LIST OF EFFECTIVE PAGES
REV DATE PRIMARY REASON FOR CHANGE
0 July 27, 1995 Initial release
1 October 6, 1995 Software version 2.10 release, error correction
2 July 10, 1996 Software version 3.00 release, error correction
3 March 10, 1997 Software version 3.10 release, error correction
4 December 17, 1997 LOGIQ 400CL release, error correction
5 June 19, 1998 Software version 3.40 and 3.41 for LOGIQ 400CL release and error correction
6 April 21, 1999 Software version 4.01y and 4.02y release and error correction
7 September 17, 1999 Additional information for SGMS Manufacturing
8 October 15, 1999 Software version 4.31y and 4.32y release and error correction
9 March 17, 2000 Software version 5.01 release and error correction
P AGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page 9 . . . . . . . . . .
GE Logo page – . . . . . .
P9030CB 2 . . . . . . . . . . .
P9030CC 4 . . . . . . . . . .
P9030CD 0 . . . . . . . . . .
A9 . . . . . . . . . . . . . . . . . .
i8 . . . . . . . . . . . . . . . . . . .
ii to x 9 . . . . . . . . . . . . . .
Chapter 1
1–1 8 . . . . . . . . . . . . . . . .
1–2 to 1–3 0 . . . . . . . . . .
1–4 2 . . . . . . . . . . . . . . . .
1–5 to 1–8 8 . . . . . . . . . .
1–9 7 . . . . . . . . . . . . . . . .
1–10 to 20 8 . . . . . . . . . .
Chapter 2
2–1 5 . . . . . . . . . . . . . . . .
2–2 to 2–4 0 . . . . . . . . . .
2–5 5 . . . . . . . . . . . . . . . .
2–6 to 2–9 0 . . . . . . . . . .
2–10 9 . . . . . . . . . . . . . . .
2–11 to 2–16 0 . . . . . . . .
Chapter 3
3–1 9 . . . . . . . . . . . . . . . .
3–2 0 . . . . . . . . . . . . . . . .
3–3 to 3–16 9 . . . . . . . . .
Chapter 4
4–1 6 . . . . . . . . . . . . . . . .
4–2 0 . . . . . . . . . . . . . . . .
4–3 2 . . . . . . . . . . . . . . . .
4–4 1 . . . . . . . . . . . . . . . .
4–5 0 . . . . . . . . . . . . . . . .
4–6 1 . . . . . . . . . . . . . . . .
4–7 0 . . . . . . . . . . . . . . . .
4–8 to 4–10 1 . . . . . . . . .
4–11 to 4–13 0 . . . . . . . .
4–14 1 . . . . . . . . . . . . . . .
4–15 0 . . . . . . . . . . . . . . .
4–16 6 . . . . . . . . . . . . . . .
4–17 0 . . . . . . . . . . . . . . .
4–18 5 . . . . . . . . . . . . . . .
4–19 1 . . . . . . . . . . . . . . .
4–20 to 4–24 5 . . . . . . .
4–25 to 4–26 9 . . . . . . .
4–27 to 4–32 5 . . . . . . .
4–33 to 4–46 6 . . . . . . .
Chapter 5
5–1 6 . . . . . . . . . . . . . . . .
5–2 0 . . . . . . . . . . . . . . . .
5–3 to 5–4 1 . . . . . . . . . .
5–5 9 . . . . . . . . . . . . . . . .
5–6 to 5–7 6 . . . . . . . . . .
5–8 to 5–9 9 . . . . . . . . . .
5–10 to 5–1 1 6 . . . . . . . .
5–12 to 5–13 0 . . . . . . .
5–14 to 5–15 2 . . . . . . .
5–16 to 5–17 6 . . . . . . .
5–18 9 . . . . . . . . . . . . . . .
5–19 8 . . . . . . . . . . . . . . .
5–20 9 . . . . . . . . . . . . . . .
Chapter 6
6–1 to 6–288 9 . . . . . . .
Chapter 7
7–1 5 . . . . . . . . . . . . . . . .
7–2 0 . . . . . . . . . . . . . . . .
7–3 4 . . . . . . . . . . . . . . . .
7–4 to 7–6 2 . . . . . . . . . .
7–7 to 7–20 5 . . . . . . . . .
Chapter 8
8–1 to 8–3 5 . . . . . . . . . .
8–4 0 . . . . . . . . . . . . . . . .
8–5 to 8–7 5 . . . . . . . . . .
