IKA EUROSTAR 200 control, EUROSTAR 200 P4 control Operating Instructions Manual

Page 1
Operating instructions EN
20000003996d
IKA® EUROSTAR 200 control
IKA® EUROSTAR 200 P4 control
EUROSTAR_102017
Page 2
2
Fig. 3
1
2
3
4
Fig. 2
Fig. 4
Fig. 7
Fig. 5 Fig. 6
H
I
J
G
G
K
L
M
T
U
Q
S
U
X
I
G
Z
H
G
W
V
RS 232
USB
Temp. Sensor
Fig. 8
Fig. 1
Q
B
M
O
N
C
L
K
A
J
I
H
G
F
E
P
D
EUROSTAR
station
Wireless Controller
(WiCo)
P
Page 3
3
Contents
Declaration of Conformity 3 Note for USA (FCC) 3 Note for Canada (IC) 3 Explication of warning symbols 4 Safety instructions 4 Correct use 4 Unpacking 4 Drive 4 Motor protection 4 Speed – normal operation 4 Speed – overload operation 7 Output shaft 7 Speed display 7
Declaration of Conformity
Source language: German
EN
Commissioning 7 Securing 7 Switching on the instrument 8 Useful information 9 Wireless Controller (WiCo) 9 Interfaces and outputs 17 Maintenance and cleaning 18 Error codes 19 Warranty 19 Accessories 19 Permitted IKA
®
stirrer tools 19
Technical data 20
Page Page
Note for USA (FCC)
This equipment complies with Part 15 of the FCC rules. Any changes or modifications not expressly approved by the Manufacturer could void the user's authority to operate the equipment. This device complies with Part 15 of the FCC rules subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept all interference received, including interference that may cause undesired operation.
NOTE:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Note for Canada (IC)
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) This device may not cause interference (2) This device must accept any interference, including interference that may cause undesired operation of the device. This device complies with Health Canada’s Safety Code 6 / IC RSS-210. The installer of this device should ensure that RF radiation is not emitted in excess of the Health Canada’s requirement.
We declare under our sole responsibility that this product corresponds to the regulations
2014/35/EU, 2006/42/EC,
2014/30/EU and 2011/65/EU and conforms with the standards or standardized documents EN 61010-1, EN 61010-2-051, EN 61326-1, EN 60529 und EN ISO 12100.
Bluetooth
®
module:
Directive: 1999/5/EC Standards:
EN 300328, EN 301489-1, EN 301489-17, EN 60950-1
Page 4
4
Explication of warning symbols
General hazard
This symbol identifies information that is of vital importance for safeguarding your health and safety. Disregarding this information can lead to health impairment and injuries.
This symbol identifies information that is of importance for the technically correct functioning of the system. Disregarding this information can result in damage to the instrument or to system components.
This symbol indicates information which is important for ensuring that the operations of the instrument are performed eciently and for using the instrument. Failure to observe this information can result in inaccurate results.
DANGER
CAUTION
WARNING
Safety instructions
• Set up the stand in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface.
• You must ensure that the stirring element is securely clamped in the chuck!
• Use stirring shaft protective equipment!
• The agitated vessels used for stirring have to be secured. Con­sider on a good stability of the entire structure.
Please pay attention to the dangerous parts of the equipment in Fig. 8.
Protect the instrument and accessories from bumping and impacting.
Check the instrument and accessories beforehand for damage each time when you use them. Do not use damaged components.
• Safe operation is only guaranteed with the accessories described in the ”Accessories” section.
• Always switch the main switch in the OFF position or disconnect the power before changing stirring element and fitting allowed accessories.
• The instrument can only be disconnected from the mains supply by pulling out the mains plug or the connector plug.
• The socket for the mains cord must be easily accessible.
• Socket must be earthed (protective ground contact).
• The voltage stated on the type plate must correspond to the mains voltage.
Please observe the permitted speed for the stirring element. Never set higher speed.
• Make certain that the unit is set at the lowest speed before com­missioning; otherwise, the unit will begin running at the speed set in last operation. Gradually increase the speed.
Pay attention when setting the speed to any imbalance of the stir­rer tools and possible spraying of the medium to be stirred.
Never operate the instrument with the stirrer
tools rotating freely. Ensure that parts of the body, hair, jewelry or items of clothing cannot be trapped by the rotating parts.
The operation of a free rotating shaft end is
dangerous. Therefore, for safety reason, only insert through the stirring tool over the upper edge of housing at standstill.
Wear your personal protective equipment in
accordance with the hazard category of the medium to be processed, otherwise there is a risk of:
- splashing of liquids
- projectile parts
-
body parts, hair, clothing and jewelry getting caught.
Read the operating instructions in full before starting up
and follow the safety instructions.
• Keep the operating instructions in a place where it can be ac­cessed by everyone.
• Ensure that only trained staff work with the instrument.
Follow the safety instructions, guidelines, occupational health
and safety and accident prevention regulations.
• Because the options for combining products, tools, stirring ves­sel, experiment and medium are nearly endless, user safety can­not be ensured simply with design requirements on the part of the product. For this reason, it may become necessary for users to take other precautionary safety measures. For example, glass device or other stirring vessels that are sensitive to mechanical stress can be damaged or shattered by an imbalance, increasing the speed too quickly or too little distance between the stirring element and the stirring vessel. Users can suffer serious injury from glass breakage or from the freely rotating stirring element.
• Uncontrolled reactions can be triggered by mixing the heated material insuciently or by the energy generated by selecting a speed that is too high. In case of these and other increased op­erational hazards, users must take additional appropriate safety precautions (e.g. shatter protection). In any case, when using critical or hazardous materials in your processes, IKA
®
recom­mends to use additional appropriate measures to ensure safety in the experiment. For example, users can implement measures that inhibit fire or explosions or comprehensive monitoring equipment. Furthermore, users must make sure that the OFF switch of the IKA
®
product can be accessed immediately, di-
rectly and without risk at any time.
If installation or positioning cannot ensure this
access at all times, an additional EMERGENCY STOP switch that can be easily accessed must be installed in the work area.
• Only process media that will not react dangerously to the extra energy produced through processing. This also applies to any ex­tra energy produced in other ways, e.g. through light irradiation.
• Do not operate the instrument in explosive atmospheres, with hazardous substances or under water.
• Process pathogenic materials only in closed vessels under a suit­able fume hood. Please contact IKA
®
application support if you
have any question.
• The instrument is not suitable for hand-held operation.
