Keithley 193A Service manual

Instruction Manual
Model 193A
System DMM
01985, Keithley Instruments, Inc.
Instrument Division
Cleveland, Ohio, U.S.A.
Document Number 193A-901-01
SAFETY PRECAUTIONS
The following safety precautions should be observed before operating the Model ~193A
This instrument is intended for use by qualified personnel who recognize shock huxds .~nd are i,lmiliar with the safety precautions required to avoid possible injury. Read WET the manual carefully before operating this instrument.
Exercise extreme caution when a shock hazard is present at the instrument’s input. The American National Standards Institute (ANSI) states that a shock hazards exists when voltage levels greater than 3OV rms or
42.4V peak are present. A good safety practice is to expect that a hazardous voltage is present in any unknown circuit before measuring.
ect the test leads for possible wear, cracks or breaks before each use. If any defects are found, replxe
Ins
.R
test leads that have the same measure of safety as those supplied with the instrument.
wt
For optimum safety do not touch the test leads or the instrument while power is applied to the circuit under test. Turn the power off and discharge all capacitors, before connecting or disconnecting the instrunwnt. Always disconnect all unused test leads from the instrument.
Do not touch any object which could provide a current path to the common side of the circuit under lest or power line (earth) ground. Always make measurements with dry hands while standing on ‘1 dry, ill­s&ted surface, capable of withstanding the voltage being measured.
Exercise extreme safety when testing high energy power circuits (AC line or mains, etc). Refer to the t ligh Energy Circuit Safety Precautions found in paragraph 2.6 (Basic Measurements).
Do not exceed the instrument’s maximum allowable input as defined in the specifications and operation section.
SPECIFICATIONS
DC VOLTS
MAXIMUM
READING RESOLUTION
RANGE (5% Digits) 6% 5%
200mV 219.999mV 10” n” 1 IrV
2 V 2.19999 V 1 $V lop”
20 V 21.9999 V
200 V 219.999 V 100 ,,V
1000 v 1000.00 v
*When properly zeroed. **Multiply digit error by 10 for h%~digit accuracy.
10 &V 100 &V
lm”
1mV 1OmV
INPUT 24
RESISTANCE 23O+l”C 23’*5T
>lCX >lCx 0.002+
1OMCl 0.003+ 1OMR 0.003+1 O.O07+l 1OMR 0.004+1
ACCURACY (5% Digits)”
* t%rdg +
Hr.,
0.003+2* 0.005+2’
1
1
counts)
90 Days,
0.005+2
0.007+1
0.007+1
TEMPERATURE
COEFFICIENT
1 Yr.,
23’f5”C O”-18’C & 28--5O~C TIME”*
0.00x+2* o.wK? in 1
0.007+2
0.009+3 (1.lx)O7+O.l
0.009+3 O.WO7
0.009+5 0.0007+0.1
“‘To (1.01% of step ch.qe.
it%rdg+countsli~~ C SETTLING ~
(l.WOS +~ 0. I
+o. I
ANALOG 1
< zms
i
< 1,115 < I”,> < lms < I”,5
NMRR: Greater than 60dB at 50 or 60Hr. MAXIMUM MEASUREMENT RATES (into internal CMRR: Greater than
(with lk0 in either lead).
120dB at dc and 50 or 6OHz
MAXIMUM ALLOWABLE INPUT: 1OOOV peak.
memory, filter and multiplex off):
3% Digit
1000 rdgis
4% Digit 5% Digit
333 rdgls 25 rdg:s
6’12 Digit
25
rdg s
TRMS AC VOLTS (Option 1930)
MAXIMUM
READING
RANGE
2" 2.19999"
2ov 2oov 219.999" 700" 700.00
*Above 2000 COUPES. **Above 20000 counts; 3%+500 typical below 20000. “‘Multiply digit erwr by ill for 6’.:-digit .~cur.xy. ~
(5% Digits)
21.9999"
v
RESOLUTION + t%rdg + counts, 1 Year, 18’.28 c
6% 5%
1 Ir”
10 /Lv
100 PV
1mV
100 flv
10 pv
1mV 1+ 100 0.25+
lOmV 1+100
20Hz-50Hz’
1 + 100
1+ 100
ACCURACY (5% Digits)‘*’
50Hz-IOkHz’
0.25+
I(10
0.25+100
100
0.35+100
IOkHz-20kHz’ ZOkHz-100kHz”
11.35 + 300
L1.35
13.31+ 3Nl
0.5 4~ SW
+ SW
I f so0 I f xx1
I _ 3ki
I + SiHl
RESPONSE:
CREST FACTOR: Rated accuracy to 3.
pulse widths > lops, peak voltage 5 1.36 x full scale.
True root mean square, ac or
ac+dc.
Specified for
AC+DC: Add 60 counts to specified accuracy. MAXIMUM INPUT: 1OOOV peak ac + dc, 2 x lO’V*Hr.
SETTLING TIME: 0.5s to within 0.1% of change
reading.
INPUT IMPEDANCE: 1Mfi shunted
by
less than 120pF.
TEMPERATURE COEFFICIENT CO’=-18°C & 28’-50°C):
Less than *(0.1x applicable accuracy specification)i”C below 50kH.z; (0.2x) for 50kHz to 100kHr.
200mv
2"
20 200 700
100 pv
ImV 2+3
v v v
1OmV 2+3
100mV 2+3
1 v
2+3
2+3
3dB BANDWIDTH: CMRR: Greater than
balance).
dBV (Ref. = 1V):
in
INPUT ZOHz-ZOkHz ZOkHz-1WkHr RESOLUTION
14 to +57dHV
(2OOmV t* 7WV rms) 0.2
-34 to 14dUV
(20mV t0 2OOmV) I .s S’
‘Typical
500kHr typical.
6OdB at 50 and 6OHr (Ikl2 1111. ~
ACCURACY +dBV
1 Year. In~-zn c
11.4
,J OldH\’
i)~illdH\
!
OHMS
MAXIMUM CURRENT
READING RESOLUTION THROUGH
RANGE (5% Digits) 6% 5’,2
200 n 219.999 n 100 po lmfl ImA
2 kO 2.19999 k!I ImIl 1Omfl ImA
20 kn 21.YYYY k0 Nllll~
200 kR 219.999 kll 1OOmR 1 n 10 I”A
2Mfl 2:19999Mn 1 R ‘IO n 14
20Mll 21.9999MQ 10 n 100 a 100 nA
200Mn 21Y.YY9Mn 100 R 1 kn 100 nA**
*When properly zeroed. ***4-terminal accuracy 200S20k range. Multiply digit error by 10 for 6’/zd accuracy.
**Nominal short circuit current.
1oomn 100 &A
UNKNOWN
ACCURACY (5% Digits)***
-t (%rdg + counts)
24
Hr.,
230
* 1°C
0.0035+2* 0.007+2* 0.010+2*
0.0035+2 0.007+2 0.010+2
0.0035+2 0.007+2 0.010+2
0.0035+2 0.007+2 0.010+2
0.005 +2 0.010+2 Ll.o10+2
0.040 +2 0.070+2 0.070~2
3.2 +2 3.2 +2 3.2 +2
90 Days, 1 Yr., 23’ i ST 230 i5T
TEMPERATURE
COEFFICIENT
*(%rdg+counts)/“C
O”lBoC & 28’-5O’C
0.001+0.7
0.001+ 0.1
0.001+0.1
0.001+0.1
0.001+0:1
0.010+0:1
0.230+0.1
CONFIGURATION: MAXIMUM ALLOWABLE INPUT: 350V
250V rms.
Automatic 2. or 4.terminal.
peak or
DC AMPS (Option 1931)
ACCURACY
(5% Digits)
(S/2 v,ges, 1
RESO- RESO- (1 Year) (1 Year) COEFFICIENT LUTION *(%rdg+counts)*(
RANGE 15% Dieits)
RANGE (5% Digits)
200 pA InA 0.09+10
20mA lOOnA
200mA l@A
LUTION *(%rdg+counts) -t(%rdg+counts)i”C
2mA 10nA 0.09+10
2 A lo+4
18*28”C
18*28”C
0.091-10 0.01+0.5
0.09+10 0.01+0.5 o.o9+1O 0.01+0.5
TEMPERATURE
0’
00-18T & 280-SOT
0.01+0.5
0.01+0.5
TRMS AC AMPS (Options 1930 and 1931)
ACCURACY*
RESO-
LUTION +(%rdg+counts) -t(%rdg+counts)/“C
RANGE (5% Digits) GO fill
2mA 1OnA
20mA IOOnA
200mA
2 A 10fiA
*Abwc 2000 counts.
