The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhaz ardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required
to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the
product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operator s are adequately trained.
Operators use the product for its i ntended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact wit h hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures ar e described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform ins tallation and service procedures.
Keithley products are designed f or use with electrical signals that are measurement, control, and data I/O connections, with low
transient overvoltages, and mus t not be directly connected to mains voltage or to voltage sources with high tr ans ient
overvoltages. Measurement Cat egory II (as referenced in IEC 60664) connections require protection for high transient
overvoltages often associated with local AC mains connections. Certai n K ei thley measuring instruments may be connected to
mains. These instruments will be mar k ed as category II or higher.
Unless explicitly allowed in the spec i fications, operating manual, and instrument labels, do not connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National S tandards Institute (ANSI) states that a shock hazard exists when voltage level s greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protec ted from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some c ases, connections must be exposed to potential
human contact. Product operators i n these circumstances must be trained to pr otect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000 V, no conductive part o f the circuit may be exposed.
Do not connect switching cards direc tly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching ca rds, install protective
devices to limit fault current and v ol tage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and j umpers for possible wear, cracks, or breaks before each use.
When installing equipment where ac cess to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be prov i ded in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or rem oving switching cards, or making int ernal changes, such as installing or r emoving jumpers.
Do not touch any object that could pr ovide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstandin g the
voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or
accessories are used in a manner not s pecified in the operating instructions, the protection provided by the equipment may be
impaired.
Do not exceed the maximum signal levels of the instruments and accessori es. Maximum signal levels are defined in the
specifications and operating inf or mation and shown on the instrument panels, test fixture panels, and switching cards.
When fuses are used in a product, replac e with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring c i rcuits, NOT as protective earth (s afety ground)
connections.
If you are using a test fixture, keep the lid closed while power is applied to t he dev i ce under test. Safe operation requires the use
of a lid interlock.
screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
If a
The
symbol on an instrument means caution, risk of hazard. The user must refer to the operating instructions located in the
user documentation in all cases where the symbol is marked on the instrument .
The
symbol on an instrument means warning, risk of electric shock. Use standard safety precautions to avoid pers onal
contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The
If this
symbol indicates a connection terminal to the equipment frame.
symbol is on a product, it indicates that mercury is present in the display l amp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains hazards that might result in personal injury or death. Always read
the associated information very car efully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the inst rument. Such damage may
invalidate the warranty.
The CAUTION heading with the
symbol in the user documentation explains hazards that could result in moderate or minor
injury or damage the instrument. Always read the associated information very carefully before performing the indicated
procedure. Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input j ac ks — must be purchased from Keithley. S tandard fuses with applicable national safety
approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrum ent may
only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchas ed from other
suppliers as long as they are equival ent to the original component (note that selected parts should be purchased on ly through
Keithley to maintain accuracy and f unctionality of the product). If you are unsure about the applicability of a replacement
component, call a Keithley office for information.
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following
environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the
instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that
consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer ) should never
require cleaning if handled acc or ding to instructions. If the board becomes contaminated and operation is aff ec ted, the board
should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of J une 2017.
Factory service ......................................................................... 1-2
Features ................................................................................... 1-1
Unpacking and inspection ........................................................ 1-1
Introduction
This section contains general information about the 2000-SCAN Scanner Card. The 2000-SCAN is a
scanner card used in the DMM6500 6½ Digit Graphic al Sampling Multimeter. The 2000-SCAN
provides 10 channels of 2-pole relay switching. The channels can also be configured for 4-pole
operation.
Section 1
General Information
If you are using this scanner card with the 2700, 270 1, or 2750, please see the 2000-SCAN Scanner
Card Instruction Manual.
Features
Features of the scanner card include:
• Ten channels of 2-pole relay input.
• Configurable channels for 4-pole operation.
• Ability to multiplex one of ten 2-pole or one of five 4-pole channel s int o a DMM6500.
Unpacking and inspection
Inspecting for damage
The 2000-SCAN is packaged in a resealable, anti-st atic bag to protect from damage due to static
discharge and contamination that could degrade it s performance. Before removing the card from the
bag, observe the following handling precautions.
Section
User's Manual
1: General Information Model 2000-SCAN Scanner Card
Handling precautions
•Always grasp the card by the side edges and cover s. Do not touch the board surfaces or
components.
•When the card is not installed in a multimeter, keep the card in the anti-static bag and store it in
the original packing carton.
•After removing the card from its anti-static bag, inspect it for any obvious signs of physical
damage. Report any such damage to the shipping agent immediately.
