Operator’s Manual
Manual de operación
Manuel d’utilisation
Stratos™Compact
Compressor Nebulizer System
Model IRC 1710
Dealer: This manual MUST be given to the end user.
User:
BEFORE using this product, read this manual and
save for future reference.
For more information regarding
Invacare products, parts, and
services, please visit
www.invacare.com
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT
FIRST READING AND UNDERSTANDING THIS
MANUAL. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS,
AND INSTRUCTIONS, CONTACT INVACARE
TECHNICAL SERVICES BEFORE ATTEMPTING
TO USE THIS EQUIPMENT - OTHERWISE
SERIOUS INJURY OR PROPERTY DAMAGE MAY
RESULT.
Stratos™Compact2Part No 1130203
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 5
Signal words are used in this manual and apply to hazards or
unsafe practices which could result in personal injury or
property damage. Refer to the table below for definitions of
the signal words.
SIGNAL
MEANING
WORD
DANGER
WARNING
CAUTION
Danger indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
Warning indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
Caution indicates a potentially hazardous situation which, if
not avoided, may result in property damage.
NOTICE
The information contained in this document is subject to
change without notice.
Part No 11302035Stratos™Compact
SPECIAL NOTES
WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
Stratos™Compact6Part No 1130203
INTRODUCTION
INTRODUCTION
The Invacare Stratos Compact Compressor Nebulizer System
is a portable medical device designed to efficiently deliver
physician-prescribed medication to the bronchial lung
passages. The nebulizer system consists of the following
components:
•Compressor
•Nebulizer
•Air tubing
•Mouthpiece
•Five extra filters
Part No 11302037Stratos™Compact
TYPICAL PRODUCT PARAMETERS
TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:115 VAC, 60Hz, Single Phase
Running Current:1.5 A @ 14 psig @ 115 volts (approx.)
Sound Level:50 to 63 dBA
Power Wattage:75 watts @ 14 psig
Electric Motor:Class B in compliance with UL2111.
Foam Intake Filter:Rayon Fiber Filter
Safety Systems:High temperature compressor shutdown.
Width:5.7 inches
Height:3.9 inches
Depth:7.5 inches
Weight:3.5 Ibs
Shipping Weight:4.5 lbs
Regulatory Listing:In compliance with UL 1431 Personal
Hygiene and Healthcare Appliances and
CSA 22.2 Electromedical Equipment.
Storage Conditions:-13°F to 158°F (-25° to 70°C); 10 to 95%
RH
Operating Conditions:50°F to 104°F (10°C to 40°C); 10 to 95%
RH
Altitude:0 to 6500 ft (2000 m) above sea level
Mode of Operation:Intermittent Use (30 min. ON / 30 min.
OFF)
NOTE: Specifications and appearance are subject to change without
notice for purposes of improvement.
NOTE: Nebulization rate and particle size are dependent on the
characteristics of the nebulizer chosen.
Part No 11302039Stratos™Compact
SUPPLEMENTAL INFORMATION
SUPPLEMENTAL
INFORMATION
Compressor
If the unit shuts down frequently, you may have a low voltage
situation.
Low voltage can also be suspected when:
•The motor does not get up to full power or speed.
•Home fuses or circuit breakers activate when starting
compressor.
•Lights dim or remain dim when compressor is started.
•Other motor operated appliances fail to operate properly
or too many motor operated appliances are on same
circuit.
Extension Cords
Make certain extension cord is in good condition. Make
certain extension cord wire is NOT LESS than 18 gauge for 25
feet.
Stratos™Compact10Part No 1130203
HANDLING
HANDLING
Shipping Instructions
The shipping container has been designed to assure
protection of this device.
If this device is to be reshipped by common carrier, it should
be packed in the same carton.
Unpacking
1. Check for any obvious damage to the carton or its
contents. If damage is evident, DO NOT use. Notify
Carrier/Invacare or your local Invacare Dealer.
2. Remove all the loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Retain carton and packing materials for use in shipping or
storing the unit until use is required.
Inspection
1. Examine exterior for nicks, dents, scratches or other
damage. Inspect all components.
Part No 113020311Stratos™Compact
SECTION 1—IMPORTANT SAFEGUARDS
SECTION 1—IMPORTANT
SAFEGUARDS
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains
important information for the safe operation and use of
this product.DO NOT use this product or any available
optional equipment without first completely reading and
understanding these instructions and any additional
instructional material such as Owner’s Manuals, Service
Manuals or Instruction Sheets supplied with this product
or optional equipment. If you are unable to understand
the Dangers, Warnings, Cautions or Instructions, contact
a healthcare professional, dealer or technical personnel
before attempting to use this equipment - otherwise,
injury or damage may occur.
