Change #1
On page 17, Table 3, replace the entry for 350.0°C Input with:
Input Lower Limits Upper Limits
350.0 °C 345.5 °C 354.5 °C
Change #2
On page 17, under Testing Temperature, replace steps 2, 4, and 5 with the following and add step and 8.
2. Set the 5520A for K-type thermocouple and an output of -10.0 °C.
4. Adjust the temperature offset, using the up and down arrow keys, until the offset on the Meter secondary display reads 0.0.
5. Enter the offset value by pressing the yellow button, then the backlight button.
8. If the Meter fails any of the Temperature Accuracy tests, perform the Temperature Calibration Adjustment procedure.