within which the
measurement lies
(over/underrange)
Measurement /
measurement point/
user number
Number of recorded
measurement data
Probe break or
short-circuit
Battery charge status
Measured
temperature
Message window
(date, time, ...)
TDA-300 (Type 702540/...)
TDA-3000 (Type 702541/...)
Handheld Thermometer
with data logger
Brief description
TDA-300 and TDA-3000 are handheld digital temperature indicators. They are used
together with precision thermocouples or RTDs to measure temperatures on surfaces, in
liquids, in melts, and in soft materials.
Interchangeable temperature probes allow rapid adaptation to different applications.
Both instruments feature a data logging function. On Type TDA-3000, the recorded
measurement data can be read out via an USB interface. No special PC program is
necessary for visualization, since the measurement data are available as ASCII files in
CSV format (evaluation via spreadsheet programs).
The plastic housing is proof against shock and breakage and resistant to corrosive
substances. Thermometers, temperature probes and accessories are all available for
delivery from stock.
Displays
Page 1/6Data Sheet 702540
Comparison of models
Data logger99 measurements9999 measurements
TAG n u mbers
(meas. point designation)5 (11 characters each)99 (11 characters each)
User199 (11 characters each)
Interfaces-USB, type Mini-B
Enclosure protectionIP67IP54
2013-12-03/00447944
TDA-300TDA-3000
Key features
k Measurement input for Pt100,
NiCr-Ni K, Fe-Con J
and Cu-Con T
k DKD calibration certificate
k Data logger for 9999 measurements
k Simple readout via USB and
evaluation of data from a PC
(TDA-3000 only)
k Limit monitoring
k Peak/bottom value acquisition
k Battery AA, Mignon LR6,
with a long service life
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, GermanyPage 2/6
Technical data
Input
Measurement inputRTD
Pt100
to EN 60751
Range limits
- resolution 1°C
- resolution 0.1°C
Measurement offset-99.9 to +99.9°C
Sampling rate0.5 seconds
Input filter1st order digital filter; filter constant adjustable from 0 — 100 seconds
Unit°C or °F
-200 to +850°C
-199.9 to +850.0°C
Thermocouple
NiCr-Ni K
to EN 60584
-200 to +1372°C
-199.9 to +999.9°C
Thermocouple
Fe-Con J
to EN 60584
-200 to +1200°C
-199.9 to +999.9°C
Accuracy
Measurement accuracy of display
at 23°C ambient temperature±(0.1% + 1 digit) or ±0.3°C; the larger value applies
Accuracy of cold junction
(thermocouple only)-
Accuracy class of
temperature probesClass AClass 2Class 2Class 2
Supply1 alkaline battery, type LR6 AA
Battery service life400 hrs continuous operation at 23°C ambient temperature
Power consumption10mW (average value)
Environmental influences
Device typeTDA-300TDA-3000
Operating temperature range-20 to +50°C
Temperature error± 0.01% of measuring span for 5 to 40°C ambient temperature
± 0.02% of measuring span for -20 to +5°C and 40 to 50°C ambient temperature
- immunity to interference
IP enclosure protectionIP67IP54
EN 61326-1
Class B
to general requirements
Housing
Dimensions (W x H x D)57 x 152 x 46mm
Weightapprox. 150g including battery
Display screen
Screen typeFSTN LCD
Display of measured temperaturethrough 4 digits
Display of memory numberthrough 4 digits
Display of total number of
measured temperaturesthrough 4 digits
Messages (date, time, ...)11 characters (68 x 8 pixels)
Data logging infothrough symbol; light is on during data recording,
flashes when ready for automatic data recording
Probe break/short-circuitthrough symbol;
light is on when the sensor is not connected or
in the event of a break or short-circuit.
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, GermanyPage 3/6
Battery charge status3-stage display
Temperature unit°C or °F (12 x 8 pixels)
Data logging function
Device typeTDA-300TDA-3000
Recording typemanual or automatic
Recording intervalany (manually), 1 — 3600 seconds (automatically)
Contents of data recordtemperature, TAG No. (meas. point designation), user,
limits, and date/time
Recording capacity99 measurements9999 measurements
Data storagein SRAM
(volatile memory)
Loss of data when battery is discharged
or has to be replaced
Data are retained for about 10 years,
memory can be rewritten about 100,000
in EEPROM
(non-volatile memory)
times.
Limit monitoring
Limit monitoringupper and lower limit can be set for each measurement point
Temperature within the limits
Temperature above or below
the limit
Measurement points
Device typeTDA-300TDA-3000
Number of measurement points599
TAG n u mber
(meas. point designation)
consisting of up to 11 characters (digits, letters, and symbols)
User
Device typeTDA-300TDA-3000
Number of users199
User names-consisting of up to 11 characters
(digits, letters, and symbols)
USB interface (Universal Serial Bus) - only for TDA-3000
SpeedUSB 2.0 (theoretically 12 Mbps max.)
