2010
• 7½-digit resolution
• 100nV rms noise floor
• 7ppm DCV repeatability
• Built-in 10-channel scanner
mainframe
• Dry circuit and low power
measurement mode
• 15 measurement functions
including support for RTD and
thermocouple temperature
measurements
• Built-in ratio measurement
function
Low Noise 7½-Digit
Ordering Information
2010 Autoranging DMM
Accessories Supplied
Model 1751 Safety Test Leads,
User Manual, Service Manual
SERVICES AVAILABLE
2000-SCAN-3Y-EW 1-year factory warranty extended to 3 years
2001-TCSCAN-3Y-EW 1-year factory warranty extended to 3 years
2010-3Y-EW 1-year factory warr anty extended to 3 years
C/2000-3Y-ISO 3 (ISO -17025 accredited) calibrations within
C/2001-3Y-ISO 3 (ISO -17025 accredited) calibrations
C/2010-3Y-ISO 3 (ISO -17025 accredited) calibrations within
*Not available in all countries
te of shipment
from da
from date of shipment
from date of shipment
3 years of purchase for Model 2000-SCAN*
within 3 years of purchase for Model 2001TCSCAN*
3 years of purchase for Mo
del 2010*
Autoranging Multimeter
The 7½-digit Model 2010 Low Noise Multi meter combines high
resolution with the high speed and accuracy needed for production applications such as testing precision sensors, transducers,
A/D and D/A con verters, regulators, references, connectors,
switches, and relays. It is based on the same high speed, low
noise A/D converter technology as the Models 2000, 2001,
and 2002.
High Measurement Flexibility
The 2010 has 15 built-in measurement functions, including DCV,
W
ACV, DCI, ACI, 2W
(with either thermocouples or RTDs), frequency, period, ratio,
continuity meas ure ment, and diode testing. This multi-functional
design minimizes added equip ment costs.
Creating a self-contained multipoint measurement solution is as
simple as plugging a 2000-SCAN or 2001-TCSCAN scanner card into the option slot in the 2010’s back
panel. This “plug-in” approach eliminates the need for a separate scanner and significantly reduces
programming and setup time in applications involving a limited number of test points. For larger
applications, the 2010 is compatible with Keithley’s Series 7000 switch matrices and cards.
Unique Resistance Measurement Functions
Characterizing the resistance, linearity, or isolation of contacts, connectors, switches, or relays
complete ly and efficiently demands an uncommon combination of ohms measurement capa bilities.
The 2010 offers:
• A low-power ohms measurement mode. Low-level resistance measurements can be made with
source current as low as 100µA, an order of magnitude lower than is possible with other DMMs,
so device self-heating is minimized. Among other benefits, this low-power measurement capability
makes the 2010 suitable for end-of-life contact testing per ASTM B539-90.
• A dry circuit test function. When measuring contact and connector resistances, it is important
to control the test voltage carefully in order to avoid puncturing any oxides or films that may have
formed. A built-in clamp limits the open circuit test voltage to 20mV to ensure dry circuit conditions.
• An offset compensated ohms function. This function eliminates thermal effects that can create
errors in low-level resistance measurements in system environments.
• An extended ohms measurement capability. The 2010 provides a 10W range for more precise
measurements of low resistances.
Optional Multiplexer Cards
Creating a self-contained multipoint measurement solution is as simple as plugging a scanner
card into the option slot on the 2010’s back
panel. This ap proach eliminates the complexities
of triggering, timing, and processing issues and
helps reduce test time significantly. For applications involving more than 10 measurement
points, the 2010 is com patible with Keithley’s
Series 7000 switch matrices and cards.
