IKA C 6000 global standards, C 6000 isoperibol Operating Instructions Manual

IKA C 6000 global standards IKA C 6000 isoperibol
20000016667
C 6000_112018
Operating instructions EN
Source language: German
1: Lift 2: RFID sensor field for decomposition vessel detection 3: Decomposition vessel 4: Touch screen
4a: Interface for USB (stick and printer)
5: Mains switch
15: Vessel group (inner vessel and outer vessel) 16: Valve group, oxygen and degassing 17: Electronics 18: Vent screw 19: Valve group, water
Front
Internal structure
Device setup
1
2
3
4
Fig. 1
5
Back
6: Interfaces for PC communication (Balance/Ethernet/
USB/Sample Rack)
7: Water filter 8: Main fuses 9: Mains socket 10: Condenser/thermostat inlet (IN) 11: Condenser/thermostat outlet (OUT) 12: Discharge hose (EMPTY) 13: Oxygen supply (IN) 14: Vent (OUT)
6
7 8
9
10
11 12 13
Fig. 2
14
15
16
17
19
Fig. 3
18
4a
Fig. 4
Dangerous spots
Page
1 Declaration of conformity 06 2 Warranty 06 3 Warning symbols 06 4 Safety instructions 07 5 Correct use 09
5.1 Use 09
5.2 Area of use 09
5.3 Recommended method of operation in working mode 09
6 Useful information 10
6.1 Determining the calorific value 10
6.2 Corrections 10
6.2.1 Acid correction 10
6.3 Note on the sample 11
6.4 Complete combustion 12
6.5 Adjustment 12
6.6 Calibration 12
6.7 System properties 13
7 Transport and unpacking 13
7.1 Transport 13
7.2 Unpacking 13
7.3 Scope of delivery 13
8 Setting up and assembly
14
8.1 Place of installation 14
8.2 Assembly of the attachments 14
8.2.1 Condenser/thermostat 14
8.2.2 Water supply line 15
8.2.3 Oxygen supply 15
8.2.4 Venting hose 15
8.2.5 Mains power supply 15
8.2.6 Peripheral devices 16
8.2.7 Mains switch 16
9 Operator panel and display 17
9.1 Explanation of the display 17
9.2 Status symbols 17
10 Commissioning 18
10.1 Switching on 18
10.2 System test 18
10.3 Switching off 18
10.4 Menu structure 19
10.5 Menu details, main menu 20
10.6 Touch screen input field 20
11 Operation 21
11.1 Procedure 21
11.2 Decomposition vessel 21
11.2.1 Creating a decomposition vessel 21
11.2.2 Editing a decomposition vessel 21
11.2.3 Calibration 22
11.3 Modules (peripheral devices) 21
11.3.1 Balance 23
11.3.2 Sample rack 24
11.3.3 Printer 25
Content
We declare under our sole responsibility that the product to which this declaration relates is in conformity with directives 2014/35/EU, 2006/42/EC, 2014/30/EU and 2011/65/EU and conforms with the following standards or normative documents: EN 61010-1, EN 61010­2-051 and EN 61326-1.
1 Declaration of conformity
3 Warning symbols
11.4 Settings 26
11.4.1 Measurement settings 26
11.4.2 Creating ignition and combustion aids 27
11.4.3 Selecting the unit 27
11.4.4 Selecting the Reference combustion value 27
11.4.5 Creating users 28
11.4.6 Evaluation standard 28
11.4.7 Setting the date and time 28
11.4.8 Setting the language 28
11.4.9 Audio settings 29
11.5 Carrying out a measurement 29
11.5.1 Creating a measurement 29
11.5.2 Editing a measurement 29
11.5.3 Simulation 30
11.5.4 Starting a measurement 30
11.5.5 Measurement procedure 31
11.6 Archive 31
11.6.1 Selecting a measurement in the archive 31
11.6.2 Editing a measurement in the archive 31
11.6.3 Opening a completed measurement 32
11.6.4 Evaluating a measurement 32
11.7 Maintenance 32
11.7.1 Maintenance programs 33
11.8 Information 34
12 Maintenance and cleaning 34
12.1 Cleaning the system 34
12.1.1 Cleaning a decomposition vessel 34
12.1.2 General maintenance and cleaning 34
12.1.3 Emptying water 35
12.2 Maintenance and cleaning of water filters 35
13 Error codes 36 14 Accessories and consumables 40
14.1 Accessories 40
14.2 Consumables 40
15 Technical data 41
In accordance with IKA warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instructions in this operating manual.
