Endress+Hauser Oxymax COS22 Series, Oxymax COS22Z, Oxymax COS22D Operating Instructions Manual

Page 1
BA00446C/07/EN/04.15 71306871
Products Solutions Services
Operating Instructions
Oxymax COS22
Analog sensor for the measurement of dissolved oxygen
Page 2

Table of contents

1 Document information .............. 3
1.1 Warnings ............................ 3
1.2 Symbols .............................. 3
2 Basic safety instructions ............ 4
2.1 Requirements for personnel ............... 4
2.2 Designated use ........................ 4
2.3 Occupational safety ..................... 4
2.4 Operational safety ...................... 5
2.5 Product safety ......................... 5
3 Device description, function ......... 6
3.1 Amperometric measuring principle .......... 6
3.2 Sensor design ......................... 6
3.3 Membrane body ....................... 6
3.4 Polarization ........................... 6
4 Incoming acceptance and product
identification ....................... 7
4.1 Incoming acceptance .................... 7
4.2 Product identification .................... 7
4.3 Scope of delivery ....................... 8
4.4 Certificates and approvals ................ 8
10 Maintenance ...................... 22
10.1 Maintenance schedule .................. 22
10.2 Maintenance tasks ..................... 22
10.3 Clean sensor ......................... 23
10.4 Wear parts and consumables ............. 23
11 Accessories ....................... 26
11.1 Assemblies (selection) .................. 26
11.2 Measuring cable ...................... 26
11.3 Zero-point gel ........................ 26
11.4 Maintenance kit ...................... 27
12 Repair ............................ 28
12.1 Spare parts and consumables ............. 28
12.2 Return .............................. 28
12.3 Disposal ............................ 28
13 Technical data .................... 29
Index .................................. 34
5 Installation ........................ 9
5.1 Installation conditions ................... 9
5.2 Mounting the sensor ................... 10
5.3 Installation examples ................... 11
5.4 Post-installation check .................. 13
6 Electrical connection .............. 14
6.1 Connecting the sensor .................. 14
6.2 Ensuring the degree of protection .......... 14
6.3 Post-connection check .................. 14
7 Calibration and adjustment ........ 16
7.1 Types of calibration .................... 16
7.2 Calibration in air ...................... 16
7.3 Calculation example for the calibration
value ............................... 17
7.4 Zero point calibration ................... 18
8 Commissioning .................... 19
8.1 Function check ....................... 19
8.2 Sensor polarization .................... 19
8.3 Sensor calibration ..................... 20
9 Troubleshooting .................. 21
2 Endress+Hauser
Page 3
Oxymax COS22 Document information

1 Document information

1.1 Warnings

Structure of information Meaning
DANGER
L
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
WARNING
L
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
CAUTION
L
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
NOTICE
Cause/situation
Consequences of non-compliance (if applicable)
Action/note
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.

1.2 Symbols

Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
Endress+Hauser 3
Page 4
Basic safety instructions Oxymax COS22

2 Basic safety instructions

2.1 Requirements for personnel

• Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided may only be carried out directly by the manufacturer or by the service organization.

2.2 Designated use

The oxygen sensor is suitable for continuous measurement of dissolved oxygen in water. The specific suitability depends on the sensor design:
• COS22-*1 (standard, measuring range 0.01 to 60 mg/l) – Measuring, monitoring and regulating the oxygen content in fermenters – Monitoring the oxygen content in biotechnology facilities
• COS22-*3 (trace measurement, measuring range 0.001 to 10 mg/l, preferred operational range 0.001 to 2 mg/l), also suitable for high CO2 partial pressure – Monitoring inertization equipment in the food industry – Monitoring the residual oxygen content in carbonated fluids of the beverage industry – Trace measurement in industrial applications such as inertizations – Monitoring the residual oxygen content in boiler feedwater – Monitoring, measuring and regulating the oxygen content in chemical processes
NOTICE
Molecular hydrogen
Hydrogen causes sensitivity in other substances and leads to false low readings or, at the worst, total failure of the sensor.
Only use the COS22-*1/3 sensor in media free of hydrogen.
Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.

2.3 Occupational safety

As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been connected in accordance with these Operating Instructions.
4 Endress+Hauser
Page 5
Oxymax COS22 Basic safety instructions

2.4 Operational safety

1. Before commissioning the entire measuring point, verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged.
2. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified: Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.

2.5 Product safety

The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and European standards have been observed.
Endress+Hauser 5
Page 6
Device description, function Oxymax COS22
1 2
3
4
5
78
9
6

3 Device description, function

3.1 Amperometric measuring principle

The oxygen molecules that diffuse through the membrane are reduced at the cathode to hydroxide ions (OH-). At the anode, silver is oxidized to silver ions (Ag+) (this forms a silver halide layer). A current flows due to the electron donation at the cathode and the electron acceptance at the anode. Under constant conditions, this flow is proportional to the oxygen content of the medium. This current is converted in the transmitter and indicated on the display as an oxygen concentration in mg/l, µg/l, ppm, ppb or Vol%, as a saturation index in % SAT or as an oxygen partial pressure in hPa.

3.2 Sensor design

A0011868
 1 COS22 1
2 3
Plug-in head Thrust collar Sensor shaft
4
O-ring 8.5 x 1.5 mm
5
Membrane body
6
Shaft sleeve
7
Glass portion with anode and cathode
8
Process seal 10.77 x 2.62 mm
9
Process connection Pg 13.5

3.3 Membrane body

The oxygen dissolved in the medium is transported to the membrane by the necessary flow. The membrane is permeable for dissolved gases only. Other substances dissolved in the liquid phase, e.g. ionic substances, will not penetrate through the membrane. Therefore, medium conductivity has no impact on the measuring signal.
The sensor is shipped with a standard membrane body, which can be used for all common applications. The membrane is pretensioned at the factory and can be installed immediately.
Electrolytes are version-specific and cannot be mixed in a single application!