‘InA
‘IpA
45Hz to
1OkHz
(5% Digits) TEMPERATURE
(1 Year) COEFFICIENT
180-28°C
0.6+300 0.04+10
0.6+300
0.6+300
0.6+300
0.6+300
0”-18”C & 28%5O’C
0.04+10
0.04+10
0.04+10
0.04+10
MAXIMUM OPEN CIRCUIT VOLTAGE: 7V
OVERLOAD PROTECTION: 2A
accessible.
fuse (25OV), externally
MAXIMUM VOLTAGE BURDEN: 0.25V on 200pV
through 20mA ranges; 0.28V on 200mA range; 1V on 2A range.
RESPONSE: CREST FACTOR:
pulse width > lms, peak current 5 1.36 x full scale.
True root mean square, ac or ac+dc.
Rated accuracy to 3. Specified for
MAXIMUM VOLTAGE BURDEN: 0.25V on 2OOpV
through 20mA ranges; 0.28V on 200mA range; 1V on 2A range.
OVERLOAD PROTECTION: 2A
accessible.
SETTLING TIME: 0.5s
reading.
to within 0.1% of change in
fuse (25OV), externally
dB (Ref. = ImA):
ACCURACY + dB
1 Year, 18O-28°C
INPUT
-34 to +66dB 20pA to 2A
-54 to -34dB
2pA t” 20/~/I
45Hz-10kHz
0.3
2
RESOLUTION
O.O~ldB
O.OldU
TEMPERATURE
THERMO-
COUPLE
TYPE RANGE
-100 to + 760°C O.l”C
J
K -100 t” +1372”C O.l”C T
R 0 to +1768”C 1 “C
-100 to + 400°C O.l”C
E - 100 to + 1000°C O.l”C kO.h”C
s 0 to c176PC 1 “C *3 “C
B
c350 t” +1821”C 1 “C
(Thermocouple) (over IEEE Blrs only)
ACCURACY=
(1 Year)
RESOLUTION 113°-28”C
10.5”C +0.5”C +0.5oc
*3 “C
1-5 “C
‘Relative to external 0°C reference junction; exclusive of ther-
mocouple errors. Junction temperature may be external.
TEMPERATURE
(RTD)
4-WIRE
RESO- ACCURACY* TEMPERATURE
RANGE LUTION 1 Yr.,
1000 to
+63O”C O.O05~C)/~C
148O to
+llOO”F 0.01 ‘F)PC
*Excluding probe errors.
O.Ol"C
O.Ol”F 10.36OF
18°-2~oC
*o.lt?"c
COEFFICIENT
* (0.0013% +
*(0.0013%+
RTD TYPE: 1000 platinum; DIN 43 760 or IPTS-68, alpha
0.00385 or 0.00392, 4.wire.
MAXIMUM LEAD RESISTANCE (each lead): 4.wire,
10R. SENSOR CURRENT: 1mA COMMON MODE REJECTION: Less than 0.005”CIV at
dc, 50Hz and 60Hz (lOOn unbalance, LO driven). MAXIMUM ALLOWABLE INPUT: 350V peak, 250V
*Ins.
IEEE-488 BUS IMPLEMENTATION
Multiline Commands: DCL, LLO, SDC, GET, GTL,
UNT, UNL, SI’E, SI’D. Uniline Commands: IFC, REN, EOI, SRQ, ATN. Interface Functions: SHl, AHl, TEO, LA, LEO, SRl, RLZ,
PPO, DCl, DTl, CO, El. I’rogrammable Parameters: Range, Function, Zero, Inte-
gration Period, Filter, dB Reference, EOI, Trigger, Ter-
minator, Delay, 500-rd P Storage, Scaling, Calibration,
Display, Multiplex Of, Status, Service Request, Self
Test, Output Format.
GENERAL
DISPLAY: 14, 0.5.in. alphanumeric LED digits \vith
decimal point and polarity. Function and IEEE bus
status also displayed. RANGING: Manual or fast autoranging. ISOLATION: Input LO to IEEE LO or power line
ground: 500V max., 5xlOSV.Hz; greater than ~10%
paralleled by 400pF. DATA MEMORY: I to 500 locations, progrumwble,
Measurement intervals selectablr lms to 999.999x-c or
triggered. BENCH READING RATE: 5 readingswr, vwpt 2O~lIl;
ZOOMR range 2 rcadingsisec.
ZERO: Control subtracts on-scale value from subsequent
readings or allows value to bc programmed. FILTER: Weighted average (exponential). Programmable
weighting: 1 to & WARMUP: 1 hour to rated accuracy.
OPERATING ENVIRONMENT: [I’,-50 C’, 0% to 809,
relative humidity up to 35’C; linearly derate 3%
RH/“C, 35°C to 50°C (except 2OOMII range: 0% TV, M1”,>
RH up to 28°C). STORAGE ENVIRONMENT: -25~~ to +65-C. POWER: 105.725V or 210.250V (internal switch selected),
50Hz or 6OHz, 40VA maximum 90.I~ItlV & 180.22O\’
versions available upon request.
CONNECTORS: Analog: Switch selectable front (lr rear,
safety input jacks. Digital: TRIGGER input and \‘(>I-.I-
METER COMPLETE output on rear panel, BSCs. DIMENSIONS, WEIGHT: 89mm high x 43Hmm wide
x 441mm deep (3% in. x ~17% in. x 171% in.). Set
weight 33kg (15 Ibs.). ACCESSORlES AVAILABLE:
Model 1600A: Hi h Voltage Probe
Model 1641: K&n Test Lead Set
Model 1651: 50.Ampere Shunt
Model 1681: Clip-On Test Lead Set
Model 1682A: RF Probe
Model 1685: Clamp-On Current Probe
Model 1751: General Purpose Test Leads
Model 1754: Universal Test Lead Kit
Model 1930: True RMS ACV Option
Model 1931: Current Option
Model 1938: Fixed Rack Mounting Kit
Model 1939: Slide Rack Mounting Kit
Model 7007-l: Shielded IEEE-488 Cable, lm
Model 7007-2: Shielded IEEE-488 Cable, 2m
Model 7008.3: IEEE-488 Cable, 3 ft. (0.9m)
Model 7008.6: IEEE-488 Cable, 6 ft. (1.8m)
Specifications subject to change without notice,
TABLE OF CONTENTS
SECTION l-GENERAL INFORMATION
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
INTRODUCTION FEATURES
..................................................................................
WARRANTY INFORMATION MANUAL ADDENDA SAFETY SYMBOLS AND TERMS SPECIFICATIONS INSPECTION USING THE MODEL 193A MANUAL, GETTING STARTED ACCESSORIES
............................................................................
.................................................................
.......................................................................
.............................................................
...........................................................................
...............................................................................
........................................................
.........................................................................
..............................................................................
SECTION 2-BASIC DMM OPERATIONS
2.1
2.2
22.1
2.2.2
2.2.3
2.2.4
2.3
2.3.1
2.3.2
2.3.3
2.3.4
2.4
2.4.1
2.4.2
2.4.3
2.5
2.6
2.6.1
2.6.2
2.6.3
2.6.4
2.6.5
2.6.6
2.6.7
2.6.8
2.6.9
2.6.10
2.6.11
2.6.12
2.6.13
2.7
2.7.1
2.7.2
2.8
2.8.1
2.82
INTRODUCTION ...........................................................................
POWER UPPROCEDURES ...................................................................
Line Power ................................................................................
rower-up Sequence ........................................................................
Factory Default Conditions User Programmed Conditions
FRONT PANEL FAMILIARIZATION
.................................................................
..............................................................
..........................................................
DisplayandIndicators .....................................................................
Controls..
................................................................................
Input Terminals ...........................................................................
Current Fuse ..............................................................................
REAR PANEL FAMILIARIZATION
Connectors and Terminals
.................................................................. 2-4
............................................................
Calibration Switch .........................................................................
Line Fuse .................................................................................
ERROR AND WARNING DISPLAY MESSAGES
BASICMEASUREMENTS ....................................................................
Warm Up Period
Zero ......................................................................................
.....................................................................................
Filter DC Voltage Measurements
..........................................................................
.................................................................
Low-Level Measurement Considerations., TRMS AC Voltage Measurements.
..........................................................
Resistance Measurements ..................................................................
Current Measurements (DC or TRMS AC) RTD Temperature Measurements dBMeasurements
......................................................................... 2.17
...........................................................
dB Measurement Considerations and Applications,, TRMS AC + DC Measurements TRMS Conslderatmns
DATA STORE
Storing Data Recalling Data
.......................................................................
.............................................................................. 2-22
............................................................................. 2-22
............................................................................
............................................................
FRONT PANEL PROGRAMS ................................................................
Cursor and Data Entry, ProgramTEMP
...........................................................................
...................................................................
............................................... 2-5
...................................................
..................................................
