JUMO 701550 User Manual

JUMO di 308
Digital Indicator
B 70.1550.0
Operating Manual
2010-04-28/00485282
Please read this operating manual before commissioning
E
)
*
the instrument. Keep the manual in a place accessible to all users at all times. Your comments are appreciated and may assist us in improving this manual.
All necessary settings are described in this operating manual. Manipulations not described in the manual or expressly forbidden will jeopardise your warranty rights. Please contact the nearest subsidiary or the head office, should you encounter problems
The manual is valid from instrument software version
217.01.01
It appears by simultaneously pressing the P and i keys (four-digit display; example: 01.01).
When accessing the inner parts of the unit and returning
.
modules, assemblies or components, please observe the regulations according to EN 61340-5-1 and EN 61340-5-2 „Protection of electrostatic sensitive devices“. Only use ESD packaging for transport.
Please note that we cannot accept any liability for damage caused by ESD.
ESD=Electro Static Discharge
Content
1 Introduction 7
1.1 Description ............................................................... 7
1.2 Typographical conventions ..................................... 9
2 Identifying the instrument version 11
2.1 Type designation .................................................... 11
2.2 Scope of delivery ................................................... 13
2.3 Accessories ............................................................ 13
3 Mounting 15
3.1 Mounting site and climatic conditions ................ 15
3.2 Dimensions ............................................................ 15
3.3 Fitting in position ................................................... 15
3.4 Removing the plug-in module .............................. 16
4 Electrical connection 17
4.1 Installation notes ................................................... 17
4.2 Electrical isolation ................................................. 19
4.3 Connection diagram .............................................. 20
4.4 Termination resistor for the RS422/485 interface 25
4.5 Connection of the PROFIBUS-DP connector ..... 26
Content
5 Operation 27
5.1 Displays and controls ............................................ 27
5.2 Level concept ......................................................... 28
5.3 Level inhibit ............................................................ 29
5.4 Entries and operator prompting ........................... 30
6 Operator level 31
7 Configuration 33
7.1 Analog inputs „INPUT“ .......................................... 35
7.2 Limit comparators „LIMITCOM“ .......................... 42
7.3 Outputs „OUTPUT“ ................................................ 51
7.4 Binary functions „BINFUNCT“ .............................. 54
7.5 Display / Operation „DISPLAY“ ............................ 56
7.6 Interfaces „INTERFCE“ ......................................... 60
8 Extra codes 63
8.1 Math and logic module ......................................... 63
8.2 Difference, humidity or ratio calculation ............. 64
Content
9 Retrofitting of modules 65
10 Appendix 67
10.1 Technical data ........................................................ 67
10.2 Alarm messages .................................................... 72
11 Index 73
Content

1 Introduction

1.1 Description

The digital indicator shows temperatures in °C or °F and standard signals in plain text.
Inputs/outputs The standard instrument is equipped with an analog
input, two binary inputs, two relay outputs, two logic outputs as well as a voltage supply for two-wire transmitters.
Optional modules
Displays The high-contrast, multi-colour LCD display for
Three extension slots can be equipped with additional inputs and outputs as well as interfaces.
process value/text and operator prompting contains a five-digit 7-segment display (showing the value or parameter setting) and an eight-digit 16-segment display with colour change (value, parameter name, channel name, process/alarm text as max. 24 character ticker or pseudo bargraph). Four additional switch position indicators are available for binary outputs (relay or logic).
Operation The instrument is operated and configured by four
keys; an optional setup program for a PC is available. The user-friendly setup program provides additional configuration possibilities (e.g. math and logic
Special functions
functions, display texts). The instrument offers 4 configurable limit
comparators and an optional math and logic module (two virtual channels).
Extensive binary functions are available for the assignment of functions to the signals of limit comparators, logic and binary inputs.
7
1 Introduction
Special functions (continued)
The computation results of both math functions can be used for the different analog parameters (e.g. as value shown in the display).
Instruments with a second (optional) analog input allow the computation of differential, humidity or ratio computations by means of default formulas.
Probes 10 types of probes (RTD temperature probe,
thermocouple, resistance transmitter, standard signals) and more than 20 linearisations are available for analog input configuration. Customer-specific linearisation with 10 interpolation points or by the entry of the polynomial coefficients is possible.
Interface and electrical connection
An optional interface (RS422/485 or PROFIBUS-DP) can be used for integration of the instrument in a data network.
Block structure
The electrical connection is made at the back of the instrument by means of screw terminals.
Analog input
2 binary inputs
Option 1
Option 2
Option 3
2 relays (change-over)
2 logic outputs
Voltage supply for 2-wire transmitters
Setup interface
8

1.2 Typographical conventions

1 Introduction
Warning signs
V
*
E
Danger
Caution
Caution
This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!
This symbol is used where special precautionary measures are required when handling components liable to damage through electrostatic discharge.
Note signs
H
v
h
Note
Reference
Action instruction
This symbol is used to draw your special attention to a remark.
This symbol refers to further information in other operating manuals, chapters or sections.
This symbol refers to a description of an action to be performed.
The individual steps are marked by this asterisk, e.g.:
EXIT
hPress
EXIT
F
F
9
1 Introduction
Representa­tion
Menu items
Blinking display
Text referring to the setup program is shown in italics, for example: „Display/Operation“.
S
E
NS
O
R
10

2 Identifying the instrument version

2.1 Type designation

(1) Basic type
701550 Digital Indicator
incl. 1 analog input, 2 binary inputs, 2 relay outputs, 2 logic outputs and 1 setup interface, Front dimension 96mm x 48mm
(2) Basic type extensions
1 Basic type
Version
8 Standard with factory settings 9 Programming to customer specification
Logic outputs (2 are available as standard)
10/12V
(3) Option slots
1. 2. 3. Option slot Max.
0 0 0 not assigned 111Analog input 2
(universal) 1 2 2 2 Relay (change-over) 2 3 3 3 2 relays (n.o. make) 2 444Analog output 2 5 5 5 2 binary inputs 2 6 6 6 Solid state relay 1A 2 7 7 7 RS422/485 interface 1 8 8 8 PROFIBUS-DP
interface 1
number
Please note: The position of the options (slot 1, 2 or 3) is freely assignable, however, the max. number must not be exceeded.
(4) Voltage supply
23 AC 110—240V -15/+10%, 48—63Hz 25 AC/DC 20—30V, 48—63Hz
(5) Extra codes
000 none 214 Math and logic module
(1) (2) (3) (4) (5)
Ordering code:
Ordering example:
701550 / 1 8 1 – 1 4 0 – 2 3 / 0 0 0
/––/
11
2 Identifying the instrument version
View of option slots
1
2
3
12
2 Identifying the instrument version

2.2 Scope of delivery

- Display instrument
- Seal
- Mounting brackets
- Operating Manual B70.1550.0 in DIN A6 format

2.3 Accessories

Mini-CD Mini-CD with demo setup program and PDF
documents (operating manual and further documentation) Sales No.: 70/00448699
PC interface PC interface with TTL/RS232 converter and adapter
(socket connector) for setup program Sales No.: 70/00350260
USB interface PC interface with USB/TTL converter, adapter (socket
connector) and adapter (pins) Sales No.: 70/00456352
Setup program
Setup program with startup function (recording and visualisation measuring data) Sales No.: 70/00493223
13
2 Identifying the instrument version
Setup program (continued)
Required hardware:
- PC Pentium IV or compatible
- 256MB RAM, 100MB free fixed disk memory
-CD ROM drive
- free serial or USB interface
Required software:
Microsoft
1
Windows 2000/XP/Vista
14
1. Microsoft is a registered trademark of Microsoft Corporation

3.1 Mounting site and climatic conditions

The conditions at the mounting site must meet the requirements specified in the technical data. The ambient temperature at the mounting site can range from 0...55°C with a maximum relative humidity of 90%.