3.4 Polarization

When the sensor is connected to the transmitter, a fixed voltage is applied between the cathode and anode. The polarization current this creates can be identified on the transmitter with a display that is initially high, but decreases with time. The sensor cannot be calibrated until the display is calibrated.
Reference value for nearly complete polarization of a sensor that was previously stored for a long time:
• COS22-*1: 2 hours
• COS22-*3: 12 hours After this time, even measurements close to the determination limit are useful. The
necessary polarization time is reduced for sensors that were in use a short time before.
6 Endress+Hauser
Page 7

Oxymax COS22 Incoming acceptance and product identification

4 Incoming acceptance and product
identification

4.1 Incoming acceptance

1. Verify that the packaging is undamaged.  Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2. Verify that the contents are undamaged.  Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3. Check the delivery for completeness.  Check it against the delivery papers and your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
 The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.

4.2 Product identification

4.2.1 Nameplate

The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Safety information and warnings Compare the data on the nameplate with your order.

4.2.2 Product identification

Product page
www.endress.com/cos22
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to the product page for your product on the Internet.
2. In the navigation area on the right-hand side, select "Check your device features" under "Device support".
 An additional window opens.
3. Enter the order code from the nameplate into the search field.  You will receive information on each feature (selected option) of the order code.
Endress+Hauser 7
Page 8
Incoming acceptance and product identification Oxymax COS22

4.3 Scope of delivery

The scope of delivery comprises:
• Oxygen sensor with watering cap (filled with tap water) for protecting the membrane
• Electrolyte, 1 bottle, 10 ml (0.34 fl.oz.)
• Tool to push out the membrane body
• Brief Operating Instructions

4.4 Certificates and approvals

4.4.1
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark.

mark

4.4.2 Material certificates

Manufacturer declaration of FDA compatibility
The manufacturer declares the use of FDA-listed materials. Ask your Sales Center for the certificates.
Product FDA certificate for
COS22-****22 Membrane, O-rings, process seal
COS22Z-*2*2 Membrane, O-rings, process seal
COS22-****23 Membrane, O-rings
COS22Z-*2*3 Membrane, O-rings
Material test certificate
A test certificate 3.1 in accordance with EN10204 is supplied depending on the version (→ Product Configurator on the product page).
EHEDG
Compliance with EHEDG's criteria for hygienic design
• TÜV Rheinland, Apeldorn, Netherlands
• Certificate type: Type EL Class I
8 Endress+Hauser
Page 9
Oxymax COS22 Installation
10° 10°
Not permissible!
Not permissible!
Permissible angle of installation

5 Installation

5.1 Installation conditions

5.1.1 Orientation

A0005584-EN
 2 Permitted orientations
The sensor must be installed at an angle of inclination of 10 to 170 ° in an assembly, bracket or appropriate process connection. Recommended angle: 45°, to prevent the formation of air bubbles.
Inclination angles other than those mentioned are not permitted. Do not install the sensor overhead.
Observe the instructions for installing sensors in the Operating Instructions for the assembly used.

5.1.2 Mounting location

• Choose a mounting location that can be easily accessed at a later stage.
• Ensure that upright posts and fittings are fully secured and vibration-free.
• Select an installation location which produces a typical oxygen concentration for the individual application.
Endress+Hauser 9
Page 10
Installation Oxymax COS22
1 2
3
4

5.2 Mounting the sensor

5.2.1 Measuring system

A complete measuring system comprises:
• An Oxymax COS22 oxygen sensor
• A transmitter, e.g. Liquisys COM2x3
• Measuring cable COK21
• Optional: an assembly, e.g. permanent installation assembly CPA442, flow assembly CPA240, or retractable assembly CPA875
 3 Example of a measuring system with COS22-*1
1 Retractable assembly CPA875 2 Transmitter Liquisys COM253 3 Measuring cable COK21 4 Oxygen sensor COS22

5.2.2 Installing a measuring point

Installation in suitable assembly is required (depending on the application)
WARNING
L
Electrical voltage
In the event of a fault, non-grounded metallic assemblies may be under voltage and then are not touchable.
When using metallic assemblies and installation equipment, observe the national
grounding provisions.
10 Endress+Hauser
A0024029
Page 11
Oxymax COS22 Installation
For a complete installation of a measuring point, proceed as follows:
1. Install a retractable or a flow assembly (if used) into the process.
2. Connect the water supply to the rinse connections (if you use an assembly with cleaning function).
3. Install and connect the oxygen sensor.
NOTICE
Installation error
Cable open circuit, loss of sensor due to cable separation, unscrewing of membrane cap
Do not install the sensor suspended from the cable.
Screw the sensor into the assembly so that the cable is not twisted.
When installing or uninstalling the sensor body, hold it tightly. Turn using only the
hexagonal nut on the armored coupling. Otherwise you might unscrew the membrane cap. This will then remain in the assembly or process. Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
Select an installation location that is easy to access for later calibrations.

5.3 Installation examples

5.3.1 Permanent installation (CPA442)

The permanent installation assembly CPA442 enables easy adaptation of a sensor to nearly any process connections from Ingold nozzles to Varivent or Tri-Clamp connections. This kind of installation is very well suited for tanks and larger pipes. You will achieve a defined immersion depth of the sensor into the medium in the simplest way.

