WATERFLUX 3000
1.1 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
This electromagnetic flowmeter is designed exclusively to measure the flow of drinking water,
raw water and irrigation water.
WARNING!
If the device is not used according to the operating conditions (refer to chapter Technical data),
the intended protection could be affected.
1.2 Certification
SAFETY INSTRUCTIONS
1
CE marking
The manufacturer certifies successful testing of the product by applying the CE marking.
This device fulfils the statutory requirements of the relevant EU directives.
For full information of the EU directives and standards and the approved certifications, please
refer to the EU Declaration of Conformity or the website of the manufacturer.
Other approvals and standards
• Measuring Instruments Directive 2014/32/EU, Annex III (MI-001), Annex VI (MI-004)
• NAMUR NE 21/04
For more information, please refer to the dedicated documentation.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
5
Page 6
1
SAFETY INSTRUCTIONS
1.3 Safety instructions from the manufacturer
1.3.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
WATERFLUX 3000
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.3.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
6
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 7
WATERFLUX 3000
1.3.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation or operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.3.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of icons as shown below.
SAFETY INSTRUCTIONS
1
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
7
Page 8
1
SAFETY INSTRUCTIONS
1.3.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
WATERFLUX 3000
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
i
RESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.4 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
8
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 9
WATERFLUX 3000
CAUTION!
The WATERFLUX 3000 can be combined with the mains powered IFC 050, IFC 100 and IFC 300
signal converter and with the battery powered IFC 070 signal converter.
This handbook only describes the combination with the IFC 050, IFC 100 and IFC 300 signal
converter.
2.1 Scope of delivery
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
The remote version will arrive in two cartons. One carton contains the converter and one carton
contains the sensor.
DEVICE DESCRIPTION
2
Figure 2-1: Scope of delivery
1 Ordered flowmeter
2 Product documentation
3 Factory calibration report
4 CD-ROM with product documentation in available languages
5 Signal cable (remote versions only)
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
9
Page 10
2
DEVICE DESCRIPTION
2.2 Device description
Electromagnetic flowmeters are designed exclusively to measure the flow and conductivity of
electrically conductive, liquid media.
Your measuring device is supplied ready for operation. The factory settings for the operating
data have been made in accordance with your order specifications.
INFORMATION!
Product specific information and extensive product specification is available using PICK, the
Product Information Center KROHNE web-tool.
PICK can be found via the service menu button on the KROHNE.com website.
The following versions are available:
• Compact version (the signal converter is mounted directly on the measuring sensor)
• Remote version (measuring sensor with connection box and a signal converter in a remote
(field) housing)
WATERFLUX 3000
10
Figure 2-2: Device versions
1 Remote version
2 Compact version with signal converter IFC 300
3 Compact version with signal converter IFC 100 (0°)
4 Compact version with signal converter IFC 100 (45°)
5 Compact version with signal converter IFC 100 (10°) Stainless steel
6 Compact version with signal converter IFC 050 (10°)
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 11
WATERFLUX 3000
2.3 Nameplate (example)
INFORMATION!
Check the device nameplate to ensure that the device is delivered according to your order.
Additional information (e.g. correct supply voltage), can be found in the documentation of the
signal converter.
Figure 2-3: Example of nameplate
1 Name and address of manufacturer
2 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
3 Meter constant, diameter, wetted materials, protection class
DEVICE DESCRIPTION
2
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
11
Page 12
3
INSTALLATION
3.1 General notes on installation
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packaging.
• Storage temperature: -50...+70°C / -58...+158°F
WATERFLUX 3000
3.3 Transport
Signal converter
• No special requirements.
Compact version
• Do not lift the device by the signal converter housing.
• Do not use lifting chains.
• To transport flange devices, use lifting straps. Wrap these around both process connections.
12
Figure 3-1: Transport
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 13
WATERFLUX 3000
3.4 Pre-installation requirements
Make sure that you have all necessary tools available:
• Allen key (4 mm)
• Small screwdriver
• Wrench for cable glands
• Wrench for wall mounting bracket (remote version only)
• Torque wrench for installing flowmeter in pipeline
3.5 General requirements
INFORMATION!
The following precautions must be taken to ensure reliable installation.
•
Make sure that there is adequate space to the sides.
•
Protect the signal converter from direct sunlight and install a sun shade if necessary.
