GE LightSpeed 16 Product Data Sheet

g
Product Data Sheet – rev 4.2 July 4th 2003 Page 1
Introduction
The LightSpeed16 CT Scanner with
TM
Xtream
GE’s CT Continuum and its award­winning LightSpeed CT platform. The LightSpeed acquires 16 sub-millimeter slices in one rotation, at a scan speed of 0.5 seconds. As with the very first LightSpeed product introduction, the LightSpeed Designed for Six Sigma (DFSS methodology in all aspects of manufacturing and engineering.
LightSpeed system or as a cost-effective upgrade to an existing LightSpeed system.
Primary Benefits
Routine use of sub-millimeter slices
Small lesion and small vessels
The 6.3 MHU Performix X-Ray Tube
technology is the next step in
16
CT Scanner routinely
16
utilizes
TM
)
16
is available both as a new
without image noise or coverage compromise.
assessments – pancreas, liver or circle of willis, renal arteries, coronary arteries and peripheral vascular arteries.
demonstrates the highest reliability and delivers the power you need for image quality. The Performix Tube now comes standard with a one-year, non pro-rated warranty.
LightSpeed16
With Xtream
CT Scanner System
TM
microVoxel
reformatted 2D resolution through the optimum choice of sub-millimeter slice thickness and reconstructed voxel size.
Unmatched Volume Image Quality specification in all planes (sagittal, coronal and axial): 15.4 lp/cm @ 0% MTF @ 0.5 sec. Rotation
HiLight
deliver consistent image quality with its 21,888 individual elements: 16 rows of
0.625mm thickness and 8 rows of 1.25mm thickness
Xtream
workflow design, breaks through existing limits on speed, image quality and flexibility to provide an optimized workflow solution from acquisition to final report.
Lowest dose acquisition using SmartmA, an automatic modulation technique and DLP display that tells the operator the dose before the scan starts.
HyperPlane
tackles the inherent issues with fast, high pitch helical scanning by reducing helical artifacts and dose, and optimizing the slice profile.
CrossBeam
solves the technical challenge of cone beam artifacts.
Large breadth of Advanced Software Applications: Advanced Lung Analysis, CT
Colonography, Advanced Vessel Analysis, Brain and Body Perfusion, Cardiac Analysis,
imaging improves 3D and
TM
Matrix II Detector is designed to
TM
technology, a revolutionary new
TM
reconstruction algorithm
TM
reconstruction algorithm
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Cardiac Function, SmartScore Pro and Advantage Sim.
Compact system design of the LightSpeed 28m² (302 square feet).
Productivity features designed for the CT Technologist: In-Room Start, Remote Gantry Tilt, Breathing Lights with countdown timer, Gantry Controls mounted on all four corners of the gantry and an Integrated IV Pole at the foot of the table.
1700 mm scannable range for full body scans.
254 GB Disk (system, image, scan disks) stores up to 250,000 512 scan data files.
Cardiac acquisition with Segment, Burst and Burst Plus enable you to scan patients
with heart rates from 40 to 110 BPM, and provide for temporal resolution up to 65 ms.
Color Coding for Kids, winner of a National Heroes Award from the Emergency Medical Services for Children, provides pediatric scan protocols based on the Broselow-Luten Color Coding system is incorporated into the protocol selection on the operator’s console and is designed to facilitate pediatric emergency care and reduce medical errors.
VariViewer
review mode that can change the slice thickness reconstruction instantaneously, resulting in the ability to “scan thin and view thick.”
16
allows for installation in only
2
images and 2881
TM
Pediatric System. This
TM
: provides an interactive axial
Clinical Performance
Chest / Abdomen / Pelvis
TM
Coverage 600 mm Rotation 0.5 sec. Mode 16 x 0.625 mm Pitch 1.75:1 mA 420 mAs 210 Speed 35 mm/s Scan Time 17.1 seconds
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 2
Peripheral Run-Off
Coverage 1,400 mm Rotation 0.5 sec. Mode 16 x 1.25 mm Pitch 1.75:1 mA 380 mAs 190 Speed 70 mm/s Scan Time 20 seconds
High Resolution Chest
Coverage 200 mm Rotation 0.5 sec. Mode 16 x 0.625 mm Pitch 1.75:1 mA 440 mAs 220 Speed 35 mm/s Scan Time 5.7 seconds
The net result is that in many cases, helical scans on the Lightspeed faster than 4-slice CT systems. With the Lightspeed second scan speed and 0.5625:1, 0.9375:1,
1.375:1, and 1.75:1 helical pitches. This added performance, at the same image quality, may allow you to: reduce contrast media, perform better thin-slice CT angiography exams, use thinner slices for most exams, and perform longer helical exams without tube cooling delays.
The LightSpeed performance through a balanced design that focuses on image quality, coverage, and exam speed.
16
Main Features
Full 360° rotation speed of 0.5 sec. provides for improved temporal resolution and image quality in areas such as cardiac imaging;
Faster scan times enable shorter breath holds and more comfortable exams for patients;
Faster scan times reduce the occurrence of re-scans by lowering the possibility of patient motion during the exam;
Routine scanning with a slice thickness of
0.625mm - optimizes lesion detection and facilitates the use of thinner images for volume presentations.
Reduced partial volume artifacts:
16
are up to 12 times
, users can routinely use a 0.5
16
delivers this level of
LightSpeed16
With Xtream
CT Scanner System
Prospective or retrospective image reconstruction of 0.625 mm, 1.25 mm,
2.5 mm, 3.75 mm, 5 mm, 7.5 mm or 10 mm images from inherently thinner acquisition data sets;
For both helical and axial acquisitions.
Image decomposition to:
Retrospective reconstruction of thin
images from data sets where thicker images were initially reconstructed;
Facilitates more detailed image analysis without need for re-scans;
Improves 3D visualization;
For both helical and axial acquisitions
Up to 2,881 multi-slice 1-sec scan rotations
stored on disk. Facilitates retrospective image decomposition during off hours.
Eight fundamental scan modes to simplify the inherently complex nature of multi-slice helical scanning.
Standard set of clinically- proven protocols derived from leading luminary
sites around the world. Up to 8,460 protocols can be edited, modified and stored on the system.
Remote Gantry Tilt from the Operator’s Console to increase exam speed, including built-in safety features to prevent accidental contact of the gantry with the patient.
Built-in patient breathing lights and digital counter provides a goal-oriented approach to coach the patients in holding their breath during an exam.
SmartStart gantry-mounted start scan button and countdown display, facilitates single-technologist operation by allowing start of scan at the gantry, with a visual reminder of time until X-ray initiation.
SmartHelical GE proprietary, non-linear interpolation algorithms, balance slice profile, helical pitch, image noise and required technique.
SmartTrack advanced hardware and software for X-ray beam tracking minimizes patient dose.
SmartTools software automates every exam task in order to increase throughput.
SmartBeam
optimizes X-ray beam filtration independently for body and head applications.
HiRes Chest software provides a single­slice, high-resolution scan mode that minimizes dose for this critical application.
TM
hardware and software
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
16
SmartPrep, standard on LightSpeed provides software for real-time monitoring of contrast enhancement.
* Direct3D, makes 3D routine and instant by building the 3D model during axial image reconstruction, reducing the need for user interaction.
* SmartScorePro, provides EKG-gated hardware and software on the Advantage Workstation for coronary artery calcium scoring.
* CardiacSnapShot, provides software and hardware to perform EKG-gated reconstructions of the heart in SnapShot Segment mode.
- SnapShot Segment is a single sector protocol using information from one heart cycle to generate an image with temporal resolution of 250ms.
- SnapShot Burst is a multi-sector protocol using up to two sectors from two different heart cycles to produce an image with temporal resolution of 125ms.
- SnapShot Burst Plus is a multi-sector protocol using up to four sectors of data from four different heart cycles to produce images with temporal resolution to 65ms.
, DLP (Dose Length Product), and
CTDI
w
Dose Efficiency display during scan
prescription provides patient dose informatbon to the operator.
,
Interactive CT Technology
The LightSpeed exceptional computer and image processing power that significantly enhances clinical productivity, building upon the strength of the LightSpeed family - true interactive CT technology.
