State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 2 of 252
Page 3
This page was intentionally left blank.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 3 of 252
Page 4
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
Important Precautions
• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING
AVERTISSEMENT
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE PRESTATAIRE DE SERVICES DU CLIENT A BESOIN DE CE MANUEL DANS
UNE AUTRE LANGUE QUE L’ANGLAIS, IL INCOMBE AU CLIENT DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL DE MAINTENANCE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST
ES AUFGABE DES KUNDEN, FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN
ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND
NACHVOLLZIEHEN KÖNNEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 4 of 252
i
Page 5
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
• ESTE MANUAL DE SERVICIO SÓLO ESTÁ DISPONIBLE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL
CLIENTE.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO SIN HABER
AVISO
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL USUARIO O EL PACIENTE SUFRAN LESIONES
PROVOCADAS POR DESCARGAS ELÉCTRICAS, PROBLEMAS MECÁNICOS O
PELIGROS DE OTRA NATURALEZA.
• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
ATENÇÃO
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE PÔR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
AVVERTENZA
HOIATUS
• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
ii -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 5 of 252
Page 6
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
• TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
VAROITUS
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
• ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ ∆ΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ.
• ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙ∆ΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ
ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ
ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ∆Ε ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝ∆ΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ.
FIGYELMEZTETÉS
VIÐVÖRUN
• EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ
EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ
ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 6 of 252
iii
Page 7
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
• TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
• V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V
JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
VÝSTRAHA
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
• V PŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
• DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
ADVARSEL
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN,
OPERATØREN ELLER PATIENTEN.
WAARSCHUWING
BRĪDINĀJUMS
• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS
VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
• ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
• ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀ
S STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
iv -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 7 of 252
Page 8
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
• ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
• NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
ĮSPĖJIMAS
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
ADVARSEL
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE
ATE NŢIE
• NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
• JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ
NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOżE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEŃ.
• ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 8 of 252
v
Page 9
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
• ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯТОЛЬКОНААНГЛИЙСКОМЯЗЫКЕ.
• ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО
НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
ОСТОРОЖНО!
• ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ
К ДАННОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ УКАЗАННЫХ ТРЕБОВАНИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ,
ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ
, ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
• TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
UPOZORNENIE
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
VAR NING
DİKKAT
• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
• BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAK
İ BİR DİLDE
OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN
SORUMLULUĞUNDADIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
vi -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 9 of 252
Page 10
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
vii
Approved Document - 5213326-100TPH_r3.pdf Page 10 of 252
Page 11
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to position ing of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Othe r conn ections between pieces o f
electrical equipment, calibrations and testing sha ll be performed by q ualified GE Healthcare person nel.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details. Mail the information to:
Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226, USA.
GE Healthcare employees should use the iTrak System to report all documentation errors or omissions.
SERVICE SAFETY CONSIDERATIONS
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNINGWARNING
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section of
the LOGIQ™ A3 Basic Service Manual (5213326-100).
viii -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 11 of 252
Page 12
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3 LOGIQ™ A3 SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 29 of 252
Page 30
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Chapter 1Introduction
Section 1-1Overview
1-1-1Purpose
This chapter describes important issues related to safely servicing the L OGIQ A3 scanner. The service
provider must read and understand all the information presented in this manual before installing or
servicing a unit.
1-1-2Table Of Contents
Table 1-1 Contents in Chapter 1
S.NoDescriptionPage Number
1
2
3
4
5
6
Overview
Safety
Important Conventions
Safety Considerations
EMC, EMI, and ESD
Customer Assistance
1-1-3Purpose of Service Manual
This Service Manual provides installation and service information for the LOGIQ A3 Ultrasound
scanning system and contains the following chapters:
1.) Chapter 1 - Introduction: Con tains a content summary and warnings.
2.) Chapter 2 - Pre-Installation: Contains pre-installation requirements for the LOGIQ A3.
4.) Chapter 4 - Functional Checks: Contains functional checks that must be performed as part of the
installation, or as required during servicing and periodic maintenance.
5.) Chapter 5 - Components and Functions: Contains block diagrams and functional explanations
of the electronics.
6.) Chapter 6 - Service Adjustments: Contains instructions on how to make availa ble adjustments to
the LOGIQ A3.
7.) Chapter 7 - Diagnostics/Troubleshooting: Provides procedure s for running diagnostic or related
routines for the LOGIQ A3.
8.) Chapter 8 - Replacement Procedures: Provides disassembly procedures and reassembly
procedures for all changeable Field Replaceable Units (FRU).
9.) Chapter 9 - Renewal Parts: Contains a complete list of replacement parts for the LO G IQ A3.
10.)Chapter 10 - Care & Maintenance:Provides periodic maintenance procedures for the
LOGIQ A3.
1-1
1-2
1-3
1-7
1-17
1-18
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 30 of 252
Chapter 1 1-1
Page 31
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
•Contractors (Some parts of Chapter - Pre-Installation)
1-1-5Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before op erating the LOGIQ A3 and also
kept near the unit for quick reference.
Section 1-2Safety
1-2-1Warnings
WARNINGWARNING
NOTICE
CAUTION
CAREFULLY READ ALL OF THE WARNINGS BELOW
1.) The operator manual should be fully read an d understood before operating the LOGIQ A3 an d kept
nearby for quick reference.
2.) Although the ultrasound energy tran sm itted from the LOGIQ A3 transducer is within AIUM/NEMA
standards, unnecessary exposure should be avoided. Only trained personnel should operate the
LOGIQ A3.
3.) To prevent electrical shock, the LOGI Q A3 sho uld be conn ec te d to a pr op e rly gr ou n de d powe r
receptacle. Do not use a three prong to two prong adapter. This defeats safety grounding.
4.) Probes are fragile, please handle with care.
5.) Concerning Outside Markings, refer to Figure 1-1 through 1-3.
6.) For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ A3 User Manual
and Caution Sheet supplied with each probe.
This medical equipment is approved, in terms of the prevention of radio wave interference, to be used
in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment
in an inappropriate environment may cause some electronic interference to radios and televisions
around the equipment. Proper handling of this equipment is required in order to avoid such trouble
according to the operator and service manuals.This equipment can be used in residential areas only
under the supervision of physicians or qualified technicians.
Improper performance possibility. Do not use the following devices near this equipment.
Cellular phone, radio transceiver, mobile radio transmitter, radio-controlled toy, etc. Use of
these devices near this equipment could cause this equipment to perform outside the published
specifications. Keep power to these devices turned off when near this equipment
1-2Section 1-2 - Safety
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 31 of 252
Page 32
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 1-3Important Conventions
1-3-1Conventions Used in Book
Icons
Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identi fied by a flag word that precedes the precautionary
message. Known or potential hazards are labeled in one of following ways:
DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL
DANGER
WARNING
CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE
IGNORED.
WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE
SEVERE PERSONAL INJURY OR PROPERTY DAMAGE IF INSTRUCTIONS ARE
IGNORED.
Caution is used to indicate the presence of a hazard that will or can cause minor personal injury
CAUTION
NOTICE
NOTE:Notes provide important information about an item or a procedure.
and property damage if instructions are ignored.
Equipment Damage Possible:
Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
Example: Disk Drive will crash.
Information contained in a NOTE can often save your time or effort.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 32 of 252
Chapter 1 1-3
Page 33
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-3-2Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within a triangle,
as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be
used to make you aware of specific types of hazards that could cause harm.
Table 1-2 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEAT
LASER
LIGHT
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-3 Standard Icons Indicating a Special Procedure Be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
1-4Section 1-3 - Important Conventions
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 33 of 252
Page 34
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-3-3Product Icons
The following table describes the purpose and locatio n of safety labe ls and other importan t information
provided on the equipment.
Table 1-4 Product Icons
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Manufacturer's name and address
Identification and Rating Plate
Date of manufacture
Model and serial numbers
Electrical ratings
Rear of console near power inlet
On each probe
Device Listing/Certification Labels
Type/Class Label
IP Code (IP68)
“CAUTION
This unit weighs...
Special care must be used to avoid..."
