HP 37718A, 37719A Installation And Verification Manual

Page 1
HP OmniBER 718/719
HP 37718A HP 37719A Installation and Verification Manual
Page 2
Copyright Hewlett­Packard Ltd.1999
All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
HP Part No. 37718-90069
First Edition, May 99
Printed in U.K.
Documentation Warranty
The information contained in this document is subject to change without notice.
Hewlett-Packardmakesno warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Hewlett-Packard shall not be liable for errors contained herein or for incidental orconsequential damages in connection with the furnishing, performance, or use of this material.
WARNING
For details of safety, see Safety information in this manual.
Hewlett-Packard Limited Telecommunications Networks Test Division South Queensferry West Lothian, Scotland EH30 9TG
Page 3
Installation and Verification Manual
HP Part Number 37718-90069
Printed in U.K. May 1999
HP OmniBER 718/719
Page 4
2
Page 5
Contents
1 General Information
Warranty 1-1
Responsibilities of the Customer 1-2 Certification 1-3 Assistance 1-3 Notice 1-3 Restricted Rights Legend 1-3
Instruments Covered By Manual 1-4 Storage and Shipment 1-4
Repackaging for Shipment 1-4
Weight and Dimensions 1-5 Safety Precautions for the Operator 1-6
Lifting/Carrying the HP 37718A/19A 1-6 Safety Symbols 1-7
Operators Maintenance 1-8
Cleaning 1-8 Power Supply Fuse Replacement 1-9
Statement of Compliance 1-10
Australian EMC Regulations 1-10 Noise Declaration (German) 1-10
Electromagnetic Compatibility 1-11
Electrostatic Discharge 1-11
2 Installation
Initial Inspection 2-1 Operating Environment 2-2 Preparation for Use 2-3
Power Requirements 2-3 Fuses 2-3 Fuse Replacement 2-3
iii
Page 6
Contents
Connecting to the Power Supply 2-4
Connecting to the Network 2-6
All Connectors 2-6 Electrical Interface Connectors 2-6 Optical Interface Connectors 2-9
Connecting Accessories 2-11
10 Base-T Lan Connection Radiated Emissions 2-11 Hewlett-Packard Interface Bus 2-12
Additional Precautions for Service Engineers 2-14
Safety Precautions 2-14 ESD Precautions 2-14 Front Panel Soft Recovery (Cold Start) 2-16
3 Performance Tests
Introduction 3-1 Equipment Required 3-1 Performance Test Record 3-1 Calibration Cycle 3-1 Option Configuration 3-1 Recall Default Settings 3-4
Self Test 3-5 PDH/DSn Internal Transmitter Clock Accuracy & Offset 3-10
Specifications 3-10 Equipment Required 3-10 E1 (2.048Mb/s) Frequency Accuracy 3-11 E2 (8.448Mb/s) Frequency Accuracy 3-12 E3 (34.368Mb/s) Frequency Accuracy 3-12 E4 (139.264Mb/s) Frequency Accuracy 3-13 DS1 (1.544Mb/s) Frequency Accuracy 3-13 DS3 (44.736Mb/s) Frequency Accuracy 3-14
PDH/DSn Transmitter Output 3-15
iv
Page 7
Contents
Specification 3-15 Equipment Required 3-16 Procedure - 34.368 Mb/s 3-19 Procedure - 8.448 Mb/s 3-21 Procedure - 2.048 Mb/s 3-23 Procedure - DSX-3 3-26 Procedure - DSX-1 3-28
PDH/DSn Receiver Equalization 3-32
Specifications 3-32 Description 3-32 Equipment Required 3-32 Procedure 3-33
PDH/DSn Receiver Monitor Levels 3-39
Specifications 3-39 Description 3-39 Equipment Required 3-39
External Mux/Demux 3-44
Specifications 3-44 Description 3-44 Equipment Required 3-44
PDH/DSn Frequency Measurement and Looped Clock 3-54
Specifications 3-54 Equipment Required 3-54
SDH/SONET Transmitter Clock Accuracy 3-58
Specification 3-58 Description 3-58 Equipment Required 3-58 Procedure 3-58
External Clock/Data Reference Inputs & Clock Reference Output 3­59
Specifications 3-59
SDH/SONET Frequency Offsets 3-64
Specifications 3-64
v
Page 8
Contents
Description 3-64 Equipment Required 3-64 Procedure 3-64
STM-0/STS-1 Transmitter Output Waveshape 3-66
Specifications 3-66 Description 3-66 Equipment Required 3-66 Procedure 3-66
STM-1/STS-3 Transmitter Output Waveshape 3-70
Specifications 3-70 Description 3-70 Equipment Required 3-70 Procedure 3-70
SDH/SONET Receiver Equalization 3-75
Specifications 3-75 Equipment Required 3-75 Procedure 3-77
SDH/SONET Receiver Monitor Levels 3-78
Specifications 3-78 Description 3-78 Equipment Required 3-78 Procedure 3-78
Multirate Optical Interfaces 3-82
Specification (1310 nm Transmitter) 3-82 Specification (1550 nm Transmitter) 3-82 Specification (Receiver) 3-82 Description 3-83 Equipment Required 3-83
PDH Transmit/Receive Jitter Amplitude Accuracy 89
Specifications 89 Description 89 Equipment Required 89 Procedure 90
vi
Page 9
Contents
TX/RX Jitter Accuracy Test Setup, Ranges (1/1.6 UIp-p) 93
External Jitter Generation and Demod Output 98
SDH Transmit/Receive Jitter Amplitude Accuracy 101
Specifications 101 Description 101 Equipment Required 102 Procedure 103
External Jitter Generation and Demod Output 117
Performance Test Record 121 Jitter Performance Test Record (options 204, 205 & 206) 131
4 Terms
ETSI/ANSI Conversion and Equivalent Terms 4-1
5 Self Test Overview
Self Tests 5-1
OVERALL FUNCTION SELF TESTS 5-1 All Tests (test time - up to 1 Hour depending on options) 5-2 SPECIFIC MEASUREMENT HARDWARE/FUNCTION TESTS 5-2 PDH Test (test time- approximately 20 minutes) 5-3 BER Test (test time- approximately 20 minutes) 5-4 JITTER Test (test time- approximately 20 minutes) 5-6
6 Option Structure/Module Configuration
Option Structure/Module Configuration 6-1
vii
Page 10
Contents
viii
Page 11
General Information

Warranty

1 General Information

Warranty
HP PRODUCT DURATION OF WARRANTY HP 37718A/19A 3 years 1 HPwarrantsHP hardware,accessoriesandsuppliesagainstdefectsin
materials and workmanship for the period specified above. If HP receives notice of such defects during the warranty period, HP will, at itsoption,eitherrepairorreplaceproductswhichprovetobedefective. Replacement products may be either new or like-new.
2 HPwarrants that HP software will not fail to execute itsprogramming
instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If HP receives notice of such defects during the warranty period, HP will replace software media which does not execute its programming instructions due to such defects.
3 HP does not warrant that the operation of HP products will be
uninterrupted or error free. If HP is unable, within a reasonable time, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the product.
4 HP products may contain remanufactured parts equivalent to new in
performance or may have been subject to incidental use.
5 The warranty period begins on the date of delivery or on the date of
installation if installed by HP. If customer schedules or delays HP installation more than 30 days after delivery, warranty begins on the 31st day from delivery.
6 Warranty does not apply to defects resulting from (a) improper or
inadequatemaintenanceorcalibration,(b)software,interfacing,parts or supplies not supplied by HP, (c) unauthorized modification or misuse, (d) operation outside of the published environmental specifications for the product, or (e) improper site preparation or maintenance.
1-1
Page 12
General Information
Warranty
7 TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE
WARRANTIESAREEXCLUSIVEANDNOOTHERWARRANTYOR CONDITION,WHETHERWRITTEN OR ORAL, IS EXPRESSED OR IMPLIED AND HP SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OF MERCHANTABILITY, SATISFACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
8 HPwillbeliablefordamage to tangible property per incident up to the
greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective HP product.
9 TOTHE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN
THIS WARRANTY STATEMENT ARE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL HP OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT, TORT, OR OTHERWISE.
FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND: THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE, RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.

Responsibilities of the Customer

The customer shall provide: 1 Access to the products during the specified periods of coverage to
perform maintenance.
2 Adequateworkingspace around the products for servicing by Hewlett-
Packard personnel.
3 Access to and use of all information and facilities determined
necessary by Hewlett-Packard to service and/or maintain the products.(Insofar as these items may contain proprietary or classified information, the customer shall assume full responsibility for
1-2
Page 13
General Information
Warranty
safeguarding and protection from wrongful use.)
4 Routine operator maintenance and cleaning as specified in the
Hewlett-Packard Operating and Service Manuals.
5 Consumablessuchaspaper,disks,magnetictapes,ribbons,inks,pens,
gases, solvents, columns, syringes, lamps, septa, needles, filters, frits, fuses, seals, detector flow cell windows, etc.
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Bureau of Standards, to the extent allowed by the Bureau's calibration facility and to the calibration facilities of other International Standards Organization members!

Assistance

Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products.

Notice

The information contained in this document is subject to change without notice.
Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material.
This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied or reproduced without the prior written consent of the manufacturer, Hewlett-Packard Ltd.

Restricted Rights Legend

Use, duplication, or disclosure by the government is subject to restrictions as set forth in subdivision (b)(3)(ii) of the Rights in Technical Data and Computer Software clause at 52.227-7013. Hewlett-Packard Company; 3000 Hanover Street; Palo Alto, California 94304.
1-3
Page 14
General Information

Instruments Covered By Manual

Instruments Covered By Manual
Attached to the rear panel of the instrument is a serial number plate. The serial number plate has a two letter reference denoting country of origin (GB = Great Britain) and an eight digit serial number. The serial number is unique to each instrument and should be quoted in all correspondence with Hewlett-Packard, especially when ordering replacement parts. Refer to Chapter 6 for instrument options covered.
Serial Number Plate

Storage and Shipment

The instrument may be stored or shipped in environments within the following limits:
Temperature -20o C to +70o C
-15o C to +50o C with lid printer Altitude Up to 15,200 meters (50,000 feet) The instrument should also be protected from temperature extremes
which could cause condensation within the instrument.

