This manual is a reference for the LOGIQ S8/LOGIQ E8 and LOGIQ S8 Vet ultrasound systems (Hereafter
listed as LOGIQ S8). All information provided in this manual is relevant for all three systems unless otherwise
specified.
Page 3
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Important Precautions
TRANSLATION POLICY
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING
(EN)
AVERTISSEMENT
(FR)
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
(DE)
DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
- i
Page 4
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
AVISO
(ES)
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.
ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
ATENÇÃO
(PT-Br)
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÃNICOS OU OUTROS.
AVISO
(PT-pt)
AVVERTENZA
(IT)
ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER
CONSULTADO E COMPREENDIDO PREVIAMENTE ESTE MANUAL.
• A INOBSERVÂNCIA DESTE AVISO PODE RESULTAR EM FERIMENTOS NO
TÉCNICO DE ASSISTÊNCIA, OPERADOR OU PACIENTE EM CONSEQUÊNCIA
DE CHOQUE ELÉCTRICO, PERIGOS DE ORIGEM MECÂNICA, BEM COMO DE
OUTROS TIPOS.
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
ii -
Page 5
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
HOIATUS
(ET)
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
VAROITUS
(FI)
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
ΤΟΠΑΡΟΝΕΓΧΕΙΡΙΔΙΟΣΕΡΒΙΣΔΙΑΤΙΘΕΤΑΙΣΤΑΑΓΓΛΙΚΑΜΟΝΟ.
• ΕΑΝΤΟΑΤΟΜΟΠΑΡΟΧΗΣΣΕΡΒΙΣΕΝΟΣΠΕΛΑΤΗΑΠΑΙΤΕΙΤΟΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΕΚΤΟΣ
ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.
EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
- iii
Page 6
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
VIÐVÖRUN
(IS)
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG
SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
•VPŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD
V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
VÝSTRAHA
(CS)
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
•VPŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
ADVARSEL
(DA)
WAARSCHUWING
(NL)
DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
iv -
Page 7
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
BRĪDINĀJUMS
(LV)
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
•ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
ĮSPĖJIMAS
(LT)
•NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SM
ŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
ADVARSEL
(NO)
OSTRZEŻENIE
(PL)
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
•JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOŻE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEUŃ.
- v
Page 8
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
ATE NŢIE
(RO)
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО
НА АНГЛИЙСКОМ ЯЗЫКЕ.
• ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО
НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
СЕ НУЖДАЕ ОТ
ЕЗИК, РАЗЛИЧЕН ОТ АНГЛИЙСКИ, ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА
ПРЕДОСТАВИ ПРЕВОДАЧЕСКА УСЛУГА.
• НЕ СЕ ОПИТВАЙТЕ ДА ИЗВЪРШВАТЕ СЕРВИЗНО ОБСЛУЖВАНЕ НА ТОВА
ОБОРУДВАНЕ, ОСВЕН ВСЛУЧАЙ, ЧЕ СЕРВИЗНОТО РЪКОВОДСТВО Е
ПРОЧЕТЕНО И СЕ РАЗБИРА.
• НЕСПАЗВАНЕТО НА ТОВА ПРЕДУПРЕЖДЕНИЕ МОЖЕ ДА ДОВЕДЕ ДО
НАРАНЯВАНЕ НА ДОСТАВЧИКА НА СЕРВИЗНИ УСЛУГИ, НА
ОПЕРАТОРА
ИЛИ ПАЦИЕНТА ВСЛЕДСТВИЕНА ТОКОВ УДАР, МЕХАНИЧНИ ИЛИ ДРУГИ
РИСКОВЕ.
OVAJ PRIRUČNIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM
JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST JE NA KLIJENTU DA PRUŽI USLUGE PREVOĐENJA.
• NEMOJTE POKUŠAVATI DA SERVISIRATE OPREMU AKO NISTE PROČITALI I
RAZUMELI PRIRUČNIK ZA SERVISIRANJE.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO POVREĐIVANJA
SERVISERA, OPERATERA ILI PACIJENTA UZROKOVANOG ELEKTRIČNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
vi -
Page 9
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI.
• ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V
DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA.
OPOZORILO
(SL)
• NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN
RAZUMELI SERVISNI PRIROČNIK.
• ČE TEGA OPOZORILA NE UPOŠTEVATE, OBSTAJA NEVARNOST
ELEKTRIČNEGA UDARA, MEHANSKIH ALI DRUGIH NEVARNOSTI IN
POSLEDIČNIH POŠKODB PONUDNIKA SERVISNIH STORITEV, UPORABNIKA
OPREME ALI PACIENTA.
OVAJ SERVISNI PRIRUČNIK DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTIJEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST KLIJENTA JE PRUŽITI USLUGE PREVOĐENJA.
UPOZORENJE
(HR)
• NEMOJTE POKUŠAVATI SERVISIRATI OPREMU AKO NISTE PROČITALI I
RAZUMJELI SERVISNI PRIRUČNIK.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO OZLJEDE
SERVISERA, OPERATERA ILI PACIJENTA PROUZROČENE STRUJNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
UPOZORNENIE
(SK)
VARNING
(SV)
TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
- vii
Page 10
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
(JA)
Traditional
Chinese
viii -
Page 11
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
(ZH-CN)
(KO)
- ix
Page 12
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
x -
Page 13
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be perfor med by qualified GE Healthcare personnel.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the sys tem type, manual
title, part number, revision number, page number and suggestion details.
Mail the information to:
Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI53226, USA.
GE Healthcare employees should use Application Lifecycle Management (ALM) to report service
documentation issues. These issues will then be in the internal problem reporting tool and
communicated to the writer.
SERVICE SAFETY CONSIDERATIONS
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNINGWARNING
USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY.
For a complete review of all safety requirements, see the Chap te r 1, Saf et y Con sid er at ion s sec tio n in
the Service Manual.
- xi
Page 14
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
PROPRIETARY TO GE HEALTHCARE
Permission to use this Advanced Service Software and related documentation (herein called the
material) by persons other than GE Healthcare emp loyees is provided only under an Advanced Service
Package License relating specifically to this Proprietary Material. This is a different agreement from the
one under which operating and basic service software is licensed. A license to use operating or basic
service software does not extend to or cover this software or related documentation.
If you are a GE Healthcare employee or a customer who has entered into such a license agreement
with GE Healthcare to use this proprietary software, you are authorized to use this Material according
to the conditions stated in your license agreement.
However, you do not have the permission of GE Healthcare to alter, decompose or reverse-assemble
the software, and unless you are a GE employee, you MAY NOT COPY the M ate ri al. T he Mat er ial is
protected by Copyright and Trade Secret laws; the violation of which can result in civil damages and
criminal prosecution.
If you are not party to such a license agre em e nt or a GE Healthcare Employee, you must exit this
Material now.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
9Februry, 2016Add UDI label information
10Februry, 2017Update adding R4 information
11June, 2017Revise with Service Note information
12October, 2017Add Installation Instruction information
13March, 2019Update for R4.2.5x release
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
- 16
Page 33
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1Purpose of this chapter
This chapter describes important issues related to safely servicing LOGIQ S8. The service provider
must read and understand all the information presented here before installing or servicing a unit.
Attention
This manual contains necessary and sufficient information for the Field Service Engineer or Biotech
Engineer to maintain and service the system safely. Advanced equipment training may be provided b y
factory trained Field Service trainers for the agreed-upon time period.
This service manual provides installation and service information for the LOGIQ S8 ultrasound system
as shown in Table 1-1 "Contents in this service manual" on page 1- 2.
NOTE:Ensure you are working with the latest version of the Proprietary Service Manual. Consult the
CDL and OnBase to confirm and to download the latest version.
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-2
Service manual overview
1-2-1Contents in this service manual
The service manual is divided into ten chapters.
In the beginning of the manual, before chapter 1, you will find the language policy for GE Healthcare’s
service documentation, legal information , a re visio n over vie w, an d th e Ta b le of Contents (TOC).
Table 1-1 Contents in this service manual
CHAPTER
NUMBER
CHAPTER TITLEDESCRIPTION
1
2
3
4
5
6
7
8
9
10
N/A
IntroductionContains a content summary and warnings.
Site preparations
System SetupContains setup procedure with procedure checklist.
Functional Checks
Components and Functions
(Theory)
Service Adjustments
Diagnostics/Troubleshooting
Replacement Procedures
Renewal Parts
Care & Maintenance
IndexA quick way to the topic you’re looking for.
Contains pre-setup requirements for the LOGIQ S8.
Contains functional checks that must be performed as part of the
installation, or as required during servicing and periodic
maintenance.
Contains block diagrams and functional explanations of the
electronics.
Contains instructions on how to make any available adjustments
to the LOGIQ S8.
Provides procedures for running diagnostic or related routines for
the LOGIQ S8.
Provides disassembly procedures and reassembly procedures
for all changeable FRUs, available option installation
instructions, and upgrade installation instructions.
