Keithley 2000 User guide

Model 2000 Multimeter
User’s Manual
2000-900-01 Rev. J / August 2010
www.keithley.com
G
A
T
E
E
A
R
AEM
R
RUS
C F
O
IF
NO
ECNED
Model 2000 Multimeter
User’s Manual
©1994-2010, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number: 2000-900-01 Rev. J / August 2010
Table of Contents

1 General Information

Feature overview..................................................................................1-2
Manual addenda ..................................................................................1-3
Safety symbols and terms....................................................................1-3
Specifications.......................................................................................1-3
Inspections...........................................................................................1-4
Options and accessories......................................................................1-5

2 Basic Measurements

Introduction ..........................................................................................2-2
Front panel summary........................................................................... 2-3
Rear panel summary............................................................................2-6
Power-up..............................................................................................2-8
Display ...............................................................................................2-17
Measuring voltage..............................................................................2-18
Measuring current ..............................................................................2-22
Measuring resistance.........................................................................2-24
Measuring frequency and period........................................................2-26
Measuring temperature......................................................................2 -28
Math...................................................................................................2-30
Measuring continuity..........................................................................2-34
Testing di odes ....................................................................................2-35

3 Measurement Options

Introduction ..........................................................................................3-2
Measurement configuration.................................................... .. ... .........3-3
Trigger operations............................................................ ....................3-8
Buffer operations............................................................................... .3-17
Limit operations.................................................................................. 3 -20
Scan operations ................................................................................. 3 -22
System operations .............................................................................3-32

4 Remote Operation

Introduction ..........................................................................................4-2
Selecting a language............................................................................4-4
RS-232 operation.................................................................................4-6
GPIB bus operation and reference.......................................................4-9
Status structure ..................................................................................4-19
Trigger model (GPIB operation).........................................................4-29
Programming syntax ............................ ... .......................................... .4-32
Common commands..........................................................................4-39

5 SCPI Command Reference

SCPI Signal oriented measurement commands ..................................5-3
SCPI command subsystems reference tables.....................................5-7
Calculate subsystem..........................................................................5-20
DISPlay subsystem............................................................................5-26
:FORMat subsystem ..........................................................................5-28
ROUTe subsystem .............................................................................5-32
[SENSe[1]] subsystem.......................................................................5-37
STATus subsystem ............................................................................5-52
:SYSTem subsystem..........................................................................5-61
:TRACe subsystem............................................................................5-68
Trigger subsystem .............................................................................5-70
:UNIT subsystem ...............................................................................5-74

B Status and Error Messages

C Example Programs

Program examples...............................................................................C-2

D Models 196/199 and 8840A/8842A Commands

E IEEE-488 Bus Overview

Introduction..........................................................................................E-2
Bus description....................................................................................E-4
Bus lines..............................................................................................E-6
Bus commands....................................................................................E-8
Interface function codes........................................ .............................E-15

F IEEE-488 and SCPI Conformance Information

Introduction..........................................................................................F-2
1
General
Information
1-2 General Information

Introduction

This section contains general information about the Model 2000 Multimeter. The informa-
tion is organized as follows:
Feature overview
Manual addenda
Safety symbols and terms
Specifications
Inspection
Options and accessories
If you have any questions after reviewing this information, please contact your local Keithley representative or call one of our Applications Engineers at 1-800-348-3735 (U.S. and Canada only). Worldwide phone numbers are listed at the front of this manual.

Feature overview

The Model 2000 is a 6½-digit high-performance digital multimeter. It has 0.002% 90-day basic DC voltage accuracy and 0.008% 90-day basic resistance accuracy. At 6½ digits, the multimeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4½ digits, it can read up to 2000 readings/sec into its internal buffer. The Model 2000 has broad measure ment ranges:
-
DC voltage from 0.1μV to 1000V.
AC (RMS) voltage from 0.1μV to 750V, 1000V peak.
DC current from 10nA to 3A.
AC (RMS) current from 1μA to 3A.
Two and four-wire resistance from 100µΩ to 120MΩ.
Frequency from 3Hz to 500kHz.
Thermocouple temperature from -200°C to +1372°C.
Some additional capabilities of the Model 2000 include:
Full range of functions — In addition to those listed above, the Model 2000 functions include period, dB, dBm, continuity, diode testing, mX+b, and percent.
Optional scanning — For internal scanning, options include the Model 2000-SCAN, a 10-channel, general-purpose card, and the Model 2001-TCSCAN, a 9-cha nnel, thermocouple card with a built-in cold junction. For external scanning, the Model 2000 is compatible with Keithley's Model 7001 and 7002 switch matrices and cards.
Programming languages and remote interfaces — The Model 2000 offers three pro­gramming language choices (SCPI, Keithley Models 196/199, and Fluke 8840A/ 8842A) and two remote interface ports (IEEE-488/GPIB and RS-232C).
Reading and setup storage — Up to 1024 readings and two setups (user and factory defaults) can be stored and recalled.
Closed-cover calibration — The instrument can be calibrated either from the front panel or remote interface.

Manual addenda

!
Any improvements or changes concerning the instrument or manual will be explained in an addendum included with the manual. Be sure to note these changes and incorporate them into the manual.

Safety symbols and terms

The following symbols and terms may be found on the instrument or used in this manual.
The symbol on the instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on the instrument shows that high voltage may be present on the termi­nal(s). Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading used in this manual explains dangers that might result in per­sonal injury or death. Always read the associated information very carefully before perform­ing the indicated procedure.
The CAUTION heading used in this manual explains hazards that could damage the in­strument. Such damage may invalidate the warranty.
1-3
1-4 General Information

Inspection

The Model 2000 was carefully inspected electrically and mechanically before shipment. After unpacking all items from the shipping carton, check for any obvious signs of physical damage that may have occurred during transit. (Note: There may be a protective film over the display lens, which can be removed.) Report any damage to the shipping agent imme diately. Save the original packing carton for possible future reshipment. The following items are included with every Model 2000 order:
Model 2000 Multimeter with line cord.
Safety test leads (Model 1751).
Accessories as ordered.
Certificate of calibration.
Model 2000 User's Manual (P/N 2000-900-00).
Model 2000 Calibration Manual (P/N 2000-905-00).
Model 2000 Support Software Disk including TestPoint run-time applications, TestPoint instrument libraries for GPIB and RS-232, and QuickBASIC examples.
If an additional manual is required, order the appropriate manual package. The manual
packages include a manual and any pertinent addenda.
-

Options and accessories

The following options and accessories are available from Keithley for use with the Model
2000.

Scanner cards

Model 2000-SCAN: This is a 10-channel scanner card that installs in the option slot of the Model 2000. Channels can be configured for 2-pole or 4-pole opera tion. Included are two pairs of leads for connection to Model 2000 rear panel inputs (Keithley P/N CA-109).
Model 2001-TCSCAN: This is a thermocouple scanner card that installs in the option slot of the Model 2000. The card has nine analog input channels that can be used for high-ac­curacy, high-speed scanning. A built-in temperature reference allows multi-channel, cold­junction compensated temperature measurements using thermocouples.

General purpose probes

Model 1754 Universal T est Lead Kit: Consists of one set of test leads (0.9m), two spade lugs, two banana plugs, two hooks, and two alligator clips.
Model 8605 High Performance Modular Test Leads: Consists of two high voltage (1000V) test probes and leads. The test leads are terminated with a banana plug with retractable sheath on each end.
Model 8606 High Performance Probe Tip Kit: Consists of two spade lugs, two alligator clips, and two spring hook test probes. (The spade lugs and alligator clips are rated at 30V RMS, 42.4V peak; the test probes are rated at 1000V.) These components are for use with high performance test leads terminated with banana plugs, such as the Model 8605.
1-5
The following test leads and probes are rated at 30V RMS, 42.4V peak:
Models 5805 and 5805-12 Kelvin Probes: Consists of two spring-loaded Kelvin test probes with banana plug termination. Designed for instruments that measure 4-terminal resistance. The Model 5805 is 0.9m long; the Model 5805-12 is 3.6m long.
Model 5806 Kelvin Clip Lead Set: Includes two Kelvin clip test leads (0.9m) with banana plug termination. Designed for instruments that measure 4-terminal resistance. A set of eight replacement rubber bands is available as Keithley P/N GA-22.
Model 8604 SMD Probe Set: Consists of two test leads (0.9m), each terminated with a surface mount device “grabber” clip on one end and a banana plug with a retractable sheath on the other end.

Low thermal probes

Model 8610 Low Thermal Shorting Plug: Consists of four banana plugs mounted to a 1­inch square circuit board, interconnected to provide a short circuit among all plugs.
1-6 General Information
Model 8611 Low Thermal Patch Leads: Consists of two test leads (0.9m), each with a banana plug with a retractable sheath at each end. These leads minimize the thermally­induced offsets that can be created by test leads.
Model 8612 Low Thermal Spade Leads: Consists of two test leads (0.9m), each terminat­ed with a spade lug on one end and a banana plug with a retractable sheath on the other end. These leads minimize the thermally-induced offsets that can be created by test leads.

Cables and adapters

Models 7007-1 and 7007-2 Shielded GPIB Cables: Connect the Model 2000 to the GPIB bus using shielded cables and connectors to reduce electromagnetic interference (EMI). The Model 7007-1 is 1m long; the Model 7007-2 is 2m long.
Models 8501-1 and 8501-2 Trigger Link Cables: Connect the Model 2000 to other instru­ments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model 8501-1 is 1m long; the Model 8501-2 is 2m long.
Model 8502 Trigger Link Adapter: Allows you to connect any of the six Trigger Link lines of the Model 2000 to instruments that use the standard BNC trigger connectors.
Model 8504 DIN to BNC Trigger Cable: Allows you to connect Trigger Link lines one (V olt­meter Complete) and two (External Trigger) of the Model 2000 to instruments that use BNC trigger connectors. The Model 8504 is 1m long.

Rack mount kits

Model 4288-1 Single Fixed Rack Mount Kit: Mounts a single Model 2000 in a standard 19-inch rack.
Model 4288-2 Side-by-Side Rack Mount Kit: Mounts two instruments (Models 182, 428, 486, 487, 2000, 2001, 2002, 6517, 7001) side-by-side in a standard 19-inch rack.
Model 4288-3 Side-by-Side Rack Mount Kit: Mounts a Model 2000 and a Model 199 side­by-side in a standard 19-inch rack.
Model 4288-4 Side-by-Side Rack Mount Kit: Mounts a Model 2000 and a 5.25-inch instru­ment (Models 195A, 196, 220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.) side-by­side in a standard 19-inch rack.

Carrying case

Model 1050 Padded Carrying Case: A carrying case for a Model 2000. Includes handles and shoulder strap.
2
Basic
Measurements
Basic Measurements
This section summarizes front panel operation of the Model 2000. It is organized as fol­lows:
Front panel summary — Includes an illustration and summarizes keys, display, and connections.
Rear panel summary — Includes an illustration and summarizes connections.
Power-up — Describes connecting the instrument to line power, the power-up se­quence, the warm-up time, and default conditions.
Display — Discusses the display format and messages that may appear while using the instrument.
Measuring voltage — Covers DC and AC voltage measurement connections and low level voltage considerations.
Measuring current — Covers DC and AC current measurement connections and cur­rent fuse replacement.
Measuring resistance — Details two and four-wire measurement connections and shielding considerations.
Measuring frequency and period — Covers frequency and period measurement con­nections.
Measuring temperature — Describes the use of thermocouples for temperature mea­surements.
Math — Covers the mX+b, percent, dBm, and dB math functions performed on single readings.
Measuring continuity — Explains setting up and measuring continuity of a circuit.
Testing diodes — Describes testing general-purpose and zener diodes.
The front panel of the Model 2000 is shown in Figure 2-1. This figure includes important
8
5
4
6
7
1
3
2
2000 MULTIMETER
RANGE
!
F
500V PEAK
FRONT/REAR
3A 250V
AMPS
HI
LO
INPUTS
350V PEAK
1000V PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1REM TALK LSTN SRQ
STAT
REL FILT
4W
BUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
INPUT
SENSE
Ω 4 WIRE
EXIT ENTER
DIGITS RATE
RELFILTER
TRIG
EX TRIG
STORE
RECALL
OPEN CLOSE
DCV
DCI
MX+B
%
dBm
ACV
ACI
Ω2 Ω4
FREQ
TEMP
dB
CONT
PERIOD TCOUPL
LIMITS ON/OFFDELAY
HOLD
SAVE SETUP
CONFIG HALT
TEST
RS232
GPIB
CAL
STEP SCAN
abbreviated information that should be reviewed before operating the instrument.
Figure 2-1 Model 2000 front panel
1 Function keys (shifted and unshifted)
Select measurement function (DC and AC voltage, DC and AC current, 2-wire and 4-wire resistance, frequency, period, temperature with thermocouples), math func
-
tion (mX+b, %, dBm, dB), or special function (continuity, diode test).
2 Operation keys
EXTRIG Selects external triggers (front panel, bus, trigger link) as the trigger
source. TRIG Triggers a measure m ent from the front panel. STORE Enables reading storage. RECALL Displays stored readings and buffer statistics (maximum, minimum, av-
erage, standard deviation). Use ▲ and ▼ to scroll through buffer; use
and to toggle between reading number and reading. FIL TER Displays digital filter status for present function and toggles filter on/off. REL Enables/disables relative reading on present function.
and Moves through selections within functions and operations. If scanner
card installed, manually scans channels. OPEN Opens all channels on internal scanner card; stops scanning. CLOSE Closes selected internal channel. STEP Steps through channels; sends a trigger after each channel. SCAN Scans through channels; sends a trigger after last channel. DIGITS Changes number of digits of resolution. RATE Changes reading rate: fast, medium, slow. EXIT Cancels selection, moves back to measurement display. ENTER Accepts selection, moves to next choice or back to measurement dis-
SHIFT Used to access shifted keys. LOCAL Cancels GPIB remote mode.
play.
3 Shifted operation keys
DELAY Sets user delay between trigger and measurement. HOLD Holds reading when the selected number of samples is within the selected
tolerance. LIMITS Sets upper and lower limit values for readings. ON/OFF Enables/disables limits; selects beeper operation for limit testing. TEST Selects built-in tests, diagnostics, display test. CAL Accesses calibration. SAVE Saves present configuration for power-on user default. SETUP Restores factory or user default configuration. CONFIG Selects minimum/maximum channels, timer, and reading count for step/
scan. HALT Turns off step/scan. GPIB Enables/disables GPIB interface; selects address and language. RS232 Enables/disables RS-232 interface; select s baud rate, flo w control, termina-
tor.
4 Range keys
Moves to higher range; increments digit; moves to next selection.
Moves to lower range; decrements digit; moves to previous selection. AUTO Enables/disables autorange.
5 Annunciators
*(asterisk) Reading being stored. (diode) Instrument is in diode testing function. (speaker) Beeper on for continuity or limits testing.
)))
(more) Indicates additional selections are available. 4W 4-wire resistance reading displayed. AUTO Autoranging enabled. BUFFER Recalling stored readings. CH 1-10 Displayed internal channel is closed. ERR Questionable reading; invalid cal step. FAST Fast reading rate. FILT Digital filter enabled. HOLD Instrument is in hold mode. LSTN Instrument addressed to listen over GPIB. MATH Math function (mX+b, %, dB, dBm) enabled. MED Medium reading rate. REAR Reading acquired from rear inputs. REL Relative reading displayed. REM Instrument is in GPIB remote mode. SCAN Instrument is in scan mode. SHIFT Accessing shifted keys. SLOW Slow reading rate. SRQ Service request over GPIB. STAT Displaying buffer statistics. STEP Instrument is in step mode. TALK Instrument addressed to talk over GPIB. TIMER Timed scans in use. TRIG Indicates external trigger (front panel, bus, trigger link) selected.
6 Input connections
INPUT HI and LO Used for making DC volts, AC volts, 2-wire resistance measure-
ments.
AMPS Used in conjunction with INPUT LO to make DC current and AC
current measurements. Also holds current input fuse (3A, 250V , fast blow, 5×20mm).
SENSE Ω4 WIRE Used with INPUT HI and LO to make 4-wire resistance mea-
sure-
HI and LO ments.
7 INPUTS
Selects input connections on front or rear panel.
8 Handle
Pull out and rotate to desired position.

