Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years from
date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries,
diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local K eithley representative, or contact K eithley headquarters in Cleveland, Ohio. You
will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility.
Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance
of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse
of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN
ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN AD VANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS
OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PR OPERTY .
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision
Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are
incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2002-900-01).........................................................................................May 1994
Addendum A (Document Number 2002-900-02)..................................................................................... May 1995
Addendum A (Document Number 2002-900-03)................................................................................October 1995
Addendum A (Document Number 2002-900-04).................................................................................. March 1996
Revision B (Document Number 2002-900-01)............................................................................... September 1997
Revision C (Document Number 2002-900-01).........................................................................................June 1998
Revision D (Document Number 2002-900-01) .....................................................................................March 1999
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand and product names are trademarks or registered trademarks of their respective holders
Safety Precautions
The following safety precautions should be observed before using
this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions
may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read the operating information
carefully before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with
hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in
the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service
personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures.
Users of this product must be protected from electric shock at all
times. The responsible body must ensure that users are prevented
access and/or insulated from every connection point. In some cases,
connections must be exposed to potential human contact. Product
users in these circumstances must be trained to protect themselves
from the risk of electric shock. If the circuit is capable of operating
at or above 1000 volts, no conductive part of the circuit may be
exposed.
As described in the International Electrotechnical Commission
(IEC) Standard IEC 664, digital multimeter measuring circuits
(e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are
Installation Category II. All other instruments’ signal terminals are
Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits.
They are intended to be used with impedance limited sources.
NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting
cables, test leads, and jumpers for possible wear, cracks, or breaks
before each use.
For maximum safety, do not touch the product, test cables, or any
other instruments while power is applied to the circuit under test.
ALWAYS remove power from the entire test system and discharge
any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal
changes, such as installing or removing jumpers.
Exercise extreme caution when a shock hazard is present. Lethal
voltage may be present on cable connector jacks or test fixtures. The
American National Standards Institute (ANSI) states that a shock
hazard exists when voltage levels greater than 30V RMS, 42.4V
peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit bef ore
measuring.
Do not touch any object that could provide a current path to the
common side of the circuit under test or power line (earth) ground.
Always make measurements with dry hands while standing on a
dry, insulated surface capable of withstanding the voltage being
measured.
The instrument and accessories must be used in accordance with its
specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or
switching card.
When fuses are used in a product, replace with same type and rating
for continued protection against fire hazard.
Chassis connections must only be used as shield connections for
measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a
lid interlock.
If a screw is present, connect it to safety earth ground using the
wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal
and common mode voltages. Use standard safety precautions to
avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might
result in personal injury or death. Alw ays read the associated infor mation very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could
damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and fire, replacement
components in mains circuits, including the power transformer, test
leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals,
may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as
long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments
to maintain accuracy and functionality of the product.) If you are
unsure about the applicability of a replacement component, call a
Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based
cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill
on the instrument. Products that consist of a circuit board with no
case or chassis (e.g., data acquisition board for installation into a
computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper
cleaning/servicing.
1.3Warranty information ......................................................................................................................................... 1-2
1.8Options and accessories ..................................................................................................................................... 1-2
2.2.1Line power connections ............................................................................................................................. 2-1
2.2.6Warm-up period ......................................................................................................................................... 2-4
2.4.1DC and AC voltage .................................................................................................................................... 2-9
2.4.2DC and AC current .................................................................................................................................. 2-23
2.4.3Two and four-wire resistance ................................................................................................................... 2-28
2.7.1Trigger model ........................................................................................................................................... 2-41
2.7.2Configuring the measure layer ................................................................................................................. 2-46
2.7.3Configuring the scan layer ........................................................................................................................ 2-48
2.7.4Configuring the arm layer ........................................................................................................................ 2-49
2.7.7Trigger Link .............................................................................................................................................. 2-53
2.8.2Configuring data storage .......................................................................................................................... 2-67
2.8.3Storing and recalling readings .................................................................................................................. 2-69
2.9.3Response time ........................................................................................................................................... 2-73
2.9.5Configuring the filters .............................................................................................................................. 2-74
2.10.4Selecting and configuring math ................................................................................................................ 2-76
3.2IEEE-488 bus connections ................................................................................................................................. 3-1
3.5General bus commands ....................................................................................................................................... 3-4
3.5.8SPE, SPD (serial polling) ........................................................................................................................... 3-5
3.6Front panel aspects of IEEE-488 operation ....................................................................................................... 3-5
3.6.1Error and status message ............................................................................................................................ 3-5
3.6.2IEEE-488 status indicators ......................................................................................................................... 3-5
3.10.10*SAV — save ........................................................................................................................................... 3-36
3.10.11*SRE <NRf> — service request enable ................................................................................................... 3-36
*SRE? — service request enable query ................................................................................................... 3-36
3.10.12*STB? — status byte query ..................................................................................................................... 3-38
ASpecifications
BInterface Function Codes
CASCII Character Codes and IEEE-488 Multiline Interface Command Messages
