A “XW Warning" identifies hazardous conditions and actions that could cause bodily harm or death. A "W Caution" identifies conditions and actions that could damage the Meter, the equipment under test,
or cause permanent loss of data.
XWWarnings and Precautions
To avoid possible electric shock or personal injury, and to avoid possible
damage to the Meter or to the equipment under test, adhere to the following
practices:
•Use this meter only as specified in this manual or the protection provided
by the Meter might be impaired.
•Before using the Meter, inspect the case. Do not use the Meter if it is
damaged. Look for cracks or missing plastic. Pay particular attention to the
insulation around the connectors.
•Inspect the test leads for damaged insulation or exposed metal. Check the
test leads for continuity. Replace damaged test leads before using the
Meter.
•Verify the Meter’s operation by measuring a known voltage. Do not use the
Meter if it operates abnormally. Protection may be impaired. When in doubt,
have the Meter serviced.
•Do not apply more than the rated voltage, as marked on the Meter, between
the terminals or between any terminal and earth ground.
•Use caution when working with voltages above 30 V ac rms, 42 V ac peak,
or 60 V dc. These voltages pose a shock hazard.
• Use the proper terminals, function, and range for your measurements.
• Do not operate the Meter around explosive gas, vapor, or dust.
• When using the probes, keep fingers behind the finger guards.
• When making connections, connect the common test lead before
connecting the live test lead; when disconnecting, disconnect the live test
lead before disconnecting the common test lead.
•Disconnect circuit power and discharge all high-voltage capacitors before
testing resistance, continuity, diodes, or capacitance.
•Before measuring current, check the Meter's fuses (see "Testing the
Fuses") and turn OFF power to the circuit before connecting the Meter to
the circuit.
• Do not operate the Meter with the case (or part of the case) removed.
• Use only a single 9 V battery, properly installed in the Meter case, to power
the Meter.
•Replace the battery as soon as the low battery indicator (b) appears. With a
low battery, the Meter might produce false readings that can lead to electric
shock and personal injury.
• Remove test leads from the Meter before opening the Meter case.
• When servicing the Meter, use only specified replacement parts.
• When replacing the fuses, use ONLY Fluke-specified fuses as listed in
Table 5.
2
Page 3
True RMS Multimeters International Electrical Symbols
International Electrical Symbols
The following international symbols appear in this document and on the Meter.
B
F
F
B
W
X
s
b
#
AC (Alternating Current)
DC (Direct Current)
AC or DC
Risk of danger. Important Information.
Refer to the manual.
Hazardous Voltage
Technischer Überwachungs Verein.
German certifying body.
Battery. Low Battery when shown on
display.
Verband Deutscher Electroniker. German
electronics association.
Test Lead Alert
Personal injury or damage to the Meter can occur if the user attempts to make a
measurement with a lead in an incorrect terminal.
J
I
P
$
T
!
;
~
XWWarning
Earth ground
Fuse
Conforms to European Union directives
Conforms to relevant Canadian Standards
Association directives
Double insulated
Underwriters Laboratories, Inc.
Meter in accordance with IEC 61010-1. 54CJ
Conforms to relevant Australian standards
Do not dispose of this product as unsorted
municipal waste. Go to Fluke’s website for
recycling information.
As a reminder to check that the test leads are in the correct terminals, LEAd is displayed when the Rotary
Switch is moved to or from the
mA or A position.
3
Page 4
175, 177, & 179
Calibration Information
Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity
at 0 % to 90 %. Accuracy specifications take the form of: ±( [ % of Reading ] + [ Counts ] ). AC conversions are
ac-coupled, true rms responding.
Maximum voltage between
any terminal and earth ground: 1000 V dc or ac rms
Surge Protection: 8 kV peak per IEC 61010
W Fuse for mA inputs:
W Fuse for A input:
Display: Digital: 6000 counts, updates 4/sec
Altitude: Operating: 2000 m; Storage: 12,000 m
Temperature: Operating: -10 °C to +50 °C;
Temperature coefficient: 0.1 X (specified accuracy) / °C
Electromagnetic Compatibility
(EN 61326-1:1997):
Relative Humidity:
440 mA, 1000 V FAST Fuse, minimum interrupt rating 10 kA
11 A, 1000 V FAST Fuse, minimum interrupt rating 17 kA
(<18 °C or >28 °C)
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature:
specified accuracy ±5 °C (9 °F)
Maximum, Noncondensing
90 % to 35 °C
75 % to 40 °C
45 % to 50 °C
Battery Life: Alkaline: 400 hrs typical
Size (H x W x L): 4.3 cm x 9 cm x 18.5 cm
Weight: 420 g
Safety Compliances: ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Measurement
Category III, 600 V Measurement Category IV
CAT III- Equipment designed to protect against transients in equipment in fixed-
equipment installations, such as distribution panels, feeders and short branch circuits,
and lighting systems in large buildings.