8–8 to 8–15 0 . . . . . . . . .
8–16 5 . . . . . . . . . . . . . . .
8–17 0 . . . . . . . . . . . . . . .
8–18 5 . . . . . . . . . . . . . . .
8–19 to 8–20 4 . . . . . . .
8–21 5 . . . . . . . . . . . . . . .
8–22 0 . . . . . . . . . . . . . . .
8–23 5 . . . . . . . . . . . . . . .
8–24 0 . . . . . . . . . . . . . . .
8–25 to 8–26 5 . . . . . . .
8–27 0 . . . . . . . . . . . . . . .
8–28 5 . . . . . . . . . . . . . . .
8–29 0 . . . . . . . . . . . . . . .
8–30 to 8–88 5 . . . . . . .
A
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 8
2127661
TABLE OF CONTENTS
SECTION TITLE PAGE
T ABLE OF CONTENTS i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 1 – INTRODUCTION
1–1 SER VICE MANUAL CONTENTS 1–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2 SAFETY 1–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–1 Warnings 1–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–2 Specifications 1–17 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3 EMC (Electromagnetic Compatibility) 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–1 EMC Performance 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–2 Notice upon Installation of Product 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–3 General Notice 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–4 Countermeasures against EMC-related Issues 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–5 Notice on Service 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–4 ADDRESS 1–20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 2 – INSTALLATION
2–1 PREINSTALLATION 2–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–1 Introduction 2–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–2 Power Line Requirements 2–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–3 Physical Specifications 2–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–1–4 Recommended Ultrasound Room Layout 2–5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2 INSTALLA TION 2–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–1 Introduction 2–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–2 Average Installation Time 2–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–3 Installation Warnings 2–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–4 Checking the Components 2–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–5 Unpacking LOGIQ 400 2–9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–6 Probe Cable Arm Installation 2–12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–7 MTZ Probe Holder Installation 2–13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–8 Transducer Connection 2–13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–9 Powering-Up Procedure 2–14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–10 Moving into Position 2–15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–11 Adjusting System Clock 2–15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–12 Product Locator Installation Card 2–16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
i
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 3 – SYSTEM CONFIGURATION
3–1 INTRODUCTION 3–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2 DIMENSIONS 3–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 ELECTRICAL SPECIFICATIONS 3–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3–1 Power Supply 3–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3–2 Facility Power Receptacle 3–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–4 ST ORAGE AND OPERATION REQUIREMENTS 3–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–5 OPTIONAL PERIPHERALS 3–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–5–1 Peripherals/Accessories Connector Panel 3–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–5–2 List of Optional Peripherals 3–11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–5–3 Power Consumption of Optional Peripherals 3–15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–6 VIDEO SPECIFICATIONS 3–16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 4 – FUNCTIONAL CHECKS
4–1 INTRODUCTION 4–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1–1 Required Equipment 4–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2 FUNCTIONAL CHECK PROCEDURES 4–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–1 Basic Controls 4–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3 DIAGNOSTICS 4–10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–1 Service Software Menu 4–10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–2 Diagnosis Test Menu 4–11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–3–3 Utility Menu 4–16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4 POWER SUPPLY ADJUSTMENTS 4–37 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4–1 Power Supply Access 4–38 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–4–2 Power Supply Adjustment Procedure 4–40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 5 – DIAGRAMS
5–1 INTRODUCTION 5–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2 LOGIQ 400 SYSTEM 5–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–3 BLOCK DIAGRAM 5–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–4 WIRING DIAGRAM 5–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–5 POWER SUPPLY BLOCK DIAGRAM 5–12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–6 POWER SUPPLY2 BLOCK DIAGRAM 5–14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–7 POWER SUPPLY2 BLOCK DIAGRAM 5–16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–8 CIRCUIT BOARD DESCRIPTION 5–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 6 – RENEWAL PARTS
6–1 RENEWAL PARTS 6–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2 DISASSEMBLY/RE–ASSEMBLY 6–69 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–1 Monitor Assy (FRU No. 100) 6–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–2 Monitor Cover Set (FRU No. 101) 6–72 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–3 Escutcheon Latch (FRU No. 103) 6–76 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–4 Escutcheon Front Door (FRU No. 107) 6–77 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–5 CRT Cap Set (FRU No. 108) 6–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–6 CRT Filter (FRU No. 109), CRT Filter Clamp Set (FRU No. 110) 6–79 . . . . . . . . . . . . . . . . . . .
