Technical content subject to change without notice.
[no content intended for this page]
Preface
Information Paragraphs
•Note paragraphs provide information that provides a deeper understanding of the situation, but is not essential to
the proper completion of the instructions.
•Important paragraphs provide information that emphasizes instructions that are essential to proper setup of the
equipment. Failure to follow these instructions carefully may cause unreliable performance.
•Caution! paragraphs provide information that alerts the operator to a hazardous situation that can cause damage to
property or equipment.
•Warning! paragraphs provide information that alerts the operator to a hazardous situation that can cause injury to
personnel. Cautionary information is also included, when applicable.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county, state and national codes,
regulations, rules and laws related to safety and safe operating conditions are met for each
installation.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As equipment
can move suddenly and without warning, do not enter the work cell of this equipment during
automatic operation, and do not enter the work envelope of this equipment during manual
operation. If you do, serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before you
perform maintenance procedures on the equipment.
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment.
Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
DigitalFlow™ XMT868i Programming Manualiii
Preface
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE)
take-back initiative, directive 2002/96/EC.
The equipment that you bought has required the extraction and use of natural resources for its production. It may
contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural
resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional
waste administration.
Visit
http://www.ge-mcs.com/en/about-us/environmental-health-and-safety/1741-weee-req.html for take-back
instructions and more information about this initiative.
The Model XMT868i flow transmitter must be properly installed and programmed, as described in the Startup Guide,
before it can provide accurate flow rate measurements. After completing the installation and initial setup, use this
chapter to program the advanced features of the Model XMT868i’s Keypad Program.
The Keypad Program provides access to the various programmable features of the Model XMT868i. Step-by-step
programming instructions are presented in this chapter. Refer to the appropriate section for a discussion of the
following menu features:
•Channel-ACTIV - activate one or both channels and select the desired measurement method
•Channel-SYSTM - enter the individual channel parameters
•Channel-PIPE - enter the pipe parameters
•Channel-I/O - set up the inputs and outputs
•Channel-SET UP - set the signal limits, response times and activating mass flow.
•GLOBL-SYSTM - enter the system units
•GLOBL-I/O - set up error handling, option cards and display
•GLOBL-COMM - set the serial port and MODBUS parameters
As a programming aid, a complete set of menu maps for the
specific figure numbers will be referenced throughout this chapter, as required.
PROG menu is included in Appendix A, Menu Maps. The
1.2Programming Methods
You can use two methods for programming the XMT868i flowmeter:
•The keypad (shown in Figure 1 on page 2), which permits direct communication with the XMT868i
IMPORTANT: The XMT868i’s keypad permits programming of the instrument through the glass faceplate without
removing the cover. Thus, all programming procedures may be performed while the unit is installed in a
hazardous area.
•PanaView™, a PC-based software program that communicates with the XMT868i using its RS232/RS485 serial
Ethernet port.
Although the displays differ, the general procedures are the same for the keypad and PanaView methods. This chapter
provides detailed keypad programming instructions. If you are using PanaView, please refer to Appendix C,
Programming with PanaView, and/or the PanaView User’s Manual for complete instructions.
DigitalFlow™ XMT868i Programming Manual1
Chapter 1. Programming Site Data
Keypad Program
Along with the 2-line, 16-character LCD, the XMT868i includes a 6-key magnetic keypad. The decal cutout for each
key contains a hall effect sensor, pushbutton switch and visible red LED. The magnetic wand used to activate a
magnetic key is found attached to the meter chassis below the front panel. An operator activates the key by pressing the
magnetic wand up to the glass lid over the desired key. The LED will light indicating a successful key press.
Note:The pushbutton switch will also act as a key press but is done with the glass lid open. Do not use the
pushbutton switch in a hazardous area where the lid needs to be installed.
Use the magnetic keypad to navigate through the user program. The menu map may be followed in sequence, or the
four arrow keys may be used to scroll through the prompt screens. Figure 1 shows the front of the XMT868i, with
magnetic keypad and magnetic wand.
Magnetic Wand
Figure 1: XMT868i Magnetic Keypad and Wand
Programming of the
ACTIV, SYSTM, and PIPE submenus of the CHx menus and the GLOBL-SYSTM menu are required
for basic operation of the Model XMT868i. Failure to accurately enter all of the necessary information will result in
unreliable flow data. Therefore, be sure to complete at least the sections of this chapter pertaining to those three
submenus.
Note:Because it is so essential, instructions for programming the ACTIV, SYSTM, and PIPE submenus of the CHx menus
and the GLOBL-SYSTM submenus are also included in the Startup Guide. If that programming has already been
completed, skip those sections in this chapter.
Except for the three submenus noted above, it is not necessary to program the Model XMT868i flowmeter in any
particular order. Therefore, the sections of this chapter need not be completed in sequence. Enter the keypad program as
described in Accessing the Keypad Program on the next page and proceed immediately to any section of interest.
Note:Be sure to record all the programming data entered in this chapter in Appendix B, Data Records.
2DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.2.1Accessing the Keypad Program
Six keys on the keypad enable users to program the XMT868i:
•[Enter] - confirms choice of a specific option and data entry within the option
•[Escape] - allows users to exit from a specific option without entering unconfirmed data
•[] and []- enable users to highlight a specific window in the display option or to scroll through a list of options
(parameters, letters, numbers) in a menu
•[] and [] - enable users to scroll to a specific option, among choices in an option, or to a character in a text
entry. Also, if the [] key is pressed when the cursor position (underlined) is at the first digit of the bottom LCD
line, the previous parameter is displayed and can be edited.
When you power up the XMT868i, the initial screen shows the model and software version:
GE
XMT868i X3A.STD
The meter then starts to display measured parameters.
CH1 VEL E1
10.00 Ft/s
To enter the Keypad Program, press the
[Escape] key, followed by the [Enter] key, and the [Escape]key again.Each
successive key must be pressed within 10 seconds of the prior key. The Keypad Program window appears:
KEYPAD PROGRAM
PROG 1
Press
[Enter] to begin programming in the Channel or Global menus.
IMPORTANT: If the keypad has not been pressed for 10 minutes, the XMT868i exits the Keypad Program and returns to
displaying measurements. The meter retains any configuration changes that were confirmed with the
[Enter] key, and restarts as if the operator had completed the programming cycle.
DigitalFlow™ XMT868i Programming Manual3
Chapter 1. Programming Site Data
1.3Entering Data in the Channel Menu
The Channel menu is used to enter data specific to each channel. Refer to Figure 28 on page 77 and remember to record
all programming data in Appendix B, Data Records.
Note:In this manual, only the programming of Channel 1 will be described. To program Channel 2 of a 2-Channel
meter, simply repeat the same procedures presented for Channel 1.
1.3.1Selecting the Channel Measurement Method
The ACTIV submenu permits selection of the desired measurement method.
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to the desired channel (CH1 or CH2) and press [Enter].
Note:In the 1-Channel XMT868i, CH2 does not appear.
3. In the Channel PROGRAM menu, scroll to ACTIV and press [Enter].
4. The option offers two choices: OFF and TRANS (Transit-Time). Scroll to the desired option for this channel and
press [Enter].
Note:In the 1-Channel XMT868i, OFF does not appear.
The XMT868i returns to the Channel PROGRAM menu. Scroll to the SYSTM option.
4DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.2Entering Data in the Channel System Option
1. From the Channel PROGRAM menu, scroll to SYSTM and press [Enter].
2. The first prompt asks for the Channel (Site) Label.Use the four arrow keys to enter the desired label (in any
numeric or text combination up to five characters), and press [Enter].
3. The next prompt asks for the Channel (Site) Message. Use the four arrow keys to enter up to 15 characters, and
press [Enter].
4. The program now asks if you want the Energy Option, which enables flow measurement in a heating or cooling
system. Scroll to Off or On, and press [Enter].
5. The next prompt asks for the Volumetric Units. Scroll through the units (from the list shown in Table 1) and press
[Enter] when you reach the desired unit.
Note:Refer to the SYSTM option of the GLOBL menu to choose between English and metric measurements.
Table 1: Available Volumetric Units
English Volumetric UnitsMetric Volumetric Units
GAL/S - U.S. Gallons per Second L/S = Liters per Second
GAL/M - U.S. Gallons per MinuteL/M = Liters per Minute
GAL/H - U.S. Gallons per HourL/H = Liters per Hour
MGD - Mega U.S. Gallons per DayML/D = Mega Liters per Day
ft3/s - Cubic Feet per Secondm^3/s = Cubic Meters per Second
ft3/m - Cubic Feet per Minutem^3/m = Cubic Meters per Minute
ft3/h- Cubic Feet per Hourm^3/h = Cubic Meters per Hour
ft3/d- Cubic Feet per Daym^3/d = Cubic Meters per Day
Mf3/d- Mega Cubic Feet per DayMm3/d = Mega Cubic Meters per Day
BBL/S - Barrels per SecondBBL/S = Barrels per Second
BBL/M - Barrels per MinuteBBL/M = Barrels per Minute
BBL/H - Barrels per HourBBL/H = Barrels per Hour
BBL/D - Barrels per DayBBL/D = Barrels per Day
MBL/D - Mega Barrels per DayMBL/D = Mega Barrels per Day
A-I/S-Acre-inches per secondN/A
A-I/M -Acre-inches per minuteN/A
A-I/H -Acre-inches per hourN/A
A-I/D -Acre-inches per dayN/A
A-F/S -Acre-feet per secondN/A
A-F/M-Acre-feet per minuteN/A
A-F/H -Acre-feet per hourN/A
A-F/D -Acre-feet per dayN/A
DigitalFlow™ XMT868i Programming Manual5
Chapter 1. Programming Site Data
1.3.2Entering Data in the Channel System Option (cont.)
6. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric flow
display, and press [Enter].
7. Scroll to the desired Totalizer Units for the totalized flow rate display (listed in Table 2), and press [Enter].
.
Table 2: Totalizer Units
EnglishMetric
GAL - U.S. GallonsL - Liters
MGAL - Mega U.S. gallonsML - Megaliters
ft^3 - Cubic Feetm^3- Cubic Meters
Mft^3 - Mega Cubic FeetMm^3- Mega Cubic Meters
BBL - BarrelsBBL - Barrels
MBBL - MegabarrelsMBBL - Megabarrels
AC-IN - Acre-inches
AC-FT - Acre-feet
8. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow rate
display, and press [Enter].
The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for
details) or the Energy Option.
•If you have activated Mass Flow, continue to step 1 on the next page.
•If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 7.
•If you have activated neither option, the XMT868i returns to the Channel PROGRAM window.
6DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.2aProgramming the Mass Flow Option
1. Scroll to the desired Mass Flow Units for flow rate display (listed in Table 3), and press [Enter].
Table 3: Available Mass Flow Units
EnglishMetric
LB = PoundsKilograms
KLB = KiloPounds (Thousands of Pounds)Tonnes = Metric Tons (1000 KG)
MMLB = MegaPounds
TONS (2000 LB)
2. Scroll to the desired Mass Flow Time Units (from seconds to days), and press
3. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate
display, and press [Enter].
4. Scroll to the desired Mass Totals (units for the totalized mass flow rate display, listed in Table 3), and press [Enter].
5. Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point) in the totalized mass
flow rate display, and press [Enter].
The program varies, depending on whether or not you have enabled the Energy Option.
[Enter].
•If you have not enabled the option, the XMT868i returns to the Channel PROGRAM window. Proceed to the PIPE
option.
•If you have enabled the option, continue to the next page.
DigitalFlow™ XMT868i Programming Manual7
Chapter 1. Programming Site Data
1.3.2Entering Data in the Channel System Option (cont.)
1.3.2bProgramming the Energy Option
1. Scroll to the desired Power Units (listed in Table 4) and press [Enter].
4. Scroll to the desired number of Energy Decimal Digits, and press [Enter].
5. If you are using a heating or cooling system, scroll to the appropriate selection (Heat or Cool) and press [Enter].
6. The final prompt asks whether you want to conduct Flow Measurement at the point of RTN (Return, where the
liquid exits) or of SPPLY (Supply, where the liquid enters). Scroll to the appropriate choice, and press [Enter].
You have completed entering data in the SYSTM option. The meter returns to the Channel PROGRAM menu.
8DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.3Entering Pipe Parameters
Enter the transducer and pipe parameters using the PIPE submenu. While following the programming instructions, refer
to Figure 29 on page 78.
1. From the
2. The first prompt asks for the Transducer Number.
Channel PROGRAM menu, scroll to the PIPE option and press [Enter].
• For a standard transducer, enter the number engraved on the transducer head in the right pane, and press [Enter].
• If there is no number engraved on the transducer head, press the right arrow key to scroll to the Special option,
enter an assigned number (from 91 to 99), and press
IMPORTANT: Special transducers, which have no engraved number on the head, are rarely used. Examine the
transducer head carefully for a number.
The menu now varies, depending on your selection in step 2.
[Enter].
•If you entered the number for a standard clamp-on transducer, proceed to the Pipe Material prompt in step 4.
•If you entered the number for a standard wetted transducer, proceed to the Pipe OD prompt in step 5.
•If you entered the number for a special transducer, proceed to step 3 below.
1.3.3aSpecial Transducers
Note:For special transducers, GE will supply a transducer data sheet with programming information.
3. For special transducers:
a. The program first asks for the Wedge Type. Press the appropriate transducer type (Rayleigh clamp-on, shear
clamp-on, or wetted).
Note:GE will supply the information required for steps b through f with the transducers.
b. Press the appropriate Frequency (from 500 kHz to 5 MHz).
c.Enter the Time Delay (Tw) value supplied by GE and press
If you have selected a special wetted transducer, the program now goes to step 5. But if you have selected a clamp-on
transducer, three more prompts appear.
d. Enter the Wedge Angle of the transducer (the angle of the ultrasonic transmission, in degrees) and press
e.Enter the Wedge Temperature and press [Enter].
f.Enter the Wedge Soundspeed and press [Enter].
[Enter].
[Enter].
DigitalFlow™ XMT868i Programming Manual9
Chapter 1. Programming Site Data
1.3.3Entering Pipe Parameters (cont.)
