GE LOGIQ 7, LOGIQ 7 Pro Advanced reference manual

Technical Publications
Direction 5116269-100 Rev. 2
GE Medical Systems
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
R4.2.0/R4.2.0Pro
Regulatory Requirement
This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices
This manual is a reference for the LOGIQ 7/LOGIQ 7 Pro. It applies to all versions of the SW R4.2.0/R4.2.0Pro and later software for the LOGIQ 7/LOGIQ 7 Pro ultrasound systems.
GE Medical Systems
GE Medical Systems: Telex 3797371 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. (Asia, Pacific, Latin America, North America)
GE Medical Systems 283, rue de la Miniere,78533 Buc Cedex France
Revision History
Table i-1: Reason for Change
REV DATE REASON FOR CHANGE
Rev. 1 Aug. 30th, 2004 Initial Release Rev. 2 February 25, 2005 R4.2.0/R4.2.0Pro Release
List of Effective Pages
Table i-2: List of Effective Pages
REVISION
PAGE NUMBER
Title Page Rev. 2 Chapter 2 Rev. 2 Revision History Rev. 2 Chapter 3 Rev. 2 Table of Contents Rev. 2 Index Rev. 2 Chapter 1 Rev. 2
NUMBER PAGE NUMBER
REVISION
NUMBER
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on GPC (GE Medical Systems Global Product Configuration). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual i-1 Direction 5116269-100 Rev. 2
This page intentionally left blank.
i-2 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Chapter 1 — Acoustic Output
Bioeffects
Concerns Surrounding the Use of Diagnostic Ultrasound - - - - - - - - - 1-2
Acoustic Output Data
Maximum output summary- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3
Chapter 2 — Measurement Formulas
Cardiac Measurement
Cardiac measurement abbreviation list- - - - - - - - - - - - - - - - - - - - - - 2-2
Measurement Formulas
Formulas–Generic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-14 Formulas-Urology - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-15 Formulas-Abdominal/Small Parts- - - - - - - - - - - - - - - - - - - - - - - - - 2-16 Formulas–OB- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-17 Formulas–GYN - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-21 Formulas–Vascular- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-23 Formulas-Cardiac - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-24

Table of Contents

Table of Contents
Chapter 3 — OB Tables
OB Tables
ASUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2 Berkowitz- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-8 Bertagnoli - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-9 Brenner - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-9 Campbell - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-10 Chitty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-10 Eriksen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 Erik-Nes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-17 Goldstein - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-19 Hadlock - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-20 Hansmann - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-27 Hellman - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-35 Hill - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-35 Hohler - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-36 Jeanty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-36 JSUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-48 Kurtz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-52 Mayden - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-53 Mercer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-54
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual TOC-1 Direction 5116269-100 Rev. 2
Merz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-55 Moore - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-65 Nelson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-65 Osaka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-66 Paris - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-70 Rempen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-73 Robinson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-78 Tokyo - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-78 Tokyo Shinozuka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-82 Williams - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-88 Yarkoni - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-88
TOC-2 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2

Chapter 1

Acoustic Output

Provides Acoustic Output information and tables, as well as possible bioeffects and prudent use.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-1 Direction 5116269-100 Rev. 2
Acoustic Output

Bioeffects

Concerns Surrounding the Use of Diagnostic Ultrasound

During a diagnostic ultrasound examination, high frequency sound penetrates and interacts with tissue in and around the area of anatomy to be imaged. Only a small portion of this sound energy is reflected back to the probe for use in constructing the image while the remainder is dissipated within the tissue. The interaction of sound energy with tissue at sufficiently high levels can produce biological effects (aka bioeffects) of either a mechanical or thermal nature. Although the generation of bioeffect is intentional with therapeutic ultrasound, it is generally undesired in diagnostic applications and may be harmful in some conditions.
NOTE: The American Institute of Ultrasound in Medicine has published
a document entitled "Medical Ultrasound Safety". This three part document covers Bioeffects and Biophysics, Prudent Use and Implementing ALARA.
Ultrasound users should read the AIUM to become more familiar with Ultrasound safety.
NOTE: Only for U.S.A. : A copy of this document is included as part of
the documentation package (Document 2163920-100).
To contact the AIUM concerning their publications:
In the USA, by telephone at 1-800-638-5352.
To write them, use the following address: AIUM 14750 Sweitzer Lane Suite 100 Laurel, MD, USA 20707-5906
1-2 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2