Interactive CT embodies a variety of design choices all striving to enhance operator and department productivity. A truly interactive CT system will:
Provide a user interface beyond “intuitive” to become purely “natural” - from the screens to the console hardware itself;
Allow two users to review cases side-by- side, with minimal interference;
Supply a truly multi-tasking environment where even advanced image processing can take place quickly and simultaneously with other processes underway;
Operate with a very high degree of automation, yet allow patient-specific
16
CT Scanner supplies
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 3
changes to be easily made, with virtually no restrictions;
Be as self-teaching as possible, and have on-line access to tutorials as needed.
One key element of this design is to combine some of the best features from several product families into one state-of-the art CT system. For example, the LightSpeed Scanner combines:
SmartTools productivity software to automate every step of the examination, critical for ensuring the highest productivity and throughput possible with the LightSpeed
Large screen interface for controlling scan acquisition easily, with virtually everything at a single glance;
Excellent simultaneity and multi-tasking performance;
Completely protocol-driven scan control with a dramatic reduction in number of screens;
Highly flexible editing tools that allow easy tailoring of the exam to the patient;
Large, 1024 color display;
Leading edge, real-time image processing
(MPR, MPVR, Volume Viewer*).
In summary, primary benefits of Interactive CT on the LightSpeed
A natural scan control user interface
Dramatic reduction in the number of
Easier and more flexible protocols
Much more flexible and intuitive graphic
View/Edit Wizard intuitively adjusts
DynaPlan Plus full screen display
Large color screen
Extensive use of picture icons and color
Large on-screen controls and attractive
16
CT Scanner;
16
Scanner:
screens; only 2 screens to set up first scan and 1 screen for real-time monitoring while scanning;
prescription process with a 1024 Localizer;
dependent parameters automatically in response to operator-initiated changes and highlights them for quick review; also alerts the operator to incompatible dependencies requiring operator intervention;
illustrates scan status pictorially, with real-time feedback.
cues enhance ease of use;
color palette provide comfortable viewing over extended periods.
16
CT
LightSpeed16
With Xtream
CT Scanner System
Enhanced multi-tasking allows operators to review more than one exam simultaneously, independently - even with AutoView and AutoFilm on/
BrightBox™ dedicated controls for image next, prior, manual paging and trackball W/L helps make two person image review practical;
Up to four 512x512 images from four different exams can be viewed on a large 1024 color display.
SmartmA™ User Interface - automatically optimizes mA to maintain constant image noise when collimation/detector configuration, scan mode, scan rotation speed, table speed, or image thickness changes. It takes the guesswork out of setting scan technique when changing parameters (note: user must select initial Noise Index as well as maximum mA setting). Noise index enable 100% Image Quality reproducibity from one patient to another and from one user to another.
Protocol Pro™ protocol manager - provides operator control of automated features (like AutoFilm, AutoStore, and AutoTransfer) on per patient basis.
Patient demographics and exam protocols can be preprogrammed in advance of patient arrival through the Schedule Patient feature.
A preprogrammed selection of AutoView and Image Review Layouts allow simple customization of the image presentation to match the anatomical area of interest ­without the complexity of free-form “windows”.
ImageWorks™ provides instant access to advanced image processing features such as MIP, MPR, MPVR, Volume Viewer*, Advanced Vessel Analysis*, CT Perfusion 2* and DentaScan*.
No waiting for network transfer or image retrieval from archive media.
Background filming allows use of the full screen for AutoView and image review/processing without interruption when auto or batch filming. Special “one touch” controls provide on-screen viewing of camera progress during AutoFilm without disrupting other image processes in progress.
ProView™ visualization algorithms available to enhance anatomical structures without additional image reconstruction time.
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Operator console convenient to locate in suite.
Computer, image processor and image reconstruction hardware completely integrated in base of console - no separate computer cabinet to site;
Split table top allows unrestricted patient viewing while still supporting 2 large color monitors;
Front and back work surfaces can be set during installation within a range of vertical heights that help accommodate a variety of siting requirements - especially surrounding the height of the console relative to the window into the scanning suite.
Work surface accommodates a large variety of working conditions and individual operator preferences.
Wide work surface has room for papers, patient charts, logbooks or other peripheral devices;
Wide area also helps make two person operation more comfortable and practical;
All console components (monitors, keyboard, mouse and BrightBox) are free-standing and can be easily moved on the front work surface;
All surfaces made of high-impact resin that hides marks and are easy to clean.
Direct network connection means a multi- suite Ethernet card is not required for a gateway out of the suite - saving costs and simplifying installation.
Learning Solutions provides an on-line Operator manual detailing system operation via a multi-media CD-ROM player integrated into the front of the operator console. Learning Solutions can also be accessed on a stand alone PC providing flexibility and productivity for on demand learning of system operation
Advanced Tube & Generator Technology
The Performix X-ray tube, with 6.3 MHU of storage and 53.2 kW operation, provides increased helical performance with greater patient throughput and virtually no tube cooling. Advanced technology in the Performix tube includes a metal-ceramic frame for long life, a high-speed bearing for sub-second scanning, a high-efficiency motor to accelerate the large anode and
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 4
efficient cooling for high throughput and superior helical performance.
Wide range of technique factors (10 to 440 mA, in 5 mA increments) gives the operator and physician flexibility to tailor technique to specific needs, optimizing patient dose, and providing the power needed to perform a broad spectrum of axial and helical examinations.
System Specifications
Scan Modes:
16
The LightSpeed virtually any clinical application due to its wide variety of scan modes. With the LightSpeed body CT studies are easier to perform and more productive than ever before.
Helical:
Continuous 360° scanning with table incrementation and no interscan delay.
Scans can be acquired in a wide variety of speeds.
Axial:
Up to 16 contiguous axial slices acquired simultaneously with each 360° rotation, with the time between scans set by the user­selected interscan delay (ISD) or intergroup delay (IGD).
Scans may be easily clustered in groups to allow multiple scans in a single breath hold.
Minimum scan-to-scan cycle time of only 1.5 sec with table moves of 10 mm (any scan time).
Scout™:
Single radiographic plane for scan
Extended range matches helical scannable
localization and graphical prescription of prospective reconstruction;
range.
Helical Scans
Slip ring technology has advanced axial scanning by enabling scans with zero interscan delay and simultaneous table movement.
CT system can perform
16
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490
LightSpeed16
With Xtream
Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
CT Scanner System
Helical Multi-slice Modes: Prospective Multiple-
Simplified scan prescriptions and easy-to-use default protocols make the LightSpeed and efficient in patient set up.
Multi-slice acquisitions and short intergroup delays significantly reduce potential mis­registration between scans by increasing the number of scans possible in a patient breath hold. Contrast agents may be better utilized as well due to significantly faster scan acquisition.
Helical protocols are almost identical to “classical” axial scan protocols. At the beginning of a study, the operator selects the
,
type of exam with the anatomical programmer, and indicates the desired scan range - either manually, or from a Scout.
After completing the prescribed exam, the system remains ready to continue with additional helical scans or a set of axial scans.
The operator may reconstruct helical scans prospectively with up to 90% overlap, and retrospectively, at any arbitrary table location in 0.1 mm increments.
Helical Multi-slice Modes
The complex nature of helical multi-slice scanning has been simplified by grouping all critical acquisition parameters within 8 basic scan modes, all optimized for image quality and speed. For eight-slice acquisition: 0.625: 1; 0.875: 1; 1.35: 1 and
1.675: 1. For sixteenslice acquisition:
0.5625: 1; 0.9375: 1; 1.375: 1 and 1.75: 1. These clinically derived multi-slice scan modes offer a wide range of selections that carefully balance acquisition speed, image thickness, artifact level and retrospective image reconstruction flexibility.
This simplified user interface guides the user in the choice of scan parameters. The user selects a pitch mode, a desired image slice thickness and table travel per rotation. The user interface also displays the resulting choice of retrospective image thicknesses available for each choice of acquisition parameters.
The 16 slice helical acquisition modes provide table speeds from 5.625 mm/rotation up to 35 mm per rotation, enabling scan speeds that are up to 12 times faster than 4-slice helical scanners.
16
fast
Thickness Reconstruction:
For all helical scan modes, the operator can choose to reconstruct images prospectively in any of the defined nominal slice thicknesses.