• The high torque developed by the EUROSTAR requires par- ticular care in the choice of stand, cross sleeve and anti-rotation element for the agitating vessel.
DANGER
DANGER
DANGER
DANGER
DANGER
Page 5
5
Beware of the risk of:
- flammable materials
-
glass breakage as a result of mechanical shaking power.
Reduce the speed if:
- the medium splashes out of the vessel because the speed is
too high
- the instrument is not running smoothly
- the instrument begins to move around because of dynamic
forces
- an error occurs.
Do not touch rotating parts during operation!
There may be electrostatic activity between the medium and the output shaft which could cause a direct danger.
• After an interruption in the power supply or a mechanical interrup­tion during a stirring process, the unit does not restart automatically.
It is important to note that the surfaces of the motor (cooling fins) and certain parts of the bearing may get very hot during operation.
• Never cover the ventilation slots or cooling fins on the motor or on the instrument.
• Avoid knocking and impacting on the lower end of the shaft and the chuck gear teeth. Even minor, invisible damage can lead to imbalance and uneven shaft action.
• Ensure that the stand does not start to move.
Imbalance of the output shaft, the chuck and in particular the stir­ring tools can lead to uncontrolled resonant vibrational behavior of the instrument and the whole assembly. Glass apparatus and stirrer containers can be damaged or shattered by this. It can cause injury to the operator, also can damage the rotating stirring tool. In this case exchange the stirring tool for one without imbalance or remedy the cause of the imbalance. If there is still imbalance, return it to the dealer or the manufac­turer along with a description of the fault.
• If the instrument is operated too long in overload or if the ambi­ent temperature is too high, the instrument switches off per­manently.
The machine must only be opened by trained specialists, even during repair. The instrument must be unplugged from the power supply before opening. Live parts inside the instrument may still be live for some time after unplugging from the power supply.
Covering or parts that are capable of being
removed from the unit without accessory equipment have to be reattached to the unit for safe operation in order to prevent, for example, the ingress of fluids, foreign matter, etc..
If during operation the battery pack RB 1 (re-
chargeable battery) becomes fully discharged, the device will continue to run or is shut down depending on the value settings for exceeding the time and safety speed. If the device is set so that it continues to run when the battery of the Wireless Controller (WiCo) is fully discharged, the only means of switching the station off are the "safe STOP" and the off switch!
Please note the following safety in­structions for the battery pack RB 1 (re­chargeable battery):
• Keep the battery pack out of reach of children at all times.
• Store the battery pack in a cool, dry place.
• Never throw the battery pack into a fire. Keep it away from di­rect sunlight and temperatures above 60 °C. High temperatures will damage the battery pack and render it unusable. Tempera­tures above 100 °C may cause it to explode.
• Never throw the battery pack into water or expose it to mois­ture. Water may lead to a short-circuit, causing the battery pack to explode.
Do not deform or crush the battery pack or damage it in any
other way. This can cause battery fluid to leak and/or the bat
-
tery pack to explode.
• When not in use, keep battery packs away from paperclips, coins, keys, nails, screws or other small metal objects which could cause the contacts to be bridged. Short-circuiting may result in an explosion.
• Explosion of a battery pack may release battery fluid and cause a fire.
• The lithium polymer battery pack must only be used and charged in IKA
®
products designed for use with this battery pack.
• When the battery pack is inserted it should slide in easily and without resistance. Do not force it.
• If the battery pack is removed for an extended period of time, store it in a sealed plastic bag to prevent short-circuiting due to moisture or contact with metal components.
• The operating temperature range of the battery pack is from 0 °C to +45 °C. Note that the battery pack capacity will be re­duced at temperatures below 20 °C.
Only the rechargeable battery types recommended in the technical
data may be used in the device!
Do not charge batteries that have leaked or that are discol­ored, deformed or damaged in any other way.
Disposal instructions:
• When disposing of the IKA® battery pack, please tape over the
contacts with adhesive tape to prevent short-circuiting due to moisture or contact with metal components. Short-circuiting may result in an explosion.
• Do not throw used battery packs into your household waste. Dis-
pose of them properly in accordance with statutory regulations.
End users are obliged by law to return all used disposable
and rechargeable batteries. Throwing them into the house­hold waste is prohibited. Disposable/rechargeable batteries containing harmful substances are marked with this symbol to indicate that they may not be disposed of as household waste.
• You can return used disposable and rechargeable batteries free of charge to your local authority collection site or to any battery retailer. In doing so you will be complying with statutory regula­tions and helping to protect the environment.
• Batteries must be disposed of in accordance with local and na­tional regulations.
DANGER
DANGER
DANGER
WARNING
DANGER
CAUTION
Page 6
6
Unpacking
• Unpacking:
- Please unpack the device carefully
- In the case of any damage a detailed report must be sent im­mediately (post, rail or forwarder).
• Delivery scope:
- EUROSTAR 200 control or EUROSTAR 200 P4 control stirrer with a Wireless Controller (WiCo) of the type ordered
- An operating instructions
- An extension arm
- A hexagonal socket screw
- A hexagon socket offset screw key
-
A chuck key
- A protective cover
- A warranty card
- OS 1.0 power supply unit
- USB cable micro A – micro B 2.0
- USB cable A – micro B 2.0
.
Correct use
• Use:
For mixing/stirring liquids with low and high viscosities by vari­ous stirring tools.
Intended use: Stand device (chuck points down).
• Range of use (indoor use only):
- Laboratories - Schools
- Pharmacies - Universities
• Wireless remote control:
Before using the wireless link between the Wireless Controller (WiCo) and the laboratory device, first check whether your re­gion is included in the radio communications approval for the device. If it is not, remote control can also be performed using a USB cable.
This instrument is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply net­work that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the instrument is operated with accessories that are not sup­plied or recommended by the manufacturer
- If the instrument is operated improperly or contrary to the manu­facture’s specifications
- if the instrument or the printed circuit board are modified by third parties.
Drive
The rotary knob (B, see Fig. 1) of the Wireless Controller (WiCo) allows the speed of the stirrer to be adjusted over the entire speed range.
OS 1.0 power supply unit (for Wireless Controller (WiCo)):
Adapter England
Adapter USA, China
Adapter Australia
Adapter Europe, Switzerland
Motor protection
The stirring instrument is suitable for continuous operation. The motor current is limited electronically. The instrument has an anti­stall and anti-overload system.
If a fault occurs, a safety circuit immediately switches off the mo­tor permanently through a relay in the power board. A fault occurs if the safe functioning of the instrument is compromised.
