Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on GPC (GE Medical Systems Global
Product Configuration). If you need to know the latest revision, contact your
distributor, local GE Sales Representative or in the USA call the GE Ultrasound
Clinical Answer Center at 1-800-682-5327 or 414-524-5698.
This product complies with the regulatory requirements of the
following:
S
Council Directive 93/42/EEC concerning medical devices:
.
For US
Only
the
to the Directive.
The location of the CE marking is shown on 2–23 of this
manual.
European registered place of business:
GE Medical Systems Europe
Quality Assurance Manager
BP 34
F 78533 BUC CEDEX France
Tel: +33 (0)1 30 70 40 40
S
Medical Device Good Manufacturing Practice Manual
issued by the FDA (Food and Drug Administration,
Department of Health, USA).
S
Underwriters’ Laboratories, Inc. (UL), an independent
testing laboratory.
S
Canadian Standards Association (CSA).
S
International Electrotechnical Commission (IEC),
international standards organizations, when applicable.
Caution: United States law restricts this device to sale or use by
or on the order of a physician.
S
General Electric Medical Systems
EN 46001 certified.
label affixed to the product testifies compliance
is ISO 9001 and
S
The original document was written in English.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Regulatory Req 1
Regulatory Requirements
NOTE: This equipment generates, uses and can radiate radio frequency energy. The
equipment may cause radio frequency interference to other medical and
non-medical devices and radio communications. To provide reasonable
protection against such interference, this product complies with emissions limits
for a Group 1, Class A Medical Devices Directive as stated in EN 60601–1–2.
However, there is no guarantee that interference will not occur in a particular
installation.
NOTE: If this equipment is found to cause interference (which may be determined by
turning the equipment on and off), the user (or qualified service personnel)
should attempt to correct the problem by one or more of the following
measure(s):
–reorient or relocate the affected device(s)
–increase the separation between the equipment and the affected device
–power the equipment from a source different from that of the affected device
–consult the point of purchase or service representative for further
suggestions
NOTE: The manufacturer is not responsible for any interference caused by using other
than recommended interconnect cables or by unauthorized changes or
modifications to this equipment. Unauthorized changes or modifications could
void the users’ authority to operate the equipment.
NOTE: To comply with the regulations on electromagnetic interference for a Class A
FCC Device, all interconnect cables to peripheral devices must be shielded and
properly grounded. Use of cables not properly shielded and grounded may
result in the equipment causing radio frequency interference in violation of the
FCC regulations.
NOTE: Do not use devices which intentionally transmit RF Signals (cellular phones,
transceivers, or radio controlled products) in the vicinity of the equipment as it
may cause performance outside the published specifications. Keep the power
to these type devices turned off when near this equipment.
The medical staff in charge of this equipment is required to instruct technicians,
patients, and other people who may be around this equipment to fully comply
with the above requirement.
This manual contains enough information to operate the system
safely. Advanced equipment training may be provided by a
factory trained Applications Specialist for the agreed upon time
period.
Read and understand all instructions in this manual before
attempting to use the LOGIQ 500 system.
Keep this manual with the equipment at all times. Periodically
review the procedures for operation and safety precautions.
LOGIQ 500 Documentation consists of three manuals:
S
The Quick Start Guide (TRANSLATED) provides a
step-by-step description of the basic features and operation
of the LOGIQ 500. It is intended to be used in
conjunction with the Basic User Manual in order to provide
the information necessary to operate the system safely.
S
The Basic User Manual (TRANSLATED) provides
information needed by the user to operate the system
safely. It describes basic functions of the system, safety
features, operating modes, basic measurements/
calculations, probes, user care and maintenance.
S
The Advanced Reference Manual (ENGLISH ONLY) is
intended for the trained, professional user. It contains all
the information found in the Quick Start Guide and Basic
User Manual, as well as information on options, advanced
customization techniques and data tables.
The LOGIQ 500 manuals are written for users who are
familiar with basic ultrasound principals and techniques. They
do not include sonography training or clinical procedures.
1–2
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Physical Principle Used
The transmission and reception of mechanical high frequency
waves through a transducer associated with a computer that
creates the image in a digital memory, are used for the creation
of medical ultrasound images. The spreading of mechanical
ultrasound waves produces echoes when body changes
density. In the case of human tissue, these echoes are created
when the signal goes from an adipose tissue (fat) region to a
muscular tissue region, among others. The echoes are
returned through the same transducer that converts them back
into electrical signals. These signals are highly amplified,
processed by filters with several frequency and time response
options, and finally scanned and stored in a digital memory.
Once in the memory, the image can be displayed in real-time on
a monitor. Several analog and digital circuits transform the
electrical high frequency signals into a flow of digital signals,
allowing the composition of the image in the memory. All the
signal reception and transmission parameters are controlled by
the main computer. Through the selection of these parameters
by the operator, the system modifies the transmission and
reception features allowing a wide range of uses, from
obstetrics to peripheral vascular examinations. As its design is
based on solid state components, the system is free from
variations over time and requires very little maintenance. All
the transducers are accurate solid state devices, allowing
control of creation of images from convex, micro-convex and
linear transducers. The use of a solid state design allows a
wide range of sweep parameters that can be optimized
resulting in a consistent creation of fine anatomical details with
excellent penetration and dynamic contrast band in the tissue.
The system features a sophisticated design, providing multiple
functions of diagnostic and function setup keys. This makes the
system user-friendly and easy to use.
System Overview
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
1–3
System Overview
General Indications for Use
The LOGIQ 500 is a general purpose ultrasound imaging
system intended for use in the dynamic evaluation of soft tissue
and vascular diseases in the following areas:
S
Head
S
Neck
S
Chest
S
Abdomen
S
Pelvis
S
Male reproductive organs
S
Female reproductive organs
S
Limbs/Extremities
S
Pregnant uterus
S
Cardiac
Indications for Fetal Doppler use
The LOGIQ 500 system can be used for fetal examination in
Pulsed Wave Doppler, Continuous Wave Doppler, Color Flow
Doppler, and Color M-Mode for the diagnosis of:
S
Structural fetal cardiac anomalies for high-risk patients.
S
Intrauterine growth retardation (IUGR) for high-risk patients
with one or more of the following known or suspected
conditions:
S
Multiple pregnancy
S
Maternal hypertension
S
Hydrops
S
Diabetes
S
Lupus
S
Placenta abnormality
1–4
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Contraindications
A
System Overview
The system is NOT intended for use in the following areas:
Ophthalmic use (or any use causing the acoustic beam to pass
through the eye).
.
Prescription Device
.
For US
Only
Pulsed Wave Doppler, Continuous Wave Doppler, Color Flow
Doppler, and Color M-Mode are not intended for routine fetal
examination or screening nor are they intended for fetal
examination in a low-risk population. The use of Doppler, even
at minimal output levels, in fetal examination must be adjunctive
with conventional fetal echocardiography and other clinical
diagnostic methods, for high risk patients only.
Caution: United States law restricts this device to sale or use by
or on the order of a physician.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
1–5
Who To Contact
Who To Contact
Contacting GE Medical Systems—Ultrasound
For additional information or assistance, please contact your
local distributor or the appropriate support resource listed
below:
USA
GE Medical SystemsTEL: (1) 800–437–1171
Ultrasound Service EngineeringFAX: (1) 414–647–4090
4855 W. Electric Avenue
Milwaukee, WI 53219
Customer Answer CenterTEL: (1) 800–682–5327
CANADA
GE Medical SystemsTEL: (1) 800–664–0732
Ultrasound Service Engineering
4855 W. Electric Avenue
Milwaukee, WI 53219
(1) 414–524–5698
Customer Answer CenterTEL: (1) 414–524–5698
LATIN & SOUTH AMERICA
GE Medical SystemsTEL: (1) 305–735–2304
Ultrasound Service Engineering
4855 W. Electric Avenue
Milwaukee, WI 53219
The LOGIQ 500 Basic User Manual is organized to provide
the information needed to start scanning right away. Detailed
information is also provided for more time-intensive studies.
S
Getting started. These sections give an overview of the
system to help the operator start scanning as soon as
possible.
