Technical content subject to change without notice.
[no content intended for this page]
ii
Preface
Information Paragraphs
Note:These paragraphs provide information that provides a deeper understanding of the situation, but is not
essential to the proper completion of the instructions.
IMPORTANT:These paragraphs provide information that emphasizes instructions that are essential to proper
setup of the equipment. Failure to follow these instructions carefully may cause unreliable
performance.
CAUTION!This symbol indicates a risk of potential minor personal injury and/or severe damage
to the equipment, unless these instructions are followed carefully.
WARNING!This symbol indicates a risk of potential serious personal injury, unless these
instructions are followed carefully.
Safety Issues
WARNING!It is the responsibility of the user to make sure all local, county, state and national
codes, regulations, rules and laws related to safety and safe operating conditions are met for each
installation. The safety of any system incorporating the equipment is the responsibility of the
assembler of the system.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working Area
WARNING!Auxiliary equipment may have both manual and automatic modes of operation. As
equipment can move suddenly and without warning, do not enter the work cell of this equipment
during automatic operation, and do not enter the work envelope of this equipment during manual
operation. If you do, serious injury can result.
WARNING!Make sure that power to the auxiliary equipment is turned OFF and locked out before
you perform maintenance procedures on the equipment.
DigitalFlow™ XMT868i Startup Guideiii
Preface
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary
equipment. Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control is an active participant in Europe’s Waste Electrical and Electronic Equipment
(WEEE) take-back initiative, directive 2012/19/EU.
The equipment that you bought has required the extraction and use of natural resources for its production. It
may contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on
the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or
recycle most of the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or
regional waste administration.
Visit www.gemeasurement.com/environmental-health-safety-ehs
information about this initiative.
To ensure safe and reliable operation of the Model XMT868i Flow Meter, the system must be installed in accordance
with the guidelines established by GE engineers. Those guidelines, explained in detail in this chapter, include the
following topics:
•Unpacking the Model XMT868i system
•Selecting suitable sites for the electronics enclosure and the flowcell/transducers
•Installing the flowcell/transducers
Note:See the appropriate Transducer Installation Guide for detailed instructions on transducer installation.
•Installing optional temperature transmitters
•Installing the electronics enclosure
•Wiring the electronics enclosure
WARNING!The Model XMT868i flow transmitter can measure the flow rate of many fluids, some
potentially hazardous. The importance of proper safety practices cannot be overemphasized.
WARNING!Be sure to follow all applicable local safety codes and regulations for installing
electrical equipment and working with hazardous fluids or flow conditions. Consult company
safety personnel or local safety authorities to verify the safety of any procedure or practice.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
1.2Unpacking
Carefully remove the electronics enclosure, the transducers, and the cables from the shipping containers. Before
discarding any of the packing materials, account for all components and documentation listed on the packing slip. The
discarding of an important item along with the packing materials is all too common. If anything is missing or damaged,
contact the factory immediately for assistance.
DigitalFlow™ XMT868i Startup Guide1
Chapter 1. Installation
1.3Site Considerations
Because the relative location of the flowcell and the electronics enclosure is important, use the guidelines in this
section to plan the XMT868i installation. Figure 1 below shows a typical installation.
Figure 1: A Typical Model XMT868i System
1.3.1Electronics Enclosure Location
The standard Model XMT868i electronics enclosure is a powder-coated, cast-aluminum, Type-7/4X explosion-proof
enclosure, and an optional stainless-steel enclosure is available. Typically, the enclosure is mounted as close as possible
to the transducers. When choosing a site, make sure the location permits easy access to the electronics enclosure for
programming, maintenance and service.
Note:For compliance with the European Union’s Low Voltage Directive, this unit requires an external power
disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
1.3.2Flowcell Location
Ideally, choose a section of pipe with unlimited access; for example, a long stretch of pipe that is above ground.
However, if the flowcell is to be mounted on an underground pipe, dig a pit around the pipe to facilitate installation of
the transducers.
2DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
1.3.3Transducer Location
For a given fluid and pipe, the Model XMT868i’s accuracy depends primarily on the location and alignment of the
transducers. In addition to accessibility, when planning for transducer location, adhere to the following guidelines:
•Locate the transducers so that there are at least 10 pipe diameters of straight, undisturbed flow upstream and
5 pipe diameters of straight, undisturbed flow downstream from the measurement point. Undisturbed flow
means avoiding sources of turbulence in the fluid such as valves, flanges, expansions, and elbows; avoiding
swirl; and avoiding cavitation.
•Locate the transducers on a common axial plane along the pipe. Locate the transducers on the side of the pipe,
rather than the top or bottom, since the top of the pipe tends to accumulate gas and the bottom tends to
accumulate sediment. Either condition will cause increased attenuation of the ultrasonic signal. There is no
similar restriction with vertical pipes. However, vertical pipes should be avoided in order to insure a full pipe at
the measurement point.
1.3.4Cable Lengths
Locate the electronics enclosure as close as possible to the flowcell/transducers, preferably directly on the flowcell.
However, GE can supply transducer cables up to 1,000 ft (300 m) in length for remote location of the electronics
enclosure. If longer cables are required, consult the factory for assistance.
1.3.5Transducer Cables
When installing the transducer cables, always observe established standard practices for the installation of electrical
cables. Do not route transducer cables alongside high amperage AC power lines or any other cables that could cause
electrical interference. Also, protect the transducer cables and connections from the weather and corrosive
atmospheres.
Note:If you are using your own cables to connect the transducers to the electronics console, they must have
electrical characteristics identical to the cables supplied by GE. For transducer frequencies up to 2 MHz, the
cables in each pair must be the same length within ±4 in. (10 cm). For transducer frequencies above 2 MHz,
the cables in each pair must be the same length within ±0.5 in. (1.25 cm).
1.4Installing a Flowcell
A flowcell is the section of pipe where the transducers are mounted. It can be created either by mounting the
transducers on the existing pipeline or by mounting them on a spoolpiece. A spoolpiece is a separately manufactured
pipe section, matched to the existing pipe, which contains ports for mounting the transducers. This approach allows the
transducers to be aligned and calibrated before inserting the spoolpiece into the pipeline.
Figure 1 on page 2 shows a typical Model XMT868i spoolpiece, with a mounting bracket to hold the electronics
enclosure. For detailed instructions on installing the transducers and/or spoolpiece, refer to the supplied drawings and
the appropriate Transducer Installation Guide.
DigitalFlow™ XMT868i Startup Guide3
Chapter 1. Installation
Transmitter
RTD
Coupling
Thermowell
Flowcell Wall
Temperature
1.5Installing Temperature Transmitters
Optional temperature transmitters may be installed as part of the flowcell, near the ultrasonic transducer ports. Be sure
to observe the siting requirements given earlier in this chapter. These transmitters should send a 0/4-20 mA signal to
the Model XMT868i. In turn, the XMT868i must be fitted with a suitable option card to process the signals and to
provide the required 24 VDC power to the transmitters. Any desired transmitters or sensors may be used, but they must
have an accuracy equal to 0.5% of the reading or better.
Note:Resistive Thermal Devices (RTDs) are a good choice for measuring the temperature.
Typically, a 1/2” NPT female threaded port is used to mount the transmitters on the flowcell. If the pipeline is insulated,
the coupling may need to be extended to provide convenient access. Of course, other types of mounting ports,
including flanged ports, may be used for the transmitters.
Figure 2 below shows a typical mounting arrangement for a temperature transmitter. The temperature transmitter
should protrude 1/4 to 1/2 way into the pipe.
4DigitalFlow™ XMT868i Startup Guide
Figure 2: Typical Temperature Transmitter Mounting
Chapter 1. Installation
1.6Mounting the XMT868i Electronics Enclosure
The standard model XMT868i electronics package is housed in a Type-4X weather-resistant enclosure suitable for
indoor or outdoor use. Refer to Figure 8 on page 21 for the mounting dimensions and the weight of this enclosure.
The Model XMT868i electronics enclosure is fitted with a mounting boss that has a single 3/4” NPTF threaded hole at
its center and four 1/4-20 tapped holes at its corners. By using this mounting boss, the electronics enclosure may be
mounted by any of the typical techniques shown in Figure 9 on page 22.
WARNING!Proper grounding of the XMT868i chassis is required to prevent the possibility of
electric shock. See Figure 9 on page 22 to locate the chassis grounding screw.
1.7Making Electrical Connections
This section contains instructions for making all the necessary electrical connections to the Model XMT868i flow
transmitter. Refer to Figure 10 on page 23 for a complete wiring diagram.
Except for the power connector, all electrical connectors are stored on their terminal blocks during shipment and may
be removed for more convenient wiring. Simply, feed the cables through the conduit holes on the side of the enclosure,
attach the wires to the appropriate connectors and plug the connectors back into their terminal blocks.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
WARNING!Always disconnect the line power from the Model XMT868i before removing either
the front cover or the rear cover. This is especially important in a hazardous environment.
Refer to Figure 3 on page 6 and prepare the Model XMT868i for wiring by completing the following steps:
1.Disconnect any previously wired power line from the unit.
2.Loosen the set screw on the rear cover.
3.Place a rod or long screwdriver across the cover in the slots provided, and rotate the cover counterclockwise
until it comes free from the enclosure.
4.Install any required cable glands in the appropriate conduit holes around the side of the enclosure.
CAUTION!To ensure safe and reliable operation of the flow meter, be sure to install and tighten
all cable glands in accordance with the gland manufacturer’s instructions.
5.Note the labels inside the rear cover to assist in wiring the power and option card connections. Also, Figure 11
on page 24 shows the rear cover option card labels for every available option card configuration.
Proceed to the appropriate section of this chapter to make the desired wiring connections.
DigitalFlow™ XMT868i Startup Guide5
Chapter 1. Installation
Slots
Power Connections
Option Card Connections
Label
Set Screw
1.7 Making Electrical Connections (cont.)
Inside ViewOutside View
Figure 3: Rear Cover with Connection Labels
1.7.1Wiring the Line Power
The Model XMT868i may be ordered for operation with power inputs of 100-120 VAC, 220-240 VAC, or 12-28 VDC.
The label on the side of the electronics enclosure lists the meter’s required line voltage and power rating. Be sure to
connect the meter only to the specified line voltage.
Note:For compliance with the European Union’s Low Voltage Directive, this unit requires an external power
disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
Note:Only use a Class 2 Rated Power supply for line power connection to a DC instrument.
Refer to Figure 10 on page 23 to locate terminal block
WARNING!Improper connection of the line power leads or connecting the meter to the incorrect
line voltage may damage the unit. It may also result in hazardous voltages at the flowcell and
associated piping as well as within the electronics enclosure.
TB5 and connect the line power as follows:
6DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
1.7.1Wiring the Line Power (cont.)
1.Prepare the line power leads by trimming the line and neutral AC power leads (or the positive and negative DC
power leads) to a length 0.5 in. (1 cm) shorter than the ground lead. This ensures that the ground lead is the last
to detach if the power cable is forcibly disconnected from the meter.
2.Install a suitable cable clamp in the conduit hole indicated in Figure 10 on page 23. If possible, avoid using the
other conduit holes for this purpose, to minimize any interference in the circuitry from the AC power line.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
3.Strip 1/4-in. of insulation from the end of each of the three line power leads.
4.Route the cable through the conduit hole and connect the line power leads to terminal block
number assignments shown in Figure 10 on page 23.
5.Leaving a bit of slack, secure the power line with the cable clamp.
WARNING!Make sure both covers, with their O-ring seals, are installed and the set screws are
tightened before applying power in a hazardous environment.
CAUTION!The transducers must be properly wired before applying power to the meter.
Proceed to the next section to continue the initial wiring of the Model XMT868i flow transmitter.
TB5, using the pin
DigitalFlow™ XMT868i Startup Guide7
Chapter 1. Installation
1.7.2Wiring Transducers
Before wiring the XMT868i transducers, complete the following steps:
1.Disconnect the main power from the electronics enclosure.
2.Remove the rear cover and install all required cable clamps.
WARNING!Before connecting the transducers, take them to a safe area and discharge any static
buildup by shorting the center conductor of the transducer cables to the metal shield on the cable
connector.
Note:For transducer frequencies below 2 MHz, the cable lengths must be within 4 in. (10 cm) of each other. If the
transducer frequency exceeds 2 MHz, the cable lengths must be within 0.5 in. (1.25 cm) of each other.
3.Locate the
CH1 transducer cables and connect them to the two CH1 transducers in the flowcell. Feed the free
ends of the cables through the selected conduit hole in the electronics enclosure.
4.If an optional lightning protector is being installed, connect it between the meter and the transducers.