DIEEE-488 Bus Overview
EIEEE-488 Conformance Information
FSCPI Conformance Information
GExample Programs
HHP3458A Emulation Mode
Figure 2-14External triggering and asynchronous trigger link input pulse specifications ......................................... 2-50
Figure 2-15Meter complete and asynchronous trigger link output pulse specifications ............................................ 2-51
Figure 2-16DUT test system ....................................................................................................................................... 2-51
Figure 2-18Trigger link connectors ............................................................................................................................ 2-53
Figure 2-19DUT test system ....................................................................................................................................... 2-54
Figure 2-20Trigger Link connections (asynchronous example #1) ............................................................................ 2-54
Figure 2-21Operation model for asynchronous trigger link example #1 .................................................................... 2-56
Figure 2-22Connections using Trigger Link adapter .................................................................................................. 2-57
Figure 2-23DUT test system (asynchronous example #2) .......................................................................................... 2-57
Figure 2-24Trigger Link connections (asynchronous example #2) ............................................................................ 2-58
Figure 2-25Operation model for asynchronous Trigger Link example #2 ................................................................. 2-60
Figure 2-26Semi-synchronous Trigger Link specifications ....................................................................................... 2-61
Figure 2-28Trigger Link connections (semi-synchronous example) .......................................................................... 2-62
Figure 2-29Operation mode for semi-synchronous Trigger Link example ................................................................ 2-63
Figure 2-30Averaging and advanced filter types ........................................................................................................ 2-72
Figure 2-31Moving and repeating filter modes .......................................................................................................... 2-73
Figure 2-32SCAN key menu structure ....................................................................................................................... 2-82
Figure 2-33SCAN key menu structure for ratio and delta .......................................................................................... 2-83
Figure 2-34Limits bar graph example ........................................................................................................................ 2-98
Figure 2-35Using limit test to start 100 Ω resistor ...................................................................................................... 2-99
Figure 2-36Digital I/O port ....................................................................................................................................... 2-100
Figure 2-37Digital I/O port simplified schematic ..................................................................................................... 2-101
Figure 3-4Model 2002 status register structure ........................................................................................................... 3-7
Figure 3-5Standard event status .................................................................................................................................. 3-9
Figure 3-6Operation event status .............................................................................................................................. 3-10
Figure 3-7Arm event status ....................................................................................................................................... 3-11
Figure 3-8Sequence event status ............................................................................................................................... 3-12
Figure 3-9Trigger event status .................................................................................................................................. 3-13
Figure 3-10Measurement event status ......................................................................................................................... 3-14
Figure 3-11Questionable event status ......................................................................................................................... 3-15
Figure 3-12Status byte and service request (SRQ) ..................................................................................................... 3-17
Figure 3-13Trigger Model (IEEE-488 bus operation) ................................................................................................ 3-19
Figure 3-18ASCII data format .................................................................................................................................... 3-79
Figure 3-19IEEE754 single precision data format (32 data bits) ................................................................................ 3-80
Figure 3-20IEEE754 double precision data format (64 data bits) .............................................................................. 3-81
Table 2-3Multiple displays by function .................................................................................................................... 2-5
Table 2-4Status and error messages .......................................................................................................................... 2-8
Table 2-6CONFIGURE DCV menu structure ........................................................................................................ 2-10
Table 2-7CONFIGURE ACV menu structure ........................................................................................................ 2-11
Table 2-8DCV and ACV integration times set-by-resolution ................................................................................. 2-12
Table 2-9DCV and ACV auto filter ........................................................................................................................ 2-13
Table 2-10DCV and ACV auto resolution ................................................................................................................ 2-13
Table 2-11CONFIGURE DCI menu structure .......................................................................................................... 2-24
Table 2-12CONFIGURE ACI menu structure .......................................................................................................... 2-24
Table 2-13DCI and ACI integration time set-by-resolution ...................................................................................... 2-25
Table 2-14DCI and ACI auto filter ........................................................................................................................... 2-25
Table 2-15DCI and ACI auto resolution ................................................................................................................... 2-26
Table 2-16CONFIGURE OHMS-2W menu structure .............................................................................................. 2-29
Table 2-17CONFIGURE OHMS-4W menu structure .............................................................................................. 2-29
Table 2-18
Table 2-19
Table 2-20
Table 2-21Trigger level range and increments .......................................................................................................... 2-32
Table 2-22CONFIGURE FREQUENCY menu structure ......................................................................................... 2-33
Table 2-23CONFIG TEMPERATURE menu structure ............................................................................................ 2-36
Table 2-24Temperature integration time set-by-resolution ....................................................................................... 2-39
Table 2-25Temperature auto resolution .................................................................................................................... 2-39
2 and Ω 4 integration time set-by-resolution ......................................................................................... 2-30
2 and Ω 4 auto filter ............................................................................................................................... 2-30
2 and Ω 4 auto resolution ....................................................................................................................... 2-31
ix
3IEEE-488 Reference
Table 3-1General bus commands and associated statements .................................................................................... 3-4
Table 3-2IEEE-488.2 common commands and queries .......................................................................................... 3-29
This section contains general information about the Model
2002 Multimeter. It is arranged in the following manner:
1.2 Features
1.3 Warranty information
1.4 Manual addenda
1.5 Safety symbols and terms
1.6 Specifications
1.7 Inspection
1.8 Options and accessories
1.2Features
Some important Model 2002 features include:
• Full range of functions Among other functions, the
multimeter can measure DC voltage (normal and peak
spikes), AC voltage (RMS, average, and peak), DC
current (normal and in-circuit), AC current (RMS and
average), two and four-wire resistance (normal and
offset-compensated), frequency (voltage and current),
and temperature (resistance temperature devices or
thermocouples).
• T wo-line display Readings and front panel messages
are shown on an alphanumeric display having a 20character top line and a 32-character bottom line.
• Multifunction measuring and display From the front
panel, you can configure the instrument to sequentially
measure and simultaneously display readings of multiple functions.
