The GE Sensing & Inspection Technologies VeriDriTM is a lowcost, loop-powered transmitter that provides accurate dew/frost
point measurements covering an overall range of –110
o
(-166
to 104oF). It can also be provided to cover a number of
moisture ranges from 0 to 10,000 PPMv.
The VeriDri is supplied with one 4 to 20 mA analog output that is
factory-configured for a specified range.
The VeriDri is easy to install, operate and maintain. This manual
contains the following sections:
o
to 40°C
• Sample System Guidelines
• Inserting the Transmitter in the Sample System/Process
• Making Wiring Connections
• Operating the Transmitter
• Cleaning the Transmitter Probe
• Specifications
VeriDri Moisture Transmitter1
January 2009
Sample System Guidelines
You can install the VeriDri transmitter into a sample system or
directly into the process. GE recommends that the unit be
installed in a sample system to protect the probe from coming
into contact with damaging elements in the process.
Before constructing a sample system, you should consult a GE
Applications Engineer and adhere to the guidelines below. See
Figure 1 on the next page for an example of a sample system.
• Consult Specifications on page 13 for dimensions and other
requirements.
• Sample systems should be kept very simple.
• The transmitter should be installed so it is perpendicular to
the sample inlet.
• The sample system should contain as few components as
possible and all or most of those components should be
located downstream of the measurement location.
• If possible, you should use stainless steel material for all
wetted parts.
• Sample system components should not be made of any
material that will affect measurement. Most common filters
and pressure regulators are not suitable for sample systems
because they have wetted parts that may absorb or release
components (such as moisture) into the sample system.
They may also allow ambient contamination to enter the
sample system.
2VeriDri Moisture Transmitter
Sample System Guidelines (cont.)
Sample
Cell
January 2009
Sample
Inlet
Vent
Figure 1: Example of a Sample System
Sample
Outlet
VeriDri Moisture Transmitter3
January 2009
Inserting the Transmitter in the Sample
System/Process
!CAUTION!
If you are mounting the VeriDri directly into
the process line, you must consult the factory
for proper installation instructions and precautions
before beginning the following procedure.
Use the steps below to install the VeriDri:
1. Make sure the sintered or sheet stainless-steel shield is in
place. The shield protects the aluminum oxide sensor from
damaging elements in the process.
Shield
2. The probe is mounted into the process via the 3/4-16 straight
male thread located on the probe. Thread the probe end of the
transmitter into the process/sample system fitting. Make sure
not to cross thread it.
3/4-16 UNF-2A
4VeriDri Moisture Transmitter
January 2009
Inserting the Transmitter in the Sample
System/Process (cont.)
3. Using a 1-1/8 in. wrench, tighten the probe securely into the
process using the probe hex nut.
!CAUTION!
Do not apply force to the transmitter
module to tighten the unit into its fitting.
Probe Hex Nut
VeriDri Moisture Transmitter5
January 2009
Making Wiring Connections
You must wire the transmitter using the factory-supplied cable
which is available in a variety of lengths.
Note: If you need to lengthen cables, refer to Table 1 below to
splice an extension onto the existing cable. Connect
positive to positive and negative to negative.
Use the following steps to wire the transmitter to your system.
1. Push the female connector end on the transmitter cable into
the mating male connector on the transmitter module. Make
sure the pins are properly aligned. Once inserted, secure the
connectors together by sliding the metal sleeve on the cable
over the connectors and turning it clockwise until it is tight.
2. Using the flying leads at the other end of the transmitt er cable,
connect the transmitter to your power supply and data
acquisition system (DAS) as shown in Figure 2 on the next
page. Refer to Table 1 below for a description of the leads in
the factory-supplied cable.
Table 1: Cable Lead Descriptions
Lead Color*Connection Description
Blue(+) 7 to 28 VDC
Brown (−) 7 to 28 VDC
Connect the shield to ground if desired.
*The blue and brown leads also produce a current output
equivalent to 4 to 20 mA.
3. Trim any unused leads back to the outer cable jacket in order
to remove the bare tinned wire and prevent accidental short
circuits.
The VeriDri is now ready for operation.
6VeriDri Moisture Transmitter
Making Wiring Connections (cont.)
January 2009
Note: The blue and brown leads also generate a
current output equivalent to 4 to 20 mA.
+
7-28 VDC
Power and
Blue
Brown
7-28 VDC
−
Circuits
Measurement
Figure 2: Wiring Connections
VeriDri Moisture Transmitter7
January 2009
Operating the Transmitter
After proper installation, the VeriDri Transmit ter is very easy to
use; simply power the unit up and you are ready to begin taking
measurements. Since the VeriDri stores moisture calibration data
in non-volatile FLASH memory, you do not have to enter data
manually or worry about losing data during a power loss.