5.Refer to the wiring diagram in Figure 10 on page 23 and connect the transducer cables to terminal block
Then, secure the cable clamp.
Note:The RED cable leads are the SIG(+) leads and the BLACK cable leads are the RTN(-) leads.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
6.For a 2-Channel XMT868i, repeat the above steps to connect the
CH2 transducers to terminal block J4. It is not
required that both channels of a 2-Channel unit be connected.
7.If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
Note:A channel must be activated before it can begin taking measurements. See Chapter 2, Initial Setup, for
instructions.
J3.
8DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
1.7.3Wiring Standard 0/4-20 mA Analog Outputs
The standard configuration of the Model XMT868i flow transmitter includes two isolated 0/4-20 mA analog outputs
(designated as outputs 1 and 2). Connections to these outputs may be made with standard twisted-pair wiring, but the
current loop impedance for these circuits must not exceed 600 ohms.
To wire the analog outputs, complete the following steps:
1.Disconnect the main power to the unit and remove the rear cover.
2.Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3.Refer to Figure 10 on page 23 for the location of terminal block
J1 and wire the analog outputs as shown.
Secure the cable clamp.
Note:Analog outputs 1 and 2 in the wiring diagram correspond to analog outputs A and B in Slot 0 in the XMT868i
software.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
4.If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
WARNING!Make sure both covers, with their O-ring seals, are installed and the set screws are
tightened before applying power in a hazardous environment.
Note:Prior to use, the analog outputs must be set up and calibrated. See Chapter 1, Calibration, of the Service
Manual for detailed instructions.
Proceed to the next section to continue the initial wiring of the unit.
1.7.4Wiring the Serial Port
The Model XMT868i flow transmitter is equipped with a built-in serial communications port. The standard port is an
RS232 interface, but an optional RS485 interface is available upon request. Proceed to the appropriate sub-section for
wiring instructions. For more information on serial communications refer to the EIA-RS Serial Communications
manual (GE document 916-054).
DigitalFlow™ XMT868i Startup Guide9
Chapter 1. Installation
1.7.4aWiring the RS232 Interface
Use the serial port to connect the Model XMT868i flow transmitter to a printer, an ANSI terminal or a personal
computer. The RS232 interface is wired as Data Terminal Equipment (DTE), and the signals available at terminal block
J1 are shown in Ta bl e 1 below. Refer to Figure 10 on page 23 and complete the following steps:
1.Disconnect the main power to the unit and remove the rear cover.
2.Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3.Use the information in Table 1 to construct a suitable cable for connecting the Model XMT868i to the external
device. If desired, an appropriate cable may be purchased from GE.
Table 1: RS232 Connection to DCE or DTE Device
DCE
J1
Pin #Signal Description
DB25
Pin #
5DTR (Data Terminal Ready)204204
6CTS (Clear to Send)4758
7COM (Ground)7575
8RX (Receive)2332
DCE
DB9
Pin #
DTE
DB25
Pin #
DTE
DB9
Pin #
9TX (Transmit)3223
Note:Signal names that imply direction (e.g., transmit and receive) are named from the point of view of the DTE
device (the GE meter is usually considered the DTE device). When the RS232 standard is strictly followed,
these signals are labeled with the same name and pin # on the DCE device side as well. Unfortunately, the
convention is not followed because the DTE and DCE side get confused. Therefore, connections that imply
direction are changed to reflect their direction on the DCE side.
4.Feed the flying leads end of the cable through the conduit hole and wire it to terminal block
J1. Connect the
other end of the cable to the printer, ANSI terminal or personal computer, and secure the cable clamp.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
After the wiring has been completed, consult the User’s Manual for the external device to configure it for use with the
XMT868i.
10DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
1.7.4bWiring the RS485 Interface
Use the optional RS485 serial port to network multiple XMT868i flow transmitters to a single computer terminal. As
an option, the standard RS232 port on the XMT868i may be configured as a two-wire, half-duplex RS485 interface.
IMPORTANT:The XMT868i must be configured at the factory for RS485 operation. The maximum cable length for
RS485 is 4000 ft (1200 m).
To wire the RS485 serial port, refer to Figure 10 on page 23 and complete the following steps:
1.Disconnect the main power to the unit and remove the rear cover.
2.Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3.Feed one end of the cable through the conduit hole, wire it to terminal block
J1 and secure the cable clamp. Use
the information in Ta bl e 2 below to wire a suitable cable for connecting the XMT868i to the external device.
Table 2: RS485 Connections
J1 Pin #Signal Description
9Data +
8Data -
7Shield
6Not Used
5Not Used
4Not Used
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
4.If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
Proceed to the next section to continue the initial wiring of the unit.
DigitalFlow™ XMT868i Startup Guide11
Chapter 1. Installation
1.7.5Wiring Option Cards
The XMT868i can accommodate one option card in Slot 1 and one option card in Slot 2. The following types of option
cards are available only in the combinations listed in Table 22 on page 69:
•Alarm Relays (Slot 1)
•Analog Inputs (Slot 1)
•Totalizer/Frequency Outputs (Slot 1)
•RTD Inputs (Slot 1)
•Analog Outputs (Slot 1)
•MODBUS Communications (Slot 2)
•MODBUS/TCP Communications (Slot 2)
•Ethernet (Slot 2)
•Foundation Fieldbus (Slot 2)
•Data logging (Slot 2) - no wiring required
1.7.5aPreparing for Wiring
Wiring any option card installed in Slot 1 requires completion of the following general steps:
1.Disconnect the main power to the unit and remove the rear cover.
2.Install a cable clamp in the chosen conduit hole on the side of the electronics enclosure and feed a standard
twisted-pair cable through this conduit hole.
3.Locate the 12-pin terminal block (
inside the rear cover (see Figure 3 on page 6 and Figure 11 on page 24). Secure the cable clamp.
IMPORTANT:Because of the attached wiring label, all rear covers must remain with their original meters.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
4.If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
Note:Prior to use, the option card must be set up and calibrated. See Chapter 1, Programming Site Data, in the
Programming Manual and Chapter 1, Calibration, in the Service Manual for detailed instructions.
J2) in Figure 10 on page 23 and wire the option card as indicated on the label
For more specific instructions on particular option cards, proceed to the appropriate sub-section(s) that follow.
12DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
C
NO
NCNC
C
NO
NCNC
C
NO
C
NO
Conventional
not triggerednot triggered
Fail-Safe
triggered or power failure
Fail-Safe
triggered
Conventional
ALARM
MONITORING
DEVICEDEVICE
MONITORING
ALARM
DEVICE
MONITORING
ALARM
DEVICE
MONITORING
ALARM
1.7.5bWiring an Alarms Option Card
Each alarms option card includes two or four general-purpose Form C relays (designated as A, B, C and D).
The maximum electrical ratings for the relays are listed in Chapter 4, Specifications. Each of the alarm relays can be
wired as either Normally Open (NO) or Normally Closed (NC).
In setting up an alarm relay, it may be wired for either conventional or fail-safe operation. In fail-safe mode, the alarm
relay is constantly energized, except when it is triggered or a power failure or other interruption occurs. See Figure 4
below for the operation of a normally open alarm relay in both conventional and fail-safe mode.
Connect each alarm relay in accordance with the wiring instructions shown on the label inside the rear cover (see
Figure 3 on page 6 and Figure 11 on page 24).
Figure 4: Conventional and Fail-Safe Operation
DigitalFlow™ XMT868i Startup Guide13
Chapter 1. Installation
+24V
INPUT +
INPUT +
+24V
+–
+ IN
– OUT
+ IN
– OUT
Sensor
Sensor
Transmitter
Transmitter
Analog Input
Analog Input
24 VDC
POWER SUPPLY
With External Power Supply
With Internal Power Supply
1.7.5cWiring a 0/4-20 mA Analog Inputs Option Card
This option card includes two or four isolated 0/4-20 mA analog inputs (designated as A, B, C and D), each of which
includes a 24 VDC power supply for loop-powered transmitters. Either input may be used to process the temperature
signal, while the other input is used to process the pressure signal.
Note:To enter programming data during operation of the meter, it will be necessary to know which input is assigned
to which process parameter. This information should be entered in Appendix B, Data Records.
The analog inputs, which have an impedance of 118 ohms, should be connected with standard twisted-pair wiring.
Power to the transmitters may be supplied either by the integral 24 VDC power supply on the analog input option card
or by an external power supply. Figure 5 below shows typical wiring diagrams, with and without an external power
supply, for one of the analog inputs.
Figure 5: Analog Input Wiring Diagram
Wire the analog inputs as shown on the label in the rear cover (see Figure 3 on page 6 and Figure 11 on page 24).
Note:The analog inputs option card can be calibrated with the Model XMT868i’s built-in analog outputs. However,
be certain that the analog outputs have been calibrated first. See Chapter 1, Calibration, in the Service Manual
for the appropriate procedures.
14DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
OUT
RTN
IN
Common
Load
Volts -
(Common)
Volts +
(Int. Pwr. Sup.)
OUT
RTN
+5V
200
:
Model XMT868Pulse Counter
Totalizer Output
Frequency CounterModel XMT868
Frequency Output
1.7.5dWiring a Totalizer and Frequency Output Option Card
Each totalizer and frequency output option card provides two or four outputs (designated as A, B, C, and D).
Wire this option card in accordance with the connections shown on the label in the rear cover (see Figure 3 on page 6
and Figure 11 on page 24). Figure 6 below shows sample wiring diagrams of a totalizer output circuit and a frequency
output circuit. Refer to Electrical Specifications on page 61 for the load and voltage requirements.
Figure 6: Totalizer and Frequency Output Wiring
DigitalFlow™ XMT868i Startup Guide15
Chapter 1. Installation
1.7.5eWiring an RTD Inputs Option Card
The Model XMT868i RTD (Resistance Temperature Device) inputs option card provides two direct RTD inputs
(designated as A and B). Each RTD input requires three wires, and should be connected as shown on the label in the
rear cover (see Figure 3 on page 6 and Figure 11 on page 24).
1.7.5fWiring a 0/4-20 mA Analog Outputs Option Card
The analog outputs option card includes two isolated 0/4-20 mA outputs (designated as A and B). Connections to these
outputs may be made with standard twisted-pair wiring, but the total current loop impedance for these circuits must not
exceed 1,000 ohms. Wire this option card with the connections shown on the label in the rear cover (see Figure 3 on page 6 and Figure 11 on page 24).
1.7.5gA Specific Example
To illustrate the procedures described in the previous sections for wiring the option cards, a specific example may
prove helpful. Assume that an XMT868i is equipped with an option card numbered 703-1223-08. Table 22 on page 69
in Appendix B, Data Records, indicates that this option card includes two analog current inputs and two standard alarm
relays.
By referring to Figure 11 on page 24, you can see that the appropriate connection diagram for this option card is the
first one from the left in the middle row. This label is also be found inside the rear cover of the XMT868i. Based on this
information, wiring of the Input/Output connections to terminal block
below.
J2 should be completed as shown in Ta ble 3
Table 3: Wiring a 703-1223-08 Option Card
J2 Pin #DescriptionConnection
1Alarm ANormally Open
2Alarm A Common
3Alarm ANormally Closed
4Alarm BNormally Open
5Alarm B Common
6Alarm BNormally Closed
7Out C+24 V to Input C
8Input CSignal (+)
9Input CReturn (-)
10Out D+24 V to Input D
11Input DSignal (+)
12Input DReturn (-)
16DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
N/C
MODBUS –
MODBUS +
1.7.5hWiring the MODBUS Option Card
The MODBUS option card uses the RS485 standard for MODBUS communication. The MODBUS card must be
installed in Slot 2. The RS485 standard allows up to 32 nodes (drivers and receivers) on one multi-drop network, at
distances up to 4,000 ft (1,200 m). GE recommends using 24-gauge (24 AWG) twisted-pair wire with a characteristic
impedance of 120 ohms and 120-ohm termination at each end of the communications line.
Note:The MODBUS option card provides its own RS485 connections. Thus, the meter can have its serial port
configured as RS232 and still provide RS485 MODBUS signals.
IMPORTANT:Because the MODBUS option card uses pins 1 and 2 on terminal block J2, only those option cards that
do not use these pins may be installed in Slot 1, while a MODBUS card is installed in Slot 2.
Specifically, only the option cards designated as “OI” and “OR” in Figure 11 on page 24 are
compatible with the MODBUS option card.
To link the XMT868i to the control system (refer to Figure 7 below):
1. Connect the MODBUS– wire to pin
2. Connect the MODBUS+ wire to pin
3. Pin #3 has no connection.
#2, the inverting or negative connection.
#1, the non-inverting or positive connection.