6–2–7 Accessory Assy (FRU No. 111) 6–80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–8 Speaker Assy (FRU No. 112) 6–82 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–9 CRT Lamp Assy (FRU No. 113), CRT Lamp (FRU No. 114) 6–84 . . . . . . . . . . . . . . . . . . . . . . .
6–2–10 CRT Assy (FRU No. 115) 6–86 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–11 CRT Cable Set (FRU No. 116) 6–88 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–12 Monitor Assy NTSC (FRU No. 150), Monitor Assy PAL (FRU No. 151) 6–90 . . . . . . . . . . . . .
6–2–13 Monitor Cover (FRU No. 152) 6–92 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–14 Escutcheon Assy (FRU No. 153) 6–93 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–15 Fixing Metal Plate (FRU No. 154) 6–96 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–16 CRT Filter (FRU No. 155), CRT Filter Clamp Set (FRU No. 156) 6–97 . . . . . . . . . . . . . . . . . .
6–2–17 Monitor Bottom Assy (FRU No. 157) 6–98 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–18 CRT Assy NTSC (FRU No. 158), CRT Assy PAL (FRU No. 159) 6–100 . . . . . . . . . . . . . . . . .
6–2–19 CRT Cable Set (FRU No. 160) 6–102 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–20 CRT Lamp (FRU No. 161) 6–103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–21 Monitor15 Assy (FRU No. 170) 6–104 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–22 Monitor15 Cover Set (FRU No. 171) 6–106 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–23 Speaker Assy (FRU No. 178) 6–109 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–24 Task Lamp (FRU No. 180) 6–110 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–25 CRT Cable Set (FRU No. 182) 6–111 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–26 Neck Assy (FRU No. 201 for Color Monitor, No. 202 for B/W Monitor) 6–114 . . . . . . . . . . . . .
6–2–27 Neck Grip (FRU No. 203) 6–117 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–28 Neck Space Plate (FRU No. 204) 6–120 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–29 Neck Space Plate 2 (FRU No. 205) 6–122 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–30 Rotation Spacer (FRU No. 206) 6–124 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–31 Side Cover Left (FRU No. 301) 6–127 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–32 Side Cover Right (FRU No. 302) 6–128 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–33 Rear Cover (FRU No. 303) 6–129 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–34 Rear Door Assy (FRU No. 304) 6–130 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–35 Front Base Cover (FRU No. 305) 6–131 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–36 Front Cover (FRU No. 306) 6–132 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–37 Keyboard Bottom Cover (FRU No. 307) 6–134 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–38 Top Cover (FRU No. 308) 6–135 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–39 Front Bumper Set (FRU No. 310) 6–136 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–40 Corner Guard (FRU No. 311) 6–137 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–41 Probe Holder (FRU No. 312) 6–138 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iii
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 6 – RENEWAL PARTS (continued)
6–2–42 Gel Holder (FRU No. 313), Gel Holder Bottom (FRU No. 314) 6–139 . . . . . . . . . . . . . . . . . . .
6–2–43 Handle (FRU No. 315) 6–140 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–44 Air Filter (FRU No. 316) 6–141 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–45 Cable Arm Assy (FRU No. 317) 6–142 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–46 Front Caster Assy (FRU No. 318) 6–143 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–47 Rear Caster Assy (FRU No. 319) 6–144 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–48 Caster Cover Assy (FRU No. 320) 6–145 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–49 Keyboard Panel Assy (FRU No. 400) 6–147 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–50 Keyboard Assy (FRU No. 401), Keyboard Cover (FRU No. 402),
Key Sheet (FRU No. 403) 6–148 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–51 Keyboard Knob Set (FRU No. 404) 6–151 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–52 Trackball (FRU No. 405) 6–152 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–53 Gain Encoder (FRU No. 410) 6–153 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–54 Keyboard Panel Assy (FRU No. 450) 6–155 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–55 Keyboard Cover Assy (FRU No. 451), SW PWB (FRU No. 462),
Rubber Key (FRU No. 452, 453, 454, and 455) 6–156 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–56 Keyboard Knob Set (FRU No. 464) 6–159 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–57 Trackball (FRU No. 457) 6–160 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–58 TGC Assy (FRU No. 458) 6–161 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–59 Freeze Key Assy (FRU No. 459) 6–162 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–60 Rotary Encoder (FRU No. 460) 6–163 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–61 Rear CONN Panel Assy (FRU No. 501) 6–166 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–62 CNTIF Assy (FRU No. 502) 6–168 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–63 AVIF Assy or BVIF Assy (FRU No. 503) 6–169 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–64 Circuit Protector (FRU No. 504: 15A, FRU No. 505: 7.5A) 6–170 . . . . . . . . . . . . . . . . . . . . . . .
