Hanna Instruments HI 255 User Manual

Instruction Manual
HI 255
Multiparameter
pH/mV/ºC
EC/TDS/NaCl Bench Meter
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of
the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or turn to the back cover for our worldwide contact list.
WARRANTY
HI 255 is guaranteed for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authori­zation number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
WARRANTY ................................................................................................................... 2
PRELIMINARY EXAMINATION ........................................................................................... 3
GENERAL DESCRIPTION .................................................................................................... 3
FUNCTIONAL DESCRIPTION ............................................................................................. 4
HI 255 SPECIFICATIONS .................................................................................................. 5
OPERATIONAL GUIDE ....................................................................................................... 7
AUTORANGING ................................................................................................................ 11
pH CALIBRATION ................................................................................................................ 12
RELATIVE mV CALIBRATION .......................................................................................... 15
ECS/TDS CALIBRATION ................................................................................................... 16
NaCl CALIBRATION .......................................................................................................18
GOOD LABORATORY PRACTICE (GLP) .....................................................................................19
LOGGING FUNCTION .......................................................................................................25
SETUP .................................................................................................................................31
TEMPERATURE CALIBRATION (FOR TECHNICAL PERSONNEL ONLY) .................................36
mV CALIBRATION (FOR TECHNICAL PERSONNEL ONLY) .............................................. 38
PC INTERFACE ............................................................................................................... 39
pH VALUES AT DIFFERENT TEMPERATURES .................................................................... 41
CONDUCTIVITY VERSUS TEMPERATURE CHART ................................................................ 42
ELECTRODE CONDITIONING & MAINTENANCE ................................................................. 43
TROUBLESHOOTING GUIDE ........................................................................................... 46
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS .............................................47
ACCESSORIES .................................................................................................................... 48
2
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
HI 1131B Glass-body Combination pH Electrode with 1 m (3.3') Cable  HI 76310 Conductivity / TDS probe  HI 7662 Temperature Probe  HI 76404 Electrode Holder  pH 4.01 & 7.01 Buffer Solutions (20 mL each)HI 7071S Electrolyte Solution  12VDC Power Adapter  Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
The HANNA HI 255 is a logging microprocessor-based pH, ORP, Conductivity (EC), TDS, NaCl and Temperature bench meter. Relative mV feature is also provided.
The pH measurements are compensated for temperature effect manually or automatically with the HI 7662 temperature probe.
pH Calibration is possible in up to three points, using five memorized buffers (4.01, 6.86, 7.01, 9.18 and 10.01).
The autoranging feature of the EC and TDS ranges automatically sets the instrument to the scale with the highest possible resolution.
The conductivity measurements are compensated for temperature effect manually or automatically with the temperature sensor inside the conductivity probe. It is also possible to disable the temperature compensation and measure the actual conductivity.
The temperature coefficient is user selectable. The GLP feature provides a guarantee of data consistency. An alarm time-out is available to alert the user that too much time elapsed
since last pH calibration. The instrument can store data in memory at the users request for later
retrieval.
3
Front Panel
Primary LCD
Secondary LCD
1
9
3
2
7
4
6
5
10
8
11
13 14 15
Rear Panel
16
POWER
12VDC
INPUT TEMP
RS232
PROBE
12
ALT FunctionsFunctions
RCL
LOG
ALT
SET
LOCK/CLR
ATC
TC
MODE
RANGE
GLP
CFM
CAL
pH/mV & EC/TDS/NaCl
HI 255 Combined Meter
TC Factor
TC
ATC
Select Temperature Compensation
Recall Log Data
RCL
SET
Set Parameters
Select Measure Mode
MODE
Confirm/GLPData
CFM/GLP
Calibration
CAL
Change Parameter Value
StoreLog Data
LOG
Lock Current Range Clear Calibration
LOCK/CLR
Select Measure Unit Scroll Log Data
RANGE
REF
/
BUFFER pH
INTVDATETIME%
ppm
S
3
2
1
BUF
WRONG
READY
CFM
NaCl
pHmVg / lLOGCAL
FUNCTIONAL DESCRIPTION
1) Liquid Crystal Display (LCD).
2) CFM/GLP key, to confirm different values or to display Good Laboratory
Practice information.
3) CAL key, to enter and exit/escape calibration mode.
4) key, to manually decrease temperature value or other parameters.
