Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instruments.
This manual will provide you with the necessary information for correct use of
the instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
These instruments are in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 2210 and HI 2211 are guaranteed for two years against defects
in workmanship and materials when used for their intended purpose
and maintained according to instructions. Electrodes and probes are
guaranteed for six months. This warranty is limited to repair or
replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair
is not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization number from the Technical
Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packed for
complete protection.
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your Dealer or the nearest Hanna
Customer Service Center.
Each instrument is supplied with:
• HI 1131B Glass-body Combination pH Electrode
• HI 7662 Temperature Probe
• HI 76404 Electrode Holder
• pH 4.01 & 7.01 Buffer Solutions, 20 mL each
• HI 7071S Electrolyte Solution
• 12VDC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The Hanna HI 2210 and HI 2211 are microprocessor-based pH and
temperature bench meters.
HI 2211 can also be used for ion concentration (ISE) and Oxidation
Reduction Potential (ORP) in the mV range.
pH measurements are compensated for temperature effect manually or
automatically with the HI 7662 temperature probe.
The instrument is equipped with a large easy-to-read LCD which shows
the pH (or mV) and temperature simultaneously, together with graphic
symbols.
The calibration process is guided step by step through clear indications on
the LCD. A stability indicator makes the calibration procedure error-free.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
Front PanelFront Panel
Front Panel
Front PanelFront Panel
Rear PanelRear Panel
Rear Panel
Rear PanelRear Panel
1) Liquid Crystal Display (LCD).
2) CAL key, to enter or exit/escape calibration mode.
3) CFM key, to confirm different values.
4)°C key and °C keys, to manually increase/decrease tem-
perature or select pH buffer.
5) RANGE key, to select measurement range (HI 2211 only).
6) MEM key, to store a value into memory.
7) MR key, to recall the stored value.
8) Secondary LCD.
9) Primary LCD.
10) ON/OFF switch.
11) Power adapter socket.
12) BNC electrode connector.
13) Temperature probe socket.
14) Electrode reference socket.
45
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
EGNAR
NOITULOSER
YCARUCCA
F°86/C°02@
noitarbilaCHp
edortcelEHp)dedulcni(B1311IH
eborperutarepmeT)dedulcni(2667IH
ecnadepmitupnI01
ylppusrewoP)dedulcni(retpadaCDV21
snoisnemiD)”9.2x1.7x4.9(mm47x281x042
thgieW
tnemnorivnE
ytnarraWsraey2
±)ylno1122IH(Vm9.993
±)ylno1122IH(Vm0002
±)ylno1122IH(Vm2.0
±)ylno1122IH(Vm1
±)C°0.001–0.0(C°5.0
noitasnepmocerutarepmeT
1±)edistuo(C°
±Hp10.0
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
POWER CONNECTIONPOWER CONNECTION
POWER CONNECTION
POWER CONNECTIONPOWER CONNECTION
Hp00.61ot00.2–
Plug the 12 VDC adapter into the power supply socket.
Notes: • These instruments use non volatile memory to retain the
pH, mV, temperature calibrations and all other settings,
even when unplugged.
C°0.021ot9.9–
Hp10.0
)ylno1122IH(Vm1.0
)ylno1122IH(Vm1
C°1.0
• Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
For pH or ORP combination electrode connect to the BNC connector on
the back of the instrument.
For electrodes with a separate reference connect the electrode’s BNC to the
BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
INSTRUMENT START-UPINSTRUMENT START-UP
INSTRUMENT START-UP
INSTRUMENT START-UPINSTRUMENT START-UP
• Turn the instrument on by pressing the ON/OFF switch located
on the rear panel.
)rorreeborpgnidulcxe(
sreffub5,noitarbilactniop2ro1
elbaliava
)10.01,81.9,10.7,68.6,10.4(
:morfcitamotuArolaunaM
)F°0.842ot2.41(C°0.021ot9.9–
• All LCD segments are displayed while the instrument performs a
self test.
pH MEASUREMENTSpH MEASUREMENTS
pH MEASUREMENTS
pH MEASUREMENTSpH MEASUREMENTS
Make sure the electrode and the instrument have been
calibrated together before taking pH measurements.
21
smho
• Submerge the electrode and the temperature
probe approximately 4 cm (1½”) into the sample
to be tested and stir gently. Allow time for the
electrode to stabilize.
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
)F°221ot23(C°05ot0
gnisnednoc-nonHR%59.xam
• The pH is displayed on the primary LCD and the temperature on
the secondary LCD.
pH
C
• If the pH reading is out of range, “----” will be displayed on the
LCD.
67
If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized water
or tap water and then with some of the next sample in order to
prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the
pH accurately, the temperature effect must be compensated for.