Rear panel summary

WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1
3 5
2
4 6
VMC
EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC 120 VAC
125mAT
(SB)
220 VAC 240 VAC
120
2
1
34 5
6
#1
1
2
34
5
6
7
8
MADE IN
U.S.A.
INPUT
500V PEAK
350V PEAK
1000V
PEAK
TRIGGER
LINK
SENSE
Ω 4W
HI
LO
!
LINE RATING
50, 60 400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
#2
EXTERNAL TRIGGER INPUT
Trigger Reading
>72μsec
TTL HI
TTL LO
Reading
Complete
VOLT METER COMPLETE OUTPUT
>10μsec
TTL HI
TTL LO
The rear panel of the Model 2000 is shown in Figure 2-2. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Figure 2-2 Model 2000 rear panel
1 Option slot
An optional scanner card (Model 2000-SCAN, 2001-SCAN, or 2001-TCSCAN) in­stalls in this slot.
2 Input connections
INPUT HI and LO Used for making DC volts, AC volts, 2-wire resistance measure-
ments and for connecting scanner card.
SENSE Ω4 WIRE Used with INPUT HI and LO to make 4-wire resistance measure-
ments
HI and LO and also for connecting scanner card.
3 TRIGGER LINK
One 8-pin micro-DIN connector for sending and receiving trigger pulses among other instruments. Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, 8504.
4 RS-232
Connector for RS-232 operation. Use a straight-through (not null modem) DB-9 ca­ble.
5 IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as Models 7007-1 and 7007-2.
6 Power module
Contains the AC line receptacle, power line fuse, and line voltage setting. The Model 2000 can be configured for line voltages of 100V/120V/220V/240VAC at line frequen cies of 45Hz to 66Hz or 360Hz to 440Hz.
2-7
-
Model 2000
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1
3 5
2
4 6
VMC EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC 120 VAC
125mAT
(SB)
220 VAC 240 VAC
120
MADE IN
U.S.A.
INPUT
500V
PEAK
350V
PEAK
1000V PEAK
TRIGGER
LINK
SENSE
Ω 4W
HI
LO
!
LINE RATING
50, 60 400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
120
240
220
100
Fuse
Spring
Window
Line Voltage Selector
Fuse Holder Assembly
2-8 Basic Measurements

Power-up

Line power connection

Follow the procedure below to connect the Model 2000 to line power and turn on the
instrument.
1. Check to see that the line voltage selected on the rear panel (see Figure 2-3) is cor­rect for the operating voltage in your area. If not, refer to the next procedure, “Setting line voltage and replacing fuse.”
CAUTION Operating the instrument on an incorrect line voltage may cause damage to
2. Before plugging in the power cord, make sure that the front panel power switch is in the off (0) position.
3. Connect the female end of the supplied power cord to the AC receptacle on the rear panel. Connect the other end of the power cord to a grounded AC outlet.
WARNING The power cord supplied with the Model 2000 contains a separate ground
the instrument, possibly voiding the warranty.
wire for use with grounded outlets. When proper connections are made, instrument chassis is connected to power line ground through the ground wire in the power cord. Failure to use a grounded outlet may result in per sonal injury or death due to electric shock.
-
Figure 2-3 Power module
4. Turn on the instrument by pressing the front panel power switch to the on (1) po sition.

Setting line voltage and replacing fuse

A rear panel fuse located next to the AC receptacle protects the power line input of the instrument. If the line voltage setting needs to be changed or the line fuse needs to be re placed, perform the following steps.
WARNING Make sure the instrument is disconnected from the AC line and other equip-
ment before changing the line voltage setting or replacing the line fuse.
1. Place the tip of a flat-blade screwdriver into the power module by the fuse holder as­sembly (see Figure 2-3). Gently push in and to the left. Release pressure on the as­sembly and its internal spring will push it out of the power module.
2. Remove the fuse and replace it with the type listed in Table 2-1.
CAUTION For continued protection against fire or instrument damage, only replace
fuse with the type and rating listed. If the instrument repeatedly blows fuses, locate and correct the cause of the trouble before replacing the fuse. See the optional Model 2000 Repair Manual for troubleshooting information.
3. If configuring the instrument for a different line voltage, remove the line voltage se­lector from the assembly and rotate it to the proper position. When the selector is in­stalled into the fuse holder assembly, the correct line voltage appears inverted in the window.
4. Install the fuse holder assembly into the power module by pushing it in until it locks in place.
2-9
-
Table 2-1
Fuse ratings
Line voltage Fuse rating Keithley P/N
100/120V 220/240V
0.25A slow-blow 5×20mm
0.125A slow-blow 5×20mm
FU-96-4 FU-91
2-10 Basic Measurements

Power-up sequence

On power-up, the Model 2000 performs self-tests on its EPROM and RAM and momen­tarily lights all segments and annunciators. If a failure is detected, the instrument momen­tarily displays an error message and the ERR annunciator turns on. (Error messages are listed in Appendix B.)
NOTE If a problem develops while the instrument is under warranty, return it to Keithley
If the instrument passes the self-tests, the firmware revision levels are displayed. An ex­ample of this display is:
REV: A01 A02
where: A01 is the main board ROM revision.
After the power-up sequence, the instrument begins its normal display of readings.
Instruments, Inc., for repair.
A02 is the display board ROM revision.

High energy circuit safety precautions

To optimize safety when measuring voltage in high energy distribution circuits, read and
use the directions in the following warning.
WARNING Dangerous arcs of an explosive nature in a high energy circuit can cause
severe personal injury or death. If the multimeter is connected to a high energy circuit when set to a current range, low resistance range, or any other low impedance range, the circuit is virtually shorted. Dangerous arcing can result even when the multimeter is set to a voltage range if the minimum volt age spacing is reduced in the external connections.
When making measurements in high energy circuits, use test leads that meet the follow-
ing requirements:
Test leads should be fully insulated.
Only use test leads that can be connected to the circuit (e.g., alligator clips, spade lugs, etc.) for hands-off measurements.
Do not use test leads that decrease voltage spacing. These diminishes arc protec­tion and create a hazardous condition.
Use the following sequence when testing power circuits:
1. De-energize the circuit using the regular installed connect-disconnect device, such as a circuit breaker, main switch, etc.
2. Attach the test leads to the circuit under test. Use appropriate safety rated test leads for this application.
3. Set the multimeter to the proper function and range.
4. Energize the circuit using the installed connect-disconnect device and make mea­surements without disconnecting the multimeter.
5. De-energize the circuit using the installed connect-disconnect device.
6. Disconnect the test leads from the circuit under test.
-
WARNING The maximum common-mode voltage (voltage between INPUT LO and the
chassis ground) is 500V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.

Power-on defaults

Power-on defaults are the settings the instrument assumes when it is turned on. The Model 2000 offers two choices for the settings: factory and user. The power-on default will be the last configuration you saved. The SAVE and SETUP keys select the two choices of power-on defaults.
To save presen t configuration as user settings:
1. Configure the instrument as desired for USER default.
2. Press SHIFT then SAVE.
3. Use the and keys to select YES or NO.
4. Press ENTER.
To restore factory or user settings:
1. Press SHIFT then SETUP.
2. Use the and keys to select FACTory or USER.
3. Press ENTER.
Since the basic measurement procedures in this manual assume the factory defaults, re­set the instrument to the factory settings when following step-by-step procedures. Table 2-2 lists the factory default settings.
Table 2-2
Factory defaults
Setting Factory default Autozero
Buffer Continuity
Beeper Digits Rate Threshold
Current (AC and DC)
Digits (AC) Digits (DC) Filter
Count
Mode Range Relative
Value
Rate (AC) Rate (DC)
Diode test
Digits Range Rate
Frequency and Period
Digits Range Relative
Value Rate
Function GPIB
Address Language
Limits
Beeper High limit Low limit
mX+b
Scale factor Offset
Percent
References
On No effect
On 4½ Fast (0.1 PLC) 10Ω
5½ 6½ On
10 Moving average Auto Off
0.0 Medium (DETector BANDwidth 30) Medium (1 PLC)
6½ 1mA Medium (1 PLC)
6½ 10V
Off
0.0 Slow (1 sec) DCV No effect (16 at factory) (SCPI at factory) Off Never +1
-1 Off
1.0
0.0 Off
1.0
Table 2-2 (cont.)
Factory defaults
Setting Factory default Resistance (2-wire and 4-wire)
Digits Filter
Count
Mode Range Relative
Value
Rate
RS-232
Baud Flow Tx term
Scanning
Channels Mode
6½ On 10 Moving average Auto Off
0.0 Medium (1 PLC) Off No effect No effect No effect Off 1-10 Internal
Temperature
Digits Filter
Count
Mode Junction
Temperature Relative
Value Rate Thermocouple Units
5½ On 10 Moving average Simulated 23°C Off
0.0 Medium (1 PLC) J °C
Triggers
Continuous Delay Source
On Auto Immediate
Table 2-2 (cont.)
Factory defaults
Setting Factory default Voltage (AC and DC)
dB reference dBm reference Digits (AC) Digits (DC) Filter
Count
Mode Range Relative
Value Rate (AC) Rate (DC)
No effect 75Ω 5½ 6½ On 10 Moving average Auto Off
0.0 Medium* Medium (1 PLC)

GPIB primary address

The GPIB primary address of the instrument must be the same as the primary address you specify in the controller’s programming language. The default primary address of the instrument is 16, but you can set the address to any value from 0 to 30 by using the following step by step instructions.
1. Press SHIFT then GPIB.
2. Use the and keys to select ADDRess. Or, press ENTER. Once you have
pressed ENTER, the unit automatically displays the address selection.
3. Use the and keys to toggle from ADDRess to the numeric entry. Notice the
values are blinking.
4. Use the and keys to change the numeric entries to the desired address.
5. Press ENTER.
See Section Four — Remote Operation for more GPIB information.

Warm-up time

The Model 2000 is ready for use as soon as the power-up sequence has completed. However, to achieve rated accuracy, all ow the instrument to warm up for one hour. If the instrument has been subjected to extreme temperatures, allow additional time for internal temperatures to stabilize.

Display

Status and error messages

The display of the Model 2000 is primarily used to display readings, along with the units and type of measurement. Annunciators are located on the top, bottom, right, and left of the reading or message display. The annunciators indicate various states of operation. See Fig ure 2-1 for a complete listing of annunciators.
Status and error messages are displayed momentarily . During Model 2000 operation and programming, you will encounter a number of front panel messages. Typical messages are either of status or error variety, as listed in Appendix B.
-

Measuring voltage

The Model 2000 can make DCV measurements from 0.1µV to 1000V and ACV measure-
ments from 0.1µV to 750V RMS, 1000V peak.

