These terms may appear in this manual or on the product:
WARNING. Warning statements identify condition or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or
other property.
The following symbols may appear in this manual or on the product:
DANGER ATTENTION Protective Earth(ground)
High Voltage refer to Manual Conductor Terminal
Terminal
i
ii
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
FOR UNITED KINGDOM ONLY
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with
the following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the
colours marking identified in your plug/appliance, proceed as follows:
USER MANUAL
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, cable of
0.75mm
2
should be protected by a 3A or 5A fuse. Larger conductors
would normally require 13A types, depending on the connection
method used.
Any moulded mains connector that requires removal /replacement
must be destroyed by removal of any fuse & fuse carrier and disposed
of immediately, as a plug with bared wires is hazardous if a engaged
in live socket. Any re-wiring must be carried out in accordance with
the information detailed on this label.
The wire which is coloured Green & Yellow must be connected to the
Earth terminal marked with the letter E or by the earth symbol
or coloured Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal
marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or
contact the supplier.
iii
iv
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declares that the below mentioned product
GOS-6112
is herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Law of Member States relating to
Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC) and Low
Voltage Equipment Directive (73/23/EEC).
For the evaluation regarding the Electromagnetic Compatibility and Low Voltage
Equipment Directive, the following standards were applied:
EN 61326-1:Electrical equipment for measurement, control and laboratory
Conducted and Radiated Emissions
EN 55011 class B: 1991
EN 55022 class B: 1994
Current Harmonic
EN 61000-3-2: 1995
Voltage Fluctuation
EN 61000-3-3: 1995
Surge Immunity
Conducted Susceptibility
Power Frequency Magnetic fie ld
Voltage Dips/ Interrupts
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Radiated Immunity
ENV 50140: 1993
Electrical Fast Transients
EN 61000-4-4: 1995
EN 61000-4-5: 1995
EN 61000-4-6: 1996
EN 61000-4-8: 1993
EN 61000-4-11: 1994
USER MANUAL
1.PRODUCT INTRODUCTION
1-1. Description
The GOS-6112 is a 100MHz, two-channel, dual-sweep, portable oscilloscope
for general purpose use. A microprocessor-based operating system controls
most of the functions of the instrument, including cursor readout and
digitized panel setting. On-sc reen alphanumeric readout and cursor fu nction
for voltage, time, frequency and phase measurement provide extraordinary
operational convenience.
The vertical deflection system has two input channels. Each channel has 11
basic deflection factors from 2mV to 5V per division. The horizontal
deflection system provides single, dual or delayed sweeps from 0.5s to 50ns
per division (delayed sweep, 50ms to 50ns per division). The trigger system
provides stable triggering over the full bandwidth of the vertical deflection
system.
v
1
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
1-2.Features
Additionally, the oscilloscope offers several other features:
1) High intensity and internal graticule CRT
The oscilloscope employs a high intensity 6-inch retangular type
cathode-ray tube with red internal graticule. It display s clear readable
traces even at high sweep speeds. Internal graticule lines eliminate
parallax-viewing error between the trace and the grat icule line .
2) Temperature compensation
The oscilloscope uses a temperature compensation circuit to reduce the
drift of base line and DC balance.
3) 20MHz bandwidth limit
When it is hard to observe or trigger a signal because a high-frequency
component is superimposed on the signal, use the 20MHz BWL
function to reduce the bandwidth of the vertical deflection system and
trigger system to 20MHz.
4) TV triggering
Exclusive TV sync separator circuit technology provides stable TV
signal measurements on fields, frames and lines.
5) Z-axis intensity modulation
For applying a blanking signal from an external source. The trace
displayed on the screen may be intensity-modulated where pulse signal
or time-scale marks are required.
USER MANUAL
6) Trigger signal output
The signal selected by the TRIGGER SOURCE is available. This
output may be used to connect to a frequency counter or other
instrument.
7) Panel setups lock
To avoid unintentional touch of the setting, the feature is extremely
useful for long term and repetitive measurements that used to be
performed under the same test condition of the oscilloscope setting.
8) LED indicator and buzzer alarm
The LED’s located in the front panel assist operation and indicated
additional information. Incorrect operation and the electrical end
position of control knobs are indicated by a warning beep.
9) SMD manufacturing technology
The instrument is built by using the most advanced SMD technology
so as to reduce the number of internal wiring and shorten the foil rout e
on the pc board. This will also greatly increase the high frequency
performance and the reliability of the product.
2
3
GOS-6112 OSCILLOSCOPE
USER MANUAL
2.TECHNICAL SPECIFICATIONS
Sensitivity 2mV~5V/DIV,11 step in 1-2-5 sequence
VERTICAL
DEFLECTION
SYSTEM
TRIGGER
SYSTEM
Sensitivity Accuracy
Vernier Vertical Sensitivity
Frequency Bandwidth(-3dB) DC ~ 100MHz (2mV/DIV:DC ~ 20MHz)
Rise Time 3.5ns (2mV/DIV:17.5ns)
Signal Delay Leading edge can be monitored
Maximum Input Voltage 400V (DC+AC) at 1kHz or less
Input Coupling AC, DC, GND
Input Impedance
Vertical Modes
CHOP Repetition Frequency Approx. 250kHz.