Federal law restricts this device to sale by or on the order of a
physician.
When using electrical products, especially when children are
present, basic safety precautions should always be followed,
including the following.
Stratos™Compact12Part No 1130203
SECTION 1—IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS BEFORE USING
DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing
to qualified service personnel.
FIRE RISK - DO NOT use this product in an oxygen enriched
environment (i.e – oxygen tent).
To Reduce the Risk of Electrocution:
ALWAYS unplug this product IMMEDIATELY after using.
DO NOT use while bathing.
DO NOT place or store product where it can fall or be pulled
into a tub or sink.
DO NOT reach for product that has fallen into water. Unplug
IMMEDIATELY.
DO NOT place in or drop into water or other liquid.
DO NOT reach for a product that has fallen into water.
Unplug IMMEDIATELY.
Part No 113020313Stratos™Compact
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
To Reduce the Risk of Burns,
Electrocution, Fire and Injury to
Persons:
This product should NEVER be left unattended when
plugged in.
Close supervision is necessary when this product is used by,
on, or near children and physically challenged individuals.
Use this product only for its intended use as described in this
manual and only with the prescribed medication. Use this
product only under physician’s direction.
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not recommended
for use with Invacare products.
NEVER operate this product with a damaged cord or plug, if
it is not working properly, if it was dropped or damaged, or if
it was dropped into water. Call qualified service personnel for
examination and repair.
Stratos™Compact14Part No 1130203
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
Keep the nebulizer tubing, cord, and unit away from heated
or hot surfaces, including space heaters, electric blankets,
stoves and similar electrical appliances.
NEVER drop or insert objects into any opening or hose.
NEVER use while sleeping or drowsy.
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air openings
may be blocked. Keep the air openings free from lint, hair and
the like. Blocked openings may cause the unit to shut down.
Contact dealer immediately.
The Stratos compressor nebulizer system uses the
surrounding air to effectively administer the prescribed
medication through the nebulizer. The user and administrator
should be aware of surroundings so as not to have
contaminated air filtered through the compressor nebulizer
system. (Examples can be air contaminated with paint or
cleaning fumes, weed killer, car exhaust, etc.)
DO NOT use the Stratos compressor nebulizer system in these
environments. Otherwise, serious bodily injury may occur.
This unit is oil-less. DO NOT lubricate.
Part No 113020315Stratos™Compact
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
This equipment is not suitable for use in the presence of a
flammable anesthetic mixture with air or with nitrous oxide.
DO NOT operate the compressor while it is still in the
carrying case.
When not administering treatment, keep out of the reach of
small children. Some of the parts of the nebulizer are small
enough for children to swallow. The tubing may present a
strangulation hazard
NEVER touch the unit with wet hands.
DO NOT pull on the power cord to unplug the unit from an
electrical outlet.
DO NOT overload wall outlets.
Pump maintenance MUST be performed by a qualified
service technician.
SAVE THESE INSTRUCTIONS
Stratos™Compact16Part No 1130203
SECTION 2—OPERATING INSTRUCTIONS
SECTION 2—OPERATING
INSTRUCTIONS
WARNING
Remove the compressor from the carrying case before
use.
DO NOT operate unit without filter.
Avoid operating this product in a dusty environment.
Otherwise, premature compressor wear could occur.
NOTE: Clean and disinfect the nebulizer and mouthpiece before
using the device for the first time. Refer to Cleaning and Disinfecting on page 22.
Setup
NOTE: For this procedure, refer to FIGURE 2.1 on page 19.
1. Place the unit on a table or other flat, stable surface. Be sure
you can easily reach the controls when you are seated.
CAUTION
Make sure the power switch is in the off “O” position.
Part No 113020317Stratos™Compact
SECTION 2—OPERATING INSTRUCTIONS
2. Remove the nebulizer kit from the carrying case.
3. Gently twist the nebulizer to separate it into two sections.
4. Be sure the pick-up funnel is in place in the bottom section
of the nebulizer.
CAUTION
DO NOT add over six ml of medication to the medication
cup.The maximum capacity of the nebulizer medication
cup is six ml.