Figure 7: The RS485 MODBUS Terminal Block Connector
DigitalFlow™ XMT868i Startup Guide17
Chapter 1. Installation
1.7.5iWiring the MODBUS/TCP Interface
Customers can also use a modified XMT868i that provides a MODBUS/TCP interface to communicate to an internal
network. An optional MODBUS/TCP card with a unique MAC (IP) address (installed only in slot 2) includes an RJ45
connector. To connect the MODBUS/TCP-enabled XMT868i to the network, insert the jack of an RJ45 cable into the
RJ45 connector, route the cable through one of the conduit holes using an appropriate cable clamp, and wire the other
end of the cable to the Ethernet network according to the manufacturer’s instructions (see Figure 12 on page 25).
Note:The MAC address for a specific XMT868i is included with customer documentation. For more information on
setting up the MAC address, refer to Chapter 6 of the Programming Manual.
1.7.5jWiring the Ethernet Interface
A modified XMT868i can use the Ethernet interface to communicate to a local area network. An optional Ethernet card
with a unique MAC (IP) address (installed only in slot 2) includes an RJ45 connector. To connect the Ethernet-enabled
XMT868i to the network, insert the jack of an RJ45 cable into the RJ45 connector, route the cable through one of the
conduit holes using an appropriate cable clamp, and wire the other end of the cable to the Ethernet network according
to the manufacturer’s instructions. An external connection is required between the Ethernet option card and the
XMT868i’s RS232 connector, as shown in Ta bl e 4 below and Figure 12 on page 25.
Note:The MAC address for a specific XMT868i is included with customer documentation. For more information on
setting up the MAC address, refer to Appendix C of the Programming Manual.
Table 4: RS232 to Ethernet Interconnections
RS232 TB
on Rear PC Board
on Ethernet Card
TB1
TXPin 1
RXPin 2
COMPin 3
18DigitalFlow™ XMT868i Startup Guide
Chapter 1. Installation
1.7.5kWiring the Foundation Fieldbus Interface
To connect the Foundation Fieldbus interface to the XMT868i, make the network connections at J8, pins 1 and 2, as
shown in drawing Y of Figure 12 on page 25. As an option, you can connect a shield to J8 pin 3, depending on the
network wiring.
No connections are made to J9 under normal operation. If you need to reset the network board to factory defaults,
proceed as follows:
1. Connect a jumper between J9 pin 2 and J9 pin 3.
2. Power cycle (i.e., turn it Off and then back On) the XMT868i.
3. Ten seconds after the power has been restored to the XMT868i, remove the jumper to return the network circuit
board to normal operation.
1.7.5lData Logging Option Card
The XMT868i can be fitted with a data logging option card, which must be installed in Slot 2. The data logging option
card increases the XMT868i’s memory by up to 2 MB.
WARNING!To meet CE Mark requirements, all cables must be installed as described in
Appendix A, CE Mark Compliance.
1.7.5m Procedure Options
After the XMT868i has been completely installed and wired, proceed to Chapter 2, Initial Setup, to program the meter
for taking flow rate measurements.
WARNING!Make sure both covers, with their O-ring seals, are installed and the set screws are
tightened before applying power in a hazardous environment.
DigitalFlow™ XMT868i Startup Guide19
Chapter 1. Installation
[no content intended for this page]
20DigitalFlow™ XMT868i Startup Guide
D
S
O
R
U
V
I
R
N
E
P
A
T
U
O
S
S
E
O
I
N
S
N
O
P
O
O
T
N
E
N
W
H
S
E
D
E
R
N
E
I
L
E
G
I
Ø6.10 (155)
8.20
2.06
51°
7 PLCS
3/4" NPTF
7 PLCS
3.86
SEE
AA
VIEW A-A
0.52 (13)
DETAIL A
3/4" NPT
1/4-20 UNC-2B
1.50
1.00
0.28 (7)
0.25 (6)
10-32 EXTERNAL GROUND SCREW
4PLCS
3/4" NPTF
7 PLCS
(52)
(98)
(208)
(38)
NOTES:
1. ALL DIMENSIONS ARE REFERENCE.
2. WEIGHT: 10 LB (4.5 KG) AL
25 LB (11.5 KG) SS
3. DIMENSIONS ARE IN INCHES (MILLIMETERS).
(25)
DETAIL A
Chapter 1. Installation
Figure 8: Model XMT868i Outline and Installation (ref. dwg #712-1318)
DigitalFlow™ XMT868i Startup Guide21
Chapter 1. Installation
SEE
7 PLCS
3/4" NPTF
2.06
AA
3.86
8.20
VIEW A-A
Ø6.10 (155)
H
R
N
O
P
A
S
E
V
U
I
R
O
D
7 PLCS
51°
P
T
O
N
O
N
E
W
0.52 (13)
N
O
S
O
S
U
N
E
T
S
I
D
E
E
N
E
L
I
R
G
I
S
E
3. DIMENSIONS ARE IN INCHES (MILLIMETERS)]
2. WEIGHT = 10 LB (4.5 KG)
1. ALL DIMENS IONS ARE REFERENCE
NOTES:
2.375" PIPE
2.830
2.830
6.00
5.00
1.40
1.08
1/4-20 x 3/4 BOLT
LOCK WASHER
WASHER
4 PLACES
1.001.000
1.250
6.00
5.00
DETAIL B
SEE DETAIL B
.375
Ø.330
4 HOLES
SEE DETAIL C
BRACKET
REF ONLY
4 PLACES
LOCK WASHER
5/16 -18 NUT
WASHER
2" PIPE U-BOLT
2 PLACES
1.250
.625
1.250
STRUCTUAL ANGLE
MATERIAL: AL 6x6 x .375 THK (REF)
.135 R
.205
.135
8 PLACES
DETAIL C
.270
.410
3/4" NPT
DETAIL A
10-32 EXTERNAL GROUND SCREW
4PLCS
1/4-20 UNC-2B
1.00 (25)
1.50
0.28 (7)
0.25 (6)
(52)
(98)
(208)
(38)
DETAIL A
DigitalFlow™ XMT868i Startup Guide22
Figure 9: Model XMT868i Outline and Installation - Right Angle Mounting Bracket (ref. dwg #712-1317)
Pin #
1
2
3
Description
Line Power
Line Neutral
Earth Ground
AC POWER INPUT
J1 - RS232/RS485 / 4-20 OUTPUT
Data Terminal Ready
Ground
Clear To Send
Description
Transmit / +
Receive / –
For compliance with the European Union's Low Voltage Directive
(2006/95/EC), this unit requires an external power disconnect device
such as a switch or circuit breaker. The disconnect device must be
marked as such, clearly visible, directly accessible and located
within 1.8 m (6 ft) of the unit.
NOTE:
J4 - CH2 TRANSDUCER CONNECTION (optional)
Upstream Transducer SIG(+)
Downstream Transducer SIG(+)
Upstream Transducer Analog RTN(-)
Downstream Transducer Analog RTN(-)
Description
DC POWER INPUT
Pin #
2
3
1
Description
Line Negative
No Connection
Line Positive
Mounting Boss
Power Cable
Inlet
Conduit Hole (1 of 7)
Grounding Jumper
Protective Conductor Terminal
Nameplate
NOTE: Enclosure is shown from the
rear with the rear cover removed.
Figure 12: Ethernet, Foundation Fieldbus and MODBUS/TCP Terminal Block Connections
DigitalFlow™ XMT868i Startup Guide25
[no content intended for this page]
Chapter 1. Installation
DigitalFlow™ XMT868i Startup Guide26
Chapter 2. Initial Setup
Chapter 2.Initial Setup
2.1Introduction
This chapter provides instructions for programming the minimum amount of data required to place the Model
XMT868i flow meter into operation. Before the Model XMT868i can begin taking measurements and displaying valid
data, the current system and pipe parameters must be entered. In addition, a 2-Channel meter requires that each channel
be activated prior to use. Additional programming options provide access to the more advanced features of the Model
XMT868i, but this information is not required to begin taking measurements.
Note:See the Programming Manual for information on those User Program features not covered in this chapter.
You can program the XMT868i using either the keypad on the lower part of the glass enclosure, or PanaView™, a
PC-based, non-resident software program that communicates with the XMT868i using its RS232 serial port. PanaView
supplements basic XMT868i functions with several additional capabilities. With PanaView, you can:
•Load and save site file data
•Create and save graph and log files
•Display text output and graphs of live measurement data
•Create custom templates for displaying text, graph and log data
•Interface with multiple GE instruments.
This chapter focuses on programming using the magnetic keypad. For information on programming the XMT868i
using PanaView, refer to Appendix C of the Programming Manual.
DigitalFlow™ XMT868i Startup Guide27
Chapter 2. Initial Setup
Magnetic Wand
ENTE RESCA PE
DigitalFlowTM XGF868i
2.2XMT868i Enclosure Keypad Program
Along with the 2-line, 16-character LCD, the XMT868i includes a 6-key magnetic keypad. The decal cutout for each
key contains a hall effect sensor, pushbutton switch and visible red LED. The magnetic wand used to activate a
magnetic key is found attached to the meter chassis below the front panel. An operator activates the key by pressing the
magnetic wand up to the glass lid over the desired key. The LED will light indicating a successful key press.
Note:The pushbutton switch will also act as a key press but is done with the glass cover removed. Do not use the
pushbutton switch in a hazardous area where the glass cover needs to be in place.
Use the magnetic keypad to navigate through the user program. The menu map may be followed in sequence, or the
four arrow keys may be used to scroll through the prompt screens. Figure 13 below shows the front of the XMT868i,
with magnetic keypad and magnetic wand.
Figure 13: XMT868i Magnetic Keypad and Wand
The following six keys on the keypad enable users to program the XMT868i:
•[Enter] - confirms choice of a specific option and data entry within the option
•[Escape] - allows users to exit from a specific option without entering unconfirmed data
•[] and [] - enable users to highlight a specific window in the display option or to scroll through a list of
options (parameters, letters, and numbers 0-9 as well as the negative sign and decimal point) in a menu
• [] and [] - enable users to scroll to a specific option, among choices in an option, or to a character in a text
entry.
When you power up the XMT868i, the display first shows the model name and software version:
XMT868i X3A.STD
28DigitalFlow™ XMT868i Startup Guide
GE
2.2 XMT868i Enclosure Keypad Program (cont.)
The meter then starts to display measured parameters.
CH1 VEL E1
10.00 Ft/s
Chapter 2. Initial Setup
To enter the Keypad Program, press the
[Escape] key, followed by the [Enter] key, and the [Escape] key again. Each
successive key must be entered within 10 seconds of the prior key.
As a guide in following the programming instructions in this chapter, the relevant portions of the Model XMT868i
menu map have been reproduced in Figure 14 on page 45 and Figure 15 on page 46. Proceed to the following sections
to enter data in the
IMPORTANT:If the keypad has not been pressed for 10 minutes, the XMT868i exits the Keypad Program and returns to
Channel or GLOBL menus.
displaying measurements. The meter retains any configuration changes that were confirmed with the
[Enter] key, and restarts as if the operator had completed the programming cycle.
2.3Entering Data in the Global Menu
To begin programming your meter, you must select the system units from the GLOBL menu as discussed below. Refer
to Figure 14 on page 45 and remember to record all programming data in Appendix B, Data Records.
Note:Refer to the Programming Manual for information on the other submenus in the
2.3.1Entering Global System Data
The GLOBL-SYSTMsubmenu is used to enter several general system parameters (i.e., English or metric units). For
units with 2 channels, this menu is also used to compute parameters such as the sum, difference or average of the
channel 1 and channel 2 signals. When calculating the
submenu is used. Any conflicting data entered in the
SUM, DIF or AVE readouts, data from the GLOBL-SYSTM
Channel Systm submenu is overridden.
GLOBL menu.
1. In the Keypad Program, scroll to
2. In the
3. In the
PROG menu, scroll toGLOBLand press [Enter].
Global PROGRAM menu, scroll to SYSTM and press[Enter].
PROGand press [Enter].
4. The first prompt asks for a Meter Message. Enter a numeric or text message (up to 16 characters) and press
[Enter].
5. Scroll to the desired System Units selection (either metric or English) and press
[Enter]. The XMT868i will
display all parameters and measurements in the designated units.
•For a single-channel XMT868i, the program returns to the Global PROGRAM menu. Scroll to CH1 or CH2 to enter
Channel menu discussed on page 34.
the
•For a 2-channel meter, proceed to step 6.
DigitalFlow™ XMT868i Startup Guide29
Chapter 2. Initial Setup
2.3.1aVolumetric Units
6. Scroll to the desired Volumetric Units for the flow rate display and press [Enter]. Ta bl e 5 below lists the
available volumetric units.