• Reading and setup storage Readings and setup data
can be stored and recalled from the front panel or over
the IEEE-488 bus. For example, the buffer can be programmed to store up to 850 readings at 4.5 digits, or up
to 250 time-stamped readings at 6.5 digits. The Model
2002 can be configured with memory options that
extend the storage capacity up to 30,000 readings and
ten setups.
• High-speed measurements The instrument is capable
of acquiring, for example, 2000 readings/second at 4.5
digits of resolution, and 215 readings/second at 6.5
digits.
• T alk-only mode From the front panel, you can set the
instrument to send readings to an IEEE-488 printer or,
with an optional adapter, to a Centronics printer.
• Digital calibration The instrument may be digitally
calibrated from either the front panel or over the bus.
• Standard IEEE-488 interface Bus operation conforms to the IEEE-488.2 and SCPI standards.
• Trigger link This is a new trigger concept that provides more versatile and precise external triggering. It
is in addition to the standard Trigger In/Measurement
Complete BNC external triggering techniques.
• Optional field-installable internal scanner This is a
10-channel scanner card, which includes eight channels
of 2-pole relay switching and two channels of 2-pole
solid-state switching. All channels can be configured
for 4-pole operation.
1-1
General Information
1.3W arranty information
Warranty information is located on the inside front cover of
this instruction manual. Should your Model 2002 require
warranty service, contact the Keithley representative or
authorized repair facility in your area for further information. When returning the instrument for repair, be sure to fill
out and include the service form at the back of this manual to
provide the repair facility with the necessary information.
1.4Manual addenda
Any improvements or changes concerning the instrument or
manual will be explained in an addendum included with the
manual. Be sure to note these changes and incorporate them
into the manual.
1.5Safety symbols and terms
The following symbols and terms may be found on an instrument or used in this manual.
!
The symbol on an instrument indicates that the user
should refer to the operating instructions located in the manual.
The symbol on an instrument shows that high voltage
may be present on the terminal(s). Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading used in this manual explains dangers that might result in personal injury or death. Always
read the associated information very carefully before performing the indicated procedure.
The CAUTION heading used in this manual explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
1.6Specifications
Full Model 2002 specifications are included in Appendix A.
1.7Inspection
The Model 2002 was carefully inspected, both electrically
and mechanically before shipment. After unpacking all items
from the shipping carton, check for any obvious signs of
physical damage that may have occurred during transit.
(Note: There may be a protective film over the display lens,
which can be removed.) Report any damage to the shipping
agent immediately. Save the original packing carton for possible future reshipment. The following items are included
with every Model 2002 order:
• Model 2002 Multimeter (with MEM1 or MEM2 memory option, if ordered) with line cord.
• Model 8605 High Performance Modular Test Leads.
• Model 2002 User’s Manual and Model 2002 Calibration Manual.
• Accessories as ordered.
• Full calibration data (conforming to MIL-STD
45662A).
If an additional manual is required, order the appropriate
manual package:
• Keithley part number 2002-900-00 for the User’ s Manual.
• Keithley part number 2002-902-00 for the Repair Manual.
• Keithley part number 2002-903-00 for the Getting
Started Manual.
• Keithley part number 2002-905-00 for the Calibration
Manual.
The manual packages include a manual and any pertinent
addenda.
1.8Options and accessories
The following options and accessories are available from
Keithley for use with the Model 2002.
Model 1050 Padded Carrying Case: A carrying case for a
Model 2002 or a Model 7001. Includes handles and shoulder
strap.
Models 2002/MEM1 and 2002/MEM2: These optional con-
figurations of the Model 2002 extend its storage capacity.
The MEM1 option has 32K-bytes for non-volatile storage of
five setups, and 7000 readings in compact format or 1400
readings in full format. The MEM2 option has 128K-bytes
for non-volatile storage of ten setups, and 30000 compact
readings or 6000 full readings.
Model 2001-SCAN: This is a 10-channel scanner card that
installs within the Model 2002. It has eight channels of 2pole relay switching and two channels of 2-pole solid-state
switching. All channels can be configured for 4-pole operation. Included are two pairs of leads for connection to Model
2002 rear panel inputs (Keithley part number CA-109).
1-2
M
General Information
Model 2001-TCSCAN: This is a thermocouple/general pur-
pose scanner card that installs in the option slot of the Model
2002. The card has nine analog input channels that can be
used for high-accuracy , high-speed scanning. A built-in temperature reference allows multi-channel, cold-junction compensated temperature measurements using thermocouples.
Model 4288-1 Single Fixed Rack Mount Kit: Mounts a sin-
gle Model 2002 in a standard 19-inch rack.
Model 4288-2 Side-by-side Rack Mount Kit: Mounts two
instruments (Models 182, 428, 486, 487, 2001, 2002, 7001)
side-by-side in a standard 19-inch rack.
Model 4288-3 Side-by-side Rack Mount Kit: Mounts a
Model 2002 and a Model 199 side-by-side in a standard 19inch rack.
Model 4288-4 Side-by-side Rack Mount Kit: Mounts a
Model 2002 and a 5
220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.) sideby-side in a standard 19-inch rack.
Models 7007-1 and 7007-2 Shielded IEEE-488 Cables:
Connect the Model 2002 to the IEEE-488 bus using shielded
cables and connectors to reduce electromagnetic interference
(EMI). The Model 7007-1 is one meter long; the Model
7007-2 is two meters long.
