Keithley 2015 Service Manual

Model 2015
THD Multimeter
Service Manual
SHIFT
LOCAL
POWER
SCAN
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 TALK LSTN SRQ SHIFT TIMER
MATH
DCV
EX TRIG
SAVE SETUP
SOURCE
CH1REM
HOLD TRIG FAST MED SLOW AUTO ERR
THD
ACV
HOLD
TRIG
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
dB
ACI
REL FILT
CONT
2 4
TEST
GPIB
DIGITS RATE
Contains Servicing Information
CAL
RELFILTER
RS232
MATH REAR
4W
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
AUTO
4 WIRE
350V PEAK
F
FRONT/REAR
SENSE
INPUTS
INPUT
HI
1000V
!
PEAK
LO
500V PEAK
R
3A 250V
AMPS
W ARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
T o e xercise this w arranty, write or call your local Keithle y representati v e, or contact K eithle y headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transporta­tion prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express writ­ten consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non­rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to fol­low instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICU­LAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REME­DIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARIS­ING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRU­MENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PER­SON, OR DAMAGE TO PROPERTY.
1/99
Model 2015 THD Multimeter
Service Manual
©1998, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Third Printing, June 1999
Document Number: 2015-902-01 Rev. C
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Le vel letter increases alphabetically as the manual under goes sub­sequent updates. Addenda, which are released between Revisions, contain important change in­formation that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Re vision is created, all Addenda associated with the previous Re vision of the manual are incorporated into the new Revision of the manual. Each ne w Revision includes a revised copy of this print history page.
Revision A (Document Number 2015-902-01).................................................................May 1998
Revision B (Document Number 2015-902-01)..................................................................July 1998
Revision C (Document Number 2015-902-01)................................................................. June 1999
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumen­tation. Although some instruments and accessories would normally be used with non-hazardous v oltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read the operating information carefully before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and
proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating, for example, setting
the line voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by ser­vice personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of products.
Only properly trained service personnel may perform installation and service procedures. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector
jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard e xists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit before measuring.
Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product users in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part
of the circuit may be exposed.
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multimeter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation Category II. All other instruments’ signal terminals are Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with imped­ance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge an y capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface ca­pable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documen­tation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the com­bined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accu­racy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water based cleaner . Clean the e xterior of the instrument only . Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and op­eration is affected, the board should be returned to the factory for proper cleaning/servicing.
Rev. 2/99
T able of Contents
1
Performance Verification
Introduction ........................................................................................ 1-2
Verification test requirements ............................................................ 1-3
Environmental conditions .............................................................. 1-3
Warm-up period ............................................................................. 1-3
Line power ..................................................................................... 1-3
Recommended test equipment ........................................................... 1-4
Verification limits ............................................................................... 1-5
Example reading limit calculation ................................................. 1-5
Calculating resistance reading limits ............................................. 1-5
Restoring factory defaults .................................................................. 1-6
Performing the verification test procedures ....................................... 1-7
Test summary ................................................................................. 1-7
Test considerations ......................................................................... 1-7
Verifying DC voltage ......................................................................... 1-8
Verifying AC voltage ....................................................................... 1-10
Verifying DC current ....................................................................... 1-12
Verifying AC current ........................................................................ 1-13
Verifying resistance ......................................................................... 1-14
Verifying temperature ...................................................................... 1-16
Verifying frequency ......................................................................... 1-17
Verifying total harmonic distortion .................................................. 1-18
Verifying function generator amplitude ........................................... 1-19
2
Calibration
Introduction ........................................................................................ 2-2
Environmental conditions .................................................................. 2-3
Warm-up period ............................................................................. 2-3
Line power ..................................................................................... 2-3
Calibration considerations ................................................................. 2-4
Calibration code ................................................................................. 2-5
Front panel calibration code ........................................................... 2-5
Remote calibration code ................................................................. 2-5
Comprehensive calibration ................................................................ 2-6
Calibration cycle ............................................................................ 2-6
Recommended equipment .............................................................. 2-6
Aborting calibration ....................................................................... 2-7