Table 5: Available Volumetric Units
EnglishMetric
GAL/S = gallons/secondL/S = Liter/sec
GAL/M = gallons/minuteL/M = Liter/min
GAL/H = gallons/hourL/H = Liter/hr
MGD = mega gallons/dayML/D = mega liters/day
ft3/s = Cubic ft/secm^3/s = Cubic m/sec
ft3/m = Cubic ft/minm^3/m = Cubic m/min
ft3/h = Cubic ft/hrm^3/h = Cubic m/hr
ft3/d = Cubic ft/daym^3/d = Cubic m/day
Mft3/d = mega cubic feet/dayMm^3/d = mega cubic meters/day
BBL/S = Barrel/secBBL/S = Barrel/sec
BBL/M = Barrel/minBBL/M = Barrel/min
BBL/H = Barrel/hrBBL/H = Barrel/hr
BBL/D = Barrel/day BBL/D = Barrel/day
MBL/D = mega barrels/dayMBL/D = mega barrels/day
A-I/S = Acre-inch/sec
A-I/M = Acre-inch/min
A-I/H = Acre-inch/hr
A-I/D = Acre-inch/day
A-F/S = Acre-feet/sec
A-F/M = Acre-feet/min
A-F/H = Acre-feet/hr
A-F/D = Acre-feet/day
7. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric
flow display, and press [Enter].
30DigitalFlow™ XMT868i Startup Guide
Chapter 2. Initial Setup
2.3.1bTotalizer Units
8. Scroll to the desired Totalizer Units for the totalized flow rate display (see Tab le 6 below), and press [Enter].
Table 6: Totalizer Units
EnglishMetric
GAL = GallonsL = Liters
MGAL = mega gallonsML = Megaliters
ft^3 = Cubic feetm^3 = Cubic meters
Mft^3 = mega cubic feetMm^3= mega cubic meters
BBL = BarrelsBBL = Barrels
MBBL = mega barrelsMBBL = mega barrels
AC-IN = Acre-inches
AC-FT = Acre-feet
9. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow
display, and press
[Enter].
The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for
details) or the Energy Option in one of the
Channel PROGRAM options.
•If you have activated Mass Flow, continue to step 1 on page 32.
•If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 33.
•If you have activated neither option, the XMT868i returns to the Global PROGRAM window. Press [Escape] to
return to the
PROG menu. Then scroll to CH1 or CH2 and press [Enter] to program the Channel menu.
DigitalFlow™ XMT868i Startup Guide31
Chapter 2. Initial Setup
2.3.1cProgramming Mass Flow Data
1. Scroll to the desired Mass Flow Units for flow rate display (see Ta bl e 7 below), and press [Enter].
Table 7: Available Mass Flow Units
EnglishMetric
LB = PoundsKG = Kilograms
KLB = KiloPounds (Thousands of Pounds) Tonne = Metric Tons (1000 KG)
MMLB = MegaPounds
TONS (2000 LB)
2. Scroll to the desired Mass Flow Time Units (from seconds to days), and press
[Enter].
3. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate
display, and press
4. Scroll to the desired Mass Totals (units for the totalized mass flow rate display and press
[Enter].
[Enter].
5. Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point) in the totalized
mass flow rate display, and press
[Enter].
The program varies, depending on whether or not you have enabled the Energy Option.
•If you have not enabled the Energy Option, the XMT868i returns to the Global PROGRAM window. Press
[Escape] to return to the PROG menu and scroll to the Channel PROGRAM.
•If you have enabled the Energy Option, continue to the next page.
32DigitalFlow™ XMT868i Startup Guide
2.3.1dProgramming the Energy Option
1. Scroll to the desired Power Units (see Table 8 below) and press [Enter].
Table 8: Available Power Units
EnglishMetric
kBTUh = Thousands of BTUs/hourkCALs = Kilocalories/second
MMBTU = Millions of BTUs/hourkCALm = Kilocalories/minute
kW = KilowattkCALh = Kilocalories/hour
TONS = Tons of refrigerant kCALd = Kilocalories/day
MCals = Megacalories/second
MCALm = Megacalories/minute
MCALh = Megacalories/hour
MCALd = Megacalories/day
kW = Kilowatt
MW = Megawatt
Chapter 2. Initial Setup
2. Scroll to the desired number of Power Decimal Digits, and press
3. Scroll to the desired Total Energy Units (see Ta bl e 9 below), and press
[Enter].
[Enter].
Table 9: Available Total Energy Units
EnglishMetric
kBTU = Thousands of BTUskCAL = Kilocalories
MMBTU = Millions of BTUsMCAL = Megacalories
kWHr = Kilowatt-hourkWHr = Kilowatt-hour
TONS = Tons (of refrigerant)MWHr = Megawatt-hour
4. Scroll to the desired number of Energy Decimal Digits, and press
[Enter].
You have completed entering selections in this option.The program returns to the
CH1 or CH2 and press [Enter]to program the Channel menu on the next page.
Global PROGRAM menu. Scroll to
DigitalFlow™ XMT868i Startup Guide33
Chapter 2. Initial Setup
2.4Entering Data in the Channel Menu
The Channel menu is used to enter data specific to each channel. Refer to Figure 14 on page 45 and remember to
record all programming data in Appendix B, Data Records.
Note:In this manual, only the programming of Channel 1 will be described. To program Channel 2 of a 2-Channel
meter, simply repeat the same procedures presented for Channel 1.
2.4.1Selecting the Channel Measurement Method
The ACTIV submenu permits selection of the desired measurement method.
1. In the Keypad Program, scroll to
2. In the
PROG menu, scroll to the desired channel (CH1 or CH2) and press [Enter].
PROG and press [Enter].
Note:In the 1-Channel XMT868i, CH2 does not appear.
3. In the
4. The option offers two choices:
Note:In the 1-Channel XMT868i,
The XMT868i returns to the
Channel PROGRAMmenu, scroll to ACTIVand press [Enter].
OFF and TRANS (Transit-Time). Scroll to the desired option for this channel
and press
[Enter].
OFF does not appear.
Channel PROGRAM menu. Scroll to the SYSTMoption.
34DigitalFlow™ XMT868i Startup Guide
Chapter 2. Initial Setup
2.4.2Entering Data in the Channel System Option
1. From the Channel PROGRAM menu, scroll to the SYSTM option and press [Enter].
2. The first prompt asks for the Channel Label.Use the four arrow keys to enter the desired label (in any numeric
or text combination up to three characters) in the right pane, and press
3. The next prompt asks for the Channel Message. Enter the desired text in the same manner as the channel label,
and press
[Enter].
4. The program now asks if you want the Energy Option, which enables flow measurement in a heating or cooling
system. Scroll to Off or On, and press
[Enter].
5. The next prompt asks for the Volumetric Units. Scroll through the available units (see Tab le 1 0 below) and
[Enter] when you reach the desired unit.
press
[Enter].
Note:Refer to the
GAL/S - U.S. Gallons per Second L/S = Liters per Second
GAL/M - U.S. Gallons per MinuteL/M = Liters per Minute
GAL/H - U.S. Gallons per HourL/H = Liters per Hour
MGD - Mega U.S. Gallons per DayML/D = MegaLiters per Day
ft3/s - Cubic Feet per Secondm^3/s = Cubic Meters per Second
ft3/m - Cubic Feet per Minutem^3/m = Cubic Meters per Minute
ft3/h- Cubic Feet per Hourm^3/h = Cubic Meters per Hour
ft3/d- Cubic Feet per Daym^3/d = Cubic Meters per Day
Mf3/d- Mega Cubic Feet per DayMm3/d = Mega Cubic Meters per Day
BBL/S - Barrels per SecondBBL/S = Barrels per Second
BBL/M - Barrels per MinuteBBL/M = Barrels per Minute
BBL/H - Barrels per HourBBL/H = Barrels per Hour
BBL/D - Barrels per DayBBL/D = Barrels per Day
MBL/D - Mega Barrels per DayMBL/D = Mega Barrels per Day
A-I/S-Acre-inches per secondN/A
A-I/M -Acre-inches per minuteN/A
A-I/H -Acre-inches per hourN/A
A-I/D -Acre-inches per dayN/A
A-F/S -Acre-feet per secondN/A
A-F/M-Acre-feet per minuteN/A
A-F/H -Acre-feet per hourN/A
A-F/D -Acre-feet per dayN/A
SYSTMoption of the GLOBL menu to choose between English and metric measurements.
Table 10: Available Volumetric Units
English Volumetric UnitsMetric Volumetric Units
DigitalFlow™ XMT868i Startup Guide35
Chapter 2. Initial Setup
2.4.2Entering Data in the Channel System Option (cont.)
6. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric
flow display, and press
[Enter].
7. Scroll to the desired Totalizer Units for the totalized flow rate display (see Tab le 11 below), and press
Table 11: Totalizer Units
EnglishMetric
GAL - U.S. GallonsL - Liters
MGAL - Mega U.S. gallonsML - Megaliters
ft^3 - Cubic Feetm^3- Cubic Meters
Mft^3 - Mega Cubic FeetMm^3- Mega Cubic Meters
BBL - BarrelsBBL - Barrels
MBBL - MegabarrelsMBBL - Megabarrels
AC-IN - Acre-inches
AC-FT - Acre-feet
8. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow
rate display, and press
The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for
details) or the Energy Option.
[Enter].
[Enter].
•If you have activated Mass Flow, continue to step 1 on the next page.
•If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 38.
•If you have activated neither option, the XMT868i returns to the Channel PROGRAM window. Proceed to the
PIPE option.
36DigitalFlow™ XMT868i Startup Guide
2.4.2aProgramming the Mass Flow Option
1. Scroll to the desired Mass Flow Units for flow rate display (see Ta bl e 1 2 below), and press [Enter].
Table 12: Available Mass Flow Units
EnglishMetric
LB = PoundsKilograms
KLB = KiloPounds (Thousands of Pounds) Tonnes = Metric Tons (1000 KG)
MMLB = MegaPounds
TONS (2000 LB)
Chapter 2. Initial Setup
2. Scroll to the desired Mass Flow Time Units (from seconds to days),and press
[Enter].
3. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate
display, and press
4. Scroll to the desired Mass Totals (units for the totalized mass flow rate display) and press
[Enter].
[Enter].
5. Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point) in the totalized
mass flow rate display, and press
[Enter].
The program varies, depending on whether or not you have enabled the Energy Option.
•If you have not enabled the Energy Option, the XMT868i returns to the Channel PROGRAM window. Proceed
to the
PIPE option.
•If you have enabled the Energy Option, continue to the next page.
DigitalFlow™ XMT868i Startup Guide37
Chapter 2. Initial Setup
2.4.2bProgramming the Energy Option
1. Scroll to the desired Power Units (see Table 13 below) and press [Enter].
Table 13: Available Power Units
EnglishMetric
kBTUh = Thousands of
BTUs/hour
MMBTU = Mega BTUs/hourkCALm = Kilocalories/minute
kW = KilowattkCALh = Kilocalories/hour
TONS = Tons (2000 LB)kCALd = Kilocalories/day
kCALs = Kilocalories/second
MCALs = Megacalories/second
MCALm = Megacalories/minute
MCALh = Megacalories/hour
MCALd = Megacalories/day
kWATT = Kilowatt
MWATT - Megawatt
2. Scroll to the desired number of Power Decimal Digits, and press
3. Scroll to the desired Total Energy Units (see Ta bl e 1 4 below), and press
[Enter].
[Enter].
Table 14: Available Total Energy Units
EnglishMetric
kBTU = Thousands of BTUs/hrkCAL = Kilocalories
MMBTU = Mega BTUs/hourMCAL = Megacalories
kWHr = Kilowatt-hourkWHr = Kilowatt-hour
TONS = Tons (2000 LB)MWHr - Megawatt-hour
4. Scroll to the desired number of Energy Decimal Digits, and press
5. Scroll to highlight if you are using a Heating or Cooling System, and press
[Enter].
[Enter].
6. The final prompt asks whether you want to conduct Flow Measurement at the point of
where the liquid exits) or of
choice, and press
[Enter].
You have completed entering data in the
to the next page to enter
PIPE parameters.
SPPLY (Supply = the point where the liquid enters). Scroll to the appropriate
SYSTM option. The meter returns to the Channel PROGRAM menu. Proceed
RTN (Return = the point
38DigitalFlow™ XMT868i Startup Guide
Chapter 2. Initial Setup
2.4.3Entering Pipe Parameters
Enter the transducer and pipe parameters using the PIPE submenu. While following the programming instructions,
refer to Figure 15 on page 46.