......................................... 2-19
1-l 1-l
I-1 l-l l-1
1-2 l-2
l-2
1-2 1-3
2-l 2-l 2-l 2-I 2-I
2.2 2-2
2-2
2.2 2-4 2-4 2-3
2-5 2-S
2-Y 2-Y
2-Y
2-11 2-12 Z-13
2.14 2-15
2-16
2.17
2-19
Z-20
2-23
2.25 2-25
2-25
i
2.8.3
2.8.4
2.8.5
2.8.6
2.8.7
2.8.8
2.8.9
2.8.10
2.8.11
2.8.12
2.8.13
2.8.14
2.8.15
2.9
2.10
2.10.1
2.10.2
2.10.3
Program AC + DC (Low Frequency AC). Program dB Program ZERO Program FILTER Program RESET Program4
Program 90 (Save) Program
Program 92 (Freq) Program 93 (Diagnostics). Program 94 (MX + B Status) Program 95 (Multiplexer)
Program 96
FRONT PANEL TRIGGERING EXTERNAL TRIGGERING
External Trigger Voltmeter Complete. Triggering Example
...........................
........................
.......................
.......................
............................
.....................
91 (IEEE Address)
.....................
(CAL)
.....................
.......................
...................
....................
.............
..............
...........
...............
...........
...............
SECTION 3-IEEE-488 PROGRAMMING
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
............................
............................
......
......
......
......
......
......
......
......
.............
.............
............. 2-26
............. Z-26
......
......
......
......
......
......
...... 2.29
...... 2.29
......
...... 2.29
......
......
......
......
......
......
...... 2-31
......
......
...... 2.32
2-25 Z-26
2-27 2-27
2.28
2-29 Z-30
2-30 2-30
2-31 2-31
3.1
3.2
3.3
3.3.1
3.3.2
3.3.3
3.4
3.4.1
3.4.2
3.4.3
3.4.4
3.4.5
3.5
3.6
3.6.1
3.6.2
3.6.3
3.7
3.7.1
3.7.2
3.7.3
3.8
3.8.1
3.8.2
3.8.3
3.8.4
3.9
3.9.1
3.9.2
3.9.3
3.9.4
3.9.5
3.9.6
3.9.7
3.9.8
3.10
INTRODUCTION BUS DESCRIPTION.. IEEE-488 BUS LINES
Data Lines Bus Management Lines Handshake Lines
BUS COMMANDS
Uniline Commands.. Universal Commands..
AddressedCommands Unaddressedcommands Device-Dependent Commands
COMMAND CODES COMMAND SEQUENCES
Addressed Command Sequence Universal Command Sequence
...........................................................................
....................................................................... 3-l
........................................................................
................................................................................
....................................................................
..........................................................................
..........................................................................
... . ..................................................................
....................................................................
.....................................................................
...................................................................
..............................................................
........................................................................
................................................................... 3-7
.............................................................
..............................................................
Device-Dependent Command Sequence .....................................................
HARDWARE CONSIDERATIONS
Typical Controlled Systems BusConnections
..........................................................................
Primary Address Programming.
SOFTWARE CONSIDERATIONS
Controller Handler Software Interface BASIC Programming Statements Interface Function Codes IEEE Command Groups
...................................................................
GENERAL BUS COMMAND PROGRAMMING.
.............................................................
.................................................................
............................................................
.............................................................
...............................................................
...................................................
..................................................................
..............................................
REN(RemoteEnable) .....................................................................
IFC(InterfaceClear) LLO(IocalLockout)
GTL (Go To Local) and Local DCL(DeviceClear)
.......................................................................
......................................................................
..............................................................
.......................................................................
SDC(Send Device Clear) ..................................................................
GET (Group Execute Trigger)
..............................................................
Serial Polling (SPE, SPD) ..................................................................
DEVICE-DEPENDENT COMMAND PROGRAMMING
.........................................
3-l
3-2 3-2
3-2 3-2 3-3 3-4 3-4 3-5
3-5 3-5 3-5
3-7 3-7 3-7 3-7 3-7
3-8
3.10
3.10 3-10
3.10 3-11
3.12 3-12
3.1~2
3.13
3.13
3.13 3-n
3-14 3-14
3.14 3-15
ii
3.10.1
3.10.2
3.10.3
3.10.4
3.10.5
3.10.6
3.10.7
3.10.8
3.10.9
3.10.10
3.10.11
3.10.12
3.10.13
3.10.14
3.10.15
3.10.16
3.10.17
3.10.18
3.10.19
3.10.20
3.10.21
3.10.22
3.10.23
3.11
3.11.1
3.11.2
3.12
3.13
3.x3.1
3.X3.2
3.13.3
3.w.4
3.x3.5
3.X3.6
3.w.7
3.13.8
3.x3.9
Execute(X) ...............................................................................
Function(F) ..............................................................................
Range(R) ................................................................................
Zero(Z) .................................................................................
Filter(P)............................................................................~
Rate(S) ..................................................................................
Trigger Mode (T) .........................................................................
Reading Mode(B) ........................................................................
Data Store Interval (Q) and Size (1) ........................................................
Value (V) and Calibration (C) Default Conditions (L)
Data Format(G) ..........................................................................
SRQ Mask (M) and Status Byte Format. ....................................................
EOI and Bus Hold-Off Modes (K) ..........................................................
Terminator (Y)
Status(U) ...............................................................................
Multiplex (A)
Delay(W) ...............................................................................
Self-Test(J)
Hit Button(H) ...........................................................................
Display(D) ..............................................................................
Reference Junction (0) ....................................................................
Exponential Filter (N)
FRONT PANEL MESSAGES .................................................................
Bus Error ................................................................................
Trigger Overrun Error .....................................................................
BUS DATA TRANSMISSION TIMES ......................................................... 3.37
TRANSLATOR SOFTWARE
Translator Format
Wild Card($)
NEW and OLD ..........................................................................
Combining Translator Words
Combining Translator Words with Keithley IEEE-488 Commands .............................. 3-40
Executing Translator Words and Keithley IEEE Commands SAVE LIST FORGET
.................................................................................... 3-41
..................................................................................... 3-41
............................................................................
.....................................................................
...............................................................................
............................................................................... 3-3’)
.................................................................................
....................................................................
.....................................................................
.........................................................................
..............................................................
... ~, 3.33
................................................................. 3-37
..............................................................
.........................
.........
3.19 3-1~ 3-19
3-20
.... 3.20
3.21
3.21
3.22
3.22
3.23
3.24 3-24
3.25
3.28
3.28
3.2’)
3.33 R-33 3-3-1
3-W 3-35
3.36
3.36
3-36
3.37
3.37 3-39
3-3’) 3-40
3.41
SECTION 4-PERFORMANCE VERIFICATION
4.1
4.2
4.3
4.4
4.5
4.5.1
4.5.2
4.5.3
4.5.4
4.55
4.5.6
INTRODUCTION ...........................................................................
ENVIRONMENTALCONDITIONS
INITIAL CONDITIONS.. ....................................................................
RECOMMENDED TEST EQUIPMENT.
VERIFICATION PROCEDURES ......................................................
DC Volts Verification.. .....................................................................
TRMS AC Volts Verification OHMS Verification DC Current Verification TRMS AC Current Verification
RTD Temperature Verification .............................................................
........................................................................
....................................................................
............................................................
........................................................
.................................................................
..............................................................
SECTION 5-THEORY OF OPERATION
5.1
INTRODUCTION ..,,...........,......,....,.,,.,...................................~....~. 5-l
4-l 1-1 4-l 4-I
........ 4-I
1-2 4-3 4-4 4-5
4-h
4-7
iii
5.2 OVERALL FUNCTIONAL DESCRIIJTION
5.3 ANALOG CIRCUITRY .......................................................................
5.3.1
5.3.2
5.3.3
5.3.4
5.4
5.5
5.6
5.6.1
5.6.2
5.7
Input Signal Conditioning Multiplexer
+Z.lV Reference Source
Input Buffer Amplifier
AIDCONVERTER ...........................................................................
CONTROLCIRCUITRY.. ...................................................................
DIGITAL CIRCUITRY .......................................................................
Microcomputer ...........................................................................
Display Circuitry
POWER SUPPLIES
............................................................................... 5-6
..........................................................................
.........................................................................
..................................................................
.................................................................... 5-8
.....................................................................
.....................................................
SECTION 6-MAINTENANCE
5-l 5-l
5-l
5-8
5-8 5-10 5-10 5-10 5-10 5-10
6.1
6.2
6.3
6.3.1 Line Fuse
6.3.2
6.4
6.5
6.6
6.6.1
6.6.2
6.6.3
6.6.4
6.6.5
6.6.6
6.6.7
6.68
6.6.9
6.6.10
6.6.11
6.6.12
6.7
6.8
6.9 TROUBLESHOOTING
6.9.1 Recommended Test Equipment
6.9.2 Power-Up Self Test .......................................................................
6.9.3 Program 93.Self Diagnostic Program.
6.9.4 Iowa Supplies ...........................................................................
6.9.5 Signal Conditioning Checks
6.9.6 Digital and Display Circuitry Checks
INTRODUCTION LINE VOLTAGE SELECTION FUSEREPLACEMENT
Current Fuse. MODEL 1930 TRMS ACV OPTION INSTALLATION MODEL 1931 CURRENT OPTION INSTALLATION
CALIBRATION
Recommended Calibration Equipment
Environmental Conditions
Warm-Up Period ..........................................................................