1.3.3bPipe Material
4. Scroll to the appropriate Pipe Material from the supplied list. For some materials (shown in Table 6), the program
displays additional selections. At the desired choice, press [Enter].
and press [Enter]. If the soundspeed is
unknown, refer to the Sound Speeds and Pipe Size Data manual (914-004).
10DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.3cPipe OD
5. To select the appropriate Pipe OD unittype from the list shown in Table 7, scroll to the right side of the screen, and
use the up and down arrow keys to step through the list. Press [Enter]. Then enter the known pipe outside diameter
or circumference on the left side and press [Enter].
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds and Pipe Size Data (914-004).
Table 7: Available Pipe OD Units
EnglishMetric
inch mm = millimeters
feetm = meters
in/PI = pipe circumference in inchesmm/PI = pipe circumference in millimeters
ft/PI = pipe circumference in feetm/PI = pipe circumference in meters
6. Enter the known Pipe Wall thickness (in inches or mm) in the right pane and press
If you are using clamp-on transducers, skip to step 9 on the next page.
[Enter].
1.3.3dPath and Axial Lengths
7. To enter the Path Length:
a. Use the [] arrow key to highlight the path length unit type at the right of the screen.Then use the [] and []
arrow keys to scroll to the desired unit type (inches or feet for English measurements, or millimeters or meters
for metric).
b. Use the [] arrow key to return to the numeric entry on the left, and enter the path length of the ultrasonic
signal. Press [Enter].
Note:If a spoolpiece was ordered with the meter, the transducer signal path length (P) and the transducer signal axial
length (L) are engraved on the flowcell and/or are included in the documentation supplied with the meter. For
on-site transducer installations, refer to Appendix C, Measuring P and L Dimensions, for instructions.
8. In the same manner, enter the appropriate Axial Length L unit type and axial length of the ultrasonic signal, and
press [Enter]. Skip to step 10 on the next page.
1.3.3ePipe Lining
9. Scroll to the appropriate choice to indicate whether you have a Pipe Lining.
a. If you have selected Ye s, then select the Lining Material from the supplied list.
• If OTHER is selected, enter the lining sound speed, and press [Enter].
b. Then enter the lining thickness and press
DigitalFlow™ XMT868i Programming Manual11
[Enter].
Chapter 1. Programming Site Data
1.3.3Entering Pipe Parameters (cont.)
1.3.3fTracking Windows
10. Press the appropriate choice to indicate whether you want Tracking Windows. These windows are used to detect the
receive signal when you are unsure of the fluid sound speed.
1.3.3gFluid Type
11. Press the appropriate Fluid Type. The selections for fluid type vary depending on whether:
• the Energy Option is ON or OFF; and
• the Tracking Window is enabled or disabled.
Refer to Table 8 if Energy Option is
OFF, or to Table 9 on page 13 if Energy Option is ON.
Note:Some of the fluid types may require additional selections, as shown in the following tables.
Table 8: Fluid Types for ENERGY OFF
Tracking Windows =
NOAdditional SelectionsYESAdditional Selections
WaterNormal Water
Sea Water
If Normal Water is selected, enter the Water
Temperature and press [Enter].
OilLube Oil
Crude Oil
Methanol
(25 C)
Ethanol
(25C)
LN2No additional selections required.
FreonNo additional selections required.
OTHEREnter the fluid soundspeed and press [Enter].
No additional selections required.Oil,
No additional selections required.OTHEREnter the Minimum Soundspeed and press
Water,
0-100C
Water,
0-260C
Tracking
No additional selections required.
No additional selections required.
No additional selections required.
[Enter]. Then enter the Maximum Soundspeed
and press [Enter].
12DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.3gFluid Type (cont.)
Table 9: Fluid Types for ENERGY ON
Tracking Windows =
NOAdditional SelectionsYESAdditional Selections
WaterEnter the Water Temperature and press [Enter].Water,
0-260C
Water,
Glycol
Mixture
OTHEREnter the Fluid Soundspeed and press [Enter].OTHEREnter the Minimum Soundspeed and press
Enter the Fluid Soundspeed and press [Enter].
Then enter the Percentage of Water and press
[Enter].
Water,
Glycol
Mixture
No additional selections required.
Enter the Percentage of Water and press [Enter].
[Enter]. Then enter the Maximum Soundspeed
and press [Enter].
1.3.3hReynolds Correction
12. Scroll to the appropriate choice to indicate whether you want Reynolds Correction.
• If Off is selected, enter the Calibration Factor and press [Enter].
• If Activ is selected, the program asks for KV Input Selection. Scroll to SIGS (Signals), SOUND (Soundspeed) or
STATC.
• If SIGS or SOUND is selected, enter the Calibration Factor and press [Enter].
• If STATC is selected, enter the Kinematic Viscosity and press [Enter]. Then enter the Calibration Factor and
press [
Enter].
•For wetted transducers, the program returns to the Channel PROGRAM menu. You have completed entering PIPE
parameters.
•For clamp-on transducers, proceed to step 13 below.
1.3.3iNumber of Traverses
13. Scroll to the appropriate Number of Traverses (from 1 to 5) and press [Enter].
14. Accept the value for the Transducer Spacing by pressing [Enter].
Note:The XMT868i automatically calculates the transducer spacing. You should not enter a different value unless
specifically instructed by the factory.
The XMT868i returns to the Channel PROGRAM menu. You have completed entering PIPE parameters.
Note:The site data programmed in this chapter is automatically stored in the meter’s non-volatile RAM (NVR). If the
power to the meter is disconnected, the same site data will be applied when the unit is next powered on.
DigitalFlow™ XMT868i Programming Manual13
Chapter 1. Programming Site Data
1.3.4Entering Input/Output Parameters
Enter the zero cutoff value and set up the temperature inputs using the I/O submenu. While programming these
parameters, see Figure 30 on page 79.
IMPORTANT: If an option card in Slot 1 fails to appear in this menu, it may be turned OFF. See the GLOBL-I/O-OPTN
section on page 29 for setup instructions.
1.3.4aZero Cutoff Value
Near a zero flow rate, the Model XMT868i’s readings may fluctuate due to small offsets caused by thermal drift or
similar factors. To force a zero display reading when there is minimal flow, enter a zero cutoff value as described in the
following steps:
1. From the Channel PROGRAM menu, scroll to the I/O option in the center pane.
2. The program asks for the Zero Cutoff. Enter a value from 0 to 1 ft/sec (0 to 0.30 m/sec) for the zero cutoff and press
[Enter].
The recommended setting is 0.1 ft/sec (0.03 m/sec).
Proceed with one of the following:
• If the ENERGY OPTION is OFF, the display now returns to the Channel menu.
• If the ENERGY OPTION is ON, refer to Temperature Input below.
1.3.4bTemperature Input
The Model XMT868i can use either a fixed temperature value or a live temperature input to calculate energy. Complete
the following steps to configure this input:
Note:The site data programmed in this chapter is automatically stored in the meter’s non-volatile RAM (NVR). If the
power to the meter is disconnected, the same site data will be applied when the unit is next powered on.
3. At the Temp Supply prompt, scroll to Fixed to enter a constant temperature value or on Slot1 to select the output in
Slot 1 that will provide the live temperature input for the supply.
Note:If Slot 1 contains an activated option card with an analog input assigned to TEMP or an RTD input, Slot 1
appears as an option at the above prompt. If the process temperature is stable, a fixed value may be used, but
most applications require a live temperature input.
14DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.4bTemperature Input (cont.)
4. One of the following two prompts will appear, depending on your selection in Step 3.
• If Fixed was chosen, enter the known fixed process temperature and press [Enter]. The meter will only accept
o
values from –328° to 1832°F (–200
to 1000oC).
• If Slot 1 was chosen, scroll to the desired temperature input (from A to D) and press [Enter].
5. At the Temp Return prompt, scroll to Fixed to enter a constant temperature value or on Slot1 to select the output in
Slot 1 that will provide the live temperature input for the return.
Note:If Slot 1 contains an activated option card with an analog input assigned to TEMP or an RTD input, Slot 1
appears as an option at the above prompt. If the process temperature is stable, a fixed value may be used, but
most applications require a live temperature input.
6. One of the following two prompts will appear, depending on your selection in Step 3.
• If Fixed was chosen, enter the known fixed process temperature and press [Enter]. The meter will only accept
o
values from –328° to 1832°F (–200
to 1000oC).
• If Slot 1 was chosen, scroll to the desired temperature input (from A to D) and press [Enter].
The program now returns to the Channel PROGRAM menu.
1.3.5Entering Setup Parameters
The signal limits and response times for the Model XMT868i are specified using the SET UP submenu. While following
the programming instructions, refer to Figure 30 on page 79. This submenu includes four options:
•SIGNL - set the parameters related to the transducer signal
•AVRG - specify the response of the meter to step changes
•INIT - initialize all parameters to default values
•ADVAN - enable mass flow, edit kinematic viscosity vs. sound speed table, activate K factors, and select transmit
code length.
To enter the SET UP submenu, scroll to the
record all programmed data in Appendix B, Data Records.
SET UP option in the Channel PROGRAM menu and press [Enter]. Remember to
DigitalFlow™ XMT868i Programming Manual15
Chapter 1. Programming Site Data
1.3.5Entering Setup Parameters (cont.)
1.3.5aThe SIGNL Option
Use this option to set the limits for the incoming signal and other parameters affecting the transducer signal. For
example, the programmed signal strength low limit may be used to determine the trigger point for an alarm.
CAUTION!The SIGNL default settings are suitable for most applications. Consult the factory
before changing any of these parameters.
1. From the SET UP option menu, scroll to SIGNL and press [Enter].
2. Press [Enter] to accept the current Signal Low Limit value or enter a new value and press [Enter].
The default value for the above parameter is 40, and values from –20 to 100 are acceptable. The E1: LOW SIGNAL error
message appears when the signal strength falls below the programmed Signal Low Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.
3. Press [Enter]to accept the current Corr. Peak Limit value or enter a new value and press [Enter].
The default value for the above parameter is 100, and values from 0 to 500 are acceptable. The E4: SIGNAL QUALITY error
message appears when the signal quality falls below the programmed Cor. Peak Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.
4. Press [Enter] to accept the current Sndspeed +– Lim. value or enter a new value and press[Enter].
The default value for the above parameter is 20%, and values from 1% to 50% are acceptable. The E2: SOUNDSPEED
error message appears when the calculated fluid soundspeed differs from the fluid soundspeed entered in the
CHx-SYSTM menu by more than the programmed Soundspeed +– Limit value. See Chapter 2, Error Codes, in the
Service Manual for a discussion of error codes.
5. Press [Enter]to accept the current Vel Low Limit value or enter a new value and press [Enter].
The default value for the above parameter is –150.0 ft/sec. (–46 m/sec) and values from –500 to 500 ft/sec (–150 to 150
m/sec) are acceptable. The E3: VELOCITY RANGE error message appears when the calculated fluid velocity is less than the
programmed Velocity Low Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error
codes.
6. Press [Enter]to accept the current Vel High Limit value or enter a new value and press [Enter].
The default value for the above parameter is 40.0 ft/sec (12.142 m/sec) and –100 to 100 ft/sec (-30.48 to 30.48 m/sec)
are acceptable values. The E3: VELOCITY RANGE error message appears when the calculated fluid velocity exceeds the
programmed Vel High Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.
16DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.5aThe SIGNL Option (cont.)
7. Press [Enter]to accept the current Accel. Limit value or enter a new value and press[Enter].
2
The default value for the above parameter is 15.0 ft/sec
are acceptable. The E6: CYCLE SKIP error message appears when the calculated fluid velocity changes by more than the
programmed Acceleration Limit value from one reading to the next. See Chapter 2, Error Codes, in the Service Manual
for a discussion of error codes.
8. Press [Enter]to accept the current Amp. Discrim Low value or enter a new value and press[Enter].
The amplitude discriminator measures the transducer signal received by the Model XMT868i. The default value for the
above parameter is 14, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error message appears when the
amplitude discriminator falls below the programmed Amp. Discrim Low value. See Chapter 2, Error Codes, in the
Service Manual for a discussion of error codes.
9. Press [Enter]to accept the current Amp. Discrim. High value or enter a new value and press [Enter].
The amplitude discriminator measures the transducer signal received by the Model XMT868i. The default value for the
above parameter is 34, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error message appears when the
amplitude discriminator exceeds the programmed AMP. DISCRIM HIGH value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.
(4.6 m/sec2) and values from 0 to 100 ft/sec2 (0 to 30 m/sec2)
10. Press [Enter]to accept the current Delta T Offset value or enter a new value and press [Enter].
An offset between the upstream and downstream transit times is specified at this prompt. The default value for the
above parameter is 0 sec, and values from –1000 to 1000 sec are acceptable.
11. Press [Enter]to accept the current % of Peak value or enter a new value and press [Enter].
The percentage of peak used to calculate the transit times and Delta T is specified at this prompt. The default value for
the above parameter is -40%, and values from -100 to 100% are acceptable.
Note:This setting is a starting point for detecting the signal. The meter will automatically adjust this value if the
calculated transit time is unacceptable. You can set the limits for this value using the MIN. PEAK% and MAX.
PEAK% discussed on the next page.
12. Scroll to the desired XMIT Sample Size (from 2 to 32) and press [Enter]
Both the upstream and downstream transducers transmit ultrasonic pulses in bursts, which consist of a series of
transmit pulses. XMIT Sample Size determines how many bursts are sent in one direction before sending in the other
direction. The default value for the above parameter is 8 and values of 2, 4, 8, 16 and 32 are available.
13. Press [Enter]to accept the current # Errors Allowed value or enter a new value (0 to 16) and press [Enter].
Use this prompt to enter the number of errors the XMT868i can record before it displays an error message. The default
value is 8.
DigitalFlow™ XMT868i Programming Manual17
Chapter 1. Programming Site Data
1.3.5aThe SIGNL Option (cont.)
14. Press [Enter] to accept the current Min Peak% Limit value or enter a new value and press [Enter].
Use this prompt to enter the minimum percent of peak that the XMT868i can use to measure transit time. The
XMT868i accepts values from -100 to 100.