3.2 Dimensions

3 Mounting

48
57.1
99 96
91.5
PC interface adapter
44.5
Panel cut-out
92
90
+0.8
15
48
+0.8
45

3.3 Fitting in position

hPlace the supplied seal
on the instrument body.
hInsert the instrument
from the front into the panel cut-out.
hFrom the panel rear,
slide the mounting
15
3 Mounting
brackets into the guides on the sides of the housing. The flat faces of the mounting brackets must make contact with the housing.
hPlace the mounting brackets against the panel rear,
and tighten them evenly with a screwdriver.
Mounting controllers back-to-back/ next to each other
Minimum spacing of panel cut-outs
without setup plug 30mm 11mm with setup plug (arrow) 65mm 11mm
Care of the front panel
The front panel can be cleaned with commercial detergents and cleaning agents. It has a limited resistance to organic solvents (e.g. methylated spirits, white spirit, P1, xylol, etc.). Do not use high-pressure cleaning equipment.

3.4 Removing the plug-in module

The plug-in module can be removed from its housing for servicing.
horizontal vertical
16
hPress together the
knurled surfaces on the front panel (left and right), and pull out the plug-in module.
H
When re-inserting the plug-in module, ensure that the latches (beneath the knurled areas) engage.

4.1 Installation notes

- The choice of cable, the installation and the electrical connection of the instrument must conform to the requirements of VDE 0100 "Regulations on the Installation of Power Circuits with Nominal Voltages below 1000V" or the appropriate local regulations.
- The electrical connection must only be carried out by qualified personnel.
- The instrument shall be operated by mains protected with a branch circuitry overcurrent protection device not more than 20 Amps. For servicing/repairing a Disconnecting Device shall

4 Electrical connection

be provided to disconnect all conductors.
- The load circuit must be fused for the maximum relay current, in order to prevent the output relay contacts becoming welded in the event of a short circuit occurring at that point.
- Electromagnetic compatibility conforms to the standards and regulations cited in the technical data.
vChapter 10.1 „Technical data“
- Run input, output and supply cables separately and not parallel to one another.
- Sensor and interface cables should be shielded cables with twisted conductors. Do not run cables close to current-carrying components or cables. Ground the shielding on one side.
- Do not connect any additional loads to the supply terminals of the instrument.
17
4 Electrical connection
- The instrument is not suitable for use in areas with an explosion hazard (Ex areas).
.
Only allow qualified personnel to
V
carry out the electrical connection.
Identify the instrument version by
H
means of the type code.
Installation information on conductor cross sections and core ferrules
Without core-end ferrule
Core-end ferrule without lip Core end ferrule with lip
up to 1.5mm Core end ferrule with lip
above 1.5mm Twin ferrule with lip
2
2
Minimum cross­section
0.34mm
0.25mm
0.25mm
1.5mm
0.25mm
2
2
2
2
2
Maximum cross­section
2.5mm
2.5mm
1.5mm
2.5mm
1.5mm
2
2
2
2
2
Min. length of core-end ferrule
10mm (stripped)
10mm 10mm
12mm
12mm
18

4.2 Electrical isolation

30 V AC 50 V DC
4 Electrical connection
3800 V AC
Input 1
»
30 V AC 50 V DC
Input 2 Solid-state relay outputs
3800 V AC
»
30 V AC 50 V DC
Relay outputs
» »
Analog outputs
»
Logic outputs
30 V AC 50 V DC
Binary inputs
Setup interface
RS422/485 PROFIBUS-DP
30 V AC 50 V DC
»
Voltage supply
»
3800 V AC
Voltage supply for 2-wire transmitters
»
19
4 Electrical connection

4.3 Connection diagram

Terminal strips on the back of the instrument:
Option 1Option 2Option 3 Terminal strip 1 (Option slots)
Terminal strip 2
Terminal strip 3
Connection diagram in the setup program
12
17
16
15
5
6
7
10
11
8
6
7
8
9
8
9
6
9
12
13
4
3
3
4
4
5
2
2
N(L+)
1
1
L1(L+)
10
11
The setup program includes a graphic connection diagram subject to updates depending on the configuration or equipment.
It also allows the preparation of a list of connections containing the hardware equipment and configuration of the connections.
Connection diagram and list of connections can be printed out.
vSetup program (Extras -> Connection diagram; or
via Toolbar „IN/OUT“)
20
4 Electrical connection
Assignment of terminal strip 3: Voltage supply and binary outputs 1+2
L1
L+
N
L-
Supply
4 5 6
8
=
U
9
11 12 13
U
L1
AC 110—240V AC/DC 20—30V
N
+
=
15.8—15.2V / 30—50mA
-
Ö
230V/3A
P S
Voltage supply for 2-wire transmitters (No-load voltage approx. 25V)
Binary output 1
L+ L-
Relays
15 16 17
Ö P S
230V/3A
Binary output 2
21
4 Electrical connection
Assignment of terminal strip 2: Analog input 1, binary inputs 1+2, and binary outputs 3+4
Terminal strip 2
RTD temp. probe
1 2 3
input 1
Analog
4
1
2
input 1
Analog
3 4
Thermo­couple
RTD temp. probe
Current Voltage
+
-
- /
I
x
Terminal
RTD temp. probe
0(2)—10V
+
U
x
-
+
~
I
~
x
Resistance transmitter
E S A
Voltage 0—1V
+
U
x
-
-
strip 1*
5
1
6
2 3 4
1
2 3 4
7 8
5 6 7 8
10 11
12
10 11
12
9
9
Binary input 1+2
6 7
1 2
8
Binary
9
10
Position of terminal strip 1 and 2 (on the back of the instrument):
Terminal strip 1
Terminal strip 2
Terminal strip 3
GND
10
11
12
Option 3
15
16
17
Binary output 3+4 (logic 12V)
13
12
10
8
9
11
9
3 (+) 4 (+)
GND (-)
5
6
7
6
7
8
8
9
4
Option 1Option 2
4
5
6
1
2
3
1
2
3
N(L+)
4
L1(L+)
* Assignment of
terminal strip 1 for analog input 2 (see also next page)
22
4 Electrical connection
Assignment of terminal strip 1 (option boards): Inputs, outputs and interfaces
1 2 3
Option 1
4
5 6 7
Option 2
8
9
10 11
Option 3
12
Analog input
Analog input 2 Connection as
analog input 1
Analog input 2 Connection as
analog input 1
Analog input 2 Connection as
analog input 1
2 binary inputs
Binary input 3+4
Binary input 5+6
Binary input 7+8
3 4 GND
5 6 GND
7 8 GND
Analog output
U
/ I
x
Analog output 5
Ux/ I
Analog output 6
Ux/ I
Analog output 7
Relay (change-over)
Ö
+
x
-
Binary output 5
+
x
-
Binary output 6
+
x
-
Binary output 7
P S
Ö P S
Ö P S
H
H
The maximum number of option boards has to be taken into account (see Chapter 2.1 „Type designation“).
Note numbering of the outputs (see „OUTPUT““
).
Chapter 7.3 „Outputs
23
4 Electrical connection
Assignment of terminal strip 1 (option boards) - continued: Inputs, outputs and interfaces
1 2 3
Option 1
4
5 6 7
Option 2
8
9
10 11
Option 3
12
2 Relays (n.o. make)
Binary output 5+8
Binary output 6+9
Binary output 7+10
Solid-state relay
5
8
6
9
7
10
Binary output 5
Binary output 6
Binary output 7
PROFIBUS
VP (+5 V) RxD/TxD-P (B) RxD/TxD-N (A) DGND
VP (+5 V) RxD/TxD-P (B) RxD/TxD-N (A) DGND
VP (+5 V) RxD/TxD-P (B) RxD/TxD-N (A) DGND
RS422
RxD + RxD ­TxD + TxD -
RxD + RxD ­TxD + TxD -
RxD + RxD ­TxD + TxD -
RS485
RxD/
TxD +
RxD/
TxD -
RxD/
TxD +
RxD/
TxD -
RxD/
TxD +
RxD/
TxD -
H
H
24
The maximum number of option boards has to be taken into account (see Chapter 2.1 „Type designation“).
Note numbering of the outputs (see „OUTPUT““
).
Chapter 7.3 „Outputs
4 Electrical connection

