HP 3325B, 3325A User Manual

*TB 9-6625-2049-35
SUPERSEDED COPY DATED 15 JANUARY 1996
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
SYNTHESIZER/FUNCTION GENERATOR
MODELS 3325A( ) AND 3325B( )
Headquarters, Department of the Army, Washington, DC
2 January 2001
Approved for public release; distribution is unlimited
REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS
You can help improve this publication. If you find any mistakes or if you know of a way to improve the procedure, please let us know. Mail your letter or DA Form 2028 to: Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-MMC-LS-LP, Redstone Arsenal, AL 35898-5230. A reply will be furnished to you. You may also send in your comments electronically to our e-mail address: ls-lp@redstone.army.mil or by FAX (256) 842-6546/DSN 788-6546
SECTION I. IDENTIFICATION AND DESCRIPTION
__________ *This bulletin supersedes TB 9-6625-2049-35, 15 January 1996.
Test instrument identification............................
Forms, records, and reports................................
Calibration description .......................................
II. EQUIPMENT REQUIREMENTS
Equipment required ............................................
Accessories required ...........................................
III CALIBRATION PROCESS
Preliminary instructions.....................................
Equipment setup .................................................
Harmonic distortion............................................
Amplitude modulation distortion........................
Square wave characteristics...............................
Ramp retrace.......................................................
Frequency accuracy............................................
Phase increment .................................................
Sine wave amplitude and flatness......................
Square wave amplitude and flatness..................
Paragraph
1 2 2 2 3 2
4 4 5 5
6 6 7 6 8 7
9 9 10 9 11 10 12 11 13 13 14 14 15 16
Page
TB 9-6625-2049-35
SECTION II
I.
Triangle and ramp amplitude ............................
Dc offset...............................................................
Dc offset with ac..................................................
High voltage output option 002 only...................
Power supply.......................................................
Final procedure...................................................
CALIBRATION PROCESS - Continued
SECTION I
Paragraph Page
16 17 17 18
18 19 19 20 20 21 21 22
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the
calibration of Synthesizer/Function Generator, Hewlett-Packard, Models 3325A( ) and 3325B( ). The manufacturer's manuals and TM 11-6625-3065-14 were used as the prime data sources in compiling these instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Option 001: High Stability Frequency Reference; Option 002: High Voltage Output; Option E04: includes Options 001, 002, and instrument front handles and chassis slides; Option 907: Front Handle Assembly; Option 908: Rack Mount Flange Kit; Option 909: Rack Mount Flange Kit/Front Handle Assembly; Option 910: Additional Operating and Service Manual. Variations among models are described in text, tables, and figures.
b. Time and Technique. The time required for this calibration is approximately 4 hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records and reports required for calibration personnel at all levels are prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which they appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are in table 1.
2
Table 1. Calibration Description
Test instrument parameters Performance specifications Harmonic distortion1 (relative to fundamental at full output)
Amplitude modulation Square wave
Ramp retrace Frequency Accuracy: 5 x 10-6 of selected value Sine wave
Square wave Triangle Positive slope ramp Negative slope ramp Phase offset
Amplitude
Amplitude accuracy with no Sine wave: .001 Hz to 100 kHz2 attenuation (attenuator range 1) into 50 load (no dc offset )
Ramps: .001 Hz to 500 Hz
Flatness with no attenuation (attenuator range 1) into a 50Ω load
Square wave: 100 kHz to 10 MHz
See footnotes at end of table.
