1.2SCOPE AND FIELD OF APPLICATION.............................................................................................................................................5
1.3IMPORTANT CONSIDERATIONS FOR THE READER......................................................................................................................6
1.4ACCESSING THIS CONFORMANCE S TATEMENT ON THE WORLD WIDE WEB.........................................................................6
2 IMPLEMENTATION MODEL .......................................................................................................................................................8
2.2SEQUENCING OF REAL-WORLD ACTIVITIES...............................................................................................................................8
3 APPLICATION ENTITY S PECIFICATIONS ...........................................................................................................................9
3.1.2 Number of Associations......................................................................................................................................................10
3.2.3 SOP Specific Conformance...............................................................................................................................................11
3.2.4 Association Acceptance Policy.........................................................................................................................................12
3.3BASIC PRINT MANAGEMENT META SOP CLASS.....................................................................................................................12
3.3.1 Basic Film Session SOP Class..........................................................................................................................................12
3.3.2 DIMSE Service N-CREATE ...............................................................................................................................................12
3.3.3 Basic Film Box SOP Class ................................................................................................................................................13
3.4.1 DIMSE Service N-SET.......................................................................................................................................................22
3.5.1 DIMSE Service N-CREATE ...............................................................................................................................................23
3.6STORE SERVICE CLASS ................................................................................................................................................................24
3.7BASIC STUDY CONTENT NOTIFICATION SOP CLASS .............................................................................................................. 26
3.7.1 Basic Study Descriptor IOD..............................................................................................................................................26
3.8.2 DIMSE Service N-GET.......................................................................................................................................................28
4 COMMUNICATION PROFILES .................................................................................................................................................30
5.1EXTENSIONS ON BASIC G RAYSCALE IMAGE BOX SOP........................................................................................................... 31
5.2EXTENSION ON THE DEFINITION OF MODALITY (0008,0060).................................................................................................31
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 3
Page 4
7 SUPPORT OF EXTENDED CHARACTER SETS ..................................................................................................................32
Study Component Management 1.2.840.10008.3.1.2.3.2 SCU
Study Management 1.2.840.10008.3.1.2.3.1 SCU
Basic Study Content Notification 1.2.840.10008.1.9 SCU
1.2 Scope and Field of Application
1.2.840.10008.5.1.1.9 SCU
1.2.840.10008.5.1.4.31 SCU
This document describes the DICOM functionality of the Kodak PACS Link Medical Image Manager 200
(MIM 200) and Kodak PACS Link Medical Image Manager 100 (MIM 100). The MIM 100 and MIM 200
function as system components (collectively referred to as MIM) for capturing exam images and patient
demographics then distributing them within and between institutions. The MIM acts as a DICOM Service
Class User (SCU) and performs transactions over a TCP/IP network via the DICOM messages exchange
protocol.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 5
Page 6
1.3 Important Considerations for the Reader
This DICOM Conformance Statement by itself is not sufficient to guarantee successful connectivity
between the MIM and equipment from other vendors. The following considerations should be made:
§ The integration of equipment from different vendors (including Kodak) goes beyond the scope of the
DICOM 3.0 standard and the DICOM Conformance Statements from Kodak and other vendors. It is
the responsibility of the user (or user’s agent) to assess the application requirements and to design a
solution that integrates Kodak equipment with equipment from other vendors.
§ When the comparison of this DICOM Conformance Statement with a DICOM Conformance
Statement from another vendor indicates that connectivity should be possible, it is the responsibility of
the user (or user’s agent) to verify this by carrying out validation tests and to check whether all
required functionality (such as cut lines) is met.
§ With regard to the future evolution of the DICOM 3.0 standard Eastman Kodak Company reserves the
right to make changes to the Kodak PACS Link Medical Image Manager 100 and the Kodak PACS
Link Medical Image Manager 200 architecture described in this document. The user (or user’s agent)
should ensure that any equipment connected via DICOM to Kodak equipment also follows the future
evolution of the DICOM 3.0 standard. Failure to do so may result in (partial) loss of connectivity.
This implementation is based on the DICOM Standard, publication PS 3.1-2004.
1.4 Accessing this Conformance Statement on the World Wide Web
As the MIM product changes, changes to this DICOM Conformance Statement are inevitable. To obtain
the most recent revision of this DICOM Conformance Statement, acces s the following URL (case
sensitive):
The following abbreviations and acronyms are used in this document.
Term Description
AE Application Entity
ASCII American Standard Code for Information Interchange
DICOM Digital Imaging and Communications in Medicine
DIMSE DICOM Message Service Element
HIS/RIS Hospital Informat ion System/Radiology Information System
IOD Information Object Definition.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 6
Page 7
Term Description
ISO International Standards Organization
LUT Look-up Table
MIM Kodak PACS Link Medical Image Manager 100/200
PDU Protocol Data Unit
PLUT Presentation Look-up Table
SC Secondary Capture
SCP Service Class Provider
SCU Service Class User
SOP Service-Object Pair
TCP/IP Transmission Control Protocol/Internet Protocol
TFT/ULUT Transfer Function Table/User Look-Up Table
UID Unique Identifier
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 7
Page 8
2 Implementation Model
This impleme ntation model uses the DICOM Basic Print Management Meta SOP Class to deliver studies to
remote printers. The SC Image Storage SOP Class is used to deliver studies to archives. Basic Worklist
Management service is used for the acquisition of patient demographics.