Calibration Switch ......................................................................... 6-5
Front Panel Calibration. IEEE-488 Bus Calibration
Calibration Sequence .......................................................................
DC Volts Calibration Resistance Calibration TRMS AC Volts Calibration DC Current Calibration.
TRMS AC Current Calibration
DISASSEMBLYINSTRUCTIONS .............................................................
SPECIAL HANDLING OF STATIC-SENSITIVE DEVICES
...........................................................................
.................................................................
.......................................................................
.................................................................................
.............................................................................
............................................
............................................
..............................................................................
.......................................................
..................................................................
....................................................................
...................................................................
....................................................................... 6-7
...................................................................... 6-8
................................................................
...................................................................
............................................................. 6-14
...................................... 6-21
...................................................................... 6-21
............................................................ 6-22
.......................................................
...............................................................
.......................................................
6-l 6-l 6-l
6-l 6-2 6-2 6-3 6-3 6-3 6-3 6-3
6-5 6-6
6-7
6-10 6-13
6-15
6-22 6-23 6-23
6.23
6.23
SECTION 7-REPLACEABLE PARTS
7.1 INTRODUCTION ............................................................................ 7-l
7.2
7.3 ORDERING INFORMATION ................................................................. 7-l
7.4 FACTORY SERVICE.. ........................................................................ 7-l
7.5
iv
PARTS LIST ................................................................................. 7-l
SCHEMATIC DIAGRAMS AND COMPONENT LOCATION DRAWINGS ......................... 7-l
LIST OF ILLUSTRATIONS
2-l 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 Z-10 2-11
3-l 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-10 3-11 3-12 3-13
4-l 4-2 4-3 4-4 4-5 4-6 4-7 4-8
Model193A Front Panel Model193ARearPanel
DC Voltage Measurements TRMS AC Voltage Measurements Two-Terminal Resistance Measurements Four-Terminal Resistance Measurements.
Current Measurements..
RTD Temperature Measurements.. ........................................................... 2-17
External Pulse Specifications Meter Complete Pulse Specifications External Triggering Example..
IEEE-488 Bus Configuration IEEE-488 Handshake Sequence
Command Groups
System Types Typical Bus Connections
IEEE-488 Connections.
Model 193A IEEE-488 Connector ........
Contact Assignments.
General Data Format
SRQ Mask and Status Byte Format .....
UO Status Word and Default Values Ul Error Status Word. Hit Button Command Numbers
Connections for DC Volts Verlflcation .......................................................... 4-2
Connections for TRMS AC Volts Verification Maximum Input TRMS AC Voltage
Connections for Ohms Verification (200%20kn) Range
Connections for Ohms Verifications (200k0-200MR Ranges).
Connections for DC Current Verification.
Connections for TRMS AC Current Verification
Connections for RTD Temperature Verification. .................................................
........................... ...... ...... 3-8
......................................................................
.......................................................................
...................................................................
............................................................
.......................................................
......................................................
....................................................................
................................................................
.........................................................
...............................................................
.............. ...... ...... ......
...........
......................
.................
...................
...................
................... ...... ...... ...... 3-26
...... ...... ...... ...... 3-30
..................
......... ...... ...... 3-34
............................................................
...... 3-3
...... ...... 3-6
...... 3-x
......
......
...... ...... 3-10
...... ......
...... ......
...................................................
..........................................
.....................................
......................................................
.................................................
...... 3-26
2-7 2-7
2.13
2.14 2-16 2-16 2-16
2-31 2-32 2-33
3-2
3-Y 3-Y
3-32
4-3 4-4 4-4 4-S 4-h
1-7
-1-7
5-l
5-2 Input Configuration During 2 and 4-Terminal Resistance Measurements
5-3 Ohms Source (2k12 Range Shown)
s-4
5-5
5-6 5-7
6-l 6-2 6-3 6-4 6-5 6-6 6-7 6-8 6-9 6-10
Overall Block Diagram ...............................................
....................................
Resistance Measurement Simplified Circuitry
JFET Multiplexer
Multiplexerl’hases .................................................
A/D Converter Simplified Schematic Models 1930 and 1931 Installation.
DC Volts Calibration Configuration (2OOmV and 2V Ranges) DC Volts Calibration Configuration (ZOV-1OOOV Ranges), Four-Wire Resistance Configuration (200%20kO Ranges) Two-Wire Resistance Calibration Configuration (200k0-ZOOM0 Ranges). Flowchart of AC Volts Calibration Procedure TRMS AC Volts Calibration Adjustments. TRMS AC Volts Calibration Configuration DC Current Calibration Configuration TRMS AC Current Calibration Configuration
....................................................
..................................
...........................................................
........................................................
..........................
.....................................
.........................................
.........................................
..................................................
.....................................................
.....................................................
..................................................
...........................
5.2
5.3 i-4 5-j S-6
j-7
5-9 h-4
b-7 6-S 6-Y
6-9
6-11
6-12
h-l3
6.14 h-15
6-11 6-12
Co”“ectors.....................................................
........................... 6-17
Model 193A Exploded View. ........................... 6-19
7-1 Jumper Board, Component Location Drawing, Dwg. No. 193.160. 7-2 Display Board, Component Location Drawing, Dwg. No. 193.110 7-3 7-4 7-5 7-6 7-7 7-8
Display Board, Schematic Diagram, Dwg. No. 193.116 Digital Board, Component Location Drawing, Dwg. No. 193.100 Digital Board, Schematic Diagram, Dwg. No. 193-106 Analog Board, Component Location Drawing, Dwg. No. 193.120 Analog Board, Schematic Diagram, Dwg. No. 193.126
Model 1,930, Component Location Drawing, Dwg. No. 1930.100 7-9 Model 1930, Schematic Diagram, Dwg. No. 1930.106 710 Model 1931, Component Location Drawing, Dwg. No. 1931-100 7-11 Model 1931, Schematic Location Drawing, Dwg. No. 1931.106
......
......
......
......
......
......
......
......
......
......
......
..............
..............
..............
..............
7-3 7-4 7-5
7-10
.............. 7-11
............. 7.22
.............
7-23
............. 7.34
.............
7-35
............. 7-38
............. 7-39
vi
LIST OF TABLES
2-l
2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9
3-l 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-10 3-11 3-12 3-n
3-14 3-15 3-16 3-17 3-18
Factory Default Conditions
Front PanelPrograms ........................................................................
Error and Warning Messages .................................................................
Internal Filter ...............................................................................
Resistance Ranges ..........................................................................
Corresponding Voltage Reference Levels for Impedance References Comparison of Average and TRMS Meter Readings. High Speed Data Store
Example MX + B Readings .................................................................
IEEE-488 Bus Command Summary.
Hexadecimal and Decimal Command Codes Typical Addressed Command Sequence Typical Device-Dependent Command Sequence
IEEE Contact Designation
BASIC Statements Necessary to Send Bus Commands
Model 193A Interface Function Codes
IEEE Command Groups.
General Bus Commands and Associated BASIC Statements
Factory Default Conditions
Device-Dependent Command Summary
Range Command Summary
High Speed Data Store
SRQ Command Parameters
Bus Hold-Off Times ........................................................................
Trigger to Reading-Ready Times (DCV Function)
Translator Reserved Words and Character.
Translator Error Messages .................................................................... 3-38
................................................................... 2-l
2.12
2.15
..............................
...........................................
.....................................................................
..........................................................
................................................... 3-5
...................................... ................ 3-7
................................................ 3-7
....................................................................
......................................... 3-11
........................................................ 3.11
.................................................................... 3.12
.................................... 3-12
..................................................................
...................................................... 3-16
.................................................................
....................................................................
..................................................................
..............................................
....................................................
2.19
2-21
2.22 Z-28
3-14
3-20 2-23 3-27 3-28 3-37 3-38
2-4 2-9
3-4
3-9
4-l 4-2 4-3 4-4 Limits for Ohms Verification 4-5 4-6 4-7
5-l
6-l 6-2 6-3 6-4 6-5 6-6 6-7 6-8 6-9 6-10 6-11 6-12 6-13
Recommeded Test Equipment ..............
Limits for DC Volts Verification .............
Limits for TRMS AC Volts Verification. ......
Limits for DC Current Verification Limits for AC Current Verification
Limits for RTD Temperature Verification ....
Ranging Information .........................................................................
Line Voltage Selection ........................................................................
Line Fuse Replacement Current Fuse Replacement Recommended Calibration Equipment
DC,Volts Calibratiqn .........................................................................
Resistance Calibration .......................................................................
