3
2
PRELIMINARY EXAMINATION
HI 2314 and HI 2315 are bench top multi-range conductivity meters for
the laboratory.
Four measurement ranges ensure the highest resolution and precision for
your measuring requirements.
Single point calibration is performed manually simply by adjusting a
knob on the front of the meter.
Temperature compensation of the readings is automatic for the HI 2315
and HI 23151 models, thanks to the temperature sensor incorporated in
probe, with userselectable temperature coefficient.
With HI 2314 the temperature compensation is manual and the temperature value can be also set with a knob.
The rugged probe with platinum 4-ring sensor responds faster than
conventional stainless steel models and can be also used for measuring
highly acidic or alkaline samples, or at high temperature.
In addition, the HI 23151 model offers analog output of 0 to 5V that
represents the full conductivity scale across all 4 ranges.
GENERAL DESCRIPTION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your dealer.
Each meter is supplied complete with:
• 4-ring conductivity probe with 1 m (3.3’) cable
• HI 76300 for HI 2314
• HI 76303 for HI 2315 and HI 23151
• 12 Vdc adapter
• Instruction manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.
TABLE OF CONTENTS
WARRANTY ................................................................................. 2
PRELIMINARY EXAMINATION ....................................................... 4
GENERAL DESCRIPTION ............................................................... 4
FUNCTIONAL DESCRIPTION HI 2314 ........................................... 5
SPECIFICATIONS HI 2314 ........................................................... 6
FUNCTIONAL DESCRIPTION HI 2315 ........................................... 7
SPECIFICATIONS HI 2315 AND HI 23151 ................................... 8
OPERATIONAL GUIDE .................................................................. 9
CALIBRATION ............................................................................ 11
CONDUCTIVITY VS. TEMPERATURE CHART................................... 14
DETERMINATION OF THE TEMPERATURE
COEFFICIENT OF A SOLUTION (HI 2315) ................................... 15
PROBE MAINTENANCE ............................................................... 16
ACCESSORIES ............................................................................ 17
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using these
instruments. This manual will provide you with the necessary
information for correct use of these instruments, as well as a precise
idea of their versatility.
If you need additional technical information, do not hesitate to
e-mail us at tech@hannainst.com or view our worldwide contact list
at www.hannainst.com.
WARRANTY
HI 2314, HI 2315 and HI 23151 are guaranteed for two years
against defects in workmanship and materials when used for their
intended purpose and maintained according to instructions.
Electrodes and probes are guaranteed for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact the dealer from whom you purchased the
instrument.
If under warranty, report the model number, date of purchase, serial
number and the nature of the problem.
If the repair is not covered by the warranty, you will be notified of the
charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization number from the Technical Service
department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packed for
complete protection.