Maximum output summary

The following tables list the typical maimum acoustic output levels achievalbe with LOGIQ 7/LOGIQ 7 Pro for all probes and operational modes. It is intended that this is information be useful in making ALARA decisions and selecting the most appropriate probe for the application . In ac co rd an ce with US FDA Guidelines, the overall maximum acoustic SPTA intensity for LOGIQ 7/LOGIQ 7 Pro is limited to 720 mW/cm limited to 1.9. Modes for which TI does not exceed 0.1 are indicated by <0.1.
The accuracy of acoustic output table of this manual : +0%/-
50%
Accuracy of TI and MI is ±50%.

Acoustic Output Data

Acoustic Output Data
2
and MI is
Operating Mode
B-mode
M-mode
(inc.B-mode)
Color Flow (inc.
B-mode)
Pulsed Doppler
(inc.B-mode)
B-Flow n/a
Contrast n/a
CW Doppler n/a n/a
Other
(specify)
*LOGIQ 7 Only
3C 3.5C 3.5CS 5C E8C BE9C 8C M7C* 7L 10L
YES YES YES YES <1 <1 <1 YES <1 YES YES YES YES YES <1 <1 <1 YES <1 YES
YES YES YES YES YES YES YES YES YES YES
YES YES YES YES YES YES YES YES YES YES
n/a n/a
Table 1-1 shows all of the probes’ operating modes. The following pages contain probe acoustic outpu t tables for those noted in Table 1-1.
Table 1-1: Summary Table
Transducer Modell
YES YES YES YES
n/a n/a
n/a n/a
n/a n/a
n/a n/a
n/a n/a
n/a n/a n/a n/a n/a n/a n/a n/a
n/a
n/a
n/a
n/a
YES YES YES YES
n/a n/a n/a n/a
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-3 Direction 5116269-100 Rev. 2
Acoustic Output
Tab le 1- 2: Summary Table
Transducer Modell
Operating Mode
B-mode
M-mode
(inc.B-mode)
Color Flow (inc. B-
mode)
Pulsed Doppler
(inc.B-mode)
B-Flow
Contrast
CW Doppler
Other
(specify)
*LOGIQ 7 Only
M12L i12L T739 3S 7S 10S 6T M3S* P2D P6D
YES YES YES YES YES YES YES <1 YES YES YES YES <1 YES YES <1
YES YES YES YES YES YES YES YES
YES YES YES YES YES <1 YES YES
YES
n/a n/a n/a
n/a
n/a
n/a
n/a
YES
n/a n/a YES YES YES YES YES YES n/a
n/a
YES
n/a
n/a n/a
n/a
n/a
n/a
n/a
n/a n/a n/a YES
n/a n/a
n/a n/a
n/a n/a
n/a n/a
n/a n/a
n/a n/a
n/a n/a
<1
n/a n/a
1-4 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Acoustic Output Data
List of Acoustic Output Parameters
Table 1-3: List of Acoustic Output Parameters
FDA IEC Meaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3
a α Acoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz) A
aprt
D
eq
d
-6
d
eq
f
c
I
pa
I
pa.3
PII I PII
.3
I
TA
(Z) I
I
TA.3
I
(Z) I
SPTA
I
(Z) I
SPTA.3
A
arpt
C
MI
D
eq
d
-6
d
eq
ƒ
awf
I
pa
I
pa,α pi
I
pi,α
Ita(z) Temporal-Average Intensity
(z) Attenuated Temporal-Average Intensity / (at depth z)
ta,α
(z) Spatial-Peak Temporal-Average Intensity
zpta
(z) Attenuated Spatial-Peak Temporal-Average Intensity
zpta,α
MI MI Mechanical Index W
o
W
(Z) P
.3
W
o1
PII p p
r
p
r.3
P Output Power / Time average acoustic power at the source
α
P
1
i
p
r
p
rα
PRF prr Pulse Repetition Rate / Pulse repetition frequency TI TI Thermal Index / (same) TIB TIB Bone Thermal Index / (same) TIC TIC Cranial-Bone Thermal Index / (same) TIS TIS Soft-Tissue Thermal Index / (same) PD t
x
-12,y-12
d
X, Y -12 dB Output Beam Dimensions / (same)
Z Z Distance from the Source to a Specified Point / (same) Z
sp
Z
bp
Z
sp
Z
b
Z
bp
Z
s
-12db Output Beam Area / Active aperture area Normalizing Coefficient Equivalent Aperture Diameter / (same) Pulse Beam Width / Beam diameter at –6 dB Equivalent Beam Diameter Acoustic Working Frequency / Center frequency Pulse-Average Intensity Attenuated Pulse-Average Intensity Pulse-Intensity Integral Attenuated Pulse-Intensity Integral