ConnectionMini-B connector
Connecting cableincluded in delivery
Supplyvia PC
PC operating systemfor Windows® 2000, XP, Vista, 7 (32-bit and 64-bit)
Additional functions
FunctionsPeak/bottom value storage,
real-time clock (date and time),
function locking and
self-diagnostics
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, GermanyPage 4/6
4Ø
100
200 (100)
Ø 1,5
4Ø
100 (150)
Temperature probes and adapters
RTD Pt100 with handle and attached connecting cable
TypeCon structionDescription
Immersion probe
702546/01-100
Insertion probe
702546/02-100
The immersion probe with handle is
particularly suitable for measuring
temperatures in liquids.
The temperature sensor is located in the
probe tip, embedded in a heat transfer
compound. The handle, which is provided
with an anti-kink spring, is made from
temperature-resistant plastic.
Max. measurement temperature: 250°C
Max. temperature of handle: 100°C
Max. temperature of cable: 180°C
Thanks to its measurement tip, this probe is
particularly suitable for measuring the core
temperature in food and other soft materials.
The silicone handle is covered by a protective
sleeve and is resistant to corrosive media
such as oil and fatty acids.
Max. measurement temperature: 250°C
Max. temperature of handle: 180°C
Max. temperature of cable: 180°C
Thermocouples NiCr-Ni K with handle and attached compensating cable
The cable length for all probes is about 1500mm. The insertion probes are protected to IP67.
Probes with type J and T thermocouples on request.
The immersion probe tip is suitable for
measuring temperatures in liquids.
Max. measurement temperature: 1150°C
Max. temperature of handle: 100°C
Max. temperature of cable: 180°C
Thanks to its measurement tip, this probe is
particularly suitable for measuring the core
temperature in food and other soft materials.
The silicone handle is covered by a protective
sleeve and is resistant to corrosive media
such as oil and fatty acids.
Max. measurement temperature: 250°C
Max. temperature of handle: 180°C
Max. temperature of cable: 180°C
The surface probe is particularly suitable for
measurement on very small and uneven
objects with poor heat conduction,
for instance electronic components, glass,
ceramic.
The thermocouple is mounted on a spring
plate so that the probe can also be applied
obliquely to the surface.
Max. measurement temperature: 400°C
Max. temperature of handle: 100°C
Max. temperature of cable: 180°C
This surface probe enables highly accurate
and easily reproducible measurements on flat
surfaces.
Since the spring bands of the probe are
transversely linked, the measurements are
largely independent of the applied pressure
and the contact angle.
Max. measurement temperature: 500°C
Max. temperature of handle: 110°C
Max. temperature of cable: 180°C
2013-12-03/00447944
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, GermanyPage 5/6
1700
1700
Adapters for existing probes
TypeDia gr amDescription
Adapter for
RTD
Pt100
702546/04-000
Adapter for
thermocouple
type K
702545/04-000
The adapter is 1700mm long and can be used
in ambient temperatures up to 100°C.
Existing RTDs can be connected up via a
connector-coupling combination
(type Mini-Flach) in copper.
The adapter is 1700mm long and can be used
in ambient temperatures up to 100°C.
Existing thermocouples can be connected up
via a thermal connector-coupling
(type Mini-Flach).
Dimensions
Carrying case
2013-12-03/00447944
Order details: Handheld thermometer with data logger
(1) Basic version
702540/TDA-300, Type 702540/88-000
702541/TDA-3000, Type 702541/88-000
(1)(2)(3)
Order code/88-000
Order example702541/88-000
Order details: Temperature probe and adapter
(1) Basic version
702545/Thermocouple K
702546/Pt100 RTD
Data Sheet 702540JUMO GmbH & Co. KG • 36035 Fulda, GermanyPage 6/6
x x01immersion probe
x x02insertion probe
x03surface probe
x x04adapter with 1500mm cable, for existing probes
x x000with probe type 04 only
x0044 mm dia. (probe type 03 only)
x01515 mm dia. (probe type 03 only)
x x100length: 100 mm (probe types 01 and 02 only)
x150length: 150 mm (probe type 02 only)
x200length: 200 mm (probe type 01 only)
Order code/-
Order example702545/01-200
(2) Probe type
(3) probe diameter/length
(1)(2)(3)
Standard accessories
- 1 Operating Manual
-1 battery
- 1 carrying hook
- 1 USB connecting cable (TDA-3000 only)
Accessories
- Silicone heat transfer compound (20 ml tube, for
temperature measurements up to 200°C,
part no. 94091460
- Carrying case for thermometer, two probes,
heat transfer compound and accessories,
part no. 00453912