The XDP Panametrics Explosion-Proof Display Package is a GE Panametrics product. GE Panametrics has
joined other GE high-technology sensing businesses under a new name—GE Sensing.
This manual is for use with XDP units
with standard software (E=1) only.
Attention!
January 2008
WarrantyEach instrument manufactured by GE Sensing, Inc. is warranted to be
free from defects in material and workmanship. Liability under this
warranty is limited to restoring the instrument to normal operation or
replacing the instrument, at the sole discretion of GE. Fuses and
batteries are specifically excluded from any liability. This warranty is
effective from the date of delivery to the original purchaser. If GE
determines that the equipment was defective, the warranty period is:
• one year for general electronic failures of the instrument
• one year for mechanical failures of the sensor
If GE determines that the equipment was damaged by misuse,
improper installation, the use of unauthorized replacement parts, or
operating conditions outside the guidelines specified by GE, the
repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of
all other warranties whether statutory, express or implied
(including warranties of merchantability and fitness for a
particular purpose, and warranties arising from course of
dealing or usage or trade).
Return PolicyIf a GE Sensing, Inc. instrument malfunctions within the warranty
period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the model
number and serial number of the instrument. If the nature of the
problem indicates the need for factory service, GE will issue a
RETURN AUTHORIZATION number (RA), and shipping instructions
for the return of the instrument to a service center will be
provided.
2. If GE instructs you to send your instrument to a service center, it
must be shipped prepaid to the authorized repair station indicated
in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine the
cause of the malfunction.
Then, one of the following courses of action will then be taken:
• If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
• If GE determines that the damage is not covered under the terms
of the warranty, or if the warranty has expired, an estimate for the
cost of the repairs at standard rates will be provided. Upon receipt
of the owner’s approval to proceed, the instrument will be repaired
and returned.
IntroductionThe XDP Explosion-proof Display Package is a general-purpose
display/control unit used for oxygen and thermal conductivity
measurement applications in Zone 1 environments. A second part of
the system is a remotely-mounted sensor/transmitter, such as the GE
Sensing XMO2, XMTC or O2X1.
Note: Refer to the documentation supplied with the specific
transmitter for complete information on the transmitter.
The XDP performs the following basic functions:
• supplies 24 VDC to power the transmitter
• accepts and processes a 4-20 mA output from the transmitter
• shows a selected process parameter on a digital display
• provides a user interface for programming the system
• provides user-configurable alarm relays and an analog output
The XDP is wall-mounted via its integral mounting plate and it is
connected to the transmitter with an electrical conduit. It is certified
for use in Zone 1 hazardous areas.
Basic FeaturesThe XDP incorporates the following basic features:
• wall-mountable, explosion-proof enclosure
• universal 100-240 VAC power input
• 24 VDC power output
• 4-20 mA analog input
• 4-20 mA/0-20 mA/0-2 V analog output
• RS-232/RS-485 serial port
• high-resolution, backlit LCD digital display
• six-key, infrared (“through-the-glass”) keypad
• five general-purpose alarm relays
• automatic calibration
See Figure 1-1 on page 1-2 for a picture of the XDP.
General Information1-1
January 2008
Basic Features (cont.)
Figure 1-1: The XDP
System OverviewIn a binary gas mixture, the transmitter measures the concentration of
a target gas in a known background gas (see the transmitter
documentation for a discussion of the specific technology employed
by the sensor mechanism). To be suitable for use in an XDP system,
the transmitter must:
• be self-powered or powered by the +24 VDC output fro m the XDP
• output a 4-20 mA signal to the XDP analog input
• comply with all performance specifications for the system
• be suitable for use in the specified environment
The XDP accepts the 4-20 mA analog output from the transmitter and
processes the information. Then, the results are displayed on the
built-in digital display window . In addition, the data may be output as
a linear analog signal to drive a data recorder or similar device.
IntroductionThis chapter provides a general description of the XDP Explosion-
proof Display Package and gives directions on how to install and wire
the unit for proper operation.
IMPORTANT:The XDP requires an external power disconnect
device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m
(6 ft) of the unit.
Be sure to observe all installation limits and precautions described in
this chapter. Pay particular attention to the ambient temperature range
of –10° to +60°C (14° to +140°F) specified for the instrument.