4.4 Termination resistor for the RS422/485 interface

Setting resistors
To ensure fault-free operation of several instruments in a line structure, their internal termination resistors must be activated at the start and end
.
hPull plug-in module out towards the front by
pressing on the knurled areas
hUsing a suitable aid (e.g. ballpoint pen), press all the
white switches into the same direction
Bus
h Push all 5 switches down termination active
No
h Push all 5 switches up bus termination (ex-factory)
hRe-insert the module into the housing
Check hPress the P + i keys
When checking the software version and the termination resistors activated, an additional decimal point appears behind the version number (top display).
Example of active: inactive:
version number 01.01
01.01.
01.01
:
25
4 Electrical connection

4.5 Connection of the PROFIBUS-DP connector

Mounting the adapter
Assignment of the 9 pole D-Sub socket
hIdentify option slot with the PROFIBUS-DP
interface by means of the type code (in the case of pre-configured instruments)
2
1
3
1
In this example, the PROFIBUS-DP interface is in option slot 1.
Pin at D-Sub socket
2 3
1 2 3 4
4
5 6
Pin at terminal strip 1: Signal
(Example for option slot 1)
6 7
Designation
L1(L+) N(L+)
4 5
6
8
6 1: VP Voltage supply,
positive
3 2: RxD/TxD-P Receive/Transmit
data, positive
8 3: RxD/TxD-N Receive/Transmit
data, negative
5 4: DGND Ground
To fit the D-Sub adapter, open the black housing
*
of the adapter board; otherwise the connection screws in the instrument back are not accessible.
It is important to note that the adapter is fitted in the position shown above to ensure correct pin assignment.
26

5.1 Displays and controls

(1)

5 Operation

(4)
(2)
(3)
(1) 7-segment display (measured value display)
five-digit, red; decimal place is configurable (automatic adjustment on display overflow)
(2) 16-segment display (24 character ticker,
parameter name, level symbols) eight-digit, green or red; decimal place configurable
(3) Indication
yellow; for four switch positions of max. four outputs (display lit = ON)
(4) Keys
The displays are configurable. vChapter 7.5 „Display / Operation „DISPLAY““
27
5 Operation

5.2 Level concept

The parameters for instrument setting are organised at different levels.
Normal display
EXIT
F
> 2 sec
or time-out
Navigation principle
EXIT
F
i
d
P
P
USER
OPERATOR
CONFIG
1. USER only
1
visible if user data is configured (setup program)
P
P
P
i
d
Time-out
H
If no key is pressed for 180 secs the instrument changes back to normal display!
vChapter 6 „Operator level“
User level
Up to eight freely chosen
parameters
Operator level
Process data
Configuration level
- INPUT (analog inputs)
- LIMITCOM (limit comparators)
- OUTPUT (outputs)
- BINFUNCT (binary functions)
- DISPLAY (display)
- INTERFCE (interfaces)
User data „USER“
vChapter 7 „Configuration“ vSetup program (Display/Operation -> Operation ->
Operation time-out)
The setup program allows the display and editing up to 8 freely chosen parameters at this level.
vSetup program (Display/Operation -> User data ->
Parameters 1...8)
The user can assign a symbol for the representation of each parameter. Otherwise, the default symbol will appear. All letters and numbers that can be presented by a 16 segment display are permissible.
28

5.3 Level inhibit

Access to the individual levels can be prevented.
Code Configuration level
0enabled 1 inhibited
5 Operation
hEnter code by pressing
P and d (simultaneously
for > 5sec).
hChange code by pressing hEnter code by pressing
P (display blinks!)
i and d. Ex-factory: all
levels enabled.
EXIT
hReturn to normal display by pressing or
EXIT
F
F
automatically after approx. 180 secs
The configuration level can also be inhibited via the binary function.
vChapter 7.4 „Binary functions „BINFUNCT““
29
5 Operation

5.4 Entries and operator prompting

When entries are made within the levels, the parameter symbol appears in the lower display.
Select
P
id/
NS
R
O
S
E
hSelect parameter by pressing hChange to the entry mode by pressing
Alter
E
S
R
O
NS
i or d.
i
d
P (lower
display blinks!)
hAlter value by pressing i and d
The value alters dynamically for as long as the key is kept pressed.
hAssign the value by pressing
P or automatically
after 2 secs
or
EXIT
hCancel the entry with .
EXIT
F
F
The value will be assigned.
To enter digits after the decimal point, the value
H
of system point must be set accordingly (see page 57).
For the display of measurement values of the analog inputs, the digits after the decimal point can be set separately (see page 38).
30
Access

6 Operator level

Normal display
EXIT
F
> 2 sec
P
or time-out
Navigation principle
EXIT
F
i
d
P
USER
1
d
P
OPERATOR
1. USER only visible if user data is configured (setup program)
Operator level
Process data
31
6 Operator level
Process data Process data is shown in the operator level in
accordance with the configuration.
Symbol Meaning
INPUT1 Measured value of analog input 1 INPUT2 Measured value of analog input 2 (only if
available)
MIN INP1 Minimum value for analog input 1
(only if function is activated)
MAX INP1 Maximum value for analog input 1
(only if function is activated)
HOLD1 Hold value for analog input 1
(only if function is activated)
MIN INP2 Minimum value for analog input 2
(only if analog input 2 is available and function activated)
MAX INP2 Maximum value for analog input 2
(only if analog input 2 is available and function activated)
HOLD2 Hold value for analog input 2
(only if analog input 2 is available and function activated)
MATHE1 Calculated result of mathematical formula 1
(only if mathematics module is available or if analog output 2 is available as a prerequisite for function „Humidity“, „Difference“ or „Ratio“)
MATHE2 Calculated result of mathematical formula 2
(Same conditions as with MATHE1)
32
Access

7 Configuration

Normal display
EXIT
> 2 sec
F
or time-out
Navigation principle
EXIT
F
i
d
H
Levels can be inhibited
P
P
vChapter 5.3 „Level inhibit“
USER
OPERATOR
CONFIG
1. USER only visible
1
d
d
P
- INPUT (analog inputs)
- LIMITCOM (limit comparators)
- OUTPUT (outputs)
- BINFUNCT (binary functions)
- DISPLAY (display)
- INTERFCE (interfaces)
if user data is configured (setup program)
Configuration level
H
H
Parameters are not displayed if the equipment level does not permit the function assigned to the parameter. Example: Analog output 2 cannot be configured if no second analog output is implemented in the instrument.
Some parameters can only be programmed through the set-up program. In the following tables, these are marked in the „Parameter“ column with „(Setup)“.
33
7 Configuration
Analog selector
With some parameters, you can choose from a series of analog values. To provide you with an overview, this selection is listed below.
Value Description
0
deactivated
1
analog input 1
2
analog input 2
3
(reserved)
4
(reserved)
5
math 1
6
math 2
7
(reserved)
8
(reserved)
9
(reserved)
10
(reserved)
11
analog marker
12
minimum value input 1
13
minimum value input 2
14
(reserved)
15
(reserved)
16
maximal value input 1
17
maximal value input 2
18
(reserved)
19
(reserved)
20
hold value input 1
21
hold value input 2
22
(reserved)
23
(reserved)
24
any analog value
25
internal Pt100
26
sampling cycle time
34

