Honeywell MultiPro User Manual

MULTIPRO PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
MultiPro
Reference Manual
Part Number 13-278
Version 3
Copyright 2012
by
Honeywell Analytics, Inc.
Lincolnshire, Illinois 60069
All rights reserved.
No page or part of this operation manual may be reproduced in any form without written
permission of the copyright owner shown above.
Honeywell Analytics reserves the right to correct typographical errors.
1
Table of Contents
CERTIFICATION INFORMATION 5 OPERATING TEMPERATURE 5
IGNAL WORDS 5
S
ARNINGS AND CAUTIONS 5
W
DESCRIPTION 8
1.
1.1 Methods of sampling 8
1.2 Multi-sensor capability 8
1.3 Calibration 8
1.4 Alarm logic 9
1.4.1 Atmospheric hazard alarms 9
1.4.2 Low battery alarms 9
1.4.3 Sensor over range alarms. 9
1.4.4 LEL response failure due to lack of O2 alarm 9
1.4.5 Security beep/flash 10
1.4.6 Latching peak alarms 10
1.4.7 Fault detection 10
1.5 Other electronic safeguards 10
1.5.1 Heartbeat 10
1.6 Sensors 10
1.7 Optional sample draw pump 10
1.7.1 Special precautions when using the MultiPro pump 11
1.8 Data storage 11
1.8.1 Black box data recorder 11
1.8.2 Event logger 11
1.9 MultiPro design components 11
1.10 MultiPro standard accessories 12
1.10.1 Alkaline MultiPro detectors 12
1.10.2 Li-Ion MultiPro detectors 12
1.11 MultiPro kits 12
1.11.1 MultiPro Confined Space Kits 12
1.11.2 MultiPro Value Packs 12
2. BASIC OPERATIONS 12
2.1Turning the MultiPro On 12
2.1.1 Start up with pump 14
2.2Operating Logic 14
2.3Turning the MultiPro Off 15
2.4 Alarms 15
2.4.1 Warning Alarms 15
2.4.2 Danger Alarms 15
2.4.3 STEL Alarms 15
2.4.4 TWA Alarms 15
2.4.5 Low battery alarms 15
2.4.6 Sensor over range alarms 16
2.4.7 LEL failure due to lack of oxygen alarm 16
2.5 PC connection via infrared port 16
2.6 Error Messages 17
2.7 Contact Cover 17
SAMPLING 17
3.
2
3.1 Manual sample draw kit 17
3.1.1 Manual sample draw kit usage 17
3.2 Motorized sample draw pump 18
3.2.1 Starting the motorized sample pump 18
3.2.2 Turning off the pump 19
3.2.3 Pump low flow alarm 19
3.3 Sample draw probe 19
CALIBRATION 19
4.
4.1 Functional (Bump) testing 20
4.2 Fresh Air/Zero Calibration 21
4.2.1Fresh air/zero calibration failure 21
4.2.2 Forced fresh air / zero calibration 21
4.3 Span Calibration 22
4.3.1Span calibration failure: Toxic and LEL sensors 22
4.3.2Span calibration failure: Oxygen sensors 23
5. MAINTENANCE 23
5.1 Cleaning 23
5.2 Storage 23
5.3 Batteries 24
5.4 Replacing alkaline batteries 24
5.5 Maintaining Li-Ion battery packs 24
5.5.1Storage guidelines for the Li-Ion versions 24
5.5.2 Charging guidelines for Li-Ion battery 25
5.5.3Charging procedure for Li-Ion battery 25
5.5.4 Charging with the pump attached. 25
5.5.5 Battery troubleshooting 25
5.6 Sensor replacement 25
5.6.1 Sensor replacement (Not Duo-Tox) 25
5.7 Sample probe assembly 26
5.7.1 Changing sample probe filters 27
5.7.2 Changing sample probe tubes (wands) 27
5.8 MultiPro Pump Maintenance 27
5.8.1 Replacing pump filters 27
6. DIRECT PROGRAMMING 27
6.1 Entering the Advanced Menu 27
6.2 Set options 28
6.3 Set time 28
6.4 Set date 29
6.5 Set alarms 29
6.6 Set cal gas 29
6.7 Set cal due 29
6.8 Set bump test due 29
6.9 Setup accept 30
PPENDICES 31
A
Appendix A Toxic gas measurement – Warning, Danger, STEL and TWA alarms 31
1. Warning and Danger Alarms 31
2. Time Weighted Average (TWA): 31
3. Short Term Exposure Limits (STEL): 31
Appendix B Calibration Frequency Recommendation 33
3
Appendix C MultiPro Sensor Information 33 Appendix D Toxic Sensor Cross-Sensitivity 33 Appendix E Basic Parts List 33
MultiPro Remote Sampling Accessories 33 MultiPro Sensors 33 MultiPro Charging & Datalogging Accessories 33
Appendix F Honeywell Analytics Standard Gas Detection Warranty 35
4
Certification Information
UL Class I Division 1 Groups A,B,C,D Temp Code T4 UL Class II, Division 1 Groups E,F,G UL Class III
CSA Class I, Division 1, Groups A,B,C,D Temp Code T4
(Per CSA C22.2 No. 152, only the combustible gas detection portion of this instrument has been assessed for performance)
ATEX Certification: II 2 G Ex ia d IIC T4 UL International DEMKO A/S 05 ATEX 0424116X
IECEx Certification: Ex ia d IIC T4 IECEx UL 07.0003X
Operating Temperature
The MultiPro’s operating temperature range is printed on the label on the back of the instrument. Use of Honeywell Analytics Gas Detectors outside of the instrument’s specified operating temperature range may result in inaccurate and potentially dangerous readings.
Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the MultiPro Reference Manual.
indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
indicates a potentially hazardous situation, which if not
avoided, may result in moderate or minor injury. CAUTION used without the safety alert symbol indicates a potentially
hazardous situation which, if not avoided, may result in property damage.
Warnings and Cautions
1. The MultiPro personal, portable gas detector has been
designed for the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously.
2.
established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
3.
whenever alkaline batteries are removed from the alkaline battery pack. Removal of the alkaline batteries from the battery pack in a hazardous area may impair intrinsic safety.
In the event of an alarm condition it is important to follow
The MultiPro must be located in a non-hazardous location
5
4.
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer
E91-LR6, Eveready EN91 batteries in the alkaline battery pack. Substitution of batteries may impair intrinsic safety.
5.
To reduce the risk of explosion, do not mix old or used
batteries with new batteries and do not mix batteries from different manufacturers.
6.
Do not charge the MultiPro with any charger other than the appropriate MultiPro charger. Standard versions of the MultiPro must be charged with the UL/CSA-approved charger, which is part number 54-49-103-1. European versions of the MultiPro must be charged with the ATEX-approved charger, which is Honeywell Analytics part number 54-49-103-5.
7.
The MultiPro must be located in a non-hazardous location
during the charging cycle. Charging the MultiPro in a hazardous location may impair intrinsic safety.
8.
MultiPro rechargeable battery packs are supplied with Panasonic CGA103450 Lithium-Ion batteries. These batteries are not replaceable by
the user. The rechargeable pack must be obtained from Honeywell Analytics and replaced as an assembly.
9.
The accuracy of the MultiPro should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. (The Canadian Standards Association (CSA) requires an accuracy check using known concentration calibration gas prior to each day’s use.)
10.
Fresh air/zero calibrations may only be performed in an atmosphere that is known to contain 20.9% oxygen, 0.0% LEL and 0 PPM toxic gas.
11.
The accuracy of the MultiPro should be checked immediately following any known exposure to contaminants by testing with known concentration test gas before further use. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
12.
The accuracy of the MultiPro should be checked immediately following any physical shock. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
13.
A sensor that cannot be calibrated or is found to be out of tolerance should be replaced immediately. An instrument that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use.
14.
Do not reset the calibration gas concentration unless you are using a calibration gas concentration that differs from the one that is normally supplied by Honeywell Analytics for use in calibrating the MultiPro. Customers are strongly urged to use only calibration materials when calibrating the MultiPro. Use of non-standard calibration gas and/or calibration kit components can lead to dangerously inaccurate readings and may void the standard Honeywell Analytics warranty.
15.
Use of non-standard calibration gas and/or calibration kit components when calibrating the MultiPro can lead to inaccurate and potentially dangerous readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas specifically developed for easy MultiPro calibration. Customers are strongly
6
urged to use only Honeywell Analytics calibration materials when calibrating the MultiPro.
16.
17.
Substitution of components may impair intrinsic safety.
For safety reasons this equipment must be operated and serviced by qualified personnel only. Read and understand this reference manual
before operating or servicing the MultiPro.
18.
A rapid up-scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the
MultiPro’s zero to 100 percent LEL detection range.