In addition to the initial reconstructed slice thickness, the operator has the option to prospectively specify additional images to be reconstructed from a single raw data set. These images can be reconstructed at any of the defined nominal slice thicknesses available for a given table speed and scan mode.
This effectively facilitates later, more detailed image analysis without additional patient scans and subsequent dose and image registration concerns.
8-SLICE HELICAL MODES
Table Speed (mm/rotation)
Slice
Thick
-ness
(mm)
1.25 6.25 8.75 13.5 16.75
3.75 6.25
Pitch
0.625:1
2.5 6.25
12.5
12.5
5 6.25
12.5
7.5 6.25
12.5
10 6.25
12.5
0.875:1
8.75
17.5
8.75
17.5
8.75
17.5
8.75
17.5
8.75
17.5
1.35:1
13.5
27.0
13.5
27.0
13.5
27.0
13.5
27.0
13.5
27.0
1.675:1
16.75
33.5
16.75
33.5
16.75
33.5
16.75
33.5
16.75
33.5
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 5
16-SLICE HELICAL MODES
Table Speed (mm/rotation)
Slice
Thick-
ness
(mm)
0.625 5.625 9.375 13.75 17.5
1.25 5.625
3.75 5.625
Helical Scan Parameters
Scan Speeds: Full 360° rotational scans in
0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.
Scan Technique
— kVp: 80, 100, 120, 140
— mA: 10 to 440 in 5 mA increments
— Power: 0.8 to 53.2 kW
— Focal Spot Selection:
Small spot for up to 24 kW
Large spot for greater than 24 kW
Single Acquisition maximum.
Helical Tilt with the gantry tilted to a maximum of 30 degree, in half degree increments.
Multiple Acquisition Maximum Multiple scans may be acquired in one series to produce up to 3,000 contiguous helical images. Up to 2,881 seconds helical coverage is possible in multiple series.
Minimum Inter-Group Delay seconds between adjacent helical scans
Pitch
0.5625
:1
11.25
2.5 5.625
11.25
11.25
5 5.625
11.25
7.5 11.25 18.75 27.5 35
10 11.25 18.75 27.5 35
: helical acquisition is possible
0.9375
9.375
18.75
9.375
18.75
9.375
18.75
9.375
18.75
:
: 120 second scan
1.375
:1
13.75
27.5
13.75
27.5
13.75
27.5
13.75
27.5
Scan Time:
(IGD): 5
:1
1. 75:1
17.5
35
17.5
35
17.5
35
17.5
35
LightSpeed16
With Xtream
CT Scanner System
Scan Fields of View
— 25 cm for adult head
— 25, 50 cm for body
— 25 cm for pediatric head
Helical Scan Enhancements
Full simultaneity allows complete image display, processing and analysis, as well as image archival and filming, concurrent with scanning and reconstruction -- even when acquiring helical images in a multi­slice mode.
Confirm Rx to X-Rays on: < 15 sec. for any state of tube and gantry; < 10 sec. with the gantry rotating
Anatomical Programmer: a ten-region anatomical selector allows quick and easy access to 90 user-programmable protocols per region. Separate selector for adult and pediatric exams. There are four selection tabs to select: GE, User, Service and Most Recent Patient. Copy/Paste is supported for easy modification and copying of protocols.
Ten user-defined regions. Each region has one default protocol displayed with the anatomical selector for very fast access to most commonly used protocols
Protocols include preset scan time, kVp, mA, scan mode, slice thickness and spacing, table speed, scan FOV, display FOV and center, recon algorithm and special image acquisition and processing options
Any scan parameters may be edited for each scan or all scans - either before or during an exam. The number of scans may also be easily changed.
AutoScan™: Fully automates longitudinal table movement and start of each scan.
AutoVoice™: 3 preset (English) and 17 user-defined messages automatically deliver patient breathing instructions; especially useful for multiple helical scanning.
Trauma Patient: Allows patient scans and image display/analysis without entering patient data before scanning.
:
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Biopsy: Simplified prescription for single or multiple scans around an arbitrary table position aids biopsy studies.
Advanced Artifact Reduction (AAR) Filter significantly reduces streaking artifacts when highly absorbent objects are in the field of view – ie: large shoulders, screws.
Helical Image Reconstruction
Reconstruction Algorithms: Soft Tissue, Standard, Detail, Bone, Bone Plus, Lung and Edge
Reconstruction Matrix
Display Matrix
Display FOV center, prospective/retrospective target selection.
CT Number Scale
Helical Reconstruction Times:
Reconstruction time as fast 6
Typical 0.167 sec. image-to-image
Maximum image-to-image cycle
Iterative bone processing
Minimum DFOV:
Minimum Pixel Size:
Queued Recon processed continuously and simultaneously with other processes on the system including scanning. Prospective recon will be prioritized over retrospective recon.
Priority Recon Queuing selection marks most recent rotation for
: Freely variable center/off-
images per second
recon in normal 16 slice recon mode.
time is ± 10% for prospective and retrospective image-to-image display. This applies for 512 matrix; any display FOV; in AutoView (all layouts); with concurrent filming and image archival for all scan modes.
increases time by 250 milliseconds. Iterative bone processing, which is always enabled for adult head scanning, reduces image artifacts in head scans stemming from X-ray beam hardening effects.
: 512
: 1024.
: -1024 to 3071 HU
9.6 cm
0.19 mm
: Requests will be
: One touch
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 6
next available recon. Available during or after scanning.
Images annotated to indicate continuous scan acquisition with table incrementation:
HE (helical) + Pitch
Table speed
Prospective Multiple Reconstruction (PMR): Up to 3 sets of reconstructions can be pre-programmed as part of the scan protocol prior to acquisition. The operator can select different start/end location, slice thickness, interval, reconstruction algorithms and display fields of view for each reconstruction. This frees the operator from sitting at the console and directly contributes to increased productivity.
Prospective Recon: Operator may initiate full recons at any table location in increments of 1/10 the image thickness; image thickness remains constant.
Retrospective Recon: Operator may initiate full recons at any table location in
0.1 mm increments; image thickness
remains constant.
Retrospective Image Decomposition: The operator has the option to retrospectively decompose the original raw data set and reconstruct additional images at any of the defined nominal image thickness available for a given table speed and scan mode.
LightSpeed16
With Xtream
CT Scanner System
Helical Scan Protocols
All protocols assume 120 kVp scans under typical clinical conditions.
Single Helical Scans:
Scan Time
10 sec. 440 20 390-440 30 330-380 40 290-350 50 260-320 60 230-310 70 210-300 80 180-280 90 170-270 100 160-260 110 150-250 120 140-230
Multiple Helical Scans
(contiguous helical coverage with 5­second IGD):
Scan Time IGD
10 sec. 5 sec. 3 340-420
20 sec. 5 sec. 2 300-360
30 sec. 5 sec. 2 230-310
40 sec. 5 sec. 2 190-290
50 sec. 5 sec. 2 170-270
60 sec. 5 sec. 2 140-240
Contiguous Helical Coverage
Multiple helical scans can be performed in succession with only 5-second delays between helical scans, providing up to 2,881 contiguous rotations (with up to 1,500 images in one series).
(The following chart assumes 120-second helical with 5-sec IGD)
Maximum mA
No.
Scans
4 300-400 5 270-390 6 240-360
3 240-330 4 180-310
3 180-280 4 140-240
3 140-230 4 110-190
3 120-200 4 90-160
3 100-170
Max
mA
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Total Scan Total Elapsed Max Time (sec.) Time (sec.) mA
150 155 100-200 200 205 80-150 450 465 50-70
Helical Scan Image Quality
Volumetric Helical Scan Image Quality
16
With LightSpeed isotropic CT scanner, it is now possible to specify coronal and sagittal image quality:
1. Visual Measurement:
Reformatted resolution is demonstrated on the Nuclear Associates AAPM High Contrast Resolution Insert #76-413.
0.5mm
0.4mm
An Effective 0.5 mm voxel size is clearly seen in coronal and sagittal views.
2. Statistic Measurement:
Volumetric MTF is computed from the X,Y and Z axis of a 0.2mm bead using sagittal and coronal images of the 8” Catphan, MTF module with bead source #CTP445.