Polarity (INV) CH2 only
Bandwidth Limit ed 20MHz
Common-mode
Rejection Ration
Dynamic Range 8 div at 60MHz, 5 div at 100MHz.
Trigger Modes AUTO, NORM, TV
Trigger Source CH1, CH2, LINE, EXT
Trigger Coupling AC, DC, HFR, LFR
Trigger Slope +/- polarity or TV sync polarity
Trigger Sensitivity
Trigger Level Range
TV Sync TV-V,TV-H
Max. External Input Voltage 400V (DC + AC peak) at 1k Hz
External Input Impedance
±3% (5 DIV at the center display )
Continuously variable to 1/2.5 or less than
panel-indicated value
1MΩ±2% // 25pF approx.
CH1,CH2,DUAL(CHOP/ALT),ADD(DIFF mode can
be established when the CH2 is in the INV mode)
50:1 or better at 50kHz
ModeFrequency INT EXT
AUTO
TV Sync signal 1DIV 200mVpp
INT : ± 4 DIV or more
EXT : ± 0.4 V or more
1MΩ± 5% ∕∕ 25pF approx.
10Hz~20MHz 0.35DIV50mVpp
20MHz~100MHz1.5DIV 150mVpp
DC~20MHz 0.35DIV50mVppNORM
20MHz~100MHz1.5DIV 150mVpp
4
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
HORIZONTAL
DEFLECTION
SYSTEM
X-Y
OPERATION
CURSOR
READOUT
FUNCTION
CRT
Z-AXIS
INPUT
USER MANUAL
Horizontal Modes MAIN(A), ALT, DELAY(B)
A (main) Sweep Time
B (delay) Sweep Time 50nS~50mS/DIV
Accuracy ± 3% (± 5% at × 10 MAG)
Sweep Magnification × 10 (maximum sweep time 5nS/DIV)
Hold Off time Variable
Delay Time 1uS~5S
Delay Jitter 1/20000 or less
Alternate Separation Variable
Sensitivity Accuracy
X-axis Bandwidth DC~500kHz (-3dB)
Phase Error
Cursor Measurement Function
Cursor Resolution 1/100 DIV
Effective Cursor Range Vertical: ± 3 DIV; horizontal: ± 4 DIV
Panel setting
Type
Phosphor P31
Accelerating Potential 12kV approx.
External intensity modulation
Coupling DC
Voltage 5V or more
Maximum Input Voltage 30V (DC+AC peak) at 1kHz or less
Bandwidth DC~5MHz
Horizontal: S/DIV (MTB, DTB), UNCAL
x 10MAG, Delay time, Hold-off.
Trigger: Source, Coupling, Slope, Level ,
TV-V/TV-H.
Others: X-Y, LOCK.
6-inch rectangular type with internal graticule
0%, 10%, 90% and 100% markers.
8 x 10 DIV (1 DIV = 1 cm)
Θ
.
TRIGGER
SIGNAL
OUTPUT
CALIBRATOR
OUTPUT
SPECIAL
FUNCTION
LINE POWER
REQUIREMENT
OPERATING
ENVIRONMENT
STORAGE
TEMPERATURE
& HUMIDITY
MECHANICAL
SPECIFICATION
ACCESSORIES
USER MANUAL
Voltage
Frequency Response DC~10MHz
Output Impedance
Waveform 1kHz ± 5%, square wave
Voltage 2Vpp ± 2%
Impedance
Panel Setups Lock Provided
Voltage
Frequency 50Hz or 60Hz
Power Consumption Approx. 90VA, 70W(max.)
Indoor use
Altitude up to 2000 m
Ambient temperature :
To satisfy specifications : 10° to 35℃ ( 50° to 95°F )
Maximum operating ranges: 0° to 40℃( 32 °to 104°F )
Relative humidity:85% RH(max.) non condensing
Installation Category II
Pollution degree 2
-10° to 70℃, 70%RH(maximum)
Dimensions 310 W × 150 H × 455 D (mm)
Weight Approx. 9kgs (19.8 lbs)
Power cord….............……….. 1
Instruction manual…………… 1
Probe (×1/×10)…………..… 2
25mV/DIV approx. in 50Ω termination
50Ω approx.
2kΩ approx.
AC 100V, 120V, 230V ± 10% selectable
5
6
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
3.PRECAUTIONS BEFORE OPERATION
3-1.Unpacking the Oscilloscope
The product has been fully inspected and tested before shipping from the
factory. Upon receiving the instrument, please unpack and inspect it to
check if there is any damages caused during transportation. If any sign of
damage is found, notify the bearer and/or the dealer immediately.