Repackaging for Shipment

Tagging for Service. If the instrument is being returned to Hewlett-
Packard for service, please complete a repair tag and attach it to the instrument.
1-4
Page 15
General Information

Weight and Dimensions

Original Packaging. Containers and materials identical to those used in factory packaging are available from Hewlett-Packard offices. If the instrument is being returned to Hewlett-Packard for servicing, attach a tag indicating the type of service required, return address, model number, and full serial number. Mark the container FRAGILE to ensure careful handling. In any correspondence, refer to the instrument by model number and full serial number.
Other Packaging. The following general instructions should be followed when repackaging with commercially available materials:
Wrap instrument in heavy paper or plastic. If the instrument is being shippedto Hewlett-Packard,attacha tag indicating thetype of service required, return address, model number and full serial number.
Use a strong shipping container. A double wall carton made of 350 pound test material is adequate.
Use a layer of shock absorbing material 70 to 100 mm (3 to 4 inch) thick, around all sides of the instrument to provide firm cushioning and prevent movement inside the container. Protect the Front Panel controls and Rear Panel connectors with cardboard.
Seal shipping container securely.
Mark shipping container FRAGILE to ensure careful handling.
In any correspondence, refer to instrument by model number and full
serial number.
Weight and Dimensions
Weight: 18 kg (40 lb) fully loaded Dimensions: 190mm (7.5in) high, 340mm (14in) wide, 420mm (17in)
deep (including cover).
1-5
Page 16
General Information

Safety Precautions for the Operator

Safety Precautions for the Operator
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Hewlett-Packard Company assumes no liability for the customer’s failure to comply with these requirements.
In particular, the operator should note the following safety information:
“Safety Symbols” on page 1-7
“Connecting to the Power Supply” on page 2-4
“Operating Environment” on page 2-2
“Fuse Replacement” on page 2-3
“Operators Maintenance” on page 1-8
“Lifting/Carrying the HP 37718A/19A” on page 1-6
DO NOT operate damaged equipment: Whenever it is possible that the safety protection features built into this product have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to a Hewlett-Packard Sales and Service Office for service and repair to ensure the safety features are maintained.

Lifting/Carrying the HP 37718A/19A

Before attempting to lift or carry the instrument consider the following basic lifting techniques to help avoid personal injury.
Using both arms to lift instrument.
Reach for the instrument - bend your knees and waist, and keep your back
straight.
GRASP the instrument firmly.
LIFT with your legs.
KEEP your shoulders level.
1-6
Page 17
General Information
Safety Precautions for the Operator

Safety Symbols

The following symbols on the instrument and in the manual indicate precautions which must be taken to maintain safe operation of the instrument
The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
Indicates the field wiring terminal that must be connected to earth ground before operating the equipment - protects against electrical shock in case of fault.
Frame or chassis ground terminal - typically connects to the equipment's metal frame.
Alternating current (AC)
Direct current (DC)
WARNING
CAUTION
Indicates hazardous voltages
Warning denotes a hazard. It calls attention to a procedure, which if not correctly performed or adhered to could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and
met. Caution denotes a hazard. It calls attention to a procedure, which if not correctly performed or adhered to could result in damage to or destruction of the instrument. Do not pro­ceed beyond a warning note until the indicated conditions are fully understood and met.
Indicates that a laser is fitted. The user must refer to the
manual for specific Warning or Caution information to
avoid personal injury or damage to the product.
1-7
Page 18
General Information

Operators Maintenance

Operators Maintenance
WARNING NO OPERATOR SERVICEABLE PARTS INSIDE. REFER
SERVICING TO QUALIFIED PERSONNEL. TO PREVENT ELECTRICAL SHOCK DO NOT REMOVE COVERS.
Maintenance appropriate for the operator is:
Cabinet cleaning
Optical Connector Cleaning
Power supply fuse replacement
Ensure ventilating fan cover is clean.

Cleaning

Cabinet Cleaning
Clean the cabinet using a dry cloth only.
Optical Connector Cleaning
It is recommended that the optical connectors be cleaned at regular intervals using the following materials:
Description HP PartNumber
Blow Brush 9300-1131 Isopropyl Alcohol 8500-5344 Lens Cleaning Paper 9300-0761 Adhesive Tape Kit 15475-68701
CAUTION Do not insert any tool or object into the IN or OUT ports of the instrument
as damage to or contamination of the optical fibre may result.
1 Recall Default settings (STORED SETTINGS 0) and remove the
1-8
Page 19
General Information
Operators Maintenance
power from the HP 37718A/19A.
2 Remove the adapters from the IN and OUT ports. 3 Usingtheblowbrushwiththebrushremovedblow through the ferrule
of the standard flexible connector and the adapter.
CAUTION If the optical fibre of the fixed connector requires further cleaning this
entails disassembly of the module which should only be carried out by suitably trained service personnel.
4 Applysome isopropyl alcohol to a piece of the cleaning paper and clean
the barrel of the adapter. Using a new piece of cleaning paper, clean the face of the adapter. Repeat this operation, using a new piece of cleaning paper each time.
5 Lightly press the adhesive side of the tape provided against the front
of the adapter,then remove it quickly - repeat twice.This removes any particles of cleaning paper which may be present.
6 Replace the adapters on the flexible connector.

Power Supply Fuse Replacement

See “Fuse Replacement” on page 2-3
1-9
Page 20
General Information

Statement of Compliance

Statement of Compliance
This instrument has been designed and tested in accordance with IEC Publication 1010-1 + A1:1992 Safety requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
The CE mark shows that the product complies with all relevant European legal Directives.
This is a symbol of an Industrial Scientific and Medical
ISM 1-A
Group 1 Class A product.
The CSA mark is a registered trademark of the Cana­dian Standards Association.

Australian EMC Regulations

The C-Tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australian EMC Framework Regulations under the terms of the Radiocommunications Act of 1992.

Noise Declaration (German)

LpA<70dB am Arbeitsplatz (operator position)
normaler Betrieb (normal position) nach DIN 45635 pt.19 (per ISO 7779)
1-10
Page 21
General Information

Electromagnetic Compatibility

Electromagnetic Compatibility
This product has been designed to meet the protection requirements of the European Communities Electromagnetic Compatibility (EMC) directives:
EN55011:1991 (Group 1, Class A) EN50082-1:1992
- IEC 1000-4-2 (1995) ESD
- IEC 1000-4-3 (1995) Radiated Susceptibility
- IEC 1000-4-4 (1995) EFT In order to preserve the EMC performance of the product, any cable
which becomes worn or damaged must be replaced with the same type and specification.

Electrostatic Discharge

When an air discharge is applied to the HP 37718A/19A in accordance with IEC 801-2 (1991), degradation in performance may be observed in the form of occasional bit errors being counted.
See also “10 Base-T Lan Connection Radiated Emissions” on page 2-11
1-11
Page 22
General Information
Electromagnetic Compatibility
Declaration of Conformity
according to ISO/IEC Guide 22 and EN45014
Manufacturer’s Name: Hewlett Packard Ltd. Manufacturer's Address:
Declares that the product
Product Name: Model Numbers: Product Options:
Conforms with the protection requirements of European Council Directive 89/336/EEC on the approximation of the laws of the member states relating to electromagnetic compatibility.
Against EMC test specifications EN 55011:1991 (Group 1, Class A) and EN 50082-1:1992
As Detailed in:
Assessed by:
Technical Report Number:6893/2200/CBR, dated 21 August 1997
Supplementary Information:
The product conforms to the following safety standards:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC, and carries the CE-marking accordingly.
Telecomms Networks Test Division South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
Communications performance Analyzer HP 37718A
This declaration covers all options of the above products as detailed in TCF A-5951-9852-01
Electromagnetic Compatibility (EMC) Technical Construction File (TCF)No. A-5951-9852-01
Dti Appointed Competent Body EMC Test Centre, GEC-Marconi Avionics Ltd., Maxwell Building, Donibristle Industrial Park, KY11 5LB Scotland, United Kingdom
IEC 61010-1:1992 EN 61010-1:1993 CSA-C22.2 No. 1010 Series M-93 EN 60825-1(1994) / IEC 825-1 (1993)
South Queensferry, Scotland
Location Date W.R. Pearson / Quality Manager
1-12
9 September
1998
Page 23
General Information
Electromagnetic Compatibility
Declaration of Conformity
according to ISO/IEC Guide 22 and EN45014
Manufacturer’s Name: Hewlett Packard Ltd. Manufacturer's Address:
Declares that the product
Product Name: Model Numbers: Product Options:
Conforms with the protection requirements of European Council Directive 89/336/EEC on the approximation of the laws of the member states relating to electromagnetic compatibility.
Against EMC test specifications EN 55011:1991 (Group 1, Class A) and EN 50082-1:1992
As Detailed in:
Assessed by:
Technical Report Number:6893/2200/CBR, dated 21 August 1997
Supplementary Information:
The product conforms to the following safety standards:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC, and carries the CE-marking accordingly.
Telecomms Networks Test Division South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
Communications Performance Analyzer HP 37719A
This declaration covers all options of the above products as detailed in TCF A-5951-9852-01
Electromagnetic Compatibility (EMC) Technical Construction File (TCF)No. A-5951-9852-01
Dti Appointed Competent Body EMC Test Centre, GEC-Marconi Avionics Ltd., Maxwell Building, Donibristle Industrial Park, KY11 5LB Scotland, United Kingdom
IEC 61010-1:1992 EN 61010-1:1993 CSA-C22.2 No. 1010 Series M-93 EN 60825-1(1994) / IEC 825-1 (1993)
South Queensferry, Scotland
Location Date W.R. Pearson / Quality Manager
9 September
1998
1-13
Page 24
General Information
Electromagnetic Compatibility
1-14
Page 25
Installation

Initial Inspection

2 Installation

Initial Inspection
WARNING TO AVOID HAZARDOUS ELECTRICAL SHOCK, DO NOT
PERFORM ELECTRICAL TESTS WHEN THERE ARE SIGNS OF SHIPPING DAMAGE TO ANY PORTION OF THE OUTER ENCLOSURE (COVERS, PANELS, METERS).
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument has been checked both mechanically and electrically. Procedures for checking electrical operation are given in Chapter 3. If the contents of the shipment are incomplete, if there is mechanical damage or defect, notify the nearest Hewlett-Packard Office. If the instrument does not pass the electrical performance tests given in Chapter 3, notify the nearest Hewlett-Packard office. If the shipping container is also damaged, or the cushioning material shows signs of stress, notify the carrier as well as the nearest Hewlett-Packard office. Keep the shipping materials for the carrier’s inspection. The Hewlett-Packard office will arrange for repair or replacement without waiting for claim settlement.
2-1
Page 26
Installation

Operating Environment

Operating Environment
This instrument is designed for Indoor use only. DO NOT operate the product in an explosive atmosphere or in the
presence of flammable gasses or fumes. This instrument may be operated in environments within the following
limits:
Temperature: 0o C to +45o C
+5oC to +40oC for Jitter operation.
+5o C to +35o C with lid printer
Altitude up to 3050m (10,000ft) Humidity up to 95% relative humidity to 40o C, but it should be
protected from temperature extremes which may cause condensation.
To ensure adequate cooling do not obstruct air vents in the instrument cabinet. Do not for example operate the instrument if it is standing on its rear feet, as air vents may be obstructed by floor covering.
CAUTION This instrument is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 644 respectively.
CAUTION VENTILATION REQUIREMENTS: When installing the instrument in a
cabinet, the convection into and out of the instrument must not be restricted. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
2-2
Page 27
Installation

Preparation for Use

Preparation for Use
WARNING FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
REPLACE FUSE ONLY WITH SAME TYPE AND RATINGS (see “Fuses” on page 2-3).
WARNING If this instrument is not used as specified, the protection
provided by the equipment could be impaired. This instrument must be used in a normal condition only (in which all means for protection are intact).

