Keithley 2000calib schematic

Model 2000Multimeter
Calibration Manual
A GREATER MEASURE OF CONFIDENCE
W ARRANTY
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defec­tive.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-recharge­able batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instruc­tions.
THIS WARRANTY IS IN LIEU OF ALL O THER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE B UYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTW ARE EVEN IF KEITHLEY INSTR UMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAM­AGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY .
Keithley Instruments, Inc. • 28775 Aurora Road • Cleveland, OH 44139 • 440-248-0400 • Fax: 440-248-6168 • http://www.k eithley.com
BELGIUM: Keithley Instruments B.V. Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02/363 00 40 • Fax: 02/363 00 64 CHINA: Keithley Instruments China Y uan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-62022886 • Fax: 8610-62022892 FRANCE: Keithley Instruments Sarl 3, allée des Garays • 91127 Palaiseau Cedex • 01-64 53 20 20 • Fax: 01-60 11 77 26 GERMANY: Keithley Instruments GmbH Landsberger Strasse 65 • 82110 Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 GREAT BRITAIN: Keithley Instruments Ltd The Minster • 58 Portman Road • Reading, Berkshire RG30 1EA • 0118-9 57 56 66 • Fax: 0118-9 59 64 69 INDIA: Keithley Instruments GmbH Flat 2B, WILOCRISSA • 14, Rest House Crescent • Bangalore 560 001 • 91-80-509-1320/21 • Fax: 91-80-509-1322 ITALY: Keithley Instruments s.r.l. Viale S. Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74 NETHERLANDS: Keithley Instruments B.V. Postbus 559 • 4200 AN Gorinchem • 0183-635333 • Fax: 0183-630821 SWITZERLAND: Keithley Instruments SA Kriesbachstrasse 4 • 8600 Dübendorf • 01-821 94 44 • Fax: 01-820 30 81 TAIWAN: Keithley Instruments Taiwan 1 Fl. 85 Po Ai Street • Hsinchu, Taiwan, R.O.C. • 886-3572-9077• Fax: 886-3572-903
9/00
Model 2000 Multimeter
Calibration Manual
©1994, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Third Printing January 2001
Document Number: 2000-905-01 Rev. C
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Le vel letter increases alphabetically as the manual under goes sub­sequent updates. Addenda, which are released between Revisions, contain important change in­formation that the user should incorporate immediately into the manual. Addenda are numbered sequentially . When a new Re vision is created, all Addenda associated with the previous Re vision of the manual are incorporated into the new Revision of the manual. Each ne w Revision includes a revised copy of this print history page.
Revision A (Document Number 2000-905-01)........................................................November 1994
Addendum A (Document Number 2000-905-02)..............................................................June 1995
Revision B (Document Number 2000-905-01)................................................................. June 2000
Revision C (Document Number 2000-905-01)............................................................ January 2001
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated in­strumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read the operating information carefully before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for
ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures
and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of prod-
ucts. Only properly trained service personnel may perform installation and service procedures. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable con-
nector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock haz­ard exists when voltage lev els greater than 30V RMS, 42.4V peak, or 60VDC are present.
practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Users of this product must be protected from electric shock at all times. The responsible body must en­sure that users are prevented access and/or insulated from every connection point. In some cases, con­nections must be exposed to potential human contact. Product users in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts,
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multi­meter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation Category II. All other instruments’ signal terminals are Installation Category I and must not be connect­ed to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power re­ceptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks be­fore each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is ap­plied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
no conductive part of the circuit may be exposed.
A good safety
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Alw ays make measurements with dry hands while standing on a dry, insulated surface ca­pable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth
ground connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documen­tation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the com­bined effect of normal and common mode voltages. Use standard safety precautions to av oid personal contact with these voltages.