Fundamental frequency: 0.1 Hz To 50 kHz Accuracy: No harmonic greater than -65 dB Fundamental frequency: 50 to 200 kHz Accuracy: No harmonic greater than -60 dB Fundamental frequency: 200 kHz to 2 MHz (200 kHz to 1 MHz for option 002) Accuracy: No harmonic greater than -40 dB Fundamental frequency: 2 to 15 MHz Accuracy: No harmonic greater than -30 dB Fundamental frequency: 15 to 20 MHz Accuracy: No harmonic greater than -25 dB AM distortion: ≤-30 dB at 80% modulation, 10 kHz, and 0 V dc offset Rise time and fall time: 20 ns, (125 ns for option 002) at full output Symmetry: .02% of period +3 ns Overshoot: 5% of p-p amplitude at full output (<10% of peak amplitude for option 002) 3 µs retrace time, positive or negative ramps
Range: 1 µHz to 20 999 999.999 Hz (option 002: .02 Hz to 25.6 kHz) Range: 1 µHz to 10 999 999.999 Hz (option 002: .02 Hz to 25.6 kHz) Range: 1 µHz to 10 999 999.999 Hz (option 002: .02 Hz to 10 kHz) Range: 1 µHz to 10 999 999.999 Hz (option 002: .02 Hz to 10 kHz) Range: 1 µHz to 10 999 999.999 Hz (option 002: .02 Hz to 10 kHz) Range: Variable ±719.9° with respect to arbitrary starting phase, or assigned zero phase Accuracy: ±0.2°
Tolerance relative to
Square wave: .001 Hz to 100 kHz Triangle: .001 Hz to 2 kHz 2 kHz to 10 kHz
500 Hz to 10 kHz
Sine wave: 100 kHz to 20 MHz
TB 9-6625-2049-35
programmed amplitude
±0.1 dB ±1.0% ±1.5% ±5% ±1.5% ±10%
Tolerance relative to
programmed amplitude at 1 kHz
±0.3 dB ±10%
3
TB 9-6625-2049-35
Table 1. Calibration Description - Continued
Test instrument parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 load Sine wave: .001 Hz to 100 kHz Square: .001 Hz to 100 kHz Triangle: .001 Hz to 2 kHz
Ramps: .001 Hz to 500 Hz
Attenuator accuracy (these errors are additive with the amplitude accuracy errors)
Tolerance relative to
±0.3 dB ±3% ±4%
2 kHz to 10 kHz
500 Hz to 10 kHz
Function and frequency range .001 Hz to 100 kHz Attenuator ranges 2 through 8
±6% ±4% ±11%
±0.1 dB
programmed amplitude
Tolerance relative to
programmed amplitude
100 kHz to 10 MHz
Attenuator ranges 2 through 8
10 MHz to 20 MHz
Attenuator ranges 2 through 4 Attenuator ranges 5 through 8
Amplitude output Range: 1.000 mV to 10.00 V p-p Amplitude (option 002) (high voltage output)
Dc offset Range: +5 V dc
Dc offset (option 002) Range: +20 V dc
Dc plus ac Range: <1 MHz
1
Not verified below 50 Hz.
2
Except lowest attenuator range where accuracy is ±20 µV.
Range: 4 mV to 40 V p-p (>500) Accuracy: +2% of full output for each range at 2 kHz Flatness: +10% of programmed amplitude
Accuracy: +0.4% of full peak output for each attenuator range2
Accuracy: + (1% +25 mV) of full output for each attenuator range
Accuracy: +1.2% Ramps: +2.4% Range: >1 MHz Accuracy: +3%
±0.2 dB
0.2 dB
±0.5 dB
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2. identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The item selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the standard and TI.
4
TB 9-6625-2049-35
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration procedure. The following peculiar accessory is also required for this calibration: 50 feedthrough termination, Hewlett-Packard, Model 11048C.
Table 2. Minimum Specifications of Equipment Required
Common name
AUDIO ANALYZER Output: 4.5 V, 10 kHz
AM distortion: ≤-30 dB at 10 kHz Distortion: -60 dB at 50 Hz
FREQUENCY COUNTER Range: 99,9950 µs to 60,000,300 Hz
Accuracy: 1.25 x 10 Capability: Time interval A to B
MEASURING RECEIVER Frequency: 100 kHz to 10 MHz
Volts: .683 to .732 V rms Accuracy: ±0.85% AM: 0 to 80%
MULTIMETER Range: -20.225 to +20.225 V dc
Accuracy: ±0.05% Range: 0.3416 to 20.4 V ac 100 Hz to 100 kHz Accuracy: ±0.3%
OSCILLOSCOPE Frequency: 1 kHz to 10 MHz
Amplitude: 1.8 to 2.2 V p-p Accuracy: ±2.5% Capabilities Duty cycle: 49.7 to 50.3% Overshoot: 5% Rise time and fall time: <20 ns
SPECTRUM ANALYZER Frequency: 50 kHz to 20 MHz
Input: -20 dBm and 40 V
Display capability: <-65 dBc SYNTHESIZER/LEVEL GENERATOR
TIME FREQUENCY WORKSTATION RESISTANCE STANDARD Range: 470 Biddle Gray, Model 71-631
Frequency: 0.1 MHz
Amplitude: 13 dBm
Reference output frequency: 10 MHz
Accuracy: ±5 x 10-8
Minimum use specifications
-8
Manufacturer and model
(part number) Boonton, Model 1120-S/10 (MIS-35954/2)
Hewlett-Packard, Model 5345A (MIS-28754/1 Type 1)
Consisting of : Measuring receiver Hewlett-Packard, Model 8902A and Sensor module Hewlett-Packard, Model 11722A
Hewlett-Packard, Model 3458A
Tektronix, Type 2430A (OS-291/G)
(AN/USM-489A(V)1)
Hewlett-Packard, Model 3335AOPT 001-K06 (MIS-35938) Autek Systems Corporation, Model 620 (MIS-38946)
(7910328)
5
TB 9-6625-2049-35
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration process. Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test requirement is not met, take corrective action before continuing with the calibration. Adjustments required to calibrate the TI are included in this procedure. Additional maintenance information is contained in TM 11-6625-3065-14 and the manufacturer's manual for this TI.
