Knight Equipment DDS-1000T Installation Manual

Page 1
DDS-1000T
INSTRUCTION MANUAL
THEORY OF OPERATION
The DDS-1000T Dry Dispensing System is designed to dispense solid, or “screen-feed” chemicals into a washer. Each dispenser includes a 24 volt water solenoid, Power Bowl Ultra dry detergent feeder (with vacuum breaker), step-down transformer, limit-timer control board, fully assembled and ready to install.
The DDS-1000T is designed as a “stand alone” dispenser that can be used on any top-load or front-load laundry machine. It can be manually activated by the operator using the external dispense switch, or connected to washer supply signals for automatic operation.
The DDS-1000T will start a feed cycle when either the Start button is pressed or when the signal input terminals receive an external 14-240 VAC trigger signal.
GETTING STARTED
Be sure to inspect the installation site prior to beginning your installation. It’s important that you locate the source for dispenser power and signals as well as hot water. To complete the installation, you may need the following parts:
Hose “Y” with 3/4" female garden hose thread
High pressure hose, 3/4" MGHT x 3/4" FGHT
#10 x 1" sheet-metal screws and wall anchors
Optional top-mount bracket kit #7600668 is av ailable for top mounting the dispenser on the laundry machine
INSTALLATION
(1) Mount the DDS-1000T on the wall in a convenient location for operators to load and unload product containers.
The dispenser should be installed above the vertical height of the injection point on the washer.
(2) Connect the 5/8" ID vinyl injection tubing supplied with the unit to the outlet fitting on the bottom of unit. Cut the
discharge tubing to desired length and secure to washer supply inlet point. A 70° elbow barb fitting is supplied for connection to washer.
(3) If an injection point is not available or desired, the injection tube can be manually placed in the machine each
load.
(4) Locate the chemical supply outputs on the washer for each product you wish to dispense. These outputs are
used to start the time feed sequence controlled by the internal KTM-500 control board.
(5) The dispenser must have constant power in order to feed when the washer signals for chemical. Locate a
constant source of power (115/208/230 VAC) and connect to the power input terminals on the primary side of the transformer (see wiring diagram on page 4). Use only approved electrical conduit and connect at bottom of dispenser using conduit fittings. Always use a voltmeter to check power, and turn power off during installation.
(6) Connect the water source to the dispenser using a 3/4" male garden hose fitting, or 1/2" FNPT x 1/2" barb brass
fitting to nylobraid tubing, or to hard-plumb use 1/2" FNPT x 1/2" sweat (copper) fitting. A 3/4" hose “Y” can be used to plumb the dispenser between the water source and the washer (see diagram on page 5).
Important:
with chemical manufacturer for recommended water temperature and optimum pressure.
Check all local plumbing codes for backflow prevention and water line connection standards. Check
0900900 Rev: D (05/07) Page 1 of 8
Page 2
(7) To maintain consistent injections, water pressure should be regulated at 25-50 PSI. Water temperature and
pressure will determine required dispenser run time.
(8) Connect the vacuum breaker to the unit using the supplied copper tubes. Tighten compression fittings on the
inlet and outlet fittings and check for leaks.
CONNECTING WASHER SUPPLY SIGNAL FOR AUTOMATIC DISPENSING
Locate the chemical supply signals in the washer. Connect the chemical signal common and hot leads to the terminals on the KTM-500 circuit board marked “SIG IN“. The KTM-500 circuit board will accept supply signal voltages from 14 to 240 VAC. The supply signal must be a minimum 5 seconds to activate chemical injection.
PROGRAMMING
Run Time:
(1) Locate the dip-switch pack on the circuit board — set switch #6 to SIGNAL, set switch #7 to RUN TIME and set
(2) Using a measuring cup or flask, press Start switch and release when feed starts. Let the unit feed until desired
(3) Set mode switch #8 to RUN MODE.
