Carefully read this operating manual before using the device for the first time.
Also observe the operating manual enclosed with the accessories.
The operating manual should be considered as part of the product and stored in a location that is easily
accessible.
When passing the device on to third parties, be sure to include this operating manual.
If this manual is lost, please request another one. The latest version can be found on our website
www.eppendorf.com (international) or www.eppendorfna.com (North America).
1.2Danger symbols and danger levels
1.2.1Hazard icons
5
Electric shockCrushing
Material damageHazard point
Heavy loadsExplosive
1.2.2Degrees of danger
The following danger levels are used in safety messages throughout this manual.
DANGERWill lead to severe injuries or death.
WARNINGMay lead to severe injuries or death.
CAUTIONMay lead to light to moderate injuries.
NOTICEMay lead to material damage.
Page 6
Operating instructions
New Brunswick
6
English (EN)
1.3Symbols used
ExampleMeaning
You are requested to perform an action.
1.
2.
• List.
™
Innova® 2300/2350 Shakers
Perform these actions in the sequence described.
References useful information.
Page 7
Safety
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
2Safety
2.1User profile
The device may only be operated by trained lab personnel who have carefully read the operating manual
and are familiar with the device functions.
CAUTION!
This equipment must be operated as described in this manual. If operational guidelines are
not followed, equipment damage and personal injury can occur. Please read the entire
Operating Manual before attempting to use this unit.
Do not use this equipment in a hazardous atmosphere or with hazardous materials for
which the equipment was not designed.
Eppendorf is not responsible for any damage to this equipment that may result from the
use of an accessory not manufactured by Eppendorf.
7
2.2Warnings for intended use
WARNING! Heavy!
Do not attempt to lift the Innova 2300/2350 Shaker by yourself.
Ask for assistance or use suitable equipment when raising or handling the device.
WARNING! Risk of electric shock and/or damage to the device!
Use a grounded power supply.
WARNING! Risk of electric shock and/or damage to the device!
Before cleaning device, turn off and unplug from mains/power supply.
WARNING! Risk of electric shock when replacing fuses!
Turn off shaker and disconnect from mains/power supply.
WARNING! Risk of explosion and injury or death!
Do not use equipment with flammable substances or organisms with flammable
by-products.
NOTICE! Risk of damage to device!
Never run shaker without a platform.
Page 8
Safety
New Brunswick
8
English (EN)
™
Innova® 2300/2350 Shakers
CAUTION!
A minimum load is required to safely operate the Innova 2300/2350
The minimum load that the unit can safely operate at under maximum speed (500 rpm) is
10.9 kg (25 lb). This includes the weight of the platform, flasks and media.
Without a minimum load there is risk of instability that may cause serious injuries.
Page 9
3Product description
3.1Main Illustration
Abb. 3-1: Innova 2300/2350 front view
1
3
™
New Brunswick
Innova® 2300/2350 Shakers
9
English (EN)
MAINT
Product description
SET
TIME
MUTE
START
STOP
SELECT
RPM
HRS
° C
*
2
Fig. 3-1:Innova 2300/2350 front view
1 Platform
3 Control panel
2 Temperature/monitor RTD probe and bracket
Optional
Unless otherwise indicated, all drawings represent both Innova 2300 and Innova 2350 models.
Page 10
10
Product description
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
Abb. 3-2: Innova 2300/2350 rear view
1
8
7
Fig. 3-2:Innova 2300/2350 rear view
1 RTD connectos
Optional
2 Fuse
Control circuit
3 On/Off switch
4 Voltage selector
234
56
5 Universal power module
6 Power cord connection
7 Chart recorder connection
Optional
8 Cover plate
3.2Features
The Innova 2300/2350 is a benchtop or floor shaker utilizing a triple eccentric counter-balanced drive to
provide horizontal plane r otary motion in a 25 (1in) or 51 m m (2 in) circular orbit. A Proportional/Integral
(PI) Microprocessor controller with instantaneous digital feedback controls the speed over a r ange of 25 5 00 rpm.
This manual is intended to provide the user with a complete understanding of:
• Installation
• Operation
• Basic components of the device
• Basic preventive maintenance and service of the device
Be sure to familiarize yourself with this manual prior to operating your Innova 2300/2350.
Page 11
Product description
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
3.2.1Operation
The shaker may be operated either continuously or in a timed mode via a programmable timer for shaking
periods of 0 hr 1 min – 99.9 hr.
A Temperature Monitoring option is available for the measurement, display and documentation of sample
temperature.