Front panel calibration ................................................................... 2-7
Preparing the Model 2015 for calibration ...................................... 2-7
Front panel short and open calibration ........................................... 2-8
DC volts calibration ....................................................................... 2-9
Resistance calibration .................................................................. 2-11
DC current calibration ................................................................. 2-12
AC voltage calibration ................................................................. 2-13
AC current calibration ................................................................. 2-14
Distortion calibration ................................................................... 2-14
Function generator calibration ..................................................... 2-15
Setting calibration dates and saving calibration .......................... 2-16
Remote calibration ....................................................................... 2-16
Preparing the Model 2015 for calibration .................................... 2-17
Short and open calibration ........................................................... 2-17
DC volts calibration ..................................................................... 2-18
Resistance calibration .................................................................. 2-19
DC current calibration ................................................................. 2-20
AC voltage calibration ................................................................. 2-21
AC current calibration ................................................................. 2-22
Distortion calibration ................................................................... 2-22
Function generator calibration ..................................................... 2-23
Programming calibration dates .................................................... 2-23
Saving calibration constants ........................................................ 2-23
Locking out calibration ................................................................ 2-23
Manufacturing calibration ............................................................... 2-24
Recommended test equipment ..................................................... 2-24
Unlocking manufacturing calibration .......................................... 2-24
Measuring synthesizer signal amplitude ...................................... 2-24
Front panel manufacturing calibration ......................................... 2-25
Remote manufacturing calibration .............................................. 2-26
3
Routine Maintenance
Introduction ....................................................................................... 3-2
Setting the line voltage and replacing the line fuse ....................... 3-2
Replacing the AMPS fuse .............................................................. 3-3
4
T roubleshooting
Introduction ....................................................................................... 4-2
Repair considerations ........................................................................ 4-3
Power-on self-test .............................................................................. 4-4
Front panel tests ................................................................................ 4-5
KEY test ......................................................................................... 4-5
DISP test ........................................................................................ 4-5
Principles of operation ...................................................................... 4-6
Power supply ................................................................................. 4-6
Display board ................................................................................. 4-8
Digital circuitry .............................................................................. 4-9
Analog circuitry ........................................................................... 4-11
Distortion digital circuitry ............................................................ 4-13
Distortion analog circuitry ........................................................... 4-15
Sine generator circuitry ................................................................ 4-16
Troubleshooting ............................................................................... 4-18
Display board checks ................................................................... 4-18
Power supply checks .................................................................... 4-19
Digital circuitry checks ................................................................ 4-20
Analog signal switching states ..................................................... 4-21
Disassembly
5
Introduction ........................................................................................ 5-2
Handling and cleaning ....................................................................... 5-3
Handling PC boards ....................................................................... 5-3
Solder repairs ................................................................................. 5-3
Static sensitive devices ................................................................... 5-4
Assembly drawings ............................................................................ 5-5
Disassembly procedures .................................................................... 5-6
Case cover removal ........................................................................ 5-6
DMM board removal ...................................................................... 5-6
DSP board removal ........................................................................ 5-7
Front panel disassembly ................................................................. 5-8
Removing power components ........................................................ 5-8
Instrument reassembly ....................................................................... 5-9
Input terminal wire connections ..................................................... 5-9
Power module wire connections .................................................... 5-9
Changing trigger link lines .............................................................. 5-10
Main CPU firmware replacement .................................................... 5-11
6
Replaceable Parts
Introduction ........................................................................................ 6-2
Parts lists ............................................................................................ 6-2
Ordering information ......................................................................... 6-2
Factory service ................................................................................... 6-2
Component layouts ............................................................................ 6-2
A
Specifications
Accuracy calculations ....................................................................... A-9
Calculating DC characteristics accuracy ....................................... A-9
Calculating AC characteristics accuracy ....................................... A-9
Calculating dBm characteristics accuracy .................................. A-10
Calculating dB characteristics accuracy ...................................... A-10