S
Introduction.
contraindications for use, who to contact and how this
documentation is organized.
S
Safety
operation of the LOGIQ 500 system.
S
Preparing the System for Use.
system for use and a map of the control layout.
Information concerning indications/
. Important information concerning the safe
How to prepare the
S
Preparing for an Exam.
information, select an exam category and application
preset.
S
Image optimization. These sections detail how to improve
image, trace, or spectral information.
S
Modes
. How to adjust and optimize B-Mode, Color
Flow, Doppler and M-Mode imaging.
S
Scanning and Display Functions.
concerning Zoom, Freeze, Cine and Annotation
functions.
How to enter patient
Information
1–10
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Manual Content (cont’d)
S
S
S
How This Book is Organized
Measurements and Reports. Shows how to do general
and exam category specific measurements and
calculations.
S
General Measurements and Calculations
on basic measurements for each mode.
S
Exam Categories.
S
Abdomen and Small Parts.
S
OB/GYN
S
Cardiology.
S
Vascular
S
Urology.
Recording Images. Explains the use of image archive and
peripheral options.
Customizing your system. Shows how to customize the
system for your particular institution, clinic, or exam type.
.
.
. Emphasis
S
Probes and Biopsy. Provides intended uses,
specifications, care and maintenance, and biopsy capability
instructions for each probe.
S
User Maintenance. Provides information concerning
system specifications, error messages, user diagnostics,
quality assurance, system care and assistance.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
1–11
How This Book is Organized
Manual Format
Information has been arranged and provided to help find
information easily and quickly.
Finding information
Tables of ContentsLocate topics in the main table of contents.
TabsChapter tabs are provided.
Headers/FootersThe section name and page number appear on the outer
corners of every page.
ReferencesSee also page references that are noted.
IndexMeant for frequent and easy reference. Extensive tool that
presents ideas, topics, terms, titles, headings, and cross
references. Also, use it to find all entries of a like topic
throughout the manual.
Text References
NotesNotes are set in
.
ReferencesReferences to other chapters appear in
IconsVarious icons highlight safety issues.
Hints
Indicates precautions or prudent use recommendations that
should be used in the operation of the ultrasound system.
Scanning hints help save time.
italics
.
W ARNINGDANGER
CAUTION
italics
.
1–12
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Safety Precautions
Precaution Levels
Icon Description
Safety Precautions
Various levels of safety precautions may be found on the
equipment and different levels of concern are identified by one
of the following flag words which precede the precautionary
statement.
DANGER
W ARNING
CAUTION
.
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions will cause:
S
Severe or fatal personal injury
S
Substantial property damage.
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions may cause:
S
Severe personal injury
S
Substantial property damage.
Indicates that a potential hazard may exist which through
inappropriate conditions or actions will or can cause:
S
Minor injury
S
Property damage.
Indicates precautions or prudent use recommendations that
should be used in the operation of the ultrasound system,
specifically:
S
Use of the ultrasound system as a prescription device,
under the order of a physician
2–2
S
Maintaining an optimum system environment
S
Using this Manual
S
Notes to emphasize or clarify a point.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Hazard Symbols
Icon Description
IconPotential HazardUsageSource
Biological
Hazard
Electrical
Hazard
Moving
Hazard
Acoustic
Output
Hazard
Safety Precautions
Potential hazards are indicated by the following icons:
S
Patient/user infection due
to contaminated
equipment.
S
Electrical micro-shock to
patient, e.g., ventricular
fibrillation initiated.
S
Electrical micro-shock to
patient/user.
S
Console, accessories or
optional storage devices
fall on patient, user, or
others.
S
Collision with persons or
objects results in injury while
maneuvering or during
system transport.
S
Injury to user from moving
the console.
S
Patient injury or tissue
damage from ultrasound
radiation.
S
Cleaning and care
instructions
S
Sheath and glove
guidelines
S
Probes
S
ECG
S
Connections to back panel
S
Moving
S
Using brakes
S
Transporting
S
ALARA, the use of acoustic
output following the as low
a
s reasonably achievable
principle
ISO 7000
No. 0659
Explosion
S
Hazard
Smoke
S
& Fire
Hazard
Non–
S
S
Ionizing
Radiation
Table 2–1. Potential Hazards
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Risk of explosion if used in
the presence of flammable
anesthetics.
Patient/user injury or
adverse reaction from fire or
smoke.
Patient/user injury from
explosion and fire.
Console failure, erratic
operation or output error
due to RF interference.
S
Flammable anesthetic
S
Replacing fuses
S
Outlet guidelines
S
RF
IEC 878
No. 03-04
2–3
Safety Precautions
Important Safety Considerations
Patient Safety
Related Hazards
W ARNING
Patient
identification
Diagnostic
information
The following topic headings (
and Personnel Safety
user aware of particular hazards associated with the use of this
equipment and the extent to which injury can occur if
precautions are not observed. Additional precautions may be
provided throughout the manual. The equipment user is
obligated to be familiar with these concerns and avoid
conditions that could result in injury.
The concerns listed can seriously affect the safety of patients
undergoing a diagnostic ultrasound examination.
Always include proper identification with all patient data and
verify the accuracy of the patient’s name or ID numbers when
entering such data. Make sure correct patient ID is provided on
all recorded data and hard copy prints. Identification errors
could result in an incorrect diagnosis.
Equipment malfunction or incorrect settings can result in
measurement errors or failure to detect details within the image.
The equipment user must become thoroughly familiar with the
equipment operation in order to optimize its performance and
recognize possible malfunctions. Applications training is
available through the local GE representative. Added
confidence in the equipment operation can be gained by
establishing a quality assurance program.
) are intended to make the equipment
Patient Safety
, and
Equipment
2–4
Mechanical
hazards
Damaged probes or improper use and manipulation of
intracavitary probes can result in injury or increased risk of
infection. Inspect probes often for sharp, pointed, or rough
surface damage that could cause injury or tear protective
barriers. Never use excessive force when manipulating
intracavitary probes. Become familiar with all instructions and
precautions provided with special purpose probes.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Related Hazards (cont’d)
Safety Precautions
Electrical
Hazard
Acoustic
Output
Hazard
Training
A damaged probe can also increase the risk of electric shock if
conductive solutions come in contact with internal live parts.
Inspect probes often for cracks or openings in the housing and
holes in and around the acoustic lens or other damage that
could allow liquid entry. Become familiar with the probe’s use
and care precautions outlined in
Ultrasound energy, even at diagnostic levels, is capable of
damaging sensitive tissues if adequate precautions are not
followed. The wrong combination of equipment settings, probe
positioning, and tissue type can result in injury. Please become
thoroughly familiar with equipment controls that affect acoustic
output levels as well as the output display.
Follow the principle of a
(ALARA) when scanning a patient. During each ultrasound
examination, the clinical user is expected to weigh the medical
benefit of the diagnostic information obtained against the risk of
potential harmful effects. Once an optimal image is achieved
the need for increasing acoustic output or prolonging the
exposure cannot be justified.
It is recommended that all users receive proper training in
applications before performing them in a clinical setting. Please
contact the local GE representative for training assistance.
ALARA training is provided by GE Application Specialists.
s low as reasonably achievable
Probes and Biopsy
.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–5
Safety Precautions
Equipment and Personnel Safety
Related Hazards
WARNING
DANGER
Explosion
Electrical
Hazard
Hazard
This equipment contains dangerous voltages that are capable
of serious injury or death.
There are no user serviceable components inside the console.
Refer all servicing to qualified service personnel only.
The concerns listed below can seriously affect the safety of
equipment and personnel during a diagnostic ultrasound
examination.
Risk of explosion if used in the presence of flammable
anesthetics.
To avoid injury:
S
Do not remove protective covers. No user serviceable
parts are inside. Refer servicing to qualified service
personnel.
S
To assure adequate grounding, connect the attachment
plug to a reliable (hospital grade) grounding outlet (having
equalization conductor
S
Do not place liquids on or above the console. Spilled liquid
may contact live parts and increase the risk of shock.
).