Power Requirements

The HP 37718A/19A Communications Performance Analyzer requires a power source of 90 V to 132 V ac or 198 V to 264 V ac at a frequency between 47 Hz and 63 Hz (nominal).
Total power consumption is 450 VA (maximum). The fuse rating for the power source is given in the following Table.

Fuses

Line Voltage Fuse Rating HP Part Number
90V to 264V 5A Timed, 250V 2110-1120

Fuse Replacement

Only the ac line fuse located at the rear of the instrument may be replaced by the operator.
WARNING ALL OTHER FUSE REPLACEMENT SHOULD ONLY BE
CARRIED OUT BY SUITABLY TRAINED SERVICE PERSONNEL AWARE OF THE HAZARDS INVOLVED.
2-3
Page 28
Installation
Preparation for Use
WARNING BEFORE REMOVING THE FUSE, THE AC LINE POWER CORD
SHOULD BE DISCONNECTED FROM THE POWER SOURCE AND THE OTHER END DISCONNECTED FROM THE INSTRUMENT.
WARNING ONLY USE A FUSE OF THE CORRECT RATING AS LISTED IN
“Fuses” on page 2-3. DO NOT use repaired fuses or short-circuited
fuseholders: For continued protection against fire, replace the line fuse(s) only with fuse(s) of the same voltage and current rating and type.
The fuse is removed by inserting a suitable flat bladed tool into the slot in the fuse cap and turning counter-clockwise. The cap and the fuse can then be removed and the fuse changed for another of the correct rating. The fuse rating and HP part number are listed in “Fuses” on page 2-3.

Connecting to the Power Supply

WARNING This is a Safety Class I instrument (providedwith a protective earthing ground,
incorporated in the powercord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited.
CAUTION Before switching on this instrument, make sure that the line supply voltage is in the
specified ranges. Range selection is automatic.
2-4
Page 29
Installation
Preparation for Use
Power Cord
WARNING TO AVOID THE POSSIBILITY OF INJURY OR DEATH, THE
FOLLOWING PRECAUTIONS MUST BE FOLLOWED BEFORE THE INSTRUMENT IS SWITCHED ON:-
WARNING (a) Note that the protection provided by grounding the
instrument cabinet may be lost if any power cable other than the three-pronged type is used to couple the ac line voltage to the instrument.
WARNING (b) If this instrument is to be energized via an auto-transformer
to reduce or increase the line voltage, make sure that the common terminal is connected to the neutral pole of the power source.
WARNING (c) The power cable plug shall only be inserted into a socket
outlet provided with a protective ground contact. The protective action must not be negated by the use of an extension cord without a protective conductor (grounding).
The power cord supplied with each instrument varies with the country of destination. The following figure illustrates the standard power plug and cord configurations that are commonly used. The part number shown beneath each plug is the part number of the appropriate power cord and plug. If the appropriate power cord is not included with the instrument notify the nearest Hewlett-Packard office and a replacement will be provided.
Power Cord Configurations and Part Numbers
2-5
Page 30
Installation

Connecting to the Network

Connecting to the Network
The network connectors are located on the modules at the side of the instrument. The connections available depend on the options fitted to your instrument.
Before Connecting, note the Warning and Caution information given.

All Connectors

CAUTION When connecting or disconnecting, ensure that you are grounded or,
make contact with the metal surface of the Mainframe with your free hand to bring you, the module, and the mainframe to the same static potential. Modules remain susceptible to ESD damage while the module is installed in the Mainframe Additional ESD information is required when servicing see “AdditionalPrecautions for Service Engineers” on page 2-14
RECEIVER 2,8,34,140 IN
TRANSMIT 2,8,34,140 OUT
DS3, 2Mb/s, 34Mb/s 75Ω IN
DS3, 2Mb/s, 34Mb/s 75ΩOUT

Electrical Interface Connectors

PDH receiver input interface. Allows the connection of 75 unbalanced data signals (all rates) and 120 balanced data signals at 2 Mb/s.
PDH transmitter output interface. Provides 75 unbalanced data output (all rates) and 120 balanced data output at 2 Mb/s
PDH/DSn receiver input interface. Allows the connection of 75 unbalanced DS3, 2 Mb/s and 34.368 Mb/s data signals.
PDH / DSn transmitter output interface. Provides 75 unbalanced DS3, 2 MB/s and 34 Mb/s data output. A “keep alive” signal is output when the transmit signal is SDH/SONET.
2-6
Page 31
Installation
Connecting to the Network
DS1 100 2Mb/s 120Ω IN
DS1 100 2Mb/s 120Ω OUT
52/155 Mb/s DATA IN
52/155 Mb/s DATA OUT
64k REF IN
1.5M REF IN 2M REF IN
REF OUT
MUX
Ω,
Ω,
PDH / DS1 receiver input interface. Allows the connection of 100 balanced DS1 and 120 balanced 2 Mb/s data signals.
PDH / DS1 transmitter output interface. Provides 100 balanced DS1 and 120 balanced 2 Mb/s data signals. A “keep alive” signal is output when the transmit signal is SDH/SONET.
SDH/SONET receiver input interface. Allows the connection of 75 unbalanced STM 0/1, STS 1/3 electrical signals.
SDH/SONET transmitter output interface. Provides a 75 unbalanced STM 0/1, STS 1/3 electrical output.
These ports allow connection of timing reference signals.
Provides a 2 MHz reference Clock Output.
Allows the insertion of an external 2 Mb/s tributary into the transmitted payload.
MUX DS1 100
MUX 2 Mb/s 75
DEMUX DS1 100
DEMUX 2 Mb/s 75
Allows an externally supplied DS1 signal to be inserted in any or all
timeslot(s) of a DS3 signal. In conjunction with the SDH/SONET module a DS1 signal can be inserted into a VT1.5 or TU-11.
Allows an externally supplied 2.048 Mb/s signal to be inserted in any or
all timeslot(s) of a 34 Mb/s signal. In conjunction with the SDH module a
2.048 Mb/s signal can be inserted into a TU-12.
Allows a DS1 signal to be dropped from any timeslot of a DS3 signal. In conjunction with the SDH/SONET module this port allows the demultiplexing of a DS1 signal carried in a VT1.5
Allows a 2.048 Mb/s signal to be dropped from any timeslot of an 8.448 Mb/s, 34.368 Mb/s or 139.264 Mb/s signal.
2-7
Page 32
Installation
Connecting to the Network
CLOCK TRIGGER
DCC
Reference SDH/SONET clock trigger output 51 MHz; nominal +/- 400 mV into 50 ohms.
Allows the Drop and Insert of Regenerator Section (192 kb/s) and Multiplexer Section (576 kb/s) Data Communication Channels (DCC).
The following connections are available:
Table 1 DCC Connector Pin-out
Pin Number Connection
1 Shield Ground 2 Data input for Tx(-) 3 Tx Clock 4 Data Output from Rx(-) 5NC 6 Rx Clock(-) 7NC 8 Signal Ground 9 Data Input for Tx(+)
PROTECTED MONITOR POINT INPUT
10 Tx Clock(+) 11 Data output from Rx(+) 12 NC 13 Rx Clock(+) 14 NC 15 NC
52/155/622 Mb/s (NRZ); Line Code NRZ; Level nominal 1 V p-p into 50 ohms; SMA connector.
2-8
Page 33
Installation
Connecting to the Network

Optical Interface Connectors

For your protection, review all laser information given in this manual before installing or using the instrument.
WARNING To prevent personal injury, avoid use that may be hazardous to
others, and maintain the module in a safe condition Ensure the information given below is reviewed before operating the module.
Laser Product Classification
All optical modules are classified as Class I (non-hazardous) laser product in the USA which complies with the United States Food and Drug Administration (FDA) Standard 21 CFR Ch.1 1040.10, and are classified as Class 1 (non-hazardous) laser products in Europe which complies with EN 60825-1 (1994).
To avoid hazardous exposure to laser radiation, it is recommended that the following practices are observed during system operation:
ALWAYS DEACTIVATE THE LASER BEFORE CONNECTING
OR DISCONNECTING OPTICAL CABLES.
• When connecting or disconnecting optical cables between the module and device-under-test, observe the connection sequences given below.
Connecting: Connect the optical cable to the input of the device-
under-test before connecting to the module’s Optical Out connector.
Disconnecting: Disconnect the optical cable from the module’s Optical
Out connector before disconnecting from the device-
under-test. Always fit the fibre optic connector dust caps over the laser aperture.
• NEVER examine or stareinto the open end of a broken, severed, ordis­connected optical cable when it is connected to the module’s Optical Out connector.
• Arrange for service-trained personnel, who are aware of the hazards involved, to repair optical cables.
2-9
Page 34
Installation
Connecting to the Network
CAUTION 1. Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous radiation exposure.
2. Always fit the fibre optic connector dust caps on each connector when not in use. Before connection is made, always clean the connector ferrule tip with acetone or alcohol and a cotton swab. Dry the connector with compressed air. Failure to maintain cleanliness of connectors is liable to cause excessive insertion loss.
Laser Warning Symbols
The front panel of the optical module has the following label:
CLASS 1 LASER PRODUCT
NOTE CLASS 1 LASER PRODUCT translates as follows:
Finnish - LUOKAN 1 LASERLAITE
Finnish/Swedish - KLASS 1 LASER APPARAT
OPTICAL IN
OPTICAL OUT
This label indicates that the radiant energy present in this instrument is non-hazardous.
Allows connection of an optical signal, wavelength 1200 to 1600 nm, at a maximum power level of -8 dBm. NEVER EXCEED +3 dBm. Accepts STM-0, STM-1, STM-4 and STM-16. Also accepts SONET signals, STS-1, STS-3, STS-12, STS-48, OC-1, OC-3, OC-12 and OC-48.
Provides an STM-0, STM-1, STM-4 or STM-16 optical signal (OC-1, OC­3, OC-12 or OC-48 SONET signals) at wavelength1280 to 1330nm, and/ or 1550 to 1565 nm, at a nominal power level of +1 dBm.
Cleaning Optical Connectors
See “Optical Connector Cleaning” on page 1-8
2-10
Page 35
Installation

Connecting Accessories

Connecting Accessories
LID
VGA
HANDSET
Printer HP-IB, RS232, PARALLEL ONLY
Remote Control HP-IB, RS232, 10 BASE -T
Provides the output for the option 602 printer which is fitted in the cover (LID) of the instrument.
Provides the output for a display monitor.
Allows connection of a telephone handset for communication across the network.
External printer connection details are given in the Users Guide. The port selected for external printer use is not available for remote control.
Remote control connection is given in the Remote Control Manual. The port selected for remote control use is not available for an external printer.