Distortion characteristics ............................................................. A-11
Calculating generator amplitude accuracy .................................. A-12
Additional derating factors ......................................................... A-12
Optimizing measurement accuracy ................................................ A-13
DC voltage, DC current, and resistance ...................................... A-13
AC voltage and AC current ......................................................... A-13
Temperature ................................................................................ A-13
Optimizing measurement speed ..................................................... A-14
DC voltage, DC current, and resistance ...................................... A-14
AC voltage and AC current ......................................................... A-14
Temperature ................................................................................ A-14
B
Calibration Reference
Introduction ...................................................................................... B-2
Command summary ......................................................................... B-3
Miscellaneous calibration commands .............................................. B-5
:CODE .......................................................................................... B-5
:COUNt? ....................................................................................... B-5
:INIT ............................................................................................. B-6
:LOCK .......................................................................................... B-6
:LOCK? ........................................................................................ B-7
:SAVE ........................................................................................... B-7
:DATE ........................................................................................... B-8
:NDUE .......................................................................................... B-8
DC calibration commands ................................................................ B-9
:STEP1 .......................................................................................... B-9
:STEP2 ........................................................................................ B-10
:STEP3 ........................................................................................ B-10
:STEP4 ........................................................................................ B-10
:STEP5 ........................................................................................ B-11
:STEP6 ........................................................................................ B-11
:STEP7 ........................................................................................ B-11
:STEP8 ........................................................................................ B-12
:STEP9 ........................................................................................ B-12
:STEP10 ...................................................................................... B-12
:STEP11 ...................................................................................... B-13
:STEP12 ...................................................................................... B-13
AC calibration commands .............................................................. B-14
:AC:STEP<n> ............................................................................. B-15
Distortion and function generator calibration commands ............. B-16
:DIST:STEP1 .............................................................................. B-16
:DIST:STEP2 .............................................................................. B-16
:FGEN:STEP1 ............................................................................ B-16
Manufacturing calibration commands ............................................ B-17
:AC:STEP<14|15> ...................................................................... B-17
:DC:STEP0 .................................................................................. B-17
Remote error reporting ................................................................... B-18
Error summary ............................................................................ B-18
Error queue .................................................................................. B-20
Status byte EAV (Error Available) bit ......................................... B-20
Generating an SRQ on error ........................................................ B-20
Detecting calibration step completion ............................................ B-21
Using the *OPC? query ............................................................... B-21
Using the *OPC command .......................................................... B-21
Generating an SRQ on calibration complete ............................... B-22
Calibration Program
C
Introduction ....................................................................................... C-2
Computer hardware requirements ..................................................... C-2
Software requirements ...................................................................... C-2
Calibration equipment ...................................................................... C-2
General program instructions ........................................................... C-3
List of Illustrations
1
Performance Verification
Connections for DC volts verification .............................................. 1-8
Connections for AC volts verification ............................................. 1-10
Connections for DC current verification ......................................... 1-12
Connections for AC current verification ......................................... 1-13
Connections for resistance verification (100 Ω -10M Ω range) ........ 1-14
Connections for resistance verification (100M Ω range) ................. 1-15
Connections for frequency verification ........................................... 1-17
Connections for total harmonic distortion verification ................... 1-18
Connections for function generator amplitude verification ............ 1-19
2
Calibration
Low-thermal short connections ......................................................... 2-8
Connections for DC volts and ohms calibration ............................... 2-9
Connections for DC and AC amps calibration ................................ 2-12
Connections for AC volts calibration .............................................. 2-13
Connections for distortion calibration ............................................. 2-15
Connections for function generator calibration ............................... 2-15
Synthesizer connections for manufacturing calibration .................. 2-25
Routine Maintenance
3
Power module .................................................................................... 3-3
T roubleshooting
4
Power supply block diagram ............................................................. 4-6
Digital circuitry block diagram ......................................................... 4-8
Analog circuitry block diagram ....................................................... 4-11
Distortion digital circuitry block diagram ....................................... 4-13
Distortion analog circuitry block diagram ...................................... 4-15