15. Press
Use this prompt to enter the maximum percent of peak that the XMT868i can use to measure transit time. The
XMT868i accepts values from -100 to 100.
After responding to the above prompt, the meter returns to the
[Enter] to accept the current Max Peak% Limit value or enter a new value and press [Enter].
SET UP option window.
1.3.5bThe AVRG Option
Use this option to specify the number of readings that occur before the meter will respond to a step change in flow rate.
In general, the smaller the number of readings, the less steady the display will appear. Complete the following steps to
set the response time:
1. From the SET UP option menu, scroll to the AVRG option and press [Enter].
2. Scroll to the desired Response Time (from 1 to 60 sec, or STATS (statistics)) and press [Enter].
For best results, select the STATS (statistics) option, as this increases the response time under steady flow conditions
while still allowing a rapid response to changes in flow rate.
After responding to the above prompt, the meter returns to the SET UP option window.
1.3.5cThe INIT Option
Use this option to initialize (reset) all of the parameters within the SIGNL menu back to their default values. Complete
the following steps to reset all of the parameters:
1. From the SET UP option menu, scroll to INIT in the center pane.
2. Scroll to No to keep the current values or on Ye s to reset all values to their default settings.
After responding to the above prompt, the meter returns to the SET UP window.
1.3.5dThe ADVAN Option
This option enables you to enable the more advanced features of the meter. In this option you can do the following:
• calculate the kinematic viscosity (KV) based on signal strength or on soundspeed
• enter a table of K-factors (based on velocity or reynolds number) that compensates for non-linear flow rates
• enable mass flow (calculated for static fluid density)
• select the size of the transducer transmission signal.
18DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.5eThe KV/xx Option
Note:To select either signal strength or soundspeed for kinematic viscosity calculations, go to Reynolds Correction
in the PIPE submenu on page 13.
Use this option to calculate the kinematic viscosity (KV) based on signal strength (SS) or on soundspeed, as selected in
the KV Input Selection option of the PIPE option. To use this option, you must enter the KV vs. SS pairs in descending
order of KV and ascending order of SS or soundspeed. For example, if pair 1 is KV = 10 centistrokes and SS = 62, then
pair 2 must be KV
enter KV and signal or soundspeed values:
1. From the ADVAN option menu, scroll to KV/xx and press [Enter].
2. Press [Enter] to accept the current # of KV/SS Pairs or enter a new value (2 to 20) and press [Enter].
3. Press [Enter] to accept the current Sig Strength (or Soundspeed) #01 value or enter a new value and press [Enter].
Enter the signal strength values in ascending order.
Note:For signal strength, the acceptable range is 50.0 to 85.0. For soundspeed, the acceptable range is 125 to 9000.
4. Press [Enter]to accept the current Kinematic Viscosity value or enter a new value (0.050 to 500) and press [Enter].
Enter the kinematic viscosity values in descending order.
10 centistrokes and SS 62. The XMT868i accepts 2 to 20 pairs. Complete the following steps to
The Sig Strength (or Soundspeed) and Kin Viscosity prompts repeat for each pair. After entering all the pairs, the meter
returns to the ADVAN option.
DigitalFlow™ XMT868i Programming Manual19
Chapter 1. Programming Site Data
1.3.5fThe MULTK Factors Option
Use this option to enter a table of K-factors. K-factors are used to create a curve for the flow range (based on velocity
or reynolds number) that compensates for non-linear flow rates. The meter accepts from 2 to 20 pairs. Complete the
following steps to enter multiple K factors for velocity or reynolds values:
1. From the ADVAN option menu, scroll to MULTK and press [Enter].
2. Scroll to Ye s to Activate Multi K or on No to disable this option, and press [Enter].
If NO was selected, the meter returns to the ADVAN window. If YES was selected, proceed to Step 3.
3. Scroll to the desired custom type: CstV (velocity) or CstR (reynolds)
4. Scroll to Ye s to Edit the K-factor Table or on No to retain the current K-factor table (and return to the ADVAN option)
and press [Enter].
Note:If the necessary velocity/reynolds vs. K-factor data was not provided with the Model XMT868i’s
documentation, the K-factor table cannot be edited.
If NO was selected, the meter returns to the ADVAN option. If YES was selected, proceed to Step 5.
5. Enter the Number Of K-factors (from 2 to 20) to be entered into the table, and press [Enter].
Note:When editing the K-factor table, the velocities must be entered in increasing order.
6. Enter the Velocity/Reynolds Value for K-factor number “X” and press [Enter].
7. Enter the K-factor corresponding to velocity/reynolds number “X” (0.333 to 3.0) and press [Enter].
The Velocity # and K Factor # prompts repeat for each pair. After entering all the pairs, the meter returns to the ADVAN
window.
20DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.3.5gThe Mass Flow Option
Use this option to calculate mass flow from a static fluid density. Complete the following steps to enter the static
density of the fluid:
1. From the ADVAN menu, scroll to MASS and press [Enter].
2. Scroll to Ye s to activate Mass from Static or on No to disable this option.
3
a. If you select Ye s, enter the Fluid Density (0.062 to 624.220 lb/ft
b. For either option, press [Enter].
After responding to the above prompt, the meter returns to the ADVAN window.
or 0.001 to 9.9675g/cc).
1.3.5hThe Code Length Option
Use this option to select the size of the transducer transmission signal. This option is helpful when measuring flow in
small pipes. You can choose from the following selections:
• AUTO - The meter will automatically determine the code length (short or long) which is based on pipe size and
transducer type.
• SHORT - is just few pulses with no code pattern built in. In some cases, the diameter of the pipe is so small that it
does not give each transducer the necessary time to send a series of signals before receiving a series of signals.
SHORT is also a good selection for pipes made of materials such as PTFE
which tend to blur the signal pattern.
• LONG - is a series of approximately a dozen pulses that have a distinctive binary pattern, easily recognizable
during auto correlation.
Complete the following steps to select the code length:
1. From the ADVAN option menu, scroll to CODEL and press [Enter].
2. Scroll to the desired Xmit Code Length and press [Enter]. AUTO is the default.
After responding to the above prompt, the meter returns to the ADVAN window. Press [Escape] three times to return to
the Channel PROGRAM, or four times to return to the Keypad Program.
DigitalFlow™ XMT868i Programming Manual21
Chapter 1. Programming Site Data
1.4Entering Data in the GLOBL Menu
The GLOBL menu is used to enter information that is not specific to any of the individual channels. Information
programmed by using this menu is used to enter several general system parameters (e.g., English or metric units). For
meters with 2 channels, this menu is also used to compute parameters such as the sum, difference or average of the
channel 1 and channel 2 signals. When calculating the SUM, DIF or AVE readouts, data from the Global-SYSTM submenu
is used. Any conflicting data entered in the Channel-SYSTM submenu is overridden.
The following submenus are included in the GLOBL menu:
•SYSTM - use to specify the units of measure used in calculations
•I/O - used to set up error handling and to configure analog inputs and outputs
•COMM - used to set up the serial communications port and MODBUS parameters
To enter the
1. From the display screen, press [Escape], [Enter], [Escape] on the keypad.
2. Press [Enter] to enter the PROG menu.
3. In the PROG menu, scroll to GLOBL and press [Enter].
Then proceed to the appropriate section of this chapter for instructions. Refer to Figure 31 on page 80 and Figure 32 on
page 81, and remember to record all programming data in Appendix B, Data Records.
GLOBL menu:
1.4.1Entering GLOBL-SYSTM Data
While completing these instructions, refer to Figure 31 on page 80.
1. From the PROG menu, scroll to GLOBL. Press [Enter].
2. Scroll to the SYSTM option. Press [Enter].
3. The first prompt asks for a Meter Message. Enter a numeric or text message (up to 16 characters) in the right pane
and press [Enter].
4. Scroll to the desired System Units selection (either metric or English) and press [Enter]. The XMT868i will display
all parameters and measurements in the designated units.
•For a single-channel XMT868i, the program returns to the Global PROGRAM menu. For a 2-channel meter, proceed
to step 5 on the next page.
22DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.1aVolumetric Units
5. Scroll to and scroll to the desired Volumetric Units for the flow rate display. Table 10 lists the available volumetric
6. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric flow
display, and press [Enter].
DigitalFlow™ XMT868i Programming Manual23
Chapter 1. Programming Site Data
1.4.1bTotalizer Units
7. Scroll to the desired Totalizer Units for the totalized flow rate display (listed in Table 11), and press [Enter].
Table 11: Totalizer Units
EnglishMetric
GAL = GallonsL = Liters
MGAL = Mega gallonsML = Megaliters
ft^3 = Cubic feetm^3 = Cubic meters
Mft^3 = Mega cubic feetMm^3= Mega cubic meters
BBL = BarrelsBBL = Barrels
MBBL = Mega barrelsMBBL = Mega barrels
AC-IN = Acre-inches
AC-FT = Acre-feet
8. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow
display, and press [Enter].
The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for
details) or the Energy Option in one of the Channel PROGRAM options.
•If you have activated Mass Flow, continue to step 1 on the next page.
•If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 1-30.
•If you have activated neither option, the XMT868i returns to the Global PROGRAM window. Press [Escape] to return
to the
PROG menu.
24DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.1cProgramming Mass Flow Data
1. Scroll to the desired Mass Flow Units for flow rate display (listed in Table 12), and press [Enter].
4. Scroll to the desired number of Energy Decimal Digits, and press
[Enter].
You have completed entering selections in this option.The program returns to the Global PROGRAM menu.
26DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.2Setting Up Inputs and Outputs
Set up the XMT868i’s inputs and outputs using the I/O submenu. While following the programming instructions, refer
to Figure 31 on page 80 and Figure 32 on page 81. Remember to record all programmed data in Appendix B, Data Records.The I/O submenu consists of the following options:
•Error Handling - program the meter’s response during an error condition
•Options - set up any option cards and the Slot 0 analog outputs
•LCD - set up the LCD display (refer to Chapter 2, Displaying Data)
To enter the
1. From the PROGRAM menu, scroll to GLOBL and press [Enter].
2. Scroll to I/O and press [Enter].
Note:In this section, Slot 1 appears as an option only if a suitable option card is installed in Slot 1.
I/O submenu:
1.4.2aSetting Up Error Handling
This menu option lets you set how the XMT868i will handle the outputs for measurements and average (two-path)
measurements during an error condition. See Chapter 2, Error Codes, in the Service Manual for a discussion of the
built-in error codes.
1. In the
2. Scroll to the desired Error Handling selection (as shown in Table 15 and Table 16 on page 28).
See Table 15 and Table 16 on page 28 for a description of the error handling options available and how the totalizers
and display respond to them for a single and two-channel meter.
Global I/O option, scroll to ERROR and press [Enter].
a. If you selected Other, enter the 4-20 mA Error Level (the number of milliamps the analog output will put out
during a fault condition). Enter a whole number between 0 and 22, and press [Enter].
Table 15: Error Options and Responses, Single-Channel Meter
OptionOutput ResponseTotalizer Response
HOLDHolds the last “good” readingHolds the last “good” reading and continues to
totalize, based on that reading
LOWForces the outputs to the low set pointStops totalizing
HIGHForces the outputs to the high set pointStops totalizing
HHIGHForces the outputs 10% above the high set pointStops totalizing
OTHERAllows user to enter 4-20 mA error level in mA.Sends analog outputs to entered mA level
when a fault occurs.
DigitalFlow™ XMT868i Programming Manual27
Chapter 1. Programming Site Data
1.4.2Setting Up Inputs and Outputs (cont .)
1.4.2aSetting Up Error Handling (cont.)
Table 16: Error Options and Responses for a 2-Channel Meter
Totalizer Response When Error Handling is
When MeasuringDisplay Response
CH1 or CH2
(vel, vol, etc.)
SUM
DIF
AVESee Error Handling for Average Measurements below.
Holds last “good” reading.Holds last “good” reading and continues
to totalize based on that “good” reading.
Adds two channels using
the last “good” reading.
Subtracts two channels
using last “good” reading.
Holds last “good” reading and continues
to totalize based on two channels.
Holds last “good” reading and continues
to totalize based on two channels.
HOLDLOW, HIGH, HHIGH
Stops totalizing.
Stops totalizing if either or
both channels go into error.
Stops totalizing if either or
both channels go into error.
For a one-channel meter, after responding to the above prompt, the meter returns to the Global I/O prompt shown on the
previous page. For a two-channel meter, proceed to step 3 below.
Error Handling for Average Measurements - AVE
The 2PATH ERROR HANDLING option is intended for applications where two sets of transducers are installed in the same
location in the same pipe to improve accuracy and the meter is operated in AVE mode. With this function enabled, the
Model XMT868i performs error handling only if both channels are in error. If this function is disabled, error handling
occurs when either channel goes into error.
3. Scroll to Ye s to enable Two-path Error Handling, or on No to disable this function, and press [Enter].
Specific responses of the display and the totalizer to the two-path error handling option available at the above prompt
are listed in Table 17.
Table 17: 2-Path Error Response Options
OptionDisplay ResponseTotalizer Response
NODisplays the average of CH1 and CH2, regardless
of the error state of either channel.
YES1. If one channel is in error, the other channel’s
value is displayed as the average.
Outputs the average of CH1 and CH2 totals,
regardless of the error state of either channel.
1. If one channel is in error, totalizing continues.
2. If both channels are in error, totalizing stops.
2. If both channels are in error, the last average
reading is held.
After responding to the above prompt, the meter returns to the GlobalI/O option.
28DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.2bSetting Up Option Cards
The Model XMT868i has two built-in analog outputs, which are assigned to Slot 0. Also, a variety of input/output
option cards may be installed in Slot 1. See Chapter 1, Installation, of the Startup Guide for a complete description of
the available option cards.
To access this submenu, scroll to OPTN in the I/O option. To set up and/or scale the inputs and outputs, complete the
following steps:
•Scroll to the desired I/O Slot to program the I/O functions in that slot, and press [Enter].
Note:If an option card is not installed in Slot 1, the
Slot1 option does not appear at the above prompt.
Proceed to the appropriate section for programming instructions specific to the type of input or output selected above.
