Technical content subject to change without notice.
[no content intended for this page]
ii
Preface
Information Paragraphs
Note:These paragraphs provide information that provides a deeper understanding of the situation, but is not
essential to the proper completion of the instructions.
IMPORTANT:These paragraphs provide information that emphasizes instructions that are essential to proper setup of
the equipment. Failure to follow these instructions carefully may cause unreliable performance.
CAUTION!This symbol indicates a risk of potential minor personal injury and/or severe
damage to the equipment, unless these instructions are followed carefully.
WARNING!This symbol indicates a risk of potential serious personal injury, unless these
instructions are followed carefully.
Safety Issues
WARNING!It is the responsibility of the user to make sure all local, county, state and national
codes, regulations, rules and laws related to safety and safe operating conditions are met for
each installation.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working Area
WARNING!Auxiliary equipment may have both manual and automatic modes of operation. As
equipment can move suddenly and without warning, do not enter the work cell of this equipment
during automatic operation, and do not enter the work envelope of this equipment during
manual operation. If you do, serious injury can result..
WARNING!Make sure that power to the auxiliary equipment is turned OFF and locked out
before you perform maintenance procedures on the equipment.
XMTC User’s Manualiii
Preface
Auxiliary Equipment (cont.)
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment.
Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control Solutions is an active participant in Europe’s Waste Electrical and Electronic Equipment
(WEEE) take-back initiative, directive 2002/96/EC.
The equipment that you bought has required the extraction and use of natural resources for its production. It may
contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural
resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional
waste administration.
Visit http://www.ge-mcs.com/en/about-us/environmental-health-and-safety/1741-weee-req.html
take-back instructions and more information about this initiative.
This chapter introduces you to the features and capabilities of the GE XMTC Thermal Conductivity Transmitter. The
following topics are discussed:
•Basic features of the XMTC thermal conductivity transmitter
•Theory of operation
•A system description of the XMTC, available options, and sample systems Information on optional
components is also provided, including a 24 VDC power supply, extra cable, and the TMO2D-TC Display.
•A brief discussion of typical XMTC applications
XMTC technical specifications can be found in Chapter 5, Specifications. Ordering information can be found in
Appendix A, Supplemental Information.
1.2Basic Features
The XMTC is a transmitter that measures the thermal conductivity of a binary (or pseudo-binary) gas mixture
•
containing hydrogen, carbon dioxide, methane or helium, and provides a 4-20 mA signal proportional to the
concentration of one of the gases in the mixture. It offers several unique design features:
•Ultra-stable thermistors and a temperature-controlled measuring cell (55°C/131°F standard, 70°C/158°F
optional) provide excellent zero and span stability, as well as tolerance of ambient temperature variations.
•The measuring cell design makes it highly resistant to contamination and flow vibrations. Since it has no
moving parts, the transmitter can handle the shock and vibration found in many industrial applications.
•A 2-port version for measurement of zero-based gas mixtures using a sealed reference gas (air or nitrogen) and
a 4-port version for measurement of zero-suppressed gas mixtures (and some other special calibrations) using a
flowing reference gas are available.
•The XMTC modular construction means that the unit can be field-calibrated quickly and easily. If desired, the
plug-in measuring cell can be replaced with a pre-calibrated spare in minutes.
•The XMTC transmitter, with weatherproof or explosion-proof packaging, is designed to be installed as close as
possible to the process sample point. It can be located up to 4000 ft (1200 m) from a display or recorder, using
inexpensive unshielded cable.
XMTC User’s Manual1
Chapter 1. Features and Capabilities
01.02.03.04.05.06.07.0
SO
2
CO
2
Air/N
2
CH
4
Ne
He
H
2
C
4
H
6
1.3Theory of Operation
The XMTC measures the concentration of a gas in a binary gas mixture by measuring the thermal conductivity of the
sample gas and comparing it to the thermal conductivity of a selected reference gas.
Two ultra-stable, glass-coated thermistors are used: one in contact with the sample gas, and the other in contact with a
selected reference gas. The thermistors are mounted so that they are in close proximity to the stainless steel walls of the
sample chamber. The entire sensor is heated to 55°C/131°F, (or 70°C/158°F) and the thermistors are heated above the
sensor temperature using a constant current source. The thermistors lose heat to the walls of the sample chamber at a
rate that is proportional to the thermal conductivity of the gas surrounding them. Thus, each thermistor will reach a
different equilibrium temperature. The temperature difference between the two thermistors is detected in an electrical
bridge circuit. It is then amplified and converted to a 4-20 mA output proportional to the concentration of one of the
constituents of the binary gas mixture. For example:
•To measure 0 to 25% H
calibration, the zero gas would be 100% N
•To measure 90-100% H
the zero gas would be 90% H
Note:The XMTC has polarity adjustment jumpers which permit the measurement of gases (such as CO
relative thermal conductivity less than air/nitrogen.