1. From the
2. The first prompt asks for the Transducer Number.
Channel PROGRAM menu, scroll to the PIPE option and press [Enter].
•For a standard transducer,enter the number engraved on the transducer head in the right pane, and press
[Enter].
•If there is no number engraved on the transducer head, press the right arrow key to scroll to the Special
option, enter an assigned number (from 91 to 99), and press
IMPORTANT:Special transducers, which have no engraved number on the head, are rarely used. Examine the
transducer head carefully for a number.
The menu now varies, depending on your selection in step 2:
[Enter].
•If you entered the number for a standard clamp-on transducer, proceed to the Pipe Material prompt in step 4.
•If you entered the number for a standard wetted transducer, proceed to the Pipe OD prompt in step 5.
•If you entered the number for a special transducer, proceed to step 3 below.
2.4.3aSpecial Transducers
Note:For special transducers, GE will supply a transducer data sheet with programming information.
3. For special transducers:
a. The program first asks for the Wedge Type. Press the appropriate transducer type (Rayleigh clamp-on,
shear clamp-on, or wetted).
Note:GE will supply the information required for steps b through f with the transducers.
b. Press the appropriate Frequency (from 500 kHz to 5 MHz).
c.Enter the Time Delay (Tw) value supplied by GE and press
Note:If you have selected a special wetted transducer, the program now goes to step 5. But if you have selected a
clamp-on transducer, three more prompts appear.
d. Enter the Wedg e A ngl e of the transducer (the angle of the ultrasonic transmission, in degrees) and press
[Enter].
e.Enter the Wedge Temperature and press
f.Enter the Wedge Soundspeed and press
[Enter].
[Enter].
[Enter].
DigitalFlow™ XMT868i Startup Guide39
Chapter 2. Initial Setup
2.4.3bPipe Material
4. Scroll to the appropriate Pipe Material from the supplied list. For some materials (see Ta bl e 15 below), the
program displays additional selections. At the desired choice, press
OTHEREnter the soundspeed of the pipe material and
press
[Enter]. If the soundspeed is unknown,
refer Sound Speeds and Pipe Size Data (GE
document 914-004).
[Enter].
40DigitalFlow™ XMT868i Startup Guide
Chapter 2. Initial Setup
2.4.3cPipe OD
5. To select the appropriate Pipe OD Unit type (see Tabl e 16 below), scroll to the right side of the screen, and use
the up and down arrow keys to step through the list. Press
or circumference on the left side and press
[Enter].
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds and Pipe Size Data (GE document 914-004).
Table 16: Available Pipe OD Units
EnglishMetric
inch mm = millimeters
feetm = meters
in/PI = pipe circumference in inchesmm/PI = pipe circumference in millimeters
ft/PI = pipe circumference in feetm/PI = pipe circumference in meters
[Enter]. Then enter the known pipe outside diameter
6. Use the arrow keys to enter the known Pipe Wall Thickness (in inches or mm) and press
[Enter].
Note:If you are using clamp-on transducers, skip to step 9 below.
2.4.3dPath and Axial Lengths
7. To enter the Path Length:
a. Use the [] arrow key to highlight the path length unit type at the right of the screen.Then use the [] and
[] arrow keys to scroll to the desired unit type.
b. Use the [] arrow key to return to the numeric entry on the left, and enter the path length of the ultrasonic
signal. Press
Note:If a spoolpiece was ordered with the meter, the transducer signal path length
axial length
[Enter].
(P) and the transducer signal
(L) are engraved on the flowcell and/or are included in the documentation supplied with the meter.
For on-site transducer installations, refer to Appendix C, Measuring P and L Dimensions, for instructions.
8. In the same manner, enter the appropriate Axial Length L unit type and axial length of the ultrasonic signal, and
[Enter]. Skip to step 10 on page 42.
press
2.4.3ePipe Lining
9. Scroll to the appropriate choice to indicate whether you have a Pipe Lining.
a. If you have selected Yes , then select the Lining Material from the supplied list.
•If OTHER is selected, enter the lining sound speed, and press [Enter].
b. Then enter the lining thickness and press
DigitalFlow™ XMT868i Startup Guide41
[Enter].
Chapter 2. Initial Setup
2.4.3fTracking Windows
10. Press the appropriate choice to indicate whether you want Tracking Windows. These windows are used to
detect the receive signal when you are unsure of the fluid sound speed.
2.4.3gFluid Type
11. Press the appropriate Fluid Type. The selections for fluid type vary depending on whether:
•the Energy Option is ON or OFF; and
•the Tracking Window is enabled or disabled.
Refer to Table 17 below if Energy Option is
OFF, or to Ta bl e 18 below if Energy Option is ON.
Note:Some of the fluid types may require additional selections, as shown in the following tables.
Table 17: Fluid Types for ENERGY OFF
Tracking Windows =
OFFAdditional SelectionsONAdditional Selections
WaterNormal Water
Sea Water
Wate r,
0-100C
No additional selections required.
Note: If Normal Water is
selected, enter the Water
Temperature and press [Enter].
OilLube Oil
Crude Oil
Methanol (25 C) No additional selectionsOil,
Wate r,
0-260C
No additional selections required.
No additional selections required.
Tracking
Ethanol (25C)No additional selectionsOTHEREnter Min. Soundspeed and press
enter Max. Soundspeed and press
LN2No additional selections
FreonNo additional selections
OTHEREnter the fluid soundspeed and
[Enter].
press
[Enter]. Then
[Enter].
Table 18: Fluid Types for ENERGY ON
Tracking Windows =
NOAdditional SelectionsYESAdditional Selections
WaterEnter Water Temperature and press
[Enter].
Water,
Glycol
Mixture
Enter Fluid Soundspeed and press
[Enter]. Then enter Percentage of
Water and press [Enter].
OTHEREnter Fluid Soundspeed and press
Wate r,
0-260C
Wate r,
Glycol
Mixture
OTHEREnter Mini. Soundspeed and press
[Enter].
42DigitalFlow™ XMT868i Startup Guide
No additional selections
Enter Percentage of Water and press
[Enter]. Then
enter Max. Soundspeed and press
[Enter].
[Enter].
Chapter 2. Initial Setup
2.4.3hReynolds Correction
12. Scroll to the appropriate choice to indicate whether you want Reynolds Correction.
•If Off is selected, enter the Calibration Factor and press [Enter].
•If Activ is selected, the program asks for KV Input Selection. Scroll to SIGS (Signals), SOUND
(Soundspeed) or STATC.
•If SIGS or SOUND is selected, enter the Calibration Factor and press [Enter].
•If STATC is selected, enter the Kinematic Viscosity and press [Enter]. Then enter the Calibration
Factor and press
[Enter].
•For wetted transducers, the program returns to the Channel PROGRAM menu. You have completed entering
PIPE parameters.
•For clamp-on transducers, proceed to step 13 below.
Number of Traverses
13. Scroll to the appropriate Number of Traverses (from 1 to 5) and press
14. Accept the value for the Transducer Spacing by pressing
Note:The XMT868i automatically calculates the transducer spacing. You should not enter a different value unless
specifically instructed by the factory.
The User Program returns to the
Note:The site data programmed in this chapter is automatically stored in the meter’s non-volatile RAM (NVR). If the
power to the meter is disconnected, the same site data will be applied when the unit is next powered On.
Channel PROGRAM menu. You have completed entering PIPE parameters.
[Enter]. Record the value.
[Enter].
2.4.3iProcedure Options
Proceed to Chapter 3, Operation, for instructions on taking measurements. Or refer to the Programming Manual for
instructions on programming the Model XMT868i’s advanced features.
DigitalFlow™ XMT868i Startup Guide43
Chapter 2. Initial Setup
[no content intended for this page]
44DigitalFlow™ XMT868i Startup Guide
[Esc][Enter][Esc]
PROGRESETCALIBCONTRAST
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
PIPEI/OSETUPSYSTMACTIV
Disabled*Transit
ENERGY OPTION
SITE/CHANNEL LABEL
SITE/CHANNEL MESSAGE
MetricEnglish
Liter/SLiter/MLiter/H
m3/hMm3/dBBL/S
MLiter/Dm3/sm3/m
BBL/MBBL/HMBL/D
VOL DECIMAL DIGITS
TOTALIZER UNITS
0
12
3
TOTAL DECIMAL DIGITS
MetricEnglish
GAL
MGALft^3Mft^3
English
LB
KLBMMLBTONSKGTONNES
/SEC/MIN/HR/DAY
MASS FLOW TIME
Metric
MGDft3/sft3/m
BBL/MBBL/HMBL/D
GAL/SGAL/MGAL/H
ft3/hMft3/dBBL/S
A-I/SA-I/M
A-F/HA-F/D
A-I/DA-F/SA-F/MA-I/H
OFFON
VOLUMETRIC UNITS
BBL
MBBLAC-INAC-FT
LiterMliterm^3
Mm^3
BBLMBBL
0
12
3
MDOT DECIMAL DIGITS
MetricEnglish
KG
TONNES
MASS DECIMAL DIGITS
0
12
3
MASS TOTALS
MASS FLOW
LB
KLBMMLB
TONS
MASS FLOW ON MASS FLOW OFF
kBTU
MMBTU kWHrTONSKCal/sMCal/sKWMW
MetricEnglish
POWER
0
123
012
3
POWER DECIMAL DIGITS
KBTU
MMBTU
KWHrTONSKCalMCalKWHrMWHr
MetricEnglish
ENERGY (TOTAL)
0
123
ENERGY OPTION ONENERGY OPTION OFF
ENERGY OPTION ONENERGY OPTION OFF
ENERGY DECIMAL DIGITS
HEATING or COOLING
FLOW MEASUREMENT
COOL
HEAT
RTNSPPLY
SYSTEM UNITS
METER MESSAGE
MetricEnglish
VOL DECIMAL DIGITS
TOTALIZER UNITS
0
12
3
TOTAL DECIMAL DIGITS
MetricEnglish
GAL
MGALft^3Mft^3
English
LB
KLBMMLBTONSKGTONNES
/SEC/MIN/HR/DAY
MASS FLOW TIME
Metric
MGDft3/sft3/m
BBL/MBBL/HMBL/D
GAL/SGAL/MGAL/H
ft3/hMft3/dBBL/S
A-I/SA-I/M
A-F/HA-F/D
A-I/DA-F/SA-F/MA-I/H
ENGMETRC
VOLUMETRIC UNITS
BBL
MBBLAC-INAC-FT
LiterMliterm^3
Mm^3
BBLMBBL
0
12
3
MDOT DECIMAL DIGITS
MetricEnglish
KG
TONNES
MASS DECIMAL DIGITS
0
12
3
MASS TOTALS
MASS FLOW
LB
KLBMMLB
TONS
MASS FLOW ON MASS FLOW OFF
kBTU
MMBTU kWHrTONSKCal/sMCal/sKWMW
MetricEnglish
POWER
0
123
012
3
POWER DECIMAL DIGITS
KBTU
MMBTU
KWHrTONSKCalMCalKWHrMWHr
MetricEnglish
ENERGY (TOTAL)
0
123
ENERGY OPTION ONENERGY OPTION OFF
ENERGY OPTION ONENERGY OPTION OFF
ENERGY DECIMAL DIGITS
HEATING or COOLING
FLOW MEASUREMENT
COOL
HEAT
RTNSPPLY
Liter/SLiter/MLiter/H
m3/hMm3/dBBL/S
MLiter/Dm3/sm3/m
BBL/MBBL/HMBL/D
2-CHANNEL UNITS
1-CHANNEL UNITS
See Figure 2-3
Chapter 2. Initial Setup
DigitalFlow™ XMT868i Startup Guide45
Figure 14: Channel-ACTIV, SYSTM and GLOBL-SYSTM Menu Map
[Esc][Enter][Esc]
PROGRESETCALIBCONTRAST
CH2*GLOBLCH1
(* for 2-Channel meter only)
I/OCOMMSYSTM
PIPEI/OSETUPSYSTMACTIV
RAYLSHEARWETTD
STDSPEC
CuAlBRASSCuNiGLASS PLSTC OTHER
See Figure 2-2See Figure 2-2
TRANSDUCER NUMBER
SPECIAL TRANSDUCER
WEDGE TYPE
1MHz2MHz5MHz500k
FREQUENCY
(RAYL/SHEAR)
STEELIRON
WEDGE SOUNDSPEED
WEDGE ANGLE
PIPE MATERIAL
10%Ni
PYREXFLINTNYLON POLYE POLYPCAST
30%Ni
DUCTCARBNSS
PVCACRYL
PIPE SOUNDSPEED
feet
cir.in
cir.ft
m
cir.mmcir.minch
mm
PIPE OD
(Metric)
(English)
feet
cir.in
cir.ft
m
cir.mmcir.minch
mm
PIPE OD
(Metric)
(English)
(WETTD)
(Clamp-on Type)(Wetted Type)
feet
mmm
inch
PIPE WALL
PATH LENGTH P
(English)
(Metric)
feet
mmm
inch
AXIAL LENGTH L
(English)
(Metric)
PIPE WALL
LINING
YES
NO
TARPYREX ASBES
RUBBR
TEFLN OTHERMORTR
LINING MATERIAL
LINING SOUNDSPEED
LINING THICKNESS
YESNO
TRACKING WINDOWS?