d. When indications specified in paragraphs 8 through 19 are not within tolerance, perform the power supply check prior to making adjustments. After adjustments are made, repeat paragraphs 8 through 19. Do not perform power supply check if all other parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup WARNING
HIGH VOLTAGE is used or exposed during the performance of this calibration. DEATH ON CONTACT may result if personnel fail to observe safety precautions. REDUCE OUTPUT(S) to minimum after each step within the performance check where applicable.
CAUTION
Before connecting TI to power source, make sure TI is set to the power source line voltage shown on rear of TI.
a. Connect TI to a 115 V ac source. Press POWER pushbutton to ON and allow at least 24 hours (72 hours for option 001) for warmup.
b. Before continuing, review (1) through (4) below: (1) The SWEEP LINEAR/LOG, ENTRY, FUNCTION, SIGNAL, and blue keys
each have an indicator denoting that it is activated. (2) When power is applied to the TI, some keys will be lit. (3) Pressing any FUNCTION key that is active will delete that function and no ac signal will be present at the output. (4) Some instructions will be repeated during programming. This is done to ensure correct output is available.
6
TB 9-6625-2049-35
8. Harmonic Distortion
a. Performance Check (1) Connect TI EXT REF IN 1, 10 MHz to spectrum analyzer 10 MHz REF
IN/OUT.
(2) Connect TI SIGNAL to spectrum analyzer INPUT 50ΩΩ. (3) Press keys and enter values using DATA keys as listed in (a) through (h) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY AMPTD. (d) -20 dBm. (e) ENTRY DC OFFSET. (f) 0 VOLT. (g) ENTRY FREQ. (h) 20 MHz.
(4) Adjust spectrum analyzer controls to display 20 MHz fundamental and at least four harmonics. All harmonics will be at least 25 dB below fundamental.
(5) Enter 14 MHz using TI DATA keys. Adjust spectrum analyzer controls to display 14 MHz fundamental and at least four harmonics. All harmonics will be at least 30 dB below fundamental.
(6) Enter 1.9 MHz using TI DATA keys. Adjust spectrum analyzer controls to display 1.9 MHz fundamental and at least four harmonics. All harmonics will be at least 40 dB below fundamental.
(7) Enter 190 kHz using TI DATA keys. Adjust spectrum analyzer controls to display 190 kHz fundamental and at least four harmonics. All harmonics will be at least 60 dB below fundamental.
(8) Enter 100 kHz using TI DATA keys. Adjust spectrum analyzer controls to display 100 kHz fundamental and at least four harmonics. All harmonics will be at least 60 dB below fundamental.
(9) Enter 40 kHz using TI DATA keys. Adjust spectrum analyzer controls to display 40 kHz fundamental and at least four harmonics. All harmonics will be at least 65 dB below fundamental.
(10) Disconnect TI from spectrum analyzer. (11) Connect TI SIGNAL to audio analyzer INPUT HIGH using 50 feedthrough
termination. (12) Press keys and enter values using DATA keys as listed in (a) through (c) below:
7
TB 9-6625-2049-35
(a) 50 Hz. (b) ENTRY AMPTD. (c) 10 VOLT.
(13) Set audio analyzer to measure distortion in dB. Audio analyzer indication will be -65 dB.