7.1 Analog inputs „INPUT“

7 Configuration
Configuration
Analog inputs
Limit comparators
Outputs Binary
functions Display /
Operation Interfaces
Sensor type
Depending on the instrument version, up to two analog inputs are available.
INPUT1 (analog input 1) ➔ ➔ INPUT2 (analog input 2)
Parameter Value/
Selection
SENSOR 0
Description
No function
1
RTD temperature probe in 3-wire circuit
2
RTD temperature probe in 2-wire circuit
3
RTD temperature probe in 4-wire circuit
4
Thermocouple
5
Resistance transmitter
7
0…20mA
8
4 ... 20mA
9
0…10V
10
2…10V
11
0 ... 1V
Factory set on analog input 2: no function
Factory settings are shown
bold
.
35
7 Configuration
INPUT1 (analog input 1) ➔ ➔ INPUT2 (analog input 2)
Linearization
Parameter Value/
Selection
LINEAR 0
10 11 12 13 14 15 16 17 18 19 20 21 22 23
Description
Linear
1
Pt100 DIN
2
Pt500 DIN
3
Pt1000 DIN
4
KTY11-6
5
Pt100 GOST
6
Pt 50 GOST
7
Cu100
8
Cu50
9
Chromel-Copel W5Re-W26Re C W3Re-W25Re D NiCr-Con E Cu-Con T Fe-Con J Cu-Con U Fe-Con L NiCr-Ni K Pt10Rh-Pt S Pt13Rh-Pt R Pt30Rh-Pt6Rh B NiCrSi-NiSi N W3Re-W26Re Customised linearization
Measurement offset
OFFSET -19999…
0
99999
For customised linearization, a maximum of 10 knee points can be implemented, or a 4th order polynomial function programmed
the setup program)
For the „KTY11-6“ linearization, the resistance
Ω
is 2k
program and with 2-wire circuit)
The measurement offset is used to correct a measured value by a certain amount upward or downward. Examples: Measured displayed value offset value
294.7 +0.3 295.0
295.3 - 0.3 295.0
To enter digits after the decimal point, the value of system point must be set accordingly (see page 57).
Special case: „2-wire circuit“ If the input is connected to an RTD temperature probe in 2-wire circuit, then the lead resistance is set in ohms here.
at 25°C
.
(setting only through the setup
(only through
.
36
Factory settings are shown
bold
.
INPUT1 (analog input 1) ➔ ➔ INPUT2 (analog input 2)
7 Configuration
Scale low point
Scale high point
Filter time constant
Parameter Value/
Selection
SCAL-LOW -19999…
0
99999
SCAL-HI -19999…
100
99999
FILTER 0.0…
0.6
100.0
Description
On transducers with standard signal and on resistance potentiometers, a display value is assigned to the physical signal (scaling).
Example: 0—20mA  0—1500°C.
The range of the physical signal can be 20 % wider or narrower without generating an out-of­range signal.
With a standard signal and customised linearization, the display range coincides with the linearization range (range of the x values). For the above example this means: Start x = 0, End x = 20, in order that the display range goes from 0 to 1500 °C. If the range of the x values is smaller, the dis­play range is reduced accordingly.
To adjust the digital input filter (time in seconds; 0.0 sec = filter off). 63% of the alterations are acquired after 2x filter time constant (2nd order filter) at a filter time step change. When the filter time constant is large:
- high damping of interference signals
- slow reaction of the process value display to process value changes
- low limit frequency (low-pass filter)
Fine adjustment begin value
Fine adjustment end value
FINEADJB -19999…
0
99999
FINEADJE -19999…
1
99999
Factory settings are shown
These parameters are factory-
(Activation in the setup program > Non documented parameters; please contact the manufacturer.)
These values cannot be accepted by another instrument.
A
If these values have been altered by mistake, this setting has to be canceled using the procedure described under „Customised fine tuning“.
See description on page 39.
bold
.
deactivated
.
37
7 Configuration
INPUT1 (analog input 1) ➔ ➔ INPUT2 (analog input 2)
Decimal point
Correction value KTY at 25°C
Parameter Value/
Selection
DECPOINT 0
(Setup) 0…
2000
4000
Factory settings are shown
Description
no digit after the decimal point
1
one digit after the decimal point
2
two digits after the decimal point
3
three digits after the decimal point
7
System point
This setting is only valid for the display of measurement value of analog input 1 or 2!
(Adjustment in the setup program under Display/Operation -> Display -> Channel name)
Resistance in ohms at 25°C/77°F for „KTY 11-6“ linearisation
Adjustment only possible in the setup program:
-> Analog inputs -> Analog input 1 or 2
bold
.
Temperature unit
Temperature unit
Mains frequency
Mains frequency
Sampling cycle time
Sampling cycle time
INPUT (analog inputs general)
Parameter Value/
Selection
UNIT
FREQUENC
CYCLE-t
Factory settings are shown
Description
01deg. Celsius
deg. Fahrenheit
Unit for temperature values
0150Hz
60Hz
Adaptation of the conversion time of the input circuitry to the supply frequency
0
50ms
1
90ms
2
150ms
3
250ms
bold
.
38
7 Configuration
Customised fine adjust­ment
The customised fine adjustment is used to correct the values displayed by the instrument. This may be necessary, for example, after a system validation, if the displayed values no longer coincide with the actual values at the point where the measurement is taken.
Using a reference measuring instrument, two measured values are determined which should be as far apart as possible (start value, end value). Ensure that the measuring conditions are stable. Enter the reference value found as the start value (FINEADJB) or end value (FINEADJE) on the instrument to be tuned.
Example The temperature inside an oven is measured with an
RTD temperature probe and displayed on an instrument. The reading on the instrument deviates from the actual temperature as a result of the sensor temperature drifting. At 20°C the instrument reads 15°C, at 80°C it shows 70°C (exaggerated example for better understanding).
Sensor
15°C / 70°C
20°C / 80°C
20°C / 80°C
Oven
Display on the instrument
Reference measuring instrument
39
7 Configuration
Procedure: hStep 1: The measurement carried out with the
reference measuring instrument shows a constant oven temperature of 20°C.
hStep 2: Enter value 20 as start value (FINEADJB) at
the instrument.
hStep 3: The oven temperature is increased to 80°C,
the temperature is still controlled by a reference measuring instrument. The temperature must remain constant.
hStep 4: Enter the value 80 as end value
(FINEADJE) at the instrument.
The following diagram shows the changes in the characteristic curve caused by the fine adjustment (point of intersection with the x axis as well as ascent)
Display (Instrument)
80
70
After fine adjustment
Before fine adjustment
20
15
Measurement (Reference)
40
20 80
7 Configuration
Special case: Offset
Repeated fine adjust­ment
If the deviation between measured value and displayed value at the low and high measuring point is identical, an offset correction is sufficient (ascent remains unchanged). In this case, fine adjustment is not required.
vMeasured value correction (offset), page 36
Reset the fine adjustment prior to repeating it. For this purpose, program start value (FINEADJB) and end value (FINEADJE) are given the same value. This automatically sets the start value to 0 and the end value to 1.
Always check the start value and the end value
*
prior to starting fine adjustment. Reset the fine adjustment, if they deviate from the factory-set values 0 (FINEADJB) and 1 (FINEADJE).
41
7 Configuration