Probes may need to be cleaned occasionally depending on the
application. Consult a GE Applications Engineer for required
cleaning intervals.
If a problem should arise with the probe, see Error Handling
below for how the transmitter reacts to error conditions.
!CAUTION!
Any attempt to open the module or remove
the sensor probe will void the warranty.
Powering Up
After the VeriDri is wired as described in the previous section,
you may apply power to the unit. The transmitter takes
approximately 20 seconds to initialize and begin normal
operation. The unit will meet specified accuracy in 3 minutes.
Error Handling
In the event of an error condition, the analog output reading is
forced to the following values:
• ≥ 22 mA to indicate a shorted probe
• ≤ 3.5 mA to indicate an open probe
8VeriDri Moisture Transmitter
January 2009
Cleaning the Transmitter Probe
!CAUTION!
Be sure to perform the probe cleaning procedure in a
well ventilated area. Observe all necessary safety
precautions when handling cleaning solvents
To clean the moisture probe, you will need the following:
• Three glass (NOT metal) containers containing the
following solvents:
• 2 containers of approximately 300 ml of reagent-grade
hexane or toluene each.
• 1 container of approximately 300 ml of distilled (NOT
deionized) water.
IMPORTANT:Make sure the containers are deep enough to
submerge the pr obe. Do not place the transmitter
module into the solvents. You should only insert
the sensor portion of the transmitter into the
solvents.
• Rubber or latex gloves.
• Oven set at 50°C ± 2°C (122°F ± 3.6°F) for drying probe.
• 1-1/8 in. wrench
Use the sections that follow to properly clean the probe.
VeriDri Moisture Transmitter9
January 2009
Removing the Transmitter
Use the following steps to remove the transmitter from the
installation site:
Note: Once the probe is cleaned, it must dry in the oven for 24
hours.
1. Using a 1-1/8 in. wrench, unthread the transmitter from the
fitting on the sample system/process using the probe hex nut
as shown below.
Probe Hex Nut
2. Record the dew point of the ambient air.
3. Disconnect the cable from the module.
4. Unscrew the stainless-steel shield from the probe mount and
carefully remove it without touching the sensor.
!CAUTION!
Any attempt to open the module or remove
the probe from the module will void the warranty.
Shield
10VeriDri Moisture Transmitter
January 2009
Soaking the Sensor and Shield
!CAUTION!
Do not place the transmitter module into the solvents.
You should only insert the sensor portion of the
transmitter into the solvents. Also, do not allow the
sensor to come into contact with the surfaces of the
cleaning containers or any other hard surface.
1. Wearing gloves, place the sensor in the first container of
hexane or toluene and allow it to soak for 10 minutes.
2. Remove the sensor from the hexane or toluene and soak it in
the container of distilled water for 10 minutes.
3. Remove the sensor from the distilled water and soak it in the
second (clean) container of hexane or toluene for 10 minutes.
4. Remove the sensor from the hexane or toluene and set it aside
until the shield has completed the cleaning cycle.
5. Repeat steps 1 to 3, above, to clean the shield. To ensure the
removal of any contaminants that may have become
embedded in the porous walls of the shield, swirl the shield in
the solvents during the soaking procedure.
6. Remove the shield from the hexane or toluene.
7. Carefully replace the shield over the exposed sensor without
touching it.
8. Place the sensor with the shield in an oven set at 50°C ± 2°C
(122°F ± 3.6°F) for 24 hours.
VeriDri Moisture Transmitter11
January 2009
Evaluating the Probe
1. Re-connect the cable to the transmitter module and measure
the dew point. Make sure you measure the same ambient air as
measured in step 2 on page 10.
2. Compare the two ambient air readings. If the new ambient air
reading is within ±2°C (±3.6°F) of the first reading, the
cleaned probe is properly calibrated and may be reinstalled. If
it is not, proceed to step 3 below.
3. If the probe is still not reading the ambient air accurately,
repeat the cleaning procedure using soaking times that are 5
times the previous cleaning sequence, until two consecutive
ambient air readings are identical.
If the above cleaning procedure does not result in accurate
readings, contact the factory for assistance.