•
Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.
•
Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-64.
INSTALLATION
3
3.5.1 Vibrations
Figure 3-2: Avoid vibrations
3.5.2 Magnetic field
Figure 3-3: Avoid magnetic field
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
13
Page 14
3
INSTALLATION
3.6 Installation conditions
CAUTION!
To prevent damage to the Rilsan
carefully. Take precautions during transport and installation to protect the in- and outlet of the
sensor.
3.6.1 Inlet and outlet
WATERFLUX 3000
®
coating, the WATERFLUX 3000 sensor must be installed
Figure 3-4: Minimal inlet and outlet
1 Inlet: t 0 DN
2 Outlet: t 0 DN
3.6.2 T-section
Figure 3-5: Distance behind a T-section
1 t 0 DN
3.6.3 Open discharge
14
Figure 3-6: Installation in front of an open discharge
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 15
WATERFLUX 3000
3.6.4 Bends
Figure 3-7: Installation in bending pipes
INSTALLATION
3
3.6.5 Pump
Figure 3-8: Installation in bending pipes
Figure 3-9: Installation behind a pump
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
15
Page 16
3
INSTALLATION
3.6.6 Control valve
Figure 3-10: Installation in front of a control valve
3.6.7 Air venting and vacuum forces
WATERFLUX 3000
Figure 3-11: Air venting
1 t 5 m
2 Air ventilation point
Figure 3-12: Vacuum
1 t 5 m
16
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 17
WATERFLUX 3000
3.6.8 Mounting position
Figure 3-13: Mounting position
• Install flow sensor in line with the pipe axis.
• Pipe flange faces must be parallel to each other.
INSTALLATION
3
3.6.9 Flange deviation
CAUTION!
Max. permissible deviation of pipe flange faces:
- L
L
max
Figure 3-14: Flange deviation
1 L
max
2 L
min
d 0.5 mm / 0.02"
min
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
17
Page 18
3
INSTALLATION
3.7 IP 68
The WATERFLUX 3000 flow sensor is rated IP68 (NEMA 4X/6P). It is suitable for submersion in
flooded measurement chambers and for subsurface installation.
WATERFLUX 3000
Figure 3-15: IP68 versions
1 Submersible
2 Buried
1
2
18
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 19
WATERFLUX 3000
3.8 Mounting
3.8.1 Torques and pressures
The maximum pressure and torque values for the flowmeter are theoretical and calculated for
optimum conditions and use with carbon steel flanges.
INSTALLATION
3
Figure 3-16: Tightening of bolts
Tightening of bolts
• Always tighten the bolts uniformly and in diagonally opposite sequence.
• Do not exceed the maximum torque value.
• Step 1: Apply approx. 50% of max. torque given in table.
• Step 2: Apply approx. 80% of max. torque given in table.
• Step 3: Apply 100% of max. torque given in table.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
19
Page 20
3
INSTALLATION
WATERFLUX 3000
Nominal size
DN [mm]
25PN 164 x M 1212
40PN 164 x M 1630
50PN 164 x M 1636
65PN 168 x M 1650
80PN 168 x M 1630
100PN 168 x M 1632
125PN 168 x M 1640
150PN 108 x M 2055
150PN 168 x M 2055
200PN 108 x M 2085
200PN 1612 x M 2057
250PN 1012 x M 2080
250PN 1612 x M 24100
300PN 1012 x M 2095
300PN 1612 x M 24136
350PN 1016 x M 2096
400PN 1016 x M 24130
450PN 1020 x M 24116
500PN 1020 x M 24134
600PN 1020 x M 27173
1 The torque values also depend on variables (temperature, bolt material, gasket material, lubricants, etc.)
outside the control of the manufacturer. Therefore these values should be regarded as indicative only.
Pressure
rating
BoltsMax. torque
1
[Nm]
20
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 21
WATERFLUX 3000
INSTALLATION
3
Nominal size
[inch]
11504 x 1/2"4
1½1504 x 1/2"11
21504 x 5/8"18
2.51508 x 5/8"27
31504 x 5/8"33
41508 x 5/8"22
51508 x 3/4"33
61508 x 3/4"48
81508 x 3/4"66
1015012 x 7/8"74
1215012 x 7/8"106
14150 212 x 1"87
16150 216 x 1"84
18150 216 x 1 1/8"131
20150 220 x 1 1/8"118
24150 220 x 1 1/4"166
1 The torque values also depend on variables (temperature, bolt material, gasket material, lubricants, etc.)
outside the control of the manufacturer. Therefore these values should be regarded as indicative only.