Delay Time:
(1) Locate the dip-switch pack on the circuit board — set switch #6 to SIGNAL, set switch #7 to DELAY TIME and
(2) Press Start switch and release when the LED begins flashing. When the desired delay time has passed, press
(3) Set mode switch #8 to RUN MODE.
Lock-Out Time:
This feature defeats consecutive dispensing of product for a pre-determined interval. Select a combination of switches 1 – 5 to program total lock-out time.
Example: For 10 minute lock-out, set switches #2 and #4 to ON with all other switches OFF. For maximum lock-out (31 min) set all switches ON. For no lock-out, set all switches OFF.
(max run time is 12 minutes and 42 seconds)
switch #8 to PROGRAM MODE.
amount of chemical is dispensed then press Start switch again to stop. The run time is now programmed. Repeat step if new volume is required.
(max delay time is 12 minutes and 42 seconds)
set switch #8 to PROGRAM MODE.
the Start switch again. The delay time is now programmed. Repeat step if new delay time is required.
(max lock-out time is 31 minutes)
OPERATION
Manual activation:
will begin counting down the delay time (if used) and will then dispense for the amount of time programmed. Once the lock-out time expires the unit will be ready to restart.
Signal activation:
seconds, the delay time (if used) will begin counting down. Then unit will dispense for the amount of time programmed. Once the lock-out time expires the unit will be ready to restart.
Relay Mode:
as long as the manual button is depressed. All other board functions (such as delay time and lock-out time) are by­passed in relay mode.
Page 2 of 8 0900900 Rev: D (05/07)
Press the Start button on the cover or on the remote switch box for 1 full second. The KTM-500
When the signal input on the circuit board receives a 14-240 VAC trigger signal for at least 5 full
Set switch #6 to RELAY. The feed will activate for as long as an external trigger signal is present, or for
Page 3
MULTIPLE UNIT CONNECTION
Connecting multiple DDS-1000 dispensers is quick and easy with the DDS-1000 Connector Kit (Part Number 7600672).
(1) Align units on the mounting surface side by side
and as close together as possible. Kit includes a 15 inch length of ½” ID Nylobraid tubing.
(2) Remove ½” plastic plug from brass “T” on bottom
of unit.
(3) Replace plug with ½” NPT x ½” Barb Elbow
fitting. Be sure to use plumbing tape or pipe sealing compound to prevent leaks.


(4) Repeat above procedure for each unit connected.
(5) Insert plastic plug into ¾” hose adapter on each
unit except the unit closest to the water source.
(6) Connect ½” Nylobraid tubing to Barb Fittings
using stainless steel hose clamps.
(7) Connect “Source” water supply hose to unit
closest to water supply.
(8) Turn on water supply, press detergent / bleach /
softener start buttons and check for leaks.

0900900 Rev: D (05/07) Page 3 of 8
Page 4
DDS-1000T WIRING DIA GRAM
DISABLING THE START BUTTON
There is a jumper marked “JP1” on the circuit board that can be used to prevent manual activation in certain applications, or to allow manual activation by
remote push-button only. This jumper only affects the on-board start button. A remote start button, or trigger signal, can always be used to activate the
chemical injection.
When the jumper is ON, the on-board start button is functional.
When the jumper OFF, the on-board start button is disabled.
Page 4 of 8 0900900 Rev: D (05/07)
Page 5
DDS-1000T ASSEMBLY DIAGRAM
0900900 Rev: D (05/07) Page 5 of 8
Page 6
OPTIONAL WATER TEMPERING VALVE AND PRESSURE REGULATOR
We recommend the use of a water tempering valve and pressure regulator for optimal dry-feed performance. Connect the tempering valve and pressure regulator before the water inlet of the DDS-1000 (see illustration below). The tempering valve will allow the water temperature to be increased or decreased by turning the adjustment dial. The adjustment range is 120° to 160° F. The pressure regulator will regulate the water pressure from 0 to 60 PSI. Most manufacturers recommend pressure of 18-25 PSI (1.2 – 1.7 BAR) at a temperature of 120º-135° F (48°-57° C).
