GE Sensing XDP Operating Manual

GE
Sensing
XDP-STD
Panametrics Explosion-Proof Display Package
User’s Manual
910-204D1 January 2008
The XDP Panametrics Explosion-Proof Display Package is a GE Panametrics product. GE Panametrics has joined other GE high-technology sensing businesses under a new name—GE Sensing.
This manual is for use with XDP units
with standard software (E=1) only.
Attention!
January 2008
Warranty Each instrument manufactured by GE Sensing, Inc. is warranted to be
free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE determines that the equipment was defective, the warranty period is:
one year for general electronic failures of the instrument
one year for mechanical failures of the sensor
If GE determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of all other warranties whether statutory, express or implied (including warranties of merchantability and fitness for a particular purpose, and warranties arising from course of dealing or usage or trade).
Return Policy If a GE Sensing, Inc. instrument malfunctions within the warranty
period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the model
number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE will issue a RETURN AUTHORIZATION number (RA), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE instructs you to send your instrument to a service center, it
must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine the
cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
If GE determines that the damage is not covered under the terms
of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iii
January 2008
Table of Contents
Chapter 1: General Information
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Basic Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Chapter 2: Installation
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Choosing an Installation Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting the XDP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Mounting the XDP Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Mounting the Remote Transmitter/Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Wiring the XDP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Wiring the Analog Input and 24 VDC Output - J6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Wiring The System Fault Alarm - J2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Wiring The XDP Alarms - J3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Wiring the Calibration Relays - J7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Wiring the System Analog Output - J5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Wiring the XDP Analog Output - J8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Wiring the Serial Port - J4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Wiring The Line Power - TB1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Chapter 3: Operation
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preventing Common Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Cleaning the Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Powering Up the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
The Enclosure Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Initial Screen Displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Starting the Sample Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Accessing the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Exiting the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
v
January 2008
Table of Contents (cont.)
Chapter 4: Setup
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Programming Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Setting Up the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Switching Display Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Configuring the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Adjusting the Display Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Setting the Display Backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Setting Up the Analog Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Entering Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Setting the Analog Input Error Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Setting Up the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Setting Up the System Fault Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Setting Up the Four XDP Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Setting Up the Analog and Digital Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Setting Up the System Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Setting Up the XDP Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Setting Up the Serial Port Digital Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Using the XDP Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Lock Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Lock All . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
vi
January 2008
Table of Contents (cont.)
Chapter 5: Calibration
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Programming Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Configuring AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Enabling AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Selecting the AutoCal Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Setting Calibration Gas Flow Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Scheduling AutoCal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Indicating an AutoCal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Initiating an AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Setting the AutoCal Drift Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Setting AutoCal Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Viewing the Drift Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Clearing the Drift Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Resetting the Verification Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Abort AutoCAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Chapter 6: Factory-Level Programming
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Programming Levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Programming Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Setup-Level Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Configuring the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Factory-Level Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
The Adv Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
The Opt>Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
vii
January 2008
Table of Contents (cont.)
Chapter 7: Specifications
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Transmitter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
The XDP Model Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Appendix A: CE Mark Compliance
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
EMC Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
LVD Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Appendix B: PanaView Programming
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Remote Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Viewing Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Logging Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Appendix C: Menu Maps
Appendix D: Factory Default Values
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Cal Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
4-20 mA Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Slot Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
AutoCAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
viii

Chapter 1

General Information

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Basic Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
January 2008

Introduction The XDP Explosion-proof Display Package is a general-purpose

display/control unit used for oxygen and thermal conductivity measurement applications in Zone 1 environments. A second part of the system is a remotely-mounted sensor/transmitter, such as the GE Sensing XMO2, XMTC or O2X1.
Note: Refer to the documentation supplied with the specific
transmitter for complete information on the transmitter.
The XDP performs the following basic functions:
supplies 24 VDC to power the transmitter
accepts and processes a 4-20 mA output from the transmitter
shows a selected process parameter on a digital display
provides a user interface for programming the system
provides user-configurable alarm relays and an analog output
The XDP is wall-mounted via its integral mounting plate and it is connected to the transmitter with an electrical conduit. It is certified for use in Zone 1 hazardous areas.