2 Warranty
DANGER
CAUTION
WARNING
NOTE
Indicates an (extremely) hazardous situation, which, if not avoided, will result in death, serious injury.
Indicates a potentially hazardous situation, which, if not avoided, can result in death, serious injury.
Indicates a potentially hazardous situation, which, if not avoided, can result in injury.
Indicates practices which, if not avoided, can result in equipment damage
4 Safety instructions
General information
Please read the instruction manual in full before use and follow the safety instructions.
Keep the instruction manual in a place where it can be ac­cessed easily.
Ensure that only trained staff use the device.
Be sure to comply with all safety instructions, directives and all matters of health, safety and accident prevention in the workplace.
Wear your personal protective equipment.
Work with the device
DANGER
Oxygen as a compressed gas is ox­idising; intensively aids combus­tion; can react violently with flam­mable materials.
Observe the relevant points of danger shown in Fig. 4.
Combustion gases are hazardous to health, therefore the venting hose must be connected to a suit­able gas cleaning system or ex­traction system.
Please observe chapter “15 Tech­nical data”.
If you are burning unknown sam­ples, leave the room or keep well away from the calorimeter.
Do not use oil or grease!
The IKA C 6000 global standards/ isoperibol calorimeter must not be used for testing explosive samples.
CAUTION
Do not use distilled or deminer­alised water (due to increased dan­ger of corrosion)!
The C 6000 must be switched off when you fit peripheral devices.
CAUTION
When handling combustion sam­ples, combustion residue and aux­iliary materials, please observe the relevant safety regulations. The following materials, for example, could pose a risk: – corrosive – highly flammable – explosive – bacteriologically contaminated – toxic.
NOTE
A constant ambient temperature is an important requirement for en­suring the high measuring accura­cy of the system. Observe the con­ditions for the place of installation.
The IKA C 6000 global standards/isoperibol calorimeter sys­tem may be used only in conjunction with the decomposition vessels C 6010 or C 6012 (Chapter 15 Technical data).
Do not operate the device in explosive atmospheres, in the presence of hazardous materials or under water.
Please observe the relevant regulations when handling oxygen.
When operating with tap water/from a tap, IKA
recom­mends that you use a standard water stop valve in the water supply line.
At the end of the work period, close the main valve for the oxygen supply.
Only change the main fuse when you have unplugged the mains power supply.
Decomposition vessel
WARNING
Perform a leakage test on the de­composition vessel before each combustion process (see operating instructions C 6010/6012).
Risk of corrosion! Substances with high halogen con­tent must not be combusted in the C 6010 decomposition vessel; instead use the C 6012 decompo­sition vessel.
NOTE
Observe the operating instructions for the decomposition vessels C 6010/C 6012.
NOTE
When using stainless steel cruci­bles thoroughly check their condi­tion after each experiment. If the material gets thinner, the crucible may catch fire and damage the de­composition vessel. Crucibles must not be used for more than 25 com­bustions for safety reasons.
Once the pressure test has been performed, a release code can be entered to enable the decomposition vessel to be used for further measurements (see decomposition vessel operat­ing instructions). The warning message will then disappear.
Please observe the maximum pressure for filling with oxygen (Chapter “15 Technical data”). Check the set pressure on the pressure reducer of your oxygen supply.
Maintenance
DANGER
If the maintenance, and especially the pressure testing, is not per­formed or is performed incorrectly, there is a risk the decomposition vessel may burst or an uncon­trolled internal fire may occur at the electrodes which could burn away the seals (oxyacetylene torch effect), thus posing a risk to life and limb.
NOTE
We recommend that you send the pressure vessel to our factory for inspection, and if necessary, repair after 1000 tests or after one year or sooner depending on use.
The declaration of conformity becomes invalid if mechanical modifications are carried out to the experiment autoclaves or if tightness can no longer be guaranteed as a result of major corrosion (e.g. pitting by halogens).