Sine generator circuitry block diagram ........................................... 4-16
5
Disassembly
Trigger link line connections ........................................................... 5-10
List of T ables
1
Performance Verification
Recommended verification equipment ........................................... 1-4
DCV reading limits ........................................................................ 1-9
ACV reading limits ....................................................................... 1-11
DCI limits ..................................................................................... 1-12
ACI limits ..................................................................................... 1-13
Limits for resistance verification .................................................. 1-15
Thermocouple temperature verification reading limits ................ 1-16
Calibration
2
Recommended equipment for comprehensive calibration ............. 2-6
Comprehensive calibration procedures .......................................... 2-8
DC volts calibration summary ...................................................... 2-10
Ohms calibration summary .......................................................... 2-11
DC current calibration summary .................................................. 2-12
AC voltage calibration summary .................................................. 2-13
AC current calibration summary .................................................. 2-14
Distortion and function generator calibration summary .............. 2-15
DC voltage calibration programming steps .................................. 2-18
Resistance calibration programming steps ................................... 2-19
DC current calibration programming steps .................................. 2-20
AC voltage calibration programming steps .................................. 2-21
AC current calibration programming steps .................................. 2-22
Distortion and function generator calibration steps ..................... 2-22
Recommended equipment for manufacturing calibration ............ 2-24
Ω 2/ Ω
Routine Maintenance
3
Power line fuse ............................................................................... 3-3
T roubleshooting
4
Power supply components .............................................................. 4-7
Display board checks ................................................................... 4-18
Power supply checks .................................................................... 4-19
Digital circuitry checks ................................................................ 4-20
DCV signal switching .................................................................. 4-21
ACV and FREQ signal switching ................................................. 4-21
2 signal switching ...................................................................... 4-22
4 signal switching ...................................................................... 4-22
4 reference switching .......................................................... 4-22
DCA signal switching .................................................................. 4-23
ACA signal switching .................................................................. 4-23
DCV signal multiplexing and gain .............................................. 4-23
ACV and ACA signal multiplexing and gain ............................... 4-23
DCA signal multiplexing and gain .............................................. 4-24
2 signal multiplexing and gain .................................................. 4-24
4 signal multiplexing and gain .................................................. 4-24
Switching device locations ............................................................ 4-25
5
Disassembly
Input terminal wire colors .............................................................. 5-9
Power module wire colors .............................................................. 5-9
6
Replaceable Parts
DMM (mother) board parts list ...................................................... 6-3
Display board parts list .................................................................. 6-9
Distortion (DSP) board parts list .................................................. 6-10
Mechanical parts list .................................................................... 6-14
B
Calibration Reference
Remote calibration command summary ....................................... B-3
DC calibration commands ............................................................. B-9
AC calibration commands ........................................................... B-14
Distortion and function generator calibration commands ........... B-16
Calibration error summary .......................................................... B-18
1
Performance
V erification
1-2 Performance Verification
Introduction
the limits stated in the instrument’s one-year accurac y specifications. Y ou can perform these ver ­ification procedures:
WARNING The information in this section is intended only for qualified service per-
Use the procedures in this section to verify that Model 2015 Multimeter accuracy is within
When you first receive the instrument to make sure that it w as not damaged during ship­ment, and that the unit meets factory specifications.
If the instrument’s accuracy is questionable.
Following calibration.
sonnel. Do not attempt these procedures unless you are qualified to do so.
NOTE If the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the correct course of action.
Performance Verification 1-3
V erification test requirements
Be sure that you perform the verification tests:
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper calibration equipment.
Using the specified reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment that has:
An ambient temperature of 18˚ to 28°C (65˚ to 82°F).
A relative humidity of less than 80% unless otherwise noted.
W arm-up period
Allow the Model 2015 Multimeter to warm up for at least one hour before conducting the v er-
ification procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2015 Multimeter requires a line voltage of 100V/120V/220V/240V, ±10% and a line frequency of 45Hz to 66Hz and 360Hz to 440Hz.
1-4 Performance Verification
Recommended test equipment
T able 1-1 summarizes recommended verification equipment. You can use alternate equipment as long as that equipment has specifications at least as good as those listed in Table 1-1. Keep in mind, however, that the calibrator will add to the uncertainty of each measurement.
Table 1-1
Recommended verification equipment
Fluke 5700A Calibrator:
AC voltage
DC voltage
100mV:±14ppm
1.0V:±7ppm 10V:±5ppm 100V:±7ppm 1000V:±9ppm
Fluke 5725A Amplifier:
AC Voltage, 50kHz: 700V, ±375ppm
Keithley 3930A or 3940 Frequency Synthesizer:
1V RMS, 1kHz, ±5ppm
Stanford Research Systems DS-360 Ultra Low Distortion Function Generator:
1kHz, .0.95V RMS sine wave, -100dB THD
Miscellaneous Equipment:
Double banana plug to double banana plug shielded cable BNC to double banana plug shielded cable
(1kHz, 50kHz) DC current
100mV:±200ppm
1.0V:±82ppm 10V:±82ppm 100V:±90ppm 700V:±85ppm
10mA:±60ppm 100mA:±70ppm 1A:±110ppm
2.2A:±94ppm
AC current (1kHZ) Resistance
1A:±690ppm
2.2A:±682ppm
100
:±17ppm
1k
:±12ppm
10k
:±11ppm
100k Ω :±13ppm 1M Ω :±18ppm 10M Ω :±37ppm 100M Ω :±120ppm
NOTE: The Fluke 5725A amplifier is necessary only if you wish to verify the 750V AC range at 50kHz.
