GW Instek GOS-6103C User Manual

GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
CONTENTS PAGE
1. PRODUCT INTRODUCTION................................................
1-1.Description………………………………………………. 1-2.Feature…………………………………………………...
1 1 2
2. TECHNICAL SPECIFICATION………………………….. 4
3. PRECAUTIONS BEFORE OPERATION…….………….. 3-1.Unpacking the instrument………………….………….. 3-2.Checking the Line Voltage…………………..…………. 3-3.Environment……………………………………..………
3-4.Equipment Installation and Operation………………....
3-5.CRT Intensity…………………………………………… 3-6.Withstanding Voltage of Input Terminals……………...
7 7 7 8 8 8 8
4. PANEL INTRODUCTION……………………..………….. 4-1.Front Panel………………………………………………. 4-2.Rear Panel……………………………………….………
9 11 31
5. OPERATION METHOD………………………………...… 5-1.Readout Display……………………………………..…... 5-2.Connecting Input Signals..……………………………… 5-3.Adjustment and Checks………………………………… 5-4.Function Check…..……………………………………… 5-5.Basic Operation………….….…………………………... 5-6.Measurement Application……………………………….
33 33 35 36 38 40 48
6. MAINTENANCE…………………………………………… 6-1.Fuse Replacement……………………………………….. 6-2.Line Voltage…………………………………………..…. 6-3.Cleaning………………………………………………….
53 53 53 54
7. BLOCK DIAGRAM……………………………………….... 55
USER MANUAL
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
WARNING. Warning statements identify condition or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
The following symbols may appear in this manual or on the product:
DANGER ATTENTION Protective Earth(ground) High Voltage refer to Manual Conductor Terminal Terminal
i 
ii
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
FOR UNITED KINGDOM ONLY
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with
the following code:
Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
USER MANUAL
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of
0.75mm
2
should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
The wire which is coloured Green & Yellow must be connected to the
Earth terminal marked with the letter E or by the earth symbol or coloured Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
iii
iv 
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
1.PRODUCT INTRODUCTION
1-1. Description
The GOS-6103C is a 100MHz, two-channel, dual-sweep, portable oscilloscope for general purpose use. A microprocessor-based operating system controls most of the functions of the instrument, including cursor readout and digitized panel setting. On-screen alphanumeric readout and cursor function for voltage, time, frequency and phase measurement provide extraordinary operational convenience. It also has the function of auto measurement for frequency, period, pulse width and duty cycle. Ten different user defined inst rument settings can be saved and recalle d without restriction.
The vertical deflection system has two input channels. Each channel has 11 basic deflection factors from 2mV to 5V per division. The horizontal deflection system provides single, dual or delayed sweeps from 0.5s to 50ns per division (delayed sweep, 50ms to 50ns per division). The trigger system provides stable triggering over the full bandwidth of the vertical deflection system.
USER MANUAL
1-2.Features
Additionally, the oscilloscope offers several other features:
1) High intensity and internal graticule CRT The oscilloscope employs a high intensity 6-inch retangular type cathode-ray tube with red internal graticule. It display s clear readable traces even at high sweep speeds. Internal graticule lines eliminate parallax-viewing error between the trace and the graticule line.
2) Temperature compensation The oscilloscope uses a temperature compensation circuit to reduce the drift of base line and DC balance.
3) 20MHz bandwidth limit When it is hard to observe or trigger a signal because a high-frequency component is superimposed on the signal, use the 20MHz BWL function to reduce the bandwidth of the vertical deflection system and trigger system to 20MHz.
4) Automatic time base range setting At a press of the AUTORANGE button, an optimum time base range is automatically set to a corresponding change in input signal period. A signal period from 1.6 to 4 cycles approx. is displayed.
5) TV triggering Exclusive TV sync separator circuit technology provides stable TV signal measurements on fields, frames and lines.
6) Z-axis intensity modulation For applying a blanking signal from an external source. The trace displayed on the screen may be intensity-modulated where pulse signal or time-scale marks are required.
1 
2 
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
7) Trigger signal output The signal selected by the TRIGGER SOURCE is available. This output may be used to connect to a frequency counter or other instrument.
8) Panel setups lock To avoid unintentional touch of the setting, the feature is extremely useful for long term and repetitive measurements that used to be performed under the same test condition of the oscilloscope setting.
9) LED indicator and buzzer alarm The LED’s located in the front panel assist operation and indicated additional information. Incorrect operation and the electrical end position of control knobs are indicated by a warning beep.