CAT IV- Equipment designed to protect user against transients from the primary
supply level, such as an electricity meter or an overhead or underground utility
service.
Continuity 600 Ω 1 Ω Meter beeps at <25 Ω, beeper turns off at >250 Ω; detects opens or shorts of
Ohms
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
AC Amps
(True RMS)
(45 Hz to 1 kHz)
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of ≤3 at full scale up to 500 V, decreasing linearly to crest factor ≤1.5 at 1000 V.
3. For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3.
4. In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2 for all models.
5. Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
6. 400.0 mA accuracy specified up to 600 mA overload.
7. >10 A unspecified.
5
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
1000 nF
10.00 μF
100.0 μF
9999 μF
60.00 mA
400.0 mA
6.000 A
10.00 A
4
6
7
0.001 V
0.01 V
0.1 V
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
1 nF
0.01 μF
0.1 μF
1 μF
0.01 mA
0.1 mA
0.001 A
0.01 A
0.15 % + 2
250 μs or longer.
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.5 % + 3
0.09 % + 2
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.5 % + 3
0.09 % + 2
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.5 % + 3
5
Page 6
175, 177, & 179
Calibration Information
Function Range1 Resolution
Accuracy ±( [ % of Reading ] + [ Counts ] )
Model 175 Model 177 Model 179
DC Amps5
Hz
(AC- or DCcoupled,
2,3,4
V or A
60.00 mA
400.0 mA
6.000 A
10.00 A
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
6
7
input )
Temperature -40 °C to +400 °C
-40 °F to +752 °F
MIN MAX
AVG
For DC functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than
350 ms in duration.
For AC functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than
900 ms in duration.
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Frequency is specified from 2 Hz to 99.99 kHz in Volts and from 2 Hz to 30 kHz in Amps.
3. For serial number 83911000 and below, and 83911241 to 84060000,frequencies <10 kHz are not specified in 600 mV ac, 60 mA ac, and 6 A
ac ranges. Noise at frequencies below 4.5 Hz and amplitude below 600 mV may exceed the frequency specification. For Serial numbers
83911001 to 83911240 and 84060001 and above, this note does not apply.
4. Below 2 Hz, the display shows zero Hz.
5. Amps input burden voltage (typical): 400 mA input 2 mV/A, 10 A input 37 mV/A.
6. 400.0 ma accuracy specified up to 600 mA overload.
7. >10 A unspecified.
8. Does not include error of the thermocouple probe.
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 °C
0.1 °F
1.0 % + 3 1.0 % + 3 1.0 % + 3
0.1 % + 1
NA NA
0.1 % + 1
0.1 % + 1
1 % + 10 8
1 % + 18
8
Function
Overload
Protection
Input
1
Impedance
(Nominal)
Common Mode Rejection
Ratio
(1 kΩ Unbalanced)
Normal Mode
Rejection
Volts AC 1000 V rms >10 MΩ <100 pF >60 dB @ dc, 50 Hz or 60 Hz
Volts DC 1000 V rms >10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
mV/T 1000 V rms 2 >10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
Open Circuit
Test Voltage
Full Scale Voltage To:
600 kΩ 50 MΩ
Short Circuit
Current
Ohms/Continuity 1000 V rms 2 <8.0 V dc <660 mV dc <4.6 V dc <1.1 mA
Continuity/Diode
test
7
1. 10
2. For circuits <0.3 A short circuit. 660 V for high energy circuits.
V-Hz maximum.
1000 V rms
2
<8.0 V dc 2.4 V dc <1.1 mA
Function Overload Protection Overload
mA Fused, 440 mA, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum
A Fused, 11 A, 1000 V FAST Fuse 20 A overload for 30 seconds maximum
6
Page 7
True RMS Multimeters Testing the Fuses
Testing the Fuses
XWWarning
To avoid electrical shock or personal injury:
•Remove the test leads and any input signals before replacing the battery or
fuses.