The
WARNING heading in a manual explains dangers that might result in personal injury or death. Always
read the associated information very carefully before performing the indicated procedure. The
CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may
invalidate the warranty. Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, replacement components in mains circuits, including the
power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments to maintain ac­curacy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Prod­ucts that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contami­nated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Rev. 10/99
T able of Contents
1
Performance Verification
Introduction..........................................................................................1-2
Verification test requirements ..............................................................1-3
Performing the verification test procedures.........................................1-6
Verifying DC voltage...........................................................................1-7
V erifying A C voltage ...........................................................................1-8
Verifying DC current .........................................................................1-10
V erifying A C current..........................................................................1-11
Verifying resistance ...........................................................................1-12
Calibration
2
Introduction..........................................................................................2-2
Environmental conditions....................................................................2-3
Calibration considerations...................................................................2-4
Calibration code...................................................................................2-5
Comprehensive calibration..................................................................2-6
Front panel calibration.........................................................................2-7
SCPI command calibration................................................................2-14
3
Calibration Command Reference
Introduction..........................................................................................3-2
Command summary.............................................................................3-4
Miscellaneous calibration commands..................................................3-6
DC calibration commands .................................................................3-10
AC calibration commands .................................................................3-16
Manufacturing calibration .................................................................3-18
Bus error reporting.............................................................................3-19
Detecting calibration step completion...............................................3-20
A
Specifications
Accuracy calculations.........................................................................A-7
Optimizing measurement accuracy...................................................A-10
Optimizing measurement speed........................................................A-11
B
Error Messages Calibration Program
C
List of Illustrations
1
Performance Verification
Connections for DC volts verification.................................................1-7
Connections for AC volts verification.................................................1-8
Connections for DC current verification...........................................1-10
Connections for AC current verification............................................1-11
Connections for resistance verification (100 Ω -10M Ω ranges).........1-12
Connections for resistance verification (100M Ω range)...................1-13
Calibration
2
Low-thermal short connections...........................................................2-8
Calibrator connections for DC volts and ohms portion of
comprehensive calibration ..............................................................2-9
Connections for DC and AC amps comprehensive calibration.........2-11
Connections for AC volts calibration................................................2-12
Synthesizer connections for manufacturing calibration....................2-21
List of T ables
1
Performance Verification
Recommended verification equipment................................................1-4
DCV reading limits..............................................................................1-7
ACV reading limits..............................................................................1-9
DCI limits ..........................................................................................1-10
ACI limits ..........................................................................................1-11
Limits for resistance verification.......................................................1-13
Calibration
2
Recommended equipment for comprehensive, DC only, or
Comprehensive calibration procedures................................................2-7
DC volts calibration summary...........................................................2-10
Ohms calibration summary................................................................2-10
DC current calibration summary .......................................................2-11
Connections for AC volts calibration.................................................2-12
AC current calibration summary........................................................2-13
DC voltage calibration programming steps.......................................2-16
Resistance calibration programming steps ........................................2-16
DC current calibration programming steps........................................2-17
AC voltage calibration programming steps.......................................2-18
AC current calibration programming steps........................................2-19
Recommended equipment for manufacturing calibration .................2-20
AC only calibration.........................................................................2-6
3
Calibration Command Reference
Calibration command summary...........................................................3-4
DC calibration commands .................................................................3-10
AC calibration commands .................................................................3-16
B
Error Messages
Error summary....................................................................................B-2
1
Performance
V erification
1-2 Performance Verification
Introduction
Use the procedures in this section to verify that the Model 2000 Multimeter accuracy is within the limits stated in the instrument’s one-year accuracy specifications. You can perform verifica­tion procedures:
W ARNING The information in this section is intended only for qualified service person-
When you first receiv e the instrument to mak e sure that it w as not damaged during ship­ment, and that the unit meets factory specifications.
When a question exists about the instrument’s accuracy.
Following calibration.
nel. Do not attempt these procedures unless you are qualified to do so.
NOTE
If the instrument is still under warranty and its performance is outside specified lim­its, contact your Keithle y repr esentative or the factory to determine the corr ect course of action.
This section includes the following information:
Verification test requirements: Explains the test requirements.
Performing the verification procedures: Provides general information about the test
procedures.
Verifying DC voltage: Provides the procedure to verify that the instrument meets its
DC voltage accuracy specifications.
V erifying AC voltage: Provides the procedure to verify AC v oltage measurement accu-
racy.
Verifying DC current: Outlines the procedure to test DC current measurement accura-
cy.
V erifying AC current: Provides the procedure to verify AC current measurement accu-
racy.
Verifying resistance: Provides the procedure to test resistance measurement accuracy.
V erification test requirements
Be sure that you perform the verification tests:
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper calibration equipment.
Using the specified reading limits.
Environmental conditions
Performance Verification 1-3
Conduct your performance verification procedures in a test environment that has:
An ambient temperature of 18-28°C (65-82°F).
