Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this instrument.
This manual will provide you with the necessary information for correct
use of this instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us
at tech@hannainst.com.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 2400 is guaranteed for two years against defects in workmanship and
materials when used for its intended purpose and maintained according to
instructions. Electrodes and probes are guaranteed for six months. This
warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair is
not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization number from the Technical Service
department and then send it with shipping costs prepaid. When shipping
any instrument, make sure it is properly packed for complete protection.
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occured during shiping. If
there is any damage, notify your Dealer or the nearest Hanna Customer
Service Center.
Each instrument is supplied with:
• HI 76407/2 DO probe with 2 m (6.7') cable
• HI 76407A membrane cap (2 pcs)
• HI 7041S electrolyte solution (30 ml)
• 12 VDC power adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in their
original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI2400 is a logging microprocessor-based DO/Temperature bench meter.
It can store up to 100 lots in memory, with up to 8000 readings. These
readings can be transferred to a computer for further analysis or permanent
storage.
Dissolved Oxygen is indicated in ppm (parts per million) or in %.
All measurements are automatically compensated for temperature.
Salinity compensation in water allows direct determination of Dissolved
Oxygen in saline waters and altitude compensation readjusts for the
altitude variance.
The Dissolved Oxygen probe has a membrane covering the polarographic
sensors and a built-in thermistor for temperature measurements and
compensation.
This permeable PTFE membrane isolates the sensor elements from the
testing solution, but allows Oxygen to pass through. When a voltage
is applied across the sensor, oxygen that has passed through the
membrane reacts causing a current flow, and hence determining a
reading.
3) CFM/GLP key, to confirm calibration selection, different setup values
or to display Good Laboratory Practice information.
4)ºC key, to manually increase temperature value or other parameters.
5)ºC key, to manually decrease temperature value or other parameters.
6) SETUP key, to enter/exit SETUP mode.
7) RANGE key, to select measurement range (% or ppm), switch to
focused data in SETUP or toggle between standard value and
temperature during calibration.
8) LOG/CLR key, to store a value into memory, or to select to delete log
records.
9) RCL key, to enter and exit view memory mode.
10) ON/OFF switch.
11) Power supply socket.
12) USB connector.
13) DO probe connector.
5
SPECIFICATIONSSPECIFICATIONS
EGNAR
mpp00.54ot00.0
%0.003ot0.0
)F°221ot0.23(Cº0.05ot0.0
NOITULOSER
mpp10.0
%1.0
C°1.0
YCARUCCA
F°86/C°02@
,tigid1±roelacslluffo%5.1±
retaergsirevehcihw
)rorreeborpgnidulcxe(Cº2.0±
noitaiveDCMElacipyT
elacslluffo%5.1±
Cº5.0±
noitarbilaCOD
%0tatniopelbuodroelgniS
(0407IH)riani(%001dna)
edutitlA
noitasnepmoC
noituloseR
)'021,31(m000,4ot0
)'823(m001
ytinilaS
noitasnepmoC
noituloseR
l/g04ot0
l/g1
noitasnepmoCerutarepmeT
Cº0.05ot0.0
)Fº221ot0.23(
eborP
2/70467IHhtiw
elbac)'6.6(m2
lavretnigniggoL
sdnoces03,01,5
,021,06,03,51,01,5,2,1ro
setunim081
noitacinummocCPBSUdetalosiotpO
ylppusrewoPretpadaCDV21
snoisnemiD)”3.4x7.8x2.9(mm901x222x532
thgieW
)bl9.2(gK3.1
)bl6.4(gK1.2redlohhtiwtik
tnemnorivnE
)Fº221-23(Cº05–0
gnisnednocnonHR%59.xam
ytnarraWsraey2
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
6
Shipping
cap
black
red
FILLFIRST
THENTAP
THENSCREW
BACKON
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
POWER CONNECTION
Plug the 12 VDC adapter into the power supply socket.
Notes: • This instrument use non volatile memory to retain the
calibration parameters and all the other settings even when
unplugged.
• Make sure a fuse protects the main line.
PROBE CONNECTION AND PREPARATION
To take measurements, connect the DO probe to the meter securely by
aligning the pins with the socket located on the back of the meter,
pushing the plug in and tightening the threaded ring.
Probes shipped from Hanna Instruments are dry. To hydrate the probe
and prepare it for use, connect it to the
meter and proceed as follows:
1. Remove the red and black plastic
cap. This cap is for shipping purposes
and can be thrown away.
2. Wet the sensor by soaking the bottom
2½ cm (1") of the probe in electrolyte
(HI 7041S) for 5 minutes.
3. Rinse the membrane cap (HI 76407A
supplied in the kit with the meter)
with electrolyte solution while shaking
it gently. Refill with clean electrolyte
solution.