•To avoid electrical shock, arc blast, or damage to the Meter, install only
fuses with the amperage, interrupt, voltage, and speed ratings specified in
Table 5.
To test the fuses (refer to Figure 1):
1. Set the Rotary Switch to e.
2. Plug a test lead into the VeG terminal and touch the probe to the
10 A terminal (to test the 11 A Fuse).
or
400 mA (to test the 440 mA Fuse)
• If the display shows a resistance value in the range shown in Figure 1, the fuse is good.
• If the display reads
OL, replace the fuse and test again.
• If the display shows any other value, have the Meter serviced. See “Service Information” earlier in
this manual.
11 A 440 mA
Ω
<12
MIN MAX
RANGEHOLD
V
OK
<.5
Ω
OKOK
OK
Figure 1. Testing the Current Fuses
aik12f.eps
7
Page 8
175, 177, & 179
Calibration Information
Replacing the Fuses
XWWarning
To avoid electrical shock, arc blast, or damage to the Meter, install only fuses
with the amperage, interrupt, voltage, and speed ratings specified in
Table 5.
To replace the Fuses (refer to Figure 2):
1. Turn the Rotary Switch to
OFF.
2. Disconnect the test leads and/or any connectors from the terminals.
3. Remove the four screws from the Case Bottom and separate the Case Top from the Case Bottom.
4. Remove the fuses by gently prying one end loose, then slide the fuse out of its bracket, and replace
with a properly rated good fuse.
5. Rejoin the Case Bottom and Case Top, and reinstall the four screws.
Replacing the Battery
XWWarning
To avoid false readings, which could lead to possible electric shock or personal
injury, replace the Battery as soon as the low battery indicator (
To replace the Battery (refer to Figure 2):
b ) appears.
1. Turn the Rotary Switch to
OFF and remove the test leads from the terminals.
2. Remove the two screws from the Battery Compartment Door, and remove it from the Case Bottom.
3. Remove the Battery from the battery connector.
4. Replace the Battery with a new 9 V battery (NEDA A1604, 6F22, or 006P).
5. Reattach the Battery Compartment Door to the Case Bottom and reinstall the two screws.
B1
F1
F2
8
Figure 2. Battery and Fuse Replacement
aik13f.eps
Page 9
True RMS Multimeters Cleaning
Cleaning
XWWarning
To avoid electrical shock:
• Remove test leads and any input signals before cleaning.
• Do not reinstall the pca until it is completely dry.
Dirt or moisture in the terminals can affect readings.
W Caution
To avoid damaging the Meter, do not use aromatic hydrocarbons or chlorinated
solvents for cleaning. These solutions will react with the plastics used in the
instruments.
Do not use detergents of any kind for cleaning the pca.
Clean the instrument case with a damp cloth and mild detergent.
The pca may be washed with isopropyl alcohol or hot deionized water and a soft brush. Remove excess
cleaning material with clean dry air at low pressure, then dry the pca at 50 °C.
Replacing the LCD
W Caution
To prevent contamination, do not handle the conductive edges of the LCD or
the LCD Elastomeric Connectors.
If the edges are contaminated, clean them with alcohol. Allow the alcohol to dry
before reassembling.
To remove and replace the LCD, perform the following procedure (refer to Figure 3):
1. Turn the Rotary Switch to
2. Remove the four screws from the Case Bottom, and separate the Case Top from the Case Bottom.
The circuit board remains attached to the Case Bottom.
3. Remove battery compartment door and disconnect the battery from the battery connector.
4. Remove the PCA Screw located under the 440 mA Fuse, and lift the pca out of the Case Bottom.
The beeper is loose in the bottom case. Exercise caution not to damage it.
5. Loosen the 4 screws on the back of the pca. This will facilitate reinstalling the LCD.
6. Insert a small, flat-head screwdriver under the LCD Mask edge and gently pry the LCD Mask from the
snaps.
Take care to not break the LCD with the screwdriver.
7. Lift out the LCD.
8. Make sure that all connector contact points are clean. Refer to “Cleaning” for more information.
9. Install a new LCD, taking care that it is correctly oriented.
10. Reattach the LCD Mask to the LCD assembly by snapping the LCD Mask into place.
11. Tighten the 4 screws on the back of the pca.
12. Lay the pca back into the Case Bottom, and screw it down.
13. Reinsert the 440 mA Fuse.
14. Reinstall the Battery.
15. Rejoin the Case Bottom, Case Top, battery compartment door and reinstall the four case screws.
OFF and remove the test leads from the Meter terminals.