3.2.2Orbit
• Triple eccentric counterbalanced drive
• 25.4 mm (1 in) diameter circular orbit or 50.8 mm (2 in) diameter circular orbit
3.2.3Alarms
The Innova 2300/2350 is equipped with audible and visible alarms that alert the user to the following
conditions:
11
• The end of a timed run
• Deviation of shaking speed outside the tolerance limits
3.2.4Platforms and accessories
To accommodate customer needs, a wide variety of platforms can be used with the Innova 2300/2350.
Dedicated platforms are available for a variety of flask sizes. Universal platforms are also available.
3.2.5Main control board
The main control board for the Innova 2300/2350 Shaker has the following functions:
• Non-volatile memory for storage of key parameters during power interruption
• Speed sensing, electronic commutation and power control for the brushless DC drive motor
• Maintains an elapsed running time clock
• Contains firmware for shaker control, as well as recognition of an expansion connector for option
modules
• Provides an operator interface via displays, audible alarm and connection to the keypad module (keys
and display graphics)
The optional temperature module is designed to piggyback onto the main board via an expansion
connector. It has the following functions:
• Control of analog power supplies.
• Signal conditioning of RTD sensor readings
• Provision of remote monitoring capabilities by supplying analog outputs for speed and temperature
which are compatible with chart recorders and analog data acquisition systems
Page 12
12
Product description
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
3.3Universal power module
The power module contains a voltage card and fuse holder which are used to select the appropriate voltage
and fusing. This universal power entry system adapts to worldwide power requirements. Voltage and fusing
has been set prior to shipment. Innova shakers are available in 100 V, 120 V, 220 V and 240 V
configurations that accommodate both 50 and 60 Hz frequencies.
WARNING! Risk of electric shock and/or damage to device!
Do not plug the shaker into a power source until you check the voltage and frequency
settings.
Check the voltage selection on the power entry module against the main identification
label or serial tag, located on the side or rear of the unit.
3.4Control panel
The control panel is located on the front of the instrument. It serves as the operator interface, and consists
of:
• A START/STOP key
• An up arrow ( )
• A down arrow ( )
• A SELECT key
• A 3-digit LED display that provides numeric values, and some letter codes
• 4 function indicator lights
• 4 status indicator lights
Page 13
Abb. 3-3: Control panel
New Brunswick
™
Innova® 2300/2350 Shakers
Product description
English (EN)
13
MAINT
MUTE
Fig. 3-3:Control panel
SET
TIME
START
STOP
2
31
RPM
HRS
° C
*
SELECT
4
1 Status indicators
2 LED display
3 Function indicators
4 Keypad
3.4.1LED display
The Innova control panel has a 3-digit LED display. During normal shaker operation, the display will
indicate:
• Shaker status (ON/OFF)
• Shaking speed
• Setpoints
• Hours remaining (timed run)
• Measured temperature (when temperature/monitor option is installed)
Page 14
14
Product description
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
3.4.2Keypad
START/STOP• Starts and stops the drive
• Activates or stops timer when timed run is
desired
Keys
SELECT• Changes the displayed parameter
• Adjusts the setpoint of a displayed parameter up
or down
• Used to enter the SET mode for setpoint
changes
3.4.3Status indicators
MAINT• Illuminates after 10,000 hours of use
• Accumulated running time is internally
monitored and may be displayed as a guideline
• Indicates need for routine maintenance (see
Routine maintenance on p. 25)
SETIlluminates to indicate that:
• Shaker is in SET mode
• Setpoints are being displayed
• Setpoints can be altered
TIME• Illuminates to indicate that the timer is in
operation
MUTE• Illuminates to indicate that the audible alarm is
disabled
3.4.4Function indicators
RPMRevolutions per minute
HRSTime remaining
°CTemperature function (only activated of the
temperature/monitor option is installed)
*Not applicable
3.5Platform assemblies
The Innova 2300 can be used with a wide variety of 46 cm x 76 cm (18 in x 30 in) platforms which will
accept a variety of clamps for flasks, test tubes, etc.
The Innova 2350 can accommodate a wide variety of 61 cm x 91.4 cm (24 in x 36 in) platforms which will
accept a variety of clamps for flasks, test tubes, etc.
Page 15
Product description
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
3.6Optional features
3.6.1Temperature/monitor option
A Temperature/Monitor Option is available for installation on the Innova 2300/2350. The temperature of
liquid in any vessel or the ambient temperature can be measured using the RTD electronics-based
measuring device supplied with this option. When the option is installed, the LED will display the measured
temperatures in 0.1 °C increments.
This option also allows the connection of a chart recorder so that temperature and shaking speed can be
documented. The analog output for shaking speed is 0 – 5 V, 1 V per 100 rpm. For temperature, the output
is 0 – 5 V with .05 V per °C. The output can also be connected to a data logging computer with an analog
data acquisition card.