(14) Perform (15) through (19) below for option 002 only. (15) Connect equipment as shown in figure 1 and set resistance standard to 470Ω.
Figure 1. Harmonic distortion - equipment setup.
(16) Press keys and enter values using DATA keys as listed in (a) through (e) below:
(a) SIGNAL on. (b) ENTRY FREQ. (c) 25 kHz. (d) ENTRY AMPTD. (e) 40 VOLT.
(17) Adjust spectrum analyzer controls to display 25 kHz fundamental and at least four harmonics. All harmonics will be at least 65 dB below fundamental.
(18) Press ENTRY FREQ key and enter 190 kHz using DATA keys. Adjust spectrum analyzer controls to display 190 kHz fundamental and at least four harmonics. All harmonics will be at least 60 dB below fundamental.
(19) Enter 1 MHz using DATA keys. Adjust spectrum analyzer controls to display 1 MHz fundamental and at least four harmonics. All harmonics will be at least 40 dB below fundamental.
b. Adjustment. No adjustments can be made.
8
TB 9-6625-2049-35
9. Amplitude Modulation Distortion
a. Performance Check
(1) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ. (d) 1 MHz. (e) ENTRY AMPTD. (f) 3 VOLT. (g) ENTRY DC OFFSET. (h) 0 VOLT. (i) Blue then AM ON.
(2) Connect equipment as shown in figure 2.
Figure 2. AM distortion - equipment setup.
(3) Set audio analyzer for a 50 (SPCL 75), 10 kHz, 2.5 V output. (4) Set measuring receiver to measure AM. (5) Adjust audio analyzer level until measuring receiver indicates 80%. (6) Set audio analyzer to measure distortion in dB. Audio analyzer indication will
be -30 dB. (7) Press TI blue key then DATA AM OFF key.
b. Adjustments. No adjustments can be made.
10. Square Wave Characteristics
a. Performance Check (1) Press keys and enter values using DATA keys as listed in (a) through (f) below:
9
TB 9-6625-2049-35
(a) SIGNAL off (option 002 not lit). (b) FUNCTION square wave. (c) ENTRY FREQ. (d) 1 MHz. (e) ENTRY AMPTD. (f) 1 V RMS.
(2) Connect TI SIGNAL to oscilloscope CH 1. (3) Set oscilloscope CH 1 COUPLING 50ΩΩ to ON. (4) Set oscilloscope controls for duty cycle measurement. Duty cycle will be between
49.7 and 50.3 percent.
(5) Enter 10 VOLT using DATA keys. (6) Set oscilloscope controls for rise time measurement. Rise time will be 20 ns. (7) Repeat (6) above for fall time. (8) Set oscilloscope controls for overshoot measurement. Overshoot will be 5
percent of peak to peak amplitude (500 mV at positive and negative peaks). (9) Perform (10) through (12) below for option 002 only. (10) Set oscilloscope CH 1 COUPLING 50ΩΩ to OFF. (11) Press TI SIGNAL key on (lit). (12)
125 ns with a
Repeat technique in (6) through (8) above. Rise time and fall time will be
n overshoot <10 percent of peak amplitude (<500 mV at positive and
negative peaks).
b. Adjustments. No adjustments can be made.
11. Ramp Retrace
a. Performance Check
(1) Connect TI SIGNAL to oscilloscope CH 1. (2) Set oscilloscope CH 1 COUPLING 50ΩΩ to ON. (3) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) SIGNAL off (option 002 not lit).
(b) FUNCTION positive ramp. (c) ENTRY FREQ. (d) 10 kHz. (e) ENTRY AMPTD. (f) 10 VOLT.
10
TB 9-6625-2049-35
(4) Set oscilloscope controls to measure ramp retrace time from the 90 to 10 percent points. Ramp retrace time will be 3 µs.
(5) Press FUNCTION negative ramp key and repeat (4) above.
b. Adjustments. No adjustments can be made.
12. Frequency Accuracy
a. Performance Check
(1) Connect TI SIGNAL to frequency counter CHANNEL A. (2) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ. (d) 20 MHz. (e) ENTRY AMPTD. (f) .99 VOLT.
(3) Set frequency counter 50ΩΩ/1MΩΩ switch to 50ΩΩ. (4) Set frequency counter controls to measure frequency. If frequency counter does
not indicate between 19.99990 and 20.00010 MHz, perform b below. (5) Press FUNCTION square wave key. Frequency counter will indicate between
9.999950 and 10.000050 MHz.
