Remove the instrument from the packing material and examine carefully to make sure that no
damage has occurred during shipment. If noticeable damage is found, immediately notify
the shipper and contact your Dealer's Service
Department.
NOTE: Conserve all packing material until the
instrument has been observed to function correctly because all defective items must be
returned to the Dealer in their original packing.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
See Table 1 and Table 2 for specifications of
each model. The capacity indicates in the
Tables refer to stirring water with a viscosity of
0.001 N.s/m
All Models:
Speed Control:
Drive:
Power Consumption:
Torque Delivered:
Power Supply:
Fuse:
Stirring Bar:
2
at 25ºC.
ON/OFF and continuously variable with
potentiometer knob
Direct drive D.C. Motor with magnet
4W
85 g. cm
100/115/240 VAC 50/60 Hz or DC 1216V depending on models
200 mA or 1A. 20 mm length x 5 mm
diameter depending on models
Magnet enclosed in PTFE
for HI 190M, HI 190MB and HI 200M,
3
4
25 mm length x 7 mm diameter
for other models, 50 mm length x 7 mm
diameter
HI 3050N stirrers with incorporated pH meter:
Range: 0.00 to 14.00 pH
Resolution:0.01 pH
Accuracy:±0.01 pH
Input Resistance:
Calibration:
Offset Regulation:
Slope:
Temperature Compensation:
Table 3 lists the electrodes for various applications that can be used with HI 3050N.
Table 1: Mini and Normal Stirrers
Capacity
(Ltr)
Model
HI 190M
HI 190MB
HI 200M
HI 300N
HI 301N
HI 302N
HI 303N
HI 304N
HI 3050N
0.0 to 100ºC
0.1ºC
±0.4ºC
12
10
Ohm
Double-point in neutral buffer solution
plus an acid or alkaline buffer solution
±1 pH
85% to 105%
Completely automatic with HI 7667 temperature probe from 0ºC to 100ºC. The
values indicated refer to temperature at
25ºC
Speed Range
(rpm)
Display
1
1
1
2.5
2.5
2.5
2.5
2.5
1
Min: 100
Max: 1000
Min: 100
Max: 1000
Min: 100
Max: 1000
Min: 100 to 500
Max: 100 to 1000
Min: 100
Max: 1000
Min: 100 to 500
Max: 100 to 1000
Min: 100
Max: 1000
Min: 100
Max: 1000
Min: 100
Max: 1000
Min: 100 to 500
Max: 100 to 1000
Min: 100
Max: 1000
Min: 100
Max: 1000
Min: 100 to 500
Max: 100 to 1000
Min: 100
Max: 1000
Min: 100
Max: 1000
Display
4-digit LCD
for rpm
4-digit LCD
for rpm
4-digit LCD
for rpm
Table 3
HI5311, HI5412, HI2110.
HI1910B
HI5311, HI5412, HI2110
HI4430 (ORP-Au)
5
Timer
Control
5 min. to
2 hr.
5 min. to
2 hr.
Size
LxBxH (in mm)
180x180x70
180x180x70
180x180x70
180x180x70
180x180x70
180x180x70
180x180x70
Weight
(Kg)
1.5
1.5
1.5
1.5
1.5
1.5
1.5
6
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The Hanna series of bench top electromagnetic
stirrers are designed with simplicity of use for
stirring and mixing solutions in a wide range of
laboratory applications. Eighteen models are
available in an extensive range covering:
• mini-stirrers
• general purpose stirrers
• dual-speed range
• inverse cycle stirring
• rpm indication
• pH/ºC measurement
• timer control
• constant speed with feedback control
All models employ a motorized magnet to rotate
a stirring bar to create a forced vortex rotational
flow in the solution with a constant angular
speed which is adjustable with an external
speed knob. Unlike conventional stirrers, all
Hanna stirrers use the concept of pulse-driven
drive to give the full power required to effectively stir the fluid regardless of the speed. This
eliminates the need to start the motor at high
speed and the user is now able to control the
speed precisely. Each stirrers is equipped with
an opto-sensor which monitor the speed and a
speed limiter circuit to prevent hazardous operation when the load is suddenly removed.