10)SMD manufacturing technology The instrument is built by using the most advanced SMD technology so as to reduce the number of internal wiring and shorten the foil rout e on the pc board. This will also greatly increase the high frequency performance and the reliability of the product.
11)Auto Measurement
A built-in 6 digits universal counter is accurate within the range of ±0.01% and can measure frequency between 50Hz and 100MHz.
USER MANUAL
2.TECHNICAL SPECIFICATIONS
VERTICAL DEFLECTION SYSTEM
TRIGGER SYSTEM
Sensitivity 2mV~5V/DIV,11 steps
Sensitivity Accuracy Vernier Vertical Sensitivity
Frequency Bandwidth(-3dB) DC ~ 100MHz (2mV/DIV:DC ~ 20MHz) Rise Time 3.5ns (2mV/DIV:17.5ns) Signal Delay Leading edge can be monitored Maximum Input Voltage 400V (DC+AC) at 1kHz or less Input Coupling AC, DC, GND Input Impedance
Vertical Modes CHOP Repetition Frequency Approx. 250kHz.
Polarity (INV) CH2 only Bandwidth Limited 20MHz Common-mode Rejection Ration Trigger Modes AUTO, NORM, TV
Trigger Source CH1, CH2, LINE, EXT Trigger Coupling AC, DC, HFR, LFR Trigger Slope +/- polarity or TV sync polarity
Trigger Sensitivity
Trigger Level Range TV Sync TV-V,TV-H
Max. External Input Voltage 400V (DC + AC peak) at 1kHz External Input Impedance
in 1-2-5 sequence ±3% (5 DIV at the center display ) Continuously variable to 1/2.5 or less than panel-indicated value
1MΩ±2% // 25pF approx. CH1,CH2,DUAL(CHOP/ALT),ADD(DIFF mode can
be established when the CH2 is in the INV mode)
50:1 or better at 50kHz
Mode Frequency INT EXT
AUTO
TV Sync signal 1DIV 200mVpp
INT : ± 4 DIV or more
EXT : ± 0.4 V or more
1MΩ± 5% ∕∕ 25pF approx.
10Hz~20MHz 0.35DIV 50mVpp
20MHz~100MHz 1.5DIV 150mVpp
DC~20MHz 0.35DIV 50mVpp NORM
20MHz~100MHz 1.5DIV 150mVpp
3 
4 
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
HORIZONTAL DEFLECTION SYSTEM
X-Y OPERATION
CURSOR READOUT FUNCTION
AUTO MEASUREMENT FUNCTION
CRT
USER MANUAL
Horizontal Modes MAIN(A), ALT, DELAY(B) A (main) Sweep Time
50ns~0.5s/DIV, continuously variable
(UNCAL) B (delay) Sweep Time 50ns~50ms/DIV Accuracy ± 3% (± 5% at × 10 MAG) Sweep Magnification × 10 (maximum sweep time 5ns/DIV) Hold Off time Variable Delay Time 1us~5s Delay Jitter 1/20000 or less Alternate Separation Variable
X-axis, Y-axis selectable Sensitivity Accuracy
X-axis: CH1, CH2 2mV~5V/DIV ± 3%
EXT 0.1V/DIV ± 5%
Y-axis: CH1,CH2 2mV~5V/DIV ± 3% X-axis Bandwidth DC~500kHz (-3dB)
Phase Error Cursor Measurement Function
3°or less at DC~50kHz
ΔV,ΔV%,ΔVdB,ΔT,1/ΔT,ΔT%
Θ
Cursor Resolution 1/100 DIV Effective Cursor Range Vertical: ± 3 DIV; horizontal: ± 4 DIV Panel setting
Vertical: V/DIV (CH1,CH2),UNCAL, ADD,
INV, P10, AC/DC/GND.
Horizontal: S/DIV (MTB, DTB), UNCAL
x 10MAG, Delay time, Hold-off.
Trigger: Source, Coupling, Slope, Level ,
TV-V/TV-H.
Others: X-Y, LOCK, SAVE/RECALL
MEM 0-9. Parameter Function
FREQ, PERIOD,±WIDTH, ±DUTY,
(+ or – polarity selected by trigger slop.) Display Digits Max. 6 digits, decimal. Frequency Range 50Hz~100MHz. Accuracy
1kHz~100MHz: ±0.01%
50Hz~1kHz : ±0.05% Measuring Sensitivity >2div(measuring source selected from CH1
and CH2 as synchronous signal source.)
Type
6-inch rectangular type with internal graticule
0%, 10%, 90% and 100% markers.