Perform servicing work only when the equipment is depressurised.
Tubes and screwed joints for oxygen, and all seals on the decomposition vessel must be kept free of grease.
The condition and function of the seals must be checked and ensured by way of a leakage test.
In particular the threads on the pressure vessel and the union nut are subject to considerable stress and must therefore be checked regularly for wear.
To prolong the life of wearing parts (o-rings, seals, etc.) we recommend that you always work with a water trap in the decomposition vessel.
Contact the IKA Maintenance Department to perform the pressure test. Comply with the safety instructions in this respect.
If the appliance is not going to be in operation for a long period of time, it is advisable to completely empty the calorimeter's water circulation. Likewise, the water must be drained out before transportation.
Please observe the maximum energy input in the decomposi­tion vessel (Chapter “15 Technical data”).
Decomposition vessels are experiment autoclaves and must be tested by a technical expert after every use.
Individual use is understood here to include a series of experi­ments performed under roughly the same conditions in terms of pressure and temperature. Experiment autoclaves must be operated in special chambers.
The decomposition vessels must undergo repeated tests (in­ternal tests and pressure tests) performed by the technical expert. The frequency of these tests is to be determined by the operator on the basis of experience, type of operation and the material used in the decomposition vessel.
Pressure vessel
CAUTION
Only technical experts may perform pressure tests and maintenance work on the pressure vessel.
• National directives and laws must be observed for
operating pressure vessels!
• Anyone operating a pressure vessel must keep it in a
proper condition, operate it properly, supervise it, carry out the necessary maintenance and repair work immedi­ately and implement the safety measures required in the circumstances.
• Pressure vessels must not be used if they have defects
which could pose a risk to staff or third parties.
Permitted media
DANGER
If the burning behaviour of a material is unknown, it must be tested before combustion in the decomposition vessel (risk of explosion).
Benzoic acid may only be combusted in its pressed form! Flammable dust and powder must be first pressed. Oven­dry dust and powder such as splints, hay, straw etc. explode when combusted! Always wet these materials first!
WARNING
Highly flammable liquids with a low vapour pressure (e. g. tetramethyl dihydrogen disiloxane) must not directly touch the cotton thread!
NOTE
When burning substances containing metals, ensure that the total energy input is not exceeded.
5 Correct use
5.2 Area of use
- laboratories - schools - universities
The device is suitable for use in residential areas and all other areas.
The safety of the user cannot be ensured:
If the device is used in conjunction with accessories not supplied or recommended by the manufacturer!
5.1 Use
The IKA C 6000 global standards/isoperibol calorimeter system is used for calorific value determination of solid and liquid substances. This is done by placing a known quantity of a substance in a decomposition vessel which is surrounded by a water bath.
5.3 Recommended method of operation in working mode
NOTE
The working mode temperature should always be in the room temperature range (+/- 2 °C) for precise measurements.
1. Read the operating instructions and get to know the device.
2. Check that your peripheral devices are compatible with the calorimeter (Chapter 11.3 Modules).
3. Select an appropriate place of installation (Chapter 8.1 Place of installation) and commission the calorimeter (Chapter 10 Commissioning).
4. Select a working mode to suit the ambient temperature and your requirements. In dynamic mode in particular, too great a difference between the selected working mode temperature and room temperature has a direct effect on the measuring accuracy of the device. Set the corresponding cooling water temperature (Chapter 15 Technical data).
Room tem­pera­ture
Cooling tempera­ture
Working mode IKA C 6000 global standard
Working mode C 6000 isoperibol
22 °C
12 °C - 20 °C Operation at water connection
17 °C - 20 °C
Opera­tion with condenser
Adiabatic 22 °C Isoperibol 22 °C Dynamic 22 °C
­Isoperibol 22 °C Dynamic 22 °C
25 °C 20 °C - 23 °C
Adiabatic 25 °C Isoperibol 25 °C Dynamic 25 °C
­Isoperibol 25 °C Dynamic 25 °C
30 °C 23 °C - 27 °C
Adiabatic 30 °C Isoperibol 30 °C Dynamic 30 °C
­Isoperibol 30 °C Dynamic 30 °C
Working mode: Adiabatic
The calorimetric decomposition vessel is ignited in a vessel filled with water (inner vessel), which in turn is in an insulating cover filled with water (outer vessel). No energy exchange takes place between the inner vessel with decomposition vessel and the outer vessel. For details, please consult the relevant international standards (e.g. DIN 51900-3).
