Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this instrument.
This manual will provide you with the necessary information for correct
use of this instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us
at tech@hannainst.com.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 2222 is guaranteed for two years against defects in workmanship and
materials when used for their intended purpose and maintained according
to instructions. The electrodes and the probes are guaranteed for a
period of six months. This warranty is limited to repair or replacement free
of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair is
not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization number from the Technical Service department
and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
RECOMMENDATIONS FOR USERS ............................................................ 3 9
2
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully
to make sure that no damage has occurred during shipping. If there is any
damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied complete with:
• HI 1048P Glass-body Combination pH Electrode with 1 m (3.3 “)
cable
• HI 7662 Temperature Probe
• HI 76404N Electrode Holder
• pH 3.00 & 7.01 Buffer Solutions (20 mL each)
• HI 700365 Cleaning Solution for wine deposits (2x20 mL)
• HI 700636 Cleaning Solution for wine stains (2x20 mL)
• HI 7082 Electrode Refiling Solution (30 mL)
• 12 VDC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI 2222 is a logging microprocessor-based pH/ORP/temperature bench
meters with Calibration Check.
Calibration Check performs a set of diagnostic tests during calibration
using the history of electrode slope and offset to detect problems that can
cause loss of accuracy.
Calibration Check Features are:
• Enhanced Calibration Messages
During calibration the user is warned if one or more parameters are not suitable
to perform an accurate calibration.
• Electrode Condition on LCD Display
Determined from the electrode offset and slope.
• Electrode response time on LCD Display
Determined from electrode performance during calibration.
Other features include: up to two point calibration with seven memorized
buffers (1.68, 3.00, 6.86, 7.01, 9.18, 10.01 and 12.45 pH), logging
up to 100 samples, GLP, calibration due alarm, pH reading with manual
or automatic temperature compensation and PC software interface.
3
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
Front Panel
Rear Panel
1) Liquid Crystal Display (LCD).
2) CAL key, to enter and exit calibration mode.
3) CFM/GLP key, to confirm calibration, different values or to display
Good Laboratory Practice information.
4) ºC key, to manually increase temperature value or other parameters.
5) ºC key, to manually decrease temperature value or other parameters.
6) SETUP key, to enter/exit SETUP mode.
7) RANGE key, to select measurement range, switch to focused data in
SETUP or toggle between buffer value and temperature during
calibration.
8) LOG/CLR key, to store a value into memory, to clear pH calibration,
or to delete log records.
9) RCL key, memory recall.
10) ON/OFF switch.
11) Power supply socket.
12) USB connector.
13) BNC electrode connector.
14)Temperature probe socket.
15) Electrode reference socket.
4
HI 2222HI 2222
HI 2222
HI 2222HI 2222
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
egnaR
noituloseR
Hp00.61ot00.2–
Vm9.996±
Vm0002±
C°0.021ot0.02–
Hp10.0
)Vm9.996±(Vm1.0
)Vm0002±(Vm1
C°1.0
Hp10.0±
ycaru
ccA
F°86/C°02@
kcehCnoitarbilaCseY
ecafretnI
retupmoCBSUdetalosi-otpO
noitarbilaCHp
gniggoLstniop001
edortcelEHp
ylppuSrewoP)dedulcni(retpadaCDV21
snoisnemiD901x222x532mm)"3.4x7.8x2.9(
thgieW
tnemnorivnE
ytnarraW2sraey
1(
noitasnepmoCerutarepmeT
P8401IHelballifernoitcnujelgnis,ydobssalg
eborPerutarepmeT2667IH)dedulcni(eborp
ecnadepmItupnI01
)Vm9.996±(Vm2.0±
)Vm0002±(Vm1±
C°2.0±
rorreeborpgnidulcxe
elbaliavasreffub7,stniop2otpU
)54.21,10.01,81.9,10.7,68.6,00.3,86.