2.1 Functional Definitions
The MIM acquires images from the connected devices and demographics via manual entry or connection
to an Information System. Studies are temporarily stored on disk. The images are then sent to the
selected destinations.
2.2 Sequencing of Real -World Activities
If a HIS/RIS service is present, the MIM establishes an association when the MIM application is started to
obtain a modality worklist. The worklist is used as a source of patient demographics. The MIM establishes
an association for each patient study to obtain the study component. The study component is used as a
source of the study description information.
The MIM establishes an association with a selected SCP when the MIM has collected sufficient
information to begin sending images. This may be before the acquisition session has ended.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 8
Page 9
3 Application Entity Specifications
The MIM provides Standard Conformance to the following SOP Classes as an SCU.
SOP Class Name SOP Class UID
Verification SOP Class 1.2.840.10008.1.1
Basic Grayscale Print Management Meta SOP
Class
Basic Color Print Management Meta SOP Class 1.2.840.10008.5.1.1.18
Basic Annotation Box SOP Class 1.2.840.10008.5.1.1.15
Basic Study Content Notification 1.2.840.10008.1.9
Detached Study Management 1.2.840.10008.3.1.2.3.1
Study Component Management 1.2.840.10008.3.1.2.3.2
3.1 Association Establishment Policies
1.2.840.10008.5.1.1.9
1.2.840.10008.5.1.4.31
3.1.1 General
3.1.1.1 Delivery
An Association may be attempted whenever a valid destination is selected and at least one image
has been acquired.
The maximum PDU size which the MIM will negotiate is 64 Kbytes.
3.1.1.2 HIS/RIS
An Association for Basic Worklist will be attempted when the MIM application is started and then
periodically thereafter.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 9
Page 10
3.1.2 Number of Associations
3.1.2.1 Delivery
Associations are initiated with the limitation that no more than three total SCU delivery
associations may be open at any given time. If more destinations are desired, the requests are
queued. The MIM will not create two associations to the same device, even if there are multiple
jobs queued for delivery.
3.1.2.2 HIS/RIS
Only one association will be initiated at a time. The MIM will open/close an independent
association after receiving a C-FIND response and each N-GET response
3.1.3 Asynchronous Nature
The MIM allows up to 1 invoked and 1 performed operation on an Association (it is synchronous, e.g.
the SCU sends only 1 Request and waits for the corresponding Response before sending the next
Request).
3.1.4 Implementation Identifying Information
The MIM provides the Implementation Class UID of “1.2.840.113564.3.1.8”.
The implementation version name attribute is of the form of “MIMyyvxxx” where yy is the Released
Year and xxx is the Version Number. (i.e. MIM03v6.0 stands for Medical Image Manager software,
released in 2003, of version 6.0)
The MIM establishes an Association using its network node name for the Calling DICOM Application
Entity Title (AE Title). The network node name is configurable through the MIM Service Application.
The MIM stores a called DICOM Application Entity Title and socket number for each DICOM compatible network
destination it knows about.
3.2 Association Initiat ion Policy
3.2.1 Associated Real -World Activity
3.2.1.1 Delivery
The MIM initiates Associations for the purpose of sending images and associated information for
printing to a Basic Grayscale Print Management SCP and archiving to an SC Image Storage
SCP.
The default well-known socket 5040 will be used for making the Association unless a different
one is configured by a product service provider.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 10
Page 11
3.2.1.2 HIS/RIS
The MIM initiates Associations for the purpose of obtaining the current Modality Worklist IOD.
3.2.2 Presentation Context Table
The MIM proposes the Presentation Contexts shown below.
Presentation Context Table
Abstract Syntax Transfer Syntax
Name UID Name List UID
Verification 1.2.840.10008.1.1 DICOM Implicit VR
Little Endian
Basic Grayscale
Print
Management
Basic Color
Print
Management
Basic
Annotation Box
Presentation
LUT
Modality
Worklist
1.2.840.10008.5.1.1.9 DICOM Implicit VR
Little Endian
1.2.840.10008.5.1.1.18 DICOM Implicit VR
Little Endian
1.2.840.10008.5.1.1.15 DICOM Implicit VR
Little Endian
1.2.840.10008.5.1.1.23 DICOM Implicit VR
Little Endian
1.2.840.10008.5.1.4.31 DICOM Implicit VR
Little Endian
Role Ext.
Negotiation
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
Secondary
Capture
Image
Storage
Basic Study
Content
Notification
1.2.840.10008.5.1.4.1.1.7 DICOM Implicit VR
Little Endian
1.2.840.10008.1.9 DICOM Implicit VR
Little Endian
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
3.2.3 SOP Specific Conformance
3.2.3.1 Verification
The MIM provides standard conformance to the DICOM Verification Service Class. When
prompted by a user, the MIM will request verification of communication to a remote DICOM AE
using the C-ECHO primitive.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 11
Page 12
3.2.3.2 Delivery
Association attempts will be retried if the SCP rejects the request with the RESULT = 2 (rejected
transient) and the REASON = 1 (temporary congestion). If all Association attempts fail, then the
user will be notified and the Film Session or Study is saved for resending or deletion. No
undelivered image files are deleted without manual user direction.
