessories, maintenance and repair

Accessories
Only the following accessories may be used with the Leica DM750 microscope:
Ϙ The Leica accessories described in this user
manual.
Ϙ Other accessories, provided that these have
been expressly approved by Leica as being technically safe in this context.
Maintenance
Ϙ The Leica DM750 microscope is basically
maintenance-free. To ensure that it always operates safely and reliably, we recommend that you take the precaution of contacting the responsible service organization.
You can arrange for periodic inspections
or, if appropriate, conclude a mainte-
nance contract with them.
Ϙ It is advisable to enter a service agreement
with Leica Service.
Ϙ For maintenance and repair, only OEM
spare parts may be used.
Repairs and service work
Ϙ Only original Leica Microsystems spare
parts may be used.
Ϙ Before opening the instruments, switch off
the power and unplug the power cable.
Ϙ Avoid contact with powered electrical
circuits, which can lead to injury.
Service address
In case of problems, please contact us as follows:
Leica Microsystems (Schweiz) AG Industry Division Max Schmidheiny Strasse 201 CH-9435 Heerbrugg (Switzerland) Phone +41 71 726 33 28
Leica DM750 Manual 15
Page 16

Electrical data and ambient conditions

Fuse Replacement
Unplug the instrument before changing any fuses. The Leica
DM750 has two fuses, which are located behind the power cord
receptacle.
Only use the following fuse types: 5×20 mm, 1 A/250V, fast-acting
fuse (# 13RFAG30003)
Electrical data
Input: 100-240 V, 50/60 Hz, 5 W (3 W LED)
General safety notes This instrument of safety class 1 has been built and tested in accordance with the following safety requirements for electrical equipment for measu­rement, control, and laboratory use:
EN 61010-1: 2002-08 EN 61010-2-101: 2008-06 IEC 61010-1: 2010-06 EN 61326-1: 2006-10 EN 61326-2-6: 2006-10
EN 55011: 2007+A2: 2010-05 EN 60825-1: 2008-05 IEC 60825-1: 2007-03 LED Class 1
In order to maintain this condition and to ensure safe operation,
the user must follow the instructions and warnings contained in
this instruction manual.
Environment
Temperature for use +10 °C … +40 °C
Storage temperature -20 °C … +52 °C
+50 °F to +104 °F
Manipulation shock 25 mm on 50 mm hard wood
Transport shock (unpacked) 100 g / 6 ms
Transport shock (packed) 800 mm free fall
Transport vibrations (unpacked) 5–200 Hz / 1.5 g
Air pressure during use and storage: 500–1060 mbar
Humidity during use and storage: 20–90%
Installation Category II (Overvoltage Category)
Pollution degree 2
Leica DM750 Manual 16
Page 17

The Leica DM750

Leica DM750 Manual 17
Page 18

Introduction

Thank you for purchasing the Leica DM750 Compound Microscope from Leica Microsys­tems. This model’s exclusive design features and full range of accessories make it a truly versatile, high quality-instrument for educatio­nal and routine clinical applications.
The instrument is to be used only as
described. Hazards to personnel may
result if used improperly.
Leica DM750 Manual 18
Page 19

Unpacking

Ϙ Carefully remove the microscope and any
components from the packing carton.
Ϙ Verify that all components are intact.
Ϙ Check the components against the planned
configuration.
Ϙ Optional items such as contrast accessories,
camera adapters, cameras, and carrying cases are not shipped as part of the stan­dard equipment. These items are delivered in separate packages.
Ϙ Please do not discard any of the packing
materials. They should be used for safely storing and transporting the instrument should the need arise.
Leica DM750 Manual 19
Page 20

Ready!

Leica DM750 Manual 20
Page 21

Substage Illumination

Substage Illumination
The Leica DM750 is available with two
types of Substage Illumination. Identify which type of Illumination you have, as this will be important to know later.
Type 1: Standard illumination
Adjustable condenser centering with provided tool.
Type 2: Koehler illumination
Adjustable condenser centering with thumbscrews and adjustable Koehler field diaphragm.
Leica DM750 Manual 21
Page 22

Attaching the viewing Tubes

Tools used
Ϙ Allen key
There are two types of viewing tube.
Identify which viewing tube you have:
Type 1: Leica EZ Viewing Tube with integrated eyepieces.
Type 2: Standard viewing tube with separate eyepieces
1. Loosen the setscrew (on top of the stand) using the Allen key provided.
2. Set the dovetail into the tripod mount and tighten the setscrew carefully (without using excessive force). This precisely posi­tions the viewing tube onto the optical axis of the microscope, regardless of the selec­ted rotation.
Leica DM750 Manual 22
Page 23