6–2–65 ECG Board Assy (FRU No. 506) 6–171 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–66 ECG Panel Assy (FRU No. 507) 6–172 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–67 Nest Fan Assy (FRU No. 511) 6–174 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–68 Probe CONN Set Assy
(FRU No. 512 for 3 slots model, FRU No. 513 for 2 slots model) 6–176 . . . . . . . . . . . . . . . . .
6–2–69 Connector Cover (FRU No. 514) 6–178 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–70 Shield Finger Long (FRU No. 515), Shield Finger Short (FRU No. 516) 6–179 . . . . . . . . . . . .
6–2–71 Probe CONN 1 Assy (FRU No. 517) or DCON Assy (FRU No. 518) 6–180 . . . . . . . . . . . . . . .
6–2–72 PRAG Assy (FRU No. 519) 6–182 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–73 MODD (FRU No. 520), MODD Fan (FRU No. 521),
MODD Holder Assy (FRU No. 522) 6–184 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–74 HDD Assy (FRU No. 523), HDD Holder Assy (FRU No. 524) 6–189 . . . . . . . . . . . . . . . . . . . . .
6–2–75 HDDB Assy (FRU No. 525) 6–195 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–76 HDD LED Assy (FRU No. 526) 6–196 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–77 Power Switch Assy (FRU No. 527) 6–197 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–78 P.C. Board(s) (FRU No. 601 through 612) 6–198 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–79 Time Keeper RAM (FRU No. 613) 6–200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–80 Time Keeper Battery (FRU No. 613B) 6–202 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–81 Time Keeper Battery (for MVME167–002B) (FRU No. 613C) 6–204 . . . . . . . . . . . . . . . . . . . . .
6–2–82 OMEM Assy (FRU No. 614) 6–206 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iv
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 6 – RENEWAL PARTS (continued)
6–2–83 Nest Mother Assy (FRU No. 615) 6–208 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–84 Mother IC (FRU No. 616) 6–210 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–85 High Voltage Assy (FRU No. 701) 6–214 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–86 HV Fan (FRU No. 702) 6–216 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–87 Low Voltage Unit (FRU No. 703) 6–218 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–88 Power Control Unit (FRU No. 704) 6–220 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–89 SSR Unit (FRU No. 705) 6–222 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–90 Power Inlet Unit (FRU No. 706 for 115V , FRU No. 707 for 220V) 6–224 . . . . . . . . . . . . . . . . .
6–2–91 PW Air Filter (FRU No. 708) 6–226 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–92 Filter Cover Set (FRU No. 709) 6–227 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–93 HV Unit (FRU No. 750) 6–230 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–94 Cooling Fan (FRU No. 751) 6–232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–95 LV2 Unit (FRU No. 752) 6–234 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–96 TRIAC Unit (FRU No. 753) 6–236 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2–97 Power Inlet Unit2 (FRU No. 754 for 115V , FRU No. 755 for 220V) 6–238 . . . . . . . . . . . . . . . .
6–2–98 LV3 Unit (FRU No. 770) 6–240 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–3 FUSE REPLACEMENT 6–243 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–3–1 Introduction 6–243 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–3–2 Replacement Procedures 6–244 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4 SYSTEM SOFTWARE INSTALLATION 6–251 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4–1 Introduction 6–251 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4–2 Preparing before Installation 6–252 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4–3 Initializing Hard Disk 6–253 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4–4 Installing Software 6–254 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4–5 Final Procedures 6–255 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–5 POWER SUPPLY REPLACEMENT 6–257 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–5–1 Introduction 6–257 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–5–2 Replacement Procedures 6–257 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–6 REPLACING 2.5–INCH HDD ASSY WITH 3.5–INCH HDD ASSY 6–263 . . . . . . . . . . . . . . . . . . . . . . . . .