Contains a complete list of replacement parts for the LOGIQ S8.
Provides periodic maintenance procedures for LOGIQ S8.
1 - 2Section 1-2 - Service manual overview
Page 35
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-2-2Typical users of the Service Manual
•GE Service Personnel (Setup, maintenance, etc.)
•Hospital Service Personnel
•Architectural Planners/Installation Planners (some parts of Chapter 2 - Site preparations)
Section 1-2 - Service manual overview1 - 3
Page 36
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Software Configurations and Hardware/Software Compatibility - Upgrade Options
Y: Original, U: Upgrade available, N: Not supported
U*:
Upgrade available. (No LCD upgrade)
Section 1-2 - Service manual overview1 - 5
Page 38
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-2-4Product description
1-2-4-1Overview of the LOGIQ S8/LOGIQ E8 ultrasound system
The LOGIQ S8/LOGIQ E8 ultrasound system is a high performance digital ultrasound imaging system
with total data management.
This system provides image generation in B-Mode, Color Doppler, Power Doppler, M-Mode, Color MMode, PW/CW and 3D/4D, Tissue Velocity imaging, Volume-Guided Ultrasound, Elastograph y, and
Contrast applications.
The fully digital architecture of the LOGIQ S8/LOGIQ E8 allows optimal usage of all scanning modes
and probe types throughout the full spectrum of operating frequencies.
Signal flows from the Probe Connector Panel to the Front End, and then over to the Back End Pro cessor
and finally to the monitor and peripherals.
System configuration is stored on the hard drive in the Back End Processor.
All necessary software is loaded from the hard drive on power up.
1 - 6Section 1-2 - Service manual overview
Page 39
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-3
Important conventions
1-3-1Conventions used in this manual
1-3-1-1Icons
The pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels, and
conventions used on the product and in the service information are described in this chapter.
1-3-1-2Safety precaution messages
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the precautionar y
message. Known or potential hazards to personnel are labeled in one of three ways:
DANGER
DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL CAUSE
SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED.
WARNINGWARNING
WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE
SEVERE PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE
IGNORED.
CAUTION
NOTICE
NOTE:Notes are used to provide important information about an item or a procedure. Be sure to read
CAUTION IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL OR CAN CAUSE
MINOR PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE IGNORED.
EQUIPMENT DAMAGE POSSIBLE.
When a hazard is present that can cause property damage, but has absolutely no personal injury risk,
a NOTICE is used.
Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
Notice Example: Disk drive may crash.
the notes; the information contained in a note can of te n sav e yo u tim e or effo rt .
Section 1-3 - Important conventions1 - 7
Page 40
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-3-2Standard Hazard Icons
Important information will always be preceded by the exclamation point (!) contained within a triangle,
or the symbols for “Danger”, “Warning” or “Caution”, as see n throughout this chapter. In addition to te xt,
several different graphical icons (symbols) may be used to make you aware of specific types of hazards
that could cause harm.
Even if a symbol isn’t used in this manual, it is included for your reference.
Table 1-5 Standard Hazard Icons
ELECTRICALBIOHAZARDRADIATION
or
LASERHEATPINCH
LASER
LIGHT
ACOUSTIC OUTPUTEXPLOSION
DANGER/WARNING/
MOVING
CAUTIONSMOKE/FIRE
or
or
1 - 8Section 1-3 - Important conventions
Page 41
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-3-2-1Standard Icons that indicate that a special procedure is to be used
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-6 Standard Icons Indicating a Special Procedure be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
WEAR HAND PROTECTIONWEAR FOOT PROTECTION
Section 1-3 - Important conventions1 - 9
Page 42
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-3-3Product Icons
The following table describes the purpose and location of safety labels and ot her important inform ation
provided on the equipment.
Table 1-7 Label Icons
Label/IconPurpose/MeaningLocation
Identification and Rating Plate
Manufacturer’s name and address
Identification and Rating Plate
Date of manufacture
Serial NumberRating Plate
Catalog NumberRating Plate
United States only
Prescription Requirement label
CE Mark
The CE Mark of Conformity indicates this equipment conforms with
the Council Directive 93/42/EEC.
Authorized European Representative addressRear of the system
Rating Plate
Rating Plate
Rear of the system
Rear of the system
IP Code (IPX8)
Indicates the degree of protection provided by the enclosure per
IEC60 529. Can be used in operating room environment.
Type B Applied Part symbol is in accordance with IEC 60878-02-03.
Type BF Applied Part (man in the box) symbol is in accordance with
IEC 60878-02-03.
Defibrillation-proof applied part type BF.ECG connector
Type CF Applied Part (heart in the box) symbol is in accordance with
IEC 60878-02-03.
1 - 10Section 1-3 - Important conventions
Footswitch
Probe marked Type B
Probe marked Type BF
eTrax Needle
Page 43
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Table 1-7 Label Icons(Continued)
Label/IconPurpose/MeaningLocation
Follow instruction for use.
Symbol indicating that the Instructions for Use are supplied in
electronic form.
“General Warning Sign”Rear of the system
“Warning” - Dangerous voltage” (the lightning flash with arrowhead)
is used to indicate electric shock hazards.
“Mains OFF” indicates the power off position of the mains power
breaker.
“Mains ON” indicates the power on position of the mains power
breaker.
Rear of the system
Probe connector
Rear Panel
Internal
Rear of the system
Rear of the system
“ON” indicates the power on position of the power button.
CAUTION: This Power button DOES NOT ISOLATE Mains Supply.
Stand-by.
The standby symbol (line partially within a broken circle), indicates a
sleep mode or low power state.
Blue: Stand-by or sleep mode
Green: Power on
“Protective Earth” indicates the protective earth (grounding) terminal.Internal
R1/R2/R3 Operator
control panel
R4 Operator control
panel
Section 1-3 - Important conventions1 - 11
Page 44
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Table 1-7 Label Icons(Continued)
Label/IconPurpose/MeaningLocation
“Equipotentiality” indicates the terminal to be used for connecting
equipotential conductors when interconnectin g (grounding) with
other equipment.
Connection of additional protective earth conductors or potential
equalization conductors is not necessary in most cases and is only
recommended for situations involving multiple equipment in a highrisk patient environment to provide assurance that all equipment is at
the same potential and operates within acceptable leakage current
limits. An example of a high-risk patient would be a special procedure
where the patient has an accessible conductive path to the heart
such as exposed cardiac pacing leads.
Rear of the system
This symbol indicates that waste electrical and electronic equipment
must not be disposed of as unsorted municipal waste and must be
collected separately . Please contact an authorized representative of
the manufacturer for information concerning the decommissioning of
your equipment.
The separate collection symbol is affixed to a battery, or its
packaging, to advise you that the battery must be recycled or
disposed of in accordance with local or country laws. The letters
below the separate collection symbol indicate whether certain
elements (Pb=Lead, Cd=Cadmium, Hg=Mercury) are contained in
the battery. To minimize potential effects on the environment and
human health, it is important that all marked batteries that you
remove from the product are properly recycled or disposed. For
information on how the battery may be safely removed from the
device, please consult the service manual or equipment instructions.
Information on the potential effects on the environment and human
health of the substances used in batteries is available at this url: http:/
/www.gehealthcare.com/euen/weee-recycling/index.html
Indicates the presence of hazardous substance(s) above the
maximum concentration value. Maximum concentration values for
electronic information products, as set by the People’s Republic of
China Electronic Industry Standard SJ/T11364-2006, include the
hazardous substances of lead, mercury, hexavalent chromium,
cadmium, polybrominated biphenyl (PBB), and polybrominated
diphenyl ether (PBDE). “10” indicates the number of years during
which the hazardous substance(s) will not leak or mutate so that the
use of this product will not result in any severe environmental
pollution, bodily injury, or damage to any assets.
Rear of the system
Probe connector
Battery Pack if contains
Pb/Cd/Hg
Probe connector
1 - 12Section 1-3 - Important conventions
Page 45
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Table 1-7 Label Icons(Continued)
Label/IconPurpose/MeaningLocation
Indicates the presence of hazardous substance(s) above the
maximum concentration value. Maximum concentration values for
electronic information products, as set by the People’s Republic of
China Electronic Industry Standard SJ/T11364-2006, include the
hazardous substances of lead, mercury, hexavalent chromium,
cadmium, polybrominated biphenyl (PBB), and polybrominated
diphenyl ether (PBDE). “20” indicates the number of years during
which the hazardous substance(s) will not leak or mutate so that the
use of this product will not result in any severe environmental
pollution, bodily injury, or damage to any assets.
Do not use the following devices near this equipment: cellular phone,
radio receiver, mobile radio transmitter, radio controlled toy,
broadband power lines, etc. Use of these devices near this
equipment could cause this equipment to perform outside the
published specifications. Keep power to these devices turned off
when near this equipment.
This product consists of devices that may contain mercury, which
must be recycled or disposed of in accordance with local, state, or
country laws. (Within this system, the backlight lamps in the monitor
display, contain mercury.)