5.3.2 Flow assembly

CPA240
The flow assembly CPA240 offers up to three installation spaces for sensors with a shaft diameter of 12 mm (0.47"), a shaft length of 120 mm (4.7"), and a Pg 13.5 process connection. It very well suited for use in pipelines or hose connections. To prevent
Endress+Hauser 11
Page 12
Installation Oxymax COS22
1
7
6
5
3
2
4
1 2
3
4
measured error with trace measurements, pay particular attention to complete ventilation of the assembly.
A0005720
 4 Flow assembly CPA240 with protective cover
 5 Bypass installation
1 Main pipe 2 Medium removal 3, 6 Manually actuated or solenoid valves 4 Sampling 5 Flow assembly with installed sensor 7 Medium return
A0005721
Flow assembly for water treatment and processes
The compact stainless steel assembly offers space for a 12-mm sensor with a length of 120 mm. The assembly has a low sampling volume and, with the 6-mm connections, it is best suited for residual oxygen measurement in water treatments and boiler feedwater. The flow comes from below.
 6 Flow assembly
1 Built-in sensor 2 Drain 3 Wall mount (clamp D29) 4 Inflow
A0014081
12 Endress+Hauser
Page 13
Oxymax COS22 Installation
5
3
41
2
3
4
2

5.3.3 Retractable assembly (CPA875 or CPA450)

The assembly is designed for installation on tanks and pipes. Suitable nozzles must be available for this.
Install the assembly at places with constant flow. The minimum pipe diameter is DN 80 (3").
A0005722-EN
 7 Permissible and impermissible sensor installation positions with retractable assembly
1 Ascending pipe, best position 2 Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming 3 Horizontal pipe, lateral installation with permissible installation angle (acc. to sensor version) 4 Down pipe, impermissible
NOTICE
Sensor not in the medium all the way, buildup, overhead installation
Results in incorrect measurements
Do not install the assembly at places where air cushions or foam bubbles can be
formed. Avoid deposits on the sensor membrane, or remove them regularly.
Do not install the sensor overhead.

5.4 Post-installation check

• Are the sensor and cable undamaged?
• Is the orientation correct?
• Is the sensor installed in an assembly and is not suspended from the cable?
• Avoid the penetration of moisture by fitting the protective cap to the immersion assembly.
Endress+Hauser 13
Page 14
Electrical connection Oxymax COS22
PK
GN / YE
BN
YE
BU
S
12
90
91
11
Assignment Terminal
External shield
Cathode
Anode
NTC temp. sensor
Active internal shield
(NTC temp. sensor)

6 Electrical connection

WARNING
L
Device is live
Incorrect connection may result in injury or death.
The electrical connection may be performed only by an electrical technician.
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein. Prior to commencing connection work, ensure that no voltage is present on any cable.

6.1 Connecting the sensor

A multi-core COK21 measuring cable is used for the electrical connection of the sensor to the transmitter.
A0005583-EN
 8 Measuring cable COK21
The polarization voltage must be set at the transmitter as follows: Standard measuring range: -650mV Trace measuring range: -550mV The voltage is applied between the working electrode (cathode) and the reference electrode (anode).

6.2 Ensuring the degree of protection

Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required, designated use, may be carried out on the device delivered.
Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC interference immunity) agreed for this product can no longer be guaranteed due, for example, to covers being left off or cable (ends) which are loose or insufficiently secured.

6.3 Post-connection check

Device condition and specifications Notes
Are the outside of the sensor, assembly, junction box, cable undamaged?
Electrical connection Notes
Are the installed cables strain-relieved and not twisted?
Visual inspection
14 Endress+Hauser
Page 15
Oxymax COS22 Electrical connection
Device condition and specifications Notes
Is a sufficient length of the cable cores stripped, and is it positioned in the terminal correctly?
Are all the screws terminals properly tightened? Tighten
Are all cable entries mounted, tightened and leak-tight? For lateral cable entries, make sure the
Are all cable entries installed downwards or mounted laterally?
Check the fit (by pulling gently)
cables loop downwards to allow water to drip off
Endress+Hauser 15
Page 16
Calibration and adjustment Oxymax COS22

7 Calibration and adjustment

Calibration is a means of adapting the transmitter to the characteristic values of the sensor.
Calibration of the sensor is required after:
• Initial commissioning
• Changing the membrane or electrolyte
• Long pauses in operation without power supply Within the framework of system monitoring and supervision, for example, the calibration
can also be cyclically monitored (at typical time intervals, depending on operating experience) or renewed.

7.1 Types of calibration

You can carry out a slope or zero point calibration for the sensor. In most applications, single-point calibration in the presence of oxygen is sufficient
(=calibration of the sensor slope). When switching from process to calibration conditions, you have to allow a longer settling time for the sensor.
The additional calibration of the zero point improves the accuracy of the measurement results at trace concentrations. You can calibrate the zero point using nitrogen (min.
99.995%) or oxygen-free water. Make sure that the sensor is polarized and the measured value is settled at the zero point (at least 20-30 minutes) to prevent later incorrect measurements at trace concentrations.
The following describes calibration of the slope in air (saturated with water vapor) as the easiest and recommended calibration method. However, this type of calibration is possible only if the air temperature is ≥ 0 °C (32 °F).

7.2 Calibration in air

1. Remove the sensor from the medium.
2. Clean the outside of the sensor with a damp cloth.
3. Allow approx. 20 minutes for the sensor temperature to adapt to the ambient air. Make sure that the sensor is not exposed to any direct ambient influences (direct sunlight, drafts) during this time.
4. When the measured value display on the transmitter is stable, carry out the calibration according to the operating instructions for the transmitter. Pay particular attention to the software settings for the stability criteria for calibration.
5. Where necessary: Adjust the sensor.
6. Then insert the sensor into the medium.
Make sure you comply with the instructions for calibration in the Operating Instructions of the transmitter.
16 Endress+Hauser
Page 17
Oxymax COS22 Calibration and adjustment
1.
2.