(6) Disconnect frequency counter from TI SIGNAL. (7) Connect frequency counter CHANNEL A to TI SYNC OUT. (8) Press TI FUNCTION triangle key. (9) Set frequency counter controls to measure period. Frequency counter will
indicate between 99.99950 and 100.00050 µs. (10) Press FUNCTION positive ramp key. Frequency counter will indicate between
99.99950 and 100.00050 µs.
(11) Press FUNCTION negative ramp key. Frequency counter will indicate between
99.99950 and 100.00050 µs.
(12) Disconnect frequency counter from TI SYNC OUT. (13) Connect frequency counter to TI AUX 21-60 MHz (rear panel). (14) Press keys and enter values using DATA keys as listed in (a) through (c) below: (a) FUNCTION sine wave. (b) ENTRY FREQ.
(c) 60 MHz.
11
TB 9-6625-2049-35
(15) Set frequency counter controls to measure frequency. If frequency counter does not indicate between 59.99970 and 60.00030 MHz, perform b below.
b. Adjustments
(1) Disconnect TI rear panel 10 MHz OVEN OUTPUT from EXT REF IN option 001 only.
(2) Connect frequency counter CHANNEL A to TI rear panel AUX 21-60 MHz. (3) Press keys and enter values using DATA keys as listed in (a) through (c) below: (a) FUNCTION sine wave. (b) ENTRY FREQ. (c) 60 MHz.
(4) Adjust REF R30 (fig. 3) until frequency counter indicates 60.000000 MHz (R).
Figure 3. Test instrument - bottom view.
(5) Disconnect frequency counter from TI rear panel AUX 21-60 MHz. (6) Connect frequency counter CHANNEL A to TI SIGNAL. (7) Enter 20 MHz using DATA keys. Frequency counter will indicate between
19.99990 and 20.00010 MHz.
(8) Disconnect frequency counter from TI. (9) Perform (10) through (14) below for option 001 only. (10) Connect time/frequency workstation OUTPUT 1 MHz to frequency difference
meter REF INPUT. (11) Connect TI rear panel 10 MHz OVEN OUTPUT to frequency difference meter
SIG INPUT.
12
TB 9-6625-2049-35
(12) Adjust A9R7 Fine Adj. (fig. 3) and A9 Coarse Adj. (fig. 3) for a minimum indication on frequency difference meter.
(13) Disconnect TI from frequency difference meter. (14) Reconnect TI rear panel 10 MHz OVEN OUTPUT to EXT REF IN.
13. Phase Increment
a. Performance Check
(1) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ. (d) 100 kHz. (e) ENTRY AMPTD. (f) 13 dBm.
(2) Connect equipment as shown in figure 4.
Figure 4. Phase increment - equipment setup.
NOTE
If TI has option 001, remove the connection from 10 MHz OUTPUT to EXT REF IN (rear panel) (fig. 4).
(3) Set synthesizer/level generator frequency to .1 MHz and amplitude to 13 dBm. (4) Set frequency counter controls for time interval A to B measurement. (5) Press keys as listed in (a) through (c) below: (a) ENTRY PHASE. (b) MODIFY until frequency counter indicates between 199 and 201 ns. (c) Blue then ASGN ZERO PHASE.
13
TB 9-6625-2049-35
(6) Adjust frequency counter sample rate control to HOLD and press RESET pushbutton. Record frequency counter indication.
(7) Press ENTRY PHASE key and enter -1 deg using DATA keys. (8) Press frequency counter RESET pushbutton and record frequency counter
indication. (9) Determine the difference between indication recorded in (6) above and indication
recorded in (8) above. The difference will be between 22.22 and 33.34 ns. (10) Enter -10 deg using DATA keys. (11) Press frequency counter RESET pushbutton and record frequency counter
indication. (12) Determine the difference between indication recorded in (6) above and indication
recorded in (11) above. The difference will be between 272.22 and 283.34 ns. (13) Enter -100 deg using DATA keys. (14) Press frequency counter RESET pushbutton and record frequency counter
indication. (15) Determine the difference between indication recorded in (6) above and indication
recorded in (14) above. The difference will be between 2722.22 and 2783.34 ns.
b. Adjustments. No adjustments can be made.