To use the Automatic Temperature Compensation feature, connect
and submerge the HI 7662 temperature probe into the sample as
close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, manual temperaturecompensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature
of 25 °C or the last recorded temperature reading
C
with the “°C” tag blinking. The temperature can now
be adjusted with the ARROW keys (from -9.9 °C to
120.0 °C).
TEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTS
TEMPERATURE MEASUREMENTS
TEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTS
Connect the HI 7662
temperature probe to the TEMP
socket and turn the instrument on.
Immerse the temperature
probe
into the sample and
allow the reading on the secondary LCD to stabilize.
MEMORY FUNCTIONMEMORY FUNCTION
MEMORY FUNCTION
MEMORY FUNCTIONMEMORY FUNCTION
Press and hold down MEM. The last reading will appear on the LCD
along with “MEM” tag until MEM is released.
Press MR and the previously memorized reading will be displayed
with “MEM” tag on the LCD.
ORP MEASUREMENTS (HORP MEASUREMENTS (H
ORP MEASUREMENTS (H
ORP MEASUREMENTS (HORP MEASUREMENTS (H
II
I
II
2 2
2
2 2
22
1111
only only
only
only only
))
)
))
2
11
22
1111
An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-reduction potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface
of the ORP electrode must be clean and smooth.
• Press RANGE to enter mV range.
• Submerge the tip of the ORP electrode 4 cm
(1½") into the sample to be tested and allow a
few seconds for the reading to stabilize.
• The instrument displays the mV reading on the primary LCD and
the temperature on the secondary LCD.
mV
C
• If the reading is out of range, “----” will be displayed on the
LCD.
89
pp
H CALIBRATIONH CALIBRATION
p
H CALIBRATION
pp
H CALIBRATIONH CALIBRATION
Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
PREPARATIONPREPARATION
PREPARATION
PREPARATIONPREPARATION
Pour small quantities of the buffer solutions into clean beakers. If
possible use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two
beakers for each buffer solution: one for rinsing the electrode and one for
calibration.
If you are measuring in the acidic range, use pH 7.01 as first buffer
and pH 4.01 as second buffer. If you are measuring in the alkaline
range, use pH 7.01 as first buffer and pH 10.01 or pH 9.18 as
second buffer.
PROCEDUREPROCEDURE
PROCEDURE
PROCEDUREPROCEDURE
Calibration has a choice of five memorized buffers: pH 4.01, 6.86,
7.01, 9.18 and 10.01.
It is recommended to perform a two-point calibration. However, onepoint calibration is also permitted by the instruments.
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
TWO-POINT CALIBRATION
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
• Submerge the pH electrode and the temperature probe approximately 4 cm (1½") into a
buffer solution and stir gently. The temperature
probe should be close to the pH electrode.
• Press CAL. The “CAL” and “
” messages will appear and
“7.01” buffer will be displayed on the secondary LCD.
pH
CAL
CAL
1
BUF
BUFFER pH
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” symbol will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer,
“READY” message will appear and “CFM” message will blink.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the second expected buffer value.
pH
CAL
2
BUF
BUFFER pH
Note: The instruments will automatically skip the buffer used for the
first point. It also skips 6.86 if 7.01 buffer was used and vice
versa. Likewise, it will skip 9.18 if 10.01 buffer was used and
vice versa.
• After the first calibration point is confirmed,
immerse the pH electrode and the temperature
probe approximately 4 cm (1½”) into the
second buffer solution and stir gently. The
temperature probe should be close to the pH
electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” symbol will blink on the LCD until the reading is stable.
• When the reading is stable, “READY” message will appear and
“CFM” message will blink.
• Press CFM to confirm calibration. The instrument
will return to measurement mode.
Notes: • If the value measured by the meter is not close to the
selected buffer, “WRONG ” and “WRONG ” messages
will blink alternately. In this case check if the correct buffer
has been used, or regenerate the electrode by following the
cleaning procedure (see page 18). If necessary, change the
buffer or the electrode.
• The “WRONG” message and temperature value are displayed blinking if the temperature reading is out of the
defined temperature range of the buffer. Calibration cannot
be confirmed in this situation.
1011
• Press RANGE to display the tem-
CFM
CAL
C
perature reading on the LCD during
calibration (HI 2211 only).
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
ONE-POINT CALIBRATION
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
• Proceed as described in “Two-point calibration” section.
• Press CAL after the first calibration point
was confirmed.
The instrument will return to measurement mode
and will memorize the one-point calibration data.
Note: Press and hold down CFM and then press CAL. The instrument
sets default calibration parameters, displays “CLR” message for a
few seconds and then returns to normal measurement mode.
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
((
for technical personnel onlyfor technical personnel only
(
for technical personnel only
((
for technical personnel onlyfor technical personnel only
))
)
))
All the instruments are factory calibrated for temperature.