12VeriDri Moisture Transmitter
Specifications
Moisture Ranges
January 2009
• –110
• –110
o
to 20°C
o
to –50°C
• –90°C to 10°C
• –80°C to –30°C
• –80° to 20°C
o
• –30
• –130
• –130
• –40
• –50
to 20°C
o
to 70°F
o
to –40°F
o
to 70°F
o
to 50°F
• -100° to 0°F
• 0 to 10 PPMv
• 0 to 100 PPMv
• 0 to 1000 PPMv
• 0 to 10,000 PPMv
Note: PPMv ranges based on constant pressure, provided at
time of order placement.
Operating Temperature
–40° to 60°C (–40° to 140°F)
Storage Temperature
70°C (158oF) maximum. The probe should be stored with the
plastic cover and desiccant packet threaded onto the probe. Store
in a cool, dry environment.
VeriDri Moisture Transmitter13
January 2009
Warm-up Time
Meets specified accuracy in 3 minutes
Calibrated Accuracy @ 77°F (25°C)
• ±3.6°F (±2°C) from -85°F to 50°F (-65°C to 10°C) dew/frost
point
• ±5.4°F (±3°C) from -112°F to -86°F (-80°C to -66°C) dew/
frost point
Repeatability
• ±0.9°F (±0.5°C) from -85°F to 50°F (-65°C to 10°C) dew/frost
point
• ±1.8°F (±1.0°C) from -112°F to -86°F (-80°C to -66°C) dew/
frost point
Response Time
Less than 5 seconds for 63% of a step change in moisture content
in either wet-up or dry-down cycle
Electronics
Power
• Supply voltage: 7 to 28 VDC (loop-powered, customer
supplied)
• Output: 4 to 20 mA
• Output Resolution: 0.01 mA
• Max. Loop R = 50 Ω × (PSV–7),
where PSV = Power Supply Voltage
Example: Given a 24 VDC Power Supply,
Max. Loop R = 50 Ω × (24–7)
= 850 Ω
• Cable: 2 m (6 ft.) standard (consult factory for custom lengths)
14VeriDri Moisture Transmitter
Mechanical
Sample Connection
• 3/4 in. (16 mm) 16 straight male thread with O-ring
• G 1/2 in with optional adapter
Operating Pressure
µHg to 5,000 psig (345 bar)
5
Enclosure
Type 4X/IP67
European Compliance
Complies with the following:
• EMC Directive 2004/108/EC and PED 97/23/EC for
DN<25
• EN 61326:1998
Class A, Annex A, Continuous Unmonitored Operation
(For EN 61000-4-3 transmitter meets performance criteria
A and in a number of frequencies, criteria B per EN 61326)
Dimensions
January 2009
• Overall: 6.76 × 1.13 in. (17.17 × 2.87 cm)
• Electronics with cable: 4.08 × 1.13 in. diameter
(10.36 × 2.87 cm)
• Weight: 5 oz (140 grams)
VeriDri Moisture Transmitter15
January 2009
Moisture Sensor
Sensor Type
Thin-film aluminum oxide moisture sensor probe
Calibration
Each sensor is individually computer-calibrated against known
moisture concentrations, traceable to NIST
Calibration Interval
Sensor recalibration at GE Sensing is recommended every six to
12 months depending on application
Calibration Data
Factory-calibrated, stored in FLASH
Flow Rate
• Gases: Static to 10,000-cm/s linear velocity at a pressure of 1
atm.
• Liquids: Static to 10-cm/s linear velocity at density of
1 g/cc
16VeriDri Moisture Transmitter
January 2009
Warranty
Each instrument manufactured by GE Sensing, Inc. is
warranted to be free from defects in material and
workmanship. Liability under this warranty is limited to
restoring the instrument to normal operation or replacing the
instrument, at the sole discretion of GE Sensing, Inc. Fuses and
batteries are specifically excluded from any liability. This
warranty is effective from the date of delivery to the original
purchaser. If GE Sensing, Inc. determines that the equipment
was defective, the warranty period is:
• one year from delivery for electronic or mechanical failures
• one year from delivery for sensor shelf life
If GE Sensing, Inc. determines that the equipment was
damaged by misuse, improper installation, the use of
unauthorized replacement parts, or operating conditions
outside the guidelines specified by GE Sensing, Inc., the repairs
are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu
of all other warranties whether statutory, express or
implied (including warranties or merchantability and fitness
for a particular purpose, and warranties arising from course
of dealing or usage or trade).
January 2009
Return Policy
If a GE Sensing, Inc. instrument malfunctions within the
warranty period, the following procedure must be completed:
1. Notify GE Sensing, Inc., giving full details of the problem,
and provide the model number and serial number of the
instrument. If the nature of the problem indicates the need
for factory service, GE Sensing, Inc. will issue a RETURN
AUTHORIZATION NUMBER (RAN), and shipping instructions
for the return of the instrument to a service center will be
provided.