ATC key (alternate function), to select EC temperature compensation mode.
5) key, to manually increase temperature value or other parameters.
TC key (alternate function), to view the temperature coefficient value.
6) LOG key, to store measured data. RCL key (alternate function), to enter/exit recall mode.
7) ALT key, to select alternate key function.
8) LOCK/CLR key, to freeze current EC range on the LCD or to clear pH calibration parameters.
SET key (alternate function), to enter/exit setup mode.
9) RANGE key, to switch measurement unit pH/mV/EC or focused data. MODE key (alternate function), to select measurement mode/pH resolution.
10) ON/OFF switch.
11) Power supply socket.
12) Conductivity socket.
13) BNC electrode connector.
14) Electrode reference socket.
15) Temperature probe socket.
16) RS232 serial communication connector.
4
HI 255
EGNAR
Hp00.61ot00.2
Hp000.61ot000.2
Vm9.996±
Vm0002±
99.92ot00.0
µ mc/S
9.992ot0.03
µ mc/S
9992ot003
µ mc/S
mc/Sm99.92ot00.3 mc/Sm0.002ot0.03
lautcamc/Sm0.005otpu
)*(
ytivitcudnoc
mpp99.41ot00.0 mpp9.941ot0.51
mpp9941ot051
l/g99.41ot05.1 l/g0.001ot0.51
lautcal/g0.004otpu
)*(
SDT
)rotcaf08.0htiw(
lCaN%0.004ot0.0
)EGNARHp(Cº0.021ot0.01
)EGNARCE(Cº0.06ot0.0
NOITULOSER
Hp10.0
Hp100.0
)Vm9.996±(Vm1.0
)Vm0002±(Vm1
10.0
µ mc/S
1.0
µ mc/S
1
µ mc/S
mc/Sm10.0
mc/Sm1.0
mpp10.0
mpp1.0
mpp1
l/g10.0
l/g1.0
lCaN%1.0
Cº1.0
YCARUCCA
Fº86/Cº02@
Hp10.0±
Hp200.0±
)Vm9.996±(Vm2.0±
)Vm0002±(Vm1±
50.0(±gnidaerfo%1±
µ romc/S
)retaergrevehcihw,tigid1
rompp30.0(±gnidaerfo%1±
1 )retaergrevehcihw,tigid
gnidaerfo%1±
Cº4.0±
)rorreeborpgnidulcxe(
SPECIFICATIONS
(*)
Actual conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.
5
egnartesffoVmleR Vm0002±
ecafretniretupmoC 232SRdetalosi-otpo
noitarbilaCHp
;noitarbilactniop3ro2,1
elbaliavasreffub5
)10.01,81.9,10.7,68.6,10.4(
noitarbilaCCE
;noitarbilacepolstniop1
:elbaliavasreffub6
3141,0.48
µ mc/S
mc/Sm8.111,0.08,88.21,00.5
:tesffotniop1
00.0
µ mc/S
noitarbilaClCaN
reffubL7307IHhtiwtniop1
)lanoitpo(
erutarepmeT
noitasnepmoc
:morfcitamotuArolaunaM
)EGNARHp(Cº0.021ot0.01
)EGNARCE(Cº0.06ot0.0
ytivitcudnocnodelbasidebnac(
)ytivitcudnoclautcaerusaemotegnar
erutarepmetytivitcudnoC
tneiciffeoc
Cº/%00.6ot00.0
)ylnoSDTdnaCErof(
Cº/%09.1sieulavtluafed
rotcafSDT
08.0ot04.0 )05.0sieulavtluafed(
edortcelEHp B1311IH
eborPCE 01367IH
eborperutarepmeT 2667IH
ecnadepmitupnI 01
21
mho
ylppusrewoP retpadaCDV21
snoisnemiD )9.2x1.7x4.9(mm47x281x042
thgieW
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
tnemnorivnE
)Fº22123(Cº050
gnisnednocnon%59HRxam
ytnarraW sraey2
HI 255
SPECIFICATIONS (cont.)
6
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12 VDC adapter into the power supply socket. Note: This instrument uses non volatile memory to retain the calibration
parameters and all other settings, even when unplugged.
Note: Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONS
For pH or ORP measurements connect an electrode with internal reference to the BNC connector on the back of the instrument.