6–6–1 Introduction 6–263 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–6–2 Time Required 6–263 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–6–3 Parts Required 6–263 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–6–4 Procedures 6–263 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–7 REPLACING LV2 UNIT WITH LV3 UNIT 6–276 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–7–1 Introduction 6–276 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–7–2 Time Required 6–276 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–7–3 Parts Required 6–276 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–7–4 Procedures 6–276 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
v
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 7 – PERIODIC MAINTENANCE
7–1 INTRODUCTION 7–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–1–1 Periodic Maintenance 7–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2 PERIODIC MAINTENANCE PROCEDURE 7–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2–1 Visual Inspection 7–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2–2 Cleaning 7–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2–3 Measurement 7–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2–4 User Data Backup 7–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–2–5 Note 7–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3 ELECTRICAL SAFETY TEST 7–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–1 Outlet Test Wiring Arrangement 7–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–2 Ground Continuity 7–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–3 Chassis Leakage Current Test 7–9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–4 Probe Leakage Current Test 7–12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–5 ECG Leakage Current Test 7–16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7–3–6 When There’s Too Much Leakage Current 7–20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 8 – OPTIONS
8–1 INTRODUCTION 8–5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2 VIDEO CASSETTE RECORDER INSTALLATION 8–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–1 Foreword 8–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–2 Tools Required 8–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–3 Time Required 8–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–4 Parts Required 8–6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–5 Functional Check–out 8–7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–6 Installing VCR on Console 8–8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–7 Installing VCR on Color Monitor 8–13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–8 Connecting Cables 8–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–9 Setting DIP Switches 8–20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–10 Operational Check-out 8–20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–11 Final Procedures 8–23 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2–12 Renewal Parts 8–24 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vi
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 8 – OPTIONS (continued)
8–3 COLOR VIDEO PRINTER INSTALLATION 8–25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–1 Foreword 8–25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–2 Tools Required 8–25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–3 Time Required 8–25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–4 Parts Required 8–25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–5 Functional Check–out 8–26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–6 Installing Color Video Printer on Console 8–26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–7 Connecting Cables 8–31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–8 Setting DIP Switches 8–33 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–9 Setting Parameters of UP–2950 Series Printers 8–33 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–10 Operational Check–out 8–36 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–11 Final Procedures 8–40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–12 Renewal Parts 8–41 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4 B/W PRINTER INSTALLATION 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–1 Foreword 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–2 Tools Required 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–3 Time Required 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–4 Parts Required 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–5 Functional Check–out 8–42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–6 Setting DIP Switches 8–43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–7 Installing B/W Video Printer on Console 8–43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–8 Connecting Cables 8–47 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–9 Operational Check–out 8–48 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–10 Final Procedures 8–51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–4–11 Renewal Parts 8–51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5 ECG OPTION INSTALLATION 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–1 Foreword 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–2 Tools Required 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–3 Time Required 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–4 Parts Required 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–5 Functional Check–out 8–52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–6 Installing ECG Board Assy 8–53 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–7 Installing ECG Panel Assy 8–55 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–8 Operational Check–out 8–57 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–9 Attaching Caution Label 8–66 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–10 Final Procedures 8–67 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–5–11 Renewal Parts 8–69 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vii
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 8 – OPTIONS (continued)
8–6 EXPANDED CINE MEMORY INSTALLATION 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–1 Foreword 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–2 Tools Required 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–3 Time Required 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–4 Parts Required 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–5 Functional Check–out 8–70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–6 Accessing Board Assy 8–71 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–7 Installing OMEM Assy 8–72 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–8 Operational Check–out 8–74 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–9 Final Procedures 8–75 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–6–10 Renewal Parts 8–77 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7 FOOT SWITCH INSTALLATION 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–1 Foreword 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–2 Tools Required 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–3 Time Required 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–4 Parts Required 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–5 Functional Check–out 8–78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–6 Connecting Foot Switch 8–79 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–7 Operational Check-out 8–80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–7–8 Final Procedures 8–80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8 SWIVEL LOCK INSTALLATION 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–1 Foreword 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–2 Tools Required 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–3 Time Required 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–4 Parts Required 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–5 Functional Check–out 8–81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–6 Installing Swivel Lock 8–82 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–7 Operational Check-out 8–84 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–8–8 Final Procedures 8–84 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
viii
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 9
SECTION TITLE PAGE
2127661
CHAPTER 8 – OPTIONS (continued)
8–9 LEFT SIDE PROBE HOLDER INSTALLATION 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–1 Foreword 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–2 Tools Required 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–3 Time Required 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–4 Parts Required 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–5 Installing Left Side Probe Holder 8–85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–6 Operational Check-out 8–86 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–9–7 Final Procedures 8–86 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10 MTZ PROBE HOLDER INST ALLATION 8–87 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–1 Foreword 8–87 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–2 Tools Required 8–87 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–3 Time Required 8–87 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–4 Parts Required 8–87 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–5 Installing MTZ Probe Holder 8–88 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–10–6 Final Procedures 8–88 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ix
T ABLE OF CONTENTS
REV 9
This page is left blank intentionally.