The power required to effectively stir a fluid is
influenced by the torque load on the stirrer
which, in turn, is dependent on the capacity,
type and viscosity of the fluid. The speed of the
stirrer decreases as the torque load increases
and vice versa, unless the power delivered
changes accordingly. For Newtonian fluids like
water, the viscosity remains constant during
stirring. The mini and normal stirrers are ideal
for Newtonian fluids.
6
istrstirR2.p65
7
The mini-stirrers (HI 30XN series) cater up to
2.5 litres. For some fluids like oil and clay, the
viscosity changes during the stirring process
and it is necessary to adjust the power to
maintain a constant speed during the stirring.
Hanna Auto-Speed stirrers (HI 31X series) are
designed with automatic feedback control to
maintain constant speed when the torque load
changes and are therefore ideal for stirring with
non-Newtonian fluids where the viscosity
changes during stirring. It is also ideal for
mixing applications where one solution is added
to another which changes the capacity and
thereby the torque load on the stirrer. The AutoSpeed stirrers maintain a constant stirring speed
during the mixing up to the full capacity of 5
litres. They can also be used with Newtonian
fluids up to a capacity of 5 litres.
Stirring Procedure:Stirring Procedure:
Stirring Procedure:
Stirring Procedure:Stirring Procedure:
1. Make sure that the SPEED knob (B) and
the ON/OFF switch (L) are in the OFF
positions.
2. Place a beaker or glass container on top of
the stirrer and place the stirring bar at the
centre of the beaker base.
3. Pour the reagent to be titrated into the
beaker. Allow sufficient space for the titrant
to be added without spillage.
4. Connect power cable and switch on the
power supply. Then switch the stirrer on
which the ON/OFF switch. Attach the pH
electrode and temperature probe on the
electrode holder (optional accessory) and
adjust the holder so that pH electrode is at
the centre of the beaker. Set the OFF/ºC/pH
to pH to display pH measurement. You
should read the pH of the reagent compensated for the sample temperature.
7
8
5. Adjust the SPEED knob slowly until the
required stirring speed is achieved and allow
sufficient time to attain uniform rotational
flow.
6. Add the titrant through a burette and observe
the changes in pH value of the sample as
the titrant is being added. Switch the OFF/
ºC/pH to ºC to display the temperature of
the mixed sample if required.
7. When the titrant is completed, turn the
SPEED knob to off. If no further stirring is
required switch off the stirrer and disconnect
the power supply.
FRONT PANELFRONT PANEL
FRONT PANEL
FRONT PANELFRONT PANEL
See Figures 2 to 7 on the following pages.
PowerPower
A.
Power indication LED lights up when there
PowerPower
is power supply
SPEEDSPEED
B.
SPEED knob for rpm setting with ON/OFF
SPEEDSPEED
and continuously variable up to the maximum.
HIGH/LOW HIGH/LOW
C.
HIGH/LOW switch for selecting low speed
HIGH/LOW HIGH/LOW
range or high speed range. See Tables 1 and
2 for speed range of each model.
rpmrpm
D.
rpm LCD displays the rpm setting by the
rpmrpm
SPEED knob.
REVERSE PERIOD REVERSE PERIOD
E.
REVERSE PERIOD knob with ON/OFF
REVERSE PERIOD REVERSE PERIOD
and continuously variable selection of inversion cycle time in minutes, from 0.5 to 3
minutes.
TIME TIME
F.
TIME knob for setting the stirring time from
TIME TIME
5 min. to 2 hours.
START/RESET START/RESET
G.
START/RESET for starting and resetting
START/RESET START/RESET
the timer control.
OFF/ºC/pHOFF/ºC/pH
H.
OFF/ºC/pH knob
OFF/ºC/pHOFF/ºC/pH
OFF:Switches display off
ºC : Switches display to temperature
8
istrstirR2.p65
9
indication in ºC
pH : Switches display to pH indication
pH/ºCpH/ºC
I.
pH/ºC LCD display the selected
pH/ºCpH/ºC
measurement depending on the position of
H.
OFFSETOFFSET
J.
OFFSET knob for calibration of pH in
OFFSETOFFSET
neutral buffer solution (pH 7.01 at 25ºC).
SLOPESLOPE
K.