8 x 10 DIV (1 DIV = 1 cm) Phosphor P31 Accelerating Potential 16kV approx. Illumination Continuous adjustable
USER MANUAL
External intensity modulation
Z-AXIS INPUT
TRIGGER SIGNAL OUTPUT
CALIBRATOR OUTPUT
SPECIAL FUNCTION
LINE POWER REQUIREMENT
.
OPERATING ENVIRONMENT
STORAGE TEMPERATURE & HUMIDITY MECHANICAL SPECIFICATION
ACCESSORIES
Coupling DC Voltage 5V or more Maximum Input Voltage 30V (DC+AC peak) at 1kHz or less Bandwidth DC~5MHz Voltage
25mV/DIV approx. in 50Ω termination
Frequency Response DC~10MHz Output Impedance
50Ω approx.
Waveform 1kHz ± 5%, square wave Voltage 2Vpp ± 2% Impedance
2kΩ approx.
TIME/DIV Auto Range Provided Panel Setting Save & Recall 10 sets Panel Setups Lock Provided Voltage
AC 100V, 120V, 230V ± 10% selectable
Frequency 50Hz or 60Hz Power Consumption Approx. 90VA, 70W(max.) Indoor use Altitude up to 2000 m Ambient temperature : To satisfy specifications : 10° to 35 ( 50° to 95°F ) Maximum operating ranges: 0° to 40( 32 °to 104°F ) Relative humidity:85% RH(max.) non condensing Installation Category II Pollution degree 2
-10° to 70, 70%RH(maximum)
Dimensions 310 W × 150 H × 455 D (mm) Weight Approx. 9kg (19.8 lbs)
Power cord….............……….. 1
Instruction manual…………… 1 Probe (×1/×10)…………..… 2
5 
6 
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
3.PRECAUTIONS BEFORE OPERATION
3-1.Unpacking the Oscilloscope
The product has been fully inspected and tested before shipping from the factory. Upon receiving the instrument, please unpack and inspect it to check if there is any damages caused during transportation. If any sign of damage is found, notify the bearer and/or the dealer immediately.
3-2.Checking the Line Voltage
The oscilloscope can be applied any kind of line voltage shown in the table below. Before connecting the power plug to an AC line outlet, make sure the voltage selector of the rear panel is set to the correct position corresponding to the line voltage. It might be damaged the instrument if connected to the wrong AC line voltage.
WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground.
When line voltages are changed, replace the required fuses shown as below:
Line voltage Range
100V 120V
108-132V
90-110V
Fuse
T 1A250V 230V 207-250V T 0.4A250V
Line voltage Range
WARNING. To avoid personal injury, disconnect the power
cord before removing the fuse holder.
Fuse
USER MANUAL
3-3.Environment
The normal ambient temperature range of this instrument is from 0° to 40°C (32° to 104°F). To operate the instrument over this specific temperature range may cause damage to the circuits. Do not use the instrument in a place where strong magnetic or electric field exists as it may disturb the measurement.
3-4.Equipment Installation, and Operation
Ensure there is proper ventilation for the vents in the oscilloscope case. If the equipment is used not according to the specification, the protection provided by the equipment may be impaired.
3-5.CRT Intensity
To prevent permanent damage to the CRT phosphor, do not make the CRT trace brighten excessively or leave the spot stay for an unreasonably long time.
3-6.Withstanding Voltages of Input Terminals
The withstanding voltages of the instrument input terminals and probe Input terminals are shown in the following table. Do not apply voltages higher than these limits.
Input terminal Maximum input voltage
CH1, CH2, inputs 400V (DC + AC peak)
EXT TRIG input 400V (DC + AC peak)
Probe inputs 600V (DC + AC peak)
Z AXIS input 30V (DC + AC peak)
CAUTION. To avoid damaging the instrument, do not apply input Voltage of the frequency over 1 kHz to the instrument.
7 
8 
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
4. PANEL INTRODUCTION
After the instrument is switched on, all the important settings are displayed in the readout. The LED’s located on the front panel assist operation and indicate additional information. Incorrect operation and the electrical end positions of control knobs are indicated by a warning beep.
Except the Power pushbutton (POWER), the Focus control (FOCUS), the Scale Illumination control (ILLUM) and the Trace Rotation control, all other controls are electronically selected, and their functions and settings can therefore be stored.
The front panel is subdivided into six sections:
Display controls Vertical controls Horizontal controls Trigger controls Measurement and SAVE/RECALL controls Input connectors
USER MANUAL
Front panel of GOS-6103C
9 
10
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
4-1.Front Panel
Display controls
The display controls adjust the on-screen appearance of the waveform and provide a probe compensation signal source.