Speed – normal operation
Speed - regulated (no variation in speed):
The speed is monitored and regulated by processor controlled. The target speed is constantly compared with the actual rota­tion speed value of the output shaft and variations corrected. This guarantees a constant speed even if the viscosity of the substance being stirred changes.
Fluctuations in mains voltage within the permitted tolerance range have no effect on the quality of regulation and constancy of speed. The speed is set using the front knob (B, see Fig. 1). During normal operations the speed value on the display (C, see Fig. 1) corresponds to the speed of the output shaft in revolutions per minute (rpm).
Page 7
7
Accessories must be assembled according to the
following assem-
bly instructions (Fig. 2 to
Fig. 7).
If above conditions are met, the instru­ment is ready for operation after plugging in the mains plug.
Speed – overload operation
In order to overcome peaks in the load such as arise when a solid or viscous medium is added, the stirrer can be run for a brief pe­riod at twice its rated power. When running in the overload range (for instance if the viscosity has increased at a particular stage of the process) the speed will be reduced to a degree sucient to keep the torque on the stirrer shaft within the rated torque of the device. The speed is continually adjusted whilst the stirrer is running to correspond to the operating conditions and so that it matches the set speed as closely as possible.
Overload status 1:
The device is already running within the overload range, whereby the actual speed does not match the set speed. This condition is maintained as long as neither the motor current nor the tempera­ture exceeds their permissible limit value. This is indicated by the torque value flashing in the display. Once the load reduces to a value within the normal range again, the torque value ceases to flash.
Overload status 2:
If the device is subjected to a fluctuating load which causes the torque to be more than double the normal torque, the actual speed of the stirrer shaft is quickly reduced to zero. Message in the display: Error 4 (see “Error codes” section).
Output shaft
The clamping chuck and output shaft permit clamping IKA® recom­mend stirrer tools (see section “Permitted IKA® Stirrer tools“).
The output shaft is designed as a hollow shaft and the opening on the top side of the housing is closed with a plastic cover. When the stirrer cover is opened, it is possible for stirring shaft to be pushed out over the top edge of the housing at standstill, e.g. during changing container. (not for the EUROSTAR 200 P4 control)
For safety reason, the stirrer cover must be pushed back into the opening of the housing so that it is properly closed. This is the only way to ensure that working with the unit is safe and that media cannot enter the instrument.
Please see section “Safety instructions“!
DANGER
Speed display
The speed is set using the rotary knob (B, see Fig.1) on the front of the Wireless Controller (WiCo).
The speed is displayed directly in revolutions per minute (rpm) in the display (C, see Fig.1) of the Wireless Controller.
Commissioning
Assemble the overhead stirrer and all necessary accessories on a stable, even, non-slip surface. The EUROSTAR overhead stir
­rer must be secured to a stable stand (e.g. R 2722 or R 2723) with a cross-sleeve (e.g. R 270). The stirring vessel must always be securely fixed for safety reasons. You must also ensure that the mounting device (stand) is set up in such a way that it is not liable to topple and does not start to move during the stirring procedure.
Securing
Securing the extension arm to the overhead stirrer
Diagram (see Fig. 2) Ensure that the extension arm is fitted securely. Vibration may cause the screw to become loose. It is therefore necessary for safe use to occasionally check that the extension arm is attached securely. Tighten the hexagonal socket screw as required.
Securing the overhead stirrer to the stand
Diagram (see Fig. 3) First secure the cross sleeve (H) to the stand (I). Then secure the extension arm (J) of the overhead stirrer in the open side of the cross sleeve which is facing upwards. Once you have set the po­sition required for the stirring procedure, firmly tighten the two clamping bolts (G). Check that the stirrer is held in position firmly prior to each use and also at regular intervals. The position of the stirrer must only be adjusted when the equipment is stationary and the power supply is disconnected.
Page 8
8
Please comply with the relevant safety instructions in the "Safety instructions" section for the RB 1 battery pack!
DANGER
Switching on the instrument
Check whether the voltage given on the type plate corresponds to the available mains voltage.
The socket used must be earthed (fitted with earth contact).
If these conditions have been met, the machine is ready for opera­tion when the mains plug is plugged in. If these conditions are not met, safe operation is not guaranteed
and the machine could be damaged.
Before the first time to use the EUROSTAR stirrer, fasten the Wireless Controller (WiCo) to the station with the screw (P, see Fig. 1), so that the rechargeable battery (RB1 battery pack) in the Wireless Controller can be charged. After the device has been switched on at the main switch (A, see Fig. 1) the device name and software version appear in the display (C, see Fig. 1) of the Wireless Controller (D, see Fig. 1). After a few seconds a signal sounds, the last setting of speed is displayed (mode B).
Once the working screen is displayed, the stirrer is ready
for operation.
If the Wireless Controller is switched on and is not attached to the stirrer (the station), the green LED bar (G, see Fig. 1) and the green Bluetooth
®
LED (F, see Fig. 1) on the stirrer (station) will light up. Make sure that the speed selected is appropriate for the trial run. If in doubt, use the rotary knob (B, see Fig. 1) to select the slowest speed. To start or stop the stirrer function, press the rotary knob (B, see Fig. 1). The controls of the Wireless Controller can be disabled by pressing the button (L)
, so that no inadvertent changes are made while
the device is running (the key symbol
appears in the display).
Pressing the button (L)
a second time enables the controls
again (key symbol
disappears from the display).
In emergency the stirrer (station) can be
switched off by pressing the “safe STOP“ button (I, see Fig.1) on the front of the stirrer. In this case the LED bar (G, see Fig.1) changes color from green to red, and flashes.
A message appears in the display, advising that the EUROSTAR stirrer (station) has been forcibly switched off. To restart the stirrer, switch the main switch (A, see Fig. 1) of the EUROSTAR stirrer (station) OFF and ON.
WARNING
Securing the stirring element using the chuck
Diagram (see Fig. 4) Slide the stirring element (M) into the chuck (L). Tighten the chuck firmly using the chuck key (K). The stirring element must only be changed when the equipment is stationary and the power supply is disconnected.
Securing the stirring shaft protector
Diagram (see Fig. 5) Use a stirring shaft protector (Q), e.g. R 301, to provide protection against injury when working with the instrument. Use the bolts (U) to attach the plastic half-shell pieces to the stirrer (T), as shown in Fig. 5. The screw (S) can be used to adjust the length of the stirring shaft protector. Check that the stirring shaft protector is held in position securely prior to each use and also at regular intervals. The position of the stirring shaft protector must only be adjusted when the equipment is stationary and the power supply is disconnected.