Connections

Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can be used; place the INPUTS button in the appropriate position.
2. Select the measurement function by pressing DCV or ACV. Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with autoranging. If you want manual ranging, use the RANGE and keys to select a measurement range consistent with the expected voltage.
4. Connect test leads to the source as shown in Figure 2-4.
CAUTION Do not apply more than 1000V peak to the input or instrument damage may
occur. The voltage limit is subject to the 8 × 107V•Hz product.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher range until an o normal reading is displayed (or press AUTO for autoranging). Use the lowest possible range for the best resolution.
6. Take readings from the display.

Crest factor

AC voltage and current accuracies are affected by the crest factor of the waveform, the ratio of the peak value to the RMS value. Table 2-3 lists the fundamental frequencies at which the corresponding crest factor must be taken into account for accuracy calculations.
Table 2-3
Crest factor limitations
Crest factor Fundamental frequency
2 3 4-5
50kHz 3kHz 1kHz
2001 MULTIMETER
Caution: Maximum Input = 750V RMS, 1000V peak, 8 x 107 V•Hz
AC Voltage
Source
Model 2000
Input Impedence = 1MΩ and 100pF
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2001 MULTIMETER
Model 2000
Caution : Maximum Input = 1010V peak
DC Voltage
Source
Input Resistance = 10MΩ on 1000V and 100V ranges ; > 10GΩ on 10V, 1V and 100mV ranges.
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
Figure 2-4 DC and AC volt­age measure­ments

Low level considerations

For sensitive measurements, external considerations beyond the Model 2000 affect the accuracy. Ef fects not noticeable when working with higher voltages are significant in micro volt signals. The Model 2000 reads only the signal received at its input; therefore, it is im­portant that this signal be properly transmitted from the source. The following paragraphs indicate factors that affect accuracy, including stray signal pick-up and thermal offsets.
-

Shielding

AC voltages that are extremely large compared with the DC signal to be measured may produce an erroneous output. Therefore, to minimize AC interference, the circuit should be shielded with the shield connected to the Model 2000 INPUT LO (particularly for low level sources). Improper shielding can cause the Model 2000 to behave in one or more of the following ways:
Unexpected offset voltages.
Inconsistent readings between ranges.
Sudden shifts in reading.
To minimize pick-up, keep the voltage source and the Model 2000 away from strong AC magnetic sources. The voltage induced due to magnetic flux is proportional to the area of the loop formed by the input leads. Therefore, minimize the loop area of the input leads and connect each signal at only one point.
NOTE Shielded cables should be used for input circuits to avoid interference caused by
conducting RF.

Thermal EMFs

Thermal EMFs (thermoelectric potentials) are generated by thermal differences between the junctions of dissimilar metals. These can be large compared to the signal that the Model 2000 can measure. Thermal EMFs can cause the following conditions:
Instability or zero offset is much higher than expected.
The reading is sensitive to (and responds to) temperature changes. This effect can be demonstrated by touching the circuit, by placing a heat source near the circuit, or by a regular pattern of instability (corresponding to changes in sunlight or the activa tion of heating and air conditioning systems).
To minimize the drift caused by thermal EMFs, use copper leads to connect the circuit to the Model 2000. A banana plug generates a few microvolts. A clean copper conductor such as #10 bus wire is ideal for this application. The leads to the input may be shielded or un shielded, as necessary. Refer to “Shielding”.
Widely varying temperatures within the circuit can also create thermal EMFs. Therefore, maintain constant temp eratures to minimize these thermal EMFs. A shielded en clo su re around the circuit under test also helps by minimizing air currents.
The REL control can be used to null out constant offset voltages.
NOTE Additional thermals may be generated by the optional scanner cards.
-
-

AC voltage offset

Displayed reading VIN()2V
OFFSET
()
2
+=
Displayed reading 100 mV()21.0m V()
2
+=
Displayed reading 0.01V()1106–V×()+=
Displayed reading 0.100005 =
The Model 2000, at 5½ digits resolution, will typically display 100 counts of offset on AC volts with the input shorted. This offset is caused by the offset of the TRMS converter. This offset will not affect reading accuracy and should not be zeroed out using the REL feature. The following equation expresses how this offset (V
Example: Range = 1VAC
Offset = 100 counts (1.0mV) Input = 100mV RMS
The offset is seen as the last digit, which is not displayed. Therefore, the offset is negli­gible. If the REL feature were used to zero the display, the 100 counts of offset would be subtracted from VIN, resulting in an error of 100 counts in the displayed reading.
See Section 3 — Measurement Options for information that explain the configuration op­tions for DC and AC voltage measurements.
) is added to the signal input (VIN):
OFFSET
The Model 2000 can make DCI measurements from 10nA to 3A and ACI measurements
2001 MULTIMETER
Model 2000
Caution: Maximum Input = 3A DC or RMS
Current
Source
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
from 1µAm to 3A RMS.
NOTE See the previous discussion about crest factor in “Measuring voltage” in this sec-
tion.
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the AMPS and INPUT LO terminals. The front inputs must be used; place the INPUTS button in the FRONT position.
2. Select the measurement function by pressing DCI or ACI.
3. Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE and keys to select a measurement range consistent with the expected current.
4. Connect test leads to the source as shown in Figure 2-5.
CAUTION Do not apply more than 3A, 250V to the input or the AMPS fuse will open-
circuit.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher range until a normal reading is displayed (or press AUTO for autoranging). Use the lowest possible range for the best resolution.
6. Take readings from the display.
Figure 2-5 DC and AC cur­rent measure­ments

AMPS fuse replacement

WARNING Make sure the instrument is disconnected from the power line and other
equipment before replacing the AMPS fuse.
1. Turn off the power and disconnect the power line and test leads.
2. From the front panel, gently push in the AMPS jack with your thumb and rotate the fuse carrier one-quarter turn counter-clockwise. Release pressure on the jack and its internal spring will push the jack out of the socket. Remove the fuse and replace it with the same type (3A, 250V, fast blow , 5 × 20mm). The Keithley part number is FU-99-1.
CAUTION Do not use a fuse with a higher current rating than specified or instrument
damage may occur. If the instrument repeatedly blows fuses, locate and correct the cause of the trouble before replacing the fuse. See the optional Model 2000 Repair Manual for troubleshooting information.
4. Install the new fuse by reversing the procedure above.
See Section 3 — Measurement Options for information that explains the configuration op-
tions for DC and AC current measurements.

Measuring resistance

The Model 2000 can make 2-wire and 4-wire resistance measurements from 100µΩ to
120MΩ.

Connections

Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the Model 2000 as follows: A. For Ω2-wire, connect the test leads to INPUT HI and LO.
B. For Ω4-wire, connect the test leads to INPUT HI and LO, and SENSE Ω4 WIRE
HI and LO. Recommended Kelvin test probes include the Keithley Models 5805 and 5806. Either the front or rear inputs can be used; place the INPUTS button in the appropriate position.
2. Select the measurement function by pressing Ω2 or Ω4. Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with autoranging. If you want manual ranging, use the RANGE and keys to select a measurement range consistent with the expected resistance.
4. Connect test leads to the resistance as shown in Figure 2-6.
CAUTION Do not apply more than 1000V peak between INPUT HI and LO or instru-
ment damage may occur.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher range until a normal reading is displayed. Use the lowest possible range for the best resolution.
6. Take a reading from the display.
2001 MULTIMETER
Model 2000
Resistance
Under Test
Shielded
Cable
Optional shield
Note: Source current flows from the INPUT HI to INPUT LO terminals.
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2001 MULTIMETER
Resistance Under Test
Shielded
Cable
Optional shield
Note: Source current flows from the INPUT HI to INPUT LO terminals.
Model 2000
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Figure 2-6 Two- and four­wire resistance measurements
resistance in a shielded enclosure and connect the shield to the INPUT LO terminal of the instrument electrically.
tions for 2-wire and 4-wire resistance measurements.
To achieve a stable reading, it helps to shield resistances greater than 100kΩ. Place the
See Section 3—Measurement Options for information that explains the configuration op-

Measuring frequency and period

The Model 2000 can make frequency measurements from 3Hz to 500kHz on voltage ranges of 100mV, 1V, 10V, 100V, and 750V. Period measurements can be taken from 2µs to 333ms on the same voltage ranges as the frequency.
The instrument uses the volts input terminals to measure frequency. The AC voltage range can be changed with the RANGE than 10% of the full-scale range.
CAUTION The voltage limit is subject to the 8 × 107V•Hz product.

Trigger level

Frequency and Period use a zero-crossing trigger, meaning that a count is taken when the frequency crosses the zero level. The Model 2000 uses a reciprocal counting technique to measure frequency and period. This method generates constant measurement resolution for any input frequency. The multimeter’s AC voltage measurement section performs input signal conditioning.
and keys. The signal voltage must be greater

Connections

2001 MULTIMETER
Model 2000
Caution: Maximum Input = 1000V peak, 8 x 107 V•Hz
AC Voltage
Source
Input Impedance = 1MΩ in parallel with <100pF
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Figure 2-7 Frequency and period measure­ments
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the INPUT HI and LO terminals of the Model 2000. Either the front or rear inputs can be used; place the INPUTS button in the appropriate position.
2. Select the FREQ or PERIOD function.
3. Connect test leads to the source as shown in Figure 2-7.
CAUTION Do not exceed 1000V peak between INPUT HI and INPUT LO or instrument
damage may occur.
4. Take a reading from the display.
See Section 3—Measurement Options for information that explains the configuration op-
tions for frequency and period measurements.

Measuring temperature

2001 MULTIMETER
Model 2000
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4CH5 CH6CH7 CH8CH9 CH1
0
HOLDTRIG FASTMED SLOW AUTOERR
OUT A HI
OUT A LO
Input
LO
Input
HI
Note: This thermocouple card must be inserted into a Keithley Model 2000.
Note: Front or rear inputs can be used.
+
-
CH 2
2001-TCSCAN
The Model 2000 measures temperature with thermocouples. The temperature measure-
ment ranges available depend on the type of thermocouple chosen.
Thermocouples can be connected to the Model 2001-TCSCAN card, which plugs into the option slot of the Model 2000, or to an external thermocouple card, such as a Model 7057A, 7402, or 7014 installed in a Model 7001 or 7002 Switch System.

Connections

Figure 2-8 Thermocouple temperature measurements

Configuration

The following information explains the various configuration options for temperature mea
surements. To select and configure the thermocouple measurement:
Press SHIFT then TCOUPL. Three choices are available using the and keys:
UNITS — C, K, F (Centigrade, Kelvin, Fahrenheit). This parameter selects the dis­played units for temperature measurements.
TYPE — J, K, T (thermocouple type).
JUNC — SIM, CH1 (simulated or referenced at Channel 1). Typically, a thermocou­ple card uses a single reference junction. The Model 2000 can simulate a reference junction temperature or use the reference junction on a switching card. Typical ref erence junction temperatures are 0°C and 23°C.
A simulated reference temperature is the temperature of the junction where the thermo­couple voltage is sensed. It is room temperature if the thermocouple wire is terminated to banana jacks and corrected directly to the multimeter. The accuracy of a temperature mea surement depends on the accuracy of the reference junction.
-
-

Math

Model 2000 math operations are divided into four categories:
mX+b and percent
dBm and dB calculations
Statistics of buffered readings
Limit testing
The first two categories are discussed here; buffered reading statistics and reading limit
testing are described in Section 3 — Measurement Options.
The procedure to select and configure a math operation is summarized as follows:
1. Press SHIFT then the appropriate math key.
2. Configure the parameters for the math operation. Press ENTER when done. (Press SHIFT then the related math function to end the calculation.)
NOTES On c e enabled for a function, the mX+b and percentage calculations are in effect
across function changes. The Model 2000 uses IEEE-754 floating point format for math calculations.

MX + B

This math operation lets you manipulate normal display readings (X) mathematically ac-
cording to the following calculation:
Y= mX + b where: X is the normal display reading
m and b are user-entered constants for scale factor and offset Y is the displayed result

Percent

Percent
Input - Reference
Reference
------------------------------------------
100%×=

Configuration

To configure the mX+b calculation, perform the following steps:
1. Press SHIFT then MX+B to display the present scale factor: M: +1.000000 ^ Enter a value and units prefix. Use the and keys to choose a numerical place
and use the and keys to increment or decrement the digits.
3. Press ENTER to confirm the M value and display the B value: B: +00.00000 m
4. Enter a value and units prefix.
5. Press ENTER to confirm the B value and display the UNITS designation: MXB
6. Scroll through the letters to change and press ENTER when done.
The Model 2000 then displays the result of the calculation.
This item selects the percentage calculation and lets you specify a reference value. The displayed reading will be expressed as a percent deviation from the reference value. The percentage calculation is performed as follows:
where: Input is the normal display reading.
Reference is the user entered constant. Percent is the displayed result.

