CE Declaration of Conformity ................... 23
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure
that no damage has occurred during shipment. If noticeable damage is found, immediately notify your Dealer.
Note: Conserve all packing material until the
instrument has been observed to function correctly because all defective
items must be returned to the Dealer
in their original packing.
GENERAL DESCRIPTION
HI 8666 is a 2-wire, 4 to 20 mA relative
humidity and temperature transmitter. The
instrument transmits electrically isolated outputs that are linear and proportional to RH%
and °C. It can be surface or panel mounted,
and used with a detachable connector. The
power must be supplied from an external
source from 12V to 24V. See the diagrams
on pages 4-6 of this instruction manual for
connections from the transmitter to other
equipment, such as recorder.
Note: After 30 minutes of measurements over
90% RH, the probe sensors could experience secondary absorption, leading to approximately +6% RH saturation. This means that the readings will
be inaccurate. To restore the probe to
a normal condition, place the probe
into the LiCl well of HI 7102 for 24
hours (see page 9).
3
FUNCTIONAL DESCRIPTION
123
1) Sintered Cap
2) Relative Humidity and Temperature Sensors
3) 11-Pin Connector
SPECIFICATIONS
RangeRH0 (4 mA) to 100% (20 mA)
Resolution RH±0.125% RH, ±0.02 mA
AccuracyRH±2% RH (5 to 80% RH)
Typical EMC RH±2.5% RH, ±0.4 mA
Deviation°C±1°C, ±0.2 mA
Output Signal4 to 20 mA
Response Time6" without sintered filter,
Power Supply12 to 24V externally supplied
Maximum Loop600 ohm
Impedance
Environment-20 to 60°C (-4 to 140°F);
Dimensions79 x 49 x 140 mm
Weight200 g (7 oz.)
°C-20 (4 mA) to 60°C (20 mA)
°C±0.1°C, ±0.02 mA
±4% RH (80 to 90% RH, 0 to 40°C)
°C±1% of span
1' with filter
5 to 95% RH
(3.1 x 1.9 x 5.5")
5
TERMINAL BOARD AND INSTALLATION
The HI 8666 Transmitter has an 11-pin connector at the rear of the instrument.
Connect it to a UNDECAL relay socket.
The terminals are:
1 +°C: Positive Terminal;
Terminals 2, 3, 5, 6, 8, 10 and 11 are not
connected.
See the following figures for recommended
installation:
1.
Passive and isolated receivers
+
•
Power
Supply
•
-
RH%
Receiver
-
+
•
•
•
4
•
+
RH%
Transmitter
-
•
7
°C
Receiver
-
+
•
•
1
•
+
°C
Transmitter
-
•
9
2. Passive and not isolated receivers
(Common positive)
RH%
+
•
+
•
Power
Supply
•
-
Receiver
-
•
4
•
+
RH%
Transmitter
-
•
7
Receiver
Transmitter
3. Passive and not isolated receivers
(Common negative)
-
•
RH%
Receiver
+
•
Receiver
°C
+
°C
°C
+
-
•
1
•
-
•
9
•
-
•
Power
Supply
•
+
7
•
-
RH%
Transmitter
+
•
4
•
7
9
•
-
°C
Transmitter
+
•
1
4. Active receivers
(both isolated or not)
RH%
Receiver
°C
Receiver
+
•
-
•
+
•
-
•
4
+
•
Transmitter
7
-
•
1
+
•
9
Transmitter
-
•
RH%
°C
CALIBRATION CHECKING PROCEDURE
All Hanna relative humidity transmitters have
been precalibrated at the factory.
Hanna Instruments uses state-of-the-art thermal humidity chambers for this purpose.
It is generally recommended to have all RH
transmitter recalibrated at least once a year.
For an accurate annual recalibration, contact
your nearest Hanna Service Center.
You can also check the status of your HI 8666
by using the Hanna mini-calibration chambers HI 7102.
The kit includes two thermally
isolated chambers, each one
equipped with a threaded cap
l
C
a
and three bottles containing
the appropriate precalibrated
N
l
i C
L
%
.4
5
7
%
.1
11
IT
K
N
IO
T
A
R
IB
L
A
C
saturated salts to produce a
known RH value.
This calibration checking procedure is written
to verify that the meter still meets the electrical calibration specifications of the factory.
PREPARING THE CALIBRATION
SOLUTIONS
•Pour approximately 26 cc
of distilled water into a
glass container.
•Immerse this container
into a bath of ice and water and shake briefly.
•Slowly add the contents
of a HI 7111 bottle containing LiCl into the glass
container while continuing
to shake.
9
O
2
H
26 cc
Li Cl
•When the salt has dissolved completely, add all
l
Li C
the content of the second
bottle of HI 7111.