7.2 Limit comparators „LIMITCOM“

Configuration
Analog inputs
Limit comparators
Outputs Binary
functions Display /
Operation Interfaces
Limit comparator functions
Hysteresis function
Limit comparators (threshold monitors, limit contacts) can be used to monitor an input variable (process value for the limit comparator) against a fixed limit value or another variable w (setpoint value for the limit comparator). When a limit value is exceeded, a signal can be output or an internal controller function initiated.
4 limit comparators are available.
Limit comparators can have different switching functions (lk1 to lk8). The switching differential HySt (HYSTERES) can be set and is, in all cases, symmetrical in relation to the limit value.
The hysteresis function (symmetrical, asymmetrical) defines the ranging of the switching differential around the limit value (adjustment in the setup program).
42
7 Configuration
Limit value AL relative to setpoint value w
lk1 symmetrical
asymmetrical, left asymmetrical, right
The limit comparator functions lk1 to lk6 monitor the input variable for a limit value AL (LIMVALUE) to be set,
the absolute value depending on setpoint value w.
ON
AL
HySt
w
Measurement
lk2 symmetrical
asymmetrical, left asymmetrical, right
ON
HySt
AL
w
Measurement
43
7 Configuration
lk3 symmetrical
asymmetrical, left asymmetrical, right
ON
HySt
AL
Measurement
w
lk4 symmetrical
asymmetrical, left asymmetrical, right
ON
HySt
AL
Measurement
w
44
lk5 symmetrical
7 Configuration
ON
AL
HySt
asymmetrical, left asymmetrical, right
lk6 symmetrical
ON
w
w
Measurement
AL
HySt
Measurement
asymmetrical, left asymmetrical, right
Example of a variable setpoint value
The measurement is monitored (analog input 1). The setpoint value w default value is manually entered via a potentiometer connected to analog input 2. For this, analog input 2 is selected as setpoint value (LCSETVAL).
45
7 Configuration
Fixed limit value AL
lk7 symmetrical
asymmetrical, left asymmetrical, right
In the case of the limit comparator functions lk7 and lk8, the measurement is monitored with respect to a fixed limit value AL (LIMVALUE).
ON
HySt
Measurement
AL
lk8 symmetrical
asymmetrical, left asymmetrical, right
ON
HySt
Measurement
AL
46
LIMITC 1 (limit comparator 1) ➔ ➔ LIMITC 2 (limit comparator 2) ➔ ➔ LIMITC 3 (limit comparator 3) ➔ ➔ LIMITC 4 (limit comparator 4)
7 Configuration
Function
Limit value
Hysteresis
Fixed limit comparator setpoint value
Parameter Value/
Selection
FUNCTION
LIMVALUE -19999…
0
99999
HYSTERES 0…1…
99999
FIXLCVAL -19999…
0
99999
Description
0
no function
1
lk1
2
lk2
3
lk3
4
lk4
5
lk5
6
lk6
7
lk7
8
lk8 Limit value to be monitored
(see limit comparator functions lk1...lk8: limit value AL)
Switching differential in respect to the limit value (see limit comparator functions lk1...lk8: hysteresis HySt)
A fixed setpoint value can be set for the limit comparator (lk1...lk6).
The limit comparator setpoint value LCSETVAL must be deactivated for the fixed setpoint value to be active.
Action/ Range response
Switch-on delay
Switch-off delay
ACT-RESP
ON DELAY0…9999 Delays the switch-on edge by a definable
OFFDELAY0…9999 Delays the switch-off edge by a definable
Factory settings are shown
0
absolute/off
1
relative/off
2
absolute/on
3
relative/on
Defines the switching action of the limit comparator and the switch status for an overrange or underrange (signal at „Out of Range“).
See description on page 49.
time period (time in seconds).
time period (time in seconds).
bold
.
47
7 Configuration
LIMITC 1 (limit comparator 1) ➔ ➔ LIMITC 2 (limit comparator 2) ➔ ➔ LIMITC 3 (limit comparator 3) ➔ ➔ LIMITC 4 (limit comparator 4)
Acknowledge­ment
Pulse time
Limit comparator actual value
Limit comparator setpoint value
Parameter Value/
Selection
ACKNOWL
PULSE-t
0
…9999 The limit comparator resets automatically
LCACTVAL (analog
selector)
Analog input 1
LCSETVAL (analog
selector)
deactivated
Description
0
no acknowledgement
1
acknowledgement only possible with the limit comparator inactive
2
acknowledgement always possible
For setting with acknowledgement, the limit comparator is latching, which means, it remains „ON“ even when the switch-on condition is no longer present. The limit comparator must be reset by pressing the
EXIT
( d + ) keys or a binary signal.
EXIT
F
F
after an adjustable time period (time in seconds).
Input variable for limit comparator (see limit comparator functions lk1...lk8: Measurement) v „Analog selector“, page 34
Setpoint value for limit comparator (see limit comparator functions lk1...lk6: setpoint value w) v „Analog selector“, page 34
Hysteresis function
(Setup)
When LCSETVAL is deactivated, parameter FIXLCVAL can be used to enter a fixed default setpoint value.
symmetrical
Switching differential ranging around the
limit value asymmetrical left
Adjustment only possible in the setup
program:
asymmetrical
-> Limit comparators -> 1 ... 4
right Factory settings are shown
bold
.
48
7 Configuration
Switching action
Switching action means: limit comparator reaction to a limit value or setpoint value change as well as to Power ON.
„absolute“ switching action:
At the time of the change, the limit comparator reacts according to its function.
„relative“ switching action:
Following Power ON, the limit comparator remains in its „OFF“ switch position, even if the process value is within the switch-on range.
If the setpoint value or the limit value is altered while the limit comparator is in its „OFF“ position, which leads to the actual value now being in the switch-on range, the limit comparator still remains in the „OFF“ switch position.
The limit comparator will only resume operation according to its function when the process value is outside of the switch-on range. In other words: it remains „OFF“ until the process value has again reached the switch-on range.
See the following example:
49
7 Configuration
Example of the switching action „relative“:
Monitoring the actual value x with function lk4, change of setpoint value w
w
1
2
a) Start situation: Limit comparator „OFF“, as actual value x not in the switch-on range (grey area).
AL
ON
OFF
w= x
1
b) Situation at the time of the setpoint value change: Limit comparator remains „OFF“, although the actual value is now in the switch-on range.
AL
ON
OFF
w
x
2
c) Actual value has left the switch-on range: Limit comparator operates according to its function again, which means it remains „OFF“ until the actual value has reached the switch-on range again.
AL
ON
50
OFF
w= x
2

7.3 Outputs „OUTPUT“

7 Configuration
Configuration
Analog inputs Limit
comparators
Outputs
Binary functions
Display / Operation
Interfaces
Numbering of the outputs
Configuration of the instrument outputs is subdivided into analog outputs (OUTANALG; max. 2) and binary outputs (OUTLOGIC; max. 10). Binary outputs are relays, solid-state relays and logic outputs. Display and numbering of the outputs depends on the assignment of the option slots.
The switching states of the binary outputs 1...4 are shown in the display (K1…K4).
Standard for all instrument versions: (Binary) output 1 = relay (Binary) output 2 = relay (Binary) output 3 = logic output (Binary) output 4 = logic output
Extended numbering of the option slots:
Option slot
Plug-in board with 1 analog output
Plug-in board with 1 binary output (relay or
Plug-in board with 2 binary outputs
(2 relays) solid-state relay)
Option 1Output 5 Output 5 Output 5+8
Option 2Output 6 Output 6 Output 6+9
Option 3Output 7 Output 7 Output 7+10
51
7 Configuration
OUTLOGIC (binary outputs)
Binary output 1
... Binary
output 10
Inversion
Parameter Value/
Selection
OUTPUT 1
... OUTPUT10
(Setup) active
inactive
Description
0
no function
1
Binary input 1
2
Binary input 2
3
Binary input 3
4
Binary input 4
5
Binary input 5
6
Binary input 6
7
Binary input 7
8
Binary input 8
9
Limit comparator 1
10
Limit comparator 2
11
Limit comparator 3
12
Limit comparator 4
13
Logic formula 1
14
Logic formula 2
15
Binary marker
16
(reserved)
17
(reserved)
18
(reserved)
Function inverted Function not inverted
Inversion also affects function „Deactivated“, i. e. the output is always activated!
Adjustment only possible in the setup program:
-> Outputs -> Binary outputs
Factory settings are shown
bold
.
52
7 Configuration
OUTANALG (analog outputs) ➔ Output 5 ➔
Output 6 ➔ ➔ Output 7
Parameter Value/
Selection
Function FUNCTION
Type of signal SIGNAL
Range error RANG ERR0…101 Output signal (in % of the value range) for
Scale low point SCAL-LOW
Scale high point
SCAL-HI -19999…
(analog selector)
deactivated
-19999…
0
99999
100
99999
Description
Function of the output v „Analog selector“, page 34
0
0...10V
1
2...10V
2
0...20mA
3
4...20mA
Physical output signal
an overrange or underrange. 101=last output signal
A value range of the output variable is assigned to a physical output signal. The ex-factory setting corresponds to an output variable with a value range of 0...100.
Example: Via an analog output (0 … 20mA), a temperature ranging between 150 … 500°C is to be output. i.e.: 150 … 500°C 0…20mA Scale low point: 150 / Scale high point: 500
Offset
(Setup) -19999…
0
99999
Factory settings are shown
The offset is used to correct the output signal by a certain amount upwards or downwards.
Examples: Original Output value Offset value
294.7 +0.3 295.0
295.3 - 0.3 295.0
To enter digits after the decimal point, the value of system point must be set accordingly (see page 57).
Adjustment only possible in the setup program:
-> Outputs -> Analog outputs
bold
.
53
7 Configuration