Appendix A, Supplemental Information, contains a table of Relative Thermal Conductivity of Common Gases. Figure 1
below shows some of these values graphically.
in N2, the reference gas would be air (2-port version, sealed reference gas), and for
2
(i.e. 0% H2) and the span gas would be 25% H2 in N2.
2
in N2, the reference gas would be 100% H2 (4-port version, flowing reference gas),
2
in N2, and the span gas would be 100% H2 (the same as the reference gas).
2
) that have a
2
Figure 1: Relative Thermal Conductivity of Some Common Gases
2XMTC User’s Manual
Chapter 1. Features and Capabilities
1.4System Description
The basic XMTC measurement system consists of an XMTC Transmitter mounted in a sample system. The sample
system is mandatory, and can either be provided by GE or constructed according to GE recommendations. The XMTC
is supplied with a standard 10 ft (3 m), 4-wire cable for power and output connections, with lengths up to 4000 ft
(1200 m) available. Optionally available from GE are a 24-VDC power supply to power the XMTC, a remote display
with programming and control capabilities, and several analyzers which can be interfaced with the XMTC.
1.4.1Packaging and Temperature Rating
The XMTC transmitter is self-contained, consisting of the thermal conductivity sensor and associated electronics. It
requires 24 VDC power (1.2 A maximum at power-up), and provides a 4-20 mA output signal proportional to the
concentration of one of the gases in the binary sample gas mixture.
The XMTC is designed to be installed in a sample system as close as possible to the process sample point. Thus, it is
available in two environmental packages:
•Weatherproof
•Explosion-proof (with the addition of flame arrestors to the sample/reference gas inlet and outlet)
Each environmental package is available in a standard 2-port (sealed reference gas) version, or an optional 4-port
(flowing reference gas) version.
The XMTC is supplied with a standard measurement cell operating temperature of 55°C (131°F). An optional 70°C
(158°F) cell operating temperature is available.
Note:The 70°C (158°F) operating temperature should be selected only for high temperature applications, because it
results in reduced sensitivity.
XMTC User’s Manual3
Chapter 1. Features and Capabilities
1.4.22-Port (Sealed Reference Gas) Version
This standard configuration (see Figure 2 below) is used for zero-based ranges with air or nitrogen at atmospheric
pressure as the balance or background gas. It utilizes air with desiccant in a factory sealed chamber as the reference gas.
The following standard ranges and gases are available:
Ranges:0-2%Gases:H
0-5%CO
0-10%SO
0-25%He in N
0-50%Argon in N
0-100%
50-100%
80-100%
90-100%
95-100%
98-100%
in N2 or air
2
in N2 or air (min. range 0-5% CO2)
2
in air (min. range 0-2% SO2)
2
or air
2
or air
2
Figure 2: 2-Port (Sealed Reference Gas) XMTC
4XMTC User’s Manual
Chapter 1. Features and Capabilities
1.4.34-Port (Flowing Reference Gas) Version
This optional configuration shown in Figure 3 below is used for zero-suppressed ranges and some other special
applications. Typically, a flowing reference gas of 100% H
are available:
or CO2 is used. The following standard ranges and gases
2
Ranges:90 to 100%Gases:H
in N
2
2
80 to 100%CO2 in N2 or air
He in N
or air
2
Note:For factory calibration pricing on the standard ranges and gases, or for pricing on other zero-suppressed
ranges and gases, please consult the factory.
Figure 3: 4-Port (Flowing Reference Gas) XMTC
XMTC User’s Manual5
Chapter 1. Features and Capabilities
1.4.4Sample System
Use of a sample system is mandatory with the XMTC. The design of the sample system depends on the conditions of
the sample gas and the requirements of the application. In general, a sample system must deliver a clean, representative
sample to the XMTC at a temperature, pressure and flow rate that are within acceptable limits. Standard XMTC sample
conditions are as follows:
•Temperature less than 50
o
C (122oF) for 55oC cell operating temperature
•Atmospheric pressure
•Flow rate 0.5 SCFH (250 cc/min)
GE offers sample systems for a wide variety of applications. Two standard sample systems for the XMTC are shown in
Chapter 2, Installation. For assistance in designing your own sample system, please consult the factory.