7.3 Calculation example for the calibration value

As a check, you can calculate the expected calibration value (transmitter display) as shown in the following example (salinity is 0).
Determine:
• the sensor temperature (air temperature for the calibration type "air", water temperature for the calibration type "air-saturated water")
• the altitude above sea level
• the current air pressure (= rel. air pressure to sea level) at the time of calibration. (If undeterminable, use 1013 hPa (407 in H2O)).
Define:
• the saturation value S acc. to Table 1
• the altitude factor K acc. to Table 2
Table 1
T
[˚C (˚F)]S[mg/l=ppm]T[˚C (˚F)]S[mg/l=ppm]T[˚C (˚F)]S[mg/l=ppm]T[˚C (˚F)]S[mg/l=ppm]
0 (32) 14.64 11 (52) 10.99 21 (70) 8.90 31 (88) 7.42
1 (34) 14.23 12 (54) 10.75 22 (72) 8.73 32 (90) 7.30
2 (36) 13.83 13 (55) 10.51 23 (73) 8.57 33 (91) 7.18
3 (37) 13.45 14 (57) 10.28 24 (75) 8.41 34 (93) 7.06
4 (39) 13.09 15 (59) 10.06 25 (77) 8.25 35 (95) 6.94
5 (41) 12.75 16 (61) 9.85 26 (79) 8.11 36 (97) 6.83
6 (43) 12.42 17 (63) 9.64 27 (81) 7.96 37 (99) 6.72
7 (45) 12.11 18 (64) 9.45 28 (82) 7.82 38 (100) 6.61
8 (46) 11.81 19 (66) 9.26 29 (84) 7.69 39 (102) 6.51
9 (48) 11.53 20 (68) 9.08 30 (86) 7.55 40 (104) 6.41
10 (50) 11.25
Table 2
Altitude
[m (ft)]
0 (0) 1.000 550 (1800) 0.938 1050 (3450) 0.885 1550 (5090) 0.834
50 (160) 0.994 600 (1980) 0.932 1100 (3610) 0.879 1600 (5250) 0.830
100 (330) 0.988 650 (2130) 0.927 1150 (3770) 0.874 1650 (5410) 0.825
150 (490) 0.982 700 (2300) 0.922 1200 (3940) 0.869 1700 (5580) 0.820
200 (660) 0.977 750 (2460) 0.916 1250 (4100) 0.864 1750 (5740) 0.815
250 (820) 0.971 800 (2620) 0.911 1300 (4270) 0.859 1800 (5910) 0.810
300 (980) 0.966 850 (2790) 0.905 1350 (4430) 0.854 1850 (6070) 0.805
350 (1150) 0.960 900 (2950) 0.900 1400 (4600) 0.849 1900 (6230) 0.801
400 (1320) 0.954 950 (3120) 0.895 1450 (4760) 0.844 1950 (6400) 0.796
450 (1480) 0.949 1000 (3300) 0.890 1500 (4920) 0.839 2000 (6560) 0.792
500 (1650) 0.943
K Height
[m (ft)]
K Height
[m (ft)]
K Height
[m (ft)]
K
Endress+Hauser 17
Page 18
Calibration and adjustment Oxymax COS22
3.
4.
Calculate the factor L:
Relative air pressure at calibration
L= -----------------------------------------------
1013 hPa
Calculate the calibration value C:
C = S . K . L
Example
• Air calibration at 18 °C (64 °F), altitude 500 m (1650 ft) above sea level, current air pressure 1009 hPa (405 in H2O)
• S = 9.45 mg/l, K = 0.943, L = 0.996
• Calibration value C = 8.88 mg/l.
You do not need factor K in the table if your measuring device returns the absolute air pressure L for calculation is then: C = S . L
(air pressure depending on altitude) as the measured value. The formula
abs
.
abs

7.4 Zero point calibration

The zero point is not so important when working with relatively high concentrations of oxygen.
This situation changes, however, once oxygen sensors are used at low concentrations and in the trace range and where they are also to be calibrated in the zero point. Zero point calibrations are demanding as the ambient medium - usually air - already has a high oxygen content. This oxygen must be blocked off for zero-point calibration of the sensor.
Calibration with zero-point gel can be used here: The oxygen-depleting gel COY8 (→  26) creates an oxygen-free medium for zero point calibration.
Prior to sensor zero point calibration, check:
• Is the sensor signal stable?
• Is the value displayed plausible?
1. If the sensor signal is stable: Calibrate the zero point.
2. If necessary: Adjust the sensor to the zero point.
The reference method (sample calibration in zero point) can also be used here if appropriate collecting vessels or reference measurement are available.
If the oxygen sensor is calibrated too early, this can result in an incorrect zero point. Rule of thumb: Operate the sensor in zero-point gel for at least 0.5 h. If the sensor was already operated in the trace range before the zero point calibration,
the time specified above generally suffices. If the sensor was operated in air, significantly more time must be factored in to also remove residual oxygen from any dead volume inherent to the design. Here a value of 2 hours applies as a general rule.
18 Endress+Hauser
Page 19
Oxymax COS22 Commissioning

8 Commissioning

8.1 Function check

Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct. If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water
or air) is connected correctly.
WARNING
L
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
Before applying compressed air to an assembly with cleaning facility, make sure the
connections are correctly fitted. Do not install the assembly in the process if you cannot make the correct connection
reliably.