2–6
Smoke
& Fire
Hazard
The system must be supplied from an adequately rated
electrical circuit. The capacity of the supply circuit must be as
specified on
3–3
.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Related Hazards (cont’d)
Safety Precautions
Biological
CAUTION
Hazard
For patient and personnel safety, beware of biological hazards
while performing invasive procedures. To avoid the risk of
disease transmission:
S
Use protective barriers (gloves and probe sheaths)
whenever possible. Follow sterile procedures when
appropriate.
S
Thoroughly clean probes and reusable accessories after
each patient examination and disinfect or sterilize as
needed. Refer to
care instructions.
S
Follow all infection control policies established by your
office, department or institution as they apply to personnel
and equipment.
Devices containing latex may cause severe allergic reaction in
latex sensitive individuals. USA customers should refer to the
FDA’s March 29, 1991 Medical Alert on latex products.
Probes and Biopsy
for probe use and
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–7
Safety Precautions
Device Labels
Label Icon Description
The following table describes the purpose and location of safety
labels and other important information provided on the
equipment.
Label/IconPurpose/MeaningLocation
Identification and Rating
Plate
Type/Class LabelUsed to indicate the degree of safety or
IP Code (IPX1)Indicates the degree of protection provided by
S
Manufacturer’s name and address
S
Date of manufacture
S
Model and serial numbers
S
Electrical ratings (Volts, Amps, phase, and
frequency)
protection.
the enclosure per IEC 529. IPX1 indicates drip
proof.
Equipment Type BF (man in the box symbol)
IEC 878-02-03 indicates B Type equipment
having a floating applied part.
Rear of console near power
inlet
Foot Switch
Probe connectors and PCG
connector
Device Listing/
Certification Labels
“DANGER – Risk of
explosion used in...”
Equipment Type CF (heart in the box symbol)
IEC 878-02-05 indicate equipment having a
floating applied part having a degree of
protection suitable for direct cardiac contact.
Laboratory logo or labels denoting
conformance with industry safety standards
such as UL or IEC.
The system is not designed for use with
flammable anesthetic gases.
“CAUTION” The equilateral triangle is usually
used in combination with other symbols to
advise or warn the user.
Table 2–2. Label Icons
ECG connector and surgical
probes
Rear of console
Various
2–8
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Label Icon Description (cont’d)
Label/IconPurpose/MeaningLocation
“ATTENTION – Consult accompanying
documents” is intended to alert the user to
refer to the operator manual or other
instructions when complete information cannot
be provided on the label.
“CAUTION – Dangerous voltage” (the lightning
flash with arrowhead) is used to indicate
electric shock hazards.
Safety Precautions
Various
“Mains OFF” Indicates the power off position of
the mains power switch.
“Mains ON” Indicates the power on position of
the mains power switch.
“ON” Indicates the power on position of the
power switch.
CAUTION: This Power Switch DOES NOT
ISOLATE Mains Supply.
“Off/Standby” Indicates the power off/standby
position of the power switch.
CAUTION: This Power Switch DOES NOT
ISOLATE Mains Supply.
“Equipotentiality” Indicates the terminal to be
used for connecting equipotential conductors
when interconnecting (grounding) with other
equipment.
CAUTION: This is only for ”FUNCTIONAL
GROUNDING”, NOT ”PROTECTIVE EARTH”.
Table 2–2. Label Icons (cont’d)
Rear of system,
adjacent to mains switch
Adjacent to On-Off/Standby
Switch
Rear of console
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–9
Safety Precautions
Classifications
*1. Class I EQUIPMENT
*2. Type BF EQUIPMENT
Type of protection against electric shock
Class I Equipment (*1)
Degree of protection against electric shock
Type BF Equipment (*2) (Except ECG)
Type CF Equipment (*3) (ECG Only)
Ordinary Equipment
Continuous Operation
EQUIPMENT in which protection against electric shock does
not rely on BASIC INSULATION only , but includes an earth
ground. This additional safety precaution prevents exposed
metal parts from becoming LIVE in the event of an insulation
failure.
TYPE B EQUIPMENT with an F-TYPE APPLIED P ART
TYPE B EQUIPMENT: EQUIPMENT providing a specified
degree of protection against electric shock, with particular
regard to allowable LEAKAGE CURRENT.
Patient leakage current
*3. Type CF EQUIPMENT
Patient leakage current
Normal ModeSingle fault condition
Less than 100 mALess than 500 mA
EQUIPMENT providing a degree of protection higher than that
for TYPE BF EQUIPMENT against electric shock particularly
regarding allowable LEAKAGE CURRENTS, and having an
F-TYPE APPLIED P ART.
Normal ModeSingle fault condition
Less than 10 mALess than 50 mA
2–10
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Classifications (cont’d)
*4. EMC (Electromagnetic Compatibility)
4.1 EMC Performance
All types of electronic equipment may characteristically cause
electromagnetic interference with other equipment, either
transmitted through air or connecting cables. The term EMC
(Electromagnetic Compatibility) indicates the capability of
equipment to curb electromagnetic influence from other
equipment and at the same time not affect other equipment with
similar electromagnetic radiation from itself.
This product is designed to fully comply with the EN60601–1–2
(IEC60601–1–2) in medical electric equipment EMC
regulations.
Proper installation following the service manual is required in
order to achieve the full EMC performance of the product.
The product must be installed as stipulated in 4.2, Notice upon
Installation of Product.
Safety Precautions
CAUTION
In case of issues related to EMC, please call your service
personnel.
Do not use the following devices near this equipment. Use of
these devices near this equipment could cause this equipment
to malfunction.
S
Devices which intrinsically transmit radio waves such as:
Cellular phone, radio transceiver, mobile radio transmitter,
radio-controlled toys, etc.
Keep power to these devices turned off when near this
equipment.
Medical staff in charge of this equipment is required to instruct
technicians, patients and other people who may be around this
equipment to fully comply with the above regulaion.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–11
Safety Precautions
Classifications (cont’d)
4.2 Notice upon Installation of Product
1. Use either power supply cords provided by GE Medical
Systems or ones designated by GE Medical Systems.
Products equipped with a power source plug should be
plugged into the fixed power socket which has the
protective grounding conductor. Never use any adaptor
or converter to connect with a power source plug (i.e.
three-prong-to-two-prong converter).
2. Locate the equipment as far away as possible from
other electronic equipment.
3. Be sure to use only the cables provided by or
designated by GE Medical Systems. Connect these
cables following the installation procedures (i.e. wire
power cables separately from signal cables).
4. Lay out the main equipment and other peripherals
following the installation procedures described in the
Option Installation manuals.
4.3 General Notice
1. Designation of Peripheral Equipment Connectable to
This Product.
The equipment indicated on
the product without compromising its EMC
performance.
Avoid using equipment not designated in the list.
Failure to comply with this instruction may result in poor
EMC performance of the product.
13–5
can be hooked up to
2–12
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Classifications (cont’d)
4.3 General Notice (cont’d)
Safety Precautions
2. Notice against User Modification
The user should never modify this product. User
modifications may cause degradation in EMC
performance.
Modification of the product includes changes in:
a. Cables (length, material, wiring, etc.)
b. System installation/layout
c. System configuration/components
d. Securing system parts (cover open/close, cover
screwing)
3. Operate the system with all covers closed. If a cover is
opened for some reason, be sure to shut it before
starting/resuming operation.
Operating the system with any cover open may affect
EMC performance.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–13
Safety Precautions
Classifications (cont’d)
*5. Patient Environmental Devices
Power Cables with Protective Earth
Peripheral Device
(1) B/W Video Printer
Signals I/O Port
Power In
Non–Imaging Probes
Front Panel
Signals I/O Port
Power Out
Rear Panel
Signals I/O Port
Power Out
Peripherals
(1) VCR
(2) B/W Video
Printer
(3) Color Video
Printer
(4) B/W Polaroid
Camera
(5) Color Polaroid
Camera
(6) Multi Image
Camera
Signals I/O Port
Power In
Imaging Probes
Imaging Probes
Probe Adaptor
ECG Cable
PCG Sensor
Probe
Ports
Signals I/O Port
Physio–Signal
Input Panel
Power Cable with Protective Earth
Foot Switch
Connector
Power In
Figure 2–1. Patient Environmental Devices
Power Cables
with Protective Earth
InSite Modem
Signals I/O Port
Power Line
Telephone Line
Foot Switch
Power Line (AC~)
Ground Line
2–14
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Classifications (cont’d)
5.1 Acceptable Devices
The devices shown in Figure 2–1 are specified to be suitable for
use within the PATIENT ENVIRONMENT.