Working mode: Isoperibol:
The calorimetric decomposition vessel is ignited in a vessel filled with water (inner vessel), which in turn is in an insulating cover filled with water (outer vessel). A specified energy exchange takes place between the inner vessel with decomposition vessel and the outer vessel. For details, please consult the relevant international standards (e.g. DIN 51900-
2).
Dynamic working mode:
A quick measuring mode developed by IKA. Its procedure and results are not subject to international standards.
5. Register the decomposition vessel during initial commissioning (Chapter 11.2 Decomposition vessel).
6. Switch the device to run hot for approx. 1 hour before starting measurements. To achieve accurate measurements you need a device that is adjusted to its ambient temperature (Chapter 10 Commissioning).
7. Every decomposition vessel that you use must be calibrated in the relevant working mode (adiabatic/isoperibol/dynamic 22 °C; 25 °C; 30 °C) during commissioning. This is done by burning a calibration substance with a known calorific value - generally benzoic acid (Chapter 6.1 Determining the calorific value). For the number of calibrations and evaluation required, see the relevant standards. You can test the stability of the measurements through control calibrations at regular intervals.
8. Select a working mode. If you are working in adiabatic mode, you must carry out an adjustment at the relevant operating temperature (22 °C; 25 °C; 30 °C). Observe the Adjustment instructions (Chapter 6.5 Adjustment). An adjustment allows you to automatically determine correct internal parameters for implementing the adiabatic principle.
With the C 6000 global standard you can now carry out adiabatic, isoperibol and dynamic measurements at specified working temperatures, and with the C 6000 iso you can carry out isoperibol and dynamic measurements at the specified working temperatures. For adaptation to individual laboratory tasks, use original IKA consumables and accessories.
The calorific value of the sample can then be calculated from the resulting increase in temperature, the sample mass and the known thermal capacity of the overall system.
Intended use: Tabletop device
If the device is not used for the intended purpose as specified by the manufacturer.
If modifications are made to the device or the PCB by third parties.
6 Useful information
The decomposition vessel C 6010/C 6012 is manufactured in accordance with the directive for pressure equipment 2014/68/EU. This is indicated by the CE symbol with the ID number of the notified body. The decomposition vessel is a category III pressure device. The decomposition vessel was subjected to an EC prototype test. The declaration of conformity confirms that this decomposition vessel corresponds to the pressure device described in the EC prototype test certificate. The decomposition vessel has undergone a pressure test with the test pressure of
6.2 Corrections
Due to the nature of the system a combustion test does not just produce the combustion heat of the sample, but also heat from external energy.
This can fluctuate considerably in relation to the heat quantity of the fuel sample.
33 MPa and a leakage test with oxygen at 3 MPa. Some materials tend to explode when combusted (e.g. due to formation of peroxide), which can cause the decomposition vessel to crack. Furthermore, toxic combustion residue in the form of gases, ashes or condensation, for example, is possible in the inner wall of the decomposition vessel.
You can obtain a copy of the Directive for pressure equipment 2014/68/EU from Beuth Verlag.
6.1 Determining the calorific value
The specific calorific value of the sample is calculated from:
Weight of fuel sample
Thermal capacity of the calorimeter system (C-value)
Temperature increase of water in the calorimeter system
For complete combustion the decomposition vessel of the calorimeter system is filled with pure oxygen (quality 3.5). The pressure of the oxygen atmosphere in the decomposition vessel should be set to 30 bar (max. possible is 40 bar). To precisely determine the calorific value of a material the combustion needs to take place under specifically defined conditions. The relevant standards are based on the following assumptions:
The temperature of the fuel before combustion depends on the set start temperature between 20 °C and 30 °C.
The water contained in the fuel before combustion and the water formed whilst combusting the hydrogenous compounds of the fuel is in fluid form after combustion.
Oxidation of the air nitrogen has not take place. The gaseous products after combustion include oxygen, nitrogen, carbon dioxide, sulphur dioxide and the oxidation products of the sample.