Configuration

To configure the percent calculation, perform the following steps:
1. Press SHIFT then % to display the present value: REF:+1.000000^ Enter a reference sign, value, and units prefix. Use the and keys to choose
a numerical place and use the and keys to increment or decrement the digits.
3. Press ENTER when done.
The Model 2000 will display the result of the calculation. The result is positive when the input exceeds the reference and negative when the input is less than the reference. Engi neering units are used to show values in the range 1 nano to 1000G . Exponential notation is used above that range.
-
dBm is defined as decibels above or below a 1mW reference. With a user-programmable
dBm = 10 log
V
2
IN
/Z
REF
⎝⎠
⎛⎞
1mW
---------------------------------
reference impedance, the Model 2000 reads 0dBm when the voltage needed to dissipate 1mW through the reference impedance is applied. The relationship between dBm, a refer ence impedance, and the voltage is defined by the following equation:
Where: VIN is the DC or AC input signal.
Z
is the specified reference impedance.
REF
NOTE Do not confuse reference impedance with input impedance. The input impedance
of the instrument is not modified by the dBm parameter.
If a relative value is in effect when dBm is selected, the value is converted to dBm then REL is applied to dBm. If REL is applied after dBm has been selected, dBm math has REL applied to it.
To set the reference impedance, perform the following steps:
1. After selecting dBm, the present reference impedance is displayed (1-9999Ω): REF: 0000
2. To change the reference impedance, use the and keys to select the numeric
position. Then use the and keys to select the desired value. Be sure to press ENTER after changing the reference impedance.
-
NOTES dBm is valid for positive and negative values of DC volts.
The mX+b and percent math operations are applied after the dBm or dB math. For example, if mX+b is selected with m=10 and b=0, the display will read 10.000 MXB for a 1VDC signal. If dBm is selected with Z
= 50Ω, the display will read
REF
130MXB.
Expressing DC or AC voltage in dB makes it possible to compress a large range of mea-
dB= 20 log
V
IN
V
REF
------------------
surements into a much smaller scope. The relationship between dB and voltage is defined by the following equation:
where: VIN is the DC or AC input signal.
V
is the specified voltage reference level.
REF
The instrument will read 0dB when the reference voltage level is applied to the input. If a relative value is in effect when dB is selected, the value is converted to dB then REL
is applied to dB. If REL is applied after dB has been selected, dB has REL applied to it.
To set the reference voltage, perform the following steps:
1. After selecting dB, the present reference voltage level is displayed: REF: +0.000000
2. To change the reference level, use the and keys to select the numeric
position. Then use the and keys to select the desired value. Be sure to press ENTER after changing the reference voltage.
NOTES The dB calculation takes the absolute value of the ratio VIN / V
The largest negative value of dB is -160dB. This will accommodate a ratio of V = 1µV and V
REF
= 1000V.
REF
IN
2001 MULTIMETER
Model 2000
Resistance
Under Test
Note: Source current flows from the INPUT HI to INPUT LO terminals.
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Figure 2-9 Continuity mea­surements
The Model 2000 uses the 1kΩ range to measure circuit continuity. After selecting conti­nuity, the unit prompts you for a threshold resistance level (1Ω-1000Ω). The Model 2000 alerts you with a beep when a reading is below the set level.
To me asure the continuity of a circuit, press SHIFT then CONT, set the threshold resis­tance level and connect the circuit.
NOTE Continuity has a non-selectable reading rate of FAST (0.1 PLC).
Connect the circuit you want to test to the INPUT HI and INPUT LO terminals of the Model
2000. The test current flows from the INPUT HI as shown in Figure 2-9.
You can define a threshold resistance from 1Ω to 1000Ω. The factory setting is 10Ω. Fol­low these steps to define the resistance level:
1. Press SHIFT then CONT. Use the and keys to choose a numerical place and use the and keys to increment or decrement the digits. Enter a value from 1 to 1000.
3. Press ENTER to confirm your setting.

Testing diodes

2001 MULTIMETER
Model 2000
General-purpose
diode
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2001 MULTIMETER
Model 2000
Zener diode
SHIFT
CH1REM TALK LSTN SRQ
STATREL FILT4WBUFFER
MATH REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Note: Source current flows from the INPUT HI to INPUT LO terminals.
With a Model 2000, you can measure the forward voltage drop of general-purpose diodes and the zener voltage of zener diodes. To test diodes, press SHIFT then current range, connect the diode, and take a reading from the display.
NOTE Diode test has a non-selectable reading rate of MEDium (1 PLC).

Connections

Connect the diode leads to the INPUT HI and INPUT LO terminals on the Model 2000. The test current flows from the INPUT HI terminal as shown in Figure 2-10.
Figure 2-10 Diode testing
, set the test

Range

Y ou can set the test current range from the front panel. The choices are 1mA, 100µA, and 10µA. The factory test current setting is 1mA. To set the test current, do following:
1. Press SHIFT then . Use the and keys to scroll through the three test current selections.
The diode test measures voltages on the 3V range for the 1mA test current and the 10V range for the 100µA and 10µA ranges. If a reading is more than 10V, the Model 2000 dis plays the “OVERFLOW” status message.
-
3
Measurement
Options
Measurement Options
This section describes the front panel features of the Model 2000. For those measure­ment options accessible only by a remote interface, refer to Sections 4 and 5. This section is organized as follows:
Measurement configuration — Describes ranging, filtering, relative readings, digits of resolution, and measurement rate.
Trigger operations — Uses a trigger model to explain trigger modes and sources.
Buffer operations — Discusses the reading storage buffer and buffer statistics.
Limit operations — Defines how to set reading limits.
Scan operations — Explains the internal and external scanning capabilities.
System operations — Gives details on setup saving and restoring, selecting a remote interface, and accessing test and calibration.
Measurement Options 3-3
The following paragraphs discuss configuring the multimeter for making measurements.
See the end of Appendix A for information about optimizing readings for speed or accuracy.
The selected measurement range affects both the ultimate digits and accuracy of the measurements as well as the maximum signal that can be measured. The range setting (fixed or auto) for each measurement function is saved when changing functions.
The full scale readings for every range on each function are 20% overrange except for the 1000VDC, 750VAC, 3ADC, 3AAC, and diode test ranges.
Input values more than the maximum readings cause the "OVERFLOW" messages to be displayed.
To select a range, simply press the RANGE or key. The instrument changes one range per keypress. The selected range is displayed for one second.
If the instrument displays the "OVERFLOW" message on a particular range, select a high­er range until an on-range reading is displayed. Use the lowest range possible without caus­ing an overflow to ensure best accuracy and resolution.
Note that the temperature and continuity functions have just one range.
To enable autoranging, press the AUTO key. The AUTO annunciator turns on when au­toranging is selected. While autoranging is selected, the instrument automatically chooses the best range to measure the applied signal. Autoranging should not be used when opti mum speed is required.
Note that up-ranging occurs at 120% of range, while down-ranging occurs at 10% of nom­inal range.
T o cancel autoranging, press AUTO or the RANGE or key . Pressing AUTO to cancel autoranging leaves the instrument on the present range.
The AUTO key has no effect on the temperature, continuity, and diode test functions.
-
FIL TER lets you set the filter response to stabilize noisy measurements. The Model 2000
Conversion #10 #9
#8 #7 #6 #5 #4 #3 #2
Conversion #1
Reading
#1
A. Type - Moving Average, Readings = 10
Conversion #11 #10
#9 #8 #7 #6 #5 #4 #3
Conversion #2
Reading
#2
Conversion #12 #11
#10 #9
#8
#7 #6 #5 #4
Conversion #3
Reading
#3
Conversion #10 #9
#8 #7 #6 #5 #4 #3 #2
Conversion #1
Reading
#1
B. Type - Repeating, Readings = 10
Conversion #20 #19
#18 #17 #16 #15 #14 #13 #12
Conversion #11
Reading
#2
Conversion #30 #29
#28 #27
#26
#25 #24 #23 #22
Conversion #21
Reading
#3
uses a digital filter, which is based on reading conversions. The displayed, stored, or trans mitted reading is simply an average of a number of reading conversions (from 1 to 100).
To select a filter:
1. Press FILTER once if the FILT annunciator is off; press twice if FILT is on.
2. Enter the number of readings.
3. Select the type of filter you want (moving average or repeating), then press ENTER. The FILT annunciator turns on. When a filter is enabled, the selected filter configuration
for that measurement function is in effect.
Pressing FILTER once disables the filter.
NOTE The filter can be set for any measurement function except frequency , period, con-
tinuity, and diode test.

Filter types

The moving average filter uses a first-in, first-out stack. When the stack becomes full, the measurement conversions are averaged, yielding a reading. For each subsequent conver sion placed into the stack, the oldest conversion is discarded, and the stack is re-averaged, yielding a new reading.
For the repeating filter, the stack is filled and the conversions are averaged to yield a read­ing. The stack is then cleared and the process starts over. Choose this filter for scanning so readings from other channels are not averaged with the present channel.
-
-
Figure 3-1 Moving average and repeating fil­ters

Relative

Response time

The filter parameters have speed and accuracy tradeoffs for the time needed to display, store, or output a filtered reading. These affect the number of reading conversions for speed versus accuracy and response to input signal changes.
The rel (relative) function can be used to null offsets or subtract a baseline reading from present and future readings. When rel is enabled, the instrument uses the present reading as a relative value. Subsequent readings will be the difference between the actual input val­ue and the rel value.
You can define a rel value for each function. Once a rel value is established for a mea­surement function, the value is the same for all ranges. For example, if 50V is set as a rel value on the 100V range, the rel is also 50V on the 1000V, 10V, 1V, and 100mV ranges.
Thus, when you perform a zero correction for DCV, Ω2, and Ω4 measurements by en­abling REL, the displayed offset becomes the reference value. Subtracting the offset from the actual input zeroes the display, as follows:
Actual Input – Reference = Displayed Reading A rel value can be as large as the highest range. Selecting a range that cannot accommodate the rel value does not cause an overflow
condition, but it also does not increase the maximum allowable input for that range. For ex­ample, on the 10V range, the Model 2000 still overflows for a 12V input.
To set a rel (relati ve) value, press REL key when the display shows the value you want
as the relative value. The REL annunciator turns on. Pressing REL a second time disables rel.
You can input a REL value manually using the mX+b function. Set M for 1 and B for any
value you want. Pressing REL enables that value to be the relative value. See Section 2 for more information on the mX+b function.

Digits

The display resolution of a Model 2000 reading depends on the DIGITS setting. It has no
effect on the remote reading format. The number of displayed digits does not affect accuracy or speed. Those parameters are controlled by the RATE setting.
Perform the following steps to set digits for a measurement function:
1. Press the desired function.
Press the DIGITS key until the desired number of digits is displayed (3½ to 6½).
NOTE Frequency and period can be displayed with four to seven digits.

Rate

The RATE operation sets the integration time of the A/D converter , the period of time the input signal is measured (also known as aperture). The integration time affects the usable digits, the amount of reading noise, as well as the ultimate reading rate of the instrument. The integration time is specified in parameters based on a number of power line cycles (NPLC), where 1 PLC for 60Hz is 16.67msec and 1 PLC for 50Hz and 400Hz is 20msec.
In general, the fastest integration time (FAST (0.1 PLC) from the front panel, 0.01 PLC from the bus) results in increased reading noise and fewer usable digits, while the slowest integration time (10 PLC) provides the best common-mode and normal-mode rejection. In­between settings are a compromise between speed and noise.
The RATE parameters are explained as follows:
FAST sets integration time to 0.1 PLC. Use FAST if speed is of primary importance (at the expense of increased reading noise and fewer usable digits).
MEDium sets integration time to 1 PLC. Use MEDium when a compromise between noise performance and speed is acceptable.
SLOW sets integration time to 10 PLC. SLOW provides better noise performance at the expense of speed.
NOTE The integration time can be set for any measurement function except frequency,
period, continuity (FAST), and diode test (MEDium). For frequency and period, this value is gate time or aperture.
For the AC functions, MEDium and SLOW have no effect on the number of power line cycles. See the discussion on “Bandwidth” that follows.

Bandwidth

The rate setting for AC voltage and current measurements determines the bandwidth set-
ting:
Slow — 3Hz to 300kHz.
Medium — 30Hz to 300kHz.
Fast — 300Hz to 300kHz.
Bandwidth is used to specify the lowest frequency of interest. When the Slow bandwidth (3Hz to 300kHz) is chosen, the signal goes through an analog RMS converter. The output of the RMS converter goes to a fast (1kHz) sampling A/D and the RMS value is calculated from 1200 digitized samples (1.2s).
When the Medium bandwidth (30Hz to 300kHz) is chosen, the same circuit is used. How­ever, only 120 samples (120ms) are needed for an accurate calculation because the analog RMS converter has turned most of the signal to DC.
In the Fast bandwidth (300Hz to 300kHz), the output of the analog RMS converter (nearly pure DC at these frequencies) is simply measured at 1 PLC (16.6ms).
Table 3-1 lists the rate settings for the various measurement functions. The FAST, MED, and SLOW annunciators are only lit when conditions in the table are met. In other case, the annunciators are turned off.
Table 3-1
Rate settings for the measurement functions
Function
Fast Medium Slow
DCV, DCI ACV, ACI
Ω2W, Ω4W
FREQ, PERIOD dB, dBm (ACV) dB, dBm (DCV) Continuity Diode test
Notes: NPLC = number of power line cycles. BW = lower limit of bandwidth (in Hz). APER = aperture in seconds. N/A = not available. X = setting ignored.
NPLC=0.1 NPLC=1, BW=300 NPLC=0.1 APER=1s NPLC=1, BW=300 NPLC=0.1 NPLC=0.1 N/A
Rate
NPLC=1 NPLC=X, BW=30 NPLC=1 APER=1s NPLC=X, BW=30 NPLC=1 N/A NPLC=1
NPLC=10 NPLC=X, BW=3 NPLC=10 APER=1s NPLC=X, BW=3 NPLC=10 N/A N/A

Trigger operations

Idle

Control
Source
Immediate
External
Event
Detection
Delay
Device Action
Output Trigger
The following paragraphs discuss front panel triggering, the programmable trigger delay,
the reading hold feature, and external triggering.