Safety Precautions
CAUTION
CAUTION
DO NOT connect any probes or accessories without approval
by GE.
Please refer to Peripheral Devices on
16–38
and verified to be compatible with the LOGIQ 500 system.
5.2 Unapproved Devices
The user takes All Responsibility for connecting unapproved
devices.
If devices are connected without the approval of GE, the
warranty will be INVALID.
Any device connected to the LOGIQ 500 must conform to one
or more of the requirements listed below:
for more details. The devices listed have been tested
1. IEC 50, IEC 65, IEC 335, IEC 348, IEC 414, IEC 820,
IEC 950, IEC 1010–1, ISO 7767, ISO 8185, ISO 8359
or IEC 60601–1.
2. The devices shall be connected to PROTECTIVE
EARTH (GROUND).
13–5
and Assistance on
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–15
Safety Precautions
Acoustic Output
Controls Affecting Output
The potential for producing mechanical or thermal bioeffects is
influenced by the controls listed below (refer to
Direct. The Acoustic Output control has the most significant
effect on Acoustic Output.
Indirect.Indirect effects may occur when adjusting the controls
listed on
Always observe the output display for possible effects.
Best practices while scanning
Hints
S
2–17
Raise the Acoustic Output only after attempting image
optimization with controls that have no affect on Acoustic
Output, such as Gain and TGC.
2–17
).
.
.
WARNING
Acoustic
Output
Hazard
NOTE: Refer to the Optimization sections of the Modes chapter
for a complete discussion of each control.
Be sure to have read and understood control explanations for
each Mode intended to be used before attempting to adjust the
Acoustic Output control or any control that can affect Acoustic
Output.
Use the minimum necessary output to get the best diagnostic
image or measurement during an examination. Begin the exam
with the probe that provides an optimum focal depth and
penetration.
Indirect. MinorApplication-dependent system preset.
The middle setting is a factory preset
determined to be a reasonable setting for all
exams. Use presets to set the output
preferred by scan mode and exam
combination.
preset.
preset.
Table 2–3. Controls Affecting Acoustic Output
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output
level, the LOGIQ 500 initiates scanning at a reduced default
output level. This reduced level is preset programmable and
depends upon exam category, application preset and probe
selected. It takes effect when the system is powered on or NewPatient is selected.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–17
Safety Precautions
Warning Label Locations
Overview
LOGIQ 500 warning labels are provided in seven different
languages (English, Japanese, German, French, Italian,
Portuguese and Spanish).
Monitor Labels
A temporary label is placed on the monitor face to warn not to
move the monitor support arm without the monitor attached.
This label is removed after installation of the monitor.
LOGIQ 500 MD
Push Open
Figure 2–2. Temporary Warning Label and Location
2–18
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Monitor Labels (cont’d)
Two caution labels are found on the back of the monitor. One
warns to only move the console with the monitor in its lowest
position; the second warns not to push the console from the
side.
Safety Precautions
12
Figure 2–3. Lowering Monitor Caution Labels and Locations
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–19
Safety Precautions
Monitor Labels (cont’d)
One caution label is found on the top of the monitor.
2–20
Figure 2–4. Top of Monitor Caution and Location
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Console Labels
Defibrillator Caution
Safety Precautions
Labels found on the back and side of the console will either be
translated to the twelve languages or be specific to the region.
LOGIQ
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Figure 2–5. Defibrillator Label Location
2–21
Safety Precautions
Ground Point
CAUTION
Figure 2–6. Signal Ground Point Location and Label
This is only for “FUNCTIONAL GROUNDING”, NOT
“PROTECTIVE EARTH”.
2–22
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Regulatory Labels (European Systems)
Safety Precautions
Figure 2–7. Regulatory Label Location (European)
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
2–23
Safety Precautions
Regulatory Labels (American Systems)
2–24
Figure 2–8. Regulatory Label Location (Americas)
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Site Requirements
Introduction
Site Requirements
.
Only qualified physicians or sonographers should perform
ultrasound scanning on human subjects for medical diagnostic
reasons. Request training, if needed.
Do not attempt to install the system alone. General Electric,
Affiliate, or Distributor Field Engineers and Application
Specialists will install and setup the system. Refer to Who To
Contact on
The LOGIQ 500 does not contain any operator serviceable
internal components. Ensure that unauthorized personnel do
not tamper with the unit.
Perform regular preventive maintenance. Refer to
maintenance instructions.
Maintain a clean environment. Turn off the system circuit
breaker before cleaning the unit. Refer to
instructions.
Never set liquids on the unit to ensure that liquid does not drip
into the control panel or unit.
1–6
.
16–11
for
16–12
for cleaning
3–2
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Site Requirements
Before the system arrives
NOTICEThis medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and
other institutions which are environmentally qualified. The use
of this equipment in an inappropriate environment may cause
some electronic interference to radios and televisions around
the equipment. This equipment can be used in residential
areas only under the supervision of physicians or qualified
technicians.
Ensure that the following is provided for the new system:
S
A separate power outlet with a 20 amp circuit breaker for
120 VAC (USA) or 7.5 amp circuit breaker for 220–240 VAC
(Europe, Latin America).
S
Take precautions to ensure that the console is protected
from electromagnetic interference.
Precautions include:
S
Operate the console at least 15 feet away from motors,
typewriters, elevators, and other sources of strong
electromagnetic radiation.
S
Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent
electromagnetic interference.
S
Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–3
Site Requirements
Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
OperationalStorageTransport (<16 hrs.)
Temperature
Humidity
Pressure
Table 3–1. System Environmental Requirements
10_- 40_ C
50_- 104_ F
30-85%
non-condensing
700-1060hPa700-1060hPa700-1060hPa
–10_- 60_ C
14_- 140_ F
30-90%
non-condensing
–40_- 60_ C
–40_- 140_ F
30-90%
non-condensing
3–4
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Console graphics
Console Overview
Console Overview
The following are illustrations of the console:
1
2
3
1Optional Probe Cable Holder (wire holder standard for Americas’ systems)
2Probe and Gel Bottle Holder (Removable for Cleaning)
3Optional Cable Clipper for Probe Cable
Figure 3–1. LOGIQ 500 System (right and left side views)
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–5
Console Overview
Console graphics (cont’d)
1
2
3
7
8
6
10
9
4
5
1Task Light Switch
2Task Light
3Optional B/W Video Page Printer
4Air Filter (front and back of system)
5Power Supply Air Filter
6Optional Probe Cable Holder (wire holder standard for Americas’ systems)
7VCR Microphone
8Release Button—to raise and lower video monitor
9Optional Physiological Input Panel
10 Swivel Lock on Video Monitor Arm
11 Hinged Peripheral Cable Access Panel Door
12 Cable Access Channel
11
12
4
ON
ON
OFF10
Figure 3–2. LOGIQ 500 System (front and back views)
3–6
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
External disk drive
Storage areas
Console Overview
The external MOD drive is a 3.5 inch 128MB/230MB MOD drive
located below the keyboard.
It can be used to perform software upgrades, image archiving
(option) and service diagnostics.
Several convenient storage areas are provided within the
console as shown by the shaded areas in Figure 3–3. Use
them to store gel, options, probe cables, accessories, etc.
1
LOGIQ 500
LOGIQ
1
1 Storage
Figure 3–3. Storage Areas
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–7
Console Overview
Peripheral/Accessory Connection
Peripheral/Accessory Connector Panel
LOGIQ 500 peripherals and accessories can be properly
connected using the rear connector panel located behind the
rear door. Only the B/W Page Printer (UP-890) can be
connected to the front accessory panel.
Located on the rear panel are video input and output, audio
input and output, camera expose, foot switch, power and
control connections for VCR, printer, MIC and service tools.
See Figure 3–5.