¼
-inch instrument (Models 195A, 196,
RMS, 42.4V peak; the test probes are rated at 1000V.) These
components are designed to be used with high performance
test leads terminated with banana plugs, such as the Model
8605 High Performance Modular Test Leads.
The following test leads and probes are rated at 30V RMS,
42.4V peak:
Models 5805 and 5805-12 Kelvin Probes: Consists of two
spring-loaded Kelvin test probes with banana plug termination. Designed to be used with instruments that measure 4terminal resistance. The Model 5805 is 0.9m long; the Model
5805-12 is 3.6m long.
Model 5806 Kelvin Clip Lead Set: Includes two Kelvin clip
test leads (0.9m) with banana plug termination. Designed for
instruments that measure 4-terminal resistance. A set of eight
replacement rubber bands for the Model 5806 is available as
Keithley P/N GA-22.
Model 8604 SMD Probe Set: Consists of two test leads (3ft),
each terminated with a surface mount device “grabber” clip
on one end and a banana plug with retractable sheath on the
other end.
Model 8610 Low Thermal Shorting Plug: Consists of four
banana plugs mounted to a 1-inch square circuit board, interconnected to provide a short circuit among all plugs.
Models 8501-1 and 8501-2 Trigger Link Cables: Connect
the Model 2002 to other instruments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model
8501-1 is one meter long; the Model 8501-2 is two meters
long.
Model 8502 Trigger Link Adapter: Allows you to connect
the Trigger Link of the Model 2002 to instruments that use
the standard BNC (In/Out) external triggering technique.
Model 8530 IEEE-488 to Centronics Printer Adapter
Cable: Translates the IEEE-488 connector pinout and signal
level to a Centronics termination. This permits a standard
Centronics parallel printer to be connected to a Model 2002
in TALK-ONLY mode.
Model 8605 High Performance Modular T est Leads: Con-
sists of two high voltage (1000V) test probes and leads. The
test leads are terminated with a banana plug with retractable
sheath on each end. (Each Model 2002 is shipped with one
set of these test leads.)
Model 8606 High Performance Probe Tip Kit: Consists of
two spade lugs, two alligator clips, and two spring hook test
probes. (The spade lugs and alligator clips are rated at 30V
Model 8611 Low Thermal Patch Leads: Consists of two
test leads (3ft), each with a banana plug with a retractable
sheath at each end. These leads minimize the thermallyinduced offsets that can be created by test leads.
Model 8612 Low Thermal Spade Leads: Consists of two
test leads (3ft), each terminated with a spade lug on one end
and a banana plug with a retractable sheath on the other end.
These leads minimize the thermally-induced offsets that can
be created by test leads.
Model 8680 RTD Probe Adapter: This adapts RTDs with
terminated and unterminated cables to instruments with
banana jacks for measuring 4-terminal resistance. It has a 4pin “T”-style connector and a 4-pin screw terminal block.
Model 8681 Miniature RTD Surface Probe: This is a low
cost platinum 4-wire-RTD with unterminated wires. It is
designed to measure the temperature of flat surfaces or free
space.
odel 8693 General Purpose/Immersion RTD Probe:
This probe has a platinum RTD sensor. It is designed for
immersion in liquids as well as other general purpose
applications.
1-3
General Information
Model 8695 Surface RTD Probe: This probe has a platinum
RTD sensor. It is designed to measure the temperature of flat
surfaces of solids.
Model 8696 Air/Gas RTD Probe: This probe has a platinum
RTD sensor. It has an exposed junction within a protective
shroud for measuring the temperature of air or gases.
1-4
2
Front Panel Operation
2.1Introduction
This section contains detailed reference information for front
panel operation of the Model 2002. It is organized as
follows:
2.2Power-up: Covers information on connecting the
instrument to line power, w arm-up period, default conditions, and the power-up sequence.
2.3Display: Covers display format and messages that
may appear while using the instrument.
2.4Functions: Describes the measurement functions of
the instrument (DC and AC voltage, DC and AC
current, 2-wire and 4-wire resistance, frequency, and
temperature).
2.5Range: Covers both manual and autoranging
operation.
2.6Relative: Gives details on using the relative feature
that can be used to null offsets or subtract a baseline
value from present and future readings.
2.7Triggering: Details types of trigger modes as well as
trigger sources that can be used.
2.8Buffer: Covers use of the reading buf fer including pro-
gramming buffer size and recalling data, time stamp,
and statistics information.
2.9Filter: Covers the use of the digital filter types that can
be used to reduce reading noise.
2.10 Math: Describes the calculations that can be per-
formed on readings in continuous operation.
2.11 Scanning: Discusses setting up internal and external
scans, along with ratio and delta calculations.
2.12 Menu: Covers selections controlled from the main
menu, such as saving instrument setups, GPIB (IEEE488 bus) configuration, calibration, self-tests, limits,
and the digital I/O port.
2.2Power-up
2.2.1Line power connections
Follow the procedure below to connect the Model 2002 to
line power and turn on the instrument.
1. The Model 2002 operates from a line voltage in the
range of 90-134V or 180-250V at a frequency of 50, 60,
or 400Hz. Check to see that the operating voltage in
your area is compatible.
CAUTION
Operating the instrument on an incorrect line voltage may cause damage to
the instrument, possibly voiding the
warranty.
2. Before plugging in the power cord, make sure the front
panel power switch is in the off (0) position.
3. Connect the female end of the supplied power cord to
the AC receptacle on the rear panel. Connect the other
end of the power cord to a grounded AC outlet.