Attenuated Output Power / Time average acoustic power derated to depth z Bounded Output Power / Power emitted from the central 1cm of aperture Pulse Pressure Squared Integral / Pulse intensity integral Peak-Rarefactional Acoustic Pressure / (same) Attenuated Peak-Rarefactional Acoustic Pressure / (same)
Pulse Duration / (same)
Depth for TIB / Depth at which the relevant index is maximum Break-Point Depth / (same) Depth for TIS / Depth at whcih the relevant index is maximum
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-5 Direction 5116269-100 Rev. 2
Acoustic Output
3C Probe
Table 1-4: Transducer Mode: 3C Operating Mode: B-Mode
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 1.4 - - - 2.7
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) 144.7 - -144.7
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.3
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.6 2.6 - - - 2.6
X(cm) 1.4 - - - 1.4 Y(cm)
PD (µsec) 1.0
1.0 - - - 1.0
prr PRF (Hz) 2927.4 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) 3.5 - - 3.5
I
at max. MI I
pa,α
PA.3
(MPa) 3.2
max
(cm) -
max
FL
(cm) 7.0 - - 7.0
Y
2
@MI
max
(W/cm
447.4
) Frequency (MHz) T4.0 T4.0 - - - T4.0 Image Depth (cm) 24.0 10.0 - - - 10.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-6 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
3C Probe (continued)
Table 1-5: Transducer Mode: 3C Operating Mode: M-Mode (inc.B-mode)
Acoustic Output Data
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 1.4 - <1 1.6 2.4
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) 144.7 - 33.5 126.1
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.3
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.3
deq(zb) deq(zsp) (cm) 0.4
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.6 2.6 - # 2.6 2.6
X(cm) 1.4 - # 1.4 1.4 Y(cm)
PD (µsec) 1.0
1.0 - # 1.0 1.0
prr PRF (Hz) 2927.4 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) 3.5 - # 3.5
I
at max. MI I
pa,α
PA.3
(MPa) 3.2
max
(cm) 0.4
max
FL
(cm) 7.0 - # 7.0
Y
2
@MI
max
(W/cm
448.9
) Frequency (MHz) T4.0 T4.0 - # T4.0 T4.0 Image Depth (cm) 24.0 10.0 - # 24.0 24.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-7 Direction 5116269-100 Rev. 2
Acoustic Output
3C Probe (continued)
Table 1-6: Transducer Mode: 3C Operating Mode: CF-Mode (inch.M-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC
non-
scan
Global Maximum: Index Value 1.6 1.4 1.1 - 3.2 2.5
IEC FDA Units
pra p P W min of [P
α
(zs, I
(MPa) 1.9
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA.3
(mW) 144.7 90.3 90.3 152.8
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) 3.3
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.3
deq(zb) deq(zsp) (cm) 0.4
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.6 2.6 2.5 - 2.6 2.6
X(cm) 1.4 0.9 - 1.4 1.9 Y(cm)
PD (µsec) 1.0
1.0 1.0 - 1.0 1.0
prr PRF (Hz) 635.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 3.3
max
(cm) 0.4
max
(cm) 3.5 2.2 - 4.8
X
FL
(cm) 7.0 7.0 - 3.5
Y
2
@MI
max
(W/cm
447.7
) Frequency (MHz) D2.5 D2.5 D2.5 - D2.5 D2.5 Image Depth (cm) 3.5 4.8 3.5 - 3.5 4.8 Vel Scale (kHz) 1.80 4.90 1.1 - 1.07 4.90 Penet On On On - On On ROI 60deg 5deg - - - 5deg
Operating Control
Conditions
PS 14 14 16.0 - 16 14
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-8 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Acoustic Output Data
3C Probe (continued)
Table 1-7: Transducer Mode: 3C Operating Mode: PWD-Mode (inch.B-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.6 1.4 1.1 - 3.2 2.2
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.5
r.3
(mW) 144.7 90.3 90.3 103.9
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) 2.6