Hanna’s temperature probes are interchangeable and no tempera-
ture calibration is needed when they are replaced.
If the temperature measurements are inaccurate, temperature
recalibration should be performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions bellow.
• Prepare a vessel containing ice and water and another one
containing hot water (at a temperature of around 50 ºC). Place
insulation material around the vessels to minimize temperature
changes.
• Use a calibrated thermometer with a resolution of 0.1 ºC as a
reference thermometer.
• With the instrument off, press and hold down the CAL & MEM
keys, then power on the instrument. The “CAL” message will
appear and the secondary LCD will show 0.0 ºC.
CAL
• Immerse the temperature probe in the vessel with ice and water
as near as possible to the reference thermometer. Allow a few
seconds for the probe to stabilize.
• Use the ARROW keys to set the reading on the secondary LCD to
that of ice and water, measured by the reference thermometer.
• When the reading is stable and close to the selected calibration
point, “READY” message will appear and “CFM” message will blink.
• To confirm press CFM. The secondary LCD will show 50.0 ºC.
1213
C
• Immerse the temperature probe in the second vessel as near as
possible to the reference thermometer. Allow a few seconds for the
probe to stabilize.
• Use the ARROW keys to set the reading on the secondary LCD to
that of the hot water.
• When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The instrument returns
to measurement mode.
Note: If the reading is not close to the selected
calibration point, “WRONG” tag will blink.
Change the temperature probe and restart
calibration.
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
p
H BUFFER TEMPERATURE
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
DEPENDENCEDEPENDENCE
DEPENDENCE
DEPENDENCEDEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions. During calibration the instrument will automatically calibrate
to the pH value corresponding to the measured or set temperature.
PMETSREFFUBHp
CºFº10.468.610.781.910.01
02310.489.631.764.923.01
51400.459.601.793.942.01
010500.429.670.733.981.01
519500.409.650.772.921.01
028600.488.630.722.960.01
527710.468.610.781.910.01
036820.458.600.741.969.9
535930.448.699.611.929.9
0440140.448.689.670.988.9
5431150.438.689.640.958.9
0522160.438.689.610.928.9
5513180.448.689.699.897.9
0604190.448.689.679.877.9
5694111.448.699.659.867.9
0785121.458.699.639.857.9
5776141.468.600.719.847.9
0867161.478.610.798.847.9
5858171.478.620.778.847.9
0949191.488.630.758.857.9
5930202.498.640.738.867.9
During calibration the instrument will display the pH buffer value at
25 ºC.
1415
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
PREPARATION PROCEDUREPREPARATION PROCEDURE
PREPARATION PROCEDURE
PREPARATION PROCEDUREPREPARATION PROCEDURE
Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes. They will disappear when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by “shaking down” the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or
HI 80300
Storage Solution for at least one hour.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1") below the
fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for
double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte
Solution for single junction electrodes.
Unscrew the fill hole screw during measurements.
®®
®
For AmpHelFor AmpHel
For AmpHel
For AmpHelFor AmpHel
®®
electrodes: electrodes:
electrodes:
electrodes: electrodes:
If the electrode does not respond to pH changes, the battery is dead
and the electrode should be replaced.
MEASUREMENTMEASUREMENT
MEASUREMENT
MEASUREMENTMEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom
4 cm /1½”) into the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.
STORAGE STORAGE
STORAGE
STORAGE STORAGE
PROCEDUREPROCEDURE
PROCEDURE
PROCEDUREPROCEDURE
To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300
or HI 80300 Storage Solution or, in its absence, Fill Solution (HI7071 or HI 8071 for single junction and HI 7082 or HI 8082 for
double junction electrodes). Follow the Preparation Procedure on page
16 before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.
PERIODIC MAINTENANCEPERIODIC MAINTENANCE
PERIODIC MAINTENANCE
PERIODIC MAINTENANCEPERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable must be intact and
well connected. No cracks should be seen on the electrode stem or
bulb. Connectors must be perfectly clean and dry. If any scratches or
cracks are present, replace the electrode. Rinse off any salt deposits
with water.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071
for single junction or HI 7082 or HI 8082 for double junction
electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
1617
CLEANING PROCEDURECLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURECLEANING PROCEDURE
•
General
Soak in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.
•
Protein
Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.
•
Inorganic
Soak in Hanna HI 7074 Inorganic Cleaning
Solution for 15 minutes.
•
Oil/grease
Rinse with Hanna HI 7077 or HI 8077 Oil and
Fat Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte (not necessary for gel-filled electrodes)
and soak the electrode in HI 70300 or HI 80300 Storage Solution for
at least 1 hour before taking measurements.
The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes more time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 25 °C.