!WARNING!
To ensure safe operation of the XDP, the unit must be
installed and operated as described in this manual. Also, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment. In addition,
all procedures should be performed by trained service
personnel.
Proceed to the next page to begin the installation.
Installation2-1
January 2008
Choosing an Installation
Site
All environmental and installation factors should have been discussed
with a GE Sensing applications engineer or field sales person at the
time the XDP was ordered. Thus, the equipment should be suited to
the application and the planned installation site. However, before
installing the unit, read these guidelines to verify that the best
installation site has been chosen for optimum instrument accuracy
and reliability.
• Make sure that the XDP and transmitter enclosures are suitable for
the environmental conditions at the installation site.
• Install the system in a location with little or no vibration.
• Mount the transmitter in accordance with the instructions provided
in its accompanying documentation.
• Make sure the ambient temperature at the XDP installation site is
–10° to 60°C (14° to 140°F).
• Use a suitable electrical conduit to interconnect the remote
transmitter and the XDP enclosure.
• Use a suitable sample system between the gas source and the
sample inlet of the transmitter.
• Observe all normal safety precautions. Specifically, do not exceed
the maximum pressure and temperature ratings of the transmitter
or the sample system.
• Install the XDP system in a location that provides ready access for
programming, testing, and servicing the unit.
• Protect all cables from excessive physical strain (bending, pulling,
twisting, etc.). Do not subject the cables to temperatures above
65°C (149°F) or below –50°C (–58°F).
• Be sure that the line voltage at the planned installation site is
within the specified power input rating for the XDP
(100-240 VAC; 50/60 Hz; 40 W max.).
2-2Installation
January 2008
Mounting the XDP
System
Mounting the XDP
Enclosure
This section explains how to mount the XDP system components and
the transmitter/sample system at the installation site.
IMPORTANT:The XDP requires an external power disconnect
device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m
(6 ft) of the unit.
IMPORTANT:This symbol indicates Caution - risk of electric
shock:
Refer to Figure 2-2 on page 2-11, and complete the following steps to
mount the XDP enclosure:
1. Select a flat surface on a vertical wall or instrument panel with
sufficient space to hold the XDP enclosure.
2. Using the mounting dimensions shown in Figure 2-2 on page
2-11, locate and prepare four (4) mounting ho les or two (2)
horizontal mounting rails on the vertical surface.
Mounting the Remote
Transmitter/Sample
System
3. Secure the XDP enclosure to the mounting surface with four (4)
sets of mounting hardware. Be sure that the enclosure is in a
vertical position, with the display window upright.
Mount the remote transmitter in accordance with the instructions in
its accompanying documentation.
Note: Most GE Sensing transmitters may be located up to
2,800 ft (850 m) from the XDP enclosure.
To provide reliable and accurate results, the transmitter is often
installed in a suitable sample system. The sample system must deliver
a clean, representative gas sample to the inlet of the transmitter at the
proper temperature, pressure and flow rate.
Note: A suitable sample system may be ordered directly from GE
Sensing. The sample system would be supplied fully assembled
on a flat plate with four (4) mounting holes. See the sample
system and/or transmitter manuals for more details.
Installation2-3
January 2008
Wiring the XDPTo completely wire the XDP, connect the following items:
• analog input (terminal J6)
• +24 VDC output (terminal J6)
• system fault alarm (terminal J2)
• XDP alarms (terminal J3)
• additional relays (terminal J7)
• system analog output (terminal J5)
• XDP analog output (terminal J8)
• serial port (terminal J4)
• line power (terminal TB1)
Refer to the wiring diagram shown in Figure 2-3 on page 2-12 and the
interconnect label (which is located inside of both covers) shown in
Figure 2-1 below, while completing the instructions in this section.
Wiring the XDP (cont.)!Attention European Customers!
To meet CE Mark requirements, install all cables as
described in Appendix A, CE Mark Compliance.
IMPORTANT:The XDP requires an external power disconnect
device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m
(6 ft) of the unit.
!WARNING!
To ensure safe operation of the XDP, the unit must be
installed and operated as described in this manual. Also, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment. In addition,
all procedures should be performed by trained service
personnel.