WATER
MIXEDOTHER
WATER TEMPERATUREFLUID SOUNDSPEEDFLUID SOUNDSPEED
Percentage of Water
W260
MIXEDOTHER
MINIMUM SOUNDSPEEDPercentage of Water
MAXIMUM SOUNDSPEED
OFFACTIVE
SOUND STATCSIGS
KV INPUT SELECTION
KINEMATIC VISCOSITY
REYNOLDS CORRECTION
CALIBRATION FACTOR
(Wetted Transducers)
(Clamp-On Transducers)
2(V)
34
1(Z)
5
NUMBER OF TRAVERSES
TRANSDUCER SPACING
Energy Option OFF
Energy Option ON
OIL
METHETHFREON OTHERWATERLN2
SEA
LUBECRUDENORML
W260
OILOTHER
W100
NOYES
TRACKING WINDOWS?
MINIMUM SOUNDSPEED
MAXIMUM SOUNDSPEED
FLUID TYPE
WATER
FLUID SOUNDSPEED
OFF
ACTIVE
SIGSSOUND STATC
KV INPUT SELECTION
REYNOLDS CORRECTION
KINEMATIC VISCOSITY
CALIBRATION FACTOR
(Wetted Transducers)(Clamp-On Transducers)
1(Z)
2(V)
4
5
3
NUMBER OF TRAVERSES
TRANSDUCER SPACING
TEMPERATURE
See Figure 2-2
FLUID TYPE
FLUID TYPE
Chapter 2. Initial Setup
Figure 15: Channel-PIPE Menu Map
DigitalFlow™ XMT868i Startup Guide46
Chapter 3. Operation
Chapter 3.Operation
3.1Introduction
See Chapter 1, Installation, and Chapter 2, Initial Setup, to prepare the Model XMT868i system for operation. When
the meter is ready to take measurements, proceed with this chapter. The following specific topics are discussed:
•Powering Up
•Using the Display
•Taking Measurements
•Pausing Measurement
Note:All inputs and outputs of the Model XMT868i are calibrated at the factory, prior to shipment. If it becomes
necessary to recalibrate any of the inputs and/or outputs, see Chapter 1, Calibration, of the Service Manual for
instructions.
WARNING!To ensure the safe operation of the Model XMT868i, it must be installed and operated
as described in this manual. In addition, be sure to follow all applicable local safety codes and
regulations for the installation of electrical equipment.
DigitalFlow™ XMT868i Startup Guide47
Chapter 3. Operation
3.2Powering Up
Because the Model XMT868i does not have an ON/OFF switch, it will power up as soon as the connected power
source is energized.
Note:For compliance with the European Union’s Low Voltage Directive, this unit requires an external power
disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
There are three methods for obtaining readings from the XMT868i:
•Built-in LCD display
•PanaView software on a computer
•Device to read the XMT868i’s analog output
At least one of the above display options must be installed in order to obtain flow rate readings from the meter.
Immediately upon power up the software version display appears. Then, the meter performs a series of internal checks,
which take about 45 seconds, prior to displaying the flow rate data.
CAUTION!The XMT868i conducts the FRAM (non-volatile RAM) test only on the first powering
up. If the screen displays the message, “Program Memory Fault Detected,” consult the factory
immediately. Otherwise, the XMT868i will not repeat the test and may not properly save
programmed values.
Note:If the Model XMT868i fails any of the internal checks (see Chapter 2, Error Codes, in the Service Manual), try
disconnecting the power and then re-powering the unit. If the meter continues to fail any of the internal checks,
contact the factory for assistance.
After successfully performing the internal checks, the Model XMT868i begins taking measurements and the software
version display is replaced by a measurement mode display. Proceed to the appropriate section for instructions on using
the LCD display and the PanaView display option.
Note:As a minimum, the system and pipe parameters (for each installed channel of a 2-channel meter) must be
entered before the Model XMT868i can display valid data. Refer to Chapter 2, Initial Setup, for specific
instructions.
48DigitalFlow™ XMT868i Startup Guide
Chapter 3. Operation
VEL
XMT868i
CH1 VOL
4500GAL/H
Parameter
Units
Channel #
Flow Rate
3.3The LCD Display
The components of the LCD display are shown in Figure 16 below, along with a typical mass flow rate readout.
Figure 16: A Typical LCD Flow Rate Display
The display screen includes the following information:
•Channel Number
•Flow Parameter
•Units of Measure
•Flow Rate Value
The example in Figure 16 above uses the default display settings. However, the first three items in the list above may
be reprogrammed to display a variety of alternative choices. See the Programming Manual for detailed instructions on
programming these parameters.
Note:The LCD backlight flashes upon signal errors. If the backlight is off when an error is detected, the display is
illuminated briefly. If the backlight is already on, the light is interrupted briefly. Error code messages may
appear in the upper right corner of the LCD display. For information about these error codes and how to
respond to them, refer to Chapter 2, Error Codes, in the Service Manual.
Proceed to the Taking Measurements section to use the XMT868i.
DigitalFlow™ XMT868i Startup Guide49
Chapter 3. Operation
3.4The Optional PanaView Display
The components of the PanaView text display appear in Figure 17 below, along with a typical flow rate readout.
Figure 17: A Typical PanaView Text Display Window
The text display window includes the following information:
•Channel Number
•Flow Parameter
•Units of Measure
•Flow Rate Value
The example in Figure 17 above is typical, but the first three items in the list above may be reprogrammed to display a
variety of alternative choices. See the next section for detailed instructions on programming these parameters.
Note:Error code messages may appear in the lower left corner of the PanaView text display window. For
information about these error codes and how to respond to them, refer to Chapter 2, Error Codes, in the
Service Manual.
Proceed to the Taking Measurements section to use the XMT868i.
50DigitalFlow™ XMT868i Startup Guide
Chapter 3. Operation
3.5Taking Measurements
The Model XMT868i is capable of displaying several different variables in a variety of formats. However, this manual
will only discuss the basic measurement displays using the LCD display or the PanaView display. Refer to Chapter 2,
Displaying Data, in the Programming Manual for instructions on setting up alternate choices. Also, refer to the
Programming Manual and/or the PanaView Installation Guide to use PanaView or the analog outputs to obtain the
flow rate data.
3.5.1Programming the LCD
Note:When you first initialize the XMT868i, the number of LCD parameters is set to OFF. You must program the
LCD to display any measured parameters.
Through the Keypad Program, you can program the LCD display to display up to four variables in sequence. Complete
the following steps to program the LCD display:
1. Power up the XMT868i and wait until it has initialized.
2. Press
3. In the Keypad Program window, scroll to
4. In the
[Escape]. [Enter], [Escape].
PROG and press [Enter].
PROGmenu, scroll to GLOBL and press [Enter].
5. Scroll to I/O and press [Enter].
6. Scroll to
7. The window now asks for the # of LCD Parameters. Scroll to the desired number (from
KEY) and press [Enter].
OFF setting switches the measurement display off, while the KEY setting enables users to change the measurement
The
display using the arrow keys, without accessing the Keypad Program. If you select
LCD and press [Enter].
OFF through 1-4 and
KEY:
•To view a parameter other than the one currently displayed, press the [] and [] keys to scroll through the
various parameters.
•To view the channel measurements on a two-channel XMT868i, press the [] or [] keys until you have
reached the desired channel and measurement.
For a 1-Channel XMT868i, the data for Channel 1 is displayed automatically, and you can skip to step 9. However, for
a 2-Channel meter, the channel data to be displayed must be specified at the following prompt.
DigitalFlow™ XMT868i Startup Guide51
Chapter 3. Operation
3.5.1Programming the LCD (cont.)
8. Scroll to the desired Channel Option from the list in Table 19 below.
Table 19: Channel Options
OptionDescription
CH1Channel 1
CH2Channel 2
SUMCH1+CH2
DIFCH1-CH2
AVE(CH1+CH2)/2
9. For each channel, select the desired Measurement Parameter from the list in Tabl e 20 below.
Table 20: Output Measurement Options
Option Bar ChoiceDescription
VEL
VOLUM
+TOTL
-TOTLReverse Totalized Volume Flow
TIME Total Flow Measurement Time
Flow Velocity
Volumetric Flow
Forward Totalized Volume Flow
MDOTMass Flow
+MASSForward Totalized Mass Flow
-
MASSReverse Totalized Mass Flow
POWEREnergy Flow Power
+ENRGForward Energy Flow
-ENRGReverse Energy Flow
SS upSignal strength for the upstream transducer.
SS doSignal strength for the downstream transducer.
SoundspeedMeasured speed of sound in the fluid.
DAC upAGC DAC count for upstream gain setting.
DAC doAGC DAC count for downstream gain setting.
T upUpstream ultrasonic signal transit time.
TdownDownstream ultrasonic signal transit time.
DELTATransit time difference between upstream and downstream signals.
REYN#Current velocity/reynolds number.
K (RE)K factor, based on the Reynolds number.
PEAK%Percentage of peak (set to +50 by default).
THETATheta 3 is the angle between the ultrasonic beam and the normal to the pipe wall. It is
calculated from measured transit time and the clamp-on parameters (clamp-ons only).
52DigitalFlow™ XMT868i Startup Guide
Table 20: Output Measurement Options (cont.)
Option Bar ChoiceDescription
QupDisplays the signal quality for the upstream transducer.
QdownDisplays the signal quality for the downstream transducer.
AMPupDisplays the value for the signal amplitude of the upstream transducer.
AMPdnDisplays the value for the signal amplitude of the downstream transducer.
P#upDisplays signal peaks for the upstream transducer.
P#dnDisplays signal peaks for the downstream transducer.
NFupDisplays the normalization factor for the upstream transducer.
NFdnDisplays the normalization factor for the downstream transducer.
CxdcrDisplays the speed of sound in the transducer (clamp-on only).
TEMPsDisplays the temperature for the supply input (energy measurement).
TEMPrDisplays the temperature for the return input (energy measurement).
Ts-TrDifference between supply input and return input measurement (energy measurement).
DENSsDisplays the density for the supply input (energy measurement).
Chapter 3. Operation
DENSrDisplays the density for the return input (energy measurement).
DELThDisplays the delta enthalpy which is the difference between the supply and return. Enthalpy is
a measure of energy contained in the fluid (energy measurement).
Note:The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in
this section. Also, when differences in one channel's programming invalidate an output previously chosen for
the other, the measurement defaults to the nearest selectable item in the parameter list.
The previous two prompts repeat until all of the specified
display parameters have been set up, the meter returns to the
[Escape] key three times.
the
# of LCD Parameters have been set up. When all of the
Global I/O window. To leave the Keypad Program, press
After leaving the Keypad Program, the XMT868i will reset itself and will begin to display the parameters specified in
this section. If more than one parameter was set up, each of the parameters will be displayed in sequence, with a pause
of several seconds between display changes.
3.5.2Using the LCD Display
To use the programmed LCD display for obtaining flow rate data, simply power on the XMT868i as described earlier in
this chapter. Then, read the flow rate directly from the display, as shown in Figure 16 on page 49.
Note:See Chapter 2, Displaying Data, in the Programming Manual to customize the contents of the LCD display.
DigitalFlow™ XMT868i Startup Guide53
Chapter 3. Operation
3.5.3PanaView Display
Note:See Chapter 2, Initial Setup, in this manual and/or Chapter 1, Programming Site Data, in the Programming
Manual for complete instructions on entering startup data using PanaView.
Power up PanaView, establish communications with the XMT868i and enter the
required startup parameters, as
described in Chapter 2, Initial Setup. Then, proceed as follows:
1. In PanaView, pull down the Output menu (see Figure 18 below) and click on the Tex t D isp lay option.
Figure 18: The Output Menu
Note:The Text Disp l ay window that appears after this step is actually stacked on top of any previously opened
windows (such as the Meter Browser window).
2. Using the Window menu, as described in the PanaView Installation Guide, arrange the open windows in the
desired format. For this discussion, Figure 19 below shows the Tex t D i sp lay window in its maximized
(full-screen) size.