10 Base-T Lan Connection Radiated Emissions

To ensure compliance with EN 55011 (1991) a category 5, FTP patch lead, RJ45 cable should be used to connect the LAN port on the processor module marked "10 Base-T".
2-11
Page 36
Installation
Connecting Accessories

Hewlett-Packard Interface Bus

The HP 37718A/19A Communications Performance Analyzer (Option 601) is connected to the HP-IB by means of an appropriate HP-IB cable.The HP-IB interconnecting cables available are listed in the following table.
HP-IB Interconnecting Cables
Length Accessory Number 1 meter HP 10833A 2 meters HP 10833B 4 meters HP 10833C
0.5 meter HP 10833D
To achieve interface design performance standards, restrictions are placed on the HP-IB system cable lengths. These restrictions allow the bus interface electronics to maintain correct line voltage levels and timing relationships.
When connecting an HP-IB system the following rules should be observed:
The total HP-IB cable length used must be less than or equal to 20 meters (65.6 feet).
The total HP-IB cable length used must be less than or equal to 2 meters (6 feet) × the total number of devices connected to the bus.
A standard HP-IB connector is provided on the instrument. The connections are shown in the following figure. The mating connector part number is HP 1251-0293 or Amphenol 57-30240.
2-12
Page 37
Installation
Connecting Accessories
HP-IB Connections
HP-IB Address Selection
The HP 37718A/19A (Option 601) HP-IB address is accessed on the OTHER display under the REMOTE CONTROL function.
The address can be set to any value between 0 and 30 inclusive.
2-13
Page 38
Installation

Additional Precautions for Service Engineers

Additional Precautions for Service Engineers

Safety Precautions

DO NOTsubstitute parts or modify equipment: Because of the danger of
introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to a Hewlett-Packard Sales and Service Office for service and repair to ensure the safety features are maintained.
DO NOT service or adjust alone: Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous electrical shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Review “Safety Precautions for the Operator” on page 1-6

ESD Precautions

CAUTION When making connections to the modules, review “Connecting to the Network” on
page 2-6
The module contains components sensitive to electrostatic discharge. To prevent component damage, carefully follow the handling precautions presented below.
The smallest static voltage most people can feel is about 3500 volts. It takes less than one tenth of that (about 300 volts) to destroy or severely damage static sensitive circuits. Often, static damage does not immediately cause a malfunction but significantly reduces the component’s life. Adhering to the following precautions will reduce the risk of static discharge damage.
• Keep the module in its conductive storage box when not installed in the Mainframe. Save the box for future storage of the module.
• Before handling the module, select a work area where potential static sources are minimized. Avoid working in carpeted areas and non-con­ductive chairs. Keep body movement to a minimum. Hewlett-Packard
2-14
Page 39
Installation
Additional Precautions for Service Engineers
recommends that you use a controlled static workstation.
• Handle the module by its front-panel. Avoid touching any components or edge connectors. When you install the module, keep one hand in contact with the protective bag as you pick up the module with your other hand. Then, before installing the module, ensure that you are grounded or make contact with the metal surface of the Mainframe with your free hand to bring you, the module, and the mainframe to the same static potential. This also applies whenever you con-
nect/disconnect cables on the front-panel.
2-15
Page 40
Installation
Additional Precautions for Service Engineers

Front Panel Soft Recovery (Cold Start)

Use the following procedure if you need to perform a front panel soft recovery (i.e. cold start) of the instrument.
Soft Recovery Procedure
1 Switch off the instrument. 2 On the instrument front panel - press and hold softkeys 0 and 4
simultaneously (the softkeys immediately below the display; key 0 is on the extreme left).
3 Power up the HP 37718A or HP 37719A while holding the softkeys
pressed.
4 Whenthe LOS LED has flashed OFF and then ON again, the keys can
be released.
5 The LOS LED will flash OFF/ON again several times (7), followed by
an audible ‘beep’ and the display indicating ‘Initializing Instrument’.
6 Once the initiations is complete the display will indicate:
‘Firmware Revision Update’ ‘Default settings assumed’ Hit any key to attempt restart’
7 Hit any key, then wait approximately 10 seconds. The instrument
should return to its default settings and normal operation.
2-16
Page 41
Performance Tests

3 Performance Tests

Introduction

The procedures given in this Section, test the HP 37718A/19A performance using the Specifications provided, as performance standards. Tests are intended to be performed in the recommended order for a full instrument calibration check. Tests can be run individually, however it will be assumed that any preceding test would meet specifications.

Equipment Required

Equipment required for the Performance Tests is given in this Chapter. Any equipment which meets or exceeds the critical specification of the equipment listed, may be substituted. Recommended models are those typically used in HP Service Centers. Alternative models are also listed.

Performance Test Record

The results of the Performance Tests may be recorded on the Test Records at the end of this Chapter.The Performance Test Record lists all the tested specifications and the acceptable limits. The results recorded at incoming inspection may be used for comparison during periodic maintenance, troubleshooting or after repair or adjustment.

Calibration Cycle

This instrument requires periodic verification of performance. Depending on use and environmental conditions, the instrument should be checked approximately once every 2 years, using these Performance Tests.
Option Configuration
Please refer to Chapter 6 and the instrument Specifications for information on option configuration.
3-1
Page 42
Performance Tests
Table 3-1 Recommended Test Equipment
Instrument Critical Specification Recommended
Model
Oscilloscope 400 MHz Bandwidth, 1 M Input Termination
Telecom mask measurement capability.
Frequency Synthesizer 75 Output, Sinewave to 80 MHz,
Amplitude to 3 V pk-pk at 10 kHz, 1 Hz resolution
Signal Generator Sinewave 700 kHz to 170 MHz, Amplitude 500 mV HP 8657A Frequency Counter Range 0 to 200 MHz, 2 channels with accuracy
<0.1ppm. (Ratio Mode) Spectrum Analyzer** Frequency >2.5 Gb/s. Resolution bandwidth 10 Hz. HP 8560E Optical Coupler** 1310 and 1510 nm. 10/90% output. HP 15477C Optical Power Meter
and Sensor Module Lightwave Converter Wavelength 1300 to 1560 nm, Conversion Gain >300
Optical Attenuator Wavelength 1200 - 1600 nm, Range 0 - 30 dB HP 8157A or
FC/PC Optical Interface Connector
Optical Cables Unique HP 11871A (Qty 2) PDH Structured Test
Set
Range -8 dBm to -15 dBm, Wavelength1270-1560 nm HP 8153A and HP
volts/watt, Frequency Response < 3 dB down at 1 GHz.
Unique HP 81000FI (Qty 4)
Unique HP37718A/37719Aor
HP 54520A opt 001 or HP 54810A opt 001
HP 3335A opt 001 or HP 3325B & HP 8657A with 50/75 matching pad
HP 5325A opt 010 or HP 53181A opt 001
81536A HP 11982A opt 012
HP 8156A opt 100
HP 37717C opt UKJ
64 kb/s Clock Generator
75 Attenuator Kit 0 to 200 MHz 3, 6, 10, 20 dB fixed attenuator pads HP 86213A Cable Simulator #E1 75 coaxial cable 6dB loss at 1 MHz 8120-0049 (120 m)* Cable Simulator #E2
Cable Simulator #E3 75 coaxial cable 12dB loss at 17 MHz 8120-0049 (120 m)* Cable Simulator #E4 75coaxial cable 12dB loss at 70 MHz 8120-0049 (80 m)*
64 kb/s ternary clock. HP 37732A
75 coaxial cable 6dB loss at 4 MHz
8120-0049 (80 m)*
3-2
Page 43
Performance Tests
Table 3-1 Recommended Test Equipment, continued
Instrument Critical Specification Recommended
Model
Cable Simulator #E1-M 75 coaxial cable 3dB loss at 1 MHz 8120-0049 (60 m)* Cable Simulator #E2-M
Cable Simulator #E3-M 75 coaxial cable 6dB loss at 17 MHz 8120-0049 (60 m)* Cable Simulator #E4-M 75coaxial cable 6dB loss at 70 MHz 8120-0049 (40 m)* Cable Simulator #DS1 75 coaxial cable. Equivalent to 655 feet ABAM cable. 8120-0049 (80m)* Cable Simulator #DS3 75Ω coaxial cable. Equivalent to 450 feet of 728A cable 8120-0049 (55m)* Cable Simulator STM-0/
STS-1 Cable Simulator STM-1/
STS-3 75
Termination 0 to 200 MHz HP 15522-80010 T Connector BNC to Dual BNC HP 1250-0781 Adaptor SMA to BNC 1250-1787 (Qty 2) Adaptor Type N to BNC HP 1250-1534 (Qty 2) Adaptor Type N to N HP 1250-1528 (Qty 2) Balanced/Unbalanced
Converter Blocking Capacitor 0.18uF 200V HP 10240B
75 coaxial cable 3dB loss at 4 MHz
coaxial cable 6dB loss at 26 MHz 8120-0049 (50 m)*
75
75
coaxial cable 12dB loss at 78 MHz 8120-0049 (75 m)*
balanced: 75Ω Unbalanced (nominal) HP 15508B (Qty 2)
110
8120-0049 (40 m)*
* Note: Cable lengths quoted are typical for the half bit rate loss. Ideally, cables should be trimmed to correct length/loss by measuring with a Network Analyzer.
** Note: Spectrum Analyzer/Optical Coupler are only required for 37718A Jitter options.
Self Test Loopback Cables and Accessories
HP E4545A 3m fibre optic cable FC/CP connectors. (supplied accessory) HP E4546A FC/CP 15 dB attenuator. (supplied accessory) HP 15525A 75 ohm BNC, 3 off. HP 15512A Siemens 3 pin (Option 010 units) HP 15670A Bantam 110 ohm (Option 011 units), 2 off. DCC 15 pin loopback plug - see figure in DCC Test. Formatted floppy disk.
3-3
Page 44
Performance Tests