TRMS AC Volts (Model 1930) Calibration DC Current Cahbratmn TRMS AC Current Calibration
Static-Sensltwe Devzes
Recommended Troubleshooting Mode Model 193A Troubleshooting Mode
Power Supply Checks
.......................................................................
......................................................................
.....................................................................
......................................................................
......
......
...............
..........
..........
...................................................................
.........................................................
..................................................... 6-13
...............................................................
........................................................
..........................................................
......
......
......
......
......
......
......
......
......
5.3 6-l
6-2
6-2
6; 6-10 6-14
6-15
6.21 6;:
h-28
vii
6.14 6-15
Digital Circuitry Checks Display Circuitry Checks
...............................................................
..............................................................
6-28 6-29
7-l 7-2 7-3 7-4 7-5 7-6
Display Board, Parts List Digital Board, Parts List Analog Board, Parts List,, Model 1930, Parts List Model 1931, Parts List.
..............................................................
..............................................................
............................................................
.................................................................
................................................................
Model 193A Mechanical Parts List
......................................................
......
......
7-2
7-7
715
7-31
7-37
7-41
vttt
SECTION 1
GENERAL INFORMATION
1.1 INTRODUCTION
The Keithley Model 193A System DMM, with the TRMS AC Volt and Current options installed, is a six function auto­ranging digital multimeter. At 6% digit resolution, the LED display can display ~2,200,OOO counts. The range of this analog-to-digital (A/D) converter is greater than the nor­mal *1,999,999 count AID converter used in many 6% digit DMMs. The built-in IEEE-488 interface makes the instru­ment fully programmable over the IEEE-488 bus. With the
TRMS ACV option and the Current option installed, the
Model 193A can make the following basic measurements:
1. DC voltage measurements from 1OOnV to lOOOV.
2. Resistance measurements from 1OOpfl to ZOOMR.
3. RTD temperature measurements from -1OOOC to 63O’C.
4. TRMS AC voltage measurements from I$/ to 700V.
5. DC current measurements from 1nA to 2A.
6. TRMS AC current measurements from 1nA to 2A.
In addition to the above mentioned measurement capa­bilities, the Model 193A can make dB and TRMS AC + DC measurements.
1.2 FEATURES
Some important Model 193A features include:
the power-up default conditions.
l Translator Software-User defined words can be used to
replace standard command strings over the IEEE488 bus.
l Thermocouple (TC) temperature measurements con be
made over the IEEE-488 bus.
1.3 WARRANTY INFORMATION
Warranty information may be found on the inside front cover of this manual. Should it become necessary to exer­cise the warranty, contact your Keithley representative or the factory to determine the pmper course of action. Keithley Instruments maintains service facilities in the United States, United Kingdom and throughout Europe. Information concerning the application, operation or ser. vice of your instrument may be directed to the applications engineer at any of these locations. Check the inside front cover for addresses.
1.4 MANUAL ADDENDA
Information concerning improvements or changes to the
instrument which occur after the printing of this manual
will be found on an addendum sheet included with this
manual. Be sure to review these changes before attemp­ting to operate or service the instrument.
l 14 Character Alphanumeric Display-Easy to read 14 seg-
ment LEDs used for readings and front panel messages.
l High Speed Measurement Rate-1000 readings per
second.
l Zero-Used to cancel offsets or establish baselines. A zero
value can be rogrammed from the front panel or over the IEEE-488 & us.
l Filter-The weighted average digital filter can be set for
1 to 99 readings from the front panel or over the bus.
l Data Store-An internal buffer that can store up to 500
readings is accessible from either the front panel or over the bus.
l Digital Calibration-The instrument may be digitally
calibrated from either the front panel or over the bus.
l User Programmable Default Conditions-Any instrument
measurement configuration can be established as
1.5 SAFETY SYMBOLS AND TERMS
The following safety symbols and terms are used in this
manual or found on the Model 193A.
The A
should refer to the operating instructions in this manual. The M
1OOW or more may be present on the terminal(s). Stand­ard safety practices should be observed when such
dangerous levels are encountered. The WARNING used in this manual explains dangers that
could result in personal injury or death. The CAUTION used in this manual explains haz.ards that
could damage the instrument.
symbol on the instrument denotes that the user
on the Instrument denotes that a potential of
l-l
GENERAL INFORMATION
1.6 SPECIFICATIONS
Detailed Model 193A specifications may be found preceding the Table of Contents of
this
manual.
1.7 INSPECTION
The Model 193A System DMM was carefully inspected, both electrically and mechanically before shipment. After unpacking all items from the shipping carton, check for any obvious signs of physical damage that may have occurred
during transit. Report any damage to Retain and use the original packing materials in case reship­ment is necessary. The following items are shipped with every Model 193A order:
Model 193A System DMM
Model 193A Instruction Manual Additional accessories as ordered.
If an additional instruction manual is required, order the manual package (Keithley Part Number 193-901-00). The manual package includes an instruction manual and any applicable addenda.
the
shipping agent.
l Section 6 contains information for servicing the instru-
ment, This section includes information on fuse replace­ment, line voltage selection, calibration and troubleshooting.
l Section 7 contains replaceable parts information.
1.9 GETTING STARTED
The Model 193A System DMM is a highly sophisticated in-
strument
and running quickly use the following procedure. For com­plete information on operating the Model 193A consult the appropriate section of this manual.
POWW-Up
1. Plug the line cord into the rear panel power jack and plug grounded power source. See paragraph 2.2.1 for more complete information.
2. Press in the POWER switch to apply power to the in­strument. The range.
with many capabilities. To get the instrument up
the
other end of the cord into an appropriate,
instrument
will power up to the 1OOOVDC
1.6 USING THE MODEL 193A MANUAL
This manual contains information necessary for operating and servicing the Model 193A System DMM, TRMS ACV option and the Current option. The information is divided into the following sections:
l Section 1 contains general information about the Model
193A including that necessary to inspect the instrument and get it operating as quickly as possible.
l Section 2 contains detailed operating information on us-
ing the front panel controls and programs, making con­nections and basic measuring techniques for each of the available measuring functions.
l Section 3 contains the information necessary to connect
the Model 193A to the IEEE-488 bus and program operating modes and functions from a controller.
l Section 4 contains performance verification procedures
for
the
instrument. This information will be helpful if you wish to verify that the instrument is operating in com­pliance with its stated specifications.
l Section 5 contains a description of operating theory.
Analog, digital, power supply, and IEEE-488 interface operation is included.
Making Measurements
1. Connect the supplied safety shrouded test leads to the front panel VOLTS HI and LO input terminals. Make sure the INPUT switch is in the out (FRONT) position.
2. To make a voltage measurement, simply connect the in­put leads to a DC voltage surge (up to 1OOOV) and take the reading from the display.
3. To change to a different measuring function, simply press the desired function button. For example, to measure resistance, press the OHMS button.
Using Data
Store
Storing Data:
1. Press the DATA STORE button. The DATA STORE in­dicator will turn on and a storage rate (in seconds) will be displayed.
2. Select an interval, other than 000.000, using the 4 and w, and data buttons.
3. Press the ENTER button. The buffer size will be displayed. Size 000 indicates that data will overwrite after 500 readings have been stored.
4. If a different buffer size is desired, enter the value us­ing the number buttons (0 through 9).
5. Press the ENTER button to start the storage process.
1-2
-
GENERAL INFORMATION
The data store mode can be exited at any time before the start of the storage process by pressing the RESET button. Once storage has commenced, the storage process can be stopped by pressing any function button. See paragraph
2.7.1 for complete information on storing data.
Recalling Data:
1. Press the RECALL button. The buffer location of last stored reading will be displayed.
2. To read the data at a different manor location, enter the value using the number buttons O-9).
3. Press the ENTER button. The reading and the memory location will be displayed.
4. The A and 7 buttons can be used to read the data in all filled memory locations.
5. To read the highest, lowest and average reading stored in the buffer, press the number 1,2 and 3 buttons respec­tively. Note that the memory location of the highest and lowest reading is also displayed. The average reading is displayed along with the number of readings averaged.
The recall mode can be exited by pressing the RESET but-
ton. See paragraph 2.7.2 for complete information on data recall.
Y
Paragraph 2.8 provides the detailed information for using the front panel programs.
1.10 ACCESSORIES
The following accessories are available to enhance the Model 193As, capabilities.
Model 1301 Temperature Probe-The Model Uol is a rug­ged low cost temperature probe designed to allo\v temper­ature measureme”ts from -55 to 150°C.
Model 16008 High Voltage Probe-The Model 1booB extends DMM measurements to 40kV.
Model 1641 Kelvin Test Lead Set-The Model 1641 has special clip leads that allow 4-terminal measurements to be made while making “nly two c”nnectims.
Model 165150.Ampere Current Shunt-The Model 1651 is
an external O.OOlR *l% 4.terminal shunt, which permits
current measurements from 0 to 50A AC or DC.