For electrodes with a separate reference connect the electrodes BNC to the BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
For EC/TDS measurements connect the probe to the 7-pin connector. Make sure the probe sleeve is properly inserted.
INSTRUMENT START-UP
 Turn the instrument on by pressing the ON/OFF switch.  All LCD tags are displayed and a beep is sounded while
the instrument performs a self test.
Note: The instrument starts in the same range and mode as it was at
power off.
The RANGE key toggles between pH, mV and EC ranges.
The ALT&MODE keys changes the displayed information for the
selected range.
pH MEASUREMENTS
Make sure the instrument has been calibrated before taking pH measurements.
 Press RANGE to enter pH range. Note: To select pH resolution press the ALT&MODE keys.
 Submerge the electrode tip and the temperature probe
approximately 4 cm (1½) into the sample to be tested and stir gently. Allow time for the electrode to stabilize.
7
 The pH is displayed on the primary LCD and the temperature on the
4cm
(1 )
1/2
pH
pH
secondary LCD.
 If the reading is out of range, the closest full-scale value will be displayed
blinking on the primary LCD.
If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample to prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must be compensated for. To use the Automatic Temperature Compensation feature, connect and submerge the
HI 7662 temperature probe into the sample as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, manual compensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature of 25 ºC or the last temperature reading with the ºC tag blinking.
The temperature can now be adjusted with the ARROW keys (from 10.0 ºC to 120.0 ºC).
mV/ORP MEASUREMENTS
An optional ORP electrode must be used to perform ORP measurements (see Accessories).
Oxidation-Reduction Potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a REDOX measurement, the surface of the ORP electrode must be clean and smooth.
 Press RANGE to enter mV range.  Submerge the tip of the ORP electrode (4 cm/1½) into
the sample to be tested and allow a few seconds for the reading to stabilize.
8
 The instrument displays the mV reading on the primary LCD and the
mV
mV
mV
temperature on the secondary LCD line.
 If the reading is out of range, the closest full-scale value will be displayed
blinking on the primary LCD.
RELATIVE mV MEASUREMENTS
Press the ALT&MODE keys simultaneously while in mV range. The rEL message will be displayed on the secondary LCD line for about one second and the mV tag will blink. After one second the temperature will be displayed on the secondary LCD line.
The reading displayed by the instrument is equal to the difference between the current mV input value and relative mV offset established in the relative mV calibration.
CONDUCTIVITY MEASUREMENTS
Connect the conductivity probe to the instrument.  Press RANGE to enter conductivity measurement range (EC).  Immerse the probe into the solution to be tested. The sleeve
holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
The conductivity value will be displayed on the primary LCD and the temperature on the secondary LCD line.
 If the reading is out of range, the full-scale value (200.0 for Mtc/Atc mode
or 500.0 for actual conductivity) will be displayed blinking.
9
 If LOCK was pressed to freeze the LCD range and the reading went out of
ALT
ATC
range, the full-scale value of the frozen range will be displayed blinking.
The EC reading is affected by temperature. Three options of compensating temperature are available, on EC measurement
mode. Note: The compensation is referenced at the selected reference temperature
(see SETUP for details, page 31).
Automatic (Atc): The EC probe has a built-in temperature sensor; the temperature value is used to automatically compensate the EC/TDS reading (from 0.0  60.0 ºC).
Manual (Mtc): The temperature value, shown on the secondary LCD, can be manually set with the ARROW keys. The ºC tag blinks when this option is active.
No Compensation (notc): The temperature value is displayed, but is not taken into account and ºC tag blinks with a higher frequency. The reading displayed on the primary LCD is the actual EC or TDS value. To select the desired option, press the ALT&ATC keys until the option is displayed on the LCD.
Note: The default compensation mode is Atc.
If no temperature probe detected, Atc mode can not be selected
and the instrument displays ---- on the secondary LCD.
If temperature compensation is active, measurements are compensated using the temperature coefficient (default value 1.90 %/ºC). To change the temperature coefficient, enter the setup mode and select the tc item (see SETUP for details, page 31). The current temperature coefficient can be quickly viewed by pressing the ALT&TC keys. The value is briefly displayed on the secondary LCD.
 If the temperature reading is out of 0.0  60.0 ºC interval and Atc option is
selected, the ºC tag will blink and the closest interval limit will be displayed.
 Press the ARROW keys to change the displayed temperature value. This
value is used to compensate the EC/TDS reading.