Recall Default Settings

The Performance Tests require the HP 37718A/19A to be set to a pre­defined (default) state at the beginning of each test.
1 Using , display softkeys, and set up the
OTHER
OTHER
STORED
SETTINGS
display as shown opposite
.
2 Press to recall the instrument default settings.The
RECALL
instrument display will blank for a few seconds while the settings are recalled and the status display will indicate stored settings number 0 recalled.
3-4
Page 45
Performance Tests

Self Test

Self Test
Description
Before carrying out the performance tests run Self Test to ascertain the integrity of the HP 37718A/19A. These tests take at least 1 hour to complete depending on the options fitted. Alternatively you can run Confidence Tests which only takes 2 to 3 minutes to complete. This is not a full verification but performs BER measurements with internal and external loopbacks fitted.
Run Confidence TESTS
1 Choose TEST TYPE on the
display.
CAUTION Safety precaution, care and connection cleanliness are essential to avoid
optical signal degradation or damage. see Operators Maintenance in chapter 1 and see Optical Interface Connectors in chapter 2 if in doubt. The use of air-gap attenuators is not recommended. Failure to attenuate the optical signal could result in damage to the optical receiver.
CONF. TESTS
OTHER
SELF TEST
3-5
Page 46
Performance Tests
Self Test
NOTE If the Optical Interface is a Dual WavelengthOption the 1550 nm IN and
OUT ports are not connected at this stage and can be tested later. If any or all of these connections are not made the HP 37718A/19A will FAIL Self Test.
2 Make the loopback connections listed below:
Connect the 15 dB optical attenuator provided at the Optical OUT and then connect the attenuator output to Optical IN using the optical cable supplied with your instrument. Connect Multirate Analyser IN to OUT Connect Transmit module 75 OUT to Receive module 75 IN Connect Transmit module 100/120 OUT to Receive module 100/ 120 IN
3 Press to activate the Self Test.TESTSTATUS RUNNING
RUN/STOP
will be displayed. The information pertaining to TEST TYPE, TEST NUMBER and SUBTEST NUMBER will change as the Self Test progresses. If the HP 37718A/19A is functioning correctly, after a time of 2 to 3 minutes, TEST STATUS PASSED is displayed. If TEST STATUS [FAIL nnn] is displayed, check the displayed error information and check loopback connections are correct. Repeat the test and if the problem persists contact your local HP representative. See Chapter 5 for information.
NOTE Each individual self test requires unique loopback connections. To obtain
a list of the connections required move the highlighted cursor to
CABLING INFO and press . The Loopbacks list will appear on the
SET
display. Also refer to Chapter 5 for Self Test connection information.
3-6
Page 47
Performance Tests
Self Test
Run ALL TESTS
1 Choose TEST TYPE on the display.
ALL TESTS
OTHER
SELF TEST
2 Insert a formatted disk into the instrument disk drive. 3 Make the loopback connections listed below:
Connect the 15 dB optical attenuator provided to Optical OUT and connect the attenuator output to Optical IN. Connect Multirate Analyser IN to OUT Connect Transmit module 75 OUT to Receive module 75 IN Connect Transmit module 100/120 OUT to Receive module 100/ 120 IN Connect Transmit module 75MUX to Receive module 75DEMUX Connect Transmit module 100/120 MUX to Receive module 100/ 120 DEMUX
4 Press to activate the Self Test.TESTSTATUS RUNNING
RUN/STOP
will be displayed. The information pertaining to TEST TYPE, TEST NUMBER and SUBTEST NUMBER will change as the Self Test progresses. If the HP 37718A/19A is functioning correctly, after a time of at least 1 hour, TEST STATUS PASSED is displayed. If TEST STATUS [FAIL nnn] is displayed, check the displayed error information and check loopback connections are correct. Repeat the test and if the problem persists contact your local HP representative. See Chapter 5 for information.
3-7
Page 48
Performance Tests
Self Test
1550 nm Dual Wavelength Tests:
If a Dual Wavelength Optical Interface is fitted, repeat the CONFIDENCE test with 1550 nm selected.
1 Select 1550 nm OPTICS wavelength 2 Connect1550 nm OUT to IN via 15dB Optical Attenuator.Ensure the
attenuator is inserted in the OUT port.
3 Run the test.
3-8
Page 49
Performance Tests
Self Test
DCC Port Tests:
1 Select BER TESTS 2 Make the following connections at the Multirate Analyzer DCC port.
3 Run the BER TESTS.
3-9
Page 50
Performance Tests

PDH/DSn Internal Transmitter Clock Accuracy & Offset

PDH/DSnInternalTransmitterClock Accuracy & Offset
Specifications
Rate Frequency
E4 E3 E2 E1 DS1 DS3
NOTE The Rates available are dependent on the HP 37718A/19A option.
139.264 MHz ± 4.5 ppm
34.368 MHz ± 4.5 ppm
8.448 MHz ±4.5 ppm
2.048 MHz ±4.5 ppm
1.544 MHz ±4.5 ppm
44.736 MHz ±4.5 ppm
Description
This test verifies that the PDH transmit data rates are within limits. These limits assume the instrument is within the calibration cycle. The Frequency Offset capability (deviation from Standard Bit Rate) is also checked. ForE1 to E4 a Frequency Counter connected to the PDH/DSn Signal Out port measures the data rate on an "all ones" pattern. This gives an indirect measure of the internal transmitter clock frequency as the data is clocked by the internal clock oscillator. Because the Frequency Counter triggers from the positive pulses only, the frequency count will, for Ternary signals E1, E2 & E3, be half the selected data rate. For DS-1 and DS-3 the Frequency Counter is connected to the CLOCK OUT port.

Equipment Required

Frequency Counter : HP 5335A Option 010 75 Termination : HP 15522-80010 T Connector : HP 1250-0781
3-10
Page 51
Performance Tests
PDH/DSn Internal Transmitter Clock Accuracy & Offset
Procedure
1 Recallthe HP 37718A/19A DEFAULT SETTINGS as shown on Page3-
4.
2 Connect the 75OUT port of the PDH/DSn module to the Frequency
Counter. Terminate the Frequency Counter input in 75 (use the T connector).
3 Set the display as shown opposite
NOTE When changing the PDH Rate or Offset value the VCXO takes time to
settle. As a consequence the frequency counter reading will not stabilize until "VCXO OUTPUT BIT RATE SETTLING" clears from the STATUS line of the display.
TRANSMIT

E1 (2.048Mb/s) Frequency Accuracy

4 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 1023995Hz and 1024005Hz.
5 Select FREQUENCY OFFSET [+50PPM]. 6 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 1024046.5Hz and 1024055.5Hz.
3-11
Page 52
Performance Tests
PDH/DSn Internal Transmitter Clock Accuracy & Offset
7 Select FREQUENCY OFFSET [-50PPM]. 8 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 1023944Hz and 1023953Hz.

E2 (8.448Mb/s) Frequency Accuracy

9 Select SIGNAL [8Mb/s]; PATTERN [ALL 1’s] on the
display.
10 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 4223981Hz and 4224019Hz.
11 Select FREQUENCY OFFSET [+30PPM]. 12 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 4224107.5Hz and 4224145.7 Hz.
13 Select FREQUENCY OFFSET [-30PPM]. 14 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 4223854Hz and 4223892Hz.
TRANSMIT

E3 (34.368Mb/s) Frequency Accuracy

15 Select SIGNAL [34Mb/s] PATTERN [ALL 1’s] on the
display.
16 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 17183922.7Hz and 17184077.4Hz.
TRANSMIT
17 Select FREQUENCY OFFSET [+20PPM] 18 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 17184266 and 17184421Hz.
19 Select FREQUENCY OFFSET [-20PPM] 20 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 17183579Hz and 17183734Hz.
3-12
Page 53
Performance Tests
PDH/DSn Internal Transmitter Clock Accuracy & Offset
21 Select FREQUENCY OFFSET [USER OFFSET] [+100PPM] 22 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 17185641Hz and 17185796Hz.
23 Select FREQUENCY OFFSET [USER OFFSET] [-100PPM] 24 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 17182205Hz and 17182359Hz.

E4 (139.264Mb/s) Frequency Accuracy

25 Select SIGNAL [140Mb/s] PATTERN [ALL 1’s] on the
display.
26 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 69631686.5Hz and 69632318.5Hz.
27 Select FREQUENCY OFFSET [+15PPM]. 28 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 69632731Hz and 69633358Hz.
29 Select FREQUENCY OFFSET [-15PPM]. 30 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 69630642Hz and 69631269Hz.
TRANSMIT

DS1 (1.544Mb/s) Frequency Accuracy

31 Connect the CLOCK OUT port of the PDH/DSn module to the
Frequency Counter.
32 Select SIGNAL [DS1 1.5Mb/s] on the display. 33 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 1543993Hz and 1544007Hz.
TRANSMIT
34 Select FREQUENCY OFFSET [+32PPM] 35 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is
3-13
Page 54
Performance Tests
PDH/DSn Internal Transmitter Clock Accuracy & Offset
between 1544042Hz and 1544056Hz.
36 Select FREQUENCY OFFSET [-32PPM] 37 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 1543944Hz and 1543958Hz.

DS3 (44.736Mb/s) Frequency Accuracy

38 Connect the CLOCK OUT port of the PDH/DSn module to the
Frequency Counter.
39 Select SIGNAL [DS3] on the display. 40 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 44735799Hz and 44736201Hz
41 Select FREQUENCY OFFSET [+20PPM] 42 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 44736693Hz and 44737096Hz.
43 Select FREQUENCY OFFSET [-20PPM] 44 Adjust the Frequency Counter ATTEN and Trigger Level to obtain a
stable reading and ensure that the frequency counter reading is between 44734904Hz and 44735307Hz.
45 Disconnect all the test equipment.
TRANSMIT
3-14
Page 55
Performance Tests

PDH/DSn Transmitter Output

PDH/DSn Transmitter Output
Specification
Rate Level Waveshape
DSX-1 DS1-LO As DSX-1 with 655' ABAM
DS3-HI 0.9V pk (nominal) DSX-3 560mV pk (nominal) Fits mask T1.102-1993 DS3-900 330mV pk (nominal) E1 Balanced
E1 Unbalanced E2 Unbalanced E3 Unbalanced E4 Unbalanced
NOTE The Rates available are dependent on the HP 37718A/19A option
Description
This test ensures the transmitter output level and pulse shape meet the required specifications at all PDH/DSn rates. The Transmitter output is connected to an Oscilloscope and the waveshape compared with the predefined masks stored in the Oscilloscope memory. The signal levels are also measured using the Oscilloscope.
3.0V 20%
Cable
3.00V 10%
2.37V 10%
2.37V 10%
1.0V 10%
0.5V 10%
Fits mask T1.102-1993
As per ITU rec G703 As per ITU rec G703 As per ITU rec G703 As per ITU rec G703 As per ITU rec G703
3-15
Page 56
Performance Tests
PDH/DSn Transmitter Output