3.2.3.3 HIS/RIS
The MIM will poll the HIS/RIS broker on power-up to initialize the local patient database on the
MIM and asynchronously on demand from the user. If the MIM does not automatically start
receiving modality worklist changes from the HIS/RIS broker, the MIM will poll the HIS/RIS broker
in 2-minute cycles when the user is in the “Study Info” or “Patient List” keypad screens.
3.2.4 Association Acceptance Policy
None.
3.3 Basic Print Management Meta SOP Class
The Meta SOP Class is defined by the following set of supported SOP Classes:
SOP Class UID Value
Basic Film Session SOP Class 1.2.840.10008.5.1.1.1
Basic Film Box SOP Class 1.2.840.10008.5.1.1.2
Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4
Basic Color Image Box SOP Class 1.2.840.10008.5.1.1.4.1
Printer SOP Class 1.2.840.10008.5.1.1.16
3.3.1 Basic Film Session SOP Class
3.3.2 DIMSE Service N-CREATE
Attribute SCU
Usage
Number of Copies U (2000,0010) 1 –> 99
Tag Possible Values
Print Priority U (2000,0020) HIGH, MED, LOW
Medium Type U (2000,0030) PAPER, CLEAR FILM, BLUE FILM, NONE
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 12
Page 13
Attribute SCU
Usage
Film Destination U (2000,0040) MAGAZINE, PROCESSOR, BIN_n (where n=1 to 9)
Film Session Label U (2000,0050) Up to 64 characters may be provided
Memory Allocation U (2000,0060) Not used
Tag Possible Values
3.3.2.1 DIMSE Service N-ACTION
The MIM uses the N-ACTION to instruct the SCP to print all films in the session. The MIM is
configurable (when the destination is installed) to issue the N-ACTION at the Film Session for
destinations known to support this optional service. If the destination is not known to support
collation, the MIM will only issue the N-ACTION on the Film Box. For Print SCPs that conform to
the N-ACTION specification in Part 4 section H.4.1.2.4 of the DICOM standard, the MIM expects
that all film boxes will be collated when printed.
3.3.2.2 DIMSE Service N-SET
All attributes supported in the N-CREATE are used with this command.
3.3.2.3 DIMSE Service N-DELETE
Once a Film Session is deleted, another will not be created on the same association. The
association will be released.
3.3.3 Basic Film Box SOP Class
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 13
Page 14
3.3.3.1 DIMSE Service N-CREATE
Attribute SCU
Usage
Image Display Format M (2010,0010) STANDARD\C,R
Referenced Film Session
Sequence
>Referenced SOP Class
UID
M (2010,0500)
M (0008,1150)
Tag Possible Values
For LANDSCAPE Film Orientation, (C,R) may =
(1,1) (2,1) (2,2) (3,2) (4,2) (3,3) (4,3) (5,3) (4,4) (5,4) (6,4) (6,5) (7,5)
For PORTRAIT Film Orientation, (C,R) may =
(1,1) (1,2) (2,2) (2,3) (2,4) (3,3) (3,4) (3,5) (4,4) (4,5) (4,6) (5,6) (5,7)
SLIDE (35 mm)
ROW\r1,r2,r3…where r1, r2, r3…is the number of images in each
row.
For PORTRAIT Film Orientation the rows are limited to 7 and the
number of images in each row is limited to 5.
For LANDSCAPE Film Orientation the rows are limited to 5 and the
number of images in each row is limited to 7.
CUSTOM\I
I = 101, 102
See Annex B for description
>Referenced SOP
Instance UID
Referenced Basic Image
Box Sequence
Referenced Basic
Annotation Box
Sequence
Film Orientation U (2010,0040) PORTRAIT, LANDSCAPE
Film Size ID U (2010,0050) 4INX6IN
M (0008,1155)
– (2010,0510) Not used.
– (2010,0520) Not used.
8INX10IN
85INX11IN (Non standard for legacy support)
10INX12IN
11INX11IN
11INX14IN
14INX14IN
14INX17IN
A4
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 14
Page 15
Configuration Information
Attribute SCU
Tag Possible Values
Usage
Magnification Type U (2010,0060) REPLICATE, BILINEAR, CUBIC, NONE
Will be constant for the entire Film Box.
Max Density U (2010,0130) 0-399
U (2010,0150) Curve Shape (CS):
000 to 999
Contrast Values (CN):
-1 to -5 Lower contrast
0 Normal
+1 to +5 Higher contrast
Pivot Density (PD):
0 to 2.4 in increments of 0.2
Perception LUT Selection (LUT):
LUT=m, n (m=string, n = 1 to 15)
Text Macros (TM):
%PRNTDAT%, %TIM%,%FOF%,
%$TIME$%,%ACTU%,%SES%
Perception LUT cannot be used with Curve Shape, Contrast or Pivot
Density.
See Annex A for description
Referenced Presentation
U (2050,0500)
LUT Sequence
>SOP Class UID U (0008,1150)
>SOP Instance UID U (0008,1155)
Annotation Display
Format ID
U (2010,0030) NONE, LABEL, 6
NONE – No annotation
LABEL – Annotation at bottom of film.