Securing the mixing vessel to the stand
Diagram (see Fig. 7) First fix the cross sleeve (H) to the stand (I). Then attach the bracket (Z) of the strap clamp in the open side of the cross sleeve which is facing upwards according to the position of stirrer and the mixing vessel, firmly tighten the two clamping bolts (G). Clamp the mixing vessel (V) with the flexible clamping band (W) and secure the flexible band (W) by means of the clamping lever (X).
Securing the Wireless Controller (WiCo) to the stirrer
Diagram (see Fig. 1) Attach the Wireless Controller (WiCo) to the charging contact provided on the station, and bolt it to the EUROSTAR station with the screw (P).
Connection of the temperature sensor, USB and RS232 cables to the stirrer
Diagram (see Fig. 6) The USB, RS 232 or temperature sensor cable should be plugged into the appropriate port after the covers have been re
-
moved as shown in Fig. 6.
Once the EUROSTAR 200 / 200 P4 control has been connected to the PC using the USB data cable, it will then transmit information to the Windows operating system to tell it which device drivers are required. Windows will then either:
- Load the driver
- Install the driver automatically, if it is not already installed
- Prompt the user to perform a manual installation Open http://www.ika.com/ika/lws/download/usb-driver.zip
Charging the RB 1 battery pack (rechargeable battery)
The battery pack of the Wireless Controller can be charged by any of the following means:
- on the EUROSTAR station
- via a USB cable at the PC or station
- via an OS 1.0 power supply unit.
Changing the RB 1 battery pack in the Wireless Controller
Page 9
9
If the Bluetooth® function of the Wireless Controller is active, the user can use the Bluetooth
®
search button (H) to search for the Wireless Controller. A beep signal will be audible even if the Wire­less Controller is switched off.
• Speed setting:
You can use the rotary knob (B, see Fig. 1) to pre-set the required speed before you start the device. If you then press the rotary knob(B, see Fig. 1) the device starts at the speed you set. When a change is made to the speed, the rated speed is shown in the display (C, see Fig. 1). When the device is stopped, the button (K, see Fig. 1) can be used to switch between the two speed ranges (I and II). When in standby mode, the display (C, see Fig. 1) shows the set speed.
The stirrer has two different speed ranges:
Range I: low speed, high torque. Range II: high speed, low torque.
• Correct procedure for changing the speed range:
- Switch the device off at the rotary knob (B, see Fig. 1)
- Switch the speed range with button (K, see Fig. 1)
- Change the speed with the rotary knob (B, see Fig. 1)
- Switch the device on at the rotary knob (B, see Fig. 1)
- The speed can be varied at any time in operation
- The speed appears in the display (C, see Fig. 1).
Useful information
The EUROSTAR 200 / 200 P4 control stirrer is controlled via a Wireless Controller (WiCo). If the Wireless Controller is attached
to the EUROSTAR station, data exchange between the stirrer (station) and Wireless Controller is performed via the contacts (E, Q, see Fig. 1). "Home" symbol
appears on the display of the Wireless Controller. If the Wireless Controller is connected to the stirrer (station) via a USB (Universal Serial Bus) cable, the symbol appears. If the Wireless Controller is neither bolted to the EUROSTAR station nor connected to the stirrer (station) via a USB cable, the data exchange between the stirrer and Wireless Controller is performed via Bluetooth
®
. In this case the Bluetooth®
symbol
is displayed. Depending on the structure of the building, the Wireless Controller can be operated at a distance of up to 150 m from the EUROSTAR station, using the Bluetooth
®
connection.
The Wireless Controller can be attached to the stirrer (station), or can be installed in a safe place easily accessible by the user while the stirrer is running. If the Wireless Controller is attached to the EUROSTAR station, the rechargeable battery is automatically recharged via the con­tacts (Q, see Fig. 1). The rechargeable battery can also be charged via the USB port of the Wireless Controller (see “Charging the RB 1 battery pack (rechargeable battery)“ in the “Securing“ section).
If the stirrer (station) generates vibration, the
Wireless Controller must be bolted to the stir- rer using the screw (P, see Fig. 1), or positioned remotely from the stirrer (station) whilst it is running.
WARNING
Wireless Controller (WiCo)
Controls of the Wireless Controller (WiCo):
Item Name M ON/OFF button: Switching the Wireless Controller on and off L key button: Disable button and rotary / push knob N Menu button: Press it once: main menu is displayed
Press it a second time: back to the working screen
B Rotary/push knob: Navigation, selecting and changing the settings in the menu O Back button: Return to the previous menu level / reset the torque K Gear button: Change the gear in different speed/torque ranges
Note: Can be activated only when the station is in stand-by mode.
B
M
O
N
L
K
Working screen at the time of delivery:
Explanation of symbols on the working screen:
The symbols displayed change depending on the status and settings of the Wireless Controller. The screen below shows the most significant symbols on the working screen.
The start screen appears for a few seconds after the Wireless Controller is switched on. The de
­vice name and the software version are displayed. After this, the following working screen appears automatically in the display.
Note: The wireless symbol appears only when the stirrer (station) is switched on.
Bluetooth®:
This symbol means the EUROSTAR station and the Wireless Controller are communicating via Bluetooth
®
.
The symbol no longer appears if no Bluetooth
®
communication is
being performed.
Key:
This symbol means that the function of the buttons and of the rotary knob for controlling the Wireless Controller are disabled. The symbol no longer appears if the functions are enabled once again by pressing the key button a second time.
B Operating mode:
This symbol indicates the operating mode currently selected (A, B, C).
B
0
0
actual rpm
set rpm
Torque:
0 Ncm
Reset Torque
Back
Timer:
00:00:00
I
B
300
99:00:00
actual rpm
rated rpm
45.0 ºC
100 Ncm
Torque:
Temp.:
Timer:
00:00
PC
controlled
LIMIT 2000
I
Page 10
10
USB:
This symbol means the EUROSTAR station is communicating via a USB cable. The symbol no longer appears if no USB cable is being used for communicating with the station.
Home:
This symbol means that the Wireless Controller is connected to the EUROSTAR station and is communicating with the EUROSTAR station via the charging contacts.
The symbol no longer appears if the Wireless Controller is removed from the EUROSTAR station.
Battery pack (rechargeable battery):
This symbol indicates the charging status of the RB 1 battery pack within the Wireless Controller.