:morfcitamotuArolaunaM
)F°0.842ot0.4–(C°0.021ot0.02–
)dedulcni(nip+CNB,llec
leetssselniats
21
mho
;)bl9.2(gK3.1
edlohhtiwtik
)bl6.4(gK1.2r
)F°221–23(C°05–0
gnisnednocnon%59HRxam
5
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
POWER CONNECTIONPOWER CONNECTION
POWER CONNECTION
POWER CONNECTIONPOWER CONNECTION
Plug the 12 VDC adapter into the power supply socket.
Notes: • These instruments use non volatile memory to retain the pH,
mV, temperature calibrations and all other settings, even
when unplugged.
• Make sure a fuse protects the mains line.
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
For HANNA P Type pH or ORP electrodes (with internal reference) connect
the electrode’s BNC to the socket on the back of the instrument and the
pin to the reference socket.
Note: Electrode condition and response information is displayed on the
bar graph gauges during the day the calibration is performed,
only if HANNA P type (PIN) electrodes are used.
If the electrode is not recognized as a HANNA P type electrode, the
bar graph gauges will blink (25 seconds OFF, 4 seconds ON, full
bar graph).
For temperature measurement and automatic temperature compensation
connect the temperature probe to the appropriate socket.
INSTRUMENT STINSTRUMENT ST
INSTRUMENT ST
INSTRUMENT STINSTRUMENT ST
• Turn the instrument on by pressing the ON/OFF switch located on
the rear panel.
• All LCD tags are displayed and a beep is sounded while the instrument
performs a self test.
ARTART
ART
ARTART
-UP-UP
-UP
-UP-UP
• The instrument will display “LoAd” message and “ ” blinking until
initialization is complete.
• The “Unscrew electrode refilling cap” message reminds the user
to loosen or remove the electrode refilling cap to improve the electrode’s
response time.
• The instrument automatically defaults to pH measurement mode
unless a HANNA P type ORP electrode is detected.
6
pH MEASUREMENTpH MEASUREMENT
pH MEASUREMENT
pH MEASUREMENTpH MEASUREMENT
Make sure the instrument has been calibrated
before taking pH measurements.
• Submerse the tip of a properly conditioned
electrode (see page 31) and the temperature
probe approximately 3 cm (1¼”) into the sample
3 cm
(1¼")
to be tested and stir gently.
Allow time for the electrode to stabilize.
• The pH is displayed on the primary LCD and the temperature on the
secondary LCD.
• The pH reading is out of range, the closest full-scale value will be
displayed blinking on the primary LCD.
• It is also possible to view the mV reading by pressing the RANGE key.
POOR GOOD
SLOW
FAST
mV
If measurements are taken successively in different samples, it is recommended
to rinse the electrode thoroughly with deionized water or tap water and
then with some of the next sample to prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the pH
accurately, this temperature effect must be compensated for. To use the
Automatic Temperature Compensation feature, connect and submerse the
HI 7662 temperature probe into the sample as close to the electrode as
possible and wait for a few minutes.
ATC
If the temperature of the sample is known, manual
temperature compensation can be used by disconnecting
the temperature probe.
MTC
The display will show the last recorded temperature
reading with the “°C” symbol blinking.
The temperature can now be adjusted with the ARROW keys (from
-20.0 ºC to 120.0 ºC).
C
C
C
7
ORP MEASUREMENTSORP MEASUREMENTS
ORP MEASUREMENTS
ORP MEASUREMENTSORP MEASUREMENTS
An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-Reduction Potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the tested sample.
The surface of the ORP electrode must be clean and smooth in order to
obtain an accurate measurement.
Pretreatment solutions are available to condition the electrode and speed
up the response time.
• The instrument automatically defaults to the
mV measurement mode if HANNA P type ORP
electrode is detected.
• Submerse the ORP electrode tip (3 cm/1¼”)
3 cm
(1¼")
into the sample. Allow a few minutes for the
reading to stabilize.
• The instrument displays the mV reading on the primary LCD.
• If the reading is out of range, the closest full-scale value will be
displayed blinking on the primary LCD.
TAKING TEMPERATURE MEASUREMENTSTAKING TEMPERATURE MEASUREMENTS
TAKING TEMPERATURE MEASUREMENTS
TAKING TEMPERATURE MEASUREMENTSTAKING TEMPERATURE MEASUREMENTS
Connect the HI 7662
instrument on. Dip the temperature
temperature probe and turn the
probe
into the sample
and allow the reading on the secondary LCD to stabilize.