2-1
Front Panel Operation
WARNING
The power cord supplied with the Model
2002 contains a separate ground wire
for use with grounded outlets. When
proper connections are made, instrument chassis is connected to power line
ground through the ground wire in the
power cord. Failure to use a grounded
outlet may result in personal injury or
death due to electric shock.
2.2.2Line fuse replacement
A rear panel fuse located below the AC receptacle protects
the power line input of the instrument. If the fuse needs to be
replaced, perform the following steps:
WARNING
Make sure the instrument is disconnected from the power line and other equipment before replacing the line fuse.
1. With the power off, place the end of a flat-blade screwdriver into the rear panel LINE FUSE holder. Push in
gently and rotate the fuse carrier one-quarter turn counterclockwise. Release pressure on the holder and its
internal spring will push the fuse carrier out of the
holder.
2. Remove the fuse and replace it with the same type
(0.5A, 250V, slow blow, 5 × 20mm). The Keithley part
number is FU-71.
CAUTION
Do not use a fuse with a higher current
rating than specified, or instrument
damage may occur. If the instrument
repeatedly blows fuses, locate and correct the cause of the trouble before
replacing the fuse. See the optional
Model 2002 Repair Manual for troubleshooting information.
2.2.3Power-up sequence
On power-up, the Model 2002 performs self-tests on its
EPROM and RAM, and checksum tests on data stored in
non-volatile memory . (See Table 2-1.) If a failure is detected,
the instrument momentarily displays an error message and
the ERR annunciator turns on. (Messages are listed in Table
2-2.)
NOTE
If a problem develops while the
instrument is under warranty, return it to
Keithley Instruments, Inc. for repair.
If the instrument passes the self-tests, the firmware revision
levels, memory option (if installed), and presently selected
IEEE-488 primary address are displayed. An e xample of this
display is shown as follows:
Model 2002
Rev . A01 A01 MEM1 IEEE Addr=16
The firmware revision levels (left to right) are for the main
microcontroller and display microcontroller. The revision
level number may be different in your particular unit. If the
MEM1 or MEM2 memory options are not present, that portion of the bottom line is left blank. The IEEE-488 address is
its default value of 16.
Next, if the unit is configured to display the calibration due
date at power-up, the unit shows the following:
Model 2002
Calibration due: mmm/dd/yy
where “mmm” is the month abbreviation, “dd” is the day,
and “yy” is the year. If no calibration date is set, the display
shows that it is due now. (See the Model 2002 Calibration
Manual to set the calibration due date and paragraph 2.12.3
of this manual to set the display option.)
After the power-up sequence, the instrument begins its normal display, for example:
3. Install the new fuse and fuse carrier into the holder by
reversing the above procedure.
2-2
000.0000 mVDC
Power-up error messages
Error messages that may be displayed during power-up are
summarized in Table 2-2. These are shown when one of the
checksum tests of Table 2-1 fails.
Table 2-1
Data checked on power-up
DataType of storageMemory option
Front Panel Operation
IEEE-488 address
Power-on default
Calibration constants
Calibration dates
Instrument setups1 in electrically-erasable PROM
Reading buffer(Volatile RAM)
Note: STD is standard memory, MEM1 is memory option #1, MEM2 is memory option #2.
Table 2-2
Power-up error messages
MessageAction
Error +515,
Calibration dates lost
Error +514,
DC calibration data lost
The cal dates are set to factory default values, but they are not stored into EEPROM.
To do this, perform a comprehensive calibration.
DC cal constants are set to factory default values, but they are not stored into
EEPROM. To do this, perform a comprehensive calibration.
4 more in non-volatile RAM
9 more in non-volatile RAM
Non-volatile RAM
STD, MEM1, MEM2
STD, MEM1, MEM2
STD, MEM1, MEM2
STD, MEM1, MEM2
STD, MEM1, MEM2
MEM1
MEM2
STD
MEM1, MEM2
Error +513,
AC calibration data lost
Error +512,
AC cal constants are set to factory default values, but they are not stored into
EEPROM. To do this, perform a comprehensive calibration.
Power-on defaults are reset to factory defaults (bench) and stored into EEPROM.
Power-on state lost
Error +511,
GPIB address is reset to factory default (16) and stored into EEPROM.
GPIB address lost
Error +510,
Reading buffer data lost
Error -314,
The reading buffer controls are reset to factory defaults, but they are not stored into
NVRAM. To do this, store readings in the buffer.
Instrument setup is reset to bench defaults and stored into EEPROM.
Save/recall memory lost
Notes:
1. Any of these error conditions may occur the first time a unit is turned on or after replacing the firmware.
2. Error +510, Reading buffer data lost, applies to units with optional memory.
3. Error +512, Power-on state lost, may occur the first time the unit is powered-up after replacing the MEM2 memory option with MEM1.
2-3
Front Panel Operation
2.2.4High energy circuit safety precautions
To optimize safety when measuring voltage in high energy
distribution circuits, read and use the directions in the following warning.
WARNING
Dangerous arcs of an explosive nature
in a high energy circuit can cause severe
personal injury, or death. If the multimeter is connected to a high energy circuit when set to a current range, low
resistance range, or any other low
impedance range, the circuit is virtually
shorted. Dangerous arcing can result
even when the multimeter is set to a
voltage range if the minimum voltage
spacing is reduced.
When making measurements in high energy circuits, use test
leads that meet the following requirements:
• Test leads should be fully insulated.
• Only use test leads that can be connected to the circuit
(e.g., alligator clips, spade lugs, etc.) for hands-off measurements.