Basic Features The XDP incorporates the following basic features:

wall-mountable, explosion-proof enclosure
universal 100-240 VAC power input
24 VDC power output
4-20 mA analog input
4-20 mA/0-20 mA/0-2 V analog output
RS-232/RS-485 serial port
high-resolution, backlit LCD digital display
six-key, infrared (“through-the-glass”) keypad
five general-purpose alarm relays
automatic calibration
See Figure 1-1 on page 1-2 for a picture of the XDP.
General Information 1-1
January 2008
Basic Features (cont.)
Figure 1-1: The XDP

System Overview In a binary gas mixture, the transmitter measures the concentration of

a target gas in a known background gas (see the transmitter documentation for a discussion of the specific technology employed by the sensor mechanism). To be suitable for use in an XDP system, the transmitter must:
be self-powered or powered by the +24 VDC output fro m the XDP
output a 4-20 mA signal to the XDP analog input
comply with all performance specifications for the system
be suitable for use in the specified environment
The XDP accepts the 4-20 mA analog output from the transmitter and processes the information. Then, the results are displayed on the built-in digital display window . In addition, the data may be output as a linear analog signal to drive a data recorder or similar device.
1-2 General Information

Chapter 2

Installation

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Choosing an Installation Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting the XDP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Wiring the XDP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
January 2008

Introduction This chapter provides a general description of the XDP Explosion-

proof Display Package and gives directions on how to install and wire the unit for proper operation.
IMPORTANT: The XDP requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
Be sure to observe all installation limits and precautions described in this chapter. Pay particular attention to the ambient temperature range of –10° to +60°C (14° to +140°F) specified for the instrument.
!WARNING!
To ensure safe operation of the XDP, the unit must be
installed and operated as described in this manual. Also, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment. In addition,
all procedures should be performed by trained service
personnel.
Proceed to the next page to begin the installation.
Installation 2-1
January 2008

Choosing an Installation Site

All environmental and installation factors should have been discussed with a GE Sensing applications engineer or field sales person at the time the XDP was ordered. Thus, the equipment should be suited to the application and the planned installation site. However, before installing the unit, read these guidelines to verify that the best installation site has been chosen for optimum instrument accuracy and reliability.
Make sure that the XDP and transmitter enclosures are suitable for
the environmental conditions at the installation site.
Install the system in a location with little or no vibration.
Mount the transmitter in accordance with the instructions provided
in its accompanying documentation.
Make sure the ambient temperature at the XDP installation site is
–10° to 60°C (14° to 140°F).
Use a suitable electrical conduit to interconnect the remote
transmitter and the XDP enclosure.
Use a suitable sample system between the gas source and the
sample inlet of the transmitter.
Observe all normal safety precautions. Specifically, do not exceed
the maximum pressure and temperature ratings of the transmitter or the sample system.
Install the XDP system in a location that provides ready access for
programming, testing, and servicing the unit.
Protect all cables from excessive physical strain (bending, pulling,
twisting, etc.). Do not subject the cables to temperatures above 65°C (149°F) or below –50°C (–58°F).
Be sure that the line voltage at the planned installation site is
within the specified power input rating for the XDP (100-240 VAC; 50/60 Hz; 40 W max.).
2-2 Installation
January 2008

Mounting the XDP System

Mounting the XDP Enclosure
This section explains how to mount the XDP system components and the transmitter/sample system at the installation site.
IMPORTANT: The XDP requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
IMPORTANT: This symbol indicates Caution - risk of electric
shock:
Refer to Figure 2-2 on page 2-11, and complete the following steps to mount the XDP enclosure:
1. Select a flat surface on a vertical wall or instrument panel with sufficient space to hold the XDP enclosure.
2. Using the mounting dimensions shown in Figure 2-2 on page 2-11, locate and prepare four (4) mounting ho les or two (2) horizontal mounting rails on the vertical surface.
Mounting the Remote Transmitter/Sample System
3. Secure the XDP enclosure to the mounting surface with four (4) sets of mounting hardware. Be sure that the enclosure is in a vertical position, with the display window upright.
Mount the remote transmitter in accordance with the instructions in its accompanying documentation.
Note: Most GE Sensing transmitters may be located up to
2,800 ft (850 m) from the XDP enclosure.
To provide reliable and accurate results, the transmitter is often installed in a suitable sample system. The sample system must deliver a clean, representative gas sample to the inlet of the transmitter at the proper temperature, pressure and flow rate.
Note: A suitable sample system may be ordered directly from GE
Sensing. The sample system would be supplied fully assembled on a flat plate with four (4) mounting holes. See the sample system and/or transmitter manuals for more details.
Installation 2-3
January 2008