14. Sine wave Amplitude and Flatness
a. Performance Check
(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 feedthrough termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) SIGNAL off (option 002 not lit). (b) ENTRY DC OFFSET. (c) 0 VOLT. (d) FUNCTION sine wave. (e) ENTRY AMPTD. (f) 3.536 V RMS. (g) ENTRY FREQ. (h) 100 Hz. (i) AMPTD CAL.
(3) Multimeter will indicate between 3.495 and 3.577 V ac. (4) Repeat technique of (2)(e) through (i) and (3) above using TI settings and
multimeter indications listed in table 3.
14
TB 9-6625-2049-35
Table 3. Amplitude
Test instrument settings Multimeter indications
ENTRY AMPTD ENTRY FREQ
3.536 V RMS 1 kHz 3.495 3.577
3.536 V RMS 100 kHz 3.495 3.577
1.061 V RMS 100 kHz 1.048 1.073
1.061 V RMS 1 kHz 1.048 1.073
1.061 V RMS 100 Hz 1.048 1.073 .3536 V RMS .3536 V RMS 1 kHz 0.3416 0.3660 .3536 V RMS 100 kHz 0.3416 0.3660
1Press ENTRY DC OFFSET key and enter 1 mV using DATA keys.
1
100 Hz 0.3416 0.3660
(V ac)
(5) Press keys and enter values using DATA keys as listed in (a) through (g) below: (a) ENTRY DC OFFSET. (b) 0 mV. (c) ENTRY FREQ. (d) 1 kHz. (e) ENTRY AMPTD. (f) .707 V RMS. (g) AMPTD CAL.
(6) Press MODIFY keys for a multimeter indication as close as possible to 0.707 V ac.
(7) Enter 100 kHz using DATA keys. Multimeter will indicate between 0.683 and 0.732 V ac. Record multimeter indication.
(8) Disconnect multimeter and 50 feedthrough termination from TI.
NOTE
If necessary, ZERO and CALIBRATE measuring receiver and sensor module.
(9) Connect measuring receiver sensor module to TI SIGNAL. (10) Set measuring receiver to measure volts at .1 MHz. (11) Press ENTRY AMPTD key. (12) Press MODIFY keys for a measuring receiver indication as close as possible to
indication recorded in (7) above. (13) Press ENTRY FREQ key and enter 500 kHz using DATA keys. (14) Set measuring receiver to measure volts at .5 MHz. Measuring receiver will
indicate between 0.683 and 0.732 V. (15) Repeat technique of (13) and (14) above using TI and measuring receiver
settings listed in table 4. Measuring receiver will indicate between 0.683 and 0.732 V.
15
TB 9-6625-2049-35
Table 4. Flatness
Test instrument
and measuring receiver
settings (MHz)
1 2 4 6
8 10 12 14 16 18 20
b. Adjustments. No adjustments can be made.
15. Square Wave Amplitude and Flatness a. Performance Check
(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 feedthrough termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) SIGNAL off (option 002 not lit). (b) ENTRY DC OFFSET. (c) 0 VOLT. (d) FUNCTION square wave. (e) ENTRY AMPTD. (f) 5 V RMS. (g) ENTRY FREQ. (h) 100 Hz. (i) AMPTD CAL.
(3) Multimeter will indicate between 4.95 and 5.05 V ac. (4) Enter 1 kHz using DATA keys. Multimeter will indicate between 4.95 and 5.05
V ac. (5) Enter 100 kHz using DATA keys. Multimeter will indicate between 4.95 and
5.05 V ac. (6) Press keys and enter values using DATA keys as listed in (a) through (d) below:
16
TB 9-6625-2049-35
(a) 1 kHz. (b) ENTRY AMPTD. (c) 1 V RMS. (d) AMPTD CAL.
(7) Multimeter will indicate between 0.99 and 1.01 V ac. (8) Disconnect TI from multimeter. (9) Connect TI SIGNAL to oscilloscope CH 1 using 50 feedthrough termination. (10) Set oscilloscope controls to measure square wave amplitude. Record oscilloscope
indication. (11) Press ENTRY FREQ key and enter 1 MHz using DATA keys. (12) Set oscilloscope controls to measure square wave amplitude. Oscilloscope
indication will be within ±10 percent of indication recorded in (10) above. (13) Enter 3 MHz using DATA keys and repeat (12) above. (14) Enter 5 MHz using DATA keys and repeat (12) above. (15) Enter 10 MHz using DATA keys and repeat (12) above.
b. Adjustments. No adjustments can be made.