Leica EZ Viewing Tube – integrated eyepieces

Leica EZ viewing tube with integrated eyepieces.
To rotate the Leica EZ viewing tube,
either loosen the set screw on the stand or replace the set screw with the (optional) captive thumbscrew.
1. In order to use the captive wingscrew, remove the setscrew that was delivered with the stand.
Make sure that the wingscrew part
of the captive wingscrew has been
completely separated before you use the stand:
2. Reinstall the Leica EZ Viewing Tube onto the stand.
3. Tighten the captive wingscrew using the wrench included in the delivery package.
Leica DM750 Manual 23
Page 24
Leica EZ Viewing Tube – integrated eyepieces (cont.)
You can now safely rotate the Leica EZ
viewing tube by loosening the captive thumbscrew, rotating the viewing tube, and tightening the captive thumbscrew again.
The eyepieces are integrated into the
Leica EZ Viewing Tube and preset; there­fore, there is no need to adjust or install the eyepieces.
Continue with the "Eyecups" section on page
26.
Leica DM750 Manual 24
Page 25

Leica Standard Viewing Tube – separated eyepieces

Standard viewing tubes; tubes do not include eyepieces yet
The standard viewing tube has a rota-
table dovetail. Therefore, you can now rotate the standard viewing tube freely in any orientation.
1. Insert the eyepieces into the tubes.
2. Secure the eyepieces in the tubes by tighte­ning the silver screws at the bottom of the tubes using a standard Phillips screwdriver (not included in the delivery package).
The eyepieces will still rotate, but they
will be captive in the eyetubes.
Leica DM750 Manual 25
Page 26

Eyecups

If you wear eyeglasses for microscope
viewing, keep the rubber eyecups folded down. If you do not wear eyeglasses, you may find it useful to unfold the rubber eyecups in order to help block out ambient room light.
If you have purchased a standard
microscope configuration, you will notice that the objectives are already installed on the nosepiece and the substage condenser is already installed on the stand. In this case, go to section "Operation" on page 30. If you purchased your Leica DM750 in the form of individual components rather than the stan­dard configuration, please continue with the "Installing Objectives" section on page 27.
Leica DM750 Manual 26
Page 27

Installing Objectives

Installing Objectives
When rotating the objective nosepiece,
always use the knurled ring on the
objective nosepiece.
As you rotate the nosepiece clockwise, attach the objectives by screwing them into the nosepiece holes starting with the lowest magnification and advancing to the highest magnification.
Leica DM750 Manual 27
Page 28

Installing the Substage Condenser

Substage Condenser
The Leica DM750 has an open substage
condenser mount, therefore the conden-
ser needs to be installed.
The substage condenser mount is open
1. Move the specimen stage upwards as far as it will go by using the coarse focusing knob on the side of the microscope.
2. Move the condenser holder into the lowest position by using the condenser focusing knob on the left side of the stage mount.
3. Unscrew the two setscrews (or wings­crews for Koehler stands) on the condenser holder.
Leica DM750 Manual 28
Page 29
Installing the Substage Condenser (cont.)
4. Push the substage condenser under the specimen stage into the holder by alig­ning the locating pin on the bottom of the condenser with the slot on the rear side of the fork.
5. Move the condenser into the highest posi­tion by using the condenser focusing knob on the left side of the stage mount.
6. Tighten the two setscrews using the tool included in the delivery package (or, for a Koehler stand, tighten the two wingscrews) so that the upper lens of the condenser is centered under the objective in working position and the substage condenser is thus roughly centered.
You will center the condenser more accu-
rately when you get to the "Complete
Condenser Centering" section on page 32.
Leica DM750 Manual 29
Page 30

Set!