Model 2000-SCAN Scanner Card
• Tenanaloginputchannels(2-pole)
• Configurableas4-pole,5-channel
, 4WW, dry circuit resistance, temperature
ACCESSORIES AVAILABLE
TEST LEADS
5804/5/6 4-Wire/Kelvin Test Lead Sets
SWITCH/SCANNER CARDS
2000 -SCAN 10-Cha nnel Scanner
2001-TCSCAN 9-Channel Thermocouple Scanner
CABLES/ADAPTERS
7007-1 Shielded IEEE-488 Cable, 1m (3.3 f t)
7007-2 Shielded IEEE-488 Cable, 2m (6.6 f t)
7009-5 RS-232 Cable
RACK MOUNT KITS
4288-1 Single Fi xed Rack Mount Kit
4288-2 Dual Fi xed Rack Mount Kit
GPIB INTERFACES
KPCI- 488LPA IEEE- 488 Interface/Control ler for the PCI Bus
KPXI-488 IEEE- 488 Interface Board for the PXI Bus
KUSB- 488A IEEE-488 USB-to-GPIB Inter face Adapter
DMM optimized for resistance measurement applications
1.888.KEITHLEY
www.keithley.com
(U.S. only)
A GREATER MEASURE OF CONFIDENCE
DIGITAL MULTIMETERS & SYSTEMS
2010
Low Noise 7½-Digit
Autoranging Multimeter
DC VOLTAGE
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
100.00000 mV 10 nV 25 + 9 37 + 9
1.0000000 V 100 nV 18 + 2 25 + 2
10.000000 V 1 µV 18 + 4 24 + 4
100.00000 V 10 µV 25 + 5 35 + 5
1000.0000 V 10 µV 31 + 6 41 + 6
RESISTANCE
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
10.000000 W 1 µW
100.00000 W 10 µW
1.0000000 kW 100 µW
10.000000 kW 1 mW
100.00000 kW 10 mW
1.0000000 MW 100 mW
10.000000 MW 1 W
100.00000 MW 10 W
40 + 9 60 + 9 10 mA
36 + 9 52 + 9 1 mA
33 + 2 50 + 2 1 mA
32 + 2 50 + 2 100 µA
40 + 2 70 + 2 10 µA
50 + 4 70 + 4 10 µA
200 + 4 400 + 4 640 nA
1500 + 4 1500 + 4 640 nA
DC CURRENT
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
10.000000 mA 1 n A 300 + 40 500 + 40 < 0.15 V
Model 2010 specifications
100.00000 mA 10 n A 300 + 40 500 + 40 < 0.18 V
1.0000000 A 100 nA 500 + 40 800 + 40 < 0.35 V
3.000000 A 1 µA 1200 + 15 1200 + 15 < 1 V
CONTINUITY 2W
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
1 kW 100 m W
100 + 100 120 + 100 1 mA
DIODE TEST
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
10.000000 V 1 µV 30 + 7 40 + 7 1 mA
4.400000 V 1 µV 30 + 7 40 + 7 100 µA
10.000000 V 1 µV 30 + 7 40 + 7 10 µA
DC OPERATING CHARACTERISTICS
Function Digits Readings/s PLCs
DCV (all ranges),
DCI (all ranges), and
Ohms (<10M range)
7½ 4 (3) 5
6½ 30 (27 ) 1
6½ 50 (44 ) 1
5½ 260 (220) 0 .1
5½ 490 (44 0) 0 .1
5½ 1000 (1000) 0.04
4½ 2000 (18 00) 0.01
SPEED AND NOISE REJECTION
Rate Digits
5 PLC 7½ 110 nV 1.2 µV 60 dB 140 d B
1 PLC 6½ 125 nV 1.4 µV 60 dB 140 dB
0.1 PLC 5½ 1.6 µV 11.5 µV — 80 dB
0.01 PLC 4½ 2.9 µV 139 µV — 80 dB
RMS Noise
100mV Range
RMS Noise
10V Range NMRR CMRR
Input
ResistanceRange Resolution 90 Day 1 Year
> 10 GW