Trigger model

The flowchart of Figure 3-2 summarizes triggering as viewed from the front panel. It is called a trigger model because it is modeled after the SCPI commands used to control trig gering. Note that for stepping and scanning, the trigger model has additional control blocks. These are described in “Scan operations” later in this section.
Figure 3-2 Front panel trig­gering without stepping/scan­ning
-
Idle
The instrument is considered to be in the idle state whenever it is not performing any mea­surements or scanning functions. From the front panel, the unit is considered idle at the end of a step or scan operation when the reading for the last channel remains displayed. To re­store triggers, use the SHIFT-HALT keys.
Once the Model 2000 is taken out of idle, operation proceeds through the flowchart.

Control source and event detection

The control source holds up operation until the programmed event occurs and is detect­ed. The control sources are described as follows:
Immediate — With this control sour ce, event detection is immediately satisfied allow­ing operation to continue.
External — Event detection is satisfied for any of three conditions:
• An input trigger via the Trigger Link line EXT TRIG is received.
• A bus trigger (GET or *TRG) is received.
• The front panel TRIG key is pressed. (The Model 2000 must be taken out of re-
mote before it will respond to the TRIG key. Use the LOCAL key or send LOCAL 716 over the bus.)

Delay

A programmable delay is available after event detection. It can be set manually or an auto delay can be used. With auto delay, the Model 2000 selects a delay based on the function and range. The AUTO settings are listed in Table 3-2.
Table 3-2
Auto delay settings
Function Range and delay
DCV
ACV
FREQ
DCI
ACI
Ω2W, Ω4W
Continuity
Diode testing
100mV 1ms 100mV 400ms 100mV 1ms
10mA 2ms
100Ω 3ms
1V 1ms 1V 400ms 1V 1ms
100mA 2ms
1kΩ 3ms 1kΩ 3ms 1mA 1ms
10V 1ms 10V 400ms 10V 1ms
1A 2ms 1A 400ms
10kΩ 13ms
100µA 1ms
100V 5ms 100V 400ms 100V 1ms
3A 2ms 3A 400ms
100kΩ 25ms
10µA 1ms
1000V 5ms 750V 400ms 750V 1ms
1MΩ 100ms
10MΩ 150ms
100MΩ 250ms
The delay function is accessed by pressing the SHIFT-DELAY keys. The present delay setting (AUTO or MANual) is displayed. Use the
and keys to select the type of delay.
If MANual is chosen, also enter the duration of the delay. The maximum is shown following:
99H:99M:99.999S
Press ENTER to accept the delay or EXIT for no change.
Changing the delay to MANual on one function changes the delays on all functions to MANual.

Device actions

The primary device action is a measurement. However, the device action block could in-
clude the following additional actions:
Filtering — If the repeating filter is enabled, the instrument samples the specified number of reading conversions to yeildl single filtered reading. Only one reading con­version is performed if the filter is disabled, or after the specified number of reading conversions for a moving average filter is reached. The output of filter feeds hold.
Hold — With hold enabled, the first processed reading becomes the “seed” reading and operation loops back within the device action block. After the next reading is pro­cessed, it is checked to see if it is within the selected window (0.01%, 0.1%, 1%, 10%) of the “seed” reading. If the reading is within the window, operation again loops back within the device action block. This looping continues until the specified number (2 to 100) consecutive readings ar e wi t hi n the window. If one of the readings is not within the window, the instrument acquires a new “seed” reading and the hold pro cess continues.
Channel closure — When stepping or scanning, the last device action is to open the previous channel (if closed) and close the next channel. Using the hold feature pro­vides an auto settling time for the scanner relays. Each open/close transition will re­start the hold process and a reading for each channel will not occur until the relay settles.

Output trigger

After the device action, an output trigger occurs and is available at the rear panel Trigger Link connector. This trigger can be used to trigger another instrument to perform an opera­tion (e.g., select the next channel for an external scan).
-

Counters

The trigger model for stepping and scanning contains additional blocks for counting sam­ples (the number of channels to scan) and counting triggers. These counters are explained in the paragraph “Scan operations” later in this section.

Reading hold (autosettle)

When a hold reading is acquired as described in “Device actions”, an audible beep is sounded (if enabled) and the reading is considered a “true measurement”. The reading is held on the display until an “out of window” reading occurs to restart the hold process.
When operating remotely or scanning, the hold process seeks a new “seed"“once it has been satisfied and the reading has been released. When operating from the front panel, the hold process does not seek a new "seed" until the held condition is removed.

External triggering

Rear Panel Pinout
1
2
34
5
6
7
8
Pin 2
External
Trigger
Input
Pin 1
Voltmeter
Complete
Output
Pin Number Description
1
2
3
4
5
6
7
8
Voltmeter Complete Output
External Trigger Input
no connection *
Signal Ground
no connection *
no connection *
no connection *
Signal Ground
* Either pin 3 or 5 may be configured as an output instead
of pin 1. Either pin 4 or 6 may be configured as an input
instead of pin 2. See the optional Model 2000 Repair
Manual for details.
Figure 3-3 Rear panel pi­nout

Hold example

1. Enable HOLD, select a window percentage and enter a count.
2. Apply test probes to a signal. Once the signal becomes stable enough to satisfy the hold condition, the reading is released, and the beeper sounds (if enabled).
3. Remove the hold condition by lifting the probes. Hold will then seek a new “seed”.
The EXT TRIG key selects triggering from two external sources: trigger link and the TRIG key. When EXT TRIG is pressed, the TRIG annunciator lights and dashes are displayed to indicate that instrument is waiting for an external trigger. From the front p anel, you can press the TRIG key to trigger a single reading. Pressing the EXT TRIG key again toggles you back to continuous triggers.
The Model 2000 uses two lines of the Trigger Link rear panel connector as External Trig­ger (EXT TRIG) input and Voltmeter Complete (VMC) output. The EXT TRIG line allows the Model 2000 to be triggered by other instruments. The VMC line allows the Model 2000 to trigger other instruments.
At the factory, line 1 is configured as VMC and line 2 as EXT TRIG. (Changing this con­figuration is described in the optional Model 2000 Repair Manual.) A connector pinout is shown in Figure 3-3.
Triggers on Leading Edge
TTL High (2V-5V)
TTL Low (0.8V)
2µs Minimum
Meter
Complete
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
10µs Minimum
Figure 3-4 Trigger link input pulse specifica­tions (EXT TRIG)
Figure 3-5 Trigger link out­put pulse specifi­cations (VMC)

External trigger

The EXT TRIG input requires a falling-edge, TTL-compatible pulse with the specifications shown in Figure 3-4. In general, external triggers can be used to control measure operations. For the Model 2000 to respond to external triggers, the trigger model must be configured for it.

Voltmeter complete

The VMC output provides a TTL-compatible output pulse that can be used to trigger other instruments. The specifications for this trigger pulse are shown in Figure 3-5. Typically, you would want the Model 2000 to output a trigger after the settling time of each measurement.

External triggering example

In a typical test system, you may want to close a channel and then measure the DUT con­nected to the channel with a multimeter. Such a test system is shown in Figure 3-6, which uses a Model 2000 to measure ten DUTs switched by a Model 7011 multiplexer card in a Model 7001/7002 Switch System.
2000 MULTIMETER
2000 Multimeter
1
DUT
#1
2
DUT
#2
10
DUT
#10
OUTPUT
Card 1
7011 MUX Card
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1
3 5
2
4 6
VMC EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC 120 VAC
125mAT
(SB)
220 VAC 240 VAC
120
MADE IN
U.S.A.
INPUT
500V
PEAK
350V
PEAK
1000V PEAK
TRIGGER
LINK
SENSE
Ω 4W
HI
LO
!
LINE RATING
50, 60 400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN USA
7001 or 7002 Switch System
2000 Multimeter
Trigger
Link
Trigger
Link Cable
(8501)
Trigger
Link
OUT
IN
Figure 3-6 DUT test system
Figure 3-7 Trigger link con­nections
The Trigger Link connections for this test system are shown in Figure 3-7. Trigger Link of the Model 2000 is connected to Trigger Link (either IN or OUT) of the Model 7001/7002. Note that with the default trigger settings on the Model 7001/7002, line #1 is an input and line #2 is an output. This complements the trigger lines on the Model 2000.
For this example, the Model 2000 and 7001/7002 are configured as follows:

Model 2000:

Factory defaults restored (accessed from SHIFT-SETUP)
External scanning, channels 1 - 10, no timer, 10 readings (accessed from SHIFT-CON­FIG)
External triggers (accessed from EXT TRIG)

Model 7001 or 7002:

Factory defaults restored
Scan list = 1!1-1!10,
Number of scans = 1
Channel spacing = TrigLink
T o run the test and store readings in the Model 2000 with the unit set for external triggers, press STEP or SCAN. The Model 2000 waits (with the asterisk annunciator lit) for an exter nal trigger from the Model 7001/7002.
-
Idle
Bypass
B
Wait for
Trigger Link
Trigger
Scan
Channel
C
Output Trigger
Trigger
D
No
Scanned
10
Channels
?
Yes
7001or 7002
Make
Measurement
Made
10
Measurements
?
2000
Press STEP to start scan
A
Wait for
Trigger Link
Trigger
E
Output Trigger
Trigger
F
No
Yes
Idle
Figure 3-8 Operation model for triggering ex­ample
Press STEP on the Model 7001/7002 to take it out of idle and start the scan. The scan­ner's output pulse triggers the Model 2000 to take a reading, store it, and send a trigger pulse. The following explanation on operation is referenced to the operation model shown in Figure 3-8.
Pressing EXT TRIG then STEP or SCAN on the multimeter places it at point A in
ABC
D
E
F
the flowchart, where it is waiting for an external trigger.
Pressing STEP takes the Model 7001/7002 out of the idle state and places opera-
tion at point B in the flowchart.
For the first pass through the model, the scanner does not wait at point B for a trig-
ger. Instead, it closes the first channel.
After the relay settles, the Model 7001/7002 outputs a Channel Ready pulse. Since the instrument is programmed to scan ten channels, operation loops back up to point B, where it waits for an input trigger.
and Remember that the Model 2000 operation is at point A waiting for a trig­ger. The output Channel Ready pulse from the Model 7001/7002 triggers the multimeter to measure DUT #1 (point E). After the measurement is complete, the Model 2000 outputs a completion pulse (point F) and then loops back to point A, where it waits for another input trigger.
The trigger applied to the Model 7001/7002 from the Model 2000 closes the next channel in the scan. This triggers the multimeter to measure the next DUT. The process continues until all ten channels are scanned and measured.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1 3 5 2
4 6
VMC EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC 120 VAC
125mAT
(SB)
220 VAC 240 VAC
MADE IN
U.S.A.
INPUT
500V
PEAK
350V
PEAK
1000V PEAK
TRIGGER
LINK
SENSE W 4W
HI
LO
!
LINE RATING
50, 60 400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
KEITHLEY
2000 Multimeter
706 Scanner
Channel
Ready
External
Trigger
Model 8503 DIN to BNC Trigger Cable
Figure 3-9 DIN to BNC trig­ger cable

External triggering with BNC connections

An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2000 to instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable has a micro-DIN connector at one end and two BNC connectors at the other end. The BNC ca­bles are labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
Figure 3-9 shows how a Keithley Model 706 Scanner can be connected to the Trigger Link of the Model 2000 using the adapter cable. With this adapter, a Model 706 could be substi tuted for the Model 7001/7002 in the previous example. With the Model 706 set for External Triggering, the test would start when the single scan mode is selected and initiated.
If the Model 2000 trigger line configuration has been changed from the factory setting, the Model 8502 Trigger Link Adapter must be used to interface with instruments having BNC trigger connections. It has two micro-DIN connectors and six BNC connectors, one for each trigger line.
-

Buffer operations

The Model 2000 has a buffer to store from two to 1024 readings and units. It also stores the channel number for scanned readings and overflow readings. In addition, recalled data includes statistical information, such as minimum, maximum, average, and standard devia tion.
The buffer fills with the requested number of readings and stops. Readings are placed in the buffer after any math operations are performed. Buffered data is overwritten each time the storage operation is selected. The data is volatile; it is not saved through a power cycle.
The following paragraphs discuss storing and recalling buffered data.

Storing readings

Use the following procedure to store readings:
1. Set up the instrument for the desired configuration.
2. Press the STORE key. Using the , , , and keys to select the number of readings desired.
4. Press ENTER. The asterisk (*) annunciator turns on to indicate a data storage oper­ation. It will turn off when the storage is finished.
-

Recalling readings

RDG NO. 10 Reading Value RDG NO. 9 Reading Value RDG NO. 8 Reading Value RDG NO. 7 Reading Value RDG NO. 6 Reading Value RDG NO. 5 Reading Value RDG NO. 4 Reading Value RDG NO. 3 Reading Value RDG NO. 2 Reading Value RDG NO. 1 Reading Value STD DEV Standard Deviation Value Average Average Value Min At XX Minimum Value Max At XX Maximum Value
RANGE
RANGE
Figure 3-10 Buffer locations
Use the following steps to view stored readings and buffer statistics:
1. Press RECALL. The BUFFER annunciator indicates that stored readings are being
displayed. The arrow annunciator indicates that more data can be viewed with the
, , , and keys.
2. As shown in Figure 3-10, use the cursor keys to navigate through the reading num­bers, reading values, and statistics. For any of the buffer statistics (maximum, mini­mum, average, statndard deviation), the STAT annunciator is on.
3. Use the EXIT key to return to the normal display.