Equipment Required

Oscilloscope : HP 54520A Option 001 110/75 Balanced to
Unbalanced Converter 75 Termination
Procedure
NOTE This performance test is written using the HP 54520A Oscilloscope with
Option 001 - TelecomMasks Application. If any other Oscilloscope is used the keystrokes given in this procedure will not apply. The Positive pulse masks are contained in the ROOT directory of the TelecomMask Application Disk and the Negative pulse masks are in the INV_MASK directory.
Installing the Telecom Masks Option
1 Insert the Disk containing the Telecom Mask into the Oscilloscope
Disk Drive.
2 Press SHIFT (Blue Key) DISK 3 Press softkey then softkey
139.264 Mb/s - All 1’s Pulse
1 Recall the HP 37718A Default Settings as shown on page 3-4.
DIRECTORY ROOT DIRECTORY
: HP 15508B
: HP 15522-80010
2 Set the display as shown below.
3-16
TRANSMIT
Page 57
Performance Tests
PDH/DSn Transmitter Output
3 Connect the PDH/DSn 75 OUT port to the Oscilloscope CHAN 1,
terminate in 75Ω.
4 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
5 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select 139BIN1 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
7 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 450 mV and 550 mV. (i.e. 900 mV-1.1V Pk-Pk)
All 0’s Pulse
9 Select PATTERN [ALL 0’s] on the display.
TRANSMIT
3-17
Page 58
Performance Tests
PDH/DSn Transmitter Output
Load the Telecom Mask
10 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select 139BIN0 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
11 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
12 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
13 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 450 mV and 550 mV. (i.e. 900 mV-1.1V Pk-Pk)
14 Select PATTERN to [ALL 0’s] on the display.
TRANSMIT
15 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
16 Repeat steps 5 to 13 to check pulses against Inverted masks.
3-18
Page 59
Performance Tests
PDH/DSn Transmitter Output

Procedure - 34.368 Mb/s

Positive Pulse
1 Recall the HP 37718A/19A Default Settings as shown on page 3-4. 2 Set the display as shown opposite.
TRANSMIT
3 Connect the PDH/DSn 75 OUT port to the Oscilloscope CHAN 1,
terminate in 75Ω.
4 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
5 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select 34MG703 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8
3-19
Page 60
Performance Tests
PDH/DSn Transmitter Output
b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
7 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 900 mV and 1.100 Volts.
Negative Pulse
9 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
10 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select 34MG703 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
11 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
3-20
Page 61
Performance Tests
PDH/DSn Transmitter Output
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
12 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
13 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 900 mV and 1.100 Volts.

Procedure - 8.448 Mb/s

Positive Pulse
1 Recall the HP 37718A/19A Default Settings as shown on page 3-4. 2 Set the display as shown opposite.
TRANSMIT
3 Connect the PDH/DSn 75 OUT port to the Oscilloscope CHAN 1,
terminate in 75Ω.
4 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
3-21
Page 62
Performance Tests
PDH/DSn Transmitter Output
Load the Telecom Mask
5 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/UseENTRY/MEASURE control to select 8MG703 in thehighlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
7 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.
Negative Pulse
9 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
3-22
Page 63
Performance Tests
PDH/DSn Transmitter Output
Load the Telecom Mask
10 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/UseENTRY/MEASURE control to select 8MG703 in thehighlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
11 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
12 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
13 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.

Procedure - 2.048 Mb/s

Positive Pulse
1 Recall the HP 37718A/19A Default Settings as shown on page 3-4.
3-23
Page 64
Performance Tests
PDH/DSn Transmitter Output
2 Set the display as shown below.
TRANSMIT
3 Connect the PDH/DSn 75 OUT port to the Oscilloscope CHAN 1,
terminate in 75Ω.
4 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
5 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS_1E in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
3-24
Page 65
Performance Tests
PDH/DSn Transmitter Output
7 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.
Negative Pulse
9 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
10 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS_1E in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
11 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
12 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on
3-25
Page 66
Performance Tests
PDH/DSn Transmitter Output
the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
13 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.

Procedure - DSX-3

Positive Pulse
1 Recall the HP 37718A/19A Default Settings as shown on page 3-4. 2 Set the display as shown below.
3 Connect the PDH/DSn 75 OUT port to the Oscilloscope CHAN 1,
terminate in 75Ω.
4 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
5 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION.
TRANSMIT
3-26
Page 67
Performance Tests
PDH/DSn Transmitter Output
b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS3_92 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
7 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to select softkey and /
MASK ALIGN
or visibly fit the waveform to the mask by adjusting the
MASK MASK AUTO
Oscilloscope vertical gain/position and horizontal delay to obtain a PASS on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is nominally 560 mV.
Negative Pulse
9 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
10 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
3-27
Page 68
Performance Tests
PDH/DSn Transmitter Output
c/ Use ENTRY/MEASURE control to select DS3_92 in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
11 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
12 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to select and / to
MASK MASK AUTO MASK ALIGN
visibly fit the waveformto the mask by adjusting the Oscilloscope vertical gain/position and horizontal delay to obtain a PASS on the Mask.
13 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is nominally 560 mV.
14 Set the signal to HIGH and check that the peak pulse
TRANSMIT
amplitude is approximately 0.9V.
15 Set the signal to LOW and check that the peak pulse
TRANSMIT
amplitude is approximately 0.3V.

Procedure - DSX-1

Positive Pulse
1 Recall the HP 37718A/19A Default Settings as shown on page 3-4. 2 Connect the DSn 100 OUT port to the Oscilloscope via the HP
15508B Balanced to unbalanced Converter and terminate in 75 at the Oscilloscope input.
3 Select the Transmitter Output page.
3-28
Page 69
Performance Tests
PDH/DSn Transmitter Output
4 Set the display as shown below.
TRANSMIT
5 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
6 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS1_NEW in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
7 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8, (Ensure CH1 Impedance is reset to 1MΩ. Select TRIGGER, SETUP and set softkey to ).
PATTERN TRIGGER Trig’d
b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
3-29
Page 70
Performance Tests
PDH/DSn Transmitter Output
NOTE Using the balanced to unbalanced converter reduces the pulse amplitude
by a factor of 1.266. This will cause the comparison with the mask to fail. Toovercome this select CHAN 1 ON and reduce the Volts/Divisionsetting (highlightedonthe softkey display). Reduced Volts/Division= highlighted value/1.266. The new value can be entered using the keypad
8 TheOscilloscope will compare the positive pulse with the mask limits.
A PASS message should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
9 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 1.896V and 2.844V.
Negative Pulse
10 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
11 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS1_NEW in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
12 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8, (Ensure CH1 Impedance is reset to 1M TRIGGER, SETUP and set softkey to ).
PATTERN TRIGGER Trig’d
. Select
b/ DEFINE MEAS
3-30
Page 71
Performance Tests
PDH/DSn Transmitter Output
c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
NOTE Using the balanced to unbalanced converter reduces the pulse amplitude
by a factor of 1.266. This will cause the comparison with the mask to fail. Toovercome this select CHAN 1 ON and reduce the Volts/Divisionsetting (highlightedonthe softkey display). Reduced Volts/Division= highlighted value/1.266. The new value can be entered using the keypad
13 The Oscilloscope will compare the negative pulse with the mask limits.
A PASS message should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
14 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 1.896V and 2.844V.
15 Set the signal to LOW and check that the peak pulse
TRANSMIT
amplitude reduces by approximately 20% and pulse shape has slower rise/fall edges.
16 Disconnect all the test equipment.
3-31
Page 72
Performance Tests

PDH/DSn Receiver Equalization

PDH/DSn Receiver Equalization
Specifications
Rate Equalization
E1- 2.048 Mb/s 6dB for root f cable at 1/2 bit rate E2 - 8.448 Mb/s 6dB for root f cable at 1/2 bit rate E3 - 34.368 Mb/s 12dB for root f cable at 1/2 bit rate E4 - 139.264 Mb/s 12dB for root f cable at 1/2 bit rate DS1- 1.544 Mb/s Single range from DS1-LO to DSX -1 DS3 - 44.736 Mb/s Automatic equalization for up to 900
feet of root f cable.
NOTE The Rates available are dependent on the HP 37718A/19A option

Description

The Receiver Equalization is checked by looping the transmitter output to receiver input through a special Cable. This simulates the specified cable loss at the rate under test. A BER measurement is run and a check made for no errors in the results page.

Equipment Required

Cable Simulator E4 80 metres of 8120-0049 Cable Simulator E3 120 metres of 8120-0049 Cable Simulator E2 80 metres of 8120-0049 Cable Simulator E1 120 metres of 8120-0049 Cable Simulator DS3 55 metres of 8120-0049 Cable Simulator DS1 80 metres of 8120-0049
110/75 Balanced to Unbalanced Converter
3-32
HP 15508B (Qty 2)
Page 73
Performance Tests
PDH/DSn Receiver Equalization

Procedure

1 Recall the HP 37718A/19A Default Settings as shown on page 3-4. 2 Select TRANSMITTER and RECEIVER
[COUPLED] on the display.
SETTINGS CONTROL
OTHER
139.264 Mb/s
3 Set the display as shown opposite
TRANSMIT
.
4 Connect Cable Simulator E4 between the PDH/DSn 75 OUT and
75 IN ports.
5 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
6 EnsurethatNOTROUBLEisdisplayedonthe display after
RESULTS
30 seconds.
7 Press to stop the measurement.
RUN/STOP
3-33
Page 74
Performance Tests
PDH/DSn Receiver Equalization
34.368 Mb/s
8 Set the display as shown below
TRANSMIT
.
9 ConnectCable Simulator E3 between PDH/DSn 75OUT and 75IN
ports.
10 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
11 Ensure that NO TROUBLE is displayed on the displayafter
RESULTS
30 seconds.
12 Press to stop the measurement.
RUN/STOP
3-34
Page 75
Performance Tests
PDH/DSn Receiver Equalization
DS3 (44.736 Mb/s)
13 Set the display as shown below.
TRANSMIT
14 Connect Cable Simulator DS3 between the DSn 75OUT and 75ΩΙΝ
ports.
15 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
16 Ensure that NO TROUBLE is displayed on the displayafter
RESULTS
30 seconds.
17 Press to stop the measurement.
RUN/STOP
NOTE If Cable Simulator DS3 is not available, it is permissible to select DS3-
900'on the displayand connect the HP 37718A/19A 75OUT
TRANSMIT
port direct to the 75 IN port. This setup assumes the DS3-900' output signal is within specification.
3-35
Page 76
Performance Tests
PDH/DSn Receiver Equalization
8.448 Mb/s
18 Set the display as shown below.
TRANSMIT
19 Connect Cable Simulator E2 between the PDH/DSn 75 OUT and
75 IN ports.
20 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
21 Ensure that NO TROUBLE is displayed on the displayafter
RESULTS
30 seconds.
22 Press to stop the measurement.
RUN/STOP
3-36
Page 77
Performance Tests
PDH/DSn Receiver Equalization
2.048Mb/s
23 Set the display as shown below.
TRANSMIT
24 Connect Cable Simulator E1 between the PDH/DSn 75 OUT and
75 IN ports.
25 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
26 Ensure that NO TROUBLE is displayed on the displayafter
RESULTS
30 seconds.
27 Press to stop the measurement.
RUN/STOP
3-37
Page 78
Performance Tests
PDH/DSn Receiver Equalization
DS1 (1.544 Mb/s)
28 Set the display as shown below.
TRANSMIT
29 Connect Cable Simulator DS1 between the 110 OUT and 110 IN
ports via two Balanced to Unbalanced Converters (HP 15508B).
30 Press then to start the
RESULTS
TROUBLE
SCAN
RUN/STOP
measurement.
31 Ensure that NO TROUBLE is displayed on the displayafter
RESULTS
30 seconds.
32 Press to stop the measurement.
RUN/STOP
NOTE If Cable Simulator DS1 is not available, it is permissible to select DS1-LO
on the display and connect the 110 OUT port direct to the
TRANSMIT
110Ω ΙΝ port. This setup assumes the DS1-LO output signal is within specification.
3-38
Page 79
Performance Tests