8.2 Sensor polarization

NOTICE
Incorrect measurements due to ambient influences
Be absolutely certain to avoid direct sunlight on the sensor.
Make sure you comply with the instructions for commissioning in the Operating
Instructions of the transmitter.
The sensor has been tested at the factory for proper function and is shipped in ready-to­operate condition.
To prepare for the calibration, carry out the following steps:
1. Pull off the sensor protection cap.
2. Expose the sensor, which should be dry on the outside, to the air atmosphere.  The air should be saturated with water vapor. Therefore, install the sensor as
close as possible to a surface of water. However, the sensor membrane must remain dry during calibration. Therefore, avoid direct contact with the surface of water.
3. Connect the sensor to the transmitter.
4. Switch on the transmitter.  When the sensor is connected to the transmitter, the polarization takes place
automatically after the power-up of the transmitter.
5. Wait for the polarization time to run out.
Endress+Hauser 19
Page 20
Commissioning Oxymax COS22

8.3 Sensor calibration

Calibrate the sensor (e.g. air calibration) immediately after the polarization time runs out. The calibration intervals depend greatly on:
• The application
• The installation position of the sensor
The following method helps you determine the necessary calibration intervals:
1. Inspect the sensor one month after commissioning. Take it out of the medium and dry it.
2. After 10 minutes, measure the oxygen saturation index in air.  Decide depending on the result: If the measured value is not 100 ±2 %SAT, you
must calibrate the sensor. Otherwise, double the length of time to the next inspection.
3. Proceed as per Point 1 after two, four and/or eight months. In this way, you can determine the optimum calibration interval for your sensor.
In any case, calibrate the sensor at least once a year.
20 Endress+Hauser
Page 21
Oxymax COS22 Troubleshooting

9 Troubleshooting

If one of the following problems is present, check the measuring system in the sequence listed here.
Problem Testing Remedial action
No display, no sensor reaction Power supply to the transmitter? Connect the power supply
Sensor cable connected correctly? Set up correct connection.
Medium flow present? Create medium flow
Coating on the membrane? Clean sensor
No electrolyte in the measuring chamber?
Displayed value too high Polarization ended? Wait for polarization time to finish
Is sensor calibrated/adjusted? recalibrate/readjust
Displayed temperature clearly too low?
Membrane visibly stretched? Replacing the membrane cap
Electrolyte dirty? Electrolyte replacement
Open the sensor and dry the electrode. Is the transmitter display now at 0?
Displayed value too low Is sensor calibrated/adjusted? recalibrate/readjust
Medium flow present? Create medium flow
Displayed temperature clearly too high?
Coating on the membrane? Clean sensor
Electrolyte dirty? Electrolyte replacement
Display value fluctuating greatly Membrane visibly stretched? Replacing the membrane cap
Open the sensor and dry the electrode. Is the transmitter display now at 0?
Refill or change electrolyte
Check sensor, if necessary send sensor in for repair.
Check the cable connection. If the problem persists, send in the sensor.
Check sensor, if necessary send sensor in for repair.
Check the cable connection. If the problem persists, send in the sensor.
Please observe the troubleshooting information in the Operating Instructions for the transmitter. If need be, carry out a test on the transmitter.
Endress+Hauser 21
Page 22
Maintenance Oxymax COS22

10 Maintenance

Take all the necessary precautions in time to ensure the operational safety and reliability of the entire measuring system.
NOTICE
Effects on process and process control
When carrying out any work on the system, take into account possible repercussions
for process control or the process itself. For your own safety, only use genuine accessories. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.

10.1 Maintenance schedule

Maintenance cycles depend to a great extent on the operating conditions. The following rule of thumb applies:
• Constant conditions, e.g. power plant = long cycles (1/2 year)
• Greatly varying conditions, e.g. daily CIP cleaning = short cycles (1 month or shorter)
The following method helps you determine the necessary intervals:
1. Inspect the sensor one month after commissioning. Take it out of the medium and dry it.
2. After 10 minutes, measure the oxygen saturation index in air.  Decide depending on the result: If the measured value is not 100 ±2 %SAT, you
must maintain the sensor. Otherwise, double the length of time to the next inspection.
3. Proceed as per Point 1 after two, four and/or eight months. In this way, you can determine the optimum maintenance interval for your sensor.
Particularly in the case of widely fluctuating process conditions, damage may occur to the membrane even within a maintenance cycle. You can recognize this by implausible sensor behavior.

10.2 Maintenance tasks

The following tasks are mandatory:
• Clean the sensor and the glass body with anode and cathode (particularly if membrane is dirty)
• Replacement of wear parts or consumables: – Electrolyte – Membrane body – – Sealing ring
• Check the measuring function:
1. Remove the sensor from the medium.
2. Clean and dry the membrane.
3. After about 10 minutes, measure the oxygen saturation index in air (without recalibration).
4. The measured value should be 100 ± 2 % SAT.
• Recalibration (if desired or required)
22 Endress+Hauser
Page 23
Oxymax COS22 Maintenance
1 2
3
4
5
78
9
6

10.3 Clean sensor

The measurement can be corrupted by sensor fouling or malfunction, e.g.:
• Buildup on the sensor membrane
causes longer response times and a reduced slope under certain circumstances.
For reliable measurement, the sensor must be cleaned at regular intervals. The frequency and intensity of the cleaning operation depend on the measuring medium.
Clean the sensor:
• before every calibration
• at regular intervals during operation as necessary
• before returning it for repairs.
Type of soiling Cleaning
Salt deposits Immerse the sensor in drinking water or in 1-5% hydrochloric acid
(for a few minutes). Afterwards, rinse it with copious amounts of water.
Dirt particles on the sensor shaft and shaft sleeve (not membrane!)
Dirt particles on the membrane or membrane body
Clean the sensor shaft and sleeve with water and a suitable brush.
Clean the membrane with water and a soft sponge.
After cleaning, rinse the sensor with copious amounts of clean water.