Figure 19: The Text Display Window
3. The left pane of the Text Display window contains the standard PanaView network tree. Expand the XMT
branch, and double-click on the desired channel. On 2-channel units, you can also display the
AVG parameters.
54DigitalFlow™ XMT868i Startup Guide
SUM, DIFF or
Chapter 3. Operation
3.5.3PanaView Display (cont.)
4. From the expanded tree, double-click on the desired flow parameter to display it in the right pane of the
window.
5. Before actual data values can be displayed in the text pane, activate one of the following data collection modes
(see Figure 19 on page 54):
•Click on the [Get Once] option button at the bottom of the right pane in the Text Display window. The
current value of the selected process parameter, as specified in the PanaView network tree, is displayed in
the right pane of the Text Display window.
or
•Enter an “interval” in the text box at the bottom of the right pane in the Te xt D is p la y window, or check the
“Max. Comm Rate” box to collect readings as fast as the system allows (1 sec). Then, click on the
[Continuous] option button to begin collecting data for display in the right pane of the Tex t D i sp lay
window.
Note:Any value entered in the “Interval” text box is overridden if the “Max. Comm Rate” box is checked.
The right pane now appears similar to Figure 17 on page 50.
6. If the
[Continuous] option was selected in Step 5 above, click on the [Stop] option button, which has replaced
the original
[Continuous] option button, to terminate data collection.
The Text Display window may be left open while other tasks are performed, or it may be closed by clicking on the
[X] control button at the far right of the menu bar.
lower
IMPORTANT:If you click on the upper [X] control button at the far right of the PanaView title bar, you will exit
PanaView completely.
DigitalFlow™ XMT868i Startup Guide55
Chapter 3. Operation
3.5.3aDisplaying Multiple Process Parameters
The procedure for displaying a single process parameter in a text screen may be repeated to simultaneously display
multiple process parameters. To do so, proceed as follows:
1. Display the first process parameter in a text screen, as described in the previous section.
2. Repeat Step 1 for any desired additional process parameters, by double clicking on them in the PanaView
network tree. PanaView automatically tiles the multiple text screens in the right pane of the Text Display
window (see Figure 20 below).
Figure 20: Multiple Text Screens in the Text Display Window
3. As in any standard Windows application, the multiple text screens may be resized by dragging their borders.
Also, the individual panes within a parameter’s text screen may be resized by dragging the borders within that
text screen.
4. To close an open text screen, right click anywhere within that screen (except in the title bar or the error section)
and click on the
[Remove] option that pops up in the context menu.
Note:After resizing or removing any of the multiple text screens, the default tiled layout may be restored by opening
the Window menu (see the PanaView Installation Guide) and clicking on the Tile Output Displays option.
56DigitalFlow™ XMT868i Startup Guide
Chapter 3. Operation
3.5.3bDisplaying Multiple Text Windows
The procedures for displaying one or more process parameters in a single Text Display window may be repeated to
open multiple Text Display windows. To do so, proceed as follows:
1. To open another Text Display window and display the desired process parameter(s) in the new window, repeat
the steps in PanaView Display.
2. Arrange the multiple Text Display windows as desired using the Window menu (see the PanaView Installation
Guide).
3.5.4Pausing Measurement
On occasion, it might become necessary for the XMT868i to stop taking measurements. Through PanaView, you can
direct the XMT868i to pause measurements without cutting off power from the meter, as follows:
1. From the meter tree in the New Meter Browser, click on the XMT entry.
2. Expand the Edit Functions option, and double-click on the Pause Measurement entry. A window similar to
Figure 21 below opens.
Figure 21: The Pause Measurement Window
3. To pause current measurements, double-click on the Stop measurement option. The window closes, and the
XMT868i stops taking measurements.
4. To restart measurement, double-click on the Pause measurement entry, and then on the Measure flow option.
The XMT868i resumes flow measurement.
DigitalFlow™ XMT868i Startup Guide57
Chapter 3. Operation
3.5.4aProcedure Options
This Startup Guide is intended to provide only those instructions necessary to get the XMT868i installed and operating.
By following the instructions in this chapter, the Model XMT868i can be set up to display the desired channel option
and the desired measurement parameter, using either the LCD display or PanaView.
To utilize the more advanced capabilities of the Model XMT868i, refer to the Programming Manual and/or the Service Manual for the instrument. Also, refer to the PanaView Installation Guide for instructions on using the PanaView
software with the XMT868i.
58DigitalFlow™ XMT868i Startup Guide
Chapter 4. Specifications
Chapter 4.Specifications
4.1General
The general specifications for the Model XMT868i flow transmitter are divided into the following categories.
4.1.1Hardware Configuration
Channel Options:
Standard: 1 Channel
Optional: 2 Channel (2 pipes or 2 paths per pipe)
Mode Options:
Transit-Time
Package Options:
Standard:
Epoxy-coated aluminum Type 7/4X, IP66
Explosion proof Class I, Div. 1, Groups B, C & D. Class II, Groups E, F & G; Class III
Flameproof II 2 GD EEx d IIC T5/T6
Optional:
Stainless steel Type 7/4X
Explosion proof Class I, Div. 1, Groups B, C & D. Class II, Groups E, F & G; Class III
Flameproof II 2 GD EEx d IIC T5/T6
Dimensions:
Standard housing: 10 lb (4.5 kg); 8.2 x 6.6 in. (208 x 168 mm)
Optional housing: 30 lb (13.6 kg); 8.2 x 6.6 in. (208 x 168 mm)
4.1.2Environmental
Operating Temperature:
-40° to 140°F (–40° to 60°C)
Storage Temperature:
-67° to 167°F (–55° to 75°C)
DigitalFlow™ XMT868i Startup Guide59
Chapter 4. Specifications
4.1.3Flow Accuracy (Velocity)
±0.5% of reading (achievable with process calibration)
Accuracy specifications assume a fully developed flow profile and a straight run of 10 pipe diameters upstream and
5 pipe diameters downstream.
Typical Clamp-On Flow Accuracy (Velocity)
Pipe Diameter (ID) > 6 in. (150 mm): ±1% to 2% of reading
Pipe Diameter (ID) <
Typical Wetted Flow Accuracy (Velocity):
±1% of reading
Note:Accuracy depends on pipe size, installation and whether measurement is one-path or two-path.
6 in. (150 mm): ±2 to 5% of reading
4.1.4Range
Transit-Time Mode:
-40 to 40 ft/s (-12.2 to 12.2 m/s)
4.1.5Rangeability (Overall)
400:1
4.1.6Repeatability
±0.1% to 0.3 of reading
4.1.7Energy Accuracy (% of Reading)
The accuracy of the energy measurement is a combination of the accuracy of the flow and temperature measurements.
Transit-Time Mode:
Accuracy of 1% is typical for calibrated systems
Note:Above specifications require a straight run of 10 pipe diameters upstream and 5 pipe diameters downstream.
4.1.8Fluid Types
Transit-Time Mode:
Acoustically-conductive fluids. This includes most clean liquids and many with entrained solids or gas bubbles (the
maximum void fraction depends on the transducer, interrogation carrier frequency, path length and pipe configuration).
Consult the factory for additional information.
60DigitalFlow™ XMT868i Startup Guide
Chapter 4. Specifications
4.2Electrical
The electrical specifications for the Model XMT868i flow transmitter are divided into the following categories.
4.2.1Power Supply
Options:
Standard: 95 to 240 VAC, 50/60 Hz, ± 10%
Optional: 12 to 28 VDC, ± 5%
Power Consumption:
20 W maximum
4.2.2Operating Modes
Flow Measurement:
Standard: Correlation Transit-Time™ Mode with clamp-on or wetted transducers
Energy Measurement (optional):
To calculate energy flow rate, use Analog Input Board for transmitters. Use RTD Input Board for direct RTDs.
4.2.3European Compliance
See the Certification & Safety Statements and the CE Declaration of Conformity at the end of this manual.
DigitalFlow™ XMT868i Startup Guide61
Chapter 4. Specifications
4.2.4Input/Output
Digital Display:
2-line x 16-character LCD display, LED backlight, configurable to display up to 4 measurement parameters in
sequence, or to scroll through measurement parameters
Digital Communications:
Standard: RS232 serial port for PC, terminal or printer
Optional: RS485 serial port for multi-user network
RS485 MODBUS communication
MODBUS/TCP
OPC Server
Ethernet
Foundation Fieldbus
Analog Outputs (built-in):
Standard: Two isolated 0/4–20 mA, 600 Ω max. load
Option Cards:
Optional Inputs/Outputs of the following types may be added:
•Analog Inputs: two isolated 0/4-20 mA inputs, 24-V loop power
•RTD Inputs: two isolated 3-wire RTD inputs;
Span –148° to 662°F (–100° to 350°C), 100 Ω Pt, Alpha 385
•Analog Outputs: two isolated 0/4–20 mA analog outputs, 1000 Ω maximum load
•Data Logging: 128 kB or 2 MB memory
•Totalizer or Frequency Outputs: two or four optically isolated pulse or frequency outputs,
100 VDC/3 A/1 W/10 kHz max.
Totalizer Mode: one pulse per defined unit of parameter (e.g., 1 pulse/gal)
Frequency Mode: pulse frequency proportional to magnitude of parameter (e.g., 10 Hz = 1 gal/min)
•Alarm Relays: two or four Form-C relays
General Purpose: 120 VAC, 28 VDC max., 5 A max., DC: 30 W max., AC: 60 VA max.
Note:The above optional inputs/outputs are available only in specific combinations. Consult GE or see Table 22 on
page 69 for details.
62DigitalFlow™ XMT868i Startup Guide
4.2.5Additional Options
PC-Interface Software:
PanaView™ software option links the XMT868i to a PC. Software package includes CD and manual.
4.3Transducer
4.3.1Wetted Transducers
Material:
Standard: 316 Stainless Steel
Chapter 4. Specifications
Optional (for isolating PanAdapta Plugs
): Hastelloy, Monel, Titanium, Duplex, CPVC, PVDF and others
Process Connection:
Standard: 1-in. NPTM, 3/8-in. NPTM
Optional: RF flange, socket weld, fuse bond and others
Maximum Pressure:
Standard: 0 to 3,000 psig (0.1013 to 20 MPa)
Optional: Higher pressures on request
Temperature Range:
Standard: –40° to 212°F (–40° to 100°C)
Optional (overall range): -310
o
to 932oF (-190 to 500oC)
Mounting:
Spoolpiece, hot tap, or cold tap
Area Classifications:
Standard: General-purpose
Optional:
Explosion proof: Class I, Div. 1, Groups B, C & D; Class II, Groups E, F and G; Class III
Flameproof: II 2 G EEx d IIC T6
Weatherproof: Type 4/IP65
Submersible
Note:Wetted transducers for special applications and flowcells are also available. Consult the factory for details.
DigitalFlow™ XMT868i Startup Guide63
Chapter 4. Specifications
4.3.2Clamp-On Transducers
Temperature Range:
Standard: –40° to 300°F (–40° to 150°C)
Optional (overall range): -310
Mounting:
Stainless steel chain or strap, welded or magnetic clamping fixtures
Area Classifications:
Standard: General-purpose
Optional:
Explosion proof: Class I, Div. 1, Groups B, C & D; Class II, Groups E, F and G; Class III
Flameproof: II 2 G EEx md IIC T6-T3
Weatherproof: Type 4/IP65
Submersible
o
to 572oF (-190o to 300oC)
Note:Clamp-on transducers for special applications and flowcells are also available. Consult the factory for details.
4.3.3Transducer Cables
Length:
Standard: Pair of coaxial cables, type RG62 aU, length to 3 ft (1 m)
Optional: Lengths up to 1000 ft (330 m) maximum
4.3.4Temperature Transducers (optional)
Type:
Loop-powered 3-wire platinum RTDs; clamp-on and wetted (thermowell) types are available
Range:
o
-4
F to 500oF (-20oC to 260oC), not all extremes of parameters can be achieved simultaneously
Accuracy:
o
0.27
F (0.15oC), wetted RTDs, matched pairs
64DigitalFlow™ XMT868i Startup Guide
4.4Pipe Size and Materials
4.4.1Wetted Transducers
Materials:
All metals and most plastics (Consult the factory for concrete, glass and cement.)
Pipe Sizes:
Transit-Time Mode: Inside diameter 1 to 200 in.
4.4.2Clamp-on Transducers
Materials:
All metals and most plastics
(Consult the factory for concrete, composite materials and highly corroded or lined pipes.)