16. Triangle and Ramp Amplitude a. Performance Check
(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 feedthrough termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) SIGNAL off (option 002 not lit). (b) ENTRY DC OFFSET. (c) 0 VOLT. (d) FUNCTION triangle. (e) ENTRY AMPTD. (f) 2.887 V RMS. (g) ENTRY FREQ. (h) 100 Hz. (i) AMPTD CAL.
(3) Multimeter will indicate between 2.843 and 2.931 V ac. (4) Enter 1.9 kHz using DATA keys. Multimeter will indicate between 2.843 and
2.931 V ac. (5) Enter 2.1 kHz using DATA keys. Multimeter will indicate between 2.742 and
3.032 V ac.
17
TB 9-6625-2049-35
(6) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and
3.032 V ac. (7) Press keys and enter values using DATA keys as listed in (a) through (c) below: (a) FUNCTION positive ramp.
(b) 100 Hz. (c) AMPTD CAL.
(8) Multimeter will indicate between 2.843 and 2.931 V ac. (9) Enter 499 Hz using DATA keys. Multimeter will indicate between 2.843 and
2.931 V ac. (10) Enter 501 Hz using DATA keys. Multimeter will indicate between 2.598 and
3.176 V ac. (11) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.598 and
3.176V ac.
b. Adjustments. No adjustments can be made.
17. Dc Offset a. Performance Check
(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 feedthrough termination.
(2) Press SIGNAL key off (option 002 not lit). (3) Press presently active FUNCTION key to remove ac output and activate
ENTRY DC OFFSET key (lit). (4) Enter 5 VOLT using DATA keys and press AMPTD CAL key. (5) Multimeter will indicate between 4.98 and 5.02 V dc. (6)
and -5.02 V dc.
Enter
-5 VOLT
using
DATA
keys. Multimeter will indicate between -4.98
(7) Repeat technique of (6) above using TI settings and multimeter indications listed in table 5.
Table 5. Dc Offset
Test instrument
DC OFFSET
settings Min Max
1.499 VOLT 1.493 1.505
-1.499 VOLT -1.505 -1.493
499.9 mV 0.4979 0.5019
-499.9 mV -0.5019 -0.4979
149.9 mV 0.1493 0.1505
-149.9 mV -0.1505 -0.1493
49.99 mV 0.04979 0.05019
-49.99 mV -0.05019 -0.04979
Multimeter indications
(V dc)
18
TB 9-6625-2049-35
Table 5. Dc Offset - Continued
Test instrument
DC OFFSET
settings Min Max
14.99 mV 0.01493 0.01505
-14.99 mV -0.01505 -0.01493
4.999 mV 0.004979 0.005019
-4.999 mV -0.005019 -0.004979
1.499 mV 0.001479 0.001519
-1.499 mV -0.001519 -0.001479
Multimeter indications
(V dc)
NOTE
Perform (8) through (11) below for option 002 only.
(8) Remove 50 feedthrough termination from equipment setup. (9) Press SIGNAL key on. (10) Enter 20 VOLT using DATA keys. Multimeter will indicate between 19.775
and 20.225 V dc. (11) Enter -20 VOLT using DATA keys. Multimeter will indicate between -19.775
and -20.225 V dc.
b. Adjustments. No adjustments can be made.
18. Dc Offset with Ac a. Performance Check
(1) Connect SIGNAL to multimeter INPUT HI and LO using 50 feedthrough termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ. (d) 20.999 999 999 MHz. (e) ENTRY AMPTD. (f) 1 VOLT. (g) ENTRY DC OFFSET. (h) 4.5 VOLT. (i) AMPTD CAL.
(3) Multimeter will indicate between 4.35 and 4.65 V dc.
19
TB 9-6625-2049-35
(4) Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.35 and
-4.65 V dc. (5) Press ENTRY FREQ key and enter 999.9 kHz using DATA keys. Multimeter
will indicate between -4.44 and -4.56 V dc. (6) Press ENTRY DC OFFSET key and enter 4.5 VOLT using DATA keys.