Verification at 220V, 50kHz using only the 5700A calibrator is adequate for most applications.
×
V erification limits
The verification limits stated in this section have been calculated using only the Model 2015 one-year accuracy specifications, and they do not include test equipment uncertainty. If a par­ticular measurement falls slightly outside the allowable range, recalculate new limits based on both Model 2015 specifications and pertinent calibration equipment specifications.
Example reading limit calculation
The following is an example of how reading limits have been calculated:
Assume you are testing the 10V DC range using a 10V input value. Using the Model 2015 one-year accuracy specification for 10V DC of ± (30ppm of reading + 5ppm of range), the cal­culated limits are:
Performance Verification 1-5
Reading limits = 10V ± [(10V
Reading limits = 10V ± (.0003 + .00005)
Reading limits = 10V ± .00035V
Reading limits = 9.99965V to 10.00035V
30ppm) + (10V × 5ppm)]
Calculating resistance reading limits
Resistance reading limits must be recalculated based on the actual calibration resistance val­ues supplied by the equipment manufacturer . Calculations are performed in the same manner as shown in the preceding example, except, of course, that you should use the actual calibration resistance values instead of the nominal values when performing your calculations.
1-6 Performance Verification
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory defaults
as follows:
1. Press
2. Using either range key , select FACT , then restore the factory def ault conditions by press-
3. Factory defaults will be set as follows:
SHIFT and then SETUP . The instrument will display the following prompt:
RESTORE: FACT.
ing
ENTER .
Speed: medium Filter: 10 readings
Performance Verification 1-7
Performing the verification test procedures
T est summary
Verification test procedures include:
DC volts
AC volts
DC current
AC current
Resistance
Temperature
Frequency
Total harmonic distortion
Function generator amplitude
If the Model 2015 is not within specifications and not under warranty , see the calibration pro­cedures in Section 2.
T est considerations
When performing the verification procedures:
Be sure to restore factory defaults as outlined above.
After restoring factory defaults and selecting the measuring function, select the SLOW integration rate with the RATE key.
Make sure that the equipment is properly warmed up and connected to the front panel input jacks. Also mak e sure that the front panel input jacks are selected with the INPUTS switch.
Do not use autoranging for any verification tests because autorange hysteresis may cause the Model 2015 to be on an incorrect range. For each test signal, you must manually set the correct range for the Model 2015 using the range keys.
Make sure the calibrator is in operate before you verify each measurement.
Always let the source signal settle before taking a reading.
Do not connect test equipment to the Model 2015 through a scanner or other switching equipment.
WARNING The maximum common-mode voltage (voltage between INPUT LO and
chassis ground) is 500V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard. Some of the procedures in this sec­tion may expose you to dangerous voltages. Use standard safety precau­tions when such dangerous voltages are encountered to avoid personal injury caused by electric shock.
1-8 Performance Verification
V erifying DC voltage
Check DC voltage accuracy by applying accurate v oltages from the DC v oltage calibrator to the Model 2015 INPUT jacks and verifying that the displayed readings fall within specified limits.
CAUTION Do not exceed 1100V peak between INPUT HI and INPUT LO because
instrument damage may occur.
Follow these steps to verify DC voltage accuracy:
1. Connect the Model 2015 HI and LO INPUT jacks to the DC voltage calibrator as sho wn in Figure 1-1.
NOTE Use shielded, low-thermal connections when testing the 100mV and 1V ranges to
avoid errors caused by noise or thermal ef fects. Connect the shield to the calibrator’ s output LO terminal.
Figure 1-1
Connections for DC volts verification
Model 2015
LOCAL
POWER
TALK LSTN SRQ SHIFT TIMER
MATH
SHIFT
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
SCAN
CH1REM
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
REL FILT
dBm
dB
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
CONT
ACI
DIGITS RATE
THD
ACV
HOLD
TRIG
MEAS
2 4
TEST
GPIB
CAL
RELFILTER
RS232
MATH REAR
4W
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
RANGE
AUTO
350V PEAK
SENSE
4 WIRE
INPUTS
F
FRONT/REAR
Input HI
INPUT
HI
!
LO
R
3A 250V
AMPS
1000V
PEAK
500V PEAK
Output HI
Input
DC Voltage Calibrator
LO
Output LO
Note: Use shielded, low-thermal cables
for 100mV and 1V ranges.