1.4.5Extra Cable (optional)
GE provides a 10 ft (3 m) length of 4-wire, color-coded cable with each XMTC to connect the power and outputs.
Optional cables are available in lengths up to 4000 ft (1200 m). If you are using your own cable, refer to Table 1 on page 13 for recommendations.
1.4.6Power Supply (optional)
The XMTC requires 24 VDC at a maximum start-up current of 1.2 A. The GE PS5R-C24 power supply converts
100/120/220/240 VAC to the required 24 VDC for the XMTC.
1.4.7TMO2D-TC Display (optional)
The GE TMO2D-TC Display provides a two-line x 24 character back-lit LCD. It also features display and option
programming, recorder outputs, alarm relays, and relays for driving sample system solenoids for automatic zero and
span calibration of the XMTC. For information on the TMO2D-TC, please contact GE.
1.4.8XDP Display (Optional)
The GE XDP Explosion-proof Display provides an integral, voltage-stabilized 24 VDC power supply, a 3-digit display
with adjustable 4-20 mA input range, two SPDT alarm relays rated for 1 A/250 VAC, and a 4-20 mA output that is
isolated from the input and adjustable to a second independent range if required. For information on the XDP, please
contact GE.
6XMTC User’s Manual
Chapter 1. Features and Capabilities
1.5Typical Applications
The XMTC can be used in a wide variety of industrial applications where it is necessary to measure the concentration
of one component of a binary gas mixture. It can also be used in pseudo-binary gas mixtures where the ratio of
concentrations of the background gas components remains constant, and in gaseous mixtures where the thermal
conductivity of the gas of interest is significantly different from that of the background gas. Some typical industries and
applications include:
•Metals Industry -
H
in heat treat furnace atmospheres
2
•Electric Power Industry -
H
in generator cooling systems
2
•Gas Production Industry -
Purity monitoring of argon, hydrogen, nitrogen, helium
•Chemical Industry -
in ammonia synthesis gas
H
2
in methanol synthesis gas
H
2
in chlorine plants
H
2
•Food Industry -
in fermentation processes
CO
2
Ethylene Oxide (ETO) sterilization
•Steel Industry -
H
in blast furnace top gas
2
•Petroleum Industry -
H
in hydrocarbon streams
2
Two very common applications are:
•H
•H
For more details on these applications, refer to Appendix B, Applications. For details on applications not shown in
Appendix B, or if you wish to discuss your own application, please contact GE.
in N2 in heat treat furnace atmospheres: zero-based 0-25% H2, 2-Port (sealed reference gas, air)
This chapter describes how to install the XMTC transmitter and its sample system. It also contains information on
connecting the XMTC to optional system components. The following topics are discussed:
•Mounting the XMTC transmitter
•Installing a GE sample system
•Wiring the XMTC transmitter
•Connecting the XMTC transmitter to optional components
2.2Mounting the XMTC Transmitter
This section applies only if you are mounting the XMTC transmitter in a sample system that has not been supplied by
GE.
Your sample system should deliver a clean, representative sample to the XMTC at the proper temperature, pressure and
flow rate. This usually means a clean, dry sample (free of solid and liquid particulates) at atmospheric pressure; a
temperature no greater than 50°C (122°F); and a flow rate of approximately 0.5 SCFH (250 cc/min). Since factory
calibration of the sensor is at atmospheric pressure and 0.5 SCFH, higher or lower operating pressure may necessitate
field-calibration adjustment.
A simple sample system for a 2-Port XMTC might have inlet and outlet flow-regulating needle valves, a flow meter,
and a pressure gauge, in addition to the XMTC transmitter.
The XMTC transmitter should be mounted in the sample system so that it is upright and level to within ±15°. Also,
provide at least 9 in. (230 mm) of clearance above the top cover of the transmitter to allow access to the transmitter
printed circuit board (PCB) for calibration and maintenance. For a 2-Port XMTC, connect the sample system Sample
Inlet and Sample Outlet ports to the appropriate XMTC port. For a 4-Port system, also connect the Reference Inlet and
Reference Outlet ports to the appropriate XMTC ports.
Note:Refer to Appendix C, Installation and Wiring Diagrams, for port locations and other information.
WARNING!Be sure your installation conforms to all safety and electrical code requirements.