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.3
deq(zb) deq(zsp) (cm) 0.5
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.6 2.6 2.5 - 2.5 2.5
X(cm) 1.4 0.9 - 0.9 0.9 Y(cm)
PD (µsec) 1.0
1.0 1.0 - 1.0 1.0
prr PRF (Hz) 1796.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 3.3
max
(cm) 0.5
max
(cm) 3.5 2.2 - 2.2
X
FL
(cm) 7.0 7.0 - 7.0
Y
2
@MI
max
(W/cm
447.4
) Frequency (MHz) D2.5 T4.0 D2.5 - D2.5 D2.5 Image Depth (cm) 3.5 10.0 2.2 - 2.2 2.2 Vel Scale (kHz) 0.64 - 1.23 - 1.23 1.23 SV 3 - 6 - 6 6 Penet on - on - on on
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-9 Direction 5116269-100 Rev. 2
Acoustic Output
3.5C Probe
Table 1-8: Transducer Mode: 3.5C Operating Mode: B-Mode
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 <1 - - - 2.3
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) #- -216.5
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.6
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.4 # - - - 2.0
X(cm) #- - -3.3 Y(cm)
PD (µsec) 0.7
#- - -1.3
prr PRF (Hz) 2383.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) #- - 4.8
I
at max. MI I
pa,α
PA.3
(MPa) 3.2
max
(cm) -
max
FL
(cm) #- - 6.6
Y
2
@MI
max
(W/cm
288.4
) Frequency (MHz) T5.0 # - - - T4.0 Image Depth (cm) 30.0 # - - - 8.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-10 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
3.5C Probe (continued)
Table 1-9: Transducer Mode: 3.5C Operating Mode: M-Mode (inc.B-mode)
Acoustic Output Data
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 < 1 - <1 1.0 2.0
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) #- 34.8 186.2
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.6
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.6
deq(zb) deq(zsp) (cm) 0.6
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.4 # - # 2.4 2.0
X(cm) #- #2.43.3 Y(cm)
PD (µsec) 0.7
#- #1.31.3
prr PRF (Hz) 2383.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) #- # 4.8
I
at max. MI I
pa,α
PA.3
(MPa) 3.2
max
(cm) 0.6
max
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
288.4
) Frequency (MHz) T5.0 # - # T5.0 T5.0 Image Depth (cm) 30.0 # - # 30.0 30.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-11 Direction 5116269-100 Rev. 2
Acoustic Output
3.5C Probe (continued)
Table 1-10: Transducer Mode: 3.5C Operating Mode: CF-Mode (inch.M-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC
non-
scan
Global Maximum: Index Value 1.7 <1 - <1 2.1 2.3
IEC FDA Units
pra p P W min of [P
α
(zs, I
(MPa) 1.8
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA.3
(mW) #- 31.5 214.2
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 5.0
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 5.0
deq(zb) deq(zsp) (cm) 0.5
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.1 # - # 3.1 2.0
X(cm) #- #3.33.3 Y(cm)
PD (µsec) 0.8
#- #1.31.3
prr PRF (Hz) 635.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 4.5
max
(cm) 0.5
max
(cm) #- # 4.8
X
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
452.8
) Frequency (MHz) D3.3 # - # D3.3 D2.0 Image Depth (cm) 4.8 # - # 4.8 4.8 Vel Scale (kHz) 1.80 # - # 1.23 7.69 Penet Off # - # Off On ROI 70deg # - # - 5deg
Operating Control
Conditions
PS 12 # - # 16 12
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-12 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Acoustic Output Data
3.5C Probe (continued)
Table 1-11: Transducer Mode: 3.5C Operating Mode: PWD-Mode (inch.B-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.7 <1 - <1 2.1 <1
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 3.0
r.3
(mW) #- 31.5 #