3.6.2Capacity upgrade option for 2300
It is possible to significantly increase the capacity of your Innova 2300 Shaker with an available capacity
upgrade package (part number M1191-9905). This option will convert an Innova 2300 to an Innova 2350
simply and easily.
15
This package consists of an appropriate counterweight, outrigger feet and hardware. Large capacity Innova
2350 platforms must be used with the Innova 2350 configuration. For a listing of available platforms, (see
Innova 2300 interchangeable platforms on p. 33) and (see Innova 2350 interchangeable platforms on p. 34).
3.6.3Quick change platform option
This accessory enables the user to snap in platforms without tools or hardware. This is especially handy for
those users who change platforms frequently. The kit includes a subplatform with spring clips, an extra
counterweight, and hardware for installation. This option is available on the Innova 2300 only. A speed
range of 25 – 400 rpm is recommended when this option is installed.
3.7Heavy duty construction
3.7.1Triple Eccentric drive
The triple eccentric drive used in the Innova shakers employs the same proven technology that has reliably
driven New Brunswick shakers for over 50 years. This drive mechanism utilizes a counterweight system to
stabilize the rotary motion produced during operation.
When the workload moves in one direction, opposing forces are generated to stabilize the shaker. This
action will help eliminate the problem of “walking” that may occur with less precisely balanced
instruments. Vibrations are minimized and the life of the unit is extended.
Innova shakers employ sealed lubricated ball bearings of the highest quality. Sealed bearings minimize the
generation of airborne particles that could be disadvantageous in clean rooms or controlled environment
areas.
3.7.3Motor
The Innova 2300/2350 Shaker uses a brushless ball bearing DC motor. This low profile motor provides high
torque along with quiet, efficient operation and low maintenance. The rugged motor has a rating of 1/8
horsepower.
Page 17
Installation
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
4Installation
4.1Unpacking
Upon unpacking the unit, inspect it carefully for any apparent damage that may have occurred during
transit. Immediately report any damage to the carrier and to the Eppendorf Service Department. Do not
discard the crate or packing material.
VERY IMPORTANT: There are 2 small plastic straps that hold the bearing housing in place for
shipping. Be sure to remove these 2 straps before operation.
4.2Checking voltage configuration
WARNING! Risk of electric shock and/or damage to device!
17
Do not plug the shaker into a power source until you check the voltage and frequency
settings.
Do not plug your shaker into the power source until you determine that voltage of your unit matches your
electrical service. This can be determined by checking the voltage indicator and label on the rear of the
unit.
4.3Space requirements
It is essential that the unit be situated in an area where there is sufficient space for the shaker and platform
to clear walls and other potential obstructions during operation.
The dimensions, including a platform but excluding glassware, are:
ModelWidthDepthHeight
Innova 2300 without capacity upgrade74 cm (29 in)56.5 cm (22 1/4 in) 17 cm (6 3/8 in)
Innova 2350, Innova 2300 with capacity
upgrade
The effective surface area required for operation is:
91 cm (36 in)61 cm (24 in)17 cm (6 3/8 in)
ModelWidthDepth
Innova 2300 without capacity upgrade81 cm (32 in)64 inch (25 in)
Innova 2350, Innova 2300 with capacity upgrade99 cm (39 in)68.5 inch (27 in)
Page 18
18
Installation
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
WARNING!
Do not position the equipment so that it is difficult to operate the disconnecting device
(main plug).
For service reasons and for cases of emergencies, sufficient space must be provided for
easy access to the main power cord/power switch (located at the back of the unit).
4.4Electrical connections
WARNING! Risk of electric shock and/or damage to device!
Use a grounded power supply.
CAUTION!
A detachable main power cord has been provided with your unit.
Only use the cord provided with the unit.
Before making electrical connections, be sure to follow these instructions:
1. Check that the power module at the rear of the unit is set to the correct voltage, corresponding to your
power source.
2. Remove the caution label from the universal power module.
3. Make sure the plastic straps have been removed from the upper bearing housing.
4. Make sure the power switch on the back of the unit is set in the OFF position.
ONLY THEN:
5. Connect the line cord to the power module and to a grounded electrical outlet.
4.5Install the platform
The Innova 2300/2350 can be used with a variety of Eppendorf platforms that will accept a wide range of
clamps for flasks, test tubes, etc. A platform, which is required for operation, is a separate item not included
with the shaker assembly (see Platform capacity on p. 29).
CAUTION! Risk of damage to the device!
Never run the shaker without a properly installed platform.
Page 19
™
New Brunswick
A platform must be installed on the device prior to use. To install a platform:
1. Set the power switch in the Off position.
2. Using the 7/32 in hex key provided, remove the 4 Allen head platform screws installed in the
subplatform of the bearing housing (under the shaker).
One of the 4 screws has a tag so you can identify the platform screws. Set them aside for reuse.