Laboratory logo or labels denoting conformance
with industry safety standards such as UL or IEC.
Used to indicate the degree of safety or protection
Indicates the degree of protection provided by the
enclosure per IEC 60529. IP68 indicates can be
used in operating room environment.
Equipment Type BF (man in the box symbol) IEC
60878 indicates B Type equipment having a floating
applied part
This precaution is intended to prevent injury that
may result if one person attempt to move the unit
considerable distances or on an incline due to the
weight of the unit.
"CAUTION" The equilateral triangle is usually used
in combination with other symbols to advise or warn
the user.
Rear of console
Foot switch
Probe connectors and PCG Connectors
On the console where easily seen during
transport
Various
ATTENTION - Consult accompanying documents "
is intended to alert the user to refer to the operator
manual or other instructions when complete
information cannot be provided on the label.
"CAUTION - Dangerous voltage" (the lightning flash
with arrowhead in equilateral triangle) is used to
indicate electric shock hazards.
"Mains OFF" Indicates the power off position of the
mains power switch.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 1 1-5
Various
Various
Rear of system adjacent to mains switch
Approved Document - 5213326-100TPH_r3.pdf Page 34 of 252
Page 35
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Table 1-4 Product Icons (Continued)
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
"OFF/Standby" Indicates the power off/standby
position of the power switch.
CAUTION
This Power Switch DOES NOT ISOLATE Mains
Supply
"Mains ON" Indicates the power on position of the
mains power switch.
CAUTION
This Power Switch DOES NOT ISOLATE Mains
Supply
Adjacent to On-Off/Standby Switch
Rear of system adjacent to mains switch
~
"Protective Earth" Indicates the protective earth
(grounding) terminal.
Alternating Current
"Non-Ionizing Radiation" indicates that the system
applies RF energy.
Do not use the following devices near this
equipment, Cellular phones, radio transceivers,
mobile radio transmitters, radio controlled toy etc.
Use of these devices could cause this equipment to
perform outside the published specifications.
Keep power to these devices turned off when near
this equipment.
This Symbol indicates that waste electrical and
electronic equipment must not be disposed of as
unsorted municipal waste and must be collected
separately. Please contact an authorised
representative for manufacturer for information
concerning the decompositioning of your
requirement.
Internal
Rear Panel, Circuit breaker label of Console and
Front Panel
Rear of console
Rear of console
Rear Panel
1-6Section 1-3 - Important Conventions
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 35 of 252
Page 36
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 1-4Safety Considerations
1-4-1Introduction
The following safety precautions must be observed duri ng all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-4-2Human Safety
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in a LOGIQ A3 Training Seminar are authorized to service the
equipment.
1-4-3Mechanical Safety
WHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE
WARNING
WARNING
EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
WARNING
CAUTION
CAUTION
NOTE:Special care should be taken when transporting the unit in a vehicle:
MAY BE DAMAGED.
Always lock the Control Console in its parking (locked) position before moving the scanner
around.
The LOGIQ A3 weights approx 70kg , depending on installed peripherals, when ready for use.
Care must be used when moving it or replacing its parts. Failure to follow the precautions listed
below could result in injury, uncontrolled motion and costly damage.
ALWAYS:
•Be sure the path way is clear.
•Use slow, careful motions.
•Use two people when moving on inclines or lifting more than 23 kg (50 lb).
•Secure the unit in an upright position.
•Lock the wheels (brake)
•DO NOT use the Control Panel as an anchor point.
•Place the probes in their carrying case.
•Eject any CD (R/W) disk from the CD(R/W) Drive
•Remove the Footswitch and place it in a secure place
•Disconnect any other Off board peripherals if used.
NOTE:Keep the Heat venting holes on the monitor unobstructed to avoid overheating of the monitor.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 36 of 252
Chapter 1 1-7
Page 37
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-4-4Electrical Safety
To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension co rd is used with the system, make sure that the total
current rating of the extension cord is greater than the system rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
CAUTION
If the Power Plug is Modified or replaced to Suit the local Conditions and regulations, Ground
continuity check should be performed between Ground Pin on the Plug and a Metal Part on the
LOGIQ A3.
1-4-5Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that p arts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an
ultrasound probe).
The purpose of the regulation is to p rotect employee s in the transport ation industry, as well as the people who will receive or open this package.
NOTE:The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purpo ses and must be
transported as a hazardous material.
1-8Section 1-4 - Safety Considerations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 37 of 252
Page 38
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 1-5Warning Label Locations
Figure 1-1 Label Location
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 38 of 252
Chapter 1 1-9
Page 39
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
Figure 1-2 Transport Label
Figure 1-3 Gender Caution Label
1-10Section 1-5 - Warning Label Locations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 39 of 252
Page 40
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
Figure 1-4 Rating Plate Label
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 1 1-11
Approved Document - 5213326-100TPH_r3.pdf Page 40 of 252
Page 41
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
Figure 1-5 Caution Label
Figure 1-6 EIAJ Label (Japan Only)
1-12Section 1-5 - Warning Label Locations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 41 of 252
Page 42
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
Figure 1-7 Gost Label
Figure 1-8 TUV Label
Figure 1-9 Circuit Breaker Label
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 42 of 252
Chapter 1 1-13
Page 43
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
Figure 1-10 Circuit Breaker Panel Rating Label
Figure 1-11 Front Panel Rating Label
Figure 1-12 Grounding Label
1-14Section 1-5 - Warning Label Locations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 43 of 252
Page 44
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
L 11
Figure 1-13 Mercury Label
L 13
Figure 1-14 ROHS Label
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 44 of 252
Chapter 1 1-15
Page 45
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Warning Label Locations (continued)
L1 (Figure 1-2)
D1: Standard LCD monitor may rotate while
transporting. Bind the system securely to prevent
damage in transportation.
D2: To avoid injury by tipping over, DO NOT PUSH
THIS UNIT FROM THE SIDES.
D3: DO NOT place a finger, hand or any object on
the joint of the monitor or monitor arm to avoid injury
when moving the monitor and monitor arm.Placing
objects on top of the monitor may cause the monitor
to tilt with the falling objects resulting in injury to the
operator.Do not place any objects on the monitor.
L2 (Figure 1-3)
D4: This machine should be used in compliance
with law. Some jurisdictions restrict certain uses,
such as gender determination.
D6: Possible shock hazard. Do not remove covers
or panels. No user serviceable parts are inside.
Refer servicing to qualified service personnel.
D7: Do not use the following devices near this
equipment: cellular phone, radio receiver, mobile
radio transmitter, radio controlled toy, etc. Use of
these devices near this equipment could cause this
equipment to perform outside the published
specifications. Keep power to these devices turned
off when near this equipment.
D8: Prescription Device: United States law restricts
this device to sale or use by or on the order of a
physicians U.S.A. Only)
D9: The CE Mark of Conformity indicates this
equipment conforms with the Council Directive 93/
42/EEC
D10: CISPR CAUTION: The LOGIQ A3 conforms
to the CISPR11, Group 1, Class A of the
international standard for Electromagnetic
disturbance characteristics.
D11: WEEE Label: This symbol indicates that the
waste of electrical and electronic equipment must
not be disposed as unsorted municipal waste and
must be collected separately. Please contact an
authorized representative of the manufacturer for
information concerning the decommissioning of
your equipment.
D12: The equipment weighs approximately 70 kg
(154 lbs). To avoid possible injury and equipment
damage when transporting from one area of use to
another:
• Be sure the pathway is clear.
• Limit movement to a slow careful walk.
• Use two or more persons to move the equipment
on inclines or long distance.
L5 (Figure 1-6)
D13: EIAJ Label (Japan Only).
L6 (Figure 1-7)
D14: Gost Label (Russia Only).
L7 (Figure 1-8)
D15: TUV Label: TUV Listing and Certification Mark
is used to designate conformance to nationally
recognized product safety standards. The Mark
bears the name and/or logo of the testing laboratory,
product category, safety standard to which
conformity is assessed, and a control number
L8 (Figure 1-9)
D16: Circuit breaker Label
L9 (Figure 1-10)
D17: Circuit breaker panel rating Label.