Model 1681 Clip-On Test Lead Set-The Model 1681 con­tains two leads, 1.2m (48’) long terminated with banana plugs and spring action clip probes.
Using Front Panel Programs Program selection is accomplished by pressing the PRGM
button followed by the button(s) that corresponds to the program number or name. For example, to select Program 91 (IEEE status), press the PRGM button and then 9 and
1 buttons. Table 2-2 lists and briefly describes the available front panel programs. Once a following general rules will
1. A displayed program condition can be entered by press­ing the ENTER button.
2. Program conditions that prompt the user with a flashin digit can be modified usmg the data buttons (0 throug
9 and +) and the cursor control buttons (manual range buttons).
3. Programs that contain alternate conditions can be dis K layed by ressin one of the manual range buttons. Eat press o between the two available conditions.
4. A program will be executed when the pressed ENTER button causes the instrument to exit the program mode.
5. A pro ram can be exited at an time and thus not ex­ecute 1 , by pressing the RESE
P P
one o these buttons toggles the display
rogram is selected the
P
app y:
if
button.
a
Model 1682A RF Probe-The Model 1682A permits voltage measurements from 1OOkHz to 250MHz. AC to DC transfer accuracy is +ldB from 1OOkHr to 250MHz at IV, peak responding, calibrated in RMS of a sine wave.
Model 1685 Clamp-On AC Probe-The .Model 1685 measures AC current by clamping on to a single conduc­tor. Interruption of the circuit is unnecessary. The Model
1685 detects currents by sensing the changing magnetic field
produced by the current flow. Model 1751 Safety Test Leads-Finger guards and shroud-
ed banana plugs help minimize the chance “1 making con-
tact with live circuitry. Model 1754 Universal Test Lead Kit-The Model 1754 is rl
12 piece test lead kit, with interchangeable plug-in ac­cessories. Included in the kit is one set of test leads (l-red,
l-black), two spade lugs, two standard banana plugs, two phone tips (0.06 DIA.), two hooks and miniature alligator clips (with boots).
Model 1930 TRMS AC Volt 0 True Root Mean Square (TRM
Model 193A. This option allows the instrument to measure the TRMS value of an AC signal. When the Model 1Y30
tion-The Model 1930 is d
! ) AC plug-in option for the
1-3
is installed, AC + DC voltage measurements can be made. Field installation or removal/replacement of the Model 1930 will require recalibration of the Model 193A and the Model
1930. Model 1931 Current 0
current option for the R
tion-The Model 1931 is a plug-in
ode1 193A. This option allows the instrument to measure DC current up to 2A. When both Models 1930 and 1931 are installed, the instrument can make TRMS AC current measurements and TRMS AC + DC current measurements. Field installation requires recalibration of the Model 193A.
Model 7007 IEEE488 Shielded Cables-The Model 7007 con­nects the Model lY3A to the IEEE-488 bus using shielded cables to reduce electromagnetic interference (EMI). The Model 7007-l is one meter in length and has a EMI shield­ed IEEE-488 connector at each end. The Model 7007-Z is identical to the Model 70071, but is two meters in length.
Model 7008 IEEE-488 Cables-The Model 7008 connects the Model 193A to the IEEE-488 bus. The Model 7008-3 is 0.9m (3 ft.) in length and has a standard IEEE-488 connector at each end. The Model 7008-6 cable is identical to the Model
7008.3, but is 1.8m (6 ft.) in length.
Model 1938 Fixed Rack Mount-The Model 1938 is a sta­tionary mount kit that allows the Model 193A to be mounted in a standard 19 inch rack.
Model 1939 Slide Rack Mount-The Model lY39 is a sliding mount kit that allows the Model 193A to be rack mounted with the added feature of sliding the instrument forward for easy access to the rear panel and top cover.
Model 8573 IEEE-488 Interface-The Model 8573 is an IEEE-488 standard interface designed to interface the IBM PC or XT computers to Keithley instrumentation over the IEEE-488 bus. The interface system contains two distinc­tive parts: an interface board containing logic to perform the necessary hardware functions and the handler software (supplied on disk) to perform the required control func­tions. These two important facets of the Model 8573 join together to give the IBM advanced capabilities over IEEE-488 interfaceable instrumentation.
1-4
SECTION 2
BASIC DMM OPERATION
2.1 INTRODUCTION
Operation of the Model 193A may be divided into two The instrument can be turned on by pressing in the front general categories: front panel operation and IEEE-488 bus panel POWER switch. The switch will be at the inner most operation. This section contains information necessary to
use the instrument from the front panel. These functions can also be programmed over the IEEE-488 bus, as describ­ed in Section 3.
2.2 POWER UP PROCEDURE
2.2.1 Line Power
Use the following procedure to connect the Model 193A to line power and power up the instrument.
1. Check that the instrument is set to correspond to the available line oower. When the instrument leaves the fac­tory, the internally selected line voltage is marked on the rear panel near the AC power receptacle. Ranges are 105Vl25V or ZloV-250V 50160Hz AC. If the line voltage setting of the instrument needs to be changed, refer to Section 6, paragraph 6.2 for the procedure. If the line frequency setting of the instrument needs to be checked and/or changed, utilize front panel Program 92 (see paragraph 2.8.11) after the instrument completes the power-up sequence.
2. Connect the female end of the power cord to the AC receptacle on the rear panel of the instrument. Connect the other end of the cord to a grounded AC outlet.
WARNING
The Model 193A is equipped with a 3-win? power cord that contains a separate ground wire and is designed to be used with grounded outlets. When proper connections are made, instrument chassis is connected to power line ground. Failure to use a grounded outlet may result in personal injury or death because of electric
2.2.2 Power-Up Sequence
position when the instrument is turned on. Upon P ower­up, the instrument will do a number of tests on itse f. Tests
are performed on memory (ROM, RAM and NVRAM). If RAM or ROM fails, the instrument will lock up. If EYROM FAILS, the message “UNCALIBRATED” will be See paragraph 6.92 for a complete description of the pw~er­up self test and recommendations to resolve failures.
2.2.3 Factory
At the factory, the Model 193A is set up so that the front panel controls and features are initially configured to cer­tain conditions on power-up and when program RESET is run. These are known as the factory default conditions and
are Iisted in Table 2.l,
Default Conditions
displayed.
Table 2-1. Factory Default Conditions
Function Range Resolution Line Frequency IEEE Address RTD Alpha Value and scale ZW” Zero Value dR dB Reference Value AC + DC Data Store Recall Filter Filter Value
DCV
1ooov
6% Digits
1 f
0.00392LC Disabled
+ooooooo
Disabled
IV, 1mA Disabled Disabled Disabled Disabled
10
shock.
‘Program 90 (save) can be used to establish the factory
CAUTION
Be sure that the power line voltage agrees with the indicated range on the rear panel of the in­strument. Failure to
ObSSNS
this
pmCautiOfl
may
default condition. However, an “UNCALIBRATED” error will set factory default to an IEEE address of 10 and a ire­quency setting of 60Hz.
result in instrument damage.
2-1
2.2.4 User Programmed Conditions
A unique feature of the Model 193A is that each function
“remembers” the last measurement configuration that it
was set up for (such as range, zero value, filter value, etc). Switching back and forth between functions will not affect the unique configuration of each function. However, the instrument will “forget” the configurations on power-down.
Certain confi urations can be saved by utilizing front panel Program 90. % n power-up, these user saved default condi-
tions will prevail over the factory default conditions. Also,
a DCL or SDC asserted over the IEEE-488 bus will set the instrument to the user saved default conditions. For more information, see paragraph 2.8.9 (Program 90).
NOTE Keep in mind that power-up default conditions can be either factory default conditions or user saved default conditions.
2.3 FRONT PANEL FAMILIARIZATION
The front panel layout of the Model 193A is shown in Figure
2-l. The following paragraphs describe the various com­ponents of the front panel in detail.
POWER-The POWER switch controls AC power to the in-
strument. Depressing and releasing the switch once turns
the power on. Depressing and releasing the switch a se­cond time turns the power off. The correct positions for on and off are marked on the front panel by the POWER switch.
INPUT-The INPUT switch connects the instrument to
either the front panel input terminals or the rear panel in-
put terminals. This switch operates in same manner as the
power switch. The front panel input terminals are selected
when the switch is in the “out” position and the rear panel
input terminals are selected when the switch is in the “in”
position.
FUNCTION GROUP-The FUNCTION buttons are used
to select the primaly measurement functions of the instru-
ment. These buttons also have secondary functions.
DCV-The DCV button places the instrument in the DC
volts measurement mode. The secondary function of this
button is to enter the number 0. See paragraph 2.6.4 for
DCV measurements.