January 2008
Wiring the Analog Input
and 24 VDC Output - J6
Remove the two side covers on the XDP enclosure. This is
accomplished by loosening the set screw in the cover and using a long
screwdriver or rod in the slots provided to unscrew the cover. Proceed
to the appropriate sub-section to wire each of the terminal blocks.
Note: A long, pin-like, plastic wiring tool is stored inside both ports
of the XDP enclosure. Insert the tool into a connector pin to
open the hole, and then insert the wire. When the tool is
removed, the wire is clamped into the connector.
Terminal block J6 contains the analog input and power connections
for the remote transmitter. To wire the transmitter, complete the
following steps:
Note: Refer to the transmitter manual for details on wiring the
transmitter’s terminal block(s).
1. Connect pin #1 to the positive terminal of the analog output from
the transmitter.
2. Connect pin #2 to the +24 VDC loop power input on the
transmitter.
3. Connect pin #3 to the return connection of the analog output from
the transmitter.
Installation2-5
January 2008
Wiring The System Fault
Alarm - J2
Wiring The XDP Alarms J3
Terminal block J2 contains the system fault alarm connections. To
wire this alarm, complete the following steps:
Note: The system fault alarm on terminal block J2 is usually factory-
configured as a “fail-safe” alarm. That is, it is energized
during normal operation and de-energized when tripped.
1. Connect pin #1 (NC) to the alarm device input.
2. Connect pin #2 (C) to the alarm device return.
3. Pin #3 (NO) is not used for fail-safe operation.
Note: For non-fail-safe operation, repr ogram the alarm or move the
pin #1 connection to pin #3.
Terminal block J3 contains connections for the four XDP alarms. To
wire these alarms, complete the following steps:
Note: The following instructions show the wiring for fail-safe
operation. For non-fail-safe operation reprogram the alarm
or move the pin #1 connection to pin #3 for that alarm.
1. Connect the XDP High alarm (alarm 1) as follows:
a. Connect pin #1 (NC) to the alarm device input.
b. Connect pin #2 (C) to the alarm device return.
c. Pin #3 (NO) is not used for fail-safe operation
2. Connect the XDP High High alarm (alarm 2) as follows:
a. Connect pin #4 (NC) to the alarm device input.
b. Connect pin #5 (C) to the alarm device return.
c. Pin #6 (NO) is not used for fail-safe operation
3. Connect the XDP Low alarm (alarm 3) as follows:
a. Connect pin #7 (NC) to the alarm device input.
b. Connect pin #8 (C) to the alarm device return.
c. Pin #9 (NO) is not used for fail-safe operation.
4. Connect the XDP Low Low alarm (alarm 4) as follows:
a. Connect pin #10 (NC) to the alarm device input.
b. Connect pin #11 (C) to the alarm device return.
c. Pin #12 (NO) is not used for fail-safe operation.
2-6Installation
January 2008
Wiring the Calibration
Relays - J7
T erminal block J7 contains connections for the four calibration relays.
To wire these relays, complete the following steps:
Note: These relays are configured for non-fail-safe operation.
1. Connect the Pr ocess/Cal relay (Relay 1) as follows:
a. Pin #1 (NC) is not used for non-fail-safe operation.
b. Connect pin #2 (C) to the alarm device return.
c. Connect pin #3 (NO) to the alarm device input.
2. Connect the Zero/Span Cal relay (Relay 2) as follows:
a. Pin #4 (NC) is not used for non-fail-safe operation.
b. Connect pin #5 (C) to the alarm device return.
c. Connect pin #6 (NO) to the alarm device input.
3. Connect the AutoCal Indication alarm (Relay 3) as follows:
a. Pin #7 (NC) is not used for non-fail-safe operation.
b. Connect pin #8 (C) to the alarm device return.
Wiring the System Analog
Output - J5
c. Connect pin #9 (NO) to the alarm device input.
4. Connect the Auto Verification alarm (Relay 4) as follows:
a. Pin #10 (NC) is not used for non-fail-safe operation.
b. Connect pin #11 (C) to the alarm device return.
c. Connect pin #12 (NO) to the alarm device input.
T e rminal block J5 contains connections for the system an alog out put.