XMTC User’s Manual9
Chapter 2. Installation
SPAN GAS
INLET
9.00
(229.5)
(MIN)
TRANSMITTER
SAMPLE GAS
INLET
ZERO GAS
INLET
SAMPLE GAS
OUTLET
2.3Mounting the Sample System
You can order a complete sample system from the factory. This includes the XMTC transmitter and all necessary
components and sample tubing mounted on a metal panel. Several standard sample systems are available, and
custom-designed sample systems can be built to your exact specifications.
Mount the sample system as close to the process sample point as possible. Once the sample system is mounted, connect
all inlet and outlet lines via the 1/4” compression fittings on the sample system. The sample line leading from the
process to the sample system should be of 1/4” stainless steel tubing, and should be as short as possible in order to
ensure a representative sample.
Following are descriptions of two standard sample systems:
2.3.1Manual, 2-Port (Sealed Reference Gas) Sample System
Figure 4 below shows a basic sample system for a 2-Port (sealed reference gas) XMTC. This sample system consists of
inlet needle valves for sample, zero, and span gases; a ball valve; a 2-port XMTC; a pressure gauge; and a flowmeter.
All components are mounted on a painted steel plate. Other components could be added for filtration (filter/coalescer),
pressure control (regulator), or flow control (pump).
Figure 4: Basic 2-Port Sample System (732-164)
10XMTC User’s Manual
Chapter 2. Installation
9.00
(229.5)
(MIN)
TRANSMITTER
2.3.2Manual, 4-Port (Flowing Reference Gas) Sample System
Figure 5 below shows a basic sample system for a 4-Port (flowing reference gas) XMTC. This sample system consists
of inlet needle valves for sample, reference, and calibration gases; a 4-port XMTC; two pressure gauges; and two flow
meters. All components are mounted on a painted steel plate. Other components could be added for filtration
(filter/coalescer), pressure control (regulator), or flow control (pump).
Figure 5: Basic 4-Port Sample System (732-028)
2.3.3Sample Systems with Automatic Switching
Also available from GE are sample systems with electrically-actuated, three-way solenoid valves. When used in
conjunction with a TMO2D or XDP Display (with the Auto Cal option), these systems allow automatic switching of
sample, zero, span, and reference gases during operation and calibration. Refer to Appendix B, Applications, for details
on sample systems designed by GE for specific applications.
XMTC User’s Manual11
Chapter 2. Installation
External Ground Screw
2.4Wiring the XMTC Transmitter
CAUTION!Always apply power to the XMTC immediately af ter installation, espe cially when it is
mounted outdoors or in a humid area.
This section describes how to wire the XMTC for 24 VDC power, RS232 communication, 4-20 mA output, and
optional devices.
2.4.1Grounding the Enclosure
WARNING!The XMTC transmitter enclosure must be properly grounded.
Connect the external ground screw on the XMTC enclosure (see Figure 6 below) to a suitable earth ground.
Figure 6: Ground Screw Location
12XMTC User’s Manual
Chapter 2. Installation
2.4.2CE Mark Compliance
WARNING!To meet CE Mark requirements, you must shield and ground all electrical cables as
described in Appendix E.
WARNING!CE Mark compliance is required for all units installed in EU countries.
WARNING!Cable entries of an approved flameproof design are required. These must be
installed according to the manufacturer’s instructions. The choice of cable entry device may limit
the overall installation category achieved.
2.4.3Cable Specifications
Table 1 below shows the transmitter wiring connections for the standard GE XMTC cable, P/N X4(10). This cable can
be used for distances up to 4000 ft (1200 m).
Table 1: GE 4-Wire XMTC Cable
LeadColorAWGTerminal
+24 VDC LineRed22TB1-1
24 VDC ReturnBlack22TB1-2
4-20 mA (+)White22TB1-3
4-20 mA (-)Green22TB1-4
If you are using your own cable to wire the XMTC, refer to Table 2 below for the cable requirements.
Table 3 below shows the connections for the GE standard 3-wire RS232 cable (P/N 704-668), which is available with a
male or female DB-9 or a DB-25 connector. This cable is available in standard lengths of 6 ft (2 m) and 12 ft (4 m).
Table 3: GE 3-Wire RS232 Cable
LeadColor AWGTerminal
RXRed22TB2-1
TXWhite22TB2-2
GNDGreen22TB2-3
Note:See EIA-RS Serial Communications (document 916-054) for detailed RS232 wiring instructions.