The charging symbol appears if the Wireless Controller
- is connected to a PC via a USB cable
- is connected to a EUROSTAR station via a USB cable
- is connected to the power supply unit OS 1.0 via a USB cable
- is connected to the EUROSTAR station via the charger contacts.
Gear ratio (speed range):
The stirrer has two different speed ranges:
Range I: low speed / high torque. Range II: high speed / low torque.
Navigation controls in
the menu
B
N O
Navigation menu and menu structure:
Navigation menu:
Navigation menu: Press the “Menu“ button (N) and turn the rotary/push knob (B) Press the "Back" button (O) or the "Menu" button (N)
Note: If you press the “Menu“ button (N), the system skips directly back to the working screen. If you press the “Back“ button (O), the system skips back to the previous display.
Menu
Tempe
Stirring
Torque Calibration
Stirring
Intermittent Mode
Torque Display
Torque Limit
Speed Limit
LIMIT Speed limit: This symbol indicates upper speed limit set for the stirrer. The setting of the EUROSTAR in delivery status is the maximum permissible speed of the stirrer.
Temperature sensor:
This symbol appears when the temperature display is active in the display.
PC
controlled
PC control:
This symbol means that either the EUROSTAR station or the Wireless Controller is connected to a computer and the stirrer is
being controlled from the computer.
Continuous mode:
This symbol indicates that the stirrer is in continuous mode and indicates the direction of rotation of the stirrer.
00:00
Intermittent mode:
This symbol indicates that the stirrer is in intermittent mode.
Press the "Menu" button (N). Select the menu by turning the rotary/push knob (B) to the right or left to select the desired menu or
sub-menu, which can then be selected by pressing the rotary/push knob.
Press or turn the rotary/push knob (B) again to select the desired menu option and edit the values or
settings, or activate/deactivate a function.
Turn the rotary/push knob (B) to "OK" and press the "Back" button (O) or "Menu" button (N) to end the
procedure and return to the previous menu or working screen.
Note: The menu option activated is highlighted yellow in the display.
Stirring
Speed Limit
Speed Limit
2000 rpm
OK
Menu
Back
Menu
Back
Back
B
0
0
actual rpm
set rpm
Torque:
0 Ncm
Reset Torque
Back
Timer:
00:00:00
I
Page 11
11
Menu structure
Stirring
Speed Limit ………………………………………………………………………… 2000 rpm EUROSTAR 200 control 530 rpm
EUROSTAR 200 P4 control
Torque Limit ……………………………………………………………..…………
200 Ncm EUROSTAR 200 control 660 Ncm EUROSTAR 200 P4 control
Torque Display ………………………………………………………………………
activated
Intermittent Mode
Run/Stop
…………………………………........ -
Interval Run Time
…..…………...
00:00 [mm:ss]
Stop Time
……………….
00:00 [mm:ss]
Torque Calibration………………………………………………………….............
-
Temperature
Probe Temperature
………………………………………………………………
-
Display
……………………………………………………………………………..
-
Timer
Set
…………………………………………………………………………………..
00:00:00 [hh:mm:ss]
Display
……………………………………………………………………………..
activated
Operating Mode
A
………………………………………………………………….........…………..
-
B
…………………………………………………………………….........………..
activated
C
…………………………………………………………………….........………..
-
Display
Torque
…………………………………………………………………….............
activated
Temperature
……………………………………………………………………...
-
Timer
…………………………………………………………………..…………..
activated
Safety
Time Out
…………………………………..…………………………...................
00:30 [mm:ss]
Safe Speed
…………………………..…………………………………................ 100 rpm EUROSTAR 200 control
50 rpm
EUROSTAR 200 P4 control
Password
……..……………………………………………………………………
000
Settings
Languages
Units
Display
Sound
Factory Settings Bluetooth Information
Version
…………………………………………
yes
Operating Mode
………………………………
yes
Safe Speed
……………………..………………
yes
Max Speed
………………………..….………..
yes
Max Torque
…………………...………………
yes
Interval Run
………………………….…………
yes
Interval Stop
………………………...…………
yes
Volume
…………………………………..…….
10%
Key Tone
………………………………..……..
-
Background
Brightness
Black
………...………....
activated
White
…………………..
-
Standard Mode
………
80%
Battery Mode
……..…..
20%
ºC
………………………………………….…… activated
ºF
……………………………………….....……
-
English
………………………………………….
activated
Deutsch
…………………………………………
-
Français
…………………………………………
-
Español
…………………………………………
-
Italiano
…………………………………………
-
日本語
…………………………………………
-
中文
………………………………………....…
-
한국의
…………………………………………
-
···
Factory settings
………………………………..……………………………………..…
activated
-
Menu
Programs
Program 1
……………………………………………………………………......
-
...
…………………………………………………………………….....................
-
Program 5
……………………………………………………………………......
-
Page 12
12
Menu (details):
Stirring:
Speed Limit:
The ”Speed Limit” menu allows the user to set the desired maxi­mum upper speed limit for the EUROSTAR 200 / 200 P4 control stirrer. The initial setting is the maximum permissible speed of the stirrer. If the user changes this setting, the Wireless Controller saves the new value for future stirring tasks. If the “Speed Limit” has been changed, then the speed can be adjusted only within the new range.
Torque Limit:
The ”Torque Limit” menu allows the user to set the desired maxi- mum deliverable torque limit. The initial setting is the maximum permissible torque of the device. If the user changes this setting, the Wireless Controller saves the new value for future stirring tasks. If the “Torque Limit“ has been changed, the stirrer can deliver an operating torque only up to the value specified as the maximum torque limit.
Note: The torque limit can be overload for about 10 seconds. This is necessary to cater for stirring tasks which require metering and feeding additives.
Torque Display:
The ”Torque Display” menu allows the user to specify that the torque is shown in the display. A tick shows that the option is activated.
Note: Pressing the "Back" button while the device is running resets the torque to 0 Ncm. The Δ symbol then appears in the display in front of the torque value.
Torque Calibration:
This menu allows the torque to be calibrated. All torque resulting from bearing friction is excluded from the torque calculation. This is performed with no stirrer element attached. A duration of 30 sec. and speed of 50 rpm are set automatically.
Note: Calibration can be performed only when a USB cable is be­ing used. For this, remove the Wireless Controller from the sta­tion (see "Interfaces and outputs" section, figure “Connection
capability: Wireless Controller to the EUROSTAR station“).
Symbol rotation
direction
Chuck rotation
direction
Graph
CW
Time
Speed
Factory setting: continuous mode
CW
Time
Speed
Function “Run / Stop“ is activated:
The run time and stop time can be set separately.