Solid materials may form (e.g. ashes).
In many cases, however, not just the combustion products referred to in the standards are produced. In such cases the fuel sample and the combustion products must be analysed, providing data for revised calculations.
The standard calorific value is then calculated from the measured calorific value and the analysis data. The calorific value Ho is calculated from the quotient of the quantity of heat released during complete combustion of a solid or liquid fuel and the weight of the fuel sample. The aqueous compounds of the fuel must be present after combustion in liquid form.
The formula for the calorific value is: Ho = (CV * dt – Qext) / m
Ho Calorific value m Mass of the sample dt Measured and corrected increase in temperature Qext All external energy originating from the ignition wire,
the ignition aids, the combustion aids and the formation of acids
CV C value (thermal capacity) of the calorimeter
The heat value Hu is the same as the calorific value, minus the condensation energy of the water contained in the fuel and formed through combustion. The heat value is the more important parameter from a technical point of view because in all major, technical applications only the heat value can be evaluated in terms of energy. The bases of calculation for the calorific and heat value can be found in the relevant standards (e. g.: DIN 51 900; ASTM D 240; ISO 1928).
Fuel
sample
Burning
aid
Igniter
Formation of
sulphuric acid
Formation of
nitric acid
Heat quantity from:
External energy
The combustion heat of the cotton thread which ignites the sample and the electric ignition energy would distort the measurement. This influence is taken into account in the calculation by way of a correction value. Materials which are dicult to ignite or combust are combusted together with a burning aid. The burning aid is weighed and put
6.3 Note on the sample
DANGER
If you are burning unknown samples, leave the room or keep well away from the calorimeter.
WARNING
Risk of corrosion! Substances with high halogen content must not be combusted in the C 6010 decomposition vessel; instead use the C 6012 decomposition vessel.
NOTE
To prolong the life of wearing parts (o-rings, seals, etc.) we recommend that you always work with a water trap.
The IKA C 6000 global standards/isoperibol calorimeter system is a high-precision measuring instrument used for routinely determining the calorific value of solid and liquid substances. However exact measurements are only possible when the individual test steps are carried out carefully. For this reason the procedure must be followed precisely.
A few points should be noted in respect of the substances to be combusted:
Normally solid combustion substances in powder form can
be combusted directly. Materials which combust quickly (e.g. benzoic acid) must not be burnt loose. Benzoic acid may only be combusted in its pressed form! Flammable dust and powder must be first pressed. Oven-dry dust and powder such as splints, hay, straw etc. explode when combusted! Always wet these materials first! Highly flammable liquids with a low vapour pressure (e.g. tetramethyl dihydrogen disiloxane) must not directly touch the cotton thread!
Highly flammable substances tend to spray. Such substances
must be pressed into tablets before combustion. The IKA pelleting press C 21, for example, is suitable for this task.
Most fluid substances can weighed out directly into the crucible. Liquids that are cloudy or that have water that may separate must be dried or homogenised before being weighed out. The water content of these samples must be determined.
Highly volatile substances are poured into combustion capsules (gelatine capsules or acetobutyrate capsules, see Accessories) and combusted together with the capsules.
For substances that are dicult to ignite or low in calories, use the burning aid (see accessories). Before filling the capsule or the combustion bag with the substance to be determined, weigh them to calculate the extra external energy added by the burning aid from the weight and the calorific value. This must be taken into account in QExtern2. You should keep the amount of burning aid used to a minimum.
6.2.1 Acid correction
Virtually all of the materials to be studied contain sulphur and nitrogen. Under the conditions in calorimetric measurements, sulphur and nitrogen combust to SO
2
, SO3 and NOx. Together with the water from combustion and moisture, sulphuric and nitric acid as well as heat of solution are produced. In order to obtain the standard calorific value, the influence of the heat of solution on the calorific value is corrected. To achieve a specified end state and to record all acids quantitatively, distilled water or another appropriate absorbing liquid is placed in the decomposition vessel in advance before the experiment
in accordance with the applicable standards. The combustion gases form acids with this absorption liquid and the combustion water. The calibration of the system must have been performed in accordance with the instructions! After the combustion the decomposition vessel is thoroughly flushed with distilled water to collect the condensate that has been deposited in the inner wall of the vessel as well. The solution produced in this way can now be examined with appropriate peripheral devices to detect its acid content. For more detailed information contact IKA or an authorised dealer.
into the crucible with the sample. The additional heat quantity can be determined from the weight of the burning aid and its known specific calorific value. You must correct the test result by this heat quantity.