2 No full rating (max. 150 psi / 10 bar).
Flange class
[lb]
BoltsMax. torque
[lbs.ft]
1
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
21
Page 22
4
ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
WATERFLUX 3000
4.2 Grounding
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
Figure 4-1: Grounding
INFORMATION!
Grounding without grounding rings. The flow sensor is equipped with a reference electrode.
4.3 Connection diagrams
22
INFORMATION!
For the connection diagrams please refer to the documentation of the applicable signal
converter.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 23
WATERFLUX 3000
5.1 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
5.2 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
5.3 Returning the device to the manufacturer
SERVICE
5
5.3.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
WARNING!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of the personnel, the manufacturer may only handle, test and repair returned devices
that have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
WARNING!
If the device has been operated with toxic, caustic, radioactive, flammable or water-endangering
products, you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that it is safe to handle and stating the
product used.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
23
Page 24
5
SERVICE
5.3.2 Form (for copying) to accompany a returned device
CAUTION!
To avoid any risk for our service personnel, this form has to be accessible from outside of the
packaging with the returned device.
Company:Address:
Department:Name:
Tel. no.:Fax no. and/or Email address:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
This medium is:radioactive
water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
WATERFLUX 3000
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the
device when it is returned.
Date:Signature:
Stamp:
5.4 Disposal
LEGAL NOTICE!
Disposal must be carried out in accordance with legislation applicable in your country.
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
According to the directive 2012/19/EU, the monitoring and control instruments marked with the
WEEE symbol and reaching their end-of-life must not be disposed of with other waste
The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or
send them back to our local organisation or authorised representative.
must not be disposed of with other waste.
must not be disposed of with other wastemust not be disposed of with other waste
24
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 25
WATERFLUX 3000
6.1 Measuring principle
An electrically conductive fluid flows inside an electrically insulated pipe through a magnetic
field. This magnetic field is generated by a current, flowing through a pair of field coils.
Inside of the fluid, a voltage U is generated:
U = v * k * B * D
U = v * k * B * D
U = v * k * B * DU = v * k * B * D
in which:
v = mean flow velocity
k = factor correcting for geometry
B = magnetic field strength
D = inner diameter of flowmeter
The signal voltage U is picked off by electrodes and is proportional to the mean flow velocity v
and thus the flow rate Q. A signal converter is used to amplify the signal voltage, filter it and
convert it into signals for totalizing, recording and output processing.
TECHNICAL DATA
6
Figure 6-1: Measuring principle
1 Field coils
2 Magnetic field
3 Electrodes
4 Induced voltage (proportional to flow velocity)
5 Rectangular cross section
The minimal height of the measuring tube decreases the distance between the field coils (1),
resulting in a stronger and more homogeneous magnetic field (2). In addition, the mean flow
velocity v increases due to the rectangular and reduced cross section. The large electrode
spacing (D) and the increased flow velocity results in a higher magnetic signal voltage, also in
the presence of a low flow rate.
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
25
Page 26
6
TECHNICAL DATA
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principleFaraday's law of induction
Application rangeElectrically conductive fluids
Measured value
Measured value
Measured valueMeasured value
Primary measured valueFlow velocity
Secondary measured
value
Design
FeaturesUnique rectangular flow tube design providing improved flow profile and
WATERFLUX 3000
Volume flow
signal to noise ratio resulting in highest accuracy and large turndown ratio
Rilsan® polymer coated flow tube approved for drinking water
No internal or moving parts
Built-in reference electrode
Modular constructionThe measurement system consists of a flow sensor and a signal converter.
Compact versionWith IFC 050 converter: WATERFLUX 3050 C
Remote versionIn wall (W)mount version with IFC 050 converter: WATERFLUX 3050 W
It is available as compact and as separate version. More information about
the signal converter can be found in the relevant documentation.