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.7
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 5.0
deq(zb) deq(zsp) (cm) 0.3
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.1 # - # 2.0 #
X(cm) #- #2.4# Y(cm)
PD (µsec) 0.8
#- #1.3#
prr PRF (Hz) 1796.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 4.5
max
(cm) 0.3
max
(cm) #- # #
X
FL
(cm) #- # #
Y
2
@MI
max
(W/cm
452.3
) Frequency (MHz) D3.3 # - # D2.0 # Image Depth (cm) 4.8 # - # 3.5 # Vel Scale (kHz) 0.64 - - # 1.23 # SV 1 - - # 4 # Penet Off - - # On #
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-13 Direction 5116269-100 Rev. 2
Acoustic Output
3.5C Probe (continued)
Table 1-12: Transducer Mode: 3.5C Operating Mode: Contrast
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 2.0 - - - 1.5
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.7
r.3
(mW) 141.8 - - 141.8
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.5
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.2 3.2 - - - 3.2
X(cm) 3.3 - - - 3.3 Y(cm)
PD (µsec) 0.4
1.3 - - - 1.3
prr PRF (Hz) 2383.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
Frequency (MHz) CHA
(MPa) 4.1
max
(cm) -
max
(cm) 4.8 - - 4.8
X
FL
(cm) 6.6 - - 6.6
Y
2
@MI
max
(W/cm
337.0
)
3.0
CHA
3.0
---CHA
3.0
Image Depth (cm) 30.0 8.0 - - - 8.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-14 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
3.5C Probe (continued)
Table 1-13: Transducer Mode: 3.5C Operating Mode: B-Flow
Acoustic Output Data
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.7 <1 - <1 2.1 1.8
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.8
r.3
(mW) #- 31.5 162.4
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.7
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 2.5
deq(zb) deq(zsp) (cm) 0.3
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.7 # - # 2.0 2.0
X(cm) #- #2.42.5 Y(cm)
PD (µsec) 1.9
#- #1.31.3
prr PRF (Hz) 300.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 3.4
max
(cm) 03
max
(cm) #- # 5.8
X
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
173.6
) Frequency (MHz) BF3.3 # - # D2.0 T4.0 Image Depth (cm) 25.0 # - # 3.5 8.0 Vel Scale (kHz) - # - # 1.23 ­SV - # - # 4 ­Penet - # - # On -
Operating Control
Conditions
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-15 Direction 5116269-100 Rev. 2
Acoustic Output
3.5CS Probe
Table 1-14: Transducer Mode: 3.5CS Operating Mode: B-Mode
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.2 <1 - - - 2.6
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 1.9
r.3
(mW) #- -189.1
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.2
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.5 # - - - 2.0
X(cm) #- - -2.0 Y(cm)
PD (µsec) 0.8
#- - -1.3
prr PRF (Hz) 2290.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) #- - 4.8
I
at max. MI I
pa,α
PA.3
(MPa) 2.6
max
(cm) -
max
FL
(cm) #- - 6.6
Y
2
@MI
max
(W/cm
144.6
) Frequency (MHz) T5.0 # - - - T4.0 Image Depth (cm) 30.0 # - - - 8.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-16 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Acoustic Output Data
3.5CS Probe (continued)
Table 1-15: Transducer Mode: 3.5CS Operating Mode: M-Mode (inc.B-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.2 <1 - <1 <1 2.3
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 1.9
r.3
(mW) #- #164.3
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) #
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.2
deq(zb) deq(zsp) (cm) #
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.5 # - # # 2.0
X(cm) #- ##2.0 Y(cm)
PD (µsec) 0.8
#- ##1.3