Rear of the system
Rear of the system
Rear of the system
UL conformity mark according to UL 60601-1 and CAN/CSA C22/2
NO. 601.1:.
UL conformity mark according to ANSI/AAMI ES60601-1:2005/
(R)2012, CAN/CSA-C22.2 No. 60601-1:14.
ISO 7010 - P007
Volume Navigation Pacemaker Warning.
No access for person with pacemaker.
How to lock Operator Panel prior to transportRear of the system.
DO NOT place a finger, hand or any object on the joint of the monitor
or monitor arm to avoid injury when moving the monitor and monitor
arm.
DO NOT push the system. Use the handle to push/pull the system,
e.g., DO NOT use the LCD. Failure to do so may cause serious injury
or system damage.
R1 to R3
Leg of the rear caster
R4 and later
Leg of the rear caster
V Nav Transmitter
Rear of the LCD
monitor.
Rear of the system
Section 1-3 - Important conventions1 - 13
Page 46
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Table 1-7 Label Icons(Continued)
Label/IconPurpose/MeaningLocation
CautionProbe connector
Neck of Monitor arm, on
Non-Ionizing Electromagnetic Radiation
equipped with Wireless
LAN
This system is for animal use only. Do not mix up with human
diagnosis.
Rear of the LS8 Vet
system
GOST Symbol. Russia Regulatory Country Clearance. Under the rating plate
EAC Mark of Conformity. Image of the Common Mark of Products
Circulation in the market of the Customs Un io n memb e r-states.
Every system has a unique marking for identification, the Unique Device
Identification (UDI) Label. The UDI label consists of a series of alpha-numeric
characters and barcode which uniquely identify the LOGIQ S8 system as a
medical device manufactured by General Electric. Scan or enter the UDI
information into the patient health record as required by country-specific laws.
Under the rating plate
Rating Plate
1 - 14Section 1-3 - Important conventions
Upgraded LOGIQ S8
systems, UDI label is on
the left-rear caster.
Probe UDI label
Page 47
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-4-3Human Safety
DANGER
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
•Operating personnel must not remove the system covers, except for removing front cover to clean
air filter.
•Servicing should be performed by authorized personnel only.
Only personnel who have participated in a LOGIQ S8 Training are authorize d to service the equipment.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT.
USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ S8, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUTDOWN MODE.
BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD,
PLACING PEOPLE IN AWKWARD POSITIONS, GE HAS LIMITED THE LIFTING
WEIGHT FOR ONE PERSON IN THE FIELD TO 16 KG (35 LBS). ANYTHING OVER 16
KG (35 LBS) REQUIRES 2 PEOPLE.
USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY.
WEAR ALL PPE INCLUDING GLOVES AS INDICATED IN THE CHEMICAL MSDS.
WARNINGWARNING
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. OPERATION
OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A
DEFINITE SAFETY HAZARD.
Section 1-4 - Safety Considerations1 - 15
Page 48
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.
ENSURE THAT THE SYSTEM IS TURNED OFF AND UNPLUGGED.
WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TEST POINTS TO VERIFY ISOLATION. THE BLUE LIGHT ON THE POWER BUTTON WILL TURN
OFF.
RISK OF ELECTRICAL SHOCK, SYSTEM MUST BE TURNED OFF AND
DISCONNECTED FROM POWER SOURCE. CORD MUST BE CONTROLLED AT ALL
TIMES.
USE EXTREME CAUTION AS LONG AS THE LOGIQ S8/LOGIQ E8 IS UN-STABLE,
NOT RESTING ON ALL FOUR CASTERS.
TILTING THE CONSOLE REQUIRES TWO PEOPLE IN ORDER TO AVOID INJURY TO
SERVICE PERSONNEL AND DAMAGE TO THE EQUIPMENT.
BEWARE OF POSSIBLE SHARP EDGES ON ALL MECHANICAL PARTS. IF SHARP
EDGES ARE ENCOUNTERED, THE APPROPRIATE PPE SHOULD BE USED TO
REDUCE THE RISK OF INJURY.
1 - 16Section 1-4 - Safety Considerations
Page 49
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-4-4Mechanical Safety
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
CAUTION
USE EXTREME CAUTION WHEN ELEVATING THE UNIT, OR IF IT IS RAISED FOR A
REPAIR OR MOVED ALONG ANY INCLINE. THE LOGIQ S8 SYSTEM MAY BECOME
UNSTABLE WHICH COULD CAUSE A TIP OVER.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT
CAN EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN
HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A
DAMAGED OR DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS
CAN RESULT IN SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR.
EVEN IF IT LOOKS OK, IT MAY BE DAMAGED.
WHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE
EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
The system weighs approximately 85 kg (LOGIQ S8) / 97 kg (LOGIQ E8) or more, depending
on installed peripherals, when ready for use.
Care must be used when moving it or replacing its parts. Failure to follow the precautions
listed could result in injury, uncontrolled motion and costly damage.
ALWAYS:
• Use the handle to move the system.
• Be sure the pathway is clear.
• Use slow, careful motions.
• Do not let the system strike walls or door frames.
Two people are required when moving on inclines or lifting more than 16 kg (35 lbs).
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
THE SYSTEM SHOULD NOT BE MOVED WITH THE OPERATOR I/O PANEL EXTENDED.
MOVE THE OPERATOR I/O PANEL TO ITS CENTERED AND LOCKED POSITION.
LOWER THE OPERATOR I/O PANEL AS MUCH AS POSSIBLE BEFORE MOVING THE
SYSTEM.
Before you move or transport the system, make sure to lock the LCD monitor firmly and flip
down the monitor to prevent damage to the system.
To avoid injury when you move the LCD monitor and the monitor arm, DO NOT put your
finger, hand, or object on the joint of the monitor or the monitor arm.
To avoid injury or damage to the monitor, make sure there is nothing within range of the LCD
before moving the monitor and monitor arm. This includes people as well as things.
Ensure that nobody touches the console arm/frogleg when moving the Operator Panel.
Section 1-4 - Safety Considerations1 - 17
Page 50
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
CAUTION
Use Protective Glasses during drilling, filing and during all other work where eyes
need protection.
CAUTION
CAUTION
NOTICE
Be careful not to pinch any of the cables.
Use Safety Shoes when doing work where there is any chance of foot damage.
Use Protective Gloves when drilling and cutting.
NOTE:Special care should be taken when transp orting the unit in a vehicle, se e 4-3-8-3 "Transporting
the System" on page 4-16.
1 - 18Section 1-4 - Safety Considerations
Page 51
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-4-5Electrical Safety
1-4-5-1Safe Practices
To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
The system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with th e system, make sure that the total
current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
WARNINGWARNING
CONNECTING A LOGIQ S8/LOGIQ E8 SCANNER TO THE WRONG VOLTAGE LEVEL
WILL MOST LIKELY DESTROY IT.
1-4-5-2Probes
All the probes for the LOGIQ S8 are designed and ma nufactured to provide trouble-free, reliable service.
To ensure this, correct handling of probes is important and the following points should be noted:
•Do not drop a probe or strike it agai nst a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
•Inspect the probe prior to each use for damage or degradation to the housing, cable strain relief,
lens, seal, connector pins and locking mechanism.
•Do not use a cracked or damaged probe. In this event, call your fiel d service representative
immediately to obtain a replacement.
•Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
•To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the machine’s wheels.
•Never immerse the probe connector or adapter into any liquid.
NOTE:For detailed information on handling probes, refer to the Basic User Manual and the care card supplied
with the probe.
Section 1-4 - Safety Considerations1 - 19
Page 52
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-4-6Auxiliary Devices Safety
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
Power Supplies for additional equipment MUST comply with IEC 60601-1.
DO NOT attempt to use different peripherals and accessories (brand and model;
connected via USB ports) other than approved and provided by GE Healthcare!
The ultrasound system is an extremely sensitive and complex medical system.
Any unauthorized peripherals may cause system failure or damage!
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1.
The LOGIQ S8 may be used with an isolation transformer to pr ovide the required separation from mains
for both, the system and the auxiliary devices.
One AUX main outlet is located at the primary power supply. It is used for connecting the two-fold splitter
whose outlets are led to the shelves intend for auxiliary devices (e.g., printer s) and the AUX main outlet
that is accessible on the back of the control console.
The IEC 60601-1-1 standard provides a guideline for safely interconnecting medical devices in systems.
“Equipment connected to the analog or digital interface must comply with the respective IEC/UL
standards (e.g. IEC 60950 / UL 60950 for data processing equipment and IEC 60601-1 / UL 60601-1
for medical equipment).
Everybody who connects additional equipment to the signal input portion or signal output portion
configures a medical system, and is therefore re sp on sib le th at the system complies with the
requirements of the system standard IEC 60601-1-1.
Special care has to be taken, if the device is connected to computer network (e.g., Ethernet), because
other devices could be connected without any control. There could be a potential difference between
the protective earth and any line of the computer network including the shield.