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
x
T ABLE OF CONTENTS
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 8
2127661
CHAPTER 1 – INTRODUCTION
TABLE OF CONTENTS
SECTION TITLE PAGE
1–1 SER VICE MANUAL CONTENTS 1–3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2 SAFETY 1–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–1 Warnings 1–4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–2 Specifications 1–17 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3 EMC (Electromagnetic Compatibility) 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–1 EMC Performance 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–2 Notice upon Installation of Product 1–18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–3 General Notice 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–4 Countermeasures against EMC-related Issues 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–5 Notice on Service 1–19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–4 ADDRESS 1–20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–1
INTRODUCTION
REV 0
This page is left blank intentionally.
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
1–2
INTRODUCTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 0
2127661
1–1 SERVICE MANUAL CONTENTS
This manual provides service information on the LOGIQ 400 Ultrasound Scanning System. It contains the following
chapters:
1. Chapter 1, Introduction: Contains a content summary and warnings;
2. Chapter 2, Installation: Contains physical and electrical requirements that must be considered prior to installation and a complete LOGIQ 400 installation procedure with installation checklist;
3. Chapter 3, System Configuration: Contains system configuration and specifications;
4. Chapter 4, Functional Checks: Contains functional checks that must be performed as part of the installation, or
as required during servicing and periodic maintenance;
5. Chapter 5, Diagrams: Contains block diagrams and functional explanations of the LOGIQ 400 electronics;
6. Chapter 6, Renewal Parts: Contains a complete list of replacement parts for the LOGIQ 400 and disassembly
procedures for all changeable FRU;
7. Chapter 7, Periodic Maintenance: Provides periodic maintenance procedures for the LOGIQ 400.
8. Chapter 8, Options: Provides installation procedures and changeable FRU for the optional devices.
1–3
INTRODUCTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 2
1–2 SAFETY
1–2–1 Warnings
2127661
W ARNING!
CAREFULLY READ ALL THE WARNINGS LISTED BELOW!
1. The operator manual should be fully read and understood before operating the LOGIQ 400 and kept nearby for
quick reference.
2. Although the ultrasound energy transmitted from the LOGIQ 400 transducer is within AIUM/NEMA standards,
unnecessary exposure should be avoided. Only trained personnel should operate the LOGIQ 400.
3. T o prevent electrical shock, the LOGIQ 400 should be connected to a properly grounded power receptacle. Do
not use a three prong to two prong adapter. This defeats safety grounding.
4. Do not use with Defibrillator when LOGIQ 400 is being operated .
5. Probes are fragile, please handle with care.
6. Concerning Outside Markings, refer to Illustration 1–1, 1–2, 1–3, 1–4, 1–5, 1–6, and 1–7.
7. For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ 400 User Manual and Caution
Sheet supplied with each probe.
NOTICE
This medical equipment is approved, in terms of the prevention of radio wave interference, to be used
in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment in an inappropriate environment may cause some electronic interference to radios and televisions around the equipment. Proper handling of this equipment is required in order to avoid such
trouble according to the operator and service manuals.
This equipment can be used in residential areas only under the supervision of physicians or qualified
technicians.
CAUTION
Improper performance possibility. Do not use the following devices near this equipment.
Cellular phone, radio transceiver, mobile radio transmitter, radio-controlled toy, etc.
Use of these devices near this equipment could cause this equipment to perform outside the
published specifications. Keep power to these devices turned off when near this equipment.
1–4
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
OUTSIDE MARKINGS OF LOGIQ 400 (For Color Monitor Models)
For further details regarding the cautions above, refer to 2–2–10 MOVING INTO POSITION in Chapter 2.
ILLUSTRATION 1–1
Note
1–5
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
OUTSIDE MARKINGS OF LOGIQ 400 (For B/W Monitor Models with S/W V3.40 or earlier)
For further details regarding the cautions above, refer to 2–2–10 MOVING INTO POSITION in Chapter 2.
B/W system is applied the color monitor from software version 4.01y.