L12 (Figure 1-11)
D17: Front panel rating Label.
L10 (Figure 1-12)
D18: Signal ground point label CAUTION: This is
only for “FUNCTIONAL GROUNDING”, NOT
“PROTECTIVE EARTH”.
L11 (Figure 1-13)
Lamp contains Mercury Dispose According to Stete/
Local Law.
L13 (Figure 1-14)
ROHS Label
NOTE:For an explanation of the symbols shown in the illustration, refer to latter pages in this chapter.
The CAUTION label for the radio influence is required to be attached on the console from April, 199 6.
The GHOST label is required to be attached on the console from June, 1998.
The Sex determination label is required to be attached on the console from September 2001, for Asia
Only
1-16Section 1-5 - Warning Label Locations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 45 of 252
Page 46
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-5-1Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
DANGER
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
EXPLOSION WARNING: DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE
WARNING
ATMOSPHERE. OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN
ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT: BECAUSE OF THE DANGER
WARNING
OF INTERDICTING ADDITIONAL HAZARDS, DO NOT INSTALL SUBSTITUTE PARTS OR
PERFORM ANY UNAUTHORIZED MODIFICATION OF THE EQUIPMENT.
Section 1-6EMC, EMI, and ESD
1-6-1Electromagnetic Compatibility
Electro Magnetic Compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interface from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio-frequency or electromagnetic interface (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cab les. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-6-2Electrostatic Discharge (ESD) Prevention
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
WARNING
1.)Always connect yourself, via an arm-wrist strap, to the advised ESD
connection point located on the rear of the scanner (to the right of the
power connector).
2.)Follow general guide lined for handling of electrostatic sensitive
equipment.
1-6-3CE Compliance
The LOGIQ A3 unit conforms to all applicable conducted and radiated emission limits and immunity
from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient
requirements.
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are
in good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 46 of 252
Chapter 1 1-17
Page 47
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 1-7Customer Assistance
1-7-1Contact Information
If this equipment does not work as indicated in this service manual or in the User Manual, or if you
require additional assistance, please contact the loca l distributor or appropriate support resource, as
listed below.
-Prepare the following information before you call:
-System ID serial number.
-Software version.
For GE Service:
Table 1-5 GE Service
LocationPhone Number
USA/CANADA:
GE Medical systems
Ultrasound Service Engineering,
4855 W. Electric A venu e Milwaukee, WI 53219
TEL: 1 800-437-1171,1-800-682-5327
1-262-524-5698,FAX: +1- 414-647-4125
LATIN AMERICA:
GE Medical systems
Ultrasound Service Engineering
4855 W. Electric Avenue Milwaukee, WI 53219
ASIA(Singapore/Japan):
GE Ultrasound Asia (Singapore)
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza Singapore 169730
ASIA(China/India/Korea/SEA):
Ultrasound Tech Support - CHINA
Ultrasound Tech Support - INDIA
Ultrasound Tech Support - KOREA
Ultrasound Tech Support - SEA
EUROPE:
GE Ultraschall Deutschland
GmbH & Co. KG
BeethovenstraBe 239
Postfach 11 05 60, D-42665
TEL: (1) 262-524-5300
TEL: (1) 262-524-5698
FAX: +1- 414-647-4125
TEL: 65-291 8528
(81) 426-48-2950
FAX: 65-272-3997
FAX: (81) 426-482902
+(81) 426482944
+(86) 8008108188
+(91) 1800114567
+(82) 262013585
+(65) 62773444
TEL:+49 212 2802 208
+49 212 2802 207
FAX: (49) 212.28.02.431
Solingen Germany
Ultrasound Tech Support - ANZ+(61) 1800647855
1-18Section 1-7 - Customer Assistance
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 47 of 252
Page 48
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-7-2System Manufacturer
Table 1-6 System Manufacturer
WIPRO GE HEALTHCARE PVT LIMITED- INDIA
WIPRO GE HEALTHCARE PVT LIMITED,
NO: 4, KADUGODI INDUSTRIAL AREA,
BANGALORE,KARNATAKA, INDIA-560067.
TEL: (91) 80-4180-1000
FAX:(91) 80-2845-2924
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 1 1-19
Approved Document - 5213326-100TPH_r3.pdf Page 48 of 252
Page 49
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
1-20Section 1-7 - Customer Assistance
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 49 of 252
Page 50
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Chapter 2 Pre-Installation
Section 2-1Overview
2-1-1Purpose
This chapter provides the information required to pl an and pr ep ar e for the installation of a LOGIQ A3 .
Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
2-1-2Table Of Contents
S.NoDescriptionPage Number
1
2
3
Overview
General Console Requirements
Facility Needs
2-1
2-2
2-6
Table 2-1 Contents in Pre-Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 50 of 252
Chapter 2 2-1
Page 51
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 2-2General Console Requirements
2-2-1Console Environmental Requirements
Table 2-2 Environmental Requirements for LOGIQ A3 Scanners
OperationStorageTransport
Temperature
Humidity
Pressure
10 - 40 degree C
50 - 104 degree F
30 - 85%
non-condensing
700 - 1060hPa700 - 1060hPa700 - 1060hPa
-10 - 60 degree C
14 - 140 degree F
30 - 85%
non-condensing
-40 - 60 degree C
-40 - 140 degree F
30 - 85%
non-condensing
Table 2-3 Environmental Requirements for an Ultrasound Room
ItemValues
Power Source
Radiation Shielding
Temperature
Humidity
Heat Dissipation
Floor Landing
Floor Condition
Weight
Refer to Table 2-4 on page 2-3.
NONE REQUIRED for ULTRASOUND ENERGY
20-26 DEG. C (68-79 DEG F) for PATIENT COMFORT
50% to 70% for PATIENT COMFORT
1366 BTU/Hr.
Approximately 680 - 800 kg/m2 without Accessories
Gradient: WITHIN 5 degrees
70 kg without Accessories
2-2-1-1Cooling
The cooling requirement for the LOGIQ A3 is 1366 BTU/hr . This figure does not include cooling needed
for lights, people, or other equipment in the room. Each person in the room places an additional 300
BTU/hr. demand on the cooling system.
2-2-1-2Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2Section 2-2 - General Console Requirements
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 51 of 252
Page 52
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
2-2-2Electrical Requirements
NOTE:GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire (not shared with any other circuit) from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire (not shared with any other circuit) from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-3Power Requirements
Electrical Specifications for LOGIQ A3.
PARAMETERAREALIMITS
Voltage Range230V230 VAC ±10% (207-253 VAC)
Voltage Range120V120 VAC ±10% (103-127 VAC)
PowerAll applicationsMAX. 425VA
Line FrequencyAll applications50/60Hz (±2Hz)
Power TransientsAll applications
Decaying OscillationAll applications
2-2-3-1Inrush Current
Inrush current is not a factor to consider due to in rush curren t limiting prope rties of th e power supplies.
2-2-3-2Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be ready accessible.
POWER OUTAGE MAY OCCUR.
The LOGIQ A3 requires a dedicated single branch circuit. To avoid circuit overload and
DANGER
possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.
Less than 25% of nominal peak voltage
for less than 1 millisecond for any type of
transient, including line frequency,
synchronous, asynchronous, or
aperiodic transients.
Less than 15% of peak voltage for less
than 1 millisecond.
Table 2-4 Electrical Specifications for LOGIQ A3
2-2-3-3Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and te st equipment needed to support this u nit must also
be present within 1 m (3.2 ft.) of the unit. Electrical ins ta llatio n mus t meet all current local, state, and
national electrical codes.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 52 of 252
Chapter 2 2-3
Page 53
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
2-2-3Power Requirements (cont’d)
2-2-3-4Unit Power Plug
If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or
the installation engineer must supply what is locally required.
2-2-3-5Power Stability Requirements
Voltage drop-out / Max 10 ms.
