GW INSTEK PART NO. 82HT-11700M01
This assembly manual contains proprietary information, which is protected by copyrights. All rights are
reserved. No part of this manual may be photocopied, reproduced or translated to another language
without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to
improve products and reserves the rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
ISO-9001 CERTIFIED MANUFACTURER
GHT-117 HV Adapter Box
The GPT-9000 series Electrical Safety Testers operate at up
to 5kVAC and 6kVDC. To prevent electrical shock, the
grounding wire of the adapter box (green/yellow wire)
must be connected to the ground connector of the GPT-9000
rear panel. Please also make sure that the GPT-9000 is
properly grounded via its AC line input. Please refer to the
GPT-9000 User Manual for more information regarding
safety and the earth ground connection.
POWER
START STOP
REMOTE
MANU/AUTO EDIT/SAVE UTILITY
ESC PAGE
GPT-9803
Tester
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
AC / DC Withstanding Voltage /
Insulation Resistance
RETURN
PASS FAIL READY TEST
DUT
GPT-9000
Rear panel
GND
GHT-117
HV
Return
Ground
GHT-117
The GHT-117 High Voltage Adapter Box is an optional accessory of the
GPT-9000 series Electrical Safety Tester aimed at providing a convenient way to
connect the DUT’s AC line socket to the GPT-9000 tester for the AC withstanding,
DC withstanding and Insulation Resistance tests.
Consideration
Connection