Analog Outputs
Complete the following steps to set up the analog outputs for any slot (refer to Figure 31 on page 80):
1. Scroll to the desired Output (A or B), and press [Enter].
Note:The set up of output A is used here as an example. Identical procedures would be used to set up output B.
2. Scroll to Off to disable Output A and return to step 1, or on 0-20 mA or 4-20 mA to specify the desired range for
output A. In either case, press [Enter].
Note:For a 1-Channel meter, proceed to step 4.
3. Scroll to the desired Channel option and press [Enter]. See Table 18 for a description of the channel options
available.
Table 18: Channel Options
OptionDescription
CH1Channel 1
CH2Channel 2
SUMCH1+CH2
DIFCH1-CH2
AVE(CH1+CH2)/2
4. Scroll to the desired Measurement Name and press
[Enter]. (Table 19 on page 30 provides a description of the
available parameters.)
DigitalFlow™ XMT868i Programming Manual29
Chapter 1. Programming Site Data
1.4.2bSetting Up Option Cards (cont.)
Analog Outputs (cont.)
Note: All the options shown below will only appear if the Mass Flow and Energy Options are activated.
Table 19: Output Measurement Options
Option Bar ChoiceDescription
VELFlow Velocity
VOLUMVolumetric Flow
+TOTLForward Totalized Volume Flow
-TOTLReverse Totalized Volume Flow
TIME Total Flow Measurement Time
MDOTMass Flow
+MASSForward Totalized Mass Flow
-MASSReverse Totalized Mass Flow
POWEREnergy Flow Power
+ENRGForward Energy Flow
-ENRGReverse Energy Flow
SS upSignal strength for the upstream transducer.
SS doSignal strength for the downstream transducer.
DAC upAGC DAC count for upstream gain setting.
DAC doAGC DAC count for downstream gain setting.
SoundspeedMeasured speed of sound in the fluid.
T upUpstream ultrasonic signal transit time.
TdownDownstream ultrasonic signal transit time.
DELTATTransit time difference between upstream and downstream signals.
REYN#Current velocity/reynolds number.
K (RE)K factor, based on the Reynolds number.
PEAK%Percentage of peak (set to +50 by default).
THETA3
Theta 3 is the angle between the ultrasonic beam and the normal to the pipe wall. It is
calculated from the measured transit time and clamp-on parameters (clamp-ons only).
QupDisplays the signal quality for the upstream transducer.
QdownDisplays the signal quality for the downstream transducer.
AMPupDisplays the value for the signal amplitude of the upstream transducer.
AMPdnDisplays the value for the signal amplitude of the downstream transducer.
P#upDisplays signal peaks for the upstream transducer.
P#dnDisplays signal peaks for the downstream transducer.
30DigitalFlow™ XMT868i Programming Manual
Table 19: Output Measurement Options (cont.)
Option Bar ChoiceDescription
NFupDisplays the normalization factor for the upstream transducer.
NFdnDisplays the normalization factor for the downstream transducer.
CxdcrDisplays the speed of sound in the transducer (clamp-on only).
TEMPsDisplays the temperature for the supply input (energy measurement).
TEMPrDisplays the temperature for the return input (energy measurement).
Ts-TrDifference between supply input and return input measurement (energy measurement).
DENSsDisplays the density for the supply input (energy measurement).
DENSrDisplays the density for the return input (energy measurement).
Chapter 1. Programming Site Data
DELTh
Displays the delta enthalpy which is the difference between the supply and return.
Enthalpy is a measure of energy contained in the fluid (energy measurement).
Note:The measurement units that appear in these prompts are those selected in the Global-System menu earlier in
this section.
5. Enter a measurement value for the Base (low) end of the analog output range and press [Enter].
6. Enter a measurement value for the Full (high) end of the analog output range and press [Enter].
After responding to the above prompt, the meter returns to the I/O option.
DigitalFlow™ XMT868i Programming Manual31
Chapter 1. Programming Site Data
1.4.2bSetting Up Option Cards (cont.)
Option Card Analog Inputs
Complete the following steps to set up the analog inputs of an option card installed in Slot 1 (refer to Figure 32 on
page 81):
1. Scroll to the desired Input (A, B, C or D) and press [Enter].
Note:The setup of Input A is used as an example. Identical procedures would be used to set up the remaining inputs.
2. Enter a Label of up to eight characters for input A and press [Enter].
3. Scroll to Off to disable Input A and return to step 1, or on Temp or Spec to designate it as a temperature or special
input, and press [Enter].
•OFF
If
OFF was selected to disable input A, the meter returns to the Input prompt in step 1. Scroll to another input, or press
[Escape] to move to another menu.
•TEMP
If
TEMP was selected to set up input A as a live temperature input, complete steps 4 and 5 below:
4. Enter a temperature value for the Base (low) end of the analog input range and press [Enter].
5. Enter a temperature value for the Full Scale (high) end of the analog input range and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.
•SPEC
If
SPEC was selected to set up Input A as a live special input, complete steps 6 through 10 below:
6. Enter a Name for input A and press [Enter].
7. Enter the Units of measurement for Input A and press [Enter].
8. Enter a value for the Base (low) end of the analog input range and press [Enter].
9. Enter a value for the Full Scale (high) end of the analog input range and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.
32DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
Option Card RTD Inputs
o
Option cards with RTD inputs have a temperature range of -148
to 660oF (–100° to 350°C). Complete the following
steps to set up the RTD inputs of an option card installed in Slot 1 (refer to Figure 32 on page 81):
1. Scroll to the desired Input (A, B, C or D) and press [Enter].
Note:The setup of RTD input A is used here as an example. Identical procedures would be used to set up the
remaining RTD inputs.
2. Enter a Label of up to eight characters for Input A and press [Enter].
3. Scroll to Off to disable Input A and return to step 1, or on Temp to enable Input A as a live temperature input, and
press [Enter].
4. Enter a temperature value for the Base (low) end of the analog input range and press [Enter].
5. Enter a temperature value for the Full (high) end of the analog input range and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.
DigitalFlow™ XMT868i Programming Manual33
Chapter 1. Programming Site Data
1.4.2bSetting Up Option Cards (cont.)
Option Card Alarm Relays
Complete the following steps to set up the alarm relays of an option card installed in Slot 1 (refer to Figure 32 on
page 81):
1. Scroll to the desired Output (A, B, C or D), and press [Enter].
Note:The setup of Alarm A is used here as an example. Identical procedures would be used to set up the additional
alarms.
2. Scroll to Off to disable Output A and return to step 1, or on High, Low or Fault to select the desired alarm type.
Press [Enter].
3. Scroll to No for standard alarm operation, or on Ye s for Failsafe operation and press [Enter]. (See Chapter 1,
Installation, of the Startup Guide for failsafe wiring instructions.)
For a 1-Channel meter, proceed to step 5.
4. Scroll to the desired Channel option, and press [Enter]. See Table 20 for a description of the channel options
available.
Table 20: Channel Options
OptionDescription
CH1Channel 1
CH2Channel 2
SUMCH1+CH2
DIFCH1-CH2
AVE(CH1+CH2)/2
Note:If
FAULT was selected as the alarm type, the next two prompts do not appear. You have completed entering data
in this option.
5. Scroll to the desired Measurement Name, and press [Enter]. (Table 19 on page 30 provides a description of the
available parameters.)
Note:The measurement units that appear in these prompts are those selected in the Global-SYSTM menu earlier in this
section.
6. Enter a value for the Trigger Point of the alarm and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.
34DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
Option Card Totalizer Outputs
This type of output issues one pulse per selected volume of flow. The meter produces a pulse each time the
programmed amount of flow passes through the pipe. Complete the following steps to set up the totalizer outputs of an
option card installed in Slot 1:
1. Scroll to the desired Output (A, B, C or D), and press [Enter].
Note:The set up of output A is used as an example. Identical procedures would be used to set up the other outputs.
2. Scroll to Off to disable Output A and return to step 1, or on TTLZR to set up Output A as a totalizer output. Press
[Enter].
For a 1-Channel meter, skip to step 4.
3. Scroll to the desired Channel option, and press [Enter]. See Table 20 on page 34 for a description of the channel
options available at the above prompt.
4. Scroll to the desired Measurement Name, and press [Enter]. (Table 19 on page 30 provides a description of the
available parameters.)
Note:The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in this
section.
5. Enter a value between 50 sec and 500,000 sec for Min. Pulse On Time (the frequency of the totalizer pulses) and
press [Enter].
Note:A complete pulse consists of equal amounts of ON and OFF times. Choose a value that is compatible with the
counter to be used.
6. Enter a value for the number of measurement units represented by each pulse, and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.
DigitalFlow™ XMT868i Programming Manual35
Chapter 1. Programming Site Data
Option Card Frequency Outputs
This type of output produces a frequency pulse that is proportional to the output measurement. Complete the following
steps to set up the frequency outputs of an option card installed in Slot 1 (see Figure 32 on page 81):
1. Scroll to the desired Output (A, B, C or D), and press [Enter].
Note:The set up of Output A is used as an example. Identical procedures would be used to set up the other outputs.
2. Scroll to Off to disable Output A and return to step 1, or on Frequency to set up Output A as a frequency output.
Press [Enter].
For a 1-Channel meter, skip to step 4.
3. Scroll to the desired Channel option. See Table 20 on page 34 for a description of the channel options available at
the above prompt.
4. Scroll to the desired Measurement Name. (Table 19 on page 30 provides a description of the available parameters.)
Note:The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in this
section.
5. Enter a flow rate value for the Base (low) end of the frequency output range and press [Enter].
6. Enter a flow rate value for the Full (high) end of the frequency output range and press [Enter].
7. Enter a value between 1 and 10,000 for the Full Scale Frequency and press [Enter].
After responding to the above prompt, the meter returns to step 1. Press [Escape]:
• once to return to the Global I/O option:
• twice to return to the Global PROGRAM option:
• three times to return to the PROG menu:
• four times to return to the Keypad Program menu:
• five times to exit the Keypad Program and return to the XMT868i display.
36DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.3Entering Communications Data
The Model XMT868i flowmeter is equipped with an RS232 or an RS485 serial interface. MODBUS (RS485),
MODBUS/TCP, Ethernet and Foundation-Bus options are available. When the MODBUS, MODBUS/TCP or
Foundation Field Bus options are present, the XMT868i may also have the standard RS232/RS485 serial interface.
When the Ethernet option is present, the RS232/RS485 serial interface is not present.
Note:To set Fieldbus communications parameters, see Appendix E, Foundation Fieldbus Communications.
The serial port is used to transmit stored data and displayed readings to a personal computer by connecting the meter’s
serial interface to the serial port of the PC. In addition, the Model XMT868i can receive and execute remote
commands, using PanaView software, using this link.
Use the COMM option to set the communications port and MODBUS parameters. While following the programming
instructions, refer to Figure 31 on page 80.
1.4.3aSetting Up the Serial Port
1. To enter this submenu from the Global PROGRAM menu, scroll to COMM and press [Enter].
2. Enter a Meter Address number between 1 and 254 and press [Enter]. The default number is 1.
A meter address is only necessary for communication with the GE PanaView software. See the PanaView User’s
Manual for more information.
IMPORTANT: If the meter address or baud rate is changed, communication with PanaView must be re-established with
the new address number.
3. Scroll to the desired Baud Rate (300 to 19,200) and press [Enter].
IMPORTANT: The Ethernet option requires that the meter’s baud rate is set to the same rate as the Ethernet option. The
factory default is 9600 baud. If a different baud rate is desired, first change the meter’s rate, and then
change the Ethernet’s baud rate setting by using the Device Discovery utility.
• If you have the RS485 MODBUS option, proceed to step 4 on the next page.
• If you have the standard RS232 serial interface or optional Ethernet interface, the meter exits the COMM submenu
and returns to the
Global PROGRAM prompt.
DigitalFlow™ XMT868i Programming Manual37
Chapter 1. Programming Site Data
1.4.3bSetting Up MODBUS (RS485) and MODBUS/TCP Communications
When equipped with an optional MODBUS output card, the XMT868i can transmit flow data serially to a flow
computer, or SCADA, using a Gould-type RTU protocol. Thus only the MODBUS function command, 3 (read multiple
registers), 6 (write multiple registers) is valid. The XMT868i uses the following format for the data exchange:
•The Send command (initiated by the host flow computer or controller) comes in the form:
Use the steps below to configure MODBUS communications:
IMPORTANT: The XMT868i MODBUS communications settings chosen in the next four steps must match those of the
MODBUS control system.
4. Scroll to the desired MODBUS Baud Rate and press [Enter].
IMPORTANT: The MODBUS./TCP interface requires a baud rate of 9600.
5. Scroll to the MODBUS Parity and press [Enter].
6. Scroll to the MODBUS Stop Bits and press [Enter].
7. Enter a MODBUS Meter Address (between 1 and 254) and press [Enter]. (The default address is 1.)
The meter returns to the Global PROGRAM prompt. Press [Escape] twice to return to the Keypad Program, and a third
time to return to the meter display.
IMPORTANT: You must reboot the XMT868i to load the new settings.
Refer to the section Requesting Parameters Using MODBUS on page 40 to retrieve data from the XMT868i using
MODBUS.
38DigitalFlow™ XMT868i Programming Manual
Chapter 1. Programming Site Data
1.4.3cSetting Up a MODBUS/TCP Connection
A modified XMT868i can use the MODBUS/TCP interface to communicate with an internal network. An optional
MODBUS/TCP card with a unique MAC (IP) address (installed only in slot 2) includes an RJ45 connector. To connect
the MODBUS/TCP-enabled XMT868i to the network, insert the jack of an RJ45 cable into the RJ45 connector, route
the cable through a conduit hole with an appropriate cable clamp, and wire the other end of the cable to the LAN
according to the manufacturer’s instructions.
To establish MODBUS/TCP communications with the XMT868i, you must install the Ruiping software utility
(available with your modified XMT868i) on a PC connected to the LAN. Once installed and running, the software
displays all Ethernet devices currently connected to the subnet. You can identify the XMT868i by its MAC address,
supplied as part of customer documentation. The default TCP Port is 502.