Power Transients
Refer Table 2-4
2-2-4EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wirin g. Ultrasound machines also generate EMI. The LOGIQ
A3 complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. These
sources include:
•medical lasers,
•scanners,
•cauterizing guns,
•computers,
•monitors,
•fans,
•gel warmers,
•microwave ovens,
•light dimmers
•portable phones.
•Lift
The presence of a broadcast station or broadcast van may also cause interference.
See Table 2-5 for EMI Prevention tips.
EMI RuleDetails
Keep the unit at least 5 meters or 15 feet away from other EMI sources. Special shielding may be
Be aware of RF sources
Ground the unit
required to eliminate interference problems caused by high frequency, high powered radio or video
broadcast signals.
Poor grounding is the most likely reason a unit will have noisy images. Check grounding of the
power cord and power outlet.
After you finish repairing or updating the system, replace all covers and tighten all screws. Any
Replace all screws, RF
gaskets, covers, cores
Replace broken RF gaskets
cable with an external connection requires a magnet wrap at each end. Install the shield over the
front of card cage. Loose or missing covers or RF gaskets allow radio frequencies to interfere with
the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket. Do not
turn on the unit until any loose metallic part is removed.
Table 2-5 EMI Prevention/Abatement
2-4Section 2-2 - General Console Requirements
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 53 of 252
Page 54
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
EMI RuleDetails
Do not place labels where
RF gaskets touch metal
Use GE specified harnesses
and peripherals
Take care with cellular
phones
Properly dress peripheral
cables
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF
leakage. Or, if a label has been found in such a position, move the label.
The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable
length, material, and routing are all important; do not change from what is specified.
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays. Loop
the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to the
frame.
Table 2-5 EMI Prevention/Abatement
2-2-5Probes Environmental Requirements
ELECTRONIC
Operation:10 to 40 degree C(50 to 104 degree F)
Storage:-20 to 50 degree C(-4 to 122 degree F)
Temperatures in degree C, conversion to degree F = degree C*(9/5) + 32)
Table 2-6 Operation and storage Temperatures for Probes
NOTE:Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
OF -20 TO + 50(-4 to 122 degree F) degrees C. WHEN EXPOSED TO LARGE TEMPERATURE
VARIATIONS, THE PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS
NOTICE
BEFORE USE.
2-2-6Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
CAUTION
Have two people available to deliver and unpack the LOGIQ A3.
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 2 2-5
Approved Document - 5213326-100TPH_r3.pdf Page 54 of 252
Page 55
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 2-3Facility Needs
2-3-1Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay,
confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
Use the Pre installation checklist to verify that all needed steps have been taken. Purchaser
responsibility includes:
•Procuring the materials required.
•Completing the preparations befo re de l iv er y of th e ultr a sou n d sys tem .
•Paying the costs for any alterations and modification s not specifically provided in the sales contract.
NOTE:All electrical installations that are preliminary to the positioning of the equipment at the site prepared for
the equipment must be performed by licensed electrical contractors. Oth er connections between pieces
of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The
products involved (and the accompanying electrical installations) are highly sophisticated and special
engineering competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified
personnel to perform electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved product further
from the system than the interface kit allows pr esents challenges to the installation team. To avoid
delays during installation, such variances should be made known to the in dividuals or group performing
the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpets are not recommended
because they collect dust and create static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact system reliability.
2-3-2Required Features
NOTE:GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
•Dedicated single branch power outlet of adequate amperage meeting all local and national codes
which is located less than 2.5 m (8 ft.) from the unit’s proposed location
•Door opening is at least 76 cm (30 in) wide
•Proposed location for unit is at least 0.3 m (1 ft.) from the wall for proper cooling of the system
•Power outlet and place for any external peripheral are within 2 m (6.5 ft) of each other with
peripheral within 1 m of the unit to connect cables.
2-6Section 2-3 - Facility Needs
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 55 of 252
Page 56
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
2-3-2 Required Features (cont’d)
NOTE:The LOGIQ A3 has three outlets in the unit. One is for the on board peripheral and two
for off board peripherals.
•Power outlets for other medical equipment
•Power outlets for test equipment and modem within 1 m (3.2 ft) of unit
•Clean and protected space to store transducers (in their cases or on a rack)
•Material to safely clean probes (done with a plastic container, never metal)
2-3-2-1Desirable Ultrasound Room Facilities
•Door is at least 92cm (3 ft.) wide
•Circuit breaker for dedicated power outlet is easily accessible
•Lab sink with hot and cold water
•Receptacle for bio–hazardous waste, like used probe sheaths
•Emergency oxygen supply
•Storage for linens and film
•Medical equipment storage
•Hospital grade equipment electrical outlet
•Nearby waiting room, lavatory, and dressing room
•Dual level lighting (bright and dim)
•Lockable cabinet for GE Software & proprietary manuals
•Trash bin.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 56 of 252
Chapter 2 2-7
Page 57
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
2-3-2-2Minimal Floor Plan Suggestion
Figure 2-1 Minimal Floor Plan
2-8Section 2-3 - Facility Needs
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 57 of 252
Page 58
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Chapter 3Installation
Section 3-1Overview
3-1-1Purpose
This chapter contains information needed to install the unit. Included are references to a procedure that
describes how to receive and unpack the equipment and how to file a damage or loss claim
3-1-2Table Of Contents
Table 3-1 Contents in Chapter 3
S.NoDescriptionPage Number
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
Overview
Receiving and Unpacking LOGIQ A3
Safety Reminders
Moving into Position
Adjusting System Clock
Product Locator Installation Card
Preparing for Installation
Completing the Installation
System Configuration
Network Connectivity
Optional Peripherals
Connectivity Installation Worksheet
Paper work
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 3 3-1
Approved Document - 5213326-100TPH_r3.pdf Page 58 of 252
Page 59
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-1-3Average Installation Time
Table 3-2 Average Installation Time
DescriptionAverage Installation TimeComments
Unpacking the scannerApproximately 0.5 hour
Scanner wo/optionsApproximately 0.5 hourDependant on the required configuration
The LOGIQ A3 including options has been designed to be installed and checked out by an experienced
service technician in approximately Four hours.
NOTE:For Installing Options average installation time is approximately between 0.5 - 2hrs depending
on the required configuration
3-1-4Installation Warnings
1.) Since the LOGIQ A3 weighs approximately 70 kg without options, preferably two people should
unpack it. Two people are also preferable for installing any additional bulky items.
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
NOTE:For information regarding packing labels, refer to LABELS ON PACKAGE.
3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hou r for each 2.5
it is below 10
o
C or above 40oC.
o
C increment in it's temperature, if
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 59 of 252
Page 60
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2Receiving and Unpacking LOGIQ A3
When a new system arrives, ensure that none of the components are damaged or in short supply. If
shipping damage or shortage occurs, contact the address shown in chapter-introduction.
The Packaging/Unpacking Procedure.
Topics discussed in the Packaging/Unpacking Procedure:
1.) Check the Shipment,
2.) Unpack the Unit,
3.) Handling Incomplete or Damaged Shipment
Please read these procedures before packing/unpacking the LOGIQ A3.
We strongly advice you to store the LOGIQ A3 packing material in undamaged condition in case of
future transportation.
Do not lift the unit by the Keyboard. This may cause equipment damage.
CAUTION
The crate with the LOGIQ A3 weighs approximately 70 kg. Be prepared for a sudden shift of
weight as the unit is removed from its base (pallet).
CAUTION
CAUTION
Inspect the Shock watch and Tilt watch. Ensure that they are not fused. (Refer to the instructions
attached on the packing Box on howto check the Shcik watch and Tiltwatch)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 60 of 252
Chapter 3 3-3
Page 61
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2 Receiving and Unpacking LOGIQ A3 (cont’d)
1.) Remove all the Metal locks to dismantle the plywood packing.
Figure 3-1 The metal bands to be removed
3-4Section 3-2 - Receiving and Unpacking LOGIQ A3
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 61 of 252
Page 62
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2 Receiving and Unpacking LOGIQ A3 (cont’d)
2.) Lift the TOP Cover up and off.