•Allow the solution obtained
to cool, and then pour it
into the chamber marked
"RH 11.1%", making sure
that no residue remains on
the walls of the glass container.
N
l
i C
L
7
%
.1
11
CALIBRATION KIT
l
C
a
%
.4
5
•Seal the chamber well when not in use,
as the LiCl solution is extremely hygroscopic and tends to capture the humidity
present in the air causing the solution to
expand in volume and to overflow from
the container.
H
2
•Pour approximately 12 cc
of distilled water into the
O
12 cc
other chamber marked "RH
75.4%".
N
l
i C
L
5
7
%
.1
11
CALIBRATION KIT
l
C
a
%
.4
•Add all the content of the
HI 7121 bottle containing
NaCl while continuously
shaking the container to
avoid the formation of
lumps. Seal this container
well when not in use.
•The calibration kit needs
approximately 4 hours for
stabilization.
Na Cl
l
i C
L
%
.1
11
CALIBRATION KIT
12
9
6
7
3
a
N
.4
5
l
C
%
4 h
CHECKING THE RH CALIBRATION
CALIBRATION
KIT
Li C
l
11.1%
N
a C
l
75.4%
•Bring the calibration kit to a temperature
of approximately 20°C.
•Remove the sintered
cap from the probe of
HI 8666.
•Connect the unit to a 12 to 24VDC power
supply and hook up a multimeter (range
of 20mA) in series with the circuit.
•Remove the cap from the
"RH 11.1%" chamber containing the LiCl solution and
insert the HI 8666 paying attention not to dip it into the
liquid.
•Wait for the measurement
to stabilize (this takes approximately 4 hours).
12
9
3
4 h
6
•Look at the multimeter display: the reading should be 5.78 mA, representing the
theoretical chamber value of 11.1% RH.
All RH to mA conversions can be made
with the tables provided (see page 12).
•Remove the HI 8666 from the chamber
and tightly seal the chamber containing
the LiCl solution.
•Remove the cap from the
"RH 75.4%" chamber containing the NaCl solution and
insert the HI 8666.
11
Li Cl
11.1%
CALIBRATION KIT
Na Cl
75.4%
12
9
•Wait for the measurement
to stabilize (approximately
3
4 h
6
4 hours)
•Insert the probe in the "RH 75.4%" chamber, the multimeter reading should be 16.06
mA: this represents the theoretical chamber value of 75.4% RH.
mA
•The relative humidity calibration checking
procedure is complete. If the theoretical
values are not within ±5%, contact your
dealer or your closest Hanna Service Center for a fast and accurate recalibration.
Spare saturation salts are also available:
HI 7111/P (LiCl) for low range humidity, and
HI 7121/P (NaCl) for high range humidity.
CHECKING THE TEMPERATURE
CALIBRATION
•Remove the sintered cap from the
probe of HI 8666.
•Connect the unit to a 12 to 24VDC power
supply and hook up a multimeter (range
of 20mA) in series with the circuit.
•Place the HI 8666 into a stable temperature environment of 25°C, preferably a
temperature chamber, and wait for the
temperature to stabilize.
•Using a NIST certified thermometer as a
reference, e.g. ChecktempC/N, verify that
the ammeter reading is matching the temperature, e.g. 13.00 mA for 25°C.
mA
All temperature to mA conversions can
be made with the tables provided (see
page 16).
•Recheck the temperature at both 10°C
(10.00 mA) and 40°C (16.00 mA): they
should both be within ±1%.
If either setting is out of range, the unit
must be repaired or recalibrated electrically. Contact your dealer or your closest
Hanna Service Center.
complete with O-ring and precalibrated saturated salts (LiCl
and NaCl)
HI 7111/PSpare saturation salts 6 x 15 g
bottles of LiCl for low range
humidity calibration
HI 7121/PSpare saturation salts 6 x 33 g
bottles of NaCl for high humidity
calibration
HI 7164Undecal connector
HI 710011Sintered cap for RH probe
21
WARRANTY
HI8666 is warranted for one year against
defects in workmanship and materials when
used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance are
not covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under warranty, report the model number, date
of purchase, serial number and the nature of
the failure. If the repair is not covered by the
warranty, you will be notified of the charges
incurred. If the instrument is to be returned
to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service department and then send it
with shipments costs prepaid. When shipping any instrument, make sure it is properly
packaged for complete protection.
All rights are reserved. Reproduction in whole
or in part is prohibited without the written
consent of the copyright owner.
Hanna Instruments reserves the right to modify
the design, construction and appearance
of its products without advance notice.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment
in which it is used.
Operation of this instrument in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps
to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instrument's EMC performance.
To avoid damages or burns, do not place this transmitter in microwave ovens.
23
06/02
www.hannainst.com
MAN8666R2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.