TDS MEASUREMENTS
Press the ALT&MODE keys while in EC range. The instrument will switch to TDS measuring range. A TDS reading will be displayed on the primary LCD and the temperature reading on the secondary LCD.
10
 If the reading is out of range, the full-scale value (100.0 for Mtc/Atc mode
or 400.0 for actual TDS) will be displayed blinking.
 If LOCK was pressed to freeze the LCD range and the reading went out of
range, the full-scale value of the frozen range will be displayed blinking.
NaCl MEASUREMENTS
Press the ALT&MODE keys while in EC range until NaCl is displayed on the LCD. The instrument will display the NaCl reading on the primary LCD and the temperature reading on the secondary LCD line.
 If the reading is out of range, the full-scale value (400.0%) will be
displayed blinking.
TEMPERATURE MEASUREMENTS
To take temperature measurements while the instrument is in pH/ORP range, connect the HI 7662 temperature probe to the appropriate socket.
Immerse the temperature probe into the sample and allow the reading on the secondary LCD to stabilize.
For EC/TDS/NaCl range, the HI 76310 probe has a built-in temperature sensor. Note:  If the instrument is in pH/ORP range and the HI7662 temperature
probe is not connected, the ºC tag will blink (manual compensation) even if EC probe is connected.
AUTORANGING
The EC and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution.
By pressing LOCK, the autoranging feature is disabled and the current range is frozen on the LCD. The Auto Off (autoranging disabled) tags will be displayed on the LCD for a few seconds. To restore the autoranging option, press LOCK again. The Auto On (autoranging enabled) tags will be displayed on the LCD for a few seconds.
Note: Autoranging is automatically restored if the range is changed, if the setup or
calibration modes are entered and if the meter is turned off and back on again.
11
pH CALIBRATION
1
BUF
BUFFER pH
1
BUF
pHCAL
Calibrate the instrument frequently, especially if high accuracy is required. The instrument should be recalibrated :  Whenever the pH electrode is replaced.  At least once a week.  After testing aggressive chemicals.  If CAL INTV tags are blinking during measurement.
Every time you calibrate the instrument use fresh buffers and perform an electrode Cleaning Procedure (see page 45).
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One for rinsing the electrode and one for calibration.
If you are measuring in the acidic range, use pH 7.01or 6.86 as first buffer and pH 4.01 as second buffer. If you are measuring in the alkaline range, use pH
7.01 or 6.86 as first buffer and pH 10.01 or 9.18 as second buffer.
PROCEDURE
Calibration has a choice of 5 memorized buffers: pH 4.01, 6.86, 7.01, 9.18 and 10.01.
It is recommended to perform a two or three-point calibration. However, the instrument also allows a one-point calibration, as described on page 14.
Select pH range by pressing RANGE.
THREE-POINT CALIBRATION
Immerse the pH electrode and the temperature probe
approximately 4 cm (1½) into a buffer solution of your choice (pH 4.01, 6.86, 7.01, 9.18 or 10.01) and stir gently. The temperature probe should be close to the pH electrode.
 Press CAL. The CAL and 
the 7.01 buffer will be displayed on the secondary LCD.
 tags will appear and
 If necessary, press the ARROW keys to select a different buffer value.
12
 The   tag will blink on the LCD until the reading is
BUFFER pH
2
BUF
pHCAL
stable.
 When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
 Press CFM to confirm calibration.  The calibrated value will be displayed on the primary LCD
and the second expected buffer value on the secondary LCD line.
Note: The instrument will automatically skip the buffer used for the first point.
It also skips 6.86 if 7.01 buffer was used and viceversa. Likewise, it will skip 9.18 if 10.01 buffer was used and viceversa.
After the first calibration point is confirmed, immerse the pH electrode and
the temperature probe approximately 4 cm (1½) into the second buffer solution and stir gently. The temperature probe should be close to the pH electrode.
 If necessary, press the ARROW keys to select a different buffer value.  The   tag will blink on the LCD until the reading is
stable.
 When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
Press CFM to confirm calibration.
The calibrated value will be displayed on the primary LCD
and the third expected buffer value on the secondary LCD line.
Note: The instrument will automatically skip the buffer used for the first and
second calibration point.
After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½) into the third buffer solution and stir gently. The temperature probe should be close to the pH electrode.