SLOPE knob for calibration of pH in acid (pH
SLOPESLOPE
4.01 at 25ºC) or alkaline (pH 10.01 at 25ºC)
solution.
SET/MEASURESET/MEASURE
L.
SET/MEASURE, selector switch for LCD
SET/MEASURESET/MEASURE
display
SET-POINTSET-POINT
M.
SET-POINT, adjust the set-point of pH
SET-POINTSET-POINT
DISPLAYDISPLAY
N.
DISPLAY, LCD display for pH measure-
DISPLAYDISPLAY
ment or pH set-point.
REAR PANELREAR PANEL
REAR PANEL
REAR PANELREAR PANEL
See Figure 1.
AC POWER CORDAC POWER CORD
K.
AC POWER CORD or battery plug for
AC POWER CORDAC POWER CORD
model HI 190MB.
ON/OFF ON/OFF
L.
ON/OFF switch for power supply.
ON/OFF ON/OFF
FUSEFUSE
M.
FUSE holder for 200 mA or 1A fuse
FUSEFUSE
depending on models.
NOTE:NOTE:
NOTE: HI 190M, HI 190MB and HI 200M are not
NOTE:NOTE:
provided with ON/OFF switch.
They are switched on when there is a power
supply input.
Fig. 1
9
10
MINI AND GENERAL PURPOSE STIRRERSMINI AND GENERAL PURPOSE STIRRERS
MINI AND GENERAL PURPOSE STIRRERS
MINI AND GENERAL PURPOSE STIRRERSMINI AND GENERAL PURPOSE STIRRERS
Models: HI 190M, HI 190MB, HI 200M, HI
300N normal stirrers and HI 310M Auto Speed
stirrer
Fig. 2
These general purpose stirrers are suitable for
daily stirring applications in the laboratory. The
HI 190M, HI 190MB and HI 200M are compact
designed mini stirrers of 1 litre capacity suitable
for use on crowed benches. The HI 190M and
HI 190MB are enclosed completely in chemical
and impact resistant plastic housing. The HI
200M, HI 300N and HI 310N are covered with
a corrosion resistant stainless steel top.
APPLICATIONSAPPLICATIONS
APPLICATIONS
APPLICATIONSAPPLICATIONS
1. Newtonian fluids up to
1 litre
2. Newtonian fluids up to
2.5 litres
3. Non-Newtonian fluids
up to 5 litres whose vis-
cosities change during
stirring or Newtonian fluids up to 5 litres or mixing
operation where load
changes during stirring.
10
USEUSE
USE
USEUSE
HI 190M or
HI 190MB or
HI 200M
HI 300N
HI 310N
istrstirR2.p65
11
OPERATIONOPERATION
OPERATION
OPERATIONOPERATION
1. Make sure that the SPEED knob (B) and
the ON/OFF switch (L) (except HI 190M,
HI 190MB and HI 200M) are in the OFF
positions.
2. Place a beaker or glass container on top of
the stirrer and place the stirring bar at the
centre of the beaker base.
3. Pour the liquids to be mixed into the beaker.
Do not fill the brim and leave sufficient
space to prevent spillage during stirring.
4. Connect power cable and switch on the
power supply. Then the stirrer on with the
ON/OFF switch (except HI 190M,
HI 190MB and HI 200M).
5. Adjust the SPEED knob slowly until the
required stirring speed is achieved. Allow
sufficient time for the mixing operation to
attain uniform rotational flow.
6. After stirring is completed, turn the SPEED
knob to off. If no further stirring is required
switch off the stirrer and disconnect the
power supply.
DUAL-SPEED RANGE STIRRERSDUAL-SPEED RANGE STIRRERS
DUAL-SPEED RANGE STIRRERS
DUAL-SPEED RANGE STIRRERSDUAL-SPEED RANGE STIRRERS
Models: HI 301N, HI 303N normal stirrers and
HI 311N, HI 313N Auto-Speed stirrers
Fig. 3
12
speed control eg. in electrolysis applications.
The selectable HIGH/LOW range allows you to
set the speed precisely.