The default IP addressing for the XMT868i is DHCP (dynamic). If you need to assign a static IP address to the
XMT868i, complete the following steps:
IMPORTANT: To assign a static IP address, the XMT868i must be connected to an Ethernet network with a dynamic
address assignment mechanism (DHCP) first.
1. Run Ruiping software, and identify the XMT868i’s current IP address.
Note:“Ruiping-e” will return the found IP addresses as well as the MAC address.
2. From a DOS command prompt, type: telnet “ip address” 10000
3. The “Welcome to Net+Works Configuration Utility” opens. Type in the login name and password.
a. Login: type root.
b. Password: type Netsilicon.
4. In the Main Menu:
a. Select IP Parameters (1). Enter the desired IP changes and return to the Main Menu.
b. Select Enable DHCP Client (3). Select N (2) to disable, then select 4 for Main Menu.
c. At the Main Menu, select 4 for Quit. Power cycle the XMT868i (i.e., turn the meter off and on) to apply the
changes.
Refer to the section Requesting Parameters Using MODBUS on page 40 to retrieve data from the XMT868i using
MODBUS.
DigitalFlow™ XMT868i Programming Manual39
Chapter 1. Programming Site Data
1.4.3dRequesting Parameters Using MODBUS
To request specific parameters from the XMT868i using the MODBUS, the control system must access the appropriate
register number, as shown in Table 21. Only registers 1–84 are available with the XMT868i for MODBUS
communications. Registers 508–512 are used by the XMT868i to store the MODBUS parameters.
:
Table 21: MODBUS Registers
MODBUS
Reg #
1
2CH1 Velocityft/s or m/s24 (32 bit integer)
4CH1 VolumetricVOL_U--4 (IEEE 32 bit)
6CH1 +TotalsTOT_URegister 104 (32 bit integer)
8CH1 -TotalsTOT_URegister 104 (32 bit integer)
10CH1 #T Digitsnone02 (16 bit integer)
11CH1 Totalizer Timesec24 (32 bit integer)
13
14CH 1 SSUPnone14 (32 bit integer)
16CH 1 SSDNnone14 (32 bit integer)
18CH 1 SNDSPft/s or m/s04 (32 bit integer)
20CH 2 Velocityft/s or m/s24 (32 bit integer)
DPR Hex
AddrDescriptionUnits
1
Clear Totalizers
2
CH1 Error Value
Scaling
(decimal places)Size in Bytes
none--2 (16 bit signed)
none02 (16 bit integer)
22CH 2 VolumetricVOL_U--4 (IEEE 32 bit)
24CH 2 +TotalsTOT_URegister 284 (32 bit integer)
26CH 2 -TotalsTOT_URegister 284 (32 bit integer)
28CH2 # T Digitsnone02 (16 bit integer)
29CH2 Totalizer Timesec24 (32 bit integer)
31
2
CH2 Error Value
none02 (16 bit integer)
32CH 2 SSUPnone14 (32 bit integer)
34CH 2 SSDNnone14 (32 bit integer)
36CH 2 SNDSPft/s or m/s04 (32 bit integer)
38
40
42
3
AVG Velocity
3
AVG Volumetric
3
AVG+Totals
ft/s or m/s24 (32 bit integer)
VOL_U--4 (IEEE 32 bit)
TOT_URegister 464 (32 bit integer)
40DigitalFlow™ XMT868i Programming Manual
MODBUS
Reg #
Table 21: MODBUS Registers (cont.)
DPR Hex
AddrDescriptionUnits
Chapter 1. Programming Site Data
Scaling
(decimal places)Size in Bytes
44
3
AVG-Totals
TOT_URegister 464 (32 bit integer)
46AVG #T Digitsnone02 (16 bit integer)
47
49
50
52
54
3
AVG Totalizer Time
4
AVG Error Value
3
AVG SSUP
3
AVG SSDN
3
AVG SNDSP
sec24 (32 bit integer)
none02 (16 bit integer)
none14 (32 bit integer)
none14 (32 bit integer)
ft/s or m/s04 (32 bit integer)
56CH 1 PowerPower_u--4 (IEEE 32 bit)
58CH 1 +EnergyEnergy_uRegister 624 (32 bit integer)
60CH 1 -EnergyEnergy_uRegister 624 (32 bit integer)
62CH 1 # Energy Digitsnone02 (16 bit integer)
63CH 1 TempS°F or °C24 (32 bit integer)
65CH 1 TempR°F or °C24 (32 bit integer)
67CH 1 TS-TR°F or °C24 (32 bit integer)
69CH 1 DELTHBtu/lb or J/gm24 (32 bit integer)
71CH 2 PowerPower_u--4 (IEEE 32 bit)
73CH 2 +EnergyEnergy_uRegister 774 (32 bit integer)
75CH 2 -EnergyEnergy_uRegister 774 (32 bit integer)
77CH 2 # Energy Digitsnone02 (16 bit integer)
78CH 2 TempS°F or °C24 (32 bit integer)
80CH 2 TempR°F or °C24 (32 bit integer)
82CH 2 TS-TR°F or °C24 (32 bit integer)
84CH 2 DELTHBtu/lb or J/gm24 (32 bit integer)
5083F6
5093F8
5103FA
5
MODBUS baud rate
6
MODBUS parity
7
MODBUS stop bits
none02 (16 bit integer)
none02 (16 bit integer)
none02 (16 bit integer)
5113FCMODBUS meter addrnone02 (16 bit integer)
5123FERESERVEDnone------
DigitalFlow™ XMT868i Programming Manual41
Chapter 1. Programming Site Data
1.4.3dRequesting Parameters Using MODBUS (cont.)
Notes:
1. Clear Totalizers: flag from the 8051 to clear totalizers in the 68332 memory.
2. Error Value: see table in XMT868i manual for error codes
3. Average:
average of channel 1 and channel 2 if both channels out of error,
channel 1 value if channel 2 is in error,
channel 2 value if channel 1 is in error,
zero if both channels in error.
4. Average Error Status:
0 = both in error
1 = chan 2 in error,
2 = chan 1 in error,
3 = both ok
Registers are written if corresponding functions are actuated by the user. Registers for unactuated functions are
initialized to zero at startup.
1.5Exiting the User Program
After completing the COMM option, the XMT868i returns to the Global PROGRAM prompt. Press [Escape] twice to return
to the Keypad Program, and a third time to return to the meter display. Then proceed to Chapter 3, Operation, of the Startup Guide for instructions on taking measurements, or refer to the appropriate chapters of this manual for detailed
instructions on using the other features of the XMT868i flow transmitter.
42DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
Chapter 2.Displaying Data
2.1Introduction
This chapter explains how to display measurement data using either of the available methods:
•LCD Display - show data on the built-in display
•PanaView - display data on a computer terminal using the optional PanaView software
2.2Displaying Data with the LCD
When equipped with the Liquid Crystal Display (LCD), the Model XMT868i may be programmed to display up to four
variables in sequence. In addition, the contrast of the LCD may be adjusted for optimum viewing. Proceed to the
appropriate section for instructions.
2.2.1Adjusting LCD Contrast
The LCD contrast may be adjusted through the Keypad Program to suit individual needs.
1. Press [Escape]. [Enter], [Escape].
2. In the Keypad Program window, scroll to CNTRS and press [Enter].
3. Scroll to DARKN or LITEN and press [Enter] until the screen shows the desired contrast.
4. When you have achieved the desired contrast, scroll to STORE to retain the setting, or to ABORT to leave the menu
without changing the setting. In either case, the XMT868i returns to the Keypad Program.
5. Then press [Escape] to return to the display screen.
DigitalFlow™ XMT868i Programming Manual43
Chapter 2. Displaying Data
2.2.2Programming the LCD
Note:When you first initialize the XMT868i, the number of LCD parameters is set to OFF. You must program the
LCD to display any measured parameters.
Through the Keypad Program, you can program the LCD display to display up to four variables in sequence. Complete
the following steps to program the LCD display:
1. Power up the XMT868i and wait until it has initialized.
2. Press [Escape], [Enter], [Escape].
3. In the Keypad Program window, scroll to PROG and press [Enter].
4. In the PROG menu, scroll to GLOBL and press [Enter].
5. Scroll to I/O and press [Enter].
6. Scroll to LCD and press [Enter].
7. The window now asks for the # of LCD Parameters. Scroll to the desired number (from OFF through 1-4 and KEY)
and press [Enter].
The OFF setting switches the measurement display off, while the KEY setting enables users to change the measurement
display using the arrow keys, without accessing the Keypad Program. If you select KEY:
•To view a different parameter, press the [] and [] keys to scroll through the various parameters.
•To scroll through the channel options on a two-channel XMT868i, press the [] and [] keys until you have
reached the desired option.
For a 1-Channel XMT868i, the data for Channel 1 is displayed automatically, and you can skip to step 9. However, for
a 2-Channel meter, the channel data to be displayed must be specified at the following prompt.
44DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
2.2.2Programming the LCD (cont.)
8. Scroll to the desired Channel option, as listed in Table 22.
Table 22: Channel Options
OptionDescription
CH1Channel 1
CH2Channel 2
SUMCH1+CH2
DIFCH1-CH2
AVE(CH1+CH2)/2
9. For each channel, select the desired Measurement Parameter, as shown in Table 19 on page 30.
Note:The measurement units that appear in these prompts are those selected in the
GLOBL-SYSTM menu earlier in this
section. Also, when differences in one channel's programming invalidate an output previously chosen for the
other, the measurement defaults to the nearest selectable item in the parameter list.
The previous two prompts repeat until all of the specified # of LCD Parameters have been set up. When all of the
display parameters have been set up, the meter returns to the Global I/O window. To leave the Keypad Program, press
the [Escape] key three times.
After leaving the Keypad Program, the XMT868i will reset itself and will begin to display the parameters specified in
this section. If more than one parameter was set up, each of the parameters will be displayed in sequence, with a pause
of several seconds between display changes.
DigitalFlow™ XMT868i Programming Manual45
Chapter 2. Displaying Data
2.3Displaying Data on a Computer Terminal
The flow rate data collected by the XMT868i may be displayed in various formats on a remote computer terminal using
the meter’s RS232/RS485 serial or Ethernet port. This requires the use of the optional PanaView software. Refer to the
instructions below to display data using PanaView.
2.3.1Preparing for PanaView Programming
Before you attempt to communicate with the XMT868i, be sure you have linked your PC to the XMT868i using its
communication interface. For details on wiring the interface, refer to Wiring the Serial Port in the Startup Guide, and to
the document EIA-RS Serial Communications (916-054). You must also install PanaView, as discussed in the
PanaView User’s Manual (910-211), and in Appendix C, Programming the XMT868i Using PanaView™.
1. Power up the XMT868i and wait until it has initialized.
2. Launch PanaView and wait until it initializes the XMT868i and displays the basic PanaView window.
3. As described in Appendix C, open the Meter Browser window and highlight the meter to be used.
At this time, any of PanaView’s comprehensive data handling capabilities may be used. These include:
•collecting and displaying live data in text format
•collecting and displaying live data in graphical format
•collecting and logging live data
•displaying a log file in text format
•displaying a log file in graphical format
To access PanaView’s data handling options, pull down the Output menu, as shown in Figure 2.
Figure 2: Output Menu
46DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
2.3.1Displaying Data on a Computer Terminal (cont.)
Proceed directly to the appropriate section for instructions on either of the following output options:
•Text Display output, discussed below
•Graphing output on page 49
Note:For information on the Logging output, see Chapter 3, Logging Data.
2.3.2The Text Display Output
To collect data from the instrument and display it on a text screen, complete the following steps:
1. Complete Steps 1-3 on the previous page.
2. Pull down the Output menu and click on the Text Display option (see Figure 2 on page 46).
Note:The Text Display window that appears after Step 2 is actually stacked on top of any previously opened
windows (such as the Meter Browser window).
3. Using the Window menu, arrange the open windows in the desired format. For this discussion, Figure 3 shows the
Text Display window in its maximized (full-screen) size.
Figure 3: Text Display Window
Note:Figure 3 has been compressed both vertically and horizontally for improved clarity.
4. The left pane of the Text Display window contains the standard PanaView network tree. Expand this tree as far as
necessary, and double click on the desired process parameter to display it in the right pane of the window.
DigitalFlow™ XMT868i Programming Manual47
Chapter 2. Displaying Data
2.3.3The Text Display Output (cont.)
5. Before actual data values can be displayed in the text pane, activate one of the following data collection modes (see
Figure 3 on page 47):
• Click on the [Get Once] option button at the bottom of the right pane in the Text Display window. The current
value of the selected process parameter, as specified in the PanaView network tree, is displayed in the right pane
of the Text Display window, or
• Enter an interval in the text box at the bottom of the right pane in the Text Display window, or check the Max.
Comm Rate box to collect readings as fast as the system allows (1 sec). Then, click on the button to begin collecting data for display in the right pane of the Text Display window.
Note:Any value entered in the Interval text box is overridden if the Max. Comm Rate box is checked.
6. If the [Continuous] option was selected in Step 5 above, click on the [Stop] option button, which has replaced the
original [Continuous] option button, to terminate data collection.
The Text Display window may be left open while other tasks are performed, or it may be closed by clicking on the
lower [X] control button at the far right of the menu bar.
[Continuous] option
IMPORTANT:If you click on the upper [X] control button at the far right of the PanaView title bar, you will exit
PanaView completely.
2.3.3aDisplaying Multiple Process Parameters
The procedure for displaying a single process parameter in a text screen may be repeated to simultaneously display
multiple process parameters. To do so, proceed as follows:
1. Display the first process parameter in a text screen, as described in the previous section.
2. Repeat Step 1 for any desired additional process parameters, by double clicking on them in the PanaView network
tree. PanaView automatically tiles the multiple text screens in the right pane of the Text Display window.
3. As in any standard Windows application, the multiple text screens may be resized by dragging their borders. Also,
the individual panes within a parameter’s text screen may be resized by dragging the borders within that text
screen.
4. To close an open text screen, right click anywhere within that screen (except in the title bar or the error section) and
click on the [Remove] option that pops up in the context menu.