3. Place the platform on top of the shaker’s subplatform.
Be sure to use the proper size platform for your particular model shaker
4. Reinstall and tighten the 4 platform screws (previously set aside) with the hex key to secure the
platform.
Innova® 2300/2350 Shakers
Installation
English (EN)
4.6Quick change platform option
If the quick change platform option is installed:
19
1. Slip the appropriately sized platform between the side guides and push the platform to the rear retainer.
2. Press down on the front edge of the platform.
The platform should snap down into place and be retained by the springs.
3. Check that the rear edge of the platform is engaged under the bend of the rear clip.
Page 20
20
Installation
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
4.7Install flask clamps
WARNING!
Do not overfill flasks (i.e., more than 50% of the flask volume).
Eppendorf flask clamps are used on a variety of shaker platforms. Flat Phillips head screws of
different lengths and thread pitch are used to secure the clamp (see Clamp hardware on p. 35).
All clamps are shipped complete with hardware.
Clamps for 2 – 6 L flasks are shipped with an additional girdle to keep the flasks in place. To install 2 – 6 L
clamps:
Abb. 4-1: Double girdle clamp
1
4
Fig. 4-1:Double girdle clamp
The instructions and illustration above also apply to 2800 ml Fernbach flask clamps.
1 Clamp mounting holes
With screws
2 Upper girdle with girdle tubes
Secures flask within the clamp
2
1. Place the clamp on the platform and secure it in
place with the correct type of screws.
2. Place the loose girdle around the upper portion of
the clamp body so that it is held in place by the
clamp legs.
3. Insert the flask into the clamp.
3
3 Lower girdle with girdle tubes
Prevents flask from spining
4 Clamp body (legs and base)
Page 21
Operation
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
5Operation
5.1Getting started
To start the instrument, turn the power switch to the ON position.
When the shaker is running, the LED display will track the speed as it accelerates to the last entered
setpoint. The shaking action may be stopped or started by pressing the START/STOP key.
5.2Continuous/unlimited run
1. If the LED displays OFF, press the START/STOP key.
2. Press the SELECT key to illuminate RPM.
3. Press either or to enter SET mode (SET indicator will illuminate).
4. Set the speed by using the or key until the desired setpoint is displayed.
Holding the or key will cause the setting to change more rapidly.
21
The setpoint may be changed at any time during a run without stopping the shaker by following steps 2 – 4.
During speed changes, the alarm may sound until the speed returns to within 5 rpm of the setpoint.
5.3Checking setpoints
To check any setpoint:
1. Press the SELECT key to illuminate the desired indicator.
2. Press either or to enter the SET mode and display the current setpoint.
Holding the or key for more than 0.5 s causes the setpoint to change. Should this occur,
resetting will be necessary.
5.4Timed functions
The shaker may be programmed to stop automatically after a preset time period of 0.1 to 99.9 hr. There
must be power to the shaker in order to set the timer. However, a timed run can be initiated while the unit
is either shaking or stopped.
5.4.1Setting the timer
1. Press the SELECT key to light HRS.
2. Press either or to enter the SET mode, then set a time period of 0.1 – 99.9 hr.
3. While SET light is illuminated, press the START/STOP key to program the time (and start the run).
Page 22
22
Operation
New Brunswick
English (EN)
The TIME indicator will illuminate and remain on for the duration of the run. At the end of the timed run:
• The display will read OFF
• The time indicator will flash
• The audible alarm will sound
The setpoint may be changed during a run without stopping the shaker by following steps 1 and 2 above.
™
Innova® 2300/2350 Shakers
5.4.2Canceling the timer
To cancel the timer without stopping the shaker:
1. Press the SELECT key to illuminate HRS.
2. Press either or to enter the SET mode, then immediately press the START/STOP key.
The Time indicator will go out and the display will read “OFF”.
5.5Alarm functions
Innova shakers have an audible alarm which is activated at predetermined times. It is possible to deactivate
and reactivate the alarm, according to your needs.
5.5.1Stopping the alarm
When the alarm sounds, you can stop it by pressing the SELECT key and changing to any other function.
5.5.2Deactivating the alarm
The alarm may be deactivated in the following way:
1. Press SELECT to illuminate HRS.
2. Simultaneously press the and keys.
The SET and MAINT indicators will flash.
3. While the SET and MAINT indicators are flashing, press the START/STOP key.
The MUTE indicator will illuminate to advise that the audible alarm is deactivated.
5.5.3Reactivating the alarm
The alarm may be reactivated in the following way:
1. Press SELECT to illuminate HRS.
2. Simultaneously press the and keys.
The SET and MAINT indicators will flash.
3. While the SET and MAINT indicators are flashing, press the START/STOP key.
Page 23
™
New Brunswick
The MUTE indicator will extinguish to advise that the audible alarm is active.