8
pp
H CALIBRATIONH CALIBRATION
p
H CALIBRATION
pp
H CALIBRATIONH CALIBRATION
Calibrate the instrument frequently, especially if high accuracy is
required. For best results and constant display of electrode condition
and electrode response on the bar graph gauges, daily calibration is
recommended.
The instrument should be re-calibrated:
• Whenever the pH electrode is replaced.
• At least once a day.
• After testing aggressive chemicals.
• If high accuracy is required.
• If “CAL DUE” message is displayed during measurement.
Every time you calibrate the instrument use fresh buffers and perform an
electrode cleaning procedure (see page 33).
PREPARATIONPREPARATION
PREPARATION
PREPARATIONPREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible,
use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two
beakers for each buffer solution. One for rinsing the electrode and one for
calibration.
PROCEDUREPROCEDURE
PROCEDURE
PROCEDUREPROCEDURE
A two-point calibration is recommended, however a one-point calibration
will be permitted. Calibration can be performed using the seven
memorized buffers:
• pH 1.68, 3.00, 6.86, 7.01, 9.18, 10.01 and 12.45.
9
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
TWO-POINT CALIBRATION
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
For most applications it is recommended that pH 7.01 or 6.86 buffers be
used as the first calibration point and pH 3.00 (for acidic samples) or pH
9.18/10.01 (for alkaline samples) as the second calibration point.
Note: The pH 12.45 buffer is not for general measurement; use only if
the sample is very alkaline to avoid sodium error.
• Submerse the pH electrode and the temperature probe approximately
3 cm (1¼”) into a buffer solution and stir gently. The temperature
probe should be close to the pH electrode.
• Press the CAL key. “CAL” and “pH” tags will be on, and the
“CLEAR CAL if new electrode” tag will blink.
CAL
CLEAR CAL if new electrode
pH
• Press the CLR key if you are using a new electrode or want to clear
the calibration history. The instrument will display the “donE”
message for a few seconds.
• Press the CAL key, or wait a few seconds to continue.
It is very important to clear the calibration history when a new electrode
is used because all error and warning messages that appear during
calibration depend on the calibration history.
Note: • The “CLEAR CAL if new electrode” will only appear if the
instrument has been previously calibrated.
• The “CAL”, “pH” and “BUFFER” tags will appear and the
“7.01” buffer will be displayed on the secondary LCD.
• Press the ARROW keys to select a different buffer value, if
necessary.
• The “ ” tag will blink until the reading has stabilized.
10
• When the reading is stable and close to the selected buffer, the
“CFM” tag will blink and if enabled, an audible signal will sound.
• Press the CFM key to confirm the calibration. The calibrated value
will be displayed on the primary LCD and the second expected buffer
value on the secondary LCD.
• After the first calibration point is confirmed, submerse the pH electrode
and the temperature probe approximately 3 cm (1¼”) into the
second buffer solution and stir gently. The temperature probe should
be close to the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
Note: The instrument will automatically skip the buffer used for the first
point. It also skips 6.86 if 7.01 was used, and vice versa.
Likewise, it will skip 9.18 if 10.01 has been used, and vice versa.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer the “CFM”
tag will blink.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the
primary LCD and and the third expected buffer
value on the secondary LCD.
11
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
ONE-POINT CALIBRATION
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
If the “Pnt” option is selected, the new calibration point overrides an
existing one. The adjacent slopes will be reevaluated.
If the “OFFS” option is selected, an electrode offset correction is performed.
The adjacent slopes will remain unchanged.
• Proceed as described in “TWO-POINT CALIBRATION” section.
• Press CAL after the first calibration point was confirmed. The
instrument will memorize the one-point calibration data and return
to measurement mode.
Notes: • To clear calibration parameters for all uncalibrated buffers
starting with current buffer, press CLR. The calibration
will continue from the current point. If this procedure is
performed while calibrating in the first calibration point, the
instrument returns to measurement mode.
• Press RANGE to toggle between pH buffer
and temperature reading.
12
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.