Buffer statistics

y
X
i
i1=
n
n
-----------------=
X
2
i
i1=
n
1 n
---X
i
i1=
n
⎝⎠
⎜⎟
⎜⎟
⎛⎞
2
⎝⎠
⎜⎟
⎜⎟
⎛⎞
n-1
--------------------------------------------------------------
y =
The MAX AT and MIN AT values are the maximum and minimum values in the buffer . The AVERAGE value is the mean of the buffered readings. The equation used to calculate the mean is:
where: xi is a stored reading
The STD DEV value is the standard deviation of the buffered readings. The equation used to calculate the standard deviation is:
n is the number of stored readings
where: xi is a stored reading
n is the number of stored readings
NOTE The Model 2000 uses IEEE-754 floating point format for math calculations.

Limit operations

Limit operations set and control the values that determine the HI / IN / LO status of sub­sequent measurements. Limits can be applied to all measurement functions except continu­ity. The limit test is performed after mX+b and percent math operations. Unit prefixes are applied before the limit test, for example:
Low limit = -1.0, High limit = 1.0 A 150mV reading equals 0.15V (IN).
Low limit = -1.0, High limit = 1.0 A 0.6kΩ reading equals 600Ω (HI).
You can configure the multimeter to beep or not when readings are inside or outside of
the limit range.

Setting limit values

Use the following steps to enter high and low limit values:
1. Press the SHIFT-LIMITS keys to view the present HI limit value: HI:+1.000000 This value represents the absolute value of that function.
2. Use the , , , and keys to enter the desired value. Move the cursor to the
rightmost position and use the and keys to move the decimal point.
3. Press ENTER to view the present LO limit value: LO:-1.000000 This value represents the absolute value of that function.
4. Enter the desired value for the low limit. Pressing ENTER returns to the normal dis­play.
^
^

Enabling limits

LO IN HI
90
Ω
LO Limit
110
Ω
HI Limit
Figure 3-11 Using limit test to sort 100Ω, 10% resistors
Use the following procedure to turn on the limits operation:
1. Press the SHIFT-ON/OFF keys to view the present beeper status: BEEP: NEVER
2. Use the and keys to change the beeper status (NEVER, OUTSIDE, INSIDE).
Press ENTER when done.
When the multimeter returns to the normal display, the HI/IN/LO status is displayed along with the reading. To disable limit testing, press SHIFT-ON/OFF again. An example of using limits to sort resistors is shown in Figure 3-11.

Scan operations

The Model 2000 can be used with an internal scanner card (Model 2000 SCAN or 2001-TC­SCAN) or with external scanner cards installed in switching mainframes such as the Models 707, 7001, and 7002. The following paragraphs discuss various aspects of using scanning with the Model 2000.

Connection precautions

WARNINGS Connection information for scanner cards is intended for qualified service

Scanning overview

A scanner lets you switch among a number of input signals to the Model 2000 for mea­surement. The channel control and scanning capabilities depend on whether an internal or external card is being used, as well as on the capabilities of the scanner card. Refer to the documentation supplied with the scanner card for specific connection information.
personnel. Do not attempt to connect the DUT or external circuitry to a scan­ner card unless qualified to do so.
To prevent electric shock that could result in serious injury or death, adhere to the following safety precautions:
• Before making or breaking connections to the scanner card, make sure the Model 2000 power is turned off and power is removed from all external circuitry.
• Do not connect signals that will exceed the maximum specifica­tions of the scanner card.
If both the front panel terminals and the scanner card terminals are connect­ed at the same time, the maximum signal ratings of the front panel terminals are reduced to the maximum signal ratings of the scanner card.
As described in the International Electrotechnical Commission (IEC) Stan­dard IEC 664, scanner cards are Installation Category I and must not be con­nected to mains.

Using an internal scanner card

The optional Model 2000-SCAN scanner card lets you step through or scan up to ten 2-
pole channels or five 4-pole channels.
The optional Model 2001-TCSCAN Thermocouple/General Purpose Scanner Card lets you multiplex one of nine 2-pole or one of four 4-pole analog signals into the Model 2000, and/or any combination of 2- or 4-pole analog signals.

Using external scanner cards

When using external channels, the switching mainframe controls the opening and closing of individual channels. To synchronize Model 2000 measurements with external channel clo sures, connect the Trigger Link lines of the multimeter and switching mainframe. Refer to “Trigger operations” earlier in this section for details and an example on using external trig gering.
-
-

Front panel scanner controls

In addition to the trigger keys discussed previously, front panel keys that affect scanner
card operation include:
and — Allow you to manually step through consecutive internal card chan-
nels.
OPEN and CLOSE — Let you selectively open and close internal card channels.
SHIFT-CONFIG — Selects internal or external scanning, scan list, time between scans, and reading count.
STEP — Starts a stepping operation of consecutive channels, where output triggers are sent after every channel closure.
SCAN — Starts a scanning operation of consecutive channels, where an output trig­ger is sent at the end of the scan list.
SHIFT-HALT — Stops stepping or scanning and restores the trigger model to a non­scanning mode.

Using and keys

The and keys can be used to manually scan through channels on the internal scanner card. With a scanner card installed in the option slot, press the key to manually increment channels or the key to manually decrement channels. The annunciator of the closed channel is lit. Hold down either key to manually scan through channels continuously. Press OPEN to open all channels.

Using OPEN and CLOSE keys

The OPEN and CLOSE keys control channels on the internal scanner card only. The keys allow you to directly:
Close a specific channel (or channel pair for 4-wire resistance).
Immediately open any internal closed channel (or channel pair for 4-wire resistance).
With a scanner card installed in the option slot of the Model 2000, the following prompt is displayed when the CLOSE key is pressed:
CLOSE CHAN:01
Use the , , , and keys to display the desired channel (1 to 10) and press EN­TER. The annunciator of the closed channel will be displayed on the front panel along with normal readings. Selecting a different channel from the one that is presently closed will cause the closed channel to open and allow a settling time before closing the selected chan nel.
Channel relays will be closed according to the presently selected function. If a 4-wire function is selected, both the selected channel relay and the matching relay pair will be closed. Fixed 4-pole relay pairs are:
1 and 6 (not available for Model 2001-TCSCAN)
2 and 7
3 and 8
4 and 9
5 and 10
Pressing the OPEN key will immediately open any closed scanner card channel or chan­nel pair for 4-wire resistance.
-

Stepping and scanning trigger model additions

Idle
Control
Source
Immediate
External
Timer
Event
Detection
Delay
Device
Action
Output Trigger
More
Readings
?
Reading
Count
(Trigger counter)
Yes
No
Figure 3-12 Front panel trig­gering with step­ping
The trigger model presented in “Trigger operations” earlier in this section has some addi-
tional capabilities when stepping or scanning. These are outlined below:
Timer — With this control source, event detection is immediately satisfied on the ini­tial pass. Each subsequent detection is satisfied when the programmed timer interval (up to 99H:99M:99.99S) elapses.
Reading counter — For both stepping and scanning, the reading count can be en­tered from SHIFT-CONFIG. (This is referred to as the trigger counter over the bus.) The reading counter can bypass the idle state. Operation will wait until the pro­grammed control source event occurs.
Channel counter — For scanning, the scan list length (maximum channel less mini­mum channel) is used to bypass the control source allowing a specified number of device actions to occur. (This counter is referred to as the sample counter over the bus.)
These additional blocks are shown in the trigger models of Figures 3-12 and 3-13. Uses of the timer control source, reading counter, and channel counter are shown in the scanning examples later in this section.
Idle
Control
Source
Immediate
External
Timer
Event
Detection
Delay
Device Action
Output Trigger
More
Readings
?
Reading
Count
(Trigger counter)
Yes
No
More
Channels
?
Yes
No
Scan List
Length
(Sample counter)
Figure 3-13 Front panel trig­gering with scan­ning

Using SHIFT-CONFIG to configure stepping and scanning

From the SHIFT-CONFIG key combination, you can select internal or external scanning, the minimum and maximum channels in the scan list, the time between scans, and the read ing count.
1. To configu re stepping or scanning, perform the following:
2. Select the desired measurement function.
3. Press the SHIFT-CONFIG keys to access the step/scan configuration. Select the type of scan (INT ernal or EXTernal) by using the and keys and press­ing ENTER.
5. Select the first channel in the scan list (MINimum CHANnel) by using the , ,
, and keys and pressing ENTER.
6. Select the last channel in the scan list (MAXimum CHANnel) and press ENTER to confirm.
7. The next selection is for timed scans. (This is the Timer control source in the trigger model.) It sets a user-specified interval for starting scans. If you choose timed scans, the Model 2000 prompts for a time interval:
00H:00M:00.000S
Use the , , , and keys to select a time interval and press ENTER to con­firm.
8. Next, you are prompted for a reading count (RDG CNT). This can be less than, equal to, or greater than the scan list length (up to 1024). It is the number of readings that will be stored in the buffer. The effects of these choices are further described in the scanning examples.
9. Press ENTER when done to return to the normal display. Note that scanned readings are always stored in the buffer, up to the setting for RDG CNT.
-
TYPE: INT MIN CHAN: 1 MAX CHAN: 10 TIMER? OFF
RDG CNT:
0010 0002
0020
STEP 20 channel closures 20 output triggers
SCAN 10 channel closures (x2) 2 output triggers
RECALL 20 Readings
STEP 10 channel closures 10 output triggers
SCAN 10 channel closures 1 output triggers
RECALL 10 Readings
STEP 2 channel closures 2 output triggers
SCAN 10 channel closures 1 output triggers
RECALL 2 Readings
Note: "Factory setup" on the Model 2000 is assumed.
SHIFT-CONFIG
Figure 3-14 Internal scan­ning example with reading count option
The following examples demonstrate the use of reading count, timed scans, delay, and
external scanning.
One of the configuration options for stepping and scanning is the reading count. The ex-
ample of Figure 3-14 shows how different settings of RDG CNT affect these operations:
With a reading count (0010) equal to the scan list length (10), a step operation con­secutively closes ten channels and sends an output trigger after each channel. A scan operation also consecutively closes ten channels but sends an output trigger only at the end of the scan.
With a reading count (0020) greater than the scan list length (10), stepping yields 20 channel closures and 20 output triggers. Scanning also goes through the scan list twice but sends an output trigger only at the end of each scan.
With a reading count (0002) less than the scan list length (10), stepping yields two channel closures and output triggers. Scanning goes through the entire scan list and sends an output trigger but only two readings are stored.
NOTE If the reading count divided by the scan list length is not an integer, it is rounded
up. For example, if the reading count is 15 and the scan list length is 10, there will be two output triggers for scanning.
The differences between stepping and scanning counters for bus commands are summa-
rized in Table 3-3.
Table 3-3
Bus commands parameters for stepping and scanning counters
Operation :SAMPle:COUNt :TRIGger:COUNt
STEP 1 reading count SCAN scan list length (reading count) / (scan list length)
Another configuration option for stepping and scanning is the timing of channel closures. The example of Figure 3-15 shows how different settings of TIMER and DELAY affect these operations. These are the Timer control source and the Delay block shown in the trigger models of Figures 3-12 and 3-13.
With the timer ON and set to five seconds and delay set to AUTO, channels are stepped through at five-second intervals with an output trigger after each closure. A scan operation yields ten channels scanned immediately with an output trigger at the end of the scan.
With the timer OFF and the delay set to MANual for five seconds, stepping and scan­ning through the channels is timed the same. The difference is in the number of out­put triggers, with stepping sending a trigger after each channel closure and scanning sending a trigger at the end of the scan.
When using both the timer and delay parameters, the timer is not started until after the delay. For example, if the timer is two minutes and the delay is ten seconds, the timer is not started until 10sec af ter p ressing SC AN. Each successive scan will occur at 2:10.0, 4:10.0, etc.
If the total delay time per scan is greater than or equal to the timer setting, the timer con­dition is already satisfied and, effectively, is ignored.
TYPE:INT MIN CHAN: 1 MAX CHAN: 10
TIMER?
TIMER? ON 00H:00M:05.000S
RDG CNT: 0010
STEP 10 channel closures at 5-second intervals 10 output triggers
SCAN 10 channel closures 1 output trigger
RECALL 10 readings
RDG CNT: 0010
DELAY: MAN 00H:00M:05.000S
STEP 10 channel closures at 5-second intervals 10 output triggers
SCAN 10 channel closures at 5-second intervals 1 output trigger
RECALLL 10 readings
OFF
ON
Note: "Factory setup" on the Model 2000 is assumed.
SHIFT-CONFIG
Figure 3-15 Internal scan­ning example with timer and delay options
The example of Figure 3-16 shows the front panel operations to configure an external
1
2
3
4
5
6
7
scan. The trigger and signal connections were shown previously in “Trigger operations”. Both instrument setups assume factory defaults. Set the Model 2000 for the desired mea­surement function.
On the Model 7001 Switch System, enter a scan list of channels 1 to 10 on card 1.
Also on the Model 7001, configure the instrument for Trigger Link triggers and one
scan of ten channels.
On the Model 2000 Multimeter, configure an external scan of the first ten channels.
Set the Model 2000 for external triggers by pressing EXT TRIG . The display will be
dashes.
Press STEP or SCAN on the Model 2000. The asterisk and STEP or SCAN annun-
ciator will light.
Press STEP on the Model 7001 to start channel closures.
After the scan, you can recall ten readings from the Model 2000 buffer.
NOTE When using an external thermocouple scanner card and channel 1 as a reference,
the Model 2000 only recognizes channel 1 when a step or scan is performed. If using a Model 7001 or 7002 to close channel 1 manually , the Model 2000 will not interpret that channel as the reference junction without a step or scan operation.
Model 7001
(from "reset setup")
SCAN CHANNELS 1!1-1!10
CONFIGURE SCAN CHAN-CONTROL CHANNEL-SPACING TRIGLINK ASYNCHRONOUS CHAN-COUNT 10 SCAN-CONTROL SCAN-COUNT 1
Model 2000
(from "factory setup")
SHIFT-CONFIG TYPE:EXT MIN CHAN: 001 MAX CHAN: 010 TIMER? OFF RDG CNT: 0010 ENTER
EX TRIG
STEP or SCAN
STEP
RECALL (10 readings) , , , EXIT
1
2
3
4
5
6
7
Figure 3-16 External scan­ning example with Model 7001

System operations

The Model 2000 has other front panel operations. Saving and restoring setup information is described in Section 2 — Basic Measurements. Selecting the remote interface and lan­guage is covered in Section 4 — Remote Operation.