APPLICATIONSAPPLICATIONS
APPLICATIONS
APPLICATIONSAPPLICATIONS
1. Newtonian fluids up to
2.5 litres
2. Repetitive stirring in
Newtonian fluids up to
2.5 litres
3. Non-Newtonian fluids
up to 5 litres whose
viscosities change
during stirring or
Newtonian fluids up to 5
litres or mixing operation where load
changes during stirring.
OPERATIONOPERATION
OPERATION
OPERATIONOPERATION
1. Make sure that the SPEED knob (B) and
the ON/OFF switch (L) are in the OFF
positions.
2. Place a beaker or glass container on top of
the stirrer and place the stirring bar at the
centre of the beaker base.
3. Pour the liquids to be mixed into the beaker.
Do not fill the brim and leave sufficient
space to prevent spillage during stirring.
4. Connect power cable and switch on the
power supply. Then the stirrer on with the
ON/OFF switch.
5. Select the low or high range with the
HIGH/LOW switch (C) as required. (See
Tables 1 and 2 for the speed range of each
stirrer.)
USEUSE
USE
USEUSE
HI 301N or
HI 303N
HI 300N
HI 310N
11
12
istrstirR2.p65
13
6. Adjust the SPEED knob slowly until the
required stirring speed is achieved. Allow
sufficient time for the mixing operation to
attain uniform rotational flow. The stirring is
indicated in the rpm LCD (D) for models HI
303N and HI 313N.
7. After stirring is completed, turn the SPEED
knob to off. If no further stirring is required
switch off the stirrer and disconnect the
power supply.
REVERSIBLE STIRRERSREVERSIBLE STIRRERS
REVERSIBLE STIRRERS
REVERSIBLE STIRRERSREVERSIBLE STIRRERS
Models: HI 302N, HI 304N normal stirrers and
HI 312N, HI 314N Auto-Speed stirrers
Fig. 4
14
APPLICATIONSAPPLICATIONS
APPLICATIONS
APPLICATIONSAPPLICATIONS
1. Newtonian fluids up to
2.5 litres
2. Repetitive stirring in
multiple batches with
the same stirring speed
3. Non-Newtonian fluids
up to 5 litres whose
viscosities change
during stirring or
Newtonian fluids up to 5
litres or mixing operation where load
changes during stirring.
TIMER CONTROL STIRRERSTIMER CONTROL STIRRERS
TIMER CONTROL STIRRERS
TIMER CONTROL STIRRERSTIMER CONTROL STIRRERS
Model: HI 322N, HI 324N Auto-Speed Stirrers
USEUSE
USE
USEUSE
HI 302N or
HI 304N
HI 304N or
HI 314N
HI 312N or
HI 314N
These reversible stirrers allow you to stir the
fluid in both directions and are ideal for powder
mixing applications to ensure a homogeneous
final product and complete dissolution of hardto-wet powders. The reverse stirring period is
adjustable from 30 sec. to 30 min. When set at
say 1 min., the fluid will rotate in one direction
for min. and then reverse its rotation for another
min. at the desired speed in a repeated cycle
until the stirrer is switched off.
13
Fig. 5
These Auto-Speed stirrers are equipped with
timer control to start and automatically stop the
stirring process after a duration which can be
preset from 5 min. to 2 hours and are ideal for
routine stirring applications where the required
stirring is known. They therefore free the
operator from constantly having to monitor the
stirring process.
14
istrstirR2.p65
15
As these are Auto-Speed stirrers with automatic feedback control, they are ideal for nonNewtonian fluids whose viscosities change
during stirring or Newtonian fluids up to 5 litres
capacity.
APPLICATIONSAPPLICATIONS
APPLICATIONS
APPLICATIONSAPPLICATIONS
1. Routine stirring application with a predetermined stirring time
2. Repetitive stirring in
multiple batches with
the same stirring
speed and predetermined stirring time.
OPERATIONOPERATION
OPERATION
OPERATIONOPERATION
1. Make sure that the SPEED knob (B) and
the ON/OFF switch (L) are in the OFF
positions.
2. Place a beaker or glass container on top of
the stirrer and place the stirring bar at the
centre of the beaker base.
3. Pour the liquids to be mixed into the beaker.
Do not fill the brim and leave sufficient
space to prevent spillage during stirring.