Note:After resizing or removing any of the multiple text screens, the default tiled layout may be restored by opening
the “Window” menu (see the PanaView User’s Manual) and clicking on the “Tile Output Displays” option.
48DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
2.3.3bDisplaying Multiple Text Windows
The procedures for displaying one or more process parameters in a single Text Display window may be repeated to
open multiple Text Display windows. To do so, proceed as follows:
1. To open another Text Display window, repeat the steps on page 2-5 and 2-6.
2. To display the desired process parameter(s) in the new window, repeat Steps 1-4 above.
3. Arrange the multiple Text Display windows as desired using the Window menu.
2.3.4The Graphing Output
To collect data from the instrument and display it graphically in a new format, complete the following steps:
2.3.4aSetting Up a Graph
1. Perform Steps 1-3 on page 46.
2. Pull down the Output menu and click on the Graphing - New option (see Figure 2 on page 46).
Note:The Graph window is stacked on top of any previously opened windows (such as the Meter Browser window).
3. Using the Window menu, arrange the open windows in the desired format. For this discussion, Figure 4 shows the
Graph window in its maximized (full-screen) size.
Figure 4: Graph Window
The Graph window shown above consists of three panes:
•left pane - PanaView network tree
•top right pane - graph with default chart style
•bottom right pane - graph controls or text display
Note:The three panes of the Graph window may be resized by simply dragging their borders.
DigitalFlow™ XMT868i Programming Manual49
Chapter 2. Displaying Data
2.3.4bGraphing Data
The PanaView network tree has already been described, and the display pane shows only a graph in the default style
with no data points. The bottom right pane enables you to set up the graph. To begin graphing your data, proceed as
follows:
1. Expand the network tree as far as necessary and double click on the process parameter to be graphed. The grid for
the selected graph appears in the display pane, with Time as the x-axis and the parameter Value as the y-axis. In
addition, a legend appears above the graph, which indicates the name, data point shape and line color for that
parameter.
2. If desired, a second parameter may be added to the graph by repeating Step 1. In this case, the parameter “Value” is
shown as a right y-axis (Y2).
Note:Although only two parameters may be graphed for any one meter channel, the same two parameters may also
be graphed for any other active meter channels.
3. To begin plotting data for the selected parameter(s), enter a time Interval for data sampling in the text box provided
(the default value is 5 seconds).
4. Click on the [Plot Continuous] control button to begin graphing the selected process parameter(s) as a function of
time, at the interval specified in Step 3.
5. While PanaView is graphing the data, the following actions may be taken from the control pane:
• change the sampling interval by entering a new value
• click on the [Zoom Out] control button to increase the size of the intervals shown on the x-axis
• click on the [Zoom In] control button to decrease the size of the intervals shown on the x-axis
Note:The zoom buttons may be clicked multiple times to enhance the effect.
• click on the Text Display tab to view the data being graphed in the text format described in the previous section
When you have finished graphing your data, click on the
Continuous] control button.
For further information on setting up graph properties, refer to Chapter 5, Displaying Data, of the PanaView User’s
Manual.
[Stop] control button, which has replaced the original [Plot
50DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
2.3.5Displaying Transducer Signals
Along with flow rate data, PanaView enables XMT868i users to read and plot transducer signals from the XMT868i.
1. From the New Meter Browser, highlight the XMT868i.
2. Right-click on the highlighted XMT868i and click on the Properties option, as shown in Figure 5.
Figure 5: Properties Option in the New Meter Browser
The Properties window opens, as shown in Figure 6.
Figure 6: Properties Window
DigitalFlow™ XMT868i Programming Manual51
Chapter 2. Displaying Data
2.3.5aReading Transducer Signals
3. To read a signal from the meter, click on the [Read Signals] button. (If the meter is a multi-channel instrument, open
the Channel drop-down menu and click on the desired channel.) After a moment, the Properties window appears
similar to Figure 7.
Figure 7: Active Signal Options
4. To select a different signal type, open the signal menu at the right (shown here with Raw Upstream highlighted)
and click on the desired signal.
2.3.5bPlotting Transducer Signals
To plot the selected signal, click on [Plot]. A graphical window opens, as shown in Figure 8.
Figure 8: Signal Graph Window
52DigitalFlow™ XMT868i Programming Manual
Chapter 2. Displaying Data
2.3.5cSaving Transducer Signals
To save the raw signal, click on [Save]. A window opens similar to Figure 9. Enter the desired name, and click on [Save]
to save the signal as a text file.
Figure 9: Save As Window
DigitalFlow™ XMT868i Programming Manual53
Chapter 2. Displaying Data
[no content intended for this page]
54DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
Chapter 3.Logging Data
3.1Introduction
The XMT868i offers two options for logging data:
•If the Model XMT868i flowmeter is equipped with an optional data logging card in Slot 2, flow rate data can be
recorded and stored in this memory in the form of a log file. After the specified data has been logged, the log file
may be uploaded to PanaView using the RS232 link.
•PanaView can also create PC log files for storage on the PC’s hard drive.
This chapter explains how to use the Model XMT868i’s data logging capability with PanaView.
3.2The Data Logging Option Card
In order to log data, the Model XMT868i must be fitted with an optional data logging option card in Slot 2. See Chapter
4, Parts Replacement, of the Service Manual for instructions on installing the card.
3.3Logging Using PanaView
PanaView is capable of creating and viewing log files of the following types:
•meter logs - log files stored in the XMT868i memory, as discussed above
• PC logs - log files stored on your PC’s hard drive (page 61)
To create or view logs of the above types, proceed to the appropriate section of this chapter.
DigitalFlow™ XMT868i Programming Manual55
Chapter 3. Logging Data
3.4Creating Meter Logs
To create Figure 11, a new meter log, complete the following steps:
1. From the New Meter Browser in PanaView, expand the network tree and open the Edit Functions option (described
on page C-6 of Appendix C, Programming the XMT868i Using PanaView). The menu will appear similar to
Figure 10.
Figure 10: Edit Functions Menu for XMT868i with Logging Card
2. Double-click on the Log Edit Menu, as shown in Figure 10. The window appears similar to Figure 11.
Figure 11: The Log Edit Menu Window
56DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.4.1Creating a Standard Meter Log
3. To create a standard meter log, complete the following steps:
a. Double-click on the Create Standard Log option.The window now appears similar to Figure 12.
Figure 12: Log Location in the Create Standard Log Option
b. PanaView first asks for the Log Location, either flash or non-volatile RAM.Double-click on the desired
location.
c.Now type the desired Log Name in the right window, and click [Next Item].
d. Type the desired Log Message in the right window, and click [Next Item].
e.PanaView now asks for the Number of Variables you wish to log. Type in the desired number (from 1 to 6), and
click [Next Item].
f.Double-click on the desired Channel for the first variable.
g. Scroll to and double-click on the desired Measurement to be logged.
h. Repeat steps f and g for the specified number of variables.
•If you select a forward or reverse total, PanaView asks if you want to Reset Totals to 0? Double-click on
Yes or No.
i.PanaView indicates a given StartTime. Double-click on OK to agree to the given time, on Now to start the log
immediately, or on Edit to change the starting time.
•If you select Edit, PanaView asks first for the desired Hour, then for the Minute, and finally for the Second.
For each parameter, type the desired number in the right window, and click
[Next Item] to step through the parameters).
[Next Item] (or simply click
DigitalFlow™ XMT868i Programming Manual57
Chapter 3. Logging Data
3.4.1Creating a Standard Meter Log (cont.)
j.PanaView indicates a given StartDate. Double-click on OK to agree to the given day, on Today to start the log
that day, or on Edit to change the starting date.
•If you select Edit, PanaView asks first for the desired Year, then for the Month, and finally for the Day. For
each parameter, type the desired number in the right window, (or scroll to and click on the desired month)
and click
k. PanaView indicates a given End Time. Double-click on OK to agree to the given time, on Now to start the log
immediately, or on Timed to indicate a specific duration.
[Next Item] (or simply click [Next Item] to step through the parameters).
•If you select Edit, PanaView asks first for the desired Hour, then for the Minute, and finally for the Second.
For each parameter, type the desired number in the right window, and click
[Next Item] to step through the parameters).
[Next Item] (or simply click
•If you select Timed, PanaView asks for the Log Time (duration). Scroll to and double-click on the desired
length of time (from 10 minutes to 24 hours.) Then go to step m.
l.PanaView indicates a given End Date. Double-click on OK to agree to the given day, on Today to start the log
that day, or on Edit to change the starting date.
•If you select Edit, PanaView asks first for the desired Year, then for the Month, and finally for the Day. For
each parameter, type the desired number in the right window, (or scroll to and click on the desired month)
and click
m. Scroll to and double-click on the desired Time Increment (from 5 seconds to 24 hours). PanaView indicates that
the log has been created, and returns to the Log Edit Menu.
If an erroneous value has been entered, PanaView displays a Log Creation Error message. Otherwise, the program
indicates that the log has been created, and returns to the Log Edit Menu.
[Next Item] (or simply click [Next Item] to step through the parameters).
58DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.4.2Creating an Error Log
To create an error meter log, complete the following steps:
1. Double-click on the Create Error Log option.
a. PanaView first asks for the Log Location, either flash or non-volatile RAM. Double-click on the desired
location.
b. Now type the desired Log Name in the right window, and click [Next Item].
c. Type the desired Log Message in the right window, and click [Next Item].
d. PanaView now asks for the Number of Variables you wish to log. Type in the desired number (from 1 to 6), and
click [Next Item].
e.Double-click on the desired Channel for the first variable.
f.Scroll to and double-click on the desired Measurement.
g. Repeat steps e and f for the specified number of variables.
•If you select a forward or reverse total, PanaView asks if you want to Reset Totals to 0? Double-click on
Yes or No.
h. PanaView asks, Is Log Circular? Double-click on Yes or No.
i.PanaView indicates a given StartTime. Double-click on OK to agree to the given time, on Now to start the log
immediately, or on Edit to change the starting time.
•If you select Edit, PanaView asks first for the desired Hour, then for the Minute, and finally for the Second.
For each parameter, type the desired number in the right window, and click
[Next Item] to step through the parameters).
j.PanaView indicates a given StartDate. Double-click on OK to agree to the given day, on Today to start the log
that day, or on Edit to change the starting date.
[Next Item] (or simply click
•If you select Edit, PanaView asks first for the desired Year, then for the Month, and finally for the Day. For
each parameter, type the desired number in the right window, (or scroll to and click on the desired month)
and click
[Next Item] (or simply click [Next Item] to step through the parameters).
DigitalFlow™ XMT868i Programming Manual59
Chapter 3. Logging Data
3.4.2Creating an Error Log (cont.)
Note:If you have selected a circular log, PanaView now indicates that the log has been created, and returns to the
Log Edit Menu.
k. PanaView indicates a given End Time. Double-click on OK to agree to the given time, on Now to start the log
immediately, or on Timed to indicate a specific duration.
•If you select Edit, PanaView asks first for the desired Hour, then for the Minute, and finally for the Second.
For each parameter, type the desired number in the right window, and click
[Next Item] to step through the parameters).
[Next Item] (or simply click
•If you select Timed, PanaView asks for the Log Time (duration). Scroll to and double-click on the desired
length of time (from 10 minutes to 24 hours.) Then go to step m.
l.PanaView indicates a given End Date. Double-click on OK to agree to the given day, on Today to start the log
that day, or on Edit to change the starting date.
•If you select Edit, PanaView asks first for the desired Year, then for the Month, and finally for the Day. For
each parameter, type the desired number in the right window, (or scroll to and click on the desired month)
and click
[Next Item] (or simply click [Next Item] to step through the parameters).
m. Scroll to and double-click on the desired Time Increment (from 5 seconds to 24 hours). PanaView indicates that
the log has been created, and returns to the Log Edit Menu.
If an erroneous value has been entered, PanaView displays a Log Creation Error message. Otherwise, the program
indicates that the log has been created, and returns to the Log Edit Menu.
3.4.3Checking Log Memory
To check on the amount of log memory available, double-click on the Memory option in the Log Edit Menu. The right
window indicates the number of bytes available in both NVR and Flash memory.
3.4.4Stop Logging
To manually terminate logging, double-click on the Stop Logging option in the Log Edit Menu. PanaView displays the
name of the currently active logs. Double-click on your choice, and then click on Yes when the program asks if you
wish to stop logging.
60DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.5Creating PC Logs
PC logs are created in a different menu from the meter logs described in the last section. To create a new PC log:
1. Open the Output menu (see Figure 13), and click on the Logging option.
Figure 13: Selections in the Logging Option
You have two choices:
• Click on the New PC Log option and proceed to step 5, or
• Click on the Log Browser option. The dialog box shown in Figure 14 appears.
Figure 14: Computer Selection
Note:If no log files have yet been created or if a log file has not yet been selected from the list, all of the option
buttons (except for
2. Double click on the computer name to advance to the dialog box shown in Figure 15 on page 62.
[Cancel]) in the Log Browser dialog box are unavailable.
DigitalFlow™ XMT868i Programming Manual61
Chapter 3. Logging Data
3.5 Creating PC Logs (cont.)
Figure 15: Log Type Selection
3. Double click on PC Logs to advance to the dialog box shown in Figure 16.
Figure 16: Create a PC Log
4. At the dialog box shown in Figure 16, click on the
[Create] option button to create a new meter log.
5. Enter a log file name (“Sample” in this case) in the dialog box shown in Figure 17. Click on the [OK] option button.
Figure 17: Log Name Dialog Box
62DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.5 Creating PC Logs (cont.)
6. At the resulting PC Log window shown in Figure 18, place a check mark in the text box to create a circular log, or
leave this box empty to create a linear log.
Figure 18: The PC Log Window
7. Enter a sampling interval in the text box shown in Figure 18.
Note:Start Time and Stop Time cannot be specified for PC logs. These options are inactive in the PC Log window.
8. Expand the network tree as far as necessary and double click on the process parameter(s) to be logged. As they are
selected, the parameters are listed in the lower right pane of the PC Log window (make sure that the Log Items tab
in this pane is selected).