/ CW...
Run-Stop-Run / CW ...
Intermittent Mode:
Fig. 9
Page 13
13
Operating Mode:
Operating Mode A:
In this operating mode, the set speed is not saved when the current run comes to an end or the device is switched off.
Operating Mode B:
In this operating mode, the set speed is saved when the current run comes to an end or the device is switched off, and the value
can be changed.
Operating mode C:
In this operating mode, the set speed is saved when the current run comes to an end or the device is switched off, and the value cannot be changed.
Temperature:
In the ”Temperature” menu the user can specify that the tem- perature sensor is displayed on the display/working screen. A tick shows that the option is activated. The precondition for this is that a temperature sensor is connected to the EUROSTAR station. If no temperature sensor is connected, or an error is present or the temperature rises above 350 °C, three dashes are shown as the temperature value.
Note:
Please comply with the temperature measurement range of the external temperature sensor specified in the "Technical data" section. The temperature display can be in either °C or °F (see the ”Settings“ menu).
Timer:
In the ”Timer” menu the user can specify that the timer is dis­played on the display/working screen. A tick shows that the option is activated. This setting allows the user to specify the actual time for the stirring procedure. A default time can also be set for the timer. This setting allows the user to start the stirring task for a standard time. The device stops automatically after expiry of the set time, and the set time used for the stirring procedure appears in the display.
Note:
The user can stop the stirring function before expiry of the set time. In this case the countdown of the timer is interrupted.
Display:
In the “Display“ menu the user can specify what information will be displayed on the main screen.
Note: If the ”Torque”option is activated, by pressing the ”Back“ button the user can reset the current torque to 0 Ncm as a reference value. At the same time the Δ appears in front of the Ncm unit.
Programs:
Under menu "Programs", 5 user-defined speed (rpm)-time pro­files can be created. In addition, the user could define the intermit­tent mode is activated or not in the programs. If the intermittent mode is activated, the run time/stop time value from "Intermittent Mode" setting will be taken.
Note: If user need to activate the intermittent mode in one segment of a program, he should set the "Run time/Stop time" in menu option "Interval", meanwhile activate the "Run/Stop" function in menu option "Intermittent Mode" (see section "Stirring"). Once a programs has been selected, the following options are avail­able.
Start: Start the selected program upon request. When the selected program is started by pressing on menu option "Start" with
rotary/push knob, screen reminds the user to confirm
the program control. Press "OK" with
rotary/push knob to start
the program control.
Edit: Edit the selected program parameters. Start to edit the selected program parameters by pressing on menu option "Edit" with rotary/push knob. The user can edit, delete or insert one selected program segment in the program.
When user edits the program time for at least one segment, tick (
)
for respective program will appear.
Delete: Deletes the selected program. If a selected program is deleted by pressing on menu option "Delete" with rotary/push knob, all the program parameters will be emptied.
The tick (
) disappears.
Details for editing the program:
When a program is being edited, following screen appears.
In this program, user can define up to 5 seg
­ments. The selected segment is highlighted. Then, the user can edit, delete or insert a segment in this program. When the "Back" button is pressed after editing, the program is save automatically.
Edit:
When the highlighted "edit" symbol appear on the right top of the screen, the user could change the speed (rpm), time value or intermittent mode setting.
Note: The speed (rpm) value can be changed during the speed limit range of the device. The time can be set from 1 minute to 10 hours.
Delete:
When delete a highlighted segment, the settings will be emptied. The section jump to next segment.
Insert:
With the insert option, a new segment will be insert below the selected segment. The default settings for the new segment are as following: rpm: 0 Time (hh:mm): 00:01 Intermittent Mode: No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
0 00:00 No
Page 14
14
View the program in graph:
After a program is being edited, press the "Back" button to save the settings. Press the "Back" button again, the screen back to follow screen.
If no key or knob is pressed or turned for 5 seconds, pictorially respec­tive program will be shown.
Press the rotary/push knob on option "Detail", the detailed set­tings of every segment can be seen.
Turn the rotary/push knob, the details screen of every segment can be switched.
Programs
Start Edit Delete
Program 1
Program 2
Program 3
Program 4
Program 5
rpm
500
1000
Seg No.: 1
rpm: 150
Time (hh:mm): 00:10
Interm. Mode: Yes
PROGRAM 1 Detail
rpm
500
1000
PROGRAM 1 Detail
rpm
500
1000
Seg No.: 2
rpm: 150
Time (hh:mm): 00:10
Interm. Mode: No
PROGRAM 1 Detail
Page 15
15
Sample for editing the program:
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
0 00:00 No
edit
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
0 00:00 No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
1
2
3
4
5
0 00:00 No
Edit Delete Insert
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:0 0
No
edit
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:0 0
No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:0 0
No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:00
No
edit
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:10
No
edit
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:10
No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:10
Yes
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
100 00:10
Yes
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150000:10
00:00
Yes
No
Programs
Start Edit Delete
Program 1
Program 2
Program 3
Program 4
Program 5
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
1
2
3
4
5
0 00:00
No
Edit Delete Insert
Back
Programs
Start Edit Delete
Program 1
Program 2
Program 3
Program 4
Program 5
Programs
Edit
Delete
Program 1
Program 2
Program 3
Program 4
Program 5
Start
Programs
Edit Delete
Program 1
Program 2
Program 3
Program 4
Program 5
Start
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150
150
500
800
1000
00:10
00:10
00:20
00:10
00:30
Yes
No
No
No
Yes
Back
rpm
500
1000
Seg No.: 1
rpm: 150
Time (hh:mm): 00:10
Interm. Mode: Yes
PROGRAM 1 Detail
rpm
500
1000
PROGRAM 1 Detail
rpm
500
1000
Seg No.: 2
rpm: 150
Time (hh:mm): 00:10
Interm. Mode: No
PROGRAM 1 Detail
Edit speed (rpm)
Edit time (hh:mm)
Edit intermittent mode (Yes/No)
Program diagram
5 s
Note: When intermittent mode is ac­tive, dashed line is used to represent the speed (rpm).
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:10
edit
Yes
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
150 00:10
edit
No
PROGRAM 1
Seg No.
rpm Time
hh:mm
Interm. Mode
Edit Delete Insert
1
2
3
4
5
100 00:10
No
Page 16
16
Safety:
Time Out:
In the “Time Out“ menu, the user can set a time limit. This time limit applies if there is a communication failure between the EUROSTAR station and the Wireless Controller or the com
­munication range is exceeded. The EUROSTAR station will continue to run at the set speed until the set time has expired. After that the EUROSTAR stirrer (station) will continue to run at the set safe speed (see the "Safe speed" menu).