2. If GE Sensing, Inc. instructs you to send your instrument to a
service center, it must be shipped prepaid to the authorized
repair station indicated in the shipping instructions.
3. Upon receipt, GE Sensing, Inc. will evaluate the instrument
to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
• If the damage is covered under the terms of the warranty,
the instrument will be repaired at no cost to the owner and
returned.
• If GE Sensing, Inc. determines that the damage is not
covered under the terms of the warranty, or if the warranty
has expired, an estimate for the cost of the repairs at
standard rates will be provided. Upon receipt of the owner’s
approval to proceed, the instrument will be repaired and
returned.
GE
Sensing
DECLARATION
OF
CONFORMITY
We,GE Sensing
declare under our sole responsibility that the
to which this declaration relates, is in conformity with the following standards:
• EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation
(For EN 61000-4-3, the above units meet performance Criteria A, and in a limited
number of frequencies, performance Criteria B per EN 61326.)
• EN 60529:1991+A1:2000
IP67
following the provisions of the 2004/108/EC EMC Directive.
The units listed above and any ancillary sample handling systems supplied with them do
not bear CE marking for the Pressure Equipment Directive, as they are supplied in
accordance with Article 3, Section 3 (sound engineering practices and codes of good
workmanship) of the Pressure Equipment Directive 97/23/EC for DN<25.
January 8, 2009
Date of Issue
1100 Technology Park Drive
Billerica, MA 01821
USA
VeriDri™ Moisture Transmitter
Mr. Gary Kozinski
Certification & Standards, Lead Engineer
CERT-DOC-H1 (August 2004)
GE
Sensing
DECLARATION
DE
CONFORMITE
Nous,GE Sensing
déclarons comme étant de notre seule responsabilité que le
sur lequel porte ce document, est conforme aux spécifications suivantes :
• EN 61326:1998, Classe A, Annexe A, Fonctionnement continu sans surveillance
(Pour EN 61000-4-3, les appareils ci-dessus sont conformes au critère A de performance
et, dans un nombre limité de fréquences, au critère B de performance, selon EN 61326.)
• EN 60529:1991+A1:2000
IP67
conformément aux dispositions de la directive 2004/108/EC (compatibilité
électromagnétique).
Les appareils indiqués plus haut et tous les accessoires d’échantillonnage fournis avec ne
portent pas la marque CE pour la directive concernant les équipements de pression, dans la
mesure où ils sont fournis conformément à l’article 3, section 3 (pratiques d’ingénierie sûres
et codes de bienfacture) de la directive concernant les équipements de pression 97/23/EC
pour DN<25.
1100 Technology Park Drive
Billerica, MA 01821
USA
Transmetteur d’humidité VeriDri™
8 janvier 2009
Date d’émission
Certification et normes, ingénieur de fil
Mr. Gary Kozinski
CERT-DOC-H1 (August 2004)
GE
Sensing
KONFORMITÄTS-
ERKLÄRUNG
Wir,GE Sensing
erklären unter alleiniger Eigenverantwortlichkeit, dass die Produkte
auf das sich diese Deklaration bezieht, die folgenden Normen erfüllen:
• EN 61326:1998, Class A, Annex A, kontinuierlicher, überwachungsfreier Betrieb
(Für EN 61000-4-3 erfüllen die obigen Geräte die Leistungskriterien A und in bei einer
begrenzten Anzahl von Frequenzen die Leistungskriterien B gemäß EN 61326.)
• EN 60529:1991+A1:2000
IP67
und dass sie die Anforderungen der EMC-Direktive 2004/108/EC einhalten.
Die oben aufgeführten Geräte und zugehörige, mitgelieferte Schallwandler (Messrohre
werden in einer separaten Konformitätserklärung behandelt) tragen keine
CE-Kennzeichnung gemäß der Druckgeräte-Richtlinie, da sie in Übereinstimmung mit
Artikel 3, Absatz 3 (gute Ingenieurpraxis) der Druckgeräte-Richtlinie 97/23/EG für DN<25
geliefert werden.
1100 Technology Park Drive
Billerica, MA 01821
USA
VeriDri™ Feuchtigkeitsmesswandler
8. Januar 2009
Außtellungsdatum
Bescheinigung und Normen, Leitungsingenieur
Hr. Gary Kozinski
CERT-DOC-H1 (August 2004)
USA
1100 Technology Park Drive
Billerica, MA 01821-4111
Web: www.gesensinginspection.com
Ireland
Shannon Industrial Estate
Shannon, County Clare
Ireland
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.