7.4 Binary functions „BINFUNCT“

Configuration
Analog inputs Limit
comparators Outputs
Binary functions
Display / Operation
Interfaces
Switching action
Binary signals of binary inputs, limit comparators and logics can be assigned functions.
Potential-free contact
ON
Switch-on edge
OFF
or switching pulse
ON - contact closed OFF - contact open
t
The following binary functions react to switch-on edges:
- Acknowledge limit comparator
- Reset min/max value
- Tare function
- Reset tare function
- Go to the next scroll parameter
All remaining binary functions react to switch-on or switch-off states.
54
7 Configuration
Binary input 1
... Binary input 8
Limit comparator 1
... Limit
comparator 4
Logic 1
Logic 2
Parameter Value/
Selection
B-FUNCT1
B-FUNCT8
LCFUNCT1
LCFUNCT4
L-FUNCT1
L-FUNCT2
Description
0
no function
1
Key inhibit
2
Level inhibit
3
Display off (keys inactive)
4
Acknowledge limit comparator
5
Hold function
6
Reset min/max value
7
Tare function
8
Reset tare function
9
Tex t disp la y
10
Go to the next scroll parameter
11
Colour change
Level inhibit: The configuration level is inhibited.
Tare function: The tare function is used to zero the display value of the analog inputs and values (math) derived from these inputs. The function is reset after Power ON.
Further functions via setup program
Text display: With the binary function active, a configurable text appears on the lower display: This text can be uniquely defined
program)
Factory settings are shown
.
(only through the setup
bold
.
Several binary functions can be combined with each other in the setup program. The text display can be configured as an information or as an alarm text (with colour change).
No information or alarm texts are shown when
*
the instrument is in the USER, OPERATOR or CONFIGURATION level.
55
7 Configuration

7.5 Display / Operation „DISPLAY“

Configuration
Analog inputs Limit
comparators Outputs Binary
functions
Display / Operation
Interfaces
Display 1 (upper display)
Display 2 (lower display)
The values to be shown, the type of presentation (e.g. text, pseudo bargraph) and the display settings (e.g. colour, brightness) can be configured under this menu item.
Furthermore, start delay after Power ON, operation time-out, level inhibit and function key assignment can be defined here.
Parameter Value/
Selection
DISPLAY1 (analog
selector)
Analog input 1
DISPLAY2 (analog
selector)
deactivated
Description
Display value for the upper display v „Analog selector“, page 34
Display value for the lower display v „Analog selector“, page 34
Display type (lower display)
Display colour (lower display)
DISPTYPE
COLOUR
0
Value
1
Channel name
2
Process display text
3
Unit and value display
4
Pseudo bargraph display
Channel name (max. 8 characters), process display text (max. 24 characters), unit (max. 2 characters) as well as bargraph scaling can only be entered through the setup program.
For better legibility we recommend the exclusive use of capitals, numbers as well as the following special characters: ° % / \ ( ) + - < > _ | , Enter a space at the end of text comprising more than 8 and less than 24 characters.
01green
red
Factory settings are shown
bold
.
56
7 Configuration
Ticker time (ticker)
Decimal point (system point)
Parameter Value/
Selection
TICKER-t
DECPOINT
Description
0
100ms (fast ticker)
1
200ms
2
300ms
3
400ms
4
500ms
5
600ms
6
700ms
7
800ms
8
900ms
9
1000ms
10
1100ms (slow ticker)
0
no digit after the decimal point
1
one digit after the decimal point
2
two digits after the decimal point
3
three digits after the decimal point
If the value to be displayed cannot be shown including the programmed decimal point, the number of digits after the decimal point are automatically reduced. If subsequently the measured value contains less digits, the reading appears with the decimal point as programmed.
Brightness
Time-out
Start delay time
Min/max mode
Hold (Value)
BRIGNESS0…5 (bright) 0...5 (dark)
TIMEOUT 0…
START-t
MIN-MAX
(Setup)
180
255
0
…3600 Start delay time in seconds after Power ON
active
not active
Time period in seconds, after which the instrument automatically returns to normal display if no key is pressed.
0
Min/max mode inactive
1
Min/max mode active for analog input 1
2
Min/max mode active for analog input 2
3
Min/max mode active for analog input 1 and 2 Hold mode for analog input 1 or 2
With the hold mode active, the current measurement can be saved with function key „F“ or the binary function. The saved value can be shown in display 1 or 2 as well as in the scroll mode.
Adjustment only possible in the setup program:
-> Display/Operation -> Display -> Min-Max/ Hold
Factory settings are shown
bold
.
57
7 Configuration
Scroll time
Function key „F“
Parameter Value/
Selection
SCROLL-t
F-KEY
0
…255 Scroll mode change-over time in seconds;
Description
0 = scroll mode inactive
255 = scroll mode stop
With the scroll mode active, keys
i and d can be used to select the next or
the previous scroll parameter. If the scroll mode was stopped, further actions are only possible with this key.
Adjustment of the scroll parameters only possible in the setup program:
-> Display/Operation -> Display > Scroll mode
The parameter names are shown in the lower display. Example: INPUT1 = Channel name, analog input 1 MIN INP1 = Min. value, analog input 1 MAX INP1 = Max. value, analog input 1 HOLD1 = Hold value, analog input 1
0
no function
1
Apply hold value
2
Tare function
3
Reset tare function
4
Reset min.-max value
5
Scroll mode stop
6
LK acknowledgement
Level inhibit
(Setup)
Keep the function key pressed for at least 2 seconds to ensure that the function will be performed.
none
Configura­tion level
Factory settings are shown
Access to the configuration level can be inhibited. The setting is independent of binary function „Level inhibit“.
Setting in the setup program:
-> Display/Operation -> Operation
See also Chapter 5.3 „Level inhibit“.
bold
.
58
7 Configuration
Bargraph scaling
Channel name
Parameter Value/
Selection
(Setup) -19999
... 0 ... +99999
-19999 ...
100
+99999
(Setup)
INPUT1 INPUT2 MATHE1 MATHE2
xxxx xxx.x xx.xx x.xxx
System point
Description
Scaling start
Scaling end
...
Adjustment only possible in the setup program:
-> Display/Operation -> Display > Lower display
Channel name for analog input 1 Channel name for analog input 2 Channel name for math 1 Channel name for math 2
no digit after the decimal point one digit after the decimal point two digits after the decimal point three digits after the decimal point
Digit after the decimal point as system point
User data
Individual channel names (max. 8 characters) can be allocated for the analog inputs and math functions.
The decimal point of the values of the analog inputs can be defined different to that of the system point.
Adjustment only possible in the setup program:
-> Display/Operation -> Display > Channel name
(The setting at the instrument is made in the menu for analog input, parameter „DECPOINT“.)
(Setup) A maximum of eight parameters from the
configuration level can be defined to be available in the user level of the instrument. The parameter name (max. 8 characters) can be user-defined. Without a user-defined entry, the name programmed in the instrument will appear.
Adjustment only possible in the setup program:
-> Display/Operation -> User data
Factory settings are shown
bold
.
59
7 Configuration