Detail Algorithm
MTF (z-axis) is 12.4 lp/cm @ 10%, 15 lp/cm @ 0% which is a 0.333 mm limiting resolution
In Plane Helical Scan Image Quality
For details of scan technique please refer to the technical reference manual.
High-Contrast Spatial Resolution
On GE Performance phantom:
Standard Algorithm
being a sub-millimeter
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 7
0.584 mm limiting resolution
4.0 lp/cm @ 50% MTF
6.5 lp/cm @ 10% MTF
8.5 lp/cm @ 0% MTF
Hi-Res Algorithm (Edge).
0.324 mm limiting resolution
8.5 lp/cm @ 50% MTF
13.0 lp/cm @ 10% MTF
15.4 lp/cm @ 0% MTF
Line pair values decrease with larger focal spot (by 5% with Standard and by 7% with Hi-Res); limiting resolution is unaffected.
Low-Contrast Detectability
On 8 inch (20 cm) CATPHAN phantom: 5 mm @ .3% at 13.3 mGy 3 mm @ .3% at 37.2 mGy
Noise:
On either an AAPM water phantom or GE Quality Assurance phantom:
0.32% +/- 0.03% at 28.5 mGy (2.85 Rad)
CTDI:
On CTDI Head and Body Dose Reference Phantoms:
expressed in mGy/100 mAs for
CTDI
vol
IEC pitch 1: (normalized to a pitch of 1).
Head 20.9 mGy/100 mAs Body 10.6 mGy/100 mAs
LightSpeed16
With Xtream
CT Scanner System
Axial Scans
Multi-slice acquisitions and short interscan delays significantly reduce potential missed registration between scans by increasing the number of scans possible in a patient breath hold. Contrast agents may be better utilized as well due to significantly faster scans.
Axial Multi-slice Prescription
Simplified scan prescriptions and easy-to-use default protocols make the Lightspeed Scanner fast and efficient in patient set-up. Axial protocols are nearly identical to helical scan protocols.
Axial Multi-slice Modes
The LightSpeed sets of 8 or 16 contiguous images in one 360° rotation.
For each rotation of the gantry, the LightSpeed data. There are five reconstruction modes available for creating images from the multi-slice scan data (1i, 2i, 4i, 8i, and 16i). By using 1i, 2i, 4i, and 8i reconstruction modes, scan data can be combined prior to image reconstruction to create slices with reduced partial-volume artifacts. This is particularly useful for posterior-fossa imaging.
1i Mode:
Produces 1 image per rotation
Nominal Thickness: 1.25, 5, 10 mm
2i Mode:
Produces 2 images per rotation
Nominal Thickness: 0.625, 2.5, 5, 7.5, 10
mm
4i Mode:
Produces 4 images per rotation
Nominal Thickness: 2.5, 3.75, 5 mm
8i Mode:
Produces 8 images per rotation
Nominal Thickness: 1.25, 2.5 mm
16i Mode:
Produces 16 images per rotation
16
acquires axial scans in
16
collects 16 rows of scan
16
CT
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Nominal Thickness: 0.625, 1.25 mm
Axial Scan Parameters
Scan Time:
0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0, 3.0 and 4.0 second full scans (360° acquisition)
Scan Technique:
kVp: 80, 100, 120, 140
mA: 10 to 440, in 5 mA increments
Power: 0.8 to 53.2 kW
Focal Spot Selection:
Small spot for up to 24 kW
Large spot for greater than 24 kW
Scan Plane Geometry:
+/- 30° gantry tilt, in 0.5° increments
Longitudinal positioning in 0.01 mm per slice
increment. Gantry display in 0.5 mm increments.
Interscan Delay (ISD):
Minimum ISD with table movements of 0 - 10 mm: 1.0 sec.
Minimum ISD with table movements of more than 10 mm and up to 20 mm: 1.3 sec
User-selectable.
Inter Group Delay (IGD):
Minimum IGD is the same as minimum ISD; also user-selectable.
Scan-to-Scan Cycle:
Minimum scan-to-scan cycle of 1.5 seconds possible for 0.5 sec. scan speed with minimum ISD's.
Scan Fields of View:
25 cm for adult head
25, 50 cm for body
25 cm for pediatric head
Scan with no table incrementation, contiguous image location, or skipped image location are possible. Overlapped axial scans are not possible.
Axial Image Reconstruction
Reconstruction Algorithms: Soft Tissue, Standard, Detail, Bone, Bone Plus, Lung and Edge.
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 8
Reconstruction Matrix: 512
Display Matrix: 1024.
Display FOV: Freely variable center/off-center, prospective/retrospective target selection.
CT Number Scale: -1024 to 3071 HU
Axial Image Reconstruction:
Reconstruction time as fast as 6 images per second.
Typical 0.167 sec. image-to-image recon in 16 slice recon mode.
Maximum image-to-image cycle time is ± 10% for prospective and retrospective image-to-image display. This applies for 512 matrix; any display FOV; in AutoView (all layouts); with concurrent filming and image archival for all scan modes.
Iterative bone processing increases time by 250 milliseconds.
Prospective Multiple Reconstruction (PMR): Up to 3 sets of reconstructions can be pre-programmed as part of the scan protocol prior to acquisition. The operator can select different reconstruction algorithms and display fields of view for each reconstruction. This frees the operator from sitting at the console and directly contributes to increased productivity.
The operator has the option to reconstruct the original raw data set at any of the defined nominal slice thicknesses.
Reconstructions can be prescribed down to 1/16 the original acquisition image thickness for images acquired in the 1i scan mode, down to 1/8 the original thickness for 2i mode, and down to 1/4 the original thickness for 4i mode.
Similarly, additional reconstruction supports partial-volume artifact reduction by reconstructing images with 4, 8, or 16 times the acquisition image thickness.
These reconstruction features effectively facilitate later, more detailed image analysis without additional patient scans and subsequent dose and image registration concerns.
The following table illustrates the retrospective reconstruction image thicknesses available for each acquisition thickness and mode:
LightSpeed16
With Xtream
CT Scanner System
Prospective Perscri pti on
Scan Mode
16 row 0.625
16row 1. 25
8 row 1. 25
8row 2. 5
Axial Scan Protocols
All protocols assume 120 kVp scans under typical clinical conditions.
Standard Scans:
Scan Time ISD mA Scans
1 sec. 1 sec. 440 18-45 0:35-1:27 1 1 400 24-55 0:47-1:49 1 1 360 32-68 1:03-2:15 1 1 340 37-76 1:13-2:31 1 1 320 43-86 1:25-2:49 1 1 300 50-97 1:37-3:11 1 1 280 58-110 1:45-3:37 1 1 260 66-122 1:55-3:59 1 1 240 74-135 2:07-4:21 1 1 220 84-152 2:21-4:49 1 1 200 94-168 2:37-5:21 2 1 200 37-77 1:13-2:33 2 1 180 42-86 1:23-2:51
Cluster Scans (All cluster protocols assume
9-second clusters of five slices, 1-second scans with 1-second interscan delays and 7 seconds between clusters).
Sl i c e
Thi ckness
Rec on
Slice Thicknesses
16i - 0.625mm
8i - 1.25mm
4i - 2. 5mm 2i - 5. 0mm
16i - 1.25mm
8i - 2. 5mm 4i - 5. 0mm
2i - 10mm
8i - 1.25mm
4i - 2. 5mm 2i - 5. 0mm
1i - 10mm 8i - 2. 5mm 4i - 5. 0mm
2i - 10mm
Acquisition
Time
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
mA
440 15-45 (3-11) 0:41-2:49 400 25-55 (5-13) 1:13-3:21 340 35-75 (7-18) 1:45-4:41 320 45-85 (9-20) 2:17-5:13 300 50-95 (10-22) 2:33-5:45 280 60-110 (12-25) 3:05-6:33 260 75-125 (13-28) 3:21-7:21 240 85-145 (14-31) 3:37-8:09 220 100-170 (16-34) 4:09-8:57 200 115-185 (18-38) 4:41-10:01 180 135-250 (22-43) 5:45-11:21
# Scans
(Clusters)
Acquisition
Time
Axial Scan Image Quality
For details of scan technique parameters, please refer to the technical reference manual.