4. Tap gently the sides of the membrane
cap with your finger tip to ensure
that no air bubbles are trapped. To
avoid damaging the membrane, do
not tap it directly on the bottom.
5. Make sure that the rubber O-ring
sits properly inside the membrane
cap.
6. With the sensor facing down,
slowly screw the cap clockwise.
Some electrolyte will overflow.
When not in use and during polarization
(see page 8), use the protective transparent
cap supplied in the kit with the meter.
7
INSTRUMENT START-UP
• Turn the instrument on by pressing the ON/OFF switch.
• All LCD tags are displayed and a beep is generated while the
instrument performs a self test.
• The instrument will display “ ” blinking until initialization is complete.
• After a few seconds “Cond” message appears on the LCD to inform
the user that the probe is in auto-conditioning (automatic polarization)
mode.
• When this message disappears, the probe is polarized and the
instrument can be calibrated.
• If the probe is disconnected, the meter will display “----”.
PROBE POLARIZATION
The probe is under polarization with a fixed voltage of approximately
800 mV.
Probe polarization is essential for stable measurements with the same
recurring degree of accuracy.
With the probe properly polarized, oxygen is continually consumed when
it passes through the sensitive diaphragm and dissolves in the electrolyte
solution contained in the probe.
If polarization is interrupted, the electrolyte solution continues to be
enriched with oxygen until it reaches an equilibrium with the surrounding
solution.
Whenever measurements are taken with a non-polarized probe, the
oxygen level revealed is both that of the tested solution, as well as that
present in the electrolyte solution. This reading is incorrect.
The calibration of this instrument is very simple.
Before proceeding with the calibration, make sure the probe is ready for
measurements (see page 7), i.e. the membrane cap is filled with electrolyte
and the probe is connected to the meter and properly polarized.
8
For an accurate calibration, it is recommended to wait at least 15 minutes
to ensure precise conditioning of the probe.
Keep the protective cap on during polarization time and remove it for
calibration and measurements. Follow the calibration procedure (see
page 10).
SALINITY AND ALTITUDE COMPENSATION
If the sample contains significant concentration of salinity or if you are
performing measurements at an altitude different from sea level, the
read out values must be corrected, taking into account the lower degree
of oxygen solubility in these situations (see pages 15-16).
Remember to set the altitude and/or the salinity before taking any DO
measurements. The meter will automatically compensate for these factors.
DO MEASUREMENTS
Make sure that the instrument has been calibrated and
the protective cap has been removed.
• Submerse the tip of the probe in the sample to be
tested. Allow approximately one minute for the reading
to stabilize.
• The Dissolved Oxygen value (in %) is displayed on the primary LCD
and the temperature on the secondary LCD.
• Press RANGE to change the reading from % to ppm and vice-versa.
For accurate Dissolved Oxygen measurements, a water movement of
0.3 m/s is required. This is to ensure that the oxygen-depleted
membrane surface is constantly replenished. A moving stream will
provide adequate circulation.
The use of a magnetic stirrer to ensure a certain fluid velocity is
recommended.
9
TEMPERATURE MEASUREMENTS
HI 7040
HI 7040
The probe has a built-in temperature sensor.
The measured temperature is indicated on the secondary LCD as shown
above.
Allow the probe to reach thermal equilibrium before taking any measurement.
This can take several minutes. The greater the difference between the
temperature at which the probe was stored and the temperature of the
sample, the longer the time will be.
Note: If “----” is displayed, the DO probe is not properly connected. This
also indicates the posibility of a broken probe cable. If temperature
displayed blinking, the temperature is out of range.
DO CALIBRATIONDO CALIBRATION
DO CALIBRATION
DO CALIBRATIONDO CALIBRATION
Calibrate the instrument frequently, especially if high accuracy is
required.
The instrument can be calibrated in maximum 2 points: 0.0% (zerocalibration) and 100.0% (slope calibration).
The zero calibration of the HI 2400 is very stable, therefore this procedure
needs to be performed only whenever the probe or the membrane is
replaced.
However, because the slope calibration is more critical, it is recommended
to perform this procedure every week.
INITIAL PREPARATION
• Pour small quantities of HI 7040 Zero Oxygen
solution into a beaker. If possible, use a plastic
beaker to minimize any EMC interferences.
• Make sure the probe is ready for measurements
(see probe preparation on page 7), i.e. the
membrane is filled with electrolyte and the probe
is connected to the meter.
• Switch the meter on by pressing the ON/OFF
switch.
• For an accurate calibration, it is recommended
to wait for at least 15 minutes to ensure precise
conditioning of the probe.
• Remove the protective cap from the DO probe.
• Set the appropriate altitude factor (see page 15). Make sure the
salinity factor is set to zero (see page 16).
10
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.