6.4 Complete combustion
NOTE
If there is any unburnt residue, the test must be repeated.
It is essential that the sample fully combusts to ensure correct determination of the calorific value. After each experiment check the crucible and all the solid residue for signs of incomplete combustion. Complete combustion is not guaranteed for materials with a tendency to spray.
6.5 Adjustment (only IKA C 6000 global standards)
NOTE
If the device is to be operated in adiabatic function, prior adjust­ment in the respective tempera­ture range (22 °C, 25 °C or 30 °C) is necessary.
Adjustment must be performed in the following cases:
When first commissioning the calorimeter and when its place
of installation has been changed.
If the measuring times for adiabatic measurements are
regularly in excess of 15-20 minutes.
If adiabatic measurements are frequently aborted because the
time limits for the pre-trial or main trial have been exceeded.
Procedure for adjustment:
In the menu select “Settings, Measurement settings, Adiabatic
working mode”.
The adjustment starting temperature is preset by selecting
the working mode/temperature range.
Start the adjustment by selecting the selection field r (e.g.
Adjustment 25 °C). Insert the decomposition vessel (without sample).
Follow the instructions.
Adjustment starts automatically and is completed within approx. 1 hour. After the adjustment has been successfully completed the adjustment value appears as an increase in temperature in the measurement record for the measurement that was performed and is automatically adopted as a system parameter.
You can see the adjustment value under the menu item “Information adjustment”.
After successful completion of adjustment the device automatically changes the mode of operation to the corresponding adiabatic mode.
Adjustment 22 a Adiabatic 22 a Adjustment 25 a Adiabatic 25 a Adjustment 30 a Adiabatic 30 a
6.6 Calibration
NOTE
Regular calibration is vital to maintaining measurement accuracy.
To guarantee accurate and reproducible measurement results, the calorimeter system is calibrated after initial start-up, after maintenance work, after parts are replaced and at specified time intervals. During calibration the thermal capacity of the calorimeter system is revised. To this end, a specified quantity of a reference substance is burned in the IKA C 6000 global standards/isoperibol under test conditions. As the calorific value of the reference substance is known, after it has been burnt it is possible to calculate the thermal capacity on the basis of the temperature increase of the calorimeter system. The reference substance for calorimetry at an international level is benzoic acid, as per the National Bureau of Standards (NBS-Standard Sample 39 J) with guaranteed calorific value. The thermal capacity is derived from the formula for the calorific value (Chapter 6.1 Determining the calorific value):
CV = (Ho * m + Qext) / dt
Depending on the standard used, determination of the thermal capacity may require performance of several measurements.
Using various statistical criteria the average value is calculated and is used as the thermal capacity for subsequent determinations of calorific values. For more detailed information on calibration, please see the relevant standards. If the IKA C 6000 global standards/isoperibol is operated with several decomposition vessels, you will need to determine the heat capacity of the system for each decomposition vessel. The parts of the decomposition vessel must not be replaced. In addition the thermal capacity depends to a small extent on the measurement procedure used. The thermal capacity must be determined for each measurement procedure that is used.
Note on calibrations The calibration must be carried out under the same conditions as the subsequent tests. If substances are used in combustion tests (e.g. distilled water or solutions), you must use exactly the same amount of this substance for calibration. For determination of calorific values the increase in temperature must be about as great as for the calibration (e.g. g.: 2 tablets = approx. 1 g benzoic acid =˜ 3 K). The optimum sample quantity must be determined by several trials where necessary.
Materials which are dicult to ignite (materials with a high mineral content, low caloric materials) can often only be fully combusted using burning aids such as combustible crucibles, combustible capsules or combustible bags (Chapter 14.2 consumables). It is also possible to use liquid burning aids such as paran oil. The ignition aids (e.g. cotton thread) must also burn completely.