Note: The initial setting for the time limit is 30 seconds. The user can set a value of up to 60 minutes for this time limit.
If the intermittent mode is activated, the
EUROSTAR station immediately switches to run at the set safe speed, or the set speed if that is less than the safe speed.
Safe Speed:
In the “Safe Speed“ menu, the user can specify a speed that is ap­propriate and safe for the stirring task. The safe speed applies if there is a communication failure between the EUROSTAR station and the Wireless Controller or the communication range is exceeded.
Note: The initial setting of the safe speed is 100 rpm (EUROSTAR 200 control) and 50 rpm (EUROSTAR 200 P4 control), and is implemented after expiry of the time limit (see "Time out").
Password:
In the “Password“ menu, the user can protect the Wireless Control- ler settings using a password (factory setting: 000).
Settings:
Languages:
The “Languages“ option allows the user to select the desired language by turning and pressing the rotary/push knob (B). A tick indicates the language that is set for the system.
Units:
The ”Units“ option allows the user to select the desired unit for displaying the temperature. The choice is between ”°C” or ”°F” and is made by turning and pressing the rotary/push knob (B). A tick indicates the unit that is set for the system.
Display:
The ”Display“ option allows the user to change the background color and brightness of the working screen.
Sound:
The ”Sound” option allows the user to activate/deactivate the key-press sound and to set the volume.
Factory Settings:
Select the ”Factory Settings” option by turning and pressing the rotary/push knob. The system requests confirmation to recre
­ate the factory settings. Pressing the ”OK” button resets all the system settings to the original standard values set at dispatch from the factory (see "Menu structure").
Bluetooth
®
:
The ”Bluetooth
®
” option allows the user to activate/deactivate
the ”Bluetooth
®
” function. A tick shows that the option is ac-
tivated.
Information:
The ”Information” option offers the user an overview of the most important system settings of the stirrer EUROSTAR 200 /
200 P4 control.
CAUTION
Page 17
17
Interfaces and outputs
The device can be operated in “Remote” mode via an RS 232 or USB interface using the laboratory software labworldsoft
®
. The RS 232 interface at the back of the device is fitted with a 9-pole SUB-D port which can be connected to a PC. The pins are assigned serial signals. The USB port at the rear of the stirrer is used for the connection between the PC and the Wireless Controller (WiCo). The Wireless Controller also has a USB port, located on its right side. It can also be used to connect to a PC for “remote control“.
Note: Please comply with the system requirements together with the operating instructions and help section included with the software.
USB interface:
The Universal Serial Bus (USB) is a serial bus for connecting the device to the PC. Equipped with USB devices can be connected to a PC during operation (hot plugging). Connected devices and their properties are automatically recognized. Use the USB interface in conjunction with labworldsoft
®
for operation in “Remote” mode
and also to update the firmware.
Installation:
First, download the latest driver for IKA® devices with USB inter­face from:
http://www.ika.com/ika/lws/download/usb-driver.zip.
Install the driver by running the setup file. Then connect the IKA
®
device through the USB data cable to the PC. The data communi­cation is via a virtual COM port. Configuration, command syntax and commands of the virtual COM ports are as described in RS 232 interface.
Serial interface RS 232 (V24)
Configuration
- The functions of the interface connections between the stirrer ma-
chine and the automation system are chosen from the signals speci­fied in EIA standard RS 232 in accordance with DIN 66 020 Part 1.
- For the electrical characteristics of the interface and the alloca-
tion of signal status, standard RS 232 applies in accordance with DIN 66 259 Part 1.
- Transmission procedure: asynchronous character transmission in
start-stop mode.
- Type of transmission: full duplex.
- Character format: character representation in accordance with
data format in DIN 66 022 for start-stop mode. 1 start bit; 7 character bits; 1 parity bit (even); 1 stop bit.
- Transmission speed: 9600 bit/s.
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to the
computer takes place only at the computer’s request.
Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master) to the stirrer machine (Slave).
- The stirrer machine sends only at the computer’s request. Even fault indications cannot be sent spontaneously from the stirrer machine to the computer (automation system).
- Commands are transmitted in capital letters.
- Commands and parameters including successive parameters are separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data) and each re­sponse are terminated with Blank CR LF (Code: hex 0x20 hex 0x0d hex 0x20 hex 0x0A) and have a maximum length of 80 characters.
- The decimal separator in a number is a dot (Code: hex 0x2E).
The above details correspond as far as possible to the recommen­dations of the NAMUR working party (NAMUR recommendations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specific IKA
®
co­mmands serve only as low level commands for communication between the stirrer machine and the PC. With a suitable terminal or communications programme these commands can be transmit­ted directly to the stirrer equipment. The IKA
®
software package,
labworldsoft
®
, provides a convenient tool for controlling stirring equipment and collecting data under MS Windows, and includes graphical entry features, for motor speed ramps for example. The following table summarises the (NAMUR) commands under­stood by the IKA
®
control equipment.
NAMUR Commands Function
IN_NAME
Read device name
IN_PV_3
Read PT1000 value
IN_PV_4
Read current speed value
IN_PV_5
Read current torque value
IN_SP_4
Read rated speed value
IN_SP_5
Read the torque limit value
IN_SP_6
Read the speed limit value
IN_SP_8
Read the safety speed value
OUT_SP_4
Adjust the rated speed value
OUT_SP_5
Adjust the torque limit value
OUT_SP_6
Adjust the speed limit value
OUT_SP_8
Adjust the safety speed value
START_4
Start the motor
STOP_4
Stop the motor
RESET
Switch to normal operating mode
OUT_MODE_n (n= 1 or 2)
Change the speed range (I/II)
IN_MODE
Read the the speed range (I/II)
PC 1.1 Cable (Station to PC):
Required for connecting the 9-pin socket to a PC.
1 2 RxD 3 TxD 4 5 GND 6 7 RTS 8 CTS
9
1
RxD 2 TxD 3
4
GND 5
6
RTS 7 CTS 8
9
PC
1 2 3 4 5
6 7 8 9
9 8 7 6
5 4 3 2 1
Page 18
18
micro B
micro A
USB micro B
USB micro A
Connection capability: Wireless Controller to the EUROSTAR station:
Connection capability: EUROSTAR station to the computer:
A
USB micro B
USB A
9-pin RS 232 9-pin RS 232
or
micro B
A
USB micro B
USB A
micro B
Maintenance and cleaning
The instrument is maintenance-free. It is only subject to the natural wear and tear of components and their statistical failure rate.