W Caution
9
Page 10
175, 177, & 179
Calibration Information
LCD Mask
Figure 3. Removing the LCD Mask
aik15f.eps
Performance Tests
XWWarning
To avoid electric shock, do not perform the performance test procedures unless
the Meter is fully assembled.
The following performance tests verify the complete operation of the Meter and check the accuracy of
each Meter function against its specifications. If the Meter fails any part of the test, calibration adjustment
and/or repair is indicated.
In the performance tests, the Meter is referred to as the unit under test (UUT).
10
Page 11
True RMS Multimeters Performance Tests
Required Equipment
A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures in
this document.
If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1.
Table 1. Calibrator Specifications
Recommended Equipment Measurement Function Accuracy
5500A Multi-Product Calibrator
(or equivalent)
Fluke 80 AK
K-type Thermocouple Adapter Accessory
DC Volts 30 mV to 1000 V
±0.0225 %
DC Current 3 mA to 9 A
±0.25 %
AC Volts 300 mV to 1000 V
±0.25 % @ 45 Hz to 1 kHz
AC Current 3 mA to 9 A
±0.375 % @ 45 Hz to 1 kHz
Resistance 25 Ω to 500 Ω
±0.225 %
19 M Ω
±0.4 %
Capacitance 900 nF
±0.30 %
Temperature -40 to 400 °C
0.25 %
Frequency 5 V, 50 kHz
±0.025 %
Temperature -
K-type Thermocouple, mini-plug on both
ends
Temperature -
Preparing for the Performance Test
XWWarning
To avoid possible electric shock or personal injury:
•Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.
•Before handling the test connections and in between tests, make sure the
calibrator is in standby mode (STBY).
To prepare for the performance test:
1. Make sure that you have the required equipment (refer to Table 1).
2. Warm up the calibrator as required by its specifications.
3. Allow the temperature of the UUT to stabilize at room temperature (23 °C ± 5 °C [73 °F ± 9 °F]).
4. Check the fuses and Battery, and replace them if necessary. Refer to "Testing the Fuses”, “Replacing
the Fuses", and "Replacing the Battery".
11
Page 12
175, 177, & 179
Calibration Information
Performance Tests
Note
When calibrating or measuring AC current, avoid the potential for errors from coupled noise, by either:
1. Shorting the VeG to the COM input
or
2. Having NO connection to the VeG input.
1. Set the UUT Rotary Switch to the position called for in Table 2 (175) or Table 3 (177/179), and for all
measurements other than current (amps), connect the calibrator to the VeG and
COM input terminals.
• If testing the milliamps function, connect the calibrator to the Meter
400 mA and COM input
terminals.
• If testing the amps function, connect the calibrator to the Meter
10A and COM input terminals.
2. Referring to Table 2 or Table 3, apply the indicated calibrator output voltages to the UUT for each test.
3. Verify that the UUT displays a reading that is within the limits shown in the Meter Response column.
Table 2. Performance Tests 175
Test
(Switch Position)
K
AC Volts
AC Volts Frequency
5500 Output
300 mV 45 Hz 296.7 mV ac 303.3 mV ac
5 V 500 Hz 4.947 V ac 5.053 V ac
5 V 1 kHz 4.897 V ac 5.103 V ac
50 V 45 Hz 49.47 V ac 50.53 V ac
50 V 1 kHz 48.97 V ac 51.03 V ac
300 V 45 Hz 296.7 V ac 303.3 V ac
500 V 500 Hz 494.7 V ac 505.3 V ac
500 V 1 kHz 489.7 V ac 510.3 V ac
1000 V 45 Hz 987 V ac 1013 V ac
1
1 V 45 Hz 44.94 Hz 45.06 Hz K Hz
5 V 50 kHz 49.94 kHz 50.06 kHz
Lower Limit Upper Limit
Meter Response
12
Page 13
True RMS Multimeters Performance Tests
Table 2. Performance Tests 175 (continued)
Test
(Switch
Position)