High Contrast Spatial Resolution:
On GE Performance phantom:
Standard Algorithm
0.584 mm limiting resolution
4.0 lp/cm @ 50% MTF
6.5 lp/cm @ 10% MTF
8.5 lp/cm @ 0% MTF
Hi-Res Algorithm
0.324 mm limiting resolution
8.5 lp/cm @ 50% MTF
13.0 lp/cm @ 10% MTF
15.4 lp/cm @ 0% MTF
Line pair values decrease with larger focal spot (by 5% with Standard and by 7% with Hi-Res); limiting resolution is unaffected.
Low-Contrast Detectability
On 8 inch (20 cm) CATPHAN phantom:
5mm @ .3% at 13.3 mGy 3mm @ .3% at 37.2 mGy
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 9
Noise:
On either an AAPM water phantom or GE Quality Assurance phantom:
0.32% +/- 0.03% at 29.3 mGy (2.93 Rad)
CTDI:
On CTDI Head and Body Dose Reference Phantoms:
CTDI expressed in mGy/100 mAs:
Head 16.8 mGy/100 mAs center
17.0 mGy/100 mAs surface
Body 5.50 mGy/100 mAs center
10.0 mGy/100 mAs surface
CTDI
expressed in mGy/100 mAs:
100
Head 18.7 mGy/100 mAs center
17.9 mGy/100 mAs surface Body 5.4 mGy/100 mAs center
11.2 mGy/100 mAs surface
CTDI
expressed in mGy/100 mAs:
vol
Head 18.4 mGy/100 mAs Body 9.3 mGy/100 mAs
Scout Scans
ScoutView™ scans provide excellent detail for anatomical localization in conjunction with scan prescription.
Scan locations may be prescribed at the operator console either graphically (via mouse), or explicitly (keyboard entry) from a Scout scan.
Prescription of scans with multiple gantry angles are also available on a single Scout.
Scout Scan Parameters
Aperture: 1.25 mm effective aperture
Table speed: 100 mm/sec.
Scan Technique:
kVp: 80, 100, 120, 140
mA: 10 to 440 in 5 mA increments
Power: 0.8 to 53.2 kW
LightSpeed16
With Xtream
CT Scanner System
Orientation: AP, RLAT, PA, LLAT (preset); or any angle from 0° - 359° (manually selected).
Axial scan prescription lines indicate scan location to nearest 1 mm table position.
Scouts longer than 1,000 mm are auto minified to fit the display.
User Interface
The Lightspeed16 Operator Console utilizes a computer workstation with the following user interface features:
Two 20-inch monitors
Scan/recon monitor for scan and recon
control with no image display
Image monitor for image display,
analysis, processing, and management
Each monitor provides a 1280 x 1024
high resolution, flicker-free display
Optional 18” LCD Monitors.
Scan control keyboard assembly with
intercom speaker, microphone and volume controls
Three button mouse with mouse pad
BrightBox (trackball assembly)
Two wide work surfaces
All these devices are free-standing and can be easily moved to accommodate a large variety of working conditions and individual operator preferences.
Split table top allows unrestricted patient viewing while still supporting 2 monitors. Each work surface can be adjusted at installation to help accommodate a variety of siting requirements.
Desktop Overview
The user interface utilizes the paradigm of managed work environments for a more intuitive clinical workflow.
Virtually all clinical operations are managed through three “virtual desktops” or applications managers: Exam Rx, ImageWorks and Learning Solutions. Operators can effortlessly move back and forth between these environments simply by clicking on an icon.
XtreamTM technology enhances multi-tasking
architecture and maintains simultaneously all processes so no work is lost or disrupted as desktops are switched.
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Exam Rx:
The Exam Rx desktop environment provides the clinical tools necessary for comfortable, efficient control of patient studies.
These tools include patient scheduling and data entry, exam protocol selection, protocol viewing and editing, scan data acquisition, image reconstruction, image display and routine analysis, AutoFilm or manual filming, AutoStore and AutoTransfer.
ImageWorks:
ImageWorks is a desktop environment designed to take advantage of the Lightspeed and image processor.
Standard features include archive, network and manual film control, as well as some advanced image processing such as multi­planar reformatting (MPR), multi-projection volume rendering (MPVR), and MR image display. It also has optional add-on packages for Volume Viewer*, CT Perfusion2* and DentaScan*.
The ImageWorks desktop also provides a gateway for DICOM 3.0 image transactions, either through a local area network, or via DICOM-formatted MOD media.
Learning Solutions:
The LightSpeed line operator manual via a multi-media CD­ROM player integrated into the front of the operator’s console. Learning Solutions is also viewable on a stand-alone PC providing flexibility and productivity for on-demand learning of system operation.
16
CT Scanner System computer
16
provides an on-screen, on-
Exam Rx
Patient Scheduling
Patient demographics and exam protocols can be pre-programmed in advance of patient arrival by selecting Schedule Patient from the scan/recon monitor. This productivity enhancement allows entry of all or some of a patient’s demographic data, as well as pre­selection of the exam protocol.
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 10
This feature is available any time a patient exam is not currently underway.
This feature uses the same interface as New Patient selection for simplified, consistent programming.
Patient information can be easily recalled to set up an immediate exam via List/Select Scheduled Patient on the scan/recon monitor. Pre-programmed patient exams can also be recalled from the New Patient screen automatically by entering the patient ID number.
Patient Data Entry:
Patient data can be entered as part of New Patient set-up, or can be recalled from the list of pre-scheduled patients.
Trauma Patient ID allows patient scans and image display/analysis without entering patient data before scanning.
Exam Protocol Selection:
One of the main contributions of the LightSpeed department productivity is its simplified exam set-up.
Exam parameter set-up has been greatly
simplified through the exclusive use of protocols
Protocols can be easily selected in one of
three convenient ways:
Two Anatomical Programmers - one for
adults and one for pediatrics - provide quick and easy access to 8,460 user­programmable protocols (total). Each programmer has ten anatomical regions with 90 protocols for each region
Default protocols have been expanded
through Protocol Pro - a “behind the scenes” protocol manager - that allows preselection of automated features like AutoVoice, AutoFilm, AutoStore and AutoTransfer on a per-exam basis.
Protocol Pro also provides preselection of
two different window/level settings per image for AutoFilm and can automatically
16
CT Scanner System to
A large, graphical Anatomical
Programmer located on the New Patient screen
A default list of the “top 10” most
commonly used protocols located near the anatomical programmer
A numerical entry
LightSpeed16
With Xtream
CT Scanner System
display the 1024 Localizer each time a new series is requested.
Default protocols also include preset scan
time, kVp, mA, slice thickness, scan mode, table speed, image interval, gantry tilt, scan field-of-view, display field-of-view and center, recon types, and breath timing parameters.
Any scan parameter can be edited for each
scan or all scans either before or during an exam. Scans can be easily added or removed from the prescription.
Scan/recon control uses only 2 screens to
set up first scan - New Patient and Protocol View/Edit.
Protocol View/Edit:
A single, full screen View/Edit table allows
fast and easy examination and modification of exam parameters before scanning begins
Exam parameters can be changed for just
one scan, or for all scans in a series
When used in conjunction with the 1024
Localizer, changes made in the View/Edit table that affect the number of scans, image interval, starting/ending locations, tilt, or display FOV are automatically shown on the 1024 Localizer
Any changes made directly on the 1024
Localizer display using the mouse and the on-screen cursor controls are also reflected automatically in the View/Edit table
View/Edit Wizard intuitively adjusts
dependent parameters automatically in response to operator-initiated changes and highlights them for quick review. It also alerts the operator to incompatible dependencies requiring operator intervention.
Tab card groupings for Timing, Recon and
Filming help organize the large number of parameters available within each protocol.
As many as 8,460 protocols can be stored
on the Operator Console.
Scan Data Acquisition:
Full-screen DynaPlan Plus illustrates scan
status graphically, with real-time feedback while the exam is underway. Scans, programmed delays (prep, breathing, inter­group), and even AutoVoice announcements are clearly shown before and during scanning.