Self-test

The TEST selections are used as diagnostic tools to isolate problems within the Model
2000. Information on using these test procedures is included in the optional Model 2000 Re pair Manual.

Calibration

The CAL selections are used to view the calibration date and next due date, to perform calibration, and to view the number of times calibration has been performed. Some of the items are password-protected to prevent unintended changing of calibration constants.
T o view the calibration dates, press SHIFT-CAL. Press ENTER at the DA TES prompt. The first date is the last time calibration was performed. The NDUE date is the calibration due date.
Running calibration is password-protected. Refer to the Model 2000 Calibration Manual for details.
To view the calib ration count, press ENTER at the COUNT prompt.
-
4
Remote
Operation
Remote Operation

Introduction

This section includes the following information:
Selecting an interface
Selecting a language
RS-232 operation
GPIB bus operation and reference
Status structure
Trigger model (GPIB operation)
Programming syntax
Common commands

Selecting an interface

The Model 2000 multimeter supports two built-in remote interfaces:
GPIB bus
RS-232 interface
Y ou can use only one interface at a time. The factory interface selection is the GPIB bus. You can select the interface only from the front panel. The interface selection is stored in non-volatile memory; it does not change when power has been off or after a remote interface reset.
Before you select a remote interface, consider the programming language you want to use. For more information about selecting programming languages, see this section.
You can connect a controller to the RS-232 interface. Some considerations for selecting
the RS-232 interface are the following:
You must define the baud rate, enable or disable software handshake XON/XOF.
You can only use the SCPI programming language with the RS-232 interface.
To select RS-232 as the remote interface, do the followi ng:
1. Access the RS-232 configuration by pressing SHIFT then RS232. You see: RS232: OFF Move to the on/off selection by pressing the key.
You see OFF selection blinking. Turn on the RS-232 interface by toggling the selection to ON using the or key
and press ENTER.
You can exit the configuration menu by pressing EXIT. For more information about the RS-232 interface, see section RS-232 operation.
The GPIB bus is the IEEE-488 interface. You must select a unique address for the Model 2000 multimeter. The address is displayed when the multimeter is turned on. At the factory, the address is set to 16.
Since GPIB is the interface selection defined by the factory, only follow these steps to se­lect the GPIB interface if you have been previously using the RS-232 remote programming interface:
1. Select the GPIB option by pressing SHIFT then GPIB. You see: GPIB: OFF Move to the on/off selection by pressing the key.
You see OFF selection blinking slowly. Turn on the GPIB interface by toggling the selection to ON using the or key and
press ENTER.
Turning off the RS-232 interface automatically selects GPIB as the remote programming
interface.
Choose one of three languages to program the Model 2000 multimeter:
SCPI (Signal Oriented Measurement Commands)
Keithley Models 196/199 Digital Multimeter
Fluke Model 8840A/8842A Digital Multimeter
The factory sets the language selection as SCPI.
You only can select a programming language from the front panel. The language selec­tion is stored in non-volatile memory, which means it does not change when power has been off or after a remote interface reset.
Table 4-1 shows the languages supported by the two available interfaces:
Table 4-1
Language support
Language GPIB RS-232
SCPI Keithley Models 196/199 Fluke Model 8840A/8842A
Yes Yes Yes
Yes No No
As you make your language selection, keep in mind that the language you select deter­mines the remote operations allowed.
To select a programming language, follow these steps:
1. Access the GPIB configuration options by pressing SHIFT then GPIB. You see GPIB:ON with GPIB blinking.
2. Select the language configuration option by pressing the ENTER key twice. You see: LANG:<name> Move to the language selection field by pressing the key.
Select the programming language you want by pressing the or key until you see the appropriate language.
The menu scrolls through these choices: SCPI, 199 (Keithley Models 196/199), and 8842 (Fluke Model 8840A/8842A).
5. Confirm your selection by pressing ENTER. The multimeter returns to the measure­ment mode.
Standard Commands for Programmable Instruments (SCPI) is fully supported by the GPIB and RS-232 interfaces. Always calibrate the Model 2000 multimeter using the SCPI language.

Keithley Models 196/199 Digital Multimeter

The Model 2000 multimeter implements virtually all commands available in the Keithley Models 196/199 digital multimeter, except for the self-test and calibration commands. The commands are listed in Appendix D.
See the Models 196/199 Digital Multimeter user manuals for more information about re­mote programming.

Fluke Model 8840A/8842A Digital Multimeter

The Model 2000 Multimeter implements virtually all commands available in the Fluke Models 8840A and 8842A digital multimeter, except for the self-test and calibration com mands. The commands are listed in Appendix D.
See the Fluke user manual for more information about remote programming.
-

RS-232 operation

Sending and receiving data

The RS-232 interface transfers data using 8 data bits, 1 stop bit, and no parity . Make sure
the controller you connect to the multimeter also uses these settings.
You can break data transmissions by sending a ^C or ^X character string to the multime-
ter. This clears any pending operation and discards any pending output.

Selecting baud rate

The baud rate is the rate at which the Model 2000 multimeter and the programming ter-
minal communicate. Choose one these available rates:
19.2k
9600
4800
2400
1200
600
300
The factory selected baud rate is 4800.
When you choose a baud rate, make sure that the programming terminal that you are connecting to the Model 2000 multimeter can support the baud rate you selected. Both the multimeter and the other device must be configured for the same baud rate. T o select a baud rate, follow these steps:
1. Access the RS-232 configuration by pressing SHIFT then RS232. You see: RS232: ON (assuming you have already selected the RS-232 interface) Go to the baud rate field by pressing the key.
You see BAUD:<rate>. Access the baud rate list by pressing the key. You see the rate selection blinking.
Scroll through the available rates by pressing the and key until you find the rate you want.
5. Confirm your selection by pressing ENTER. The multimeter prompts you to define signal handshaking. Continue on for information about handshaking. Y ou can return to measurement mode by pressing EXIT.
Signal handshaking between the controller and the instrument allows the two devices to communicate to each other regarding being ready or not ready to receive data. The Model 2000 does not support hardware handshaking (flow control).
Software flow control is in the form of X__ON and X__OFF characters and is enabled when XonXoFF is selected from the RS232 FLOW menu. When the input queue of the Mod­el 2000 becomes more than 3/4 full, the instrument issues an X_OFF command. The control program should respond to this and stop sending characters until the Model 2000 issues the X_ON, which it will do once its input buffer has dropped below half-full. The Model 2000 rec ognizes X_ON and X_OFF sent from the controller. An X_OFF will cause the Model 2000 to stop outputting characters until it sees an X_ON. Incoming commands are processed after the <CR> character is received from the controller.
If NONE is the selected flow control, then there will be no signal handshaking between the controller and the Model 2000. Data will be lost if transmitted before the receiving device is ready.
Perform the following steps to set flow control:
1. Access the RS-232 configuration by pressing SHIFT and then RS232. You see: RS 232: ON (assuming you have already selected the RS-232 interface). Go to the flow control field by using the or key. You see FLOW: <control>. Access the flow control options by pressing the key. You see the flow control se­lection blinking. Use the or key to display the desired flow control (NONE or XonXoFF) and press ENTER. Y ou will then be prompted to set the terminator . Continue on for infor­mation about the terminator. You can return to the measurement mode by pressing EXIT.
-
The Model 2000 can be configured to terminate each program message that it transmits to the controller with any combination of <CR> and <LF>. Perform the following steps to set the terminator:
1. Access the RS-232 configuration by pressing SHIFT and then RS232. You see: RS 232: ON (assuming you have already selected the RS-232 interface). Go to the terminator field by using the or key.
You see TX TERM: <terminator>. Access the terminator options by pressing the key. You see the terminator selection blinking. Use the or key to display the desired terminator (LF, CR or LFCR) and press
ENTER. The instrument will return to the measurement mode.
Rear Panel Connector
54321
9876
RS232
Figure 4-1 RS-232 interface connector
The RS-232 serial port can be connected to the serial port of a controller (i.e., personal computer) using a straight through RS-232 cable terminated with DB-9 connectors. Do not use a null modem cable. The serial port uses the transmit (TXD), receive (RXD) and signal ground (GND) lines of the RS-232 standard. It does not use the hardware handshaking lines CTS and RTS. Figure 4-1 shows the rear panel connector for the RS-232 interface, and Ta ble 4-2 shows the pinout for the connector.
If your computer uses a DB-25 connector for the RS-232 interface, you will need a cable or adapter with a DB-25 connector on one end and a DB-9 connector on the other, wired straight through (not null modem).
Table 4-2
RS-232 connector pinout
Pin number Description
1 2 3 4 5 6 7 8 9
1
CTS and RTS signals are not used.
no connection TXD, transmit data RXD, receive data no connection GND, signal ground no connection CTS, clear to send1
RTS, ready to send1 no connection
-
See Appendix B for RS-232 error messages.

GPIB bus operation and reference

Introduction

This section contains information about connecting to and using the GPIB (IEEE-488)
bus. The information is organized as follows:
GPIB bus standards
GPIB bus connections
Selecting the primary address
QuickBASIC 4.5 programming
General bus commands
Front panel GPIB operation

GPIB bus standards

The GPIB bus is the IEEE-488 instrumentation data bus with hardware and programming
standards originally adopted by the IEEE (Institute of Electrical and Electronic Engineers) in
1975. The Model 2000 multimeter conforms to these standards:
IEEE-488-1987.1
IEEE-488-1987.2
This standard defines a syntax for sending data to and from instruments, how an instru­ment interprets this data, what registers should exist to record the state of the instrument, and a group of common commands.
SCPI 1991 (Standard Commands for Programmable Instruments)
This standard defines a command language protocol. It goes one step farther than IEEE­488-1987.2 and defines a standard set of commands to control every programmable aspect of an instrument.

GPIB bus connections

Instrument
Controller
InstrumentInstrument
Figure 4-2 IEEE-488 con­nector
Figure 4-3 IEEE-488 con­nections
To connect the Model 2000 multimeter to the GPIB bus, use a cable equipped with stan-
dard IEEE-488 connectors as shown in Figure 4-2.
To allow many parallel connections to one instrument, stack the connector. Two screws are located on each connector to ensure that connections remain secure. Current standards call for metric threads, which are identified with dark-colored screws. Earlier versions had different screws, which were silver-colored. Do not use these types of connectors on the Model 2000 multimeter, because it is designed for metric threads.
Figure 4-3 shows a typical connecting scheme for a multi-unit test system.
To avoid possible mechanical damage, stack no more than three connectors on any one unit.
NOTE To minimize interference caused by electromagnetic radiation, use only shielded
IEEE-488 cables. Available shielded cables from Keithley are models 7007-1 and 7007-2.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1 3 5 2
4 6
VMC
EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC 120 VAC
125mAT
(SB)
220 VAC 240 VAC
MADE IN
U.S.A.
INPUT
500V PEAK
350V PEAK
1000V PEAK
TRIGGER
LINK
SENSE
W 4W
HI
LO
!
LINE RATING
50, 60
400HZ
22 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
KEITHLEY
Figure 4-4 IEEE-488 con­nector location
To connect the Model 2000 multimeter to the IEEE-488 bus, follow these steps:
1. Line up the cable connector with the connector located on the rear panel. The con­nector is designed so that it will fit only one way. Figure 4-4 shows the location of the IEEE-488 connector.
2. Tighten the screws securely, making sure not to over tighten them.
3. Connect any additional connectors from other instruments as required for your appli­cation.
4. Make certain that the other end of the cable is properly connected to the controller. Most controllers are equipped with an IEEE-488 style connector, but a few may re­quire a different type of connecting cable. See your controllers instruction manual for information about properly connecting to the IEEE-488 bus.
NOTE You can only have 15 devices connected to a IEEE-488 bus, including the control-
ler. The maximum cable length is either 20- meters or two meters times the num­ber of devices, whichever is less. Not observing these limits may cause erratic bus operation.

Selecting the primary address

The Model 2000 multimeter ships from the factory with a GPIB address of 16. When the multimeter powers up, it momentarily displays the primary address. You can set the address to a value of 0-30. Do not assign the same address to another device or to a controller that are on the same GPIB bus.
Usually controller addresses are 0 or 21, but see the controllers instruction manual for de­tails. Make certain that the address of the controller is the same as that specified in the con­trollers programming language.
To change the primary address, follow these steps:
1. Access the GPIB configuration settings by pressing SHIFT then GPIB. You see: GPIB:ON, with GPIB blinking Go to Address choice by pressing the key.
You see: ADDR:16 Go to the numeric field by pressing the key.
Enter a new address from 0-30 by using the and ; press ENTER.
5. Return to the main display by pressing EXIT.

QuickBASIC 4.5 programming

Programming examples are written in Microsoft QuickBASIC 4.5 using the Keithley KPC-
488.2 (or Capital Equipment Corporation) IEEE interface and the HP-style Universal Lan­guage Driver (CECHP).