PDH/DSn Receiver Monitor Levels

PDH/DSn Receiver Monitor Levels
Specifications
Bit Rate Nominal
Loss
E1 - 2.048 Mb/s 20, 26, 30dB 3dB E2 - 8.448 Mb/s 20, 26, 30dB 3dB E3 - 34.368 Mb/s 20, 26dB 6dB E4 - 139.264 Mb/s 20, 26dB 6dB DS1 - 1.544 Mb/s 20, 26, 30dB 3dB
DS3 - 44.736 Mb/s 20, 26dB 6dB
NOTE The Rates available are dependent on the HP 37718A/19A option
Equalization at 1/2 Bit Rate

Description

The signal from the HP 37718A/19A PDH Transmitter is applied to the PDH Receiver after attenuating by an amount equal to the selected Receiver Flat Loss plus the specified Cable Loss. The Flat Loss is obtained from a 75Ω, attenuator kit. The Cable Loss is supplied by inserting the correct Cable Simulators for each bit rate in the attenuation path.

Equipment Required

Cable Simulator E4-M 40 metres of 8120-0049 Cable Simulator E3-M 60 metres of 8120-0049 Cable Simulator E2-M 40 metres of 8120-0049 Cable Simulator E1-M 60 metres of 8120-0049 Cable Simulator DS3 55 metres of 8120-0049 Cable Simulator DS1 80 metres of 8120-0049 75Attenuator Kit HP 86213A Type N/BNC adaptors HP 1250-1534 & -1528 (2 off each)
3-39
Page 80
Performance Tests
PDH/DSn Receiver Monitor Levels
Procedure
1 Recallthe HP 37718A/19A DEFAULT SETTINGS as shown on page 3-
4.
2 Connect the equipment as shown in Figure 3-1.
Figure 3-1 Receiver Monitor Input Test Setup
2.048 Mb/s (E1)
1 Press and set up the display as shown
TRANSMIT
MAIN SETTINGS
opposite.
2 Press and set up the display as shown
RECEIVE
MAIN
SETTINGS
3-40
Page 81
Performance Tests
PDH/DSn Receiver Monitor Levels
below.
3 Press 4 Press to start the measurement.
RESULTS
RUN/STOP
5 Ensurethat NO TROUBLE is displayed on the display after
TROUBLE SCAN
RESULTS
30 seconds.
6 Press to stop the measurement. 7 Press and select EQUALIZATION [ON] GAIN [20 dB].
RUN/STOP RECEIVE
8 Connect Cable Simulator E1-M between the PDH OUT Port and the
attenuator.
9 Press then to start the
RESULTS
TROUBLE SCAN
RUN/STOP
measurement.
10 Ensure that NOTROUBLEis displayed on the displayafter
30 seconds. Press to stop the measurement.
RUN/STOP
RESULTS
11 Connect the 6dB Fixed Attenuator in the signal path to give a total
path attenuation of 26dB.
12 Press and select MONITOR LEVEL [26dB]. 13 Press ; then to start the
RECEIVE RESULTS
TROUBLE SCAN
RUN/STOP
measurement.
14 Ensure that NOTROUBLEis displayed on the displayafter
30 seconds. Press to stop the measurement.
RUN/STOP
RESULTS
15 Remove Cable Simulator E1-M from the signal path.
3-41
Page 82
Performance Tests
PDH/DSn Receiver Monitor Levels
16 Press and select EQUALIZATION [OFF]. 17 Press ; then to start the
RECEIVE RESULTS
TROUBLE SCAN
RUN/STOP
measurement.
18 Ensure that NOTROUBLEis displayed on the displayafter
30 seconds. Press to stop the measurement.
RUN/STOP
RESULTS
19 Remove the 6dB fixed attenuator and connect the 10dB Fixed
Attenuator in the signal path to give a total path attenuation of 30dB.
20 Press and set MONITOR LEVEL [30dB]. 21 Press then to start the
RECEIVE RESULTS
TROUBLE SCAN
RUN/STOP
measurement.
22 Ensure that NOTROUBLEis displayed on the displayafter
30 seconds. Press to stop the measurement.
23 Press and select EQUALIZATION [ON].
RECEIVE
RUN/STOP
RESULTS
24 Connect Cable Simulator E1-M between the PDH OUT Port and the
attenuator.
25 Press . Press to start the
RESULTS
TROUBLE SCAN
RUN/STOP
measurement.
26 Ensure that NOTROUBLEis displayed on the displayafter
30 seconds. Press to stop the measurement.
RUN/STOP
RESULTS
8.448 Mb/s (E2)
27 Repeat steps 1 to 26 (20dB, 26dB and 30dB tests) with the HP 37718A/
19A and displays set to SIGNAL [8 Mb/s] and
TRANSMIT
RECEIVE
Cable Simulator E2-M fitted in place of Cable Simulator E1-M.
34.368 Mb/s (E3)
28 Repeat steps 1 to 18 (20dB and 26dB tests) with the HP 37718A/19A
TRANSMIT
and displaysset to SIGNAL[34 Mb/s] and Cable
RECEIVE
Simulator E3-M fitted in place of Cable Simulator E1-M.
139.264 Mb/s (E4)
29 Repeat steps 1 to 18 (20dB and 26dB tests) with the HP 37718A/19A
TRANSMIT
and displays set to SIGNAL [140 Mb/s] and
RECEIVE
Cable Simulator E4-M fitted in place of Cable Simulator E1-M.
3-42
Page 83
Performance Tests
PDH/DSn Receiver Monitor Levels
1.544 Mb/s (DS-1)
30 Repeat steps 1 to 26 (20dB, 26dB and 30dB tests) with the HP 37718A/
19A and displays set to SIGNAL [DSX-1].
TRANSMIT
RECEIVE
Connect Cable Simulator DS1 between two Balanced to Unbalanced converters and fit in place of Cable Simulator E1-M.
44.736 Mb/s (DS-3)
31 Repeat steps 1 to 18 (20dB and 26dB tests) with the HP 37718A/19A
TRANSMIT
and displays set to SIGNAL [DSX-3] and Cable
RECEIVE
Simulator DS3 fitted in place of Cable Simulator E1-M.
32 Disconnect all the test equipment.
3-43
Page 84
Performance Tests

External Mux/Demux

External Mux/Demux
Specifications
Rate Interface Source
2.048 Mb/s (E1) Meets ITU-T Rec. G.703for unbalanced coaxial pair.
1.544 Mb/s (DS-1) Meets T1.102-1993 for Balanced pair.
NOTE The Mux/Demux Rate available is dependent on the HP 37718A/19A
option.

Description

This test verifies operation of the PDH/DSn Mux and demux hardware and confirms the output characteristics of the external demux port on the PDH/DSn Receiver. The PDH/DSn Test Set is set up to transmit an unframed 2Mb/s/DS-1 pattern. This is applied to the HP 37718A/19A INSERT Port. The 2Mb/s/ DS-1 signal is multiplexed into a 34Mb/s/DS-3 data stream. The HP 37718A/19A transmitter and receiver are looped. The unframed 2Mb/s/ DS-1 signal is Demultiplexed from the 34Mb/s/DS-3 data stream to the PDH/DSn Test Set via the HP 37718A/19A DROP port. A BER test is performed to verify the integrity of the 2 Mb/s/DS-1 signal. The output from the DROP Port is then applied to an Oscilloscope, and the waveform characteristics are checked to ensure they meet specifications.
Acceptsa 2 Mb/s Unbalanced signal conforming to ITU-T Rec. G.703.
Accepts a DS-1 balanced signal conforming to T1.102-
1993.