AutoScan: Fully automates longitudinal
table movement and the start of each scan
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
AutoVoice: Preset (English) and user-
recorded messages automatically deliver patient breathing instructions, especially useful for multiple or multi-pass helical scans
Full Simultaneity allows scan and recon to
work concurrently with image display, processing and analysis (including computationally intensive features such as MPR, MPVR and 3D*/MIP) while still running image archival, filming and networking processes.
Dose Computation & Display
CTDI
(CTDI volume), DLP (Dose Length
vol
Product), and Dose and display during scan prescription provides patient dose information to the operator.
CTDI
is a dose index defined by IEC 601-2-
vol
44. This index is computed automatically by the Lightspeed the Exam Rx screen. CTDI number consisting of 2/3 of the CTDI peripheral dose plus 1/3 of the CTDI dose that is divided by the helical or axial pitch factor.
CTDI
is a dose index based upon CTDI
100
dose measurements over a 100 mm volume, as defined in IEC 601-2-44.
Dose Length Product (DLP) is given in
mGy*cm and is computed and displayed for each group prior to the scan. Additionally, an accumulated DLP is displayed for the entire exam, as the exam prescription progresses. The final exam accumulated DLP provides a convenient measure for maintaining patient or procedure dose management statistics.
Dose Efficiency is automatically
computed and displayed on the Exam Rx screen. The dose efficiency is a measure of how much of the Z-axis X-ray beam is used by the system, as defined in IEC 601-2-44.
Efficiency computation
16
CT System and reported on
is a single
vol
100
central
100
AutoView Layouts:
Eight powerful AutoView layouts provide
exceptional flexibility in tailoring the 1,024 image display to the user or the application at hand - without the complexity of free-form “windows.”
AutoView Layouts include:
1024 AutoView image
768 AutoView image (matches the image
size shown on the HiSpeed Advantage
2.X Series OC monitor)
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 11
512 AutoView image + 512 Localizer Scout with cut lines automatically showing the location of the AutoView image on the Scout
Two 512 AutoView images (same image
but at different window/level settings) + 512 Localizer Scout with cut lines automatically showing the location of the AutoView images on the Scout
512 AutoView image + 512 AutoFilm
image
Last two 512 AutoView images
Last four 512 AutoView images
AutoLink which links the current series to
a view port
Basic image review features such as
window/level, magnification and flip/rotate are available for AutoView images.
Any window not used for AutoView is
available to independent, simultaneous review of other exams.
Special BrightBox, a three-button trackball
device, provides independent control of image next, prior, manual paging and trackball window/level for any review images in focus. This helps make two person operation practical.
Regardless of the AutoView Layout used,
AutoFilm viewing is available anytime via an on-image selection - without disrupting other image processes in progress. Background filming allows full use of the image display monitor for AutoView and image review/processing without interruption during AutoFilm.
Image Review Layouts:
Five flexible Image Review Layouts are
provided for those applications where greater than 512 image display may be desired and AutoView is not required.
Image Review Layouts include:
Note: uses short notation for screen options
1024 single image display
768 single image display
Two 512 image display, horizontal format
Two 512 image display, vertical format
Four 512 image display
Each image display window can be further
subdivided into four more images, increasing the total number of images that can be displayed at once to 16.
BrightBox image control is also available for
Image Review Layouts.
LightSpeed16
With Xtream
CT Scanner System
Image Access:
Point and click interface along with a
pictorial directory (browser) allows for easy selection by exam, series or image
Routine Image Display:
Image display features provided within
Exam Rx:
Zoom/Roam
Flip/Rotate
Display Normal
List/Select
Ellipse ROI
Measure Distance
Grid On/Off
Cross Reference
User Annotation
Exam/Series Page
Hide Graphics
Erase
Screen Save
Gray Scale Enhancement
ProView visualization algorithms are available to enhance anatomical structures without additional reconstruction time:
Four Selections for enhancement of high
Three ways are provided to adjust window/ level of images in focus in order to meet a variety of clinical work environments and user preferences:
Explicit Magnify
ProView
contrast objects where fine detail is required without aliasing (such as lungs)
Three Selections for modifying perceived levels of noise and low contrast discrimination
Six user-programmable keys on the scan control keyboard (F6 - F11), plus one key for returning to prior setting (F5)
On-image through middle mouse button
BrightBox trackball
Routine Measurements:
Image measurement features provided
within Exam Rx:
Box ROI
Ellipse ROI
Trace ROI
Measure Distance
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Measure Angle
Grid On/Off
Hide Graphics
Erase
Screen Save
MIROI (Multiple Image ROI)
Report Pixels
Display Preferences:
Display settings available to tailor the overall
display (settings apply to all images in all exams):
Annotation Levels
Inverse Video
Next/Prior Each View Port
Next/Prior Series Binding
Continuous Report Cursor
Auto Image Management:
The Exam Rx work environment conveniently provides for selection of AutoFilm, AutoStore (to local or remote MOD), and AutoTransfer (across a network).
An AutoFilm Composer provides a simple programming interface for automated filming set-up.
Batch Filming is accomplished through a single keystroke which automatically prints an entire series at a time.
Manual Image Filming:
On-screen filming is available for any analog
or digital camera using a 3M-952 protocol.
Images may be individually filmed manually
via “drag and drop” to the on-screen Film Composer.
Print Series permits automatic printing of an
entire series with one keystroke.
Page filming permits creation of an entire
film with one keystroke.
Multiple image formatting allows filming of
multiple images in a single film frame.
Film formats supported are 1:1, 2:1, 4:1, 6:1,
8:1, 9:1, 12:1, 15:1, 16:1, 20:1, 24:1 and 35­mm slide
Important note: The Lightspeed comes standard with a DICOM Print Interface configurable for multiple DICOM Print destinations. Connections with cameras that
16
CT Scanner
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 12
do not support DICOM Print may require a filming interface (purchased separately).
To save further filming cost, the Operator Console can directly print to a postscript printer such as the GE Color Printer available as an option.
ImageWorks
ImageWorks software is designed to take advantage of the Lightspeed computer and image processor. This desktop environment includes image management and networking.
Because some of the image analysis and display features of ImageWorks replicate those in Exam Rx, the next section describes only features that are incremental or significantly different.
Image Analysis
Multi-Projection Volume
Reconstruction
way to generate volumetric images for CT angiography without thresholding data or removing unwanted anatomy. An entire volume is used to generate images in any plane, creating real-time frames of reference at the same time;
• Clinical utility is extended via two additional modes:
MIPS - enhances contrast and improves visualization of calcifications
Average - generates 2D radiographic images;
Multi-planar Reformation (MPR):
Provides real-time assessment of anatomy in offaxis planes. Sagittal, coronal, oblique and curved planar reformations available;
Batch reformatting can also be defined
and executed for later viewing if desired;
• Image Addition and Subtraction: Includes image addition of more than two images at a time;
*Volume Viewer
Volume Viewer is an innovative and powerfull suite of productivity enhanvers
(MPVR): Quick and easy
16
CT Scanner’s
LightSpeed16
With Xtream
CT Scanner System
(Volume Rendering, Volume Analysis and Navigator) that includes :
Dynamic Volume Review™ for Fast Screening
Curved Volume Of Interest
Protocol Management and Loading
Review Layout Presets
Multiple VR Objects Merge
Pseudo Surface Shading Mode
Predefined Cut Planes
Volume Rendered Navigator views
VR Preset save/recall
3D Rendered Lumen View
Automatic Path Tracking
Path Bridging (in case of occlusions)
SmartCursor™ for Easy Navigation
Synchronized Reformatted Views
Cut visualization mode
*Advanced Vessel Analysis
Advanced Vessel Analysis is the ultimate tool to assess and quantify vascular structures, including stenosis analysis, stent planning procedures, post stenting or vascular surgery follow-up.
protocol driven tools to perform quick,
flexible and accurate quantitative analysis of vascular anatomy
provides maximum, minimum and mean
intraluminal diameter measurements
provides cross-sectional areas of true
orthogonal sections of the aortoiliac systems at selected anatomical points
clinical benefits include: stenosis sizing, pre-
and post- surgical assessment, stent planning
Measurements in % stenosis or mm of
stenosis, and measurement of length and dimension of stenosis.
Image Display
Magnifying Glass allows quick 2X mag
window that can be moved over an image.
Image Scroll moves an image within its’ own
window.