7.6 Interfaces „INTERFCE“

Configuration
Analog inputs Limit
comparators Outputs Binary
functions Display /
Operation
Interfaces
Protocol PROTOCOL
Baud rate
The interface parameters for the RS422/485 or PROFIBUS-DP interface have to be configured in order to communicate with PCs, bus systems and peripheral devices.
RS422485 (Modbus)
Parameter Value/
BAUD RATE
Selection
Description
01Modbus
Modbus integer
0
9600 bps
1
19200 bps
2
38400 bps
Data format
Device address
Min. response time
DFORMAT
ADDRESS 0…1…
255
(Setup)
0
…500 Time period in milli-seconds that elapses
Factory settings are shown
0
8 data bits, 1 stop bit, no parity
1
8 data bits, 1 stop bit, odd parity
2
8 data bits, 1 stop bit, even parity
3
8 data bits, 2 stop bits, no parity Address in data network
between the request of a device in the data network and the response of the display instrument.
Adjustment only possible in the setup program:
-> Interfaces -> RS422/RS485
bold
.
60
PROFIBUS (PROFIBUS-DP)
7 Configuration
Parameter Value/
Protocol PROTOCOL
Device address ADDRESS
Analog marker
(analog value)
Binary marker
(binary value)
ANA-VAL
BIN-VAL
For further information, please refer to the
v
separate interface descriptions:
- B70.1550.2.0 (Modbus)
Description
Selection
0
Intel
1
Motorola
2
Intel integer
0…
125
Address in data network
255
-19999…
0
99999
0
…255 Binary value
Factory settings are shown
Analog value
bold
.
- B70.1550.2.3 (PROFIBUS-DP)
61
7 Configuration
62

8.1 Math and logic module

Prerequisite: The „Math“ extra code must be enabled.
vSetup program (Extras -> Enable extra codes)
The Setup program can be used to carry out two mathematical calculations or logical combinations of various signals and process variables from the controller. The formula is created by means of a formula editor.
vSetup program (Math/Logic)
With math formulae, the calculated result is presented through the two signals „Math 1“ and „Math 2“ of the analog selector. With logic formulae, the result of the

8 Extra codes

Entering formulae
logical combinations is available through the signals „Logic 1“ and „Logic 2“ of the binary selector when configuring the binary functions.
vChapter 7.4 „Binary functions „BINFUNCT““
- The string of characters in the formula consists of ASCII characters. It can have a maximum length of 60 characters.
- The formula can only be entered in the setup program.
- Formulae can be freely entered according to the standard mathematical rules.
- In the string of characters of the formula, spaces can be inserted as required. Spaces are not permitted within function designations, variables names and constants.
63
8 Extra codes

8.2 Difference, humidity or ratio calculation

The controller can be configured through the Setup program such that a difference, humidity or ratio
calculation is carried out by means of a default formula. Analog input 2 must be available. The functions need not be enabled.
vSetup program (Math/Logic)
Difference The difference of the measurements is formed from
analog input 1 (E1) and analog input 2 (E2). Difference: E1-E2
Humidity The relative humidity is determined by means of a
psychrometric humidity sensor, through the mathematical combination of the wet bulb and dry bulb temperature.
Relative humidity: (E1, E2) E1 - dry bulb temperature via analog input 1
E2 - wet bulb temperature via analog input 2
Ratio The math module forms the ratio of the
measurements from analog input (E1) and analog input 2 (E2).
Ratio: E1/E2
Result The result is under „Math 1“ or „Math 2“ and can be
used as analog value for various parameters.
64
vAnalog selector, Page 34
Safety notes
A

9 Retrofitting of modules

Only qualified personnel are permitted to carry out module retrofits.
Removing the controller module
Identifying the module
E
Risk of damage to the modules by electrostatic discharge. For this reason, avoid electrostatic charge during fitting and removal. Carry out retrofitting on a grounded workbench.
hPress together the knurled
surfaces on the front panel (left and right), and pull out the controller module.
hIdentify the module by the sales number pasted on
the packaging
Modules Code Sales
No.
View of boards
Analog input 2 1 70/00442785
1 relay (changeover)
2 relays (make, N/O)
1 analog output 4 70/00442788
2 binary inputs 5 70/00442789
2 70/00442786
3 70/00442787
65
9 Retrofitting of modules
Retrofitting of modules
Modules Code Sales
No.
1 solid-state relay 230V/1A
Interface RS422/485
PROFIBUS-DP 8 70/00442791
6 70/00442790
7 70/00442782
hSelect slot for the option
1
2
3
View of boards
66
hPush the module into the slot until the plug
connector engages
hPush the module into the housing until the lugs
engage in their slots