 If necessary, press the ARROW keys to select a different buffer value.  The   tag will blink on the LCD until the reading is
stable.
 When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
Press CFM to confirm calibration.
 The instrument stores the calibration value and returns
to normal measurement mode.
13
Note:  If the value measured by the instrument is not
2
BUF
close to the selected buffer, WRONG 
and WRONG   tags will blink alternately. In this case, check if the correct buffer has been used or regenerate the electrode by following the Cleaning Proce­dure (see page 45). If necessary, change the buffer or the electrode.
 If WRONG, Buffer pH tags and Old tag on secondary LCD are
displayed blinking, an inconsistency between new and previous (old) calibration is detected. Clear calibration parameters by pressing LOCK/CLR and proceed with calibration from the current calibration point (the instrument will keep all confirmed values during current calibration).
 If LOCK/CLR is pressed at the first calibration point, the instru-
ment clears all calibration parameters and returns to measurement mode.
 The WRONG tag and temperature value are displayed blinking
if temperature reading is out of the defined temperature range of the buffer. Calibration cannot be confirmed in this situation.
 Press RANGE to display the temperature reading
on the LCD during calibration.
TWO-POINT CALIBRATION
 Proceed as described in Three-point calibration section.  Press CAL after the second calibration point was confirmed. The instrument will return to measurement mode and will memorize the
two-point calibration data.
ONE-POINT CALIBRATION
 Proceed as described in Three-point calibration section.  Press CAL after the first calibration point was confirmed. The instrument will return to measurement mode and will memorize the
one-point calibration data. Note: Calibration parameters are evaluated taking in consideration the new
values for the confirmed buffers in current calibration and the old values, if existing, for the other buffers. To clear old calibration values, press LOCK/CLR before exiting calibration.
14
RELATIVE mV CALIBRATION
 Press CAL when the instrument is in RELATIVE mV measurement mode.
The mV and   tags will blink. Absolute mV is displayed on the primary LCD and AbS message is displayed on the secondary LCD.
 When the reading is stable and inside the offset window (±200 0 mV), the
instrument asks for confirmation.
 If the reading is out of the offset window, WRONG tag will be displayed.  Press CFM to confirm absolute value. The instrument will display 0.0 mV
on the primary LCD and rEL message on the secondary LCD line. In this moment the relative mV offset is equal to absolute mV reading.
 Use the ARROW keys if you want to change the displayed value.  Press CFM to confirm relative mV calibration. The instrument will return to
measurement mode.
15
EC/TDS CALIBRATION
S
BUF
CAL
S
BUF
CAL
READY
CFM
EC calibration is a one-point procedure. Selectable calibration points are 0.00 µS for offset and 84.0 µS, 1413 µS, 5.00 mS, 12.88 mS, 80.0 mS, 111.8 mS for slope.
Rinse the probe with calibration solution or deionized water. Immerse the probe into the solution. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. To enter EC calibration, select the EC range and press CAL.
Note: TDS reading is automatically derived from the EC
reading and no specific calibration for TDS is needed. Pressing CAL when TDS range is selected has no effect.
For zero calibration, just leave the dry probe in the air. This calibration is performed in order to correct the reading around 0.00 µS point. The slope is evaluated when the calibration is performed in any other point.
The BUF and CAL tags are displayed. The primary LCD will display the EC reading. The secondary LCD will display the buffer value. The   and ~ tags will blink.
Select the desired value with the ARROW keys, if necessary.
When the reading is stable, READY tag is displayed and CFM tag starts blinking on the LCD, asking for confirmation.
Press CFM to confirm calibration. The instrument stores the calibration value and returns to
measurement mode. Note: If the uncalibrated reading is too far from the expected value, the
WRONG and   tags will blink. Calibration can not be confirmed.
In this case check if the correct buffer has been used or clean the probe
by following the Cleaning Procedure (see page 45).
16
If the meter is in Atc mode and the buffer temperature is out of
WRONG
CAL
0.0  60.0 ºC interval, WRONG ºC tags and the temperature will be displayed blinking.
For best results choose an EC buffer value close to the sample to be
measured.
In order to minimize any EMC interference, use plastic or glass
beakers.
It is possible to set the cell constant value directly, without following
the calibration procedure. To set the cell constant, enter SETUP mode and select CEL (see SETUP for details, page 31).
17
Loading...
+ 39 hidden pages