With IFC 100 converter: WATERFLUX 3100 C
With IFC 300 converter: WATERFLUX 3300 C
In wall (W)mount version with IFC 100 converter: WATERFLUX 3100 W
In field (F), wall (W) or rack ( R) mount version with IFC 300 converter:
WATERFLUX 3300 F, W or R
26
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 27
WATERFLUX 3000
Measuring accuracy
Maximum measuring
error
RepeatabilityDN25...300 / 1...12" ; ±0.1% (v >0.5m/s / 1.5 ft/s)
Calibration / VerificationStandard:
MID Annex MI-001
(Directive 2014/32/EU)
TECHNICAL DATA
IFC 050: down to 0.5% of the measured value ±1 mm/s
IFC 100: down to 0.3% of the measured value ±1 mm/s
IFC 300: down to 0.2% of the measured value ±1 mm/s
The maximum measuring error depends on the installation conditions.
For detailed information refer to
DN350...600 / 14...24"; ±0.2% (v >0.5m/s / 1.5 ft/s)
Standard:
Standard:Standard:
2 point calibration by a direct volume comparison.
Option (for DN25...600):
Option (for DN25...600):
Option (for DN25...600):Option (for DN25...600):
Verification to Measurement Instrument Directive (MID), Annex MI-001.
Standard: Verification at Ratio (Q3/Q1) = 80, Q3 t 2 m/s / 6.6 ft/s
Optional: Verification at Ratio (Q3/Q1) > 80
Only in combination with the IFC 300 signal converter
EC-Type examination certificate to MID Annex MI-001
EC-Type examination certificate to MID Annex MI-001
EC-Type examination certificate to MID Annex MI-001EC-Type examination certificate to MID Annex MI-001
Only in combination with the IFC 300 signal converter
Diameter range: DN25...600 / 1...24"
Minimum straight inlet flow: 0 DN
Minimum straight outlet flow: 0 DN
Forward and reverse (bi-directional) flow
Orientation: any
Ratio (Q3/Q1) up to 640
Liquid temperature range: +0.1°C / 50°C
Maximum operating pressure: d DN200 / 8" : 16 bar / 232 psi, t DN250 / 10":
10 bar / 145 psi
For detailed information refer to
Measurement accuracy
Legal metrology
on page 31.
on page 34.
6
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
27
Page 28
6
TECHNICAL DATA
Operating conditions
Temperature
Temperature
TemperatureTemperature
Process temperature-5...+70°C / +23...+158°F
Ambient temperatureStandard
Protect electronics against self-heating at ambient temperatures above 55°C / 131°F
Storage temperature-50…+70°C / -58…+158°F
Measurement range
Measurement range-12...+12 m/s / -40...+40 ft/s
Measurement rangeMeasurement range
Pressure
Pressure
PressurePressure
Operating pressureUp to 16 bar / 232 psi for DN25...00 / 1...12"
Vacuum load0 mbar / 0 psi absolute
Pressure lossFor detailed information refer
Chemical properties
Chemical properties
Chemical propertiesChemical properties
Physical conditionWater: drinking water, raw water, irrigation water.
Electrical conductivityt 20 PS/cm
WATERFLUX 3000
Standard: DN25...200: remote or compact version with aluminium converter
StandardStandard
housing and with stainless steel flanges:
-40…+65°C / -40…+149°F
compact version with stainless steel converter housing and with stainless
steel flanges
-40…+55°C / -40…+130°F
Standard
Standard, DN250...600: remote version with stainless steel converter
StandardStandard
housing and with carbon steel flanges:
-20…+65°C / -4…+149°F
compact version with stainless steel converter housing and with carbon
steel flanges
-20…+55°C / -4…+130°F
Option
Option, DN250...600 with low temperature carbon steel or stainless steel
OptionOption
flanges :
-40…+65°C / -40…+130°F .
Up to 10 bar / 150 psi for DN50...600 / 14...24"
Pressure loss
For salt water please contact the factory.
to on page 40.
Installation conditions
InstallationAssure that the flow sensor is always fully filled.
For detailed information refer to
Flow directionForward and reverse
Arrow on flow sensor indicates forward flow direction.
Inlet runt 0 DN
For detailed information refer to
Outlet runt 0 DN
For detailed information refer to
Dimensions and weightsFor detailed information refer to
Installation
Measurement accuracy
Measurement accuracy
Dimensions and weights
on page 12.
on page 34.
on page 34.
on page 36.