Leica DM750 Manual 30
Page 31

Turning on the microscope

Work Surface
Always use your microscope on a hard,
stable surface.
Power Cord
If the power cord is not already atta-
ched, attach it securely to the back of
the microscope.
Setting the Illumination Intensity
Set the illumination to the lowest
setting to start with, using the control­ler on the bottom left of the stand. The illumi­nation control knob allows you to adjust the intensity of light produced by the illumination system.
Plug in and Turn on the Microscope
1. Plug the power cable of the microscope into a corresponding grounded socket. A grounded 3-wire cord is provided.
2. Switch on the microscope using the switch at the bottom right of the microscope stand.
Leica DM750 Manual 31
Page 32

Complete Condenser Centering

If you have purchased a standard Leica
DM750 configuration, the condenser has already been pre-centered by Leica Microsys­tems.
1. Open the condenser aperture by rotating
the knurled ring on the condenser to the right.
Make sure that the condenser is in the
highest position.
2. Write an "X" on a piece of paper in the size of a business card and place it onto the light output of the microscope stand in such a way that the "X" is centered over the illumination.
3. Look at the X through the eyepieces and center it in the field of view by tightening the setscrews with the tool included in the delivery package (or, for a Leica DM750 with Koehler illumination, tighten the knurled screws).
Not centered
Centered
Leica DM750 Manual 32
Page 33

Using the Condenser

Using the Condenser
The condenser is furnished with an iris
diaphragm, which can be adjusted to match the effective numerical aperture of the objective
1. To open and close this diaphragm, simply turn the knurled ring on the condenser to the right or left so that the line on the ring is aligned with the objective magnification used.
Match the line of the rotating ring with the objective magnification in use.
2. Open the iris diaphragm of the condenser completely at first by turning the conden­ser ring all the way to the right.
Leica DM750 Manual 33
Page 34

Prepare to View a Specimen Slide

1. Position a specimen slide on the specimen stage by sliding it under the slide grips.
Slide grips hold the slide in place.
Slide grips
2. Using the X/Y-stage control, position the specimen slide such that a part of the speci­men is under the objective used.
Leica DM750 Manual 34
Page 35

Focusing

1. Rotate the objective nosepiece (using the knurled ring) in such a way that the objec­tive with the lowest magnification level is rotated into the working position.
2. Move the specimen stage upwards by turning the coarse focusing knob as far as it will go to the maximum position.
Coarse focus adjustment knob
3. Look into the eyepieces and adjust the illu­mination intensity to a level that is comfor­table for your eyes.
The stand of the Leica DM750 has been
calibrated at the factory so that the focus can be found from this position within 1.5 rota­tions of the fine focus.
4. Bring the specimen into focus using the
fine focusing knob.
Leica DM750 Manual 35
Page 36

Viewing Tube Adjustment

Adjust the eyetubes
1. Adjust the tubes to your interpupillary distance. Fold or unfold the eyetubes to decrease or increase the distance between the eyepieces until you see one illuminated circle.
The viewing tubes maintain a constant
tube length for all interpupillary settings. This means that a change of interpupillary distance does not affect parfocality, magnifica­tion, or calibrations that depend on magnifica­tion.
Leica EZ Viewing Tube
Ϙ If you are using a Leica EZ viewing tube,
which has the eyepieces integrated with the eyetubes, no additional adjustments are necessary. Be sure you are wearing your eyeglasses or contact lenses.
Ϙ If you have a Leica DM750 with stan-
dard illumination (without Koehler field diaphragm), continue with the "Oil Immer­sion Technique" section on page 40.
Ϙ If you have a Leica DM750 with Koehler
field diaphragm, continue with the "Koeh­ler Configuration" section on page 38.
Standard Viewing Tube with Two Fixed Eyepieces
Ϙ If you are using a standard viewing tube
with two fixed eyepieces (no focusing eyepieces), no additional adjustments necessary. Be sure you are wearing your eyeglasses or contact lenses.
Ϙ If you have a Leica DM750 with stan-
dard illumination (without Koehler field diaphragm), continue with the "Oil Immer­sion Technique" section on page 40.
Ϙ If you have a Leica DM750 with Koehler
field diaphragm, continue with the "Koeh­ler Configuration" section on page 38.
Standard Viewing Tube with One or Two Focusing Eyepieces
If you are using a standard viewing tube with one or two focusing eyepieces, you need to make some adjustments.
1. Set the focusing eyepieces to "0".
Leica DM750 Manual 36
Page 37
Viewing Tube Adjustment (cont.)
If you are comfortable wearing your
corrective lenses (contact lenses or eyeglasses) for microscope viewing, leave them on and your adjustments will be minimal.
2. Using the fine focus adjusting knob, focus
on the specimen while looking through only one of the eyepieces (when using one eyepiece that is focusable and one that is not focusable, look through the one that is not focusable). To help focus, cover or close the other eye.
3. Now look with the other eye just through the other eyepiece (focusing eyepiece). This time, focus the specimen by using the focu­sing capability in the focusing eyepiece.
When doing so, do not change the
height of the specimen stage.
4. Grip the knurled ring on the focusing eyepiece with one hand and rotate the top of the eyepiece with the other hand until the specimen is in focus for this eye and this focusing eyepiece. This corrects for any
vision differences between your right eye and left eye.
5. Now, switch to an objective with a high magnification level (no oil objective) and bring the microscope into focus while looking through the eyepieces with both eyes.
The higher magnifications have a shallo-
wer depth of field. Therefore, after focus­ing with a high magnification, you will find that when you change to lower magnifications, you only have to adjust the fine focus slightly, if at all.
Ϙ If you use a DM750 with standard illumi-
nation, continue with the "Oil Immersion Technique" section on page 40.
Ϙ If your DM750 has a Koehler configuration,
please continue with the next section, "Koehler Configuration", on page 38.
Leica DM750 Manual 37
Page 38
Koehler Configuration
If your Leica DM750 is fitted with a field diaphragm for Koehler Illu-
mination, use the following procedure to assure condenser cente-
ring and focus.
1. Connect the Koehler field diaphragm to the base of the microscope so that the diaphragm leaves are in the field of view when you look through the eyepieces.
Close the Koehler field diaphragm
Close the Koehler field diaphragm
2. Focus the leaves of the field diaphragm using the condenser focusing knob on the left side of the stage mount.
Condenser focusing knob
Leaves are in sharp focus
Leica DM750 Manual 38
Page 39
Koehler Configuration (cont.)
3. Turn the condenser centering wingscrews simultaneously to center the image of the field diaphragm.
4. Open the field diaphragm until the diaphragm leaves are just outside of the field of view.
Leica DM750 Manual 39
Page 40