Connect to DDS-1000
Water Inlet
Cold Water
Pressure gauge assembly can be purchased at your local hardware store (see parts list below) 1 - 1/2” MNPT x 1/4” Brass Reducer 1 - 1/4” Pressure Gauge 2 - 1/4” Brass Close Nipple
7407118 Pressure
Regulator
Hot Water
7407119 Water Tempering
Valve
1 - 1/4” MNPT x 3/8” Barb 1 - 1/4” FPT Brass Tee
OPERATION
Insert a chemical capsule into the bowl of the DDS-1000. Run a few test cycles and inspect the operation of both the washer and DDS-1000 to ensure proper chemical feed is achieved. Check all plumbing for leaks.
CALIBRATION OF SOLID DETERGENTS
As general rule of thumb program 10 seconds of run time for one ounce (28.3 grams) of detergent. For 2 ounces (56.7 grams) set the run time for 20 seconds. First weigh the capsule (electronic scale may be needed) then start or signal the dispenser. Weigh the capsule again after the run time is completed to calculate amount dispensed. Run the capsule 2 or 3 times to see if the amount is consistent.
Page 6 of 8 0900900 Rev: D (05/07)
Page 7
TROUBLESHOOTING
Water leaking from inlet fittings
Check for proper connection at fittings.
Check for holes or splits on plumbing lines.
Chemical is not feeding
Check washer signal for proper operation.
Check to see if solenoid is opening properly.
Check to see if water supply is on.
Check inside of vacuum breaker for any debris or blockage.
Check for proper water pressure.
Check operation of Power Bowl Ultra water valve.
Check operation of Power Bowl Ultra spray-jet.
Concentration too high or too low
Check water temperature. Consult with solid chemical manufacturer for recommended temperature.
Check water supply for fluctuations in pressure.
Check feed time of control and adjust as needed.
Power Bowl Ultra is over-flowing
Output hose is blocked, or plugged up.
The end of the output hose is above the mounting height of the DDS-1000.
0900900 Rev: D (05/07) Page 7 of 8
Page 8
DISCLAIMER
Knight LLC does not accept responsibility for the mishandling, misuse, or non-performance of the described items when used for purposes other than those specified in the instructions. For hazardous materials information consult label, MSDS, or Knight LLC. Knight products are not for use in potentially explosive environments. Any use of our equipment in such an environment is at the risk of the user, Knight does not accept any liability in such circumstances.
WARRANTY
All Knight controls and pump systems are warranted against defects in material and workmanship for a period of ONE year. All electronic control boards have a TWO year warranty. Warranty applies only to the replacement or repair of such parts when returned to factory with a Knight Return Authorization (KRA) number, freight prepaid, and found to be defective upon factory authorized inspection. Bearings and pump seals or rubber and synthetic rubber parts such as “O” rings, diaphragms, squeeze tubing, and gaskets are considered expendable and are not covered under warranty. Warranty does not cover liability resulting from performance of this equipment nor the labor to replace this equipment. Product abuse or misuse voids warranty.
FOOTNOTE
The information and specifications included in this publication were in effect at the time of approval for printing. Knight, LLC reserves the right, however, to discontinue or change specifications or design at any time without notice and without incurring any obligation whatsoever.
Knight Headquarters
Tel: 949.595.4800 Fax: 949.595.4801
USA Toll Free
Tel: 800.854.3764
Fax: 800.752.9518
KNIGHT LLC,
Tel: 905.542.2333
Fax: 905.542.1536
A Unit of IDEX Corporation (www.knightequip.com)
Knight Canada
Knight Europe
Tel: 44.1293.615.570
Fax: 44.1293.615.585
Knight Australia
Tel: 61.2.9725.2588
Fax: 61.2.9725.2025
Knight N. Asia
Tel: 82.2.3481.6683
Fax: 82.2.3482.5742
Knight S. Asia
Tel: 65.6763.6633
Fax: 65.6764.4020
Page 8 of 8 0900900 Rev: D (05/07)
Loading...