ILLUSTRATION 1–2
Note
Note
1–6
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
Possible Injury. Placing objects on top of the monitor may cause the monitor to tilt with the
falling objects resulting in injury to the operator. Do not place any objects on the monitor.
CAUTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
OUTSIDE MARKINGS OF LOGIQ 400 (For Color Monitor Models)
ILLUSTRATION 1–3
1–7
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
For the symbols shown in the illustration above, refer to latter pages in this chapter.
The CAUTION label for the radio influence is attached on the console from April, 1996.
OUTSIDE MARKINGS OF LOGIQ 400 (For USA)
ILLUSTRATION 1–4
Note
1–8
INTRODUCTION
REV 7
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
For the symbols shown in the illustration above, refer to latter pages in this chapter.
The CAUTION label for the radio influence is attached on the console from April, 1996.
The GOST label is attached on the console from June, 1998.
OUTSIDE MARKINGS OF LOGIQ 400 (For Europe)
ILLUSTRATION 1–5
Note
1–9
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
Labels including English, Russian, Swedish, Danish, Greek, and Turkish
EUROPEAN CAUTION LABELS FOR MAIN CAUTION LABEL
The labels shown in ILLUSTRATION 1–6 are supplied with the consoles for Europe. They shall be
attached on the console over the existing labels as necessary. Refer to the installation instructions
supplied with the labels.
[supplied with European Console]
ILLUSTRATION 1–6
Note
1–10
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
For the symbols shown in the illustration above, refer to latter pages in this chapter.
The CAUTION label for the radio influence is attached on the console from April, 1996.
The Japanese EMC label is attached on the console wiith the software version 4.10 or later instead of
the VCIM label.
OUTSIDE MARKINGS OF LOGIQ 400 (For Japan)
ILLUSTRATION 1–7
Note
1–11
INTRODUCTION
REV 8
1–2–1 Warnings (continued)
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
ILLUSTRATION 1–8 shows the labels attached on the console for Korea.
OUTSIDE MARKINGS OF LOGIQ 400 (For Korea)
ILLUSTRATION 1–8
Note
1–12
INTRODUCTION
REV 8
1–2–1 Warnings (Continued)
Do not use a Defibrillator simultaneously with the ECG, as its excessive voltage will damage
the signal input block of the ECG unit.
CAUTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
OUTSIDE MARKINGS OF LOGIQ 400 (For Units with ECG)
ILLUSTRATION 1–9
Note
This label is attached only on the LOGIQ 400 console with the optional ECG unit.
Labels including English, Russian, Swedish, Danish, Greek, and Turkish
[supplied with European Console]
EUROPEAN LANGUAGE LABEL ON ECG LABEL
ILLUSTRATION 1–10
Note
The labels shown in ILLUSTRA TION 1–10 are supplied with the consoles for Europe. They shall be
attached on the console over the existing labels as necessary. Refer to the installation instructions
supplied with the labels.
1–13
INTRODUCTION
REV 8
1–2–1 Warnings (Continued)
HAZARDOUS VOLTAGE. 140VDC CAN CAUSE A SEVERE INJURY OR DEATH, OR THE
POWER SUPPLY TO BE DAMAGED. TURN OFF THE POWER AND CHECK THE RESIDUAL
VOL TAGE OF CAPACITORS BEFORE ACCESSING THE POWER SUPPLY UNIT. CAREFULLY
WORK WHILE ACCESSING THE POWER SUPPLY UNIT.
WARNING!
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
OUTSIDE MARKINGS OF LOGIQ 400 (ON POWER SUPPLY BOX)
Same labels are attached on both left and right outside of the power supply unit.
ILLUSTRATION 1–11
Note
1–14
INTRODUCTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 8
2127661
1–2–1 Warnings (Continued)
The following table describes the purpose and location of safety labels and other important information provided on the
equipment.
Label/Symbol Purpose/Meaning Location
Identification and Rating
Plate
Type/Class Label
Device Listing/
Certification Labels
”DANGER – Risk of explosion
used in...”
• Manufacturer’s name and address
• Date of manufacture
• Model and serial numbers
• Electrical ratings
Used to indicate the degree of safety
or protection.
Equipment Type BF (man in the box
symbol) IEC 878–02–03 indicates B
Type equipment having a floating applied part.
Equipment Type CF (heart in the box
symbol) IEC 878–02–05 indicates
equipment having a floating applied
part having a degree of protection
suitable for direct cardiac contact.
Laboratory logo or labels denoting
conformance with industry safety
standards such as UL or IEC.