Wiring the XDP To completely wire the XDP, connect the following items:

analog input (terminal J6)
+24 VDC output (terminal J6)
system fault alarm (terminal J2)
XDP alarms (terminal J3)
additional relays (terminal J7)
system analog output (terminal J5)
XDP analog output (terminal J8)
serial port (terminal J4)
line power (terminal TB1)
Refer to the wiring diagram shown in Figure 2-3 on page 2-12 and the interconnect label (which is located inside of both covers) shown in Figure 2-1 below, while completing the instructions in this section.
TB 1
L
N
AC
-
+
RELAY 3
RELAY 4
12 11 10 9 8 7 6 5 4 3 2 1 3 2 1
J7
RELAY 1
RELAY 2
NO C NC NO C NC NO C NC NO C NC
24VDC
ANALOG IN
RTN +24 IN
J6
BOTH PORTS SHOWN ON LABEL
ANALOG OUT 1
ARTN AOUT
3 2 1
2 1
- +
NO C NC NO C NC NO C NC NO C NC NO C NC
J5 J2 J4
ALARM 4
SYSTEM
FAULT
12 11 10 9 8 7 6 5 4 3 2 1 3 2 1
ALARM 3
J3
FUSE F1
ALARM 2
ALARM 1
XDP
4-20mA OUT
RTN OUT
2 1
RS232/485
COMM
RTN RX TX
- +
J8
Figure 2-1: Interconnect Label (Drawing #420-790E)
2-4 Installation
Wiring the XDP (cont.) !Attention European Customers!
To meet CE Mark requirements, install all cables as
described in Appendix A, CE Mark Compliance.
IMPORTANT: The XDP requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
!WARNING!
To ensure safe operation of the XDP, the unit must be
installed and operated as described in this manual. Also, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment. In addition,
all procedures should be performed by trained service
personnel.
January 2008
Wiring the Analog Input and 24 VDC Output - J6
Remove the two side covers on the XDP enclosure. This is accomplished by loosening the set screw in the cover and using a long screwdriver or rod in the slots provided to unscrew the cover. Proceed to the appropriate sub-section to wire each of the terminal blocks.
Note: A long, pin-like, plastic wiring tool is stored inside both ports
of the XDP enclosure. Insert the tool into a connector pin to open the hole, and then insert the wire. When the tool is removed, the wire is clamped into the connector.
Terminal block J6 contains the analog input and power connections for the remote transmitter. To wire the transmitter, complete the following steps:
Note: Refer to the transmitter manual for details on wiring the
transmitter’s terminal block(s).
1. Connect pin #1 to the positive terminal of the analog output from the transmitter.
2. Connect pin #2 to the +24 VDC loop power input on the transmitter.
3. Connect pin #3 to the return connection of the analog output from the transmitter.
Installation 2-5
January 2008
Wiring The System Fault Alarm - J2
Wiring The XDP Alarms ­J3
Terminal block J2 contains the system fault alarm connections. To wire this alarm, complete the following steps:
Note: The system fault alarm on terminal block J2 is usually factory-
configured as a “fail-safe” alarm. That is, it is energized during normal operation and de-energized when tripped.
1. Connect pin #1 (NC) to the alarm device input.
2. Connect pin #2 (C) to the alarm device return.
3. Pin #3 (NO) is not used for fail-safe operation.
Note: For non-fail-safe operation, repr ogram the alarm or move the
pin #1 connection to pin #3.
Terminal block J3 contains connections for the four XDP alarms. To wire these alarms, complete the following steps:
Note: The following instructions show the wiring for fail-safe
operation. For non-fail-safe operation reprogram the alarm or move the pin #1 connection to pin #3 for that alarm.
1. Connect the XDP High alarm (alarm 1) as follows: a. Connect pin #1 (NC) to the alarm device input. b. Connect pin #2 (C) to the alarm device return. c. Pin #3 (NO) is not used for fail-safe operation
2. Connect the XDP High High alarm (alarm 2) as follows: a. Connect pin #4 (NC) to the alarm device input. b. Connect pin #5 (C) to the alarm device return. c. Pin #6 (NO) is not used for fail-safe operation
3. Connect the XDP Low alarm (alarm 3) as follows: a. Connect pin #7 (NC) to the alarm device input. b. Connect pin #8 (C) to the alarm device return. c. Pin #9 (NO) is not used for fail-safe operation.
4. Connect the XDP Low Low alarm (alarm 4) as follows: a. Connect pin #10 (NC) to the alarm device input. b. Connect pin #11 (C) to the alarm device return. c. Pin #12 (NO) is not used for fail-safe operation.
2-6 Installation
January 2008
Wiring the Calibration Relays - J7
T erminal block J7 contains connections for the four calibration relays. To wire these relays, complete the following steps:
Note: These relays are configured for non-fail-safe operation.
1. Connect the Pr ocess/Cal relay (Relay 1) as follows: a. Pin #1 (NC) is not used for non-fail-safe operation. b. Connect pin #2 (C) to the alarm device return. c. Connect pin #3 (NO) to the alarm device input.
2. Connect the Zero/Span Cal relay (Relay 2) as follows: a. Pin #4 (NC) is not used for non-fail-safe operation. b. Connect pin #5 (C) to the alarm device return. c. Connect pin #6 (NO) to the alarm device input.