(1).POWER – Pushbutton and symbols for ON(1) and OFF(0)
When switch on the oscilloscope to have all LEDs lighted and the software version will be displayed on the screen. After the Internal test is completed successfully, the normal operation mode is present. Then the last settings become activated and the LED indicates the ON condition.
(2).INTEN – TRACE/READOUT & READOUT ON/OFF– Control
knob with associated pushbutton and readout display. The control knob is used for adjusting the traces and readout intensity. Turning the knob clockwise to increase the intensity and turning it counterclockwise to decrease the intensity. The TRACE/READOUT pushbutton is for selecting the intensity function and indicates the letter “TRACE INTEN” or “READOUT INTEN” in the readout. Press the pushbutton briefly for the following sequences: “TRACE INTEN” — “READOUT INTEN” — “TRACE INTEN”
READOUT ON/OFF
Pressing and holding the TRACE/READOUT pushbutton to switch on
or off the readout.
USER MANUAL
(3).TRACE ROTATION
The TRACE ROTATION is for aligning the horizontal trace in parallel with graticule lines. This potentiometer can be adjusted with a small screwdriver.
(4).FOCUS
The control knob effects both the trace and the readout sharply.
(5).ILLUM
The knob controls the graticule illumination brightness.
(6).CAL
The terminal provides a reference signal of 2Vp-p at 1kHz for probe adjustment.
Vertical controls
The vertical controls select the displayed signals and control the
amplitude characteristics.
11
12
GOS-6103C OSCILLOSCOPE
GOS-6103C OSCILLOSCOPE
USER MANUAL
(7).20MHz BWL – Pushbutton with indicator LED.
Briefly pressing the pushbutton, the bandwidth is reduced to approx. 20MHz, and the measurement is made by eliminating undesired high frequency signal from the waveform. Also the high frequency component over 20MHz is eliminated from the trigger signal.
(8).CURSOR POS - V1/2 — Pushbutton with double function and
associated indicator LED.
The function of Cursor Position or CH1/CH2 Position can be selected only after the pushbutton of Cursor Function is pressed to appear enabling their cursor measurement. Press the pushbutton once briefly to have the related LED lighted, the CH1/CH2 POSITION control knob is then operated as CURSOR 1/CURSOR 2 POSITION control.
V1/2
The function is required and available only in DUAL mode in combination with V (Voltage) measurement. Pressing and holding the pushbutton, then switch between CH1 and CH2, the measured result will be displayed by the readout with “V1…” or “△V2… ” providing the defection coefficient is calibrated. The settings of the cursors must be related to the signal of the selected channel.
(9).CH1 POSITION – C1 — Control knob with double function.
The vertical trace position of channel 1 can be set with the control knob, which is also operated as CURSOR 1 position control in cursor measurement mode.
(10)CH2 POSITION – C2 — Control knob has several functions.
The vertical trace position of channel 2 can be set with the cont rol knob, which is also operated as CURSOR 2 position control in cursor measurement mode. In alternate time base mode, this control knob can be used to separate the DELAY time base trace from the MIAN time base trace. Please note TRACE SEP (11).
(11)TRACE SEP
The instrument contains a trace separate function which is required in the alternate time base mode to separate the DELAY time base trace (s) from the MAIN time base in vertical direction. Consequently this function is only available in alternate time base mode. Press the pushbutton once to have the related LED lighted, the CH1 POSITION control knob is then operated as vertical position control for the trace(s) of the DELAY time base.
(12)ALT/CHOP/ADD-INV
The pushbutton has sever al function s, which are requ ired and avail able only when both channels are active. ALT– Displays in the readout, indicates alternate channel switching. After each time base sweeps the instrument internally, switches over from channel 1 and channel 2 and vice versa. CHOP– Indicates chopper mode. The channel switching occurs constantly between channel 1 and channel 2 during each sweep. ADD– Displays in the readout, indicates additional mode. Whether the algebraic sum (addition) or the difference (subtraction) of both input signals is displayed, depends on the phase relationship and the INV setting. As a result, both signals are displayed as one signal. For correct measurements, the deflection coefficients for both channels must be equal. INV—Pressing and holding the pushbutton to set the channel 2 invert function on or off. The invert on condition is indicated with a horizontal bar above “CH2” in the readout. The invert function causes the signal display of channel 2 to be inverted by 180
USER MANUAL
o
.
13
14