prr PRF (Hz) 2290.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) #- # 4.8
I
at max. MI I
pa,α
PA.3
(MPa) 2.6
max
(cm) #
max
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
144.6
) Frequency (MHz) T5.0 # - # # T5.0 Image Depth (cm) 30.0 # - # # 30.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-17 Direction 5116269-100 Rev. 2
Acoustic Output
3.5CS Probe (continued)
Table 1-16: Transducer Mode: 3.5CS Operating Mode: CF-Mode (inch.M-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC
non-
scan
Global Maximum: Index Value 1.2 <1 - <1 1.6 1.7
IEC FDA Units
pra p P W min of [P
α
(zs, I
(MPa) 1.9
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA.3
(mW) #- 50.1 119.6
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 4.3
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.2
deq(zb) deq(zsp) (cm) 0.6
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.5 # - # 3.3 2.0
X(cm) #- #2.02.0 Y(cm)
PD (µsec) 0.8
#- #1.31.3
prr PRF (Hz) 2290.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 2.6
max
(cm) 0.6
max
(cm) #- # 4.8
X
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
144.6
) Frequency (MHz) D3.3 # - # D3.3 D2.0 Image Depth (cm) 4.8 # - # 4.8 4.8 Vel Scale (kHz) 1.80 # - # 1.23 7.69 Penet Off # - # Off On ROI 70deg # - # - 5deg
Operating Control
Conditions
PS 12 # - # 16 12
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-18 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Acoustic Output Data
3.5CS Probe (continued)
Table 1-17: Transducer Mode: 3.5CS Operating Mode: PWD-Mode (inch.B-mo de)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.2 <1 - <1 1.6 1.3
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 1.9
r.3
(mW) #- 50.1 84.8
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.4
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.2
deq(zb) deq(zsp) (cm) 0.7
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.5 # - # 2.0 2.0
X(cm) #- #1.41.4 Y(cm)
PD (µsec) 0.8
#- #1.31.3
prr PRF (Hz) 2290.0 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 2.6
max
(cm) 0.7
max
(cm) #- # 4.8
X
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
144.6
) Frequency (MHz) D3.3 # - # D2.0 D2.0 Image Depth (cm) 4.8 # - # 3.5 3.5 Vel Scale (kHz) 0.64 - - # 1.23 1.23 SV 1 - - # 4 4 Penet Off - - # On On
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-19 Direction 5116269-100 Rev. 2
Acoustic Output
3.5CS Probe (continued)
Table 1-18: Transducer Mode: 3.5CS Operating Mode: Contrast
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.5 1.3 - - - 2.4
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.2
r.3
(mW) 193.5 - - 193.5
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 4.5
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.1 2.1 - - - 2.1
X(cm) 2.5 - - - 2.5 Y(cm)
PD (µsec) 1.1
1.3 - - - 1.3
prr PRF (Hz) 2288.9 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
Frequency (MHz) CHA
(MPa) 3.2
max
(cm) -
max
(cm) 5.1 - - 5.1
X
FL
(cm) 6.6 - - 6.6
Y
2
@MI
max
(W/cm
288.3
)
3.0
CHA
3.0
---CHA
3.0
Image Depth (cm) 30.0 6.0 - - - 6.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-20 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
3.5CS Probe (continued)
Tab le 1-19: Transducer Mode: 3.5CS Operating Mode: B-Flow
Acoustic Output Data
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.7 <1 - <1 1.6 1.1
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.8
r.3
(mW) #- 50.1 81.3
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.4
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 4.5
deq(zb) deq(zsp) (cm) 0.7
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 2.7 # - # 2.0 2.7