Am-With the ACV option installed, the ACV button places
the instrument in the AC volts measurement mode. The
secondary function of this button is to enter the number
1. See paragraph 2.6.6 for ACV measurements.
2.3.1 Display and Indicators
Display-The 14 character, alphanumeric, LED display is used to display numeric conversion data, range and func­tion mnemonics (i.e. mV) and messages.
Status Indicators-These three indicators apply to instru­ment operation over the IEEE-488 bus. The REMOTE indi-
cator shows when the instrument is in the IEEE-488 remote
state. The TALK and LISTEN indicators show when the in-
strument is in the talk and listen states respectively. See
Section 3 for detailed information on operation over the
bus.
2.3.2 Controls
All front panel controls, except the POWER and INPUT switches, are momentary contact switches. Indicators are located above certain feature buttons to show that they are enabled. Included are AUTO (autorange), ZERO, FILTER, RECALL and DATA STORE. Some buttons have secondary functions that are associated with front panel program operation. See paragraph 2-8 for detailed Information on front panel programs.
OHMS-The OHMS button places the instrument in the
ohms measurement mode. The secondary function of this
button is to enter the number 2. See paragraph 2.6.7 for
resistance measurements.
ACA-With the ACV option and current option installed,
the ACA button places the instrument in the AC amps
measurement mode. The secondary function of this but-
ton is to enter the number 3. See paragraph 26.8 for ACA
measurements.
DCA-With the current option installed, the DCA button
places the instrument in the DC amps measurement mode. The secondary function of this button is to enter the
number 4. See paragraph 2.6.8 for DCA measurements. TEMP-The TEMP button places the instrument in the RTD
temperature measurement mode. The secondary functions
of this button are to select the TEMP program (select alter-
nate alpha value and thermometric scale) and to enter the
number 5. See paragraph 2.6.9 for RTD temperature
measurements.
RANGE GROUP-The Aand vbuttons are used for manual ran
ing. These %
ing and the AUTO button is used for autorang-
uttons also have secondary functiolw.
2-2
OPERATION
Manual-Each time the A button is pressed, the instru­ment will move up one range, while the v button will move the instrument down one range each time it is pressed. Pressing either of these buttons will cancel autorange, if it was previous selected. The secondary functions of these buttons are associated with front panel program operation.
AUTO-The AUTO button places the instrument in the autorange mode and turns on the AUTO indicator. While in this mode, the instrument will go to the best range to measure the applied signal. Autoranging is available for all
functions and ranges. Autoranging may be cancelled by pressin f the AUTO button or one of the manual range but­tons. T e secondary function of this button is to enter the
* sign.
MODIFIER GROUP-The MODIFIER buttons activate features that are used to enhance the measurement ca
bilities of the Model 193A. These features in effect ma Ify
a-
1 the selected function. In addition to their primary tasks, these buttons have secondary functions.
ZERO-The ZERO button turns on the ZERO indicator and causes the displa ed reading to be subtracted from subse­quent readings. -F, his feature allows for zero correction or storage of baseline values. The secondary function of this button is to select the ZERO program. Refer to paragraph
2.6.2 for detailed information on the zero feature. dB-The dB button places the instrument in the dB
measurement mode and may be used with the ACV and ACA functions. Under factory default conditions, measure­ments are referenced to 1V or ImA. However, the dB pro­gram may be used to change the reference level. The secon­dary function of this button is to select the dB program. See paragraph 2.6.10 for dB measurements.
FILTER-The FILTER button turns on the FILTER indicator and causes the instrument to start weighted averaging a number (l-99) of readings. The factory default value is 10, but may be changed using the FILTER program (see paragraph 2.8.6). See paragraph 2.6.3 for filter operation. Selecting the FILTER tions of this button. TK
rogram is one of the secondary func-
e other secondary function is to enter
the number 6. AC + DC-With the appropriate options installed, the AC
+ DC button (with ACV selected) places the instrument in the AC + DC measurement mode. With the ACV o installed, VAC + DC measurements can be made.
tion
%ith both the ACV and current option installed, AAC + DC measurements can be made. See paragraph 2.6.12 for AC
+ DC measurements. The secondary functions of this but­ton are to select the AC + DC program (low frequency TRMS measurements) and to enter the number 7.
CONTROL GROUP-The CONTROL buttons are features that allow for the control and manipulation of various aspects of instrument operation. All of these buttons, ex­cept PRGM, have a secondary function.
RESOLN-The RESOLN button allows for the selection of the number of digits of display resolution. Each press of the RESOLN button increases resolution by one digit. Pressing the RESOLN button after the maximum resolu-
tion is reached will revert the display back to the lowest resolution. Dis can be selecte B
lay resolution of 3%, 4’12, 5% or 6% digits
for DCV and ACV. Display resolution of 4% or 5’/2 digits can be selected for DCA and ACA. On OHMS, Y/z, 4% 5% and 6% digit resolution is available on the ZOOR through 200kfl ranges. On the 2MQ and 20MR
ranges, 5% and 6% digits can be selected. On the 200MR ran e, only 5%d resolution is available. The RESOLN but­ton a as no effect on low frequency AC + DC (Program AC
+ DC), TEMP or dB measurements. The secondary func-
tion of this button is to enter the decimal point (.).
TRIGGER/ENTER-The TRIGGER/ENTER button is used
as a terminator for data entry when the instrument is in the front panel program mode and as a front panel trigger when the data store is active.
STATUS/RESET STATUS-Instrument status can be displayed when the in-
strument is in the normal measurement mode or logRing readings. When the STATUS button is first pressed, the following current instrument conditions can be displayed with the use of the A and v buttons:
Software revision level
IEEE address
Line frequency setting Multiplexer status (on/off) MX+B status (on/off) MX+B values
dB reference value
Filter value (OO=filter disabled) Zero status (on/off) Zero value
Pressing the STATUS button a second time takes the in-
strument out of the status mode.
RESET-The RESET button is used to reset the instrument
back to the previously entered parameter. Keyed in
parameters are only entered after the ENTER button is
pressed. If RESET is pressed with the last parameter of a
program displayed, the program will be exited and the in-
strument will return to the previous operating state. This
button aborts back to normal operation when it is in one
of the following modes:
2-3
1. The data store is prompting for parameters (interval or size).
2. The instrument is in the RECALL mode.
3. A front panel program has been selected (except Program AC+DC which is treated as a normal measurement func­tion (see STATUS).
Program RESET-Program RESET returns the instrument to the factory default conditions. See paragraph 2.8.7 for information on using this program.
DATA STORE-The DATA STORE button selects the 500 point data store mode of operation. Paragraph 2.7 contains a complete description of data store operation. The secon­dary function of this button is to enter the number 9.
2.3.3 Input Terminals
The input terminals are intended to be used with safety shrouded test leads to help minimize the possibility of con­tact with live circuits. Note that the terminals are duplicated
on the rear panel and that the INPUT switch determines which set of terminals is active.
VOLTS OHMS HI and LO-The VOLTS OHMS HI and LO
terminals are used for making DC volts, AC volts and two-
wire resistance measurements.
AMPS and LO-The AMPS and LO terminals are used for making DC current and AC current measurements.
RECALL-The RECALL button recalls and displays readings stored in the data store. Paragraph 2.7.2 provides a detailed procedure for recalling data. The secondary func­tion of this button is to enter the number 8.
I’RGM-The PRGM button places the instrument in the front panel program mode. Table 2-2 lists the available pro­grams. Paragraph 2.8 contains descriptions and detailed operating procedures for each front panel program.
LOCAL-When the instrument is in the IEEE-488 remote state (REMmE indicator on), the LOCAL button will return
the instrument to front panel operation. However, if local lockout (LLO) was asserted over the IEEE-488 bus, the LOCAL button will be inoperative. See Section 3 for infor­mation on operating the instrument over the IEEE-488 bus.
Table 2-2. Front Panel Programs
AC+DC
L
TEMP
dB
ZERO
FILTER RESET
4 90 91 92
93 94 95 96
Set RTD value and scale. Low Frequency TRMS AC + DC. Recall/modify dB reference value. Recall/modify zero value. Recall/modify number of readings aver­aged (filter value). Reset internal conditions to factorv default. MX + B select. save current front panel setup. Recall/modify IEEE address.
Recall/modify line frequency setting
(50160Hz). Self-test
Set values for MX + B program Multiplexer on/off. Digital calibration.
i
OHMS SENSE HI and LO-The OHMS SENSE HI and LO terminals are used with the VOLTS OHMS HI and Lx) ter-
minals to make four-wire resistance measurements and four-wire RTD temperature measurements.
2.3.4 Current Fuse
The current fuse protects the Model 1931 from input cur-
rent overloads. The instrument can handle UD to 2A con­tinuously or 2.2A for less than one minute. l&fer to para­graph 6.3.2 for the current fuse replacement procedures.
2.4 REAR PANEL FAMILIARIZATION
The rear panel of the Model 193A is shown in Figure 2-2.