In this case the only way to operate the system safely is to use an isolated signal link with minimum
4mm creepage distance, 2.5mm air clearance of the isolation device. For computer networks there are
media converters available which convert the electrical to optical signals. Please consider that this
converter has to comply with IEC xxx standards* and is battery operated or connected to the isolation
mains output of the LOGIQ S8.
* IEC xxx stands for standards such as:
•IEC 60601 for medical devices
•IEC 60950 for information technology equipment etc.
CAUTION
1 - 20Section 1-4 - Safety Considerations
The leakage current of the entire system including any / all auxiliary equipment must not exceed
the limit values as per EN 60601-1-1:1990 (IEC 60601-1-1) respectively other valid national or
international standards. All equipment must comply with UL, CSA and IEC requirements.
Page 53
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
CAUTION
CAUTION
CAUTION
NOTICE
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
Line Printers are no medical devices, they have to be located outside of the patient environment
(according to IEC 60601-1 / UL 60601-1).
Auxiliary equipment must only be connected to the main console with the special main outlet
provided for the electrical safety of the system.
Auxiliary equipment with direct main connection requires galvanic separation of the signal and/
or control leads.
.
The system integrator (any person connecting the medical device to other devices) is responsible
that the connections are safe.
If in doubt, consult the technical service department or your local representative.
NOTICE
NOTICE
All peripherals mounted on the LOGIQ S8 system chassis must be firmly secured in position.
Each signal output (DVI, D-SUB (VGA), S-VIDEO, COMPOSITE (BNC), COMPOSITE (RCA)) of UVC
(Universal Video Converter) is insulated.
Section 1-4 - Safety Considerations1 - 21
Page 54
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-4-7Battery Safety
NOTE:The LOGIQ S8 ultrasound system is supplied with a lithium ion battery in the power supply
module, as option.
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such as watch batte ries) .
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:Regulations vary for different countries. Dispose of a used battery in accordance with local
regulations.
WARNINGWARNING
CAUTION
CAUTION
THE LOGIQ S8 BATTERY HAS A SAFETY DEVICE.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH THE
LOGIQ S8/LOGIQ E8 ULTRASOUND SYSTEM.
•DO NOT short-circuit the battery by directly connecting the negative terminals with metal objects.
•DO NOT heat the battery or discard it in a fire.
•DO NOT expose the battery to temperatures over 50 °C. Keep the battery away from fire and other
heat sources.
•DO NOT charge the battery near a heat source, such as, a fire or heater.
•DO NOT leave the battery in direct sunlight.
•DO NOT pierce the battery with a sharp object, hit it, or step on it.
•DO NOT use a damaged battery.
•DO NOT apply solder to a battery.
•DO NOT connect the battery to an electrical power outlet.
TO PREVENT THE BATTERY BURSTING, IGNITING, OR FUMES FROM THE BATTERY CAUSING
EQUIPMENT DAMAGE, ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
•DO NOT immerse the battery in water or allow it to get wet.
•DO NOT place the battery into a microwave oven or pressurized container.
•If the battery leaks or emits an odor, remove it from all possible flammable sources.
•If the battery emits an odor or heat, is deformed or discolored, or in a way appe ars abnormal during
use, recharging or storage, immediately remove it and stop using it.
NOTE:If you have any questions about the battery, consult your local GE representative.
•Storage of the battery pack:
Short-term (less than 1 month): between 0 °C (32 °F) and 50 °C (122 °F)
Long-term (more than 3 months): between 10 °C (50 °F) and 35 °C (95 °F)
When charging the battery for the first time after long-term storage, recover the battery pack to original
performance through repeating several cycles of full charging and discharging.
1 - 22Section 1-4 - Safety Considerations
Page 55
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-5Labels Locations
1-5-1The labels on the cover
The LOGIQ S8 comes equipped with product labels and icons.
These labels and icons represent pertinent information regarding the operation of the unit.
NOTE:The rating label is located on the rear of R1/R2/R3 system.
19” Monitor
LOGIQ S8/LOGIQ E8
LOGIQ S8 Vet
23”/22” Wide Monitor
Figure 1-1 Label location - R1/R2/R3 system
Section 1-5 - Labels Locations1 - 23
Page 56
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Figure 1-2 Label location - R4 and later system
1. LCD Caution label
2. LOGIQ S8 rating label
3. Gender Caution (only for India, China, Korea)
4. Multi Caution label
5. LOGIQ S8 Rating label (only for China, Korea, Japan)
6. UL label
NOTE:* Depending on country, “Gender Label” is attached.
1 - 24Section 1-5 - Labels Locations
Page 57
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Figure 1-4 WLAN Label location (R4 and later system)
Section 1-5 - Labels Locations1 - 25
Page 58
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1-5-2Rating plate
The following information prints on the rating plate.
1. Catalogue number
2. Manufacturing date
3. Manufacturer
4. System Serial Number
Figure 1-5 Rating plate and UL label - example
5. Power Supply Voltage
6. Power Input
7. Power Supply Frequency
8. Model Name
Rating plateUL label
NOTE:The rating label is located on the rear of R1/R2/R3 system.
Figure 1-6 Rating plate and UL label - R4 example
1 - 26Section 1-5 - Labels Locations
Page 59
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-6
Dangerous Procedure Warnings
Warnings, such as the examples below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGER
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE
PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN
HANDLING, TESTING AND ADJUSTING.
IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ S8/LOGIQ E8, SOME
METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD
IF TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. OPERATION
OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A
DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT.
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.
Section 1-6 - Dangerous Procedure Warnings1 - 27
Page 60
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-7
Lockout/Tagout (LOTO) Requirements
Follow OSHA Lockout/Tagout requirements (USA) or local Lockout/Tagout requirements by ensuring
you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Control all stored and residual energy.
6.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
NOTICE
Energy Control and Power Lockout for LOGIQ S8
When servicing parts of the system where there is exposure to voltage greater than 30 Volts:
1. Remove the front cover and confirm the power status LED. If ON, then turn off the system.
2. Unplug the system.
3. Maintain control of the system power plug.
4. Wait for at least 20 seconds for capacitors to discha rge as there are no test po ints to verify isolation.
The Power On/Off button LED turns off.
5. Confirm the Power Status LEDs are OFF.
- If the LEDs are turned ON, attempt to shut down the system again.
- If unable to turn off the system, press and hold the POWER on/off button to force shutdown .
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-8
Returning/Shipping System, Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEHC policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEHC employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
and ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE:The US Department of Transportation (DOT) has ruled that “items what were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purpose and must be
transported as a hazardous material.
When returning or shipping the LOGIQ S8 system in the original packaging:
1.) Adjust the LCD monitor and control panel using Up/Down control and Swivel control to their
centered and lowest positions.
2.) Flip down the LCD monitor and lock the monitor arm.
See 4-3-8 "Moving and Transporting the LOGIQ S8" on page 4-13 for more information.
NOTICE
Equipment being returned must be clea n and free of blood and other infectious substances.
1-8-1For LOGIQ S8 Vet system
Return used/unused spare parts from a veterinary environment with the purple recycling label
(regardless of its actual condition) and add a description on the label stating that the items were
removed from a LOGIQ S8 Vet in a veterinary environment.
This apply for Probes and covers labeled as Vet used.
If purple recycling label is not used in your region, use local recycling label.
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-9
Electromagnetic Compatibility (EMC) and
1-9-1What is EMC?
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as inte nded due interferen ce from its environment or
when the device produces unacceptable levels of emission to its environmen t. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to ele ctromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
For applicable standards please refer to Chapter 2 in the Basic User Manual of the each system.
1-9-2Compliance
The LOGIQ S8 conforms to all applicable conducted and radiated emission limits and to immunity from
electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient
requirements as mentioned in IEC 60601-1-2.
Electrostatic Discharge (ESD)
NOTE:For CE Compliance, it is crit ical that all covers, screws, shielding, gaskets, mesh, clamps, ar e in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-9-3Electrostatic Discharge (ESD) Prevention
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE POWER
CONNECTOR).
FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
DO NOT touch any boards with integrated circuits prior to taking the necessary ESD
precautions:
1.) When installing boards, ESD may cause damage to a board.
ALWAYS connect yourself, via an arm-wrist strap, to the advised ESD connection
point located on the rear of the system (to the right of the power connector).
2.) Follow general guidelines for handling of electrostatic sensitive equipment.
Risk of electrical shock: System must be turned off.
Avoid all contact with electrical contacts, conductors and components.
Always use non-conductive handles designed for the removal and rep lacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged
or isolated before making contact.
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 1-10
Customer Assistance
1-10-1Contact Information
If this equipment does not work as indicated in this ser vice manual or in the Basic User Manual, or if you
require additional assistance, please contact the local distributor or appropriate support resource, as
listed below.
Before you call, identify the following information, and acquire image (Alt+D) to send to the Customer
Care team:
1.) System ID serial number (also visible on label on back of the system).
2.) Software version.
3.) Date and time of occurrence.