3. Connect the AutoCal Indication alarm (Relay 3) as follows: a. Pin #7 (NC) is not used for non-fail-safe operation. b. Connect pin #8 (C) to the alarm device return.
Wiring the System Analog Output - J5
c. Connect pin #9 (NO) to the alarm device input.
4. Connect the Auto Verification alarm (Relay 4) as follows: a. Pin #10 (NC) is not used for non-fail-safe operation. b. Connect pin #11 (C) to the alarm device return. c. Connect pin #12 (NO) to the alarm device input.
T e rminal block J5 contains connections for the system an alog out put. To wire the system analog output, complete the following steps:
1. Connect pin #1 to positive input on the device.
2. Connect pin #2 to the return line from the device.
3. Pin #3 has no connection.
Installation 2-7
January 2008
Wiring the XDP Analog Output - J8
Terminal block J8 contains connections for the XDP analog output. To wire the XDP analog output, complete the following steps:
1. Connect pin #1 to positive input on the device.
2. Connect pin #2 to the return line from device.
3. Pin #3 has no connection.
Wiring the Serial Port - J4 Terminal block J4 contains connections for the RS-232/RS-485
(RS-485 supports multi-drop installations) serial port. To wire the serial port, complete the following steps:
1. Connect pin #1 to the transmit pin on the PC serial port.
2. Connect pin #2 to the receive pin on the PC serial port.
3. Connect pin #3 to the return pin on the PC serial port.
Note: This connection may be made with a GE Sensing
#704-668 cable. If this cable is used, connect the red wire to pin #1, the white wire to pin #2, and the green wire to pin #3.
See EIA-RS Serial Communications (document 916-054) for a detailed discussion of serial port wiring.
Wiring The Line Power ­TB1
Terminal block TB1 on the power supply board contains connections for the line power to the XDP. Make these connections only with wire that meets the following specifications:
individual conductor gauge of 18 AWG minimum (maximum
current capacity of 12 Amps)
voltage rating of 600 Vo lts minimum
insulation temperature rating of 105°C minimum
!WARNING!
The rating of the wire insulation must be at least 15°C
above the expected ambient temperature.
2-8 Installation
January 2008
Wiring The Line Power - J1 (cont.)
Refer to Figure 2-3 on page 2-12 and complete the following steps:
!WARNING!
Before proceeding with the section, verify that the line
power has been turned off at the external disconnect
device.
1. Attach a cable or conduit with the three line power conductors to a
cable entry port on the XDP enclosure.
Caution!
Be sure that the input voltage complies with the input
power rating for the XDP (see Chapter 7, Specifications).
2. Connect the line power leads to terminal block J1 as follows: a. Connect the line power lead (black or brown) to pin
L on the
power connector.
b. Connect the neutral power lead (white or blue) to pin
N on the
power connector.
c. Connect the ground power lead (green or green/yellow) to the
chassis ground screw shown in Figure 2-2 on page 2-10.
3. Make sure that the factory-installed jumper wire at pin
G on the
power connector is secure.
This completes the wiring of the XDP. Proceed to Chapter 3, Operation, for instructions on using the meter.
Installation 2-9
January 2008
8.28
9.05
35°
3.77
A
XDP
6.44
6.13
4.88
5.857.10
Ø.53 THRU
MTG HOLES
Ø5.60
4X
(7.10)
.50
A
Figure 2-2: XDP Mounting Dimensions (Drawing #712-1095B)
Installation 2-11
January 2008
SYSTEM ANALOG OUTPUT - J5
Pin No.
Description
Analog Output (AOUT)
1 2
Return (ARTN) No Connection
3
SYSTEM FAULT ALARM - J2
Description
Pin No.
Normally Closed (NC)
1 2
Common (C) Normally Open (NO)
3
Wiring Tool
This unit requires an external power
NOTE:
disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible and located within 1.8 m (6 ft) of the XDP.
Factory Jumper
POWER INPUT - TB1
Pin No.
L/+
N/-
Gnd
100-240 VAC
Line Neutral Ground
+24 VDC
Positive Negative Ground
LEFT SIDE CONNECTIONS RIGHT SIDE CONNECTIONS
XDP ALARMS - J3
Description
Pin No.
1 2 3 4 5 6 7 8
9 10 11 12 Return (RTN)
Alarm
High
NC
C
Alarm
(Alarm 1)
NO
High High
NC
Alarm
C
(Alarm 2)
NO NC
C NO NC
C NO
Low
Alarm
(Alarm 3)
Low Low
Alarm
(Alarm 4)
[* wire colors for Panametrics
#704-668 serial cable]
RS232/485 COMM - J4
Description
Pin No.
Transmit (TX)
1 2
Receive (RX)
3
Color*
White Green
Red
CALIBRATION RELAYS - J7
Description
Pin No.
1
NC
2
C
3
NO
4
NC
5
C
6
NO
7
NC
8
C
9
NO
10
NC
11
C
NO
12
Alarm
Process/
Cal
(Relay 1)
Zero/Span
Cal (Relay 2) AutoCal
Indication
(Relay 3)
Auto
Verification
(Relay 4)
ANALOG INPUT/LOOP POWER - J6 Pin No.
Analog Input (IN)
1 2
Power Output (+24)
3
Return (RTN)
Rating
Voltage
Size Type
Ground Terminal
XDP ANALOG OUT PUT - J8
Pin No.
DescriptionDescription
1
Analog Output (OUT)
2
Return (RTN)
3
No Connection
FUSE
1.25 Amp
250 V 5 x 20 mm T
Figure 2-3: XDP Wiring Connections (from Drawing #442-790E)
Installation 2-12