5. Add the prescribed amount of medication to the cup of the
nebulizer using an eyedropper, pre-measured dose or
ampoule (FIGURE 2.1).
6. Gently twist the top and bottom sections together to
assemble the nebulizer. Be sure the two sections fit
securely.
Stratos™Compact18Part No 1130203
SECTION 2—OPERATING INSTRUCTIONS
Eyedropper
Nebulizer Cup
FIGURE 2.1 Adding Medication
7. Attach one end of the air tubing to the air tubing connector
located on the front of the compressor.
8. Attach the other end of the air tubing to the base of the
nebulizer.
WARNING
In the case of high humidity, condensation may form
inside the internal tubing. Run the unit for two minutes
before connecting the nebulizer to the tubing.
9. Attach the mouthpiece to the top section of the nebulizer.
NOTE: Before starting a cold unit, let the unit reach room
temperature in a heated area.
10. With the power switch off, plug the power cord into an
appropriate electrical outlet.
Part No 113020319Stratos™Compact
SECTION 2—OPERATING INSTRUCTIONS
CAUTION
This device is for intermittent use: 30 min. ON/30 min.
OFF. Failure to operate within these limits could damage
the device beyond repair.
Treatment Instructions
NOTE: For this procedure, refer to FIGURE 2.2.
1. Turn the unit on, pressing
the ON/OFF switch to the
“-” position.
2. Place mouthpiece
between your teeth, close
your mouth and tilt the
nebulizer toward your
mouth (FIGURE 2.2).
FIGURE 2.2 Using the
Nebulizer
3. As aerosol begins to flow, inhale deeply, slowly, and hold
breath for ten seconds. Then exhale slowly through the
mouthpiece.
NOTE: If you need to rest during a treatment, place the nebulizer in
the round holder located within the handle opening.
NOTE: Tighten caps on medication jars.
Stratos™Compact20Part No 1130203
SECTION 2—OPERATING INSTRUCTIONS
4. When treatment is complete, turn the unit off by setting
the switch to the “O” position and unplug the unit from
the electrical outlet.
5. Before storing supplies between treatments, clean and dry
the nebulizer and accessories. Refer to Cleaning and
Disinfecting on page 22.
Part No 113020321Stratos™Compact
SECTION 3—CLEANING AND DISINFECTING
SECTION 3—CLEANING AND
DISINFECTING
Following the cleaning instructions after each use will prevent
any remaining medication in the bottle from drying and
causing the device to nebulize ineffectively.
It is recommended that the disposable nebulizer and the
mouthpiece be thoroughly rinsed with hot tap water after
each use and cleaned with a mild detergent after the last
treatment of the day. The same pieces MUST also be
disinfected daily by soaking in a vinegar/water solution or a
commercially available disinfectant. If your physician or
respiratory therapist specifies a different cleaning procedure,
follow their instructions
Cleaning the Outer Case
WARNING
Unplug unit before cleaning.
DO NOT submerge in water to clean. Clean outer case
with damp cloth.
Stratos™Compact22Part No 1130203
SECTION 3—CLEANING AND DISINFECTING
DANGER
DO NOT remove outer case of this unit. Removing the
case creates a serious electrical shock hazard. All
disassembly and maintenance of this unit MUST be done
by a qualified technician.
Cleaning the Nebulizer
WARNING
To prevent possible risk of infection due to
contamination, follow these instructions:
• Clean the nebulizer before initial operations.
• Clean the nebulizer after each aerosol treatment.
• ALWAYS use fresh cleaning solution.
• DO NOT towel dry the nebulizer components after
cleaning.
NOTE: Your physician and/or an Invacare dealer may specify
certain cleaning procedures. If so, follow their recommendations. If
not, clean according to the following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.
Part No 113020323Stratos™Compact
SECTION 3—CLEANING AND DISINFECTING
2. Disassemble nebulizer chamber by turning
counterclockwise and separating.
3. Fill two plastic containers or bowls:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar
to three parts water).
4. Thoroughly clean the nebulizer and mouthpiece in hot
water/detergent solution.
5. Remove from the water/detergent solution and rinse with
clear hot tap water.
6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.
NOTE: If using medical disinfectant cleaners, follow
manufacturer’s instructions carefully.
8. Keep the outer surface of the tubing dust-free by wiping
regularly.
NOTE: The nebulizer air tubing does not have to be washed
internally because only filtered air passes through the tubing.