3-2.Checking the Line Voltage
The oscilloscope can be applied any kind of line voltage shown in the
table below. Before connecting the power plug to an AC line outlet, make
sure the voltage selector of the rear panel is set to the correct position
corresponding to the line voltage. It might be damaged the instrument if
connected to the wrong AC line voltage.
WARNING. To avoid electrical shock the power cord
protective grounding conductor must be connected to
ground.
When line voltages are changed, replace the required fuses shown as below:
Line voltage Range
100V
120V
108-132V
WARNING. To avoid personal injury, disconnect the power
Fuse Line voltageRange
90-110V
cord before removing the fuse holder.
T 1A 250V 230V 207-250V T 0.4A 250V
Fuse
USER MANUAL
3-3.Environment
The normal ambient temperature range of this instrument is from 0° to
40°C (32° to 104°F). To operate the instrument over this specific
temperature range may cause damage to the circuits.
Do not use the instrument in a place where strong magnetic or electric
field exists as it may disturb the measurement.
3-4.Equipment Installation, and Operation
Ensure there is proper ventilation for the vents in the oscilloscope case. If
this equipment is used not according to the specification, the protection
provided by the equipment may be impaired.
3-5.CRT Intensity
To prevent permanent damage to the CRT phosphor, do not make the
CRT trace brighten excessively or leave the spot stay for an unreasonably
long time.
3-6.Withstanding Voltages of Input Terminals
The withstanding voltages of the instrument input terminals and probe
Input terminals are shown in the following table. Do not apply voltages
higher than these limits.
Input terminalMaximum input voltage
CH1, CH2, inputs 400V (DC + AC peak)
EXT TRIG input 400V (DC + AC peak)
Probe inputs 600V (DC + AC peak)
Z AXIS input 30V (DC + AC peak)
CAUTION. To avoid instrument damage, do not apply exceeding
maximum input voltage of 1 kHz the frequency to the instrument.
7
8
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
4. PANEL INTRODUCTION
After the instrument is switched on, all the important settings are displayed
in the readout. The LED’s located on the front panel assist operation and
indicate additional information. Incorrect operation and the electrical end
positions of control knobs are indicated by a warning beep.
Except the Power pushbutton (POWER), the Focus control (FOCUS), and
the Trace Rotation control, all other controls are electronically selected,
and their functions and sett ings can t herefor e be store d.
The front panel is subdivided into six sections:
z Display controls
z Vertical controls
z Horizontal controls
z Trigger controls
z Measurement control
z Input connectors
USER MANUAL
Front panel of GOS-6112
9
10
GOS-6112 OSCILLOSCOPE
GOS-6112 OSCILLOSCOPE
USER MANUAL
4-1.Front Panel
Display controls
The display controls adjust the on-screen appearance of the waveform and
provide a probe compensation signal source.
FOCUS
CAL
2Vp-p
1 KHz
5
ROTATION
43
TRACE INTEN
OFF
(1).POWER – Pushbutton and symbols for ON(1) and OFF(0).
When switch on the oscilloscope to have all LEDs lighted and the
software version will be displayed on the screen. After the Internal test
is completed successfully, the normal operation mode is present. Then
the last settings become activated and the LED indicates the ON
condition.
(2).INTEN – TRACE/READOUT – Control knob with associated
pushbutton and readout display.
The control knob is used for adjusting the traces and readout intensity.
Turning the knob clockwise to increase the intensity and turning it
counterclockwise to decrease the intensity.
The TRACE/READOUT pushbutton is for selecting the intensity
function and indicates the letter “TRACE INTEN” or “READOUT
INTEN” in the readout. Press the pushbutton briefly for the following
sequences:
“TRACE INTEN” — “READOUT INTEO” — “TRACE INTEN”
TRACE/
POWER
READOUT
21
USER MANUAL
(3).TRACE ROTATION
The TRACE ROTATION is for aligning the horizontal trace in
parallel with graticule lines. This potentiometer can be adjusted with a
small screwdriver.
(4).FOCUS
The control knob effects both the trace and the readout sharply.
(5).CAL
The terminal provides a reference signal of 2Vp-p at 1kHz for probe adjustment.
Vertical controls
The vertical controls select the displayed signals and control the
amplitude characteristics.
910
VERTICAL
13
15
POSITION
C1
8
20MHz
7
AC/DC
CURSOR
BWLPOS
V1/2
VOLTS/DIV
2mV
5V
CH1
VAR
GND
Px10
17191820
C2
ALT/
CHOP/
ADD
5V
AC/DC
POSITION
INV
VOLTS/DIV
VAR
Px10
TRACE
2mV
CH2
GND
SEP
11
12
14
16
11
12
Loading...
+ 21 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.