4.) Sequence of events leading to issue.
5.) Is the issue repeatable?
6.) Imaging mode, probe, preset/application.
7.) Media brand, speed, capacity, type.
8.) Save secondary image capture, cine loop.
NOTE:Restart the application before resuming clinical scanning.
Table 1-8 Phone Numbers for Customer Assistance
LocationPhone Number
USA
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive (RP-2123)
Wauwatosa, WI 53226, USA
Canada1-800-668-0732
Latin America
Europe
GE Ultraschall Deutschland GmbH
Beethovenstraße 239
Postfach 11 05 60, D-42655 Solingen
Germany
EAGM
Asia (Singapore)
GE Ultrasound Asia Service Department -
Ultrasound
1 Maritime Square #13-01
HarbourFront Centre
Singapore 099253
Service On-site
Service: Parts
Applications support
Service
Applications support
Support Phone: +49 (0) 212-2802-652
Support Fax: +49 (0) 212-2802-431
Egypt Service Center
UAE Service Center
Tel: 800 1012882
Fax: +65 6291-7006
1-800–437–1171
1-800-558-2040
1-800-682-5327 or 1-262-524-5698
1-800-524-5300
1-262-524-5698
00202 2322 1252
00971 8003646
ANZ -Service Support1 800 647 855
Japan Support Center
Phone: 81-42-648-2940
Fax: 81-42-648-2905
Section 1-10 - Customer Assistance1 - 31
Page 64
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
The location of the CE marking is shown in the Safety chapter of this manual.
Authorized EU Representative European registered place of business:
GE Medical Systems Information Technologies GmbH (GEMS IT GmbH)
Munzinger Strasse 3, D-79111 Freiburg, GERMANY
+49 761 45 43 -0 /
+49 761 45 43 -233
1 - 32Section 1-10 - Customer Assistance
Page 65
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Chapter 2
Site preparations
Section 2-1
Overview
2-1-1Purpose of this chapter
This chapter provides the informatio n re qu ir ed to plan and prepare for the setup of a LOGIQ S8.
Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
The cooling requirement for the LOGIQ S8 is 680 BTU/hr. This figure does no t include coolin g needed
for lights, people, or other equipment in the room.
Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
o
C /50 - 95 oF-10 - 50 oC/14 - 122 oF-10 - 50 oC/14 - 122 oF
2-2-1-2Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimme rs can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2 - 2Section 2-2 - General console requirements
Page 67
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-2-2Electrical Requirements
2-2-2-1General requirements
NOTE:GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
system. This dedicated power shall originate at the last distribution panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the Ground from the main
facility's incoming power source to the Ultrasound system is only a conduit.
2-2-2-2Electrical requirements
In the table below, the electrical specifications for LOGIQ S8 include LCD and on board peripherals.
Table 2-2 Electrical Specifications
ModelVoltageTolerancesCurrentFrequency
LOGIQ S8
Power Consumption nominal 900 VA including all on-board peripherals.
2-2-2-3Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-4Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
CAUTION
!! CAUTION:
POWER OUTAGE MAY OCCUR.
The LOGIQ S8 requires a dedicated single branch circuit. To avoid circuit overload and possible
loss of critical care equipment, make sure you DO NOT have any other equipment operating on
the same circuit.
2-2-2-5Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cord s. Other adequ ate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical insta llatio n mus t meet all current local, state, and
national electrical codes.
The LOGIQ S8 is supplied with a main power plug, as standard.
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Section 2-2 - General console requirements2 - 3
Page 68
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-2-3EMI Limitations
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. Ultrasound systems also generate EMI. The LOGIQ
S8 complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a de fect. These
sources include:
•Medical lasers
•Scanners
•Cauterizing guns
•Computers
•Monitors
•Fans
•Gel warmers
•Microwave oven
•Light dimmers
•Mobile phones
•In-house wireless phones (DECT phones)
•Wireless computer keyboard and mouse
•Air conditioning system
•High frequency (HF) surgery equipmen t
•General AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.
2 - 4Section 2-2 - General console requirements
Page 69
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
See table below for EMI Prevention tips.
Table 2-3 EMI Prevention/Abatement
EMI RuleDetails
Keep the unit at least 5 meters (16.4 feet) away from other EMI sources. Special shielding may be
Be aware of RF sources.
required to eliminate interference problems caused by high frequency, high powered radio or video
broadcast signals.
Ground the unit.
Replace and/or
reassemble all screws,
RF gaskets, covers and
cores.
Replace broken RF
gaskets.
Do not place labels where
RF gaskets touch metal.
Use GE- specified
harnesses and
peripherals.
Take care with cellular
phones.
Properly dress peripheral
cables.
Poor grounding is the most likely reason a unit will have noisy images.
Check grounding of the power cord and power outlet.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF leakage.
In case a label has been found in such a location, move the label to a different, appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding.
Cable length, material, and routing are all important; do not make any changes that do not meet all
specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop the excess length for peripheral cables inside the peripheral bays.
Attach the monitor cables to the frame.
Section 2-2 - General console requirements2 - 5
Page 70
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-2-4Probe Environmental Requirements
Probes should be operated, stored, or transported within the parameters outlined below.
NOTE:See attached TEE probe user manual for the environmental requirements of 6Tc-RS probe.
NOTE:See attached FibroScan user manual for the environmental requirements of FibroScan probe.
WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT
IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
NOTE:Refer to Table 3-2 Acclim ate Time on page 3-4 to determine the needed settlement time.
2-2-5Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the LOGIQ S8.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury and/or damage to the system.
2 - 6Section 2-2 - General console requirements
Page 71
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 2-3
Facility needs
2-3-1Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay,
confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
Purchaser responsibility includes:
•Procuring the materials required.
•Completing the preparations before deliver y of th e ultr a so un d sys tem .
•Paying the costs for any alterations and modifications not specifically provided in the sales contract.
NOTE:All electrical installations that are preliminary to the positioning of the equipment at the site pr epared for
the equipment must be performed by licensed electr ical contractors. Other connections between p ieces
of electrical equipment, calibrations, and testing must also be performed by qualified personnel.
The products involved (and the accompanying electrical installations) are highly sophisticated and
special engineering competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified
personnel to perform electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved prod uct further
from the system than the interface kit allows pres ents challenges to the installation team. To avoid
delays during installation, such variances should be m ade known to the individuals or group performing
the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended
because it collects dust and creates static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact system reliability.
Section 2-3 - Facility needs2 - 7
Page 72
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-3-2Required Facility Needs
NOTE:GE Healthcare requires a dedicated power and gr ound for the proper operation of its Ultrasound
system. This dedicated power shall originate at the last distribution panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in Section 2-3-4 "Desirable features" .
•Dedicated single branch power outlet of adequate amperage (see Table 2-2 Electrical
Specifications on page 2-3) that meets all local and national codes and is located less than 2.5 m
(8.2 ft.) from the LOGIQ S8 proposed location.
•Door opening of at least 76 cm (2.5 ft.) in width.
•Proposed location for the LOGIQ S8 is at least 0.2 m (0.67 ft.) from the walls, to enable cooling.
•Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•Power outlets for other medical equipment.
•Clean and protected space for storage of probes (either in their case or on a rack).
•Material to safely clean probes (perform ed usin g a pla stic con tain e r, ne ve r me ta l).
NOTE:The LOGIQ S8 has one outlet for on board peripherals.
In case of network option:
•An active network outlet in the vicinity of the ultrasound unit.
•A network cable of appropriate length (regular Pin-to-Pin network cable).
•An IT administrator who will assist in configuring the unit to work with your local network.
A fixed IP address is required. Refer to the form provided in Figure 3-61 Worksheet for DICOM
Network Information on page 3-96 for network details that are required.
NOTE:All relevant preliminary network port installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
2-3-3Privacy and Security
Since the LOGIQ systems are integrated into your IT-network, GE wants to make sure that you are
aware of the proactive measures we are taking to secure the system.
For more information on privacy and security, refer to LOGIQ S8 Priva cy and Se curity manual in eDoc
CD.
2 - 8Section 2-3 - Facility needs
Page 73
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-3-4Desirable features
The followings are (optional) site recommendations.
•Door opening of at least 76 cm (2.5 ft.) in width.
•Accessible circuit breaker for a dedicated power outlet.
•Sink with hot and cold running water.
•Receptacle for bio–hazardous waste, for example, used probe sheaths.
•Emergency oxygen supply.
•Storage area for linens and equipment.
•Nearby waiting room, lavatory, and dressing room.
•Dual level lighting (bright and dim).
•Lockable cabinet for software and manuals.
2-3-5Minimal floor plan suggestions
Dedicated Power Outlets
LOGIQ S8
Cabinet for
Software and Manuals
(optional)
Figure 2-1 Minimal Floor Plan 4.3m x 5.2m (14ft x 17ft)
Hospital Network
Section 2-3 - Facility needs2 - 9
Page 74
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-3-6Networking setup requirements
For detail of Connectivity setup information, refer to the Basic User Manual supplied with the each
system.