Performance Verification 1-9
2. Select the DC volts function by pressing the DCV key, and set the Model 2015 to the 100mV range. Select the SLOW integration rate with the RATE key.
3. Set the calibrator output to 0.00000mV DC, and allow the reading to settle.
4. Enable the Model 2015 REL mode. Leave REL enabled for the remainder of the DC volts verification tests.
5. Source positive and negative and full-scale voltages for each of the ranges listed in Table 1-2. For each voltage setting, be sure that the reading is within stated limits.
Table 1-2
DCV reading limits
Range Applied DC voltage* Reading limits (1 year, 18°-28°C)
100mV 1V 10V 100V 1000V
* Source positive and negative values for each range.
100.0000mV
1.000000V
10.00000V
100.0000V
1000.000V
99.9915 to 100.0085mV
0.999963 to 1.000037V
9.99965 to 10.00035V
99.9949 to 100.0051V
999.949 to 1000.051V
1-10 Performance Verification
V erifying AC voltage
Check AC voltage accuracy by applying accurate AC voltages at specific frequencies from the A C v oltage calibrator to the Model 2015 inputs and verifying that the displayed readings f all within specified ranges.
CAUTION Do not exceed 1100 V peak between INPUT HI and INPUT LO, or 8 × 10
V•Hz input, because instrument damage may occur.
Follow these steps to verify AC voltage accuracy:
1. Connect the Model 2015 HI and LO INPUT jacks to the A C voltage calibrator as sho wn in Figure 1-2.
7
Figure 1-2
Connections for AC volts verification
Note: Amplifier required only
for 700V, 50kHz output.
Model 2015
LOCAL
POWER
TALK LSTN SRQ SHIFT TIMER
MATH
SHIFT
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
SCAN
CH1REM
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
REL FILT
HOLD TRIG FAST MED SLOW AUTO ERR
dBm
dB
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
CONT
ACI
2 4
TEST
GPIB
DIGITS RATE
THD
ACV
HOLD
TRIG
MEAS
RELFILTER
RS232
CAL
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
EXIT ENTER
MATH REAR
4W
TEMP
RANGE
AUTO
RANGE
SENSE
INPUT
4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V PEAK
INPUTS
F
R
FRONT/REAR
3A 250V
AMPS
Shielded cable
Input HI
Output HI
Input LO
Output LO
AC Voltage Amplifier
AC Voltage Calibrator
Performance Verification 1-11
2. Select the AC volts function by pressing the ACV key, then choose the SLOW integra­tion rate with the RATE key.
3. Set the Model 2015 for the 100mV range; make sure that REL is disabled.
4. Source 1kHz and 50kHz AC voltages for each of the ranges summarized in Table 1-3, and make sure that the respective Model 2015 readings fall within stated limits.
Table 1-3
ACV reading limits
ACV range
100mV 1V 10V 100V 750V
* If the 5725A amplifier is not available, change the 700V @ 50kHz step to 220V @
50kHz. Reading limits for 220V @ 50kHz = 219.36 to 220.64V.
Applied AC voltage
100.0000mV
1.000000V
10.00000V
100.0000V
700.000V*
1kHz reading limits (1 year, 18°C-28°C)
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
699.36 to 700.64V
50kHz reading limits (1 year, 18°C-28°C)
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
698.79 to 701.21V
1-12 Performance Verification
V erifying DC current
Check DC current accuracy by applying accurate DC currents from the DC current calibrator to the AMPS input of the Model 2015 and v erifying that the displayed readings f all within spec­ified limits.
Follow these steps to verify DC current accuracy:
1. Connect the Model 2015 AMPS and INPUT LO jacks to the calibrator as shown in Figure 1-3.
Figure 1-3
Connections for DC current verification
LOCAL
POWER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
TALK
LSTN
SRQ SHIFT TIMER
HOLD TRIG FAST MED SLOW AUTO ERR
MATH
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
ACV
HOLD
TRIG
MEAS
THD
SHIFT
DC Current Calibrator
Model 2015
SCAN
CH1REM
REL FILT
dBm
dB
ACI
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
CONT
2 4
TEST
GPIB
DIGITS RATE
CAL
RELFILTER
RS232
MATH REAR
4W
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
RANGE
SENSE
INPUT
4 WIRE
HI
350V
!
PEAK
LO
INPUTS
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
1000V PEAK
500V PEAK
Input LO
Output HI
Amps
Output LO
Note: Be sure calibrator is set for
normal current output.