Figure 3-2 Removing the top cover
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 62 of 252
Chapter 3 3-5
Page 63
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2 Receiving and Unpacking LOGIQ A3 (cont’d)
3.) Remove the Front plywood...
Figure 3-3 Removing the front plywood
3-6Section 3-2 - Receiving and Unpacking LOGIQ A3
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 63 of 252
Page 64
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2 Receiving and Unpacking LOGIQ A3 (cont’d)
4.) Move the slide plywoods to make space for placing the ramp.
5.) Remove the internal foams carefully.
6.) Place the top cover as shown in the figure 3-4 below to form the ramp to put the console down.
Figure 3-4 Position Top Cover
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 64 of 252
Chapter 3 3-7
Page 65
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-2 Receiving and Unpacking LOGIQ A3 (cont’d)
7.) Unlock the brakes on the front castors, then carefully put the console off the Palette.
8.) Remove the Plastic Wrapping around the LOGIQ A3.
Figure 3-5 Moving the LOGIQ A3 from the crate
3-8Section 3-2 - Receiving and Unpacking LOGIQ A3
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 65 of 252
Page 66
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-2-1Safety Reminders
DANGER
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate
to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use
a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
Do not operate this unit unless all board covers and frame panels are securely in place.
System performance and cooling require this.
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQ A3 and
kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the LOGIQ A3 probe is within AIUM/NEMA
standards, avoid unnecessary exposure. Ultrasound energy can produce heat and
mechanical damage.
NOTE:Check the shipping container for special instructions. Verify that the container is intact. In some cases
a secondary container may be used. If so, ask the carrier for unpacking instructions.
Figure 3-6 LABELS ON PACKAGE
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 66 of 252
Chapter 3 3-9
Page 67
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-2-2Moving into Position
Do not lift the unit by the Keyboard.
Do not tilt the unit more than 5 degrees to avoid tipping it over.
CAUTION
Equipment Damage Possibility. Lifting the console by holding covers may damage the covers.
Do not lift the console by holding any covers.
CAUTION
In general, a single adult can move the LOGIQ A3 along an even surface with no steep grades. At least
two people should move the machine when large humps, grooves, or grades are encountered. (It is
better to pull from the rear rather than push from the front of the unit). Before moving, store all loose
parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.
Although LOGIQ A3 is a compact and mobile machine, two people should move it over rough surfaces
or up and down grades.
3-2-3Adjusting System Clock
Set the system clock for the LOGIQ A3 to the local time. For procedure of adjusting the system clock,
refer to Software Configuration Checks, in Chapter FUNCTIONAL CHECKS.
3-2-4Product Locator Installation Card
Fill out proper customer Information on the Product Locator Installation Card. Mail this Installation Card
“Product Locator” to the address corresponding to your pole.
NOTE:The Product Locator Installation Card shown may not be same as the provided Product Locator card.
PRODUCT LOCATOR INSTALLATION CARD
3-10Section 3-2 - Receiving and Unpacking LOGIQ A3
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 67 of 252
Page 68
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-3Preparing for Installation
3-3-1Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-3-2Physical Inspection
3-3-2-1Systems Voltage Settings
Verify that the recieved scanner is set to the correct voltage.
The Voltage ratings for the LOGIQ A3 Scanner is found on Rating label near the Circu it Breaker at the
rear of the system
WARNINGWARNING
CONNECTING A LOGIQ A3 SCANNER TO THE WRONG VOLTAGE LEVEL WILL
MOST LIKELY DESTROY THE SCANNER.
3-3-2-2Video Formats
Check that the video format is set to the locally used video standard, NTSC or PAL.This can be done
in the setup menu.(Refer to Chapter 13 of the Basic Users Manual for the pr oced ure for chan ge in the
settings)
3-3-3EMI Protection
This unit has been designed to minimize the effects of Electo-Ma gnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
3-3-4Checking the Components
When a new system arrives, check that nonne of the components are damaged or in short supply. If
shipping damage or shortage occurs, contact the address shown in Chapter-Introduction.
If the Power Plug is Modified or replaced to Suit the local Conditions and regulations, Ground
continuity check should be performed between Ground Pin on the Plug and the Metal Part on
CAUTION
the LOGIQ A3. Refer Section 10-6-4 for more details.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 68 of 252
Chapter 3 3-11
Page 69
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-4Completing the Installation
3-4-1System Specifications
3-4-1-1Physical Dimensions
The physical dimensions of the LOGIQ A3 unit are summarized in Table 3-4 on page 12 .
The Size of LOGIQ A3, with monitor and peripherals is shown in Figure 3-7 on page 12
Table 3-4 Physical Dimension of LOGIQ A3
HeightWidthDepthUnit
12541.659cm
Note: Tolerance of +/- 1.5 cm
125 cm
41.6cm
Figure 3-7 OVERALL DIMENSIONS
59cm
3-12Section 3-4 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 69 of 252
Page 70
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-4-1-2Weight with Monitor and without Peripherals
The Weight of LOGIQ A3 with monitor, without peripherals
Table 3-5 Weight of LOGIQ A3
ModelWeight [Kg]Weight [lbs]
LOGIQ A370154
3-4-1-3Acoustic Noise Output:
Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2.
3-4-2Electrical Specifications
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before
installing the LOGIQ A3 console
Table 3-6 Electrical Specification for LOGIQ A3
SystemVoltageCurrentFrequency
1120 ~ VAC4.25A50-60 Hz
2230 ~ VAC1.94A50-60 HZ
3-4-3Probe (Transducer) Connection
1.) Connect a transducer to the upper transducer receptacle as follows:
A.) Ensure that the transducer twist lock lever points towards the 9 o'clock position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle
mating surface.
C.) Twist the transducer twist lock lever to the 2 o'clock position to lock it in place. Twist the
lever to the 9 o'clock position to disconnect the transducer.
NOTE:It is not necessary to turn OFF power to connect or disconnect a transducer.
Connect the main power cable to a hospital grade power receptacle with the proper rated voltage
checked during pre installation. Never use a three-to-two prong adapter; this defeats the safety ground.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 70 of 252
Chapter 3 3-13
Page 71
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-4-4Power On / Off Process
NOTE:After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-4-4-1Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power re ceptacle with the proper rate d voltage.
Never use an adapter that would defeat the safety ground.
3.) Switch ON the Circuit Breaker at the rear of the system.
A
B
Figure 3-8 Circuit Breaker and Power Cable on Back of Scanner
A- Circuit Breaker
B- Power cable
When power is applied to the scanner and the rear Circuit Breaker is turn ed ON, power is distributed to
the Transformer Assembly
3-14Section 3-4 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 71 of 252
Page 72
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-4-4-2System Boot Up
Press the Power On/Off switch on the Control Panel once.
1
Figure 3-9 Power On/Off Switch for Back-end Processor Boot Up
1-Power ON/OFF Switch
When the Power On/Off switch on the Control Panel is pressed once, the Power is distributed to the
Front-End, to the Peripheral outlets and to the Master Board at Back End. Master Board at Back End
starts and the software is initiated to start the scanner.
No status messages are displayed during this process.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 72 of 252
Chapter 3 3-15
Page 73
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-4-5Power Off/ Shutdown
NOTE:After turning off a system, wait at least ten seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
3-4-5-1System Power Down
To power off the system:
1.) Lightly press the Power On/Off switch at the front of the system once. The Shutdown window
is displayed.
NOTE:DO NOT press and hold down the Power On/Off switch to shutdown the system. Instead, lightly
press the Power On/Off switch and select Shutdown.
2.) Using the Trackball, select Shutdown.
The shutdown process takes a few seconds and is completed when the control panel
illumination shuts down.
3.) Disconnect the probes.
Clean or disinfect all probes as necessary. Store them in their shipping cases to avoid
damage.
4.) If daily maintenance is to be performed, turn off th e circ uit br e ake r .
NOTE:Do not pull the power cable without turning off the circuit breaker.