•Ethernet network connection (R1.x.x/R2 and later)
•Up to 900 Mbps WLAN (R2 and later, optional)
•10/100/1000 Mbps Ethernet/DICOM network connection is required.
2-3-6-3InSite Requirements
Need internet access available to be able to connect to Insite.
Insite requires an Ethernet connection with a 10/100 Mbps or 10/100/1000 Mbps interface.
2-3-6-4Purpose of the DICOM network function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or tra nsferri ng
images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board LCD and peripherals,
enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-6-5DICOM Option Pre-installation Requirements
To configure the LOGIQ S8 to work with other network connections, the site’s network administrator
must provide information to complete the form in Figure 2-2 Worksheet for DICOM Networ k Information
on page 2-11. Ensure that there are no spaces in any field of the form.
Entries must include:
•A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ S8.
•The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
•The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ S8 for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the
revision of the device, is also included. This information may be useful for error solving.
2 - 10Section 2-3 - Facility needs
Page 75
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Figure 2-2 Worksheet for DICOM Network Information
Site System Information
Site:
Dept:
LOGIQ™ S8
SN:
CONTACT INFORMATION
Name
Type:
TCP/IP Settings
System IP Settings
Name - AE Title:
IP Address:
Subnet Mask:
Default Gateway:
Title
Floor:
Room:
REV:
Phone
Remote Archive Setup
Name - AE Title:
IP Address:
Subnet Mask:
Default Gateway:
Server Name:
Remote DB User Name:
Comments:
E-Mail Address
Services (Destination Devices)
Device Type
1
2
3
4
5
6
7
8
9
10
11
12
Manufacturer
Name
IP Address
Port
AE Title
Section 2-3 - Facility needs2 - 11
Page 76
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 2-4
Environmental Dangers
Commercial devices such as laser cameras, printers, VCRs and external monitors, usually exceed
allowable leakage current limits and, when plugge d in to sep arate AC ou tlet s, are in viola tion of pa tie nt
safety standards. Suitable electrical isolation of such external AC outlets, or providing the device with
extra protective earth, will be required in order to meet UL60601-1 and IEC60601-1 / IEC60601-1-1
standards for electrical leakage.
2-4-1Patient Vicinity UL60601-1 (USA)
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to
be contacted by the patient or attendant who can touch the patient. This encloses a space within the
room 1.83 m (6 ft.) beyond the perimeter of the bed (examination table, dental chair, treatment booth,
and the like) in its intended location, and extending vertically 2.29 m (7.5 ft.) above the floor.
2 - 12Section 2-4 - Environmental Dangers
Page 77
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2-4-2Patient Environment IEC60601-1 (IEC60601-1-1) and ANSI AAMI ES60601-1
Sub Clause 2.202 and figure 201 (IEC60601-1-1:2000)
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out. It is
very difficult to attach unique dimensions to the PATIENT ENVIROMENT. In practice, a distance of 2,5
m (8.2 ft.) above the floor on which the medical personnel stand and a horizontal distance of 1,5 m (4.9
ft.) have justified themselves as indicative of the dimensions of the Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure. See example below.
Section 2-4 - Environmental Dangers2 - 13
Page 78
GE HEALTHCARE
DIRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
This page was intentionally left blank.
2 - 14Section 2-4 - Environmental Dangers
Page 79
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Chapter 3
System Setup
Section 3-1
Overview
3-1-1Purpose of this chapter
This chapter contains information needed to set up the LOGIQ S8. Included is a procedure that
describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual setup, and how to check and test the unit, probes, and
external peripherals for electrical safety are included in this procedure.
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-2
Setup reminders
3-2-1Average Setup Time
Once the site has been prepared, the average installation time required is shown in table below.
Table 3-1 Average Installation Time
DescriptionAverage Setup TimeComments
Unpacking the scanner
Setup the scanner / options / printers
DICOM Option (connectivity)
Install Insite/iLink
0.5 hours
0.5 to 1.5 hoursDependent on the configuration
0.5 - 1.5 hoursDependent on the configuration
0.5 hours
3 - 2Section 3-2 - Setup reminders
Page 81
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-2-2Setup warnings
!! WARNING:
!! CAUTION:
!! CAUTION:
!! CAUTION:
DANGER
WARNINGWARNING
WARNINGWARNING
CAUTION
CAUTION
CAUTION
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T
TOUCH THE UNIT!
DO NOT PULL OR LIFT THE SYSTEM WITH THE FRONT HANDLE OF THE USER
INTERFACE (OPERATOR PANEL).
THE SYSTEM SHOULD NOT BE MOVED WITH THE OPERATOR CONTROL PANEL
EXTENDED. MOVE THE OPERATOR CONTROL PANEL TO IT’S CENTERED AND
LOCKED POSITION. REFER TO Section 8-2-3 "Returning/Shipping Probes and
Repair Parts" on page 8-4.
TO PREVENT ELECTRICAL SHOCK, CONNECT THE UNIT TO A PROPERLY GROUNDED
POWER OUTLET. DO NOT USE A THREE TO TWO PRONG ADAPTER. THIS DEFEATS
SAFETY GROUNDING.
DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS AND MORE
THAN 30 V PEAK IS PRESENT.
DO NOT OPERATE THIS ULTRASOUND SYSTEM UNLESS ALL BOARD COVERS AND
FRAME PANELS ARE SECURELY IN PLACE. SYSTEM PERFORMANCE AND COOLING
REQUIRE THIS.
(When covers are removed, EMI may be present).
CAUTION
CAUTION
OPERATOR MANUAL(S)
THE USER MANUAL(S) SHOULD BE FULLY READ AND UNDERSTOOD BEFORE
OPERATING THE LOGIQ S8/LOGIQ E8 AND KEPT NEAR THE UNIT FOR QUICK
REFERENCE.
ACOUSTIC OUTPUT HAZARD
ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM THE LOGIQ S8 PROBE IS
WITHIN AIUM/NEMA STANDARDS, AVOID UNNECESSARY EXPOSURE. ULTRASOUND
ENERGY CAN PRODUCE HEAT AND MECHANICAL DAMAGE.
•The system weighs approximately 85 kg (LOGIQ S8) / 97 kg (LOGIQ E8) without peripherals, two
people are required to unpack it.
•There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
Section 3-2 - Setup reminders3 - 3
Page 82
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-2-2-1If the unit is very cold or hot
CAUTION
!! CAUTION:
Turning the system ON after arrival at the site - without allowing time for acclimation - may
cause system damage!
Following transport, the system may be very cold, or ho t. Allow time for the system to acclimate before
o
being switched ON. Acclimation requires 1 hour for each 2.5
system is below 10
NOTE:After a long period of storage, or after transportation of the system with the monitor in the fold ed-
down position, it is highly recommended to place the monitor in the upright position - and to
leave it in this position for a period of longer than 1 hour before use. This will enable it to properly
adjust to the environmental conditions.
3 - 4Section 3-2 - Setup reminders
Page 83
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-3
Receiving and unpacking the equipment
3-3-1Receiving and unpacking warnings
WARNINGWARNING
CAUTION
CAUTION
DO NOT PULL OR LIFT THE SYSTEM WITH THE FRONT HANDLE OF THE USER
INTERFACE (OPERATOR PANEL).
Two people are needed to unpack the unit because of its weight. Attempts to move the unit
considerable distances or on an incline by one person could result in injury or damage or
both.
Two people are required whenever a part weighing 16 kg (35 lbs.) or more must be lifted.
Remember to use relevant personal protecting equipment (pre) during packing/unpacking.
Check with your local EHS representative.
Section 3-3 - Receiving and unpacking the equipment3 - 5
Page 84
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-3-2Receiving the LOGIQ S8
3-3-2-1Overview
Improper handling during transportation may harm the equipment inside the package even if the
package itself is undamaged.
3-3-2-2Examine package
Examine package closely at time of delivery.
Table 3-3 Examine all packages
STEPTASK
1
Is damage apparent?
If YES, continue with the instruction in Section 3-3-2-3 "Damage in
transportation" on page 3-6.
If NO, continue with the next step.
2
Before cutting the straps, check the tilt indicator.
Is the tilt Indicator red colored (1) inside the middle of the indicator?
•If YES: The tilt Indicator has been activated. Continue with the
instructions in Section 3-3-2-3 "Damage in transportation" on page 3-6.
•If NO: continue with the instruction in Section 3-3-4 "Unpacking the
LOGIQ S8" on page 3-9.
3-3-2-3Damage in transportation
Follow this procedure if damage is apparent:
Table 3-4 Damage in transportation
STEPTASK
1
Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted
or “signed for“ by a GE representative or hospital receiving agent.
2
Report the damage to the carrier.
•Whether noted or concealed, damage MUST be reported to the carrier immediately upon
discovery, or in any event, within 14 days after receipt, and the contents and containers held for
inspection by the carrier.
•A transportation company will not pay a claim for damage if an inspection is not requested within
this 14 day period.