2. Select the DC current measurement function by pressing the DCI key, then choose the SLOW integration rate with the RATE key.
3. Set the Model 2015 for the 10mA range.
4. Source positive and negati ve full-scale currents for each of the ranges listed in Table 1-4, and verify that the readings for each range are within stated limits.
Table 1-4
DCI limits
DCI range Applied DC current* Reading limits (1 year, 18°C-28°C)
10mA 100mA 1A 3A
10.0000mA
100.0000mA
1.000000A
2.20000A
9.99460 to 10.00540mA
99.9100 to 100.0900mA
0.999160 to 1.000840A
2.197315 to 2.202685A
*Source positive and negative currents with values shown.
V erifying AC current
Check AC current accuracy by applying accurate AC voltage current at specific frequencies from the A C current calibrator to the Model 2015 input and v erifying that the displayed readings fall within specified limits. Follow these steps to verify AC current:
1. Connect the Model 2015 AMPS and INPUT LO jacks to the calibrator as shown in Figure 1-4.
Figure 1-4
Connections for AC current verification
Model 2015
SENSE
SHIFT
LOCAL
POWER
MATH
DCV
EX TRIG
SOURCE
TALK LSTN SRQ SHIFT TIMER
SAVE SETUP
THD
SCAN
CH1REM
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
dBm
THD
ACV
DCI
HOLD
LIMITS ON/OFFDELAY
TRIG
STORE
RECALL
CONFIG HALT
MEAS
STEP SCAN
BUFFER
REL FILT
dB
CONT
ACI
2 4
TEST
CAL
RELFILTER
RS232
GPIB
DIGITS RATE
MATH REAR
4W
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
RANGE
4 WIRE
350V PEAK
INPUTS
F
AUTO
FRONT/REAR
Amps
Performance Verification 1-13
INPUT
HI
!
LO
R
3A 250V
AMPS
1000V PEAK
500V
PEAK
Output HI
Input LO
Output LO
AC Current Calibrator
2. Select the AC current function by pressing the ACI key , then choose the SLOW integra­tion rate with the RATE key.
3. Set the Model 2015 for the 1A range.
4. Source 1A and 2.2A, 1kHz full-scale AC currents as summarized in Table 1-5, and verify that the readings are within stated limits.
Table 1-5
ACI limits
ACV range Applied AC voltage Reading limits @ 1kHz (1 year, 18°C-28°C)
1A 3A
1.000000A
2.20000A
0.99860 to 1.00140A
2.1949 to 2.2051A
1-14 Performance Verification
V erifying resistance
Check resistance by connecting accurate resistance values to the Model 2015 and verifying
that its resistance readings are within the specified limits.
CAUTION Do not apply more than 1100V peak between INPUT HI and LO or more
than 350V peak between SENSE HI and LO, or instrument damage could occur
Follow these steps to verify resistance accuracy:
1. Using shielded 4-wire connections, connect the Model 2015 INPUT and SENSE jacks to the calibrator as shown in Figure 1-5.
.
Figure 1-5
Connections for resistance verification (100
Sense HI
Model 2015
SENSE
LOCAL
POWER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
TALK
LSTN
SRQ SHIFT TIMER
HOLD TRIG FAST MED SLOW AUTO ERR
MATH
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
ACV
HOLD
TRIG
MEAS
THD
SHIFT
SCAN
CH1REM
REL FILT
dBm
dB
CONT
ACI
2 4
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
DIGITS RATE
MATH REAR
4W
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
TEST
CAL
RELFILTER
RS232
GPIB
EXIT ENTER
RANGE
RANGE
4 WIRE
350V PEAK
INPUTS
F
AUTO
FRONT/REAR
Sense LO
2. Set the calibrator for 4-wire resistance with external sense on.
3. Select the Model 2015 4-wire resistance function by pressing the the SLOW integration rate with the RATE key.
4. Set the Model 2015 for the 100
5. Recalculate reading limits based on actual calibrator resistance values.
-10M
ranges)
Sense HI
INPUT
HI
!
LO
R
3A 250V
AMPS
Input HI
1000V PEAK
500V PEAK
Input LO
Output HI
Output LO
Resistance Calibrator
Sense LO
Note: Use shielded low-thermal cables to
minimize noise. Enable or disable calibrator external sense as indicated in procedure.