6 – Six annotation positions on two lines, centered at bottom of film.
Smoothing Type U (2010,0080) NORMAL (minimum cubic convolution error)
ENHANCED, ENHANCED1 (Valid only for Magnification Type
CUBIC)
0-15 (Valid only for Magnification Type CUBIC)
Border Density U (2010,0100) BLACK, WHITE, i,
where i may = 1-399
Empty Image Density U (2010,0110) Not used
Min Density U (2010,0120) 0-399 (Value must be less than Max Density (2010,0130))
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 15
Page 16
Attribute SCU
Usage
Illumination MC (2010,015E) Positive integer in units of cd/m2
Reflective Ambient Light MC (2010,0160) Positive integer in units of cd/m2
Trim U (2010,0140) YES, NO
Tag Possible Values
3.3.3.2 DIMSE Service N-ACTION
The MIM uses the N-ACTION to instruct the SCP to print the current film in the session.
3.3.3.3 DIMSE Service N-SET
This service is not used.
3.3.3.4 DIMSE Service N-DELETE
This service is not used.
3.3.4 Basic Grayscale Image Box SOP Class
3.3.4.1 DIMSE Service N-SET
Attribute & Usage SCU
Usage
Image Position M (2020,0010) All values within the range of Image Display Format
Preformatted Grayscale
Image Sequence
>Samples Per Pixel M (0028,0002) 1
>Photometric
Interpretation
>Rows M (0028,0010) Minimum Value 64
M (2020,0110)
M (0028,0004) MONOCHROME2, MONOCHROME1
Tag Supported Values
Maximum Values:
Known for all Kodak printers, configurable for others. The aspect ratio
is used with the printer’s page extents, display format, etc., to
calculate this value.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 16
Page 17
Attribute & Usage SCU
Tag Supported Values
Usage
>Columns M (0028,0011) Minimum Value 64
Maximum Values:
Known for all Kodak printers, configurable for others. The aspect ratio
is used with the printer’s page extents, display format, etc., to
calculate this value.
>Pixel Aspect Ratio MC (0028,0034) R\C
R, C = 1 to 9999 (Integer)
Note: This attribute is always included, even if it is 1\1. It’s v alue will
always be 1\1 if Magnification Type is NONE.
>Bits Allocated M (0028,0100) 8, 16
>Bits Stored M (0028,0101) 8, 10, 12
>High Bit M (0028,0102) Bits Stored -1
>Pixel Representation M (0028,0103) 0000H (unsigned integer)
>Pixel Data M (7FE0,0010) All values consistent with Bits Stored
Polarity U (2020,0020) NORMAL, REVERSE
Magnification Type U (2010,0060) REPLICATE, BILINEAR, CUBIC, NONE
Note: Is always the same as the Magnification Type specified for the
Film Box.
Smoothing Type U (2010,0080) NORMAL, ENHANCED, ENHANCED1
0-15 (Valid only for Magnification Type CUBIC)
Note: Is always the same as the Smoothing Type specified for the
Film Box.
Configuration Information
U (2010,0150) Setting these values is expected to override the film box se ttings for
this image position.
Curve Shape (CS):
000 to 999
Perception LUT Selection (LUT):
LUT=m, n (m=string, n = 1 to 15)
Curve Shape and Perception LUT are mutually exclusive.
See Annex A for description
Requested Image Size U (2020,0030) Row length in fractional mm up to the size of the printable image,
which is a function of Image Display Format and Film Size ID.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 17
Page 18
Attribute & Usage SCU
Usage
Referenced Presentation
LUT Sequence
>SOP Class UID U (0008,1150)
>SOP Instance UID U (0008,1155)
Body Part Examined U (0018,0015) ABDOMEN, ANKLE, ARM, BREAST, CHEST, CLAVICLE, COCCYX,
MA, MR, MS, NM, OT, PT, RF, RG, ST, TG, US, XA
OT = Other
See section 4 for more information.
See section 4 for more information.
3.3.5 Basic Color Image Box SOP Clas s
3.3.5.1 DIMSE Service N-SET
Attribute & Usage SCU
Usage
Image Position M (2020,0010) All values within the range of Image Display Format
Basic Color Image Sequence M (2020,0111)
>Samples Per Pixel U (0028,0002) 3
>Photometric Interpret ation U (0028,0004) RGB
>Planar Configuration M (0028,0006) 000H or 001H
Tag Supported Values
000H – pixels arrive in R1G1B1R2G2B2R3G3B3 order
001H – pixels arrive in R1R2R3, G1G2G3, B1B2B3 order
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 18
Page 19
Attribute & Usage SCU
Tag Supported Values
Usage
>Rows M (0028,0010) Minimum Value 64
Maximum Values:
Known for all Kodak print ers, configurable for others. The aspect
ratio is used with the printer’s page extents, display format, etc.
to calculate this value.
>Columns M (0028,0011) Minimum Value 64
Maximum Values:
Known for all Kodak printers, configurable for others. The aspect
ratio is used with the printer’s page extents, display format, etc.
to calculate this value.