6.7 System properties
CAUTION
Operation is only permitted with decomposition vessels C 6010 and C 6012.
The system has the following properties:
Reduction of routine tasks thanks to automated measuring procedure
Integral oxygen filling/degassing
Automatic vessel detection
Operation without cooling unit: Connects to tap with pressure reducer IKA C 25; temperature range 12 °C - 20 °C; water consumption per measurement approx. 4 L; max. pressure 1 bar to 1.5 bar (Chapter 15 Technical data).
Operation with active cooling unit e.g. IKA RC 2 (Chapter
5.3. Recommended method of operation in working mode)
Measurement and determining of calorific value and calculation of the heat value as per DIN (Chapter 15 Technical data).
Measuring range: max. 40,000 J (This corresponds to an increase in temperature in the decomposition vessel of approx. 5 K).
Operation possible with IKA PC-Software CalWin
®
C 6040
Can be connected to sample rack C 5020
Includes printer connection (USB, network, RS 232 (Fig. 2, 6))
USB drive connection (Fig. 1, 4a)
7 Transport und Unpacking
7.1 Transport
CAUTION
The appliance must be completely emptied before storing and transportation.
The system must be protected against mechanical impact, vibrations, dust deposits and corrosive ambient air during transportation and storage. It is also important to ensure that the relative humidity does not exceed 80 %.
7.2 Unpacking
Unpack the device carefully.
Any damage should be notified immediately to the shipping agent (post oce, railway network or transport company).
Hoses:
Emptying aid
Anschlussrohr
Entleerhilfe / Ablassschlauch
SW17SW10
Discharge hose
1.5 m (EMPTY)
Water inflow pipe (IN):
Water return pipe (OUT):
Condenser
Venting hose (OUT)
Connecting pipe O
2
(IN)
SW 8
SW 8
SW 10
7.3 Scope of delivery
Calorimeter IKA C 6000 global standards/isoperibol
Attachment set
Tool
C 60.1012 Organizer
C 6000.1 water protect
C 723 Benzoic acid
Double-end/single-end spanner
Sealing disc
Filter spanner
Screwdriver
Power supply cable
USB drive
Operating instructions
Warranty card
8 Setting up and assembly
8.1 Place of installation
NOTE
A constant ambient temperature is an important requirement for en­suring the high measuring accura­cy of the system. Observe the con­ditions for the place of installation.
No direct solar radiation
No draughts (e.g. beside windows, doors, air conditioning)
Nucient distance from radiators and other heat sources
the minimum distance between the wall and the back of the device must not be less than 25 cm.
Laboratory additions such as shelves, cable ducts, ring lines etc. must not be build above the system.
The room temperature must remain constant.
The system must be installed on a level surface.
For operation of the system the following must be available at the place of installation:
A power supply corresponding to the type plates of the system components,
An oxygen supply (99.95 % pure oxygen, quality 3.5; pressure 3 MPa) with pressure gauge.
There must be a shut-off device for the oxygen supply. Observe the instructions for oxygen (Chapter 4 Safety instructions).
8.2 Assembly of the attachments
8.2.1 Condenser/thermostat
CAUTION
Do not use distilled or demineralised water (due to increased danger of corrosion)!
In normal operation the discharge hose at the “EMPTY” connection (Fig. 2, 12) must not be inserted.
Make sure the filter casing is always securely closed.
NOTE
Observe the operating instructions for the condenser/thermostat.
Tap water of drinking quality is recommended. Mix in the water bath additive supplied (max. 1 ml for 4-5 L of water). This improves the usable life of the water.
IKA recommends that you use the system with the recirculating condenser RC 2 at a speed of 2800 1/min (place of installation downstream of the C 6000: 32001/min).
1. Insert the inflow pipe into the “IN” port until it clicks home in the “IN” inlet (Fig. 2, 10).
2. Connect the other end of the pipe to the “OUT” port of the condenser (water pressure max 1.5 bar).
Insert the return pipe into the “OUT” port until it clicks home (Fig. 2, 11) and connect the other end of the pipe to the “IN” port of the condenser. The “EMPTY” connection is intended only for emptying the device, e.g. for transport purposes (Chapter 12.1.3 Emptying water).
Loading...
+ 30 hidden pages