4 key, then choose
range, and make sure the FILTER is on.
6. Source the nominal full-scale resistance values for the 100 Ω -10M Ω ranges summarized in Table 1-6, and verify that the readings are within calculated limits.
7. Connect the Model 2015 INPUT and SENSE jacks to the calibrator as shown in Figure 1-6.
8. Disable external sense on the calibrator.
9. Set the Model 2015 for the 100M
10. Source a nominal 100M Ω resistance value, and verify that the reading is within cal­culated limits for the 100M Ω range.
Figure 1-6
Connections for resistance verification (100M
range)
range.
Performance Verification 1-15
1k Ω
Model 2015
2 4
TEST
GPIB
CAL
RELFILTER
RS232
MATH REAR
BUFFER
STAT
2015 THD MULTIMETER
PERIOD TCOUPL
FREQ
EXIT ENTER
4W
TEMP
POWER
TALK LSTN SRQ SHIFT TIMER
MATH
SHIFT
DCV
LOCAL
EX TRIG
SAVE SETUP
SOURCE
THD
SCAN
CH1REM
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
REL FILT
dBm
dB
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
CONT
ACI
DIGITS RATE
THD
ACV
HOLD
TRIG
MEAS
Sense LO
Table 1-6
Limits for resistance verification
Nominal
Range
100 Ω
resistance
100 1k
10k Ω 100k 1M 10M 100M
10k 100k 1M 10M 100M
Sense HI
RANGE AUTO RANGE
SENSE
INPUT
4 WIRE
HI
350V
!
PEAK
LO
INPUTS
F
R
FRONT/REAR
3A 250V
AMPS
Input HI
1000V
PEAK
500V PEAK
Input LO
Output HI
Output LO
Resistance Calibrator
Note: Use shielded cables to minimize noise.
Disable calibrator external sense mode.
Nominal reading limits (1 year, 18°C-28°C) Recalculated limits*
99.9860 to 100.0140
0.999890 to 1.000110k
9.99890 to 10.00110k
99.9890 to 100.0110k
0.999890 to 1.000110M
9.99590 to 10.00410M
99.8470 to 100.1530M
__________ to __________ __________ to __________ k __________ to __________ k __________ to __________ k __________ to __________ M __________ to __________ M __________ to __________ M
* Calculate limits based on actual calibration resistance values and Model 2015 one-year accuracy
specifications. See Verification limits.
1-16 Performance Verification
V erifying temperature
Thermocouple temperature readings are derived from DC volts measurements. For that rea­son, it is not necessary to independently verify the accuracy of temperature measurements. As long as the DC volts function meets or exceed its specifications, temperature function accuracy is automatically verified. Howe ver, temperature verification procedures are provided below for those who wish to separately verify temperature accuracy.
1. Connect the DC voltage calibrator output terminals to the Model 2015 INPUT jacks using low-thermal shielded connections. (Use 2-wire connections similar to those sho wn in Figure 1-1.)
2. Configure the Model 2015 for °C units, type J temperature sensor, and 0°C simulated ref­erence junction as follows:
A. Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C. (If necessary, use the cursor and range keys to select °C units.) B. Press ENTER. The unit then displays the thermocouple type: TYPE: J. C. Select a type J temperature sensor, then press ENTER. The unit then displays the
reference junction type: JUNC: SIM. D. Make certain that the simulated reference junction type is selected, then press
ENTER. The unit then displays the current simulated reference junction tempera-
ture: SIM: 023. E. Using the cursor and range keys, set the reference junction temperature to 0°C, then
press ENTER to complete the temperature configuration process.
3. Select the temperature function by pressing the TEMP key.
4. Source each of the voltages summarized in Table 1-7, and verify that the tempera­ture readings are within limits. Be sure to select the appropriate thermocouple type for each group of readings. (See step 2 above.)
Table 1-7
Thermocouple temperature verification reading limits
Thermocouple type
J
K
* Voltages shown are based on ITS-90 standard using 0°C reference
junction temperature. See text for procedure to set reference junction temperature.
Applied DC voltage*
-7.659mV 0mV
42.280mV
-5.730mV 0mV
54.138mV
Reading limits (1 year, 18°C-28°C)
-190.6 to -189.4°C
-0.5 to +0.5°C
749.5 to 750.5°C
-190.6 to -189.4°C
-0.5 to +0.5°C
1349.2 to 1350.8°C
Loading...
+ 128 hidden pages