CAUTION
DO NOT turn off the circuit breaker before the Power On/Off switch LED is
amber.
Data may be lost or system software damaged if the circuit breaker is turned off
before the Power On LED is amber.
(or)
1.) Press the Power On/Off switch on the Control Panel , and hold it till power goes off from the
system.
3-16Section 3-4 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 73 of 252
Page 74
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-4-5-2Scanner Shutdown
A
B
Figure 3-10 Circuit Breaker and Power Cable on Back of Scanner
A- Circuit Breaker
B- Power cable
1.) Switch OFF the Circuit Breaker at the back of the system.
2.) Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 74 of 252
Chapter 3 3-17
Page 75
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-5System Configuration
3-5-1System Configuration
3-5-1-1System Settings
Table 3-1 System Configuration
Configuration CategoryDescription
Enables the user or service personnel to set the
Settings
1.) Power on the scanner.
2.) Press Utilities button
3.) Select “System” tab to display system settings menu.
date, time, unit, language, basic information about
the organization such as the institution name and
department.
4.) Set the Hospital name, Department, Date format, Language, and Units.
5.) Click on “Save” button.
6.) Click on Exit or Utilities to terminate the utility function.
NOTE:After making changes reboot the system when message appears on the screen.
3-18Section 3-5 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 75 of 252
Page 76
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-6Network Connectivity
You use Connectivity functionality to set up the connection and communication protocols for the
ultrasound system. This page gives an overview of each of the connectivity functions.
This configuration category enables users with rights to set the network connection for the system and
connected remote PC to store images data. The windows XP remote PC with “Internet Information
Services” enabled should be required.
NOTE: The remote PC with Windows XP Operating System should be required, and also make sure
Internet Information Services Installed on PC.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 76 of 252
Chapter 3 3-19
Page 77
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-6 Network Connectivity (cont’d)
Check the availability of Internet Information Services by following below steps:
1.) Select “Start->Control Panel”
Figure 3-1. Control Panel
2.) Select “Administrative Tools” from control panel window.
Figure 3-2. Administrative Tools
3-20Section 3-6 - Network Connectivity
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 77 of 252
Page 78
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-6 Network Connectivity (cont’d)
3.) Select “Internet Information Services” from Administrative tools.
Figure 3-3. Internet Information Services
4.) Double click on Internet Information Services. Select “FTP sites” under “local computer” of
“Internet Information Services” window.
Check for the “Default FTP site” state as “Running”.
Figure 3-4. FTP site
NOTE: Default FTP site state as “Running” indicates that the system is ready for file transfer.
5.) Select the “Network Connection” from “Control Panel” to setup static IP Address. From
“Network Connections”, select “Local Area Network”.
The “Local Area Connection properties” window displays on the screen. Select “Internet Protocol
(TCP/IP)”-> ”Properties”. Select “Use the following IP Address” from Internet Protocol (TCP/IP)
properties.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 78 of 252
Chapter 3 3-21
Page 79
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-6 Network Connectivity (cont’d)
Figure 3-5. Static IP Address Setup
Enter IP address (for eg:3.245.71.173), and enter subnet mask as “255.255.255.0”.
6.) Now connect ethernet cable between the LOGIQ A3 System and remote PC.
To enable the network connectivity, select “Utility-> Measurement->Enable Network” option.
It allows to enter the following Data:
A.) Type the name of the Ultrasound system in the AE Title of Devices section.
B.) type port number (Default:“21”) in port field of Devices section.
C.) In the IP settings section, identify the ultrasound system to the rest of the network by one
of the following:
*DO NOT enable DHCP.
*Type the IP-Address (acquire unique static IP address from hospital network
administrator, for e.g.: 3.245.71.170), Subnet Mask as 255.255.255.0, and Default
Gateway (if applicable).
*Enable DHCP, the system acquires dynamic IP Address from DHCP server.
D.) In the External Host section, type the IP Address (acquire unique static IP address from
hospital network administrator, for e.g.: 3.245.71.173).
E.) On pressing “Ping” button tests the network connectivity.
Ping successful on the “ping status box” indicates the system is successfully connected to the
network, ping failure message on the “ping status box” indicates the network connection
failed.
Reboot the system once the network connection is established.
3-22Section 3-6 - Network Connectivity
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 79 of 252
Page 80
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-7Storage And Operation Requirements
The LOGIQ A3 is shipped in a single container excluding PROBES. Shipping weight is approximately
70 kg. The size of the container is B120 cm x W66 cm x H139 cm. (47 in. x 26 in. x 55 in). Table 3-2
provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping,
installation, and operation.
Table 3-1 STORAGE AND OPERATION REQUIREMENTS
ParameterStorageOperation
Temperature (Deg C)
(Deg F)
Atmospheric Pressure
(hPa)
Humidly (%)
(non-condensing)
-10 to 60
14 to 140
700 to 1060700 to 1060
30 to 8530 to 85
Section 3-8Optional Peripherals
3-8-1Optional Peripherals/Peripheral Connection
See the Internal and External I/O description in Chapter 5.
3-8-1-1Approved peripherals
The below table shows the suggested optional peripherals for LOGIQ A3. CD Drive is the only Onboard peripheral. All other peripherals are Off-board.
Table 3-2 List of Recording Devices
DeviceManufacturerModel
B/W Video PrinterSONY
Deskjet PrinterHPHP-1160C H47002LMNTSC/PAL
UP-897MDW
UP-D897MD
10 to 40
50 to 104
Catalog Number
H47012LW
Video Signal
NTSC/PAL
Table 3-3 USB Devices
Device
Device
USB
USB
USB HDD
Name
JetFlashTS256
M
CruzerSanDiskDigital
Silver Series
Portable Hard
Drive
ManufacturerDevice TypeMemory Size
TranscendDigital512MB USB 2.0 Flash Memory
iomegaDigital<= 120GB Hard disk
NOTE:See each option installation instructions for installation and connection procedures.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 80 of 252
Chapter 3 3-23
512MB USB 2.0 Flash Memory
Page 81
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-8-1-2Connecting Cables
Equipment damage possibility. Be sure to use the following recommended connecting cables to
connect recording devices and a network with LOGIQ A3 console.Failure to do so would result
CAUTION
in unstable system behaviour and can possibly damage the Equipment
Table 3-4 LIST OF CONNECTING CABLES
NamePart No.FigureNOTE
Use only approved IEC Couplers
Power CableP9509EE
for peripherals connection to
System Rear panel
P9509MH
Mini-Plug CableP9509BE
3-8-1-3Power Consumption of Optional Peripherals
Table 3-5 Power Consumption of Optional Recording Devices
DeviceManufacturerModel
B/W Video PrinterSONY
Deskjet PrinterHPHP-1160C50
UP-897MDW
UP-D897MD
NOTE:At Any point of time, only one peripheral should be activated.
Used only for installing Peripherals
on front panel of the console
For control signals (used only for B/
W video printer)
Power Consumption
(VA)
133
133
3-24Section 3-8 - Optional Peripherals
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 81 of 252
Page 82
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-8-2Available Probes
See in specification in the LOGIQ A3 User Manual for Probes and intended use.
See Chapter - Renewal Parts for Part Numbers to be used when ordering new or repla cemen t p robes.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 85 of 252
Page 86
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 3-9Paper work
NOTE:During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for
the peripheral units must be kept as part of the original system documentation. This will ensure
that all relevant safety and user information is available during the operation and service of the
complete system.
3-9-1Product Locator Installation
NOTE:The Product Locator Installation Card shown may not be same as the provided Product
Locator card.
Figure 3-1 Product Locator Installation Card
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 3 3-29
Approved Document - 5213326-100TPH_r3.pdf Page 86 of 252
Page 87
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
3-9-2GE Cares Sticker
Stick GE CARES sticker after Installation. Refer Figure 3-12 for details.
NOTE:InSite is not supported in Logiq A3, hence is not applicable.