A relative humidity of less than 80% unless otherwise noted.
Line power
The Model 2000 Multimeter requires a line voltage of 100V/120V/220V/240V, ±10% and a
line frequency of 45Hz to 66Hz, or 360Hz to 440Hz.
W arm-up period
Allow the Model 2000 Multimeter to warm up for at least one hour before conducting the v er-
ification procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufac­turer.
1-4 Performance Verification
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. Use the Fluke Model 5700A
Calibrator (or the equivalent) to verify Model 2000 Multimeter measurement accuracy.
You can use alternate equipment as long as that equipment has specifications at least as good as those listed in T able 1-1. K eep in mind, howe ver , that the calibrator will add to the uncertainty of each measurement. T able 1-1 lists the uncertainties of the recommended Fluk e 5700A at each source value.
Table 1-1
Recommended verification equipment
Fluke 5700A Calibrator:
DC voltage
100mV: ±14ppm
1.0V: ±7ppm 10V: ±5ppm 100V: ±7ppm 1000V: ±9ppm
Fluke 5725A Amplifier:
AC Voltage, 50kHz: 700V, ±375ppm
*1kHz specifications shown. 5725A amplifier required to source 700V @ 50kHz. 90-day, 23°C ±5°C specifications shown.
NOTE: The Fluke 5725A amplifier is necessary only if you wish to verify the 750V AC range at 50kHz. Verifica­tion at 220V, 50kHz using only the 5700A calibrator is adequate for most applications.
AC voltage
(1kHz, 50kHz)*
100mV: ±200ppm
1.0V: ±82ppm 10V: ±82ppm 100V: ±90ppm 700V: ±85ppm
DC current
10mA: ±60ppm 100mA: ±70ppm 1A: ±110ppm
2.2A: ±94ppm
AC current
(1kHz)
1A: ±690ppm
2.2A: ±682ppm
Resistance
100 1k
: ±12ppm 10k 100k 1M 10M 100M
: ±17ppm
: ±11ppm
: ±13ppm
: ±18ppm
: ±37ppm
: ±120ppm
×
V erification limits
The verification limits stated in this section have been calculated using only the Model 2000 one-year accuracy specifications, and they do not include test equipment uncertainty. If a par­ticular measurement falls slightly outside the allowable range, recalculate new limits based on both Model 2000 specifications and pertinent calibration equipment specifications.
Example reading limit calculation
The following is an example of how reading limits have been calculated:
Assume you are testing the 10V DC range using a 10V input value. Using the Model 2000 one-year accuracy specification for 10V DC of ± (30ppm of reading + 5ppm of range), the cal­culated limits are:
Performance Verification 1-5
Reading limits = 10V ± [(10V
Reading limits = 10V ± (.0003 + .00005)
Reading limits = 10V ± .00035V
Reading limits = 9.99965V to 10.00035V
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory defaults as follows:
1. Press SHIFT and then SETUP. The instrument will display the following prompt:
RESTORE: FACT
NOTE Pressing either range key toggles the RESTORE selection between USER and FACT.
2. Restore the factory default conditions by pressing ENTER.
3. Factory defaults will be set as follows: Speed: medium
Filter: 10 readings
30ppm) + (10V × 5ppm)]
1-6 Performance Verification
Performing the verification test procedures
T est summary
Verification test procedures include:
DC volts
AC volts
DC current
AC current
Resistance
If the Model 2000 is not within specifications and not under warranty , see the calibration pro-
cedures in Section 2.
T est considerations
When performing the verification procedures:
Be sure to restore factory defaults as outlined above.
Make sure that the equipment is properly warmed up and connected to the front panel input jacks. Also mak e sure that the front panel input jacks are selected with the INPUTS switch.
Do not use autoranging for any v erification tests because autorange hysteresis may cause the Model 2000 to be on an incorrect range. For each test signal, you must manually set the correct range for the Model 2000 using the range keys.
Make sure the calibrator is in operate before you verify each measurement.
Always let the source signal settle before taking a reading.
Do not connect test equipment to the Model 2000 through a scanner or other switching equipment.
W ARNING The maximum common-mode voltage (voltage between INPUT LO and chas-
sis ground is 500 V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard. Some of the procedures in this section may expose you to dangerous voltages. Use standard safety precautions when such dangerous voltages are encountered to avoid personal injury caused by electric shock.