Groupings allow application of window/level
values, magnification/minification, image
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
scroll or flip and rotate to a user-defined image set.
Save State stores user-selected image
orientation and window/level with each data set.
Window/Level values may be:
Preset to provide six on-screen instant
window/level settings
Set independently for up to 16 images on
the screen
User-modified in discrete or variable
steps
Adjusted real-time on-image by holding
down the middle mouse button and moving the mouse
Cine mode provides paging in up to 4 view
ports of up to 128 previously-stored CT or MR images at full selected display frame rate. For more than 128 images, display frame rate may be reduced.
Cine mode also provides temporal, spatial
or manual playback loops.
Text Page
Image Annotation
Image annotation and cursor are shadowed
to permit ease in reading.
Image Management
Images may be stored and retrieved via
Magnetic Optical Drive (MOD) media using DICOM 3.0 format. This allows interchange with other imaging systems supporting DICOM 3.0 MOD media. Not all vendors implementation of DICOM 3.0 are identical, so please check with the manufacturer for compatibility.
Off-line retrieval of all image files. Images
may be viewed as soon as they are restored from MOD.
Image retrieval time is approximately 30 images 512 per second.
VariViewerTM:
VariViewer is an interactive axial review
mode that can change the slice thickness reconstruction instantaneously.
The user selects the volume to be
specifically analyzed and chooses the slice thickness to be displayed as axial images. The user can then save a number of recon images sets, viewing a large number of slices for pure axial review and filming.
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 13
Batch film can typically enable you to reduce filming images by 50%, thickening the slab from 1.25 to 2.5mm with no information loss.
Image Networking:
Exams can be selected and moved from MR Signa Advantage, CT HiSpeed Advantage, Advantage Independent Console, Advantage Windows Workstation to the Lightspeed Console. This is performed via the AdvantageNet (GenesisNet) point-to-point protocol. This protocol supports send, receive and pull/query (no broadcast). NOTE: Because the Lightspeed images in pure DICOM format, exams
not be moved from the Lightspeed
Scanner System to MR Signa Advantage, CT HiSpeed Advantage or the Advantage Independent Console systems.
Exams can also be selected and moved between the Lightspeed System and any imaging system supporting the DICOM 3.0 protocol for network send, receive and pull/query. GE systems that support this interoperability include CT HiSpeed CT/i, X/i, NX/i, QX/i, MR Signa LX, and Advantage Windows
1.2/2.x/3.x/4.x.
Image transfer time using either AdvantageNet or DICOM 3.0 protocols is approximately 0.1 second per 512 image on a 100baseT Network.
Computer Based Training
This innovative feature provides on­screen, on-line support of Exam Rx and ImageWorks functions via a multi-media CDROM player, integrated into a tower on operator console for easy access. Use of rich graphics and text to provide truly useful system help.
A well-indexed table of contents helps speed access to information. The CD can be viewed on the scanner or on a PC.
Industry Standards
The LightSpeed
complies with a wide variety of industry standards to facilitate more rapid adoption of features and performance improvements as the computing and medical imaging industry evolves.
16
CT Scanner’s Operator
16
creates
16
CT
16
CT Scanner
16
CT Scanner System
can
LightSpeed16
With Xtream
CT Scanner System
DICOM Conformance Standards:
DICOM 3.0 Storage Service Class
Service Class User (SCU) for image
send
Service Class Provider (SCP) for image
receive
DICOM 3.0 Query/Retrieve Service Class
DICOM 3.0 MOD Media Service Class on
1.2- and 2.3-GB MOD media
DICOM 3.0 Storage Commitment Class
Push
DICOM 3.0 Modality Worklist*
DICOM 3.0 Modality Performed Procedure
Step*
DICOM 3.0 Print
Filming Protocol:
3M-952 Standard
System Components
Gantry:
Advanced slip ring design continuously rotates generator, tube, detector and data acquisition system around the patient.
Aperture: 70 cm
Tilt: ± 30°
Tilt Speed: 1°/sec.
Focus to Detector: 95 cm
Focus to Isocenter: 54 cm
Maximum SFOV: 50 cm
Rotational Speeds: 360° in 0.5, 0.6, 0.7 0.8,
0.9, 1.0, 2.0, 3.0 and 4.0 sec.
Remote Tilt from Operator’s Console.
Integrated breathing lights and countdown
timer.
Integrated start scan button with countdown
to X-ray on.
Scan plane toward front of gantry for improved positioning access.
Biopsy and interventional studies have been facilitated through a more streamlined gantry shroud, and bilateral table/gantry controls and gantry display that maximize maneuverability while working next to the gantry.
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Laser Alignment Lights:
Define both internal and external scan
planes to ± 1 mm accuracy.
Operate over full range of gantry tilt;
activated any time during exam (with tube stationary).
Coronal light remains perpendicular to axial
light as gantry tilts.
Visual readout is easy to read from the table side or from the operator console.
Gantry tilt controls are located on the side of the gantry.
Table
Single table, cantilever design with wide
height range
Vertical Range: 51.6 cm to 99.1 cm
Vertical Scannable Range: 80.6 cm to 99.1
cm
Elevation Speeds: 5 mm/sec and 40
mm/sec
Horizontal Range: 170 cm
Horizontal Scannable Range: 170 cm
metal-free (axial) and 160 cm metal-free (helical & Scout)
Horizontal Speed: Up to 100 mm/sec
Table automatically recenters on scan plane
with changes in vertical position (after setting internal landmark with alignment lights on)
Table Load Capacity:
180 kg (400 lb) with +/- 0.25 mm
positional accuracy
205 kg (450 lb) maximum allowed with
normal operation and +/- 1 mm positional accuracy
Controls on gantry for elevation and cradle
incrementation. Foot pedals on both sides of table for fast elevation. Cradle position controlled from OC for prescribed scans.
IV Pole integrated at the foot-end of the
table prevents IV lines from becoming crossed and tangled, and ensures that the lines stay securely in place on the patient.
X-Ray Tube
Performix ADV Metal-Ceramic Tube Unit. Design optimized for exams requiring a large number of scans without tube cooling.
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 14
Heat Storage Capacity: 6.3 MHU 16 x 1.25 mm (uses all 24 rows)
Heat Dissipation:
Anode (max) 840 KHU/min
Casing (cont) 300 KHU/min
Tube Unit: 6.9 kW continuous for 10
minutes
Dual Focal Spots:
Small Focal Spot:
0.7 mm (W) x 0.6 mm (L) nominal value (IEC 336/93)
0.9 mm (W) x 0.7 mm (L) traditional methodology
Large Focal Spot:
0.9 mm (W) x 0.9 mm (L) nominal value (IEC 336/93)
1.2 mm (W) x 1.2 mm (L) traditional methodology
Maximum Power: 53.2 kW
Beam collimated to 55° fan angle.
Average time to replace tube: <
10 hours
High Voltage Generation
High-frequency on-board generator.
Continuous operation during scan.
53.2 kW output power.
kVp: 80, 100, 120, 140
mA: 10 to 440 mA, 5 mA increments
Maximum mA for each kVp selection:
KVp Max mA
80 400 100 420 120 440 140 380
HiLight Matrix II Detector:
21,888 individual elements composed by: 8 rows of 1.25 mm thickness and 16 rows of
0.625 mm thickness, each containing 888 active patient elements; 24 reference elements.
4 typical modes of data output:
8 x 1.25 mm (uses center 16 rows)
8 x 2.5 mm (uses all 24 rows)
LightSpeed16
With Xtream
CT Scanner System
16 x 0.625 mm (uses center 16 rows )
70% geometric efficiency; 98% absorption efficiency.
Data Acquisition System
12,288 available input channels.
1,640 Hz maximum sample rate.
Effective analog to digital conversion range greater than two million to one.
Scan/Control Unit
Located in base of Operator Console.
Host Computer
Dual SMP 2.66.GHz Intel Xeon processors
with 512KB L2 cache.
Intel Hyper-threading technology.
2GB DDR266 Dual Channel Memory with a
throughput of 4.2GB/sec
Image Processor:
Nvidia Quadro4 980XGL AGP 8X graphics
with 128MB Memory
Graphics Processor Unit (GPU) Clock
300Mhz
Graphics Memory Clock 325Mhz
Image Reconstruction Engine (GRE)
Custom-designed special purpose
CT Image Generator
Custom CT back projection hardware
accelerates 2D & 3D back projection.