4. Connect power cable and switch on the
power supply. Then the stirrer on with the
ON/OFF switch.
5. Set the TIME knob (F) to the desired
stirring time from 5 min. to 2 hours. Then set
the START/RESET (G) switch to START.
The stirrer will only start stirring when the
START switch is depressed.
6. Adjust the SPEED knob slowly until the
required stirring speed is achieved. Allow
HI 322N
HI 324N
USEUSE
USE
USEUSE
16
sufficient time for the mixing operation to
attain uniform rotational flow. The stirring is
indicated in the rpm LCD (D) for models HI
324N. If you need to restart the timer, press
START/RESET switch to RESET. This will
only reset the timer. Then press the switch
to START to begin stirring again.
7. The stirrer will automatically stop when the
preset time is reached. Turn the SPEED
knob to off. If no further stirring is required
switch off the stirrer and disconnect the
power supply.
STIRRERS WITH INCORPORATEDSTIRRERS WITH INCORPORATED
STIRRERS WITH INCORPORATED
STIRRERS WITH INCORPORATEDSTIRRERS WITH INCORPORATED
Model: HI 305N normal stirrer
These unique stirrers incorporate a temperature
compensated pH meter into the housing of the
stirrer and allow you to take pH and temperature
measurements while thoroughly mixing the
samples. They are indispensable for accurate
potentiometric titrations where reliable and reproducible results are required.
pH METERpH METER
pH METER
pH METERpH METER
and HI 315N Auto-Speed stirrer
Fig. 6
15
16
istrstirR2.p65
17
APPLICATIONSAPPLICATIONS
APPLICATIONS
APPLICATIONSAPPLICATIONS
1. Newtonian fluids up to
2.5 litres.
2. Non-Newtonian fluids
up to 5 litres whose
viscosities change
during stirring or
Newtonian fluids up to 5
litres or mixing operation where load
changes during stirring
The pH meter incorporated in the stirrer should
be calibrated before using.
Calibration:Calibration:
Calibration:
Calibration:Calibration:
1. Connect the pH electrode and temperature
probe to the stirrer. For refillable electrode,
lower the sleeve from the pH electrode to
reveal the refill hole.
2. Connect power cable and switch on power
supply. Then switch the stirrer on with the
ON/OFF switch (L).
3. Immerse the electrode and temperature
probe into pH 7.01 buffer solution and wait
one minute. Turn the OFF/ºC/pH knob (H) to
ºC to display temperature and note the
temperature of the buffer solution.
4. Turn the OFF/ºC/pH to display pH
measurement and adjust the OFFSET knob
(J) until the LCD (I) displays the pH value at
the noted temperature in accordance to
Table 4, eg. if the temperature of the buffer
is 20ºC, adjust the OFFSET knob to display
7.03 pH.
5. Rinse the pH electrode thoroughly in water
and immerse in pH 4.01 or pH 10.01 buffer
solution.
HI 305N
HI 315N
USEUSE
USE
USEUSE
18
6. Wait one minute and adjust the SLOPE
knob (K) until the buffer value corresponding
to the noted temperature of the buffer is
displayed.
The calibration of pH is now completed and the
stirrer is ready for use.
STIRRERS WITH pH METER FOR MANUALSTIRRERS WITH pH METER FOR MANUAL
STIRRERS WITH pH METER FOR MANUAL
STIRRERS WITH pH METER FOR MANUALSTIRRERS WITH pH METER FOR MANUAL
The HI 3050N is specially developed for manual
titration, particularly used for the educational
and field analysis. It combines an accurate
ATC pH meter and a magnetic stirrer. When the
pH reading is within 0.5 pH of the user-selected
set point, an audible alarm will alert the operator, allowing time to decrease the dosage of the
reagent.
Model: HI 3050N
TITRATIONTITRATION
TITRATION
TITRATIONTITRATION
17
18
istrstirR2.p65
19
Table 4
Calibration:Calibration:
Calibration:
Calibration:Calibration:
1. Connect the HI 1217D electrode to the rear
input DIN socket of HI 3050N.