2.4.4Wiring the Signal Connections
The XMTC power input, analog output, and RS232 connections are made to terminal blocks TB1 and TB2, which are
accessed by removing the XMTC cover. See Figure 7 below for the location and pin designations for terminal blocks
TB1 and TB2. Also refer to Appendix C, Installation and Wiring Diagrams.
CAUTION!Do not make any connections to unassigned or unused terminals.
Figure 7: XMTC Signal Wiring Connections
14XMTC User’s Manual
Chapter 2. Installation
2.4.4Wiring the Signal Connections (cont.)
Use the following steps to make the proper wiring connections:
WARNING!Cable entries of an approved explosion-proof design are required. These must be
installed according to the manufacturer’s instructions. The choice of cable entry device may limit
the overall installation category achieved.
1.Install the selected cable entry device, such as conduit or seal-off, in accordance with the manufacturer’s
instructions.
Note:If installation of the cable entry device is only partially complete, GE recommends tagging the device to ensur e
the safety of subsequent users.
2.Route the cable into the XMTC.
3.Unplug the TB1 and TB2 connectors by pulling them straight off the PCB, and loosen the screws on the sides
of the connectors.
4.Connect the power leads:
CAUTION!Connecting the +24 VDC line (red) lead to any terminal except TB1-1 will damage
the XMTC.
a.Insert the 4-wire cable +24 VDC line (red) lead into pin TB1-1 and tighten the screw.
b. Insert the 4-wire cable 24 VDC return (black) lead into pin TB1-2 and tighten the screw.
5.Connec t the an alog output leads:
a.Insert the 4-wire cable + 4-20 mA (white) lead into pin TB1-3 and tighten the screw.
b. Insert the 4-wire cable – 4-20 mA (green) lead into pin TB1-4 and tighten the screw.
IMPORTANT: You can use either an RS232 serial port (discussed in Step 6) or an RS485 serial port (discussed in Step 7),
but not both.
6.Connect the RS232 serial port leads:
a.Insert the 3-wire cable RX (red) lead into pin TB2-1 and tighten the screw.
b. Insert the 3-wire cable TX (white) lead into pin TB2-2 and tighten the screw.
c.Insert the 3-wire cable GND (green) lead into pin TB2-3 and tighten the screw.
XMTC User’s Manual15
Chapter 2. Installation
Jumper J7
Jumper J8
RX 1
TX 2
GND 3
1 TD OUT
2
3
4 RD IN
XMTC (TB2)RS485 Converter
Red
White
Green
2.4.4Wiring the Signal Connections (cont.)
7.Connect the RS485 serial port leads.
a.Locate the jumpers J7 and J8 on the main PCB, using Figure 8 below as a guide. Move the jumpers from
the left (RS232) side to the right (RS485) side.
b. Connect the other end of the cable to the RS485 converter, as shown in Figure 8 below.
Figure 8: Jumper Locations and RS485 Connections
8.Carefully plug the TB1 and TB2 connectors back onto the PCB, and reinstall the cover on the XMTC.
9.Connect the other ends of the cables to the 24 VDC power supply, the 4-20 mA input of the display device, and
the serial port of the computer or terminal (see the instruction manuals for those devices for details).
16XMTC User’s Manual
Chapter 2. Installation
Line
AC2
Neut
AC1
PS5R-C24
24 Volt
Power Supply
+24
VDC
GND
GND
GND
Neut
AC1
Line
AC2
AC Input
XMTC Transmitter
Red
Output
Device
Green
White
Black
TB1-1
+Vin
Black
TB1-2
RTN
White
TB1-3
4-20+
Green
TB1-4
4-20-
Red
TB2-1
RX
White
TB2-2
TX
Green
TB2-3
GND
RS232
Terminal
or PC
2.5Connecting to Other Components
This section gives interconnection details for using other GE devices in conjunction with your XMTC.
2.5.1PS5R-C24 Power Supply
The GE 24-Volt power supply converts 100/120/220/240 VAC to 24 VDC for use with the XMTC. Figure 9 below
shows an interconnection diagram for the XMTC and the PS5R-C24 power supply.
Figure 9: Interconnection Diagram
XMTC User’s Manual17
Chapter 2. Installation
2.5.2TMO2D Display
The GE TMO2D Display provides a two-line x 24 character back-lit LCD. It also features display and option
programming, recorder outputs, alarm relays, and optional relays for driving sample system solenoids for automatic
zero and span calibration of the XMTC. See Figure 72 on page 83 for an interconnection diagram for the XMTC and
the TMO2D, and refer to the TMO2D User’s Manual (910-084) for details on its operation.