The shaker may be stopped or started by pressing the START/STOP key. When starting, the
unit will automatically return to the last function and speed setting.
The audible alarm will sound until the speed is within 5 rpm of the setpoint.
The alarm will not sound while the shaker is accelerating immediately following turning on the power.
Innova® 2300/2350 Shakers
Operation
English (EN)
5.6Total running time
The control modules of the Innova shakers keep track of the time the shaker has been on, tracking the
hours of usage.
To display the accumulated running time:
1. Select HRS using the SELECT key.
23
2. Simultaneously press the and keys.
The SET and MAINT indicators will flash and the accumulated running time will be displayed in
hundreds of hours (i.e., “02” equals 200 hours; “102” equals 10,200 hours). This display will continue
for 10 seconds and then default to the previous mode readout.
After 10,000 hours of operation, the MAINT indicator will light. Preventive maintenance is
recommended at this point. The Eppendorf Service Technician will deactivate the light when
the technician performs the required maintenance.
5.7Temperature/monitor option
This option consists of an internal electrical interface, an RTD temperature probe and an analog output for
chart recorder or data aquisition system. When this option is installed, either the ambient temperature or
the temperature of any vessel on the shaker platform can be measured with the probe:
1. Remove the probe from its holder and insert it into the vessel to be monitored.
2. Use the SELECT key to indicate °C.
The °C indicator will only function when the temperature/monitor option is installed.
Since the temperature/monitor option does not provide temperature control, any attempt to enter a
temperature setpoint results in the display of Err on the control panel.
Page 24
24
Operation
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
5.8Recorder adaptation
To record speed or temperature, an auxiliary recorder (not supplied, but available from Eppendorf) can be
used. The recorder should have the following capabilities:
Abb. 5-1: Recorder connector as seen from the rear of the unit
4
5
Fig. 5-1:Recorder connector as seen from the
Pin numberSignal nameScale
6Speed1 V = 100 rpm
2Ground
7Temperature1 V = 20 °C
3Ground
3
7
8
9
rear of the unit
21
6
• For speed, each channel should have signal
conditioning that accepts 0 – 5 V input.
• For temperature, each chanel should have a
signal conditioning that accepts 0 – 5 V input
• A mating connector is required on the recorder
cable (not supplied, but available from
Eppendorf). This is a 9-pin male D subminiature
connector, AMP Amplimite HDP-20 series or
equivalent.
Page 25
Maintenance
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
6Maintenance
WARNING!
Before performing any service, turn off the power, using the ON/OFF switch on the rear
panel, and disconnect the power cord.
6.1Routine maintenance
No routine maintenance is required for the Innova 2300/2350 Shaker.
The MAINT indicator light illuminates at the end of 10000 hours of use. At that time, contact your local
Eppendorf Service Engineer, or call the Eppendorf Service Department. This periodic maintenance will
keep your unit in premium condition.
25
6.2Cleaning external and internal surfaces
WARNING! Personnel injury and equipment damage!
When cleaning the unit, always turn off the shaker and disconnect the power cord from the
power supply.
The unit may be cleaned using a damp cloth or any standard household or laboratory cleaner to wipe down
its outer surfaces. Do not use abrasive or corrosive compounds to clean this instrument, as they may
damage the unit.
6.3Fuse replacement
WARNING! Risk of electric shock when replacing fuses!
Turn off shaker and disconnect from mains/power supply.
CAUTION!
A detachable main power cord has been provided with your unit.
Only use the cord provided with the unit.
Page 26
26
Maintenance
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
To replace fuses (without changing the fusing arrangement):
1. Disconnect the unit from the power source.
2. Using a small screwdriver, remove the cover/fuse block located on the rear of the unit.
3. Remove the old fuse.
4. Insert a new one of the same type.
5. Replace the cover/fuse block into the power module.
Spare fuses are supplied with the unit.
Abb. 6-1: Power entry module (2300/2350)
321
Fig. 6-1:Power entry module (2300/2350)
1 Housing
2 Voltage selector card
3 Indicator pin
5
4
4 Cover
5 Fuse block
Page 27
Technical data
English (EN)
New Brunswick
™
Innova® 2300/2350 Shakers
7Technical data
7.1Specifications
7.1.1Innova 2300
Tab. 7-1: Shaking
Speed•25 rpm – 500 rpm• 25 rpm – 300 rpm
Motion• 25.4 mm (1 in) diameter circular orbit• 50.8 mm (2 in) diameter circular
orbit
Indication• LED digital electric display
•1 rpm increments
Setpoint and
control
Accuracy•±1 rpm
• Digital adjustment with PI microprocessor control and instantaneous visual
feedback
27
At 25 – 400 rpm, the unit will perform to specifications with up to ±10 % line voltage
fluctuation. To attain speed accuracy at 401 – 500, the line voltage cannot be lower than 5 %
of the rated voltage.