X(cm) #- #1.42.0 Y(cm)
PD (µsec) 1.9
#- #1.31.3
prr PRF (Hz) 2285.2 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 3.9
max
(cm) 0.7
max
(cm) #- # 4.5
X
FL
(cm) #- # 6.6
Y
2
@MI
max
(W/cm
282.4
) Frequency (MHz) BF3.3 # - # D2.0 BF3.3 Image Depth (cm) 30.0 # - # 3.5 5.0 Vel Scale (kHz) - # - # 1.23 ­SV - # - # 4 ­Penet - # - # On -
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-21 Direction 5116269-100 Rev. 2
Acoustic Output
5C Probe
Table 1-20: Transducer Mode: 5C Operating Mode: B-Mode
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.3 1.2 - - - 1.6
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) 76.9 - -76.9
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) -
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 2.9
deq(zb) deq(zsp) (cm) -
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.5 3.5 - - - 3.5
X(cm) 1.3 - - - 1.3 Y(cm)
PD (µsec) 0.5
0.9 - - - 0.9
prr PRF (Hz) 2927.4 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) 4.5 - - 4.5
I
at max. MI I
pa,α
PA.3
(MPa) 3.1
max
(cm) -
max
FL
(cm) 6.0 - - 6.0
Y
2
@MI
max
(W/cm
645.8
) Frequency (MHz) 4.0 4.0 - - - 4.0 Image Depth (cm) 24.0 15.0 - - - 15.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-22 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
5C Probe (continued)
Table 1-21: Transducer Mode: 5C Operating Mode: M-Mode (inc.B-mode)
Acoustic Output Data
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC non­scan
Global Maximum: Index Value 1.3 1.2 <1 - <1 1.4
IEC FDA Units
pra p P W
min of [P
(zs, I
α
(MPa) 2.4
r.3
(mW) 76.0 # #66.7
.3(Z1),ITA.3
ta,α(zs
o
)] [(W
(z1)]) -
zs z1(cm) ­zbp zbp(cm) ­zb zsp(cm) #
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 2.9
deq(zb) deq(zsp) (cm) #
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.5 3.5 # - # 3.5
X(cm) 1.3 # - # 1.3 Y(cm)
PD (µsec) 0.5
0.9 # - # 0.9
prr PRF (Hz) 2927.4 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FLX(cm) 4.5 # - 4.5
I
at max. MI I
pa,α
PA.3
(MPa) 3.1
max
(cm) #
max
FL
(cm) 6.0 # - 6.0
Y
2
@MI
max
(W/cm
645.8
) Frequency (MHz) 4.0 4.0 # - # 4.0 Image Depth (cm) 24.0 15.0 # - # 24.0
Operating Control
Conditions
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual 1-23 Direction 5116269-100 Rev. 2
Acoustic Output
5C Probe (continued)
Table 1-22: Transducer Mode: 5C Operating Mode: CF-Mode (inch.M-mode)
Index Label MI
scan
TIS TIB
non-scan
Aaprt< Aaprt>1
TIC
non-
scan
Global Maximum: Index Value 1.8 1.2 - <1 1.6 1.2
IEC FDA Units
pra p P W min of [P
α
(zs, I
(MPa) 3.2
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA.3
(mW) 76.9 - 37.8 51.4
(z1)]) #
zs z1(cm) # zbp zbp(cm) # zb zsp(cm) 3.1
Assoc Acoustic Paramete
z at max. Ipi,α z
sp
(cm) 3.3
deq(zb) deq(zsp) (cm) 0.6
ƒ
awf
Dim of A
t
d
aprt
f
c
(MHz) 3.9 3.5 - # 3.9 4.0
X(cm) 1.3 - # 1.4 0.9 Y(cm)
PD (µsec) 0.6
0.9 - # 0.9 0.9
prr PRF (Hz) 635.3 pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Other infromation
Focal Length FL
I
at max. MI I
pa,α
PA.3
(MPa) 4.8
max
(cm) 0.6
max
(cm) 4.5 - # 2.1
X
FL
(cm) 6.0 - # 6.0
Y
2
@MI
max
(W/cm
460.1
) Frequency (MHz) D4.0 D4.0 - # D4.0 D4.0 Image Depth (cm) 3.3 2.1 - # 3.3 2.1 Vel Scale (kHz) 1.80 8.53 - # 2.13 8.53 Penet On On - # On On ROI 60deg 10deg - # - 10deg
Operating Control
Conditions
PS 14 14 - # 16 14
NOTE:
a. This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3) b. This probe is not intended for transcranial or neonatal cephalic uses. c. This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
1-24 LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
Loading...
+ 214 hidden pages