>Pixel Aspect Ratio MC (0028,0034) R\C
R, C = 1 to 9999 (Integer)
>Bits Allocated U (0028,0100) 8
>Bits Stored U (0028,0101) 8
>High Bit U (0028,0102) 7
>Pixel Representation M (0028,0103) 0000H (unsigned integer)
>Pixel Data M (7FE0,0010) All values consistent with Bits Stored
Polarity U (2020,0020) NORMAL, REVERSE
Magnification Type U (2010,0060) REPLICATE, BILINEAR, CUBIC, NONE
Smoothing Type U (2010,0080) NORMAL
ENHANCED, ENHANCED1 (Valid only for Magnification Type
CUBIC)
0-15 (Valid only for Magnification Type CUBIC)
Configuration Information U (2010,0150) Setting these values is expected to override the film box settings
for this image position.
Curve Shape (CS):
000 to 999
Perception LUT Selection (LUT):
LUT=m, n (m=string, n = 1 to 15)
Curve Shape and Perception LUT are mutually exclusive.
See Annex A for description
Requested Image Size U (2020,0030) Row length in fractional mm up to the size of the printable image,
which is a function of Image Display Format and Film Size ID.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 19
Page 20
Attribute & Usage SCU
Usage
Color Profile U (2011,0160) DEFAULT1, DEFAULT2, DEFAULT3, DEFAULT4, DEFAULT5,
Tag Supported Values
DEFAULT6
See section 4 for more information.
3.3.6 Printer SOP Class
3.3.6.1 DIMSE Service N-EVENT-REPORT
The MIM will process the indication of the N-EVENT-REPORT operation. Any string sent by the
SCP is accepted and displayed on the MIM user interface. Some outflanking links require printer
status notification so the MIM will translate some common Printer Status Info values to
appropriate outflanking notifications. In this translation, all characters that are not space
characters or in the ASCII range “A” - “Z” are stripped.
The MIM translates Attributes as described in the following table. Other strings are not translated
but may be displayed on a MIM keypad.
Attribute SCU
Usage
Tag Expected Values
Printer Status U (2110,0010) NORMAL
WARNING
FAILURE
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 20
Page 21
Attribute SCU
Tag Expected Values
Usage
Printer Status Info U (2110,0020) BAD RECEIVE MGZ
BAD SUPPLY MGZ
CALIBRATING
CALIBRATION ERR
CHECK CHEMISTRY
CHECK INK CART
CHECK PRINTER
CHECK PROC
CHECK PROCESSOR
CHECK RIBBON
CHECK R MAG
CHECK SORTER
CHECK SUPPLY MAG
CHEMICALS EMPTY
CHEMICALS LOW
COVER OPEN
ELEC CONFIG ERR
ELEC DOWN
ELEC SW ERROR
EXPOSURE FAILURE
FATAL
FATAL ERROR
FILM JAM
FILM TRANSP ERR
FINISHER EMPTY
FINISHER ERROR
FINISHER LOW
INK OUT
INSUFFIC MEMORY
INVALID PAGE DES
NO RECEIVE MGZ
NO RESPONSE
NO RIBBON
NO SUPPLY MGZ
NO STATE
NO TONER
NORMAL OFFLINE
PRINTER BUSY
PRINTER DOWN
PRINTER INIT
PRINTER NOT RDY
PRINTER OFFLINE
PRINTER STOPPED PROCESSOR DOWN
PROC DOWN
PROC INIT
PROC NOT READY
PROC NOT RDY
PROC OVERFLOW FL
PROC OVERFLOW HI
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 21
Page 22
Attribute SCU
Usage
Printer Status Info (continued)
Printer Name U (2110,0030) Any valid string
Printer Manufacturer U (0008,0070) Any valid string
Printer Manufacturer Model
Name
Printer Device Serial Number U (0018,1000) Any valid string
QUEUED
U (0008,1090) Any valid string
Tag Expected Values
RECEIVER FULL
REQ MED NOT INST
REQ MED NOT AVAI
RIBBON ERROR
RIBBON MISMATCH
STATE UNKNOWN
SUPPLY EMPTY
SUPPLY LOW
TIME OUT
UNKNOWN
UNKNOWN STATUS
Software Version U (0018,1020) Any valid string
Date of Last Calibration U (0018,1200) Ignored
Time of Last Calibration U (0018,1201) Ignored
3.4 Basic Annotation Box SOP Class
3.4.1 DIMSE Service N-SET
The Basic Annotation Box SOP Instance is created at the time the Basic Film Box SOP Instance is
created, based on the value of the Annotation Display Format ID attribute (2010,0030) of the Basic
Film Box.
Attribute & Usage SCU
Usage
Annotation Position M (2030,0010) 0 for FormatID LABEL,
Text String M (2030,0020) Up to 64 characters
Tag Supported Values
1..6 for FormatID 6
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 22
Page 23
3.5 Presentation LUT SOP Class
3.5.1 DIMSE Service N-CREATE
The Presentation LUT SOP Instance is created by the SCU prior to the creation of the Basic Film Box
SOP Instanc e. Multiple Presentation LUT instances are supported in an association, but only one
instance will be supported for each image.