L
DC Volts
DC Volts
Frequency
1
E
DC Millivolts
2
Ohms
E
Capacitance
Continuity
Meter Response
5500 Output
Lower Limit Upper Limit
5 V 4.990 V dc 5.010 V dc
300 V 299.3 V dc 300.7 V dc
1000 V 996 V dc 1004 V dc
-1000 V -1004 V dc -996 V dc
3 V 45 Hz 44.94 Hz 45.06 Hz D Hz
30 V 50 kHz 49.94 kHz 50.06 kHz
30 mV 0 Hz 29.8 mV dc 30.2 mV dc
-300 mV -300.7 mV dc -299.3 mV dc
600 mV 598.9 mV dc 601.1 mV dc
500 Ω 495.3 Ω 504.7 Ω e
19 MΩ 18.68 MΩ 19.32 MΩ
1
Apply 0.90 μF 887 nF 913 nF
25 Ω Beeper On R
250 Ω Beeper Off
G
1
Diode
F
AC Milliamps
Apply 2.0 V 1.978 V dc 2.022 V dc
3 mA 45 Hz 2.92 mA ac 3.08 mA ac
50 mA 1 kHz 49.22 mA ac 50.78 mA ac
400 mA 1 kHz 393.7 mA ac 406.3 mA ac
4.0 A 45 Hz 3.937 A ac 4.063 A ac ?
AC Amps
mI
DC Milliamp
1
9.0 A 1 kHz 8.83 A ac 9.17 A ac
3 mA, 0 Hz 2.94 mA dc 3.06 mA dc
50 mA 49.47 mA dc 50.53 mA dc
-400 mA -404.3 mA dc -395.7 mA dc
DC Amps
1. Press the YELLOW button to access this function.
1
4.0 A 3.957 A dc 4.043 A dc A
-9.0 A -9.12 A dc -8.88 A dc
2. Does not include test lead resistance.
13
Page 14
175, 177, & 179
Calibration Information
Table 3. Performance Tests 177/179
Test 5500 Output
K
AC Volts
300 mV 45 Hz 296.7 mV ac 303.3 mV ac
5 V 500 Hz 4.947 V ac 5.053 V ac
5 V 1 kHz 4.897 V ac 5.103 V ac
50 V 45 Hz 49.47 V ac 50.53 V ac
50 V 1 kHz 48.97 V ac 51.03 V ac
300 V 45 Hz 296.7 V ac 303.3 V ac
500 V 500 Hz 494.7 V ac 505.3 V ac
500 V 1 kHz 489.7 V ac 510.3 V ac
1000 V 45 Hz 987 V ac 1013 V ac
K
Hz
AC Volts
Frequency
L
DC Volts
1
1 V 45 Hz 44.94 Hz 45.06 Hz
5 V 50 kHz 49.94 kHz 50.06 kHz
5 V 4.993 V dc 5.007 V dc
300 V 299.5 V dc 300.5 V dc
1000 V 997 V dc 1003 V dc
-1000 V -1003 V dc -997 V dc
3 V 45 kHz 44.94 Hz 45.06 Hz L Hz
DC Volts
Frequency
1
30 V 50 kHz 49.94 kHz 50.06 kHz
Meter Response
Lower Limit Upper Limit
E
DC Millivolts
30 mV 0 Hz 29.8 mV dc 30.2 mV dc
-300 mV -300.5 mV dc -299.5 mV dc
600 mV 599.3 mV dc 600.7 mV dc
T
Temperature
(179 only)
1,3,4
0 °C -1.0 °C 1.0 °C
-40 °C -41.4 °C -38.6 °C
400 °C 395.0 °C 405.0 °C
2
Ohms
N
Capacitance
1
500 Ω 495.3 Ω 504.7 Ω e
19 MΩ 18.68 M Ω 19.32 M Ω
0.90 μF 887 nF 913 nF
25 Ω Beeper On R
Continuity
O
1
Diode
1. Press the YELLOW button to select this function.
2. Does not include test lead resistance.
3. Be sure to use correct cable.
4. To ensure accuracy, the Meter and thermocouple adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures.
250 Ω Beeper Off
2.0 V 1.978 V dc 2.022 V dc
14
Page 15
True RMS Multimeters Calibration Adjustment
Table 3. Performance Tests 177/179 (continued)