V erifying DC voltage
Check DC voltage accuracy by applying accurate DC v oltages from the calibrator to the Mod­el 2000 INPUT jacks and verifying that the display reads within specified limits. Follow these steps to verify the DC voltage:
1. Connect the Model 2000 HI and LO INPUT jacks to the DC v oltage calibrator as sho wn in Figure 1-1.
NOTE Use shielded, low-thermal connections when testing the 100mV and 1V ranges to
avoid errors caused by noise or thermal ef fects. Connect the shield to the calibrator’ s output LO terminal.
Performance Verification 1-7
Figure 1-1
Connections for DC volts verifica­tion
5700A Calibrator (Output DC Voltage)
Model 2000
2000 MULTIMETER
Input HI
Output HI
!
R
Input LO
Output LO
Note : Use shielded, low-thermal cables for 100mV and 1V ranges.
2. Select the DC volts function by pressing the DCV key, and set the Model 2000 to the 100mV range.
3. Set the calibrator output to 0.00000mV DC, and allow the reading to settle.
4. Enable the Model 2000 REL mode. Leave REL enabled for the remainder of the DC volts verification tests.
5. Source positive and negative full-scale voltages for each of the ranges listed in Table 1-
2. For each voltage setting, be sure that the reading is within stated limits.
Table 1-2
DCV reading limits
DCV
Range
Applied DC
voltage*
Reading limit
(1 year, 18°C-28°C)
99.9915 to 100.0085mV
0.999963 to 1.000037V
9.99965 to 10.00035V
99.9949 to 100.0051V
999.939 to 1000.061V
1V
100.0000mV
1.000000V
10.00000V
100.0000V
1000.000V
100mV
10V
100V
1000V
* Source positive and negative values for each range.
1-8 Performance Verification
V erifying AC voltage
Check AC voltage accuracy by applying accurate AC voltages at specific frequencies from the calibrator to the Model 2000 inputs and verifying that the display reads within specified lim­its.
CAUTION Do not exceed 1000 V peak between INPUT HI and INPUT LO, or 8
V•Hz input, because instrument damage may occur.
Follow these steps to verify AC voltage accuracy:
1. Connect the Model 2000 HI and LO INPUT jacks to the AC voltage calibrator as shown in Figure 1-2.
7
×
10
Figure 1-2
Connections for AC volts verifica­tion
5725 Amplifier (Connect to calibrator)
Note: Amplifier required only for 700V, 50kHz output.
Model 2000
2000 MULTIMETER
Input HI
Output HI
!
R
Input LO
Shielded cable
Output LO
5700A Calibrator (Output AC Voltage)
2. Select the AC volts function by pressing the ACV key.
3. Set the Model 2000 for the 100mV range; make sure that REL is disabled.
4. Source 1kHz and 50kHz AC voltages for each of the ranges summarized in Table 1-3, and make sure that the respective Model 2000 readings fall within stated limits.
Table 1-3
ACV reading limits
Performance Verification 1-9
ACV
Range
100mV
10V 100V 750V
*If the 5725A amplifier is not available, change the 700V @ 50kHz step to 219V @ 50kHz. (Reading limits for 219V @ 50kHz = 218.362 to 219.638V.)
1V
Applied AC
voltage
100.0000mV
1.000000V
10.00000V
100.0000V
700.000V*
Reading limits (1 year, 18°C-28°C)
1kHz 50kHz
99.9100 to 100.0900mV
0.999100 to 1.000900V
9.99100 to 10.00900V
99.9100 to 100.0900V
699.355 to 700.645V
99.8300 to 100.1700mV
0.998300 to 1.001700V
9.98300 to 10.01700V
99.8300 to 100.1700V
698.785 to 701.215V
1-10 Performance Verification
V erifying DC current
Check DC current accuracy by applying accurate DC currents from the calibrator to the AMPS input of the Model 2000 and verifying that the display reads within specified limits. Fol­low these steps to verify DC current accuracy:
1. Connect the Model 2000 AMPS and INPUT LO jacks to the calibrator as shown in Fig­ure 1-3.
2. Select the DC current measurement function by pressing the DCI key.
3. Set the Model 2000 for the 10mA range.
4. Source positive and negative full-scale currents for each of the ranges listed in Table 1­4, and verify that the readings for each range are within stated limits.