Stratos™Compact24Part No 1130203
SECTION 3—CLEANING AND DISINFECTING
Changing the Nebulizer
The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of
inactivity or when it is obstructed by dry medication, dust,
etc. The reusable nebulizer should last from six months to one
year with general use. The disposable nebulizer should last 30
days. For the nebulizer that is disposable, it is recommended
that an extra nebulizer be kept at all times
Cleaning the Compressor
CAUTION
Unplug unit before cleaning. DO NOT submerge in water
to clean.
DO NOT clean using abrasive cleaners.
1. Clean the casing of the main unit daily by using a soft cloth
moistened with water or a mild detergent.
2. Wipe the casing and immediately dry using a soft clean
cloth.
Part No 113020325Stratos™Compact
SECTION 3—CLEANING AND DISINFECTING
Cleaning the Mouthpiece
1. Boil the mouthpiece in water for a maximum of five
minutes or autoclave at a maximum of 121° C for ten
minutes.
2. Allow the mouthpiece to air dry.
3. Once dry, the mouthpiece may be stored inside the
nebulizer top.
Stratos™Compact26Part No 1130203
SECTION 4—MAINTENANCE
SECTION 4—MAINTENANCE
DANGER
DO NOT remove outer case of this unit. Removing the
case creates a serious electrical shock hazard. All
disassembly and maintenance of this unit MUST be done
by a qualified technician.
Routine Care
To keep your device in the best condition and protect the unit
from damage, follow these directions:
•DO NOT subject the device or the parts to any strong
shocks such as dropping the device on the floor.
•DO NOT store the device in extreme hot or cold
temperatures, high humidity or under direct sunlight.
Storage
1. Put the nebulizer and mouthpiece in the carrying case.
2. Coil the air tubing.
3. Wrap the AC cord in a coil.
4. Zip the bag closed.
Part No 113020327Stratos™Compact
SECTION 4—MAINTENANCE
5. Store the device in a safe area.
NOTE: Store the repackaged device in a dry area. DO NOT place
other objects on top of the repackaged device.
Nebulizer Change
The nebulizer MUST be replaced after a long period of
inactivity, or when it is obstructed by dry medication, dust,
etc. The reusable nebulizer should last from six months to one
year with general use. The disposable nebulizer should last 30
days.
Filter Change
1. The filter MUST be replaced after approximately 30 hours
of use or when it turns grey.
2. Open the filter holder to replace the filter with a new one.
CAUTION
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate unit without an air filter.
Stratos™Compact28Part No 1130203
SECTION 5—TROUBLESHOOTING
SECTION 5—TROUBLESHOOTING
PROBLEMSOLUTION
Device does not
switch ON.
Device does not
nebulize, or it
nebulizes weakly.
1. Make sure the plug is firmly fitted to the wall socket.
2. Try plugging into another wall socket that is known
to work.
3. Contact your Invacare representative.
1. Make sure that the ends of the air tubing are fitted
tightly onto the main unit and onto the nebulizer.
2. Check whether the nebulizer is empty of medication
or has too much medication (MAX six ml).
3. Check whether the nebulizer nozzle is obstructed;.
4. Check whether the funnel is inserted.
If these possible solutions DO NOT work, contact your
medical dealer.
Part No 113020329Stratos™Compact
LIMITED WARRANTY
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN
DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE
TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who
purchases this product when new and unused from Invacare
Corporation or a dealer. This warranty is not extended to any
other person or entity and is not transferable or assignable to
any subsequent purchaser or owner. Coverage under this
warranty will end upon any such subsequent sale or other
transfer of title to any other person. This warranty gives you
specific legal rights and you may also have other legal rights
which vary from state to state.
Invacare warrants this product when purchased new and unused
to be free from defects in materials and workmanship for a
period of five years from date of purchase from Invacare or a
dealer, with a copy of the seller’s invoice required for coverage
under this warranty. If within such warranty period any such
product shall be proven to Invacare’s satisfaction to be defective,
such product shall be repaired or replaced, at Invacare’s option.
This warranty does not apply to normal wear and tear or failure
to adhere to the product instructions, nor does it include
shipping charges incurred in replacement part installation or
repair of any such product. Invacare’s sole obligation and your
exclusive remedy under this warranty shall be limited to such
repair and/or replacement.
For warranty service, please contact the Invacare service
department during normal business hours at the toll-free
telephone number identified on the back cover of this manual.
Stratos™Compact30Part No 1130203