• Do not use test leads that decrease voltage spacing. This
diminishes arc protection and creates a hazardous
condition.
Use the following sequence when testing power circuits:
1. De-energize the circuit using the regular installed
connect-disconnect device, such as a circuit breaker,
main switch, etc.
2. Attach the test leads to the circuit under test. Use appropriate safety rated test leads for this application.
3. Set the multimeter to the proper function and range.
4. Energize the circuit using the installed connectdisconnect device and make measurements without
disconnecting the multimeter.
5. De-energize the circuit using the installed connectdisconnect device.
6. Disconnect the test leads from the circuit under test.
2.2.5Power-on default conditions
Power-on default conditions are those conditions the instrument assumes when it is first turned on. You can change these
power-on default conditions (e xcept the primary address) by
using the save setup feature that is av ailable with the MENU
key, as described in paragraph 2.12.1. Depending on the
installed memory option, either one, five, or ten user-defined
setups can be stored, any one of which could be selected as
the power-on default.
Table 2-42 in paragraph 2.12.1 lists the default conditions
that are set at the factory to optimize bench and GPIB (IEEE-
488) operation.
2.2.6Warm-up period
The Model 2002 can be used within one minute after it is
turned on. However, the instrument should be turned on and
allowed to warm up for at least four hours before use to
achieve rated accuracy.
2.2.7IEEE-488 primary address
The IEEE-488 primary address of the instrument must be the
same as the primary address you specify in the controller’s
programming language. The default primary address of the
instrument is 16, but you can set the address to any value
from 0 to 30 by using the MENU key. Refer to paragraph
2.12.2 for step-by-step instructions on setting the primary
address. Section 4 contains details on using the Model 2002
over the IEEE-488 bus.
2.3Display
The display of the Model 2002 is primarily used to display
readings along with the units and type of measurement.
When not displaying readings, it is used for informational
messages, such as menu headings and selections. At the top
of the display are annunciators to indicate various states of
operation. The Getting Started manual covers the display
basics.
2-4
WARNING
The maximum common-mode voltage
(voltage between INPUT LO and chassis ground) is 500V peak. Exceeding this
value may cause a breakdown in insulation, creating a shock hazard.
2.3.1Multiple displays
Each measurement function has its own set of “multiple displays” shown on the bottom line of the front panel display.
The PREVious and NEXT DISPLAY keys scroll through the
selections for the present function.
Front Panel Operation
Some of the multiple displays are for multiple functions,
where different functions are measured sequentially from the
same set of test leads. The readings are shown simultaneously, such as:
• T op line sho ws a DC v oltage measurement; bottom line
shows positive and negative peak spike measurements.
• Top line shows an AC RMS voltage reading; bottom
line shows an AC frequency measurement and a crest
factor calculation.
Also, the multiple displays can sho w a reading in a dif ferent
form, or give additional information about the reading, for
example:
• Top line shows a reading; bottom line shows a zerocentered bar graph with adjustable limits.
Table 2-3
Multiple displays by function
FunctionNext displayParagraph
All
Bar graph
Zero-centered bar graph
Maximum and minimum values
Relative and actual values
Calculated and actual values (see Note 1)
Limits bar graph (see Note 1)
Adjacent channel readings (see Note 2)
DC voltage
DC volts, AC ripple voltage and frequency
Positive peak spikes and highest value
Negative peak spikes and lowest value
Positive and negative peak spikes
AC voltage
AC RMS voltage, frequency, and crest factor
AC RMS, average, and peak voltages
DC current
AC current
(none specific to function)
AC RMS (or average) current and frequency
AC RMS and average current
2-wire resistance
Source current
Voltage drop across DUT
4-wire resistance
Source current
Voltage drop across DUT
Lead resistance
Frequency
Period calculation
Trigger level
Temperature
Celsius, Fahrenheit, and Kelvin units
RTD resistance (or thermocouple voltage)
Reference junction (thermocouples only)
Data storage buffer
Maximum and minimum values
Average and standard deviation
Notes:
1. Multiple displays for calculated values and limits bar graph are not available for the frequenc y function.
2. The multiple display for adjacent channel readings is not available for the DC and AC current functions.
• Top line shows a frequency measurement; bottom line
shows the adjustable trigger level.
To scroll through the multiple displays available for each
measurement function, repeatedly press and release the
NEXT DISPLAY key. The same action with the PREVious
DISPLAY key does a reverse scroll through the displays. To
return to the default reading display, just press and hold
either key.
Multiple displays that are specific to a particular function or
operation are discussed later in this section, such as the peak
spikes displays in DC voltage, and the calculations display in
math. (See T able 2-3 for paragraph references.) Displays that
are common to most of the measurement functions are discussed here.
2.3.1
2.3.1
2.3.1
2.6
2.10
2.12.5
2.11
2.4.1
2.4.1
2.4.2
2.4.2
2.4.3
2.4.3
2.4.4
2.4.5
2.8
2-5
Front Panel Operation
Bar graph
The “normal” bar graph, with a zero at the left end, is a
graphical representation of a reading as a portion of a range.
(See Figure 2-1.) The vertical lines displayed along the bar
designate 0%, 25%, 50%, 75%, and 100% of full scale. Each
full segment of the bar represents approximately 4% of the
range limit.
-11.96859 VDC
0-20V
Full
Range
25% of
full range
50% of
full range
75% of full range
Figure 2-1
Bar graph (zero-at-left) multiple display
For measurement functions with a range (voltage, current,
and resistance), the right endpoint of the bar graph is plus full
scale of the present range for positive readings, and minus
full scale for negative readings. When the 100% line changes
to an arrow, the reading exceeds the present range.