28
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
29
Page 30
6
TECHNICAL DATA
Electrical connections
Signal cable
Signal cable (remote versions only)
Signal cableSignal cable
Type A (DS)In combination with the IFC 050, IFC 100 and the IFC 300 signal converter
Type B (BTS)Only in combination with the IFC 300 signal converter
I/OFor full details of I/O options, including data streams and protocols, see
WATERFLUX 3000
For detailed information refer to the documentation of the relevant signal
converter.
In combination with the IFC 050, IFC 100 and the IFC 300 signal converter
In combination with the IFC 050, IFC 100 and the IFC 300 signal converterIn combination with the IFC 050, IFC 100 and the IFC 300 signal converter
Standard cable, double shielded.
Max. length: 600 m / 1950 ft
(depends on electrical conductivity and measuring sensor).
For detailed information refer to the documentation of the relevant signal
converter.
Only in combination with the IFC 300 signal converter
Only in combination with the IFC 300 signal converterOnly in combination with the IFC 300 signal converter
Optional cable, triple shielded.
Max. length: 600 m / 1950 ft
(depends on electrical conductivity and measuring sensor).
For detailed information refer to the documentation of the relevant signal
converter.
technical datasheet of the relevant converter.
Approvals and certificates
CE
CE
CECE
This device fulfils the statutory requirements of the EU directives. The manufacturer certifies successful
testing of the product by applying the CE mark
For full information of the EU directives & standards and the approved
certifications, please refer to the EU Declaration of Conformity or the
manufacturer website.
Custody transferOnly in combination with the IFC 300 signal converter.
Directive:2014/32/EU MID Annex III (MI-001) and MID Annex VI (MI-004) type
examination certificate ( DN25...600 / 1...24")
Innerstaatliche Bauartzulassung als Kältezähler (For Germany, Switzerland
and Austria)
Other approvals and standards
Other approvals and standards
Other approvals and standardsOther approvals and standards
Drinking water approvalsACS, DVGW W270, NSF / ANSI Standard 61, TZW, KIWA (ATA), KTW, WRAS
Protection category acc.
to IEC 60529
Corrosion protection class
acc.
to ISO 12944-2
Shock testIEC 68-2-27
Vibration testIEC 68-2-64
Standard:
Standard:
Standard:Standard:
IP66 / 67 (NEMA 4/4X/6)
Option:
Option:
Option:Option:
IP68 factory (NEMA 6P)
IP68 field (NEMA 6P)
IP68 is only available for separate design
Standard: C3 medium
Option: off-shore paint C5-I high - C5-M high
30 g for 18 ms
f = 20...2000 Hz, rms = 4.5 g, t = 30 min
30
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 31
WATERFLUX 3000
6.3 Legal metrology
INFORMATION!
MID Annex MI-001 is only
6.3.1 MID Annex III (MI-001)
All new designs of water meters that are to be used for legal purposes in Europe require
certification under the Measurement Instrument Directive (MID) 2014/32/EU.
Annex III (MI-001) of the MID applies to water meters intended for the measurement of the
volume of clean, cold or heated water in residential, commercial, and light industrial use. An ECtype examination certificate is valid in all countries of the European Union.
The WATERFLUX 3300 has an EC-type examination certificate and can be verified to the
MID Annex III (MI-001) for water meters with diameter DN25...DN600. The conformity
assessment procedure followed for the WATERFLUX 3300 is Module B (Type Examination) and
Module D (Quality Assurance of the Production Process).
The maximum permissible error on volumes delivered between Q2 (transitional) flow rate and
Q4 (overload) flow rate is ±2%.
The maximum permissible error on volumes delivered between Q1 (minimum) flow rate and Q2
(transitional) flow rate is ±5%.
TECHNICAL DATA
only available in combination with the IFC 300 signal converter!
onlyonly
6
Q1 = Q3 / R
Q2 = Q1 * 1.6
Q3 = Q1 * R
Q4 = Q3 * 1.25
Figure 6-2: ISO flow rates added to figure as comparison towards MID
X:
X: Flow rate
X:X:
Y [%]:
Y [%]: Maximum measuring error
Y [%]:Y [%]:
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
31
Page 32
6
TECHNICAL DATA
MI-001 certified flow characteristics
WATERFLUX 3000
DNSpan (R)
Flow rate [m3/h]
Q3 / Q1
Minimum Q1Transitional Q2Permanent Q3Overload Q4
256400.0250.0401620.0
406400.06250.1004050.0
506300.1000.1606378.75
656350.15750.252100125.0
806400.250.400160200.0
1006250.400.640250312.5
1256400.6251.00400500.0
1506301.001.60630787.5
2005081.5752.528001000
2504002.504.0010001250
3004004.006.4016002000
35016015.62525.025003125
40016025.0040.040005000
45016025.0040.040005000
50016039.37563.063007875
60010063.00100.863007875
32
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 33
WATERFLUX 3000
6.3.2 Verification to MI-001
INFORMATION!