Chapter 3

Operation

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preventing Common Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Cleaning the Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Powering Up the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Accessing the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Exiting the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
January 2008

Introduction The XDP is an easily operated monitoring/control device. Make sure

that the system has been installed in accordance with the instructions given in Chapter 2, Installation, before applying power. This chapter includes discussions of the following topics:
preventing common problems
powering up the system
programming the meter
taking measurements
!WARNING!
To ensure safe operation of the XDP, it must be installed
and operated as described in this manual. In addition, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment.

Preventing Common Problems

The complexity of accurately measuring the target gas mixtures dictates that some basic precautions be observed in using the XDP. Failure to observe these simple procedures is often the cause of some common problems. Compliance with the following points will help to eliminate such problems:
The transmitter must be properly connected to the XDP before
applying power to the system. Connecting or disconnecting the transmitter while the XDP is powered up may generate a 24 Volt Power Fault message. If this should occur, cycle the power to the system OFF and then ON to clear the problem.
The automatic calibration feature should be enabled and scheduled
to be run on a regular basis. The optimum calibration interval depends on the application (consult the factory).
Never use thread sealant on any connection in the sample gas flow
path. Thread sealant emits combustible vapors that can cause reading errors.
The infrared keypad may not function properly if direct sunlight
strikes the face of the XDP or if the XDP is installed in an area that has other infrared devices in operation. To minimize such factors, provide suitable shading at the installation site.