Note that the normal bar graph is not available when the
ACV units are dB or dBm.
For functions without a range (frequency and temperature),
the right endpoint of the bar graph is user-programmable by
pressing either CONFIG-NEXT DISPLAY or CONFIGPREV DISPLAY. Note that these configuration menus are
context-sensitive. If the unit is in any function except frequency or temperature, CONFIG-NEXT DISPLAY results
in the zero bar graph configuration display.
Perform the following to view or change the range of the bar
graph:
1. From the frequency or temperature function, press the
CONFIG key and then the NEXT or PREV DISPLAY
key. The following menu is displayed:
BARGRAPH TYPE
ZERO-AT-LEFT ZERO-CENTERED
2. Use the cursor keys ( and ) to place the cursor on
ZERO-AT -LEFT and press ENTER. You will access one
of the following menus:
For frequency:
FREQ BARGRAPH RANGE
2Hz 20Hz 200Hz 2kHz 20kHz
200kHz 2MHz 15MHz
For temperature:
BARGRAPH:0 to 0040°C
3. Change the frequency range by highlighting one of the
selections and pressing ENTER. For the temperature
range, use the cursor keys and the RANGE ▲ and ▼
keys to enter a numeric value (0 - 9999°C). Press ENTER when done.
Zero-centered bar graph
The zero-centered bar graph is a graphical representation of
a reading with plus and minus limits. (See Figure 2-2.) The
limits are expressed in a user-selectable percentage of range
for voltage, current, and resistance, and a user-selectable value for frequency and temperature.
-05.95770 VDC
-50%+50%
-50% of range
-25% of range
0%
+50% of range
25% of range
Figure 2-2
Zero-centered bar graph multiple display
The vertical lines displayed along the bar designate the plus
and minus limits, zero, and halfway to either limit. There are
ten full segments between zero and each end, so each full
segment represents 10% of the limit. When a line at the limit
changes to an arrow, the reading exceeds the programmed
range.
The plus and minus percentage of range that is programmed
(0.01 - 99.99%) applies to all voltage, current, and resistance
functions. Because of rounding, values greater than 99.5%
are shown as 100% and, likewise, values greater than 1%
(such as 1.67%) are shown rounded to the nearest integer
percent.
Note that the zero-centered bar graph is not available when
the ACV units are dB or dBm.
2-6
Figure 2-3
Maximum and minimum multiple display
Minimum
value
-15.82867 VDC
Max = -05.74602Min = -15.82867
Maximum
value
Front Panel Operation
Perform the following to view or change the plus and minus
percentage of range:
1. From a voltage, current, or resistance function, press
CONFIG and then NEXT or PREV DISPLAY. The following is displayed:
ZERO-BARGRAPH±50.00%
2. Change the percentage by using the cursor keys and the
RANGE ▲ and ▼ keys to enter a numeric value (0.01 -
99.99%). Press ENTER when done. The same percentage of range is used for voltage, current, and resistance
measurements.
Perform the following to view or change the plus and minus
value limit:
1. From the frequency or temperature function, press
CONFIG and then NEXT or PREV DISPLAY. The following menu is displayed:
BARGRAPH TYPE
ZERO-AT-LEFT ZERO-CENTERED
2. Use the cursor keys ( and ) to place the cursor on
ZERO-CENTERED and press ENTER. You will access
one of the following menus:
For frequency:
FREQ ZEROBARGRAPH(±)
1Hz 10Hz 100Hz 1kHz 10kHz
100kHz 1MHz 10MHz 15MHz
• Pressing the present function key.
• Leaving the display by changing function or entering a
menu.
The resolution, units, and prefix on the bottom line are the
same as shown for top line reading. If necessary, the bottom
line values automatically change to scientific notation,
rounded to 4.5 digits.
2.3.2Information messages
Press the INFO key to view context-sensitive information
from most of the displays. An arrow ( or ) on the bottom line indicates that there is more information. Use the
cursor keys ( and ) to view the complete line. To exit
an INFO display, just press INFO, ENTER, EXIT, or a function key.
2.3.3Status and error messages
For temperature:
ZERO-BARGRAPH±0002°C
3. Change the frequency limits by highlighting one of the
selections and pressing ENTER. For the temperature,
use the cursor keys and the RANGE ▲ and ▼ keys to
enter a numeric value (0 - 9999°C). Press ENTER when
done.
Maximum and minimum
The maximum and minimum multiple display shows the
maximum and minimum readings since the display was entered. (See Figure 2-3.) The maximum and minimum values
are reset by the following:
During Model 2002 operation and programming, you will
encounter a number of front panel messages. Typical
messages are either of status or error variety, as listed in
Table 2-4.
The most recent status or error messages can be momentarily
displayed. Just enter a configuration menu or the main menu,
and press the PREV key . (The display is blank if no message
is queued.)