> 10 GW
> 10 GW
10 MW ±1%
10 MW ±1%
Test CurrentRange Resolution 90 Day 1 Year
Burden
VoltageRange Resolution 90 Day 1 Year
Test CurrentRange Resolution 90 Day 1 Year
Test CurrentRange Resolution 90 Day 1 Year
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
Accuracy (1 Year)
23°C ±5°C
±(% of reading + % of range)
Range Resolution
100 mV to 750 V 0.1 µV to 1 mV
Frequency
Range
3 Hz–10 Hz 0.35 + 0.03
10 Hz–20 kHz 0.06 + 0.03
20 kHz–50 kHz 0.12 + 0.05
50 kHz–100 kHz 0.60 + 0.08
100 kHz–300 kHz 4 + 0.5
AC OPERATING CHARACTERISTICS
Function Digits Readings/s Rate Bandwidth
ACV (all ranges), and
ACI (all ranges)
6½ 2s/reading SLOW 3 Hz–300 kHz
6½ 1.4 MED 30 Hz–300 kHz
6½ 4.8 MED 30 Hz–300 kHz
6½ 2.2 FAST 300 Hz–300 kHz
6½ 35 FAST 300 Hz–300 kHz
FREQUENCY AND PERIOD CHARACTERISTICS
ACV
Range
100 mV to
750 V
Frequency
Range
3 Hz to
500 kHz
Period
Range
333 ms to
2 µs
Resolution
Gate
Time
±(ppm of
reading)
1 s 0.3 0. 01
Accuracy
90 Day/1 Year
±(% of reading)
TEMPERATURE CHARACTERISTICS
Thermocouple Accuracy
Type Range Resolution
J –200 to + 760°C 0.001°C ±0 .5°C ±0.65°C
K –200 to + 1372°C 0.001° C ±0.5 °C ±0.7 0°C
N –200 to + 1300°C 0 .001°C ± 0.5°C ± 0.70° C
T –200 to + 400°C 0.0 01°C ±0.5 °C ±0.68°C
4-WIRE RTD Accuracy
–100° to +100°C 0.001° C ±0.08°C ±0.1 2°C
–200° to +630°C 0.001°C ±0.14° C ±0.18 °C
TEMPERATURE NOTES
1. For temp eratures <–100 °C, add ±0.1°C and >9 00°C add ±0.3°C.
2. Speci fications appl y to channels 2–6. Add 0.0 6°C/channel fro m channel 6.
3. Exclud ing probe err ors.
90 Day/1 Year (23°C ± 5°C)
Relative to
Reference Junction
90 Day/1 Year
1
(23°C ± 5°C)
3
USING
2001-TCSC AN
Accuracy
2 Years
(23°C ± 5°C)Range Resolution
GENERAL
POWER SU PPLY: 100V / 120V / 220V / 240V ±10%.
LINE FREQUENCY: 45Hz to 66Hz and 360Hz to 440Hz, automatically sensed at power-up.
POWER CONSUMPTION: 22 VA.
OPERATING ENVIRONMENT: Specified for 0°C to 50°C. Speci fied to 80% R.H. at 35°C.
STORAGE ENVIRONMENT: –40°C to 70°C.
SA FET Y: Designed to IEC-1010-1.
EMC: Complies with European Union Directive 89/336/EEC (CE marking requirements), FCC
part 15 class B, CTSPR 11, IEC 801-2, IEC 801-3, IEC 801-4.
VIBRATION: MIL-T-28800E Type III, Class 5.
WAR MUP: 2 hours to rated accuracy.
DIMENSIONS:
Rack Mounting: 89mm high × 213mm wide × 370mm deep (3
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 370mm
1
⁄8 in × 93⁄8 in × 149⁄16 in).
deep (4
SHIPPING WEIGHT: 5kg (11 lbs).
VOLT HERTZ PRODUCT: ≤8 × 10
7
V·Hz .
½
in × 83⁄8 in × 149⁄16 in).
2
3
1.888.KEITHLEY
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
(U.S. only)
A GREATER MEASURE OF CONFIDENCE