JUMO TDA-300 and JUMO TDA-3000
Light is on during active
data recording
Shows the range
within which the measurement lies (over/underrange)
Measurement /
measurement point/
user number
Number of recorded
measurement data
Probe break or short-circuit
Battery charge status
Measured temperature
Message window (date, time, ...)
TDA-300 (Type 702540/...)
TDA-3000 (Type 702541/...)
Handheld Thermometer with data logger
Brief description
TDA-300 and TDA-3000 are handheld digital temperature indicators. They are used together with precision thermocouples or RTDs to measure temperatures on surfaces, in liquids, in melts, and in soft materials. Interchangeable temperature probes allow rapid adaptation to different applications. Both instruments feature a data logging function. On Type TDA-3000, the recorded measurement data can be read out via an USB interface. No special PC program is necessary for visualization, since the measurement data are available as ASCII files in CSV format (evaluation via spreadsheet programs). The plastic housing is proof against shock and breakage and resistant to corrosive substances. Thermometers, temperature probes and accessories are all available for delivery from stock.
Displays
Page 1/6Data Sheet 702540
Comparison of models
Data logger 99 measurements 9999 measurements TAG n u mbers
(meas. point designation) 5 (11 characters each) 99 (11 characters each) User 1 99 (11 characters each) Interfaces - USB, type Mini-B Enclosure protection IP67 IP54
2013-12-03/00447944
TDA-300 TDA-3000
Key features
k Measurement input for Pt100,
NiCr-Ni K, Fe-Con J and Cu-Con T
k DKD calibration certificate
k Data logger for 9999 measurements
k Simple readout via USB and
evaluation of data from a PC (TDA-3000 only)
k Limit monitoring
k Peak/bottom value acquisition
k Battery AA, Mignon LR6,
with a long service life
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, Germany Page 2/6
Technical data
Input
Measurement input RTD
Pt100
to EN 60751
Range limits
- resolution 1°C
- resolution 0.1°C Measurement offset -99.9 to +99.9°C Sampling rate 0.5 seconds Input filter 1st order digital filter; filter constant adjustable from 0 — 100 seconds Unit °C or °F
-200 to +850°C
-199.9 to +850.0°C
Thermocouple
NiCr-Ni K
to EN 60584
-200 to +1372°C
-199.9 to +999.9°C
Thermocouple
Fe-Con J
to EN 60584
-200 to +1200°C
-199.9 to +999.9°C
Accuracy
Measurement accuracy of display at 23°C ambient temperature ±(0.1% + 1 digit) or ±0.3°C; the larger value applies
Accuracy of cold junction (thermocouple only) -
Accuracy class of temperature probes Class A Class 2 Class 2 Class 2
±0.5°C for 5 to 40°C ±1°C for -20 to +5°C
and for 40 to 50°C
±0.5°C for 5 to 40°C ±1°C for -20 to +5°C
and for 40 to 50°C
Thermocouple
Cu-Con T
to EN 60584
-50 to +400°C
-50.0 to +400.0°C
±0.5°C for 5 to 40°C ±1°C for -20 to +5°C
and for 40 to 50°C
Measuring circuit monitoring
Probe short-circuit, probe/cable break, wrong connection
Symbol in display
Electrical data
Supply 1 alkaline battery, type LR6 AA Battery service life 400 hrs continuous operation at 23°C ambient temperature Power consumption 10mW (average value)
Environmental influences
Device type TDA-300 TDA-3000 Operating temperature range -20 to +50°C Temperature error ± 0.01% of measuring span for 5 to 40°C ambient temperature
± 0.02% of measuring span for -20 to +5°C and 40 to 50°C ambient temperature
Climatic conditions rel. humidity 95% annual average, no condensation Electromagnetic compatibility (EMC)
- interference emission
- immunity to interference IP enclosure protection IP67 IP54
EN 61326-1
Class B
to general requirements
Housing
Dimensions (W x H x D) 57 x 152 x 46mm Weight approx. 150g including battery
Display screen
Screen type FSTN LCD Display of measured temperature through 4 digits Display of memory number through 4 digits Display of total number of
measured temperatures through 4 digits Messages (date, time, ...) 11 characters (68 x 8 pixels) Data logging info through symbol; light is on during data recording,
flashes when ready for automatic data recording
Probe break/short-circuit through symbol;
light is on when the sensor is not connected or
in the event of a break or short-circuit.
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, Germany Page 3/6
Battery charge status 3-stage display Temperature unit °C or °F (12 x 8 pixels)
Data logging function
Device type TDA-300 TDA-3000 Recording type manual or automatic Recording interval any (manually), 1 — 3600 seconds (automatically) Contents of data record temperature, TAG No. (meas. point designation), user,
limits, and date/time Recording capacity 99 measurements 9999 measurements Data storage in SRAM
(volatile memory)
Loss of data when battery is discharged
or has to be replaced
Data are retained for about 10 years,
memory can be rewritten about 100,000
in EEPROM
(non-volatile memory)
times.