Figure 3-12 GE CARES Sticker Location
3-9-3User Manual(s)
1.) User: Check that the correct User Manual(s) for the system and software revision, is
included with the installation. Specific language versions of the User Manual may also be
available. Check with your GE Sales Representative for availability.
3-30Section 3-9 - Paper work
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 87 of 252
Page 88
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Chapter 4Functional Checks
Section 4-1Overview
4-1-1Purpose
This chapter provides procedures for quickly checking major functions of the LOGIQ™ A3 scanner
diagnostics by using the built-in service software, and power supply adjustments.
4-1-2Table Of Contents
Table 4-1 Contents in Chapter 4
S.NoDescriptionPage Number
NOTICE
1
2
3
4
5
6
7
8
Most of the information pertaining to this Functional Checks chapter is found in the LOGIQ™ A3
Overview
Required Equipment
General Procedure
Functional Checks
Software Configuration Checks
Peripheral Checks
Safety Issues
Site Log
Basic User Manual (Direction number 5191391-100).
Look for the letters (BUM) after a section in the Table of Contents to determine if the information
is in this chapter or in the Basic User Manual.
Section 4-2Required Equipment
•An empty (blank) CD R/W disk
•At least one transducer. (normally all the transducers used on the system should be checked).
4-1
4-1
4-2
4-3
4-11
4-11
4-13
4-14
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 88 of 252
Chapter 4 4-1
Page 89
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 4-3General Procedure
CAUTION
SYSTEM REQUIRES ALL COVERS
Operate this unit only when all board covers and frame panels are securely in place. The
covers are required for safe operation, good system performance and cooling purposes.
NOTICE
Lockout/Tagout Requirements (For USA only)
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable
on the system.
TAG
&
LOCKOUT
Date
Signed
4-3-1Power On/Boot Up
For Power On/Off process , Refer Power On / Off Process on page 3 - 14
4-3-2Adjusting the Display Monitor
4-3-2-1Brightness and Contrast
To adjust the brightness:
1.) Press the menu button of the display.
2.) Press the “+” button to increase brightness.
Press the “-” to decrease brightness.
The amount of brightness is shown on a slider on the screen.
To adjust the contrast:
1.) Press the menu button of the display. Select the contrast from the popup menu.
2.) Press the “+” button to increase contrast.
Press the “-” to decrease contrast.
The amount of the contrast is shown on a slider on the screen.
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug.
NOTICE
TAG
&
LOCKOUT
Date
Signed
4-2Section 4-3 - General Procedure
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Lockout/Tagout Requirements (For USA / Europe only)
Follow OSHA Lockout/Tagout requirements by ensuring you are in to ta l control of the Power Ca ble
on the system.
Approved Document - 5213326-100TPH_r3.pdf Page 89 of 252
Page 90
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 4-4Functional Checks
4-4-1Basic Controls
4-4-1-1Alpha Numeric Keyboard
Table 4-2 Keyboard
TaskExpected Result
EscExit current display screen.
4-4-1-2Keyboard Controls
Table 4-3 Keyboard Controls
TaskExpected Result
PatientEnter Patient Demographic data screen.
B-Mode (Scan)
Report
UtilityActivates the machine’s configuration.
ProbeActivates probe selection menu.
Returns machine to scanning state and select
scan mode.
Activates default report and touch screen of
report choices.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 90 of 252
Chapter 4 4-3
Page 91
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-2B Mode Checks
4-4-2-1Preparations
1.) Connect one of the probes listed in Available Probes of Chapter - Installation to the system probe
connector.
2.) Turn ON the scanner (if it isn’t turned on already))
3.) Perform the “Screw driver Test” on the system, to ensure that that B-mode image is fine.
Figure 4-1 LOGIQ™ A3 B Mode Controls
4-4Section 4-4 - Functional Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 91 of 252
Page 92
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-2-2B Mode OP Panel Controls
Table 4-4 B Mode Control Panel Controls
1
2
3
4
5
6
7
8
9
10
S.No
TaskExpected Result(s)
Press B Mode keyB Mode Starts
Adjust Depth
Adjust Gain
Adjust Time Gain Compensation (TGC)
Activate Auto Optimize
Adjust Zoom
Adjust Focus
ReverseToggles the left/right orientation of the scan image.
Harmonics (Activate Tissue Harmonics)
Dynamic Range
Adjust the field of view. Increasing the depth may view larger/deeper structures rates, and
decreasing the depth may view near the skin line.Turn Depth right/left to increase/decrease.
Depth displays on the monitor in cm.
Controls the amount of echo information displayed in an image. Turn B Mode dial to the left/
right to increase/decrease Gain. Gain displays on the monitor in G (dB).
Amplifies the returning signals to correct for the attenuation caused by tissues at increasing
depth. TGC slide pots spaced proportional to the depth. Move the slide pots to the left/right to
decrease/increase TGC. A TGC curve appears on the display.
Optimize the image based upon a specified region of interest or anatomy. Press the right Auto
control to activate.
invokes the ROI on the display, according to ROI movement by using trackball thye zoomed
image with respect to ROI will be displayed on the image area of the screen.
Changes the location of the focal point(s). A triangular focus marker indicates the depth of the
focal point.
Diminishes low frequency high amplitude noise and improves imaging. Enhances near field
resolution for improved small parts imaging as well as far field penetration. Press Harmonics
Key to activate.
Dynamic Range controls how echo intensities are converted to shades of gray, thereby
increasing the adjustable range of contrast. This Control is available on the Top Menu
Table 4-5B ModeTop and Sub menu Controls
StepTaskExpected Result(s)
1
Image InvertFlips the image top/bottom by 180 degrees.
2
Edge Enhance
3
Frequency
4
Biopsy GuidelineBiopsy guidelines appears.
5
Grey MapImage brightness will be adjusted as per grey map selection.
6
Frame AverageSmooths the image by averaging frames.
Enhance the gray scale differences corresponding to the edges of structures. Press and turn Edge Enhance
left/right to cycle through the settings.
Enables the adjustment of the probe’s operating frequency. Rotate Frequency and select desired value. The
selected frequency is displayed in the status window.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 4 4-5
Approved Document - 5213326-100TPH_r3.pdf Page 92 of 252
Page 93
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-3M Mode Controls
4-4-3-1Preparations
1.) Connect one of the probes listed in Available Probe of Chapter-Installation to the system probe
connector.
2.) Turn ON the scanner (if it isn’t turned on already)
Figure 4-2 Controls available in M Mode
4-6Section 4-4 - Functional Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 93 of 252
Page 94
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-3-2M Mode OP Panel Controls
Table 4-6 M Mode OP Panel Controls
StepTaskExpected Result(s)
1
Press M Mode keyM Mode Starts
Controls the amount of echo information
2
Adjust Gain
displayed in an image. Turn M Mode dial to
the left/right to increase/decrease Gain. Gain
displays on the monitor in G (dB).
Displays the M-Mode cursor on the B-Mode
image. Press M/D Cursor and Trackball to
position M-Mode Cursor.
4-4-3-3
3
Display M-Mode Cursor
M ModeTop and Sub menu Controls
Table 4-7 M Mode Top and Sub Menu Controls
StepTaskExpected Result(s)
1
Sweep Speed
2
RejectionAffects border sharpness.
3
MGrey MapImage brightness will be adjusted as per grey map selection.
Changes the speed at which timeline is swept. Rotate Sweep Speed left/right to increase/decrease
the value.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 4 4-7
Approved Document - 5213326-100TPH_r3.pdf Page 94 of 252
Page 95
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-4Measurement and Multi Image Checks
4-4-4-1Performing B Mode Measurements Checks
Figure 4-3 Measurement Keys
Table 4-8 Measurement Checks
StepTask Expected Result(s)
1.
Generate the B image.B image is active on the screen
2.
Press FreezeFreeze the image to measure.
The measurement soft menu for the current
Press MEASURE on the Control Panel to
3.
enter the Assign and Measure modality.
4
Trackball the cursor
5.
Press Set.Anchor the start point of the measurement.
6.
Trackball the cursor
exam category is displayed on the Monitor.