The system is not designed for use
with flammable anesthetic gases.
• ”CAUTION” The equilateral
triangle is usually used in combination with other symbols to advise or
warn the user.
Rear of console near power
inlet
Probe connectors
and PCG connector
ECG connector and surgical
probes
Rear of console
Rear of console
Various
• ”ATTENTION – Consult accompanying documents ” is intended to alert
the user to refer to the operator
manual or other instructions when
complete information cannot be provided on the label.
• ”WARNING – Dangerous voltage”
(the lightning flash with arrowhead) is
used to indicate electric shock hazards.
1–15
Various
Left and right side of power
supply unit
INTRODUCTION
REV 8
1–2–1 Warnings (Continued)
Label/Symbol Purpose/Meaning Location
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
2127661
• ”Mains OFF” Indicates the power
off position of the mains power
switch.
• ”OFF/Standby” Indicates the power off/standby position of the power
switch.
CAUTION
This Power Switch DOES NOT ISOLATE Mains Supply
• ”Mains ON” Indicates the power
on position of the mains power
switch.
• ”ON” Indicates the power on position of the power switch.
CAUTION
This Power Switch DOES NOT ISOLATE Mains Supply
• ”Protective Earth” Indicates the
protective earth (grounding) terminal.
• ”Equipotentiality” Indicates the terminal to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment.
Rear of system
Adjacent to mains switch
Adjacent to
On–Off/Standby Switch
Rear of system
Adjacent to mains switch
Adjacent to
On–Off/Standby Switch
Not used
Rear of console
• ”Non-Ionizing Radiation” indicates
that the system applies RF energy.
1–16
Rear of console near power
inlet
INTRODUCTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 8
1–2–2 Specifications
Type of protection against electric shock: Class I EQUIPMENT (*1)
Degree of protection against electric shock: Type BF EQUIPMENT (*2) (Except ECG)
Type CF EQUIPMENT (*3) (ECG Only)
Ordinary Equipment
Continuous Operation
*1. Class I EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULA TION only, but which
includes an additional safety precaution in that means are provided for the connection of the EQUIPMENT to
the protective earth conductor in the fixed wiring of the installation in such a way that ACCESSIBLE MET AL
PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION.
*2. Type BF EQUIPMENT
TYPE B EQUIPMENT with an F–TYPE APPLIED P ART
TYPE B EQUIPMENT: EQUIPMENT providing a particular degree of protection against electric shock, particularly regarding:
2127661
– allowable LEAKAGE CURRENT ;
Normal mode Single failure mode
Patient leakage current Less than 100µ A Less than 500µA
*3. Type CF EQUIPMENT
EQUIPMENT providing a particular degree of protection higher than that for TYPE OF BF EQUIPMENT
against electric shock particularly regarding allowable LEAKAGE CURRENT, and having an F–TYPE APPLIED PART.
– allowable LEAKAGE CURRENT ;
Normal mode Single failure mode
Patient leakage current Less than 10µ A Less than 50µA
1–17
INTRODUCTION
LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS
REV 8
2127661
1–3 EMC (Electromagnetic Compatibility)
1–3–1 EMC Performance
All types of electronic equipment may characteristically cause electromagnetic interference with other equipment,
either through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the
equipment, which curbs electromagnetic influence from other equipment and at the same time does not affect other
equipment with similar electromagnetic radiation from itself.
This product is designed to fully comply with the EN60601–1–2 (IEC601–1–2), in Medical electrical equipment EMC
regulations.
Proper installation following this service manual is required in order to achieve the full EMC performance of the product.
The product must be installed as stipulated in 1–3–2, Notice upon Installation of Product.
In case of issues related to EMC, please follow procedures stated in 1–3–4, Countermeasures against EMC-related
Issues.
1–3–2 Notice upon Installation of Product
1) Use either power supply cords provided by GEMS or ones designated by GEMS. Products equipped with
power source plug should be plugged into the fixed power socket which has the protective grounding conductor.
Connect a three-pole plug to a three-pole socket without using a three-pole-to-two-pole converter.
2) Locate the equipment as far as possible from other electronic equipment.
3) Be sure to use either any cables provided by GEMS or ones designated by GEYMS. Wire these cables follow-
ing these installation procedures.
(Example) Wire power cables separately from signal cables.
4) Lay out the main equipment and other peripherals following the installation procedures described in Chapter2,
INST ALLATION.
1–18
INTRODUCTION