Equipment Required

Oscilloscope : HP 54520A Option 001 PDH Test Set : HP 37718A/19A
75 Termination T Connector : HP 1250-0781
3-44
: HP 15522-80010
Page 85
Performance Tests
External Mux/Demux
NOTE The TestSet used in this procedure is an 37718A/37719A. Any other PDH
TestSet, capable of generating and measuring at 2 Mb/s/DS-1 unframed, can be used.
Procedure 2Mb/s Mux/Demux
1 Recallthe HP 37718A/19A DEFAULT SETTINGS as shown on page 3-
4.
2 Connect the equipment as shown in Figure 2.
Figure 2 External Mux/Demux Test Setup
3-45
Page 86
Performance Tests
External Mux/Demux
3 Press on the Test Set and set up the display as shown
TRANSMIT
below.
4 Press on the Test Set and set up the display as shown below.
RECEIVE
3-46
Page 87
Performance Tests
External Mux/Demux
5 Press on the HP 37718A/19A and set up the
6 Press on the HP 37718A/19A and set up the
TRANSMIT
MAIN SETTINGS
TRANSMIT
STRUCTURED
display as shown below.
SETTINGS
display as shown below.
3-47
Page 88
Performance Tests
External Mux/Demux
7 Press on the HP 37718A/19A and set up the
8 Press on the HP 37718A/19A and set up the
SETTINGS
RECEIVE
MAIN SETTINGS
RECEIVE
display as shown below.
display as shown below.
STRUCTURED
9 Press on the TestSet. Press on
RUN/STOP
RESULTS
TROUBLE SCAN
the Test Set and ensure that NO TROUBLE is displayed.
10 Press SINGLE error add key on the TestSet and ensure the Bit Error
count increments by one each time the key is pressed.
3-48
Page 89
Performance Tests
External Mux/Demux
2.048 Mb/s Positive Pulse
11 Disconnect the UUT 75DROP port from the Test Set and connect to
the Oscilloscope. Terminate in 75 at the Oscilloscope input.
12 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
13 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS_1E in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
14 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
15 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
16 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.
2.048 Mb/s Negative Pulse
17 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK
3-49
Page 90
Performance Tests
External Mux/Demux
b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
18 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS_1E in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
19 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8 b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
20 The Oscilloscope will automatically display and compare an isolated
positivepulsewith the mask limits.A PASSmessage should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
21 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.
DS-1 Mux and Demux
1 Connect the UUT balanced DS-1 Insert and Drop ports to the DS-1
Balanced Ports of the Test Set and repeat steps 3 through 10 substituting DS-1 for 2Mb/s and DS-3 for 34Mb/s.
3-50
Page 91
Performance Tests
External Mux/Demux
DS-1 Positive Pulse
2 Disconnect the UUT 100 DROP port from the Test Set and connect
to the Oscilloscope via the HP 15508B Balanced to unbalanced Converter. Terminate in 75 at the Oscilloscope input.
3 Select CHAN 1 on the Oscilloscope and set to 1M input impedance.
Load the Telecom Mask
4 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS1_NEW in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
5 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8, (Ensure CH1 Impedance is reset to 1MΩ. Select TRIGGER, SETUP and set softkey to ).
PATTERN TRIGGER Trig’d
b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
NOTE Using the balanced to unbalanced converter reduces the pulse amplitude
by a factor of 1.266. This will cause the comparison with the mask to fail. Toovercome this select CHAN 1 ON and reduce the Volts/Divisionsetting (highlightedonthe softkey display). Reduced Volts/Division= highlighted value/1.266. The new value can be entered using the keypad
6 TheOscilloscope will compare the positive pulse with the mask limits.
A PASS message should appear on the Oscilloscope.
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
3-51
Page 92
Performance Tests
External Mux/Demux
7 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 1.896V and 2.844V.
DS-1 Negative Pulse
8 On the Oscilloscope, use the following key sequence to access the
inverted mask. a/ SHIFT (Blue Key) DISK b/ softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK d/ softkey
CHANGE DIRECTORY
Load the Telecom Mask
9 Onthe Oscilloscope, use the following key sequence to select and load
the required Telecom Mask: a/ SHIFT (Blue Key) APPLICATION. b/ softkey.
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS1_NEW in the highlighted MSK FILES window. d/ softkey.
LOAD SETUP 8 M1M2
Compare Pulse with Mask
10 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask: a/ RECALL 8, (Ensure CH1 Impedance is reset to 1MΩ. Select TRIGGER, SETUP and set softkey to ).
PATTERN TRIGGER Trig’d
b/ DEFINE MEAS c/ softkey
COMPARE
d/ softkey
TEST ON
e/ softkey
M1M2
f/ softkey
AFTER FAIL CONTINUE
g/ RUN
NOTE Using the balanced to unbalanced converter reduces the pulse amplitude
by a factor of 1.266. This will cause the comparison with the mask to fail. Toovercome this select CHAN 1 ON and reduce the Volts/Divisionsetting (highlightedonthe softkey display). Reduced Volts/Division= highlighted value/1.266. The new value can be entered using the keypad
11 The Oscilloscope will compare the negative pulse with the mask limits.
A PASS message should appear on the Oscilloscope.
3-52
Page 93
Performance Tests
External Mux/Demux
NOTE It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS on the Mask.
12 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 1.896V and 2.844V.
13 Disconnect all test equipment.
3-53
Page 94
Performance Tests

PDH/DSn Frequency Measurement and Looped Clock

PDH/DSn Frequency Measurement and Looped Clock
Specifications
Accuracy 4.5 ppm Measured Offset 100 ppm
NOTE The Rates available are dependent on the HP 37718A/19A option
Description
This test verifies that the HP 37718A/19A Receiver Frequency Measurement is within specified limits. These limits assume the instrument is within the calibration cycle.
A Synthesizer is used to generate a sinewave at half the data rate. This is applied to the HP 37718A/19A Receiver Signal In port. As this signal corresponds to an All Ones Ternary Signal, the HP 37718A/19A receiver should sync up with no errors if set to PATTERN [ALL ONES]. The Frequency Measurement accuracy of the HP 37718A/19A can be determined by comparison with the frequency displayed on the Synthesizer. Frequency Offset Measurement is also verified during this test as the HP 37718A/19A will display deviation from the expected Signal In frequency in ppm. The PDH transmitter recovered clock function is also verified at 2 Mb/s using the frequency counter in RATIO mode.

Equipment Required

Synthesizer Frequency Counter : HP 5335A Option 010
Procedure
1 Recallthe HP 37718A/19A DEFAULT SETTINGS as shown on page 3-
4.
2 Connect up the equipment as shown in Figure 3-3.
3-54
: HP 3335A Option 001 (75)
Page 95
Performance Tests
PDH/DSn Frequency Measurement and Looped Clock
Figure 3-3 Receiver Frequency Measurement Test Setup
3 Set the Synthesizer to:
FREQUENCY - 1024.000 KHz sinewave AMPLITUDE - +10dBm.
4 Set the frequency counter to RATIO A/B. 5 Press and set up the display as shown below.
RECEIVE
3-55
Page 96
Performance Tests
PDH/DSn Frequency Measurement and Looped Clock
6 Press and set up the display as shown below.
7 Press and set up the display as shown below.
TRANSMIT
RESULTS
8 Verify that the FREQUENCY displayed is between 2047991 Hz and
2048009 Hz.
9 Verify that the Offset displayed is between +4.5 ppm and -4.5 ppm. 10 Set the synthesizer frequency to 1024.102 kHz and verify that the
3-56
Page 97
Performance Tests
PDH/DSn Frequency Measurement and Looped Clock
frequencydisplayedon the display is between 2048.194 kHz
RESULTS
and 2048.214 kHz.
11 Verify that the Offset displayed is between 95.5 ppm and 104.5 ppm.
The frequency counter should read 1.00.
12 Set the synthesizer frequency to 1023.898 kHz and verify that the
frequencydisplayedon the display is between 2047.786 kHz
RESULTS
and 2047.806 kHz.
13 Verify that the Offset displayed is between -95.5 ppm and -104.5 ppm.
The frequency counter should read 1.00.
14 Set the synthesizer level to +10 dBm, synthesizer frequency and the
HP 37718A/19A to the settings given in Table 3-1 and verify the displayed Offset at each point.
Table 3-1 PDH/DSn Offsets Receive
Frequency
Synthesizer Frequency
Displayed Offset
8 Mb/s (E2) 4224,000 Hz -4.5 to +4.5 ppm 8 Mb/s (E2) 4223,578 Hz -95.5 to -104.5 ppm 8 Mb/s (E2) 4224,422 Hz +95.5 to +104.5 ppm 34 Mb/s (E3) 17,184,000 Hz -4.5 to +4.5 ppm 34 Mb/s (E3) 17,182,282 Hz -95.5 to -104.5 ppm 34 Mb/s (E3) 17,185,718 Hz +95.5 to +104.5 ppm 140 Mb/s (E4) 69,632,000 Hz -4.5 to +4.5 ppm 140 Mb/s (E4) 69,625,036 Hz -95.5 to -104.5 ppm 140 Mb/s E4) 69,638,963 Hz +95.5 to +104.5 ppm
1.544 Mb/s (DS-1) 772,000 Hz -4.5 to +4.5 ppm
1.544 Mb/s (DS-1) 771,923 Hz -95.5 to -104.5 ppm
1.544 Mb/s (DS-1) 772,077 Hz +95.5 to +104.5 ppm
44.736 Mb/s (DS-3) 22,368,000 Hz -4.5 to +4.5 ppm
44.736 Mb/s (DS-3) 22,365,763 Hz -95.5 to -104.5 ppm
44.736 Mb/s (DS-3) 22,370,237 Hz +95.5 to +104.5 ppm
3-57
Page 98
Performance Tests

SDH/SONET Transmitter Clock Accuracy

SDH/SONET Transmitter Clock Accuracy
Specification
Bit Rate Accuracy
51.84 Mb/s
4.5 ppm

Description

The test uses a Frequency Counter connected to the Multirate Analyzer module Clock Trigger output port. This output is derived from the Clock module internal 10MHz clock oscillator providing a 51.84 MHz that is directly related to all SDH/SONET output rates.

Equipment Required

Frequency Counter : HP 5335A Option 010

Procedure

1 Recallthe HP 37718A/19A DEFAULT SETTINGS as shown on page 3-
4.
2 Set the Transmitter Output to SDH/SONET. 3 Connect the HP 37718A/19A Multirate Analyzer module CLOCK
TRIGGER port to the Frequency Counter Input A, set the input termination to 50Ω.
4 Adjustthe Frequency Counter Trigger Level toobtain a stable reading
and ensure that the Frequency Counter reads between 51.839670 MHz and 51.840230 MHz.
5 Disconnect all the test equipment.
3-58
Page 99
Performance Tests

External Clock/Data Reference Inputs & Clock Reference Output

External Clock/Data Reference Inputs & Clock Reference Output
Specifications
Clock Rate Description
2.048 Mb/s MTS Accepts timing reference as per ITU-T G.811 10 MHz Reference Accepts 10 MHz timing reference
1.544 Mb/s BITS Accepts DS-1 timing reference as per TA-TSY­000378
64 kb/s Accepts 64 kb/s timing reference as per ITU-T G.703
Section 1.2.2
STM-1/STS-3 Receive Recovers clock from received STM/STS input signal.
NOTE The Rates available are dependent on the HP 37718A/19A option
Description
This test verifies that signal integrity is maintained when an EXTERNAL clock is used as a reference.
Equipment Required
Synthesizer PDH/DSn Test Set : HP 37718A/19A 110/75 Balanced to
Unbalanced Converter Frequency Counter : HP 5335A Opt 010 64kb/s Test Set : HP 37732A
: HP 3335A option 001 (75)
: HP 15508B
3-59
Page 100
Performance Tests
External Clock/Data Reference Inputs & Clock Reference Output
Procedure MTS Clock - Clock Format
1 Connectthe 52/155 Mb/s DATAOUT port to the 52/155 Mb/s DATAIN
port.
2 Recallthe HP 37718A/19A Default Settings as shown on page 3-4 and
set the display as shown below.
OTHER
SETTINGS CONTROL
3 Press and set up the display as shown below.
TRANSMIT
4 Set the HP 3335A to 2.048 MHz at +10 dBm.
3-60
Loading...