2.4.1 Connectors and Terminals
AC Receptacle-Power is applied through the supplied power cord to the 3-terminal AC receptacle. Note that the selected supply voltage is marked on the rear panel near the connector.
Input Terminals-The rear panel input terminals perform the same functions as the front panel input terminals.
Paragraph 2.3.3 contains the description of the input
terminals.
IEEE-488 Connector-This connector is used to connect the
instrument to the IEEE-488 bus. IEEE interface functions
are marked above the connector. External Trigger Input-This BNC connector is used to ap-
ply pulses to trigger the Model 193A to take one or more readings, depending on the selected trigger mode.
2-4
OPERATION
Voltmeter Complete Output-This BNC output connector provides a pulse when the Model 193A has completed a reading. It is useful for triggering other instrumentation.
2.4.2 Calibration Switch
Calibration of the Model 193A can only be done if the calibration switch is in the unlock position.
2.4.3. Line Fuse
‘The line fuse provides protection for the AC power line
input. Refer to paragraph 6.3.1 for the tine fuse replacement procedure.
2.5 ERROR AND WARNING DISPLAY MESSAGES
Table 2-3 lists and explains associated with incorrect front panel operation of the 193A. Also included is a warning to the user that sent on the input terminals.
hazardous voltages (4OV or more) are pre-
the
various display messages
message that
indicates
Model
2-512-6
CONTROL
OPERATION
0
Figure 2-1. Model 193A Front Panel
Figure 2-2. Model 193A Rear Panel
2-112-8
OPERATION
Table 2-3. Error and Warning Messages set to voltage range if the minimum voltage
spacing is reduced.
.....-“J’ -..y.“..“.*-..
NEED 1930
Selected option not installed.
NEED 1931
NEED 1930.1931
“H”
NO PROGRAM
1.VERFLO KOHM Overrange-Decimal point position
I(
TRIG-OVERRUN Trigger received while still pro-
CONFLICT Trying to calibrate with instru-
<OT ACV or ACA Selecting AC+DC or dB with in-
SHORT-PERIOD Instrument, as currently con-
High Voltage: 4OV or more on input.
Invalid entry while trying to select program.
and mnemonics define function and range (Zkfl range shown).
The number of characters in the “OVERFLO” message defines the display resolution (6%d resolu­tion shown).
cessmg reading from last trigger. ment in an improper state. strument not presently in ACV or
ACA. figured, cannot run fast enough
to store readings at the selected interval.
2.6 BASIC MEASUREMENTS
The following paragraphs describe the basic procedures for making voltage, resistance, current, temperature, dB, and AC + DC measurements. An ACV option must be install­ed for ACV measurements, the current option must be in­stalled for DCA measurements and both options must be installed for ACA and AAC + DC measurements.
When making measurements in high energy circuits use test leads that meet the following requirements:
l Test leads should be fully insulated. l Only use test leads that can be connected to the circuit
(e.g. alligator clips, spade lugs, etc.) for hands-off measurements.
l Do not use test leads that decrease voltage spacing. This
diminishes arc protection and creates a hazardous condition.
Use the following sequence when testing power circuits:
1. De-energize the circuit using the regular installed connect-disconnect device such as the circuit breaker,
main switch, etc.
2. Attach the test leads to the circuit under test. Use ap­propriate safety rated leads for this application.
3. Set the DMM to the proper function and range.
4. Energize the circuit using the installed conncct­disconnect device and make measurements without disconnecting the DMM.
5. De-energize the circuit using the installed connect­disconnect device.
6. Disconnect the test leads from the circuit under lest.
WARNING
The maximum common-mode input voltage (the voltage between input LO and chassis ground) Is 500V peak. Exceeding this value may create a shock hazard.
2.6.1 Warm Up Period
High Energy Circuit Safety Precautions To optimize safety when measuring voltage in high energy
distribution circuits, read and use the directions in the following warning.
WARNING
Dangerous arcs of an explosive nature in a high energy circuit can cause severe personal injury or death. If the meter is connected to a high energy circuit when set to a current range, low
resistance range or any other low impedance
range, the circuit is virtually shorted. Danger­ous arcing can result even when the meter is
The Model 193A is usable immediately when it is first turn-
ed on. However, the instrument must be allowed to warm up for at least one hour to achieve rated accuracy.
2.6.2
The zero feature sewes as a means of baseline suppression
by allowing a stored offset value to be subtracted from subsequent readings. When the ZERO button is the instrument takes the currently displayed rea mg as a baseline value. All subsequent readings represent the dif­ferences between the applied signal level and the stored baseline.
Zero
rewed,
R
2-9
OPERATION
A baseline level can be established for any or all measure­ment functions and is remembered by each function. For example, a XIV baseline can be established on DCV, a 5V baseline can be established on ACV and a 1OkQ baseline can be established on OHMS. These levels will not be cancelled by switchin back and forth between functions. Once a baseline is esta hshed for a measurement function, %, that stored level will be the same regardless of what range the Model 193A is on. For example, if 1V is established as the baseline on the 2V range, then the baseline will also be 1V on the 2OV through 1OOOV ranges. A zero baseline level can be as large as full range.
NOTE
The following discussion on dynamic range is based on a display resolution of 6% digits. At 5%
digit resolution, the number of counts would be
reduced by a factor of 10. At 4%d resolution, counts
would be reduced by a factor of 100 and 3%d reso­lution would reduce counts by a factor of 1000.
By design, the dynamic measurement range of the Model 193A, at 6% digit resolution, is 4400000 counts (excluding the 1OOOVDC and 700VAC ran es). With zero disabled, the
displayed reading range of t e mstrument is +2200000 counts. With zero enabled, the Model 193A has the capabili­ty to display &IO0000 counts. This increased display range ensures that the dynamic measurement range of the instru­ment is not reduced when using a zero baseline value. The following two exam les will use the maximum allowable zero values (+2200 00 counts and -2200000 counts) to CT show that dynamic measurement range will not be re-
duced. It is important to note that the increased display range does not increase the maximum allowable input level to the instrument. For example, on the 2V range, the Model
193A will always overrange when more than +2.2V is con­nected to the input.
Example l-The instrument is set to the 2VDC range and
a maximum -2.2OOOOOV is established as the zero value. When -2.2OOOOOV is connected to the in ut of the Model
193A. the disolav will read O.OOOOOOV.
is connected &I tGe input, the display will read +4.4OOOOOV. Thus, the dynamic measurement range of the Model 193A
is OV to 4.4V, which is 4400000 counts.
Example Z-The instrument is still set to the 2VDC range, but a maximum +2.2OOOOOV is the zero level. When
+2.2OOOOOV is connected to the input of the Model 193A, the display will read O.OOOOOOV. When -2.2OOOOOV is con­nected to the input, the display will read -4.4COOOOV. Thus the d
namic measurement range of the instrument is -4.4V
9 to 0
, which is still 4400000 counts.
%
&en +2.2ooooov
Zen, Correction-The Model 193A must be properly zeroed when using the 2OOmV DC or the 2003 range in order to achieve rated accuracy specifications. To use ZERO for zero correction, perform the following steps:
1. Disable zero, if button. The Z
2.
Select the 2OOmV DC or the 2000 range.
3. Connect the test leads to the input of the Model 193A and short them together. If four-wire resistance measurements are to be made, connect and short all four leads together.
Note: At 5% and 6% digit resolution, low measurement techniques need to be employed. Use Kelvin test leads or shielded test leads. See paragraph 2.6.5 for low level
measurement considerations.
4. Press the ZERO button. The display will read zero.
5.
Remove the short and connect the test leads to the signal or resistance to be measured.
Note: Test lead resistance is also compensated for when zeroing the 2000 range with the above procedure.
Baseline Levels-Baseline values can be established by
either applying baseline levels to the instrument or by set­ting baseline values with the front panel ZERO program.
Paragraph 2.8.5 contains the complete procedure for using
the ZERO program. To establish a baseline level by apply­ing a level to the Model 193A, perform the following steps:
1. Disable zero, if presently enabled, by pressing the ZERO button. The ZERO indicator will turn off.
2. Select a function and range that is appropriate for the anticipated measurement.
3. Connect the desired baseline level to the in Model 193A and note that level on the P
4. Press the ZERO button. The display will zero and the ZERO indicator will be enabled. The previously
displayed reading will be the stored baseline. The zero baseline value will also be stored in Program ZERO, replacing the previous zero value.
resently enabled, by pressing the ZERO
IT
RO indicator will turn off.
ut of the
disp ay.
WARNING
If *4OV or more is present on the input ter­minals, the Model 193A will display the mnemonic “H” to indicate the presence of hazardous voltage. For example, the display “OO.OOOOHVDC” indicates than k4OV or more is present on the input.
5. Disconnect the stored signal from the input and connect the signal to be measured in its place. Subsequent readings will be the difference between the stored value and the applied signal.
2-10
Loading...
+ 152 hidden pages