Figure 1-3
Connections for DC current verifi­cation
Model 2000
2000 MULTIMETER
Input LO
!
R
Output HI
Amps
Output LO
Table 1-4
DCI limits
DCI
Range
10mA
100mA
* Source positive and negative currents with values shown.
1A 3A
Applied DC
current*
10.0000mA
100.0000mA
1.000000A
2.20000A
Reading limits
(1 year, 18°C-28°C)
9.99460 to 10.00540mA
99.9100 to 100.0900mA
0.999160 to 1.000840A
2.19732 to 2.20268A
5700A Calibrator (Output DC Current)
Note: Be sure calibrator is set for
normal current output.
V erifying AC current
Check A C current accurac y by applying accurate AC current at specific frequencies from the calibrator to the Model 2000 input and verifying that the display reads within specified limits. Follow these steps to verify the AC current:
1. Connect the Model 2000 AMPS and INPUT LO jacks to the calibrator as shown in Fig­ure 1-4.
Figure 1-4
Connections for AC current verifi­cation
Model 2000
Performance Verification 1-11
Output HI
!
2000
MULTIMETER
R
Input LO
Amps
Output LO
5700A Calibrator (Output AC Current)
2. Select the AC current function by pressing the ACI key.
3. Set the Model 2000 for the 1A range.
4. Source 1A and 2.2A, 1kHz AC currents as summarized in T able 1-5, and verify that the readings are within stated limits.
Table 1-5
ACI limits
ACI
Range
1A 3A
Applied AC
current
1.000000A
2.20000A
Reading limits
(1 year, 18°C-28°C)
1kHz
0.998600 to 1.001400A
2.19490 to 2.20510A
1-12 Performance Verification
V erifying resistance
Check resistance by connecting accurate resistance values from the calibrator to the Model
2000 and verifying that its resistance readings are within the specified limits.
Follow these steps to verify resistance accuracy:
1. Using shielded 4-wire connections, connect the Model 2000 INPUT and SENSE jacks to the calibrator as shown in Figure 1-5.
2. Set the calibrator for 4-wire resistance with external sense on.
3. Select the Model 2000 4-wire resistance function by pressing the
4. Set the Model 2000 for the 100
5. Recalculate the limits in Table 1-6 based on actual calibrator resistance values.
6. Source the nominal full-scale resistance values for 100 Table 1-6, and verify that the readings are within stated limits.
4 key.
range, and make sure the FILTER is on.
-10M Ω ranges summarized in
Figure 1-5
Connections for resistance verifi­cation (100 10M
ranges)
-
5700A Calibrator
Sense LO
Model 2000
2000
Sense LO
Sense HI
MULTIMETER
!
R
Sense HI
Input HI
Output HI
InputLOOutput
LO
Note : Use shielded low-thermal cables to minimize noise. Enable or disable calibrator external sense as indicated in procedure.
7. Connect the Model 2000 INPUT and SENSE jacks to the calibrator as shown in Figure 1-6.
8. Disable external sense on the calibrator.
9. Set the Model 2000 to the 100M
10. Source a nominal 100M the 100M
range listed in Table 1-6.
resistance value, and verify the reading is within the limits for
range.
Table 1-6
Limits for resistance verification
1k Ω
Performance Verification 1-13
1M Ω
1k Ω
1M Ω
Figure 1-6
Connections for resistance verifi­cation (100M
range)
Range
100 Ω
10k Ω
100k Ω
10M Ω
100M Ω
Nominal
applied
resistance
100 Ω
10k Ω
100k Ω
10M Ω
100M Ω
Model 2000
Sense LO
Sense HI
2000
MULTIMETER
Nominal reading limits
(1 year, 18°C-28°C)
99.9860 to 100.0140
0.999890 to 1.000110k
9.99890 to 10.00110k
99.9890 to 100.0110k
0.999890 to 1.000110M
9.99590 to 10.00410M
99.8470 to 100.1530M
5700A Calibrator (Output 2-wire Resistance)
Input HI
!
R
Output HI
InputLOOutput
LO
Note : Use shielded cables to minimize noise. Disable calibrator external sense mode.
