Carefully read this operating manual before using the device for the first time.
Also observe the operating manual enclosed with the accessories.
The operating manual should be considered as part of the product and stored in a location that is easily
accessible.
When passing the device on to third parties, be sure to include this operating manual.
If this manual is lost, please request another one. The latest version can be found on our website
www.eppendorf.com (international) or www.eppendorfna.com (North America).
1.2Danger symbols and danger levels
1.2.1Hazard icons
5
Electric shockCrushing
Material damageHazard point
Heavy loadsExplosive
1.2.2Degrees of danger
The following danger levels are used in safety messages throughout this manual.
DANGERWill lead to severe injuries or death.
WARNINGMay lead to severe injuries or death.
CAUTIONMay lead to light to moderate injuries.
NOTICEMay lead to material damage.
Operating instructions
New Brunswick
6
English (EN)
1.3Symbols used
ExampleMeaning
You are requested to perform an action.
1.
2.
• List.
™
Innova® 2300/2350 Shakers
Perform these actions in the sequence described.
References useful information.
Safety
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
2Safety
2.1User profile
The device may only be operated by trained lab personnel who have carefully read the operating manual
and are familiar with the device functions.
CAUTION!
This equipment must be operated as described in this manual. If operational guidelines are
not followed, equipment damage and personal injury can occur. Please read the entire
Operating Manual before attempting to use this unit.
Do not use this equipment in a hazardous atmosphere or with hazardous materials for
which the equipment was not designed.
Eppendorf is not responsible for any damage to this equipment that may result from the
use of an accessory not manufactured by Eppendorf.
7
2.2Warnings for intended use
WARNING! Heavy!
Do not attempt to lift the Innova 2300/2350 Shaker by yourself.
Ask for assistance or use suitable equipment when raising or handling the device.
WARNING! Risk of electric shock and/or damage to the device!
Use a grounded power supply.
WARNING! Risk of electric shock and/or damage to the device!
Before cleaning device, turn off and unplug from mains/power supply.
WARNING! Risk of electric shock when replacing fuses!
Turn off shaker and disconnect from mains/power supply.
WARNING! Risk of explosion and injury or death!
Do not use equipment with flammable substances or organisms with flammable
by-products.
NOTICE! Risk of damage to device!
Never run shaker without a platform.
Safety
New Brunswick
8
English (EN)
™
Innova® 2300/2350 Shakers
CAUTION!
A minimum load is required to safely operate the Innova 2300/2350
The minimum load that the unit can safely operate at under maximum speed (500 rpm) is
10.9 kg (25 lb). This includes the weight of the platform, flasks and media.
Without a minimum load there is risk of instability that may cause serious injuries.
3Product description
3.1Main Illustration
Abb. 3-1: Innova 2300/2350 front view
1
3
™
New Brunswick
Innova® 2300/2350 Shakers
9
English (EN)
MAINT
Product description
SET
TIME
MUTE
START
STOP
SELECT
RPM
HRS
° C
*
2
Fig. 3-1:Innova 2300/2350 front view
1 Platform
3 Control panel
2 Temperature/monitor RTD probe and bracket
Optional
Unless otherwise indicated, all drawings represent both Innova 2300 and Innova 2350 models.
10
Product description
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
Abb. 3-2: Innova 2300/2350 rear view
1
8
7
Fig. 3-2:Innova 2300/2350 rear view
1 RTD connectos
Optional
2 Fuse
Control circuit
3 On/Off switch
4 Voltage selector
234
56
5 Universal power module
6 Power cord connection
7 Chart recorder connection
Optional
8 Cover plate
3.2Features
The Innova 2300/2350 is a benchtop or floor shaker utilizing a triple eccentric counter-balanced drive to
provide horizontal plane r otary motion in a 25 (1in) or 51 m m (2 in) circular orbit. A Proportional/Integral
(PI) Microprocessor controller with instantaneous digital feedback controls the speed over a r ange of 25 5 00 rpm.
This manual is intended to provide the user with a complete understanding of:
• Installation
• Operation
• Basic components of the device
• Basic preventive maintenance and service of the device
Be sure to familiarize yourself with this manual prior to operating your Innova 2300/2350.
Product description
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
3.2.1Operation
The shaker may be operated either continuously or in a timed mode via a programmable timer for shaking
periods of 0 hr 1 min – 99.9 hr.