Limit monitoring
Limit monitoring upper and lower limit can be set for each measurement point Temperature within the limits
Temperature above or below the limit
Measurement points
Device type TDA-300 TDA-3000 Number of measurement points 5 99 TAG n u mber
(meas. point designation)
consisting of up to 11 characters (digits, letters, and symbols)
User
Device type TDA-300 TDA-3000 Number of users 1 99 User names - consisting of up to 11 characters
(digits, letters, and symbols)
USB interface (Universal Serial Bus) - only for TDA-3000
Speed USB 2.0 (theoretically 12 Mbps max.) Connection Mini-B connector Connecting cable included in delivery Supply via PC PC operating system for Windows® 2000, XP, Vista, 7 (32-bit and 64-bit)
Additional functions
Functions Peak/bottom value storage,
real-time clock (date and time),
function locking and
self-diagnostics
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, Germany Page 4/6
100
200 (100)
Ø 1,5
100 (150)
Temperature probes and adapters
RTD Pt100 with handle and attached connecting cable
Type Con struction Description Immersion probe
702546/01-100
Insertion probe
702546/02-100
The immersion probe with handle is particularly suitable for measuring temperatures in liquids. The temperature sensor is located in the probe tip, embedded in a heat transfer compound. The handle, which is provided with an anti-kink spring, is made from temperature-resistant plastic. Max. measurement temperature: 250°C Max. temperature of handle: 100°C Max. temperature of cable: 180°C
Thanks to its measurement tip, this probe is particularly suitable for measuring the core temperature in food and other soft materials. The silicone handle is covered by a protective sleeve and is resistant to corrosive media such as oil and fatty acids. Max. measurement temperature: 250°C Max. temperature of handle: 180°C Max. temperature of cable: 180°C
Thermocouples NiCr-Ni K with handle and attached compensating cable
Type Con struction Description Flexible immersion probe
(mineral-insulated thermocouple)
702545/01-...
Insertion probe
702545/02-...
Surface probe
702545/03-004
Surface probe
702545/03-015
The cable length for all probes is about 1500mm. The insertion probes are protected to IP67. Probes with type J and T thermocouples on request.
The immersion probe tip is suitable for measuring temperatures in liquids. Max. measurement temperature: 1150°C Max. temperature of handle: 100°C Max. temperature of cable: 180°C
Thanks to its measurement tip, this probe is particularly suitable for measuring the core temperature in food and other soft materials. The silicone handle is covered by a protective sleeve and is resistant to corrosive media such as oil and fatty acids. Max. measurement temperature: 250°C Max. temperature of handle: 180°C Max. temperature of cable: 180°C
The surface probe is particularly suitable for measurement on very small and uneven objects with poor heat conduction, for instance electronic components, glass, ceramic. The thermocouple is mounted on a spring plate so that the probe can also be applied obliquely to the surface. Max. measurement temperature: 400°C Max. temperature of handle: 100°C Max. temperature of cable: 180°C
This surface probe enables highly accurate and easily reproducible measurements on flat surfaces. Since the spring bands of the probe are transversely linked, the measurements are largely independent of the applied pressure and the contact angle. Max. measurement temperature: 500°C Max. temperature of handle: 110°C Max. temperature of cable: 180°C
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, Germany Page 5/6
1700
1700
Adapters for existing probes
Type Dia gr am Description
Adapter for RTD Pt100
702546/04-000
Adapter for thermocouple type K
702545/04-000
The adapter is 1700mm long and can be used in ambient temperatures up to 100°C.
Existing RTDs can be connected up via a connector-coupling combination (type Mini-Flach) in copper.
The adapter is 1700mm long and can be used in ambient temperatures up to 100°C.
Existing thermocouples can be connected up via a thermal connector-coupling (type Mini-Flach).
Dimensions
Carrying case
2013-12-03/00447944
Order details: Handheld thermometer with data logger
(1) Basic version
702540/ TDA-300, Type 702540/88-000
702541/ TDA-3000, Type 702541/88-000
(1) (2) (3)
Order code / 88 - 000 Order example 702541 / 88 - 000
Order details: Temperature probe and adapter
(1) Basic version
702545/ Thermocouple K
702546/ Pt100 RTD
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, Germany Page 6/6
x x 01 immersion probe x x 02 insertion probe x03surface probe x x 04 adapter with 1500mm cable, for existing probes
x x 000 with probe type 04 only x 004 4 mm dia. (probe type 03 only) x 015 15 mm dia. (probe type 03 only) x x 100 length: 100 mm (probe types 01 and 02 only) x 150 length: 150 mm (probe type 02 only) x 200 length: 200 mm (probe type 01 only)
Order code /- Order example 702545 / 01 - 200
(2) Probe type
(3) probe diameter/length
(1) (2) (3)
Standard accessories
- 1 Operating Manual
-1 battery
- 1 carrying hook
- 1 USB connecting cable (TDA-3000 only)
Accessories
- Silicone heat transfer compound (20 ml tube, for temperature measurements up to 200°C, part no. 94091460
- Carrying case for thermometer, two probes, heat transfer compound and accessories, part no. 00453912
2013-12-03/00447944
Loading...