The SCU shall send either Presentation LUT Sequence or the Presentation LUT Shape. These
values are mutually exclusive. The pres ence of the Presentation LUT instance overrides any data set
in the Configuration Information attribute (2010,0150) of the Film Box or Image Box.
The Presentation LUT values are calculated using the Perception LUT (TFT) values and Contrast
values or the Cu rve Shape and Contrast values set at the MIM keypad or autofilming links.
Attribute &
Usage
Presentation LUT
Sequence
>LUT Descriptor MC (0028,3002) The first value is the number of entries in the lookup table.
>LUT Explanation U (0028,3003)
>LUT Data MC (0028,3006) The LUT Data shall be stored in a format equivalent to 16
Presentation LUT
Shape
SCU
Usage
MC (2050,0010)
MC (2050,0020) Enumerated values IDENTITY and LIN OD.
Tag Supported Values
The number of entries shall be equal to the number of
possible values in the input. (For 8 bit input it will be 256
entries, for 12 bit input it will be 4096 entries)
.
The second value is the first input value mapped, and shall
always be 0.
The third value specifies the number of bits for each entry in
the LUT Data. It shall be between 10 and 16 inclusive.
bits allocated where the high bit is equal to bits stored - 1,
where bits stored is the third value of the LUT Descriptor.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 23
Page 24
3.6 Store Service Class
Secondary Capture Images are sent to the SC Storage SCP.
3.6.1 Secondary Capture IOD
The IOD sent from the MIM minimally contains the following attributes:
Attribute Name Tag DICOM Type MIM Type
Patient
Patient Name (0010,0010) 2 2
Patient ID (0010,0020) 2 2
Patient Birth Date (0010,0030) 2 2
Patient Birth Time (0010,0032) 3 2
Patient Sex (0010,0040) 2 2
General Study
Study Instance UID* (0020,000D) 1 1
Study ID** (0020,0010) 2 2
Referring Physician (0008,0090) 2 2
Accession Number (0008,0050) 2 2
Study Description*** (0008,1030) 3 3
Study Date (0008,0020) 2 1
Study Time (0008,0030) 2 1
Requested Procedure Code Sequence (0032,1064) 3 3
General Series
Modality (0008,0060) 1 1
Series Number (0020,0011) 1 1
Series Instance UID (0020,000E) 1 1
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 24
Page 25
Attribute Name Tag DICOM Type MIM Type
Series Description (0008,103E) 3 3
Body Part Examined (0018,0015) 3 3
Protocol Name (0018,1030) 3 3
General Equipment
Manufacturer (0008,0070) 2 2
Institution Name (0008,0080) 3 3
Station Name (0008,1010) 3 3
Manufacturer Model Name (0008,1090) 3 3
SC Equipment
Conversion Type (0008,0064) 1 1
General Image
Image Number (0020,0013) 1 1
Image SOP Class UID (0008,1150) 1 1
Image Instance UID (0008,1155) 1 1
SC Image
Conversion Type (0008,0064) 1 1
Image Pixel
Samples per Pixel (0028,0002 1 1
Photometric Interpretation (0028,0004) 1 1
Rows (0028,0010) 1 1
Columns (0028,0011) 1 1
Bits Allocated (0028,0100) 1 1
Bits Stored (0028,0101) 1 1
High Bit (0028,0102) 1 1
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 25
Page 26
Attribute Name Tag DICOM Type MIM Type
Pixel Representation (0028,0103) 1 1
Pixel Data (7FE0,0010) 1 1
Planar Configuration (0028,0006) 1C 1C
Pixel Aspect Ratio (0028,0034) 1C 1C
SOP Common
SOP Class UID (0008,0016) 1 1
SOP Instance UID (0008,0018) 1 1
Notes:
*The Study Instance UID is normally provided by the HIS/RIS Broker. In the case where the Broker is not
present, a unique UID will automatically be generated.
**The StudyID will contain either the Scheduled Procedure Step Description or the Requested Procedure
Description or the Patient Name in that order of availability from HIS/RIS server.
***The Qualified Service Provider can configure the system to replace the content s of the Study
Description tag (0008, 1030) with information from the Requested Procedure Description tag (0032,
1060).
3.7 Basic Study Content Notification SOP Class
Basic Study Content Notification is used to notify the Store SCP that all images in a study have been
sent.
3.7.1 Basic Study Descriptor IOD
The IOD sent from the MIM minimally contains the following attributes:
Attribute Name Tag DICOM Type MIM Type
Patient Summary
Patient Name (0010,0010) 2 2
Patient ID (0010,0020) 2 2
Study Content
Study ID (0020,0010) 2 2
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 26
Page 27
Attribute Name Tag DICOM Type MIM Type
Study Instance UID (0020,000D) 1 1
Referenced Series Sequence (0008,1115) 1 1
>Series Instance UID (0020,000E) 1 1
>Retrieve AE Title (0008,0054) 2C 2C
>Referenced Image Sequence (0008,1140) 1 1
>>Referenced SOP Class UID (0008,1150) 2 2
>>Referenced SOP Instance UID (0008,1150) 2 2
SOP Common
SOP Class UID (0008,0016) 1 1
SOP Instance UID (0008,0018) 1 1
Specific Character Set (0008,0005) 1C 1C
3.8 Basic Worklist Service
The C-FIND request for a Modality Worklist sends an Identifier object that contains all the attributes of the
Modality Worklist Information Model. The Matching Key attributes that may optionally contain a non-NULL
value in the request are:
Attribute Tag
Accession Number (0008, 0050)
Patient’s Name (0010, 0010)
Patient’s ID (0010, 0020)
Scheduled Station Name (0040, 0010)
Scheduled Procedure Step Start Date (0040, 0002)
Scheduled Procedure Step Start Time (0040, 0003)
This is intended to produce a series of responses from the Worklist SCP for all matching Scheduled
Procedures on the said Station.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 27
Page 28
3.8.1 Modality Worklist IOD
For additional information on the Modality Worklist Information Model, refer to the DICOM
specification, Part 4, Table K.6-1. The MIM will only accept the ISO registration number ISO -IR 6 or
ISO-IR 100 character sets for the Specific Character Set attribute (0008, 0005).