3 - 6Section 3-3 - Receiving and unpacking the equipment
Page 85
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-3-2-4The Tilt & Shock indicators (if applied)
NOTE:Tilt and Shock Indicators are removed in later product shipments.
3-3-2-4-1Overview
Improper handling during transportation may harm the equipment inside the package even if the
package itself is undamaged.
To make it easier to detect if the handling during transportation has been improper, a set of Tilt & Shock
indicators have been attached to the transportation box.
3-3-2-4-2Position of the Tilt & Shock indicators
Inspect the Drop and Tilt Indicator for evidence of accidental shock or tilting during transit (damage
incident, as illustrated in the figure below.
Figure 3-1 Tilt & Shock indicators
NOTE:Before cutting the straps, check the Shock and Tilt Tags to make sure they have not been triggered.
If damaged, report it to the carrier. If not, then cut the straps around the crate.
Section 3-3 - Receiving and unpacking the equipment3 - 7
Page 86
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-3-3Transportation Box Label
Figure 3-2 Label 1
Figure 3-3 Label 2
Figure 3-4 Label 3
3 - 8Section 3-3 - Receiving and unpacking the equipment
Page 87
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-3-4Unpacking the LOGIQ S8
The instruction manual describes the best method for unpacking the LOGIQ S8 ultrasound scanning
unit. Images are ONLY for reference; wear proper PPE when handling packaging (gloves, safety shoes,
etc....).
Table 3-5 Unpacking Procedure
StepTask
1.Cut the straps around the crate.
2.Remove the top cover.
3.Remove three P-Joint, then remove the out sleeve.
4.Remove the option box.
5.Carefully remove foam packing material and plastic bag
from the ultrasound unit and monitor
6.Remove the box bottom assy..
Section 3-3 - Receiving and unpacking the equipment3 - 9
Page 88
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Table 3-5 Unpacking Procedure
StepTask
7.Unlock the casters and align the orientation of the 4 caster's
direction (refer to 4-3-7 "Caster" on page 4-11) and then
using the incline plane of the box bottom assy., slowly move
the system from the box.
Note: Place all of the filling inside the Transportation Box. Close the box, and store the filling for possible future use.
Section 3-4
Packing materials - recycling information
The packing materials for LOGIQ S8 are recyclable:
•The Transportation Box is made of cardboard.
•The inner reinforcements are made of Ethafoam (Polyethylene foam).
•The plastic foil is made of LDPE (Low Density Polyethylene).
3 - 10Section 3-4 - Packing materials - recycling information
Page 89
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-5
Preparing for setup
3-5-1Verify Customer Order
Compare all items listed on the packing slip (delivery note) with those received.
Report any items that are missing, back ordered, or damaged.
The envelope with delivery address, packing list and invoice is located on the
front panel of the crate.
Check whether delivery is complete (according to packing list) and check visual
damage!
Figure 3-5 envelope at front panel of the crate
3-5-2Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to 3-3-2-3 "Damage in transport ation" on pa ge 3-6 in the
beginning of this manual.
3-5-3EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Ma ny of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secu red
before the unit is put into operation.
Refer to Section 2-2-3 "EMI Limitations" on page 2-4.
Section 3-5 - Preparing for setup3 - 11
Page 90
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-6
Completing the Set Up
3-6-1System Specifications
3-6-1-1System requirements verification
•Verify that the site meets the requirements listed in (see: Section 2-3 "Facility needs" on p age 2-7).
•Verify that the specifications below don’t conflict with any on-site conditions.
3-6-1-2Physical Dimensions
Table 3-6 LOGIQ S8 R1 to R3 (without Peripherals)
HeightWidthDepthWeight
1720 mm, 67.7 in (max)
1150 mm, 45.3 in (min) *
* with low cabinet option.
Table 3-7 LOGIQ S8 R4 and later (without Peripherals)
500 mm, 19.7 in (Keyboard)
620 mm, 24.4 in (Caster)
865 mm, 34.1 in (max)
795 mm, 31.3 in (Caster)
85 kg
(187.4 lbs.)
HeightWidthDepthWeight
1760 mm, 69.3 in (max)
1150 mm, 45.3in (min, 23” LCD)
1100 mm, 43.3in (min, 22” OLED)
Table 3-8 LOGIQ E8 R2 and R3 (without Peripherals)
HeightWidthDepthWeight
1750 mm (max)
1250 mm (min)
Table 3-9 LOGIQ E8 R4 and later (without Peripherals)
HeightWidthDepthWeight
1760 mm (max)
1240 mm (min)
3-6-1-3Acoustic Noise Output
max. 50 dB(A)
520 mm, 20.5 in (Keyboard)
530 mm, 20.9 in (Caster)
561 mm, 22.1 in (23” LCD Monitor)
527mm, 20.7in (22” OLED Monitor)
500 mm (Keyboard)
620 mm (Caster)
520 mm (Keyboard)
530 mm (Caster)
527mm (22” OLED Monitor)
865 mm, 34.1 in (max)
795 mm, 31.3 in (Caster)85 kg
(187.4 lbs.)
865 mm (max)
795 mm (Caster)
865 mm (max)
795 mm (Caster)
97 kg
97 kg
3 - 12Section 3-6 - Completing the Set Up
Page 91
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-6-2Electrical specifications
WARNINGWARNING
CONNECTING A LOGIQ S8 TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY THE UNIT.
3-6-2-1Verification of the LOGIQ S8’s voltage setting
Verify that the mains voltage specified for the unit is available on-site.
The voltage setting for the unit is found on a label on the back of the system on lower r ear frame of the
LOGIQ S8.
3-6-2-2Electrical Specifications
In the table below, the electrical specifications for LOGIQ S8 include LCD and on board peripherals.
Table 3-10 Electrical Specifications
ModelVoltageTolerancesCurrentFrequency
LOGIQ S8
Power Consumption nominal 900 VA including all on-board peripherals.
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-6-3Connection the I/O Rear Panel
Accessory equipment connected to the analog an d d igital interfaces must be certified according to the
respective IEC standards (i.e., IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everyone who connects add itional equipment to the signal in put part or signal
output part of the LOGIQ S8 system configures a medical system, and is therefore resp onsible to ensure
that the system complies with the requirements of the valid version of IEC60601-1-1. If in doubt, consu lt
the technical service department or your local GE Healthcare representative.
Figure 3-6 External I/O Connectors - on Rear of System
Table 3-11 External I/O Connector - Description
ItemConnector NameSection numberDescription
USB Port
1
2
HDMI Connector
3
Ethernet
4
AUDIO
5-2-3-2
5-2-4-1
5-2-4-2
5-2-4-3
R1/R2: USB2.0 Full Speed (Isolated)
R3/R4: USB2.0 Full Speed (No isolated)
Connector for external Monitor.
LAN for InSite Connection (RJ45).
Connector for External Speaker .
Audio line out (3.5mm pin jack)
3 - 14Section 3-6 - Completing the Set Up
Page 93
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-6-4Connecting Probes
3-6-4-1Introduction
Probes can be connected at any time, regardless of whether the unit is on or off. To ensure that the
ports are not active, place the system in the image freeze condition
.
LOGIQ S8/LOGIQ S8 Vet
LOGIQ E8
a. Active probe port
b. Parking probe port
a
c. Pencil probe port
b
Figure 3-7 Probe port
3-6-4-2Cable Handling
Take the following precautions with probe cable s:
•Keep free from wheels
•Do not bend the cable acutely
•Avoid crossing cables between probes.
3-6-4-3Connect a probe
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
DO NOT use a probe which appears damaged until functional and safe performance is
verified.
Fault conditions can result in electric shock hazard. Do not touch the surface of probe
connectors which are exposed when the probe is removed.
DO NOT allow the probe head to hang free. Impact to the probe head could result in
irreparable damage.
1.)Before connecting the probe:
a.) Do a visual check of the probe pins and system sockets.
b.) Remove any dust or foam rests from the probe pins.
c.) Inspect the probe for damage or degradation to the housing, strain relief, lens, seal, cable and
connector.
2.)Put the probe in the pr obe holder. Use the integrated cable management hook to wrap the cord.
3.)Ensure that the probe locking handle is at horizontal position.
4.)Hold the probe connector vertically with the cable pointing upward.
5.)Align the connector with the probe port and carefully push into place.
6.)Turn the connector locking handle clockwise (to vertical position) to secure the probe connector.
7.)Carefully position the probe cable so it is free to move and is not resting on the floor.
Section 3-6 - Completing the Set Up3 - 15
Page 94
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-6-4-4Connecting the CW Pencil Probe (LOGIQ S8/LOGIQ S8 Vet)
Insert the probe connector into the pencil probe port all the way seated in. Carefully position the probe
cable so it is free to move and is not resting on the floor.
3-6-4-5Connecting the TEE Probe
Type of this connector is “RS”, connect to the system by TEE probe adapter.