To wire the system analog output, complete the following steps:
1. Connect pin #1 to positive input on the device.
2. Connect pin #2 to the return line from the device.
3. Pin #3 has no connection.
Installation2-7
January 2008
Wiring the XDP Analog
Output - J8
Terminal block J8 contains connections for the XDP analog output.
To wire the XDP analog output, complete the following steps:
1. Connect pin #1 to positive input on the device.
2. Connect pin #2 to the return line from device.
3. Pin #3 has no connection.
Wiring the Serial Port - J4Terminal block J4 contains connections for the RS-232/RS-485
(RS-485 supports multi-drop installations) serial port. To wire the
serial port, complete the following steps:
1. Connect pin #1 to the transmit pin on the PC serial port.
2. Connect pin #2 to the receive pin on the PC serial port.
3. Connect pin #3 to the return pin on the PC serial port.
Note: This connection may be made with a GE Sensing
#704-668 cable. If this cable is used, connect the red wire to
pin #1, the white wire to pin #2, and the green wire to pin #3.
See EIA-RS Serial Communications (document 916-054) for a
detailed discussion of serial port wiring.
Wiring The Line Power TB1
Terminal block TB1 on the power supply board contains connections
for the line power to the XDP. Make these connections only with wire
that meets the following specifications:
• individual conductor gauge of 18 AWG minimum (maximum
current capacity of 12 Amps)
• voltage rating of 600 Vo lts minimum
• insulation temperature rating of 105°C minimum
!WARNING!
The rating of the wire insulation must be at least 15°C
above the expected ambient temperature.
2-8Installation
January 2008
Wiring The Line Power - J1
(cont.)
Refer to Figure 2-3 on page 2-12 and complete the following steps:
!WARNING!
Before proceeding with the section, verify that the line
power has been turned off at the external disconnect
device.
1. Attach a cable or conduit with the three line power conductors to a
cable entry port on the XDP enclosure.
Caution!
Be sure that the input voltage complies with the input
power rating for the XDP (see Chapter 7, Specifications).
2. Connect the line power leads to terminal block J1 as follows:
a. Connect the line power lead (black or brown) to pin
L on the
power connector.
b. Connect the neutral power lead (white or blue) to pin
N on the
power connector.
c. Connect the ground power lead (green or green/yellow) to the
chassis ground screw shown in Figure 2-2 on page 2-10.
3. Make sure that the factory-installed jumper wire at pin
G on the
power connector is secure.
This completes the wiring of the XDP. Proceed to Chapter 3,
Operation, for instructions on using the meter.
disconnect device such as a switch or
circuit breaker. The disconnect device
must be marked as such, clearly visible,
directly accessible and located within 1.8
m (6 ft) of the XDP.
IntroductionThe XDP is an easily operated monitoring/control device. Make sure
that the system has been installed in accordance with the instructions
given in Chapter 2, Installation, before applying power. This chapter
includes discussions of the following topics:
• preventing common problems
• powering up the system
• programming the meter
• taking measurements
!WARNING!
To ensure safe operation of the XDP, it must be installed
and operated as described in this manual. In addition, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment.
Preventing Common
Problems
The complexity of accurately measuring the target gas mixtures
dictates that some basic precautions be observed in using the XDP.
Failure to observe these simple procedures is often the cause of some
common problems. Compliance with the following points will help to
eliminate such problems:
• The transmitter must be properly connected to the XDP before
applying power to the system. Connecting or disconnecting the
transmitter while the XDP is powered up may generate a 24 Volt Power Fault message. If this should occur, cycle the power to the
system OFF and then ON to clear the problem.
• The automatic calibration feature should be enabled and scheduled
to be run on a regular basis. The optimum calibration interval
depends on the application (consult the factory).
• Never use thread sealant on any connection in the sample gas flow
path. Thread sealant emits combustible vapors that can cause
reading errors.
• The infrared keypad may not function properly if direct sunlight
strikes the face of the XDP or if the XDP is installed in an area that
has other infrared devices in operation. To minimize such factors,
provide suitable shading at the installation site.
Cleaning the EnclosureIf the glass window or the case of the XDP becomes soiled, use a soft
cloth dampened with water for cleaning. Never use solvents or
detergents to clean the XDP.