Attributes For The Modality Worklist Information Model
Attribute Tag
Accession Number (0008,0050)
Modality (0008,0060)
Referring Physician’s Name (0008,0090)
Patient’s Name (0010,0010)
Patient’s ID (0010,0020)
Patient’s Birth Date (0010,0030)
Patient’s Sex (0010,0040)
Study Instance UID (0020, 000D)
Requested Procedure Description (0032,1060)
Requested Procedure Code Sequence (0032,1064)
Schedule Procedure Step Sequence (0040,0100)
3.8.2 DIMSE Service N-GET
Once configured to retrieve Study Description from a HIS/RIS Server, a Study N-GET is initiated to
obtain the Referenced Study Component sequence using the Study Instance UID attribute returned
from the Modality Worklist C-FIND. A Study Component N-GET is initiated to obtain the actual Study
Component information using the Referenced Study Component sequence returned from the Study
N-GET. The Study Description tag (0008,1030) is extracted and used with DICOM Store Service
Class.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 28
Page 29
3.8.2.1 Study Management
A Study N-GET is initiated to obtain the Referenced Study Component sequence using the Study
Instance UID attribute returned from the Modality Worklist C-FIND.
Attributes from Study Management N-GET:
Attribute Tag
Referenced Study Component Sequence (0008, 1111)
3.8.2.2 Study Component Management
A Study Component N-GET is initiated to obtain the actual Study Component information using
the Referenced Study Component sequence returned from the Study N-GET. The Study
Description tag (0008,1030) is extracted and used with DICOM Store Service Class.
Attributes from Study Management N-GET:
Attribute Tag
Study Descriptions (0008, 1030)
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 29
Page 30
4 Communication Profiles
4.1 Supported Communications Stacks
The MIM provides TCP/IP Network Communication Support as defined in Part 8 of the DICOM standard.
The MIM normally issues a network ping prior to delivering a job to a qualified DICOM Print or DICOM
Store destination to ensure that the device is active. The MIM does not issue a network ping to
destinations for which this capability has not been previously verified by the Kodak DICOM Verification
and Validation group.
If a destination does not support a network ping, or if the customer has disabled this service on the
network, the MIM configuration must be changed for each destination as appropriate to disable the
issuing of the network ping.
4.2 Physical Media
The MIM supports Ethernet with the following physical connectors:
Standard Twisted pair (10BaseT and 100BaseT)
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 30
Page 31
5 Extensions/Specializations/Privatizations
The following extensions to DICOM SOP classes are supported by the MIM. The extensions are all
optional attributes.
5.1 Extensions on Basic Grayscale Image Box SOP
Three new attributes have been added to the standard SOP class:
• Body Part Examined (0018, 0015) may be used to specify the body part corresponding to the
image being transmitted.
• Modality (0008,0060) may be sent to specify the modality type corresponding to the image being
transmitted.
• A new private element of Image Tone Adjustment (2011, 0170) may also be used to specify tone
scaling for the image.
5.2 Extension on the definition of Modality (0008,0060)
The defined terms for Modality (0008,0060) have been extended to include KCR (Kodak DirectView CR
Systems), FCR (Fuji Computed Radiography Systems), DN (Dental), and FD (Kodak LS Digitizers).
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 31
Page 32
6 Configuration
The following attributes are configurable by a qualified service provider:
n IP address
n Subnet Mask
n Local Network Host Name (MIM AE Title)
n Router Address (Gateway)
n SCP DICOM Called Application Entity Ti tle
n SCP port number
n SCP DICOM Service(s) available
n SCP Film Sizes available
n Other destination properties as indicated in this document.
7 Support of Extended Character Sets
The MIM supports the ISO -IR 100 Latin 1 character set as well as the ISO-IR 6 default character set.
8 Error Handling
The MIM has limited recovery from communication errors. Some specific warnings will be logged locally
and communication will continue. All other DICOM status codes (warning or error) will result in an aborted
session (A -ABORT sent).
Codes that will NOT cause A-ABORT:
0x0000 SUCCESS
0xb602 Film session does not contain Image
0xb603 Film Box does not contain Image
0x0210 Duplicate Invocation
0x0107 Attribute List Error. Interpreted as Value out of range, default used
0x0116 Attribute value is out of range, default used.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 32
Page 33
Annex A: Configuration Information
The Configuration Information attribute contains the list of Kodak -specific values. These attributes are not DICOM
standard attributes.