Test 5500 Output
F
AC Milliamps
J
AC Amps
mI
DC Milliamps
I
DC Amps
1. Press the YELLOW button to select this function.
2. Does not include test lead resistance.
3. Be sure to use correct cable.
4. To ensure accuracy, the Meter and thermocouple adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures.
1
1
3 mA 45 Hz 2.92 mA ac 3.08 mA ac
50 mA 1 kHz 49.22 mA ac 50.78 mA ac
400 mA 1 kHz 393.7 mA ac 406.3 mA ac
4.0 A 45 Hz 3.937 A ac 4.063 A ac
9.0 A 1 kHz 8.83 A ac 9.17 A ac
3 mA, 0 Hz 2.94 mA dc 3.06 mA dc
50 mA 49.47 mA dc 50.53 mA dc
-400 mA -404.3 mA dc -395.7 mA dc
4.0 A 3.957 A dc 4.043 A dc
-9.0 A -9.12 A dc -8.88 A dc
Lower Limit Upper Limit
Meter Response
Calibration Adjustment
Perform the calibration adjustment procedures if the Meter fails the performance test. The Meter buttons behave as follows when the calibration mode is enabled:
H
Press and hold this button to test the present function. This measurement is uncalibrated and
may be inaccurate. This is normal.
M
R
Press and hold this button to display the required input. Press this button to skip the present calibration step without accepting the input signal.
Note
If a calibration step is skipped, the remaining calibration procedure may be invalidated.
Note
This feature is not available for units with serial numbers between 83911001 – 83911240
and 84060001 and above.
B
Press this YELLOW button to store the calibration value and advance to the next step. This
button is also used to exit calibration mode after the calibration adjustment sequence is
complete.
15
Page 16
175, 177, & 179
Calibration Information
Calibration Adjustment Procedure
Use the following steps to adjust the Meter’s calibration:
For Meters with a serial number of 84060000 and below, switch the Meter to volts
AC (C). For
Meters with a serial number of 83911001 – 83911240 and 84060001 and above, switch the Meter to
DC.
E
1. Turn the Meter over and find the Calibration Seal located near the top of the Meter (refer to Figure 4).
2. With a small probe, break the Calibration Seal and press the Calibration Button for 1 second. The
Meter will beep and change to the calibration mode. The display reads C-01, designating the first
calibration step. The Meter remains in calibration mode until the Rotary Function Switch is turned off.
3. Proceed through the calibration steps by entering the input value listed in the table for each step.
Note
When calibrating or measuring AC current, avoid the potential for errors from coupled noise, by either:
1.Shorting the VeG to the COM input or
2.Having NO connection to the VeG input.
Note
After pressing the YELLOW button, wait until the step number advances before changing the calibrator
source or turning the Rotary Function Knob.
4. After each input value is applied, press the YELLOW button to accept the value and proceed to the
next step (C-02 and so forth).
5. When the last step in a function is reached, turn the Rotary Function Knob to the next required
function. The Meter will not allow a step to be completed if the Rotary Function Knob is turned to the
wrong function.
Note
If the calibration adjustment procedure is not completed correctly, the Meter will not operate
correctly. When calibration adjustment is not performed correctly, the Meter displays the messages
“CAL” and “Err”and the Meter must be recalibrated.
The Meter is damaged and requires service if:
• “CAL” and “Err” messages continue to appear after a proper recalibration.
• “EEPr” and “Err”messages are alternating on the display.
• “EEPr” message appears on the display.
16
Page 17
True RMS Multimeters Calibration Adjustment
Calibration Seal
Calibration Button
aik14f.eps
Figure 4. Calibration Access
Table 4. Calibration Steps
If the Meter has a serial number of 84060000 or below, or 83911241 to 84060000, use the following steps.
Function
(Switch Position)
C
(AC Volts)
L
(DC Volts)
TE
DC Millivolts
(temperature - 179 only)*
e
(Ohms)
R
(Continuity)
mA
(Milliamps)
A
(Amps)
* To ensure accuracy, the Meter and Thermocouple Adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures
.
Calibration Step Input Value
C-01 6.000 V, 900.0 Hz
C-02
C-03 6.000 V, 100 Hz
C-04 6.000 V dc
C-05 60.00 V dc
C-06 600.0 V dc
C-07 1000 V dc
C-08
C-09 0.0 °C
C-09 (175/177)
C-10 (179)
C-10 (175/177)
C-11 (179)
C-11 (175/177)
C-12 (179)
C-12 (175/177)
C-13 (179)
600.0 mV, 100 Hz
600.0 mV dc
6.000 kΩ
600 Ω
600.0 mA dc
6.000 A dc
17
Page 18
175, 177, & 179
Calibration Information
Table 4. Calibration Steps (continued)
If the Meter has a serial number from 83911001 to 83911240 or 84060001 and above, or if the model number
flashes on the display at turn on, use the following steps.*
* If the Meter is not connected correctly, or if the rotary switch is in the wrong position, the Meter will beep 2 times to alert the user.