CAUTION
CAUTION
Each outer (case) ground line of peripheral/accessory
connectors are Earth Grounded.
Signal ground lines are Not Isolated, except the Service Port.
All of the signal lines (including the signal ground) of the
Service Port are Isolated.
Service
Figure 3–4. Service Port
Use only approved probes, peripherals or accessories.
3–8
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Peripheral/Accessory Connector Panel (cont’d)
Console Overview
12
1Old Rear Panel2New Rear Panel
100V 5.0A Max.
Including front printer panel
120V 4.1A Max.
Including front printer panel
220–240V 2.0A Max.
Including front printer panel
Figure 3–5. Peripheral/Accessory Connector Panel
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–9
Console Overview
Foot Switch (option)
Freeze
Record
Foot Switch
Use only the recommended optional multi-functional Foot Switch
in parallel with or as an alternative to the Freeze and Record 1
controls to:
S
Freeze a real-time image (left switch).
S
Send an image to the hard copy device (right switch).
The Foot Switch connection is located at the back of the
console on the left-hand side of the back panel.
LOGIQ 500
1
2
3
4
1Accessory Panel for Foot Switch Connector
2Store Foot Switch Here
3Console Air Filter Screen
4Power Supply Air Filter
Figure 3–6. Foot Switch Storage and Connectors
Other Peripherals/Accessories (options)
Refer to the
information.
Recording Images
3–10
chapter of this manual for more
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
System Positioning/Transporting
Moving the System
System Positioning/Transporting
When moving or transporting the system, follow the precautions
below to ensure the maximum safety for personnel, the system,
and other equipment.
Before moving the system:
1. Turn the system power switch off.
2. Unplug the power cord.
3. All cables from off-board peripheral devices (IIE
camera, external printer, VCR, etc.) must be
disconnected from the console.
4. Ensure that no loose items are left on the console.
5. Loop the cord around the handle on the back of the
system or wrap the cord in a bundle and store it behind
the rear storage panel.
To prevent damage to the Power Cord, DO NOT pull
excessively on the cord or make sharp bends while wrapping.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–11
System Positioning/Transporting
Moving the System (cont’d)
1Handle
2Storage Area Behind Door
.
Figure 3–7. Location of Storage Area
6. Connect all probes to be used while off site. Ensure
that probe cables are out of the way from the wheels
and not protruding beyond the console.
NOTE: If more than three (3) probes are intended to be used,
store the additional probes securely in the front storage
area.
7. Store all other probes in their original cases or in soft
cloth or foam to prevent damage.
8. Store sufficient gel, optical disks, and other essential
accessories in the provided space.
9. Adjust the monitor to its lowest position possible.
Ensure that the monitor arm is locked in place.
10. Unlock the front wheels.
3–12
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Moving the System (cont’d)
When moving the system:
1. Always use the rear handle grips to move the system.
2. Take extra care when moving the system long
distances and on inclines. Ask for help if necessary.
Avoid ramps that are steeper than ten degrees to avoid
tipping over the
System Positioning/Transporting
system.
..
CAUTION
NOTE: Wheel chair ramps are usually less than five degrees.
Utilize additional care and personnel when moving on
steep incline (>5°) or loading into a vehicle for
transport.
NOTE: DO NOT attempt to move the console using any cables
or fixtures, such as the probe connectors.
3. Use the brake, located on the bottom of the system in
the front, when necessary.
4. Do not let the system strike walls or door frames.
5. Use extra care when crossing door or elevator
thresholds.
6. Once the destination is reached, lock the wheels.
The system weighs approximately 180 kg (397 lbs). To avoid
possible injury and equipment damage:
S
Be sure the pathway is clear.
S
Limit movement to a slow careful walk.
S
Use two or more persons to move the system on inclines or
long distances.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–13
System Positioning/Transporting
Transporting the System
Use extra care when transporting the system using vehicles. In
addition to the instructions used when moving the system (refer
to
3–11
), do the following:
1. Only use vehicles that are designed for transport of the
LOGIQ 500 system.
2. Load and unload the system to a vehicle parked on a
level surface.
3. Ensure that the transporting vehicle can handle the
weight of the system plus the passengers.
4. Ensure that the load capacity of the lift (a minimum of
180 kg [397 lbs] is recommended) is capable of
handling the weight of the system.
5. Ensure that the lift is in good working order.
6. Secure the system while it is on the lift so that it cannot
roll. Use either wood chocks, restraining straps, or
other similar types of constraints. Do not attempt to
hold it in place by hand.
.
WARNING
.
NOTE: Strap the system below its handle so that the system
does not break loose.
Never ride on the lift with the system. A person’s
weight coupled with the weight of the system may
exceed the load capacity of the lift.
7. Employ two to three persons to load and unload safely
from a vehicle.
8. Load the unit aboard the vehicle carefully and over its
center of gravity. Keep the unit still and upright.
NOTE: Do not lay the unit down.
9. Ensure that the system is firmly secured while inside
the vehicle. Any movement, coupled with the weight of
the system, could cause it to break loose.
10. Secure system with straps or as directed otherwise to
prevent motion during transport.
11. Prevent vibration damage by driving cautiously. Avoid
unpaved roads, excessive speeds, and erratic stops or
starts.
3–14
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Wheels
System Positioning/Transporting
.
Front wheels
Back wheels
.
Setting the lock
Examine the wheels frequently for any obvious defects that
could cause them to break or bind.
The front wheels swivel, pivot, and lock.
The back wheels swivel and pivot but do not lock.
NOTE: For the USA version console, the back wheels do not
pivot.
To engage the wheel lock:
Press down on the lock pedal (located at the front of
the wheels). The pedal remains depressed.
To release the lock:
Press down on the pedal again. The pedal returns to
its normal position.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–15
Powering On the System
Powering On the System
Connecting and Using the System
To connect the system to the electrical supply:
1. Ensure that the wall outlet is of the appropriate type.
2. Ensure that the power switch is turned off.
3. Unwrap the power cable. Make sure to allow sufficient
slack in the cable so that the plug is not pulled out of
the wall if the system is moved slightly.
4. Push the power plug securely into the wall outlet.
W ARNING
Plug
Plug and Outlet Configuration
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet. See
the system arrives
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital
grade” or “hospital only” grounded power outlet.
on
3–3
for rating information.
Outlet
120 VAC, 1350 V A
(USA)
Figure 3–8. Example Plug and Outlet Configurations
220–240 VAC, 1350 V A
Plug and Outlet Configuration
Site Requirements, Before
Plug
Outlet
(Europe)
3–16
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Acclimation Time
Powering On the System
After being transported, the unit requires one hour for each 2.5
increment its temperature is below 10_ C or above 40_ C.
_
C
_
F
hours
_
C
_
F
hours
6055504540353025201510
140131122113104958677685950
86420000000
50–5–10–15–20–25–30–35–40
413223145–4–13–22–31–40
2468101214161820
Table 3–2. System Acclimation Time Chart
Power On/Off Standby
CAUTION
Press the top portion of the Power On/Off Standby switch to
turn the power on. The circuit breaker, on the rear of the unit,
must also be in the on position (see
3–22
_
).
Figure 3–9. Power Switch Location
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–17
Powering On the System
Power Up Sequence
S
The monitor and console power indicator light up.
S
The system is initialized. During this time:
SSS
Start of
diagnostic
run
1
Two beeps sound during the sequence.
All lighted buttons on the keyboard light.
System diagnostics run. Its status is reflected on the
monitor by the graphics in
Figure 3–10
.
XXXXXXX–X
2
Version X.XX
End of
diagnostic
run
Version X.XX
1Current SBC Assembly Part Number2Current Software Version
Figure 3–10. Power Up Graphic Sequence
XXXXXXX–X
Version X.XX
3–18
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Power Up Sequence (cont’d)
Powering On the System
.
Hints
.
Password Protection
NOTE: If errors occur, an error message appears at the bottom
of the screen. See User Maintenance, Troubleshooting
for more information.
If problems occur, freeze the image and take a picture
for reference. This will help if there is a need to call for
service.
S
Probes are initialized for immediate operation.
NOTE: If no probes are connected, the system goes into
standby mode.