10.4 Wear parts and consumables

Parts of the sensor are subject to wear during operation. By taking suitable measures, you can restore the normal operating function.
Measure Reason
Sealing ring replacement Visible damage to a sealing ring
Electrolyte replacement Unstable or implausible measuring signal or fouling of the electrolyte
Membrane body replacement Membrane is damaged or can no longer be cleaned (hole or overstretching)
A0011868
 9 COS22 1
Plug-in head
2
Thrust collar
3
Sensor shaft
4
O-ring 8.5 x 1.5 mm
5
Membrane body
6
Shaft sleeve
7
Glass portion with anode and cathode
8
Process seal 10.77 x 2.62 mm
9
Process connection Pg 13.5
Endress+Hauser 23
Page 24
Maintenance Oxymax COS22

10.4.1 Replacing sealing rings

The sealing ring must be replaced if visibly damaged. For replacement, use only original sealing rings.
The following O-rings can be replaced:
• Sealing ring for shaft sleeve: item 4 →  23
• Sealing ring for process: item 8
If the sealing ring on the membrane body (item 5) is damaged, you must replace the entire membrane body.

10.4.2 Replacing electrolyte

The electrolyte is used up slowly during operation. This is caused by electrochemical substance reactions. In de-energized state, no substance reactions take place, and the electrolyte is not used up. The operating time of the electrolyte is shortened by diffusion of dissolved gases such as H2S, NH3 or high concentrations of CO2.
Theoretical operating time at pO2 = 210 mbar and T=25 ˚C (77 ˚F) COS22-*1 (standard sensor): > 1.5 years COS22-*3 (trace sensor): > 3 months
CAUTION
L
The standard electrolyte is a strong irritant
Danger of severe skin and eye irritation
Be absolutely certain to observe the corresponding workplace safety regulations.
Wear protective clothing, gloves and goggles when handling the electrolyte.
In case of eye contact: Remove contact lenses, flush eyes with water for a few minutes
and contact a doctor. In case of skin contact: Take off wet clothing immediately, wash the skin or take a
shower.
Generally, the following applies:
• Changing the electrolyte is mandatory if the membrane body is detached.
• Sensors operated close to the zero point consume hardly any chemical electrolyte. The electrolyte does not have to be replaced for a long period.
• Sensors operated at high partial oxygen pressures (> 100 hPa) consume a significant amount of electrolyte. The electrolyte has to be replaced frequently.
• 25 ml electrolyte are enough to fill the membrane body approx. 15 times.
Draining the electrolyte
1. Remove the sensor from the medium.
2. Clean the outside of the sensor.
3. Hold the sensor vertically and unscrew the shaft sleeve.  The membrane body is either in the shaft sleeve or is still on the glass portion
with the anode and cathode.
4. Remove the membrane body. For this purpose, use the tool provided to push out the membrane body.
5. Drain the membrane body and flush it with potable water.
Top up the electrolyte and install the membrane body
1. Fill fresh electrolyte from the supply bottle into the membrane body.
2. Remove all air bubbles from the electrolyte by tapping the side of the membrane body (using a pen or pencil, for example).
3. Hold the sensor vertically and carefully push the membrane body, filled with electrolyte, on the glass portion.
24 Endress+Hauser
Page 25
Oxymax COS22 Maintenance
4. Carefully screw on the shaft sleeve as far as it will go.
After replacing, the sensor must be repolarized and recalibrated. Then insert the sensor into the medium and check that no alarm is displayed on the transmitter.

10.4.3 Replacing membrane body

Removing the membrane body
1. Remove the sensor from the medium.
2. Clean the outside of the sensor.
3. Hold the sensor vertically and unscrew the shaft sleeve.  The membrane body is either in the shaft sleeve or is still on the glass portion
with the anode and cathode.
4. Remove the membrane body. For this purpose, use the tool provided to push out the membrane body.
5. Dispose of the old membrane body and the old electrolyte.
6. Take a new membrane body out of its packaging.
Top up the electrolyte and install the membrane body
1. Fill fresh electrolyte from the supply bottle into the membrane body.
2. Remove all air bubbles from the electrolyte by tapping the side of the membrane body (using a pen or pencil, for example).
3. Hold the sensor vertically and carefully push the membrane body, filled with electrolyte, on the glass portion.
4. Carefully screw on the shaft sleeve as far as it will go.
After replacing, the sensor must be repolarized and recalibrated. Then insert the sensor into the medium and check that no alarm is displayed on the transmitter.

10.4.4 Replacing glass body with cathode

NOTICE
Polishing the cathode can cause impaired function or total failure of the sensor.
Do not clean the cathode mechanically.
If the cathode is coated, replace the glass body:
1. Hold the sensor vertically and unscrew the shaft sleeve: item 6 .
2. If the membrane body (item 5) remains on the glass body (item 7) and not in the shaft sleeve, remove it from the glass body.
3. Flush the glass body, along with the anode and cathode, using distilled water.
4. Pull the used glass body out of the holder.
5. Dry the inside of the electrode holder.
6. Plug a new glass body (from the membrane kit) into the holder so that it fits. Ensure that you do not damage the electrical contact pins.
7. Fill the membrane body with electrolyte then screw the shaft sleeve back on.
Endress+Hauser 25
Page 26
Accessories Oxymax COS22

11 Accessories

The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office.

11.1 Assemblies (selection)

Cleanfit CPA875
• Retractable process assembly for sterile and hygienic applications
• For in-line measurement with standard 12 mm sensors for parameters such as pH, ORP and oxygen
• Product Configurator on the product page: www.endress.com/cpa875
Technical Information TI01168C
Flowfit CPA240
• pH/redox flow assembly for processes with stringent requirements
• Product Configurator on the product page: www.endress.com/cpa240
Technical Information TI00179C
Unifit CPA442
• Installation assembly for food, biotechnology and pharmaceutics
• With EHEDG and 3A certificate
• Product Configurator on the product page: www.endress.com/cpa442
Technical Information TI00306C
Cleanfit CPA450
• Manual retractable assembly for installing 120 mm sensors in tanks and pipes
• Product Configurator on the product page: www.endress.com/cpa450
Technical Information TI00183C
Flow assembly
• For sensors with Ø 12 mm and length 120 mm
• Compact stainless steel assembly with low sampling volume
• Order No. Order No.: 71042404