C-15 400.0 mA ac, 60 Hz
C-16 6.000 A dc A
C-17 6.000 A ac, 60 Hz
18
Page 19
True RMS Multimeters User-Replaceable Parts and Accessories
User-Replaceable Parts and Accessories
User-replaceable parts and accessories are shown in Table 5 and Figure 5.
Table 5. User-Replaceable Parts and Accessories
Item Description Part No Qty
-- (Not shown) TL75 Test Lead Set** 855705 1
--
-- Calibration Information (Not Shown) 1600476 1
1 Case Top (Model 175)
2 Shock Absorber 428441 1
3 Keypad (177, 179)
4 LCD Mask (175)
5 LCD 1560856 1
6 LCD Light Pipe (177 & 179)
7 WFuse, 11 A, (fast acting), 1000 V ac/dc, minimum interrupt rating 17 kA * 943121 1
8 WFuse, 440 mA (fast acting), 1000 V ac/dc , minimum interrupt rating 10 kA * 803293 1
9 PCA Screw
10 Shield, AC
11 AC Shield screw and Internal Cover screws 448456 5
12 Case Bottom serial numbers 83911001 to 83911240 and 84060001 and above (includes
13 Battery, 9 V (Alkaline, 9 V, 0-200 mA) 614487 1
14 Battery Compartment Door 1564799 1
15 Case Screws 832246 4
16 Tilt Stand 648961 1
17 LCD Elastomeric Connector 650264 2
18 Internal Cover 1564786 1
19 Battery Connector, serial number 8391100 and below, and 83911241 to 84060000 825976 1
20 Battery Connector, serial numbers 8391101 to 83911240 and 84060001 and above 1988201 1
21 Beeper, serial numbers 8391101 to 83911240 and 84060001 and above 2041050 1
22 Switch Detent Spring 822643 1
23 Switch Knob 648706 1
24 Calibration Button 1564889 1
* WTo avoid shock, injury, or damage to the meter, use exact replacement only.
** Fluke accessories are available from your authorized Fluke distributor.
Bottom Shield and Calibration Button, Beeper not included)
1564823
1580498
1580480
1580500
1591675
1589129
1589118
1560052
1560842
1587908
1587896
1587881
1564806
519116
648755
2095692 1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
19
Page 20
175, 177, & 179
Calibration Information
14
13
21
15
4X
11
5X
16
24
12
18
2X
17
6
5
4
3
22
2
19
20
1
10
23
9
8
7
aik16f.eps
Figure 5. User-Replaceable Parts and Accessories
20
Page 21
True RMS Multimeters User-Replaceable Parts and Accessories
Lifetime Limited Warranty
Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As
used herein, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty
period shall be at least ten years from the date of purchase. This warranty does not cover fuses, disposable batteries,
damage from neglect, misuse, contamination, alteration, accident or abnormal conditions of operation or handling, including
failures caused by use outside of the product’s specifications, or normal wear and tear of mechanical components. This
warranty covers the original purchaser only and is not transferable.
For ten years from the date of purchase, this warranty also covers the LCD. Thereafter, for the lifetime of the DMM, Fluke
will replace the LCD for a fee based on then current component acquisition costs.
To establish original ownership and prove date of purchase, please complete and return the registration card
accompanying the product, or register your product on http://www.fluke.com
replace or refund the purchase price of a defective product purchased through a Fluke authorized sales outlet and at the
applicable international price. Fluke reserves the right to charge for importation costs of repair/replacement parts if the
product purchased in one country is sent for repair elsewhere.
If the product is defective, contact your nearest Fluke authorized service center to obtain return authorization information,
then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB
Destination). Fluke assumes no risk for damage in transit. Fluke will pay return transportation for product repaired or
replaced in-warranty. Before making any non-warranty repair, Fluke will estimate cost and obtain authorization, then
invoice you for repair and return transportation.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY
CAUSE OR THEORY. AUTHORIZED RESELLERS ARE NOT AUTHORIZED TO EXTEND ANY DIFFERENT
WARRANTY ON FLUKE’S BEHALF. Since some states do not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held
invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity
or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherlands
. Fluke will, at its option, repair at no charge,
2/02
21
Page 22
175, 177, & 179
Calibration Information
22
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.