2.5.3XDP Display
The XDP Explosion-proof Display Package provides an integral, voltage-stabilized 24 VDC power supply, a 3-digit
display with adjustable 4-20 mA input range, two SPDT alarm relays rated for 1 A/250 VAC, and a 4-20 mA output
that is isolated from the input and adjustable to a second independent range if required. The XDP is supplied in a
weatherproof and explosion-proof enclosure that is rated for EEx d IIC T6 and IP66. See Figure 72 on page 83 for
interconnection diagrams for the XMTC and the MIS-1, MIS-2 and MMS-3 analyzers, and refer to the XDP User’s Manual (910-204) for details on its operation.
2.5.4Moisture Series Analyzers
The GE Moisture Image Series 1 (MIS-1) and Moisture Monitor Series 3 (MMS-3) analyzers accept inputs from a
variety of sensors (including the XMTC) and offer new graphical and digital user interfaces, improved performance,
and low range calibration. See Figure 72 on page 83 for interconnection diagrams for the XMTC and the MIS-1 and
MMS-3 analyzers, and refer to the appropriate User’s Manual for details on operating the MIS-1 (910-108) or MMS-3
(910-110) analyzer.
18XMTC User’s Manual
Chapter 3. Operation and Programming
Chapter 3.Operation and Programming
3.1Introduction
This chapter provides information on operating the XMTC transmitter. The following topics are discussed:
•Powering up the XMTC
•Basic sample gas considerations
•Programming the XMTC with GE Instrument Data Manager (IDM™) software
If you have not already done so, please read Chapter 2, Installation, for details on mounting and wiring the XMTC and
the sample system.
3.2Powering Up the XMTC
WARNING!It is the responsibility of the user to ensure that all cable entry devices and covers
are properly installed and secure prior to applying power to the XMTC.
The XMTC does not have a power switch. It begins operating as soon as it is connected to a 24 VDC power source.
Because the XMTC is controlled at a constant 55°C (131°F) operating temperature, allow 30 minutes for the unit to
warm up and reach temperature stability. During this time, you can establish a sample gas flow through the sample
system.
3.3Starting the Sample Gas
Open the necessary valves to establish a sample gas flow of 0.5 SCFH (250 cc/min) at atmospheric pressure. Make sure
that nothing obstructs the flow of sample gas, thereby causing a pressure buildup in the sensing chamber. For proper
operation, the XMTC should be vented to atmosphere.
Note:Unless otherwise specified, the XMTC is factory calibrated at atmospheric pressure and 0.5 SCFH
(250 cc/min) and should therefore be operated at atmospheric pressure. Operating the XMTC at any other
pressure will necessitate a field calibration at that pressure in order to maintain accuracy. See Chapter 4,
Calibration, for more information.
If you are using the 4-Port (flowing reference gas) configuration, open the necessary valves to establish a reference gas
flow of 0.5 SCFH (250 cc/min) at atmospheric pressure.
Note:If desired, you can use a reference gas flow as low as 5 cc/min to conserve gas.
XMTC User’s Manual19
Chapter 3. Operation and Programming
3.4Programming with IDM™
The XMTC is factory-programmed and ready for immediate use. However, if you wish to check or change the
calibration, you can access the XMTC programming from your PC, using GE Instrument Data Manager (IDM)
software. IDM also allows you to upload or download site files, display data, and log and view real-time data and
diagnostic data in numeric, bar chart or line chart formats. For further information on the display and logging functions,
refer to the Instrument Data Manager User’s Manual (910-185).
Note:Be sure to install Instrument Data Manager on your PC before attempting to program the XMTC.
3.5The Edit Functions Menu
To access the XMTC calibration, you must open the Edit Functions menu in the Instrument window. See Chapter 5,
Using the Instrument Menu, in the Instrument Data Manager User’s Manual for more information on the Instrument
Menu. The menu consists of the five commands shown in Figure 10 below. To access a command, select it from the list
of options.
Note:While following the programming instructions, refer to Figure 51 on page 41 and Figure 52 on page 42.
Figure 10: Edit Functions Menu
Three buttons appear on the right side of all menu windows: Previous Item, Next Item/Enter, and Exit Page (see
Figure 11 on page 21):
•Clicking on Previous Item returns you to the previous window (either the command menu or the previous
parameter entered).
•Next Item/Enter confirms the selection or data entered, and either opens the next window or returns you to the
command menu (depending on your position in the program).
•Exit Page returns you to the command menu.
20XMTC User’s Manual
Loading...
+ 96 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.