Select the appropriate exam category and
measurement kind.
Move the cursor to the start point of the
measurement
Move the cursor to the measurement end
point.The current distance value is displayed
in the Measurement result table and is
instantaneously updated when moving the
cursor
4-8Section 4-4 - Functional Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 95 of 252
Page 96
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-5Basic Measurements
NOTE:The following instructions assume that you first scan the patient and then press Freeze.
4-4-5-1Distance and Tissue Depth Measurements
1.) Press MEASURE once; an active caliper displays.
2.) To position the active caliper at the start point (distance) or the most anterior point (tissue
depth), move the TRACKBALL.
3.) To fix the start point, press Set. The system fixes the first caliper and displays a second active
caliper.
4.) To position the second active caliper at the end point (distance) or the most posterior point
(tissue depth), move the TRACKBALL.
5.) To complete the measurement, press SET. The system displays the distance or tissue depth
value in the measurement results window.
Before you complete a measurement:
To toggle between active calipers, press MEASURE.
To erase the second caliper and the current data measured and start the measurement again,
press CLEAR once.
NOTE:After you complete the measurement, to erase all data that has been measured to this
point, but not data entered onto worksheets, press CLEAR.
4-4-5-2Circumference/Area (Ellipse) Measurement
1.) Press MEASURE once; an active caliper displays.
2.) To position the active caliper, move the TRACKBALL.
3.) To fix the start point, press SET. The system fixes the first caliper and displays a second active
caliper.
4.) To position the second caliper, move the TRACKBALL.
5.) Turn the ELLIPSE control; an ellipse with an initial circle shape appears.
NOTE:Be careful not to press the Ellipse control as this activates the Body Pattern.
6.) To position the ellipse and to size the measured axes (move the calipers), move the
TRACKBALL.
7.) To increase the size, turn the ELLIPSE control in a clockwise direction. To decrease the size,
turn the ELLIPSE control in a counterclockwise direction.
8.) To toggle between active calipers, pr es s MEASURE.
9.) To complete the measurement, press SET. The system displays the circumference and area
in the measurement results window.
Before you complete a measurement:
-To erase the ellipse and the current data measured, press CLEAR once. The original caliper
is displayed to restart the measurement.To exit the measurement fu nction without completing
the measurement, press CLEAR a second time.
4-4-5-3Report Pages
Measurements/Calculations can also be displayed on Report Pages.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 96 of 252
Chapter 4 4-9
Page 97
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-6Probe/Connectors Usage
4-4-6-1Connecting a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to th e probe head could result in irreparable
damage.
4.) Turn the connector locking handle counterclockwise.
5.) Align the connector with the probe port and carefully push into place.
6.) Turn the connector locking handle cloc kwise to secure the probe connector.
7.) Carefully position the probe cable in the probe cord holder spot so it is free to move, but not
resting on the floor.
4-4-6-2Activating the probe
Press “Probe” Key available on the control panel. The next available probe will be actiavetd and
presets will be loaded automatically.
When deactivating the probe, the probe is automatically placed in standby mode.
1.) Press the Freeze key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for
complete probe cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
4-4-6-3Disconnecting the probe
Probes can be disconnected at any time. However, the probe should not be selected as the active
probe.
1.) Move the probe locking handle counterclockwise. Pull the pr obe and connector str aight out of
the probe port.
2.) Carefully slide the probe and connector away from the pr obe port and a round the right side of
the keyboard. Ensure the cable is free.
3.) Be sure that the probe head is clean before placing the probe in its storage box.
4-10Section 4-4 - Functional Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 97 of 252
Page 98
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
4-4-7Using Cine
4-4-7-1Activating CINE
Press FREEZE, Then cine guage appears on the bottom of submenu. Use “Trackball” to move frame
by frame or Gain key. Press “Numeric Key 2” to start cine loop.
4-4-8Image Management (QG)
For Image Management functionality refer to the LOGIQ™ A3 Quick Guide. It talks about several topics:
•Clipboard
•Internal Memory
•Starting an Exam
4-4-9CF Checks
If all the previous tests have been passed successfully, the CF is most likely OK. If the system seems
to be operating erratically, Please refer to Chapter7, Diagnosti c/Troubleshooting.
4-4-10Software Configuration Checks
Table 4-9 Software Configuration Checks
StepTask to doExpected Result(s)
1.
Check Date and Time settingDate and Time are correctAdjust the Date and Time setting
Check that Location (Hospital Name and
2.
Department) is correct
3.
Check Language settingLanguage is properSet the proper Language
4.
Check Units settingUnits are properSet the local units
5
Check assignment of Printer Keys
Check that all of the customer’s options are
6
set up correct
Location Name is correctRe-enter the correct Location Name
Print1-2 Keys are assigned as desired by the
customer
All authorized functions are enabled
Section 4-5Peripheral Checks
Check that peripherals work as described below:
Table 4-10 Peripheral checks
StepTask to do Expected Result(s)
1.
Press(FREEZE
)Stop image acquisition.
2.
Press (PRINT 1
3.
Press B MODE
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
) or (PRINT2) on the Control panel
on the Control Panelto return to the scanning mode
Chapter 4 4-11
The image displayed on the screen is printed on B&W,
depending on the key assignment configuration.
Approved Document - 5213326-100TPH_r3.pdf Page 98 of 252
Page 99
GE HEALTHCARE
DIRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 4-6Presets Backup/Restore Checks
Check that preset Backup/Restore feature work as described below:
1.) Presets Backup: It allows to save presets on memorystick.
2.) Presets Restore: It allows to copy the presets from memory stick to the system.
4-6-1Presets Backup
1.) Create a folder "imgsettingsbackup" on Memory stick
2.) Create a file "unopresetbackup" on Memory stick
3.) Switch Off the system and insert memory stick on top panel memory port
4.) Switch ON the system and allow it to boot.
The system copies the Imaging Presets onto the Memory stick.
Rename the file on Memory stick “unopresetbackup” as "unopresetbackup_ " to have a backup
before restoring in to the system.
4-6-2Presets Restore
1.) Create a file “unopresetbackup” as "unopresetrestore" onto memory stick
2.) Switch Off the system and insert memory stick on top panel memory port
3.) Switch ON the system and allow it to boot.
The presets from memory stick get restored on the System at bootup.
4-12Section 4-6 - Presets Backup/Restore Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 99 of 252
Page 100
GE HEALTHCARE
D
IRECTION 5213326-100, REVISION 3LOGIQ™ A3 SERVICE MANUAL
Section 4-7Safety Issues
4-7-1Probe/Connectors Check
Take the probes and check them as described below:
Table 4-11 Probe and connectors check
StepTask to doExpected Result(s)
1.
Test Each delivered Probe it will display pictorial data on the screen
2.
Test each probe in each connector slotIt will display pictorial data each time
Hold the probe connector vertically with the cable pointing
upward.
Turn the connector locking handle to the horizontal position.
Align the connector with the probe port and carefully push into
place.
3.
Rotate the locking handle to the full vertical position to lock in
place. Position the probe cable so that it is not resting on the floor
CAUTION:
Do not allow the probe head to hang freely. Impact to the probe
head may result in irrepairable damage.
Press the “Probe” key available on the control panel to activate
4.
second probe connected.
Able to connect a probe
Second probe details loaded on the screen automatically.
Rotate the lock handle counter-clockwise to the horizontal
position to unlock the connector.
5.
Remove the connector from the port
Ensure that the probe head is clean before placing the probe in its
storage case.
WARNINGWARNING
Probes can be connected at any time, whether the unit is on or off
Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously, including
ultrasound probe connectors.
WARNINGWARNING
Take the following precautions with the probe cables: Keep free from the wheels. Do
not bend.Do not cross cables between probes.
4-7-2Power Supply Adjustment
There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a
component connected to that specific power outlet.
To disconnect probes:
The probes that are not connected to the unit should be stored in
their storage case.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Chapter 4 4-13
Approved Document - 5213326-100TPH_r3.pdf Page 100 of 252
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.