10.1 Technical data

Thermocouple input

10 Appendix

Designation Measuring range Measuring
accuracy
Fe-Con „L“ Fe-Con „J“ Cu-Con „U“ Cu-Con „T“ NiCr-Ni „K“ NiCr-Con „E“ NiCrSi-NiSi „N“ Pt10Rh-Pt „S“ Pt13Rh-Pt „R“ Pt30Rh-Pt6Rh „B“ W5Re-W26Re „C“ W3Re-W25Re „D“ W3Re-W26Re Chromel-Copel
Cold junction Pt 100 internal
1
incl. measuring accuracy at the cold junction
2
in the range from 300…1820°C
EN 60584
EN 60584 EN 60584 EN 60584 EN 60584 EN 60584 EN 60584 EN 60584
GOST R
8.585-2001
-200 to +900°C
-200 to +1200°C
-200 to +600°C
-200 to +400°C
-200 to +1372°C
-200 to +1000°C
-100 to +1300°C 0 to +1768°C 0 to +1768°C 0 to +1820°C 0 to +2320°C 0 to +2495°C 0 to +2400°C
-200 to +800°C
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
RTD temperature probe input
2
1,3
Ambient temperature error
100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K 100ppm/K
Designation Connection
circuit
Pt100 DIN EN
60751
Pt500 DIN EN
60751
Pt1000 DIN EN
60751
Pt50 GOST
6651-94
Pt100 GOST
6651-94
Cu50 GOST
6651-94
Cu100 GOST
6651-94
KTY11-6 2-wire -50 to +150°C
3
The accuracy refers to the max. measurement range span. The linearization accuracy is
reduced with short spans.
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
Measuring range
-200 to +850°C
-200 to +850°C
-200 to +850°C
-200 to +850°C
-200 to +850°C
-50 to +200°C
-50 to +200°C
Measuring accuracy
3-/4­wire
0.05%≤ 0.4% 50ppm/K
0.2%
0.1%
0.1%
0.05%≤ 0.4% 50ppm/K
0.2%
0.1%
−≤
3
2­wire
0.4% 100ppm/K
0.2% 50ppm/K
0.8% 50ppm/K
1.6% 50ppm/K
0.8% 50ppm/K
2.0% 50ppm/K
Ambient tempera­ture error
67
10 Appendix
RTD temperature probe input (continued)
Sensor lead resistance
Measuring current approx. 250µA Lead compensation Not required for 3-wire and 4-wire circuit. For a 2-wire circuit, the
max. 30Ω per lead for 3-wire/4-wire circuit
lead resistance can be compensated in the software by correcting the actual value.
Standard signals input
Designation Measuring range Measuring
accuracy
Voltage 0(2)—10V
0—1V Input resistance RIN > 100k
Current 0(4)—20mA
Voltage drop ≤ 1.5V
Resistance transmitter min. 100Ω, max. 4k
3
The accuracy refers to the max. measurement range span. The linearization accuracy is
reduced with short spans.
Ω±
Ω
0.05%
0.05%
0.05% 100ppm/K
4
Ω
Ambient
3
temperature error
100ppm/K 100ppm/K
100ppm/K
Binary inputs
Floating contacts open = inactive; short-circuited to GND = active
Measuring circuit monitoring
In the event of a fault, the outputs change to defined statuses (configurable).
Sensor Measuring overrange /
underrange
Thermocouple - • RTD temperature probe • Voltage 2—10V
0—10V 0—1V
Current 4—20mA
0—20mA
Resistance transmitter - -
• = detected - = not detected
Probe or lead short-circuit
-
-
-
Probe or lead break
-
-
-
68
10 Appendix
Outputs
Relay (change-over)
Contact rating Contact life
Relay (changeover (option))
Contact rating Contact life
Relay (n.o. make (option))
Contact rating
Contact life Logic output 0/12V / 25mA max. (sum of all output currents) Solid-state relay (option)
Contact rating
Protection circuitry Voltage (option)
Output signals
Load resistance
Accuracy Current (option)
Output signals
Load resistance
Accuracy Voltage supply for
2-wire transmitter
350,000 operations at rated load/750,000 operations at 1A
100,000 operations at rated load/350,000 operations at 3A
350,000 operations at rated load/900,000 operations at 1A
15.8—15.2V / 30—50mA (no-load voltage approx. 25V)
5A at 230VAC resistive load
8A at 230VAC resistive load
3A at 230VAC resistive load
1A at 230V
Varistor
0—10V / 2—10V
R
0—20mA / 4—20mA
R
electrically isolated, not stabilised
Load
Load
500
0.5%
≤ 500
0.5%
Ω
Ω
A/D converter
Resolution dynamic up to 16 Bit Sampling cycle time 50ms, 90ms, 150ms, 250ms (configurable)
Display
Type LCD with background lighting Display 1 7-segment display, 18mm high, 5 digits, color: red Function of display 1 measurement display and parameter setting Display 2 16-segment display, 7mm high, 8 digits, color: red/green
(switchable)
Function of display 2 24-character running text display (alarms), display of
measurements or parameter names
Display 3 4 switching status indicators (K1 to K4), 3mm high
69
10 Appendix
Electrical data
Supply voltage (switch­mode PSU)
Electrical safety acc. to EN 61010, part 1
Power consumption max. 13VA Data backup EEPROM Electrical connection at the back via screw terminals,
Installation information on conductor cross-sections and core-end ferrules
Without core-end ferrule
Core-end ferrule without lip
Core end ferrule with lip up to 1.5mm2
Core end ferrule with lip above 1.5mm2
AC 110—240V -15/+10%, 48—63Hz
AC/DC 20—30V, 48—63Hz
Overvoltage category III, pollution degree 2
conductor cross section up to max. 2.5mm2
with core-end ferrule (length: 10mm)
min. cross­section
0.34mm22.5mm210mm
0.25mm 2.5mm210mm
0.25mm21.5mm210mm
1.5mm
max. cross­section
2
2.5mm212mm
Min. length of core-end ferrule
(stripped)
Electromagnetic compatibility
Interference emission
Interference immunity
Twin ferrule with lip
meeting industrial requirements
0.25mm21.5mm212mm
EN 61326-1
Class B
70
10 Appendix
Housing
Housing type Plastic housing for panel mounting acc. to IEC 61554 Depth behind panel 90mm Ambient/storage
temperature range Climatic conditions rel. humidity ≤ 90% annual average, no condensation Operating position horizontal Enclosure protection acc. to EN 60529, front IP 65, back IP 20 Weight (fully equipped) approx. 380g
Interface
Modbus
Interface type RS422/RS485 Protocol Modbus, Modbus-integer Baud rate 9600, 19200, 38400 Device address 0—255 Max. number of
nodes
0 to 55°C / -30 to +70°C
32
PROFIBUS-DP
Device address 0—255
Approvals/approval marks
Approval mark Testing agency Certificate/
certification number
c UL us Underwriters
Laboratories
E 201387 UL 61010-1
Test basis valid for
CAN/CSA-C22.2 No. 61010-1
JUMO di 308 Type 701550/...
71
10 Appendix

10.2 Alarm messages

Display Cause Fault remedy
-19999 (blinking!)
99999 (blinking!)
all displays on
PROF-ERR PROFIBUS error Can be suppressed by setting the
OPT-ERR Hardware configuration
Underrange for the value being displayed.
Overrange for the value being displayed.
Watchdog or power ON trigger initialization (reset).
error
(test/repair/replace)
Is the medium being measured within the range (too hot? too cold?)
Check probe for break and probe short-circuit.
Check the probe connection and the terminals.
Check cable. Replace unit if initialization takes
longer than 5s.
PROFIBUS address to "0". Check which option boards are
installed in the slots.
Overrange / underrange covers the following events:
- Probe break or short-circuit
- Measured value outside the probe measuring range
- Display overflow
72

11 Index

A
Accessories 13 Acknowledgement Action - limit comp.
48
47
Actual value - limit comp. Analog input Analog selector Analog value
35
34
61
B
Bargraph scaling 59 Baud rate Binary functions Binary output Binary value
60
54
52
61
48
Displays
27, 56
E
Electrical isolation 19 Entering formulae Entering values Extra codes
63
63
30
F
Filter time constant 37 Fine calibration Fitting in position Function - analog output Function - limit comp. Function key
37
15
42, 47
58
53
Brightness of the display
57
C
Care of the front panel 16 Channel name Configuration level Connection diagram Controls
27
Correction value KTY
59
33
20
38
D
Data format 60 Decimal point - analog input Description of the instrument Device address Difference control Dimensions Display / Operation Display colour
6061
64
15
56
56
38
7
H
Hold mode 57 Humidity control
64
Hystereses - limit comp. Hysteresis function
42, 48
I
Installation notes 17 Instrument version Interfaces
60
11
Inversion - binary output
L
Level concept 28 Level inhibit Limit comparator Limit value Linearisation
29, 58
42
47
36
47
52
Display text
56
73
11 Index
M
Mains frequency 38 Math and logic module Measurement offset Measuring range Min/max mode Modbus
60
Mounting site
53
57
15
63
36
O
Offset - analog output 53 Operation time-out Operator level Option slots Outputs
51
31
12
57
S
Sampling cycle time 38 Scaling Scope of delivery Scroll mode Sensor type Setpoint value - limit comp. Setup program Special characters Start delay time Switch status - limit comp. Switching action - binary func-
Switch-off delay Switch-on delay System point
37, 53
13 58 35
48
13
56
57
47
tion 54
47
47
57
P
PC interface 13 Process data PROFIBUS-DP Protocol
6061
Pulse time - limit comp.
32
61
48
R
Ratio control 64 Removing the controller module
16
Response time Retrofitting of modules RS422/485
60
65
60
T
Tare function 55 Temperature unit Ticker
57
Type designation
38
11
Type of signal - analog output
U
USB interface 13 User data
28, 59
V
Variable setpoint value 45
53
74
JUMO GmbH & Co. KG
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Telefax: 315-697-5867 E-mail: info@jumo.us Internet: www.jumo.us
Loading...