Note:To add all the channel parameters to a given log, right-click on the desired channel. A pop-up button, “Add
Group to Log,” appears. Click on the button, and PanaView adds the entire set of channel parameters to the
log.
9. Click on the [Apply] option button to record the log setup information, and then click on the [Start] option button to
begin logging data immediately. (The Start Time automatically appears in the PC Log window.)
DigitalFlow™ XMT868i Programming Manual63
Chapter 3. Logging Data
3.6Viewing Meter Log Files
After one or more meter log files have been created, the logs may be viewed by using the New Meter Browser as
follows:
1. From the New Meter Browser in PanaView, expand the network tree and click on the Meter Logs option. If you
have created one or more logs, the tree will appear similar to Figure 19.
Figure 19: An Expanded Meter Logs Option
2. Double-click on the desired log. A window opens for the individual log, as shown in Figure 20.
Figure 20: Data Window for an Individual Meter Log
•Click on the Log Items tab (see Figure 20) for a list of the parameters specified for logging.
•Click on the Contents tab (see Figure 21 on page 65) for a list of all the data points logged to date for the parameter
currently highlighted on the Log Items tab.
•Click on the [Graph] option button to display a graphical representation of the data logged to date.
64DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.6 Viewing Meter Logs (cont.)
Note:The graph is displayed in its own window, which is opened on top of the Meter Log window. For instructions
on using the Graph Log window, refer to the Graphing Output section in Chapter 2 of the PanaView User’s
Manual.
Remember that only two parameters per channel may be graphed, and that the same two parameters must be graphed
for each channel of a multi-channel graph display. When you have finished viewing the graph, click on the [Close]
option button to close the Graph window and leave the log running.To review log details, click on the Contents tab. The
window now appears similar to Figure 21.
Figure 21: Typical Log Contents
DigitalFlow™ XMT868i Programming Manual65
Chapter 3. Logging Data
3.7Viewing PC Log Files
After one or more PC log files have been created, the logs may be viewed by using PanaView as follows:
1. You can access PC logs in two ways:
• From the New Meter Browser in PanaView, expand the network tree and click on the PC Logs option. If you have
created one or more logs, the tree will appear similar to Figure 22. Double-click on the log name to open the log,
or
• Pull down the Output menu (see Figure 13 on page 61), and click on the Log Browser option. Advance to the PC
Logs dialog box as described in previous sections. A screen similar to that shown in Figure 23 appears.
Figure 22: The Expanded PC Logs Option
Figure 23: Selecting a Log File from the Log Browser
Notice that each of the log files in Figure 22 and Figure 23 has a traffic light icon next to its name. These icons indicate
the current status of the log as follows:
•red light - log has not run to its specified stop time but has been manually stopped
•green light - log is currently running but is not yet complete
66DigitalFlow™ XMT868i Programming Manual
Chapter 3. Logging Data
3.7 Viewing PC Logs (cont.)
2. In the Log Browser, after you highlight the name of the desired log file, the following option buttons become
available:
• [Start] - resumes logging if the log is currently stopped
or
[Stop] - stops logging, if the log is currently running
• [Delete] - permanently deletes the log file
• [Open] - opens the log in the PC Log window
3. Click on one of the option buttons listed in Step 2 above.
You can monitor the progress of an ongoing log in several ways:
•Click on the Log Items tab (see Figure 21 on page 65) for a list of the parameters specified for logging.
•Click on the Contents tab (see Figure 21 on page 65) for a list of all the data points logged to date for the parameter
currently highlighted on the Log Items tab.
•Click on the [Graph] option button to display a graphical representation of the data logged to date.
Note:The graph is displayed in its own window, which is opened on top of the PC Log window. For instructions on
using the Graph Log window, refer to the Graphing Output section in Chapter 2 of the PanaView User’s
Manual.
• Click on the [Refresh] option button to update the information shown on the Contents tab and in the Graph Log
window. Any data logged since the last use of the
Remember that only two parameters per channel may be graphed, and that the same two parameters must be graphed
for each channel of a multi-channel graph display. When you have finished viewing the graph, click on the [Close]
option button to close the Graph window and leave the log running.
4. To terminate the logging process, simply click on the [Stop] option button, which has replaced the original [Start]
option button. (The Stop Time automatically appears in the PC Log window.)
Because a PC log has no specific “Stop Time,” the log will continue to run until it is manually stopped (unless the PC is
turned off or you run out of hard drive space).
[Refresh] option button is added to the list and to the graph.
DigitalFlow™ XMT868i Programming Manual67
Chapter 3. Logging Data
[no content intended for this page]
68DigitalFlow™ XMT868i Programming Manual
Chapter 4. Printing Data
Chapter 4.Printing Data
4.1Data Types for Printing
The Model XMT868i flowmeter has no ability to print any of its data directly. However, any of the data stored in its
memory may be printed using the built-in RS232 communications port, using a computer terminal. In order to use the
capability, the XMT868i must be linked to the computer terminal with the optional PanaView software.
Note:See Chapter 1, Installation, of the Startup Guide for instructions on wiring the RS232 serial port. For
additional information, refer to the EIA-RS Serial Communications Manual (916-054).
After making the hardware connections and installing PanaView software, the following data may be output to a printer
connected to the personal computer:
•live data in numeric or graphical format
•log file in numeric or graphical format
•site file in tabular format
•transducer signal array data in tabular format
For detailed instructions on printing any of the data types listed above, consult the PanaView User’s Manual.
DigitalFlow™ XMT868i Programming Manual69
Chapter 4. Printing Data
[no content intended for this page]
70DigitalFlow™ XMT868i Programming Manual
Chapter 5. Clearing Data
Chapter 5.Clearing Data
5.1Introduction
This chapter explains how to purge totalized measurements, site data and/or log files from the XMT868i’s memory.
Note:For detailed information on creating a log file, see Chapter 3, Logging Data. For detailed information on
programming site data, see Chapter 1, Programming Site Data.
IMPORTANT: None of the clearing procedures described in this chapter can be undone. Be very sure that the exact
consequences of a selected option are thoroughly understood before proceeding.
5.2Clearing the XMT868i’s Memory
If the XMT868i’s available memory becomes nearly full, it may be necessary to purge some or all of the existing data
from memory, before any additional data can be stored. In order to accomplish this task, proceed to the appropriate
section for step-by-step instructions.
Note:Depending on the current configuration of the XMT868i, it may be possible to add additional memory by
installing an option card in Slot 2. See Chapter 4, Parts Replacement, of the Service Manual for details.
5.2.1Clearing Site Data
This feature has not yet been implemented, because the XMT868i currently has no capability to store site files in its
own memory.
DigitalFlow™ XMT868i Programming Manual71
Chapter 5. Clearing Data
5.2.2Clearing Log Files
To clear log files from the XMT868i or from the PC’s memory, complete the following steps:
1. From PanaView, click on the Logging option in the Output menu. Then click on the Log Browser option, as shown
in Figure 24.
Figure 24: Log Browser in the Logging Option
2. Double-click on the computer name to move to the network level.
3. Double-click on PC Logs to access the stored PC logs, or on the communication port name and the meter name to
access the stored meter logs. In either case, the Log Browser appears similar to Figure 25.
Figure 25: Log Browser Window
4. Highlight the log you wish to delete, and click on the [Delete] option button.
72DigitalFlow™ XMT868i Programming Manual
Chapter 5. Clearing Data
5.2.3Clearing the Totalizers
You can clear the XMT868i’s totalizers (i.e., reset specified totalized flow rates to zero) from either the LCD display or
from PanaView. To reset totals from the display, complete the following steps:
1. Press
[Escape], [Enter], [Escape].
2. In the Keypad Program, scroll to RESET and press [Enter].
3. The XMT868i asks Reset Totals? Scroll to NO or YES (2) and press [Enter] at the desired choice.
The XMT868i totalizers are reset to 0, and the meter returns to the Keypad Program. Press [Escape] to return to the data
display.
To reset the totals from PanaView:
1. From the New Meter Browser in PanaView, highlight the XMT868i.
2. Open the Edit menu and select the Properties option, as shown in Figure 26.
Figure 26: Properties Option in the Edit Menu
3. The Properties window appears similar to Figure 27. To clear the XMT868i totalizers, click on the
button. The XMT868i totalizers are reset to 0.
Figure 27: Properties Window
[Clear Totalizers]
DigitalFlow™ XMT868i Programming Manual73
Chapter 5. Clearing Data
[no content intended for this page]
74DigitalFlow™ XMT868i Programming Manual
Appendix A.Menu Maps
Appendix A. Menu Maps
DigitalFlow™ XMT868i Programming Manual75
Appendix A. Menu Maps
[no content intended for this page]
76DigitalFlow™ XMT868i Programming Manual
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual77
[Esc][Enter][Esc]
PROGRESETCALIBCNTRS
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
PIPEI/OSETUPSYSTMACTIV
OffTransit
ENERGY OPTION
CHANNEL LABEL
SITE/CHANNEL MESSAGE
MetricEnglish
Liter/SLiter/MLiter/H
m3/hMm3/dBBL/S
MLiter/Dm3/sm3/m
BBL/MBBL/HMBL/D
VOL DECIMAL DIGITS
TOTALIZER UNITS
0
12
3
TOTAL DECIMAL DIGITS
MetricEnglish
GAL
MGALft^3Mft^3
English
LB
KLBMMLBTONSKGTONNES
/SEC/MIN/HR/DAY
MASS FLOW TIME
Metric
MGDft3/sft3/m
BBL/MBBL/H
GAL/H
Mft3/dBBL/SA-I/M
A-I/DA-F/SA-I/H
OFFON
VOLUMETRIC UNITS
BBL
MBBLAC-INAC-FT
LiterMliterm^3
Mm^3
BBLMBBL
0123
MDOT DECIMAL DIGITS
MetricEnglish
KG
TONNES
MASS DECIMAL DIGITS
0
12
3
MASS TOTALS
MASS FLOW
LB
KLBMMLBTONS
MASS FLOW ON MASS FLOW OFF
KBTU
MMBTU
kWTONS
KcalMcal
KWHrMWHr
MetricEnglish
POWER
0
123
0123
POWER DECIMAL DIGITS
kBTU
MMBTUkWTONSKCal/sKCal/m KCal/hKCal/d
MetricEnglish
ENERGY (TOTAL)
0
123
ENERGY OPTION ONENERGY OPTION OFF
ENERGY OPTION ON
ENERGY OPTION OFF
ENERGY DECIMAL DIGITS
HEATING or COOLING
FLOW MEASUREMENT
COOL
HEAT
RTNSPPLY
ZERO CUTOFF
FIXED TEMP.
FIXED
SLOT1
ABCD
See Figure A-2
MCal/s MCal/m MCal/h MCal/d
KWMW
See Figure A-3
Energy Option ON
TEMP INPUT SUPPLY
ANALOG IN
FIXED TEMP.
FIXED
SLOT1
ABCD
TEMP INPUT RETURN
ANALOG IN
(See Service Manual)
NO
YES
Reset Totals
Energy Option Off
DARKN
LITEN
DARKNLITEN
GAL/SGAL/Mft3/h
A-I/S
ft3/d
BBL/DMBL/D
A-F/MA-F/HA-F/D
Figure 28: Channel X - ACTIV, SYSTM and I/O Menu Map
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual78
[Esc][Enter][Esc]
PROGRESETCALIBCONTRAST
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
PIPEI/OSETUPSYSTMACTIV
RAYLSHEARWETTD
STDSPEC
CuAlBRASSCuNiGLASS PLSTC OTHER
See Figure A-1See Figure A-1
TRANSDUCER NUMBER
SPECIAL TRANSDUCER
WEDGE TYPE
1MHz2MHz5MHz500k
FREQUENCY
(RAYL/SHEAR)
STEELIRON
WEDGE SOUNDSPEED
WEDGE ANGLE
PIPE MATERIAL
10%Ni
PYREXFLINTNYLON POLYE POLYPCAST
30%Ni
DUCTCARBNSS
PVCACRYL
PIPE SOUNDSPEED
feet
cir.in
cir.ft
m
cir.mmcir.minch
mm
PIPE OD
(Metric)
(English)
feet
cir.in
cir.ft
m
cir.mmcir.minch
mm
PIPE OD
(Metric)
(English)
(WETTD)
(Clamp-on Type)(Wetted Type)
feet
mmm
inch
PIPE WALL
PATH LENGTH P
(English)
(Metric)
feet
mmm
inch
AXIAL LENGTH L
(English)
(Metric)
PIPE WALL
LINING
YES
NO
TARPYREX ASBES
RUBBR
TEFLN OTHERMORTR
LINING MATERIAL
LINING SOUNDSPEED
LINING THICKNESS
YESNO
TRACKING WINDOWS?
WATER
MIXEDOTHER
WATER TEMPERATUREFLUID SOUNDSPEEDFLUID SOUNDSPEED
Percentage of Water
W260
MIXEDOTHER
MINIMUM SOUNDSPEEDPercentage of Water
MAXIMUM SOUNDSPEED
OFFACTIVE
SOUND STATCSIGS
KV INPUT SELECTION
KINEMATIC VISCOSITY
REYNOLDS CORRECTION
CALIBRATION FACTOR
(Wetted Transducers)
(Clamp-On Transducers)
2(V)
34
1(Z)
5
NUMBER OF TRAVERSES
TRANSDUCER SPACING
Energy Option OFF
Energy Option ON
OIL
METHETHFREON OTHERWATERLN2
SEA
LUBECRUDENORML
W260
OILOTHER
W100
NOYES
TRACKING WINDOWS?