Install the universal language driver

Before any programming example can be run, the Universal Language Driver must first
be installed. To install the driver, from the DOS prompt, enter this command:
cechp
If you include the CECHP command in your AUTOEXEC.BAT file, the driver will automat-
ically be installed every time you turn on your computer.

About program fragments

Program fragments are used to demonstrate proper programming syntax. As the name
implies, only a fragment of the whole program is used to avoid redundancy.
At the beginning of each program, driver files have to be opened. The input terminator
should be set for CRLF. For example:
OPEN "ieee" FOR OUTPUT AS #1 OPEN "ieee" FOR INPUT AS #2 PRINT #1, "interm crlf"
A typical program fragment includes an OUTPUT command and an ENTER command. The OUTPUT command sends a program message (command string) to the Model 2000 multimeter. If the program message includes a query command, then the ENTER command is required to get the response message from the Model 2000 multimeter. The ENTER com mand addresses the Model 2000 multimeter to talk. The following example program frag­ment demonstrates how OUTPUT and ENTER commands are used. Note that the commands assume address 16, which is the factory-set address of the Model 2000 multim­eter.
PRINT #1, "output 16; :func 'volt:ac'; func?"
PRINT #1, "enter 16"
If you wish to display the response message on the CRT, the computer will have to read the message and then “print” it to the CRT display as follows:
LINE INPUT #2, A$
PRINT A$
The following programming example shows how all the above statements are used to­gether. The program fragment is shown in bold typeface.
OPEN "ieee" FOR OUTPUT AS #1 'Open driver
OPEN "ieee" FOR INPUT AS #2 'Open driver
PRINT #1, "interm crlf" 'CRLF terminator
PRINT #1, "output 16;:func 'volt:ac'; func?"
'Select ACV and query
PRINT #1, "enter 16" 'Get response message
LINE INPUT #2, A$ 'Read response message
PRINT A$ 'Display message
-
General commands are those commands, such as DCL, that have the same general meaning regardless of the instrument. T able 4-3 lists the general bus commands along with the programming statement for each command, which use the Keithley KPC-488.2 IEEE in terface and the HP- style Universal Language Driver. Note that the commands requiring that the primary address be specified assume that the address is the factory-set address of 16.
Table 4-3

General bus commands and associated statements

-
Command
REN IFC LLO GTL
DCL SDC GET SPE, SPD
Programming statement
REMOTE 16 ABORT LOCAL LOCKOUT LOCAL 16 LOCAL CLEAR CLEAR 16 TRIGGER 16 SPOLL 16
Effect on Model 2000 Multimeter
Goes into effect when next addressed to listen. Goes into talker and listener idle states. LOCAL key locked out. Cancel remote; restore front panel operation for the 2000. Cancel remote; restore front panel operation for all devices. Return all devices to known conditions. Returns Model 2000 to known conditions. Initiates a trigger. Serial Polls the Model 2000.

REN (remote enable)

The remote enable command is sent to the Model 2000 by the controller to set up the in­strument for remote operation. Generally, the instrument should be placed in the remote mode before you attempt to program it over the bus. Simply setting REN true does not ac­tually place the instrument in the remote state. You must address the instrument to listen af­ter setting REN true before it goes into remote.
Note that the instrument need not be in remote to be a talker.
Program fragment
PRINT #1, "remote 16" 'Place the Model 2000 in re-
mote; turn on REM annunciator
Note that all front panels controls except for LOCAL (and POWER) are inoperative while the instrument is in remote. You can restore normal front panel operation by pressing the LOCAL key.
The IFC command is sent by the controller to place the Model 2000 multimeter in the lo­cal, talker, listener idle states. The unit responds to the IFC command by canceling front pan­el TALK or LSTN lights, if the instrument was previously placed in one of those states.
Note that this command does not affect the status of the instrument; settings, data, and
event registers are not changed.
T o send the IFC command, the controller need only set the IFC line true for a minimum of
100µs.
Program fragment
PRINT #1, "output 16; *idn?" 'Send query command PRINT #1, "enter 16" 'Read data; turn on TALK annun-
ciator SLEEP 3 'Wait 3 seconds PRINT #1, "abort" 'Talker idle state; turn off
TALK annunciator

LLO (local lockout)

Use the LLO command to prevent local operation of the instrument. After the unit receives
LLO, all its front panel controls except the POWER are inoperative. In this state, pressing the LOCAL will not restore control to the front panel. The GTL command restores control to the front panel.
Program fragment
PRINT #1, "remote 16" 'Place 2000 in remote PRINT #1, "local lockout" 'Lock out front panel (includ-
ing LOCAL key) SLEEP 6 'Wait 6 seconds PRINT #1, "local 16" 'Restore front panel operation
Use the GTL command to put a remote-mode instrument into local mode. The GTL com-
mand also restores front panel key operation.
Program fragment
PRINT #1, "remote 16" 'Place 2000 in remote SLEEP 3 'Wait 3 seconds PRINT #1, "local 16" 'Place 2000 in local mode
Use the DCL command to clear the GPIB interface and return it to a known state. Note that the DCL command is not an addressed command, so all instruments equipped to im­plement DCL will do so simultaneously.
When the Model 2000 multimeter receives a DCL command, it clears the Input Buf fer and Output Queue, cancels deferred commands, and clears any command that prevents the processing of any other device command. A DCL does not affect instrument settings and stored data.
Program fragment
PRINT #1, "clear" 'Clear all devices
The SDC command is an addressed command that performs essentially the same func­tion as the DCL command. However, since each device must be individually addressed, the SDC command provides a method to clear only selected instruments instead of clearing all instruments simultaneously, as is the case with DCL.
Program fragment
PRINT #1, "clear 16" 'Clear 2000
GET is a GPIB trigger that is used as an arm, scan and/or measure event to control op­eration. The Model 2000 multimeter reacts to this trigger if it is the programmed control source. The control source is programmed from the SCPI: TRIGger subsystem.
With the instrument programmed and waiting for a GPIB trigger, the following program fragment will provide the GET:
Program fragment
PRINT #1, "trigger 16" 'Trigger 2000 from over the bus
This sends IEEE-488 commands UNT UNL LISTEN 16 GET. When the command is ex­ecuted, the trigger event occurs. (The command TRIGGER just sends GET. Any other lis­teners are triggered when the command is executed.)

SPE, SPD (serial polling)

Use the serial polling sequence to obtain the Model 2000 serial poll byte. The serial poll byte contains important information about internal functions, (see “status structure”). Gener­ally, the serial polling sequence is used by the controller to determine which of several in­struments has requested service with the SRQ line. However, the serial polling sequence may be performed at any time to obtain the status byte from the Model 2000 multimeter.
Program fragment
PRINT #1, "spoll 16" 'Serial poll the 2000
INPUT #2, S 'Read serial poll byte
PRINT S 'Display the decimal value of
the serial poll byte

Front panel GPIB operation

This section describes aspects of the front panel that are part of GPIB operation, including
messages, status indicators, and the LOCAL key.

Error and status messages

See Section 2 for a list of error and status messages associated with IEEE-488 program­ming. The instrument can be programmed to generate an SRQ, and command queries can be performed to check for specific error conditions.

GPIB status indicators

The REM (remote), TALK (talk), LSTN (listen), and SRQ (service request) annunciators show the GPIB bus status. Each of these indicators is described below.
REM — This indicator shows when the instrument is in the remote state. REM does not necessarily indicate the state of the REM line, as the instrument must be ad­dressed to listen with REM true before the REM indicator turns on. When the instru­ment is in remote, all front panel keys, except for the LOCAL key, are locked out. When REM is turned off, the instrument is in the local state, and front panel operation is restored.
TALK — This indicator is on when the instrument is in the talker active state. Place the unit in the talk state by addressing it to talk with the correct MTA (My Talk Ad­dress) command. TALK is off when the unit is in the talker idle state. Place the unit in the talker idle state by sending an UNT (Untalk) command, addressing it to listen, or sending the IFC (Interface Clear) command.
LSTN — This indicator is on when the Model 2000 Multimeter is in the listener active state, which is activated by addressing the instrument to listen with the correct MLA (My Listen Address) command. LSTN is off when the unit is in the listener idle state. Place the unit in the listener idle state by sending UNL (Unlisten), addressing it to talk, or sending IFC (Interface Clear) command over the bus.
SRQ — Y ou can program the instrument to generate a service request (SRQ) when one or more errors or conditions occur. When this indicator is on, a service request has been generated. This indicator stays on until the serial poll byte is read or all the conditions that caused SRQ have ceased to exist. See “status structure” for more in formation.
-

LOCAL key

The LOCAL key cancels the remote state and restores local operation of the instrument. Pressing the LOCAL key also turns off the REM indicator and returns the display to nor-
mal if a user-defined message was displa yed.
If the LLO (Local Lockout) command is in effect, the LOCAL key is also inoperative.
0
2 3
5 6
Cal
7
9 10 11 12 13
14
15
Questionable
Condition
Register
(Always Zero)
0
2 3
5 6 7
9 10 11 12 13
14
15
Questionable
Event
Register
0
11 1
2 3
5 6 7
9 10 11 12 13
14
15
Questionable
Event
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
3
Trig
7 8
Idle 11 12 13
15
Operation Condition
Register
(Always Zero)
3
7 8
Idle 11 12 13
15
Operation
Event
Register
3
7 8
Idle
11 12 13
15
Operation
Event Enable Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
Idle
EAV QSB MAV ESB
RQS/MSS
OSB
Status
Byte
Register
1 EAV QSB MAV
ESB
6
OSB
Service Request Enable Register
&
&
&
&
&
&
&
Logical
OR
*STB?
*SRE *SRE?
Master Summary Status (MSS)
MSB = Measurement Summary Bit EAV = Error Available QSB = Questionable Summary Bit MAV = Message Available ESB = Event Summary Bit RQS/MSS = Request for Service/Master Summary Staus OSB = Operation Summary Bit
Error Queue
Output Queue
Trig Trig
Note : RQS bit is in serial poll byte, MSS bit is in *STB? response.
1
14 14
OPC
QYE DDE
EXE CME URQ
PON
8 9
8 11 12
13
15
Standard
Event
Status
Register
8 9
8 11 12
13
15
Standard
Event Status
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
(Always Zero)
Operation Complete
Query Error
Device Specific Error
Execution Error
Command Error
User Request
Power On
OPC
QYE DDE
EXE CME URQ
PON
*ESR? *ESE
*ESE?
MSB MSB
12 13 14 15
(Always Zero)
12 13 14 15
Measurement
Event
Register
12 13 14 15
Measurement
Event
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
Measurement
Condition
Register
ROF
LL HL
3 4
RAV
BAV
BHF
BFL
ROF
LL HL
RAV
BAV
BHF BFL
ROF LL1 HL1
RAV
BAV
BHF BFL
Reading Overfolw
Low Limit
High Limit
Reading Available
Buffer Available
Buffer Half Full
Buffer Full
11
Cal Cal
Calibration Summary
Command Warning
Warn Warn Warn
222
Temperature Summary
Temp Temp Temp
6
10
66
10 10
999
3 4
3 4
666
0 1
0 1
0 1
Meas Meas Meas
11 11 11
Measuring Triggering

Status structure

See Figure 4-5 for the Model 2000 Multimeters status structure. Instrument events, such as errors, are monitored and manipulated by four status register sets. Notice that these sta tus register sets feed directly into the Status Byte Register. More detailed il lustrations of these register sets are provided by Figures 4-5 through 4-9.
Figure 4-5 Model 2000 sta­tus register struc­ture
-

Condition registers

As Figure 4-5 shows, all status register sets have a condition register. A condition register is a real-time, read-only register that constantly updates to reflect the current operating con ditions of the instrument. For example, while a measurement is being performed, bit B4 (Meas) of the Operation Condition Register is set. When the measurement is completed, bit B4 clears.
Use the :CONDition? query commands in the STATus Subsystem to read the condition registers. See Section 5 for more information.

Event registers

As Figure 4-5 shows, each status register set has an event register. An event register is a latched, read-only register whose bits are set by the corresponding condition register, Once a bit in an event register is set, it remains set (latched) until the register is cleared by a specific clearing operation. The bits of an event register are logically ANDed with the bits of the corresponding enable register and applied to an OR gate. The output of the OR gate is applied to the Status Byte Register.
Use the *ESR? Common Command to read the Standard Event Register. All other event registers are read using the :EVENt? query commands in the STATus Subsystem. See Sec tion 5 for more information.
An event register is cleared when it is read. The following operations clear all event reg­isters:
Cycling power
Sending *CLS
-
-

Enable registers

As Figure 4-5 shows, each status register set has an enable register. An enable register is programmed by you and serves as a mask for the corresponding event register. An event bit is masked when the corresponding bit in the enable register is cleared (0). When masked, a set bit in an event register cannot set a bit in the Status Byte Register (1 AND 0 = 0).
To use th e Status Byte Register to detect events (i.e., serial poll), you must unmask the events by setting (1) the appropriate bits of the enable registers.
To program and query the Standard Event Status Register, use the *ESE and *ESE? Common Commands respectively. All other enable registers are programmed and queried using the :ENABle and :ENABLe? commands in the STATus Subsystem. See Section 5 for more information.
An enable register is not cleared when it is read. The following operations affect the en­able registers:
Cycling power - Clears all enable registers
:STATus:PREset clears the following enable registers: Operation Event Enable Register Questionable Event Enable Register Measurement Event Enable Register *ESE 0 - Clears the Standard Event Status Enable Register.
Loading...