2. Connect power cable and switch on the
power supply. Then switch the stirrer "ON"
with the "ON/OFF" switch. (Fig. 1, L)
3. Immerse the electrode into pH 7.01 buffer
solution with an accurate thermometer,
measured the solution temperature. Wait
for one minute, select the "SET/MEA" switch
(Fig. 7, L) to "MEA" position.
4. Note down the temperature and pH value
shown on the LCD display (Fig. 7, N). Adjust
the "OFFSET" knobs (Fig. 7, J) until the
LCD shows the correct pH value in accordance to Table 4. Eg. If temperature is
30ºC, adjust "OFFSET" knob for LCD reading pH 7.00.
5. Rinse the electrode and immerse in pH
4.01 or 10.01 reference buffer.
6. Wait for one minute and adjust "SLOPE"
knob (Fig. 7, K) until the LCD showed value
in accordance to the noted temperature of
the buffer. (See Table 4)
20
7. The calibration is now completed and the
stirrer is ready for use.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
All Hanna Instruments meters are warranted
for two years against defects in workmanship
and materials when used for their intended
purpose and maintained according to instructions.
The Probes are warranted for a period of six
months.
This warranty is limited to repair or replacement
free of charge. Damage due to accidents,
misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of
purchase, serial number and the nature of the
failure. If the repair is not covered by the
warranty, you will be notified of the charges
incurred. If the instrument is to be returned to
Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service department and then send it with
shipments costs prepaid. When shipping any
instrument, make sure it is properly packaged
for complete protection.
To validate your warranty, fill out and return the
enclosed warranty card within 14 days from
purchase date.
Hanna Instruments reserves the right to
modify the design, construction and appearance of its products without advance notice.
19
20
istrstirR2.p65
21
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
22
OPTIONAL ACCESSORIESOPTIONAL ACCESSORIES
OPTIONAL ACCESSORIES
OPTIONAL ACCESSORIESOPTIONAL ACCESSORIES
HI 76502/PHI 76502/P
HI 76502/P Micro stir bar
HI 76502/PHI 76502/P
HI 76503/PHI 76503/P
HI 76503/P Standard stir bar
HI 76503/PHI 76503/P
HI 7004LHI 7004L
HI 7004L pH 4.01 Buffer Solution
HI 7004LHI 7004L
HI 7006LHI 7006L
HI 7006L pH 6.86 Buffer Solution
HI 7006LHI 7006L
HI 7007LHI 7007L
HI 7007L pH 7.01 Buffer Solution
HI 7007LHI 7007L
HI 7009LHI 7009L
HI 7009L pH 9.18 Buffer Solution
HI 7009LHI 7009L
HI 7010LHI 7010L
HI 7010L pH 10.01 Buffer Solution
HI 7010LHI 7010L
HI 70300HI 70300
HI 70300 Electrode storage solution
HI 70300HI 70300
HI 7061HI 7061
HI 7061Electrode cleaning solution
HI 7061HI 7061
HI 7071HI 7071
HI 7071Reference electrode solution
HI 7071HI 7071
HI 7072HI 7072
HI 7072Electrolyte solution
HI 7072HI 7072
HI 7073HI 7073
HI 7073Pepsin and HCl solution
HI 7073HI 7073
HI 7074HI 7074
HI 7074Thiourea solution
HI 7074HI 7074
HI 76405HI 76405
HI 76405 Electrode holder
HI 76405HI 76405
HI 75012HI 75012
HI 75012 VDC-DC 12V Car lighter adapter
HI 75012HI 75012
HI 75110HI 75110
HI 75110 VAC-AC 110V power adapter
HI 75110HI 75110
HI 75220HI 75220
HI 75220 VAC-AC 220V power adapter
HI 75220HI 75220
21
22
istrstirR2.p65
23
OTHER PRODUCTS FROM HANNA
•Chemical Test Kits
•Colorimeters
•Conductivity and TDS Meters
•Dissolved Oxygen Meters
•Dosing Pumps
•Electrodes (pH and ORP)
•Hygrometers
•pH Meters
•Magnetic Stirrers
•Process Instrumentation
•Solutions and Accessories
•Testers (pH, µS/cm, pNa, TDS, ORP)
•Thermometers
24
ISTRSTIRR2
11/00
HTTP://WWW.HANNAINST.COM
23
24