11.2 Measuring cable

11.2.1 Cable for COS22

COK21
• Cable length 3 m (9.8 ft) Order No. 51505870
• Cable length 10 m (33 ft) Order No. 51505868

11.3 Zero-point gel

COY8
Zero-point gel for oxygen sensors
• Oxygen-depleting gel for test purposes
• Product Configurator on the product page: www.endress.com/coy8
Technical Information TI01244C
26 Endress+Hauser
Page 27
Oxymax COS22 Accessories

11.4 Maintenance kit

COS22Z
• Service Kit, COS22 and COS22D
• Ordering information: www.endress.com/cos22d under "Accessories/spare parts"
Endress+Hauser 27
Page 28
Repair Oxymax COS22

12 Repair

12.1 Spare parts and consumables

COS22Z
• Service Kit, COS22 and COS22D
• Ordering information: www.endress.com/cos22d under "Accessories/spare parts"

12.2 Return

The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and conditions at www.endress.com/support/return-material.

12.3 Disposal

The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
28 Endress+Hauser
Page 29
Oxymax COS22 Technical data

13 Technical data

13.1 Input

Measured values Dissolved oxygen [mg/l, µg/l, ppm, ppb or % SAT or hPa]

Temperature [˚C, ˚F]

Measuring ranges Measuring ranges apply for 20 ˚ (68 ˚F) and 1013 hPa (15 psi)

Measuring range Optimum operational range
COS22-*1 0.01 to 60 mg/l
0 to 600 % SAT 0 to 1200 hPa (0 to 6 psi) 0 to 100 Vol%
COS22-*3 0.001 to 10 mg/l
0 to 120 % SAT 0 to 250 hPa (0 to 6 psi) 0 to 25 Vol%
1) Applications in this range guarantee a long service life and minimum maintenance
0.01 to 20 mg/l 0 to 200 % SAT 0 to 400 hPa (0 to 6 psi) 0 to 40 Vol%
0.001 to 2 mg/l 0 to 20 % SAT 0 to 40 hPa (0 to 6 psi) 0 to 4 Vol%

13.2 Performance characteristics

Response time From air to nitrogen at reference operating conditions:

• t90 : < 30 s
• t98 : < 60 s

reference operating conditions

Reference temperature: 25 ˚C (77 ˚F) Reference pressure: 1013 hPa (15 psi) Reference application: Air-saturated water
1)

Signal current in air COS22-*1 (standard sensor): 40 to 100 nA

COS22-*3 (trace sensor): 210 to 451 nA

Zero current COS22-*1 (standard sensor): < 0.1 % of the signal current in air

COS22-*3 (trace sensor): < 0.03 % of the signal current in air

Measured value resolution

COS22-*1 (standard sensor): 10 ppb in aqueous, 0.2 hPa or 0.02 Vol% in
gaseous media
COS22-*3 (trace sensor): 1 ppb in aqueous, 0.02 hPa or 0.002 Vol%
in gaseous media
Corresponds to the recommended measured value resolution at the transmitter

Maximum measured error COS22-*1 (standard sensor): ≤ ±1 % of measuring range + 10 ppb

COS22-*3 (trace sensor): ≤ ±1 % of measuring range + 1 ppb
*
at reference operating conditions
*
*
Endress+Hauser 29
Page 30
Technical data Oxymax COS22

Long-term drift < 4 % per month in reference operating conditions

≤ 1 % per month in operation with reduced oxygen concentration (< 4 Vol% O2)

Influence of the medium pressure

Polarization time

Intrinsic oxygen consumption

Operating time of the electrolyte

Temperature compensation COS22

Pressure compensation not required
COS22-*1 (standard sensor): < 30 min for 98% signal value, 2 h for
COS22-*3 (trace sensor): < 3 h for 98% signal value, 12 h for 100%
COS22-*1 (standard sensor): Approx. 20 ng/h in air at 25 ˚C (77 ˚F) COS22-*3 (trace sensor): Approx. 100 ng/h in air at 25 ˚C (77 ˚F
→  24
Compensation of the membrane properties depending on the transmitter, recommended:
2.4 % per K

13.3 Environment

100%

Ambient temperature range

Storage temperature –5 to +50 ˚C (20 to 120 ˚F) at 95% relative humidity, non-condensing

Degree of protection IP 68 (10 m (33 ft) head of water at 25 ˚C (77 ˚F) over 45 days, 1 mol/l KCl)

Humidity 0 to 100%, not condensating in area of T-82 connection

-5 to +135 °C (23 to 275 °F), non-freezing
NOTICE
Danger of sensor drying out
Store the sensor with the watering cap only (filled with tap water).

13.4 Process

Process temperature

Process pressure Ambient pressure ... 12 bar (... 174 psi) absolute

-5 to +135 °C (23 to 275 °F), non-freezing
30 Endress+Hauser
Page 31
Oxymax COS22 Technical data
p
T
[bar]
[°C]
6
12
135
80
[psi]
174
87
T
[°F]
176
275
-5230
32

Temperature-pressure ratings

A0028771-EN

Minimum flow COS22-*1 (standard sensor): 0.02 m/s (0.07 ft/s)

COS22-*3 (trace sensor): 0.1 m/s (0.33 ft/s)

Chemical resistance Parts in contact with the medium are chemically resistant to:

• Diluted acids and alkalis
• Hot water and superheated steam up to max. 135 ˚C (275 ˚F)
• CO2 up to 100 %, only with trace sensor COS22-*3
NOTICE
Hydrogen sulfide and ammonia shorten the operating life of the sensor.
Do not use the sensor in applications where it is exposed to hydrogen sulfide or
ammonia vapors.

Cross-sensitivity

COS22-*1/3 Molecular hydrogen causes false low readings and can, in a worst-case scenario, result in
total failure of the sensor.