S
Peripheral devices are activated on power up.
After initialization has been completed, the new patient entry
menu will be displayed.
A maximum of eight personal IDs and associated passwords
can be preset in the LOGIQ 500 (see
If IDs and passwords have been entered and the Password Ask
parameter is on, the following occurs in the power up sequence:
S
The message USERID:__________ appears in the middle
of the start up sequence.
14–61
).
S
Type in a User ID and press Return.
S
Keyboard lights go out.
S
The message PASSWORD:__________ is highlighted.
S
Type in the password that corresponds to the User ID and
press Return.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–19
Powering On the System
Password Protection (cont’d)
USER ID :
PASSWORD:
Figure 3–11. System Startup Screen with Password Ask on
If the correct User ID/Password is entered, the system
continues with the power up sequence.
Version X.XX
.
If an incorrect User ID/Password is entered, the system allows
four additional attempts. After the fifth incorrect entry, the
message “Login incorrect” appears. The power up sequence
will not continue. The system must be turned off, then on, to
begin again.
NOTE: If User ID/Passwords are registered and the Password
Ask function is off, press Return at the ID__________ prompt.
The system continues the power up sequence.
3–20
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Power Off
Powering On the System
When switching off the system:
S
Move the ON, OFF/STAND-BY switch to the OFF position.
The message “WARNING: NOW STARTING THE POWER
OFF PROCESS” appears at the bottom of the display.
S
The LOGIQ 500 takes a few seconds to save current
scan parameter data in the temporary files to the hard drive
before turning the power off.
During this time a message flashes on the screen:
“Do not pull Power Cable.
Do not turn off Breaker.”
Pulling the power cable or turning off the circuit breaker,
while the hard drive is working, may corrupt the system
operating software on the hard drive.
If the system has not turned off five minutes after pressing
the power switch off, listen for hard drive activity. If there is
no hard disk drive activity, the circuit breaker on the bottom
of the power supply can be used to turn off the system. Do
NOT turn off the circuit breaker while the hard disk is
working.
S
Disconnect the probes.
Clean or sanitize all probes as necessary. Store them in
their shipping cases to avoid damage.
S
If daily maintenance is to be performed, turn off the circuit
breaker in the back of the system.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–21
Powering On the System
Circuit breaker
1
0
.
The Circuit Breaker is located on the back of the console, at the
bottom of the system. On supplies main power to all internal
systems. Off removes main power from all internal systems.
The circuit breaker automatically shuts off power to the system
in case of a power overload.
If a power overload occurs:
1. Turn off all peripheral devices.
2. Turn off the Main Power Switch to the console.
3. Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the On position; DONOT hold the switch in the On position. If the Circuit Breaker
switch remains On, follow the
described.
NOTE: If the Circuit Breaker switch does not remain in the On
position or trips again:
1. Disconnect the Power Cable.
2. Call Service immediately .
DO NOT attempt to use the system.
Power On
procedure previously
3–22
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Circuit breaker (cont’d)
ÏÏ
Powering On the System
1
ON
ON
OFF10
1Circuit Breaker
Figure 3–12. Location of Circuit Breaker
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–23
Adjusting the Monitor
Adjusting the Display Monitor
Rotate, tilt, raise and lower the monitor
The monitor position can be adjusted for easy viewing.
S
The monitor can be rotated around it’s central pivot point.
S
The monitor can be tilted for the optimum viewing angle.
S
The monitor arm can swing forward or backwards.
S
The monitor arm can be raised or lowered for the best
viewing height.
CAUTION
3–24
Figure 3–13. Display Monitor Movement
Movement of the monitor swing arm or height adjustment
requires the release of the locking mechanism. After an
adjustment is made, ensure that the mechanism is locked to
prevent unexpected motion.
When moving the LOGIQ 500 system, lower the monitor to its
lowest possible position to improve stability.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Brightness and Contrast
Adjusting the monitor’s contrast and brightness is one of the
most important factors for proper image quality. If these
controls are set incorrectly, the Gain, TGC, Dynamic Range and
even Acoustic Output may have to be changed more often than
necessary to compensate.
The proper setup displays a complete gray scale. The lowest
level of black should just disappear into the background and the
highest white should be bright, but not saturated.
To adjust the Brightness and Contast:
1. Select the Set Up Top Menu. From the Set Up
Adjusting the Monitor
Sub-Menus, select Utility. The Utility Menu appears as
shown in Figure 3–14.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Figure 3–14. Utility Menu
3–25
Adjusting the Monitor
Brightness and Contrast (cont’d)
2. Enter menu number “11” (Monitor Adjustment) in the
Utility Menu and press Return. The Monitor Adjustment
Menu appears as shown in Figure 3–15.
Figure 3–15. Monitor Adjustment Menu
3. Enter the number of a gray scale map for adjustment
according to the brightness of the exam room and the
exam category and press Return.
1: Dark Room—for a dimly lit room with a color
monitor.
2: Bright Room—for a well lit room with a color
monitor.
3: Cardiac—for a Cardiac Exam with a color monitor.
4: Dark Room (B/W)—for a dimly lit room with a black
and white monitor.
5: Bright Room (B/W)—for a well lit room with a black
and white monitor.
6: Cardiac (B/W)—for a Cardiac Exam with a black
and white monitor.
3–26
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Brightness and Contrast (cont’d)
The adjustment maps’ gray scale bar appears on the
monitor as shown in Figure 3–16.
4. Access the Brightness/Contrast controls by pushing on
the top center of the control panel door, located under
the display screen. Push Brightness and Contrast
slidepots to the maximum (to the right).
5. Decrease the Brightness by degrees until “B” in the
gray scale bar is not visible.
6. Decrease the Contrast by degrees until “A” in the gray
scale bar is not visible.
7. Push Brightness slidepots to the maximum (to the
right) again.
8. Decrease the Brightness by degrees until “C” in the
gray scale bar is not visible.
9. Press the Utility rocker switch to return to the previous
scan mode.
3
5
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–27
Adjusting the Monitor
Brightness and Contrast (cont’d)
Generally speaking, do not change the controls once they have
been set. Once set, the display then becomes the reference for
the hard copy device(s).
..
Speakers
NOTE: After readjusting the monitor’s Contrast and Brightness,
readjust all preset and peripheral settings.
NOTE: Monitor degaussing (demagnetizing) is done
automatically when the system is turned on.
Stereo audio, provided by speakers located on the sides of the
Display Monitor, is standard for:
S
Audio Doppler operation (left side is blood flow away/right
side is blood flow toward)
S
Audio playback of videotaped scan sessions
S
Audio error notification
1
23
3–28
1Speakers
2Toward
3Away
Figure 3–17. Display Monitor Speakers
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Introduction
Probes
Probes
Only use approved probes.
All imaging probes can be plugged into any of the three probe
ports.
For more
information
Refer to the
Connecting the Probe
Probes can be connected at any time, regardless of whether
the console is powered on or off. To ensure that the ports are
not active, place the system in the image freeze condition.
To connect a probe:
CAUTION
Probes
1. Place the probe’s carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cord.
3. DO NOT allow the probe head to hang free. Impact to
the probe head could result in irreparable damage.
4. Turn the connector locking handle counterclockwise.
5. Align the connector with the probe port and carefully
push into place.
chapter.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–29
Probes
Connecting the Probe (cont’d)
Figure 3–18. Probe Connector Panel
6. Turn the connector locking handle clockwise to secure
the probe connector.
7. Carefully position the probe cord so it is free to move
and is not resting on the floor.
3–30
Figure 3–19. Connecting a Probe
LOGIQ 500 Basic Users Manual
lock
unlock
2233658–100 Rev . 0
Cable Handling
Take the following precautions with probe cables:
SSS
Activating the Probe
To activate the probe, press the Probe Select key that
corresponds to the probe port to which the desired probe is
connected. The Single CWD key is used to activate the
dedicated continuous wave Doppler (CWD) probe.
Probes
Keep free from wheels
Do not bend the cable acutely
Avoid crossing cables between probes.
Figure 3–20. Probe Select Keys
Key Light Indicators:
S
Dark—no probe attached to probe port.