Factory service
To return the scanner card to Keithley Instruments for repair:
•Call the Repair Department at 1-800-833-9200 or send an email to RMAREQUEST@tektronix.com for a Return Material Authorization (RMA) n um ber.
• Carefully pack the instrument in the original packi ng carton.
• Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label.
1-2 2000-SCAN-900-01 Rev. A / April 2018
Card installation and removal ................................................. 2-10
This section includes information on making conne ct i ons t o the 2000-SCAN Scanner Card and
installing the scanner card in a DMM6500.
Section 2
Card connections and installation
The information in this topic is intended only for qualified service personnel. Some of the
procedures may expose you to hazardous voltages that could result in personal injury or
death. Do not attempt to perform these procedures unless you are qualified to do so.
Installation handling precautions
To maintain high-impedance isolation between cha nnels, avoid contamination from foreign materials
as body oils when handling the scanner card. Contamination can substantially lower leakage
resistances, degrading card performance. To avoid possible contamination, always hold the scanner
card by the side edges or covers. Do not touch board surfaces, components, or areas adjacent to
electrical contacts.
Dirt build-up over a period of time is another possible source of contamination. To avoid this problem,
operate the multimeter and scanner card in a clean env i ronment.
Board cleaning
The circuit board should be thoroughly cleaned with pure water after soldering to remove all
contaminants. After cleaning with water, swab t he area with pure methanol and allow the board to dry
for several hours in a 50 °C environment before use.
Section
User's Manual
2: Card connections and installation Model 2000-SCAN Scanner Card
Connections
This section provides information on connecti ng your external test circuitry to the scanner card.
Failure to observe the following precautions may result in degraded card performance.
Connection precautions
Connection and wiring procedures in this document are intended for use by qualified
personnel only. Do not perform these procedures unless qualified to do so. Failure to
recognize and observe normal safety precautions could result in personal injury or death.
Do not attempt to perform this procedure unless qualified to do so. Failure to recognize and
observe normal safety precautions could result in personal injury or death.
Do not exceed the maximum specifications for the 2000-SCAN Scanner Card. Refer to the
specifications provided in the data sheet. Failure to recognize and observe normal safety
precautions could result in personal injury or death.
As described in the International Electrotechnical Commission (IEC) Standard IEC 664 ,
scanner cards are Installation Category I and must not be connected to mains.
To prevent electric shock that could result in serious injury or death, make sure the DMM
power is turned off and power is removed from all external circuitry before making or
breaking connections to the scanner card.
Card configurations
The following figure shows a simplified diagram of t he 2000-SCAN. The scanner card has 10 input
channels and two outputs. Channels 1 to 10 are all switched by the relays.
4-pole paired channels are as follows:
• Channels 1 and 6
• Channels 2 and 7
• Channels 3 and 8
• Channels 4 and 9
2-2 2000-SCAN-900-01 Rev. A / April 2018
•Channels 5 and 10
Model 2000
Card connections and installation
-SCAN Scanner Card User's Manual Section 2:
Figure 1: Simplified schematic
2000-SCAN-900-01 Rev. A / April 2018 2-3
Section
User's Manual
2: Card connections and installation Model 2000-SCAN Scanner Card
Card connectors
The following figure shows the input and output connectors for the card. Card connectors include:
•CH 1 to 10 (channels 1 to 10): HI and LO input terminals prov ided for each of the 10 channels on
the card.
•OUT A: HI and LO output connections for all 10 channels in 2-pole mode or channels 1 to 5 in
4-pole mode.
•OUT B: HI and LO output connections for channels 6 t o 10 i n 4-pole mode.
To access the connectors, open the plastic shield by pressing in the locking tab, and swing the shield
away from the circuit board.
Figure 2: Card connectors
2-4 2000-SCAN-900-01 Rev. A / April 2018
Model 2000
Card connections and installation
-SCAN Scanner Card User's Manual Section 2:
Output connections
When making output connections from your 2000 -SCAN to a DMM6500, use the supplied test leads
for scanner output connections.
To make output connections to your DMM6500:
1. Connect the red leads to the output (OUT A and OUT B) HI terminals.
2. Connect the black leads to the output LO terminals. See the following figure for details.
3. Dress the output test leads through the cable clam p.
4. After all wires are connected and secure, close t he pl astic shield, and secure it with the locking
tab.
Figure 3: Output connections
Wiring procedures
Before connecting or disconnecting wires, all power must be off and stored energy in
external circuitry discharged.
2000-SCAN-900-01 Rev. A / April 2018 2-5
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.