Multimeter will indicate between 4.44 and 4.56 V dc. (7) Press FUNCTION square wave key. Multimeter will indicate between 4.44 and
4.56 V dc. (8) Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.44 and
-4.56 V dc. (9) Press ENTRY FREQ key and enter 9.9999 MHz using DATA keys.
Multimeter will indicate between -4.35 and -4.65 V dc. (10) Press FUNCTION triangle key and enter 9.9 kHz using DATA keys. (11) Multimeter will indicate between -4.44 and -4.56 V dc. (12) Press FUNCTION positive ramp key. Multimeter will indicate between -4.38
and -4.62 V dc.
b. Adjustments. No adjustments can be made.
19. High Voltage Output Option 002 only a. Performance Check
(1) Connect SIGNAL to multimeter INPUT HI and LO. (2) Press keys and enter values using DATA keys as listed in (a) through (i) below: (a) FUNCTION sine wave. (b) ENTRY FREQ. (c) 2 kHz. (d) SIGNAL on. (e) ENTRY DC OFFSET. (f) 0 VOLT. (g) ENTRY AMPTD. (h) 14.14 V RMS. (i) AMPTD CAL.
(3) Multimeter will indicate between 13.86 and 14.42 V ac. (4) Press keys and enter values using DATA keys as listed in (a) through (e) below:
20
TB 9-6625-2049-35
(a) ENTRY FREQ. (b) 100 kHz. (c) ENTRY AMPTD. (d) 10 V RMS. (e) AMPTD CAL.
(5) Multimeter will indicate between 9 and 11 V ac. (6) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) FUNCTION square wave. (b) ENTRY FREQ. (c) 2 kHz. (d) ENTRY AMPTD. (e) 20 V RMS. (f) AMPTD CAL.
(7) Multimeter will indicate between 19.6 and 20.4 V ac. (8) Press FUNCTION triangle key and enter 11.5 V RMS using DATA keys. (9) Press
AMPTD CAL
key. Multimeter will indicate between 11.27 and 11.73 V ac.
(10) Press FUNCTION positive ramp key and repeat (9) above.
b. Adjustments. No adjustments can be made.
20. Power Supply NOTE
Do not perform power supply check if all other parameters are within tolerance.
a. Performance Check
(1) Connect multimeter INPUT HI to -15 V TP (fig. 5) and LO to ground side of C9 (fig. 5). Adjust R22 (fig. 5) for a multimeter indication of -14.97 to -15.03 V dc (R).
(2) Connect multimeter INPUT HI to +15 V TP (fig. 5). Readjust R22 (fig. 5) if multimeter does not indicate between 14.9 and 15.1 V dc.
(3) Connect multimeter INPUT HI to +5 V TP (fig. 5). Readjust R22 (fig. 5) if multimeter does not indicate between 5.01 and 5.05 V dc.
(4) Repeat (1) through (3) above for best in-tolerance condition on all test points.
21
TB 9-6625-2049-35
Figure 5. Power supply assembly A2 – test instrument top view.
(5) Press presently active FUNCTION key to remove ac output and activate ENTRY DC OFFSET key (lit).
(6) Enter 0 VOLT using DATA keys and press AMPTD CAL key. (7) Connect multimeter INPUT HI to TP AMP OUT (fig. 3) and LO to circuit board
ground.
NOTE
The voltages measured in (1) through (3) above may be adjusted out of tolerance by (8) through (11) below. This is not a cause for concern.
(8) Adjust R40 (fig. 3) until digital multimeter indicates less than 5 mV dc (R). (9) Connect multimeter INPUT HI and LO to TI SIGNAL. Do not use 50
feedthrough termination. (10) Enter 5 VOLT using DATA keys. (11) Press AMPTD CAL key. Adjust R22 (fig. 5) until multimeter indicates 10.000
V dc. (12) Enter -5 VOLT using DATA keys. (13) Multimeter will indicate between -9.985 and -10.015 V dc.
21. Final Procedure
a. Deenergize and disconnect all equipment. b. Annotate and affix DA label/form in accordance with TB 750-25.
22
TB 9-6625-2049-35
By Order of the Secretary of the Army:
Official: ERIC K. SHINSEKI
General, United States Army Chief of Staff
JOEL B. HUDSON
Administrative Assistant to the Sec retary of the Army
DISTRIBUTION:
To be distributed in accordance with IDN 342200, requirements for calibration
procedure TB 9-6625-2049-35.
0029703
PIN: 052316-000
Loading...