Oil Immersion Technique

1. Search for the area on the specimen slide that you want to study.
2. Move the specimen stage down to the lowest position using the coarse adjust­ment knob.
3. Add a drop of immersion oil from Leica to the area of the specimen slide you are studying.
4. Rotate the oil immersion objective (the objective labeled "OIL") to the working position.
Leica DM750 Manual 40
Page 41
Oil Immersion Technique (cont.)
5. Slowly move the specimen stage upwards using the coarse adjustment knob until the oil drop on the slide comes into direct contact with the lens of the oil immersion objective.
6. Hold the knurled ring on the objective turret and swing the objective back and forth to eliminate air bubbles. Then bring the oil objective into the final position, so that the oil drop is between the front lens of the objective and the specimen slide.
7. Look through the microscope and slowly rotate the fine adjustment knob, so that the specimen stage moves upwards until the specimen is in focus.
8. If you are finished working with the oil immersion objective, clean the front side of the objective, the specimen slide, and all other surfaces that have come into contact with the oil, following the instructions in the "Care of the Microscope" section on page 46.
Leica DM750 Manual 41
Page 42
Time Delay Shutoff
The Leica DM750 is equipped with a time
delay shutoff capability, which automati­cally turns the illumination off after 2 hours of no changes in the intensity control.
Ϙ All four-position nosepiece stands have
the time delay shutoff enabled as a default (mostly educational applications).
Ϙ All five-position nosepiece stands have the
time delay shutoff disabled as a default (mostly clinical applications).
Changing the status of the time delay shutoff
1. Rotate the intensity regulator to the lowest level.
2. Switch on the instrument.
3. Rotate the intensity regulator to the high­est level and then back to the lowest level within five seconds.
Ϙ The LED Illumination will flash to indicate
the time delay shutoff status was changed.
Ϙ The LED Illumination will flash two times
slowly then stay on when you disable the Time Delay Shutoff.
Ϙ The LED Illumination will flash three times
quickly then stay on when you enable the Time Delay Shutoff.
When you turn the power off and then
turn the power on again, the system will be in the last Time Delay Shutoff status (Enabled or Disabled) and you will not see any blinking.
Leica DM750 Manual 42
Page 43
Go!
Leica DM750 Manual 43
Page 44

Ready! Set! Go!