A Temperature Monitoring option is available for the measurement, display and documentation of sample
temperature.
3.2.2Orbit
• Triple eccentric counterbalanced drive
• 25.4 mm (1 in) diameter circular orbit or 50.8 mm (2 in) diameter circular orbit
3.2.3Alarms
The Innova 2300/2350 is equipped with audible and visible alarms that alert the user to the following
conditions:
11
• The end of a timed run
• Deviation of shaking speed outside the tolerance limits
3.2.4Platforms and accessories
To accommodate customer needs, a wide variety of platforms can be used with the Innova 2300/2350.
Dedicated platforms are available for a variety of flask sizes. Universal platforms are also available.
3.2.5Main control board
The main control board for the Innova 2300/2350 Shaker has the following functions:
• Non-volatile memory for storage of key parameters during power interruption
• Speed sensing, electronic commutation and power control for the brushless DC drive motor
• Maintains an elapsed running time clock
• Contains firmware for shaker control, as well as recognition of an expansion connector for option
modules
• Provides an operator interface via displays, audible alarm and connection to the keypad module (keys
and display graphics)
The optional temperature module is designed to piggyback onto the main board via an expansion
connector. It has the following functions:
• Control of analog power supplies.
• Signal conditioning of RTD sensor readings
• Provision of remote monitoring capabilities by supplying analog outputs for speed and temperature
which are compatible with chart recorders and analog data acquisition systems
12
Product description
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
3.3Universal power module
The power module contains a voltage card and fuse holder which are used to select the appropriate voltage
and fusing. This universal power entry system adapts to worldwide power requirements. Voltage and fusing
has been set prior to shipment. Innova shakers are available in 100 V, 120 V, 220 V and 240 V
configurations that accommodate both 50 and 60 Hz frequencies.
WARNING! Risk of electric shock and/or damage to device!
Do not plug the shaker into a power source until you check the voltage and frequency
settings.
Check the voltage selection on the power entry module against the main identification
label or serial tag, located on the side or rear of the unit.
3.4Control panel
The control panel is located on the front of the instrument. It serves as the operator interface, and consists
of:
• A START/STOP key
• An up arrow ( )
• A down arrow ( )
• A SELECT key
• A 3-digit LED display that provides numeric values, and some letter codes
• 4 function indicator lights
• 4 status indicator lights
Abb. 3-3: Control panel
New Brunswick
™
Innova® 2300/2350 Shakers
Product description
English (EN)
13
MAINT
MUTE
Fig. 3-3:Control panel
SET
TIME
START
STOP
2
31
RPM
HRS
° C
*
SELECT
4
1 Status indicators
2 LED display
3 Function indicators
4 Keypad
3.4.1LED display
The Innova control panel has a 3-digit LED display. During normal shaker operation, the display will
indicate:
• Shaker status (ON/OFF)
• Shaking speed
• Setpoints
• Hours remaining (timed run)
• Measured temperature (when temperature/monitor option is installed)
14
Product description
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
3.4.2Keypad
START/STOP• Starts and stops the drive
• Activates or stops timer when timed run is
desired
Keys
SELECT• Changes the displayed parameter
• Adjusts the setpoint of a displayed parameter up
or down
• Used to enter the SET mode for setpoint
changes
3.4.3Status indicators
MAINT• Illuminates after 10,000 hours of use
• Accumulated running time is internally
monitored and may be displayed as a guideline
• Indicates need for routine maintenance (see
Routine maintenance on p. 25)
SETIlluminates to indicate that:
• Shaker is in SET mode
• Setpoints are being displayed
• Setpoints can be altered
TIME• Illuminates to indicate that the timer is in
operation
MUTE• Illuminates to indicate that the audible alarm is
disabled
3.4.4Function indicators
RPMRevolutions per minute
HRSTime remaining
°CTemperature function (only activated of the
temperature/monitor option is installed)
*Not applicable
3.5Platform assemblies
The Innova 2300 can be used with a wide variety of 46 cm x 76 cm (18 in x 30 in) platforms which will
accept a variety of clamps for flasks, test tubes, etc.
The Innova 2350 can accommodate a wide variety of 61 cm x 91.4 cm (24 in x 36 in) platforms which will
accept a variety of clamps for flasks, test tubes, etc.