Keypad timer• Programmable shaking periods from 0 hr 1 min – 99.9 hr by a digital timer
• Timer shuts off at the end of shaking period and energizes status light
• Timer counts down and digital display indicates remaining time
• Can be deactivated for continuous operation
• Unit will display total accumulated running time for service information
Operating ambient
environment
Self-diagnostic
status
Remote speed
monitoring
(optional)
Automatic restart• Unit will automatically restart after undesired power interruption
Motor• 1/8 HP
Electrical service• 100 V, 120 V, 220 V, 240 V
Electrical
protection
•0 °C – 60 °C
• Up to 90 %, non-condensing
• Up to 2000 m
• Warning signal (audible and visible) indicates when shaking speed deviates
more than 5 rpm
• Warning signal (audible and visible) indicates when timer operation has expired
• Audible alarm can be deactivated/reactivated by the operator
• Chart recorder output for speed 0.5 V, 1 V per 100 rpm
• Accuracy ±25 mV
• Setpoints are maintained by non-volatile memory
• Interruption is indicated by a flashing LED
• 3 phase
• Brushless ball bearing DC motor
• All voltages 50/60 Hz 150 VA
• Universal power entry system adapts to U.S. or international requirements
• Main fuse(s) in power entry module
• Control circuits provided with separate fuse
Page 28
28
Technical data
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
Contruction• Heavy gauge steel
• Phosphate coated
• Textured painted frame
WeightNet• 49 kg (108 lb)
Gross• 66 kg (145 lb)
Dimensions
Width74 cm (29 in)
Depth56.5 cm (22 1/4 in)
Height17 cm (6 3/8 in)
Platform dimensions
Width46 cm (18 in)
Depth76 cm (30 in)
7.1.2Innova 2350
Tab. 7-2: Shaking
Speed•25 rpm – 400 rpm• 25 rpm – 300 rpm
Motion• 25.4 mm (1 in) diameter circular orbit• 50.8 mm (2 in) diameter circular
orbit
Indication• LED digital electric display
•1 rpm increments
Setpoint and
control
Accuracy•±1 rpm
At 25 – 400 rpm, the unit will perform to specifications with up to ±10 % line voltage
fluctuation. To attain speed accuracy at 401 – 500, the line voltage cannot be lower than 5 %
of the rated voltage.
Keypad timer• Programmable shaking periods from 0 hr 1 min – 99.9 hr by a digital timer
Operating ambient
environment
• Digital adjustment with PI microprocessor control and instantaneous visual
feedback
• Timer shuts off at the end of shaking period and energizes status light
• Timer counts down and digital display indicates remaining time
• Can be deactivated for continuous operation
• Unit will display total accumulated running time for service information
•0 °C – 60 °C
•90 %
• Non-condensing
• Up to 2000 m
Page 29
New Brunswick
™
Innova® 2300/2350 Shakers
Technical data
English (EN)
29
Self-diagnostic
status
Remote speed
monitoring
(optional)
Automatic restart• Unit will automatically restart after undesired power interruption
Motor• 1/8 HP
Electrical service• 100 V, 120 V, 220 V, 240 V
Electrical
protection
Contruction• Heavy gauge steel
WeightNet• 51 kg (112 lb)
• Warning signal (audible and visible) indicates when shaking speed deviates
more than 5 rpm
• Warning signal (audible and visible) indicates when timer operation has expired
• Audible alarm can be deactivated/reactivated by the operator
• Chart recorder output for speed 0.5 V, 1 V per 100 rpm
• Accuracy ±25 mV
• Setpoints are maintained by non-volatile memory
• Interruption is indicated by a flashing LED
• 3 phase
• Brushless ball bearing DC motor
• All voltages 50/60 Hz 150 VA
• Universal power entry system adapts to U.S. or international requirements
• Main fuse(s) in power entry module
• Control circuits provided with separate fuse
• Phosphate coated
• Textured painted frame
Gross• 68 kg (150 lb)
Dimensions
Width91 cm (36 in)
Depth61 cm (24 in)
Height17 cm (6 3/8 in)
Platform dimensions
Width61 cm (24 in)
Depth91 cm (36 in)
7.2Platform capacity
Universal platforms have multiple holes enabling you to mount an assortment of flask clamps or other
accessories on a single platform. The capacities shown in this section represent the maximum number of
flasks in a given size that will fit on the platform in a balanced pattern. Universal platform, clamps and
accessories are sold separately.
When just one size flask will be used on the shaker (i.e. 250 mL flasks), dedicated platforms come with flask
clamps already mounted. Dedicated platforms generally will hold a greater number of flasks than the
universal platform, but do not offer the versatility.