MINIMUM SOUNDSPEED
MAXIMUM SOUNDSPEED
FLUID TYPE
WATER
FLUID SOUNDSPEED
OFF
ACTIVE
SIGSSOUND STATC
KV INPUT SELECTION
REYNOLDS CORRECTION
KINEMATIC VISCOSITY
CALIBRATION FACTOR
(Wetted Transducers)(Clamp-On Transducers)
1(Z)
2(V)
4
5
3
NUMBER OF TRAVERSES
TRANSDUCER SPACING
TEMPERATURE
See Figure A-4
FLUID TYPE
FLUID TYPE
See Figure A-4See Figure A-4
See Figure A-1See Figure A-3
DARKN
LITEN
STORE ABORT
WEDGE TEMPERATURE
FLUID TYPE
NO
YES
Figure 29: Channel X - PIPE Menu Map
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual79
[Esc][Enter][Esc]
PROGRESETCALIBCONTRAST
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
PIPEI/OSETUPSYSTMACTIV
NOYES
See Figure A-1See Figure A-1
Press YES to Default
See Figure A-4See Figure A-4See Figure A-4
See Figure A-1
DARKN
LITEN
STOREABORT
% of Peak
DELTA T OFFSET
AMP. DISCRIM HIGH
AMP. DISCRIM LOW
ACCELERATION LIMIT
VELOCITY HIGH LIMIT
VELOCITY LOW LIMIT
SOUNDSPEED +– LIMIT
COR. PEAK LIMIT
SIGNAL LOW LIMIT
42
8
1632
XMIT SAMPLE SIZE
See Figure A-2
SIGNLAVRGINITADVAN
10
5
30
21
STATS60
RESPONSE TIME
Max Peak %
Min. Peak %
# of Errors
NOYESNOYESAUTOSHORT
CstVCstR
MULTK
MASSCODELKV/xx*
NOYES
LONG
Multi K Factors?
Custom Type
Edit Table?
Enter # of K Factors
Reynolds/Velocity #*
K Factor #*
*Repeat for the # of K Factors entered previously.
Mass from Static Dens?
Fluid Density
Xmit Code Length# of KV/SS pairs
Sig Strength
(or Soundspeed) #**
Kin Viscosity #*
*To choose between signal strength
and soundspeed for kinematic
viscosity, refer to Reynolds Correction
in Figure A-1.
**Repeats for # of pairs ente
red
previously. Enter in ascending order.
NOYES
Figure 30: Channel X - SET UP Menu Map
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual80
[Esc][Enter][Esc]
PROGRESETCALIBCONTRAST
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
SYSTEM UNITS
METER MESSAGE
MetricE nglish
VOL DECIMAL DIGITS
TOTALIZER UNITS
0
12
3
TOTAL DECIMAL DIGITS
MetricEnglish
GAL
MGALft^3Mft^3
English
LB
KLBMMLBTONSKGTONNES
/SEC/MIN/HR/DAY
MASS FLOW TIME
Metric
MGDft3/sft3/m
BBL/MBBL/HBBL/D
GAL/SGAL/MGAL/Hft3/h
Mft3/dBBL/SA-I/S
A-I/MA-I/DA-F/SA-F/MA-I/H
ENGMETRC
VOLUMETRIC UNITS
BBL
MBBLAC-INAC-FT
LiterMliterm^3
Mm^3
BBLMBBL
0123
MDOT DECIMAL DIGITS
MetricEnglish
KG
TONNES
MASS DECIMAL DIGITS
0
12
3
MASS TOTALS
MASS FLOW
LB
KLBMMLBTONS
MASS FLOW ON MASS FLOW OFF
kBTU
MMBTUkWTONSKCal/s KCal/m KCal/hKCal/d
MetricEnglish
POWER
0
123
0
12
3
POWER DECIMAL DIGITS
KBTU
MMBTU
kWHrTONSKCalMCalKWHrMWHr
MetricEnglish
ENERGY (TOTAL)
0
12
3
ENERGY OPTION ONENERGY OPTION OFF
ENERGY OPTION ONENERGY OPTION OFF
ENERGY DECIMAL DIGITS
Liter/SLiter/MLiter/H
m3/hMm3/dBBL/S
MLiter/Dm3/sm3/m
BBL/MBBL/HMBL/D
2-CHANNEL UNITS
1-CHANNEL UNITS
See Figures A-1, A-2 and A-3
ERROROPTNLCD
HOLDLOWHIGHHHIGHOTHER
YESNO
Error Handling
2PATH ERROR HANDLING
SLOT0
OFF0-20m4-20m
AB
Slot 0 Outputs
SLOT1
See Figure A-5
CH1CH2SUMDIFAVE
Slot 0 Output A
Channel
(appears only for 2-Channel meter)
VELVOLUM+TOTL–TOTLTIMEMDOT
+MASS–MASSPOWER+ENRG–ENRGDIAG*
BASE
(*all diagnostic parameters appear here)
FULL
OFF12
3
4KEY
Measurement Name
CH1CH2SUMDIFAVE
VELVOLUM+TOTL–TOTLTIMEMDOT
+MASS–MASSPOWER+ENRG–ENRGDIAG*
(*all diagnostic parameters appear here)
(until all channels set up )(appears only for 2-Channel meter)
Channel 1
300600120024004800960019200
240048009600
NONEODDEVEN
1 BIT2 BITS
XXX
MODBUS PARITY
MODBUS BAUD RATE
MODBUS STOP BITS
MODBUS ADDRESS
BAUD RATE
Meter Address
MODBUS Only
NOYES
DARKNLITEN
STOREABORT
Reset Totals
(See Service Manual)
ft3/d
MBL/D
A-F/H
A-F/D
MCal/s MCal/m MCal/h MCal/d
KWMW
Figure 31: GLOBL - SYSTM, SLOT0 and COMM Menu Map
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual81
[Esc][Enter][Esc]
PROGRESETCALIBCNTRS
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
See Figure A-4
(See Service Manual)
NO
YES
Reset Totals
DARKN
LITEN
DARKNLITEN
ERROROPTNLCD
SLOT0SLOT1
ABCDABCDABCD
OFFTTLZROFF
FREQ
LOW
FAULT
OFFHIGH
CH1CH2
+TOTL–TOTL+MASS
SUMDIFAVE
–MASS+ENRG–ENRG
Totalizer Outputs
Slot 1 Outputs
Channel
2 Channel meter)
Measurement Name
PULSE ON TIME
UNITS/PULSE
See Figure A-4
See Figure A-4
Slot 1
appears only if option card installed
(Option Card)
CH1CH2
VELVOLUM+TOTL
SUMDIFAVE
–TOTLTIMEMDOT
Channel
2 Channel meter)
Measurement Name
(*all diagnostic
+MASS–MASS+ENRG–ENRGPOWERDIAG*
(appears only for(appears only for
FULL SCALE FREQUENCY
BASE
FULL
parameters
Slot 1 Outputs
Frequency Outputs
NO
YES
CH1CH2
VELVOLUM+TOTL
SUMDIFAVE
–TOTLTIMEMDOT
Channel
2 Channel meter)
FAULT
(*all diagnostic parameters
+MASS–MASS+ENRG–ENRGPOWERDIAG*
(appears only for
Trigger point
appear here)
Failsafe?
HIGH/LOW
Slot 1 Outputs
Alarm Relays
appear here)
Measurement Name
A
BCD
OFF
0-20m
CH1CH2
VELVOLUM+TOTL
SUMDIFAVE
–TOTLTIMEMDOT
Channel
2 Channel meter)
Measurement Name
(*all diagnostic
+MASS–MASS+ENRG–ENRGPOWERDIAG*
(appears only for
BASE
FULL
parameters
Slot 1 Outputs
Analog Outputs
appear here)
4-20m
ABCD
OFFTEMP
RTD Inputs
LABEL
Input Measurement
ZERO Value
FULL Scale Value
A
BCD
OFFTEMP
Analog Inputs
LABEL
Input Measurement
FULL Scale Value
SPEC
INPUT NAME
BASE Value
INPUT UNITS
See Figures A-1, A-2 and A-3
Figure 32: GLOBL - I/O Menu Map
Appendix A. Menu Maps
DigitalFlow™ XGM868i Programming Manual82
[no content intended for this page]
Appendix B. Data Records
Appendix B.Data Records
B.1Available Option Cards
The XMT868i can accommodate one option card in Slot 1 and one in Slot 2. The available configurations are listed in
Table 23.
Table 23: Option Card Configurations
Card #703-Slot #Configuration
1215-021FF - 4 Frequency Outputs
1215-03TT - 4 Totalizer Outputs
1215-04FT - 2 Frequency Outputs/2 Totalizer Outputs
1215-05FO - 2 Frequency Outputs
1215-06TO - 2 Totalizer Outputs
1215-07AA - 4 Standard Alarms
1215-09FA - 2 Frequency Outputs/2 Standard Alarms
1215-11TA - 2 Totalizer Outputs/2 Standard Alarms
1473-02OI - 2 Current Inputs
1473-03OR - 2 RTD Inputs
1473-04TI - 2 Current Inputs/2 Totalizer Inputs
1473-05TR - 2 RTD Inputs/2 Totalizer Inputs
1473-06FI - 2 Current Inputs/2 Frequency Inputs
1473-07FR - 2 RTD Inputs/2 Frequency Inputs
1473-08AI - 2 Current Inputs/2 Standard Alarms
1473-10AR - 2 RTD Inputs/2 Standard Alarms
1225-13CO - 2 Current Outputs
1225-14CF - 2 Current Outputs/2 Frequency Outputs
1225-15CT - 2 Current Outputs/2 Totalizer Outputs
1225-16CA - 2 Current Outputs/2 Standard Alarms
1430-03RR - 4 RTD Inputs
1430-04IR - 2 RTD Inputs/2 Current Inputs
1233-02II - 4 Current Inputs
1272-02CI - 2 Current Outputs/2 Current Inputs
1272-03CR - 2 Current Outputs/2 RTD Inputs
1272-04CIR - 2 Current Outputs/1 Current Input/1 RTD Input
1146-02/1332-022128 KB Memory
1146-03/1332-032 MB Memory
1310-02/1345-04MODBUS Communications Protocol
1477-032MODBUS/TCP Communication
1477-012Ethernet Communication
1475-012Foundation Fieldbus
DigitalFlow™ XMT868i Programming Manual83
Appendix B. Data Records
B.2Option Cards Installed
Whenever an option card is installed or changed in the XMT868i flow transmitter, record the type of card and any
additional setup information in the appropriate row of Table 24.
Table 24: Option Cards Installed
Slot #Type of Option CardAdditional Setup Information
0Analog Outputs (A, B)
1
2
84DigitalFlow™ XMT868i Programming Manual
Appendix B. Data Records
B.3Site Data
After the Model XMT868i flow transmitter has been installed, specific site data must be entered using the User
Program, prior to operation. Record that information in Table 25.
Table 25: Site Data
General Information
Model #Serial #
Software Vers.Setup Date
GLOBAL-SYSTEM
Meter MessageVol. Dec. Digits
System UnitsEnglishMetricTotalizer Units
Vol. UnitsTot. Dec. Dig.
GLOBAL-I/O-ERROR
Error Handling2-Path ErrorNoYes
GLOBAL-COMM
Meter AddressMOD. Parity
Baud RateMOD. Stop Bits
MOD. Baud RateMOD. Address
Channel x-ACTIV
Channel 1Channel 2 (if applicable)
Channel Status
Off
1
Channel 1Channel 2 (if applicable)
Channel LabelChannel Label
Chan. Message
2
Energy OptionOnOffEnergy OptionOnOff
Vol. UnitsVol. Units
Vol. Dec. DigitsVol. Dec. Digits
Totalizer UnitsTotalizer Units
Tot. Dec. Dig.Tot. Dec. Dig.
Mass FlowMass Flow
Mass Flow TimeMass Flow Time
MDOT Dec. Dig.MDOT Dec. Dig.
Mass TotalsMass Totals
Mass Dec. DigitsMass Dec. Digits
PowerPower
Power Dec. Dig.Power Dec. Dig.
Energy (Total)Energy (Total)
Energy Dec. Dig.Energy Dec. Dig.
Heat. or Cool.Heat. or Cool.
Flow Meas.Flow Meas.
1
not available for 1-Channel meter, 2”Site Message” for 1-Channel meter
Code LengthAuto Short LongCode LengthAuto Short Long
88DigitalFlow™ XMT868i Programming Manual
Appendix C. Programming the XMT868i Using PanaView™
Appendix C.Programming the XMT868i Using PanaView™
C.1Introduction
The Model XMT868i flow transmitter must be properly installed and programmed, as described in the Startup Guide,
before it can provide accurate flow rate measurements. After completing the installation and initial setup, use this
chapter to program the advanced features of the Model XMT868i’s using the PanaView™ software program. Refer to
the appropriate section for a discussion of the following menu features:
•Channel-Status - activate one or both channels and select the desired measurement method
•Channel-System - enter the individual channel parameters
•Channel-Pipe - enter the pipe parameters
•Channel-I/O - set up the inputs and outputs
•Channel-Setup - set the signal limits, response times and activating mass flow.
•Global-System - enter the system units
•Global-I/O - set up error handling, option cards and display
•Global-Comm - set the serial port and MODBUS parameters
As a programming aid, a complete set of menu maps for the
Maps. The specific figure numbers will be referenced throughout this chapter, as required.
PROG menu is included in Appendix D, PanaView Menu
C.2Programming Using PanaView™
You can program the XMT868i using PanaView™, a PC-based, non-resident software program that communicates
with the XMT868i through its RS232/RS485 serial or Ethernet port.
C.2.1 Preparing for PanaView Programming
Before you attempt to communicate with the XMT868i, be sure you have linked your PC to the XMT868i using its
communication interface. For details on wiring the interface, refer to Wiring the Serial Port in the Startup Guide, and to
the document EIA-RS Serial Communications (916-054). You must also install PanaView, as discussed in the
PanaView User’s Manual (910-211).
DigitalFlow™ XMT868i Programming Manual89
Appendix C. Programming the XMT868i Using PanaView™
C.2.2 Setting Up the Serial Communications Port
Use the steps below to establish PanaView communications with the XMT868i.
1. Launch PanaView, as discussed in Chapter 3, Initial Setup, of the PanaView User’s Manual.
2. Open the New Meter Browser window in the File menu and expand the network tree. Then, highlight the My
Computer (Name) branch by clicking on it.
3. Pull down the Edit menu by clicking on it in the menu bar.
4. Click on the New menu option to select it, and a submenu opens with two choices on it (see Figure 33).
Figure 33: Edit Menu
90DigitalFlow™ XMT868i Programming Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.