CIP compatibility Yes (COS22-*1/3)

SIP compatibility Yes, max. 140 °C (284 °F) (COS22-*1/3)

Autoclavability Yes, max. 140 °C (284 °F), max. 30 min. (COS22-*1/3)

Endress+Hauser 31
Page 32
Technical data Oxymax COS22
Pg 13,5
54 (2.13)
120 / 225 / 360 / 420 (4.72 / 8.86 / 14.2 / 16.5)*
12 - 0.1
(0.47)
28 (1.10)
Ø6 (0.24)
39 (1.54)
160 (6.30)
30
(1.18)
Ø6 (0.24)
Ø 29 x 1.5 (1.14 x 0.06)

13.5 Mechanical construction

Dimensions

A0011886
 10 Dimensions in mm (inch)

Flow assembly for sensors with Ø 12 mm (accessories)

 11 Dimensions in mm (inch)

Weight Depending on the design (length)

0.2 kg (0.44 lbs) to 0.7 kg (1.54 lbs)

Materials

Parts in contact with medium
Sensor shaft (depending on the version) Stainless steel 1.4435 (AISI 316L)
Electrode combination COS22-*1/3: silver / platinum Process seal VITON (FDA-compliant) Process seal for ATEX/FM/CSA/NEPSI/TIIS VITON (not FDA-compliant) Seals/O-rings
Membrane body, sealing ring for shaft sleeve Membrane Silicone (FDA-compliant, in compliance

Process connection Pg 13.5

A0015019
Titanium Alloy C22
VITON (FDA-compliant) Perfluoroelastomer with USP88 Class VI
with USP87/88 class VI), PTFE, steel mesh
32 Endress+Hauser
Page 33
Oxymax COS22 Technical data

Surface roughness Ra < 0.38 µm

Temperature sensor NTC 22 kΩ

Electrolyte

COS22-*1 (standard sensor): Slightly alkaline electrolyte COS22-*3 (trace sensor): Neutral electrolyte
Endress+Hauser 33
Page 34

Index

Index
A
Accessories ................................ 26
Adjustment ................................16
Ambient temperature range ................... 30
Amperometric measuring principle ............... 6
Assemblies ................................ 26
Autoclavability ............................. 31
C
Calibration
Calculation example ....................... 17
In air .................................. 16
Types of calibration ....................... 16
Zero point calibration ...................... 18
Cathode .................................. 25
Check
Connection ..............................14
Function ............................... 19
Installation ............................. 13
Chemical resistance ..........................31
CIP compatibility ............................ 31
Connection
Check ..................................14
Ensuring the degree of protection ............. 14
Cross-sensitivity ............................ 31
Cust. tag number ............................10
D
Declaration of Conformity ...................... 8
Degree of protection
Ensuring ............................... 14
Technical data ........................... 30
Designated use .............................. 4
Device description ............................ 6
Disposal .................................. 28
E
EHEDG .................................... 8
Electrical connection ......................... 14
Electrolyte
Operating time ...........................24
Properties .............................. 33
Replacement ............................ 24
Environment ...............................30
F
FDA compatibility ............................ 8
Function check ............................. 19
G
Glass body .................................25
H
Humidity ..................................30
I
Incoming acceptance ......................... 7
Influence of the medium pressure ............... 30
Installation
Check ..................................13
Examples ............................... 11
Orientation .............................. 9
Sensor ................................. 10
Installation instructions ........................9
Intrinsic oxygen consumption .................. 30
L
Long-term drift ............................. 30
M
Maintenance schedule ........................22
Maintenance tasks .......................... 22
Material test certificate ........................ 8
Materials ................................. 32
Maximum measured error ..................... 29
Measured value resolution .....................29
Measured values ............................ 29
Measuring cable ............................ 26
Measuring principle ...........................6
Measuring ranges ........................... 29
Measuring system ........................... 10
Medium pressure ........................... 30
Membrane body
Description ...............................6
Replace ................................ 25
Minimum flow ............................. 31
Mode of operation ............................6
N
Nameplate ................................. 7
O
Occupational safety ........................... 4
Operational safety ............................5
Orientation ................................. 9
P
Performance characteristics ....................29
Polarization ................................ 6
Polarization time ............................30
Pressure temperature load curve ................ 31
Process ................................... 30
Process connection .......................... 32
Process pressure ............................ 30
Process temperature ......................... 30
Product identification ......................... 7
Product safety ............................... 5
R
reference operating conditions ..................29
Repair ....................................28
Replacing sealing rings ....................... 24
Response time ..............................29
Return ................................... 28
34 Endress+Hauser
Page 35
S
Safety
Occupational safety ........................ 4
Operation ................................5
Product ................................. 5
Safety instructions ............................4
Scope of delivery ............................. 8
Sensor
Calibration ..............................20
Cleaning ............................... 23
Connecting ..............................14
Design .................................. 6
Mounting ...............................10
Polarization ........................... 6, 19
Sensor design ............................... 6
Signal current in air ..........................29
SIP compatibility ............................ 31
Spare parts ................................ 28
Storage temperature ......................... 30
Surface roughness ........................... 33
Symbols ................................... 3
Index
T
Technical data
Environment ............................ 30
Input .................................. 29
Mechanical construction ....................32
Performance characteristics ................. 29
Process ................................ 30
Temperature compensation .................... 30
Temperature sensor ......................... 33
Temperature-pressure ratings .................. 31
Troubleshooting ............................ 21
U
Use .......................................4
W
Warnings .................................. 3
Wear parts and consumables ................... 23
Weight ................................... 32
Z
Zero current ............................... 29
Zero solution
Application ............................. 18
Zero-point gel .............................. 26
Endress+Hauser 35
Page 36
*71306871*
71306871
www.addresses.endress.com
Loading...