Page 30
30
Technical data
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
7.2.1Innova 2300 platform capacities
For universal platforms that measure 46 cm (18 in) × 76 cm (30 in):
DescriptionAmount
10 mL flask183
25 mL flask92
50 mL flask92
125 mL flask39
250 mL flask30
500 mL flask18
1 L flask12
2 L flask8
2.8 L flask6
4 L flask6
5 L flask6
6 L flask4
Large TT rack7
Medium TT rack9
Small TT rack9
Microplate rack (stack)16
Microplate rack (1 layer)4
For dedicated platforms that measure 46 cm (18 in) × 76 cm (30 in):
DescriptionAmount
50 mL flask108
125 mL flask60
250 mL flask40
500 mL flask24
1 L flask15
2 L flask12
2.8 L flask6
4 L flask6
6 L flask4
7.2.2Innova 2350 platform capacities
Page 31
New Brunswick
For universal platforms that measure 61 cm (24 in) × 91 cm (36 in):
DescriptionAmount
125 mL flask75
250 mL flask50
500 mL flask30
1 L flask24
2 L flask15
2.8 L flask12
4 L flask10
5 L flask10
6 L flask8
Large TT rack12
Medium TT rack13
Small TT rack18
Microplate rack (stack)27
Microplate rack (1 layer)8
™
Innova® 2300/2350 Shakers
English (EN)
Technical data
31
For dedicated platforms that measure 61 cm (24 in) × 91 cm (36 in):
DescriptionAmount
125 mL flask96
250 mL flask70
500 mL flask40
1 L flask24
2 L flask15
2.8 L flask12
4 L flask12
6 L flask6
Page 32
32
Technical data
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
Page 33
New Brunswick
™
8Ordering information
8.1Service parts (2300/2350)
Part numberDescriptionQuantity
P0380-37100.16A / 250V / Time Delay - Fuse1
P0380-35301.6A / 250V / Time Delay - Fuse1
P0420-161010VA Transformer1
M1191-5300130VA Transformer Assembly1
P0320-03502100μF Capacitor1
P0460-4791Diode Bridge1
P0360-4040130V Varistor2
M1191-4000Large Motor Assembly1
M1190-9941Main Control P.C.B.1
M1190-5000Membrane Switch Panel1
P0460-2200Power Entry Module1
P0720-2053Power Cord 120V 10A1
P0720-2021Power Cord 220V1
P0180-0281Bearing Shield Upper3
P0180-0282Bearing Shield Lower6
R-336Belt1
M1191-6331Bearing Housing Assembly,
1-inch Circular Orbit
M1194-8000Stainless Steel RTD Assembly
(Optional)
M1191-0050Innova 2300/2350 Operations
Manual
M1191-6332Bearing Housing Assembly.,
2-inch Circular Orbit
1
1
1
1
Ordering information
Innova® 2300/2350 Shakers
English (EN)
33
8.2Innova 2300 interchangeable platforms
Catalog number46 cm × 76 cm (18 in × 30 in) plarform size
Clamp quantitySize of glassware
M1250-9920Universal Platform
M1191-990810850mL Erlenmeyer Flask
M1191-990960125mL Erlenmeyer Flask
M1191-991040250/300mL Erlenmeyer Flask
M1191-991124500mL Erlenmeyer Flask
AG-1151L Erlenmeyer Flask
AG-2122L Erlenmeyer Flask
Page 34
34
Ordering information
New Brunswick
English (EN)
Catalog number46 cm × 76 cm (18 in × 30 in) plarform size
AG-464L Erlenmeyer Flask
AG-646L Erlenmeyer Flask
AG-2862800mL Fernbach Flask
™
Innova® 2300/2350 Shakers
Clamp quantitySize of glassware
8.3Innova 2350 interchangeable platforms
The Innova 2350 is counterweighted and balanced for a 61 x 91 cm (24 in x 36 in) platform. It has a speed
range of 25 rpm – 500 rpm. Maximum recommended speed may be limited to 400 rpm under certain
conditions.
Catalog number60 cm × 91 cm (24 in × 36 in) platform size
Clamp quantitySize of glassware
PTL-386Universal Platform
M1191-991296125mL Erlenmeyer Flask
M1191-991370250/300mL Erlenmeyer Flask
M1191-991440500mL Erlenmeyer Flask
AG-21-1241L Erlenmeyer Flask
AG-21-2152L Erlenmeyer Flask
AG-21-4124L Erlenmeyer Flask
AG-21-666L Erlenmeyer Flask
AG-21-8122800mL Fernbach Flask
8.4Accessory flask clamps
All clamps listed are made of stainless steel.