Operation3-1
January 2008
Powering Up the SystemCheck the wiring connections and close both the XDP and transmitter
enclosures before applying power. Energize the external disconnect
device to power up the XDP. Allow the transmitter to warm up for at
least thirty minutes, before taking measurements.
IMPORTANT:The XDP requires an external power disconnect
device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m
(6 ft) of the unit.
The Enclosure WindowThe window at the top of the XDP enclosure includes the components
shown in Figure 3-1 below.
Display Window
ESC Key
LEFT Arrow
UP Arrow
ENTER Key
RIGHT Arrow
DOWN Arrow
Figure 3-1: The Enclosure Window
The XDP is factory-configured to display O2 ppm/pcnt with a
resolution of 0.01 as the default. However, changing the gas label, the
gas units, and the display resolution are easily accomplished. See
Chapter 4, Setup, for programming instructions.
IMPORTANT:The XDP’s infrared keypad permits programming of
the instrument through the glass faceplate without
removing the cover. Thus, all programming
procedures may be performed while the unit is
installed in a hazardous area.
3-2Operation
January 2008
Initial Screen DisplaysImmediately upon power up, the digital display shows the following
sequence of informational screens, as the meter performs its internal
systems checks.
BSeries Loader v1.05
Boot is Flash.
Program CRC valid.
Booting from Flash
If the meter passes all of its internal system checks, the following
initial data screen appears:
This is a typical example of
the first initialization screen.
This is a typical example of
the second initialization
screen.
This is a typical example of
the final initialization screen.
1 O2HH:MM
0.00
Note: The padlock icon at the upper right hand corner of the above
display indicates that the User Program is locked with
password protection.
±
pcnt
LL
Notice that the absence of a
valid signal has triggered the
Low Low alarm (the LL symbol in the lower right corner).
Operation3-3
January 2008
Starting the Sample GasStart the sample gas flow to the transmitter and allow a few minutes
for the system to reach equilibrium. The following data screen should
appear if the oxygen level is above the Low alarm setpoint and below
the High alarm setpoint.
Accessing the User
Program
1 O2HH:MM±
pcnt
X.XX
Note: Allow approximately 30 minutes for the sensor to reach
temperature equilibrium, to ensure accurate readings
Notice that the display shows the slot number, the ga s label, and the
current time on the top line. The current data value along with the
units are displayed across the bottom of the screen.
Note: Slot 0 is the main circuit board, slot 1 holds the XDP circuit
board, and slots 2-4 are normally not used.
To configure the XDP for a specific application, it is necessary to
leave measurement mode and enter the User Program as follows (see
the menu maps in Figures C-1 and C-2 on pages C-1 and C-2):
1 O2HH:MM±
pcnt
X.XX
This is a typical display. The
actual oxygen concentration of
your sample gas will be shown
here.
While in measurement mode,
press the
[ESC], [ENTER], and [ESC] keys
in sequence.
[ENTER], [ENTER],
Enter the operator-level password (2719) at the next screen to access
the menus available to the end user. (A different password is required
to access the factory-level menus.)
Password
Enter Password
XXXX
s
CalOptDisp
3-4Operation
Use [W] and [X] to position the
cursor under the desired character. Then, use [
increment the value. Press
[ENTER] when done.
This screen shows the highest
level of the User Program,
which is referred to as the
Main Menu.
S] and [T] to
January 2008
Exiting the User ProgramT o leave the User Program, proceed as follows from the main menu:
CalOptDisp
1 O2HH:MM
pcnt
Press the [ESC] key.
You are now back in normal
measurement mode.
X.XX
Notice that the padlock is no longer visible at the upper right corner
of the display window. Once the User Program has been accessed
with a valid password, the menus at that programming level remain
unlocked when you return to measurement mode. With the menus
unlocked, you may reenter the User Program by simply pressing the
[ESC] key only. To re-lock the menus, see the instructions in Chapter
4, Setup.
Note: If the unit is powered down, the User Program will be locked
when the unit is powered up again.
To program a specific feature of the XDP, proceed to the appropriate
section of Chapter 4, Setup.
Operation3-5
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.