Verification to MI-001 , standard at the following values for R, Q1, Q2 and Q3.
Verification at other values for R and Q3 available on request.
Verification to MID Annex III (MI-001)
TECHNICAL DATA
6
DNSpan (R)
Flow rate [m3/h]
Q1 Q2Q3
25800.0500.084
40800.1250.2010
50800.2000.3216
65800.3130.5025
80800.5000.8040
100800.7881.2663
125801.2502.00100
150802.0003.20160
200803.1255.00250
250805.0008.00400
300807.87512.60630
3508020.0032.01600
4008031.2550.02500
4508031.2550.02500
5008050.0080.04000
6008078.751266300
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
33
Page 34
6
TECHNICAL DATA
6.4 Measurement accuracy
Every electromagnetic flowmeter is calibrated by direct volume comparison. The wet calibration
validates the performance of the flowmeter under reference conditions against accuracy limits.
The accuracy limits of electromagnetic flowmeters are typically the result of the combined effect
of linearity, zero point stability and calibration uncertainty.
Figure 6-3: Flow velocity vs. accuracy
XXXX [m/s]: flow velocity
YYYY [%]: deviation from the actual measured value
Converter typeAccuracyCurve
IFC 0500.5% of mv +1 mm/s3
IFC 1000.3% of mv +1 mm/s2
IFC 3000.2% of mv +1 mm/s1
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 35
WATERFLUX 3000
TECHNICAL DATA
6.4.1 WATERFLUX 3050, 3100 and 3300 without straight inlet and outlet sections
Disturbed flow profiles, such as those that occur behind elbows, tee pieces, reducers or valves
installed in front of a flowmeter, affect the measuring performance. Therefore it is usually
recommended to fit a straight inlet length in front of and straight outlet length behind a
flowmeter.
As a result of the unique WATERFLUX flow sensor design, whereby the mean flow velocity and
flow profile are optimized within the rectangular and reduced cross section, the additional
uncertainty for upstream disturbances are drastically reduced. Therefore the requirements for
straight length and in front of and behind a meter are reduced.
EC-type certificate according MID Annex III (MI-001)
• In combination with the signal converter IFC 300
• Diameter range DN25...600
• Minimum straight inlet and outlet pipe length of 0 DN
• Bi-directional flow
6
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
35
Page 36
6
TECHNICAL DATA
WATERFLUX 3000
6.5 Dimensions and weights
Remote version
Remote versiona = 88 mm / 3.5"
Remote versionRemote version
b = 139 mm / 5.5"
c = 106 mm / 4.2"
Total height = H + a
Compact version with :
Compact version with :
Compact version with :Compact version with :
IFC 300
IFC 300
IFC 300IFC 300
a = 155 mm / 6,1"
b = 230 mm / 9.1"
c = 260 mm / 10.2"
Total height = H + a
1
1
36
www.krohne.com07/2018 - 4001626605 - HB WATERFLUX 3000 R06 en
Page 37
WATERFLUX 3000
Compact version with:
Compact version with:
Compact version with:Compact version with:
IFC 100 (0
IFC 100 (0°))))
IFC 100 (0IFC 100 (0
Compact version with:
Compact version with:
Compact version with:Compact version with:
IFC 100 (45
IFC 100 (45°))))
IFC 100 (45IFC 100 (45
TECHNICAL DATA
a = 82 mm / 3.2"
b = 161 mm / 6.3"
c = 257 mm / 10.1"
Total height = H + a
a = 186 mm / 7.3"
b = 161 mm / 6.3"
c = 184 mm / 2.7"
Total height = H + a
1
1
6
Compact version with:
Compact version with:
Compact version with:Compact version with:
stainless steel IFC 100 (10