19.
The MultiPro is not designed for use in oxygen enriched atmospheres.
20.
Do not use the MultiPro pump for prolonged periods in an atmosphere containing a concentration of solvent or fuel that may be greater than 50% LEL.
7
1. Description
The MultiPro is a multi-sensor gas detector that can be configured to meet a wide variety of user requirements. This chapter provides an overview of many of the features of the MultiPro. More detailed descriptions of the specific features of the MultiPro are contained in the subsequent chapters of this manual.
1.1 Methods of sampling
The MultiPro may be used in either diffusion or sample­draw mode. In either mode, the gas sample must reach the sensors for the instrument to register a gas reading. The sensors are located at the bottom of the instrument. There are three distinct sensor ports that allow the sample to reach the individual sensors.
The sensor ports must be
kept free of obstruction. Blocked sensor ports can lead to inaccurate and potentially dangerous readings.
In diffusion mode, the atmosphere being measured reaches the sensors by diffusing through the vents at the bottom of the instrument. Normal air movements are enough to carry the sample to the sensors. The sensors react quickly to changes in the concentrations of the gases being measured. Diffusion-style operation monitors only the atmosphere that immediately surrounds the detector.
The MultiPro can also be used to sample remote locations with the hand-aspirated sample-draw kit or with the motorized, continuous sample draw pump (both available separately). During remote sampling, the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing. Remote sampling operations only monitor the atmosphere at the end of the sample draw probe.
Use of the hand-aspirated sample draw kits is covered in section 3.1.
Use of the motorized sample draw pump is covered in section 3.2.
A detailed description of the MultiPro probe assembly is given in section 5.7.
1.2 Multi-sensor capability
The MultiPro can be configured to simultaneously monitor oxygen, carbon monoxide, hydrogen sulfide and combustible gases and vapors. All sensors are replaceable in the field. Each of the MultiPro’s sensor channels is configured for a specific type of sensor.
Note: It is necessary to verify the accuracy of the MultiPro by calibration with known concentration test gas whenever a change is made to the sensors installed in the instrument.
Calibration procedures are discussed in detail in Chapter 4.
The MultiPro uses electrochemical toxic gas sensors that have been designed to minimize the effects of common interfering gases. These sensors provide accurate, dependable readings for toxic gases commonly encountered during confined space entry and other industrial applications.
Different measurement units are used depending on the gas being measured.
Type of Hazard Measurement unit
Oxygen (O2) Percentage by
volume
Combustible gas Percentage of lower
explosive limit
(%LEL) Carbon Monoxide, Hydrogen Sulfide
Table 1.2. Units of Measurement.
Parts per million
(PPM)
1.3 Calibration
The MultiPro detector features one-button fresh air and span calibration.
The accuracy of the MultiPro should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. (The Canadian Standards Association (CSA) requires an accuracy check using known concentration calibration gas prior to each day’s use.)
8
Calibration procedures are discussed in detail in Chapter 4.
Recommended calibration frequency is discussed in Appendix B.
1.4 Alarm logic
MultiPro gas alarms can be adjusted with BioTrak Software through a PC with an IrDA port or directly with the MODE button (see Chapter 6 for direct programming instructions). Alarms may be set anywhere within the nominal range of the specific sensor type. When an alarm set point is exceeded a loud audible alarm sounds, and the bright red LED alarm lights flash.
1.4.1 Atmospheric hazard alarms
The combustible gas alarm is activated when the percent LEL (Lower Explosive Limit) gas concentration exceeds any pre-set alarm level.
Two oxygen alarm set points have been provided; a danger alarm for low concentrations associated with oxygen deficiency and a warning alarm for high concentrations associated with oxygen enrichment.
Four alarm set points have been provided for each toxic gas sensor: Warning, Danger, STEL (Short Term Exposure Limit) and TWA (Time Weighted Average).
S E N S O R
CO 35 100 100 35
H2S 10 20 15 10
Default Toxic Sensor Alarm Levels
Appendix A discusses alarm levels and factory default alarm settings.
detectors have been designed for the detection of deficiencies of oxygen, accumulations of flammable gases and vapors, and accumulations of specific toxic gases. An alarm condition indicating the presence of one or more of these potentially life-threatening hazards should be taken very seriously.
condition it is important to follow established procedures. The safest course
W
A R N
I
N
D A N G E R
S T E L
G
MultiPro portable gas
In the event of an alarm
T
W
A
of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
A rapid up-scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the MultiPro’s zero to 100 percent LEL detection range.