TEE RS-DLP probe adapter has 2 connection ports. On e is “RS” type for TEE pr obe, the other is “DLP”
type for LOGIQ S8 system.
DLP Connector lock knob
RS Connector
(Probe side)
DLP Connector
(System side)
RS Connector lock knob
Figure 3-8 DLP-RS Connector
1.)Hold the probe connector and TEE DLP-RS adapter with the cable pointing DLP connector lock
knob.
2.)Turn the RS connector lock knob to release side.
Release
RS Connector
lock knob
Lock
Lock
DLP Connector
lock knob
Release
Figure 3-9 Lock/Release position
3.)Align the connector with the RS port and carefully push into place.
4.)Turn the RS connector lock knob to the lock side.
5.)Hold the adapter + probe connector vertically with the cable pointing upward.
6.)Turn the DLP connector lock knob to lock side to secure the probe connecto r.
7.)Carefully position the probe cord so it is free to move and is not resting on the floor.
3 - 16Section 3-6 - Completing the Set Up
Page 95
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-7
Power On/Off
DANGER
ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE
PROTECTIVE GROUNDING CONNECTOR.
DANGER
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE
SAFETY GROUND.
DANGER
ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION
NOTICE
NOTICE
NOTE:If the battery option is installed (R4 or later)
LOGIQ S8 REQUIRES ALL COVERS.
OPERATE THIS UNIT ONLY WHEN ALL BOARD COVERS AND FLAME PANELS ARE SECURELY
IN PLACE. THE COVERS ARE REQUIRED FOR SAFE OPERATION, GOOD SYSTEM
PERFORMANCE AND COOLING PURPOSES.
User only power supply cords, cables and plugs provided by or designated by GE Healthcare.
After turning off the system, wait at least 10 seconds before turning it on again. The system may not
be able to boot if power is recycled too quickly.
The extended battery/small battery is charged automatically when the LOGIQ S8 is connected to the
wall outlet via the power cable and circuit breaker i s turned on wh ether the LOGIQ S8 is turned on or in
standby-mode.
But you can turn off the circuit breaker and disconnect the power cable when the battery is charging.
•While the extended battery/small battery charging, a fan is running. a fan stops automatically when
battery is fully charged.
•While the extended battery charging, the LED is blinking and turn off when the battery is fully
charged.
Section 3-7 - Power On/Off3 - 17
Page 96
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-7-1Connect AC (mains) Power to the LOGIQ S8
Connecting AC Power to the LOGIQ S8 involves preliminary checks of the power cord, voltage level
and compliance with electrical safety requirements.
WARNINGWARNING
CAUTION
POWER OUTAGE MAY OCCUR. THE ULTRASOUND UNIT REQUIRES A DEDICATED
SINGLE BRANCH CIRCUIT. TO AVOID CIRCUIT OVERLOAD AND POSSIBLE LOSS OF
CRITICAL CARE EQUIPMENT, MAKE SURE YOU DO NOT HAVE OTHER EQUIPMENT
OPERATING ON THE SAME CIRCUIT.
Disconnect the plug from the wall outlet in case an emergency should occur. Ensure easy
access to the power outlet.
1.) Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
2.) Uncoil the power cable, allowing sufficient slack so that the unit can be moved slightly.
3.) Verify that the power cable is without any visible scratches or any sign of damage.
4.) Verify that the on-site mains voltage is within the limits indicated on the rating label on the right side
of the rear panel.
5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the unit.
6.) Lock the plug in position with the Retaining Clamp.
7.) Verify that the Mains Power Circuit Breaker is in OFF position, if not, switch it OFF.
Circuit Breaker
Screw
Pull Out Protection
(Metal Bracket)
Mains Power Cable
Screw
Figure 3-10 Circuit breaker and main power cable at rear of system
8.) Connect the Power Cable’s other end (male plug) to a hospital grade mains power outlet with the
proper rated voltage, and the unit is ready for Power ON/Boot Up.
NOTE:The power for on-board peripheral auxiliary equipment are commonly switched with the power button.
The power switch of any attached printer(s) needs to be in ON position before starting the system.
However, be aware some auxiliary equipment may switch itself to standby mode (e.g., Color video
printer) and must therefore be switched on separately.
3 - 18Section 3-7 - Power On/Off
Page 97
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-7-2Power ON
NOTE:The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the ON/OFF standby button. So the auxiliary equipment need not to be switched ON/OFF
separately.
1.) Switch ON the Mains Power Circuit Breaker at the rear of the system. The power button turns on a
light as blue.
2.) Press once the power button on the Operator Contro l Panel to boot the system . Th e powe r b utton
turns on a light as green.
Off/Stand-by
Figure 3-11 Power Button
3.) The System (including the Back-end Processor) starts and the operating system is loaded which
then leads the application software to activate the scanner.
During a normal boot, you may observe
-Power is distributed to Peripherals, Operator Panel (control panel), Monitor, Front-End and
Back-End Processor.
-The Back-End Processor and rest of the scanner starts with the sequence listed in following
steps:
a.) “Boot Screen” is displayed.
b.) Back-End Processor is turned ON and starts to load the software.
c.) The Start Screen is displayed on the monitor.
d.) Start-up progress bars indicating software loading procedures, are displayed on the
monitor.
e.) The software initiates and sets up the Front-End electronics and the rest of the scanner
(incl. the clicking sound of the relays on the PID board).
f.) The Keyboard backlight is lit.
4.) As soon as the software has been loaded, the system enters B-Mode with the probe connected to
port nearest to a user. Total time used for start-up is about 2 minutes. The power button on the
Control Panel and indicator color.
ON (Power Assistant mode)
Section 3-7 - Power On/Off3 - 19
Page 98
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-7-3Login
At login, you are notified that “You are accessing a diagnostic medical device that is provided by
authorized usage only. Data stored on this device may be subject to various regulations including but
not limited to regulations which govern disclosure and privacy of this data. By using this device you are
acknowledging that you are authorized to do so and are trained in appropriate use and regulatory
guidelines.”
NOTICE
The System Administrator manages system groups, users, and permissions. After you have been
added as a valid user, the System Administrator assigns you a temporary password . When you log into
the system for the first time, you will be prompted to change your password.
Figure 3-12 Login Window (Example of R4.2.5x)
After you have established your password, follow these steps to login.
1.) Type your user name in the Operator field.
2.) Type your password in the Password field.
3.) Press OK to login, or Cancel to cancel login.
3 - 20Section 3-7 - Power On/Off
Page 99
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-7-4Logoff/Shutdown/Exit
1.) When you lightly press the Power On/Off switch once. The System-Exit window is displayed.
System Exit Window - R1 System Exit Window - R2 and later
2.) The SYSTEM - EXIT menu, used when power off the unit, gives you these choices:
•Logoff
Use this button to log off the current user. The system remains ON and ready fo r a new user to log
on. If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.
•Shutdown
Use this button to shut down the system. The entire system will shut down. The power button turns
on blue indicated Standby- mode (refer to Section Figure 3-11 "Power Button" on page 3-19).
For the system with R4.2.5x and later, press Power On/Off button three (3) times to shutdown.
NOTE:If the system has Small/Extended battery (Option), the system automatically starts battery-charging
when goes to Standby-mode.
•Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
•Exit (Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop. Refer to Section Section 7-3 "Exit
to Windows Desktop" on page 7-5.
•Sleep (R1 to R4)
Select this button to go to SLEEP mode. Refer to Section 3-7-6 "Sleep Mode (R1 to R4)" on page
3-23.
Section 3-7 - Power On/Off3 - 21
Page 100
GE HEALTHCAREPROPRIETARY TO GE
D
IRECTION 5394227, 12LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3-7-5Complete shutdown
NOTICE
!! NOTICE:
After turning off the system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
NOTE:The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the power
button. So the auxiliary equipment need not to be switched ON/OFF separately.
1.) Enter the scan screen and lightly press the Power On/Off switch once. The System-Exit window is
displayed.
2.) Select Shutdown. The system performs an automatic full shutdown sequence.
NOTICE
The
NOTICE
The system will go to Power Assistant Mode (Optional, R2 or later) if above procedures are skipped.
Be sure to wait with the next step until the system has finished its shut-down. Failing to do so may
destroy data on the Hard Disk Drive, making the system fail later.
3.) Switch OFF the Circuit breaker on the rear panel of the system.
4.) After complete power down, disconnect the main po wer cable from the system or unplug it from the
AC wall outlet socket.
WARNINGWARNING
Confirmation of LED status is necessary when repairing the system.
Confirm the Power Status LEDs are OFF.
- If the LEDs are turned ON, attempt to shut down the system again.
- If unable to turn off the system, press and hold the Power on/off button to force shutdown.
NOTE: For R4 or later system with Extended Battery option, the extended battery is charged when the
LOGIQ S8 in Standby-mode. While battery charging, LED is blinking and main fan is operating. But you
can turn off the circuit breaker and disconnect the power cable when the battery is charging.
3 - 22Section 3-7 - Power On/Off
Power Status LED
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.