Cleaning the Enclosure If the glass window or the case of the XDP becomes soiled, use a soft

cloth dampened with water for cleaning. Never use solvents or detergents to clean the XDP.
Operation 3-1
January 2008

Powering Up the System Check the wiring connections and close both the XDP and transmitter

enclosures before applying power. Energize the external disconnect device to power up the XDP. Allow the transmitter to warm up for at least thirty minutes, before taking measurements.
IMPORTANT: The XDP requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
The Enclosure Window The window at the top of the XDP enclosure includes the components
shown in Figure 3-1 below.
Display Window
ESC Key
LEFT Arrow
UP Arrow
ENTER Key
RIGHT Arrow
DOWN Arrow
Figure 3-1: The Enclosure Window
The XDP is factory-configured to display O2 ppm/pcnt with a resolution of 0.01 as the default. However, changing the gas label, the
gas units, and the display resolution are easily accomplished. See Chapter 4, Setup, for programming instructions.
IMPORTANT: The XDP’s infrared keypad permits programming of
the instrument through the glass faceplate without removing the cover. Thus, all programming procedures may be performed while the unit is installed in a hazardous area.
3-2 Operation
January 2008
Initial Screen Displays Immediately upon power up, the digital display shows the following
sequence of informational screens, as the meter performs its internal systems checks.
BSeries Loader v1.05 Boot is Flash. Program CRC valid. Booting from Flash
Scanning Hardware Found Image Slot 1
Downloading Slot 1
Scanning Hardware Initializing Machine Initializing UI
If the meter passes all of its internal system checks, the following initial data screen appears:
This is a typical example of the first initialization screen.
This is a typical example of the second initialization screen.
This is a typical example of the final initialization screen.
1 O2 HH:MM
0.00
Note: The padlock icon at the upper right hand corner of the above
display indicates that the User Program is locked with password protection.
±
pcnt
LL
Notice that the absence of a valid signal has triggered the Low Low alarm (the LL sym­bol in the lower right corner).
Operation 3-3
January 2008
Starting the Sample Gas Start the sample gas flow to the transmitter and allow a few minutes
for the system to reach equilibrium. The following data screen should appear if the oxygen level is above the Low alarm setpoint and below the High alarm setpoint.

Accessing the User Program

1 O2 HH:MM±
pcnt
X.XX
Note: Allow approximately 30 minutes for the sensor to reach
temperature equilibrium, to ensure accurate readings
Notice that the display shows the slot number, the ga s label, and the current time on the top line. The current data value along with the units are displayed across the bottom of the screen.
Note: Slot 0 is the main circuit board, slot 1 holds the XDP circuit
board, and slots 2-4 are normally not used.
To configure the XDP for a specific application, it is necessary to leave measurement mode and enter the User Program as follows (see the menu maps in Figures C-1 and C-2 on pages C-1 and C-2):
1 O2 HH:MM±
pcnt
X.XX
This is a typical display. The actual oxygen concentration of your sample gas will be shown here.
While in measurement mode, press the
[ESC], [ENTER], and [ESC] keys
in sequence.
[ENTER], [ENTER],
Enter the operator-level password (2719) at the next screen to access the menus available to the end user. (A different password is required to access the factory-level menus.)
Password
Enter Password
XXXX
s
Cal Opt Disp
3-4 Operation
Use [W] and [X] to position the cursor under the desired char­acter. Then, use [ increment the value. Press
[ENTER] when done.
This screen shows the highest level of the User Program, which is referred to as the Main Menu.
S] and [T] to
January 2008

Exiting the User Program T o leave the User Program, proceed as follows from the main menu:

Cal Opt Disp
1 O2 HH:MM
pcnt
Press the [ESC] key.
You are now back in normal measurement mode.
X.XX
Notice that the padlock is no longer visible at the upper right corner of the display window. Once the User Program has been accessed with a valid password, the menus at that programming level remain unlocked when you return to measurement mode. With the menus unlocked, you may reenter the User Program by simply pressing the
[ESC] key only. To re-lock the menus, see the instructions in Chapter
4, Setup.
Note: If the unit is powered down, the User Program will be locked
when the unit is powered up again.
To program a specific feature of the XDP, proceed to the appropriate section of Chapter 4, Setup.
Operation 3-5
Loading...
+ 68 hidden pages