1.4.2 Low battery alarms
The MultiPro may be equipped with either a rechargeable Lithium Ion (Li-Ion) or an alkaline battery pack. The MultiPro includes multi-staged low battery alarms to let the user know that the battery is running low.
For detailed information concerning the low battery alarms, see section 2.4.5.
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer E91­LR6, Eveready EN91 batteries. Substitution of batteries may impair intrinsic safety.
1.4.3 Sensor over range alarms.
The MultiPro will go into alarm if a sensor is exposed to a concentration of gas that exceeds its established range. In the case of an LEL reading that exceeds 100% LEL, the LEL channel will be automatically disabled by the instrument and the instrument will remain in constant alarm until it is turned off, brought to an area that is known to be safe, and then turned back on. The display will show “OL” in place of the sensor reading for any channel that has gone into over range alarm.
See section 2.4.5 for further details on sensor over range alarms.
A sensor range chart is provided in Appendix C.
In the event of an LEL overrange alarm the MultiPro must be turned off, brought to an area that is known to be safe and then turned on again to reset the alarm.
1.4.4 LEL response failure due to lack of
alarm
O
2
The MultiPro features automatic warning against LEL sensor response failure due to lack of oxygen. See section 2.4.7 for details.
9
1.4.5 Security beep/flash
The MultiPro includes a security beep function that is designed to notify the user that the instrument is powered up and running. Once enabled the MultiPro will emit a short audible beep and give a short flash on the LED at a user-defined interval. The security beep/flash function may be enabled and the interval may be changed with BioTrak software or directly with the MODE button (see Chapter 6 for direct programming instructions).
1.4.6 Latching peak alarms
The MultiPro’s alarms are self-resetting unless the alarm latch is enabled. With the MultiPro’s alarm latch enabled, the audible and visible alarms will continue to sound after the atmospheric hazard has cleared. Press the MODE button to reset the alarms. If the alarm latch is disabled and the alarm condition is no longer present, the instrument will automatically return to normal operation, and the visible and audible alarms cease without further input from the user.
1.4.7 Fault detection
MultiPro software includes a number of additional alarms designed to ensure the proper operation of the instrument. When the MultiPro detects that an electronic fault or failure condition has occurred, the proper audible and visible alarms are activated and an explanatory message or message code is displayed.
The MultiPro is designed to detect potentially life threatening atmospheric conditions. Any alarm condition should be taken seriously. The safest course of action is to immediately leave the affected area, and return only after further testing determines that the area is once again safe for entry.
1.5 Other electronic safeguards
Each time the detector is turned on, the MultiPro automatically tests the LED alarm light, audible alarm, internal memory and pump status (if so equipped). The battery is monitored continuously for proper voltage. The MultiPro also monitors the connection of sensors that are currently installed. The detection of any electronic faults causes the activation of the audible and visible alarms and causes the display of the appropriate explanatory message.
1.5.1 Heartbeat
At the center of the display is a heart symbol that will blink every few seconds to show that the instrument is functioning normally.
1.6 Sensors
The MultiPro can be configured to simultaneously monitor oxygen, carbon monoxide, hydrogen sulfide and combustible gases and vapors. The sensor configuration of the MultiPro may be specified at the time of purchase, or changed in the field by appropriately trained personnel.
Note: A CF Value may be required when replacing the 54-49-24 CO/H2S sensor. Instructions are included with the replacement sensor.
Replacement sensor part numbers and sensor ranges are given in Appendix B.
Sensor cross-sensitivity figures are given in Appendix C.
A sensor that cannot be
calibrated or is found to be out of tolerance must be replaced immediately. An instrument that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use.
Calibration procedures are discussed in detail in Chapter 4.
1.7 Optional sample draw pump
A motorized sample-draw pump is available for the MultiPro for situations requiring continuous "hands free" remote monitoring.
Use only part number 54­49-102 sample draw pump with the UL/CSA-approved version of the MultiPro. Use only part number 54-49-102-5 sample draw pump with the ATEX-approved version of the MultiPro.
The pump contains a pressure sensor that detects restrictions in airflow caused by water or other obstructions being drawn into the unit and immediately acts to turn the pump off in order to protect the sensors, pump, and other MultiPro components from damage.
Pump status is continuously monitored by the MultiPro microprocessor. When the pump is
10
Loading...
+ 25 hidden pages