Since the resistance of the pH electrode is in the range of 50-200 Mohms,
the current across the membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and
static discharges are detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature1- 3 years
90 °CLess than 4 months
120 °CLess than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is called
alkaline error and causes the pH to be underestimated. Hanna’s glass
formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 °C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
H SENSITIVE GLASSH SENSITIVE GLASS
p
H SENSITIVE GLASS
pp
H SENSITIVE GLASSH SENSITIVE GLASS
+
+
13.00
13.50
14.00
12.50
13.00
13.50
14.00
2021
0.10
0.14
0.20
0.10
0.18
0.29
0.40
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
pH BUFFER SOLUTIONS
HI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcs
HI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcs
HI 70010P pH 10.01 Buffer Sachets, 20 mL, 25 pcs
HI 7004L pH 4.01 Buffer Solution, 500 mL bottle
HI 7006L pH 6.86 Buffer Solution, 500 mL bottle
HI 7007L pH 7.01 Buffer Solution, 500 mL bottle
HI 7009L pH 9.18 Buffer Solution, 500 mL bottle
HI 7010L pH 10.01 Buffer Solution, 500 mL bottle
HI 8004L pH 4.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8006L pH 6.86 Buffer Sol. in FDA approved bottle, 500 mL
HI 8007L pH 7.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8009L pH 9.18 Buffer Sol. in FDA approved bottle, 500 mL
HI 8010L pH 10.01 Buffer Sol. in FDA approved bottle, 500 mL
ELECTRODE STORAGE SOLUTIONS
HI 70300L Storage Solution, 460 mL bottle
HI 80300L Storage Solution in FDA approved bottle, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Rinse Sachets, 20 mL, 25 pcs
HI 7061L General Cleaning Solution, 460 mL bottle
HI 7073L Protein Cleaning Solution, 460 mL bottle
HI 7074L Inorganic Cleaning Solution, 460 mL bottle
HI 7077L Oil & Fat Cleaning Solution, 460 mL bottle
HI 8061L General Cleaning Sol. in FDA approved bottle, 460 mL
HI 8073L Protein Cleaning Solution in FDA approved bottle, 460 mL
HI 8077L Oil & Fat Cleaning Sol. in FDA approved bottle, 460 mL
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI 70713.5M KCl+AgCl Electrolyte, 4x50 mL, for single junction
electrodes
HI 70721M KNO3 Electrolyte, 4x50 mL
HI 70823.5M KCl Electrolyte, 4x50 mL, for double junction
electrodes
HI 80713.5M KCl+AgCl Electrolyte in FDA approved bottle,
4x50 mL, for single junction electrodes
HI 80721M KNO3 Electrolyte in FDA approved bottle, 4x50 mL
HI 80823.5M KCl Electrolyte in FDA approved bottle, 4x50 mL
95
ORP PRETREATMENT SOLUTIONS
HI 7091LReducing Pretreatment Solution, 460 mL bottle
HI 7092LOxidizing Pretreatment Solution, 460 mL bottle
pH ELECTRODES
All electrodes part numbers ending with B are supplied with BNC
connector and 1 m (3.3') cable, as shown below:
Plastic-body (PEI), gel-filled, combination platinum ORP electrode.
Use: general purpose.
12 mm
0.5"
HI 3230
120 mm
"S"VERSION
4.7"
HI 4430B
Plastic-body (PEI), gel-filled, combination gold ORP electrode. Use:
general purpose.
12 mm
0.5"
HI 4430
120 mm
"S"VERSION
4.7"
Consult the Hanna General Catalog for more electrodes with screwtype or BNC connectors.
FC 911
110 mm
4.3"
2425
EXTENSION CABLE FOR SCREW-TYPE ELECTRODES
(SCREW TO BNC ADAPTER)
HI 7855/1Extension cable 1 m (3.3') long
HI 7855/3Extension cable 3 m (9.9') long
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
OTHER ACCESSORIES
HI 710005 Voltage adapter from 115 VAC to 12 VDC (USA plug)
HI 710006 Voltage adapter from 230 VAC to 12 VDC (European plug)
HI 710012 Voltage adapter from 240 VAC to 12 VDC (UK plug)
HI 710013 Voltage adapter from 230 VAC to 12 VDC (South Africa plug)
HI 710014 Voltage adapter from 230 VAC to 12 VDC (Australia plug)
HI 76405 Electrode holder
RECOMMENDATIONS FOR USERS
Before using these products, make sure they are entirely suitable for
the environment in which they are used.
Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment, requiring the
operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may
degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltages
at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in
microwave ovens.
HI 8427pH and ORP electrode simulator with 1 m (3.3') coaxial
cable ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 7662Temperature probe with 1 m (3.3') cable
Windows® is registered Trademark of "Microsoft Co."
2627
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. (662) 619.0708.11 • Fax (662) 619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and
Technical offices, please see www.hannainst.com
28
MAN2211R1
02/08
.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.