S
Half Bright—probe attached to probe port but not active.
S
Brightly Lit—probe attached to probe port and is active.
The probe activates in the currently selected operating mode.
The probe’s default settings for the mode and selected
application are used automatically.
CWD Mode, split Crystal or Doppler-Only CW probes are
options available on the LOGIQ 500.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–31
Probes
Probe Name Menu
If the Display Probe Name preset in the Setup/System
Parameters page one is ON, by pressing a half bright or brightly
lit key causes the probe names for each port to be displayed in
the soft menu. The desired probe can then be selected from
the soft menu or by pressing the Probe Select keys. Press the
Mode Top Menu key to exit the probe name display.
C364
(1)
Deactivating the Probe
When deactivating the probe, the probe is automatically placed
in standby mode.
To deactivate a probe:
"yA
SSS
S
PresetB
L739
(2)
Set UpECG
????
(3)
Figure 3–21. Probe Name Menu
Press the Freeze key.
Gently wipe the excess gel from the face of the probe.
Carefully slide the probe around the right side of the
keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
3–32
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Disconnecting the Probe
Probes can be disconnected at any time. However, the probe
should not be selected as the active probe.
S
Move the probe locking handle counterclockwise.
S
Pull the probe and connector straight out of the probe port.
S
Carefully slide the probe and connector away from the
probe port and around the right side of the keyboard.
S
Ensure the cable is free.
S
Be sure that the probe head is clean before placing the
probe in its storage box.
Transporting Probes
Secure the probe in its holder for moving short distances.
When transporting a probe a long distance, store it in its
carrying case.
Probes
Storing the Probe
It is recommended that all probes be stored in the carrying case
provided.
S
First place the probe connector into the carrying case.
S
Carefully wind the cable into the carrying case.
S
Carefully place the probe head into the carrying case. DO
NOT use excessive force or impact the probe head.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–33
Operator Controls
Control Panel Map
Operator Controls
Controls are grouped together by function for ease of use. See
the callouts for this illustration on the following page.
123
4
10
↓
q
98
Figure 3–22. Control Panel
5
"yA
→
→
"yA
6
7
3–34
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Control Panel Map (cont’d)
Operator Controls
Key Illumination
1. Alphanumeric Keyboard (see
2. Patient Information (see
3. Probe Selection (see
4. Soft Menu (see
5. Doppler CFM (see
6. User Define (see
7. Mode Display Record (see
8. Measurements/Annotations (see
9. TGC and Acoustic Output (see
10. VCR (see
The front panel keys are all illuminated according to their
availability.
S
Disabled or unavailable selections have the key
illumination turned off.
S
Enabled or available selections have the key illumination
half lighted (back lit).
3–44
3–37
14–97
).
3–29
).
5–28
4–3
and
).
).
).
5–50
3–40
3–34
).
5–10
).
).
3–42
).
).
S
Active or cancelable selections have the key illumination
fully lighted.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–35
Operator Controls
Keyboard
Red
Blue
The standard alphanumeric keyboard has some special
functions.
The Cursor Home key brings the alphanumeric cursor to the
very upper left corner of the available field.
Control is used in conjunction with other keys to activate special
keyboard functions.
Tab is used to move forward or backwards through the text one
word at a time, or eight characters at a time.
Red Shift is used to activate the special characters highlighted
in red on the keys to the right side of the keyboard. See
for details.
Blue Shift activates the VCR controls on the keyboard for the
approved Sony SVO-9500MD. When Blue Shift is activated
and the VCR is in play mode, the left/right pointers control
searching reverse or forward. The up/down pointers control
shifting backwards/forwards one frame at a time while the VCR
is paused.
6–22
User Define Function
3–36
If the VCR is not in play mode, the left/right pointers will cause
the VCR to rewind or fast forward.
Alphanumeric keys (A~Z and 1~0) can also be used for the
User Define Key function. Refer to
14–97
LOGIQ 500 Basic Users Manual
for details.
2233658–100 Rev . 0
Soft Menu Control Panel
Additional functionality, not available as a control or key on the
front panel, can be found via the Soft Menus. Different soft
menus appear depending on the mode, special function or
calculation package selected.
The Soft Menu consists of Top Menu Select keys andSub-Menu Select rocker switches.
T op Menu Select
Operator Controls
TOP MENU
Current
Available
SUB-MENU
Current
Available
Current Top
Menu Page
Available Top
Menu Pages
1
2
2
4
Sub-Menu Select
PresetECG
Frame Imaging
Average
Freq
OFF
3 MHz
Set UpB
Image
Softner
Color
3D
Mode
Figure 3–23. Soft Menu Control Panel
The Soft Menu Display is in the center of the Soft Menu
Controls. The display is divided into twelve sections, four top
menu categories and eight sub-menu categories.
TOPTOPTOPTOP
SUBSUBSUBSUB
SUBSUBSUBSUB
Figure 3–24. Top Menu Page Display
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–37
Operator Controls
Top Menu Organization
The Top Menu Select key cycles through the top level menu
page selections. The far left side top menu is the default
selection and its sub-menus are automatically displayed.
The top menu groups are divided into two pages. The first top
menu page displays:
Mode DefaultPresetSet UpECG
Figure 3–25. Mode Default Top Menu (page one)
S
S
S
Mode Default Menu is the current highest priority active
mode. The five possibilities are “B”, “M”, “PWD”, “CWD”
and “CFM”.
Preset
is for user programmable application and factory
exam parameter preset selections.
Set Up enables the system customization sub-menus.
S
ECG (option) is used to adjust the ECG waveform and ECG
synchronized scanning.
The second top menu page shows:
ArchiveDICOMAuto SeqCINE
Figure 3–26. Mode Default Top Menu (page two)
S
Archive
images to MOD.
S
DICOM
S
Auto Sequence
measurement sequences.
S
CINE
accumulated image data.
is used with the Image Archive option for storing
(option) displays sub-menus for DICOM setup.
displays the user programmed
is used when the image is frozen for review of the
3–38
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Top Menu Organization (cont’d)
The four top menu select keys select and highlight the
corresponding top menu and display its associated
sub-menus.
Four keys on the front panel automatically disable all top menus
and display specific sub-menu selections.
S
Body Pattern displays the available graphic selections in
the sub-menu area.
S
Comment displays the available selections from the
comment library.
S
Measurement displays the available measurements
suitable for the current exam category, image format or
individual user sequences.
S
Image Recall displays information pertaining to images
temporarily stored in system memory.
Operator Controls
.
NOTE: To return functions to the Top-Menus, press Clear.
Sub-Menu Organization
Different Sub-Menus are displayed according to the front panel
key pressed or Top Menu selected.
Each selection or parameter in the Sub-Menus relates directly
to the Top Menu that is fully illuminated.
The Sub-Menu Select rocker switch cycles to the next
( ") or previous (A ) sub-menu page.
The Sub-Menu rocker switches allow for the
increase/decrease of a parameter value (i.e. Dynamic
Range) or enabling/disabling of a parameter (i.e. Image
Softener).
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–39
Operator Controls
Mode, Display and Record
This group of controls provides various functions relating to the
display mode, display orientation, image recording/saving,
freeze, gain and cine scroll.
The Mode Controls select the desired display mode or
combinations of display modes.
During dual display modes the L and R keys activate the Left or
Right displayed image.
IncreaseDecrease
→
→
The Depth knob controls the image display depth.
The Reverse key toggles the left/right orientation of the scan
image.
The Image Memory key stores the current frozen displayed
image in system image memory.
Maximum number of B-Mode images is 8.
Maximum number of Timeline images is 4.
All images are erased when the New Patient key is pressed or
there is a loss of system power.
The Image Recall key displays a menu of the images stored in
memory. After pressing Image Recall, select the desired
image from the Soft Menu for display.
3–40
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Mode, Display and Record (cont’d)
Record 1 and Record 2 are used to activate/print the
designated recording device (i.e. video page printer,
multi-image camera, image archive option).
The Freeze key is used to stop the acquisition of ultrasound
data and freeze the image in system memory.
"yA
Pressing Freeze a second time continues live image data
acquisition.