“Questionable T emperature”
“Questionable Calibration”
“Excessive temp drift during cal”
“Low level calibration data lost”
“Preamp calibration data lost”
“Install option id lost”
“Calibration dates lost”
“DC calibration data lost”
“AC calibration data lost”
“Power-on state lost”
“GPIB address lost”
“Reading buffer data lost”
“Parameter Error”
“Settings conflict”
“Parameter data out of range”
“Too much data”
“Illegal parameter value”
“Data corrupt or stale”
“Hardware missing”
“Expression Error”
“Cannot create program”
“Illegal program name”
“Save/recall memory lost”
“Configuration memory lost”
“Self T est failed”
“Queue overflow”
“Query interrupted”
“Query unterminated”
“Query deadlocked”
“Query unterminated after indefi-
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
nite response”
SE = Status event
EE = Error event
2-8
▼
Front Panel Operation
2.3.4Navigating menus
There are basically two types of menu structures; the Main
Menu and the Configure menus. The Main Menu accesses
items for which there are no dedicated keys, and Configure
menus are used to configure measurement functions and other instrument operations.
Use the following rules to navigate through the menu structure:
1. The top level of the Main Menu is accessed by pressing
the MENU key. A Configuration menu is accessed by
pressing CONFIG and then the desired function (DCV,
ACV, etc.) or operation (TRIG, STORE, etc.).
2. A menu item is selected by placing the cursor on it and
pressing ENTER. Cursor position is denoted by the
blinking menu item or parameter. The cursor keys (
and ) control cursor position.
3. A displayed arrow ( and ) on the bottom line indicates that there are one or more additional items (messages) to select from. Use the appropriate cursor key to
display them.
4. A numeric parameter is keyed in by placing the cursor
on the digit to be changed and using the RANGE ▲ or
key to increment or decrement the digit.
5. A change is only executed when ENTER is pressed.
Entering an invalid parameter generates an error and the
entry is ignored.
6. The EXIT key is used to back out of the menu structure.
Any change that is not entered is cancelled when EXIT
is pressed. The EXIT key has additional actions and are
summarized in Table 2-5.
“remembers” its own unique setup configuration. Setup configuration parameters that are remembered include:
• Range
• Relative
• Integration time
• Filtering
• Display resolution
• Multiple displays
To access the configuration menus for the measurement
functions, press the CONFIG key and then a function key
(DCV, ACV, DCI, ACI, Ω 2, Ω 4, FREQ, TEMP). Rules for
navigating the menu structures are covered in paragraph
2.3.4.
2.4.1DC and AC voltage
The Model 2002 can make DCV measurements from 1nV to
1100V and ACV measurements from 10nV to 775V. The
basic measurement procedures for DCV and ACV are contained in the Getting Started manual.
Voltage configuration
The following information explains the various configuration options for DC and AC voltage measurements. The configuration menus are summarized in Tables 2-6 and 2-7. The
menus are accessed by pressing CONFIG and then DCV or
ACV. Paragraph 2.3.5 summarizes the rules for navigating
through menus.
2.4Functions
To minimize the need to reprogram setup conditions every
time you switch from one function to another, each function
Table 2-5
EXIT key actions
ConditionEXIT key action
T emporary message displayed
(e.g., TRIGGERS HAL TED)
INFO message displayed
Reading display hold
Scanning
Data storage
Cancels display of temporary message.
Cancels INFO message, returns to menu or normal reading display.
Cancels reading display hold, resumes normal reading display.
Disables scanning. Also stops data storage if enabled.
Stops data storage. Temporary message STORAGE INTERRUPTED is displayed.
Note that a function does not have to be selected in order to
be configured. When the function is selected, it will assume
the programmed status.
2-9
Front Panel Operation
Table 2-6
CONFIGURE DCV menu structure
Menu itemDescription
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
FILTER
AUTO
AVERAGING
ADVANCED
AVERAGING-MODE
RESOLUTION
AUTO
3.5d, 4.5d, 5.5d, 6.5d,
7.5d, 8.5d
Measurement speed (integration time) menu:
Select 1 PLC (power line cycle, 16.67msec for 60Hz, 20msec for 50Hz and 400Hz).
Select 0.01 PLC.
Select 0.1 PLC.
Select 10 PLC.
Set integration time in PLC (0.01-50).
Default to setting appropriate for resolution.
Digital filter menu:
Default to filter appropriate for integration time.
Program a simple average filter (1-100 readings).
Program a simple average filter (1-100 readings) with a noise tolerance window (0-
100% of range).
Select moving average or repeating average mode.
Display resolution menu:
Default to resolution appropriate for integration time.
Select a specific resolution.
2-10
Table 2-7
CONFIGURE ACV menu structure
Menu itemDescription
Front Panel Operation
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
FILTER
AUTO
AVERAGING
ADVANCED
AVERAGING-MODE
RESOLUTION
AUTO
3.5d, 4.5d, 5.5d, 6.5d,
7.5d, 8.5d
UNITS
VOLTS
dB
dBm
COUPLING
AC
AC+DC
Measurement speed (integration time) menu:
Select 1 PLC (power line cycle, 16.67msec for 60Hz, 20msec for 50Hz and 400Hz).
Select 0.01 PLC.
Select 0.1 PLC.
Select 10 PLC.
Set integration time in PLC (0.01-50).
Default to setting appropriate for resolution.
Digital filter menu:
Default to filter appropriate for integration time.
Program a simple average filter (1-100 readings).
Program a simple average filter (1-100 readings) with a noise tolerance
window (0-100% of range).
Select moving average or repeating average mode.
Display resolution menu:
Default to resolution appropriate for integration time.
Select a specific resolution.
Display units menu:
Select volts.
Select dB and set voltage reference level.
Select dBm and set reference impedance.
Coupling menu:
Select AC coupled measurements.
Select DC coupled measurements.
AC-TYPE
RMS
AVERAGE
PEAK
LOW-FREQ-RMS
POSITIVE-PEAK
NEGATIVE-PEAK