Recalculated limits
Low limit High limit
___________ ___________ ___________ ___________ ___________ ___________ ___________
___________ ___________ ___________ ___________ ___________ ___________ ___________
2
Calibration
2-2 Calibration
Introduction
Use the procedures in this section to calibrate the Model 2000. Calibration procedures in-
clude:
Comprehensive calibration: Calibrate DC and AC voltages, DC and AC currents, and
resistance values.
Manufacturing calibration: Usually only performed at the factory.
W ARNING This information in this section is intended only for qualified service person-
nel. Do not attempt these procedures unless you are qualified to do so.
All the procedures require accurate calibration equipment to supply precise DC and AC volt­ages, DC and A C currents, and resistance v alues. Comprehensiv e A C and DC calibration can be performed any time by a technician either from the front panel, or by using the SCPI commands sent either over the IEEE-488 bus or the RS-232 link.
NOTE Manufacturing calibration is required in the field only if the Model 2000 has been re-
paired.
This section includes the following information:
Environmental conditions: Explains the type of environment needed for calibration.
Calibration considerations: Summarizes test conditions to observe when performing cali-
bration.
Calibration code: Explains how to enter the calibration code to unlock Model 2000 calibra-
tion.
Comprehensive calibration: Summarizes the calibration cycle and also lists recommended
comprehensive calibration equipment.
Front panel calibration: Provides the calibration procedures using the front panel.
SCPI command calibration: Provides the detailed procedures for calibrating the Model
2000 using SCPI commands.
Manufacturing calibration: Explains the manufacturing calibration procedure using both
the front panel and SCPI commands.
Environmental conditions
Conduct the calibration procedures in a location that has:
An ambient temperature of 23°C ±5°C
A relative humidity of less than 80% unless otherwise noted
W arm-up period
Allow the Model 2000 Multimeter to warm up for at least one hour before performing cali-
bration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated in the above section) allow e xtra time for the instrument’ s internal temperature to stabilize. Typ­ically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified tempera­ture range.
Calibration 2-3
Also, allow the test equipment to warm up for the minimum time specified by the manufac­turer.
Line power
The Model 2000 Multimeter requires a line voltage of 100V/120V/220V/240V, ±10% and a line frequency of 45Hz to 66Hz or 360Hz to 440Hz.
2-4 Calibration
Calibration considerations
When performing the calibration procedures:
Make sure that the equipment is properly warmed up and connected to the appropriate input jacks. Also make sure that the correct input jacks are selected with the INPUTS switch.
Make sure the calibrator is in operate before you complete each calibration step.
Always let the source signal settle before calibrating each point.
Do not connect test equipment to the Model 2000 through a scanner or other switching equipment.
If an error occurs during calibration, the Model 2000 will generate an appropriate error message. See Section 3 and Appendix B for more information.
W ARNING The maximum common-mode voltage (voltage between INPUT LO and chas-
sis ground) is 500 V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard. Some of the procedures in this section may expose you to dangerous voltages. Use standard safety precautions when such dangerous voltages are encountered to avoid personal injury caused by electric shock.
Calibration code
Comprehensive calibration code
Before performing comprehensive (user) calibration, you must first unlock calibration by en-
tering the appropriate calibration code.
Front panel calibration code
For front panel calibration, follow these steps:
1. Access the calibration menu by pressing SHIFT then CAL, and note that the instrument displays the following:
CAL: DATES
2. Use the up or do wn range keys to scroll through the av ailable calibration parameters until the unit displays RUN, then press ENTER.
3. The Model 2000 then prompts you to enter a code. (The f actory default code is 002000.) Use the left and right arrow keys to move among the digits; use the up range key to in­crement numbers, and press the down range key to specify alphabetic letters. Confirm the code by pressing ENTER.
4. The Model 2000 allo ws you to define a new calibration code. Use the up and do wn range keys to toggle between yes and no. Choose N if you do not want to change the code. Choose Y if you want to change the code. The unit then prompts you to enter a ne w code. Enter the code, and press ENTER.
Calibration 2-5
Programming the calibration code
If you are performing calibration over the IEEE-488 bus or the RS-232 link, send this com-
mand to unlock the calibration lock:
:CAL:PROT:CODE <up to 8-character string> The default code command is: :CAL:PROT:CODE 'KI002000'
Manufacturing calibration lock
T o unlock manufacturing calibration, press and hold the front panel OPEN k ey while turning
on the power. See Manufacturing calibration at the end of this section for procedures.
Loading...
+ 61 hidden pages