Controls the gain of the displayed echoes during B-Mode
scanning. Controls the gain of the displayed timeline echoes
during B/M- and M-Mode. During B/M-Mode, B-Mode gain can
be controlled by the Doppler/CFM Angle Control.
Operator Controls
The gain value displayed on the monitor is:
B-ModeB-Mode Gain
M-ModeM-Mode Gain
B/M-ModeM-Mode Gain
When the image is frozen it controls scrolling forwards and
backwards through the cine loop images in temporary memory.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–41
Operator Controls
Measurement and Annotation
This group of controls performs various functions related to
making measurements, annotating and adjusting the image
information.
The Comment key enables the image text editor and displays
the annotation library soft menu. For more details, refer to
6–16
.
The Body Pattern key enables the body pattern Soft Menu and
displays the default pattern on the screen. For more details,
refer to
The M/D Cursor key enables Trackball control of the M-Mode or
Doppler cursor line (not angle correction) or the CFM window in
real-time Color Flow Mode.
The Scan Area key enables Trackball control of the B-Mode
image area size and position in B/W, the CFM window size and
position in Color mode and the Zoomed area size and position
in Zoom.
6–24
.
The Measurement key is used in all types of basic
measurements. It displays the measurement soft menu for the
current exam category.
3–42
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Measurement and Annotation (cont’d)
The Ellipse rocker switch is used to activate the ellipse
measurement function after the first distance measurement has
↑
↓
been set. It also toggles which cursor is the movable cursor
during the ellipse adjustment.
The Set key is used for various functions, but is generally used
to fix or finish an operation (i.e. to fix a measurement cursor or
exit scan area size/position in B-Mode and zoom).
The Clear key is generally used to erase or exit functions such
as annotations/comments, measurements and zoom. This will
return the system to the basic mode top menu.
Press the Zoom key to activate the zoom function. For more
details, refer to
6–2
.
Operator Controls
Zoom Size
Rotation
This knob is used to control the zoom size with the zoom
function in real-time or rotate the probe position indicator in the
body pattern function.
The Trackball is used with almost every key function in this
group. Trackball control depends on the last key function
pressed.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
3–43
Operator Controls
VCR Controls
↓
These controls are used in conjunction with the approved
optional video cassette recorder.
Press the External Video key to allow for the display of an
external video signal (i.e. from a VCR) on the system monitor.
Pressing Play automatically selects external video.
Press the key a second time to return to LOGIQt 500 video
display.
Press this key to enable or disable the on board microphone to
allow for audio to be added to the taping of system video.
"
J
/
/
.
F
Pressing the Stop/Play key either places the VCR in the play
mode or stops the play or record mode.
Pressing the Pause/Record key either places the VCR in the
record mode or causes the VCR to pause if it is already in the
record or play mode.
NOTE: See Keyboard on 3–36 for additional VCR controls.
3–44
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Beginning an Exam
Introduction
Beginning an Exam
Begin an exam by entering new patient information.
The operator should enter as much information as possible,
such as:
S
Exam category
S
Patient name
S
Patient ID
S
Comments
S
Exam Information
The patient’s name and ID number is retained with each
patient’s image and transferred with each image during
archiving or hard copy printing.
4–2
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Beginning a New Patient
The New Patient key should be pressed at the beginning of
each patient study. Pressing this key automatically erases all
patient data, annotations, measurements, calculations and
summary report pages. A preset parameter can be set to
prompt the user if they wish to erase all patient data or not. The
system also resets every parameter (i.e. Zoom, 3D Mode,
Triplex Mode, etc.).
Beginning an Exam
.
.
CAUTION
NOTE: Oper ID and Ref MD will not be erased.
Enter patient data with the alphanumeric keyboard.
The Soft Menu automatically defaults to the last exam preset
top menu when New Patient is pressed. An exam applications
preset can be selected from the sub-menu to provide a suitable
starting point for system scan parameters.
NOTE: After a specific time interval the Soft Menu defaults to
the “Mode” Top Menu.
To avoid patient identification errors, always verify the
identification with the patient. Make sure the correct patient
identification appears on all screens and hard copy prints.
The Patient Entry Menu appears on the display monitor.
Figure 4–1. Patient Entry Menu
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
4–3
Beginning an Exam
Beginning a New Patient (cont’d)
The first Data Entry field is presented in reversed display.
S
Trackball to the exam category selection. Select from 7
examination categories: Radiology/Abdominal, Obstetrics,
Gynecology, Cardiology, Vascular, Urology and Small Parts.
Input the appropriate number. The patient information input
menu changes. Information pertinent to the selected exam
category appears in the menu.
To navigate through the Patient Entry Menu, use the Return
key or the Trackball to move the cursor.
S
Input the Patient Name (29 characters maximum).
S
Input Patient ID Number (14 characters maximum).
S
Input any desired Note (30 characters maximum).
S
Input the desired Operator ID (four digits maximum).
.
NOTE: Patient Name, Patient ID, Notes, Oper ID, Ref MD, and
Comments are common to all exam category menus.
The Patient Name and Patient ID will appear on the image
screen. All other information is automatically entered into the
appropriate exam report page.
Information in the Exam Category patient entry menus is
considered necessary for that type of exam. Fill in all
information possible.
S
The display units of measure for items such as weight or
height can be selected from the Set Up/Preset program
menu page 10. Choose the priority and unit of measure on
this Preset Menu page.
S
Input Ref MD (16 characters maximum).
S
Input comment field (2 lines of 50 characters each).
S
When all patient data entries have been completed,
highlight Exit and press Return or the New Patient key.
4–4
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Beginning a New Patient (cont’d)
Beginning an Exam
..
OB/GYN Exam Categories
NOTE: If patient information needs to be edited or the exam
category changed, use the ID Name key. Pressing ID Name
allows for modification of the Patient Entry Menus without
erasing accumulated patient images, measurements,
annotations, calculations and summary reports.
NOTE: Patient age entry information (years, months, weeks,
days) is selected in Set Up/Preset Program page 10, “Display
Unit Age”. Patient height is “Display Unit Height” and Patient
weight is “Display Unit Weight”. Choose the unit values that are
to appear on the patient entry menu.
BBT is a pregnancy origin data selection choice that appears in
the Tokyo University, Osaka University and European OB
formats only. LMP, EDC and GA are the only choices in the
USA version.
If the Multigestational option is installed, “Fetus Number:” will
appear to the right of Ref. MD:. If there is more than one fetus,
enter the correct number of fetuses. If the Multigestational
option is not installed, each fetus must be scanned separately.
Re-enter the patient’s information for each fetus.
ID/Name
The ID/Name key functions the same as the New Patient key
except it does NOT erase previous patient data (i.e.
measurements, calculations, etc.) Use the ID/Name key to
enter or replace patient data without changing the current status
of the system. One common reason might be to change the
exam category.
Pressing ID/Name enables the Patient Entry Menu.
Edit the menu as described in
No other function can be started until ID/Name is completed.
To complete the ID/Name process, press ID/Name a second
time or move the highlighted cursor to Exit and press Return.
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Beginning a New Patient
.
4–5
Beginning an Exam
Helpful hints
Hints
If power is lost during the ID/Name function, any data that was
added or modified will not be saved.
Patient information/setup is saved to the system hard drive at
power off only if the System Parameters (page 1) preset “Power
On Status” is set to
it’s power down state when power is turned on.
The following rules apply when filling in the New Patient menu:
S
Press Caps Lock to type uppercase letters. Press Caps
Lock again to type lowercase letters.
S
Press Back Space to erase characters and correct errors.
S
To change information, press Return or use the Trackball
to move to the field, then type over the existing information
with correct information.
S
Press Return to move to the next field.
S
Use the Trackball to move the reversed cursor to the
desired item.
S
When pressing Return at the last data entry field, the
system returns to real-time scanning.
S
Standard keyboard keys repeat when held down.
Keep-Latest
. The setup is then returned to
S
To start over, press New Patient.
Remember, user and factory
dependent upon the exam category selected when filling in
the New Patient menu.
-defined presets are
4–6
LOGIQ 500 Basic Users Manual
2233658–100 Rev . 0
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.