The Configuration Information value is an ordered list. The attribute is specified using the ASCII two character key
prefix in the following sequence:
1. Curve Shape, Contrast, Pivot Density, or Perception LUT
2. Text Macros.
The Film Box Curve Shape value applies to all images in the Film Box except when Curve Shape or Perception
LUT is specified for the image in the Image Box.
ATTRIBUTE USAGE DESCRIPTION DEFAULT
Curve Shape designated
by the ASCII two
character prefix:
CS
Contrast designated by
the ASCII two character
prefix:
CN
Pivot Density designated
by the ASCII two
character prefix:
PD
Perception LUT
Selection designated by
the ASCII threecharacter prefix:
LUT
U/M 000 to 999
Note: 000 = linear, 999 = highest curvature
Curve Shape is a tone scale adjustment used to
optimize the image on film compared to the image
on the operator console monitor. Curve shape is not
valid when a Perception LUT is specified.
U/M -5 to 5
Note: Integer values only.
Negative Contrast settings are lower contrast where
the amount of data that is represented by medium
film densities is increased.
Positive settings are higher contrast where the
amount of data that is represented by high and low
densities is increased.
U/M
0.0 to 2.4
Note: Value must be specified in increments of 0.2.
Densities above and below the pivot density will be
adjusted up and down by an amount that is a
function of the difference between the code value
and the pivot density code value.
U/M LUT=m, n
Allows selection of the LUT that best suits the user’s
images. M is the name of the TFT set and N
specifies a contrast setting within the group. Curve
shape will be ignored if the LUT parameter is used.
m=string (0=default group)
n=0 to 15 (0=use default value)
Film Box:
Value set in the Printer by
the user
Image Box:
Basic Film Box Curve Shape
Value set in the Printer by
the user
Value set in the Printer by
the user
m=0, n=0
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 33
Page 34
ATTRIBUTE USAGE DESCRIPTION DEFAULT
Text Macros designated
by the ASCII twocharacter key prefix:
TM
M/M %PRNTDAT% Date of Printing DD-MMM -YY
%TIM% Date of Printing HH: (HH=0-23)
%FOF% Film of Film Count NN/MM
%$TIME$% Time of Printing HH:MM:SS
%SES% Film Session Label AAAAAA
(1-64 chars from the Film Session SOP Class)
Note: The text macros will be printed on the bottom
of the film and will be truncated if necessary. These
are only used when Annotation Box SOP is not
supported by Print SCP.
None
Examples
“CS333”
The curve shape is set to 1/3 of the printer’s tone scale range and defaults are applied to contrast and pivot
density.
”CS500\ CN3\PD2.2”
The curve shape is set to 1/2 the printer’s tone scale range, Contrast is set to 3, and pivot density is set to 2.2.
”PD2.0”
The pivot density is set to 2.0., and defaults are applied to curve shape and contrast.
”CS333\ CN3\PD2.2\TM%PRNTDAT%%TIM%%FOF%”
The curve shape is set to 1/3 of the range, Contrast is set to 3, and pivot density is set to 2.2.
The following text macros will be printed on the bottom of the page:
Date of Printing, Time of Printing, and Film of Film count.
“LUT=Ver693c0.w87,3”
The Perception LUT TFT set is “Ver693c0.w87” and the Contrast Setting is 3.
“LUT=0,3\ TM%PRNTDAT%%TIM%%FOF%”
The Perception LUT TFT set is 0 (default) and the Contrast Setting is 3.
The following text macros will be printed on the bottom of the page:
Date of Printing, Time of Printing, and Film of Film count.
“TM%PRNTDAT%%TIM%%FOF%”
The following text macros will be printed at the bottom of the page:
Date of Printing, Time of Printing, and Film of Film count.
”PD2.0\CN4\CS333”
This is invalid because the attributes are out of order, curve shape must precede pivot density and contrast, and
contrast must precede pivot density. It should be “CS333\CN4\PD2.0”.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 34
Page 35
”CS333\PD1.2\LUT=0,3”
This is invalid because Curve Shape and Pivot Density cannot be mixed with Perception LUT. In this case, the
Perception LUT setting will be used.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 35
Page 36
Annex B: Custom Formats
The following formats are expected to be supported by printers manufactured by Kodak and will only be specified
to these printers.
Format ID 101
This format consists of 7 image positions, 1 large image in the upper sec tion of the page and 6 smaller images in
the lower section of the page. The size and positioning of the images are defined in terms of the standard formats
2 and 12.
Upper Section: 1 frame of a 2-up format.
Lower Section: 6 frames of a 12 -up format.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 36
Page 37
Format ID 102
This format consists of 11 image positions, 2 large images in the upper section of the page and 9 smaller images
in the lower section of the page. The size and positioning of the images are defined in terms of the standard
formats 6 and 15.
Upper Section: 2 frames occupying top 1/3 of media.
Lower Section: 9 frames occupying bottom 2/3 of media.
Kodak PACS Link Medical Image Manager 100/200 SCU v. 6.1.1
DICOM Conformance Statement, Document # 2E0380, Rev. Q 37