Intel Hyper-threading Technology.
32-bit floating point data format
2GB DDR226 ECC Dual Channel Memory Standard (4.2 GB/sec).
Software Architecture:
Software architecture based on industry
standards and client-server design
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Peripherals:
Total of 254 GB system:
Main system (host) disk drive:
High Performance Drive
36 GB, 3.5 inch form factor
15,000 RPM
Ultra320 SCSI interface
Assigned to applications software and
image files
2 system disk drives (Image Disk)
High Performance Drive
73 GB, 3.5 inch form factor each.
15,000 RPM
Ultra320 SCSI interface
Assigned to image files only
- 250,000 uncompressed 512 images
Scan data disk drive:
High Performance Drive
36 GB, 3.5 inch form factor
UltraSCSI interface
Scan data disk drive:
High Performance Drive
36 GB, 3.5 inch form factor
UltraSCSI interface
Standard MOD drive:
Magnetic Optical Disk Drive
Erasable, rewritable media
2.3 GB, 3.5 inch form factor
Assigned to DICOM 3.0 format image
file.
Stores 4,700 lossless JPEG compressed
512 image files per side
Off-line retrieval of image. Images may
be viewed as soon as they are restored from MOD
DVD Ram:
4.7 GB per side, 5.25" half height form
factor
Transfer rate 2.7MB/sec
Assigned to scan data file and protocol
file storage/retrieval.
Color monitors:
21 inch diagonal width
1280 x 1024 dot resolution
Non-interlaced, flicker-free presentation
76 kHz Horizontal deflection frequency
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 15
72 Hz Vertical deflection frequency Easy-to-use calcium scoring software
Sync on green
LCD Color Monitor*: 18” diagonal width
Scan control keyboard (English language)
assembly with intercom speaker, microphone and volume controls.
Global modem to allow InSite connectivity
3-Button Mouse
3-Button Trackball
Image Networking:
Standard auto-configuring
100BaseT/10BaseT Ethernet (UTP connection)
Direct network connection; multi-suite
ethernet card not required for gateway out of suite
Protocols supported:
DICOM 3.0 network send (one IP
address at a time) and receive, pull/query, and storage commitment push;
AdvantageNet (GenesisNet) point-to-
point send, receive and pull/query (no broadcast);
InSite point-to-point;
TCP/IP (for system administration)
Compatible Options
A DICOM Print Interface is standard on the
system.
English Keyboard (B7800KE)
French UIF & Keyboard (Z62401F)
German UIF & Keyboard (Z62401G)
Italian UIF & Keyboard (Z62401H)
Spanish UIF & Keyboard (Z62401S)
Portuguese UIF & Keyboard (Z62401P)
ConnectPro HIS/RIS Interface with
Performed Procedure Step (PPS) M4 compatibility (B7500LN)
Bar Code Reader (B7540RB)
Volume Viewer (B7540LD)
DentaScan (B7540LS)
SmartScorePro (B7850PL) – Gating SW
Prospective or retrospective EKG-gated
cardiac imaging in a single breath-hold
250-300 msec temporal resolution
LightSpeed16
With Xtream
CT Scanner System
with integrated patient report generation
Snapshot Option (B7710LS)
CT Perfusion 2(B7700PL)
Smart Step (B7868WL) – SW
Smart Step HW (B7868SM & B7868SN)
Advanced Vessel Analysis (B7868LA)
CT Colonoscopy (B7868LC)
CardIQ2 (B7710LR)
Remote 21 inch Color Monitor (B7530RC)
Exide UPS (B7999PS)
Metal Free Table Interface for 3
Attachments (B7999PM)
LCD Monitor (B7858LC)
GE Color Printer (E7014LA)
Table Tray
Siting Requirements
Minimum Suite Size:
Overall Suite:
15.5 feet (4.74 m) wide x 19.5 feet (5.96 m) deep x 9 feet (2.75 m) high.
Recommended Suite Size:
Overall Suite:
19.5 feet (5.96 m) wide x 24.5 feet (7.49 m) deep x 9 feet (2.75 m) high.
Exam Room: 13 feet (3.97 m) wide x 19.5 feet (5.96 m) deep.
Control Room: 11 feet (3.36 m) wide x 12.5 feet (3.82 m) deep.
Equipment Room: Not required. Power Distribution Unit heat and noise output under all operating conditions with all available options is low enough to allow inclusion in the patient scanning suite with the table and gantry.
Temperature and Humidity:
Exam and Control Rooms: 60°-75°F (15°­26°C) at 30%-60% relative humidity (non­condensing).
Equipment Room: If a separate equipment room is used to house the PDU, the allowable temperature range is 60°-84°F (15°-29°C) at 30%-60% relative humidity (non-condensing).
rd
Party
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Temperature Rate of Change: 3°C/hour max.
Relative Humidity Rate of Change: 5% RH/hour max.
Power Requirements:
The only facility input to the system is a 380 to 480 V nominal, 3 phase Delta or Wye, 50/60 Hz, 90 kVA service, 20 kVA average power; main disconnect to be located within 5 feet (1.5 m) of the PDU. The facility must also provide a protective disconnect device with low voltage, low energy local and multi-point remote capability, in the line feeders to the PDU.
Complete, detailed specifications of all power requirements are available upon request. For most installations, the Lightspeed Scanner System does not require any power conditioning equipment to be used in conjunction with the PDU. Regulators are not recommended for use with this system. For those sites with known large power line transients, a suppresser filter for the system computer and peripherals may be useful. In general, suppresser filters are not recommended.
Compatible UPS options are available if desired.
Cooling Requirements:
The cooling requirements do not include cooling for the room lighting, personnel or non-CT equipment present. Cooling requirements are listed by subsystem to allow planning for each room of the CT suite.
Cooling requirements are given for minimum, recommended and growth allowance scenarios.
The minimum cooling figures assume
patient throughput of 3 patients per hour and 75 scan rotations per patient.
The recommended cooling requirements assume patient throughput limited by the tube cooling algorithm.
The suite cooling can be sized for future developments by using the growth allowance figures. This cooling will accommodate more patients per hour and/or potential future system enhancements.
16
CT
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 16
Minimum Allowance
Subsystem Watts BTU/hr
Gantry 5,440 18,600** Table 200 700 PDU 1,000 3,400 Operator
Console Optional Laser
Camera
Recommended cooling values should not be used for calculating system input power requirements.
** Recommended Allowance: 7,150/24,400 Growth Allowance: 9,200/31,400
2,165 7,400
800 2,730
CT Scanner System
System Components Dimensions
mm inches
Gantry:
Width: 2,230 87.6
Depth: 1,007 39.6
Height: 1,887 74.1
Table:
Width: 610 24
Length: 5,050 199
Height: 1,120 44
Power Distribution Unit:
Width: 762 30
Depth: 585 23
Height: 1,270 50
Operator Console:
LightSpeed16
With Xtream
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
Warranty
The published Company warranty in effect on the date of shipment shall apply. The Company reserves the right to make changes.
All specifications are subject to change.
Regulatory Compliance
This product is designed to comply with applicable standards under the Radiation Control for Health and Safety Act of 1968.
Laser alignment devices contained within this product are appropriately labeled according to the requirements of the Center for Devices and Radiological Health.
This product is a CE-compliant device which satisfies regulations regarding Electro­Magnetic Compatibility (EMC) and Electro­Magnetic Interference (EMI), pursuant to IEC-
601.
Width: 1,238 48.7
Depth: 1,288 50.7
Height: 1,257 49.5
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 17
LightSpeed16
With Xtream
CT Scanner System
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
* Option
g
Product Data Sheet – rev 4.2 July 4th 2003 Page 18
Minimum Room Layout
Typical Room Layout
LightSpeed16
With Xtream
CT Scanner System
GE Medical Systems-America Milwaukee, USA - Fax: 1 414 544 3384 GE Medical Systems-Asia: Tokyo, Japan - Fax: 81 425 85 5490 Hong Kong - Fax: 852 2559 3588 GE Medical Systems-Europe
* Option
Loading...