Now all you need to do is change objectives, set the condenser aperture (and field diaphragm if you have a DM750 Koehler Stand) appropriately for the objective magnification you are using, and enjoy the view!
Leica DM750 Manual 44
Page 45

Care of the Microscope

Leica DM750 Manual 45
Page 46

General maintenance

Storage and handling
Always use both hands to carry the
microscope. There is a handle on the back of the microscope and an undercut in the front for this purpose.
Ϙ The cord wrap allows you to wrap the cord
in such a way that only the length you need is extended.
Ϙ Keep all optical components clean. Clean-
liness is important for maintaining good optical performance.
Ϙ The microscope should always be covered
with the plastic dust cover (provided with the instrument) when it is not in use.
Instructions for cleaning
Ϙ If any optical surface becomes coated with
dust or dirt, clean the surface by blowing it off with a syringe or brushing it off with a camel hair brush before attempting to wipe the surface clean.
Ϙ Optical surfaces should be cleaned using
a lint-free cloth, lens cloth or cotton swab moistened with a commercially available glass cleaner.
Ϙ It is very important to avoid the excessive
use of solvents, so use them sparingly. The lint-free cloth, lens tissue or cotton swab should be moistened with solvent, but not be wet enough for the solvent to seep around the lens.
Leica DM750 Manual 46
Page 47
General maintenance (cont.)
Ϙ No part of the microscope is quite so vulne-
rable to collecting dirt, dust, and oil as the front lens of the objective. Whenever you encounter lack of contrast, cloudiness or poor definition, carefully check the condi­tion of the front lens with a magnifier.
Ϙ The lower magnification objectives have
fairly large front lenses and can be cleaned with a cloth or lens tissue wrapped around a finger and moistened with methanol.
Ϙ Cleaning 40× and 100× objectives requires
more care. Note: To attain the typical flat­ness for objectives with higher magnifica­tion, the objective has a small concave front lens with a very small radius and slight curvature. The surface of this front lens can be readily cleaned with a toothpick cove­red with a cotton tip, or with a small cotton swab. Soak the cotton with isopropanol and squeeze it out until it is almost dry again. Wipe the front lens lightly without applying undue force or scrubbing action.
Make sure that the cotton tip contacts the concave lens surface. Check the objective with a magnifier after cleaning.
Ϙ If you need to remove the microscope’s
viewing body, be careful not to accidentally touch the outer lens surface (located on the underside of the body). Fingerprints on this surface will reduce image clarity. This lens can be cleaned in the same manner as objectives and eyepieces.
Mechanical Maintenance
Ϙ Use the dust cover. It is the most impor-
tant way to keep the microscope in good mechanical and physical condition.
Ϙ The Leica DM750 has a durable finish. All
non-optical parts of the instrument can be cleaned with isopropanol or soap and water. Avoid the use of all other organic solvents when cleaning the instrument.
Illumination
Ϙ The Leica DM750 uses LED illumination.
Therefore, no lamp changing is required for the life of the microscope.
Leica DM750 Manual 47
Page 48

Troubleshooting

Leica DM750 Manual 48
Page 49

Troubleshooting

Stand
Make sure that the socket has voltage.
The microscope does not respond.
Focus
The specimen cannot be brought into focus.
Dark field
No definite DF contrast is possible.
The image is not uniformly illuminated. The objective magnification is too weak. Use a higher magnification. Unwanted stray light. Clean the specimen and neighboring lens surfaces.
Polarization
Check the cable connections. Make sure that the stand is connected to the power supply. Check whether the fuse is defective and replace it if necessary (see page 16).
Use the correct immersion medium. Lay the specimen with the cover glass toward the top. Make sure that the cover slip thickness is correct and that it meets the specifica­tions on the objective.
Make sure that a DF objective is being used. The objective aperture is too high (maximum 0.75/ 1.10); if necessary, reduce the objective aperture through the iris diaphragm on the objective. Check the condenser centering. Open the aperture diaphragm completely.
Leica DM750 Manual 49
Page 50
Troubleshooting (continued)
Polarization contrast cannot be adjusted.
Phase contrast
Phase contrast cannot be adjusted.
Specimen Stage
Positioning range of the stage in the x-direc­tion decreases after working a long time.
Bring the polarizer and analyzer into cross position until they reach maximum darkness (without specimen).
The specimen is too thick, too thin, or too brightly stained. Refractive indexes of the mounting medium and specimen are identical, so that there is no phase jump. The cover glass is not placed uniformly. Check that the correct light ring is positioned. Check the centering of the light rings. Check the condenser centering. Open the aperture diaphragm completely.
Move the mechanical stage with coaxial drive all the way to the left. Manually press the screw that holds the mechanical stage even farther to the left, as far as it will go. Then move the mechanical stage with coaxial drive all the way to the right. Manually press the screw that holds the mechanical stage even farther to the right, as far as it will go.
Leica DM750 Manual 50
Page 51

Dimensions

Leica DM750 Manual 51
Page 52
Leica DM750 Manual 52
Page 53
Leica DM750 Manual 53
Page 54
Leica DM750 Manual 54
Loading...