Catalog numberClamp type
ACE-10510 mL Erlenmeyer
ACE-25525 mL Erlenmeyer
ACE-50550 mL Erlenmeyer
ACE-125S125 mL Erlenmeyer
ACE-250S250 mL Erlenmeyer
ACE-500S500 mL Erlenmeyer
ACE-1000S1.0 L Erlenmeyer
ACE-2000S2.0 L Erlenmeyer
ACE-4000S4.0 L Erlenmeyer
Page 35
New Brunswick
Catalog numberClamp type
ACE-6000S6.0 L Erlenmeyer
ACFE-2800S2.8 L or 2800 mL Fernbach
ACSB-500S500 mL Media Bottles
ACSB-1000Sfor 1 L Media Bottles
™
Ordering information
Innova® 2300/2350 Shakers
English (EN)
8.5Clamp hardware
Eppendorf flask clamps are used on a variety of shaker platforms. Flat Phillips and flat head screws of
different lengths and thread pitch are used to secure the clamp. The following tables identify the proper
screw for your shaker application by reference to the head style.
Tab. 8-1: 10 – 500 mL clamp hardware application chart
DescriptionPart numberQty.Application
10-24 x 5/8 (15.87 mm) flat
Phillips (+) head screw
10-24 x 5/16 (7.9 mm) flat
Phillips (+) head screw
10-32 x 5/16 (7.9 mm) flat
slotted (-) head screw
S2116-31011 19.05 mm (3/4 in) thick wood platform
S2116-30511
S2117-30501All stainless steel platforms
7.9 mm (5/16 in) thick aluminum,
phenolic and stainless steel platforms
10-24 x 5/8 (15.87 mm)
flat Phillips (+) head screw
10-24 x 5/16 (7.9 mm) flat
Phillips (+) head screw
10-32 x 5/16 (7.9 mm) flat
slotted (-) head screw
The 1 – 6 liter chart also applies to 2800 mL Fernbach flask clamps.
S2116-31015 19.05 mm (3/4 in) thick wood platform
S2116-30515 7.9 mm (5/16 in) thick aluminum,
phenolic and stainless steel platforms
S2117-30505All stainless steel platforms
Page 36
36
Ordering information
New Brunswick
English (EN)
™
Innova® 2300/2350 Shakers
8.6Optional kits
8.6.1Temperature/monitor option
The temperature and speed monitoring kit (part number M1191-9924) allows an Innova 2300 or 2350 to
measure and display sample or ambient temperature and to record it on a remote chart recorder or
computer. This kit consists of:
• An internal electrical interface
• An RTD temperature probe
• An analog output for chart recorder or computer
The package does not include a chart recorder.
This option must be installed by an authorized service technician.
8.6.2Quick change platform kit
The easy-loading Quick Change Platform kit (part number M1191-9904) enables users :of the Innova 2300
(without capacity upgrade) to change or mount any 46 cm x 76 cm (18 in x 30 in) platform without the use
of tools or hardware.
This kit must be installed by an authorized service technician.
8.6.3Capacity upgrade option
With this kit (part number M1191-9905), an Innova 2300, which has an 46 cm x 76 cm (18 in x 30 in)
platform, can be converted to an Innova 2350, with a 61 cm x 91 cm (24 in x 36 in) platform.
The kit consists of:
• Counterweighting
• Hardware
• Outrigger feet
Platforms must be purchased separately.
This option must be installed by a qualified service engineer.
Page 37
Transport, storage and disposal
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
9Transport, storage and disposal
9.1Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations
based on EU Directive 2012/19/EU pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business
sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To
document this, they have been marked with the following identification:
37
Because disposal regulations may differ from one country to another within the EU, please contact your
supplier if necessary.
In Germany, this is mandatory from March 23, 2006. From this date, the manufacturer has to offer a
suitable method of return for all devices supplied after August 13, 2005. For all devices supplied before
August 13, 2005, the last user is responsible for the correct disposal.
Page 38
38
Transport, storage and disposal
™
New Brunswick
Innova® 2300/2350 Shakers
English (EN)
Page 39
10Certifications
Declaration of Conformity
The product named below fulfills the requirements of directives and standards listed. In the case of
unauthorized modifications to the product or an unintended use this declaration becomes invalid.
Product name:
Innova® 2300 and 2350
including accessories
New Brunswick
™
Innova® 2300/2350 Shakers
English (EN)
Certifications
39
Product type:
Benchtop open air shaker
Relevant directives / standards:
2006/95/EC: EN 61010-1
2004/108/EC:EN 61000-6-1, EN 61000-6-4
2011/65/EU
2012/19/EU
Date: October 28, 2013
tnemeganaMoiloftroPdraoB tnemeganaM
Your local distributo r: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany
eppendorf@eppendorf.com