Pipe Sizes:
Transit-Time Mode:
Outside Diameter: 0.5 to 300 in. (12.7 mm to 7.6 m) and larger
1” NPTM for 6 in. or greater pipe sizes; 3/8” NPTM for 2 to 5 in. sizes
Interrogation:
Standard: Conventional 45
o
tilted-diameter; one or more traversals
Optional: Mid-radius chord; multi-chords including multiple paths
66DigitalFlow™ XMT868i Startup Guide
Appendix A. CE Mark Compliance
Appendix A.CE Mark Compliance
A.1Introduction
For CE Mark compliance, the Model XMT868i flow transmitter must be wired in accordance with the instructions in
this appendix.
IMPORTANT:CE Mark compliance is required for all units intended for use in EU countries.
A.2Wiring
The Model XMT868i must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Tab le 21 below for the specific requirements.
Table 21: Wiring Requirements
ConnectionCable TypeGround Termination
TransducerArmored RG62 a/UGrounded using a cable gland
Input/Output Armored 22 AWG, shielded with armored material added to
outside of jacket
PowerArmored 14 AWG, 3 conductorGrounded using a cable gland
Grounded using a cable gland
Note:If the Model XMT868i is wired as described in this appendix, the unit will comply with the EMC Directive.
DigitalFlow™ XMT868i Startup Guide67
Appendix A. CE Mark Compliance
[no content intended for this page]
68DigitalFlow™ XMT868i Startup Guide
Appendix B. Data Records
Appendix B.Data Records
B.1Available Option Cards
The XMT868i can accommodate one option card in Slot 1 and one in Slot 2. The available configurations are listed in
Tab le 2 2 below.
Table 22: Option Card Configurations
Card #713-Slot #Configuration
1215-021FF - 4 Frequency Outputs
1215-03TT - 4 Totalizer Outputs
1215-04FT - 2 Frequency Outputs/2 Totalizer Outputs
1215-05FO - 2 Frequency Outputs
1215-06TO - 2 Totalizer Outputs
1215-07AA - 4 Standard Alarms
1215-09FA - 2 Frequency Outputs/2 Standard Alarms
1215-11TA - 2 Totalizer Outputs/2 Standard Alarms
1223-02/1473-02OI - 2 Current Inputs
1223-03/1473-03OR - 2 RTD Inputs
1223-04/1473-04TI - 2 Current Inputs/2 Totalizer Inputs
1223-06/1473-06FI - 2 Current Inputs/2 Frequency Inputs
1223-07/1473-07FR - 2 RTD Inputs/2 Frequency Inputs
1223-08/1473-08AI - 2 Current Inputs/2 Standard Alarms
1223-10/1473-10AR - 2 RTD Inputs/2 Standard Alarms
1225-13CO - 2 Current Outputs
1225-14CF - 2 Current Outputs/2 Frequency Outputs
1225-15CT - 2 Current Outputs/2 Totalizer Outputs
1225-16CA - 2 Current Outputs/2 Standard Alarms
1430-03RR - 4 RTD Inputs
1430-04IR - 2 RTD Inputs/2 Current Inputs
1146-02/1332-022128 KB Memory
1146-03/1332-032 MB Memory
1345-04Modbus Communications Protocol
1477-03MODBUS/TCP/IP
1477-01Ethernet
1475-01Foundation Fieldbus
DigitalFlow™ XMT868i Startup Guide69
Appendix B. Data Records
B.2Option Cards Installed
Whenever an option card is installed or changed in the XMT868i flow transmitter, record the type of card and any
additional setup information in the appropriate row of Tab le 2 3 below.
Table 23: Option Cards Installed
Slot #Type of Option CardAdditional Setup Information
0Analog Outputs (A, B)
1
2
70DigitalFlow™ XMT868i Startup Guide
Appendix B. Data Records
B.3Site Data
After the Model XMT868i flow transmitter has been installed, specific site data must be entered via the Keypad
Program, prior to operation. Record that information in Ta bl e 24 below.
Table 24: Site Data
General Information
Model #Serial #
Software Vers.Setup Date
GLOBL-SYSTEM
Meter MessageVol. Dec. Digits
System UnitsEnglishMetricTotalizer Units
Vol. UnitsTot. Dec. Dig.
GLOBL-I/O-ERROR
Error Handling2-Path ErrorNoYes
GLOBL-COMM
Meter AddressMOD. Parity
Baud RateMOD. Stop Bits
MOD. Baud RateMOD. Address
CHANNEL X-ACTIV
Channel 1Channel 2 (if applicable)
Channel Status
Channel 1Channel 2 (if applicable)
Channel LabelChannel Label
Chan. Message
2
Energy OptionOnOffEnergy OptionOnOff
Vol. UnitsVol. Units
Vol. Dec. DigitsVol. Dec. Digits
Totalizer UnitsTotalizer Units
Tot. Dec. Dig.Tot. Dec. Dig.
Mass FlowMass Flow
Mass Flow TimeMass Flow Time
MDOT Dec. Dig.MDOT Dec. Dig.
Mass TotalsMass Totals
Mass Dec. DigitsMass Dec. Digits
PowerPower
Power Dec.Dig.Power Dec. Dig.
Energy (Total)Energy (Total)
Energy Dec. Dig.Energy Dec. Dig.
Heat. or Cool.Heat. or Cool.
Flow Meas.Flow Meas.
1
not available for 1-Channel meter, 2”Site Message” for 1-Channel meter
If you are using wetted transducers, the XMT868i requires you to enter the path length (P) and the axial dimension (L),
P is the transducer face-to-face distance and L is the axial projection of P in the flow stream.
where
To determine
in Figure 22 below, if possible. If such a measurement is not possible, consult the factory.
L, physically measure the distance between the center of the transducer ports at the inside wall as shown
To determine
•Pipe inside diameter (ID)
•Wall thickness (WT)
•Installed pipe coupling length (CL)
•Transducer face depth (FD)
•Mounting angle (MA)
DigitalFlow™ XMT868i Startup Guide75
Figure 22: 180
P, you need to know the following:
o
Transducer Installations - Top View
Appendix C. Measuring P and L Dimensions
FD
CL
MA = 45°
C.1 Measuring P and L (cont.)
Use Figure 23 below to properly measure the coupling length. Typically, the transducer face is positioned just outside
the inside diameter (ID) of the pipe or slightly retracted inside the coupling.
Figure 23: Determining the Pipe Coupling Length
Use the following formula to determine the P dimension:
[ID + 2(WT)]/(cos MA) + 2(CL) - FD) = P Dimension
For example, given the following:
•Inside diameter (ID) = 48”
•Wall thickness (WT) = 3/8”
•Installed coupling length (CL) = 2.0”
•Transducer face depth (FD) = 1.75”
•Mounting Angle (MA) - 45
The P dimension would be [48 + 2(3/8)]/(0.7071) + 2(2.0 - 1.75) = 69.4”
o
76DigitalFlow™ XMT868i Startup Guide
Warranty
Warranty
Each instrument manufactured by GE Sensing is warranted to be free from defects in material and workmanship.
Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at
the sole discretion of GE Sensing. Fuses and batteries are specifically excluded from any liability. This warranty is
effective from the date of delivery to the original purchaser. If GE Sensing determines that the equipment was
defective, the warranty period is:
•One year from delivery for electronic or mechanical failures
•One year from delivery for sensor shelf life
If GE Sensing determines that the equipment was damaged by misuse, improper installation, the use of unauthorized
replacement parts, or operating conditions outside the guidelines specified by GE Sensing, the repairs are not covered
under this warranty.
The warranties set forth herein are exclusive and are in lieu of all other warranties whether
statutory, express or implied (including warranties or merchantability and fitness for a
particular purpose, and warranties arising from course of dealing or usage or trade).
Return Policy
If a GE Sensing instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE Sensing, giving full details of the problem, and provide the model number and serial number of the
instrument. If the nature of the problem indicates the need for factory service, GE Sensing will issue a
RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to
a service center will be provided.
2. If GE Sensing instructs you to send your instrument to a service center, it must be shipped prepaid to the
authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Sensing will evaluate the instrument to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
•If the damage is covered under the terms of the warranty, the instrument will be repaired at no cost to the owner
and returned.
•If GE Sensing determines that the damage is not covered under the terms of the warranty, or if the warranty has
expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s
approval to proceed, the instrument will be repaired and returned.
WARNING - DO NOT OPEN WHEN ENERGIZED
AVERTISSEMENT - Ne pas ouvrir sous tension
WARNING - DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE IS PRESENT
AVERTISSEMENT - Ne pas ouvrir une atmosphere eplosive gazeuse est presente
Measurement & Control
Certification &
Safety Statements
Certification & Safety Statements for GE Measurement & Control Ultrasonic Flow Transmitters
When installing this apparatus, the following requirements must be met:
• Field wiring shall be rated at least 10°C above 70°C.
• Connecting cables shall be mounted securely and protected from mechanical damage, pulling and twisting.
• Cable entries are ¾” NPT.
• Cable glands of an approved flameproof design are required. These must be installed according to the manufacturer’s
instructions. Where the cable glands are provided by GE, the manufacturer’s instructions, as supplied, to GE, will be
included in the documentation.
• Unused cable entries must be sealed using a certified threaded plug.
• Modifications to the flameproof enclosure are not permitted.
• The apparatus should be de-energized before opening.
• Installation should comply with IEC/EN 60079-14.
• Equipment is of type flameproof “d” design and complies with: EN 60079-0:2009, EN 60079-1:2007,
EN 60529:1991 +A1:2000, IEC 60079-0:2011, IEC 60079-1:2007, IEC 60529:2001.
• The product contains no exposed parts which produce surface temperature infrared, electromagnetic ionizing, or
non-electrical dangers.
• The product must not be subjected to mechanical or thermal stresses in excess of those permitted in the certification
documentation and the instruction manual.
• The product cannot be repaired by the user; it must be replaced by an equivalent certified product. Repairs should only
be carried out by the manufacturer or by an approved repairer.
• Only trained, competent personnel may install, operate and maintain the equipment
• The product is an electrical apparatus and must be installed in the hazardous area in accordance with the
requirements of the EC Type Examination Certificate. The installation must be carried out in accordance with all the
appropriate international, national and local standard codes and practices and site regulations for flameproof
apparatus and in accordance with the instructions contained in the manual. Access to the circuitry must not be made
during operation.
Special Conditions for Safe Use:Consult the manufacturer if dimensional information on the flameproof joints is
necessary.
Markings: Markings shall appear on the product as shown below:
1100 Technology Park Drive, Billerica, MA 01821, U.S.A.
Sensing House, Shannon Free Zone East, Shannon, County Clare, Ireland
Telephone: 978-437-1000 or 800-833-9438
CSS-0001, Rev. A
June 2012
Telephone: +353 61 470200
[no content intended for this page]
GE
Sensing
We,GE Sensing
1100 Technology Park Drive
Billerica, MA 01821
USA
declare under our sole responsibility that the
DigitalFlow™ XGF868i Ultrasonic Flare Gas Flow Transmitter
DigitalFlow™ XGM868i Ultrasonic Gas Flow Transmitter
DigitalFlow™ XGN868i Ultrasonic Natural Gas Flow Transmitter
DigitalFlow™ IGM878i Ultrasonic Industrial Gas Flow Transmitter
EU DECLARATION
OF
CONFORMITY
DOC-0003, Rev. D
to which this declaration relates, are in conformity with the following standards:
• EN 60079-0: 2012/A11:2013
• EN 60079-1: 2014
• II 2 G Ex d IIC T5; FM12ATEX0059X (FM-NoBo xxxFM Approvals Ltd, Windsor, Berkshire, UK – NoBo 1725)
(Baseefa – NoBo xxxBaseefa, Buxton, Derbyshire, UK-NoBo 1180)
• EN 61326-1: 2013
• EN 61010-1: 2012, Overvoltage Category II
following the provisions of the 2014/36/EU EMC and 2014/34/EU ATEX directives.
Where products were initially assessed for compliance with the Essential Health and Safety Requirements of the
ATEX Directive 94/9/EC using earlier harmonized standards, a subsequent review has determined that “technical
knowledge” is unaffected by the current harmonized standards listed above.
16 June 2016
IssuedMr. Chris Frail
Engineering Manager
GE Measurement Solutions
Billerica, MA U.S.A.
[no content intended for this page]
Customer Support Centers
U.S.A.
The Boston Center
1100 Technology Park Drive
Billerica, MA 01821
U.S.A.
Tel: 800 833 9438 (toll-free)
978 437 1000
E-mail: sensing@ge.com
Ireland
Sensing House
Shannon Free Zone East
Shannon, County Clare
Ireland
Tel: +353 (0)61 470200
E-mail: gesensingsnnservices@ge.com