Cleaning:
For cleaning disconnect the main plug!
Use only cleaning agents which have been approved by IKA® to clean IKA
®
instruments.
Dirt Cleaning agent
Dyes isopropyl alcohol Construction materials
water containing tenside/isopropyl alcohol
Cosmetics
water containing tenside/
isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA
®
application support.
Wear protective gloves during cleaning the instruments.
Electrical instruments may not be placed in the cleansing agent for the purpose of cleaning.
Do not allow moisture to get into the instrument when cleaning.
Before using another than the recommended method for cleaning or decontamination, the user must ascertain with IKA
®
that this
method does not destroy the instrument.
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
-
item and designation of the spare part see
www.ika.com,
spare parts diagram and spare parts list
- Software version.
Repair: Please send in instrument for repair only after it has been cleaned and is free from any materials which may consti­tute a health hazard.
For this, you should request the “Decontamination Certifi­cate” from IKA®, or use the download printout of it from the IKA
®
website www.ika.com.
If you require servicing, return the instrument in its original pack­aging. Storage packaging is not sucient. Please also use suitable transport packaging.
or
A
9-pin RS 232 9-pin RS 232
USB micro B
USB A
micro B
USB A
USB A
USB A
Page 19
19
Warranty
In accordance with IKA® warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or mainte­nance not carried out in accordance with the instructions in this operating manual.
Error codes
The fault is shown by an error message in the display (C) as following if the error occurs, e.g. Error 4.
Proceed as follows in such cases:
Switch the device switch (A) off.
Remove the stirrer tool and remove the instrument from the assembly.
Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error Cause Effect Solution Error 2 Motor current sensor fault motor off - Switch off the instrument Error 3 Temperature inside instrument is too high motor off - Switch off the instrument and allow it to cool down Error 4 Motor blockage or overload motor off - Switch off the instrument
- Decrease the load on the motor and restart again
Error 8 Speed sensor fault or overload motor off - Switch off the instrument Error 21 Safety relay fault motor off - Switch off the instrument
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
Accessories
R 2722 H-stand R 2723 Telescopic stand R 270 Cross sleeve R 271 Cross sleeve RH 5 Strap clamp R 301 Stirrer shaft protection R 301.1 Support holder PC 1.1 Cable
H 70 Extension cable H 62.51 Stainless steel sensor H 66.51 Stainless steel sensor, glass-coated RB 1 Battery pack OS 1.0 Power supply USB cable micro A – micro B 2.0 USB cable A – micro B 2.0 Protective cover
R 1342 Propeller stirrer ≤ 2000 R 1345 Propeller stirrer ≤ 800 R 1381 Propeller stirrer ≤ 2000 R 1382 Propeller stirrer ≤ 2000 R 1385 Propeller stirrer ≤ 800 R 1388 Propeller stirrer ≤ 400 R 1389 Propeller stirrer, PTFE ≤ 800 R 1311 Turbine stirrer ≤ 2000
Permitted IKA® stirrer tools
max. speed (rpm)
R 1312 Turbine stirrer ≤ 2000 R 1313 Turbine stirrer ≤ 800 R 1375 Paddle stirrer ≤ 800 R 1376 Paddle stirrer ≤ 800 R 1330 Anchor stirrer ≤ 1000 R 1331 Anchor stirrer ≤ 1000 R 1333 Anchor stirrer ≤ 800
max. speed (rpm)
Page 20
20
EUROSTAR 200 control EUROSTAR 200 P4 control
EUROSTAR station + Wireless Controller (WiCo)
Speed range I (high torque) Speed range II (high speed)
rpm 0 / 6 – 400
0 / 30 – 2000
0 / 4 – 110
0 / 16– 530 Speed adjustment Stepless Speed display TFT / Wireless Controller Speed setting accuracy rpm ± 1 Deviation – speed measurement Speed < 300 rpm: ± 3 rpm / Speed > 300 rpm: ± 1% Max. torque at stirrer shaft Speed range I
Speed range II
Ncm 200
40
660
130 Torque trend measurement yes Torque trend display yes Deviation – torque measurement Speed range I
Speed range II
Ncm ± 20
± 6
± 60
± 10 Max. stirring quantity (water) ltr 100 Max. viscosity mPas 100000 150000 Intermittent operation yes Reversible direction of rotation no Port for ext. temperature sensor yes Temperature display yes Timer function yes Permitted on time % 100 Nominal voltage VAC 230 ± 10% (EURO) / 115 ± 10% (USA) Frequency Hz 50 / 60 Max. input power W 130 134 Max. output power at stirring shaft W 84 76 Protection class according to DIN EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload yes / motor current limitation Fuse (on mains plate) A T 4 A (IKA
®
Ident. No. 2585100) Ambient temperature °C + 5 to + 40 Ambient humidity (rel.) % 80 Drive Brushless motor Clamping chuck – clamping range mm 0,5 – 10 Hollow shaft internal diameter mm 10.3 no Extension arm (Ø x L) mm 16 x 220 Housing Coated aluminium casting and thermoplastic plastic Dimensions (W x D x H), without extension arm, with Wireless Controller
mm 91 x 231 x 294 91 x 231 x 379
Weight (with extension arm and clamping chuck) kg 4.9 5.8 Operation at a terrestrial altitude m max. 2000 USB interface yes RS 232 interface yes Temperature measurement resolution K 0.1 Measurement range, temperature °C - 10 to + 350 Limiting deviation, temperature sensor PT 1000 DIN EN 60751 class A
K ≤ ± (0.15 + 0.002 x ITI)
Measurement accuracy, temperature K ± 0.5 + tolerance PT 1000 (DIN EN 60751 class A) max. communication range (dependent on the building) m 40 – 150 Dimension (W x D x H) – Wireless Controller mm 71 x 74 x 151 Weight – Wireless Controller kg 0.28 USB interface – Wireless Controller yes
RB 1 Battery pack
Voltage V 3.7 Battery capacity mAh 2000 Charging time h 4.5 Working time h 15 Battery type Lithium-polymer
Subject to technical changes!
Technical data
Page 21
IKA
®
-Werke GmbH & Co.KG
Janke & Kunkel-Str. 10 D-79219 Staufen Tel. +49 7633 831-0 Fax +49 7633 831-98 sales@ika.de
www.ika.com
Loading...